In-situ measurements of P11 (VSF), P12 and P22 elements of the scattering Mueller
matrix from 0.1-155o (nominal) in water
VSF (P11) only at small angles, 0.1 to 15o (nominal) in 32 logarithmic steps in angles
Integration of 0.1-155o VSF provides a good estimate of total scattering coefficient b.
Beam attenuation measured with LISST-100X optics.
Roving EyeballTM optics permit 1o resolution in scattering angles between 15 -155
o
Approximately 4 sec per measurement set [involves 2 turns of Eyeball with vertical,
and horizontal polarized laser].
Daylight rejection by laser modulation.
Data from small and large angles in a single data stream, including depth,
temperature,date and time.
This document is copyrighted by SEQUOIA SCIENTFIC, INC.
LISST-VSF
SPECIFICATIONS:
Parameters measured
Small-angle VSF in 32 log-spaced angles, from 0.1 to 15o(nominal)
VSF, P12 and P22 over 15-150o in 1o steps (nominal)
Temperature from –5° to 50°C with 10 mdeg resolution
Operational depth (50 m max depth @ 8 cm resolution)
Operating Concentration range
Beam attenuation from 0.13 to 20 m
Technology
Fiber Coupled TE-cooled Laser Diode @ 515 nm
Ring Detector for small-angle VSF
Roving Eyeball and PMT for 15-150o (nominal)
Mechanical and Electrical
LISST-VSF Instrument
Dimensions 13.3 cm (5.25”) diameter, 114.0 cm (44.9”) L
Weight: 15.6 kg (34.4 lbs) in air; in water TBD
Depth rating: 300 m survival depth (50 m operational depth)
External power input: 12VDC @ 5A, min 9VDC, max 20VDC
Sampling rate: 4 seconds for a full measurement of P11, P12, P22
Power drain: 700 mA measuring / 170mA quiescent
Data storage: 128GB, equivalent to 40,000? measurements
Battery Housing
Dimensions 13.3 cm (5.25”) diameter, 71.1 cm (28.0”) L
Weight: 10.3 kg (22.8 lbs]) in air; in water TBD
Capacity: 14.8V, 39 A-hr.
-1
(based on >=30% transmission)
Welcome to the LISST-VSF!
Using this manual
This manual is divided into two sections.
Section One contains an introduction to the LISST-VSF instrument and the principles of its operation.
Section Two provides detailed instructions for operation
Appendices provide further details …
Instrument Specific Constants
These are included in a MATLAB file, named
Cal_Factors_xxxx.mat, where xxxx is the serial number of your
instrument.
Technical assistance
For technical assistance please contact your local Distributor or
Sequoia. Please be sure to include the instrument Serial
Number with any correspondence.
Sequoia Scientific, Inc. contact information:
Telephone: +1(425) 641-0944
Email: info@sequoiasci.com
Waste Electrical and Electronic Equipment
Smaltimento di apparecchiature elettriche ed elettroniche da rottamare
Thank you for purchasing a LISST-VSF instrument, and congratulations on
your new purchase.
The LISST-VSF delivers a powerful suite of measurements for marine optical
scientists, all from one package. It delivers the Volume Scattering Function
(VSF) from small forward angles to wide angles, depolarization parameters,
and the beam attenuation coefficient c. An excellent estimate of the beam
scattering coefficient b, and by difference, the beam absorption coefficient a
can also be derived from the data. All properties are measured at a laser
wavelength of 532nm. The forward angles in water, from 0.0936 to 15.05, at
which VSF is measured are spaced logarithmically. The intermediate and
large angles (>15-deg) are linear, with 1-deg resolution. Because of the high
angular resolution, this instrument permits examination of enhanced
scattering by bubbles in the ~80-deg region.
Light Scattering
and VSF
The Volume Scattering Function (VSF) describes the distribution of scattered
light energy as a function of scattering angle. VSF is defined for unpolarized
light source and no discrimination of polarization in the scattered light.
However, scattering of light by particles always produces changes in
polarization state of the scattered rays. The state of polarization is described
by the Stokes vector I, with elements I,Q,U,V. The Stokes vector of the
scattered light and that of the incident light Ii are related through a matrix
product:
Is = P I
i
where P is the scattering Mueller matrix. P is a 4 x 4 matrix. It contains
information regarding particle scattering characteristics. The elements of P
are often denoted by Pij . The first element, P11 is identical to the volume
scattering function, VSF. Elements [1,2] and [2,2] relate to changes in the
linear polarization properties I and Q. For example, element P12 produces
depolarization. Element P22 is known to be an indicator of sphericity of
particles; so that P22 =1 everywhere implies that particles are spheres. With
the LISST-VSF, it is assumed that symmetry forces elements P
13
and P
14
to
zero, and that P
12
and P21 are equal. Thus only 3 unknown are produced: P11,
P12, and P22,
At small forward angles, scattered light maintains its original polarization, so
that depolarization is small. This instrument does not measure depolarization
at angles smaller than 15o, because there the VSF is measured by a set of
ring detectors, identical to our familiar LISST-100X instrument. A different
instrument, LISST-Stokes has been developed for small-angle depolarization
studies.
Section I: LISST-VSF Instrument
I.1 Introduction
LISST-VSF User’s Guide 13
Theory and
Mathematical
Description
Appendix A and B provide detailed description of optics of the instrument,
and of data processing method. Appendix C describes MATLAB data
processing functions. If data are good, a single command make_P processes
a set of files and produces elements of the P matrix, together with beam
attenuation and scattering coefficients. The function name make_P derives
from the scattering matrix P.
Instrument
Overview
The LISST-VSF instrument employs two optical systems which combine to
produce the full VSF. The small angle measurements (<15o) are done with
ring detectors, identical to our LISST-100X instrument. We will often refer to
data collected with these detectors as ring data, or rings data. Scattering at
larger angles (>15
o
) is measured using a rotating ‘eyeball’ (see photo on
opposing page). This data is referred to as eyeball data.
For ring data collection optics, please refer to details in one of several
publications (e.g. Agrawal & Pottsmith, 2000; included on the ship disk). A
simplified explanation of the eyeball optics for large-angle measurements is
offered here.
Typical volume scattering function sensors, or for polarization studies, typical
polarimeters observe a common sample volume with a multiple set of
detectors. In some of these systems, a single viewing telescope rotates
around the sample volume. The former system can only have a small number
of detectors, and each needs to be carefully calibrated. In the latter, rotating a
telescope requires cumbersome and slow mechanical systems that don’t lend
themselves to deep submerged usage. For these reasons, the present
instrument employs a rotating eyeball. Instead of viewing a common sample
volume, the eyeball scans the length of a laser beam, thus obtaining angular
information.
In the LISST-VSF instrument, the scattered light entering the eyeball is split
into its two polarization components, and each is sensed with a dedicated
photomultiplier (PMT). A set of measurements comprises the recording of
these two PMT outputs over 2 rotations of the eyeball. Each rotation is for a
fixed polarization of the laser; e.g. the first rotation is with perpendicular laser
polarized, and a second rotation is with parallel polarization. To rotate laser
polarization between eyeball turns, a half-wave plate is mechanically inserted
in the laser beam before it enters water. Data are stored in a single datafile
that combines rings and eyeball data.
The small eyeball makes the LISST-VSF instrument compact, autonomous,
and manageable underwater. As a result, this instrument may be left on a
tripod or mooring.
As explained in Appendix B, due to the rotating eyeball receiver, the stored
scattered light signals are a mix of the 3 parameters P11, P12 and P22. The
convenience of the eyeball thus involves a price – solving for these quantities.
We provide software to perform this function. For details, please see
Appendix C.
14 LISST-VSF User’s Guide
Electro-Optic
Components
The instrument employs a single 515nm TE-cooled diode laser for both optical
systems – rings and eyeball optics. To reject ambient light, the beam is
modulated at ~3 kHz. Two miniature, compact PMT’s are used to sense scattered light. Electronics control the gain of these two PMT’s. The gain can
be adjusted with software commands. The output of PMT’s is digitized and
stored in data files. See Appendix B.
Depth and
Temperature
Sensors
The depth sensor is mounted to the instrument connector endcap. The
temperature sensor is also thermally bonded to this endcap. Please note that
the temperature sensor is therefore slow.
Communication
A cable is provided to connect the underwater connector on the instrument
endcap to a PC. Communication is at 9600baud, except when downloading
data, it automatically switches to 115Kbaud.
Internals
The cylindrical case with the flat endcap contains all the transmitting optics
and the 2 PMT’s. This includes the laser, a beam-expander, the polarization
rotator, and laser power (output) sensor (called Laser Reference throughout
this manual). The cylindrical case with the connector endcap contains the
small-angle measuring ring-detector optics, i.e. a 50-mm focal length lens, the
ring-detectors with its built-in sensor for laser power transmitted through water
(called Transmission sensor throughout) and related electronics. The space
around the eyeball in the middle is the test section, or the sample volume. The
water being measured is the water in this space. For laboratory work, you will
form a test chamber to hold water around the eyeball.
Two pressure windows can be seen on the two cases on either side of the
eyeball. One of these windows is clear. This is the transmit side window. The
other is dark. The dark window has an ND 1 filter cemented to it. This filter
reduces backscatter (glow) from optics on the small-angle optics side,
permitting measurements with the eyeball to about 155°.
A scheduling computer is placed alongside eyeball electronics.
Communication with this computer is over a serial line. The LISST-VSF
instrument can be programmed from a PC to execute a specific data
collection sequence. Software for programming the instrument and recovery
of data is provided. Data download is also carried out using the provided
software.
Measurement
Section
The eyeball is set to one side of the sample volume. This positioning permits
the eyeball to view scattering from about 10o onward to about 155°. The data
storage electronics are programmed to start capturing scattered light data
when the eyeball is positioned to view 5o scattering. (The precise angle varies
from instrument to instrument due to the setting of the optical encoder. The
offset in this angle is included in your calibration files). Ring detectors and the
eyeball, both cover the 5-15-degree scattering angle range. The overlap
permits calibrating the eyeball data with the outermost ring detector, thus
providing absolute calibration for photomultiplier signals. The 5-15 degree
range of data from the eyeball also includes the laser spot on the transmit
window. This spot glows brightly, thus offering a fixed target to further
calibrate/verify eyeball viewing angles.
LISST-VSF User’s Guide 15
The rectangular box below the eyeball in the photo is the housing for the
optical encoder which keeps track of eyeball angular position.
Figure 1 – In laboratory work, make sure the rectangular box is at top!
Laboratory and
Field Use
The instrument can be used in the laboratory or in the field. It can be operated
with an External Power Supply, or using the supplied rechargeable battery.
In battery-powered field usage, the instrument can be used from a wire in a
profiling mode, or it can be mounted on a tripod or mooring for a time-series
observation.
NEVER INSERT ANYTHING REFLECTIVE IN THIS SPACE. THE LASER
POWER IS APPROXIMATELY 10mW. IT WILL DAMAGE EYES.
The windows and endcap faces should be clean if doing laboratory work. If
deploying in the ocean, make sure both windows are clean. Use lens cleaner
or soap and water. DO NOT USE ABRASIVE CLEANERS.
Rechargeable LiIon Battery Pack
The instrument power source is a rechargeable lithium pack. With all systems
on, the instrument draws about 0.7 A current. The fully charged battery can
power continuous operation of the instrument for up to ~40 hours. The
instrument checks the state of battery before capturing each set of data. If the
battery is low, a graceful shutdown is executed.
A battery charger is provided. While the battery is being charged, it cannot power the instrument.
A separate external power supply, with an underwater connector that mates
with the instrument housing is provided for use in the laboratory.
16 LISST-VSF User’s Guide
I.2 How It All Works
Overview
After connecting power (battery or external power) and the communications
cable, the instrument becomes ready for operation. When commanded to
capture data, the following sequence happens: the laser is turned on for a
period to reach stable power. beam emerges from the transmit window and
the eyeball starts rotating. Each rotation takes about 2 seconds. During this
time eyeball data are captured. The sequence is as follows: beginning when
the optical encoder reads 15-degrees, the two PMT data are captured at 1degree intervals. Between 15 degree and 50-degrees of eyeball position, the
laser is dimmed by the laser controller. This is done because this scattering is
stronger; dimming keeps it within the PMT dynamic range. At 51 degrees
encoder position laser power is returned to full. Eyeball data are recorded
from the start angle for a specified number of angles, typically 150.
Ambient Light
Rejection
The eyeball continues the ‘blind’ part of its rotation. During this time, rings
data are captured. After ring data have been captured, the combined ring and
eyeball data are written to data file. Immediately, a half-wave plate is inserted
in the laser beam, so that during the next turn, PMT’s data would be captured
with laser polarization rotated by 90o. During the time allocated for the halfwave plate insertion and settling of its mechanism (see figure), the laser is
turned off briefly to capture an ambient light background from rings. The cycle
repeats for the second rotation, only now the half-wave plate is removed. This
is illustrated in Fig.2.
.
Figure 2 – Laser power through 1 set of 2 rotations (shading shows laser
modulation during eyeball portion of data capture), HW is half-wave plate to
rotate polarization of laser, Note that laser power is reduced for the 5-50o part
of eyeball data capture, and laser is extinguished to read dark rings (ambient
light on rings) after activating HW.
Also stored in each rotation are auxiliary parameters - depth and temperature,
battery voltage, and date and time.
Each data set requires two rotations of the eyeball. The number of sets of
data is controlled from the command window or from the operating modes
menu. This is described in later chapters.
Black Glass
Receive Window
Why is an ND filter used on the small-angle optics window? This is because
internal glow from the small-angle optics contaminates particle scattering at
large angles, i.e. when the eyeball is viewing scattering at around 130o or
greater. The ND filter attenuates the glow.
LISST-VSF User’s Guide 17
Background
Light
Measurement
As with the LISST-100X instrument, so also with this instrument, a
background measurement is necessary. Background light for ring detectors
originates as scattering from optical surfaces. For eyeball data, background
light from scattering by water itself can be significant. Ambient light does not
contribute to background as it is rejected by laser chopping. A background
reading is required before any data collection. The background is measured
with flitered clean water in the sample volume. The measured background is
subtracted from particle data in data processing.
Dynamic Range
Challenge
Changing PMT
sensitivity
Because the VSF typically requires a wide dynamic range, it is not reasonable
to build a fully general purpose system. By using a combination of 12-bit A/D
for PMT outputs, and a factor of ~25 extension of the upper limit on scattered
light by dimming the laser at small scattering angles, we achieve about 16 bits
dynamic range. This is sometimes not sufficient. If so, it is possible to
increase or decrease sensitivity of PMT’s by changing the control voltage.
This is possible from the command window. The instrument is shipped with a
setting that is found suitable by us. Because saturation of the A/D is possible,
we recommend that you always view the raw data and look for saturation. A
function view_rawfile is provided for this purpose.
PMT sensitivity may be changed from the Terminal Window by typing
~c
The instrument will display a range of settings. Just hit enter for all, until the
PMT setting is displayed. Move it up or down 50 counts at a time to,
respectively, increase or decrease sensitivity.
Low Battery
Cutoff
The instrument checks for low-battery during sampling. If the battery is low, it
shuts off the high-current components such as the eyeball motor and laser in
an orderly fashion. When shut off in this mode, notification is printed to the
communications port, and residual battery power is used to return the system
to the command prompt.
Excess Depth
Shut Off
The instrument, when deployed in profiling mode, automatically checks for
depth exceeding maximum permissible for eyeball operation. If the critical
depth is exceeded, the eyeball is powered down and data collection pauses
until depth is again within specifications.
Data Processing
Data processing is performed in MATLAB. Software is provided to perform
these functions: (i)view any data file or background file with view_rawfile, (ii)
construct a mean background file from recently acquired background (pure
water) data, with make_zsc, and, finally, (iii) to produce the desired end
result: P11, P12, and P22, beam_c and beam_b with make_P. For convenience,
a few other functions are provided, e.g. to simulate eyeball data from Mie
theory. The expected measurement from, say, single size polystyrene beads
can be computed using this function for comparison with actual data.
Appendix B explains the methods. A listing of provided functions is in
Appendix C. The use of MATLAB permits flexibility to the scientific user.
18 LISST-VSF User’s Guide
A Tweak-able
Parameter
The relative gain of the two photomultipliers is required in solving for P11 etc.
This quantity cannot be reliably measured directly. Fortunately, the outputs of
the two PMT’s are theoretically exactly equal at scattering angles of 45
o
and
135o [see Appendix B, Eq.2b]. This permits estimating this ratio, which we call
from the data. However, error in estimation of can result in spikes in
estimates of P22. Thus, adjusting may be required. The value of at which
the spikes are minimized is relatively easy to find.
It is not correct to use the same value of at all times since it will depend on
PMT voltage, and we do not know if it drifts significantly over time.
NOTE: A version of the LISST-VSF MATLAB software tools that allows
simple modification of during data processing is under development!
A Quick Look at
End Results
For details on data processing, see the following Data Processing Overview
section.
If all goes well, you should see a composite VSF as shown below: a log-log
plot and a semi-log plot. Note that when working with small particles, the VSF
at the smallest angles (below left) may not be smooth. This is because the
amount of light on the smallest ring-detectors is very small, hence noisy. The
example below shows 10 sets (curves) from 1.6 micron polystyrene beads.
Figure 3 – Laboratory data from 1.6 micron polystyrene beads.
In general, field work will not produce such difficulty with smallest angles. If
you see this difficulty, and if turbulent scintillation is unlikely, it is possible to
use symmetry and flatten the curves. We make this suggestion only. Please
use your discretion on what is reasonable.
LISST-VSF User’s Guide 19
The software will also produce a display of P12, as well as an estimate of P22
which has a problem around 45 and 135-degrees, explained in Appendix B.
P22 is unity for spherical particles, which the example below shows
reasonably.
The processed data is summarized in a 2x2 plot containing the P11, P12, P22 as
well as the near-forward VSF derived from the ring detectors.
Figure 4 – Summary of LISST-VSF processed data
See Appendix B for explanation of spikes in result and how to minimize
them.
050100150
10
-2
10
-1
10
0
10
1
Angle [deg]
Eyeball P
11
[m
-1
sr
-1
]
10
-2
10
0
10
2
10
0
10
1
10
2
10
3
Angle [deg]
Near-forward VSF [m
-1
sr
-1
]
050100150
-0.5
0
0.5
1
Angle [deg]
Eyeball P
12
[m
-1
sr
-1
]
50100150
-1
0
1
2
3
Angle [deg]
Eyeball P
22
[m
-1
sr
-1
]
20 LISST-VSF User’s Guide
I.3 General Precautions
Handling the
Instrument
The LISST-VSF is a delicate instrument. It contains precision, highly
aligned optics, and electronics. It is designed for the rigors of field use,
however it should be handled with care at all times.
AVOID SHOCK at all times.
Do Not Open
Instrument Case
You should never have to open the LISST-VSF or battery pack pressure
cases. Doing so may void the Manufacturer’s Warranty!
Connecting Cables
Always connect this cable to the instrument first, then connect the other
end to the battery pack, bench-top power source, or PC.
Use Care When
Charging Battery
When closing the charging port after finishing recharging batteries,
make sure the plug is fully tight.
A BATTERY CHARGER IS PROVIDED WITH THE LISST-VSF.
NEVER USE ANY OTHER CHARGER FOR BATTERY CHARGING.
THIS MAY DAMAGE THE BATTERY AND VOID THE WARRANTY.
EXTREME CAUTION IS ADVISED IN CLOSING THE CHARGING
PORT AFTER BATTERY CHARGING.
CAUTION
RISK OF FIRE
WARNING
The LISST-VSF uses a powerful green laser emitting a
maximum of 10 mW of visible light at a wavelength of 515 nm.
The laser beam under normal circumstances is not a threat.
However, if objects are placed in the path of the laser beam, the
light could be reflected into an eye causing permanent damage.
CAUTION
LASER RADIATION
LISST-VSF User’s Guide 21
I.4 Quick Start
Contents of
shipping case
So, now you are ready to go!
Open the shipping case containing the instrument. You should see the
following items inside (see also Appendix H):
1. Lexan ZSCAT Sheet
2. ZSCAT clamps (2)
3. Communications Cable
4. Battery Cable
5. Battery Charger
6. Power Supply
7. Stands (4)
8. Installation Disk
9. Manual
10. Sediment Samples
11. Spare O-rings
12. Mounting Clamps (4 sets)
13. Battery housing
14. LISST-VSF instrument
15. Test dusts in small quantities
Set Up on the
Bench Top
1. First place the instrument stands approximately 1m apart.
2. Remove the instrument and cradle it on the stands.
3. Do the same with the battery case.
Do not connect the two instrument cases with the power cable yet! Let
us become familiar with the various parts first.
No end-caps ever need to be opened by the user. Only recharging
a battery requires opening a small access port on the battery
case. The user should never try to loosen or disassemble any of
the other components attached to either endcap. Doing so could
immediately void any warranty.
Instrument and
Battery Connector
Endcaps
There are 3 connectors on the
endcap. The 3-pin connector is for
power (from battery or line power
supply). The 5-pin connector is for
communication. The 6-pin
connnector is for auxiliary use.
See Appendix E.
22 LISST-VSF User’s Guide
Battery Housing
and Endcap
There is only one connector on the
battery housing endcap. This is used
for powering the instrument or for
charging the battery. The instrument
can not be operated with the battery
while the battery is being charged.
See Appendix E for pin-outs..
Check for Clean
Windows
In preparation for getting started, at this time, check the optical windows to
make sure that they are clean. There are three windows: Two are on the inner
endcaps, and one is on the eyeball.
The best way to check the windows is by using a flashlight. By shining light
from one side, almost parallel to the window surface, and viewing from the
other, the surface of the windows can be easily checked for cleanliness. If
there is dirt or fingerprints on the windows clean them first by rinsing them
with lukewarm water and a mild soap solution (e.g. mild hand soap, liquid dish
soap) and then rinsing off all soap residue with clean, particle free water such
as de-ionized water, MilliQ water or distilled water. The windows can also be
wiped clean with a soft cloth (e.g. a lens cloth) and alcohol. It is not
recommended to use stronger solvents, such as acetone or toluene. Also, do
not use any abrasive cleaners or wipes. Treat the windows as you would an
expensive camera lens.
Attach
Communications
Cable
Remove the Communications cable from the
shipping crate. It is the 2 meter cable with the 9pin DB-9 connector on one end and the 5-pin
underwater connector on the other. Remove the
underwater cap from the Communications
connector. The connectors will all look similar.
The Communication connector is the only 5-pin
connector. It is located below the engraved
serial number. After removing the cap install the
cable making sure that proper alignment of the
cable is maintained, so that the connectors are
not bent.
Install LISST-VSF
Software
The software is similar to our LISST-100X instrument, with a few changes.
Simply insert the CD in your PC’s CD drive (or connect the USB disk) and
follow along. A program icon will appear on your desktop and under the
Sequoia folder in your All Programs menu.
The MATLAB *.m files provided on the installation CD (or USB drive) will be
copied to your working folder for data processing.
LISST-VSF User’s Guide 23
Form the
Laboratory Test
Chamber
The LISST-VSF is provided with a bendable clear plastic sheet with sealing
gaskets and quick-release band clamps. This chamber has the advantages of
quick and easy installation and the ability to inspect the sample being
measured. Install with rectangular encoder housing facing up. Fill with clean,
particle-free water (0.2 micron filtered).
Check out the Sequoia Scientific YouTube channel for a video showing
how to assemble the flexible test chamber.
http://www.youtube.com/user/SequoiaScientific
NOTE: An improved test chamber for bench-top and flow-through use is in
development.
Start LISST-VSF
Program
Launch the software by double-clicking the icon on your desktop. If
the icon does not exist, check that the software installed properly.
Connect Power
Connect the LISST-VSF to the battery pack with the supplied underwater
cable, or to the bench-top power supply. The instrument is now live! It will now
respond to <enter> command from the PC.
Open Terminal
Window
Under the Communications menu, select Serial Settings and verify that the
serial port listed is the port to which the instrument is connected.
On the menu bar, click on the icon that resembles a terminal window. This
will open a window within the software window. If you hit <enter> on the
PC, the instrument will respond with a LISST-VSF> prompt.
Note: The instrument communicates at 9600 baud through this window.
However, when downloading data, the transfer rate automatically changes
to 115k.
Basic Commands
There are basic commands for turning on the laser, powering the eyeball
(so it starts turning) etc. For help, type:
L-VSF:> he
The responses are self-explanatory.
24 LISST-VSF User’s Guide
Acquiring
Background
Scattering File
Now try to acquire a background file. The Laboratory Test Chamber should
be nearly full with clean water (so that both of the windows on inner endcaps are submerged, and the strut receiving the upper eyeball shaft is
partially submerged). Make sure that the water is well-mixed (give it a stir)
else scintillation will affect the small-angle measurements. At the L-VSF:>
prompt, type the command
L-VSF:> za
and hit <enter>. The laser will turn on along with the eyeball. Laser warm
up will take about 15 seconds. Then, 16 sets of eyeball and ring-detector
data will be acquired (recall that each set is 2 turns of the eyeball). When
the instrument is finished collecting the data, it will shut all systems off and
the L-VSF:> prompt will be displayed in the terminal window.
The instrument will save a background file with name ZDDDHHMM.DAT.
The background has 16 sets of data, each set being two turns of the
eyeball, one with perpendicular laser polarization, the second with parallel.
At this point, you may download the background file, or move on to collect
another file, e.g. some particle data. Let’s take some particle data.
Collecting Particle
Data
Drop some particles in the test chamber (use fine particles, e.g. 0.3 micron
or 1 micron polystyrene beads from Duke Scientific, Inc.). Mix the water
well to homogenize the suspension. This is important!
In the command window, type the commands:
L-VSF:> pd 16
L-VSF:> go
This command means grab 16 sets of scans (again, recall each set is two
rotations of the eyeball). When finished, the instrument will shut down, and
you will know that because the laser will be out. The command window will
return to the L-VSF:>prompt.
To collect data in a field deployment, you will need to program the
instrument, i.e. set the starting delay, sample rate, length of data collection
etc. This is described in section.III.2.
Downloading Data
Downloading procedure is identical to our LISST-100X. From the windows
program, click on the download button. A listing of stored files will appear.
Select the ones to download and click OK.
Cleaning the
Instrument
Before putting the instrument away, clean the inner endcaps and windows.
Drain the test chamber, wipe and dry the windows.
LISST-VSF User’s Guide 25
Put Instrument to
Sleep and Waking
it Up
The instrument consumes some battery power when powered on
and not in low power mode. It can be put into a low power mode by
pressing the sleep button on the software menu bar.
Wake up the instrument by clicking the adjacent button.
Note that once put into sleep mode, it can take up to a minute to
wake up the instrument.
26 LISST-VSF User’s Guide
Section II: Operation
Section
Organization
This section is divided into these subheadings:
1. Instrument mounting, storage and maintenance
2. Charging and Connecting the Lithium battery pack
3. Bench-testing and collecting background data
4. Offloading and erasing data files from instrument memory
5. Recording and storing a sample data file
6. Data Processing
LISST-VSF User’s Guide 27
Horizontal
Mounting
It is recommended that the instrument be mounted horizontally while in use or
storage. This ensures that particles will not settle on windows and degrade
data quality.
Precautions for
mounting
Do not mount the instrument in a way that over-constrains it. No
bending moments should arise on the instrument. Bending moments
will degrade alignment, causing loss of data at small angles.
Isolate from other
Metals
To reduce corrosion of aluminum parts a zinc anode is attached to the
Connector endcap. This anode must be exposed to the water for it to be
effective.
When mounting the instrument be sure to electrically isolate the instrument
from all other metal. Any contact with other metal can greatly increase the rate
of corrosion. Isolate the instrument with rubber or plastic to keep the LISSTVSF from becoming the sacrificial anode for the mounting hardware. Failure
to properly isolate the instrument from all other metal will void the warranty.
A set of stainless steel clamps with rubberized interiors ships are part of the
instrument package. Use these or the plastic clamps for mounting, and make
sure that no metal is in direct contact with the pressure housing or other
components of the instrument. If you lose the spare clamps, replacement
clamps are available for purchase from Sequoia. See Appendix H for:
Accessories.
Storage Notes
The LISST-VSF is a sensitive instrument. When not in use, the instrument
should be stored in its shipping case.
Clean and dry the instrument before storage.
Avoid storing the instrument in excessive heat or humidity. Do not store the
instrument in freezing temperatures, or in environments with wide temperature
fluctuations.
When storing the instrument, unplug the battery pack. It is recommended that
the venting plug on the battery case be left open for any gases to escape.
Low Power Sleep
Mode
To put the instrument into low power sleep mode choose Put LISST to Sleep
from the LISST menu or click the button.
II.1 Instrument Mounting, Storage and Maintenance
28 LISST-VSF User’s Guide
Cleaning the
optical windows
The condition of the windows is critical to the performance of the LISST-VSF.
Care must be taken when cleaning the windows. The windows and the
instrument should be rinsed thoroughly with fresh water after each
deployment to ensure that salts and sediments will not scratch the window
during the cleaning process. The windows should be cleaned with a soft cloth
or lens tissue, using a single direction wipe across the window. DO NOT
SCRUB. Detergents such as Liquinox and water may be used to increase the
efficiency of cleaning. For removing grease spots, finger prints, etc, alcohol
may be used. Do not use stronger solvents such as Acetone or Toluene.
Abrasive powders must never be used near the optics windows;
they will scratch the windows and degrade instrument
performance.
O-rings
User should not need to open any endcap on LISST-VSF
Calibrations and
adjustments
The LISST-VSF does not require any adjustment or calibrations. The pressure
and temperature sensors can be re-calibrated if desired. The pressure sensor
can be zero’ed at the surface before deployment (see ZD command) The
performance of the instrument can be checked with the use of a sample of
particles of a known size distribution.
Standard particles can be obtained from manufacturers of standard particles,
such as:
During prolonged usage, it is recommended that a fan be used on the VSF
housing to reduce heat buildup, or operate the instrument in a water bath.
LISST-VSF User’s Guide 29
About the
Recharge-able
Battery Pack
The LISST-VSF is supplied with a submersible, recharge-able, Lithium-Ion
battery back.
Dimensions 13.3 cm (5.25”) diameter, 71.1 cm (28.0”) L
Weight: 10.3 kg (22.8 lbs) in air
Capacity: 14.8V, 39 A-hr.
The battery pack connects to the LISST-VSF by way of a 3-pin wet mate-able
connector (bulkhead connection is MCBH-3F). This same connector is also
used to connect the battery pack to the charger.
The battery pack bulkhead also contains a zinc anode and special screw-in
plug for venting while charging.
Charging the
Battery Pack
The Lithium-Ion cells contained in
the pack must be vented while being
charged. Unscrew the vent plug until
the small vent hole is exposed. The
vent plug cannot be unscrewed
completely from the unit due to a
mechanical stop, and an internal
switch will prevent the batteries from
charging if the plug is not unscrewed
to the venting position.
Connect the supplied battery charger to the battery
pack and allow 8-10 hours of charge time.
Lithium-Ion batteries prefer partial discharge to deep
discharge, so avoid depleting the battery fully. LithiumIon batteries do not have a “memory”, so the cells are
not harmed with partial charge. Deep discharge can
damage the Li-Ion cells.
Lithium-ion batteries age. Typical lifetime is ~3 yrs.
Completion of charging is indicated by LED status indicators on the battery
charger.
A BATTERY CHARGER IS PROVIDED WITH THE LISST-VSF.
NEVER USE ANY OTHER CHARGER FOR BATTERY CHARGING. THIS
MAY DAMAGE THE BATTERY AND VOID THE WARRANTY.
USE OF OTHER CHARGERS OR IMPROPER VENTING CAN RESULT IN
DAMAGE TO BATTERIES OR FIRE. USE CAUTION AND RESPECT
LITHIUM-ION BATTERIES!
II.2 Charging and Connecting the Lithium-Ion Battery Pack
30 LISST-VSF User’s Guide
After charging is complete, the vent plug must be closed before the battery
can be used.
EXTREME CAUTION IS ADVISED IN CLOSING THE CHARGING PORT
AFTER BATTERY CHARGING. FAILURE TO CLOSE PROPERLY CAN
RESULT IN FLOODING, PERMANENT DAMAGE, AND EVEN FIRE.
LISST-VSF User’s Guide 31
Bench Setup
The LISST-VSF instrument should be set up on the two white plastic
mounting blocks, placed roughly between the middle of each pressure case
on either side of the eyeball.
Make sure that you do not add bending moments on the pressure case.
Bending will cause loss of alignment, and loss of data at small angles. Photos
below show instrument on a test bench, with one plastic support block
showing on left.
Install Sample
Chamber
The LISST-VSF is provided with a bendable clear plastic sheet with sealing
gaskets and quick-release band clamps. This chamber has the advantages of
quick and easy installation and the ability to inspect the sample being
measured.
Wrap the plastic sheet around the sample volume and install clamps over the
gasket strips to secure the chamber to the LISST-VSF instrument. Verify that
that there are no leaks. Adjust clamps and/or replace gasket material if
needed.
II.3 Bench Testing and Collecting Background Data
32 LISST-VSF User’s Guide
Fill with Clean
Water
Fill the sample volume with organic-free, degassed, particle-free water.
Clean water should reach a level approximately 1-2 cm above the optical
windows. Sequoia recommends using distilled or reverse-osmosis water
which has undergone 0.2 micron (or smaller) filtration. Superior lab results
have been obtained by continuously circulating (pumping) the clean water
contained in the sample volume through a 0.2 micron capsule filter (e.g.,
Whatman PTFE, PALL Supor, GE Memtrex) for ~1 hour before sampling.
Check Instrument
Alignment and
Water Cleanliness
Connect communications cable. Connect battery or benchtop power supply.
Run LISST-VSF software and open Terminal Window.
Connect to the instrument by clicking on the terminal window icon
or selecting Terminal Window from the Communications menu.
If the data acquisition program is still running, stop its execution
with a <CTRL-C> or pressing the Stop button in the terminal
window. The instrument will display a L-VSF:> prompt.
Select Check Instrument Alignment from the LISST menu or by
pressing the corresponding button on the toolbar. After selecting
the factory background scatter file to use for comparison the
window will be displayed. Note: this is similar to the Collect
Background functionality of the 100X, but is used in the LISSTVSF only as a simple way to verify instrument operation,
alignment, and clean water quality.
LISST-VSF User’s Guide 33
Make sure that the water is well-mixed (give it a stir) otherwise scintillation will
affect the small-angle measurements.
Press the BEGIN Collect button. 20 samples will be displayed to the screen
as they are acquired. After all 20 are collected the average will be displayed.
The values can be saved to a file for later reference. Enter the filename in the
box and press the Accept and Save button to save the file for future
reference.
If there is a problem with the instrument or if the water or windows are not
clean, error messages will be displayed.
Dirty water or windows will generally cause higher values across the
middle rings.
Large bubbles or particles in the water can cause higher values on the
inner rings or left hand side of the display.
High values on the inner rings combined with a lower Laser power value
can also be an indication of optical misalignment.
Record and Store
Background
Scattering
“ZSCAT”
Connect communications cable. Connect battery or benchtop power supply.
Run LISST-VSF software and open Terminal Window. Verify instrument
communications by issuing DS command in the terminal window.
Clean instrument, install test chamber, fill with clean filtered water as
described in Check Instrument Alignment and Water Cleanliness (page
33).
The LISST-VSF uses a dark ring detector reading to reduce sensitivity to
ambient light. However, it is suggested that the sample volume be covered
with a felt sheet or other opaque barrier during sampling.
Issue ZA command to start acquisition of background scattering file. After
brief warm-up period, 16 sets of scans will be saved to a file Zdddhhmm.DAT,
where ddd, hh, and mm, are the day of year, hour, and minute, respectively.
34 LISST-VSF User’s Guide
Connect
Instrument and
Start LISST-VSF
Software
Connect communications cable. Connect battery or benchtop power supply.
Run LISST-VSF software and open Terminal Window.
Connect to the instrument by clicking on the terminal window icon
or selecting Terminal Window from the Communications menu.
If the data acquisition program is still running, stop its execution
with a <CTRL-C> or pressing the Stop button in the terminal
window. The instrument will display a L-VSF:> prompt.
Select Offload from the LISST menu or press the Offload button.
Select Files to
Offload
Choose the files to be offloaded by clicking on the file name on the list. Multiple
files can be selected by hold down the CTRL key while clicking on files. Use the
SHIFT key to select a range of files.
Press the OK button to accept the current selection.
Select Download
Location
A dialog box with a path for storing the downloaded data will appear. Edit the
path or press on the Browse button to select a new path. Press OK to begin the
offloading.
II.4 Offloading and Erasing Data Files from Instrument
LISST-VSF User’s Guide 35
The file names will remain the same as on the compact flash card. The file
creation date and time will not be preserved.
A Status bar will be displayed for each file offloaded. Text in the lower left
corner will display the current file being offloaded.
Changing Offload
Baud Rate
The standard offload baud rate is 115K. If you are using a long offload cable
(i.e., >20 m cable), this value may be too high and cause errors during
offloading. If an offload error occurs, you can change the offload baud rate as
follows:
Go to Settings in the File Menu and click the Serial Port tab. In the File Offload
Baud Rate drop down menu, choose a lower baud rate and try again.
Deleting Files
from LISST-VSF
Memory
When connected to instrument as described above:
Select Erase from the LISST menu or press the Erase button. A
list of files on the LISST-VSF datalogger will appear.
36 LISST-VSF User’s Guide
Select the files to be deleted, then click the OK button. Use the CTRL key to
select multiple files or the SHIFT key to select a range of files.
Confirm the files to be deleted by
clicking the Yes button.
WARNING: Once a file has been
deleted there is no way to recover
the data. Make sure that the file
has been properly offloaded
before deleting any files
LISST-VSF User’s Guide 37
In the Laboratory
Connect communications cable. Connect battery or benchtop power supply.
Run LISST-VSF software and open Terminal Window. Verify instrument
communications by issuing DS command in the terminal window.
Clean instrument, install test chamber, fill with clean filtered water as
described in Check Instrument Alignment and Water Cleanliness (page
33).
The LISST-VSF uses a dark ring detector reading to reduce sensitivity to
ambient light. However, it is suggested that the sample volume be covered
with a felt sheet or other opaque barrier during sampling.
Set the sampling parameters to start immediately on command, save a
specified number of measurements, and stop sampling by issuing the
following commands:
Command
Response
ST 5<ENTER>
New Start Condition Setting: Delay Start
TD 0<ENTER>
New Start Condition data = 0
Start Condition: Delay Start with 0 minute
delay
SP 5<ENTER>
New Stop Condition Setting: Fixed Number Stop
PD 20<ENTER>
New Stop Condition data = 20
Stop Condition: Fixed Number Stop at 20
samples
This configures the LISST-VSF to collect 20 sets of scattering data (each
including both laser polarization states). The number specified with the PD
command can be changed to collect a different number of sets.
Finally, issue the GO command to start acquisition of background scattering
file. After brief warm-up period, 20 sets of scans will be saved to a file
Zdddhhmm.DAT, where ddd, hh, and mm, are the day of year, hour, and
minute, respectively.
Alternately, issue the command “GO nn”, where nn is the number of sets to
save.
Offloading and
Processing the
Acquired
Scattering File
The data offloading procedure is identical to that described in Section II.4.
For data processing, see the following, Section II.6.
Configuring
Sampling
Parameters for
Field Deployment
See Section III.2.
II.5 Recording and Storing a Sample Data File
38 LISST-VSF User’s Guide
LISST-VSF User’s Guide 39
II.6 Data Processing
Datafile Format
The datafile consists of sets of data. Each set contains data for two rotations of
the eyeball. Each rotation contains the following variables:
Variable no. Variable Name
1-32 Rings
33 Laser transmission
34 Battery voltage
35 PMT control voltage
36 Laser Transmission
37 Depth
38 Temperature
39-40 Date and Time
41-791 150 pairs of [angle PMT1on PMT1off PMT2on PMT2off]
For more information, see Appendix D.
Data
Processing
Sequence
The MATLAB functions provided for processing LISST-VSF data files will
prompt you for filenames. In addition to the background and data files, you will
be prompted for a Factory background file, and a calibration factors file. These
files have the names, respectively,
Factory_zsc_xxxx.asc and cal_factorsXXXX.mat
where XXXX is instrument serial number.
These files should be in the same folder where your MATLAB software is
installed. The data processing involves these steps (this description is intended
to help you follow the logic):
.
1. Reading the datafile;
2. Reading the corresponding background file;
3. Subtracting the background from the data;
4. Find the laser power change factor at 50th data in eyeball rotation
5. Find the relative gain of the two PMT’s
6. Calculate the eyeball p11,p12,and p22;
7. Calculate ring-detectors net scattering (similar to LISST-100X)
8. Apply ring area and vignetting corrections to ring data
9. Match the rings and eyeball data at 15-deg scattering angle
10. Compute composite P11, as [small angle and eyeball angle]
11. Plots of VSF, p12 and p22.
Although the software is tested, you may be required to fine-tune the PMT
relative gain factor as was noted earlier.
40 LISST-VSF User’s Guide
One-Step
Processing
A single command in MATLAB is set up to process datafiles and produce the
VSF and depolarization parameters. This assumes that the datafile is good.
The command is:
make_P.m
It outputs the eyeball VSF(p11), the angles corresponding to p11, the full VSF,
P11 (combined from eyeball and rings data), and the beam attenuation and
scattering coefficients respectively beam_c and beam_b. Also output are depth,
temperature, battery voltage, and PMT control voltage, along with the date and
time stamps on the data.
Several plots are output during the processing, and the make_P function is
described in more detail in the In Depth Processing section below.
In Depth Data
Processing
The following sub-sections describe a more detailed processing of LISST-VSF
data.
The following command reads and displays a raw LISST-VSF data file. The raw
data are also passed as output parameters from the function.
view_rawfile.m
When this script is run, a dialog box will appear for selecting the LISST-VSF
Data File (*.DAT) downloaded from the instrument
The output variables rp, pp, rr, and pr represent PMT signal outputs for the two
laser polarizations, at the 150 angles also output by view_rawfile. Note that the
angles contained in the raw data are offset from the true angles. The naming
convention is that the first letter indicates laser polarization and the second
denotes the PMT (analyzer) polarization. Letter r is used for perpendicular, and
p for parallel.
These PMT signals are also displayed by view_readfile:
Figure 5 – Log-plots of net scattering from both PMT’s, both laser polarizations.
This is the net scattering seen by the PMT’s (i.e. PMT_on – PMT_off) for the
perpendicular polarized laser (top pair) and parallel polarized laser. Note the
low amount of scattered light at ~90-100° for rr and pp.
PMT Saturation?
Saturation of PMT’s is indicated with red dots in these plots (not shown).
Data with saturated PMT’s will produce erroneous results. If some of your data
are found to be saturated, we do not know of any recourse but to extrapolate
the missing (saturated) data.
Saturation can be avoided by lowering the PMT control voltage. This is done by
using the ~c command in the Terminal window. A lot of variables are displayed.
When the PMT control is displayed, adjust is up or down in steps of 50.
42 LISST-VSF User’s Guide
Ring Detector
Output
The raw ring scattering for the 32 detectors for each laser polarization is output
in the rings1 and rings2 variables.
The left and right plots are for perpendicular and parallel polarization of the
laser. Minor differences are due to change in concentration and density
fluctuations in the sample volume, and due to slightly weaker laser when the
half-wave plate is inserted in the beam.
Figure 6 – Output of ring detectors for the two polarizations of laser.
Auxiliary Data
(i) Depth (top left), (ii) laser reference (red) and transmission (bottom, left), (iii)
temperature (top, right) and (iv) the ratio of laser transmission/laser reference
(bottom right). The last plot serves as indicator of homogeneity of the
suspension.
Figure 7 – Auxiliary data (top) and laser reference and transmission.
LISST-VSF User’s Guide 43
read_rawfile.m
Similar to view_rawfile, this command reads a raw datafile, passes the
scattering, angle and aux data as output parameters, but does not provide
output plots of the output data (most commonly, this command is used in the
make_P function to read in a file without displaying extra plots of the raw input
data).
Usage:
Depending on a variable within the function (plot_raw_onoff), it can display
diagnostic plots of the raw data measured by the PMT which is used to
calculate the net scattering signals rp, rr, pp, and pr.
If plot_raw_onoff is enabled, the raw data showing laser on and laser off values
for all signals will be shown. Each eyeball rotation (laser polarization) is shown
in a separate plot (only the first for rp and rr is shown below). In addition, it
shows the net scattering (i.e. PMT_on – PMT_off). This view shows a
discontinuity in PMT outputs at 50-degree nominal eyeball angle where the
laser power is un-dimmed.
Figure 8 – Outputs of two PMT’s for two laser polarizations.
This function reads a user-specified background raw datafile, and saves means
of rr, rp, pp, pr and rings1 and rings2 to a .mat file. The key purpose of this
function is to display current backgrounds in contrast to the factory
backgrounds.
Usage:
The output parameters are the median values of the data contained . These are
saved in a file of the same name as the input background file, but of type *.mat.
The make_zsc function will prompt for the LISST-VSF raw data file to be
processed. Only files matching Z*.DAT will be shown by default. All *.DAT files
can be displayed in the Open Background Scattering File dialog by selecting it
from the “Files of type” list.
Following selection of the background file to be processed, the function will
prompt for a factory background file for comparison. The factory background file
(Factory_zsc_NNNN.mat) file is provided with the LISST-VSF MATLAB
processing tools.
If no comparison of the current background to the factory background is
needed, this dialog may be cancelled.
LISST-VSF background scattering will be read from the input file and displayed.
Four signals (rp, rr, pp, pr) are displayed, along with factory values for
comparison.
Figure 9 – Comparison of current and factory eyeball background light.
Such a comparison (eyeball outputs above) guides the user if the water is not
clean enough and if another background file is warranted. The spike at small
angles is where the laser spot emerges into water at the window. Remaining
values are typically <100 or so out of a possible maximum value of 32,768.
Some small negative values may exist in the mean background. It is not a
problem.
Detailed
Walkthrough of
make_P.m
You may run the make_P function in debug mode to see details of processing.
It is here that the jump in raw data at 50o (seen with the view_rawfile command)
is removed; laser power variation is normalized out, sample volume stretching
is compensated, etc.
Several input files must be provided to make_P, selectable with typical
File/Open dialog boxes:
050100150
10
0
10
2
10
4
10
6
Angle [deg]
Measured rp [counts]
050100150
10
0
10
2
10
4
10
6
Angle [deg]
Measured rr [counts]
050100150
10
0
10
2
10
4
10
6
Angle [deg]
Measured pp [counts]
050100150
10
0
10
2
10
4
10
6
Angle [deg]
Measured pr [counts]
Current
Factory
46 LISST-VSF User’s Guide
Binary Data File to Process:
This is the raw file downloaded from the LISST-VSF containing scattering data.
By default, files named (V*.DAT) saved from the GO command are listed. Use
the Files of type dropdown box to select any file *.DAT.
Background Scattering File:
This file is the clean water background scattering file corresponding to the
particle scattering file previously selected. This is the background scattering file
saved via the ZA command. Processed or raw file formats can be used to load
the background scattering data:
LISST-VSF Processed Background File (*.MAT) – Processed median
background file saved with the make_zsc function
LISST-VSF Unprocessed Background File (Z*.DAT) and LISST-VSF
Binary Data File (*.DAT) – Raw scattering file saved using the ZA
command, downloaded and not processed with make_zsc
LISST-VSF User’s Guide 47
FACTORY Background Scattering File:
This is the factory background scattering file provided by Sequoia, named
Factory_ZSC_XXXX.mat, where XXXX is the serial number of the instrument.
FACTORY Calibration Factors File:
This is the calibration factors file provided by Sequoia, named
Cal_factorsXXXX.mat, where XXXX is the serial number of the instrument.
48 LISST-VSF User’s Guide
Depending on the settings in read_rawfile and the selection of input files,
several plots may be shown in addition to those described below.
Estimates of for all sets of rotations of eyeball: (a small spread is ideal)
Figure 10 – Estimates of from data.
Recall that alpha is the ratio of sensitivity of the two PMT’s. In principle, it
should be fixed. Until we know better, we find the ratio from the data for each of
the sets of rotations, and take a median value.
A check of how well the chosen value of applies is shown next.
How well the estimated matches PMT’s at 45 and 135 (should be unity)
Figure 11 – How well PMT’s match at 45 and 135 with the selected .
LISST-VSF User’s Guide 49
Corrected PMT Signals (after all adjustments): Signals shown are rp(top
left), pp (bottom left), rr(top right) and pr (bottom right). The rising left edge on
all is the part of eyeball that is not seeing particle scattering.
Figure 12 – Corrected data after removing laser attenuation for <50°
The VSF, p12, and p22: The eyeball estimate of VSF is at top left, the rings
VSF is top right, p12 is bottom left and p22 is bottom right.
Figure 13 - Eyeball data after all corrections, before merging with rings.
Note that p22 is unstable at 45 and 135o as explained in Appendix B.
50 LISST-VSF User’s Guide
Rings data are processed in exactly the same way as with our LISST-100X
data. The VSF constructed with rings data is computed. Finally, a factor is
found that scales eyeball data at 15-deg to rings data at 15-degree. That
completes calculation of VSF and p12 and p22.
The Composite VSF (rings and eyeball): log-log axis, and log-linear.
Figure 14 –Composite VSF, combining rings and eyeball data on log-log and
log-linear axes. Note: for small particles, rings data are noisy at smallest
angles.
LISST-VSF User’s Guide 51
Adjusting PMT
Relative Gain (α)
“tweaking alpha”
In make_P, an automated procedure is used to find the relative gain () of the
two PMT’s. Noisy data may require adjusting the relative gain. This is the only
adjustable parameter in the LISST-VSF instrument.
To adjust the relative gain, open make_P.m in the MATLAB Editor, and enable
debug mode by uncommenting the keyboard statement to enable a breakpoint,
then save the file (<CTRL>+S). Make sure that cell mode is enabled: select
Enable Cell Mode in the Cell menu.
% keyboard
%% SET BREAKPOINT BY UNCOMMENTING keyboard COMMAND ABOVE
% alpha = 1.8
Execute the make_P function as described previously. Execution will stop, the
MATLAB command prompt will be “K>>” and the current statement will be
highlighted by in the editor. Place the cursor in the cell below, for example on
the % alpha = 1.8 line and execute the current cell with <CTRL>+<ENTER>.
Examine the PMT ratio plots and P12, P22 results of the estimated alpha. If
“tweaking” of the gain is desired, uncomment the alpha = 1.8 (or similar).
Modify the value for alpha, and then re-execute the current cell with
<CTRL>+<ENTER>.
The effects of changing alpha can also be examined by using the
increment/decrement value function in the MATLAB editor. Place the cursor on
the alpha = 1.8 line, set the increment/decrement by value to the desired
amount (0.1 works well), then use the and buttons on the toolbar to
modify the value of alpha, and automatically recalculate and display the plots of
PMT signal ratio, P11,P12, and P22.
When finished, press <F5> to continue execution of the script beyond the
breakpoint and current cell.
RE-COMMENT THE keyboard ANDalpha = … LINES
WHEN FINISHED, AND BE SURE TO SAVE CHANGES TO
MAKE_P!
52 LISST-VSF User’s Guide
Other Software
We have provided other software that makes it easy to compare your data with
Mie theory, when working with single size particles. Here’s a list:
sim_data.m
This software calculates the simulated data from the photomultipliers for
particles of known size and refractive index. The output is over 0:0.1:180
degrees. Uage (for diameter 1.06 microns and index 1.566):
This software is provided to compare measured P11 p12 and p22 with Mie
theory for single-size particles.
SUMMARY
To recap:
1. A single step command make_P can be used to process a set of
datafiles comprising of (a) particle data; and (b) background data. The
software asks for the Factory background file and the instrument
calibration factors file.
2. It is possible to view each datafile using the view_file or read_file
functions. Each displays raw data in different details.
3. The break in p22 at 45 and 135o is caused by the nature of these data
with the rotating eyeball.
LISST-VSF User’s Guide 53
Section Organization
The LISST-VSF instrument programming for field use is very similar to
the more familiar LISST-100X. In this section, we first describe this
same method again, i.e. how to program (set up) the instrument before a
field deployment, or even a laboratory run.
In the next section, we describe commands that permit greater control,
at a deeper level.
Serial Port Settings
Communication with the LISST-VSF is via an RS-232C link. A cable
that connects the instrument to a PC has been provided. This cable
connects the 5-pin underwater connector on the instrument to a DB-9
serial port connector. If required, DB-9 to DB-25 pin adapters are
available.
The RS232 link communicates at 9600 baud, 8 data bits, No parity, and
1 stop bit. Data file transfer is done using a YMODEM transfer at 115K
baud. The transfer rate can be changed as described in Changing
Offload Baud Rate in Section II.4.
Communication
Software
Installing the LISST-VSF software will create an icon on your desktop to
launch the software. Once connected, you will normally communicate
with the LISST-VSF using this Windows software that also permits easy
programming and data offload.
We have also provided Motocross software specifically for the purpose
of uploading updates to the instrument firmware which may be provided
from time to time.
Section III Instrument Programming and Command Details
54 LISST-VSF User’s Guide
Using the
Terminal Window
Opening the Terminal Window in the LISST-VSF, or alternate terminal
software such as Motocross with the proper COM port and communications
settings will allow the user to communicate with the instrument. If a program is
already in execution, its progress will be displayed in the terminal window.
The instrument firmware is retained in on-board flash memory and will start
executing when power is applied. After a power up, pressing the ENTER key
will cause an echo of the L-VSF:> prompt. The Stop button will also send the
commands to turn off the main power to the laser and electronics. Pressing
the Stop button when the instrument is not running will not hurt the instrument.
It is recommended to press the Stop button multiple times after stopping a
running program to make sure that the main power is properly shut off.
Pressing the Start button will begin the collection of data using the current
configuration.
Two Letter
Commands
Type “HE” at the L-VSF:> prompt to see a list of two letter commands. Under
most situations the LISST-VSF Windows software terminal window will be
used to configure and operate the LISST-100X. However, there are times
when the LISST-VSF may be interfaced with another data logger or custom
program. For this purpose a set of two letter commands is available to
operate the instrument. See the LISST-VSF Command Summary (Section
III.2) for detailed descriptions of each command.
Start Condition
The LISST-VSF Windows software (Operating Modes under the LISST
menu) or two letter commands (in a terminal window or other RS-232
link) can configure the LISST-VSF with one of five Start conditions:
Depth, Time, External Mechanical Switch, External Digital Input, and
Time Delay. The details of each condition are described below.
Depth Start
The built-in depth sensor of the LISST-VSF is used to check the current
depth to determine if the desired start depth has been exceeded. The
instrument is powered up and 5 measurements of the depth are
averaged over a two-second period. If the depth exceeds the threshold
the program will proceed to the data collection routine. If the depth
does not exceed the threshold the instrument will power down and wait
28 seconds before checking the depth again. The program will
continue checking until the depth is exceeded or until the program is
stopped.
Time Start
The program will check the current time every second and compare it
to the Start Time. If the Start Time is equal to or earlier than the current
time the program will go directly to the data collection routine. It will
continue checking the time until the Start Time is reached or until the
program is stopped.
III.1 Programmed Operation for Field or Laboratory Use
LISST-VSF User’s Guide 55
External
Mechanical Switch
Start
The LISST-VSF has a white plastic lever on the endcap. This lever has
a magnet imbedded in it. This magnet can activate a switch inside the
pressure case. The base program looks at the status of this switch
once a second. If the switch is in the on or “1” positions the program
will go directly to the data collection routine. It will continue checking
the switch status until the switch is moved to the “0” position or until the
program is stopped. When in the “0” position the lever is up against the
zinc anode.
External Digital
Input Start
The LISST-VSF is equipped with auxiliary inputs on the 6-pin
underwater connector. Pin 2 of the 6-pin connector is the DIG 1 input
(see Engineering Drawings). The program will check the status of the
digital input once a second. If the voltage at the DIG 1 input is greater
than 2 volts (relative to Digital Ground, Pin1) the program will go
directly to the data collection routine. It will continue checking the status
of the digital input until voltage exceeds 2 volts or until the program is
stopped.
WARNING: The maximum permissible input voltage on the digital
input line is 3.3 volts.
Time Delay Start
The time delay start condition will cause the program to wait the
specified number of seconds before continuing on to the data collection
routine.
Stop Condition
The LISST-VSF can be configured with one of six Stop conditions:
Depth, Time. External Mechanical Switch, External Digital Input, Fixed
number of samples, and Maximum Memory or Low Battery. The Stop
conditions are checked after each measurement set (two rotations of
the eyeball). The details of each condition are described below.
Depth Stop
The built-in depth sensor of the LISST-VSF is used to check the current
depth to determine if it is less than the desired Stop depth. The
averaged depth from the last sample acquired is used as the current
depth. If the depth is less than the threshold the sampling will stop. If
the depth is not less than the threshold the program will continue
sampling as per the configuration. If the Start Condition is a Depth Start
the program will wait 30 seconds and then return to looking for the
Depth Start Conditions. The delay is to keep the instrument from
starting and stopping too quickly as the instrument is moving up and
down. If the Start Condition is Depth Start the Base program will return
to checking for the Start Conditions. For all other Start Conditions the
when the current depth is less than the threshold the program will stop
and return to the L-VSF:> prompt.
Time Stop
The program will check the current time after each sample or burst and
compare it to the Start Time. If the Start Time is equal to or later than
the current time the program will stop and return to the L-VSF:> prompt.
56 LISST-VSF User’s Guide
External
Mechanical Switch
Stop
After each measurement set the status of the Switch lever is checked.
If the switch lever is in the off or “0” position sampling will stop. If the
Start Condition is a Switch Start the program will return to checking the
start condition. For all other Start conditions the program will stop and
return to the L-VSF:> prompt.
External Digital
Input Stop
The status of the Dig 1 input is checked after each sample or burst. If
the voltage at the input is less than 0.7 volts the sampling will stop. If
the Start Condition is a Digital Input Start the program will return to
checking the start condition. For all other Start conditions the program
will stop and return to the L-VSF:> prompt.
Fixed Number of
Samples Stop
The program will acquire a fixed number of samples before stopping.
The number of samples saved is checked after each measurement set
is saved. When the number of sample to be saved has been reached
the program will stop and return to the L-VSF:> prompt.
Maximum Memory
or Low Battery
Stop
The Maximum Memory or Low Battery Stop condition will continue to
sample until the memory capacity has been reached or when the
battery voltage has dropped to less than 6 volts. The program will
continue to sample until one of these conditions is met. It will then
return to the L-VSF:> prompt.
LISST-VSF User’s Guide 57
III.2 LISST-VSF Command Summary
Display Commands
DB
Display Current Battery Voltage
Display current Disk Directory
DS
Display current status information
DT
Display Current Time and Date
HE
Display general help messages and command list
Setup Commands
OM x
Set Operating mode.
ST x
Set Start Condition.
TD x
Set Start Condition Data
SP x
Set Stop Condition
PD x
Set Stop Condition Data
SI x
Set Sample interval
MA x
Set samples per average equal to x
SC mm/dd/yy hh:mm:ss
Set Clock with time and date, where mm=month, dd=day, yy=year,
hh=hour(24 format), mm=minute, ss=seconds, Example: ST
01/05/2005 21:05:03
AS
Autostart Setting
SD
Store Current Settings as Default
SM
Store Mode Setting
ZD
Reset Depth Sensor Offset
Acquisition/Action Commands
DL filename
Delete File
GO
Start Data collection using current parameters
XR
Display Raw data to screen when average saved
YS
YMODEM offload of file at 115K Baud
ZA
Acquire 16 samples (ring and eyeball) and save to Zdddhhmm.DAT
file.
ZS
Acquire 16 samples and transmit them.
Used by Check Alignment Function.
ZZ
Go into deep sleep mode (minimum power consumption)
58 LISST-VSF User’s Guide
Display Commands
DB
Display Battery Voltage
Syntax:
DB or db
Description:
The current battery voltage is displayed to the screen.
Note that the instrument will turn on the eyeball motor, laser, and AOM in order
to determine the battery voltage under load.
Example:
input: DB
output: Current Battery voltage: 8.90 volts
DD
Display Disk Directory
Syntax:
DD or dd
Description:
Display current disk directory in DOS type format. Includes total bytes used and
bytes available.
Example:
input: DD
output:
L-VSF:>dd
LISST-100X Disk Directory
Volume in drive C is NONAME
Volume Serial Number is 778B-155F
The instrument settings and status are displayed to the screen.
Example:
input: DS
output:
LISST-VSF Current Status and Settings
Serial number = 1425
Firmware Version 0.920 Jul 23 2012 17:34:17
Operating Mode: Fixed Rate Mode
Start Condition: Delay Start with 0 minute delay
Stop Condition: Fixed Number Stop at 20 samples
Measurements per Average: 10
Sample Interval: 3
Start Degree = 5
Offset Index = 0
LISST-VSF User’s Guide 59
Number of Degrees = 150
Depolarizer Servo Number = 250 [removed in 2/2012].
No Polarizer Servo Number = 0
Half Wave Plate polarizer Servo Number = 500
Motor Controller Setpoint = 45
PMT Sensitivity = 500
AOM Analog Value = 200
Battery Voltage is 10.75
Current Date/Time: Tuesday, August 07, 2012 16:18:36
Current Day of the Year: 220
input: DS 1
output:
LISST-VSF Current Status and Settings
SN = 1425
OM = 3
ST = 5
TD = 0
SP = 5
PD = 20
MA = 10
SI = 3
BI = 1
SB = 1
BB = 0
CT = 0
IC = 0
Current Time = 08/07/2012 16:21:36
Battery= 1079
Switch= 0
Memory= 128245760
HE
Display general help messages and command list
Syntax:
HE or he
Description:
Displays the list of command to the screen.
Example:
input: HE
output: LISST-100X Commands (followed by 30 lines of text)
Setup Commands
MA
Set measurements per average
Syntax:
MA x or ma x
Where x = number of samples per average
Description:
Each recorded or displayed measurement is based on an average of
60 LISST-VSF User’s Guide
measurements. The number of measurements per average is set using the SA
command.
If no value follows command, prompts will be displayed for the value.
Example:
input: MA 10
output: New Measurements per Average: 10
Cautions:
None
ST
Set Start Condition
Syntax:
ST x or st x, where x is the start condition code described below
Description:
The ST command sets the start condition to be used when the GO command is
issued. The start condition options are:
If no value follows command, prompts will be displayed for the value.
Example:
input: ST 5
output: New Start Condition Setting: Delay Start
TD
Set Start Condition Data
Syntax:
TD x or td x, where x is the start condition data described below
Description:
The TD command sets the start condition data to be used when the GO
command is issued. The start condition data is used with the Start Condition
setting as follows:
If the Start Condition is Depth Start (option 1) the input will be start depth
in meters.
If the start condition is set to Time/Date Start (option 2) the input for TD
will be the start date and time.
If the Start Condition is Delay Start (option 5) the input will be time delay
in seconds.
The TD setting is ignored for Mechanical Switch Start (option 3) or Digital
Input Start(option 4).
If no value follows command, prompts will be displayed for the value.
Example:
input: TD 3 (if Start Condition = 1 (Depth Start))
output: New Start Condition data = 3
Start Condition: Depth Start at 3 meters
input: TD 12/31/05 23:59:59 (if Start Condition = 2 (Time/Date Start))
output: New Start Condition data = 12/31/05 23:59:59
Start Condition: Time Start at 12/31/05 23:59:59
Input: TD 2 (if Start Condition = 5 (Delay Start))
Output: New Start Condition data = 2
Start Condition: Delay Start with 2 minute delay
SP
Set Stop Condition
Syntax:
SP x or sp x, where x is the stop condition code described below
LISST-VSF User’s Guide 61
Description:
The SP command sets the stop condition to be used when collecting data. The
stop condition options are:
1 = Depth Stop
2 = Time/Date Stop
3 = Mechanical Switch Stop
4 = Digital Input Stop
5 = Fixed Number of Samples Stop
6 = Maximum memory or Low Battery Stop
If no value follows command, prompts will be displayed for the value.
Example:
input: SP 5
output: New Stop Condition Setting: Fixed Number Stop
PD
Set Stop Condition Data
Syntax:
PD x or pd x, where x is the stop condition data as described below
Description:
The PD command sets the stop condition data to be used when the collection
data. The stop condition data is used with the Stop Condition settings as
follows.
If the Stop Condition is Depth Stop (option 1) the input will be stop depth
in meters.
If the stop condition is set to Time/Date Stop (option 2) the input for PD
will be the stop date and time.
If the Stop Condition is Fixed Number Stop (option 5) the input will be the
number of samples to collect before stopping.
The PD setting is ignored for Mechanical Switch Stop (option 3), Digital
Input Stop (option 4) or Maximum memory or Low Battery Stop (option
6).
Example:
input: PD 3 (if Stop Condition =1 (Depth Stop))
output: New Stop Condition data = 3
Stop Condition: Depth Stop at 3 meters
input: PD12/31/05 23:59:59 (if Stop Condition = 2 (Time/Date Stop))
output: New Stop Condition data = 12/31/05 23:59:59
Stop Condition: Time Stop at 12/31/05 23:59:59
input: PD 2 (if Stop Condition = 5 (Fixed Number of Samples Stop))
output: New Stop Condition data = 2
Stop Condition: Fixed number Stop at 2 samples
SI
Set Sample Interval
Syntax:
SI x or si x, where x is the number of seconds between samples, from 1 to
10,000.
Description:
In Fixed Rate Mode (OM = 3) or Burst Rate Mode (OM = 2), the sample interval
is the number of seconds between two consecutive samples, each composed as
an average of a number of measurements (specified by the MA command).
Example:
input: SI 5
output: New Seconds between Samples: 5
BI
Set Burst Interval
Syntax:
BI x or bi x, where x is the number of seconds between the start of 2
consecutive bursts, from 1 to 10,000 seconds.
Description:
In Burst Mode (OM = 2), the burst interval is the number of seconds between two
62 LISST-VSF User’s Guide
consecutive bursts, each burst composed of a number of samples per burst
(specified by the SB command).
Example:
input: BI 900
output: New Seconds between Bursts: 900
SB
Set Samples per Burst
Syntax:
SB x or sb x, where x is the number of samples per burst.
Each sample is taken at the sample interval (in seconds) set by the SI command.
If no value follows command, prompts will be displayed for the value.
Example:
input: SB 10
output: New Samples per Burst: 10
SC
Set Clock with time and date
Syntax:
SC mm/dd/yy hh:mm:ss or sc mm/dd/yy hh:mm:ss
Where mm=month, dd=day, yy=year, hh=hour (24 hour format), mm=minute,
ss=seconds
If no values follow the “SC” or “sc” command, prompts for entering the time and
date will be displayed.
Example:
input: SC 01/05/2001 21:05:03
output: Command Data in SC is: 01/05/01 21:05:03
Current Date/Time: Friday, January 05, 2001 21:05:03
AS
Autostart Setting
Syntax:
AS x or as x, where x is 1 (yes) or 0 (no)
Description:
With Autostart enabled, the firmware will immediately start the sampling program
that is stored in the EEPROM of the Persistor datalogger upon power up of the
instrument.
Use the SD command to store the current settings as default values to be used
in conjunction with the AS command.
Example:
input: AS 1
output: AutoStart will occur upon power up!
Cautions:
1) If Autostart is enabled, the user cannot talk to the instrument when powering it
up – it will immediately start sampling according to the SD defaults. In order to
stop sampling and establish normal communication, the user must issue a stop
command, by either pressing the STOP button or issuing a CTRL-C command in
the terminal window when the instrument is powered up.
2) Make sure the current program settings are as desired and saved as default
settings (see SD).
SD
Store Current Settings as Default
Syntax:
SD or sd
Description:
Issue the SD command to verify the current settings. Then confirm (1 = yes, 0 =
no) that these settings should be burned into the EEPROM to be used with the
AutoStart setting.
Example
input: SD
output: LISST-100X Current Status and Settings
LISST-VSF User’s Guide 63
Serial number = 1335
Firmware Version 1.997 Feb 16 2011 14:18:45
Operating Mode: Fixed Rate Mode
Start Condition: Delay Start with 3 minute delay
Stop Condition: Fixed Number Stop at 2 samples
Measurements per Average: 10
Sample Interval: 5
Battery Voltage is 7.28
Current Date/Time: Wednesday, August 17, 2011 12:24:14
Current Day of the Year: 229
Do you wish to set current settings to be defaults? (1=yes,0=no): [0] ? 1
Current Settings being saved to EEPROM...
Current Settings saved as defaults!!!
Cautions:
If you change any of the settings after issuing the SD command you must
reissue
the SD command to save the updated settings.
SM
Store Mode Setting
Syntax:
SM x or sm x, where x is 1 or 0
Description:
The Store Mode setting disables (x = 1) or enables (x = 0) data storage on the
internal drive. No Ldddhhmm.DAT file will be created if store mode is disabled.
It is intended to be used for long-term real-time deployments, where it may not
be desirable to have the data stored on the datalogger to avoid it filling up.
Example:
input: SM 1
output: Data will NOT be stored after each sample!!!
input: SM 0
output: Data will be stored after each sample!!!
Cautions:
Be very careful that the correct store mode is selected before a
deployment where it is the intent to store the data internally on the
instrument datalogger!
Issue a DS command to verify the status of the store mode setting. A
warning will be displayed as part of the DS status if the store mode is
disabled:
WARNING: Data storage disabled. Data will not be saved!!!
ZD
Reset Depth Sensor Offset
Syntax:
ZD or zd
Description:
The ZD (or zd) command resets the depth sensor offset so that the sensor reads
a depth of 0m at zero depth (in air). You must issue the ZD command, then
select 1 (yes) or 0 (no) to reset depth sensor
Instrument must be at zero depth and similar temperature to field conditions.
64 LISST-VSF User’s Guide
Do you wish to reset Depth Sensor offset? (1=yes,0=no): [0] ? 1
Previous offset was -14.03. New offset is -13.87.
Previous Depth was -0.15 meters.
New Depth using corrected offset is 0.00 meters.
Cautions:
None
Acquisition/Action commands
DL
Delete file from Compact Flash Module
Syntax:
DL filename or dl filename where filename is the name of the file to be deleted.
Description:
DL command is used to delete file from the Compact flash module. Wildcards
such as *.* can be used.
Example:
Input: DL L159*.dat
Output: Are you sure (Y/N)…
Cautions:
WARNING: Make sure that the file being delete has already been offloaded
before deleting the file. Once the file is delete it can not be recovered.
GO
Start Data Collection using current Settings
Syntax:
GO or go
Description:
Starts Fixed Rate or Burst Mode Data collection using current settings.
Example:
Input: GO
Output: Waiting for start conditions…
Cautions:
To stop data acquisition before it is complete press the Stop button or CTRL-C.
XR
Display Raw data to screen when average saved
Syntax:
XR x or xr x, where x =1 (yes) or x = 0 (no)
Description:
If display of raw data is enabled, the RAW data will be printed to the serial
port between curly brackets in 40 rows:
{
…. 32 ring values + 8 Aux parameters
}
Example:
input: XR 1
output: Raw data WILL be transmitted when data is stored
Cautions:
None
YS
YMODEM Offload
Syntax:
YS filename or ys filename where filename is the name of the file to be
offloaded.
Description:
Offloads file from Compact Flash Module to PC using YMODEM offload at
115 Kbaud
Example:
Input: YS L1391205.dat
Output: Starting…
LISST-VSF User’s Guide 65
Notes:
The YS will initiate the YMODEM offload on the instrument. A YMODEM
file transfer must be started on the PC to accept the data. HyperTerminal
can accept YMODEM file transfers at 115K baud.
ZA
Collect Ring and Eyeball Background Scattering
Syntax:
ZA or za
Description:
Acquires 16 measurement sets and saves the results to Zdddhhmm.DAT.
The ZA command should be used to collect the background scattering
data used in processing LISST-VSF data.
Example:
Input: ZA
Output: <Status information during sampling>
Notes:
The ZS command does not store the sample to a datafile, but average
detector ring values may be written to a text file using the LISST-VSF
software.
ZS
Grab 20 samples and transmit
Syntax:
ZS or zs
Description:
Acquires 20 averaged sample and displays the result to the screen. The
ZS command is used by the LISST-VSF program for the Check Alignment
function.
Example:
Input: ZS
Output: {
…. 32 ring values + 8 Aux parameters
}
repeat 20 times
Notes:
The ZS command does not store the sample to a datafile, but average
detector ring values may be written to a text file using the capture
functions of the Motocross terminal software.
ZZ
Go into deep sleep mode (minimum power consumption)
Syntax:
ZZ or zz
Description:
Sends LISST-VSF in to low power sleep mode. Instrument will wake up
every 10 seconds and wait for a response. If there is no response within 2
seconds, the instrument will return to sleep mode.
Example:
Input: ZZ
Output: Deep Sleep… Enter zz to wake upDeep Sleep… Enter zz to wake upDeep Sleep… Enter zz to wake up
Cautions:
The background scattering data and other settings are not affected by the
ZZ command.
66 LISST-VSF User’s Guide
LISST-VSF User’s Guide 67
Appendix A:Details of LISST-VSF Instrument
Fundamentals: The Volume Scattering Function (VSF) is an inherent optical property of water. It
describes how an incident photon is scattered into different directions. The integral of the VSF
produces the total scattering coefficient, usually denoted as b. The attenuation of light is described
with a beam attenuation coefficient, c. It is an axiom of optics that attenuation equals the sum of
absorption, denoted with a and scattering, i.e. c = a + b. The present instrument measures c directly
with an attenuation sensor, b indirectly via the VSF, thus by subtraction, also permitting an indirect
estimate of absorption a.
The scattering of light is almost always a polarizing event, i.e. the state of light scattered by particles
is altered from the state of the incident light. The scattering properties of particles are described by a
4 x 4 scattering Mueller matrix. The present instrument measures the top left set of these elements.
The relation between the incident state and scattered state of polarization is described by Mueller
algebra:
S
where I, Q, U and V are conventionally symbols used to denote the elements of the Stokes Vector of
light, S. For more details, see Bohren and Huffman(2004).
The Mueller matrix has these simplifications due to symmetry:
P13 = P14 = P23 = P24 = 0; and
P31 = P32 = P41 = P42 = 0;
For spherical particles, further simplifications result:
P12 = P21
P11 = P
It follows that:
(i) When the light source is unpolarized, with its Stokes vector being [1 0 0 0]’, the
(ii) When the light source is verticallypolarized, with its Stokes vector being [1 1 0 0]’, the
(iii) When the light source is horizontally polarized, with its Stokes vector being [1 -1 0 0]’,
scattered
= P S
22
(1)
incident
scattered light has a Stokes vector [P11 P21 0 0].
scattered light has a Stokes vector [P11+P12 P21+P22 0 0].
the scattered light has a Stokes vector [P11-P12 P21-P22 0 0].
The 3 unknowns P11 P21 and P22 can thus be extracted with these measurements. In particular, the
total scattered light power in case (i) above is simply the VSF. In this case, the difference in the two
polarizations, measured by two PMT’s corresponds to M
. Cases (ii) and (iii) above permit
12
extraction of P22 with some redundancy to get a best estimate. These ideas are incorporated in the
LISST-VSF instrument.
In addition to the ‘roving eyeball’ element of this instrument, we have incorporated the small-angle
forward scatter VSF measuring optics, commonly used in the LISST-100X instrument manufactured
by this company. In this manner, this instrument is designed to measure the VSF and 2 other
elements of the scattering matrix from 0.1-degree to 165 degrees (nominal).
The Key Idea: It is common practice to use a discrete number of distinct detectors, all viewing a
common illuminated volume of water, to get the VSF at those distinct angles. Our essential
68 LISST-VSF User’s Guide
innovation was to not be constrained by looking at a fixed point along a beam. Instead, we have
chosen to look from a fixed receiver that rotates to view different portions of a laser beam. This
reduces the receiving optics to a small ‘roving eyeball’ (later). This key innovation makes the
instrument totally submersible, autonomous, and field deployable.
The beam transmission sensor has a very low acceptance angle, which improves estimates of
beam-c (Boss and Slade, 2009).
Ambient light rejection is important while measuring weak scattering from marine particles. We
employ an acousto-optic cell to modulate the laser and use synchronous detection to reject ambient
light. We also exploit this ability to modulate laser power to extend the dynamic range of VSF
measurement. This is done by dimming of laser power while measuring near-forward scattering (to
50 degrees), and then using full power for larger angles. This permits extension of VSF data from 12
bits to 16 bits.
Simplified Optics Diagram:
The LISST-VSF optical design consists of a 1 mm diameter green (532 nm) laser beam originating
from a TEC laser module (1), then folded by a prism (3) onto the primary instrument optical axis. The
default laser polarization is determined by rotation of the laser module and a polarizing beamsplitter
(4). Only the diffracted (modulated) AOM beam is passed by an iris (5) and then expanded to 3 mm
by a beam expander (6). The expanded beam is sampled (7) to measure reference laser power by a
photodiode (8), calibrated to incident laser power entering the sample volume. A half-wave plate (9)
can be mechanically inserted into the laser beam to change the incident polarization between
perpendicular and parallel. The beam then passes through the pressure window (10) into water
within the sample volume. Behind the receive side pressure window covered in ND glass (12) is the
receive lens (13). This lens focuses the scattered light on to the ring detector (14), and the focused
beam passes through a central hole to a photodiode (15) which measures beam transmission. The
eyeball (11) receives scattered light from the sample volume, and is also shown in a side view.
Scattered light viewed by the eyeball passes through a spatial filter (16) to limit the viewing angle of
the eyeball, and interference filter (17) allowing only light at the laser wavelength. Received
scattered light is then split into parallel and perpendicular polarization components by the
beamsplitter (18) and measured by the two PMT modules (19) and (20).
Engineering Design
LISST-VSF User’s Guide 69
Opto-mechanical Design: CAD drawings of the instrument layout is shown in Figure A-15. The top
figure shows the open center section for water. The eyeball is barely visible. The first level of detail is
in the middle drawing, On the left is the transmit optics of Error! Reference source not found..
n the right is the optics to sense small-angle scattering, comprising a receive lens, an ND filter to
attenuate the powerful laser, and the ring detector. An attempt is made to show all components in
the lowest figure.
Electronics:
Subsystems: The following subsystems are involved:
Power for (i) laser; (ii) eyeball motor; (iii) photomultipliers, and (iv) signal processing and
data storage electronics.
Microcomputer for control and data storage.
Photomultipliers, amplifiers, and control voltage connection.
The master clock is derived from the rotating eye-ball and an encoder. This ensures perfect
synchronism of all data capture electronics with angle location of the eyeball.
Upon power up, the laser deliver laser power in water. The eyeball motor is energized to generate a
timing clock for data capture. This system employs a phase-lock loop. The system tolerates
fluctuations in motor speed arising from ‘O’-ring friction. Based on control setting, a high-voltage
control sets PMT sensitivity individually. Data are captured in the following sequence:
1. First rotation: Perpendicularly polarized laser propagates in water; VSF signals are captured
from both PMTs, then ring-detector signals captured.
2. : Half wave plate is inserted, VSF data is captured with PMT’s, then ring-detector data is
captured.
3. Data file is written to memory.
4. If selected, another set of eyeball rotations are begun, and so on.
For each data acquisition sequence, a background scattering is measured. This is done with
0.2micron filtered water. This background is subtracted from particle data, to retain contribution of
particles alone. The background from pure water was verified to appear similar to Rayleigh
scattering.
70 LISST-VSF User’s Guide
Figure A-15 – Laser (1) emits light toward left, and is reflected by prism (3) to go to right. An iris (4)
rejects undiffracted beam, and a beam-expander (5) changes beam diameter to 3mm. The beam
passes through a pellicle beam sampler (6) which splits the beam for measurement by a photodiode;
then through either a ½-λ plate (7), and enters water through transmit window (9). The beam and
forward scattered light enter the receive window(10) and are focused by lens (11) through an ND
filter (12) to ring detector (13). The large-angle scattering is sensed by the eye (14) and the scattered
light beam is then steered to a polarization separator and twin-PMT assembly. This figure is out of
date.
LISST-VSF User’s Guide 71
Appendix B: Method of Extracting P11, P12 and P22
The Signals:
The forward model is that, following Figure 2, the scattered light, with Mueller scattering matrix P is first
multiplied with the Stokes vector of the laser. Then, the light reflected by the lower prism is rotated by the
same angle as the scattering angle to make it’s polarization axes normal to the second prism. A final matrix
product with the Mueller matrix of a linear polarizer is performed to separate the two components of this
light.
The Laser Stokes vectors are: (note only the second element changes)
S= [1
±1
0
0].
The Mueller matrix of scattering is (following data displayed by Voss and Fry, 1984, Figure 1, we set the off
block diagonal elements to zero):
So that, in a manner identical to eq. B3, another estimate of the photomultiplier relative gain is obtained
from:
PMT1V: b = P11+P12 at and (B4c)
PMT2V: d = {P11+P12} at and (B4d)
Extracting VSF (P11):
It follows from summing Eqs. B1a,b
a + b +[ c +d]/ = 4 P11, (B6)
Extracting P12:
From B2 and B4, it follows that:
[b - a +(d - c )/ ] = 4 P12 (B7)
Extracting P22:
Note that P22 always occurs in combination with P12 in Eqs. B2a,b and B4a,b. Thus, only a best estimate of
P22 is extracted. Usually, this involves fine-tuning the relative gain factor . Also, note from Eqs.2a-d, at 45
and 135o , the measurement contains no information on P22. In reverse, solving for P22 blows up at these
angles.
Extracting P22 does sometimes require the tweeking of the relative gain of the photomultipliers. This can be
done within the make_P function. Experienced MATLAB users will know how. We will provide a description
shortly.
These equations are implemented in the MATLAB function make_P.
LISST-VSF User’s Guide 73
Absolute Calibration of P11: Since P12 and P22 are conventionally displayed after normalization by P11, only
the absolute calibration of P11 is required.
The calibration is performed by matching P11 from PMT data with the P11 from Ring 32 of the forward small-
angle VSF. The latter is absolutely calibrated as reported in: Agrawal YC, Mikkelsen O. A (2009):
Empirical forward scattering phase functions from 0.08 to 16 deg. for randomly shaped terrigenous
1-21 µm sediment grains. Optics Express 17:8805–8814.
74 LISST-VSF User’s Guide
Appendix C: MATLAB Software
The MATLAB functions perform these functions:
read_rawfile Reads a raw data file, displays raw data (optional), and outputs uncalibrated
variables;
view_rawfile Reads a raw data file, displays PMT and ring signals and aux data, and outputs
uncalibrated variables;
make_zsc Reads a raw data file acquired as a background file, outputs mean uncalibrated
background file. Useful to compare with factory background.
make_P: Reads particle data file and background data file(.DAT or .MAT produced by
make_zsc) and outputs the parameters P11, P12 and P22, depth,
temperature, date, etc.. P11 includes small angle VSF for angles of Appendix D.
sim_abcd Simulates Mie theory estimate of a,b,c,d for given particle size and complex
refractive index, to compare with data seen with read_vsf.
fastmie.m Function to compute Mie scattering for arbitrary sphere diameter and relative
refractive index (may be complex)
pol_vsf.m Function to compute Mueller matrix elements for particle of arbitrary diameter and
relative refractive index (may be complex). Uses bhmie.m. Outputs at 0.1 degree
intervals.
Overview of MATLAB Software
read_vsf.m This function opens the data file, displays the raw data with laser on and off,
assigns variables such as PMT1plus, PMT2plus, etc., and depth, temperature, PMT setting
etc. The display of raw data (laser on and off) helps in spotting errors, or glitches.
make_zsc This function reads a file that is taken with clean water. It edits out bad points
that may arise due to glitches or floating particles, and generates a mean value of the output
of rings, and of PMT’s. The function uses the median values, instead of means, to
reject the occasional floating particle. Additionally, this function also shows a comparison of
the current background and the factory background. This is then, the health-check of the
instrument as well as a check on the quality of the background. The mean background is
saved as an .asc file.
make_P: This function reads the background data file (binary or ascii) and the particle
data file and outputs the 3 desired variables of P for all 182 angles, along with the angles.
Corrections include: attenuation correction of scattering signals, subtraction of backgrounds
from particle data, and correction for laser reference drift. Absolute sensitivity of PMT’s is
adjusted to match the VSF estimated from ring 32. Estimates of the relative gain parameter
are obtained from data. The output is not averaged over scans. VSF and depolarization can
then be plotted against depth.
LISST-VSF User’s Guide 75
Auxiliary Files: In addition to the functions above, the calibration factors and others:
cal_factorsXXXX.mat Calibration factors for sensors for laser power, depth, temperature.
Each calibration factor is identified with a slope and offset; e.g. the calibration for laser
reference sensor is called LREF_slope and
REF_offset, etc.
Variable Names:
The datafile contains two rotations of 40 + 5*150 16-bit values, per set. In each rotation, the first 40
variables are identical in format to our LISST-100X, i.e these contain the output of ring detectors for
forward scattering, laser transmission, battery voltage and PMT control voltage (new for LISSTVSF), laser reference, depth, temperature, and date and time. Then for each of 150 angles, a 5
parameter subset follows. Each subset contains PMT1_ON, PMT1_OFF, PMT2_ON, PMT2_OFF,
ANGLE. The ‘on’ and ‘off’ refer to PMT outputs in digital counts summed for several cycles of laser
chopping. See Appendix D for full details.
variable name laser polarization PMT polarization
rp perpendicular PMT1: parallel
rr perpendicular PMT2: perpendicular
P11 Composite VSF, includes small-angle and eyeball data
angles, Angles over which eyeball data is useful (in degrees)
Angles Composite angles at which P11 is presented
beam_b Beam scattering coefficient, (m-1) over 0.09 to 150 degrees.
beam_c Beam attenuation coefficient, (m-1)
Bat Battery voltage
date Date and time, two variables, [day*100+hr] and [min*100+sec].
depth Depth in 0.01 m steps
p11 VSF seen by eyeball (over eyeball angles), units m-1 sr
-1
p12, Depolarization
p22, Depolarization
PMT PMT Control voltage (in mV); sets PMT gain.
rings1 40 variables from 32 rings etc. for perpendicular laser polarization
rings2 40 variables from 32 rings etc. for horizontal laser polarization temp Water
temperature, 0.01 oC steps, at endcap on ring-detector side.
These variable names appear in data processing MATLAB software, and in function calls.
76 LISST-VSF User’s Guide
Angles in Water at which VSF is measured
The following Table shows 32 angles at which small-angle VSF is measured, and eyeball angles
beyond. ‘Lower’ and ‘Upper’ refer to angle range covered by each ring detector. These data are
provided an .asc file: Observation_Angles_in_Degrees angles.asc and also as a MATLAB file, Observation_Angles_in_Degrees angles.mat. These angles are in water.
Ring #
Lower
Upper
Median
1
0.0862
0.1017
0.0936
2
0.1017
0.1200
0.1104
3
0.1200
0.1416
0.1303
4
0.1416
0.1671
0.1538
5
0.1671
0.1972
0.1815
6
0.1972
0.2327
0.2142
7
0.2327
0.2746
0.2525
8
0.2746
0.3240
0.2983
9
0.3240
0.3823
0.3520
10
0.3823
0.4512
0.4153
11
0.4512
0.5324
0.4901
12
0.5324
0.6283
0.5784
13
0.6283
0.7414
0.6825
14
0.7414
0.8748
0.8054
15
0.8748
1.0323
0.9503
16
1.0323
1.2181
1.1214
17
1.2181
1.4372
1.3231
18
1.4372
1.6957
1.5611
19
1.6957
2.006
1.8419
20
2.006
2.3600
2.1729
21
2.3600
2.7837
2.5631
22
2.7837
3.2827
3.0229
23
3.2827
3.8703
3.5644
24
3.8703
4.5613
4.2016
25
4.5613
5.3731
4.9506
26
5.3731
6.3250
5.8296
27
6.3250
7.4384
6.8591
28
7.4384
8.7365
8.0614
29
8.7365
10.2430
9.4598
30
10.2430
11.9804
11.0777
31
11.9804
13.9674
12.9358
32
13.9674
16.2149
15.0493
Angles 33 to 182
6-155
Appendix D: Observation Angles, Data Storage Format
and Variable Names
LISST-VSF User’s Guide 77
Raw Data Storage Format
The values in the binary raw data file (.DAT
extension ) are stored in the order shown in the
table below.
Elements
Parameter
Record 1:
First eyeball rotation with laser polarized perpendicular
1:32
Light intensity of ring detectors, 1-32
33
Laser transmission sensor (used in beam-c computation)
34
Battery voltage
35
PMT Control Voltage [e.g. 550 means 0.550V; max can be 1V]
36
Laser reference sensor (monitors laser emission), raw counts
37
Pressure, in digital counts, each count is 0.1m
38
Temperature in units of 1/100th of 1o C
39
(Day*100 + Hour) at which data taken
40
(Minutes*100 + seconds) at which data taken
41-790
150 sets of [angle, PMT1_on, PMT1_off, PMT2_on, PMT2_off]
Record 2:
Second eyeball rotation with laser polarized parallel
1:32
Light intensity of ring detectors, 1-32
33
Laser transmission sensor (used in beam-c computation)
34
Battery voltage
35
PMT Control Voltage [e.g. 550 means 0.550V; max can be 1V]
36
Laser reference sensor (monitors laser emission), raw counts
37
Pressure, in digital counts; each count is 0.1m
38
Temperature in units of 1/100th of 1o C
39
(Day*100 + Hour) at which data taken
40
(Minutes*100 + seconds) at which data taken
41-790
150 sets of [angle, PMT1_on, PMT1_off, PMT2_on, PMT2_off]
The raw data file is binary, of type *.dat. Each VSF requires two rotations of the eyeball, so
that two records are stored per measurement of VSF. One record is with laser polarization
being vertical, the second is with polarization rotated horizontal.
[Note: vertical and horizontal terms are used here simply to donate polarizations with respect to the instrument
optical bench; they are not meaningful with respect to true horizontal!.]
Each record represents one full turn of the eyeball. It contains these data: 40 variables for
rings data and aux measurements, and 150 sets of 5 measurements each for each degree
of eyeball rotation. Each set being: PMT1_on, PMT1_off, PMT2_on, PMT2_off, angle. The ‘on’ and
‘off’ refer to laser being modulated on or off. Both ‘on’ and ‘off’ values are stored so that
saturation of A/D may be recognized as an error.
78 LISST-VSF User’s Guide
Output Variable
Sizes from
read_VSF.m
For N sets of VSF (i.e. 2N rotations of the eyeball) the following are
variable sizes output by function read_vsf.m
Elements
Parameter
rp
N x 150
pp
N x 150
rr
N x 150
pr
N x 150
rings1
N x 40
rings2
N x 40
LP
N x 2
(laser power is recorded for each rotation of the
eyeball)
LREF
N x 2
(laser reference output is recorded for each rotation
of the eyeball)
PMT
N x 2
(even though fixed for all rotations in a file)
The data format is the same for background files, and for particle-data files.
A file can be opened in MATLAB using the command:
Refer to Data Processing section for further details.
The variable names are:
rings1 and rings2 refer to ring data (40 variable format) corresponding to the two rotations of the
eyeball for each measurement of VSF.
LP: laser power transmitted through water, 1 measurement for each of 2 turns per VSF
LREF: laser reference power, used for normalization and compensation of laser drift
PMT: Control voltages of PMT. The PMT high-voltage supply is controlled by a 0-1V setting. This
setting is stored. PMT gain depends on actual high-voltage. Mainly used to distinguish data at different
settings. This data is not used otherwise.
LISST-VSF User’s Guide 79
The LISST-VSF has 3
separate underwater
connectors: A 5-pin, a 6-pin,
and a 3-pin connector. The
photograph shows the
placement of each connector.
The following text describes
the detailed wiring for each
connector. xxx
Appendix E: Connector Pinouts for LISST-VSF
80 LISST-VSF User’s Guide
Connector Pin #
Use
1
Serial Ground
2
No Connection
3
Serial Ground
4
Serial Out (to DB-9 Pin 2)
5
Serial In (to DB-9 Pin 3)
Communications Connector (5 pin connector)
Connector Manufacturer: Impulse Enterprise, Inc. San Diego, CA, USA
Connector Part Number (Bulkhead): MCBH (WB)-5-MP Stainless Steel
Mating Cable Part Number: MCIL-5-FS xxx
LISST-VSF User’s Guide 81
Connector Pin #
Use
1
Digital Ground
2
Digital In #1
3
Digital In #2
4
No Connection
5
Analog In (0 to 2.50V max)
6
Analog Ground
Connector Pin #
Use
1
Power Ground
2
Custom Lithium Battery Charger Input, 14.8V nom.
3
Power Out (12V nom. 9V-16.9V actual)
Auxiliary Input Connector (6 pin connector)
Connector Manufacturer: Impulse Enterprise, Inc. San Diego, CA, USA
Connector Part Number (Bulkhead): MCBH (WB)-6-MP Stainless Steel
Mating Cable Part Number: MCIL-6-FS xxx
Battery/Power Connector (3 pin connector)
Connector Manufacturer: Impulse Enterprise, Inc. San Diego, CA, USA
Connector Part Number (Bulkhead): MCBH (WB)-3-MP Stainless Steel
Mating Cable Part Number: MCIL-3-FS xxx
82 LISST-VSF User’s Guide
a. Lexan ZSCAT Sheet
b. ZSCAT clamps (2)
c. Communications Cable
d. Battery Cable
e. Battery Charger
f. Power Supply
g. Stands (2)
h. Installation Disk
i. Manual
j. Sediment Samples
k. Spare O-rings
l. Mounting Clamps (4 sets)
Appendix H: LISST-VSF Accessories
LISST-VSF User’s Guide 83
This Statement of Limited Warranty applies to all Sequoia Scientific, Inc. (“SEQUOIA”) products ("Products").
Any additional or different terms, including any terms in any purchase order, will be of no effect unless agreed
to in writing by an authorized representative of SEQUOIA as reflected in a written SEQUOIA quotation.
1. Limited Warranty
SEQUOIA warrants that upon delivery by SEQUOIA (a) the Products will be free from defects in materials and
workmanship, (b) the Products will perform substantially in accordance with SEQUOIA's applicable
specifications, and (c) any Products (or components or parts thereof) that are manufactured by SEQUOIA do
not infringe any U.S. patent or copyright.
2. Correction of Non-Compliance
If, during the twelve months after delivery (the “Warranty Period”), any Product does not comply with the
warranties set forth in 1(a) and 1(b) above, SEQUOIA will, at its option, either (a) repair the Product, (b)
replace the Product, or (c) refund the purchase price paid by Customer to SEQUOIA for the Product; provided
that Customer gives SEQUOIA written notice of the noncompliance within the Warranty Period and ships the
Product to SEQUOIA within one month after the end of the Warranty Period. As to any Product repaired or
replaced by SEQUOIA, the Warranty Period will end upon the later of the end of the original Warranty Period
or 90 days after SEQUOIA's delivery of the repaired or replacement Product to Customer. Any Product,
component, part or other item replaced by SEQUOIA becomes the property of SEQUOIA. SEQUOIA may use
refurbished components in the repair of Products supplied hereunder.
SEQUOIA's warranties shall be void and not apply if the Product has been subjected to misuse or alteration or
repaired by a party not approved by SEQUOIA or the serial number on a product (if applicable) has been
altered or defaced. SEQUOIA shall not be liable for normal wear and tear (such as replacement of
consumables), nor for defects or failure caused by maintenance, misuse, negligence or failure resulting from
non-compliance with SEQUOIA’s specifications, operating or maintenance manuals.
3. Infringement Indemnification
If any Product does not comply with the warranty set forth in 1(c) above, SEQUOIA will defend and indemnify
Customer against any third-party claim asserted in any proceeding against Customer based on this
noncompliance; provided that Customer gives SEQUOIA prompt written notice of the claim, SEQUOIA has
exclusive control over the defense and settlement of the claim, Customer provides such assistance as
SEQUOIA may request in connection with the defense and settlement of the claim (in which event SEQUOIA
will reimburse the reasonable out-of-pocket costs incurred by Customer to provide such assistance), Customer
does not settle the claim without the prior written consent of SEQUOIA and, upon SEQUOIA's request,
Customer returns the Non-Complying Product to SEQUOIA for modification, replacement or a refund of the
purchase price paid by Customer to SEQUOIA for the Non-Complying Product, less a reasonable allowance
for Customer's use prior to return.
4. Exclusive Warranties
THE WARRANTIES SET FORTH IN PARAGRAPH 1 ABOVE ARE EXCLUSIVE AND IN LIEU OF ALL
OTHER WARRANTIES, EXPRESS OR IMPLIED. SEQUOIA DISCLAIMS ANY AND ALL WARRANTIES,
EXPRESS OR IMPLIED (INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, AND ANY IMPLIED WARRANTY
ARISING FROM COURSE OF PERFORMANCE, COURSE OF DEALING, OR USAGE OF TRADE) OTHER
THAN THOSE SPECIFICALLY SET FORTH IN PARAGRAPH 1.
5. Exclusive Remedies
CUSTOMER'S RIGHTS AND REMEDIES SET FORTH IN PARAGRAPHS 2 AND 3 ABOVE ARE EXCLUSIVE
AND IN LIEU OF ANY AND ALL OTHER RIGHTS AND REMEDIES FOR ANY BREACH OF OR OTHER
FAILURE TO COMPLY WITH ANY WARRANTY WITH REGARD TO ANY PRODUCT.
6. No Consequential Damages
SEQUOIA will not be liable for any indirect, incidental, special or consequential damages, any cover, or any
loss of revenue, profit, data or use.
7. Limitations of Liability
Warranty
STATEMENT OF LIMITED, EXTENDED WARRANTY
84 LISST-VSF User’s Guide
SEQUOIA's liability (whether in contract, tort, or otherwise; and notwithstanding any fault, negligence, strict
liability or product liability) with regard to any Product (including, but not limited to, any breach of or default by
SEQUOIA) will in no event exceed the purchase price paid by Customer to SEQUOIA for such Product.
Further, SEQUOIA will not be liable for, or be in breach of or default on account of, any delay or failure to
perform as a result of any cause, condition or circumstance beyond SEQUOIA's reasonable control.
8. Indemnification by Customer
Customer acknowledges that the Products are designed and manufactured for use in non-critical, monitoring
situations. If Customer chooses to purchase a Product or Products for use in applications that could result in
damages in excess of the price of the Product if the Product does not operate properly or otherwise fails,
Customer acknowledges and agrees that it is Customer’s responsibility to provide for redundancy and/or other
safety or back-up measures sufficient to assure that failure of a Product(s) will not cause such damages.
Customer agrees that it will defend and hold SEQUOIA harmless from any and all claims and costs (including
but not limited to attorney’s fees and other costs of defense against such claims) in excess of the price of the
Products arising directly or indirectly from such Customer’s use of the Products. Such indemnification is a
critical part of the consideration being provided by Customer (over and above the price paid for the Product(s))
for the right to use the Products for such purposes and Customer shall not use a Product or Products for such
purposes if it is unwilling or unable to provide such indemnification.
9. Statute of Limitations
Customer will not commence any action based on breach of warranty with respect to any Product more than
18 months after SEQUOIA's delivery of such Product.
10. Software
The Products may include or be delivered with certain computer programs, databases or other software that is
proprietary to SEQUOIA. SEQUOIA hereby grants Customer a nonexclusive license to use such software
solely for the purpose of operating Products. Customer will not: use any such software for any other purpose;
modify, adapt, translate, or create derivative works based on any such software; or disassemble, decompile or
reverse engineer any such software. No title to or ownership of any software or intellectual property rights are
transferred to Customer.
11. U.S. Government Restricted Rights Notice
All software, data, technical information, and written materials provided by SEQUOIA are provided with
restricted rights. Use, duplication, or disclosure by the government is subject to restrictions as set forth in
subparagraph (c)(1)(ii) of the Rights in Technical Data and Computer Software clause at 48 C.F.R. § 252.2277013 or in subparagraph (c)(2) of the Commercial Computer Software-Restricted Rights clause at 48 C.F.R. §
52.227-19, as applicable. Portions of these items may be unpublished. SEQUOIA reserves all rights under
applicable copyright laws.
12. Controlling Document
In the event of any conflict or inconsistency between any provision of this Statement of Limited Warranty and
any other provision of the Order, the provision of this Statement of Limited Warranty will control.
13. Controlling Law
This Statement of Limited Warranty will be governed by the laws of the State of Washington without reference
to its rules relating to choice of law for the purpose of applying another jurisdiction’s law. The U.N. Convention
on Contracts for the International Sale of Goods will not apply.
LISST-VSF User’s Guide 85
Revision
Date
Description
Author
A
2012/08/28
Draft document
YCA/WHS/RS
B
2013/03/08
Updated for non-AOM units
RS/WHS
Document History
86 LISST-VSF User’s Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.