Sanyo MCO-19M UV Installation Manual

Page 1
INSTRUCTION MANUAL
MCO-19M(UVH) MCO-19M(UV)
Multi Gas Incubator
MCO-19M
Page 2
CONTENTS
INTRODUCTION P. 3
PRECAUTIONS FOR SAFE OPERATION P. 4
LABELS ON THE INCUBATOR P. 8 ENVIRONMENTAL CONDITIONS P. 8 INCUBATOR COMPONENTS P. 9
Control panel and keypad P. 12
Remote alarm terminals P. 13 INSTALLATION SITE P. 14 INSTALLATION P. 15 Connection of CO Connection of N Connection of gas injection nozzle P. 18 Using the inner door and gastight split doors P. 18 PREVENTING CONTAMINATION P. 19 PRECAUTIONS FOR CULTURES P. 20
Using the unlock key P. 21
CORRECT OPERATION P. 21 LCD PANEL P. 22 BASIC OPERATIONS ON CONTROL PANEL P. 24 BASIC PARAMETERS Setting the chamber temperature, CO
Setting the key lock  P. 26 Releasing the key lock P. 27
Setting the key lock password P. 28 Automatic N Setting the upper limit alarm temperature P. 32
ALARM PARAMETERS  P. 33
UV LAMP PARAMETERS P. 34 Using the UV lamp P. 34
Precautions when using the UV lamp P. 35
Setting the UV lamp ON period P. 36 Lighting the UV lamp for 24 hours P. 37 OTHER PARAMETERS Setting the date, time, and log interval P. 38 Initial settings (LCD/DAQ parameters) P. 40 DISPLAYING THE LOG P. 41 Transferring data P. 43 WATER LEVEL SENSOR P. 44 ROUTINE MAINTENANCE Cleaning the chamber and inner attachments P. 45 Filling the humidifying pan P. 48 H
DECONTAMINATION P. 49
2O2
H Precautions when handling H ALARMS, SAFETY, AND SELF-DIAGNOSIS P. 55 CALIBRATION Temperature/CO TROUBLESHOOTING P. 60 DISPOSING OF THE MULTI-GAS INCUBATOR P. 62 AUTOMATIC CO AUTOMATIC CO STACKING INCUBATORS P. 72
decontamination P. 51
2O2
gas cylinder P. 16
2
(or O2) gas cylinder P. 17
2
and CO2 density P. 25
2
(or O2) gas cylinder changeover P. 30
2
decontamination reagent P. 54
2O2
calibration P. 58
2/O2
CYLINDER CHANGEOVER P. 67
2
AND O2 DENSITY CALIBRATION P. 69
2
1
Page 3
CONTENTS
SPECIFICATIONS P. 74 PERFORMANCE P. 75 SAFETY CHECK SHEET P. 76
2
Page 4
INTRODUCTION
Read this manual carefully before using the Product and follow the instructions for safety operation.
Sanyo disavows any responsibility for safety if the Product is used for other than the intended use or
used with any procedures other than those given in this manual.
Keep this manual in a suitable place so that it can be referred to as necessary.
The contents of this manual are subject to change without notice for improvement of performance or
functions.
Contact a Sanyo sales representative or agent if any page of the manual is lost or the page order is
incorrect.
Contact a Sanyo sales representative or agent if any point in this manual is unclear or if there are any
inaccuracies.
No part of this manual may be reproduced in any form without the expressed written permission of
Sanyo.
CAUTION
SANYO guarantees the product under certain warranty conditions. SANYO in no way shall be responsible for any loss of content or damage of content.
3
Page 5
PRECAUTIONS FOR SAFE OPERATION
It is imperative that the user complies with this manual as it contains important safety advice.
Items and procedures are described so that you can use this unit correctly and safely. If the precautions advised are followed, this will prevent possible injury to the user and any other person.
Precautions are illustrated in the following way:
WARNING
Failure to observe WARNING signs could result in a hazard to personnel possibly resulting in serious injury or death.
CAUTION
Failure to observe CAUTION signs could result in injury to personnel and damage to the unit and associated property.
Symbol shows;
Be sure to keep this manual in a place accessible to users of this unit.
< Label on the unit >
this symbol means caution.
this symbol means an action is prohibited.
this symbol means an instruction must be followed.
This mark is labeled on the cover in which the electrical components of high voltage are enclosed to prevent the electric shock. The cover should be removed by a qualified engineer or a service personnel only.
WARNING
As with any equipment that uses CO2 gas, there is a likelihood of oxygen depletion in the vicinity of the equipment. It is important that you assess the work site to ensure there is suitable and sufficient ventilation. If restricted ventilation is suspected, then other methods of ensuring a safe environment must be considered. These may include atmosphere monitoring and warning devices.
4
Page 6
WARNING
Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to
rain water.
Only qualified engineers or service personnel should install the unit. The installation by
unqualified personnel may cause electric shock or fire.
Install the unit on a sturdy floor and take an adequate precaution to prevent the unit from turning over. If the floor is not strong enough or the installation site is not adequate, this may result
in injury from the unit falling or tipping over.
Never install the unit in a humid place or a place where it is likely to be splashed by water.
Deterioration of the insulation may result which could cause current leakage or electric shock.
Never install the unit in a flammable or volatile location. This may cause explosion or fire.
Never install the unit where acid or corrosive gases are present as current leakage or electric
shock may result due to corrosion.
Always ground (earth) the unit to prevent electric shock. If the power supply outlet is not
grounded, it will be necessary to install a ground by qualified engineers.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such
grounding may cause electric shock in the case of an incomplete circuit.
Connect the unit to a power source as indicated on the rating label attached to the unit. Use
of any other voltage or frequency other than that on the rating label may cause fire or electric shock.
Never store volatile or flammable substances in this unit if the container cannot be sealed. These
may cause explosion or fire.
Do not insert metal objects such as a pin or a wire into any vent, gap or any outlet on the unit.
This may cause electric shock or injury by accidental contact with moving parts.
Use this unit in safe area when treating the poison, harmful or radiate articles. Improper use
may cause bad effect on your health or environment.
Turn off the power switch (if provided) and disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent electric shock or injury.
Do not touch any electrical parts (such as power supply plug) or operate switches with a wet hand. This may cause electric shock.
5
Page 7
WARNING
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of
maintenance. These may be harmful to your health.
Never splash water directly onto the unit as this may cause electric shock or short circuit.
Never put containers with liquid on the unit as this may cause electric shock or short circuit when the liquid is spilled.
Never bind, process, or step on the power supply cord, or never damage or break the power supply plug. A broken supply cord or plug may cause fire or electric shock.
Do not use the supply cord if its plug is loose. Such supply cord may cause fire or electric shock.
Never disassemble, repair, or modify the unit yourself. Any such work carried out by an
unauthorized person may result in fire, or electric shock or injury due to a malfunction.
Disconnect the power supply plug if there is something wrong with the unit. Continued
abnormal operation may cause electric shock or fire.
When removing the plug from the power supply outlet, grip the power supply plug, not the cord.
Pulling the cord may result in electric shock or fire by short circuit.
Disconnect the power supply plug before moving the unit. Take care not to damage the power
cord. A damaged cord may cause electric shock or fire.
Disconnect the power plug when the unit is not used for long periods. Keeping the connection
may cause electric shock, current leakage, or fire due to the deterioration of insulation.
If the unit is to be stored unused in an unsupervised area for an extended period, ensure that
children do not have access and that doors cannot be closed completely.
The disposal of the unit should be accomplished by appropriate personnel. Remove doors to
prevent accidents such as suffocation.
Do not put the packing plastic bag within reach of children as suffocation may result.
Use the reagent specified by Sanyo for H
may result in explosion or damage to the Incubator.
When performing H
door. Failure to do so may be harmful to health due to leakage of H
During H to do so may be harmful to health due to leakage of H
decontamination, plug the access hole with the silicon cap that is provided. Failure
2O2
decontamination, securely close gastight split doors, inner door and outer
2O2
decontamination. Using a different H
2O2
gas.
2O2
gas.
2O2
2O2
solution
6
Page 8
Use a dedicated power source (a dedicated circuit with a breaker) as indicated on the rating label
attached to the unit. A branched circuit may cause fire resulting from abnormal heating.
Connect the power supply plug to the power source firmly after removing the dust on the plug.
A dusty plug or improper insertion may cause a heat or ignition.
Never store corrosive substances such as acid or alkali in this unit if the container cannot be
sealed. These may cause corrosion of inner components or electric parts.
Check the setting when starting up of operation after power failure or turning off of power switch. The stored items may be damaged due to the change of setting.
Be careful not to tip over the unit during movement to prevent damage or injury.
Prepare a safety check sheet when you request any repair or maintenance for the safety of service
personnel.
Wear rubber gloves when handling the H
result in inflammation of the skin.
H
decontamination can be performed only for the chamber and chamber attachments with
2O2
standard specifications, and not for any other objects.
Perform H
Arranging them in a different way may result in insufficient decontamination.
After H
to wipe off the residual H decontaminated, and the bottoms of ducts.
2O2
decontamination with the chamber attachments arranged as specified by Sanyo.
2O2
decontamination has been completed, wear rubber gloves and use a non-woven cloth
2O2
CAUTION
reagent. Direct contact with the H
2O2
fluid from the bottom of the chamber, any objects that were
reagent may
2O2
7
Page 9
LABELS ON THE INCUBATOR
Warning and caution labels are attached to the Incubator. The following table describes the labels.
This label is attached to covers that access high-voltage electrical components to prevent electric shock. Only a qualified engineer or service personnel should be allowed to open these covers.
This symbol indicates an ultraviolet light (UV) caution.
This symbol indicates that caution is required. Refer to product documentation for
details.
This symbol indicates a hot surface.
This symbol indicates an earth.
This symbol means “ON” for a power switch.
This symbol means “OFF” for a power switch.
ENVIRONMENTAL CONDITIONS
This equipment is designed to be safe at least under the following conditions (based on the IEC 61010-1):
Indoor use;
Altitude up to 2000 m;
Ambient temperature 5
Maximum relative humidity 80% for temperature up to 31
o
at 40
C;
Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage;
Other supply voltage fluctuations as stated by the manufacturer;
Transient overvoltages according to Installation Categories (Overvoltage Categories) II; For mains
supply the minimum and normal category is II;
Pollution degree 2 in accordance with IEC 60664.
o
C to 40oC
o
C decreasing linearly to 50% relative humidity
8
Page 10
INCUBATOR COMPONENTS
(
)
)
1
18
Handles
Handles
Rear Right Side
3
6
For MCO-19M(UVH)/19M(UV) or when MCO-19UVS is installed.
12
2
11
13
1524
21
14
9
10(inside
25
4
19
Rear Left Side
5 7
8
16, 17
For MCO-19M(UVH)/19M(UV) or when MCO-19UVS is installed.
inside
22
23
20
9
Page 11
1. Outer door: The outer door is held to the frame with a magnetic seal. A door heater is installed in the
door panel. The door opening is reversible. Contact a Sanyo representative or agent to change the door hinge from left to right or vice versa.
2. Inner door: Open when take the trays out or clean inside.
3. Gastight split door: Made of tempered glass, however avoid excessive impact on the glass.
4. Leveling feet: The leveling feet can be turned to adjust the height. Adjust the feet so that the Incubator
is level.
5. Trays:
tray out.
6. Tray supports: The tray supports can be removed by lifting the front side and pulling toward you.
7. Side supports: The right and left side supports can be removed for disinfection. Refer to page 45.
8. Duct: The duct for the path for circulating air. It is removable.
9. Fan cover: The fan cover serves as the inlet for circulating air. It is removable.
10. Fan (inside the duct): The fan is made from polypropylene resin. It can be disinfected in an autoclave.
11. Sample air outlet: The sample air outlet also functions as an internal gas outlet. Normally, cover this
outlet with the sample air outlet cap.
12. Sample air outlet cap: Always attach this cap except when using the sample air outlet.
13. Door switch: Detects the door opening/closing and stops the circulating fan and electromagnetic
valve for CO
14. Humidifying pan: Fill the humidifying pan with sterile distilled water. Install the pan properly so that it
can be covered with the pan cover.
15. Humidifying pan cover: The cover prevents UV light entering the chamber. When filling the pan, lift
the front side and take out the pan. Refer to page 48 for details.
Set trays at the same level as each gastight split door. Open a inner door when take the
and N2/O2 when door is open. UV lamp is also deactivated by door opening*.
2
16. UV lamp*: This Sanyo UV lamp does not generate ozone. Never look directly at the UV light. For
replacement, contact a Sanyo representative or agent.
17. Water level sensor for humidifying pan: This sensor detects the water level in the humidifying pan.
Refer to page 44 for details.
18. Remote alarm terminals: Refer to page 13.
19. Access port: Place the silicon caps on the port both outside and inside when the port is not being
used.
20. Power switch: This is the main switch for the Incubator. It also functions as an overcurrent breaker.
21. Glow starter*: The glow started if for the UV lamp (model FG-7P)
22. Connecting port A/B for CO
Changeover System is installed, both ports A and B are available. If the MCO-21GC is not installed, only port A is available. Refer to page 16 for gas cylinder connection. Ensure that the gas pressure is set at
0.03 MPaG (0.3 kgf/cm
2
G, 4.3 psiG). Refer to page 67 for automatic CO2 cylinder changeover.
gas pipe: When the optional MCO-21GC Automatic CO
2
Cylinder
2
10
Page 12
23. Connecting port A/B for N
the gas pressure is set at 0.05 MPaG (0.5 kgf/cm gas cylinder changeover.
24. Attachment location for electric key: This is the attachment location for the electric key included in
the optional Component H decontamination. Refer to the installation procedure for MCO-HL for details.
25. Gas injection nozzle
* MCO-19M(UVH)/19M(UV) or when an optional UV System Kit (MCO-19UVS) is installed.*
2O2
gas pipe: Refer to page 17 for gas cylinder connection. Ensure that
2/O2
Decontamination Kit (MCO-HL). This kit must be attached to perform H2O2
2
G, 7.1 psiG). Refer to page 30 for automatic N2 (or O2)
11
Page 13
Control panel and keypad
LCD panel
2
1
4
9
1. Upper limit regulator:
This regulator is used to set the upper
temperature limit.
2. OVERHEAT indicator:
This indicator lights when the chamber
temperature reaches the upper limit.
3. MENU Key:
Press this key to access the menu.
4. Contrast knob:
Turn this knob to adjust the contrast of the
LCD.
5. BUZZER Key: Press this key to silence the buzzer.
Note:
buzzer for an upper limit temperature alarm.
The following optional components must be installed to perform decontamination.UV Lamp Add-on Kit (MCO-19UVS) (This is provided as standard equipment for the MCO-19M(UV).) HH
5
It is not possible to silence the
Decontamination Kit (MCO-HL)
2O2
Generator (MCO-HP)
2O2
6. CE Key:
Press this key to clear the entered value
when entering a setting.
7. Cursor Keys (Up, Down, Left, and Right):
Use these keys to move the cursor on the
LCD panel.
8. ENTER Key:
Press this key to select a menu command.
When entering a setting, pressing this key to
move to the next parameter.
9. Numeric Keys
10. H
Press this key to start H
Refer to page 49 for details.
2O2
Key:
decontamination.
2O2
10
3
7
8
6
12
Page 14
Remote alarm terminal
The remote alarm terminal is located at the rear right side of the incubator.
The alarm is outputted from this terminal. Contact capacity is DC 30 V, 2 A.
Contact output:
between COM. and N.O. between COM. and N.C.
At normal Open Close
At abnormal Close Open
Note:
The alarm is actuated when the power cord is disconnected from the outlet or the power switch is OFF.
• The remote alarm cannot be stopped by pressing the alarm buzzer stop key (BUZZER) since the remote
alarm is not conjunct with the alarm buzzer stop key (BUZZER).
Remote alarm
terminal
Remote alarm terminal
13
Page 15
INSTALLATION SITE
For correct operation of the Incubator, install it in a location with the following conditions.
WARNING
When using CO small room without adequate ventilation may cause gas poisoning or oxygen deprivation. In addition, when opening the Incubator doors, do not directly inhale the air in the chamber.
Si l’appareil est utilisé dans un evdroit restreint, le niveau de la densite CO être nocif aux humains. Evitez d’aspirer l’air provenant de l’inérieur de l’appareil quand vous ouverz la porte.
Normal air environment
z
Install the Incubator in an environment with normal air.
Do not expose to direct sunlight
z
Do not install the Incubator in a location where it will be exposed to direct sunlight. If the Incubator is operated in direct sunlight, performance will be adversely affected.
Separate from heat sources
z
Do not install the Incubator near significant heat sources, such as heaters, boilers, ovens, or autoclaves. Heat will adversely affect the performance of the Incubator.
Ambient temperature at least 5°C lower than set temperature
z
The control temperature of the Incubator is at least 5°C higher than the ambient temperature. For example, if the chamber is controlled at 37°C, the ambient temperature must normally be no more than 32°C. Do not allow the ambient temperature to become too high.
Strong and level floor
z
Select a site with a strong and level floor. If the floor is uneven or the installation is not level, the Incubator will be unstable and this may cause accident or injury. To avoid vibration and noise, always make sure that the installation is stable. An unstable surface may result in vibration or noise.
gas for control, make sure that there is adequate ventilation. Using CO
2
de l’air peut s’élever et peut
2
gas in a
2
WARNING
Install the Incubator at a location that can support the weight. If the floor is not strong enough or if the
installation is insufficient, the Incubator may fall over and cause injury.
Always make sure that the floor is strong, even, and level, and that the Incubator will not tip over.
An insufficient installation may result in injury due to water leakage or the Incubator falling over.
Low humidity
z
Select a site with a relative humidity of 80% or lower. Using the Incubator in high humidity may result in current leakage or electric shock.
WARNING
Do not use the Incubator outdoors. If the Incubator is exposed to rain water, it may result in current
leakage or electric shock.
Never install the Incubator in a moist location, such as near a sink or water line, or where it is likely to be exposed to water. In addition, do not install it near water or steam pipes. Moisture can
the insulation to deteriorate, which may result in current leakage or electric shock.
cause
No inflammable or corrosive gas
z
Never install the Incubator in a location where it will be exposed to inflammable or corrosive gas. Doing so may result in explosion or fire. In addition, insulation may deteriorate due to corrosion of protective casing, resulting in current leakage or electric shock.
No falling objects
z
Do not install the Incubator in a location where there is the possibility of objects falling from above. Doing so may result in damage or accident.
14
Page 16
INSTALLATION
1. Remove the packing tape and clean up.
Remove all the tape that is securing the doors and inner attachments. Open the doors for ventilation. If
the outer panels are dirty, dampen a cloth with a diluted neutral detergent and wipe them. (Using an
undiluted solution may damage the plastic. Follow the diluting instructions for the detergent that is used.)
Wipe off the residual detergent with a wet cloth and then wipe off any moisture.
WARNING
Do not leave plastic wrapping bags within reach of children. If a bag is placed over a child’s head, it
can block the mouth and nose and cause suffocation.
2. Adjust the leveling feet.
Adjust the leveling feet by turning them counterclockwise to level the Incubator. (See the illustration on
the right.)
3. Ground the Incubator.
Ground the Incubator during installation to prevent electric shock in case the insulation is not sufficient. If
there is no ground wire at the location, consult with qualified service personnel.
When a ground must be installed
z
If a grounded 3-pole outlet is not available, then a ground must be installed. Consult with qualified service
personnel.
WARNING
To prevent electric shock, always ground the Incubator. If grounding is not possible, then have a ground
installed by qualified personnel. If the Incubator is not grounded, it may result in electric shock.
Never connect the ground wire to a gas pipe, water pipe, lightning rod, or telephone ground wire.
Doing so may cause electric shock
Installing a ground fault circuit breaker
z
If using the Incubator in a location with moisture or humidity cannot be avoided, then it is recommended
that a ground fault circuit breaker be installed in the power supply circuit (i.e., the power supply at the
Incubator). Have the circuit breaker installed by qualified service personnel.
.
CAUTION
Do not climb on the Incubator or place objects on top of it. Doing so may damage it or cause it to fall
over, resulting in injury. If it is to be stacked, refer to page 72 and stack it securely.
When the Incubator is not in use
z
Empty the water from the humidifying pan and remove moisture from the chamber. Make sure that the
chamber is completely dry before closing the doors. Failure to do so may result in damage.
Before moving the Incubator
z
Before moving the Incubator, empty the water from the humidifying pan, disconnect the power supply plug
from the outlet, and make sure that the cord will not be damaged. Failure to do so may result in electric
shock or fire.
15
Page 17
Connection of CO2 gas cylinder
WARNING
When connecting a gas cylinder to the Incubator, confirm the gas type. Confirm that the connections are secure and that no gas will escape. Be sure to use the specified pressure. Using an incorrect
gas or pressure may result in explosion or fire, or in gas poisoning or oxygen deprivation due to escaping
gas.
Install the Incubator in a location with adequate ventilation. If adequate ventilation cannot be
provided, then install an alarm system using
CO2 and O2 densitometers.
1. Use a liquefied CO
gas cylinder (at least 99.5% pure). The siphon (dip tube) type cannot be used.
2
2. Install the Gas Pressure Regulator (MCO-100L, purchased separately) specified by Sanyo on the CO
gas cylinder. Otherwise, use a CO
for the primary pressure, and 0.2 MPaG (2 kgf/cm
gas pressure regulator rated at 25 MPaG (250 kgf/cm2G, 3600 psiG)
2
2
G, 30 psiG) for the secondary pressure.
2
3. Using the gas supply pipe that is provided, connect the CO
gas pressure regulator to the CO2 gas pipe
2
inlet (located at the lower left-hand side on the rear panel of the Incubator).
Note:
If CO
is supplied to multiple CO2 Incubators from a single gas cylinder, a CO2 solid will be formed in the
2
gas pressure regulator. The gas pressure regulator safety valve will operate, and there may an explosive
sound.
When the MCO-21GC is not mounted
Using the gas supply pipe that is provided, connect the CO
cylinder to CO
gas pipe connection port A on the Incubator. After connecting the pipe, check to make
2
gas pressure regulator for the CO2 gas
2
sure that no gas is leaking.
When the MCO-21GC is mounted
Connect a pair of CO
gas cylinders with CO2 gas pressure regulators independently. The gas supply line
2
will be switched automatically. Connect the main cylinder to port A and the reserve cylinder to port B. After
connecting the cylinders, check to make sure that no gas is leaking.
For details on installing the optional automatic CO
z
Cylinder Changeover System (MCO-21GC), refer to
2
the MCO-21GC installation guide. For details on using the MCO-21GC, refer to page 67.
4. Set the CO
pressure increases, the CO
gas on the secondary side to 0.03 MPaG (0.3 kgf/cm2G, 4.3 psiG) for gas injection. As the
2
gas concentration control range will increase. Excessive pressure may
2
cause gas supply lines inside the Incubator to come loose, which may result in gas poisoning or oxygen
deprivation due to the escaping of gas. If gas lines come loose, the Incubator must be repaired.
The gas lines connected to the Incubator will degrade over time. If any deterioration or abnormalities
z
are found during inspection, replace the lines immediately.
16
Page 18
Connection of N2 (or O2 ) gas cylinder
WARNING
Check the gas type and ensure that it is fit for the purpose. Make sure that all pipes are connected correctly and are not liable to become disconnected. Ensure that the gas pressure is set at the specified value. Improper connection of the gas pipe or use of incorrect gas pressure may result in leakage of gas. Elevated level of gas can be hazardous to health and may lead to asphyxiation and risk of death.
This incubator needs N
When the setting of O
When the setting of O
O
density in the atmosphere is about 20%. For the control of O2 density in the chamber, O2 gas is
2
diluted by N
gas when the setting of O2 density is less than that of the atmosphere. On the contrary, O
2
gas is added when the setting of O2 density is more than that of the atmosphere.
1. Install a gas pressure regulator (optional accessory MCO-100L) on N
regulator rated at 25 MPaG (250 kgf/cm
30 psiG) on the secondary side.
2. Using the gas tube provided, connect the pressure regulator to the connecting port A/B for N
pipe located at the rear left side of the incubator. Connect the main cylinder to the connecting port A for
N
gas pipe and connect the sub-cylinder to the connecting port B for N2/O2 gas pipe so that the gas
2/O2
supply can be switched automatically when one gas cylinder is empty. Connect the same gas type to
both connecting port A/B for N
connecting port A/B for N
3. Set the N
or O2 gas pressure on the secondary side to 0.05 MPaG (0.5 kgf/cm2G, 7.1 psiG) (at gas
2
injection). The higher gas pressure makes control range of O
pressure may cause disconnection of internal pipes inside the incubator, which will result in leakage of N
or O
gas into the atmosphere. Elevated level of CO2 gas or O2 gas can be hazardous to health and
2
may lead to asphyxiation and risk of death or fire by gas leakage. The repair of the incubator will be
necessary if the internal pipe is disconnected.
4. Check that no gas is leaking at any point where the pipe connects with the gas regulator or the
incubator.
The gas lines connected to the Incubator will degrade over time. If any deterioration or abnormalities
z
are found during inspection, replace the lines immediately.
or O2 gas depending on the setting of O2 density. The selection is as follows:
2
density is less than 18%: N2 gas
2
density is more than 22%: O2 gas
2
or O2 gas cylinder. Or use a
2
G, 3600 psiG) on the primary side and 0.2 MPaG (2.0 kgf/cm2G,
gas pipe. Never connect N2 gas and O2 gas at the same time to the
2/O2
gas pipe respectively.
2/O2
density wider. In addition, excessive
2
2
2/O2
2
gas
2
WARNING
gas increases the susceptibility of substances to burn. Take care of the handling of flame in a room
O
2
where the incubator is installed.
17
Page 19
Connection of gas injection nozzle
In the case of control of O
port by using the connecting pipe enclosed (inner diameter; 5 mm, outer diameter; 9 mm, length; 400
mm). This helps faster recovery of humidity after opening of the inner door.
Fill the humidifying pan with sterile distilled water so that the injection nozzle can be under the water level.
                  
density in the chamber, connect the gas injection nozzle to the gas injection
2
Gasinjectionnozzle
Using the inner door and gastight split doors
zTo effectively use gastight split doors, set three trays (standard equipment) at the same level as bottom
side of each gastight split door.
              
At the same level as bottom side of each gastight split doors
18
Page 20
PREVENTING CONTAMINATION
To prevent contamination of the chamber, select a suitable installation site.
Avoid locations with high temperatures or humidity.
z
Avoid locations with high temperatures or humidity, because of a greater presence of microorganisms in
the air.
Avoid locations with passers-by or drafts.
z
Avoid locations near doors, air conditioners, fans, etc., where passers-by or drafts can facilitate the entry
of microorganisms into the chamber.
If possible, use a cleanroom.
z
To achieve a better culture, it is recommended that a cleanroom be used if one is available.
Use clean containers.
z
The greatest cause of contamination is dirty containers for cultures. Be careful not to get containers or
trays dirty when taking them in and out.
Keep the chamber clean.
z
Wipe off any fingerprints. If water spills from the humidifying pan, or if the doors are left open for a long
time, condensation may form on the inside of the doors. If that occurs, wipe off the condensation with a
dry sterile gauze. In particular, clean and disinfect the chamber if the culture medium is spilled. For details,
refer to Routine Maintenance on page 45.
Use sterile distilled water in the humidifying pan.
z
Always use sterile distilled water in the humidifying pan. Do not use ultrapure water, because it may
contain red rust-like suspended particles.
A water level alarm is displayed in the status display area on the control panel. Quickly refill the water in
the pan when the water level alarm is displayed. Adding low-temperature water will significantly lower the
temperature in the chamber. Clean the humidifying pan once a month.
Keep the Incubator out of direct airflows from air conditioners or fans.
z
Cool airflow from an air conditioner may cause condensation and lead to possible contamination.
19
Page 21
PRECAUTIONS FOR CULTURES
Leave space between culture containers.
z
Always leave space for ventilation between culture containers (Petri dishes, flasks, etc.). Inadequate
spacing may result in uneven temperature distribution and CO
Do not place harmful materials in the chamber.
z
Never place samples that release acidic, alkali, or corrosive gas in the chamber. Doing so may cause
damage resulting from discoloration or corrosion.
Close the inner door and all gastight split doors.
z
Always close the inner door and all gastight split doors before closing the outer door. Failure to close the
inner door or any gastight split door will adversely affect performance even if the outer door is closed.
Open and close the doors gently.
z
Always open and close the doors gently. Closing the doors forcefully may cause spillage of the culture
medium, incomplete closing, or damage to the gasket. Before opening the inner door, check through the
glass to confirm that the UV lamp is OFF (if the MCO-19M(UVH)/19M(UV) or the optional MCO-19UVS is
installed).
Be careful when closing the outer door.
z
Use the handle when closing the outer door. Holding the door in other places may cause injury by getting
fingers caught in the door. Do not lean on the outer door. Doing so may result in injury from the outer door
coming loose or the Incubator falling over, or it may cause current leakage or electric shock.
Be careful of the inside of the outer door.
z
The inside of the outer door may become hot.
Avoid using excessive force on the inner door.
z
Do not put your hand on the glass, poke it with sharp objects, or apply strong force. Doing so may result
in injury from breaking the glass.
Check the cause of any alarm buzzer.
z
If an alarm buzzer sounds while the Incubator is in use, immediately check the cause of the alarm. For
details on what may cause an alarm buzzer to sound, refer to page 55.
gas concentration.
2
20
Page 22
Using the Unlock Key
Unlocking when power is interrupted
z
If power is interrupted to the MCO-19M(UVH)/19M(UV) with an MCO-HL installed, the outer door is
electrically locked. To unlock the outer door while the power is interrupted, use the unlock key that is
provided. To lock the outer door again, turn the unlock key to the lock direction while the door is open.
After the door has been locked condition manually, then close the door.
Note:
The outer door cannot be locked by using the unlock key while the door is closed. Lock the door while it is
open. Attempting to turn the unlock key while the door is closed may damage the electric lock system.
WARNING
For the MCO-19M(UVH)/19M(UV) with an MCO-HL installed, the outer door is electrically locked during
decontamination. The chamber is filled with H
during decontamination.
gas during decontamination. Do not unlock the door
2O2
CORRECT OPERATION
Use the following procedure to start trial operation or actual operation of the Incubator.
1. Install the Incubator correctly, referring to Installation on page 15.
2. Remove the packing materials from the chamber and inner attachments. Clean and disinfect the
chamber and inner attachments, referring to Routine Maintenance on page 45.
3. Add approximately 1.5 L of sterile distilled water to the humidifying pan. (Refer to page 48.)
4. Turn ON the power supply switch on the left side of the rear panel of the Incubator.
Note:
The humidity in the Incubator chamber is adjusted to the optimum setting. To prevent condensation on
the surface inside chamber and the inner door, there is a low-temperature area under the humidifying pan
in the bottom of chamber to recondense evaporated moisture. Condensation may occur around the
humidifying pan at the bottom of the chamber, but this does not indicate a problem.
21
Page 23
LCD PANEL
The following display (called the Top Display) will appear when the power switch is turned ON. The
default temperature is 37.0
and time are preset at the factory. Refer to page 38 to change the date and time.
      
2
   
10
    
1. Display of set value of temperature
The set value of chamber temperature is displayed.
2. Display of current temperature
The current chamber temperature is displayed.
3. Display of set value of CO
The set value of the chamber CO
4. Display of current CO
The current chamber CO
5. Display of current CO
A and B will be displayed if the optional Automatic CO installed. The port of the CO Nothing will be displayed if the MCO-21GC is not installed.
6. Display of set value of O
The set value of the chamber O
7. Display of current O2 density
The current chamber O
8. Display of current N
The connecting port A/B for N reverse video.
1
01 0
1
2
3
4
5
6
7
2
T e
mp 3 7 . 0 o C C O 2 5 . 0 % A B O 2 5 . 0 % A B
37.
S t a t u s U V R H P
O K
2
o
C, the default CO2 density is 0% and the default O2 density is 20%. The date
3
11 12 1314151617181920212223242526272829 30 31 32 33 34 35 363738394
.
A N D o o r : O p e n
4
density
2
density is displayed.
2
density
2
density is displayed.
2
cylinder
2
Cylinder Changeover System (MCO-21GC) is
2
gas pipe that is currently supplying CO2 will be displayed in reverse video.
2
density
2
density is displayed.
2
density is displayed.
cylinder
2/O2
gas pipe that is currently supplying N2 or CO2 will be displayed in
2/O2
65
.
.
7
8
0
22
Page 24
9. Status display field
Various status or alarms are displayed.
When UV lamp is lit: “UV” is displayed in reverse video. (If the MCO-19M(UVH)/19M(UV) or the optional
MCO-19UVS is installed).
When humidifying water is low: “RH PAN” is displayed in reverse video and blinks. When the door is open: Door: Open” is displayed in reverse video.
10. Message display field
A message is displayed when fault occurs. The message is displayed alternately in normal characters
and reverse video. Refer to pages from 55 to 57 for alarm details. “OK” is displayed during normal operation.
    
23
Page 25
BASIC OPERATIONS ON CONTROL PANEL
The following operation are possible through control panel: 
z Setting the temperature・・・The chamber temperature can be set (page 25).
z Setting the CO
z Setting the O
z
Setting the key lock・・・Changing the temperature and CO
z Setting alarms・・・The temperature alarm, CO
limit alarm temperature (page 32) can be set.
z Setting UV parameters・・・ON/OFF parameters and the ON period of the UV lamp can be set for the
MCO-19M(UVH)/19M(UV) or when the Incubator is equipped with the optional MCO-19UVS (page 36).
z Setting other parameters・・・The initial settings of the date, time, and log cycle (page 38), and the
LCD display and baud rate (page 40) can be set.
z Displaying the operation log and transferring data・・・A graph of past operation data can be
displayed (page 41) and operation data can be transferred to a PC (page 43).
z H
decontamination・・・H
2O2
performed when the optional H
z Automatic N
30). The CO
System (MCO-21GC) is installed (page 67).
z Setting the standard gas density・・・The standard gas density can be set when the optional
Automatic Standard Gas Calibration Kit (MCO-SG) is installed. (page 69)
density・・・The chamber CO
2
density・・・The chamber O
2
decontamination of the chamber and internal attachments can be
2O2
Generator (MCO-HP) is installed (page 49).
2O2
(or O2) gas cylinder changeover・・・The N
2
gas cylinder can be switched when the optional automatic CO2 Cylinder Changeover
2
density can be set (page 25).
2
density can be set (page 25).
2
density can be disabled (page 26).
2
density alarm, O2 density alarm (page 33) and upper
2
(or O2) gas cylinder can be switched (page
2
24
Page 26
BASIC PARAMETERS

Setting the chamber temperature, CO
The setting procedure for the chamber temperature, CO
settings:Chamber temperature: 37
The Incubator automatically starts operation using these settings after the power is turned ON.
1. From the Top Display, press the MENU Key to access the menu. Select Set and press the ENTER
Key.
 
1
 
2
3
4
5
6
7
 
2. The Stand-By Setting Display will be displayed. Set the parameters.  
1
 
2
3
4
5
  
6
The setting ranges of the parameters are as follows:
Temperature: 0 to 50
CO
O
3. Press the MENU Key to display the menu after setting the parameters. Select OK and press the
ENTER Key. The settings will be saved.
Note:
When starting the Incubator for the first time or after not using it for an extended period of time, allow at
least 4 hours for the chamber temperature, humidity, CO
setting the desired chamber temperature at a 0% CO
setting to the desired CO
T e m p 3 7 . 0o C C O 2 5 . 0 %
37.
S t a t u s
O K
S t a n d - by S e t t ing T e m p e r a t u r e
C O 2 D e n s i t y 0 . 0 % O 2 D e n s i t y 2 0 . 0 %
K e y L o ck 0 (0.Unlock 1 . L o c k) H i g h L i m i t 5 2 . 0
o
C. To 37.0oC, enter 370.
density: 0 to 20%. To set 5.0%, enter 050.
2
density: 1 to 18 and 22 to 80%. To set 5.0%, enter 050.
2
and O2 density.
2
o
C, CO2 density: 0%, O
.
3 7 . 0oC ( 0 . 0oC - 5 0 . 0 o C)
o
and O2 density are given below. (Default
2
density: 20%)
2
( 0 . 0 % - 2 0 . 0 % ) ( 1 . 0 % - 1 8 . 0 % ) ( 2 2 . 0 % - 8 0 . 0 % )
C
sensor and O2 sensor to become stable after
2
density and a 20% O2 density. Then change the
2
and O2 density
2
O 2 5 . 0 %
M E N U
S e t
.
.
L o g T o o l s
M E N U
O K C a n c e l
25
Page 27
Setting the key lock
1. To set the key lock, change the value of the key lock parameter from 0 to 1 on the Stand-by Setting Display and press the ENTER Key. The buzzer will sound briefly and the keys will be locked.
 
1
 
2
3
4
5
  
6
S t a n d - by S e t t ing T e m p e r a t u r e 3 7 . 0
C O 2 D e n s i t y 5 . 0 % O 2 D e n s i t y 5 . 0 %
K e y L o ck H i g h L i m i t 5 2 . 0
2. “Key Lock” will be displayed in reverse video at the right side of “Stand-by Setting”, and changing the
temperature and CO
density settings will be disabled.
2
3. Press the MENU Key to display the menu, select OK and press the ENTER Key. The settings will be
saved.
o
C ( 0 . 0oC - 5 0 . 0 o C)
( 0 . 0 % - 2 0 . 0 % ) ( 1 . 0 % - 1 8 . 0 % ) ( 2 2 . 0 % - 8 0 . 0 % )
1 ( 0. Unl ock 1 . L o c k)
o
C
M E N U
O K C a n c e l
26
Page 28
Releasing the key lock
1. To remove the key lock, change the value of the key lock parameter from 1 to 0 on the Stand-by
Setting Display and press the ENTER Key.
 
1
 
2
3
4
5
  
6
S t a n d - b y S e t t i n g K e y L o c k T e m p e r a t u r e 3 7 . 0
C O 2 D e n s i t y 5 . 0 % O 2 D e n s i t y 5 . 0 %
K e y L o ck H i g h L i m i t 5 2 . 0
2. The cursor will move to the Password Field. Input the 4-digit password (default: 0000) and press the
ENTER Key. When key lock is released, the buzzer will sound briefly, and “Key Lock” will disappear from
the Stand-by Setting Display.
  
1
2
3
4
5
  
6
S t a n d - b y S e t t i n g T e m p e r a t u r e 3 7 . 0
C O 2 D e n s i t y 5 . 0 % O 2 D e n s i t y 5 . 0 %
K e y L o c k 0 H i g h L i m i t 5 2 . 0
3. Press the MENU Key to display the menu, select OK and press the ENTER Key. The settings will be
saved.
Note: The buzzer will sound for a long time if the wrong password is entered. Enter the correct password. The password for releasing the key lock must be shared and administered among all users of the
Incubator. The password is set to 0000 when the Incubator is shipped from the factory. The procedure for
changing the password is given next.
 
o
C ( 0 . 0oC - 5 0 . 0 o C)
( 0 . 0 % - 2 0 . 0 % ) ( 1 . 0 % - 1 8 . 0 % ) ( 2 2 . 0 % - 8 0 . 0 % )
0 ( 0 . U n l o c k 1 . L o c k )
o
C
K e y L o c k
o
C ( 0 . 0oC - 5 0 . 0 o C)
( 0 . 0 % - 2 0 . 0 % ) ( 1 . 0 % - 1 8 . 0 % ) ( 2 2 . 0 % - 8 0 . 0 % )
Password
o
C
27
Page 29
Setting the key lock password
1. From the Top Display, press the MENU Key to display the menu, select Tools, and press the ENTER
Key.
 
1
2
 
3
4
5
6
 
7
2. Select Key Lock PW Setting from the Select Tools Display, press the MENU Key to display the menu, select OK, and press the ENTER Key.
          
3. Input the current user password (4 digits), select OK, and press the ENTER Key. (The default user
password is 0000 when the Incubator is shipped from the factory.)
          
T e m p 3 7 . 0o C C O 2 5 . 0 %
37.
S t a t u s
O K
S e l e c t T o o l s ( 1 / 2 ) T e m p / C O 2 C a l i b r a t i o n
A l a r m Se t t i n g L C D / D A Q S e t t i n g U V S e t t i n g K e y L o c k P W S e t t i n g D a t e T i m e
K e y L o c k P W S e t t i n g
C u r r e n t U s e r P a s s w o r d * * **
.
O 2 5 . 0 %
.
.
M E N U S e t L o g
T o o l s
M E N U
O K
S v C C a n c e l
28
Page 30
4. Enter the new user password (4 digits), select OK, and press the ENTER Key.
 
K e y L o c k P W S e t t i n g
  
N e w U s e r P a s s w o r d * * **
      
5. Enter the new user password again, select OK, and press the ENTER Key to save. After that press the MENU Key to display the menu, and select Cancel from the menu.
 
K e y L o c k P W S e t t i n g
  
N e w U s e r P a s s w o r d ****
 
R e E n t e r U s e r P a s s w o r d * ** *
    
Note:
Be careful not to forget the key lock password. If you have forgotten the password and cannot release the
key lock, contact a Sanyo sales representative or agent.
  
29
Page 31
Automatic N2 (or O2) gas cylinder changeover
This incubator switches the gas supply lines when one N
Note: An automatic switcher for the CO
accessory.
Perform the following procedure
1. Connect 2 gas tubes to connecting ports A/B for N pipes are located on the left back of the Incubator. (Refer to page 9).
2. Connect a N
page 17 for information on connecting the gas cylinders. (“Gas cylinder A” is the N
connected to the connecting port A for N
connected to the connecting port B for N
3. Open the valves of both gas cylinders.
4. The current N
    
1
2
3
 
4
5
6
7
 
5. The incubator detects that there is no more N
chamber does not increase for a while even after opening the N
switches the gas supply line when it detects that there is no more gas. At that time, the following
indications of switching the gas cylinder are provided.
When N2 (or O2) gas gas supply line
switches automatically.
To silence the buzzer, press the BUZZER Key. The alarm condition is cleared and normal operation
continues.
T e m p 3 7 . 0o C C O 2 5 . 0 % A B O 2 5 . 0 % A B
37.
S t a t u s
O K
(or O2) gas cylinder equipped with a gas pressure regulator to each gas tube. Refer to
2
(or O2) gas supply line is displayed on the Top Display.
2
gas supply lines (MCO-21GC) is available as an optional
2
gas pipe, and “gas cylinder B” is the N2 (or O2) gas cylinder
2/O2
gas pipe.)
2/O2
.
2
Message display field Display of gas supply
Err02: N2/O2 gas Empty is
displayed.
(or O2) gas cylinder becomes empty.
2
gas pipe. Connecting ports A/B for N2/O2 gas
2/O2
(or O2) gas cylinder
2
Current N2/O2 gas supply line
.
.
(or O2) gas in a cylinder when the O2 density in the
gas valve in the Incubator. It
2/O2
The character for gas supply line
that is empty blinks.
The gas supply line that is
currently being used is displayed
in reverse video.
30
Page 32
6. When gas cylinder A is switched to B, remove gas cylinder A and replace it with new one. Note: Be careful when handling the empty N
cylinder.
When gas cylinder B becomes empty, the line is switched to gas cylinder A again.
The incubator detects that there is no more N
chamber does not increase for a while even after opening the N
supply line may also be switched for other reasons even though there is gas left in the cylinder: blocking
or restricting of gas tube, reduction of N
cylinder. Always check the gas quantity in the cylinder before disconnecting it.
Use the following procedure to manual switch the line between N
procedure shows switching the line from N
1. From the Top Display, press the Up Cursor Key for 3 seconds.
1
 
2
3
4
5
 
6
7
2. The N
 
1
2
3
 
4
5
6
 
7
(When the line needs to be switched from N
3 seconds.)
T e m p 3 7 . 0o C C O 2 5 . 0 % A B O 2 5 . 0 % A B
37.
S t a t u s
O K
(or O
2
T e m p 3 7 . 0o C C O 2 5 . 0 % A B O 2 5 . 0 % A B
) gas supply line will be switched from cylinder B to A
2
37.
S t a t u s
O K
(or O2) gas pressure, or improper opening of N2 (or O2) gas
2
.
.
(or O2) gas cylinder. Some gas may still be left in the
2
(or O2) gas in a cylinder when the O2 density in the
2
(or O2) gas valve in the Incubator. The
2
(or O2) gas cylinders A and B. This
2
(or O
2
) gas cylinder B to A as an example.
2
.
.
.
.
(or O2) gas cylinder A to B, press the Down Cursor Key for
2
31
Page 33
Setting the upper limit alarm temperature
An upper limit temperature alarm is provided with the Incubator. The alarm temperature can be changed
by using the following procedure.
1. In the Stand-by Setting Display, turn the upper limit regulator on the control panel using a small
screwdriver to set the desired upper limit alarm temperature. The upper limit temperature can be set
between 35 and 51
 
1
 
2
3
4
5
  
6
S t a n d - by S e t t ing T e m p e r a t u r e 3 7 . 0
C O 2 D e n s i t y 5 . 0 % O 2 D e n s i t y 5 . 0 %
K e y L o ck 0 ( 0 . U n l o c k 1 . L o c k ) H i g h L i m i t 5 2 . 0
Note:
Set the upper limit alarm temperature (High Limit) to at least 5
temperature.
Set the upper limit alarm temperature to at least 50
2. After setting the parameter, press the MENU Key to display the menu, select OK, and press the
ENTER Key. The alarm temperature will be saved.
Refer to the tables of alarms and safety functions for culture operations on page 55 for details.
Note:
The alarm temperature will be changed at any time by turning the upper limit regulator even if the
Stand-by Setting Display is not displayed.
Set the upper limit alarm temperature after the chamber reaches the set temperature for operation.   
o
C.
o
C ( 0 . 0oC - 5 0 . 0 o C)
( 0 . 0 % - 2 0 . 0 % ) ( 1 . 0 % - 1 8 . 0 % ) ( 2 2 . 0 % - 8 0 . 0 % )
o
C
o
C higher than the chamber set
o
C when H2O2 decontamination is performed.
32
Page 34
ALARM PARAMETERS
MEN
1. Press the MENU Key from the Top Display to display the menu, select Tools, and press the ENTER
Key.
 
1
2
 
3
4
5
6
 
7
T e m p 3 7 . 0o C C O 2 5 . 0 %
37.
S t a t u s
O K
2. Select Alarm Setting from the Select Tools Display, press the MENU Key to display the menu, select OK, and press the ENTER Key.
01 02 03 04 05 06 07 08 0910 11 12 13 14 15 1617181920212223242526272
S e l e c t T o o l s ( 1 / 2 )
T e m p / C O2 / O 2 C alibration
 
A l a r m S e t t i n g
L C D / D A Q S e t t i n g
U V S e t t i n g
K e y L o ck P W S etting
D a t e T im e
 
3. The Alarm Setting Display will appear. The temperature alarm, CO
back time, and door delay can be set on this display. The alarm buzzer can be silenced by pressing the
BUZZER Key. The buzzer will sound again after the specified ring back time if the condition that caused
the alarm continues. The ring back time can be set.
0
1 02 03 04 05 06 07 08 09 10 11 12 13 14 15 1617181920212223242526272829303132 33 34 35 36 37 383940
A l a r m Se t t i n g
1
T e m p A l a r m ± 1 . 0oC ( ±1. 0°C ± 5 . 0 ° C
3
C O 2 A l a r m ± 1 . 0 % ( ± 0 . 5 % ±5 . 0 %
O 2 A l a r m ± 1 . 0 % ( ± 0 . 5 % ± 5 . 0 %
4
A l a r m De l a y 15min (0-15min)
5
R i n g Ba c k 30min (0.OFF 1− 9 9 m i n
D o o r De l a y 2min (1-30min)
6
Temperature alarm: ±1.0 to ±5.0CO
alarm, O2 alarm: ±0.5 to ±5.0% (Default: ±1.0%)
2
Alarm delay: 0 to 15 minutes (Default: 15 minutes) Ring back time: 1 to 99 minutes, or OFF (Default: 30 minutes) Door alarm delay: 1 to 30 minutes (Default: 2 minutes)
.
o
C (Default: ±1.0oC)
O 2 5 . 0 %
.
.
U
OK
Cancel
2303132 33 34 35 36 37 383940
8
alarm, O2 alarm, alarm delay, ring
2
M E N U S e t L o g
T o o l s
M E N U
O K
S v c C a n c e l
M E N U
O K
C a n c e l
33
Page 35
UV LAMP PARAMETERS
A UV lamp is used with the MCO-19M(UVH)/MCO-19M(UV) or the MCO-19M with an MCO-19UVS UV
Lamp Expansion Kit installed. Use the following procedures to make the settings.
Using the UV lamp
A UV lamp is located inside the duct to sterilize the water in the humidifying pan and the air circulating in
the chamber. Observe the following points to use the UV lamp correctly.
When all chamber attachments are installed correctly, only the inside of the duct and the inside of the
z
humidifying pan cover are exposed to UV light.
Correctly install all of the chamber attachments when starting a cell culture, and never turn ON the UV
z
lamp when the humidifying pan cover is removed.
Always install the humidifying pan cover even when using the Incubator without turning ON the UV
z
lamp. Leaving the cover uninstalled will affect the chamber temperature distribution and the humidity
recovery performance.
The UV lamp stays lit for a preset period of time after the outer door is closed. The default setting is for
z
five minutes.
If the outer door is not opened for at least 12 consecutive hours, the UV lamp will light for the preset
z
time period every 12 hours.
The recommended replacement time for the UV lamp (i.e., when the UV output ratio drops to 60% to
z
70% of its initial value) is when the accumulated ON time reaches 1,000 hours. When the accumulated
ON time reaches approximately 1,000 hours, the “UV” will flash in the status display field on the LCD
panel (when the UV lamp is not lit). It is recommended that the UV lamp be quickly replaced at this point.
Consult a Sanyo sales representative or agent for information on replacing the UV lamp.
If the UV lamp burns out, Err18:UV Lamp Abnormal will be displayed in the message display field. If
z
this occurs, replace the UV lamp. When replacing the UV lamp, replace the glow starter (type FG-7P) at
the same time. Consult a Sanyo sales representative or agent for information on replacing the UV lamp.
If the UV lamp burns out (Err18:UV Lamp Abnormal will be displayed in the message display field), it
z
will not be possible to perform H
decontamination. Replace the UV lamp and the glow starter.
2O2
34
Page 36
Precautions when using the UV lamp
WARNING
Do not look directly at UV light. UV light is harmful to the eyes.
Always use the humidifying pan and humidifying pan cover.
z
The humidifying pan and cover prevent UV light from escaping. Always use them even when not
humidifying. To check whether the UV lamp is lit, open the outer door and then press the door switch with
the inner door still closed. Visible blue light can be confirmed from the front of the humidifying pan cover.
UV light is harmful to the eyes, so do not light the UV lamp when the inner door or humidifying pan cover
is open.
Be careful when handling the UV lamp.
z
There is a UV lamp inside of the chamber duct. Be careful not to damage it when installing or removing
chamber attachments, the humidifying pan, or the H
 
Generator.
2O2
35
Page 37
MEN
Setting the UV lamp ON period
Use the following procedure to change the setting of the UV lamp ON period.
1. Press the MENU Key from the Top Display to display the menu, select Tools, and press the ENTER
Key.
 
1
2
 
3
4
5
6
 
7
2. Select UV Setting from the Select Tools Display, press the MENU Key to display the menu, select OK,
and press the ENTER Key.
         
3. The UV Setting Display will appear. Set the UV Timer for the UV lamp ON period.
 
1
 
3
 
4
5
  
6
4. Press the MENU Key to display the menu, select OK, and press the ENTER Key. The parameter will
be saved.
 
T e m p 3 7 . 0o C C O 2 5 . 0 %
37.
S t a t u s
O K
01 02 03 04 05 06 07 08 0910 11 12 13 14 15 1617181920212223242526272
S e l e c t T o o l s ( 1 / 2 ) T e m p / C O2 / O 2 C alibration
A l a r m Se t t i n g L C D / D A Q S e t t i n g U V S e t t i n g K e y L o ck P W S etting D a t e T im e
U V S e t t i n g U V T i m er 5 mi n ( 0 - 3 0 mi n )
U V L i f e 0 % (100%=1000h) U V T i m e r E x t + 0 %
U V 2 4 h M o d e 0 ( 0 . O F F 1 . O N )
.
O 2 5 . 0 %
.
.
U
OK
Cancel
8
9303132 33 34 35 36 37 383940
M E N U S e t L o g
T o o l s
M E N U
O K
S v c C a n c e l
36
Page 38
The UV timer can be set from 0 to 30 min. The default setting is for 5 min. When the UV timer is set to zero, the UV lamp will not light.
If the outer door is opened while the UV lamp is lit, the lamp will turn OFF. Then, when the door is closed, the lamp will light for the preset period of time.
If the UV lamp ON time is set for longer than 10 minutes or if only the outer door is repeated opened and closed, condensation may occur in the chamber and may affect temperature distribution. It will also shorten the service life of the UV lamp.
When replacing the UV lamp, consult with a Sanyo representative or agent.
Note: To compensate for the drop in UV ray output along with increased accumulated ON time of UV
lamp, the Incubator automatically extends the UV lamp ON time according to the accumulated ON time.
The automatically extended time is displayed as a percentage under UV Timer Ext. in the UV Setting Screen.
Lighting the UV lamp for 24 hours
If the chamber has been contaminated by dirt or by spilling the medium, use the following procedure to decontaminate the chamber by lighting the UV lamp for 24 hours.
1. Remove all attachments from the chamber, including the trays, tray supports, side supports, fan cover,
duct, fan, humidifying pan, humidifying pan cover, N attachments in an autoclave or with disinfectant alcohol.
2. Clean and wipe off inside the chamber with disinfectant alcohol.
3. Set the CO
 
U V S e t t i n g
1
U V T i m er 5 min (0-30min)
3
U V L i f e 0 % (100%=1000h)
 
U V T i m e r E x t + 0 %
4
5
U V 2 4 h M o d e 1 (0.OFF 1.ON)
6
 
4. Press the MENU Key to display the menu, select OK, and press the ENTER Key. The parameter will
be saved.
This procedure must be preformed only with the outer door closed and the UV lamp turned OFF. The UV lamp will light continuously for 24 hours after this parameter is set. The setting is canceled if the
outer door is opened. Perform the above procedure to set the 24-hours lighting mode again if you open the door.
Install the attachments after completion of the 24-hour lighting mode.
Note: Set the upper limit alarm temperature to at least 10
when using the 24-hour lighting mode of UV lamp. The 24-hour lighting mode may cause an automatic
chamber temperature alarm because the temperature of the chamber will increase.
density to 0% and O2 density to 20%. Set the UV 24h Mode to 1 on the UV Setting Display.
2
or O2 gas injection nozzle, and. Clean all the
2
o
C higher than the chamber set temperature
37
Page 39
OTHER PARAMETERS
Setting the date, time, and log interval
1. Press the MENU Key from the Top Display to display the menu, select Tools and press the ENTER Key.
 
1
2
 
3
4
5
6
 
7
2. Select Date Time from the Select Tools Display, press the MENU Key to display the menu, select OK,
and press the ENTER Key.
                          
T e m p 3 7 . 0o C C O 2 5 . 0 %
37.
S t a t u s
O K
S e l e c t T o o l s ( 1 / 2 ) T e m p / C O 2 C a l i b r a t i o n
A l a r m Se t t i n g L C D / D A Q S e t t i n g U V S e t t i n g K e y L o ck P W S etting D a t e T i m e
O 2 5 . 0 %
.
.
.
M E N U S e t L o g
T o o l s
M E N U
O K
S v c C a n c e l
38
Page 40
3. The Date Time Display will appear. Set the date, time and log interval.
 
D a t e T i m e
1
3
D a t e 1 0 / 1 0 / 0 1 ( Y Y / M M / D D )
 
4
5
 
6
  ・Entering the date Example for October 1, 2010: Enter 101001 in the Date Field.
Entering the time
Example for 10:05:00: Enter 100500 in the Time Field.
Entering the log interval
Example for 6 minutes: Enter 6 in Log Interval Field.
Note:
The default is 6 minutes.
The log interval can be set to between 2 and 30 minutes.
Relation between the log interval and the period that can be saved
1: Log interval of 2 minutes -- About 5 days
2: Log interval of 6 minutes -- About 14 days
3: Log interval of 30 minutes -- About 70 days
For the data beyond the above period, the older data is deleted to save the newer data.
 
T i m e 1 0 : 0 5 : 0 0 ( h h m m s s )
L o g I n t e r v a l 6 m i n ( 2 - 3 0 m i n )
39
Page 41
Initial settings (LCD/DAQ parameters)
1. Select LCD/DAQ Setting from the Select Tools Display, press the MENU Key to display the menu, select OK, and press the ENTER Key.
S e l e c t T o o l s ( 1 / 2 )
 
T e m p / C O 2 C a l i b r a t i o n
A l a r m Se t t i n g
L C D / D A Q S e t t i n g
U V S e t t i n g
K e y L o ck P W S etting
 
D a t e T i m e
 
2. The LCD/DAQ Setting Display will appear. Set the initial setting for each parameter as necessary.
 
L C D / D A Q S e t t i n g
1
L C D B a c k C o l o r 1 ( 1 . B l u e 2 . W h i t e )
3
D A Q S p ee d 0 ( 0. 2400 2. 96 0 0 )
 
D A Q I D 0 ( 0 . O F F 1 - 2 5 0 )
4
D A Q M o d e 0 ( 0 . L o c a l 1 . R e m o t e )
5
R e m o t e A l a r m 1 ( 0 . O F F 1 . A c t i v e )
6
 
LCD Back Color: Setting of the background color (1: Blue, 2: White)
Note:
DAQ is an external monitoring system of the chamber status. It is necessary to set the DAQ speed, DAQ
ID, and DAQ mode to use the optional communications software. Communications softwareis an optional accessory. Contact a Sanyo sales representative or agent for details.
 
M E N U
O K
S v c C a n c e l
40
Page 42
DISPLAYING THE LOG
A log of the past chamber temperatures, CO
1. Press the MENU Key from the Top Display to display the menu, select Log and press the ENTER Key.
  
1
 
2
3
4
5
 
6
7
T e m p 3 7 . 0o C C O 2 5 . 0 %
37.
S t a t u s
O K
densities and O2 densities can be displayed on a graph.
2
.
O 2 5 . 0 %
.
.
M E N U S e t
L o g
T o o l s
41
Page 43
2. The log will be displayed with dots. Press the Up Cursor Key and Down Cursor Key to switch between
the temperature, CO
scroll the displayed data (Left Cursor Key: older data, Right Cursor Key: newer data).
 
Temperature
5 0 ° C 2 0 10/10/01 T e mp
1
3
 
4
5
  
0 1 2 24
6
    
CO
density
2
2 0 % 2 0 10/10/01 C O 2
1
3
 
4
5
  
6
1 2 24
0
    
O2 density
9 0 % 2 0 10/10/01 O 2
1
3
 
4
5
  
6
1 2 24
0
    
density and O2 density displays. Press the Left Cursor Key and Right Cursor Key to
2
Switched with Up and Down Cursor
Switched with Up and Down Cursor
42
Page 44
Transferring data
Use the following procedure to transfer the log data to a PC. 
1. To transfer the log data for one day, press the MENU Key to display the menu, select PC 1D, and press the ENTER Key. To transfer all of the log data, select PC All and press the ENTER Key.
 
5 0 ° C 2 0 10/10/01 T e mp
1
3
 
4
5
  
6
2. A Progress Display will appear. On a HyperTerminal on PC, specify a transfer, text capture, and the
name of the file to save. Use TXT or CSY as the file name extension. Press the MENU Key to display the
menu, select Start, and press the ENTER Key.
The transfer will be started. “Finished” will be displayed when the transfer has been completed. Select
Cancel from the menu after the transfer has been completed and press the ENTER Key.
 
1
3
  
4
5
 
6
Setting in PC side for transmission of log data (For Windows 2000 and Windows XP)
1. From the Windows Start Button, select Program
HyperTerminal to start the HyperTerminal. (If the HyperTerminal is not available from the Start Menu,
execute the following file: C:¥Program Files¥Windows NT¥hypertrm.exe.)
2. In the HyperTerminal Window, set a new connection, the name (for example, Sanyo), connection
settings, method of connection, COM1, properties of COM1, and port.
Baud rate: 9,600, Data bits: 8, Parity: None, Stop bits: 1, Flow control: Xon/Xoff.
(The communications condition of the Incubator will be set automatically to the above settings when the
Progress Display appears.)
Note: An optional RS232C/RS485 Interface MTR-480 (RS-232C/RS-485) is required to transfer data.
0
1 2 24
P r o g r e ss
S e n d l og d a t a to PC.
L o g D a t a 2 0 1 0 / 1 0 / 0 1
F i n i s h e d .
M E N U
P C 1 D
-
Accessories
P C A l l C l e a r C a n c e l
M E N U
S t a r t
C a n c e l
-
Communications
-
43
Page 45
WATER LEVEL SENSOR
This Incubator is equipped with a water level sensor for the humidifying pan. The sensor is set
automatically when the humidifying pan is installed. Take care not to damage the sensor when removing
or installing the humidifying pan.
When the humidifying pan is removed (side view)
When the humidifying pan is installed (side view)
Note: Make sure that the humidifying pan is all the way to the back and that the sensor comes down
when you install the humidifying pan.
Note:
• Lift the sensor before installing the humidifying pan if the sensor is in the lower position after
maintenance.
• When installing the humidifying pan, make sure that the pan is set properly and that sensor comes down
into the pan. “RH PAN” will be displayed in reverse video in the status display field on LCD panel if the
sensor does not come down completely. If necessary, set the pan again in the proper location.
• The sensor detects the water level every 30 minutes and just after the outer door is closed. It takes
several seconds to detect the water level. Therefore, “RH PAN” may displayed in reverse video several
times in the status display field on LCD panel after the outer door is closed even when the humidifying pan
is full.
Pan
Pan
Sensor
Sensor
CAUTION
Foreign particles on the water surface can adhere to the water level sensor and fittings by capillary action
because the sensor is always in the water. The adhered foreign particles degrade sensor performance
and “RH PAN” may displayed in reverse video in the status display field on LCD panel even though there
is sufficient water in the humidifying pan. Be sure to wipe OFF any dirt on the water level sensor with
disinfectant alcohol whenever you change the humidifying water. When cleaning the sensor, take care
not to apply excessive force to the lead wires.
44
Page 46
ROUTINE MAINTENANCE
Cleaning the chamber and inner attachments
When using the Incubator for the first time or when performing H
Chamber Attachments on the following page and remove and clean the inner attachments.
WARNING
Before performing any repairs or maintenance, turn OFF the power switch and unplug the Incubator.
Failure to do say may result in electric shock or injury.
Be careful not to inhale chemicals, vapors or aerosols when cleaning the Incubator. Doing so may
be harmful to health.
decontamination, refer to Removing
2O2
CAUTION
Wear rubber gloves when performing maintenance on the chamber. Failure to wear gloves may
result in cuts or abrasions from sharp edges or corners.
Note:
Be careful not to damage the humidifying pan water level sensor or the UV lamp in the chamber duct (if
the MCO-19M(UVH)/19M(UV) or the optional MCO-19UVS is installed).
Do not use detergents or antiseptic solutions with acid, alkali, or chlorine. Doing so may cause
discoloration, corrosion, or rusting.
1. Close the valve of the CO
2. Open the outer and inner doors and pull out the all trays. (See Fig. 1.) Note: Open the inner door after closing all gastight split doors.
3. Lift the front of the tray supports and pull them out. (See Fig. 2.)
4. Lift the humidifying pan cover off from the pins on the rear side. (See Fig. 3.)
and N2/O2 gas cylinder and turn off the power to the incubator.
2
Tray suppor
Fig. 1 Fig. 2 Fig. 3
Humidifying pan cover
45
Page 47
5. Pull out the N
6. Pull out the humidifying pan. (See Fig. 5.)
7. Loosen the two screws securing the fan cover and take off the fan cover. (See Fig. 6.)
gas injection nozzle and N2/O2 gas injection nozzle tube. (See Fig. 4.)
2/O2
Gas injection nozzle tube
8. Lift the duct and remove it from the pins on the rear side. (See Fig. 7.)
9. Remove the chamber circulation fan by pulling out the central spring and then by pulling out the fan.
(See Fig. 8.)
Gas injection nozzle
Fig. 4
Humidifying pan
Fig. 5
Fan cover
Fig. 6
10. Remove the screw securing the clamp for the side support and remove the clamp. (See Fig. 9.)
11. Lift the tray brace off of the pins. (See Fig. 10.)
Duct
Clamp
Fig. 7
Side
support
Fig. 9
Fan
Fig. 8
Fig. 10
46
Page 48
12. Clean all the attachments with a diluted neutral detergent, and then rinse them thoroughly with
distilled water.
13. Wipe the trays, the inner attachments such as the chamber circulation fan, and the chamber sides with sterilizing alcohol. Be careful not to leave any residual alcohol.
14. Wipe the water level sensor with sterilizing alcohol. Be careful not to leave any residual alcohol. When cleaning the sensor, take care not to apply excessive force to the lead wires.
15. To reinstall all the attachments, perform the procedure in reverse order from step 14.
Note:
• When installing the fan, insert it securely on the motor shaft (See Fig.11). Lightly turn the fan manually
to make sure that it does not strike the rear panel. Improper insertion may cause poor performance.
• As shown in Fig. 12, set the tray with only the front edge bent down. If the tray is set in the wrong
direction, it may not be level and may become unstable.
Spring
Fan
Fig. 11
Bent down
Fig. 12
47
Page 49
Filling the humidifying pan
Use the following procedure to fill the humidifying pan or to replace the water.
1. Lift the humidifying pan cover. (See Fig. 1.)
2. Pull the humidifying pan forward. (See Fig. 2.)
3. Dispose of the remaining water in the pan and
clean the pan with a neutral dishwashing detergent.
Then rinse it thoroughly with distilled water and wipe
it with sterilizing alcohol.
4. Wipe all moisture from the bottom of the chamber.
5. Return the pan to the chamber and add sterile
distilled water (approx. 1.5 L, preheated to 37°C).
(See Fig. 3.)
6. Set the pan with the inner side flush against the
back, and replace the cover. Close all gastight split
doors, inner door and outer door, and confirm that
RH PAN is not displayed in reverse video in the
status display area.
Note:
• Preheat to 37°C the water to be added to the
humidifying pan. Adding low-temperature water will
lower the temperature and humidity in the chamber.
• If the water level alarm indication “RH PAN” is
displayed, use the above procedure to replenish the
water.
Fig. 1
Fig. 2
CAUTION
When refilling the water in the humidifying pan,
always wipe off any dirt from the water level sensor with sterilizing alcohol. While doing that,
be careful not to apply excessive force to the
sensor lead wire.
Fig. 3
48
Page 50
H2O2 DECONTAMINATION
When the chamber has been contaminated, or when cleaning the chamber prior to starting a culture,
H
decontamination can be performed using the H2O2 Generator.
2O2
The following products must be purchased separately in order to perform H
z
performing H
decontamination, check to make sure that these are correctly installed.
2O2
UV Lamp Expansion Kit (MCO-19UVS) (Not required for the MCO-19M(UV).)
H
H
Decontamination Kit (MCO-HL)
2O2
Generator (MCO-HP)
2O2
The H
z
decontamination function decontaminates the chamber and the inner attachments. Do not
2O2
place any objects other than the specified attachments in the chamber during decontamination.
Before performing H
z
decontamination, clean the chamber and the inner attachments with sterilizing
2O2
alcohol.
The chamber temperature during H
z
decontamination reaches 45°C. Before performing H2O2
2O2
decontamination, set the upper limit temperature alarm to 50°C or higher.
For the H
z
reagent, use one bottle of the H2O2 Decontamination Reagent specified by Sanyo
2O2
(MCO-H202).
After decontamination, dilute the residual H
z
Decontamination Reagent in the H2O2 Generator with a
2O2
large quantity of water and dispose of it, and rinse with distilled water. (Do not wash either the inside or
outside of the H
Generator with alcohol.)
2O2
decontamination. Before
2O2
WARNING
Use the reagent specified by Sanyo for H2O2 decontamination. Using a different H
solution may
2O2
cause explosion or damage to the Incubator, or insufficient decontamination.
Do not use chemicals other than the H
to the H
Generator.
2O2
reagent, such as alcohol. Doing so may result in damage
2O2
WARNING
When performing H2O
door are securely closed. During H
provided. Failure to do so may be harmful to health due to leakage of H
decontamination, make sure that gastight split doors, inner door and outer
2
decontamination, plug the access hole with the silicon cap that is
2O2
gas.
2O2
CAUTION
H2O2 decontamination can be performed only for the chamber and chamber attachments with standard
specifications, and not for any other objects.
49
Page 51
CAUTION
Perform H2O2 decontamination with the chamber attachments arranged as specified by Sanyo.
Arranging them in a different way may result in insufficient decontamination.
CAUTION
Wear rubber gloves when handling the H2O2 reagent. Direct contact with the H
in inflammation of the skin.
CAUTION
After H2O2 decontamination has been completed, residual H2O2 solution will remain on the bottom of the
chamber, the H
gloves, wipe it off with a non-woven cloth. Failure to do so may result in a deficient culture.
Generator, and the bottom of the duct. Wearing protective glasses and rubber
2O2
reagent may result
2O2
50
Page 52
H2O2 decontamination
Use the following procedure to perform H
1. Take out all trays, tray supports, side supports, the humidifying pan cover, the N
nozzle, the N
or O2 gas injection nozzle tube, the humidifying pan, the fan cover, and the duct from the
2
chamber. Dispose of the water in the humidifying pan, wipe the inside walls of the chamber with a gauze containing water or alcohol for sterilization.
2. Attach the duct and side supports, and set the tray supports in the 2nd, 5th, 9th holes from the top of
the side supports. Then set the trays.
3. Pour one bottle of H
pins on the H
Generator in the 2 holes on the lower left side of the duct (Fig. 1).
2O2
Decontamination Reagent (MCO-H2O2) into the H2O2 Generator. Set the two
2O2
4. Connect the H
Generator and connector on the bottom right of the far side of the chamber with the
2O2
enclosed cable. Be sure to keep the connector cap (Fig. 1).
5. Set the humidifying pan cover, humidifying pan, fan cover, the N
or O
gas injection nozzle tube as shown in Fig. 2, and close the inner door and outer door .
2
Note: The trays included as accessory are designed to be appropriate for decontamination. If Half Trays
(MCO-25ST, optional) or Trays for previous models are used, decontamination may not be sufficiently effective.
Front
H2O2 Generator
Ultrasound oscillator
Front
2O2
UV lamp
decontamination using the H2O2 Generator (MCO-HP).
or O2 gas injection
2
or O2 gas injection nozzle and the N2
2
Fron
Connector
Humidifying pan cover
Fan cover
N2 or O2 gas injection nozzle N
or O2 gas
2
injection nozzle tube
Humidifying pan
H
2O2
Generator
Fig. 1
Fig. 2
51
Page 53
6. Press the H
H 2 O 2 D e c o n t a m ination
1
3
4
5
6
7. If the system is normal, the following display will appear. (If the system is not normal, refer to Table 2 on page 57.) Select OK and press the ENTER Key to start H
be performed automatically to step 10. The buzzer will sound when decontamination has been completed
(It takes about 1 hour 40 minutes after warming-up).
Note: The outer door will be locked with an electric lock for safety until the decontamination has been completed.
H 2 O 2 D ec o n t a m ination
1
3
4
5
6
8. During the H Note: H
2O2
H 2 O 2 D e c o n UV Resolve S t e p:4
1
2
3
4
5
6
7
Key for 3 seconds. The system check will start.
2O2
H 2 O 2 D e c o n S y s tem C h e c k ・ ・・ ・ ・
decontamination. Decontamination will
2O2
M E N U
R e a d y t o S t a rt?
mist generation, “H2O2 Decon” will flash at the top left corner of the screen.
2O2
O K
C a n c e l
mist generates from the H2O2 Generator and then quickly changes to a gas.
min
90min
T e m p : 4 5 . 0 D o o r : L o c k e d
℃
52
Page 54
9. After completion of H
resolution by UV light will start.
1
2
3
4
5
6
7
H 2 O 2 D e c o n U V R e s o l v e S t e p:7
0min
10. After H
gas resolution, H2O2 decontamination has been completed. The following display will
2O2
appear. Open the outer door and disconnect the cable of H
chamber. Then take out the H
Note: Always put on protection glasses and rubber gloves when performing decontamination.
1
2
3
4
5
6
7
H 2 O 2 D e c o n t a m i n a t i o n
D e c o n F i n i s h e d .
P l e a s e R e m o v e H 2 O 2 U n
11. The following display will appear. Press the MENU Key to display the menu, select OK, and press the
ENTER Key to return to the Top Display.
1
2
3
4
5
6
7
H 2 O 2 D e c o n t a m i n a t i o n S t e p:8
D e c o n F i n i s h e d .
P U S H M E N U O K
mist generation, “UV Resolve” will flash at the top of the screen, and H2O2 gas
2O2
89min
T e m p : 4 5 . 0 D o o r : L o c k e d
Generator from the connector in the
2O2
Generator.
2O2
S t e p:8
T e m p : 4 5 . 0
i t D o o r : U n l o c k
T e m p : 4 5 . 0 ℃ D o o r : U n l o c k
℃
℃
53
Page 55
12. Dilute the remaining H
it. Rinse and wash the H
environment outside of the chamber.
Note: After H
13. After H
bottom part of the H
less), so put on protective glasses and rubber gloves and wipe it up with a non-woven cloth.
14. Ventilate the chamber sufficiently and place all the attachments back into the chamber. Note: Be sure to attach the connector cap.
CAUTION
The electric lock will remained locked if power fails during H2O2 decontamination. After the power comes
back ON, the H
decontamination again because the decontamination will be incomplete.
decontamination, cover the connector on the chamber side with the connector cap.
2O2
decontamination, surplus H2O2 liquid will remain at the bottom of the chamber and in the
2O2
2O2
gas resolution process will start execution and finish automatically. Execute the
2O2
reagent in the H2O2 Generator with a large volume of water and dispose of
2O2
Generator with distilled water. Then keep the H2O2 Generator in a clean
2O2
Generator duct. This solution contains H2O2 at a low concentration (about 1% or
WARNING
Never open the door by unlocking it with the unlock key during H2O2 decontamination or H2O2 gas
resolution with the UV lamp. H
gas leakage is potentially harmful to health.
2O2
Precautions when handling H2O2 decontamination reagent
Observe the following precautions when handling the H
Handling Precautions
z
Wear protective equipment, such as protective glasses and rubber gloves.
Do not use fire in the area where the reagent is being handled.
Do not leave any reagent in the container after it has been used or while it is being used.
Do not place inflammable or combustible materials near the area where the reagent is handled.
Precautions for Storage
z
Always close the container cover securely to prevent impurities from becoming mixed in the reagent.
Check the container to make sure that there is no damage, corrosion, or cracking.
Store the container with the inlet facing upwards, and make sure that the container will not tip or be
knocked over.
If possible, store the reagent in a location where water is available and where the floor can be easily
washed.
Precautions for Disposal
z
Dispose according to the rule in your country.
Decontamination Reagent (MCO-H2O2).
2O2
54
Page 56
ALARMS, SAFETY, AND SELF-DIAGNOSIS
)
The Incubator supports the following alarms, safety functions, and self-diagnostic functions.
Table 1: Alarms and Safety Functions for Culture Operations
Alarm or safety
function
Upper limit temperature alarm
Automatic set temperature alarm
Automatic set CO2 concentration alarm Automatic set O2 concentration alarm
Auto-return
The chamber temperature exceeds the upper limit alarm temperature set value.
The chamber temperature is out of the automatic set temperature alarm setting range (±1.0 to ±5.0
The chamber CO2 concentration is out of the automatic set CO alarm setting range (±0.5% to ±5.0%).
The chamber O2 concentration is out of the automatic set CO alarm setting range (±0.5% to ±5.0%).
There are no key operations for approx. 90 s in a setting mode.
Key lock The key lock is ON.
Door alarm The outer door is open.
CO2 gas cylinder empty
The CO2 concentration does not increase when the CO opened.
Conditions Display Buzzer Safety operation
Heater OFF Remote alarm operates
Remote alarm operates after delay of 0 to 15 min.
Remote alarm operates after delay of 0 to 15 min.
Remote alarm operates after delay of 0 to 30 min.
The setting mode is canceled.
Setting changes are prohibited. The CO closed. The N closed.
valve is
2
2/O2
valve is
The heater turns OFF after 1 min.
Remote alarm operates.
o
C).
concentration
2
concentration
2
valve is
2
OVERHEAT indicator lights. Continuous tone
Warning :High Temp or Warning :Low Temp
is displayed in the message display field.
Warning :High CO Warning :Low CO
in the message display field.
Warning :High O Warning :Low O
in the message display field.
Density or
2
Density is displayed
2
Density or
2
Density is displayed
2
Intermittent tone after set alarm delay time (0 to 15 min) has elapsed Intermittent tone after set alarm delay time (0 to 15 min) has elapsed Intermittent tone after set alarm delay time (0 to 30 min) has elapsed
Top screen is displayed. ----­Key Lock is displayed on the standby
operation setting screen.
-----
Intermittent tone
Open is displayed in the
Door
message display field.
after set alarm delay time (1 to 30 min) has elapsed
Err01: CO
the message display field.
Gas Empty is displayed in
2
Intermittent tone
N2 (or O
cylinder empty
) gas
2
CO2 line changeover
N2 (or O
changeover
) line
2
Connection mistake of CO2 line and N
(or O
2
line
2
Selection Mistake of N2 gas and O2 gas.
Chamber temperature sensor error
Sensor box temperature sensor error
Ambient temperature sensor error
CO2 sensor error
The O2 concentration does not increase when the N
valve is opened.
2/O2
The CO2 gas supply is switched automatically to the reserve cylinder (only when the MCO-21GC is installed).
The N2/O2 gas supply is switched automatically to the reserve cylinder.
The CO
concentration keeps
2
increasing when the CO closed.
concentration keeps keep
The O
2
changing into the opposite direction when the N
valve is opened.
2/O2
valve is
2
The chamber temperature sensor is disconnected.
The chamber temperature sensor is short-circuited.
The sensor box temperature sensor is disconnected.
The sensor box temperature sensor is short-circuited.
The ambient temperature sensor is disconnected.
The ambient temperature sensor is short-circuited.
The Vref or Vgas output voltage for the CO
sensor is abnormal.
2
Err02: N in the message display field.
Err01: CO
the message display field. The text flashes for the empty CO supply line. Err02: N in the message display field. The text flashes for the empty N gas supply line.
Err03: CO displayed in the message display field.
Err04: N displayed in the message display field.
Err05
Gas Empty is displayed
2/O2
Gas Empty is displayed in
2
Gas Empty is displayed
2/O2
Gas Connect Abnormal is
2
Gas Connect Abnormal is
2/O2
Temp Sensor Open
2/O2
gas
2
is displayed in the message display field.
Temp Sensor Short
Err06
is displayed in the message display field.
Err07: CO
is displayed in the message display
S Box Temp Sensor Open
2
field. Err08:CO2S Box Temp Sensor Short is
displayed in the message display field. Err09: AT Sensor Open is displayed in
the message display field. Err10: AT Sensor Short is displayed in
the message display field.
Err11: CO Err12: CO
displayed in the message display field.
Sensor Vref Abnormal or
2
Sensor Vgas Abnormal is
2
Intermittent tone
Remote alarm operates.
The CO is switched.
gas supply
2
Remote alarm operates.
2/O2
gas
The N supply is switched. Remote alarm operates. The CO closed. The N closed. The CO closed. The N closed.
valve is
2
2/O2
valve is
2
2/O2
valve is
valve is
The heater turns OFF. Remote alarm operates. The heater turns OFF. Remote alarm operates. CO2 valve is closed. Remote alarm operates.
valve is closed.
CO
2
Remote alarm operates.
Remote alarm operates.
Remote alarm operates.
valve is closed.
CO
2
Remote alarm operates.
55
Page 57
Alarm or safety
function
O2 sensor error
Main heater error
Bottom heater error
Door heater error
Sensor box heater error
Heater SSR burnout Low humidifying water
UV lamp failure
New UV lamp recommended
The measured voltage for the O
Main heater burnout occurs or the main heater SSR is short-circuited when the upper limit temperature alarm operates.
Bottom heater burnout occurs or the bottom heater SSR is short-circuited.
Door heater burnout occurs or the door heater SSR is short-circuited.
Sensor box heater burnout occurs or the sensor box heater SSR is short-circuited. Main, bottom, door, or sensor box heater SSR burnout occurs. The water in the humidifying pan drops to approx. 600 ml. The UV lamp burns out (only for the MCO-19M(UVH)/19M(UV) or when the optional MCO-19UVS is installed). The accumulated ON time reaches approx. 1,000 hours (only for the MCO-19M(UVH)/19M(UV) or when the optional MCO-19UVS is installed).
Conditions Display Buzzer Safety operation
N
2/O2
sensor is abnormal.
2
Err19: O
displayed in the message display field.
Err13: Main Heater Abnormal is displayed in the message display field.
Err14: Humidity Heater Abnormal
is displayed in the message display field.
Err15: Door Heater Abnormal is displayed in the message display field.
Err16: CO
displayed in the message display field. Err17: Heater SSR Open is displayed
in the message display field. RH PAN flashes in reverse video in the status display field.
Err18: UV Lamp Abnormal is displayed in the message display field.
Warning :UV Life is displayed in the message display field.
Sensor Abnormal is
2
S Box Heater Abnormal is
2
Intermittent tone
----- -----
Intermittent tone
----- -----
closed. Remote alarm operates
Remote alarm operates.
Remote alarm operates.
Remote alarm operates.
Remote alarm operates.
Remote alarm operates.
Remote alarm operates.
The buzzer can be stopped by pressing the BUZZER Key, but remote alarms will not stop. In addition, a
buzzer that sounds because of the upper limit temperature alarm cannot be stopped by pressing the
BUZZER Key.
Err01 recovery is achieved automatically by correctly connecting the gas and pressing the BUZZER Key
to stop the buzzer. In addition, when the MCO-21GC is installed and the gas supply is switched to the
reserve gas cylinder, press the BUZZER Key and replace the gas cylinder.
Err02 recovery is achieved automatically by correctly connecting the gas and pressing the BUZZER Key
to stop the buzzer. In addition, when the gas supply is switched to the reserve gas cylinder, press the
BUZZER Key and replace the gas cylinder.
For Err03, confirm whether a CO2 gas cylinder is correctly connected to a connecting port A/B for CO
gas pipe. Refer to page 16
For Err04, confirm whether a N
N
gas pipe. Refer to page 17
2/O2
(or O2) gas cylinder is correctly connected to a connecting port A/B for
2
If an error from Err05 to Err19 occurs (except for the upper limit temperature alarm in Err13 to Err16),
then consult a Sanyo sales representative or agent.
valve is
2
56
Page 58
Table 2: Alarm and Safety Functions for H
Alarm and safety
function
System check error at start of H2O2
decontamination
Failure during H2O2
decontamination
Failure during H2O gas analysis by UV
The H2O2 Generator is not connected.
There is no H2O2 solution or the H failed (or is disconnected).
A door is not closed.
The volume of H2O2 generated is abnormal.
Power was interrupted during H
The UV lamp failed during H
2
Power was interrupted during H
Condition Display Buzzer Safety operation
level sensor has
2O2
decontamination.
2O2
gas analysis by UV.
2O2
gas analysis by UV.
2O2
Decontamination
2O2
H2O2 Decon System Error is displayed in the screen center. Err31:H2O2 Unit No Connect is displayed in the message display field.
H2O2 Decon System Error is displayed in the screen center. Err32: H2O2 Level is Low is displayed in the message display field.
H2O2 Decon System Error displayed in the screen center. Err33: Door is Open is displayed in the message display field.
Err34:H2O2 Volume is displayed in the message display field.
After power is restored, Err35: Power Failed is displayed in the screen message display field. After completion, Decon Failed is displayed in the screen center.
Err36: UV Lamp Failure is displayed in the message display field.
After power is restored, Err37: Power Failed is displayed in the message display field.
-----
-----
-----
Intermittent tone with 15min delay
Intermittent tone
Intermittent tone
-----
Decontamination mode is cancelled.
Decontamination mode is cancelled.
Decontamination mode is cancelled.
Moves to UV analysis. Remote alarm operates with 15 min delay.
Electrically locked during power interruption. After power is restored, moves to UV analysis. Remote alarm operates.
UV analysis time is extended. Remote alarm operates.
Electrically locked during power interruption. After power is restored, UV analysis is repeated.
Table 3: Alarm and Safety Functions for automati c CO2 and O2 density calibration
Alarm and safety
function
System check error at start/operating of automatic CO2 and O2 density calibration
The gas pressure for the standard gas calibration CO gas line is less than the specified pressure.
Condition Display Buzzer Safety operation
Std Gas Calib Error is displayed in the screen center.
2
Err41: CO2 Std Gas Empty is displayed in the message display field.
Intermittent tone
The standard gas calibration mode is cancelled. Remote alarm operates.
Note
For models with electric locking, an unlock key is provided in order to unlock the door during a power
interruption or in case the electric lock fails. Always store this key in a safe place. It is recommended that
you make a note of the key symbol and number in case the key is lost.
WARNING
Do not use the unlock key to unlock the door during H2O2 decontamination or during H2O2 gas analysis by UV. Doing so may cause harm to health from H
gas leakage.
2O2
57
Page 59
CALIBRATION
Temperature/CO2/O2 calibration
1. Press the MENU Key from the Top Display to display the menu, select Tools, and press the ENTER
Key.
 
1
2
 
3
4
5
6
 
7
2. Select Temp/CO2/O2 Calibration from the Select Tools Display, press the MENU Key to display the menu, select OK, and press the ENTER Key.
        
3. The Temp/CO2/O2 Calibration Display will appear.
Example
Enter the value shown below if the displayed chamber temperature is 37.0
is 36.5
Temperature span: Enter 365.
 
1
 
3
 
4
5
  
6
T e m p 3 7 . 0o C C O 2 5 . 0 %
37.
S t a t u s
O K
S e l e c t T o o l s ( 1 / 2 ) T e m p / C O 2 / O 2 C a l i b r a t i o n
A l a r m Se t t i n g L C D / D A Q S e t t i n g U V S e t t i n g K e y L o ck P W S etting D a t e T i m e
o
C.
T e m p / C O 2 / O 2 C a l i b r a t i o n P V
T e m p S p a n 3 6 . 5oC37 . 0oC 0 . 5°C C O 2 Z e r o 3 . 9 1 V 5 0 0 C O 2 S p a n 0 . 0 % 5 . 0 % 3 . 1 6 V 7 0 0 O 2 Z er o 5 0 0 O 2 S p a n 0 . 0 % 5 . 0 % 5 8 m V 5 0 0
.
O 2 5 . 0 %
.
.
Volt
M E N U S e t L o g
T o o l s
M E N U
O K
S v c C a n c e l
o
C but the actual temperature
58
Page 60
Example
Enter the value shown below if the displayed chamber CO
density is 5.0% but the actual CO2 density is
2
4.5%.
CO
density span: Enter 045.
2
 
T e m p / C O 2 / O 2 C a l i b r a t i o n
1
 
P V
3
T e m p S p a n 0 . 0oC37 . 0oC 0 . 5°C
C O 2 Z e r o 3 . 9 1 V 5 0 0
4
C O 2 S p a n 4 . 5 % 5.0% 3.16 V 7 0 0
5
O 2 Z er o 5 0 0
 
O 2 S p a n 0 . 0 % 5 . 0 % 5 8 m V 5 0 0
6
Volt
Example
Enter the value shown below if the displayed chamber O
density is 5.0% but the actual O2 density is
2
5.5%.
O
density span: Enter 055.
2
 
T e m p / C O 2 / O 2 C a l i b r a t i o n
1
 
P V
3
T e m p S p a n 0 . 0oC37 . 0oC 0 . 5°C
C O 2 Z e r o 3 . 9 1 V 5 0 0
4
C O 2 S p a n 0 . 0 % 5 . 0 % 3 . 1 6 V 7 0 0
5
O 2 Z er o 5 0 0
 
O 2 S p a n 5 . 5 % 5 . 0 % 5 8 m V 5 0 0
6
Volt
4. Press the MENU Key to display the menu, select OK, and press the ENTER Key.
5. After the Select Tools Display is displayed, press MENU Key to display the menu, select Cancel, and
press the ENTER Key to return to the Top Display.
Note:
It is important to accurately measure the temperature inside the Incubator when calibrating the
temperature. Particularly, the temperature gauge used must have an accuracy of JIS 0.5 Class or better.
The temperature must be measured at several points.
The temperature setting can not change by more than ±1.0
o
C during calibration. If these values are
exceeded, an error buzzer will sound, the input data will be ignored, and the Incubator will return to the
o
display mode. If it is necessary to change the temperature by more than 1.0
C, perform calibration in
several stages over a period of time.
59
Page 61
TROUBLESHOOTING
If the Incubator does not seem to be working properly, check the following items before calling for service.
Symptom Items to check and countermeasures
The Incubator does not operate at all.
Is the Incubator plugged in?
Is there a power outage, or has a circuit breaker interrupted the
power?
The key is not working.
Is the key lock function turned OFF?
An alarm has occurred. When starting operation:
Does the chamber temperature match the set value?
Does the CO
gas concentration in the chamber match the set
2/O2
value?
(1) Is the secondary pressure for the CO
specified value of 0.03 MPaG (0.3 kgf/cm
(2) Is the secondary pressure for the N
specified value of 0.05 MPaG (0.5 kgf/cm
(3) Is the gas tube properly connected?
During operation:
Is the upper limit alarm temperature set at least 1°C higher than the chamber set temperature?
Has the temperature setting been changed? Has the door been left open for a long time? Has a low-temperature object been placed in the chamber? If any of these is the case, the alarm will be automatically cleared if you wait.
Has the gas tube come loose, or is there a gas leak?
gas pressure regulator at the
2
2
G. 4.3 psiG)?
gas pressure regulator at the
2/O2
2
G, 7.1 psiG)?
The chamber temperature does not match the set value.
The gas concentration does not match the set value.
The chamber temperature does not rise.
Has the CO
/O2 gas concentration setting been changed?
2
Is the gas cylinder empty? Check the primary pressure of the gas cylinder once a week. (When the primary pressure is 3.8 MPa (38 kgf/cm
2
G) or lower, it is a sign that there is little gas remaining.
Replace the cylinder soon.)
Is the ambient temperature less than 5°C different from the set value for the chamber temperature?
Is the outer door closed with the gastight split doors or inner door left open?
Is the secondary pressure for the CO specified value of 0.03 MPaG (0.3 kgf/cm
Is the secondary pressure for the N specified value of 0.05 MPaG (0.5 kgf/cm
gas pressure regulator at the
2
2
G, 4.3 psiG)?
gas pressure regulator at the
2/O2
2
G, 7.1 psiG)?
Is the gas tube blocked?
Is there enough water in the humidifying pan? (Be sure to use sterile
distilled water.)
60
Page 62
Symptom Items to check and countermeasures
A large quantity of gas is being consumed.
Are the doors being frequently opened and closed?
Check whether gas is leaking from connectors due to deterioration
of the gas tube, or whether there may be any pinhole leaks. The gas tube is a replaceable part, and it is recommended that it be
replaced once a year.
Is the packing seal for gastight split doors and inner door defective?
Is the access hole open?
Normal cultures are not possible, and the gas concentration is suspect.
The gas concentration is slow to recover.
Is the ambient air environment around the Incubator normal? Is there a source of polluted gas in the vicinity?
A HEPA filter is used for the Incubator gas piping. If gas concentration is slow to recover when the CO
gas concentration is
2
normal, it is possible that the HEPA filter may be clogged. Consult a Sanyo representative or agent.
Is there little gas remaining in the gas cylinder?
Is the secondary pressure for the CO
specified value of 0.03 MPaG (0.3 kgf/cm
Is the secondary pressure for the N the specified value of 0.05 MPaG (0.5 kgf/cm
gas pressure regulator at the
2
2
G, 4.3 psiG)?
/CO2 gas pressure regulator at
2
2
G, 7.1 psiG)?
Is the gas tube blocked?
The gas is not being injected.
The control method for the Incubator is the ON-OFF method. Gas is intermittently injected as the gas concentration in the chamber approaches the set value. Injections may be stopped for periods of approximately 60 seconds, but that is not an error.
The outer door does not open. MCO-19M(UVH), or when the optional MCO-HL is installed:
When the power is OFF, the electric lock is locked and the outer door does not open. Either turn ON the power or use the accessory unlock key to override the electric lock.
During decontamination the outer door is electrically locked and does not open.
H2O2 decontamination cannot be performed.
If the MCO-19M(UV)/19M is being used, is the optional MCO-HL installed?
If the MCO-19M is being used, are the optional MCO-HL and MCO-19UVS installed?
Is the UV lamp burned out? If the UV lamp is burned out, H decontamination will not be possible.
Is the H
Has the entire bottle of H
Has
the upper limit temperature alarm been set to 50°C or higher?
Generator cable properly connected?
2O2
Decontamination Reagent been used?
2O2
Note: If the problem still has not been solved after trying the above checks and countermeasures, or for
any problems not covered here, contact a Sanyo representative or agent.
61
2O2
Page 63
DISPOSING OF THE MULTI-GAS INCUBATOR
When disposing of the multi-gas incubator, contact a Sanyo sales representative or agent.
WARNING
The incubator must be dismantled and disposed of by qualified personnel only. If the incubator is
left where outsiders enter, it may result unexpected accident (for example, children to become locked
inside).
Before disposing a multi-gas incubator with biohazardous danger, decontaminate the multi-gas incubator
to the extent possible by the user.
62
Page 64
Note: This symbol mark and recycle system apply only to EU countries. They do not applied to the countries in other areas of the world.
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC
(English)
This SANYO product is designed and manufactured with high-quality materials and components that can be recycled and reused.
This symbol means that the electrical and electronic equipment, at their end-of-life, should be disposed of separately from household wastes.
Dispose of this product at your local community waste collection/recycling center.
In the European Union there are separate collection systems for used electrical and electronic products.
Please help us to conserve the environment we live in.
(German)
Ihr SANYO Produkt wurde entworfen und hergestellt mit qualitativ hochwertigen Materialien und Komponenten, die recycelt und wiederverwendet werden können.
Dieses Symbol bedeutet, daß elektrische und elektronische Geräte am Ende ihrer Nutzungsdauer von Hausmüll getrennt entsorgt werden sollen.
Bitte entsorgen Sie dieses Gerät bei Ihrer örtlichen kommunalen Sammelstelle oder im Recycling Centre.
In der Europäischen Union gibt es unterschiedliche Sammelsysteme für Elektrik- und Elektronikgeräte.
Helfen Sie uns bitte, die Umwelt zu erhalten, in der wir leben!
63
Page 65
(French)
Votre produit Sanyo est conçu et fabriqué avec des matèriels et des composants de qualité supérieure qui peuvent être recyclés et réutilisés.
Ce symbole signifie que les équipements électriques et électroniques en fin de vie doivent être éliminés séparément des ordures ménagères.
Nous vous prions donc de confier cet équipement à votre centre local de collecte/recyclage. Dans l’Union Européenne, il existe des systèmes sélectifs de collecte pour les produits électriques et électroniques usagés.
Aidez-nous à conserver l’environnement dans lequel nous vivons !
Les machines ou appareils électriques et électroniques contiennent fréquemment des matières qui, si elles sont traitées ou éliminées de manière inappropriée, peuvent s’avérer potentiellement dangereuses pour la santé humaine et pour l’environnement. Cependant, ces matières sont nécessaires au bon fonctionnement de votre appareil ou de votre machine. Pour cette raison, il vous est demandé de ne pas vous débarrasser de votre appareil ou machine usagé avec vos ordures ménagères.
(Spanish)
Los productos SANYO están diseñados y fabricados con materiales y componentes de alta calidad, que pueden ser reciclados y reutilizados.
Este símbolo significa que el equipo eléctrico y electrónico, al final de su ciclo de vida, no se debe desechar con el resto de residuos domésticos.
Por favor, deposite su viejo “televisor” en el punto de recogida de residuos o contacte con su administración local.
En la Unión Europea existen sistemas de recogida específicos para residuos de aparatos eléctricos y electrónicos.
Por favor, ayúdenos a conservar el medio ambiente!
64
Page 66
(Portuguese)
O seu produto SANYO foi concebido e produzido com materiais e componentes de alta qualidade que podem ser reciclados e reutilizados.
Este símbolo significa que o equipamento eléctrico e electrónico no final da sua vida útil deverá ser descartado separadamente do seu lixo doméstico.
Por favor, entregue este equipamento no seu ponto local de recolha/reciclagem.
Na União Europeia existem sistemas de recolha separados para produtos eléctricos e electrónicos usados.
Por favor, ajude-nos a conservar o ambiente em que vivemos!
(Italian)
Il vostro prodotto SANYO è stato costruito da materiali e componenti di alta qualità, che sono riutilizzabili o riciclabili.
Prodotti elettrici ed elettronici portando questo simbolo alla fine dell’uso devono essere smaltiti separatamente dai rifiuti casalinghi.
Vi preghiamo di smaltire questo apparecchio al deposito comunale. Nell’Unione Europea esistono sistemi di raccolta differenziata per prodotti elettrici ed elettronici.
Aiutateci a conservare l’ambiente in cui viviamo!
65
Page 67
(Dutch)
Sanyo producten zijn ontwikkeld en gefabriceerd uit eerste kwaliteit materialen, de onderdelen kunnen worden gerecycled en weer worden gebruikt.
Het symbool betekent dat de elektrische en elektronische onderdelen wanneer deze vernietigd gaan worden , dit separaat gebeurt van het normale huisafval.
Zorg ervoor dat het verwijderen van de apparatuur bij de lokaal erkende instanties gaat gebeuren. In de Europese Unie wordt de gebruikte elektrische en elektronische apparatuur bij de daarvoor wettelijke instanties aangeboden.
Alstublieft help allen mee om het milieu te beschermen.
(Swedish)
Din SANYO produkt är designad och tillverkad av material och komponenter med hög kvalitet som kan återvinnas och återanvändas.
Denna symbol betyder att elektriska och elektroniska produkter, efter slutanvändande, skall sorteras och lämnas separat från Ditt hushållsavfall.
Vänligen, lämna denna produkt hos Din lokala mottagningstation för avfall/återvinningsstation.
Inom den Europeiska Unionen finns det separata återvinningssystem för begagnade elektriska och elektroniska produkter.
Vänligen, hjälp oss att bevara miljön vi lever i!
66
Page 68
AUTOMATIC CO2 CYLINDER CHANGEOVER
An automatic CO
system switches the gas supply lines when one CO
Note: The installation of the MCO-21GC must be performed only by qualified service personnel.
Install the MCO-21GC and then perform the following procedure
1. Connect 2 gas tubes to connecting ports A/B for CO located on the left back of the Incubator. (Refer to page 9). Note: Only port A can be used when the MCO-21GC is not installed. There will be 2 ports, ports A and B, after the MCO-21GC is installed.
2. Connect a CO
for information on connecting the gas cylinders. (“Gas cylinder A” is the CO2 gas cylinder connected to
the connecting port A for CO
connecting port B for N
3. Open the valves of both gas cylinders.
4. The current CO
(“A” and “B” will be displayed only when the MCO-21GC is installed. Nothing will be displayed if
MCO-21GC is not installed.)
    
1
 
2
3
4
5
 
6
7
5. The MCO-21GC detects that there is no more CO
chamber does not increase for a while even after opening the CO
the gas supply line when it detects that there is no more gas. At that time, the following indications of
switching the gas cylinder are provided.
When CO2 gas supply line switches
automatically (Only when MCO-21GC
is installed)
To silence the buzzer, press the BUZZER Key. The alarm condition is cleared and normal operation
continues.
T e m p 3 7 . 0o C C O 2 5 . 0 % A B O 2 5 . 0 % A B
37.
S t a t u s
O K
cylinder changeover system (MCO-21GC) is available as an optional accessory. This
2
gas cylinder becomes empty.
2
gas pipe. Connecting ports for CO2 gas pipes are
2
gas cylinder equipped with a gas pressure regulator to each gas tube. Refer to page 16
2
gas pipe, and “gas cylinder B” is the CO2 gas cylinder connected to the
2
gas pipe.)
2/O2
gas supply line is displayed on the Top Display.
2
Current CO2 gas supply line
.
gas in a cylinder when the CO2 density in the
2
2
Message display field Display of gas supply
Err01: CO
displayed.
Gas Empty is
2
.
.
gas valve in the Incubator. It switches
The character for gas supply line
that is empty blinks.
The gas supply line that is
currently being used is displayed
in reverse video.
67
Page 69
6. When gas cylinder A is switched to B, remove gas cylinder A and replace it with new one. Note: Be careful when handling the empty CO
When gas cylinder B becomes empty, the line is switched to gas cylinder A again.
The MCO-21GC detects that there is no more CO
does not increase for a while even after opening the CO
also be switched for other reasons even though there is gas left in the cylinder: blocking or restricting of
gas tube, reduction of CO
quantity in the cylinder before disconnecting it.
Use the following procedure to manual switch the line between CO
MCO-21GC is installed). This procedure shows switching the line fromCO example.
1. From the Top Display, press the Left Cursor Key for 3 seconds.
  
1
2
3
4
 
5
6
7
 
2. The CO
 
1
2
 
3
4
5
6
 
7
(When the line needs to be switched from cylinder A to B, press the Right Cursor Key for 3 seconds.)
T e m p 3 7 . 0o C C O 2 5 . 0 % A B O 2 5 . 0 % A B
37.
S t a t u s
O K
gas supply line will be switched from cylinder B to A
2
T e m p 3 7 . 0o C C O 2 5 . 0 % A B O 2 5 . 0 % A B
37.
S t a t u s
O K
gas pressure, or improper opening of CO2 gas cylinder. Always check the gas
2
.
.
gas cylinder. Some gas may still be left in the cylinder.
2
gas in a cylinder when the CO2 density in the chamber
2
gas valve in the Incubator. The supply line may
2
cylinders A and B (only when
2
cylinder B to A as an
2
.
.
.
.
68
Page 70
AUTOMATIC CO2 AND O2 DENSITY CALIBRATION
To calibrate the CO
By connecting a standard gas cylinder that has a density that is the same as the set CO
(usually 5% CO
the CO
Setting connected standard gas density
1. Press the MENU Key from the Top Display to display the menu, select Tools, and press the ENTER
Key.
          
2. Press the Down Cursor Key from the Select Tools (1/2) Display to display the Select Tools (2/2) Display, and then select Standard Gas Setting, press the MENU Key to display the menu, select OK, and press
the ENTER Key.
                       
and O2 density display as necessary.
2
1
2
3
4
5
6
7
T e m p 3 7 . 0o C C O 2 5 . 0 %
37.
S t a t u s
O K
S e l e c t T o o l s ( 1 / 2 ) T e m p / C O2 / O 2 C alibration
A l a r m Se t t i n g L C D / D A Q S e t t i n g U V S e t t i n g K e y L o ck P W S etting D a t e T i m e
S e l e c t T o o l s ( 2 / 2 ) S t a n d a r d G a s S e t t i n g
S t a n d a r d G a s C a l i b r a t i o n
and O2 density, a Semi-automatic one point Gas Calibration Kit (MCO-SG) is available.
2
and O2 density
2
and 5% O2), the CO2 sensor and O2 sensor can be calibrated to maintain the accuracy of
2
.
O 2 5 . 0 %
.
.
M E N U S e t L o g
T o o l s
M E N U
O K
S v c C a n c e l
M E N U
O K
S v c C a n c e l
69
Page 71
3. Input the CO
select OK, and press the ENTER Key.
 
S t a n d a r d G a s S e t t i n g
1
 
3
S t d G a s C O 2 D e n s i t y 5 . 0 % ( 4 . 0 % - 2 1 . 0 % )
( N o Calibrat i o n : 0 . 0 % )
4
S t d G a s O 2 D e n s i t y 5 . 0 % ( 0 . 5 % - 1 8 . 0 % )
5
( 2 2 . 0 % - 8 1 . 0 % )
( N o Calibration:18.1 % - 2 1 . 9%)
6
Ex. Standard gas density input
Standard gas density
Calibrate both of CO2 and O2 density CO2:5% / O2:5% 050 050
Calibrate only CO2 density CO2:5% / O2:20% 050 200
Calibrate only O2 density CO2:0% / O2:5% 000 050
Calibration method for CO
1. Select Standard Gas Calibration from the Select Tools (2/2) Display, press the MENU Key to display the menu, select OK, and press the ENTER Key.
 
S e l e c t T o o l s ( 2 / 2 )
S t a n d a r d G a s S e t t i n g
S t a n d a r d G a s C a l i b r a t i o n
      
2. The system check will start.
 
S t a n d a r d G a s C a l i b r a t i o n S t e p : 1
1
3
S t d G a s C a l i b S y s t e m C h e c k ・・・・・
 
4
CO2 5. 5% O 2 4 . 5 %
5
STG 5. 0% STG : 5 . 0 %
 
6
 
and O2 density of standard gas to be used, press the MENU Key to display the menu,
2
M E N U
O K
C a n c e l
and O2 gas density
2
(For example)
Input value of
CO
density
2
Input value of
O2 density
M E N U
O K
S v c C a n c e l
70
Page 72
3. If the system is normal, the following display will appear. Select OK and press the ENTER Key to start
calibrating the CO
                                           
S t a n d a r d G a s C a l i b r a t i o n S t e p 2
1
3
S t d G a s C alib Start?
4
CO2 5. 5% O 2 4 . 5 %
5
STG 5. 0% ST G : 5 . 0 %
6
S t a n d a r d G a s C a l i b r a t i o n S t e p 3
1
3
S t d G a s C a l i b r a t i n g ・・・・・
4
CO2 5. 5% O 2 4 . 5 %
5
STG 5. 0% ST G : 5 . 0 %
6
1
2
3
4
5
6
7
T e m p 3 7 . 0 o C C O 2 5 . 0 % A B O 2 5 . 0 % A B
37.
S t a t u s
O K
and O2 density. After calibration, return to normal cultivating operation.
2
M E N U
O K C a n c e l
.
.
.
71
Page 73
STACKING INCUBATORS
Use the following procedure to stack Incubators. This work is potentially dangerous, so contact a Sanyo
representative or agent.
CAUTION
Select a floor that is strong enough to support the stacked incubators.
Never stack three or more incubators. Doing so is dangerous.
1. Take out two stacking plates A, two stacking plates B, four protective stickers, and four screws from
accessory bag.
Note:
When stacking two incubators, use two stacking plates (A and B) for each one. Only one pair of
stacking plates is provided per incubator, so take out the stacking plates (A and B) from both incubators.
Similarly only 2 protective stickers are provided per incubator
2. Unplug both of the Incubators that are to be stacked.
3. Remove the caps at the two places in the front of the top surface of the Incubator that is to be on the
bottom, attach the protective stickers to the four corners on the top surface of the Incubator that is to be
on the bottom. (See Fig. A.)
4. Secure stacking plate A with the screws that are provided.
5. Remove the front panel (5 screws) from the Incubator that is to be on the top.
6. Place the upper Incubator on top of the lower one, being careful not to pinch your fingers.
7. Make the lower Incubator level by adjusting the legs, and then level the upper Incubator in the same
way.
8. Secure the stacking plate A fixed to the lower incubator to the upper Incubator with the accessory
screws.
9. Remove from the back of the lower Incubator the two hooks (4 screws) and remove from the upper
Incubator the 2 screws on the lower rear panel.
10. Use the 6 screws that were removed in step 9 to secure stacking plate B to the rear panels of the
upper and lower Incubators.
11. Replace the front panel (5 screws) to the upper Incubator.
12. Use the hooks on the back of the upper Incubator to prevent the Incubators from falling over.
72
Page 74
Front panel
Front
Stacking plate A
Stacking plate B
Protective sticker
Hook
Fig.A
73
Page 75
SPECIFICATIONS
Name Multi-gas Incubator
Models MCO-19M MCO-19M(UV) MCO-19M(UVH)
External dimensions W620 x D710 x H900 mm (W24.4 x D28.0 x H35.4 inch)
Internal dimensions W490 x D523 x H665 mm (W19.3 x D20.6 x H26.2 inch)
Interior volume 162 L (5.72 cu.ft.)
Exterior Painted steel
Interior Stainless steel containing copper
Outer door Painted steel Inner door Stainless steel with gasket
Gastight split door 4 doors made of tempered glass
Trays 3 trays made of stainless steel containing copper
W450 x D450 x H12 mm (W17.7 x D17.7 x H0.47 inch), Maximum load: 7 kg/tray Access port Inner diameter: 30 mm (1.18 inch), On the back side
Insulation Rigid polyurethane foamed-in place
Heating system DHA system (heater jacket + air jacket system)
Heater 314 W
Humidifying system Natural evaporation with humidifying pan
Temperature controller PID control system
Temperature display Digital display
CO2 controller PID control system
CO2 density display Digital display
O2 controller PID control system
O2 density display Digital display
Air circulation Fan assisted
Air filter 0.3 µm, Efficiency: 99.97% or higher
UV lamp
Water level sensor
Alarms
Remote alarm contacts Allowable contact capacity: 30 VDC, 2 A
CO2 inlet connection 4 to 6 mm (0.157 to 0.236 inch) diameter tube can be connected
CO2 inlet pressure 0.03 MPaG (0.3 kgf/cm2G, 4.3 psiG)
O2 inlet connection 4 to 6 mm diameter tube can be connected
O2 inlet pressure 0.05 MPaG (0.5 kgf/cm2G, 7.1 psiG)
Weight 94 kg
Automatic set temperature alarm, Automatic set CO2 density alarm
Automatic set O2 density alarm, Upper limit temperature alarm
----- 4W X 1 (ozone-free emission)
Optical
various gas/sensor/heater alarms
74
Page 76
Name Multi-gas Incubator
Models MCO-19M MCO-19M(UV) MCO-19M(UVH)
3 trays, 3 sets of tray supports, 3 gas tubes, 1 humidifying pan,
Accessories
1 pair of stacking plates A and B, 2 protective stickers, 6 tube bands
H
2O2
UV Lamp Add-on Kit
(MCO-19UVS)
Optional accessories
Half Trays (MCO-25ST), Automatic CO
Roller Base (MCO-18RB), UV lamp replacement kit (MCO-20UV)
Stacking spacer (MCO-18SB, when stacking on the MCO-175),
Stacking spacer (MCO-21SB, when stacking on the MCO-20AIC)
Ethernet (LAN) Interface (MTR-L03), RS232C/RS485 Interface (MTR-480)
Note: Designs and specifications are subject to change without notice.
1 gas injection nozzle, 1 gas injection nozzle tube
H
Generator (MCO-HP)
2O2
Decontamination Kit (MCO-HL)
Standard equipment
H2O2 Decontamination Reagent (MCO-H2O2),
Automatic CO
Standard Gas Calibration Kit (MCO-SG),
2
Gas Pressure Regulator (MCO-100L),
Extra Trays (MCO-47ST, includes 2 tray supports),
Cylinder Changeover System (MCO-21GC),
2
SANYO Data acquisition software (MTR-5000)
4-20mA Interface (MCO-420MA, USA only)
Standard equipment
PERFORMANCE
Temperature control range
Temperature distribution
Temperature variation
CO2 control range
CO2 variation
O2 control range
O2 variation
Chamber humidity
Applicable environment condition
Noise level
Power consumption
Heat emission Max. 1,274 kJ/h Rated voltage, frequency Amperage
Note: The unit with CE mark complies with EC directives.
Based on our measuring method.
Designs and specifications are subject to change without notice.
Ambient temperature+5 to 50
o
± 0.25
o
± 0.1
C (ambient temperature: 25oC, setting: 37oC, CO2: 5%, O2: 5%, no load)
± 0.15% (ambient temperature: 25
± 0.2% (ambient temperature: 25
Max. 354 W Max. 335 W
110 to 120 V, 60 Hz 220 V, 60 Hz 220 to 240 V, 50 Hz
Max. 3.1 A Max. 1.5 A Max. 1.5 A
o
C (ambient temperature: 5 to 35oC)
C (ambient temperature: 25oC, setting: 37oC, CO2: 5%, O2: 5%, no load)
0 to 20%
o
C, setting: 37oC, CO2: 5%, O2: 5%, no load)
1 to 18% and 22 to 80%
o
C, setting: 37oC, CO2: 5%, O2: 5%, no load)
95% ± 5%
Temperature: 5 to 35
o
C, Humidity: 80% max.
(The designed performance may not be obtained
If the ambient temperature is equal or less than 15
o
C)
30 dB (A scale)
Max. 1,206 kJ/h
75
Page 77
CAUTION
Please fill in this form before servicing.
Hand over this form to the service engineer to keep for his and your safety.
Safety check sheet
1. Unit contents Risk of infection: Risk of toxicity: Risk from radioactive sources:
(List all potentially hazardous materials that have been stored in this unit.) Notes :
2.
Contamination of the unit
Unit interior No contamination Decontaminated Contaminated Others:
3. Instructions for safe repair/maintenance of the unit a) The unit is safe to work on b) There is some danger (see below) Procedure to be adhered to in order to reduce safety risk indicated in b) below.
Date : Signature :
Address, Division :
Telephone :
Yes No
Yes NoYes No
Yes NoYes No
Yes No
Yes NoYes NoYes No
Yes NoYes No
Product name :
Multi-gas incubator
Please decontaminate the unit yourself before calling the service engineer.
Model :
MCO-19M(UVH)
MCO-19M(UV)
MCO-19M
Serial number : Date of Installation :
76
Page 78
7FB6P101569000
(
26 August 2010)
5-5, Keihan-Hondori 2-Chome Moriguchi City, Osaka 570-8677 Japan
Printed in Japan
Loading...