Sanyo MDF-U537D, MDF-U537, MDF-U333 Service Manual

Service Manual Biomedical freezer
SANYO Electric Biomedical Co., Ltd.
MDF-U333 MDF-U537
MDF-U537D
MDF-U537
Effective models
This service manual is effective following models.
Model name Model code
Voltage and Frequency
MDF-U537 823 168 51 115V 60Hz
823 168 52 220V 50Hz 823 168 53 220V 60Hz 823 168 54 230V 50Hz 823 168 55 240V 50Hz 823 168 56 115V 60Hz 823 168 57 110V 60Hz 824 168 58 230V 50Hz 824 168 59 220V 50Hz 824 168 60 240V 50Hz 825 168 61 825 168 62
MDF-U537D 823 169 51 115V 60Hz
823 169 52 220V 50Hz 823 169 53 220V 60Hz 823 169 54 230V 50Hz 827 169 55 240V 50Hz 823 169 56 240V 50Hz 823 169 57 230V 50Hz 823 169 58 115V 60Hz 823 169 59 110V 60Hz 823 169 60 220V 50Hz 823 169 61 823 169 62
MDF-U333 823 170 51 115V 60Hz
823 170 52 220V 50Hz 823 170 53 220V 60Hz 823 170 54 230V 50Hz 823 170 55 240V 50Hz 823 170 56 110V 60Hz 823 170 57 230V 50Hz 823 170 58 240V 50Hz 823 170 59 220V 50Hz 823 170 60 823 170 61 823 170 62
Contents
Feature 1
Specification MDF-U 333 3
MDF-537 5
MDF-537D 7
Dimensions MDF-333 9
MDF-537 10
MDF-537D 11
Refrigeration circuit 12
Refrigeration circuit principle diagram MDF-U333 15
Refrigeration circuit welding points MDF-333 16
Refrigeration circuit principle diagram MDF-U 537 17
Refrigeration circuit welding points MDF-537 18
Refrigeration circuit principle diagram MDF-U 537D 19
Refrigeration circuit welding points MDF-537D 20
・ Electrical Parts 21
Specification of sensor 24
・ Control specification 25
Wiring diagram 34
Circuit diagram 37
Components on P.C.B. 38
Test data 39
・ Parts loyout 49
User’smanual 50
Features
<MDF-U333>
<MDF-U333>
<MDF-U333><MDF-U333>
HFC refrigerant
Refrigerant circuit contains HFC refrigerant that effect to environment little.
Latest cooling system reduces negative factor to global environment. With ambient temperature at 35, while temperature of inside chamber maintains -30℃.
PCM structure
PCM structure reduces amount of organic solvent.
Temperature control function Unified temperature control range for same categories realized to set -20℃, which
is suitable for reagent preservation.
Specification unified
Unified specification for same categories, which also realized to unify performance for high/low temperature alarm (SV is changeable in the range ±5~±15℃), power failure
alarm, remote alarm terminal, setting memorization by non-volatile memory.
Adopt new microprocessor system for display panel and keypad makes operation easily.
Self diagnosis function
Abnormal condition for temperature sensor is indicated by self-diagnosis system.
Validation
Control panel enables zero “0” adjust for validation.
New door latch mechanism
New door latch is adopted in order to close door surely for reagents.
It is also available to use individual lock mechanism.
Insulation wall thickness
Unit has necessary insulation performance for each temperature range.
Combination of PCM and frame form makes 610mm for frame size commonly.
Option
Internal temperature observation system is option.
Unit has communication function corresponding to it.
−1−
Features
<MDF-U537,537D>
<MDF-U537,537D>
<MDF-U537,537D><MDF-U537,537D>
HFC refrigerant
Refrigerant circuit contains HFC refrigerant that effect to environment little.
Latest cooling system reduces negative factor to global environment. With ambient temperature at 35, while temperature of inside chamber maintains -30℃.
PCM structure
PCM structure reduces amount of organic solvent.
Temperature control function New cooling system realizes to set -20, which is suitable for reagent preservation.
Specification unified
Unified specification for same categories, which also realized to unify performance for high/low temperature alarm (SV is changeable in the range ±5~±15℃), power failure
alarm, remote alarm terminal, setting memorization by non-volatile memory.
Adopt new microprocessor system for display panel and keypad makes operation easily.
Self diagnosis function
Abnormal condition for temperature sensor is indicated by self-diagnosis system.
Validation
Control panel enables zero “0” adjust for validation.
New door latch mechanism
New door latch is adopted in order to close door surely for reagents.
It is also available to use individual lock mechanism.
Draw box for reagent preservation
100mm for depth of box makes things easily to arrange and take in & out.
(Adopted in bottom chamber of MDF-U537/U537D)
Compressor protection function
Sensor detects compressor’s warming up to control unit running.
2

Specification

MDF-U333

Items Specification
Model description Biomedical freezer
Model Number MDF-U333
Outer dimension Inner dimension W 490 × D 485 × H 1290 mm
Effective chamber capacity 274 lit.
Exterior Zinc galvanized steel w/powder polyester coating (PCM)
Interior Polystyrene resin
Insulation Rigid polyurethane foam
Outer door 1 pc, Zinc galvanized steel w/powder polyester coating (PCM)
Latch 1 pc
Door lock 1 pc
Basket
Accessories
Others Defrosting spatula 1 pc, key 1 set, 4 card holders
Caster 4 pcs
Leveling legs 2 pcs (with caster) Access hole φ 30mm. 1 place in left side.
Weights 81 Kg
Evaporator Tube on sheet type (function both evaporator and shelf)
Condenser Wire and tube type
Compressor Hermetic type 225W
W 614 × D 733 × H 1620 mm
(Max)
(L) 4 pcs Inner dimension: W446×D369×H220 mm (S) 1 pc (for bottom column) Inner dimension:
W
487×D221×H155mm
Refrigerant
Cooling performance -30 (internal 1/2h air temperature, AT:35℃, no load)
Internal temperature control range
Power supply Local voltage
HFC refrigerant
R-134a
-20~-30 (AT:5℃~35℃、no load)
3
MDF-U333
Items Specification
Temperature controller
Microprocessor controlled system with non-volatilized memory. Temp. set range : -18℃~-40℃
Temperature sensor Thermistor
Temperature display
High/Low Temperature
Alarm
Power failure
Remote alarm contact
LED digital display (green colored) by 1℃. Range : -50℃~50℃ Changeable SV±5℃~15℃. (Initial SV±10℃) ALARM lamp brinks and buzzer sounds intermittently
(about 15 min. delay) Remote alarm contact: turns ON during power failure alarm. ALARM lamp brinks, buzzer sounds intermittently, remote alarm contact turns ON during power failure alarm. Remote alarm terminal 3P: contact capacity DC30v, Max2A, N.O. –COM. (During power failure, temperature alarm: ON) ALARM lamp BUZZER: Buzzer key ALARM TEST: Alarm test key
Control panel
Operation panel
SET: To switch the SV and PV
: To shift the set digit : To change the SV
DEF: Defrost key
Rear
Key lock function
Power switch (Breaker switch) Battery switch: Power failure alarm switch
Press key for five seconds L0: Unlock L1: Lock When each sensor is in fault,
Self diagnosis function
Error code and internal temperature are displayed alternately. Remote alarm contact turns ON with buzzer sounds.
Power switch Breaker switch
Compressor protection Overload relay
Chart recorder: MTR-G85
Option
Mounting kit: MPR-S7(for bottom mounting) : MDF-S740T(for top mounting)
4
Specification

MDF-U537

Items Specification
Model description Biomedical freezer
Model Number MDF-U537 Outer dimension (Max) W 804 × D 772 × H 1802 mm Inner dimension W 658 × D 607 × H 1272 mm
Effective chamber capacity 482 lit.
Exterior Zinc galvanized steel w/powder polyester coating (PCM)
Interior Polystyrene resin
Insulation Rigid polyurethane foam
Outer door
Latch 2 pcs
Door lock 1 pc
Basket
Accessories
Others Defrosting spatula 1 pc, key 1 set, 14 card holders
Caster 4 pcs
Zinc galvanized steel w/powder polyester coating (PCM) 2 pcs (upper and lower compartment)
(M) 6 pcs (upper compartment)
W
Inner dimension:
290×D536×H136 mm
(S) 8 pcs (lower compartment)
W
Inner dimension:
290×D536×H100mm
Leveling legs 2 pcs (with caster) Access hole φ 30mm. 1 place on rear.
Weights 131 Kg
Evaporator Tube on sheet type (function both evaporator and shelf)
Condenser Wire and tube type
Compressor Hermetic type 350W
HFC refrigerant
Refrigerant
R-407D
Cooling performance -30 (internal 1/2h air temperature, AT:35℃, no load)
Internal temperature control range
-20~-30 (AT:5℃~35℃、no load)
Power supply Local voltage
5
MDF-U537
Items Specification
Temperature controller
Microprocessor controlled system with non-volatilized memory. Temp. set range : -18℃~-40℃
Temperature sensor Thermistor 2P
Temperature display
High/Low Temperature
Alarm
Power failure
Remote alarm contact
LED digital display (green colored) by 1℃. Range : -50℃~50℃ Changeable SV±5℃~15℃. (Initial SV±10℃) ALARM lamp brinks and buzzer sounds intermittently
(about 15 min. delay) Remote alarm contact: turns ON during power failure alarm. ALARM lamp brinks, buzzer sounds intermittently, remote alarm contact turns ON during power failure alarm. Remote alarm terminal 3P: contact capacity DC30v, 2A, N.O. –COM. (During power failure, temperature alarm: ON) ALARM lamp BUZZER: Buzzer key ALARM TEST: Alarm test key
Control panel
Operation panel
SET: To switch the SV and PV
: To shift the set digit : To change the SV
DEF: Defrost key
Rear
Key lock function
Power switch (Breaker switch) Battery switch: Power failure alarm switch
Press key for five seconds L0: Unlock L1: Lock When each sensor is in fault,
Self diagnosis function
Error code and internal temperature are displayed alternately. Remote alarm contact turns ON with buzzer sounds.
Power switch Breaker switch
Overload relay Detect temperature near compressor by thermistor sensor to control
Compressor protection
compressor during overload. Alarm: Error code and internal temperature are displayed alternately. Remote alarm contact turns on with buzzer sounds. Chart recorder: MTR-G85
Option
Mounting kit: MPR-S7(for bottom mounting) : MDF-S740T(for top mounting)
6
Specification

MDF-U537D

Items Specification
Model description Biomedical freezer
Model Number MDF-U537D
Outer dimension
Inner dimension
W 804 × D 772 × H 1802 mm
(Max)
Upper chamber W 658 × D 607 × H 589 mm Lower chamber W 658 × D 607 × H 603 mm
Effective chamber capacity 452 lit.
Exterior Zinc galvanized steel w/powder polyester coating (PCM)
Interior Polystyrene resin
Insulation Rigid polyurethane foam
Outer door
Zinc galvanized steel w/powder polyester coating (PCM) 2 pcs (upper and lower compartment)
Latch 2 pcs
Door lock 1 pc
(M) 6 pcs (upper compartment)
W
290×D536×H136 mm
W
290×D536×H100mm
Accessories
Basket
Inner dimension: (S) 8 pcs (lower compartment) Inner dimension:
Others Defrosting spatula 1 pc, key 1 set, 14 card holders
Caster 4 pcs
Leveling legs 2 pcs (with caster) Access hole φ 30mm. 2 places on rear.
Weights 136 Kg
Evaporator Tube on sheet type (function both evaporator and shelf)
Condenser Wire and tube type
Compressor Hermetic type 350W
HFC refrigerant
Refrigerant
R-407D
Cooling performance -30 (internal 1/2h air temperature, AT:35℃, no load)
Internal temperature control range
-20~-30 (AT:5℃~35℃、no load)
Power supply Local voltage
7
MDF-U537D
Items Specification
Temperature controller
Microprocessor controlled system with non-volatilized memory. Temp. set range : -18℃~-40℃
Temperature sensor Thermistor 2P (upper and lower compartment)
LED digital display (green colored) by 1℃.
Temperature display
Range : -50℃~50℃ Selectable to display upper or lower compartment by switch.
High/Low Temperature
Alarm
Power failure
Remote alarm contact
Changeable SV±5℃~15℃. (Initial SV±10℃) ALARM lamp brinks and buzzer sounds intermittently (about 15 min. delay) Remote alarm contact: turns ON during power failure alarm. ALARM lamp brinks, buzzer sounds intermittently, remote alarm contact turns ON during power failure alarm. Remote alarm terminal 3P: contact capacity DC30v, 2A, N.O. –COM. (During power failure, temperature alarm: ON) ALARM lamp BUZZER: Buzzer key ALARM TEST: Alarm test key
Control panel
Operation panel
SET: To switch the SV and PV
: To shift the set digit : To change the SV
DEF: Defrost key TOP/BOTTOM: Switch key to display upper or lower compartment.
Rear
Key lock function
Power switch (Breaker switch) Battery switch: Power failure alarm switch
Press key for five seconds L0: Unlock L1: Lock When each sensor is in fault,
Self diagnosis function
Error code and internal temperature are displayed alternately. Remote alarm contact turns ON with buzzer sounds.
Power switch Breaker switch
Overload relay Detect temperature near compressor by thermistor sensor to control
Compressor protection
compressor during overload. Alarm: Error code and internal temperature are displayed alternately. Remote alarm contact turns on with buzzer sounds. Chart recorder: MTR-G85
Option
Mounting kit: MPR-S7(for bottom mounting) : MDF-S740T(for top mounting)
8

Dimensions

<<<<MDF-U333>>>>

−9−
Dimensions

<<<<MDF-U537>>>>

−10−
Dimensions

<<<<MDF-U537D>>>>

−11−

Refrigeration circuit

MDF-U333 Unit:mm

Parts description Specification
Evaporator
Dehydrator
Capillary tube
Condenser assembly A
Condenser assembly B
Frame pipe
Evacuation and vacuum pressure
Refrigerant and oil
Refrigeration circuit spec. and welding points
Type Tube on sheet Material Aluminum (φ 8.0)
Type of molecular sieves
A-XH9134 stamped
Charged q’ty 8±1g Resistance 0.686MPaG (7Kgf/㎠ G
Dimension φ 1.8×(φ 0.65)×2000mm Resistance indicator color White Type Wire and tube Dimension
w
480×P40×30 columns
Type Wire and tube Dimension
w
480×P40×30 columns
Note: Lower 2 columns are used in desuper circuit.
Material C1220T-1/2H Dimension
φ 4.0×
0.5
Evacuation time: over than 20min. by 2 ports (high pressure and low pressure side). (N2 exchange) When charged, vacuum pressure is lower than 1.0torr pump capacity 300L/min
Refrigerant : R-134a Charged q’ty : 226±5g Oil : CF32 Charged q’ty : 230±10cc Oil additive: n-pentane Charged q’ty : 14g
Refer to following pages
−12−

Refrigeration circuit

MDF-U537 Unit:mm

Parts description Specification
Evaporator
Upper (1) Lower (1) Bottom (3)
Dry core
Type 9lines×1column×5pcs
Material Aluminum (φ 8.0)
Type D-S032T Manufacturer SANYO Solid core 50cm
Capillary tube (Total number: 2)
Resistance 0.725MPaG (7.4Kgf/㎠ G) Dimension φ 1.8×L4500 Resistance indicator color Blue
Condenser assembly front (Total number: 4)
Frame pipe
Type Wire and tube Dimension
Material C1220T Dimension
Evacuation and vacuum pressure
Evacuation time: over than 20min. by 2 ports (high pressure and low pressure side). (N2 exchange) When charged, vacuum pressure is lower than 1.0torr pump capacity 300L/min
Refrigerant and oil
Refrigerant : R-407D Charged q’ty : 301±5g Oil : CF32 Charged q’ty : 280±10cc Oil additive: n-pentane Charged q’ty: 19g
Refrigeration circuit spec. and welding points
Refer to following pages
3
w
480×P40×6 columns
L
φ 4.0×
0.5×
7470
−13−
Refrigeration circuit

MDF-U537D Unit:mm

Parts description Specification
Evaporator assembly
Upper (1)
Type 9lines×1column×5pcs
Lower (1) Bottom (2)
Material Aluminum (φ 8.0)
Lower entrance (1)
Dry core
Type D-S032T Manufacturer SANYO Solid core 50cm
Capillary tube
Upper compartmnt Lower compartmnt Resistance 10.3Kg/G 9.9Kg/G Dimension φ 1.8×L4000 φ 1.8×L3000 Resistance indicator color Yellow Red
Condenser assembly front (Total number: 4)
Frame pipe
Type Wire and tube Dimension
Material C1220T Dimension
Evacuation and vacuum pressure
Evacuation time: over than 20min. by 2 ports (high pressure and low pressure side). (N2 exchange) When charged, vacuum pressure is lower than 1.0torr pump capacity 300L/min
Refrigerant and oil
Refrigerant : R-407D Charged q’ty : 357±5g Oil : CF32 Charged q’ty : 280±10cc Oil additive: n-pentane Charged q’ty : 23g
Refrigeration circuit spec. and welding points
Refer to following pages
3
w
480×P40×6 columns
φ 4.0×
0.5×L7470
−14−
Refrigeration circuit principle diagram

<<<<MDF-U333>>>>

−15−
Refrigeration circuit welding points

<<<<MDF-U333>>>>

−16−

Refrigeration circuit principle diagram

<<<<MDF-U537>>>>

−17−
Refrigeration circuit welding points

<<<<MDF-U537>>>>

−18−
Refrigeration circuit principle diagram

<<<<MDF-U537D>>>>

−19−
Refrigeration circuit welding points

<<<<MDF-U537D>>>>

−20−

Electrical Parts

MDF-U333

Compressor
Starting relay
Overload relay
Starting capacitor Rating 100μF,160VAC 100μF,160VAC 30μF,300VAC Running capacitor Rating 25μF,220VAC 20μF,220VAC 5μF,400VAC
Temp. control relay
Switching power supply
Temperature sensor
P.C.B.
Battery
Battery switch
Breaker switch
Type C-2SN220L0W C-2SN220L2W C-2SN220L5W
Code 80771120 80771122 80771125
Rated voltage (50/60Hz) 100 / 100 , 110 -- / 115 220,240 / 220 Winding resistance C-S(Aux) 5.8Ω 7.2Ω 30.6Ω C-R(Main) 1.6Ω 2.0Ω 7.8Ω
Type AMVL-180A AMVL-180A AMVL-300A Pick up voltage(50Hz) AC91V〜111V AC91V〜111V AC185V〜217V (60Hz) AC108〜128V AC108〜128V AC215〜232V Drop out voltage(50Hz) AC30〜65V AC30〜65V AC60〜120V (60Hz) AC30〜75V AC30〜75V AC69〜132V
Type P24KU P19KU P10MU
Action to the temp. (No current)
OFF 150±10 150±10 130±8
ON 85±10 85±10 69±10
Action to the current (AT25℃) 24.0A 19.5A 10A
Operation time 615 seconds 615 seconds 615 seconds
Non-Action to the temp. (80℃)
Non-Action
Action 8.0A within 15 minutes 7.4A within 15 minutes 3.4A within 15 minutes
Type G4F-11123T G4F-11123T G4F-11123T
Contact capacity AC220V,20A AC220V,20A AC220V,20A
Coil DC12V DC12V DC12V
Type ZWS10-12/J ZWS10-12/J ZWS10-12/J
Input
Rated output DC12V,0.85A DC12V,0.85A DC12V,0.85A
Type 502AT-1 502AT-1 502AT-1 Rating 5kΩ,25℃ 5kΩ,25℃ 5kΩ,25℃
Type DF-333 DF-333 DF-333
Type 5HR-AAC 5HR-AAC 5HR-AAC
Rating 6V 1100MAH 6V 1100MAH 6V 1100MAH
Type SLE6A2-5 SLE6A2-5 SLE6A2-5
Rating AC250V 4A AC250V 4A AC250V 4A
Type BAM215131 BAM215131 BAM215131
Rating AC250V 15A AC250V 15A AC250V 15A
AC110V,60Hz AC115V,60Hz AC220-240V , 50/60Hz
6.1A more than 30 minutes
AC100-240V 50/60Hz,0.3A
5.4A more than 30 minutes
AC100-240V 50/60Hz,0.3A
2.5A more than 30 minutes
AC100-240V 50/60Hz,0.3A
−21−
g
Electrical Parts

MDF-U537

Compressor Type C-2SN350L0R C-2SN350L5S C-2SN350L5S
Starting relay Type AMVL-180A AMVL-300A AMVL-300A
Overload relay Type P18AU P10MU P10MU
Starting capacitor Rating 100μF,160VAC 30μF,300VAC 30μF,300VAC Running capacitor Rating 25μF,220VAC 5μF,400VAC 5μF,400VAC Temp. control relay Type G4F-11123T G4F-11123T G4F-11123T
Solenoid valve Type G2R-1A-T G2R-1A-T G2R-1A-T control relay Contact capacity AC250V,10A AC250V,10A AC250V,10A
Switching power supply Type ZWS10-12/J ZWS10-12/J ZWS10-12/J
Temperature sensor Type 502AT-1 502AT-1 502AT-1
Solenoid valve control Type 502AT-1 502AT-1 502AT-1 sensor Rating 5kΩ,25℃ 5kΩ,25℃ 5kΩ,25℃ Compressor protect sensor Type 502AT-1 502AT-1 502AT-1
P.C.B. Type DF-537 DF-537 DF-537 Battery Type 5HR-AAC 5HR-AAC 5HR-AAC
Battery switch Type SLE6A2-5 SLE6A2-5 SLE6A2-5
Breaker switch Type BAM215131 BAM215131 BAM215131
Solenoid valve coil Type NEVAC120V NEVAC220V NEVAC240V
Condensin
fan motor
Code 807764201 807763251 807763251 Rated voltage (50/60Hz) Winding resistance C-S(Aux) 5.8Ω 30.6Ω 30.6Ω C-R(Main) 1.6Ω 7.8Ω 7.8Ω
Pick up voltage(50Hz) AC91V〜111V AC185V〜217V AC185V〜217V (60Hz) AC108〜128V AC215〜232V AC215〜232V Drop out voltage(50Hz) AC30〜65V AC60〜120V AC60〜120V (60Hz) AC30〜75V AC69〜132V AC69〜132V
Action to the temp. (No current)
OFF 120±8 130±8 130±8
ON 69±10 69±10 69±10
Action to the current (AT25℃) 18.5A 10A 10A
Operation time 615 seconds 615 seconds 615 seconds
Non-Action to the temp. (80℃)
Non-Action
Action 5.5A within 15 minutes 3.4A within 15 minutes 3.4A within 15 minutes
Contact capacity AC220V,20A AC220V,20A AC220V,20A Coil DC12V DC12V DC12V
Coil DC12V DC12V DC12V
Input
Rated output DC12V,0.85A DC12V,0.85A DC12V,0.85A
Rating 5kΩ,25℃ 5kΩ,25℃ 5kΩ,25℃
Rating 5kΩ,25℃ 5kΩ,25℃ 5kΩ,25℃
Rating 6V 1100MAH 6V 1100MAH 6V 1100MAH
Rating AC250V 4A AC250V 4A AC250V 4A
Rating AC250V 15A AC250V 15A AC250V 15A
Rating 120V 50/60Hz 220V 50/60Hz 240V 50/60Hz Type FU2-A051B1MP FU2-A051B5MP FU2-A051B5MP Rating AC110-115V 50/60Hz 220-240V 50/60Hz 220-240V 50/60Hz Thermal fuse 130 130 130
AC110-115V,60Hz AC220V,50/60Hz AC230V-240V,50Hz
4.0A more than 30 minutes
AC100-240V 50/60Hz,0.3A
2.5A more than 30 minutes
AC100-240V 50/60Hz,0.3A
2.5A more than 30 minutes
AC100-240V 50/60Hz,0.3A
−22−
g
Electrical Parts

MDF-U537D

Compressor Type C-2SN350L0R C-2SN350L5S C-2SN350L5S
Starting relay Type AMVL-180A AMVL-300A AMVL-300A
Overload relay Type P18AU P10MU P10MU
Starting capacitor Rating 100μF,160VAC 30μF,300VAC 30μF,300VAC Running capacitor Rating 25μF,220VAC 5μF,400VAC 5μF,400VAC Compressor relay Type G4F-11123T G4F-11123T G4F-11123T
Temp. control relay Type G2R-1A-T G2R-1A-T G2R-1A-T
Switching power supply Type ZWS10-12/J ZWS10-12/J ZWS10-12/J
Temperature sensor Type 502AT-1 502AT-1 502AT-1
Solenoid valve control Type 502AT-1 502AT-1 502AT-1 sensor Rating 5kΩ,25℃ 5kΩ,25℃ 5kΩ,25℃ Compressor protect sensor Type 502AT-1 502AT-1 502AT-1
P.C.B. Type DF-537D DF-537D DF-537D Battery Type 5HR-AAC 5HR-AAC 5HR-AAC
Battery switch Type SLE6A2-5 SLE6A2-5 SLE6A2-5
Breaker switch Type BAM215131 BAM215131 BAM215131
Solenoid valve coil Type NEVAC120V NEVAC220V NEVAC240V
Condensin
(2 pcs.) Rating 120V 50/60Hz 220V 50/60Hz 240V 50/60Hz
fan motor
Code 807764201 807763251 807763251 Rated voltage (50/60Hz) Winding resistance C-S(Aux) 5.8Ω 30.6Ω 30.6Ω C-R(Main) 1.6Ω 7.8Ω 7.8Ω
Pick up voltage(50Hz) AC91V〜111V AC185V〜217V AC185V〜217V (60Hz) AC108〜128V AC215〜232V AC215〜232V Drop out voltage(50Hz) AC30〜65V AC60〜120V AC60〜120V (60Hz) AC30〜75V AC69〜132V AC69〜132V
Action to the temp. (No current)
OFF 120±8 130±8 130±8
ON 69±10 69±10 69±10
Action to the current (AT25℃) 18.5A 10A 10A
Operation time 615 seconds 615 seconds 615 seconds
Non-Action to the temp. (80℃)
Non-Action
Action 5.5A within 15 minutes 3.4A within 15 minutes 3.4A within 15 minutes
Contact capacity AC220V,20A AC220V,20A AC220V,20A Coil DC12V DC12V DC12V
Contact capacity AC250V,10A AC250V,10A AC250V,10A Coil DC12V DC12V DC12V
Input
Rated output DC12V,0.85A DC12V,0.85A DC12V,0.85A
Rating 5kΩ,25℃ 5kΩ,25℃ 5kΩ,25℃
Rating 5kΩ,25℃ 5kΩ,25℃ 5kΩ,25℃
Rating 6V 1100MAH 6V 1100MAH 6V 1100MAH
Rating AC250V 4A AC250V 4A AC250V 4A
Rating AC250V 15A AC250V 15A AC250V 15A
Type FU2-A051B1MP FU2-A051B5MP FU2-A051B5MP Rating AC110-115V 50/60Hz 220-240V 50/60Hz 220-240V 50/60Hz Thermal fuse 130 130 130
AC110-115V,60Hz AC220V,50/60Hz AC230V-240V,50Hz
4.0A more than 30 minutes
AC100-240V 50/60Hz,0.3A
2.5A more than 30 minutes
AC100-240V 50/60Hz,0.3A
2.5A more than 30 minutes
AC100-240V 50/60Hz,0.3A
−23−
Loading...