Refrigerant circuit contains HFC refrigerant that effect to environment little.
Latest cooling system reduces negative factor to global environment. With ambient
temperature at 35℃, while temperature of inside chamber maintains -30℃.
■ PCM structure
PCM structure reduces amount of organic solvent.
■ Temperature control function
Unified temperature control range for same categories realized to set -20℃, which
is suitable for reagent preservation.
■ Specification unified
Unified specification for same categories, which also realized to unify performance for
high/low temperature alarm (SV is changeable in the range ±5~±15℃), power failure
alarm, remote alarm terminal, setting memorization by non-volatile memory.
Adopt new microprocessor system for display panel and keypad makes operation easily.
■ Self diagnosis function
Abnormal condition for temperature sensor is indicated by self-diagnosis system.
■ Validation
Control panel enables zero “0” adjust for validation.
■ New door latch mechanism
New door latch is adopted in order to close door surely for reagents.
It is also available to use individual lock mechanism.
■ Insulation wall thickness
Unit has necessary insulation performance for each temperature range.
Combination of PCM and frame form makes 610mm for frame size commonly.
■ Option
Internal temperature observation system is option.
Unit has communication function corresponding to it.
−1−
Page 5
Features
<MDF-U537,537D>
<MDF-U537,537D>
<MDF-U537,537D><MDF-U537,537D>
■ HFC refrigerant
Refrigerant circuit contains HFC refrigerant that effect to environment little.
Latest cooling system reduces negative factor to global environment. With ambient
temperature at 35℃, while temperature of inside chamber maintains -30℃.
■ PCM structure
PCM structure reduces amount of organic solvent.
■ Temperature control function
New cooling system realizes to set -20℃, which is suitable for reagent preservation.
■ Specification unified
Unified specification for same categories, which also realized to unify performance for
high/low temperature alarm (SV is changeable in the range ±5~±15℃), power failure
alarm, remote alarm terminal, setting memorization by non-volatile memory.
Adopt new microprocessor system for display panel and keypad makes operation easily.
■ Self diagnosis function
Abnormal condition for temperature sensor is indicated by self-diagnosis system.
■ Validation
Control panel enables zero “0” adjust for validation.
■ New door latch mechanism
New door latch is adopted in order to close door surely for reagents.
It is also available to use individual lock mechanism.
■ Draw box for reagent preservation
100mm for depth of box makes things easily to arrange and take in & out.
(Adopted in bottom chamber of MDF-U537/U537D)
■ Compressor protection function
Sensor detects compressor’s warming up to control unit running.
−2−
Page 6
Specification
MDF-U333
Items Specification
Model description Biomedical freezer
Model Number MDF-U333
Outer dimension
Inner dimension W 490 × D 485 × H 1290 mm
Leveling legs 2 pcs (with caster)
Access hole φ 30mm. 1 place on rear.
Weights 131 Kg
Evaporator Tube on sheet type (function both evaporator and shelf)
Condenser Wire and tube type
Compressor Hermetic type 350W
HFC refrigerant
Refrigerant
R-407D
Cooling performance -30℃ (internal 1/2h air temperature, AT:35℃, no load)
Internal temperature control
range
-20℃~-30℃
(AT:5℃~35℃、no load)
Power supply Local voltage
−5−
Page 9
MDF-U537
Items Specification
Temperature controller
Microprocessor controlled system with non-volatilized memory.
Temp. set range : -18℃~-40℃
Temperature sensor Thermistor 2P
Temperature display
High/Low
Temperature
Alarm
Power failure
Remote alarm
contact
LED digital display (green colored) by 1℃.
Range : -50℃~50℃
Changeable SV±5℃~15℃. (Initial SV±10℃)
ALARM lamp brinks and buzzer sounds intermittently
(about 15 min. delay)
Remote alarm contact: turns ON during power failure alarm.
ALARM lamp brinks, buzzer sounds intermittently, remote alarm
contact turns ON during power failure alarm.
Remote alarm terminal 3P: contact capacity DC30v, 2A, N.O. –COM.
(During power failure, temperature alarm: ON)
ALARM lamp
BUZZER: Buzzer key
ALARM TEST: Alarm test key
Control panel
Operation
panel
SET: To switch the SV and PV
>: To shift the set digit
∧: To change the SV
DEF: Defrost key
Rear
Key lock function
Power switch (Breaker switch)
Battery switch: Power failure alarm switch
Press > key for five seconds
L0: Unlock L1: Lock
When each sensor is in fault,
Self diagnosis function
Error code and internal temperature are displayed alternately.
Remote alarm contact turns ON with buzzer sounds.
Power switch Breaker switch
Overload relay
Detect temperature near compressor by thermistor sensor to control
Compressor protection
compressor during overload.
Alarm: Error code and internal temperature are displayed alternately.
Remote alarm contact turns on with buzzer sounds.
Chart recorder: MTR-G85
Option
Mounting kit: MPR-S7(for bottom mounting)
: MDF-S740T(for top mounting)
−6−
Page 10
Specification
MDF-U537D
Items Specification
Model description Biomedical freezer
Model Number MDF-U537D
Outer dimension
Inner dimension
W 804 × D 772 × H 1802 mm
(Max)
Upper chamber W 658 × D 607 × H 589 mm
Lower chamber W 658 × D 607 × H 603 mm
Leveling legs 2 pcs (with caster)
Access hole φ 30mm. 2 places on rear.
Weights 136 Kg
Evaporator Tube on sheet type (function both evaporator and shelf)
Condenser Wire and tube type
Compressor Hermetic type 350W
HFC refrigerant
Refrigerant
R-407D
Cooling performance -30℃ (internal 1/2h air temperature, AT:35℃, no load)
Internal temperature control
range
-20℃~-30℃
(AT:5℃~35℃、no load)
Power supply Local voltage
−7−
Page 11
MDF-U537D
Items Specification
Temperature controller
Microprocessor controlled system with non-volatilized memory.
Temp. set range : -18℃~-40℃
Temperature sensor Thermistor 2P (upper and lower compartment)
LED digital display (green colored) by 1℃.
Temperature display
Range : -50℃~50℃
Selectable to display upper or lower compartment by switch.
High/Low
Temperature
Alarm
Power failure
Remote alarm
contact
Changeable SV±5℃~15℃. (Initial SV±10℃)
ALARM lamp brinks and buzzer sounds intermittently
(about 15 min. delay)
Remote alarm contact: turns ON during power failure alarm.
ALARM lamp brinks, buzzer sounds intermittently, remote alarm
contact turns ON during power failure alarm.
Remote alarm terminal 3P: contact capacity DC30v, 2A, N.O. –COM.
(During power failure, temperature alarm: ON)
ALARM lamp
BUZZER: Buzzer key
ALARM TEST: Alarm test key
Control panel
Operation
panel
SET: To switch the SV and PV
>: To shift the set digit
∧: To change the SV
DEF: Defrost key
TOP/BOTTOM: Switch key to display upper or lower compartment.
Rear
Key lock function
Power switch (Breaker switch)
Battery switch: Power failure alarm switch
Press > key for five seconds
L0: Unlock L1: Lock
When each sensor is in fault,
Self diagnosis function
Error code and internal temperature are displayed alternately.
Remote alarm contact turns ON with buzzer sounds.
Power switch Breaker switch
Overload relay
Detect temperature near compressor by thermistor sensor to control
Compressor protection
compressor during overload.
Alarm: Error code and internal temperature are displayed alternately.
Remote alarm contact turns on with buzzer sounds.
Chart recorder: MTR-G85
Option
Mounting kit: MPR-S7(for bottom mounting)
: MDF-S740T(for top mounting)
−8−
Page 12
Dimensions
<<<<MDF-U333>>>>
−9−
Page 13
Dimensions
<<<<MDF-U537>>>>
−10−
Page 14
Dimensions
<<<<MDF-U537D>>>>
−11−
Page 15
Refrigeration circuit
MDF-U333 Unit:mm
Parts description Specification
Evaporator
Dehydrator
Capillary tube
Condenser assembly A
Condenser assembly B
Frame pipe
Evacuation and vacuum
pressure
Refrigerant and oil
Refrigeration circuit spec.
and welding points
Type Tube on sheet
Material Aluminum (φ 8.0)
Type of molecular sieves
4A-XH9(134 stamped)
Charged q’ty 8±1g
Resistance 0.686MPaG (7Kgf/㎠ G)
Dimension φ 1.8×(φ 0.65)×2000mm
Resistance indicator color White
Type Wire and tube
Dimension
w
480×P40×30 columns
Type Wire and tube
Dimension
w
480×P40×30 columns
Note: Lower 2 columns are used
in desuper circuit.
Material C1220T-1/2H
Dimension
φ 4.0×
t
0.5
Evacuation time: over than 20min. by 2 ports (high pressure and low
pressure side). (N2 exchange)
When charged, vacuum pressure is lower than 1.0torr
(pump capacity: 300L/min)
Resistance 0.725MPaG (7.4Kgf/㎠ G)
Dimension φ 1.8×L4500
Resistance indicator color Blue
Condenser assembly
front (Total number: 4)
Frame pipe
Type Wire and tube
Dimension
Material C1220T
Dimension
Evacuation and vacuum
pressure
Evacuation time: over than 20min. by 2 ports (high pressure and low
pressure side). (N2 exchange)
When charged, vacuum pressure is lower than 1.0torr
(pump capacity: 300L/min)
Upper compartmnt Lower compartmnt
Resistance 10.3Kg/㎠ G 9.9Kg/㎠ G
Dimension φ 1.8×L4000 φ 1.8×L3000
Resistance indicator color Yellow Red
Condenser assembly
front (Total number: 4)
Frame pipe
Type Wire and tube
Dimension
Material C1220T
Dimension
Evacuation and vacuum
pressure
Evacuation time: over than 20min. by 2 ports (high pressure and low
pressure side). (N2 exchange)
When charged, vacuum pressure is lower than 1.0torr
(pump capacity: 300L/min)
TypeAMVL-180AAMVL-180AAMVL-300A
Pick up voltage(50Hz)AC91V〜111VAC91V〜111VAC185V〜217V
(60Hz)AC108〜128VAC108〜128VAC215〜232V
Drop out voltage(50Hz)AC30〜65VAC30〜65VAC60〜120V
(60Hz)AC30〜75VAC30〜75VAC69〜132V
TypeP24KUP19KUP10MU
Action to the temp. (No current)
OFF 150±10℃150±10℃130±8℃
ON 85±10℃85±10℃69±10℃
Action to the current (AT25℃)24.0A19.5A10A
Operation time 6~15 seconds6~15 seconds6~15 seconds
Non-Action to the temp. (80℃)
Non-Action
Action 8.0A within 15 minutes7.4A within 15 minutes3.4A within 15 minutes
Code807764201807763251807763251
Rated voltage (50/60Hz)
Winding resistance C-S(Aux)5.8Ω30.6Ω30.6Ω
C-R(Main)1.6Ω7.8Ω7.8Ω
Pick up voltage(50Hz)AC91V〜111VAC185V〜217VAC185V〜217V
(60Hz)AC108〜128VAC215〜232VAC215〜232V
Drop out voltage(50Hz)AC30〜65VAC60〜120VAC60〜120V
(60Hz)AC30〜75VAC69〜132VAC69〜132V
Action to the temp. (No current)
OFF 120±8℃130±8℃130±8℃
ON 69±10℃69±10℃69±10℃
Action to the current (AT25℃)18.5A10A10A
Operation time 6~15 seconds6~15 seconds6~15 seconds
Non-Action to the temp. (80℃)
Non-Action
Action 5.5A within 15 minutes3.4A within 15 minutes3.4A within 15 minutes
Code807764201807763251807763251
Rated voltage (50/60Hz)
Winding resistance C-S(Aux)5.8Ω30.6Ω30.6Ω
C-R(Main)1.6Ω7.8Ω7.8Ω
Pick up voltage(50Hz)AC91V〜111VAC185V〜217VAC185V〜217V
(60Hz)AC108〜128VAC215〜232VAC215〜232V
Drop out voltage(50Hz)AC30〜65VAC60〜120VAC60〜120V
(60Hz)AC30〜75VAC69〜132VAC69〜132V
Action to the temp. (No current)
OFF 120±8℃130±8℃130±8℃
ON 69±10℃69±10℃69±10℃
Action to the current (AT25℃)18.5A10A10A
Operation time 6~15 seconds6~15 seconds6~15 seconds
Non-Action to the temp. (80℃)
Non-Action
Action 5.5A within 15 minutes3.4A within 15 minutes3.4A within 15 minutes
The following show temperature and resistance characteristics on each thermistor sensor
(type 502).
Temperature
(℃)
-50 154.50 -7 17.92 12 8.17
-45 116.50 -6 17.16 13 7.85
-40 88.85 -5 16.43 14 7.55
-35 68.15 -4 15.74 15 7.27
-30 52.84 -3 15.08 16 6.99
-25 41.19 -2 14.45 17 6.73
-20 32.43 -1 13.86 18 6.48
Resistance
(kΩ)
Temperature
(℃)
Resistance
(kΩ)
Temperature
(℃)
Resistance
(kΩ)
-19 30.92 0 13.29 19 6.24
-18 29.50 1 12.74 20 6.01
-17 28.14 2 12.22 25 5.00
-16 26.87 3 11.72 30 4.18
-15 25.65 4 11.25 35 3.51
-14 24.51 5 10.80 40 2.96
-13 23.42 6 10.37 45 2.51
-12 22.39 7 9.96 50 2.14
-11 21.41 8 9.57 55 1.83
-10 20.48 9 9.20 60 1.57
-9 19.58 10 8.84
-8 18.73 11 8.49
−24−
Page 28
Controlspecification
Controlspecification
ControlspecificationControlspecification
<<<<MDF-U333・・・・537・・・・537D>>>>
Keyandswitch
Keyandswitch
1.
1.
KeyandswitchKeyandswitch
1.1.
BZ :When alarm lamp is blinking and buzzer sounds, buzzer and remote alarm
Output …………Force to turn off
In F25, setting value (time) is except for 000;buzzer will turn ON after the
setting time has passed. (This RING BACK function would be carried out in
s/n 110XXXXX or bigger)
:Buzzer is not turned off during alarm test. :When alarm lamp is blinking and buzzer is off, buzzer will not turn on. :When press this key during power failure, currentinternal temperature is
displayed for 5 seconds.
SET :Press once to setting mode and press again to memorize the set value.
(It works as ENTER key)
DEF :In PV display keep pressing the key over 5 seconds to start defrosting.
During defrosting, press the key to quit defrosting. During defrosting, internal temperature and “dF” is displayed alternately. During defrosting, high-low temperature alarm is cancelled.
When sensor error is occurred during defrosting, error code and internal
temperature is displayed alternately, and “dF” display is turned off.
During defrosting, only buzzer key is acceptable.
ALARM TEST:Press the key in PV display, alarm test will be start.
Press again then quit alarm test. During alarm test, alarm lamp will blink and buzzer will sound intermittently. Digital display will be turned off. During alarm test, only alarm test key is acceptable.
:During setting mode, shift between the first digit and the second digit.
In PV display, keep pressing the key over 5 seconds to enter the key lock
mode. (“L0”display)
:During setting mode, count up the blinking digit. In PV display, keep
pressing the key over 5 seconds to enter the function mode (“F00” display)
TOP/BOTTOM :Thiskey is available only in MDF-U537D.
Initially TOP (upper compartment) temperature is displayed.
Press key to display BOTTOM (lower compartment) temperature.
Press this key during power failure to display compartment temperature for 5
seconds, that was not selected before power failure.
−25−
Page 29
2.
2.Temperature control and High, Low temperature alarm
2.2.
<Temperature control>
Setting range : -18℃~-40℃
Display range : -50℃~+50℃
Setting method :Press SET key and set the required value with key and
Key. Press SET key again to memorize the set value,
And then return to PV display.
Out of setting range : If change to out of setting range value by mistake and press
SET key , buzzer sound continuously for 1 second.
<Alarm>
High temperature alarm :
When the internal temperature is higher than SV (+5℃~+15℃), alarm lamp and
digital display is blinking. After 15min.later, buzzer will sound and remote alarm
contact will turn ON. (initial value: +10℃)
Low temperature alarm :
When the internal temperature is lower than SV(-5℃~-15℃), alarm lamp and digital
display is blinking. After 15min.later, buzzer will sound and remote alarm contact will
turn ON. (initial value: -10℃)
Setting method: Keep pressing key over 5 seconds to enter function mode (“F00”).
Then press key to count up value. Change “F01” to input high temperature alarm
value, “F02” to input low temperature alarm value.
Set either “F01” or “F02” then press SET key to turn alarm temperature setting mode
to set required value with key and key.
Press SET key again to memorize value.
Out of setting range :
If change to out of setting range valueby mistakeand press SET key, sound buzzer
continuously for 1second andremain in setting mode to maintain the setting value.
3.
3. Defrosting
3.3.
This function is different by each model code (products) .
a. In PV display keep pressing DEF key over 5 seconds, defrosting will be started.
Model code 001 (MDF-U333) --- Force to turn compressor OFF.
Model code 002 (MDF-U537) --- Force to turn compressor & solenoid valve OFF. (Use 1 solenoid valve)
Model code 003 (MDF-U537D)
--- Single compartment → Force to turn OFF solenoid valve at the ROOM lamp illuminated.
--- Both compartments → Force to turn OFF 2 solenoid valves & compressor.
b. Press again DEF key to quit defrosting.(Manual start and finish)
−26−
Page 30
4.
4. Self diagnosis function
4.4.
E01 : Temperature sensor (upper compartment temperature sensor for MDF-U537D)
is open circuit
E02 : Temperature sensor (upper compartment temperature sensor for MDF-U537D)
is short circuit
E03 : Lower compartment temperature sensor is open circuit (for MDF-U537D only)
E04 : Lower compartment temperature sensor is short circuit (for MDF-U537D only)
E05 : Compressor protect sensor is open circuit (for MDF-U537,U537D)
E06 : Compressor protect sensor is short circuit (for MDF-U537,U537D)
E07 : Solenoid valve control sensor is open circuit (for MDF-U537 only)
E08 : Solenoid valve control sensor is short circuit (for MDF-U537 only)
E09 : Buttery switch OFF with ALARM TEST
E10 : Compressor temperature is abnormal (for MDF-U537, U537D)
a. Model code setting (for code: 001,002,003)
When temperature sensor is open circuit, display shows alternately E01 with –50℃.
When temperature sensor is short circuit, display shows alternately E02 with +50℃.
b. Model code setting (for code: 002)
When solenoid valve control sensor is open circuit, display shows alternately E07 with
–50℃.
When solenoid valve control sensor is short circuit, display shows alternately E08 with
+50℃.
c. Model code setting (for code: 003)
When TOP lamp is displayed, display shows same as a. in upper compartment.
When BOTTOM lamp is displayed, display shows lower compartment temperature.
When lower compartment temperature sensor is open circuit, display shows alternately
E03 with –50℃.
When lower compartment temperature sensor is short circuit, display shows alternately
E04 with +50℃.
d. Error diagnosis
Self diagnosis of open circuit is done when it is lower than -60℃.
Self diagnosis of short circuit is done when it is higher than +60℃.
E09 is functioned only during ALARM TEST is performed.
In E10 is displayed: When condenser temperature is higher than 85℃, it diagnoses as
abnormal and indicate E10. When condenser temperature is higher
than 100℃ force to turn compressor OFF.
When condenser temperature is lower than 65℃, it diagnoses as
normal and cancel to force compressor turn OFF.
When numbers of error happened simultaneously, error code of smallest number will be
displayed. With exception for MDF-U537D, alternative display of error code and PV,which
is selected compartment internal temperature .
During power failure, display shows alternately E09 with blank.
−27−
Page 31
5.
5. Key lock function
5.5.
In PV display、Keep pushing key over 5 seconds to enter key lock mode displaying with
“L0”.
Push key to 1 ・・・・・key lock ON 0 ・・・・・key lock OFF
then press SET key to memorize set value then return to PV display.
Note) With key lock ON to enter SET mode, but unable to change value.
With key lock ON to perform Function and ALARM TEST, while unable to perform
DEF.
6.
6. Power failure alarm (Buzzer and display isn’t functioned without buttery power
6.6.
supplied)
When power failure is occurred, power failure alarm will be turned on by battery inside units.
During power failure alarm, alarm lamp is blinking, digit display turns off, buzzer sounds
intermittently and remote alarm contact turns ON.
Press BZ key to turn power failure alarm buzzer OFF.
Press BZ key to display PV for 5 seconds.
During power failure alarm, only BZ key (TOP/BOTTOM key for MDF-U537D) is available.
7.
7. Alarm test
7.7.
The function is to check performance (failure) for alarm lamp, buzzer and remote alarm.
In PV display, press ALARM TEST key to blink alarm lamp, to sound buzzer intermittently, to
turn remote alarm contact on (Without time delay). Digital display will be turned off.
During alarm test, only ALARM TEST key is available.
Note: During alarm test, with buttery switch turned OFF to display error code E09.
Displaying error code E09 takes priority over other error codes before ALARM TEST is
performed.
8.
8. Auto return function
8.8.
Function: During setting mode or key lock mode or function mode with any key not pressed
for 90 seconds, value will not be memorized to return to PV mode automatically.
Note: In F09 or F10, auto return function will not be performed.
9.
9. Compressor delay start time
9.9.
During cycle time, normally force to turn compressor OFF after 2 minutes passed compressor
turned off.
When main power is supplied (microcomputer is RESET), changeable to set compressor time
lag for 1 ~15 minutes. (See F05)
−28−
Page 32
10.
10.Function mode
10.10.
Functions are as follow.
F00 : To return to PV display
F01 : To change temperature of high temperature alarm
F02 : To change temperature of low temperature alarm
F03 : To change temperature of high temperature alarm
(Applied for lower compartment of MDF-U537D)
F04 : To change temperature of low temperature alarm
(Applied for lower compartment of MDF-U537D)
F05 : To set compressor delay start time
F06 : To input service code (384)
F07 : To adjust zero for temperature sensor
(Upper compartment sensor for MDF-U537D)
F08 : To adjust zero for temperature sensor
(Not available for MDF-U333)
(Solenoid valve control sensor for MDF-U537)
(Lower compartment sensor for MDF-U537D)
F09 : Not available
F10 : Not available
F11 : Not functioned. Return to PV display.
F12 : To display temperature of temperature sensor (per 0.1℃)
F13 : To display temperature of MDF-U537/U537D compressor protect sensor (per 1℃)
F14 : To display temperature of temperature sensor (per 0.1℃)
(Solenoid valve control sensor for MDF-U537)
(Lower compartment sensor for MDF-U537D)
F15 : Not available
F16 : Not functioned. Return to PV display.
F17 : To set or check model code
model code 001: MDF-U333
model code 002: MDF-U537
model code 003: MDF-U537D
F18 : Preset value for second offset performance: -26℃ (for MDF-U537/U537D)
F19 : Not functioned. Return to PV display.
F20 : Not functioned. Return to PV display.
F21 : To set communication ID(000~255)
F22 : To set communication mode
F23,24,26~29 : With input each function code, press SET key to sound buzzer 1 second
continuously to maintain input status.
F25 : To set alarm ring back time
Setting value: 000,010,020,030,040,050,060
Setting method ::::
In PV display, keep key pressing over 5 seconds to enter function mode and will be
displayed “F00”. Press key to input required function code then press
SET key again.
F00 : Not functioned. In F00 displayed, press SET key to return PV display.
−29−
Page 33
F01 : To set temperature of high temp. alarm performance in the range of +5℃~+15℃.
(initial value : +10℃)
F02 : To set temperature of low temp. alarm performance in the range of -5℃~-15℃.
(initial value : -10℃)
F03 : To set temperature of high temp. alarm performance in the range of +5℃~+15℃.
(initial value : +10℃)
Return to PV display except for lower compartment of MDF-U537D.
F04 : To set temperature of low temp. alarm performance in the range of -5℃~-15℃.
(initial value : -10℃)
Return to PV display except for lower compartment of MDF-U537D.
F05 : To set compressor delay start time
(performed only when main power supplied or microcomputer reset)
The function is to reduce compressor’s start up failure or breaker’s blown out that
caused by numbers of units simultaneous start up after power failure.
The setting range of time lag is 1~15 minutes. (Initial value is 1 minute on factory
outgoing)
SV before doing factory’s performance test, is 3 seconds or less. Therefore, the time
lag is 0 (3 seconds) when PCB replacement isdone in service. It is recommended to
input 1~15 for value when PCB is replaced.
F06 : Input service code with key and key when F07 and following functions,
except F21, F22 and F25 would be performed. Press SET key to memorize service
code to enter function mode again to perform F07 and following functions.
Service code : 384
Note)Service code is memorized unless input “000” again in ”F06”, or turn power OFF.
(Turn OFF = buttery switch and power supply switch are turned OFF)
Note) After using service code on service, input “000” again in “F06”.
F07 :To adjust 0 of temperature sensor (setting range: -9.9℃~+9.9℃)
Setting method: In ”F07” displayed, press SET key to display ”00.0”(initial value).
Set temperature with key and key and press SET key
again to memorize set value. Input is done by differential value.
Ex) When actual temperature of 1/2h is shown –28.5℃and digital display is
shown -30℃, add +1.5 to value in F07.
Calculation = (-28.5-(-30)) = +1.5
key enables to change only “0” and “-“ in second digit and to change
“0”~”9” in first digit and first decimalplace.
key enables to shift the second digit ⇔ the first digit ⇔
the first decimal place⇔the second digit.
−30−
Page 34
F08 :To adjust 0 of temperature sensor (setting range: -9.9℃~+9.9℃)
In F17: It is functioned in model code 002(MDF-U537), 003(MDF-U537D).
Return to PV display in model code 001(MDF-U333)
Setting method: In ”F08” displayed, press SET key to display ”00.0”(initial value).
Set temperature with key and key and press SET key
again to memorize set value. Input is done by differential value.
Ex) When actual temperature of 1/2h chamber is shown –28.5℃and digital
display is shown -30℃, add +1.5 to value in F08.
Calculation = (-28.5-(-30)) = +1.5
key enables to change only “0” and “-“ in second digit and to change
“0”~”9” in first digit and first decimalplace.
key enables to shift the second digit ⇔ the first digit ⇔
the first decimal place⇔the second digit.
F12:Temperature of temperature sensor is displayed as per 0.1℃.
From second digit to first decimal point is shown in display.
It is not shown “-“ in display when temperature is lower than -20℃.
(Ex. Actual temperature -35℃ → Display shows 35.0℃)
When temperature is within the range of -19.9℃~-0.1℃, ”-“ is displayed.
Range : -72.0~+83.0℃(±9.9℃)
F13 :Temperature of compressor protect sensor is displayed as per 1℃.
Range: -72~+163℃
F14 :Temperature sensor (solenoid valve control sensor for MDF-U537, lower compartment
temperature control sensor for MDF-U537D)
Temperature of each sensor is displayed as per 0.1℃
From second digit to first decimal point is shown in display.
It is not shown “-“ in display when temperature is lower than -20℃.
(Ex. Actual temperature -35℃ → Display shows 35.0℃)
When temperature is within the range of -19.9℃~-0.1℃, ”-“ is displayed.
Range : -72.0~+83.0℃
−31−
Page 35
F17 :Model code setting or checking
001:MDF-U333 002:MDF-U537 002:MDF-U537D
Followings are initial setting values
Internal temp. SV : -30℃ (for U537D : both compartments)
Key lock mode : 0 (OFF)
High temperature alarm : SV+10℃(for U537D : both compartments)
Low temperature alarm : SV-10℃(for U537D : both compartments)
0 adjustment:+0℃(for temperature sensor)
Compressor start delay time : 1 minute
Communication ID:000 Communication mode:000
Alarm ring back time : 030 (30 minutes)
Set model code as MDF-U537, MDF-U537D to activate compressor protect function.
F21 : To set serial communication ID.
Input range: 000~255 (000 means that communication is OFF)
F22 : To set serial communication mode.
rd
Control mode (3
digit) 0 : Local (initial value)
1 : Remote
nd
Baud rate (2
digit) 0 : 2400bps (initial value)
1 :4800bps
2 :9600bps
st
1
digit is not used.
Note) Set control mode as “remote” unable to change SV and to activate defrosting.
F25 : To set value for alarm recovery (ring back) time.
Input value: 000,010,020,030,040,050,060 (In 000, recovery function is OFF)
11.
11.Differential (COMP. ON/OFF point on program)
11.11.
COMPON permit: Activate at +0.0℃ higher than SV COMPOFF: Activate at -0.6℃ lower than SV
− 32 −
Page 36
12.
12.Offset value
12.12.
The function is to correct differential value between temperature of temperature sensor and
of internal 1/2H (centre of chamber).
Initial offset
001 (MDF-U333): Shift –1.5℃from temperature of temperature control sensor (PV)
002 (MDF-U537): Shift +1.0℃from temperature of temperature control sensor (PV)
(For upper, lower compartment)
003 (MDF-U537D): Shift +0.0℃from temperature of temperature control sensor (PV)
(For upper, lower compartment)
Second offset
001 (MDF-U333): None
002 (MDF-U537):-18℃≧SV≧-25℃--- shift –0℃from PV (For upper, lower compartment)
-30℃>SV --- shift –0℃from PV (For upper compartment)
*Note) For lower compartment, add 0℃ to PV.
13.
13.Display offset
13.13.
In display, temperature that is lower than SV is shown for a little time. With this case, display
offset –0.5℃should be done. (Round off numbers to the first decimal point)
14.
14.Remote alarm
14.14.
During normal operation … … … Remote alarm contact is OPEN
During alarm or power failure … … … Remote alarm contact is CLOSE
15.
15.Others
15.15.
MDF-U537: In case of : -18℃≧SV≧–25℃ with PV-SV+5℃≧0
Solenoid valve is: Forced to turn OFF
MDF-U537: In case of : PV-(SV-0.6℃)≧0 with PV
TOP
- PV
BOTTOM
-1>0
Solenoid valve is: Forced to turn OFF
(In PV-(SV-0.6)<0, quit to be forced to turn OFF)
− 33 −
Page 37
−
Wiring diagram
−34
Page 38
−
−35
Page 39
−
−36
Page 40
−
Circuit diagram
<Main PCB>
−37
Page 41
Components on PCB
<Main PCB>
−38−
Page 42
Test data
・MDF-U333 Pull down 40
・MDF-U333 Pull up 40
・MDF-U333 Cooling unit pressure 41
・MDF-U333 Power consumption 41
・MDF-U333 Temperature uniformity and fluctuation 42
・MDF-U537 Pull down 43
・MDF-U537 Pull up 43
・MDF-U537 Cooling unit pressure 44
・MDF-U537 Power consumption 44
・MDF-U537D Pull down 45
・MDF-U537D Pull down 45
・MDF-U537D Pull up 46
・MDF-U537D Cooling unit pressure 47
・MDF-U537D Power consumption 47
・MDF-U537/D Temperature uniformity and fluctuation 48
−39−
Page 43
40
MDF-U333 Pull down
AT35℃,noload
30
20
10
0
-10
Internal temperature [℃]
-20
-30
-40
0246810
Time scale [H]
50Hz60Hz
MDF-U333 Pull up
20
10
0
-10
-20
Internal temperature [℃]
-30
-40
02468
Time scale[H]
AT35℃,無負荷
−40−
Page 44
1600
1400
1200
1000
Pressure [kPa]
MDF-U333 Cooling unit pressure
Pd(50Hz)Ps(50Hz)Pd(60Hz)Ps(60Hz)
AT35℃,no load
800
600
400
200
0
01234567
Time scale [h]
MDF-U333 Power consumption
W(50Hz)W(60Hz)
250
AT35℃,no load
200
150
100
Power consumption[W]
50
0
01234567
Time scale [h]
−41−
Page 45
MDF-U333
Temperature uniformity and fluctuation
(Test condition)
Mains: 1φ100V50/60Hz
AT: 35℃
SV: -25℃・-20℃・-30℃
No loadMeasure with type T sensor
Measuring point:Center point of each container
1
2
3
4
5
(1).SV=-25℃
Measuring point
Top-21.8
1
2
Middle-24.3± 2.0
3( 1/2 center air)
4
Bottom-24.2
Uniformity
5
Max./Min-21.8/-24.6
Range(deg)2.8(±1.4)
Temperature(℃)Fluctuation(℃)
-23.2
-24.6
(2).SV=-20℃
Measuring point
Top-17.1
1
2
Middle-19.1±2.6
3( 1/2 center air)
4
Bottom-19.0
Uniformity
5
Max./Min-17.1/-19.4
Temperature(℃)Fluctuation(℃)
-18.3
-19.4
Range(deg)2.3(±1.15)
(3).SV=-30℃
Measuring point
Top-26.3
1
2
Middle-29.3±1.7
3( 1/2 center air)
4
Bottom-29.1
Uniformity
5
Max./Min-26.3/-29.4
Temperature(℃)Fluctuation(℃)
-28.1
-29.4
Range(deg)3.2(±1.6)
note)These data are reference only.
−42−
Page 46
MDF-U537 Pull down
50Hz60Hz
40
30
20
10
0
-10
Internal temperature[℃]
-20
-30
-40
024681012
Time scale [H]
AT 35℃,no load
MDF-U537 Pull up
AT 35℃,no load
20
10
0
-10
-20
Internal temperature [℃]
-30
-40
0246810
Time scale [H]
−43−
Page 47
2000
1500
1000
Pressure [kPa]
500
0
024681012
MDF-U537Cooling unit pressure
Pd(60Hz)Ps(50Hz)Pd(60Hz)Ps(60Hz)
AT 35℃,no load
Time scale [H]
MDF-U537 Power consumption
W(50Hz)W(60Hz)
300
AT 35℃,no load
200
100
Power consumption[W]
0
024681012
Time scale [H]
−44−
Page 48
40
MDF-U537D Pull down
30
20
10
-10
Internal temperature[℃]
-20
-30
-40
AT35℃,50Hz,no load
Top chamber
Bottom chamber
0
024681012
Time scale[H]
40
30
20
10
-10
Internal temperature[℃]
-20
-30
-40
MDF-U537D Pull down
AT35℃,60Hz,no load
Top chamber
Bottom chamber
0
024681012
Time scale[H]
−45−
Page 49
MDF-U537D Pull up
30
AT35℃,no load
20
10
0
-10
Top chamber
Internal temperature[℃]
-20
-30
-40
0246810
Time scale[H]
Bottom
chamber
−46−
Page 50
2500
2000
1500
1000
Pressure[kPa]
500
0
024681012
MDF-U537D Cooling unit pressure
Pd(50Hz)Ps(50Hz)Pd(60Hz)Ps(60Hz)
AT35℃,no load
Time scale [H]
MDF-U537D Power consumption
W(50Hz)W(60Hz)
500
AT35℃,no load
400
300
200
Power consumption[W]
100
0
024681012
Time scale[H]
−47−
Page 51
MDF-U537 / D temperature uniformity and fuluctuation
Mains : 100V50Hz
AT : 35℃
SV :-20℃・-25℃・-30℃
No load Mesure with type T sensor
Measureing point : Center point of each container /chamber
MDF-U537
1
2
3
4
5
6
7
《 MDF-U537 》
Measuring point
Uniformity [deg.]
Fluctuaation[℃]
MDF-U537D
1
2
8
SV -20℃SV -25℃SV -30℃
1
2
3
4
5
6
7
8
-19.7-24.6-30.8
-20.9-25.7-31.7
-21.1-25.4-31.7
-19.6-24.7-29.4
-19.9-24.4-29.6
-21.2-25.1-30.7
-18.7-22.8-27.8
-19.9-24.1-29.7
2.52.63.9
±2.0 ±1.4 ±1.5
3
5
6
7
8
4
9
《 MDF-U537D 》
Measuring point
1
2
3
4
Uniformity [deg.]
Fluctuaation[℃]
5
6
7
8
9
Uniformity [deg.]
Fluctuaation[℃]
note)These data are reference only
SV -20℃SV -25℃SV -30℃
-17.7-22.4-27.4
-19.9-25.1-30.1
-18.8-24.9-30.8
-19.8-24.9-30.2
2.22.73.4
±3.1±2.1±0.9
-22.4-25.8-29.8
-21.3-25.3-29.5
-20.5-25.8-30.4
-17.9-23.3-28.1
-21.7-26.2-30.4
4.52.52.3
±3.2 ±2.6 ±1.3
−48−
Page 52
<Memorandum>
Page 53
Parts layout
MDF-U537
MDF-U537
MDF-U537MDF-U537
Inside of control panel
Electrical parts on top
Top view of productAccess port for recorder wires
Connecter for buzzer
Connecter for display P.C.B.
Back view of product
Door latch
Inner parts
of door latch
Remote alarm terminalBreaker switch
・Temperature control sensor (MDF-U537)
・Temperature control sensor for upper chamber (MDF-U537D)
・Solenoid valve control sensor (MDF-U537)
・Temperature control sensor for lower chamber (MDF-U537D)
Connecter for tenperature recorder power sourece
−49−
Page 54
User’s manual
・ This section is extracted and printed from User’s Manual.
・ If you find out “Refer to page ●●” in them, this page means not page in service manual
but page in the lower corner of each page in the extract from user’s manual.
This page number is not corresponded with serial number in Service manual.