Sencore SC3100 User Manual

Now Touch And Test Any Circuit Test Point
And Make Autoranged Error Free Measurements
The SC3100 offers:
A Complete Waveform And Circuit Analyzing System
Measure circuit parameters and view all of the waveforms shown in any service literature with one complete unit. The
SC3100 is guaranteed to increase your
analyzing capabilities with the push of a button.
Auto-Tracking Digital Readout Of Waveform Voltage And Frequency
Measure the key parameters of any waveform with one probe connection, at the push of a button, for fast and accurate troubleshooting.
Integrated Measurements Of All Circuit Parameters
There’s no need for a separate DVM to analyze the rest of the circuit para-
meters Measure ohms and current
with an integrated, complete circuit
analyzer that provides you with
troubleshooting answers.
In A Fraction Of The Time!
Full Performance, 100 MHz, Dual Trace Oscilloscope
View any waveform quickly, easily, and
more accurately. The “fiddle free” trigger controls provide rock solid viewing of any
signal and include a special TV mode for
complex video waveforms. No signal is too large or too small with our exclusive 2 mV to 2 kV input range.
Exclusive Autoranged Timebase And Vertical Attenuators
No more time wasted turning knobs. Simply set the timebase and channel attenuators to Auto and view the waveform without resetting the controls as you step through the circuit. This allows you to concentrate on the circuit - not
the equipment.
Digital Delta Measurements To Analyze Every Portion Of Any Waveform
Highlight any part of a waveform with Sencore’s exclusive Delta Bar and analyze the amplitude, absolute DC, time, or frequency. No more wasting time on graticule counting or setting cursors.
All Functions Microprocessor
Integrated For Ease Of Use
The SC3100’s analyzing speed will increase your servicing capability. All measurements are based on digital circuits, not the analog CRT, for fast, easy, and accurate readings. There are no hidden menus, no multiple function buttons, no complicated setups, and no confusing on-screen displays. Just push a button and read the results on the LCD display. The ‘AUTO TRACKER"™ eliminates any chance of measurement errors.
SENCORE
Call 1-800-SENCORE (736-2673)
SC3100 SPECIFICATIONS:
Vertical Amplifiers:
DISPLAY MODES: Channel A, inverted channel A (-A), channel B, dual trace (A&B). algebraic sum (A+B) or difference (B-A), vector (X-Y). CALIBRATION ACCURACY:
FREQUENCY RESPONSE (100 MHz). AC coupled: to 100 MHz, useable to 150 MHz SENSITIVITY: 20 mV/div. to 200 V/div. with supplied 396292 mV/div. to 20 V/div. with (optional) DP270 Direct Probe.
MAXIMUM INPUT PROTECTION: Supplied 396292 10X Probe: 2500 volts breakdown (DC + Peak AC).
Horizontal Sweep:
SWEEP RATES: 100 milliseconds/division to 20 nanosecond/division. Autorange automatically selects sweep rate to show approximately 2-5 cycles of waveform. ACCURACY:
Trigger Circuits:
TRIGGER SOURCE, CH A, CH B, AC power line, or external. TRIGGER MODES. NORM. AUTO, TV
AUTO-TRACKING™ Digital Tests DC Volts:
DCV FUNCTION: Provides direct reading of DC voltage on selected channel.
ACCURACY ±0.5% ±2 digits
Peak-to-Peak Volts:
VPP FUNCTION: Provides direct reading of peak-to-peak voltage on selected
channel
ACCURACY: ±2% ±4 counts.
FREQUENCY RESPONSE: ±0.5dB from 20 Hz to 30 MHz. ≤-3dB at 100 MHz.
AC Volts:
ACV FUNCTION. Calculates RMS sinewave value from PPV measurement.
dBm:
dBm FUNCTION: Calculates dBm measurement from PPV sinewave
measurement, referencing 1 mW across 600W (0 dBm +.7746 volts RMS).
Frequency:
FUNCTION: Automatically displays the frequency of the signal on selected
channel.
RANGES: 10.00 Hz to 150 MHz.
ACCURACY: .001% ±1 digit.
Delta Peak-to-Peak:
FUNCTION: Measures amplitude of intensified area on selected channel.
Delta Time:
FUNCTION: Measures time of intensified waveform portion.
±3%.
with either X10 or direct probes.
±3% at 1 kHz.
±dB
of 1 kHz level from 10 Hz
10X
probe; 2
1/Delta Time:
FUNCTION: Converts Delta lime reading to equivalent frequency.
Delta DC Volts
FUNCTION: Measures DC voltage level of marked waveform point in respect to ground using the PPV and DCV functions. MARKER: Fully adjustable over entire range of waveform.
Digital Meter Tests: Ohms
FUNCTION: Provides in- or out-of-circuit ohms. RANGES: 0 00 to 100 MW ACCURACY: 0.2% ±2 digits.
Continuity Test
FUNCTION: Provides audible tone of continuity RANGE: 0 to 199W Audible tone turns on if resistance is <10W and turns off if resistance is >15W ±2W
DC Current
FUNCTION: Provides measurement of DC current RANGES: .001 to 1 ACCURACY: 0.3%
General:
SIZE: 7.25” x 13.75” x 15” HWD (18.4 x 34.9 x 38.1cm) WEIGHT: 25 Ibs (9.33 kg.)
Accessories: Supplied:
2 ea: 396292 10X Low Capacity probe with 3” ground, 12” ground and ground insulator ring, color ID bands and retractable tip.
1 ea: 396295 DC voltage probe.
Optional:
HP200: 50 KV High Voltage Probe PL207: RF Pickup Loop TP212: 10 KV Transient Protector Probe PC259: Protective Cover DP270: 1:1 Direct Probe 396293: 250 MHz Demodulator Probe 396303: Probe Accessory Kit
amp
±
digits
Covered by Patents #4,473,857, #4,564,805 and 4,323,972
Specifications
subject to change without notice.
Sencore’s Exclusive After-The-Sale Support
All Sencore Instruments Are Backed By Our Exclusive 30 Day Money Back Guarantee
“If you are not completely satisfied with any Sencore instrument, you may return it during the first 30 days, and we’ll give you a full refund,
including freight, no questions asked.”
Toll Free Access To The Entire Factory
Sencore does not stop providing assistance after you purchase one of our instruments. We provide free technical assistance on any
instrument use questions, future purchase assistance, industry trend information, instrument service, or any question you may have on test instruments. Just call Sencore at our toll free
1-800-SENCORE
Technician will be happy to assist you.
Made Right Guarantee
In addition to the 1 year warranty on parts and labor (less than 2% “AUTO TRACKER"™ 100 MHz Waveform
are returned for warranty service), Sencore offers you a 100% Lifetime Made Right Guarantee. We guarantee your instrument was engineered right the first time. If there’s a problem, we’ll make it right,
for the lifetime of the instrument.
New!
Extended Service Agreement Programs
(ESA)
Sencore announces our new Extended Service Agreement Programs for your SC3100. These programs can be an extension of your warranty coverage or simply a method for you to have annual calibrations at a reduced rate.
programs, please contact your Area Sales Engineer or the Service Department.
number. Your Area Sales Engineer, an Application Engineer, or a Service
Here’s How To Order Your SC3100
&
Circuit Analyzer
Ordering the
SC3100 "AUTO
Circuit Analyzer is as simple as picking up your phone. Call 1-800-SENCORE and talk directly with your Area Sales
Engineer about the many programs available to help with your instrument investment.
TRACKER” 100 MHz Waveform
For further information regarding these
&
Call 1-800-SENCORE
SENCORE
Form #5129
(736-2673)
Loading...