Sartorius 1002MP9, 1006MP9 Installation And Operation Instructions Manual

Page 1
Electronic Toploader
1002MP9, 1006MP9
Installation and Operation Instructions
Page 2
Electronic Toploader
1002MP9 and 1006MP9
Model 1002MP9
Model 1006MP9
Page 3
Complete consignment:
Electronic toploader, Pan; power supply component
Connect power supply component (1).
Balance calibration: Remove cap from calibration button (8). Tare balance: display: 0,0 g / 0 g (g = stability indication)
Switch balance on (4). First, the auto-check codes are display followed by automatic “0,0 g” by 1002MP9, “0 g” by 1006MP9
Load test weight (see accessories). If balance display deviates from test weight, use the calibration button (8) to set display to test weight value. Repeat procedure to assure accuracy.
Page 4
Load battery / NC battery: switch balance off, open battery compt. (3). Insert battery / NC battery.
Important: make sure you have correct polarity, check for proper setting of the battery / NC battery selector (2).
Environmental conditions:
switch balance off. Switch 3 in upper position = stable conditions, in lower position = unstable environment
Operating mode switches (5): pull open the cover. switch 3 = stable operating environment, switch 4 = automatic switch-off is off.
Automatic switch-off:
Switch balance off. Switch 4 in upper position = off, in lower position = on
Page 5
Duty button
When the ON/OFF switch (4) is set to "on", the balance can also be switched off and back on again with the duty button (6). lf there is no weight change for at least 0.5 minutes, the display will read "---" followed by switch-off.
To switch on: depress button (6) - display will show multiple eight-check-
To switch off: depress button (6) for approx. 3 secs., after which display will read "---" confirming balance is switching off.
Accessories
Calibration weight (1x500 g) cat. no. 707216
Carrying case cat. no. 606707 NC battery* cat. no. 629809
The manufacturer's guaranty does not cover balance damage resulting from battery / NC battery leakage.
Data lnput with 4 function keys cat. no. 609001
* Balance is suitable for use with commercially sold NC batteries (e.g. 6x501 RS Mignon) or batteries (e.g. 6x1.5 V Mignon cells).
Warning! Use only NC batteries with a minimum permanent Charge current of 50 mA. Disregard of this specification may result in leakage.
Technical Data
Model 1002 MP9 1006 MP9
Weighing range 505 g 5050 g
Readability 0,1 g 1 g
Standard deviation (reproducibility)
±0,1 g
±1 g
Max. deviation from linearity
±0,1 g
±1 g Taring range (by subtraction) 505 g 5050 g
Stabilization time (typical) ~ 2s
Integration time 400/800 ms
Display update
200 ms Ambient temperature range 273 K ... 313 K (0oC ...+40oC) Sensitivity drift (ambient temperature
range 283 K … 303 K)
±710
-5
/K
±410-5/K
Power supply from power supply component 220 V/110 V
Consumption 0,35 W
Batteries 6x 1,5 V Mignon cells
Service life approx. 30 hrs. in continuous operation
NC batteries six 501 RS Mignon
Service life approx. 18 hrs. in continuous operation
Page 6
Sartorius AG
B 37070 Göttingen P Weender Landstraße 94–108, 37075 Göttingen T (0551) 308-0, F (0551) 308-3289
Internet: http://www.sartorius.com Copyright by Sartorius AG, Göttingen, Deutschland.
All rights reserved. No part of this publication may be printed or translated in any form or by any means without the prior written permission of Sartorius AG.
Sartorius AG reserves the right to make change to the technology, features, specification and design of the equipment without notice.
Loading...