Sanyo MCO-80IC User Manual

Page 1
Reach-in CO2 Incubator MCO-80IC
SANYO’s large capacity, Reach-in CO2 incubator has
the capacity and flexibility to grow with your culturing
needs while providing a precise and repeatable
temperature, humidity and CO
environment.
2
Model: MCO-80IC, 30.1 cu.ft. (851L)
Page 2
MCO-80IC Reach-in CO2 Incubator
The MCO-80IC is ideal for culturing large volumes of patient samples, performing short-term studies, and
working with large volume cell culture apparatus. It includes SANYO’s exclusive incubator technologies such as
inCu saFe™ interiors, UV decontamination option, IR CO2 sensor with P.I.D. control, and features exceptionally
low CO2 gas consumption.
Superior CO2 and Temperature Control
• Infrared (IR) CO2 sensor with P.I.D. microprocessor
control and forced air circulation system delivers fast CO recovery characteristics.
gas consumption rates, less
2
Usability
• 30.1 cu.ft. (851L) large capacity cabinet allows
flexibility in usage.
• Cabinet will accommodate a roller bottle apparatus, 5 bottles wide x 7 bottles high (requires mounting ramp kit, MCO-80RBS).
• Full view, double paned glass door allows easy
observation of cultured samples.
• Large LED digital display and keypad for greater
visibility and ease of set-up.
• Exceptionally low CO
than half of similar competitive units.
• P.I.D. temperature control with deviation of +/- 0.1°C
Anti-contamination Measures
• Interior, plenums and shelving constructed of SANYO’s
exclusive inCu saFe™, germicidal, copper-enriched stainless steel.
• Optional UV sterilization system for humidity reservoir
• Extra heaters positioned on the outer glass door to
eliminate condensation.
Industry Leading Warranty
• The MCO-80IC incubator carries SANYO’s world class
warranty of 3 years parts and labor, the best in the industry.
2
Easy to reach control panel
Adjustable inCu saFe™ Germicidal
Shelves & Interior
Page 3
No Door
Openings
per day
2 x 30 Second
Door Openings
per day
2 x 60 Second
Door Openings
per day
No Door
Openings
per day
2 x 30 Second
2 x 60 Second
CO2 Consumption
280 Liters/day
440 Liters/day 457 Liters/day
597 Liters/day
728 Liters/day 752 Liters/day
30 kg CO2 Cylinder
Retention Time*
㪁㪫㪼㫊㫋㩷㪚㫆㫅㪻㫀㫋㫀㫆㫅㫊㪑㩷㪩㫆㫆㫄㩷㪫㪼㫄㫇㪅㪔㪉㪇㷄㪃㪫㪼㫄㫇㪅㩷㪪㪭㪔㪊㪎㷄㪃㪚㪦2㩷㪪㪭㪔㪌㩼 㪁㩷㪘㫃㫃㩷㫍㪸㫃㫌㪼㫊㩷㪸㫉㪼㩷㪸㪺㫋㫌㪸㫃㩷㫋㪼㫊㫋㩷㫍㪸㫃㫌㪼㫊㩷㪽㫆㫉㩷㫉㪼㪽㪼㫉㪼㫅㪺㪼㩷㫆㫅㫃㫐㩷㪸㫅㪻㩷㪺㪸㫅㫅㫆㫋㩷㪹㪼㩷㪾㫌㪸㫉㪸㫅㫋㪼㪼㪻㩷㫀㫅㩷㫆㫇㪼㫉㪸㫋㫀㫆㫅㪅
SANYO MCO-80IC
Competitive Model
60 days
39 days37 days 28 days 23 days 23 days
Door Openings
per day
Door Openings
per day
Interior Laminar Air Flow Plenums
CO2 Recovery After a 60 second Door Opening
for the SANYO MCO-80IC
Temperature Recovery After a 30 Second Door
Opening for the SANYO MCO-80IC
5%
4%
3%
2%
1%
37°C
36°C
35°C
34°C
33°C
10 min. 20 min.
10 min. 20 min.
CO2 Recovery After a 60 second Door Opening
for the SANYO MCO-80IC
CO
2 Recovery After a 60 second Door Opening
for the Competitive Model
Temperature Recovery After a 30 Second Door
Opening for the SANYO MCO-80IC
5%
4%
3%
2%
1%
37°C
36°C
35°C
34°C
33°C
5%
4%
3%
2%
1%
10 min. 20 min.
10 min. 20 min.
10 min. 20 min.
CO2 Recovery After a 60 second Door Opening
for the SANYO MCO-80IC
5%
4%
3%
2%
1%
10 min. 20 min.
Horizontal Laminar Airflow System
SANYO’s reach-in incubator’s cross-shelf directed air flow system promotes
optimum temperature uniformity throughout the chamber and contributes to
quick temperature recovery after door openings. Utilizing SANYO’s exclusive
inCu saFe™ chamber material in the perforated side plenums helps minimize
contamination concerns and direct positive and negative pressure air flow.
Optional UV Sterilization and Humidity Selection
SANYO’s patented and
laboratory proven Safe Cell
UV™ sterilization system
is employed to sterilize the
humidifying water reservoir and help
eliminate contamination concerns. The
unit can be set to both nominal and
high humidity setpoints.
Humidity reservoir heaters are
located on the outside walls of the
reservoir and are not susceptible to
corrosion and scaling from water as
competitive systems are. An optional
auto ll secondary tank (Model MCO-
80AS) system is also available to
ensure continuous water supply to
the humidity reservoir. This system
employs a large tank (4.8 gallons),
with electronic water level sensor and
autofill solenoid valve.
Faster CO2 Recovery &
Lower CO
SANYO’s large capacity reach-in incubator was designed specifically for critical
applications in pharmaceutical, biotechnology and clinical investigation.
Large chamber capacity applications require special consideration of gas usage and recovery. SANYO’s proprietary IR sensor with P.I.D. CO2 control algorithm are paramount to the industry’s best design. An optional inner door system (MCO-80ID) is also available to enhance these results further.
SANYO MCO-80IC
Consumption
2
Competitor
Page 4
Specifications
Mode l
MCO-80IC
Exterior dimensions
W38.8.” x D33.6” x H80.3” (W986 X D853 X H2040m m)
Interior dimensions
W31.7” x D27.3.” x H60.0” (W806 X D693 X H1524m m)
Exterior cabinet
Electrogalvanized steel (Acrylic resin baking painted finish)
Interior volume
30.1cu.ft ( 851 L)
Interior cabinet
Copper alloy Stainless Steel (except humidifying reservoir)
External door
Double paned glass door
with outer door latch
Inner door system (option)
Available as an option (Polycarbonate, 5 doors set ), Model MCO-80ID
Shelves
Copper alloy Stainless Steel, 5 standard
Dimensions W30.5” x D25.9” x H0.4” (W776 X D659 X H10m m)
Load Capacity: 30kg (66 lbs.) per shelf
Access ports
Φ40 mm(1.57”), 2 locations (Right and left sides) with silicone rubber stopper
Humidifying reservoir
Integrated with an inner chamber (Stainless Steel, SUS304)
Front accessible drainage valve
Humidifying water sterilization system (option)
UV light over the humidifying reservoir (available as an option), Model MCO-80UVS
Heating method
Heater with fan air circulation, Cross shelf laminar air flow
Temp. control system
Microprocessor P.I.D. with thermistor sensor, Temp. fluctuatio
n range ±0.1ºC
Temp. display
Digital display (0.1ºC increments)
Ambient temperature condition
20ºC to 35ºC
Temp. controllable range
(Ambient temperature + 5
o
C) to 50ºC
Temp. distribution
±0.5ºC (AT䇭25ºC, SV 37ºC, CO
2
5%, No load, 9 point measurement)
Temp. fluctuation margin
±0.1㷄 (AT䇭25㷄, SV 37ºC,CO
2
5%, No load)
Humidifying method
Heated vaporization with water in humidity pan (High humidity mode available)
Chamber humidity
Over 80%RH (High humidity mode: over 90%RH)
High humidity mode
Selectable by function mode (for over 90%RH)
Humidifying water retention days (Reference)
15 days (Reference)
(Under a condition of AT20ºC, SV 37ºC, CO
2
5%, no load, no door opening, and 80%RH)
Water drainage
Drainage valve (lower side of front frame)
Drainage into a drainage tray/bottle (provided)
Water Fill
Water fill located at the front side of interior bottom
Automatic water supply system (Option)
Water supply connection (chamber bottom rear side)
Tank connection only (Direct connection with public water line is
not advised),
Solenoid valve opens in response to low water level sensor
Overflow
Drain outlet on front lower side
CO
2
gas control
Microprocessor P.I.D. with infrared sensor
CO
2
level display
Digital display (0.1% increments)
CO
2
gas range
0% to 20%
CO
2
level fluctuation range
±0.15%
AT 25ºC,SV 37ºC, CO
2
5%, no load
CO
2
gas inlet pressure
15 psi (0.1 MPa) at injection
Net unit weight
605 lbs. (275 kg.)
Inner circulation system
Forced air circulation, Horizontal laminar cross shelf airflow
Power requirements
115V, 20 Amps, 60 Hz, NEMA 5-20P Plug
Options
Built-in type gas automatic switcher kit (MCO-80GC)
Roller bottle apparatus mounting ramp kit (MCO-80RBS)
Inner Door Kit
MCO-80ID
5 partition doors, made with acrylate resin
UV sterilization system (MCO-80UVS)
Sterilization of humidifying water resevoir only
Automatic water supply system kit (MCO-80AS)
Includes control box, water tank, water supply hose, 18L (4.8 gallon) tank
Additional shelf (MCO-80ST)
4 to 20 mA analog output module (MCO-420MA)
Interior/Exterior convenience receptacle
Interior Duplex: Vapor Proof. 115V, 3 Amp. Max. Rating - Exterior Duplex: 115V, 1Amp. Max. Rating
Remote alarm contacts
N.O. (normally open) and N.C. (normally closed) contacts included, rating DC 30V, 2A
CO
2
tank regulator, (MCO-100L)
Alarms
High/Low temp. alarm, CO
2
density alarm, Upper limit temp. alarm, Door ajar alarm
My life. My work. My choice.
www.sanyobiomedical.com
A Division of SANYO North America Corporation
1300 Michael Drive, Wood Dale, IL 60191 USA
Toll Free USA 800-858-8442 • Fax 630-238-0074
1- 300 Applewood Crescent, Concord, Ontario L4K 5C7
SANYO Commercial Solutions
www.sanyobiomedical.com
905-760-4025 • Fax 905-760-9945
SANYO Canada, Inc.
LR-80IC-1.08
Loading...