sebaKMT Hydrolux HLE 500, Hydrolux HLE 5000 Operating Instructions Manual

Operating instruction
Hydrolux® Leak locator HLE 500
Produktbild wie in den Prospekten ohne Rahmen Ansprechpartner Frau Badziura
HLE 5000
Version 1.0
Mess- und Ortungstechnik Measuring and Locating Technologies
Elektrizitätsnetze Power Networks
Kommunikationsnetze Communication Networks
Leitungsortung Line Locating
Edition: 02 (2005/02)
Operating instructions HLE 500/5000
II
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
III
Quality certificate
The quality control system used by Seba Dynatronic® and sebatel® meet the highest requirements of the
international quality standard DIN ISO 9001 and the European standard EN 29001. This quality system was approved by the German Association for the Certification of Quality Systems with the numbers EN 19677 and DQS 19677-01.
How do you contact us?
SebaKMT
Mess- und Ortungstechnik GmbH
Dr.-Herbert-Iann-Str. 6
96148 Baunach, Germany
Telephone: +49 (0) 9544 680
Fax: +49 (0) 9544 2273
Email: sales@sebakmt.com
MAN_HLE500_HLE5000_eng_02
Internet: www.sebakmt.com
Operating instructions HLE 500/5000
IV
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
V
Table of contents
1 Description 1
1.1 General 1
1.2 Construction 2
1.3 Technical data 2
1.4 Delivery includes 3
1.4.1 5000/500 equipment 3
1.4.2 Selectable microphone sets 3
1.4.3 Optional accessories 4
2 Getting to know the HLE 5000/500 5
2.1 HLE 5000/500 controls 5
2.2 Connections at the side 5
2.3 Battery chamber 6
2.4 Automatic battery monitoring 7
2.5 Illumination of the LCD display 7
2.6 Microphones 7
2.6.1 PAM W-1 8
2.6.2 PAM U 9
2.7 Headphones 10
3 Practical work 11
3.1 Headphones and microphone connection 11
3.2 Switching on
3.3 Volume and amplification 12
3.3.1 Setting the volume 12
3.3.2 Setting the gain 13
3.4 The amplification display 13
3.5 Filter setting (only HLE5000) 14
3.5.1 Adjusting the cut-off frequencies 14
3.6 Filter selection in practice 15
3.6.1 Filter selection A (Ground microphone PAM W-1, PAM U with 3-point foot) 16
3.6.2 Filter selection B: Sensor rod (Pam U with spike) 16
3.7 Mute button
3.8 The histogram 16
3.9 Long-term measurement (HLE 5000 only) 18
3.10 Location of pipes using the RSP-3 or PWG (HLE 5000 only) 19
3.11 Switching off

11
X
16

19
4 Faults 20
4.1 Can't switch on 20
4.2 Battery monitor does not react 20
4.3 No sound can be heard 20
4.4 Scratching sounds in the headphones 20
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
VI
List of illustrations
Fig 1 : HLE 5000/500 control panel...................................................5
Fig 2 : Side view left and right with connections................................6
Fig 3 : Base plate with locking screws...............................................7
Fig 4 : PAM U with sensor rod...........................................................9
Fig 5 : PAM U with magnet................................................................9
Fig 6 : PAM U with 3-point foot........................................................10
Fig 7 : Start image with version number..........................................11
Fig 8 : Opening menu after switch-on..............................................12
Fig 9 : Setting the volume................................................................12
Fig 10 : Setting the gain.....................................................................13
Fig 11 : Current and minimum value .................................................14
Fig 12 : Filter settings........................................................................14
Fig 13 : Adjusting the lower filter cut-off frequency............................15
Fig 14 : Adjusting the upper filter cut-off frequency...........................15
Fig 15 : Dual segment analysis (DSA)...............................................17
Fig 16 : Histogram measurement ......................................................17
Fig 17 : Long-term measurement ......................................................18
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
1 Description
1.1 General
The new HLE 5000/500 leak locator comprehensively deals with finding leaks in pipe networks for the supply of drinking water. The equipment can of course be used on other pipe systems, provided that the liquid comes out of the pipe under pressure and that the resulting noise carries to the surface of the ground. With the HLE 5000/500 it is possible to both pre­locate and to pinpoint the location of the fault. With a special Dual Segment Analysis (DSA) display, both the minimum value of the constant noise as well as the leak noise can be recognised. With the “Mute” option, used when moving the ground microphone, a new minimum display results. A comparison of the measured values is thus always possible. In the “HISTOGRAM” function, each measurement location is stored, one by one. This equipment is the first instance where computer supported reduction of extraneous noise is used, where impulse type disturbances are acoustically suppressed. Preferably, only constant noise (as produced by a leaking pipe) is displayed as a minimum value.
When locating plastic pipes with the assistance of an RSP3 “pipe pecker”, picking up impulses is then desirable and improves the results. For this reason, the HLE 5000 has a special pipe location mode that is activated, after switching on, by pressing the symbol button.
Another important function of the HLE 5000 is the noise level recording with which the course of the development of the noise is shown on the LCD display as a graph over time.
1
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
1.2 Construction
The leak location equipment is in a splash-proof housing made of robust plastic. The few control buttons are so arranged that they can also be operated wearing gloves. The display is equipped with backlighting. That means that work is also possible under poor lighting conditions, or at night.
The batteries required to supply the power are housed in the base of the equipment behind a cover which is easy to open to quickly exchange the batteries. The connections for the microphone and headphones are on the two sides of the equipment and can be quickly plugged in or removed.
2
1.3 Technical data
Analysis bandwidth: 0 Hz – 4000 Hz Filter cut-off frequencies: 0 - 70 Hz, 106 Hz, 160 Hz,
Histogram recordings: 9 dual displays Storage of noise levels: 3 – 10 – 30 min Display 130 x 36 mm: LCD display LCD illumination: available Power supply 1: Power supply 2 (option):
Operating time: > 35 h (battery), Storage: 9 measurements
Mute button: available Operating temperature: -10 to +50 °C Storage temperature: -10 to +70 °C Protection class when in operation: Dimensions (L x W x D): 215 x 95 x 110 mm Weight: HLE 5000 Weight: PAM W-1 Weight: PAM U
240 Hz, 360 Hz, 540 Hz, 800 Hz, 1200 Hz, 1800 – 4000 Hz
10 x AA batteries, (1.5 Volt) 12 x rechargeable batteries (1.2 Volt)
> 15 h (rechargeable battery)
IP 54 1200 g (with batteries)
3.5 kg (with carrying pole) 500 g
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
1.4 Delivery includes
1.4.1 5000/500 equipment Leak location equipment with batteries HLE 5000 or HLE 500
Headphones (extraneous noises filtered) Carrying strap Case HLK Operating instructions BED HLE 5000
1.4.2 Selectable microphone sets
Professional set consisting of:
Piezo ground microphone (with wind shield) Connection cable to PAM W-1 VK65 3-point foot adapter PAM W-1D Carrying pole for PAM W-1 Universal PAM U microphone PAM U Magnetic adapter Sensor rod 3-point foot for PAM U PAM U-D Extension rod VST T-1
Ground microphone (with wind shield) consisting of:
Piezo ground microphone (with wind shield) Connection cable to PAM W-1 VK65 3-point foot adapter PAM W-1D Carrying pole for PAM W-1
Ground microphone
Piezo ground microphone PAM B-1 Connection cable to PAM W-1 VK65 Carrying pole for PAM B-1
Universal microphone
Universal microphone PAM U Magnetic adapter Sensor rod 3-point foot adapter for PAM U PAM U-D
KR 2
PAM W-1
PAM W-1
3
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
1.4.3 Optional accessories
4
Radio module for wireless transmission between microphone and amplifier Sensor rod extension VST T-1 Special headphones (highly insulated from extraneous noise) Sliding adapter 42 mm AD S-42 Sliding adapter 20 mm AD S-20 Magnetic adapter
PAM 868
KM2
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
I/O
HYDROLUX
®
with backlight
2 Getting to know the HLE 5000/500
Before putting the equipment into operation, you should get to know all of the controls, the devices which can be connected and the power supply.
2.1 HLE 5000/500 controls
5
ON/OFF/
backlight
button
Mute button
2.2 Connections at the side
LCD display
X
Fig 1 : HLE 5000/500 control panel
Volume control
Soft keys
The sockets for the microphone and the headphones are on the two sides of the equipment. The connections are shown in fig 2.
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
connection
6
Headphones
connection
2.3 Battery chamber
Microphone
Fig 2 : Side view left and right with connections
The sockets are only suitable for headphones and microphones from SebaKMT. Connecting other headphones or microphones can lead to equipment failure or damage to the HLE 5000/500.
After opening the base flap (fig. 3) the battery chamber is accessible. When fitting the batteries make sure that the polarity is correct.
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
Locking screws
Fig 3 : Base plate with locking screws
2.4 Automatic battery monitoring
7
When working with the leak detection equipment, the state of the batteries is monitored continuously. Should the battery voltage drop too far, this will be indicated by a flashing battery symbol in the top right of the display. From that point, there is about 4 hours of battery life remaining. When you have to change the batteries, change all 10 batteries at one time!
2.5 Illumination of the LCD display
By briefly pushing the on/off button equipment is turned on, the illumination for the display will be activated / deactivated.
2.6 Microphones
The microphones are connected to the leak location equipment via a cable. The socket for this connection is on the right hand side of the equipment (see fig. 2). To record the sound of leaks there are various sensors/ground microphones available. A sensor rod microphone is used to listen to sounds from directly accessible pipe parts, such as hydrants or valves. Each sensor is used as follows:

when the
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
2.6.1 PAM W-1
The PAM W-1 is an active piezo ground microphone for roads and hard ground coverings. The microphone is particularly well shielded from the wind, using a bell-shaped guard. The carrying pole can be removed with a quarter turn to the left. This is particularly useful in reducing extraneous noise to a minimum when there is a strong wind.
For this ground microphone a VK 65 connection cable is required.
Important note: Piezo-electric microphones should not be subject to shocks so place them gently in position!
8
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
2.6.2 PAM U
PAM U was developed primarily as a sensor rod or contact microphone, but can be used universally by exchanging the contact point for various adapters.
Sensor rod variant:
In this variant the PAM U is suitable for listening to valves, hydrants or even directly on a pipe. It is particularly good when used as a ground microphone on soft ground (soil, meadow etc.). Push the point as deep as possible into the ground to get optimum acoustic contact to the source of the noise. When using an extension rod VST 1 the working posture is more comfortable but, due to the extension, wind and surrounding sources of noise are more intrusive. Take care: When pulling the microphone out, always pull the body of the microphone and not the cable, otherwise you may damage the cable.
9
Fig 4 : PAM U with sensor rod
Magnetic variant:
When you hold the microphone, even the smallest movement creates loud noises that can interfere with the measurement. It is thus ideal if you can let go of the microphone during a measurement. The magnet screwed in at the front is suited to that. Due to the high holding force of the magnet an excellent acoustic coupling is achieved. This is however only possible when in contact with ferromagnetic materials, not with plastic, some stainless steels, etc.
Fig 5 : PAM U with magnet
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
Take care: When removing the microphone from a valve, always pull the body of the microphone and not the cable, otherwise you may damage the cable.
3-point foot variant: By screwing in a 3-point foot, the PAM U can also be used as a ground microphone on a hard surface. Surrounding noise, particularly wind, is not so well screened off as with the ground microphone PAM W-1, but nevertheless very good results can be achieved with this variant.
10
Fig 6 : PAM U with 3-point foot
2.7 Headphones
The KR 2 headphones with ambient sound insulation are supplied as standard. These electrodynamic headphones reproduce leak sounds well. Other models of headphone should be avoided if at all possible, as hearing protection to VBG 121 can not be guaranteed. According to para. 10 VBG 121 the headphone volume may not exceed 85 dB.
The KR 2 headphones with particularly effective ambient sound insulation are available as an accessory and can also be used on the HLE 5000/500.
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
3 Practical work
Now that you have got to know the HLE 5000/500 leak detector, you can start with the practical use.
3.1 Headphones and microphone connection
Before switching on the HLE 5000/500, connect both the headphones and the microphone. The reverse is true when switching off: Always switch the equipment off first and then remove the microphone and headphones.
11
3.2 Switching on
By pushing the on/off button turned on. The start image then appears, together with the current version number and the battery status.
Fig 7 : Start image with version number
After a few seconds, the opening menu appears with the last equipment settings.


the equipment will be
Battery status
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
0
headphone volume
setting
100 200
150
50
Setting the
0 – 4000 Hz
12
Fig 8 : Opening menu after switch-on
3.3 Volume and amplification
The HLE 5000/500 is equipped with separate controls for volume and amplification. Thus, every user can make the best settings for their personal characteristics.
3.3.1 Setting the volume
Using the two buttons on the right (see fig. 9) the headphone volume can be set. To start a measurement, medium volume
- 3 scale units - should be selected.
Amplification
Fig 9 : Setting the volume
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
3.3.2 Setting the gain
The amplification of the microphone signal, the gain, is set via the two soft keys, bottom right (see fig. 10). The gain is shown via a horizontal bar. On top of this bar the gain values of 1 to 8 can be seen.
13
Fig 10 : Setting the gain
To start a measurement, medium gain e.g. 3-4 should be selected. Changing the gain affects the level bar and the volume in the headphones.
3.4 The amplification display
The bar display (fig. 11) shows both the current value of the sound picked up and the amplified sound.
The lower, thicker bar shows the minimum value of the measurement. When considering the nature of a leak sound, which is a continuous noise, the display of this value provides a much better result and is much less susceptible to pulses of interference. This minimum value is recalculated after the mute button has been pushed.
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
0
1800
4000
0 –
70
Curre
nt value
Minimum value
cut-off
cut-off
14
100 200
150
50
Fig 11 : Current and minimum value
0 – 4000 Hz
3.5 Filter setting (only HLE5000)
Filter settings on the HLE 5000 are very easy. Use the soft key button in the main menu, to get in to the filter settings. As you can see in fig. 12, in the centre of the display, there are 9 vertical level bars. Underneath, there is a horizontal bar, which identifies the selected filter range. The lower and upper cut-off is shown to the left and right of the bars respectively.
Lower
0 - 70Hz
106Hz
240Hz
160Hz
540Hz
360Hz
1200Hz
800Hz
1800 ­4000Hz
Upper
Fig 12 : Filter settings
The following 9 cut-off frequencies can be set: 0 - 70, 106, 160, 240, 360, 540, 800, 1200, 1800 -4000 Hz
3.5.1 Adjusting the cut-off frequencies
To adjust the lower cut-off frequency, use the two soft key buttons underneath , as you can see in fig. 13.
MAN_HLE500_HLE5000_eng_02
Filter range
Operating instructions HLE 500/5000
1800
HYDROLUX
0 –
®
HYDROLUX
Fig 13 : Adjusting the lower filter cut-off frequency
The upper cut-off frequency is adjusted in a similar way, using the two soft key buttons to the right, underneath
(see fig. 14).
4000 Hz
®
15
Hz
70
Fig 14 : Adjusting the upper filter cut-off frequency
For the two cut-off frequencies, you can set the 9 frequencies specified above and can thus easily set the special filter range for every measurement. After the filter range has been correctly set, use the soft key button “ESC” to return to the main menu.
3.6 Filter selection in practice
The decision about the filter settings has to be made by the user. In general, one can say that, for measurements with ground microphones, lower frequency ranges should be chosen. For measurements directly on the pipe or valve using the sensor rod, higher frequency ranges should be
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
chosen. Should a measurement not be successful, then a broadband setting is to be recommended, to ensure that none of the leak sound is filtered away.
3.6.1 Filter selection A (Ground microphone PAM W-1, PAM U with
3-point foot) As a standard setting, a filter from 240 Hz - 540 Hz should be
used. Particularly with plastic pipes and pipes at low operating pressures, lower frequencies can occur.
3.6.2 Filter selection B: Sensor rod (Pam U with spike)
For measurements directly on the pipe with the sensor rod, the frequency range should be set to 540 Hz - 1200 Hz.
16
X
3.7 Mute button
To move the ground microphone, you should first of all push the mute button
the headphones to be interrupted and the current level will be “frozen”. This means that your hearing is protected as the loud contact noises are suppressed. After you have chosen a new location for the ground microphone, the mute button
X
should be pushed again. This switches the headphones back on, the level display is updated and the minimum is recalculated.
3.8 The histogram
The “Histogram” function is used to be able to compare a series of sequentially recorded measurements. This can be used when pre-locating a burst pipe with a sensor rod microphone as well as when pinpointing the actual leak. For the HLE 5000/500 a maximum of 9 measurements are displayed.
X
(see fig. 1). This causes the sound in
After pressing the soft key button (see fig. 1) the LCD display switches to the histogram view and the first histogram is activated.
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
The individual histogram displays consist of a dual segment analysis (DSA), which show the current value and the minimum value. This DSA is shown in fig. 15. The narrow segment stands for the current value and the wide segment shows the minimum value, which is of particular significance when looking for burst pipes.
17
Fig 15 : Dual segment analysis (DSA)
X
With the first push of the mute button measurement is stored and simultaneously displayed. In this way, nine values can be stored, one after the other. When more than nine values are measured, the first value will be deleted and all other values will be shifted one position to the left. The last nine values are thus always available.
Fig 16 : Histogram measurement
the current
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
The histogram measurement as shown in fig. 16 shows very different current values. On the other hand the 5th DSA shows a clear maximum for the minimum values. The burst pipe is thus in the vicinity of the 5th measurement location. Additionally a small graphical symbol indicates the maximum measurement for easy recognition. To leave the histogram measurement, press the “ESC” soft key.
3.9 Long-term measurement (HLE 5000 only)
This function is intended to record sounds over a programmable period and then display as a graph. With this method the identity of a water pipe can be established by recording the flow noise of a valve. To do that, the microphone is put onto the pipe and the long-term measurement is started. Then, you close the valve for a certain time (at least 2 minutes) and then open it again. If the pipe at the listening location is identical to the shut-off pipe, then this should be visible on the sound level curve. In fig. 17 a sound level curve like this is shown.
18
Fig 17 : Long-term measurement
To set this function, from fig. 1 the soft key is pushed. The menu for the long-term measurement will appear as in
fig. 17. To specify the duration press the soft key (time) repeatedly until the required measurement time is set, as shown top right in the display. Recording durations of 3 - 10 ­30 minutes are available. After setting the duration, push the
(start) soft key. The measurement will start and could be
stopped early by pressing the (stop) soft key.
MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
3.10 Location of pipes using the RSP-3 or PWG (HLE 5000 only)
To switch the HLE 5000 into this mode, after switching on, soft key 2 (underneath the impulse symbol ) should be pressed whilst the welcome image is displayed. The HLE 5000 is now in pipe location mode, shown by the impulse symbol at the top edge of the display. This mode of operation is particularly suited to display impulse noise, such as the tapping of the pipe pecker or the PWG. The horizontal level bars respond particularly sensitively to impulses, the bar is shown much larger and is delayed more and the filters are set to lower frequencies Of course, if required, the filter frequencies can be altered as usual. The HLE5000 stays in pipe location mode until it is switched off. After being turned on again, it will be in normal mode.
19
3.11 Switching off
The equipment is switched off by a longer press on the button. Any measurements stored will be lost. After 35 minutes in operation, the equipment switches off automatically.
In histogram mode the equipment will also switch off automatically after 35 minutes, as long as the mute and the illumination button are not used.


MAN_HLE500_HLE5000_eng_02
Operating instructions HLE 500/5000
4 Faults
4.1 Can't switch on
Presumably the headphones are not plugged in or headphones are being used which do not belong to the system.
4.2 Battery monitor does not react
One or more batteries are incorrectly fitted. Open base flap and check the polarity of the batteries. See fig. 3. If the polarity of all of the batteries is OK, the state of every single battery must be checked.
20
4.3 No sound can be heard
With working equipment there are two possibilities: a. Headphones not fitted, or fitted incorrectly
X
b. The mute button
headphones are switched off. Press the mute button
X
again to restore the sound.
4.4 Scratching sounds in the headphones
This is usually caused by a poor contact. Check all contacts.
is active, which means that the
MAN_HLE500_HLE5000_eng_02
Loading...