Sartorius arium 611VF Installation And Operation Manual

Page 1
Installation and Operation Manual
arium 611VF
Water Purification System
85030-513-68
Page 2
2
About This Manual
arium Systems 5 Control Panel and Display 6
Initial Operation 13 Cartridge Installation 13 System Settings 14 Setting the Date and Time 14 Measurement Units 14 Set Point 15 Language 15 Initial Flush and Sanitization 16 Flushing and Purging Air from the Cartridge Packs 16 System Sanitization 16 Ultrafilter Flush 17 Reset UV Timer Prior to Initial Operation 17 Final Filter Installation 17
Further Options 18 Print Data 18 Pump Protection 18 Time|Volume Dispense 19 Set Cell Constant 20 Calibration of the System 21
Start Up| Operation 22 Normal Operation Mode 22 Dispensing Product Water 22 Standby Mode 22 System Inactive Mode 23
Maintenance and Servicing 24 Sanitization 24 Cartridge Replacement 25 Ultrafilter Replacement 26 Replacing the Final Filter 26 Cell Cleaning 27 UV Bulb Replacement 28 Activate the UV Timer 29 Fuse Replacement 30
Appendix 31 Troubleshooting Guide 31 Specifications 32 Accessories and Replacement Parts 33
This manual instructs you on how to prepare and operate the arium 611VF water purification system.
Sartorius has designed the arium 611 water purification system for reliability, economy and safe operation. To ensure this, you must read this instruction manual carefully before attempting to operate the system. The manual will give you important instructions that will help you avoid potential hazards and ensure the reliability of the arium 611VF.
Please read the Safety Information on page 3.
If you have any questions about the correct use of arium 611VF, please contact us at the address below or your local Sartorius office:
Sartorius AG Life Science Business Unit - Lab Division Weender Landstrasse 94–108 37075 Goettingen Phone +49.551.308.0 Fax +49.5 51.308.3289
Table of Contents
Page 3
Please read the following safety information thoroughly and follow the instructions exactly. This information is designed to ensure your own safety and will prevent damaging the arium 611VF unit.
The following symbols are used in this manual:
Warning!
Warnings alert you to a possibility of personal injury or property damage.
Caution!
Caution signs alert you to a possibility of damage to the equipment.
Note
Notes alert you to pertinent facts and conditions that are important for economic operation of the equipment.
Warning!
Danger of personal injury! Servicing and repairs may only be performed by trained and qualified personnel.
Warning!
Severe electrical shock hazard or danger of electrocution!
– Use a properly grounded electrical
outlet of correct voltage and current handling capacity (100 to 240 V ~, 50/60 Hz) to plug in the arium 611VF unit.
– Do not locate arium 611VF on top
of electrical equipment. Routine maintenance of this unit may involve water spillage and subse­quent electrical shock hazard if improperly located.
– Remove the plug from the electrical
outlet prior to maintenance and servicing of arium 611VF.
3
Safety Information
Warning!
Danger of fire or explosion!
– Do not use in the presence of
flammable or combustible materi­als; fire or explosion may result. This device contains components which may ignite such materials.
– The arium 611VF is to be used
with water feeds only. Sanitizing| cleaning agents should only be used according to the instructions in this manual.
Warning!
Danger of injury to eyes and skin!
– Avoid splashing sanitants or
disinfecting solutions on clothing or skin.
– Ensure all tubing connections are
tight to avoid chemical leakage.
– Turn off feed water and push
the draw-off lever to the right to depressurize system prior to changing cartridge packs, sanitizing or performing any service on the 611 system.
– Allow a defective UV bulb to cool
off before removing it.
– Carefully follow the manufacturer’s
safety instructions on labels of chemical containers and filter cartridges.
Caution!
Danger of irreversibly damaging arium 611VF components! Be sure to replace defective fuses with those of the same type and rating. When installing a new UV bulb, do not touch the bulb with your bare hands. Finger prints may damage the bulb.
!
!!
!!
!
!
!!
Page 4
The arium 611VF water purification system is designed exclusively to provide reagent-grade water for the laboratory from water pretreated by distillation, deionization or reverse osmosis. To ensure that this unit works properly, use only the filter media and other auxiliary media that are listed in this manual. Using this system for any other purpose shall be considered improper usage.
– arium 611VF may only be operated
by trained personnel.
– Operate arium 611VF only using
original accessories or replacement parts. If you modify this water purification system on your own without consulting Sartorius, the performance and operating safety of the system are no longer guaran­teed, thus constituting a safety hazard for the operator.
– Please take all pertinent precautions
to prevent accidents and observe the generally valid technical and occupational safety rules and regulations of your country.
– Use only Sartorius-listed materials
(such cartridges, connectors, gaskets, tools and sanitizing and disinfecting agents).
4
Intended Usage
Page 5
arium Systems
The Sartorius arium 611VF is a water purification system designed to provide pyrogen-free, low TOC reagent grade water that exceeds ASTM Type I and NCCLS Type I standards. The system uses a deionization process with mixed bed resins and activated carbon combined with UV oxidation, ultra­filtration and a 0.2-micron final filter to produce laboratory reagent grade water. The 611VF system is capable of producing pyrogen-free (< 0.001 EU/ml), low TOC (< 1 ppb) water with a resistivity of up to
18.2 MegOhms + cm. Water resistivity is continuously sensed by a resistivity cell and displayed on a digital display.
Item No. Description
1 Left and right door of
arium 611VF
2 Product water outlet with
0.2-micron filter and bell
assembly 3 Display and control panel 4 Recessed grip for opening
the right door 5 Draw-off lever for
opening and closing
the outlet valve
(shown in open position) 6 Outlet for a remote dis-
play|dispenser unit, TOC
instrument or a dispenser 7 15-pin D-sub port for
connecting a remote
display|dispenser unit 8 Outlet for time|volume
dispense 9 Reject water outlet 10 Fuse drawer 11 Main power switch 12 Power cord receptacle 13 9-pin D-sub port for
connecting a printer 14 Connector for pump
interlock that prevents the
pump from running dry 15 Feed water inlet
5
Product Description
Front view
Side view, left
3
2
1
10
11
12
13
14
15
4
5
6
9
8
7
Page 6
Control Panel and Display
Operate arium 611VF using the control panel that incorporates four function keys and two control keys for the cursor.
STANDBY Key
Press the Standby key to place the unit into Standby Mode. In this mode the water in the system is recirculated for 15 minutes per hour.
OPERATE|STOP Key
Press the OPERATE|STOP key to activate or deactivate the unit. During Normal Operation the system will recirculate. When the system is stopped the pump turns off and water will not recirculate.
MENU Key
This key provides access to system options and menus.
Cursor keys
These two cursor control keys are used to select individual menu items of the different menus. The Y key moves the cursor one line up, the y key one line down.
ENTER Key
Use this key to select a menu item you have highlighted with the cursor.
Display
– During Normal Operation the
display will show the water quality reading – according to the pre-selected measurement unit options: (MO+cm or µS/cm) The letter “C” indicates that the water quality measurement is automatically compensated to 25°C.
– When the system is in Standby
Mode the display will show “Standby”. The UV bulb will turn on periodically to make certain the TOC is kept at an acceptable level. An automatic flush will take place every 24 hours. When the system is flushing the display shows the following:
for 15 minutes every hour.
– The display will show the following
when the system is stopped:
System Inactive
Standby
Auto UF Flush
18.2 MO * c m c
14.32 20.02.01
6
Page 7
Control panel and display of arium 611VF
Flow chart of arium 611VF
7
Page 8
Unpacking
Remove arium 611VF from its packaging. You will find the accessories inside the arium housing and in the top cover of the packing. The equipment supplied includes the following parts:
Display|Dispenser Unit Installation
The Display | Dispenser unit has been designed to mount in 3 ways: At the top of the right door for bench systems, at the bottom of the right door for wall mount systems or in a remote location from the system (additional kit required).
To install:
• If necessary, remove the rubber cover mounted at the bottom of the right door (for wall mounting).
• Route the supply cables through the large opening of the right door (top or bottom) and the two water tubes from the outside through the small opening.
• Use the 4 screws and washers provided to mount the Display| Dispenser Unit to the door by routing the screws from inside the door through the holes in the door to the outside and screw to the assembly.
• Route the water tubing through the 4 holes inside the door and connect to the color coded connectors. Follow the course of dotted lines in the figure: Positions (1) and (3): data and power cables Positions (2) and (4): water tubes
• Put rubber cover into the remaining hole (top or bottom) in the door.
8
Unpacking and Installation
Part Description Number arium 611VF 1 Dispenser unit with display 1 Screws and washers for the Display | Dispenser unit 4 Blue cover with manufacturer's label (VF) 1 Blue cover with model name of arium (for dispenser unit) 1 Ultrafilter cartridge 1 Reject water tubing 1 Tubing adapter for inlet water (1”, English and 4”, NPT) 2 Tube for inlet water with quick disconnect insert 1 Adapter with O-ring for cartridge pack 1 End cap for the sanitization syringe adapter 1 Sanitization syringe 1 Wall mounting bracket 1 Power cord 1 Installation and operation manual 1 Timed Dispense Tube, PE 4 inch OD X 8 feet 1 Tubing adapter, 4 inch OD tube to 4 NPT 1
Note
The cartridge packs are supplied separately. They are not included in the arium system packaging.
Page 9
Mounting the upper display | dispenser unit
Mounting the lower display | dispenser unit
9
1
2
4
3
Page 10
Ultrafilter Installation
Caution!
Danger of irreversibly damaging arium 611VF components and the filter cartridges! Tighten all union nuts made of plastic by hand. Do not use a wrench or pliers.
Before starting up arium 611VF, you need to install the ultrafilter module in the housing cover. Proceed as follows:
• Make sure to unplug the power cord from the electrical outlet (“O”).
• Open both doors of the unit.
• Remove ultrafilter from packaging and remove protective covers from fittings.
• Connect the tubing connectors to the ultrafilter module as shown in the Figure: (1): green connector for feed water - Bottom (2): blue connector for product water - Upper left (3): red connector for reject water - Top
• Push in the ultrafilter module all the way into the holders so that the connector fitting on the side is facing left.
• Close both doors.
10
!!
Attach the three tubing connectors to the ultrafilter module
1
2
3
Page 11
Bench Mounting
Warning!
Severe electrical shock hazard or danger of electrocution! Do not place arium 611VF on top of electrical equipment. Water may spill when using the system.
Warning!
Danger of fire or explosion! Do not use in the presence of flammable or combustible materi­als; fire or explosion may result. The device contains components which may ignite such materials.
• Place the arium 611VF on a flat surface.
• Ensure that feed water and a 100-240 V electric socket are accessible.
Note
The outlet of a gravity feed storage reservoir must be at least 10 cm higher than the inlet of the arium 611VF.
11
Wall Mounting
Caution!
Wall composition, condition and construction, as well as fastener type must be considered when mounting this unit. The mounting surface and fasteners selected must be capable of supporting a minimum of 25 kg. Inadequate support and | or fasteners may result in injury to the operator and|or damage to the equipment.
A wall bracket (2) is included in the shipment to enable you to attach the system securely to the wall so that it occupies a minimum amount of space. A clear wall area of 63 + 63 cm is required to mount the arium system. Two sturdy keyhole drill holes (1) are located at the top of the back of the housing enabling you to hang the unit on the two pins (5) on the wall bracket. Use the predrilled holes in the bracket to fasten the bracket to the wall. The lower metal profile (6) on the arium housing functions as a spacer.
• Ensure that water, a 100 - 240 V electric socket and an atmospheri­cally vented drain are accessible.
• Fasten the bracket to the wall with suitable screws and dowels.
• Lower the keyhole drill holes of the arium 611VF onto the two pins on the wall bracket.
• Tighten the screw (3) of the metal support (4) on the right side of the bracket with a screwdriver until the metal support rests against the arium housing.
!
!
!
1
2
3
4
5
6
Lower the arium housing onto the wall-mounting bracket
Page 12
12
Water Inlet Connection
Feed water is supplied to the system through this inlet (1). The arium system is supplied with two water supply adapters – one with 1” British thread and one with 4”-NPT.
Caution!
Do not draw water from the top of an open container or carboy to supply water to the 611 system. Damage to the system may result.
Note
We recommend that a customer supplied shutoff valve be installed in your feed water line. If a gravity tank is used to supply feed water to the 611 system, the feed water tube must be connected to the bottom of the tank with a ridged or hard plumbed connection. The connection at the bottom of the gravity must be level with, or preferably, slightly above the inlet water connection of the 611 system.
• Connect the free end of the tube to the appropriate water supply adapter. Push the tube all the way into the adapter to seat it properly (approx. 20 mm).
• Install the tubing adapter onto your feed water source.
• Cartridges must be installed prior to supplying feed water to the system. See cartridge installation instructions.
• Insert the other end with a quick disconnect insert into the system’s counter coupling (1) until it audibly clicks into place.
Reject Water Connection
The reject water is discarded through this connection (2). The reject water tubing has a 4-inch O.D. Push the tube all the way (approx. 20 mm) into the reject water connector (2)) to seat it properly. Direct the other end of the reject tubing to an atmospheric drain.
Note
Ensure there are no kinks in the tubing and that it proceeds in a downward plane.
!!
Connector for feed water tubing
Connector for reject water tubing
1
2
Page 13
Cartridge Installation
• Open both doors of the system.
• Remove both cartridge packs from their packaging.
One cartridge pack has a label with a blue dot and the number 1, the other cartridge has a red dot and the number 2. In the housing to the lower left side you will find a blue labeled cartridge adaptor, on the lower right side a red cartridge adaptor.
• Remove the red plugs from the top of the cartridge pack and push the blue adapter onto the cartridge pack with the blue dot. Press vigorously. Be sure the label with the blue dot faces you. The adapter displays “front”.
Two screws each are seated in the cartridge cover. The two screw heads must rise from the drillings of the safe guard slides with the upper ring of the spacers.
• Push both safe guard slides all the way under the upper ring of the spacers, as shown in the figure opposite.
• Place the cartridge pack with the blue dot left into the housing and push until properly seated. Be sure the label faces you.
• Proceed with the cartridge pack with the red dot in the same manner on the right side of the unit and connect the red coupler.
• Close both doors.
Note
– Moisten the O-rings of the
cartridge pack adapter with water to facilitate connection.
– Before initial operation the
cartridge packs have to be flushed, see Cartridge Flush section.
– Screw the final filter onto the NPT
fitting of the dispensing valve only after flush cycle and sanitization of cartridge packs and ultrafilter.
13
Initial Operation
Press the adapter onto the cartridge (e.g. red cartridge pack)
Secure the adapter by pushing the two safe guard slides forward over the security screws
Page 14
14
System Settings
Prior to initial operation of the arium 611VF, set date and time, the language for the display text and the desired measurement units. To set these functions:
Setting the Date and Time
• The main power switch on the left hand side of the housing must be on (“I”).
The arium 611VF will then perform the system check. The display will show the following information:
The display will show the following when the system check is complete
• Press Menu to get in to the Main Menu
• Select “Setup”. Press ENTER to confirm.
• Select “Time & Date”. Press ENTER to confirm.
Display will show the following:
TIME/DATE Standard
U.S.
SETUP Time & Date
Set Point Meas. Units
MENU Utilities
Maintenance Setup
MENUOMain Menu OPEROOperation ENTRONew Cart. Flush
Sartorius System VF
S/N XXXXXXXX
System Check OK
V1.XX
– When “Standard” is selected,
the date will be displayed in the DD.MM.YY format and time in the 24-hour format.
– When “U.S.” is selected, the date
will be displayed in the MM.DD.YY format and the time in the 12-hour format.
• Choose a format and press ENTER to confirm.
• Select Time & Date and use the arrow keys Y or y to change current Time & Date settings. Press ENTER after each digit is set to move to the next digit.
• Press ENTER to confirm the Time & Date setting.
The system will then switch to the previous Operation Mode.
Measurement Units
This feature allows the user to select how measurement units are displayed. The user can choose between MO+cm, µS/cm, with the option to display readings compensated to 25°C or uncom­pensated, or temperature. To set the measurement units:
• Press MENU to get into the Main Menu:
• Select “Setup”. Press ENTER to confirm.
• Select “Measurement Units”. Press ENTER to confirm. The display will show the following:
Use the arrow keys Y or y to select the desired option and press ENTER to confirm.
Note
In some menus not all menu items are visible on the display. Repeatedly press the arrow key y to display menu items listed further below.
The system will then switch to the previous Operation Mode.
MEAS. MO * cm comp.
µS/cm comp. M
O * cm uncomp.
SETUP Time & Date
Set Point Meas. Units
MENU Utilities
Maintenance Setup
Page 15
15
Language
This feature allows the user to select one of six languages for the display texts for convenient operation. The options are German, English, French, Spanish, Italian and Japanese. To select a language:
• Press MENU to get into the Main Menu:
• Select “Setup” and press ENTER to confirm.
• Use the arrow key y to scroll down, select “Language” and press ENTER to confirm. The display will show the following:
• Use the arrow keys Y or y to select “Language” and press ENTER to confirm.
The system will return to the previous Operation Mode.
LANG. English
French German
SETUP Time & Date
Set Point Meas. Units
MENU Utilities
Maintenance Setup
Set Point
The option “Set Point” allows the user to select a defined quality level for the product water in MO+cm or µS/cm. When the water quality falls below this Set Point, cartridge packs should be replaced. The factory setting for the Set Point is 10.5 MO+cm. To select the Set point:
• Press MENU to get into the Main Menu:
• Select “Setup” and press ENTER to confirm.
• Select “Set Point” and press ENTER to confirm. The display will show the following:
• Select current Set Point or use the arrow keys Y or y to change to the desired Set Point. Press ENTER after each digit is set to move to the next digit.
• Press ENTER to confirm the Set Point.
The system will switch to the previous Operation Mode.
When the water quality drops below the defined Set Point, in the Normal Operation Mode the following display is shown blinking with the current measurement unit:
Check Set Point
14:32 20.03.01
Set Point
10.5 M
O * cm
SETUP Time & Date
Set Point Meas. Units
MENUOMain Menu OPEROOperation ENTRONew Cart. Flush
Page 16
16
Initial Flush and Sanitization
Prior to initial operation, cartridge packs and the ultrafilter must be flushed to remove air and rinse wetted parts.
Flushing and Purging Air from the Cartridge Packs
Note
The final filter must not be fitted onto the dispenser fittings at this time.
Proceed as follows:
• Press MENU to get into the Main Menu:
• Select “Maintenance” and press ENTER to confirm.
• Select “New Cartr. Flush” and press ENTER to confirm. The display will show the following:
• Press ENTER.
The New Cartridge Flush will begin and the display will indicate the remaining flush time:
When this time has elapsed the following display is shown with the word “Finished” blinking:
• Open the dispensing valve and discard water for approx. 3 minutes.
The New Cartridge Flush is complete, and the final filter can now be installed. Refer to filter installation instructions on page 17.
AIRPURGE/FLUSH CYCLE Finished Push OPERO Operation
AIRPURGE/FLUSH CYCLE
19 min
AIRPURGE/FLUSH CYCLE
Press ENTER to Start
MAINTENANCE
UV Life Timer Sanitization
MENU Utilities
Maintenance Setup
System Sanitization
To sanitize the system:
• Press MENU to get into the Main Menu:
• Select “Maintenance” and press ENTER to confirm.
• Select “Sanitization” and press ENTER to confirm.
• Follow the instructions on the next few screens and press ENTER after each is complete.
Note
Make sure system is depressurized before attempting to inject sanitization liquid.
Inject the sanitization liquid with the syringe as follows (see the figure below):
Inject the sanitization liquid at the red adapter fitting
Open Product Outlet Collect Outlet Water Inject Sanit. Liquid Push ENTER
Disconnect Feed Line Remove End Filter
Push ENTER
MAINTENANCE
UV Life Timer Sanitization
MENU Utilities
Maintenance Setup
• Unscrew the luer cap from the connection of the red cartridge adapter.
• Inject the sanitization liquid through the connection into the system and then slowly remove the syringe. Some solution may exit.
• Reattach luer end cap.
After pressing ENTER the pump will run for some seconds. When the pump stops, the display will show the following:
After pressing ENTER, the 30-minute active sanitization time will begin. Time will start to count down on the display. When the 30 minutes expires, the system will flush for 6 minutes automatically. When the time expires, the Saniti­zation Purity Circulation will begin automatically and time will count (60 minutes). The system will enter Standby Mode automatically when time expires.
• To complete the Sanitization Procedure, press the OPERATE| STOP button and flush 0.5 liters of water to drain through the dispensing valve.
The final filter can now be installed.
Close Product Outlet Reconnect Feed Line
Push ENTER
Page 17
17
Ultrafilter Flush
Prior to initial operation, the Ultrafilter module has to be flushed. The system offers a short (1 min) and a long UF flush (5 min).
Proceed as follows:
• Press MENU to get into the Main Menu:
• Select “Maintenance” and press ENTER to confirm.
• Select “UF Flush” and press ENTER to confirm.
Two flush options are shown. The Short Flush runs 1 minute, the Long Flush runs 5 minutes.
• Select “Long” and press ENTER to confirm. Display will show the type of flush selected:
• The flush cycle begins – on the display, time will begin to count down. When the time has elapsed, the system will return to its previ­ous Operation Mode. If necessary press OPERATE | STOP to get into the Operation Mode.
• Then open the dispensing valve and flush 5 liters of water to drain.
• Close the dispensing valve.
LONG FLUSH 5 min
Push ENTER to Start
UF-FLUSH
Short Long
MAINTENANCE
UV Life Timer Sanitization
MENU Utilities
Maintenance Setup
Reset UV Timer Prior to Initial Operation
Prior to initial operation the UV Timer must be reset. See the Activate the UV Bulb Timer section on page 28.
Final Filter Installation
• Unpack a new final filter.
• Make sure the dispensing valve is closed.
• Apply two to three wraps of Teflon tape to threaded NPT fitting of filter.
• Screw filter clockwise into valve NPT fitting until snug.
• Remove the protective cap from the bell assembly and flush 6 liters of water to drain through final filter.
After this step, the system is now ready for operation.
Note
Flush an additional 50 ml water to drain prior to collecting water for analysis.
Page 18
18
Further Options
Print Data
If you have connected an appropri­ate printer to the printer port (1) on your arium system, you have two print options:
– Print Interval Option
With the Print Interval Option, the system can be set up to print data at regular intervals when in Normal Operation.
– Print Screen Option
The Print Screen Option will only print the last data recorded.
Note
The arium system has a 9-pin connection for a serial printer. The printer must use a separate power supply and be configured as fol­lows: 19,200 baud, 8-bit, no parity.
To select a print option:
• Press MENU to get into the Main Menu:
• Select “Utilities” and press ENTER to confirm.
• Select “Print” and press ENTER to confirm.
• Select the desired option and press ENTER to confirm.
– With the Print Interval Option, the
system can be set up to print data at regular intervals. If you have selected “Print Interval”, you can use the control keys for the cursor to set the desired length of interval in minutes when prompted by the display.
– “Print Screen”
The Print Screen Option will only print the last data recorded. You can generate a printout of the screen by pressing ENTER once. After printing, the system returns to the previously activated Operate Mode.
PRINT
Print Interval Print Screen
UTILITIES
Print Timer Contr.
MENU Utilities
Maintenance Setup
Pump Protection
If you are using a separate tank for the feed water supply, you can connect the water tank and the arium unit via port (2) with a pump protection for tank feed systems. When the tank is empty, the pump is turned off and water will not recirculate (only with arium tank).
Printer Port (1) and pump interlock (2)
2
1
Page 19
Time|Volume Dispense
Note
The arium 611 System must be equipped with a Time | Volume Dispense port to function. The Time|Volume Dispense port is located on the right side of the system (1).
Warning!
Be sure a 4” outlet tube is connected to the Time | Volume Dispense port located on the right side of the system. When time dispense is activated, water will exit the system via this port. The end of the outlet tube should be placed in a filling device.
Note
In order to use the Time | Volume Dispense feature, you must first measure the flow rate of the water from the timed dispense outlet. Install the supplied 4” tube to the dispense outlet port. Follow the menu instructions below to operate the timed dispense control. When asked to enter a time, use the up| down arrows and enter a 1-minute. Place the open end of the tubing in a graduated container of at least 2-liter volume. Activate the timer and allow water to collect in the container. When the 1-minute timed dispense has stopped, measure the volume of water col­lected. Use the volume collected in 1-minute (flow rate) to calculate your dispense time requirement. For example, you have measured a 1-liter volume in 1-minute. Your measured flow rate is 1l/min. If you need 10 liters to fill a container, 10 l/1l per minute = 10 minutes dispense time.
Note
If you are using a filter with the Timed Dispense feature, attach the filter and rinse the filter for 5 minutes to drain before measur­ing the flow rate. Measure the flow rate each time you change filters or cartridges.
• Press MENU.
• Use Y or y to select “Utilities”. Press ENTER to confirm.
• Use Y or y to select “Timer Contr.”. Press ENTER to confirm.
• Use Y or y to change the time. Press ENTER after each digit is set to move to the next digit. When the time has been entered the following screen is displayed.
• Be sure an outlet tube is connected to the time dispense port on the right side of the system. To begin time dispense press ENTER.
TIMER CONTROL
To Start Filling
Push ENTER
TIMER CONTROL
10.0 min
UTILITIES
Print Timer Contr.
MENU Utilities
Maintenance Setup
• The system will begin to dispense water through the time dispense port. Time will count down as shown on the screen below.
• When time expires, the time dispense port will close and water will continue to recirculate through the system. Filling is now finished. The screen below will show on the display.
Filling Finished
18.2 MO * c m c
11.25 10.01.02
10 min
18.2 MO * c m c
11.15 10.01.02
19
!
1
Page 20
Set Cell Constant
The resistivity and temperature cell constants are factory set! If you have to replace the cell, you will find the resistivity and temperature cell constants listed on the cell cable (1) located inside the arium system. To provide a more accurate reading, the cell constants should be entered in the system memory by the user. Follow the steps below to enter the cell constants:
• Power off the arium system (“O”) and disconnect it from the AC power outlet.
• Open the front doors and place the red cartridge pack on the side.
• Locate the white label on the gray cell cable (1) underneath the resistivity cell. Record the resistivity and temperature cell constants.
• Place the red cartridge back in the system, close the doors, connect the system to the AC outlet and power on the arium water system (“I”).
• Press MENU to get into the Main Menu:
• Use Y or y to select “Utilities”. Press ENTER to confirm.
• Use Y or y to select “Set Cell Constant”. Press ENTER to confirm.
• Use Y or y to enter the resis- tivity (Res) and temperature (Temp) cell constants previously recorded. Press ENTER after each digit is set to move to the next digit.
When the last digit is entered and you have pressed ENTER to confirm, the values will be recorded in the system memory and the system will enter Operate Mode last activated.
Note
If a digit is entered in error, press the MENU button to return to the Main Menu and begin the process again.
Cell Constants Res 0.1000 Temp +1.0000
UTILITIES
Print Timer Contr.
MENU Utilities
Maintenance Setup
20
Cell constants listed on the cell cable (1)
1
Page 21
Calibration of the System
The system can be calibrated using an optional N.I.S.T. calibration module available from Sartorius. Proceed as follows:
• Turn the power switch on the arium System to off (“O”) and disconnect power cord from the unit.
• Disconnect the inlet water line from the arium system. Open outlet valve to depressurize the system.
• Open the front doors and move the red cartridge pack #2 to the side by rotating the pack to the right.
• Pull up the yellow sealing hood from the gray resistivity cell cable.
• Disconnect the cell from the gray cable. Attach the calibration module where the cell cable was connected.
• Close the front doors, reattach the power cord and turn the power switch on (“I”) – wait for the unit to complete the system check.
Caution!
Danger of water spillage.
• Press MENU to get into the Main Menu.
• Use the cursor to select “Mainte­nance”. Press ENTER to confirm.
• Use the DOWN arrow y to select “Calibration”. Press ENTER to confirm.
• After pressing ENTER, the calibration will begin.
See Owners Manual for Details.
Push ENTER to Start
MAINTENANCE
UV Life Timer Sanitization
MENU Utilities
Maintenance Setup
21
This will take a few minutes to complete:
When calibration is successful, the display will show the following:
• Turn the power switch on the arium System to off (“O”) and remove the power cord.
• Open the front doors and unplug the calibration module from the cell.
• Reattach the gray cable to the cell.
• The yellow sealing hood should form a tight seal around the connection.
• Close the doors and reattach the power cord.
• Connect the inlet water line and power on the system (“I”).
The calibration is now complete. The unit is ready for normal operation.
Calibration
Finished
Calibration
In Process
!
Page 22
Note
If your laboratory uses several liters of water or more per hour, it is rec­ommended that the system be left in the Operate Mode to maintain the highest quality water. During times of low water consumption it is recommended that the system be put in the Standby Mode to conserve energy. To dispense water from the standby mode, press the operate button and allow the system to circulate for several minutes prior to drawing water. After drawing water the system may be put back in the Standby Mode. The system should be put in the Standby Mode overnight and for weekends and other off duty hours.
Normal Operation Mode
Once the new cartridge flush procedure and the sanitization procedure are completed, you can start Normal Operation with your arium 611VF. Turn on the power switch (“I”), connect the feed water source. When the internal system check is complete, the display will show the following options:
• At this point the system Main Menu can be accessed and Normal Operation started by pressing the OPERATE|STOP button.
MENUOMain Menu OPEROOperation ENTRONew Cart. Flush
Dispensing Product Water
In Normal Operation Mode the display will show the quality of the treated water according to the preselected measurement units:
When the required quality of the product water is displayed, you can dispense the water at the outlet by operating the draw-off lever:
• Place a water container under the outlet.
• Wait until the display shows the desired resistivity reading of the water.
• Remove the protective cap on the bell assembly of the final filter and
• Push the draw-off lever to the right to dispense the product water.
• After dispensing the water, pull the draw-off lever forward to close the valve.
• Place the protective cap back on the bell assembly.
Note
For all low detection level work, i.e. with a low content of TOC or endotoxins, flush an additional 50 to 100 ml water to drain prior to collecting water for analysis. (This corresponds to the volume of the supply to the final filter and the final filter capsule).
18.2 MO * c m c
14.32 20.02.01
Standby Mode
The “Standby Mode” is designed to conserve energy while maintaining water purity during periods of little or no purified water demand. At the end of the working day or before the weekend press the Standby button to engage the “Standby Mode”.
The display shows “Standby” when in Standby Mode.
When the system is in the Standby Mode, water will recirculate for 15 minutes per hour with the UV lamp on. The next morning or working day set the unit back into service by pressing the Operate| Stop button.
While in the “Standby Mode” an “Automatic UF-Flush Cycle” will occur once in every 24 hours. During the “Automatic UF-Flush Cycle”, UF concentrate water is rap­idly flushed to drain for 30 seconds. The first Automatic UF Flush Cycle will occur approximately 3 hours after the system is placed in the “Standby Mode”.
The display shows “Standby Auto UF-Flush Cycle” during the UF flush cycle.
Note
To conserve energy, the Standby Mode is recommended during non-working hours or when water consumption is less than 2 liters per hour. If water consumption is greater than 2 liters per hour, the system should be placed in the Normal Operation Mode.
Standby
Auto UF-Flush Cycle
Standby
22
Start Up|Operation
Page 23
23
System Inactive Mode
If the Operate|Stop button is pressed while the system is on Nor­mal Operation the arium 611VF is deactivated. This mode is intended purely for troubleshooting and should not be used for prolonged periods of time. Otherwise the entire system will have to be flushed and sanitized or cartridge packs replaced. When the system is inactive, the display shows:
If you need to shut down arium 611VF for an extended period of time, you should completely drain it, and remove all cartridge packs and the ultrafilter to prevent the growth of bacteria. If the system has remained inactive and full of water for more than 96 hours, you must drain the unit, install new cartridges and sanitize the unit prior to use.
System Inactive
Open product outlet valve
Push the draw-off lever forwards to close the product outlet valve.
Page 24
Warning!
Severe electrical shock hazard or danger of electrocution!
– Only qualified personnel may
service and repair arium 611VF.
– Disconnect arium 611VF from
power supply outlet before maintenance and servicing.
– Avoid splashing disinfecting
solutions on clothing or skin.
– Ensure all tubing connections are
tight to avoid chemical leakage.
– Turn off feed water and depres-
surize the unit by pushing the draw-off lever to the right. Only then should the housing doors be opened.
– Allow the UV bulb to cool off
before removing it. Be sure to wear protective gloves to remove the lamp to avoid leaving finger prints on bulb or sleeve.
– Carefully follow manufacturer’s
safety instructions on labels of chemical containers and material safety data sheets.
Sanitization
Sanitization is necessary to control bacteria growth and minimize or prevent the build-up of a bio-film within the cartridges or piping. Without Sanitization, the TOC level in the water can increase significantly.
The frequency with which you will need to clean your unit and replace the cartridges packs depends on your feed water’s characteristics, your purity requirements and your usage. Sanitize your arium 611VF and replace the cartridges when the product water purity drops below acceptable purity levels.
To sanitize the arium 611VF:
• Press MENU to get into the Main Menu:
• Select “Maintenance” and press ENTER to confirm.
• Select “Sanitization” and press ENTER to confirm.
• Follow the instructions on the next few screens and press ENTER after each is complete.
Open Product Outlet Collect Outlet Water Inject Sanit. Liquid Push ENTER
Disconnect Feed Line Remove End Filter
Push ENTER
MAINTENANCE
UV Life Timer Sanitization
MENU Utilities
Maintenance Setup
The Sanitization Liquid is injected using the syringe into the connec­tor ports of the adapter on the red cartridge adapter (see figure on page 16):
• Unscrew luer end cap from the injection port of the red adapter
• Inject the sanitization liquid through the port and then slowly remove the syringe.
• Screw luer end cap back on and close doors.
The 30-minute active sanitization time will begin. Time will start to count down on the display. When the 30 minutes expires, the system will flush for 6 minutes automati­cally. When the time expires, the Sanitization Purity Circulation will begin automatically and time will count (60 minutes). The system will enter Standby Mode automatically when time expires.
• To complete the Sanitization Procedure, press the OPERATE| STOP button and flush 0.5 liters of water to drain through the dispensing valve.
• Close the dispensing value and install a new final filter on the dispensing valve.
Close Product Outlet Reconnect Feed Line
Push ENTER
24
Maintenance and Servicing
!
Page 25
Cartridge Replacement
Cartridge pack life is directly related to feed water quality and water consumption. Cartridge packs need to be replaced when product water quality drops below the require­ments established by the user.
To replace the cartridges:
• Turn unit off using the operate| stop button and the main power switch on the rear panel (“O”) and remove the power plug.
• Make sure the inlet water is discon­nected from the feed water supply or from the arium unit disconnect valve.
• Push the draw-off lever to the right to release the system pressure.
• When water flow stops, pull draw­off lever forwards again and open the front doors of the housing.
• Pull the safe guard slides on the blue adapter towards you.
• Lift the adapter up out of the cartridge pack.
• Remove the cartridge pack.
• Place a 500-ml container under the adapter; lower the adapter until water flow stops.
• Repeat these steps with the second cartridge adapter.
• Unpack the new cartridge packs.
One cartridge pack is marked with a blue label, the other with a red label, (#2). In the housing to the lower left side you will find a blue-labeled connector, on the right a red connector.
• To facilitate connection of the adapter, moisten O-rings on the adapter for the cartridge pack.
• Place the blue adapter on to the blue-labeled cartridge pack (#1) and push firmly into place (see figure on page 13).
25
Two screws are seated in the cover of each cartridge pack. The two screw heads must rise from the drillings of the safe guard slides with the upper ring of the spacers.
• Push both safe guard slides all the way under the upper ring of the spacers (see figure on page 13).
• Place the blue-labeled cartridge pack in the housing on the left, push all the way home. The label must be facing you.
• Connect the red adapter to the red-labeled cartridge pack (#2) in the same manner on the right side of the unit.
• Close both doors.
• Open the feed water line or connect the line to the arium unit.
• Connect arium power cable and switch on the main switch on the left side of the housing (“I”).
The new cartridge packs have to be flushed prior to initial use (see Flushing the Cartridge Packs section on page 16).
Note
Do not attach the final filter to the NPT fitting of the dispensing valve until after the new cartridge flush has been completed.
Page 26
Ultrafilter Replacement
The length of your ultrafilter’s life will depend to some extent on conditions and use in your lab. When you are unable to completely sanitize your system (i.e. you can no longer obtain pyrogen-free water even after sanitizing your system), replace the ultrafilter cartridge as follows:
• Turn off arium 611VF and discon­nect it from the AC power outlet.
• Disconnect the water inlet tube from the arium System.
• Push the draw-off lever to the right to release the pressure form the system.
• When water stops draining, push the draw-off lever back to the front. Open the front doors of the system.
• Remove the two cartridge packs. Refer to the Section Cartridge Replacement on page 25) or push to one side.
• Unscrew the three screw fittings from the old ultrafilter (see figure on page 10) and remove the ultrafilter from the holder. To do so, hold your index finger under the lower fitting of the ultrafilter to keep any water left in the filter from running out. Empty the ultrafilter to drain.
• Unpack the new ultrafilter module.
• Connect the tubes that were unscrewed earlier to the new ultra­filter module: (1): Inlet is green - bottom (2): Product water adapter is blue­Upper left (3): Reject adapters are red -Top
Caution!
Danger of leaks. Make sure that the O-ring on the blue adapter is seated in a vertical position before connecting the ultrafilter.
• Push the ultrafilter module back into the holder so that it engages. The connection fitting on the side should be on the upper left.
• Re-install the two cartridge packs or push them back in place (see Cartridge Replacement, page 25).
• Close the two doors.
• Open the inlet water or reconnect the arium System to the feed water supply.
• Connect arium power cable and switch on the main switch on the left side of the housing (“I”).
• Perform a “long” UF Flush (5 minutes) (for details see UF Flush Section on page 17).
Note
1 or 2 additional short flush cycles may be required to remove air from the UF module.
Replacing the Final Filter
Replace the final filter whenever
the following conditions occur: – Every 30 days – The product water flow rate is
reduced, or – Bacteria break through is detected.
The final filter is shipped assembled
with a bell. To replace the final
filter assembly:
• Remove the old final filter assembly
by turning it counterclockwise.
• Remove the new final filter assem-
bly from its bag.
• Apply two to three wraps of Teflon
tape to the threaded end of the
filter.
• Screw the filter clockwise into the
outlet until snug.
• Rinse approx. 6 liters of water
through the filter to drain prior
to using the product water.
Note
If a newly installed final filter clogs
rapidly after installation, arium
611VF may need to be sanitized
to remove bacterial contaminants.
26
!!
Page 27
Cell Cleaning
Warning!
Severe electrical shock hazard or danger of electrocution!
– Only qualified personnel may
service and repair arium 611VF.
– Disconnect arium 611VF from
the AC power outlet before maintenance and servicing.
– Depressurize the unit by turning
off feed water and pushing the draw-off lever to the right before opening the front doors to replace the cartridges.
Caution!
Malfunction of the resistivity cell!
– The cell electrodes (4) are etched
to improve wetting characteristics. Do not mechanically abrade or damage the surface (i.e. do not clean with a wire brush, sandpaper, etc.).
– Do not immerse the entire cell
assembly in cleaning solution, only the electrode portion.
• Turn off arium 611VF (“O”) and disconnect it from the AC power outlet.
• Disconnect the feed water supply or the line to the arium unit.
• Push the draw-off lever to the right in order to depressurize the system.
• When water flow stops, pull the draw-off lever back forwards and open the front doors.
• Remove the red-labeled cartridge pack from the housing.
• Remove the gray cell cable (2) under the yellow hood covering.
• Remove the cell (1) from the black socket and unscrew it.
• Carefully remove the O-ring (3) before cleaning the cell.
• Clean the cell in a mild detergent solution using a soft brush or immerse for 10 min in a 1% hydrochloric or sulfuric acid solution.
Warning!
Danger of acid burns! Carefully follow the acid manufac­turer’s warnings and recommended handling procedures found on pack labels and material safety data sheets. Wear protective gloves and glasses.
27
• Thoroughly rinse the cell in deionized or distilled water follow­ing the detergent or acid cleaning.
• Apply two wraps of Teflon tape to threaded end of cell.
• Reinstall the cell well and hand­tighten.
• Reconnect the gray cable.
• Reinstall the right cartridge pack into the housing.
• Close the front doors.
• Open the feed water line or reconnect it to the arium system.
• Reconnect the unit to the AC power outlet and turn the unit back on using the main power switch on the left side of the housing (“I”). Press the Operate | stop button to start system.
• Open draw off valve and flush water to drain for 2 minutes.
• Close valve.
• The unit is now ready to operate again.
Clean the metal portion (4) of the resistivity cell
2
1
!
!
!!
3
4
Remove the gray cable from the resistivity cell
Page 28
Note
Make sure that the O-Ring on the new lamp is not dislocated.
• Replace the black plastic cover over the bulb holder and carefully screw into stainless steel chamber. After O-ring begins to seat tighten only a further 4 turn!
• Reconnect the power plug (1) to the bulb holder in the black plastic cap (2).
• Push the recessed shield over the power cord and screw on to the stainless steel chamber.
• Put the cartridge pack back into the unit.
• Close the doors.
• Open the inlet water line or reconnect it to the arium system.
• Reconnect the AC power cable und turn the unit back on using the main power switch on the left side of the housing (“I”).
• Perform a 5 minute UF flush cycle to purge air from the system.
Note
After installing a new UV bulb reset the timer, which reminds you to replace the UV bulb. Refer to the following section.
• Press the OPERATE| STOP key to put the system into OPERATE Mode.
28
To replace the ultraviolet bulb:
• Turn off arium 611VF using the main power switch (“O”) and disconnect it from the AC power outlet.
• Shut-off or disconnect the feed water supply.
• Push the draw-off lever back in order to depressurize the system.
• When water stops draining from the system, pull the draw-off lever back towards you and open the front doors.
• Take the left cartridge pack with connected adapter out of the housing or slide it to the right towards the center.
• Remove the screw from the metal shield or plastic cap.
• Remove the power plug (1) from the black plastic cap (2).
• Unscrew the top plastic cover from the stainless steel chamber of the UV bulb (see top figure on the next page).
• Hold a beaker (300 ... 500 ml size) under the black plastic cap (2) and slowly unscrew the cap from the stainless steel chamber (by turning it counterclockwise); collect any draining water in the beaker.
If the plastic cap is seated too tightly (2), insert a screwdriver into both grooves and carefully lift up the cap.
• Then remove the old lamp (4) from the stainless steel chamber.
• Insert new lamp with new O-Ring (3) into the stainless steel chamber.
UV Bulb Replacement
The UV bulb consists of a quartz glass tube with integrated mercury bulb. The quartz glass tube and the mercury bulb glass deteriorate during exposure to UV radiation and will eventually absorb the most desirable wavelengths. If only the mercury bulb is replaced, there is no guarantee that the desired UV radiation will pass the quartz sleeve in the reaction chamber.
Warning!
Danger of personal injury!
– Depressurize the system prior
to opening front doors.
– Allow a defective UV bulb to cool
off before removing it. Do not leave finger prints on a new bulb.
– Ultraviolet radiation is harmful to
the eyes and skin. Do not attempt to observe the lamp directly when energized.
Caution
Danger of irreversibly damaging UV bulb! Do not touch the glass portion of the lamp. It is recommended that lint free gloves be worn when handling the lamp. The glass portion must be free of fingerprints, perspiration, etc. A single fingerprint will reduce the effectiveness of the lamp. Clean the lamp with isopropyl alcohol and a lint free cloth.
!
!!
Page 29
Remove screw from metal shield
Remove UV bulb (4) from stainless steel chamber
29
Activate the UV Timer
The UV bulb should be replaced approximately every six months. A timer can be activated to display a message every 6 months to remind you that the UV bulb is ready to be replaced.
To activate the timer after installing a new UV bulb:
• Press MENU to get into the Main Menu:
• Select “Maintenance” and press ENTER to confirm.
• Select “UV Timer” by using y and press ENTER to confirm.
• Press ENTER to reset the timer.
The internal timer is reset and will remind you to replace the UV bulb after 6 months.
• Alternatively press the MENU key to return to the Main Menu without resetting the timer.
After 6 months the display will show the following:
18.2 MO * c m c Change UV Bulb
09.48 23.04.01
UV LIFE TIMER
Timer Reset Escape
MAINTENANCE
UV Life Timer Sanitization
MENU Utilities
Maintenance Setup
1234 5
Page 30
Fuse Replacement
arium 611VF has two fuses for the unit.
Warning!
Severe electrical shock hazard or danger of electrocution!
– Only qualified personnel may
service and repair arium 611VF.
– Disconnect arium 611VF from the
AC power outlet before mainte­nance and servicing.
– Ensure all tubing connections are
tight to avoid leakage.
– Depressurize the unit by pushing
back the draw-off lever to the right before opening the front doors to exchange the cartridges.
• Turn off arium 611VF using the main power switch (“O”) and disconnect it from the AC power outlet.
The fuse drawer with two safety fuses is located on the left side of the housing above the main power switch.
• Gently press the locking device of the fuse drawer upwards and pull out the fuse drawer (see figure).
Caution!
Danger of damage to components! First determine why the fuse blew and eliminate the cause. Then replace any defective fuses with new ones.
• Remove the old fuses and replace with fuses of the same type and rating.
• Place fuse drawer back into the housing until it clicks into place.
• Reconnect the unit to the AC power outlet and turn the unit back on using the main power switch (“I”).
The unit is now ready to operate again.
30
!
!!
Page 31
31
Appendix
Problem Possible Causes Solutions
Pump not operating, control panel not lit No electrical power to arium 611VF. Ensure that the arium 611VF power cord
is connected to a live power source and completely plugged into electrical outlet.
The power entry plug fuse is defective. Identify and eliminate problem, replace
the fuse (see section on Fuse Replacement
on page 29). The internal AC adapter fuse is Contact Sartorius Customer Service. defective.
Pump runs, but no display Connector from main PCB to display Disconnect unit from AC power.
PCB not plugged in. Check and reconnect.
Pump does not run. Display lit. Pump worn out or defective. Replace pump. System leakage Fitting is not tight or broken. Change or tighten fitting. High TOC UV bulb is not working. Replace UV bulb (see page 27).
Exhausted cartridge packs Replace the cartridge packs (see page 25). The system is contaminated. Perform sanitization (see section on System
Sanitization on page 16).
Check feed water quality.
Resistivity too low. Exhausted cartridge packs Replace the cartridge packs (see Cartridge
Pack Replacement section on page 25). The system is contaminated. Replace the cartridge packs (see section on
Cartridge Pack Replacement on page 25)
and perform sanitization (see section on
System Sanitization on page 16)
Low resistivity. Resistivity cell dirty. Clean cell (see section on Cell Cleaning
on page 30). The system is not calibrated. Perform calibration (see section on System
Calibration on page 21).
High endotoxin level. The system is contaminated. Replace the cartridge packs (see section on
Cartridge Pack Replacement on page 25),
perform sanitization (see section on System
Sanitization on page 16) and replace
the ultrafilter (see section on Ultrafilter
Replacement on page 26). The ultrafilter is contaminated Sanitize or replace the ultrafilter or defective. (see section on Ultrafilter Replacement
on page 26).
Reduced or no product water flow The final filter is clogged. Replace the final filter (see section on from the final filter. Final filter Replacement on page 26).
The system is contaminated. Replace the cartridge packs (see section on
Cartridge Pack Replacement on page 25)
and perform sanitization (see section on
System Sanitization on page 16). Pump defective Replace pump.
Short cartridge life. Feed water quality is low. Check the feed water source.
Increased water consumption. Replace cartridge packs with fresh ones.
Display shows “Replace UV bulb”. UV bulb burnt out. Replace UV bulb (see section on UV bulb
Replacement on page 27).
No Display Display cable not connected Connect display cable.
Troubleshooting Guide
Immediately after you discover any malfunction, use the main power switch to turn off arium 611VF. Then use the troubleshooting chart below to try to find out what the problem is and take the suggested action.
Page 32
32
Specifications
Dimensions W + D + H 43.2 + 48.3 + 34.3 [cm] Empty weight 15.9 kg Operating weight 23.6 kg Clearance requirements Sides: 10 cm minimum for cable and tubing connections
Front: 40 cm minimum for opening the doors
Inlet water requirements arium 611VF requires water pretreated by:
- distillation
- deionization or
- reverse osmosis Distilled water: > 250 KO · cm (< 4 µS/cm) RO water: TDS < 25 ppm CaCO
3
> 20 KO · cm (< 50 µS/cm)
Deionized water: TDS < 10 ppm CaCO
3
> 50 KO · cm (<20 µS/cm),
All: Turbidity < 1 N.T.U.
Silica < 1000 ppb TOC < 1000 ppb
Pressure: From atmospheric tank inlet to a maximum inlet pressure of 7 bar
Product water capabilities Quality of the
purified water: 18.2 MO + cm at 25°C
RNase-/DNase-/DNA-free Bacteria: < 1 CFU/100 ml TOC: < 1 ppb Endotoxin: < 0,001 EU/ml
Flow rate Up to 1.5 l/min maximum at minimum inlet feed water pressure of 2.0 bar at 50 HZ and
with a new final filter.
Environmental conditions Operating: 5 °C - 28°C; 80% relative humidity, non-condensing
Storage: 5 °C - 45°C; 80% relative humidity, non condensing
Electrical requirements 100-240 V~, 50/60 Hz, 1 phase Fuses power entry 2 fuses 5 + 20 mm, time lag, 250 V, 1 A
Installation Category II (over-voltage) in accordance with IEC 664, Pollution Degree 2 in accordance with IEC 664. Altitude limit: 5,000 meters.
Complies with the EMV: EN 50081-1, EN 50082-1 following standards Safety: IEC 1010-1-92
Page 33
Accessories and Replacement Parts
Ordering Replacement Parts
When requesting any customer service or replacement parts or in any correspon­dence about arium 611VF, be sure to state the complete model number and the serial number that are on the Specification Plate label on the rear panel of the unit.
You can obtain any replacement parts listed in this manual from Sartorius or from your local Sartorius dealer.
Order No. Description
611VF Complete Unit arium 611VF
611CKRU Consumable kit with
– 611CPR for RO or Distilled inlet water if application
requires low TOC. Will be used on the left (blue) side of the system.
– 611CPU for low TOC level application in systems with an
UV lamp chamber. Will be used on the right (red) side of the system.
– 5441307H4—NO Sartopore 2 150 Final Filter (2+).
611CKDU Consumable kit with
– 611CPD for DI water if application requires low TOC levels.
Will be used on the left (blue) side of the system and Pack.
– 611CPU for low TOC level application in systems with an
UV lamp chamber. Will be used on the right (red) side of the system.
– 5441307H4—NO Sartopore 2 150 Final Filter (2+).
611CDU5 Replacement Ultrafilter Cartridge arium 611
5441307H4—NO—B Sartopore 2 150 Final Filter (Pack of 5)
611CEL1 Replacement UV lamp arium 611
611CEF1 Replacement fuses power inlet 2 fuses
611CDS2 Sanitization Kit 2 syringes
611CDS6 Sanitization Kit 6 syringes
611AMA1 Inlet feed water adapter 4” NPTF 1”
611AEC1 Calibration module
611AMB1 Wall Bracket
33
Page 34
Page 35
Page 36
Printed in Germany on paper that has been bleached without any use of chlorine W4A000 · KT Publication No.: SLG6010-e02042 Order No.: 85030-513-68
Sartorius AG Weender Landstrasse 94-108 37075 Goettingen, Germany
Telephone +49.551.308.0 Telefax +49.551.308.3289 www.sartorius.com
Copyright by Sartorius AG, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG.
The status of the information, specifications and illustrations in this manual is indicated by the date given below. Sartorius AG reserves the right to make changes to the technology, features, specifications, and design of the equipment without notice.
Status: April 2002, Sartorius AG, Goettingen, Germany
Loading...