SeQual Eclipse 1000 User Manual

Page 1
Model 1000 User Manual
Page 2
Page 3
Important Safety Instructions 2
Symbols Used in User Manual and on Oxygen System 2
Indications For Use 5
Contraindications 5
Safety Guidelines 6
Oxygen Therapy Safety 6
Operating Safety 7
Power Cartridge Safety 9
Introduction To Your ECLIPSE Oxygen System 10
Understanding the Controls and Front Panel 11
Description of the ECLIPSE 11
Front Panel Icons 13
System Status Indicators 13
Recommended Operating Environments 15
The Proper Location 16
Operating Instructions 17
Before Operating 17
Powering Your ECLIPSE System 18
Operating Your ECLIPSE System 25
Alarm Conditions and Troubleshooting 30
Accessories 32
Cleaning, Car
e and Routine Maintenance
35
Technical Data 38
Summary of All Warnings and Cautions 40
Electromagnetic Compatibility 48
Disposal of Equipment and Accessories 4
9
Model 1000 User Manual
Contents
Page 4
Important Safety Instructions are Shown Under the Following Headings:
SeQual®Technologies Inc. (SeQual) recommends that this User Manual be given to each patient receiving an ECLIPSE
TM
Oxygen System. This User Manual describes all
the necessary information in order for a patient or caregiver to operate the ECLIPSE.
WARNING Important safety information for hazards that might cause
serious injury.
CAUTION Important information for preventing damage to the ECLIPSE
Oxygen System.
DO NOT Information to which you should pay special attention.
Important! Clarifying information, specific instructions, commentary, sidelights,
or interesting points.
Symbols Used in User Manual and on Oxygen System
Refer to this section for an explanation of the symbols and the warnings that accompany them.
“Read Instructions”: This symbol, when used alone or in conjunction
with any of the following symbols, indicates the need to consult the operating instructions provided with the product. A potential risk exists if the operating instructions are not followed.
Attention or Caution”:
Consult accompanying documents. T
his
symb
ol indicates important information that requires the user to take
special precautions when certain conditions are present.
Important Safety Instructions
2|ECLIPSETMModel 1000 User Manual
Page 5
“No Smoking or Naked Flames”: The use of oxygen therapy
requires that special care be taken to reduce the risk of fire. Materials that burn in air, and even some materials that will not burn in air, are easily ignited and burn rapidly in high concentrations of oxygen. For safety concerns, all possible sources of ignition must be kept away from the oxygen concentrator and preferably out of the room in which it is being used.
Smoking in the proximity of an operating oxygen concentrator is dangerous, and can permanently damage the device and void the warranty.
“Use No Oil or Grease”: Indicates important information about the
possibility of spontaneous and violent ignition that may occur if oil, grease or other petroleum based substances come into contact with oxygen under pressure.
“Drip Proof Equipment-IPX0”: The ECLIPSE enclosure does not
provide protection against the harmful effects of liquid ingress. (IPX0, per IEC 60529)
“Drip Proof Equipment-IPX1”: The ECLIPSE AC and DC Power
Supplies provide protection against the harmful effects of liquid ingress. (IPX1, per IEC 60529)
“IEC Icon for DC Power (Input)”
“T
ype B Equipment”:
Equipment providing a particular degree of
protection against electric shock, particularly regarding:
- Allowable LEAKAGE CURRENT;
- Reliability of the protective earth connection.
“Amperes”
IPX
0
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|3
IPX
1
Page 6
“Alternating Current”
“Audible Alarm”:
An audible indicator is provided to announce a reduction in oxygen concentration or when the power has been interrupted or lost.
“ON/OFF”: Turns the unit ON and OFF but does not directly
disconnect the mains power.
“Increase Flow Setting”: Increases the flow setting in 0.5
increments each time the button is pushed.
“Decrease Flow Setting”: Decreases the flow setting in 0.5
increments each time the button is pushed.
“Flow Mode”: Activates either Continuous Mode or Pulse Flow
Mode Operation.
“External Power Present Indicator”: Indicates the presence of
external power to operate the concentrator.
“Power Cartridge Status”: Shows the amount of charge remaining in
the Power Cartridge.
“Power Cartridge Symbol”: The symbol on the Power Cartridge
pack.
“Oxygen Output”: The port in which the oxygen flows out and where
the tubing is attac
hed.
Symbols (continued)
4|ECLIPSETMModel 1000 User Manual
Page 7
THE ECLIPSE OXYGEN SYSTEM IS INDICATED FOR THE ADMINISTRATION OF SUPPLEMENTAL OXYGEN. THE DEVICE IS NOT INTENDED FOR LIFE SUPPORT NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES.
YOUR PHYSICIAN HAS PRESCRIBED A SPECIFIC OXYGEN FLOW RATE SETTING TO MEET YOUR INDIVIDUAL NEEDS.
RECOMMENDED OXYGEN FLOW RATES SHOULD BE ADJUSTED ONLY UNDER THE ADVICE OF A PHYSICIAN.
WARNING FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE
ORDER OF A PHYSICIAN.
WARNING THE ECLIPSE IS NOT INTENDED FOR LIFE SUPPORTING OR LIFE
SUSTAINING APPLICATIONS, NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES.
WARNING IN CERTAIN CIRCUMSTANCES, THE USE OF NON-PRESCRIBED
OXYGEN CAN BE HAZARDOUS. THIS DEVICE SHOULD ONLY BE USED WHEN PRESCRIBED BY A PHYSICIAN.
WARNING FOR ADULT USE ONLY. NOT QUALIFIED FOR USE ON PEDIATRIC
PATIENTS.
W
AR
N
I
NG
N
O
T FO
R US
E IN THE PRESENCE OF FLAMMABLE ANESTHETICS.
Contraindications
Indications for Use
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|5
WARNING AS WITH ANY ELECTRICALLY POWERED DEVICE, THE USER MAY
E
X
PERIENCE PERIODS OF NON-OPERATION AS THE RESULT OF ELECTRICAL POWER INTERRUPTION, OR THE NEED TO HAVE THE OXYGEN CONCENTRATOR SERVICED BY A TECHNICIAN. AN OXYGEN CONCENTRATOR IS NOT APPROPRIATE FOR ANY PATIENT WHO WOULD EXPERIENCE ADVERSE HEALTH CONSEQUENCES AS THE RESULT OF SUCH TEMPORARY INTERRUPTION.
Page 8
Safety Guidelines
6|ECLIPSETMModel 1000 User Manual
This section of the manual provides warnings and cautions that apply to hazards or unsafe practices that could result in personal injury or property damage.
Oxygen Therapy Safety
“Read Instruction Manual”: DO NOT operate this oxygen system
without first reading and understanding this manual. If you are unable to understand the warnings and instructions, contact your authorized SeQual dealer/provider before attempting to use this oxygen system.
“Caution or Attention”: In certain circumstances, oxygen therapy can
be hazardous; therefore, it is advisable to seek medical advice before using the oxygen system.
DO NOT deviate from prescribed flow rates. Adjust the flow only to your prescribed
flow rate, unless otherwise instructed by your physician.
WARNING “No Smoking or Naked Flames”: The use of oxygen therapy requires
that special care be taken to reduce the risk of fire. Oxygen enhances the combustion of flammable material. For safety concerns, all possible sources of ignition must be kept away from the oxygen system and preferably out of the room in which it is being used. Smoking in the proximity of an operating oxygen concentrator is dangerous, and can permanently damage the device and void the warranty.
W
AR
N
I
NG “Use No Oil or Grease”:
A spontaneous and violent ignition may occur if oil, grease or other petroleum substances come into contact with oxygen under pressure. Keep these substances away from the oxygen system, tubing and connections and any other oxygen source.
DO NOT use any petroleum based or other lubricants.
Page 9
Operating Safety
Keep oxygen system and cord away from hot surfaces.
DO NOT use the oxygen system if it has a damaged cord or plug; contact your
dealer/provider.
Your oxygen system should be located in a well-ventilated area to allow for adequate air intake and to avoid the intake of airborne pollutants and fumes.
ONLY use this device with accessories designed or recommended by SeQual
Technologies Inc.
If your oxygen system has been dropped, damaged or exposed to water, contact your dealer/provider for examination or repair.
DO NOT drop or insert any object into any opening.
DO NOT block the air inlet or the exhaust vent of the ECLIPSE when it is on
a soft surface, such as a bed or couch or car seat.
DO NOT cover unit with a blanket, towel, quilt, etc.
DO NOT overfill the humidifier, if used. Fill with water only to the level shown
by the manufacturer.
W
hen using ECLIPSE in a vehicle, make sure it is properly secured or restrained.
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|7
Page 10
Operating Safety (continued)
WARNING DO NOT remove the cover, no serviceable parts inside. Qualified
service personnel should only remove the cover.
WARNING DO NOT operate device or accessories on wet surfaces or in standing
water and
DO NOT submerse or expose to water. The ECLIPSE
enclosure does not provide protection against the harmful effects of liquid ingress. Electrical shock or damage to the unit may result.
WARNING Protect electrical cords from sharp edges. Electrical shock and serious
physical injury may result.
WARNING Use only the SeQual AC Power Supply and DC Power Supply supplied
with your ECLIPSE Oxygen System. Use of a non-SeQual Power Supply may be hazardous, cause damage to your oxygen system and will void the warranty.
CAUTION Availability of an alternate source of oxygen is recommended in
case of power outage or mechanical failure.
CAUTION DO NOT expose unit to temperature outside of specified operating
or storage temperatures as this may damage the unit.
C
A
UTION
L
ocate o
xygen tubing and Power Supply cords to prevent tripping
hazards.
CAUTION DO NOT place ECLIPSE in a small, enclosed space, such as a
bathroom or closet, with cannula leading out of the enclosed space.
Safety Guidelines (continued)
8|ECLIPSETMModel 1000 User Manual
Page 11
Power Cartridge Safety
WARNING This Power Cartridge may explode and cause potential injury
if exposed to or disposed of in a fire.
WARNING DO NOT short-circuit the Power Cartridge’s metal contacts with
metallic objects, such as keys or coins. It may cause sparks or excessive heat.
WARNING DO NOT disassemble, puncture, or crush the Power Cartridge. Power
Cartridge electrolytes may be toxic if swallowed and can be harmful to skin and eyes. Keep the Power Cartridge away from children.
WARNING Use of a damaged Power Cartridge may cause personal injury.
WARNING Exposing the Power Cartridge to water or other liquids may cause
personal injury.
CAUTION DO NOT expose the Power Cartridge to temperatures above
140°F (60°C), such as in a vehicle parked in the sun or on a hot day.
CAUTION Charging the Power Cartridge below 50 F° (10° C) or above
104° F (40° C) may degrade the performance.
CAUTION DO NOT drop the Power Cartridge or expose it to mechanical shock.
CAUTION Only use the Power Cartridge for its intended purpose.
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|9
Page 12
Introduction to Your ECLIPSE Oxygen System
10|ECLIPSETMModel 1000 User Manual
This User Manual will inform you about the use and care of the ECLIPSE Oxygen System and its accessories. Thoroughly read all the information in this manual in its entirety before operating your ECLIPSE. Contact your dealer/provider for proper training in the use of this device.
YOUR PHYSICIAN HAS PRESCRIBED SUPPLEMENTAL OXYGEN AS PART OF A TREATMENT PLAN. LONG TERM OXYGEN THERAPY IS NOW WIDELY ACCEPTED AS THE STANDARD TREATMENT FOR CHRONIC HYPOXIA CAUSED BY CHRONIC OBSTRUCTIVE PULMONARY DISEASE. OXYGEN THERAPY HAS BECOME MORE
VERSATILE IN RECENT YEARS DUE TO THE DEVELOPMENT OF OXYGEN CONCENTRATORS. THE ECLIPSE MAKES IT EASIER FOR PHYSICIANS TO SELECT A MODE OF OXYGEN THERAPY THAT SUITS THE PATIENT’S LIFESTYLE.
THE ECLIPSE MAY BE USED IN HOME, INSTITUTION, AND MOTOR VEHICLES.
USE OF THE ECLIPSE IN HOSPITALS SHOULD INCLUDE THE USE OF A
“H
O
S
PITAL GRADE” POWER CORD.
YOUR PHYSICIAN HAS PRESCRIBED A SPECIFIC OXYGEN FLOW RATE SETTING TO MEET YOUR INDIVIDUAL NEEDS. THIS FLOW RATE IS DETERMINED BY THE AMOUNT OF OXYGEN YOU NEED. RECOMMENDED OXYGEN FLOW RATES SHOULD BE ADJUSTED ONLY UNDER THE ADVICE OF A PHYSICIAN.
Page 13
Description of the ECLIPSE
The key features of the ECLIPSE Oxygen System and control panel are important items to become familiar with.
Understanding the Controls and Front Panel
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|11
I NPUT/OUTPUT CONNECTIONS
Oxygen Outlet Port Extension tubing or a cannula is attached to this port.
Air Inlet Filter Ambient air is drawn into the unit through the air inlet located on the
top/bac
k of the unit. T
his fi
lter prevents dust and debris from entering
the ECLIPSE and should be cleaned regularly.
Service P
ort
T
his connection allows service personnel to diagnose and monitor
ECLIPSE performance. It is not for patient use.
External Power Receptacle Either the ECLIPSE AC or DC Power Supply output cords may be
connected to this receptacle.
Exhaust Vent Exhaust air from the ECLIPSE is expelled from this vent.
BACK
FRONT
Service Port
(Not for Patient Use)
Air Inlet Filter
Cart Insert Screw
Exhaust Vent
Power Cartridge Latch
Main/Rating Label (Under Base)
Power Cartridge
Handle
Control Panel
Oxygen Outlet Port
Simplified User Instructions (Under Base)
External Power Receptacle
Page 14
Control Panel Features
The ECLIPSE displays important operating information using the icons, numbers and visual indicators on the control panel. This section will help you understand these features.
Understanding the Controls and Front Panel (continued)
12|ECLIPSETMModel 1000 User Manual
000.6
No Smoking Icon
ON/OFF Button
ON/OFF Indicator (Green)
Alert Indicator (Yellow )
Flow Rate Setting (2 Digits) (Green)
Alarm Indicator (Red)
Increase Flow Rate Button
Decrease Flow Rate Button
P
ower Cartridge St
atus
Gauge (Gree
n)
External Power Present Indicator (Gree
n
)
Pulse Flow Mode Indicator (Green)
Flow Mode Button
Page 15
User Controls
This section describes the control buttons and function for your ECLIPSE Oxygen System.
FRONT PANEL ICONS
ON/OFF Button and Indicator:
This button turns the system ON or
OFF. The Green Indicator is lit when the unit is ON.
Increase or Decrease Flow Rate Button: Use these buttons to set the
flow rate to your prescribed level.
Flow Mode Button and Indicator: The ECLIPSE contains a button to
toggle between Continuous Flow and Pulse Flow mode. Pulse Flow mode allows a significant increase in the amount of time the system can operate from the Power Cartridge. When the Pulse Flow Mode is activated, the green Flow Mode indicator goes on and a pulse of oxygen is delivered upon inhalation. When the pulse of oxygen is delivered, the indicator briefly goes off. Only change your flow settings at the recommendation of your physician.
SYSTEM STATUS INDICATORS
The ECLIPSE control panel displays important operating information. This section will help you understand these features.
Green, Yellow, and Red indicators on the front panel indicate the operating condition of the system. For more information on the indicators, see the Alarm Conditions and Troubleshooting table.
ON
/OFF (Green) Indicator - System OK:
Indicates that the EC
LIPSE is ON and operating properly.
ALE
RT (Y
ellow) Indicator - Low and Medium Priority Alerts:
When steadily illuminated, this indicates a low priority awareness condition or Caution. Continue to use your system and refer to the
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|13
Page 16
Troubleshooting section for the proper response. A flashing yellow indicates a medium priority alert. Prompt response is needed to deal with a condition. Refer to the Troubleshooting section.
ALARM (Red) Indicator - High Priority Alarms:
Indicates a high priority alarm condition. Immediate response is necessary to deal with the condition. Refer to the Troubleshooting section and contact your dealer/provider.
Buzzer: An audible alarm (or buzzer) is used to alert you of the
condition of the unit, either a warning or failure, and to confirm a valid key press.
Flow Rate Setting Indicator (Green): This is the main focus on your
control panel. Your dealer/provider will set up your prescribed flow rates for your Continuous Flow Mode (LPM) and your Pulse Flow settings, if pulse flow has been prescribed. DO NOT change these settings unless it is under the order of your physician. Each time you turn the unit on, the last flow mode and flow setting will be used.
Power Cartridge Status Gauge: This indicator displays the charge
remaining in the Power Cartridge. When the Power Cartridge is fully charged all 10 bars are illuminated. Each bar represents approximately 10% of the total Power Cartridge charge. When the Power Cartridge is being charged, the charge indicator bars will blink in a waterfall fashion.
If the Power Cartridge is not installed, or if it is improperly installed, the Status Gauge will not light.
External Power Present Indicator: When the ECLIPSE system is
properly plugged in using the AC or DC Power Supply, this indicator will appear on the Control Panel.
Understanding the Controls and Front Panel (continued)
14|ECLIPSETMModel 1000 User Manual
Page 17
The following information recommends certain boundaries that will allow the ECLIPSE Oxygen System to operate at its full potential.
Operating Temperature 50º F to 104º F (10º C to 40º C)
Operating Humidity 10% - 95% at an 82.4°F (28°C) dew point
Transport/Storage Temperature -4º F to 140º F (-20º C to 60º C)
Humidity: Up to 95% Non-Condensing
Electrical Use no extension cords or electrical
outlets controlled by a switch.
Altitude 0 – 13,123 feet (0 – 4000 meters)
Placement Do Not block air inlet or bottom exhaust vent
(keep at least 3 inches away from walls, draperies, furniture, etc.)
Environment Smoke, pollutant, and fume free.
Operating Time Up to 24 hours per day when connected
to external power.
When moving the ECLIPSE from an extreme environment, allow time for your ECLIPSE to acclimate to the recommend operating environment.
WARNING Operating your ECLIPSE outside the recommended operating
environment may impact performance, cause damage and will void the warranty.
CAUTION Locate the unit in a well-ventilated room that provides adequate air
ow and c
hec
k if air is being prevented from entering or exiting the
system by furniture, draperies, bedding, clothing or pets.
Recommended Operating Environments
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|15
Page 18
Recommended Operating Environments (continued)
16|ECLIPSETMModel 1000 User Manual
The Proper Location
Select a location that avoids fumes and pollutants and allows the oxygen system to draw in room air through the air inlet filter on the back of the cabinet and exhaust air out the vent at the bottom of the case without being restricted.
Locate the unit such that the alarms can be heard.
Route oxygen tubing such that it does not kink or is subject to occlusion.
DO NOT cover unit with a blanket, towel, quilt, or other covering, as the unit may
overheat.
DO NOT locate near flammable materials or cleaning agents or in the direct path of
any heat source, such as a heat register or car heater.
Keep the oxygen system at least five (5) feet (1.5 m) away from hot, sparking objects or open sources of flame.
Page 19
Before Operating
Your SeQual dealer/provider or authorized service representative will demonstrate how to operate the oxygen system. This manual serves as a supplemental reference to help you operate and maintain the unit. If you have any further questions or problems call your SeQual dealer/provider or authorized service representative.
Important! DO NOT attempt to operate the ECLIPSE System without first reading
the “Safety Guidelines” section of this manual. Follow all operating instructions. Observe all warnings on the product and in the operating instructions. To reduce the risk of fire, bodily injuries, and damage to the oxygen system, observe all safety precautions.
WARNING Protect the ECLIPSE and AC/DC Power Supplies from fluid spills or
drips to avoid possible shock hazards.
CAUTI O N Always check to see that the Air Inlet and the Exhaust Vent are not
blocked and the air inlet filter is clean before using your system.
Operating Instructions
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|17
Air Inlet and
Air Inlet Filter
Exhaust Vent
Page 20
Powering Your ECLIPSE System
The ECLIPSE is a lightweight, portable oxygen system capable of being operated directly from three different power sources: (
1) an AC Power Supply, (2) a DC Power
Supply,
or (3) a rechargeable Power Cartridge.
The ECLIPSE charges the optional Power Cartridge when external power (AC or DC) is present and when the Power Cartridge temperature is below a safe charging temperature. When external power is disconnected, the system will automatically switch over to the Power Cartridge, if installed. When external power is restored (by plugging into AC or DC outlets) the system will accept power from the external source and replenish the Power Cartridge while operating, if adequate power is available.
Connect the AC or DC Power Supply by aligning the arrow on the plug locking ring with the white dot on the receptacle of the unit. Push until the locking ring twists slightly counter-clockwise, then snaps back into alignment with the dot on the receptacle. Disconnect by grasping the plug, twist counter-clockwise and pull. Connect the AC Power Cord to a grounded power outlet.
Operating Instructions (continued)
18|ECLIPSETMModel 1000 User Manual
(3) RECHARGEABLE POWER CARTRIDGE
(1) AC POWER SUPPLY
(2) DC POWER SUPPLY
Page 21
WARNING DO NOT use any cord if damaged. Improper use of the Power Cord
may result in a fire or electrical shock hazard.
WARNING DO NOT touch the Power Cartridge cavity connectors with the Power
Cartridge removed while the unit is connected to an external power source. This may result in electrical shock.
(1) AC POWER SUPPLY: The ECLIPSE System includes a universal AC Power Supply
for use at home or where standard AC power is available. To plug into AC, make sure the AC Power Supply cord is securely inserted in the power receptacle on the side of the unit, and plug the AC Power Supply into an AC outlet. When the unit is plugged in properly, a green indicator on the AC Power Supply will light and the
External Power Present Indicator, will appear on the control panel.
CAUTION The AC Power Cord used with the ECLIPSE Oxygen System must
meet the requirements of the country where the product is used.
CAUTION Use with grounded, 3-pin receptacle. DO NOT remove ground pin from
cord. Use only with SeQual supplied cords.
CAUTION DO NOT connect the ECLIPSE to an extension cord or electrical
outlet controlled by a switch.
(2) DC POWER SUPPLY: A DC Power Supply allows the system to operate from
DC accessory outlets, such as those found in motor vehicles.
You need to
determine the Fuse rating in your motor vehicle, either 15 AMP or 20 AMP.
Consult your vehicle’s Operator’s Manual or look at your fuse b
o
x in the vehicle. Set the switch on the DC Power Supply to the appropriate one. Use the 10 AMP position on the D
C P
ower Supply for a 15 AM
P vehicle fuse and use the 15 AMP
position for a 20 AM
P or greater vehicle fuse.
Start your vehicle. Insert the D
C cord in the proper receptacle on the side of the ECLIPSE. Next, insert the DC Power Supply plug into the DC accessory outlet in your vehicle. When the system is properly plugged in and receiving power from the DC source, a green indicator light on the Power Supply, and the
External Power Present indicator,
on the control panel will appear.
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|19
Page 22
WARNING Secure the ECLIPSE and the DC Power Supply in your vehicle and
make sure the air inlet and exhaust airflow are not blocked. This may impact performance.
WARNING DO NOT leave the ECLIPSE or DC Power Supply plugged into the
vehicle without the engine running or attempt to start the vehicle while the DC Power Supply is connected to the vehicle. This may drain the vehicle’s battery.
A buzzer in the DC Power Supply is intended to warn user when low vehicle power is present. Immediately disconnect the DC Power Supply from the vehicle for low voltage alarm conditions.
WARNING If the vehicle power source (vehicle accessory connector) drops below
11.5 Volts (brownout) the ECLIPSE will revert to Power Cartridge operation (if present). The External Power Present Indicator may remain on.
TYPICAL FLOW RATES AVAILABLE AT EACH FUSE SETTING
10 AMP 15 AMP
Setting Continuous Pulse Continuous Pulse
0.5 Yes Yes
1.0 Yes Yes Yes Yes
1.5
Y
es Ye s Ye s Ye s
2.0
Y
es
Ye s Ye s Ye s
2.5
N
O
Y
es
Y
es Yes
3.0 NO Yes Yes Yes
3.5 Yes Yes
4.0
Ye s Ye s
4.5 Yes Yes
5.0
Y
es Yes
5.5 Yes Yes
6.0 Yes Yes
Actual performance may vary dependent upon usage conditions and vehicle.
Operating Instructions (continued)
20|ECLIPSETMModel 1000 User Manual
Page 23
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|21
Latch
P
ower Cartridge
When operating in the 10 AMP switch position, flow rate settings above 2.0 LPM continuous flow will automatically switch to 2.0 LPM continuous flow when the DC Power Supply is connected.
(3) RECHARGEABLE POWER CARTRIDGE: The ECLIPSE system can also be
powered by a rechargeable Power Cartridge.
Installing the Power Cartridge: Align the Power Cartridge so that it will slide into the
Power Compartment. Push the Power Cartridge into the Power Compartment until it is secure and flush with the back of the system. When the Power Cartridge is properly inserted and the ECLIPSE is
ON, the Power Cartridge Status Gauge will appear on
the Control Panel. (If you
DO NOT have a Power Cartridge installed, the Status Gauge
will not light up).
To remove the Power Cartridge, push up on the latch on the Power Cartridge. The Power Cartridge will slide out so that you can grasp it for removal.
Power Cartridge Capacity: The capacity of the Power Cartridge is lit up on the control
panel. A variety of factors, such as flow rate and Pulse or Continuous Flow Mode operation determine the duration of operating time. The following table provides estimates of amount of time that the ECLIPSE can operate on a fully charged Power Cartridge given certain flow rates and operating conditions.
Page 24
TYPICAL NEW POWER CARTRIDGE OPERATION TIME
Power Cartridge
Flow Rate Setting
Continuous Flow Mode (LPM) Pulse Flow Mode (Setting)
0.5 4.4 hours
1.0 3.7 hours 4.4 hours
2.0 2.4 hours 3.6 hours
3.0 1.3 hours 3.0 hours
4.0 2.6 hours
5.0 2.3 hours
6.0 2.1 hours
Operating time will degrade with Power Cartridge use and age.
CAUTION Store your Power Cartridge in a cool, dry location. This will help to
assure the longevity of your Power Cartridge.
CAUTION DOT/UN Regulations require the removal of the Power Cartridge
for international travel.
CAUTION There are no serviceable parts inside the Power Cartridge.
DO NOT open.
CAUTION Replace the Power Cartridge with only SeQual approved Power
Cartridge.
C
A
U
T
I
O
N
T
he P
ower Cartridge used in this device may present a risk of fi
re
or chemical burn if mistreated.
DO NOT disassemble, incinerate,
or heat ab
ove 14
0
° F (60° C).
C
A
U
T
I
ON
Return your P
ower Cartridge to your dealer/provider for proper
disposal.
CAUTION Keep your Power Cartridge away from children.
Operating Instructions (continued)
22|ECLIPSETMModel 1000 User Manual
Page 25
Typical Power Cartridge Recharge Time
The typical time to recharge your Power Cartridge to achieve 80% capacity from a fully discharged Power Cartridge is 1.4 hours to 5.0 hours, dependent upon the flow setting.
If the Power Cartridge is too warm, charging will not begin until it sufficiently cools.
The Power Cartridge may not charge or the charge time may increase when operating from the DC Power Supply set in the 10 AMP position.
INITIAL POWER CARTRIDGE OPERATION:
The new Power Cartridge packaged with your ECLIPSE Oxygen System is not fully charged. Before using your ECLIPSE Oxygen System for the first time, you must fully discharge the Power Cartridge.
1. Turn your ECLIPSE ON and allow the Power Cartridge to completely discharge.
Unit will sound a low Power Cartridge alert until the unit completely shuts down. This could continue for up to approximately 20 minutes.
2. Connect the AC Power Supply and plug in the AC Power Cord to completely
charge the Power Cartridge. This may require up to five hours of uninterrupted charging. This completes the calibration of the Power Cartridge.
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|23
Page 26
WARNING DO NOT tamper with, disassemble, crush or heat the Power Cartridge
above 140°F (60°C). The Power Cartridge may present a risk of fire or explosion and will void the warranty.
CAUTION Store the Power Cartridge in a cool, dry place when not in use.
CAUTION DO NOT leave your ECLIPSE or your Power Cartridge in your vehicle
or in the trunk during a hot day.
CAUTION The ECLIPSE system can only work with a SeQual Power Cartridge.
Use of another Power Cartridge or Battery may damage the unit, present a risk of fire or explosion and will void the warranty.
In the event of a power interruption, the ECLIPSE Oxygen System will automatically switch to the Power Cartridge operation. When power is restored, the Power Cartridge will automatically start recharging. If the Power Cartridge is not present during a power interruption the ECLIPSE Oxygen System will stop operation. When power is restored, turn the system back ON to resume oxygen production.
Operating Instructions (continued)
24|ECLIPSETMModel 1000 User Manual
Page 27
Operating Your ECLIPSE
STEP 1: LOCATING YOUR ECLIPSE
Place the ECLIPSE in a well-ventilated place so that the air inlet and exhaust have clear access.
Position the ECLIPSE so that any alarm can be heard.
Check to see if the air inlet filter is in place before using. If the air inlet filter needs to be cleaned, wash with soapy water and allow drying before using. If necessary, replace with a new air inlet filter.
Plug unit into AC Power, DC Power, or make sure there is a charged Power Cartridge installed.
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|25
Page 28
STEP 2: TURN ON THE UNIT AND ALLOW TO WARM UP
Press and hold the “ON/OFF” Button for one (1) second to turn ON your ECLIPSE. A brief self-test will happen. Verify that all the indicators are lit and a 3-second buzzer sounds, indicating the system is in working order.
Warm-up Period
Oxygen Monitoring - The ECLIPSE Oxygen System has an oxygen monitoring system with a yellow (alert) and red (alarm) visual indicator and buzzer. The buzzer will activate in the event of an internal malfunction or power outage (external power loss with no Power Cartridge installed). The red, yellow, and green (all OK) indicators will initially light up and then the red and yellow indicators sequentially go out when the oxygen purity is within specification.
• After initially turning ON the unit, allow up to five (5) minutes for the unit to reach its specified performance. The control panel will indicate when the unit is done warming up. When the green ON indicator is lit, the system is ready for use.
Operating Instructions (continued)
26|ECLIPSETMModel 1000 User Manual
Red light
(alarm)
Yellow light
(alert)
Green light
(ready)
000.6
Page 29
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|27
STEP 3: CONNECT THE TUBING TO THE OXYGEN OUTLET AND THE CANNULA
PER THE INSTRUCTIONS FROM YOUR DEALER/PROVIDER.
Replace the cannula and tubing on a regular basis. Consult your dealer/provider for cannula and tubing cleaning and replacement information.
STEP 4: SELECT FLOW MODE
P
ress the
Flow Mode button to select the fl
ow delivery mode.
Continuous Flow Mode Operation
W
hen operating in Continuous Flow Mode, a continuous supply of o
xygen
will fl
ow through your tubing and nasal cannula.
WARNING When operating in Continuous Flow Mode, to ensure adequate flow,
D
O NO
T
use a tubing/cannula e
xtension more than 50’ (15.2m) long.
Page 30
Pulse Flow Mode Operation
• Pulse Flow Mode should only be used under the direction of a qualified clinician or physician.
• When operating in Pulse Flow Mode, an alarm will beep after 15 seconds if inspiration is not detected. If another 15 seconds elapses and no inspiration is detected, the unit will switch to Continuous Flow Mode at the last Continuous Flow Mode rate setting used.
WARNING Pulse Flow settings should be determined for each patient individually.
Settings from Continuous Flow applications may not be applicable to Pulse Flow Mode.
WARNING When operating in Pulse Flow Mode, to ensure adequate breath
detection, the length of the tubing/cannula must not exceed 7 feet (2.1m).
WARNING As with all conserving devices, the ECLIPSE may not be able to detect
some respiratory efforts.
STEP 5: ADJUST FLOW RATE TO PRESCRIBED LEVEL
Using the Flow Rate Setting buttons, adjust the flow rate to the prescribed setting supplied by your physician.
WARNING DO NOT change from prescribed flow rate setting without consulting
with a qualifi
ed clinician or physician.
28|ECLIPSETMModel 1000 User Manual
Page 31
STEP 6: BEGIN USE
Breathe normally through the cannula.
You will get longer operating time on your Power Cartridge if you operate your ECLIPSE in Pulse Flow.
STEP 7: CART OPERATION
The Cart is designed for use on pedestrian surfaces.
Align the locator holes on the bottom of the unit to the tabs on the cart.
Align the screw to the insert on the back of the ECLIPSE and hand-tighten the knob on the cart.
Push the button on the cart handle
to adjust the height.
STE
P 8:TUR
N OFF THE UNIT
Press and hold the “ON/OFF” Button for two (2) seconds to turn OFF your ECLIPSE.
Store the ECLIPSE unit in a cool, dry location.
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|29
Page 32
WARNING AVAILABILITY OF AN ALTERNATE SOURCE OF OXYGEN IS
RECOMMENDED IN CASE OF POWER OUTAGE OR MECHANICAL FAILURE. CONSULT YOUR PHYSICIAN FOR THE TYPE OF BACK-UP SYSTEM REQUIRED.
DO NOT IGNORE ALARMS.
TYPICAL
CONDITIONS
UNIT N EED S ATTENTIONSHUTDOWN CONDITIONS
The table shows possible alarm conditions and suggests appropriate responses. If any other alarms occur, call your authorized SeQual dealer/provider.
Green Yellow Red Audible
Condition Indicator Indicator Indicator Alarm What to do
“Warming Up. Please Wait.” On On On None Wait. The system typically takes
3–5 minutes to reach specified performance.
“2.0” (or flow setting) On Off Off None Nothing. The system is operating
properly at the specified flow rate.
“Low or Warm Power On Blinking Off Beeps Plug into external power or replace Cartridge” with a charged Power Cartridge.
“O2 Concentration < 85%” Off Blinking Off None Continue Use. Check or clean air
intake filter. If condition persists, contact dealer/provider.
“O
2
Concentration < 70%” Off Off Blinking Beeps Continue Use. Check or clean air
intake filter. If condition persists, contact dealer/provider.
“Flow Rate Error/ Off Blinking Off Beeps Check tubing and/or humidifier Blocked Flow” for obstruction. Check/clean
air intake filter. If condition persists past 10 minutes, contact dealer/provider.
“No Inspiration Detected” On On Off Beeps Make sure tubing does not (in Pulse Flow Mode only) exceed 7 ft (2.1m) when using
Pulse Flow Mode. Audible alarm will beep after 15 seconds if no inspiration is detected. If no inspiration detected 15 seconds after that, the ECLIPSE will automatically switch to Continuous Flow Mode.
“Power Cartridge Shutdown Off Off Blinking On Seek external power source. or hot Power Cartridge” (with If Power Cartridge malfunction no external power present)
persists, contact dealer/ provider.
“Loss of External Power” Off Off Blinking On Seek an alternative oxygen (without charged Power supply until external power is Cartridge installed) restored or install a charged
Power Cartridge.
“System Malfunction” Off Off On On Contact dealer/provider. Use
bac
k up o
xygen supply
.
Alarm Conditions and Troubleshooting
30|ECLIPSETMModel 1000 User Manual
Page 33
Symptom Probable Cause Remedy
ECLIPSE does not power on No Power Cartridge installed Install charged Power Cartridge when ON/OFF button is pressed or plug into external power
Power Cartridge is discharged Plug into external power
Malfunction Contact your dealer/provider
No Oxygen Blocked or kinked tubing Check tubing or replace
Filter Blocked Clean Air Inlet Filter
Humidifier Check humidifier attachment
and tubing
ECLIPSE not ON Turn ECLIPSE ON
Tubing or cannula is not properly Check tubing, cannula and connected or kinked connections
Power Cartridge is discharged Install charged Power Cartridge or no external power is present or attach external power
Malfunction Contact your dealer/provider
Low Oxygen Concentration Restriction in tubing or humidifier Repair or replace tubing
or humidifier
Filter restricted Clean or replace air inlet filter
Malfunction Contact your dealer/provider
Low Oxygen Flow Restriction in humidifier or tubing Repair or replace humidifier
or tubing
Filter Blocked Clean or replace air inlet filter
Malfunction Contact your dealer/provider
No Oxygen delivered in Tubing/cannula longer than Attach 7 foot (2.1m) Pulse Mode
7 feet (2.1m) tubing/cannula
Humidifier attached Remove humidifier
No inspiration detected Contact your dealer/provider
Power Cartridge Status Gauge Power Cartridge needs Fully discharge the Power never indicates fully charged calibration Cartridge and recharge
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|31
Page 34
WARNING Use of certain oxygen administration accessories not specified for
use with this oxygen system may impair the performance.
ONLY use
recommended oxygen delivery accessories.
There are many different types of oxygen tubing, cannula, and humidifiers. The following accessories are recommended to be used with your ECLIPSE Oxygen System. Refer to the “Maintenance” section of this manual for accessory maintenance instructions.
Humidifi
er - Salter Labs® Bubble Humidifi
er, Part Number 7600 or equivalent:
If your physician has prescribed a humidifier, follow the manufacturer’s instructions for use. A
ttac
h the humidifier to the oxygen outlet port of the oxygen system. The
O
2
symbol marks the oxygen outlet port on the concentrator. Use of humidifiers not
specified for this oxygen system may impair performance and void the warranty.
D
O NO
T
use a humidifi
er in P
ulse Flow Mode. The ECLIPSE will not detect an
inspiration and will default to Continuous Flow Mode.
Humidifi
er A
dapter - SeQual Part Number 3509:
If your physician has prescribed a
humidifi
er, you may need to use the SeQual Humidifier Adapter. Follow the instructions for use. Attach the Humidifier Adapter to the oxygen outlet port of the oxygen system and then to the humidifier. Attach cannula or extension tubing to humidifier outlet.
Accessories
32|ECLIPSETMModel 1000 User Manual
ECLIPSE Model 1000
DC Power Supply Model 6105
Power Cartridge Model 2400
Cart
AC Power Supply
Model 6104
Page 35
Oxygen Tubing - Salter Labs®Series 2000 or equivalent tubing (internal diameter of the tubing should be no less than 3/16” (0.48 cm):
Connect the oxygen tubing to the outlet port of the humidifier or directly to the oxygen outlet port if you do not use a humidifier. Connect the other end of the tube to the nasal cannula, if not already attached. Tubing not specified for use with this oxygen system may impair the performance.
WARNING The tube fitting may be tight. DO NOT use oil or grease to lubricate.
WARNING If tubing becomes disconnected during operation, ECLIPSE does not
alarm. Check flow at the cannula to verify oxygen delivery.
CAUTION To ensure there is adequate flow to deliver oxygen, the length of the
tubing must not exceed 50 feet (15.2m) when using Continuous Flow Mode, and must not exceed 7 feet (2.1m) when operating in Pulse Flow Mode for adequate breath detection.
Cannula – Salter Labs®Part Number 1600 Series or equivalent: Your physician will
have prescribed a cannula to deliver oxygen. In most cases they are already attached to the oxygen tubing. If not, follow the instructions included with the cannula to attach it to the oxygen tubing. Use of cannula not specified for use with this oxygen system may impair the performance.
Power Options: You have various options to provide power to your ECLIPSE.
A Power Cartridge, AC Power Supply and DC Power Supply.
Power Cartridge, Model 2400: The rechargeable Power Cartridge supplied with
your EC
LIPSE is capable of providing you with battery operation depending on your
prescribed flow rate.
Your Power Cartridge will be recharged when the ECLIPSE is connected to the AC P
ower Supply or the DC Power Supply, if adequate power is available. The Power Cartridge will take longer to recharge when connected to the DC Power Supply and will depend on the flow setting and the amperage switch setting.
Power Compartment: The Power Compartment is available when Power Cartridge
operation is not required. When using the Power Compartment, your ECLIPSE will
ONLY operate on the AC and DC Power Supplies.
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|33
Page 36
AC Power Supply, Model 6104: Your AC Power Supply enables the ECLIPSE to be
connected to a grounded AC electrical outlet with the country-specific power cord. Please contact your dealer/provider for the country-specific power cord.
DC Power Supply, Model 6105: Your DC Power Supply enables the ECLIPSE to be
operated in an automobile or other vehicle with an accessory outlet. Contact your dealer/provider to set up the DC Power Supply for proper use in your own automobile.
The indicator on the AC and DC Power Supplies will be either green or red. Green indicates normal operation and red indicates abnormal operation.
WARNING DO NOT expose the AC and DC Power Supplies to water. Electrical
shock or serious injury may result.
Cart: Your ECLIPSE Oxygen System can be easily transported while attached to the
Cart. The cart has wheels and a telescoping handle for easy mobility.
WARNING DO NOT drip, spill or expose the ECLIPSE to water during portable
use. The ECLIPSE enclosure does not protect against the harmful effects of liquid ingress. Electrical shock or damage to the unit may result.
CAUTION When operating the ECLIPSE on the cart, DO NOT bang, drop or
bump the unit while in transport.
Accessories (continued)
34|ECLIPSETMModel 1000 User Manual
Page 37
Cleaning, Care and Routine Maintenance
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|35
Routine Maintenance
A trained technician will carry out all routine maintenance at least once a year or as needed.
Your dealer/provider will perform inspections and maintenance of the compressor, filters, alarms, 9V battery, Power Cartridge and other internal parts as needed.
WARNING DO NOT disassemble the ECLIPSE or attempt to perform any routine
maintenance other than what is described in the Patient Cleaning and Care section below. Disassembling the oxygen system may create a hazard of electrical shock and void your warranty. Contact your dealer/provider for any maintenance required.
Patient Cleaning and Care
Cannula Replacement: Replace your tubing and cannula on a regular basis as
recommended by your dealer/provider. Your physician or dealer/provider will provide you with cleaning, disinfection and replacement information.
Air Inlet Filter: Ambient air is drawn into the unit through the air inlet located on the
top/back of the unit. Cleaning the air inlet filter is the most important activity you will perform to keep your oxygen system performing properly. Check this air filter daily and clean it at least once a week.
1. Remove the fi
lter
.
2. Wash the filter in warm water
using a mild detergent solution.
3. Rinse the filter thoroughly and
squeeze out the excess water.
4. Allow the filter to air dry.
5. Reinsert the filter in the cabinet.
Page 38
CAUTION The filter should be free of liquid water and dry to the touch before
reinstallation, as excess moisture may impair the proper operation of the oxygen system.
CAUTION When the filter is removed, the Service Port is exposed. DO NOT
attach any cables to this Service Port. It is for Service ONLY.
CAUTI O N DO NOT poke objects through the vent holes.
The air inlet filter should be replaced as needed. If the oxygen system is used in a dusty environment, the filter may need to be replaced more often. You should not operate the ECLIPSE without a filter installed for longer than 30 minutes.
Cabinet, AC Power Supply, DC Power Supply and Power Cord: Turn OFF the
ECLIPSE and disconnect from AC or DC power before any cleaning or disinfection activity. Use a damp (not soaking wet) cloth or sponge and a mild detergent solution to clean the cabinet and power supplies. To disinfect the ECLIPSE use Lysol®Brand II disinfectant. Proceed as directed by the manufacturer.
WARNING Unplug Power Cords, AC or DC Power Supplies before cleaning the
exterior cabinet.
DO NOT use denatured alcohol or apply liquid spray
or aerosol cleaners.
Humidifier: If used, clean your humidifier daily to reduce possibility of contamination.
Replace humidifier monthly or as recommended by the manufacturer or your dealer/provider.
Cart: Before cleaning the cart, remove the ECLIPSE. Use a damp (not soaking wet)
cloth or sponge and mild detergent solution to clean the Cart.
36|ECLIPSETMModel 1000 User Manual
Page 39
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|37
Power Cartridge Cleaning, Care and Maintenance
The Power Cartridge in the ECLIPSE Oxygen System requires special care to assure a longer life and higher performance. The SeQual Power Cartridge is the only approved Power Cartridge to be used.
Once a month, completely discharge your Power Cartridge and fully recharge to calibrate the Power Cartridge.
Use a damp cloth or sponge and mild detergent to clean the Power Cartridge case and latch.
WARNING DO NOT tamper with or try to repair the Power Cartridge.
No serviceable parts inside.
Power Cartridge Disposal: Your Power Cartridge is rechargeable and can
be recycled. Always return it to your dealer/provider for proper disposal.
Power Cartridge Storage: Your Power Cartridge should be stored in a cool,
dry location.
WARNING Exposing the Power Cartridge to water or other liquids may cause
personal injury.
Page 40
Flow Rate Settings 1 to 6 setting in Pulse Flow Mode
0.5 to 3.0 LPM in Continuous Flow Mode
Oxygen Concentration 90% +/– 3% for all flow settings
Oxygen Concentration Indicator Green Light = Normal Operation
Yellow Light = Warning or Caution, less than 85% +/– 3% Red Light Flashing = Abnormal Operation, less than 70% +/– 5%
Maximum System Pressure 15 psig (103.5 kPa)
Oxygen Outlet Pressure 5.0 psig (34.5 kPa) nominal
Electrical Power:
AC Operation 100 – 240 VAC, 50/60 Hz
DC Operation 12 V
Power Cartridge 195 W-Hrs
AC Power Supply 200 Watts Max
DC Power Supply 150 Watts Max
Operating Temperature 50° F to 104° F (10° C to 40° C)
Operating Humidity 10% - 95% at 82.4° F (28° C) dew point
Nominal Sound Level 48 dBA at 3.0 LPM in Continuous Flow Mode
40 dBA at 3.0 LPM in Pulse Flow Mode
Dimensions (H x W x D) 19.3 inches x 12.3 inches x 7.1 inches
(49.0 cm x 31.2 cm x 18.0 cm)
Weight 17.9 pounds (8.1 kg) with Power Cartridge
14.5 pounds (6.6 kg) without Power Cartridge
Continuous Flow Indication Expressed in liters per minute (LPM)
Audible Alarm Indicators Loss of power/Hot Power Cartridge
Low Battery/Warm Power Cartridge Low Oxygen Output O
2
ow outside normal limits No inspiration detected in Pulse Mode Unit Malfunction
Back-Up Alarm Power 9V Internal Battery
Filters Air Inlet, HEPA
Device Classification IEC Class I, Type B Applied Part, IPX0
Transport/Storage Requirements Temperature: -4° F (-20°C) to +140° F (+60° C)
Humidity: up to 95% Non-Condensing
Back pressure Effect on Flow Indicator None
Altitude 0 –13,123 feet (0 – 4,000 meters)
Technical Data
*
38|ECLIPSETMModel 1000 User Manual
Page 41
Pulse Flow Mode
Principle of Operation Pulse Flow Mode is designed to deliver 16 ml boluses of oxygen
per incremental flow setting for each patient inspiration detected. Bolus volume is held constant regardless of breath rate within a given setting. Delivery of the bolus is triggered by detection of the end of inspiration followed by the start of a subsequent inspiration. Sensitivity is dealer/provider adjustable. Based on this inspiration trigger signal, a “pulse” of oxygen is released to the patient by an electronically controlled flow valve. There is no mixing of oxygen with room air.
Range of Operation — Settings Increments of 1– 6 times (x) the bolus of 16 ml
(16 to 96 ml per breath)
Accuracy of Delivered Bolus ± 15%
Breathing Frequency range 6 to 25 breaths per minute average, sustainable
Minimum time between breaths 1.25 seconds (maximum 3 consecutive breaths)
Response to missing breaths ECLIPSE switches to Continuous Flow Mode after 30 seconds
without breath detection.
* Specifications are subject to change without notice.
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|39
Page 42
Summary of All Warnings and Cautions
40|ECLIPSETMModel 1000 User Manual
WARNING FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE
ORDER OF A PHYSICIAN.
WARNING THE ECLIPSE IS NOT INTENDED FOR LIFE SUPPORTING OR LIFE
SUSTAINING APPLICATIONS, NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES.
WARNING IN CERTAIN CIRCUMSTANCES, THE USE OF NON-PRESCRIBED
OXYGEN CAN BE HAZARDOUS. THIS DEVICE SHOULD ONLY BE USED WHEN PRESCRIBED BY A PHYSICIAN.
WARNING FOR ADULT USE ONLY. NOT QUALIFIED FOR USE ON
PEDIATRIC PATIENTS.
WARNING NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS.
WARNING AS WITH ANY ELECTRICALLY POWERED DEVICE, THE USER MAY
EXPERIENCE PERIODS OF NON-OPERATION AS THE RESULT OF ELECTRICAL POWER INTERRUPTION, OR THE NEED TO HAVE THE OXYGEN CONCENTRATOR SERVICED BY A TECHNICIAN. AN OXYGEN CONCENTRATOR IS NOT APPROPRIATE FOR ANY PATIENT WHO WOULD EXPERIENCE ADVERSE HEALTH CONSEQUENCES AS THE R
ES
U
LT OF SUCH TEMPORARY INTERRUPTION.
WARNING “No Smoking or Naked Flames”: The use of oxygen therapy requires
that special care be taken to reduce the risk of fire. Oxygen enhances the combustion of fl
ammable material. F
or safety concerns, all possible sources of ignition must be kept away from the oxygen system and preferably out of the room in which it is being used. Smoking in the proximity of an operating oxygen concentrator is dangerous, and can permanently damage the device and void the warranty.
Page 43
WARNING “Use No Oil or Grease”: A spontaneous and violent ignition may
occur if oil, grease or other petroleum substances come into contact with oxygen under pressure. Keep these substances away from the oxygen system, tubing and connections and any other oxygen source.
DO NOT use any petroleum based or other lubricants.
WARNING DO NOT remove the cover, no serviceable parts inside. Qualified
service personnel should only remove the cover.
WARNING DO NOT operate device or accessories on wet surfaces or in standing
water and
DO NOT submerse or expose to water. The ECLIPSE
enclosure does not provide protection against the harmful effects of liquid ingress. Electrical shock or damage to the unit may result.
WARNING Protect electrical cords from sharp edges. Electrical shock and
serious physical injury may result.
WARNING Use only the SeQual AC Power Supply and DC Power Supply supplied
with your ECLIPSE Oxygen System. Use of a non-SeQual Power Supply may be hazardous, cause damage to your oxygen system and will void the warranty.
W
AR
N
ING
T
his P
ower Cartridge may e
xplode and cause potential injury if
exposed to or disposed of in a fire.
WARNING DO NOT short-circuit the Power Cartridge’s metal contacts with
metallic objects, suc
h as keys or coins. It may cause sparks or
excessive heat.
WARNING DO NOT disassemble, puncture, or crush the Power Cartridge. Power
Cartridge electrolytes may be toxic if swallowed and can be harmful to skin and eyes. Keep the Power Cartridge away from children.
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|41
Page 44
WARNING Use of a damaged Power Cartridge may cause personal injury.
WARNING Exposing the Power Cartridge to water or other liquids may cause
personal injury.
WARNING Protect the ECLIPSE and AC/DC Power Supplies from fluid spills or
drips to avoid possible shock hazards.
WARNING DO NOT use any cord if damaged. Improper use of the Power Cord
may result in a fire or electrical shock hazard.
WARNING DO NOT touch the Power Cartridge cavity connectors with the Power
Cartridge removed while the unit is connected to an external power source. This may result in electrical shock.
WARNING Secure the ECLIPSE and the DC Power Supply in your vehicle and
make sure the air inlet and exhaust airflow are not blocked. This may impact performance.
WARNING DO NOT leave the ECLIPSE or DC Power Supply plugged into the
vehicle without the engine running or attempt to start the vehicle while the DC Power Supply is connected to the vehicle. This may drain the vehicle’s battery.
W
AR
N
I
N
G
If the vehicle power source (vehicle accessory connector) drops below
11.5 Volts (brownout) the ECLIPSE will revert to Power Cartridge operation (if present). T
he External P
ower P
resent Indicator may
remain on.
WARNING When operating in Continuous Flow Mode, to ensure adequate flow,
DO NOT use a tubing/cannula extension more than 50’ (15.2m) long.
42|ECLIPSETMModel 1000 User Manual
Page 45
WARNING Operating your ECLIPSE outside the recommended operating
environment may impact performance, cause damage and will void the warranty.
WARNING Pulse Flow settings should be determined for each patient individually.
Settings from Continuous Flow applications may not be applicable to Pulse Flow Mode.
WARNING When operating in Pulse Flow Mode, to ensure adequate breath
detection, the length of the tubing/cannula must not exceed 7 feet (2.1m).
WARNING As with all conserving devices, the ECLIPSE may not be able to detect
some respiratory efforts.
WARNING DO NOT change from prescribed flow rate setting without consulting
with a qualified clinician.
WARNING AVAILABILITY OF AN ALTERNATE SOURCE OF OXYGEN IS
RECOMMENDED IN CASE OF POWER OUTAGE OR MECHANICAL FAILURE. CONSULT YOUR PHYSICIAN FOR THE TYPE OF BACK-UP SYSTEM REQUIRED.
DO NOT IGNORE ALARMS.
W
AR
N
ING
Use of certain o
xygen administration accessories not specifi
ed for
use with this oxygen system may impair the performance.
ONLY use
recommended oxygen delivery accessories.
W
AR
NING
T
he tube fi
tting may be tight.
D
O NO
T
use oil or grease to lubricate.
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|43
Page 46
WARNING If tubing becomes disconnected during operation, ECLIPSE does
not alarm. Check flow at the cannula to verify oxygen delivery.
WARNING DO NOT tamper with, disassemble, crush or heat the Power Cartridge
above 140° F (60° C). The Power Cartridge may present a risk of fire or explosion and will void the warranty.
WARNING DO NOT expose the AC and DC Power Supplies to water. Electrical
shock or serious injury may result.
WARNING DO NOT disassemble the ECLIPSE or attempt to perform any routine
maintenance other than what is described in the Patient Cleaning and Care section below. Disassembling the oxygen system may create a hazard of electrical shock and void your warranty. Contact your dealer/provider for any maintenance required.
WARNING Unplug Power Cords, AC or DC Power Supplies before cleaning
the exterior cabinet.
DO NOT use denatured alcohol or apply liquid
spray or aerosol cleaners.
WARNING DO NOT tamper with or try to repair the Power Cartridge. No
serviceable parts inside.
W
AR
N
ING
Exposing the P
ower Cartridge to water or other liquids may cause
personal injury.
W
AR
N
ING
D
O NO
T
drip, spill or e
xpose the EC
L
IPSE to water during portable
use. T
he EC
LIPSE enclosure does not protect against the harmful effects of liquid ingress. Electrical shock or damage to the unit may result.
44|ECLIPSETMModel 1000 User Manual
Page 47
CAUTION Availability of an alternate source of oxygen is recommended in case
of power outage or mechanical failure.
CAUTION DO NOT expose unit to temperature outside of specified operating or
storage temperatures as this may damage the unit.
CAUTION Locate oxygen tubing and Power Supply cords to prevent tripping
hazards.
CAUTION DO NOT place ECLIPSE in a small enclosed space, such as a
bathroom or closet, with cannula leading out of the enclosed space.
CAUTION DO NOT expose the Power Cartridge to temperatures above 140° F
(60° C), such as in a vehicle parked in the sun or on a hot day.
CAUTION Charging the Power Cartridge below 50 F° (10° C) or above 104° F
(40° C) may degrade the performance.
CAUTION DO NOT drop the Power Cartridge or expose it to mechanical shock.
CAUTI O N Only use the Power Cartridge for its intended purpose.
CAUTION Locate the unit in a well-ventilated room that provides adequate air
flow and check if air is being prevented from entering or exiting the system by furniture, draperies, bedding, clothing or pets.
C
A
U
T
I
ON
Always c
hec
k to see that the Air Inlet and the Exhaust V
ent are not
blocked and the air inlet filter is clean before using your system.
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|45
Page 48
CAUTION The AC power cord used with the ECLIPSE Oxygen System must
meet the requirements of the country where the product is used.
CAUTION Use with grounded, 3-pin receptacle. DO NOT remove ground pin
from cord. Use only with SeQual supplied cords.
CAUTION DO NOT connect the ECLIPSE to an extension cord or electrical
outlet controlled by a switch.
CAUTI O N Store your Power Cartridge in a cool, dry location. This will help to
assure the longevity of your Power Cartridge.
CAUTION DOT/UN Regulations require the removal of the Power Cartridge
for international travel.
CAUTION There are no serviceable parts inside the Power Cartridge.
DO NOT open.
CAUTION Replace the Power Cartridge with only SeQual approved Power
Cartridge.
CAUTION The Power Cartridge used in this device may present a risk of fire or
chemical burn if mistreated.
DO NOT disassemble, incinerate, or heat
above 140° F (60° C).
CAUTION Return your Power Cartridge to your dealer/provider for proper
disposal.
46|ECLIPSETMModel 1000 User Manual
Page 49
CAUTION Keep your Power Cartridge away from children.
CAUTION Store the Power Cartridge in a cool, dry place when not in use.
CAUTION DO NOT leave your ECLIPSE or your Power Cartridge in your vehicle
or in a trunk during a hot day.
CAUTION The ECLIPSE system can only work with a SeQual Power Cartridge.
Use of another Power Cartridge or Battery may damage the unit, present a risk of fire or explosion and will void the warranty.
CAUTION To ensure there is adequate flow to deliver oxygen, the length of the
tubing must not exceed 50 feet (15.2m) when using Continuous Flow Mode, and must not exceed 7 feet (2.1m) when operating in Pulse Flow Mode for adequate breath detection.
CAUTION When operating the ECLIPSE on the cart, DO NOT bang, drop or
bump the unit while in transport.
CAUTION The filter should be free of liquid water and dry to the touch before
reinstallation, as excess moisture may impair the proper operation of the oxygen system.
CAUTION When the filter is removed, the Service Port is exposed. DO NOT
attac
h any cables to this Service P
ort. It is for Service
ON
L
Y
.
CAUTION DO NOT poke objects through the vent holes.
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|47
Page 50
Guidance and Manufacturer’s declaration - electromagnetic emissions
The SeQual Technologies ECLIPSE Oxygen System is intended for use in the electromagnetic environment specified below. The customer or the user of the SeQual Technologies ECLIPSE Oxygen System should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions CISPR 11 Group 1 The SeQual Technologies ECLIPSE Oxygen
System uses RF energy only for its internal function. T
herefore, its R
F emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The SeQual Technologies ECLIPSE Oxygen
System is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-volt
age power supply network that supplies buildings used for domestic purposes.
Harmonic emissions Class A IEC 61000-3-2
Voltage fluctuations/ Complies flicker emissions IEC 61000-3-3
This equipment has been tested and found to comply with the limits for medical devices to the IEC60601-1-2: 2001-09 Electromagnetic Compatibility standard. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed according with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving device.
Increase the separation distance between the equipment.
Connect the equipment into an outlet on a circuit different from that which the other device(s) are connected.
Consult the manufacturer or service technician for help.
Electromagnetic Compatibility
48|ECLIPSETMModel 1000 User Manual
Page 51
Disposal of Equipment and Accessories
Model 1000 User Manual
ECLIPSETMModel 1000 User Manual|49
Follow local governing ordinances and recycling plans regarding disposal of device components.
DO NOT dispose of the Power Cartridge. Power Cartridges should be
returned to your dealer/provider.
MANUFACTURED BY:
SeQual Technologies Inc. 11436 Sorrento Valley Road San Diego, CA 92121 USA (858) 202-3100 (800) 826-4610
www.sequal.com
EUROPEAN UNION AUTHORIZED AGENT:
MeddiQuest Limited Business & Technology Centre Bessemer Drive Stevenage SG1 2DX United Kingdom Tel: + 44 1763 222505 Fax: + 44 1763 222049
DEALER/PROVIDER NAME AND ADDRESS:
SeQual®and ECLIPSETMare trademarks of SeQual Technologies Inc., San Diego, CA 92121
Salter Labs®is a registered mark of Salter Labs, Arvin, C
A 923
03
Lysol
®
is a registered trademark of Reckitt Benckiser, UK.
PHYSICIAN NAME AND ADDRESS:
Page 52
ECL
I
P
S
E
TM
Model 1000 User Manual
PN 6112 03/06