■ Read this manual carefully before using the appliance and follow the instructions for safety operation.
■ Sanyo never guarantee any safety if the appliance is used for any objects other than intended use or
used by any procedures other than those mentioned in this manual.
■ Keep this manual in an adequate place to refer to it as necessary.
■ The contents of the manual will be subjected to change without notice due to the improvement of
performance or functions.
■ Contact Sanyo sales representative or agent if any page of the manual is lost or page order is incorrect.
■ Contact Sanyo sales representative or agent if any point in this manual is unclear or if there are any
inaccuracies.
■ No part of this manual may be reproduced in any form without the expressed written permission of
Sanyo.
CAUTION
SANYO guarantees the product under certain warranty conditions. SANYO in no way shall be
responsible for any loss of content or damage of content.
- 51 -
2
Page 4
PRECAUTIONS FOR SAFE OPERATION
It is imperative that the user complies with this manual as it contains
important safety advice.
Items and procedures are described so that you can use this unit correctly and safely.
If the precautions advised are followed, this will prevent possible injury to the user and
any other person.
Precautions are illustrated in the following way:
WARNING
Failure to observe WARNING signs could result in a hazard to personnel
possibly resulting in serious injury or death.
CAUTION
Failure to observe CAUTION signs could result in injury to personnel and
damage to the unit and associated property.
Symbol shows;
this symbol means caution.
this symbol means an action is prohibited.
this symbol means an instruction must be followed.
Be sure to keep this manual in a place accessible to users of this unit.
< Label on the unit >
This mark is labeled on the cover in which the electrical components of high voltage
are enclosed to prevent the electric shock.
The cover should be removed by a qualified engineer or a service personnel only.
WARNING
As with any equipment that uses CO2 gas, there is a likelihood of oxygen depletion in the vicinity
of the equipment. It is important that you assess the work site to ensure there is suitable and
sufficient ventilation. If restricted ventilation is suspected, then other methods of ensuring a
safe environment must be considered. These may include atmosphere monitoring and
warning devices.
3
- 52 -
Page 5
WARNING
Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to
rain water.
Only qualified engineers or service personnel should install the unit. The installation by
unqualified personnel may cause electric shock or fire.
Install the unit on a sturdy floor and take an adequate precaution to prevent the unit from
turning over. If the floor is not strong enough or the installation site is not adequate, this may result
in injury from the unit falling or tipping over.
Never install the unit in a humid place or a place where it is likely to be splashed by water.
Deterioration of the insulation may result which could cause current leakage or electric shock.
Never install the unit in a flammable or volatile location. This may cause explosion or fire.
Never install the unit where acid or corrosive gases are present as current leakage or electric
shock may result due to corrosion.
Always ground (earth) the unit to prevent electric shock. If the power supply outlet is not
grounded, it will be necessary to install a ground by qualified engineers.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such
grounding may cause electric shock in the case of an incomplete circuit.
Connect the unit to a power source as indicated on the rating label attached to the unit. Use
of any other voltage or frequency other than that on the rating label may cause fire or electric shock.
Never store volatile or flammable substances in this unit if the container cannot be sealed. These
may cause explosion or fire.
Do not insert metal objects such as a pin or a wire into any vent, gap or any outlet on the unit.
This may cause electric shock or injury by accidental contact with moving parts.
Use this unit in safe area when treating the poison, harmful or radiate articles. Improper use
may cause bad effect on your health or environment.
Turn off the power switch (if provided) and disconnect the power supply to the unit prior to
any repair or maintenance of the unit in order to prevent electric shock or injury.
Do not touch any electrical parts (such as power supply plug) or operate switches with a wet
hand. This may cause electric shock.
- 53 -
4
Page 6
WARNING
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of
maintenance. These may be harmful to your health.
Never splash water directly onto the unit as this may cause electric shock or short circuit.
Never put containers with liquid on the unit as this may cause electric shock or short circuit when
the liquid is spilled.
Never bind, process, or step on the power supply cord, or never damage or break the power
supply plug. A broken supply cord or plug may cause fire or electric shock.
Do not use the supply cord if its plug is loose. Such supply cord may cause fire or electric shock.
Never disassemble, repair, or modify the unit yourself. Any such work carried out by an
unauthorized person may result in fire, or electric shock or injury due to a malfunction.
Disconnect the power supply plug if there is something wrong with the unit. Continued
abnormal operation may cause electric shock or fire.
When removing the plug from the power supply outlet, grip the power supply plug, not the cord.
Pulling the cord may result in electric shock or fire by short circuit.
Disconnect the power supply plug before moving the unit. Take care not to damage the power
cord. A damaged cord may cause electric shock or fire.
Disconnect the power plug when the unit is not used for long periods. Keeping the connection
may cause electric shock, current leakage, or fire due to the deterioration of insulation.
If the unit is to be stored unused in an unsupervised area for an extended period, ensure that
children do not have access and that doors cannot be closed completely.
The disposal of the unit should be accomplished by appropriate personnel. Remove doors to
prevent accidents such as suffocation.
Do not put the packing plastic bag within reach of children as suffocation may result.
5
- 54 -
Page 7
CAUTION
Use a dedicated power source (a dedicated circuit with a breaker) as indicated on the rating label
attached to the unit. A branched circuit may cause fire resulting from abnormal heating.
Connect the power supply plug to the power source firmly after removing the dust on the plug.
A dusty plug or improper insertion may cause a heat or ignition.
Never store corrosive substances such as acid or alkali in this unit if the container cannot be
sealed. These may cause corrosion of inner components or electric parts.
Check the setting when starting up of operation after power failure or turning off of power
switch. The stored items may be damaged due to the change of setting.
Be careful not to tip over the unit during movement to prevent damage or injury.
Prepare a safety check sheet when you request any repair or maintenance for the safety of service
personnel.
LABELS ON THE INCUBATORS
Warning and caution labels are attached to the incubator. The following table describes the labels.
This label is attached to covers that access high-voltage electrical components to
prevent electric shock. Only a qualified engineer or service personnel should be
allowed to open these covers.
This symbol indicates an ultraviolet light (UV) caution.
This symbol indicates that caution is required. Refer to product documentation for
details.
This symbol indicates a hot surface.
This symbol indicates an earth.
This symbol means “ON” for a power switch.
This symbol means “OFF” for a power switch.
- 55 -
6
Page 8
ENVIRONMENTAL CONDITIONS
This equipment is designed to be safe at least under the following conditions (based on the IEC
61010-1):
■ Indoor use;
■ Altitude up to 2000 m;
■ Ambient temperature 5
■ Maximum relative humidity 80% for temperature up to 31
o
at 40
C;
■ Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage;
■ Other supply voltage fluctuations as stated by the manufacturer;
■ Transient overvoltages according to Installation Categories (Overvoltage Categories) II; For mains
supply the minimum and normal category is II;
■ Pollution degree 2 in accordance with IEC 60664.
o
C to 40oC
o
C decreasing linearly to 50% relative humidity
7
- 56 -
Page 9
INCUBATOR COMPONENTS
Handles
5
13
7
8
9 (inside)
6
2
1
19,20 (inside)
For MCO-18AC(UV) or
when an optional UV system kit
MCO-18UVS3 is installed.
18
15
Handles
4
10, 11
For MCO-18AC(UV) or
when an optional UV system kit
MCO-18UVS3 is installed.
14
21
3
17
16
12
Rear Right Side
- 57 -
Rear Left Side
8
Page 10
1. Outer door: Sticks to frame with magnetic seal. Door heater is installed in the outer door panel.
The door opening is reversible. Contact Sanyo representative or agent to change the door hinge from
left to right or vice versa.
2. Inner door: Made of tempered glass, however avoid excessive impact on the glass.
3. Leveling foot: Screw type for adjusting the height. Adjust the foot so that the incubator can be
level.
4. Tray: Can be pulled toward you.
5. Tray support: Can be removed by lifting the front side and pulling toward you.
6. Side support: Right and left side supports can be removed for disinfection. See page 27, 28 and
29.
7. Fan cover: Inlet of circulating air. Removable.
8. Duct: Flow path for circulating air. Removable.
9. Fan (inside the rear duct): Made from polypropylene resin. Can be disinfected by an autoclave.
10. Sample air outlet: This also functions as an internal gas outlet. Normally, cover this outlet with
the sample air outlet cap.
11. Sample air outlet cap: Always attach this cap except at the time of using of sample air outlet.
Note: Operation without sample air outlet cap may result low humidity or condensation on the outer door.
12. Connecting port for CO
gas pipe: When an optional component MCO-21GC (automatic CO
2
2
cylinder changeover system) is installed, both A and B are available. If MCO-21GC is not used, only A
is available. Refer to page 14 for gas cylinder connection. Ensure that the gas pressure is set at 0.03
MPaG (0.3 kgf/cm
2
G, 4.3 psiG). Refer to page 39 for automatic CO2 cylinder changeover system.
13. Door switch: Detects the door opening/closing and stops the circulating fan, UV lamp* and
electromagnetic valve for CO
when door is open.
2
14. Humidifying pan: Use sterile distilled water to fill the pan. (MCO-18AC(UV) or when an optional
UV system kit MCO-18UVS3 is installed) Install the pan properly so that it can be covered with the pan
cover.
15. Remote alarm terminal: Refer to page 11.
16. Access port: When not in use, cap with the silicon caps on both outside and inside.
17. Power switch: Main switch of the incubator. Also functions as an over-current breaker. The
switch is covered by a switch cover to prevent the accidental push. To turn on or off the switch, remove
the switch cover by loosening the screw.
18. Humidifying pan cover*: Prevents UV light being exposed to the chamber. When filling the pan,
lift the front side and take out the pan. See page 30 for details.
19. UV lamp*: Sanyo UV lamp does not generate ozone. Never look at the UV light directly. For
replacement, contact Sanyo sales representative or agent.
20. Water level sensor for humidifying pan*: Detects the water level in the humidifying pan. See
page 24 for details.
21. Glow starter*: For UV lamp (model; FG-7P)
*MCO-18AC(UV) or when an optional UV system kit MCO-18UVS3 is installed.
9
- 58 -
Page 11
Control panel and keypad
1
2
3
4
5
6
7
8
9
10
11
12
13
14
151617
1. Digital temperature indicator (TEMPERATURE
o
C): Normally, this indicator shows the chamber
temperature. In the setting mode, it shows the set value of the chamber temperature. If the self
diagnostic function detects any abnormality, an error code will be displayed.
2. Heater lamp (HEAT): This lamp lights when the heater is energized.
3. UV indicator (UV): This lamp lights when the UV lamp is ON. The blink of this indicator
recommends the replacement of UV lamp. See page 19 for the details. MCO-18AC(UV) or when an
optional UV system kit MCO-18UVS3 is installed.
4. Door lamp (DOOR): This lamp lights when the outer door is open.
5. Water level alarm lamp (RH PAN): This lamp blinks when the water in the humidifying pan is less
than approximately 0.8 liter. MCO-18AC(UV) or when an optional UV system kit MCO-18UVS3 is
installed.
6. Upper limit regulator: This regulator is used to set the upper temperature limit.
7. Over heat lamp (OVER HEAT): This lamp lights when the chamber temperature reaches the upper
limit set value.
8. CO
inject lamp (INJECT): This lamp lights when CO2 gas is being injected.
2
9. Digital CO
chamber. In the setting mode, it indicates the set value of the CO
mode, the decimal point in the digital CO
density indicator (CO2 %): Normally, this indicator shows the CO2 concentration in the
2
concentration. In the calibration
2
density indicator blinks.
2
10. Set key (SET): Pressing this key to enter the setting mode, and the digits to be set will blink.
11. Calibration key (CAL): Pressing this key to enter the CO
calibration pre-conditioning mode.
2
Normally when the outer door is opened, the fan motor will stop operating. This mode allows the fan
motor to continue to circulate the chamber environment to ensure accurate CO
uniformity inside the
2
chamber. At this time, by pressing this key for approximately 5 seconds, the incubator enters the
temperature calibration input mode. And then pressing this key again, the incubator enters the CO
2
calibration input mode. See page 31 and 32 for the details. Also, used to change the UV lamp ON
period. See page 21 for the details.
Note:
This operation affects the basic performance. When calibrating the CO
density, contact a Sanyo sales
2
representative or agent.
- 59 -
10
Page 12
12. Alarm buzzer stop key (BUZZER): Press this key to silence the buzzer when the alarm operates
and the buzzer sounds (The remote alarm is not stopped).
13. CO
provided that MCO-21GC automatic CO
14. CO
an automatic CO
is empty, the CO
gas supply line indicator (A/B): The lamp for the supply line currently in use lights up
2
cylinder changeover system (option) is installed.
2
gas supply line switching key : This key to select CO2 gas supply line is available only when
2
cylinder changeover system MCO-21GC (option) is installed. When one CO2cylinder
2
is supplied by the other cylinder automatically.
2
15. Enter key (ENT): Pressing this key memorizes the set value in the controller.
16. Numerical value shift key ( ): Pressing this key in the setting mode causes the numerical value
to shift. In key lock mode, pressing this key makes key lock ON or OFF. See page 23 for details.
17. Digit shift key ( ): Pressing this key in the setting mode causes the changeable digit to shift.
Pressing this key more than 5 seconds enters key lock mode. See page 23 for the key lock.
Remote alarm terminal
The terminal of the remote alarm is installed at the rear right side of the incubator. The alarm is
outputted from this terminal. Contact capacity is DC 30 V, 2 A.
Contact output:
between COM. and N.O. between COM. and N.C.
At normal Open Close
At abnormal Close Open
Note:
The alarm is actuated when the power cord is disconnected from the outlet or the power switch is OFF.
• The remote alarm cannot be stopped by pressing the alarm buzzer stop key (BUZZER) since the
remote alarm is not conjunct with the alarm buzzer stop key (BUZZER).
Remote alarm terminal
11
- 60 -
Page 13
INSTALLATION SITE
To operate this incubator properly and to obtain maximum performance, install the incubator in a location
with the following conditions:
WARNING
Ventilate a room air occasionally when using CO2 gas for control. The gas density will increase in
an enclosed small room and high level of gas density can be hazardous to health. In addition, avoid
inhaling the chamber air directly when opening the door if CO
Si l’appareil est utilisé dans un evdroit restreint, le niveau de la densite CO
être nocif aux humains. Evitez d’aspirer l’air provenant de l’inérieur de l’appareil quand vous ouverz la
porte.
■ A location not subjected to direct sunlight
Do not install the incubator under direct sunlight. Installation in a location subjected to direct sunlight
cannot obtain the intended performance.
■ A location away from heat generating sources
Avoid installing the incubator near heat-emitting appliances such as a heater or a boiler etc. Heat can
decrease the intended performance of the incubator.
gas is used.
2
de l’air peut s’élever et peut
2
o
■ 5
C higher than the ambient temperature
The chamber temperature must be at least 5
chamber temperature is set to 37
ambient temperature in adequate range.
■ A location with a sturdy and level floor
Always install the incubator on a sturdy and level floor. The uneven floor or tilted installation may cause
failure or injury. Install the incubator in stable condition to avoid the vibration or noise. Unstable
condition may cause vibration or noise.
o
C, the ambient temperature must be less than 32oC. Keep the
o
C higher than the ambient temperature. For example, the
WARNING
Install the incubator on a sturdy floor. If the floor is not strong enough or the installation site is not
adequate, this may result in injury from the incubator falling or tipping over.
Select a level and sturdy floor for installation. This precaution will prevent the incubator from tipping.
Improper installation may result in water spillage or injury from the incubator tipping over.
■ A location not prone to high humidity
Install the incubator in the ambient of 80% R.H. or less humidity. Installation under high humidity may
cause current leakage or electric shock.
WARNING
Do not use the incubator outdoors. Current leakage or electric shock may result if the incubator is
exposed to rain water.
Never install the incubator in a humid place or a place where it is likely to be splashed by water.
Deterioration of the insulation may result which could cause current leakage or electric shock.
■ A location without flammable or corrosive gas
Never install the incubator in a flammable or corrosive location. This may cause explosion or fire or may
result in the current leakage or electric shock by the corrosion of the electrical components.
- 61 -
12
Page 14
INSTALLATION
1. Remove the packaging materials and tapes
Remove all transportation packaging materials and tapes. Open the doors and ventilate the incubator.
If the outside panels are dirty, clean them with a neutral detergent and wipe it up with a wet cloth.
WARNING
Do not put the packing plastic bag within reach of children as suffocation may result.
2. Adjust the leveling feet
Extend the leveling feet by rotating them counterclockwise to contact them to the floor. Ensure the
incubator is level.
3. Fix the incubator
Two fixtures are attached to the rear of the frame. Fix the frame to the wall with these hooks and rope or
chain.
4. Ground (earth)
WARNING
Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet is
not grounded, it is necessary to install a ground by qualified engineers.
Never ground the incubator through a gas pipe, water main, telephone line or lightning rod. Such
grounding may cause electric shock in the case of an incomplete circuit.
● Stacked module
When the product is used in stacked module, make sure to fix the incubator by using the stacking kit
WARNING
Do not climb onto the incubator and do not put articles on the incubator. This may cause injury by
tipping or damage to the incubator. When stacking the incubator, follow the procedure shown on page
40.
● If not used
When the incubator is not used, dispose the water in the humidifying pan and remove any moisture in the
chamber completely. Check that the chamber is completely dry before closing the doors.
● Before moving the incubator
Empty the humidifying pan completely before moving the incubator. Spilled or splashed water may
cause current leakage or electric shock.
13
- 62 -
Page 15
Connection of CO2 gas cylinder
WARNING
Check the gas type and ensure that it is fit for the purpose. Make sure that all pipes are
connected correctly and are not liable to become disconnected. Ensure that the gas pressure is
set at the specified value. Improper connection of the gas pipe or use of incorrect gas pressure
may result in leakage of CO
lead to asphyxiation and risk of death.
gas. Elevated level of CO2gas can be hazardous to health and may
2
1. Use a liquefied CO
gas cylinder, not a siphon (dip tube) type. The CO2gas should be 99.5% or more
2
pure.
2. Install a gas pressure regulator (optional accessory MCO-100L) on the cylinder. Use a regulator
rated at 25 MPaG (250 kgf/cm
2
G, 3600 psiG) on the primary side and 0.2 MPaG (2.0 kgf/cm2G, 30 psiG)
on the secondary side.
3. Using the gas supply pipe provided, connect the CO
CO
gas pipe located at the rear left side of the CO2 incubator.
2
gas pressure regulator to the connecting port for
2
Note:
If CO
is supplied to multiple CO2 incubators from a single gas cylinder, a CO2 solid will be formed in the
2
gas pressure regulator. The gas pressure regulator safety valve will operate, and there may an explosive
sound.
When the MCO-21GC is not mounted
Using the gas supply pipe that is provided, connect the CO
cylinder to CO
gas pipe connection port A on the incubator. After connecting the pipe, check to make
2
gas pressure regulator for the CO2 gas
2
sure that no gas is leaking.
When the MCO-21GC is mounted
Connect a pair of CO
gas cylinders with CO2 gas pressure regulators independently. The gas supply line
2
will be switched automatically. Connect the main cylinder to port A and the reserve cylinder to port B. After
connecting the cylinders, check to make sure that no gas is leaking.
4. Set the CO
Excessive pressure may cause disconnection of internal pipes inside the CO
in leakage of CO
pressure on the secondary side to 0.03 MPaG (0.3 kgf/cm2G, 4.3 psiG) (at gas injection).
2
incubator which will result
2
gas into the atmosphere. Elevated level of CO2gas can be hazardous to health
2
and may lead to asphyxiation and risk of death. The repair of the incubator will be necessary if the
internal pipe is disconnected.
4. Check that no gas is leaking at any point where the pipe connects with the CO
regulator or the CO
2
incubator.
Note:
The incubator, including the gas supply pipes and services must be examined at frequent intervals to
ensure they are safe. Ensure that items such as pipes are replaced if there is any sign of deterioration.
- 63 -
2
14
Page 16
PREVENTING CONTAMINATION
To prevent contamination of the chamber, select an appropriate location for installation as well as the
complete disinfection of the chamber components.
● Avoid hot and humid location
Avoid location with high temperature and/or humidity as the presence of bacteria in the air is greater than
in normal environment.
● Avoid drafty location and location with many passers-by
Avoid locations near doors, air conditioners, fans, etc., where slight breezes can facilitate the entry of
bacteria into the chamber.
● Installation in a sterile room
To get the cultivation more efficiently, install the incubator in a sterile room.
● Use clean containers
The contamination is mainly caused by the containers such as Petri dishes or bottles stored in the
chamber. Always keep the containers clean.
● Always keep the chamber clean
The condensation may be caused on the inside of the door by spilled water form humidifying pan or
opening of outer door for long period. Wipe off the condensation completely with a sterile dry gauze.
Especially when the culture medium is spilled, clean and disinfect the chamber immediately. Refer to
page 27 “Routine maintenance” for details.
● Keep the inside panels dry
To protect the inside of the incubator from contamination, the inside panels should always be kept dry. If
water is spilled from a humidifying pan or if the door is kept open for a long period, condensation will form
on the panels, allowing germs to breed. In such a case, wipe away the water with a dry sterile gauze.
Particularly, if the medium is spilled, wipe it up immediately and sterilize the area.
● Fill the humidifying pan with sterile distilled water
Always use sterile distilled water to fill the pan. The water level alarm lamp (RH PAN) on the control
panel blinks when the water level is low. Refill the sterile distilled water to the pan when the water level
alarm lamp (RH PAN) blinks. Note that when low temperature sterile distilled water is poured, the
chamber temperature drops significantly. Clean the humidifying pan once a month.
● Do not place the incubator in the direct air flow from an air conditioning system
Air from an air conditioning system may cause condensation and lead to possible contamination.
15
- 64 -
Page 17
PRECAUTIONS FOR CULTURES
● Do not subject to direct air flow
Do not allow the air for air conditioning to hit the incubator or door directly. Direct hit may cause
condensation or contamination.
● Allow adequate space between the cultures
When storing cultures in the chamber, keep the Petri dishes or bottles containing the cultures sufficiently
apart from each other to allow adequate air circulation. Inadequate space may result in uneven
temperature distribution and CO
● Stored materials
Never place acid or alkaline materials or materials that release corrosive gas in the chamber. Such
materials can cause failure resulting from discoloration or corrosion.
● Always shut the inner door
Shut the inner door completely, and then shut the outer door. If the inner door is not closed completely,
even if the outer door is closed, the incubator will fail to exhibit its maximum performance. And close the
doors gently. Rude closing may cause spillage of medium, incomplete closing, or damage of gasket.
concentration in the chamber.
2
● Open/close the doors gently
Ensure you close the doors gently. Robust closing may cause spillage of medium, incomplete closing,
or damage of gasket. Also, before opening the inner door, check that the UV light is OFF
(MCO-18AC(UV) or when an optional UV system kit MCO-18UVS3 is installed).
● Take care of the inside of the outer door.
It may get hot.
● Never lean or press on the glass.
Broken glass due to intentional force may cause injury.
● Alarm
Always investigate the cause and fix the alarm condition immediately when the alarm is activated. Refer
to page 25 for alarm details.
● Fix the tray supports and trays securely
Incomplete installation may cause injury or damage.
- 65 -
16
Page 18
START-UP OF INCUBATOR
When start the test operation or the operation, follow these steps as below.
1. Install the incubator referring to “INSTALLATION” on page 13.
2. Remove all transportation packaging materials and tapes on the attachments in the chamber. Then
clean and sterilize the chamber and inner attachments. Refer to “Cleaning the chamber and inner
attachments” on page 27.
3. Add approximately 1.5 L of sterile distilled water to the humidifying pan. (Refer to page 30.)
4. Turn ON the power switch on the rear left side.
Note:
The humidity in the incubator chamber is adjusted to the optimum setting. To prevent condensation on
the surface inside chamber and the inner door, there is a low-temperature area under the humidifying pan
in the bottom of chamber to recondense evaporated moisture. Condensation may occur around the
humidifying pan at the bottom of the chamber, but this does not indicate a problem.
17
- 66 -
Page 19
SETTING THE CHAMBER TEMPERATURE AND CO
DENSITY
2
Table below shows the basic procedure for setting the chamber temperature and CO
density. The
2
upper limit alarm temperature setting is also shown in the table. Perform key operations in the sequence
indicated in the table. The example in the table is based on the assumption that the desired temperature
o
is 37
C and CO2 density is 5%. Adjustment of the upper limit regulator should be executed after the
chamber temperature reaches the stable condition.
Note:
• The incubator is set at the factory so that the chamber temperature is 37
o
C and CO2 density is 0%.
• When starting the incubator for the first time or after not using it for an extended period of time, allow at
least 8 hours for the chamber temperature, humidity, and CO
desired chamber temperature at 0% CO
density. Then change the setting to the desired CO2 density.
2
sensor to become stable after setting the
2
Basic operation sequence (Example: Chamber temperature: 33oC, CO2 density: 5%)
Description of operation Key operatedIndication after operation
1 Turn the power switch ON. ----
The current chamber temperature is displayed
in the digital temperature indicator.
2 Press set key. SET The left digit blinks.
By pressing numerical value shift
3
key and digit shift key,
set the figure to 37.0.
When pressed,
the changeable digit is shifted.
When pressed,
the figure of settable digit changes.
Set temperature is memorized.
4 Press enter key. ENT
By pressing digit shift key and
5
numerical value shift key, set the
When pressed, the changeable digit is shifted.
figure to 05.0.
Left digit in the digital
CO
density indicator blinks.
2
When pressed,
the figure of settable digit changes.
6 Press enter key. ENT Set CO
(Executed after the chamber
temperature reaches the stable
condition)
7
Adjust upper limit regulator so that
the alarm temp. is 1
o
C higher than
----
set temperature.
8 Press enter key. ENT
In the digital CO
In the digital temperature indicator, upper limit
temp. is displayed.
The upper limit temp. can be changed
by turning upper limit regulator.
This is the end of set mode and the indicators
display current temperature and CO
density is memorized.
2
density indicator, HI is displayed.
2
density.
2
Note:
● In each set mode, if the change of the setting is not necessary, pressing set key (SET) skips to next set
mode.
● When the CO
density is set to 00.0, the control is OFF regardless of chamber density.
2
● The upper limit temperature set value will change when the regulator is turned even if the incubator is
not in set mode, because the alarm circuit is an independent circuit.
● In each set mode, the indicator returns to the current temperature and CO
density display mode
2
automatically when 90 seconds has passed without any key operation.
- 67 -
18
Page 20
UV LAMP
The clauses below are applicable when the model is MCO-18AC(UV) or an optional UV system kit
MCO-18UVS3 is installed.
A UV lamp is located inside the duct to sterilize the water in the humidifying pan and air circulating in the
chamber.
Following shows precautions and instructions about the UV lamp.
z When all chamber attachments are installed correctly, only the inside of the duct and the inside of the
humidifying pan cover are exposed to UV light.
z Correctly install all of the chamber attachments when starting a cell culture, and never turn ON the UV
lamp when the humidifying pan cover is removed.
z Always install the humidifying pan cover even when using the incubator without turning ON the UV
lamp. Leaving the cover uninstalled will affect the chamber temperature distribution and the humidity
recovery performance.
(MCO-18AC(UV) or MCO-18AC with MCO-18UVS3)
z The UV lamp stays lit for a preset period of time after the outer door is closed. The default setting is for
5 minutes. The period can be changed when necessary as shown in the page 21.
z If the outer door is not opened for at least 12 consecutive hours, the UV lamp will light for the preset
time period every 12 hours. The period can be changed when necessary as shown in the page 21.
● When checking the UV lamp operation, open the outer door and push the door switch with the inner
door closed. The visible blue light can be checked under the humidifying pan cover. The UV light is
harmful to eyes. Never turn on the UV light with the inner door or humidifying pan cover opened.
z The recommended replacement time for the UV lamp (i.e., when the UV output ratio drops to approx.
70% of its initial value) is when the accumulated ON time reaches 1,000 hours. When the accumulated
ON time reaches approximately 1,000 hours, the UV indicator blinks (when the UV lamp is not lit). It is
recommended that the UV lamp be quickly replaced at this point. Contact a Sanyo sales representative or
agent for information on replacing the UV lamp.
z If the UV lamp burns out, E18 is displayed on the digital temperature indicator. If this occurs, replace
the UV lamp. When replacing the UV lamp, replace the glow starter (type FG-7P) at the same time.
Contact a Sanyo sales representative or agent for information on replacing the UV lamp.
19
- 68 -
Page 21
Precautions when using the UV lamp
z
z
The clauses below are applicable when the model is MCO-18AC(UV) or an optional UV system kit
MCO-18UVS3 is installed.
WARNING
Do not look directly at UV light. UV light is harmful to the eyes.
Always use the humidifying pan and humidifying pan cover.
The humidifying pan and cover prevent UV light from escaping. Always use them even when not
humidifying. To check whether the UV lamp is lit, open the outer door and then press the door switch with
the inner door still closed. Visible blue light can be confirmed from the front of the humidifying pan cover.
UV light is harmful to the eyes, so do not light the UV lamp when the inner door or humidifying pan cover
is open.
Be careful when handling the UV lamp.
There is a UV lamp inside of the duct. Be careful not to damage it when installing or removing chamber
attachments or the humidifying pan.
- 69 -
20
Page 22
Setting the UV lamp ON period
The clauses below are applicable when the model is MCO-18AC(UV) or an optional UV system kit
MCO-18UVS3 is installed.
Follow the procedure below when changing the setting of the UV lamp ON period.
Basic operation sequence (Example: change of UV lamp ON period from 5 minutes to 3 minutes)
Description of operation Key operatedIndication after operation
Press calibration key
1
for 5 seconds.
By pressing numerical value shift
2
key and digit shift key,
set the figure to F01.
3 Press enter key. ENT
By pressing numerical value shift
4
key and digit shift key,
set the figure to 003.
5 Press enter key. ENT
x The UV timer can be set from 0 to 30 min. The default setting is for 5 min. When the UV timer is set to
zero, the UV lamp will not light.
x If the outer door is opened while the UV lamp is lit, the lamp will turn OFF. Then, when the door is
closed, the lamp will light for the preset period of time.
x If the UV lamp ON time is set for longer than 5 minutes or if only the outer door is repeated opened and
closed, condensation may occur in the chamber and may affect temperature distribution. It will also
shorten the service life of the UV lamp.
x When replacing the UV lamp, contact a Sanyo sales representative or agent.
CAL
The left digit in the digital
temperature indicator is blinked.
When pressed,
the figure of settable digit changes.
When pressed, the changeable digit
is shifted.
The current setting is displayed
in the digital CO
When pressed, the figure of
settable digit changes.
Set value is memorized and the display return to
normal display mode.
density indicator.
2
Pressing calibration key (CAL) for about 5 seconds leads the calibration mode. In the calibration mode,
the calibration of temperature and CO
performance. Never touch any other keys on the control panel in the event of pressing calibration key
(CAL) accidentally. After about 90 seconds, the incubator returns to chamber temperature display mode
automatically.
21
density is possible. Wrong key operation affects the basic
2
- 70 -
Page 23
Lighting the UV lamp for 24 hours
The clauses below are applicable when the model is MCO-18AC(UV) or an optional UV system kit
MCO-18UVS3 is installed.
If the chamber has been contaminated by dirt or by spilling the medium, use the following procedure to
decontaminate the chamber by lighting the UV lamp for 24 hours.
1. Remove all attachments from the chamber, including the trays, tray supports, side supports, fan cover,
duct, fan, humidifying pan, and humidifying pan cover. Clean all the attachments in an autoclave or with
disinfectant alcohol.
2. Clean and wipe off inside the chamber with disinfectant alcohol.
3. Set the CO
density to 0% and set the UV 24-hour lighting mode.
2
Description of operation Key operatedIndication after operation
1 Press calibration key for 5 seconds. CAL
By pressing numerical value shift
2
key and digit shift key,
When pressed,
set the figure to F02.
3 Press enter key. ENT
4 By pressing numerical value shift
When pressed,
key, set the figure to 001.
5 Press enter key. ENT
The left digit in the temperature
indicator blinks.
When pressed, the figure of settable digit changes.
the changeable digit is shifted.
“000” is displayed
in the digital CO
density indicator.
2
the figure of settable digit changes.
Set value is memorized and the display returns to
normal display mode.
・This procedure must be preformed only with the outer door closed and the UV lamp turned OFF.
・The UV lamp will light continuously for 24 hours after this parameter is set. The setting is canceled if the
outer door is opened. Perform the above procedure to set the 24-hours lighting mode again when
opening the outer door.
・The incubator keeps running with set temperature and CO
density during UV lighting.
2
・Install the attachments after completion of the 24-hour lighting mode.
o
Note: Set the upper limit alarm temperature to at least 10
C higher than the chamber set temperature
when using the 24-hour lighting mode of UV lamp. The 24-hour lighting mode may cause an automatic
chamber temperature alarm because the temperature of the chamber will increase.
Pressing calibration key (CAL) for about 5 seconds leads the calibration mode. In the calibration mode,
the calibration of temperature and CO
density is possible. Wrong key operation affects the basic
2
performance. Never touch any other keys on the control panel in the event of pressing calibration key
(CAL) accidentally. After about 90 seconds, the incubator returns to chamber temperature display mode
automatically.
- 71 -
22
Page 24
KEY LOCK FUNCTION
This incubator is provided with a key lock function. When the key lock is ON, change of temperature or
CO
density setting through the key pad is not available.
2
Note: The key lock is set in OFF mode (L0) at the factory.
Display Mode Function
Key lock is OFF Enable to change of temperature and CO
Key lock is ON Disable to change of temperature or CO
setting
2
setting
2
Procedure for key lock setting (change from key lock OFF to key lock ON)
Description of operation Key operatedIndication after operation
1 ----
2 Press digit shift key for 5 seconds.
Press numerical value shift key
3
and scroll the figure to 1.
L0 is displayed in the temperature
When pressed, the figure of settable
4 Press enter key. ENT
The current chamber temperature and
CO
density are displayed.
2
indicator.
digit changes.
The key lock is set to ON.
The current chamber temperature is displayed.
Note:
●The key lock function is available for temperature and CO
density setting.
2
● To cancel the key lock, set to L0 in the above procedure.
23
- 72 -
Page 25
WATER LEVEL SENSOR
Water level sensor for the humidifying pan is used with the MCO-18AC(UV) or the MCO-18AC with an
optional UV system kit MCO-18UVS3 installed. The sensor is set in active position with respect to the
installation of the humidifying pan. Take care not to damage the sensor at the time of removal or
installation of the humidifying pan.
When the humidifying pan is removed (side view)
Sensor
Humidifying pan
(MCO-18AC(UV) or MCO-18AC with MCO-18UVS3)
When the humidifying pan is installed (side view)
Sensor
Humidifying pan
Note: At the installation, check the humidifying pan is located far depth and the sensor is down.
Note:
• Lift the sensor before installing the humidifying pan if the sensor is in lower position after maintenance.
• At the time of installation of humidifying pan, check that the pan is set properly and sensor is down in
the pan. The water level alarm lamp (RH PAN) blinks if the sensor is not down completely. In this case,
set the pan again in adequate location.
• The sensor detects the water level every 30 minutes. In addition, detection is executed after operation of
the outer door. It takes several seconds to detect the water level. Therefore, the water level alarm
lamp may blink a few times after the outer door is closed with the humidifying pan filled sufficiently.
CAUTION
The foreign particles on the water surface adhere to the water level sensor and its fittings by the capillary
action because the sensor is always in the water. The adhered foreign particles degrade the sensor
performance and the water level alarm lamp may blink with the enough water in the humidifying pan.
At the cleaning, take care not transmit the excessive force to the sensor lead wire.
- 73 -
24
Page 26
ALARMS, SAFETY, AND SELF-DIAGNOSIS
y
y
y
g
y
y
y
y
y
y
y
y
y
y
y
y
r
y
p
The incubator supports the following alarms, safety functions, and self-diagnostic functions.
Alarms, safety and self-diagnosis for culture operations
Alarm or safety
function
Upper limit
temperature alarm
Automatic set
temperature alarm
Automatic set CO
density alarm
Door check The outer door is open. Door lamp lights -----
Auto-return
lock When the key lock is “ON”. ---------- The setting is disabled.
Ke
gas cylinder
CO
2
empty
Gas line changeover
Chamber
temperature sensor
abnormality
Ambient temperature
sensor abnormality
CO
sensor
2
abnormality
Main heater
abnormality
Bottom heater
abnormality
Door heater
abnormality
Disconnection of
sensor for each
heater
Low humidifying
Wate
*
UV lamp failure* When the UV lamp is burned out.
Recommendation of
new UV lam
*
If the chamber temperature exceeds
the upper limit alarm temperature set
value.
If the chamber temperature deviates
from the set temperature by ±1
more.
If the chamber CO
2
from the set value b
When there is no key pressing in
each settin
If the CO2 density does not increase
when the gas valve is opened.
When the gas supply line is switched.
(only when MCO-21GC is installed)
If the temperature sensor is
disconnected.
If the temperature sensor is short
circuited.
If the ambient temperature sensor is
disconnected.
If the ambient temperature sensor is
short circuited.
If the output voltage of the CO2 sensor
is abnormal.
If the upper limit alarm temperature
alarm operates, or if the main heater is
open circuit, or the main heater relay
short circuit.
If the bottom heater goes open circuit,
or the bottom heater relay short circuit.
If the door heater goes open circuit,
or the door heater relay short circuit.
If the relay of main heater, bottom
heater or sensor box heater goes open
circuit.
If the water in the humidifying pan is
about 0.8 liter.
The accumulated ON time is over about
1,000 hrs.
*MCO-18AC(UV) or when an optional UV system kit MCO-18UVS3 is installed.
Conditions Indication Buzzer Safety operation
Over heat lamp lights.
density deviates
2
±1% or more.
mode for 90 seconds.
o
C or
E12 or E16 and chamber
temperature are displayed
alternatel
All digits on the
temperature indicator blink.
All digits on the digital CO
density indicator blink.
Normal display mode. -----
E01 is displayed alternatel
with the temperature on the
digital temperature
indicator.
E01 is displayed alternatel
with the temperature on the
digital temperature
indicator.
CO2 gas supply line
indicatorof empty blinks.
E05 is displayed alternatel
with the temperature on the
digital temperature
indicator.
E06 is displayed alternatel
with the temperature on the
digital temperature
indicator.
E09 is displayed alternatel
with the temperature on the
digital temperature
indicator.
E10 is displayed alternatel
with the temperature on the
digital temperature
indicator.
E11 is displayed alternatel
with the temperature on the
digital temperature
indicator.
E12 is displayed alternatel
with the temperature on the
digital temperature
indicator.
E13 is displayed alternatel
with the temperature on the
digital temperature
indicator.
E14 is displayed alternatel
with the temperature on the
digital temperature
indicator.
E16 is displayed alternatel
with the temperature on the
digital temperature
indicator.
Water level alarm lamp(RH
PAN)blinks.
E18 is displayed alternatel
with the temperature on the
digital temperature
indicator.
UV indicator blinks
when UV lamp is OFF.
.
Continuous tone
Intermittent tone with
15 minutes delay.
Intermittent tone with
2
15 minutes delay.
Intermittent tone Remote alarm
Intermittent tone
Intermittent tone
Intermittent tone Remote alarm
Intermittent tone
Intermittent tone Remote alarm
Intermittent tone Remote alarm
Intermittent tone Remote alarm
Intermittent tone Remote alarm
----- -----
Intermittent tone Remote alarm
----- -----
Heater OFF
Remote alarm
Remote alarm with 15
minutes delay
Remote alarm with 15
minutes dela
Fan motor OFF
CO
valve is closed
2
UV lamp OFF*
(60 seconds after)
Heater OFF
The setting mode is
canceled.
Gas supply line is
altered.
Remote alarm
Heater OFF
Remote alarm
CO
valve is closed
2
Remote alarm
• The alarm can be canceled by pressing the alarm buzzer stop key (BUZZER), but the remote alarm cannot be
stopped. And the upper limit temperature alarm cannot be silenced with the alarm buzzer stop key (BUZZER).
• E01 is cleared automatically when the gas is connected correctly and the buzzer is silenced with the alarm buzzer
stop key (BUZZER). When MCO-21GC is installed, press the alarm buzzer stop key (BUZZER) to silence the alarm
after changeover of gas supply line.
• If one of E05 to E18 (except for E12, E13, and E14) is displayed, contact a Sanyo sales representative or agent.
25
- 74 -
Page 27
SETTING OF ALARM RESUME TIME
The alarm buzzer is silenced by pressing alarm buzzer stop key (BUZZER) on the control panel during
alarm condition (The remote alarm is not stopped).
The buzzer will be activated again after certain suspension if the alarm condition is continued (The
remote alarm continues). The suspension time can be set by following the procedure shown in the table
below.
The example in the table is based on the assumption that the desired duration is 20 minutes.
Note: The duration is set in 30 minutes at the factory.
Table Changing procedure for alarm resume time (Ex: change from 30 minutes to 20 minutes)
Description of operation Key operatedIndication after operation
1 ----
Press calibration key
2
for 5 seconds.
Set the figure to F25
with the numerical value shift key
3
and digit shift key.
CAL The left digit blinks.
The current chamber temperature is
displayed.
When pressed,
the figure of settable digit changes.
The settable digit is shifted.
4 Press enter key. ENT
Set the figure to 020 with the
5
numerical value shift key.
6 Press enter key. ENT
• The settable alarm resume time are 0, 10, 20, 30, 40, 50, or 60 minutes (The setting is 000, 010, 020,
030, 040, 050 or 060 respectively). The buzzer would not reset if the reset time is set in 000.
• It is recommended to set the alarm resume time when the incubator is not under alarm condition. The
setting during alarm condition is effective on the next alarm condition.
• The setting cannot be changed during power failure.
• The remote alarm during power failure cannot be stopped.
• The set mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing enter key is not memorized.
Pressing calibration key (CAL) for about 5 seconds leads the calibration mode. In the calibration mode,
the calibration of temperature and CO
performance. Never touch any other keys on the control panel in the event of pressing calibration key
(CAL) accidentally. After about 90 seconds, the incubator returns to chamber temperature display mode
automatically.
density is possible. Wrong key operation affects the basic
2
The current setting is displayed.
The middle digit blinks.
When pressed,
the figure of settable digit changes.
The setting is memorized and
the current chamber temperature is
displayed.
Operation after power failure
The set value is memorized by nonvolatile memory. Accordingly, the incubator resumes the operation
with setting before power failure.
- 75 -
26
Page 28
ROUTINE MAINTENANCE
Cleaning the chamber and inner attachments
When the chamber of the incubator is contaminated, the chamber and internal attachments should be
cleaned and sterilized as follows.
WARNING
Before performing any repairs or maintenance, turn OFF the power switch and unplug the incubator.
Failure to do say may result in electric shock or injury.
Be careful not to inhale chemicals, vapors or aerosols when cleaning the incubator. Doing so may
be harmful to health.
CAUTION
Wear rubber gloves when performing maintenance on the chamber. Failure to wear gloves may
result in cuts or abrasions from sharp edges or corners.
Note:
Be careful not to damage the water level sensor or the UV lamp in the chamber duct (if the
MCO-18AC(UV) or an optional UV system kit MCO-18UVS3 is installed).
Do not use detergents or antiseptic solutions with acid, alkali, or chlorine. Doing so may cause
discoloration, corrosion, or rusting.
1. Close the valve of the CO
2. Open the outer door. Remove 2 pins fixing the inner door hinge and remove the inner door.
3. Pull out the all trays. (See Fig. 1.)
4. Lift the front of the tray supports and pull them out. (See Fig. 2.)
5. (MCO-18AC(UV) or when an optional UV system kit MCO-18UVS3 is installed)
Lift the humidifying pan cover off from the pins on the rear side. (See Fig. 3.)
gas cylinder and turn off the power to the incubator.
2
Tray support
Humidifying pan cover
27
Fig. 1 Fig.2Fig.3
- 76 -
Page 29
6. Pull out the humidifying pan. (See Fig. 4.)
7. Loosen the two screws securing the fan cover and take off the fan cover. (See Fig. 5.)
8. Lift the duct and remove it from the pins on the rear side. (See Fig. 6.)
Fan cover
Humidifying pan
Fig.4 Fig.5 Fig.6
9. Remove the fan by pulling out the central spring and then by pulling out the fan. (See Fig. 7.)
10. Remove the screw securing the clamp for the side support and remove the clamp. (See Fig. 8.)
11. Lift the side support off of the pins. (See Fig. 9.)
Fan
Clamp
Side
support
Duct
Fig.7 Fig.8 Fig.9
12. Clean all the attachments with a diluted neutral detergent, and then rinse them thoroughly with
distilled water.
13. Wipe the trays, the inner attachments such as the fan, and the chamber sides with sterilizing alcohol.
Be careful not to leave any residual alcohol.
14. Wipe the inside wall of the chamber with a gauze containing alcohol for sterilization and then wipe off
with a dry gauze.
- 77 -
28
Page 30
15. (MCO-18AC(UV) or when an optional UV system kit MCO-18UVS3 is installed)
Wipe the water level sensor with sterilizing alcohol. Be careful not to leave any residual alcohol. When
cleaning the sensor, take care not to apply excessive force to the lead wires.
16. To reinstall all the attachments, perform the procedure in reverse order from step 11.
17. Fill the humidifying pan with sterile distilled water.
Note:
• When installing the fan, insert it securely on the motor shaft (See Fig.10). Lightly turn the fan manually
to make sure that it does not strike the rear panel. Improper insertion may cause poor performance.
• As shown in Fig. 11, set the tray with only the front edge bent down. If the tray is set in the wrong
direction, it may not be level and may become unstable.
Fan
Spring
①②
Fig. 10
Bent down
Fig. 11
29
- 78 -
Page 31
Filling the humidifying pan
Use the following procedure to fill the humidifying pan or to replace the water.
1. (MCO-18AC(UV) or when an optional UV system
kit MCO-18UVS3 is installed)Lift the humidifying pan
cover. After that pull the humidifying pan forward.
(See Fig. 1.)
2. Dispose of the remaining water in the humidifying
pan and clean the humidifying pan with a neutral
dishwashing detergent. Then rinse it thoroughly with
distilled water and wipe it with sterilizing alcohol.
3. Wipe all moisture from the bottom of the chamber.
4. Return the humidifying pan to the chamber and
add sterile distilled water (approx. 1.5 L, preheated
to 37qC). (See Fig. 2.)
Note:
Preheat to 37qC the water to be added to the
humidifying pan. Adding low-temperature water will
lower the temperature and humidity in the chamber.
Fig. 1
Fig. 2
5. Set the humidifying pan with the inner side flush against the back, and close the inner and outer doors.
(MCO-18AC(UV) or when an optional UV system kit MCO-18UVS3 is installed)
Set the humidifying pan with the inner side flush against the back, and replace the humidifying cover
(Refer to page 24). Close the inner and outer doors, and confirm that water level alarm lamp (RH PAN)
on the control panel is off.
Note:
(MCO-18AC(UV) or when an optional UV system kit MCO-18UVS3 is installed)
If the water level alarm lamp(RH PAN) blinks, use the above procedure to replenish the water.
CAUTION
(MCO-18AC(UV) or when an optional UV system kit MCO-18UVS3 is installed)
When refilling the water in the humidifying pan, always wipe off any dirt from the water level sensor
with sterilizing alcohol. While doing that, be careful not to apply excessive force to the sensor lead
wire.
- 79 -
30
Page 32
CALIBRATION
Temperature calibration
1. Press the calibration key (CAL) for approximately 5 seconds again to enter the temperature calibration
input mode.
2. The left digit of the digital temperature indicator blinks, and the digital CO
3. Set the present correct temperature with the digit shift key ( ) and numerical value shift key ( ) ,
then press the enter key (ENT).
4. The incubator automatically reverts to the display mode.
[Example]
If the displayed chamber temperature is 37.0
1. Press the calibration key (CAL) for about 5 seconds.
2. The “3” on the digital temperature indicator blinks, and the digital CO
3. Adjust the set value to the actual value of 36.8
key ( ), then press enter key (ENT).
4. The incubator automatically reverts to the display mode.
density indicator goes out.
2
o
C (set value) and the actual temperature is 36.8oC.
density indicator goes out.
o
C with the digit shift key ( ) and numerical value shift
2
31
- 80 -
Page 33
CALIBRATION
CO2 calibration
Span setting
Span setting should be done under stable condition of temperature, humidity, and CO
1. Press the calibration key (CAL) once to enter the CO
2. Confirm blinking the decimal point in the digital CO
3. Open the outer door and remove the sample air outlet cap on the inner door.
4. Measure the CO
5. Press the calibration key (CAL) for about 5 seconds again to enter the temperature calibration input
mode.
6. The left digit on the digital temperature indicator blinks, and the digital CO
7. Press the calibration key (CAL) once again to enter the CO
8. The left digit on the digital CO
9. Set the present correct CO
then press the enter key (ENT).
10. Replace the sample air outlet cap.
11. The incubator automatically reverts to the display mode.
Note:
This calibration is available when the setting of CO
density in the chamber through the sample air outlet on the inner door.
2
density.
2
calibration pre-conditioning mode.
2
density indicator.
2
density indicator goes out.
2
calibration input mode.
2
density indicator blinks, and the temperature indicator goes out.
2
density with the digit shift key ( ) and numerical value shift key ( ),
2
density is 2% or more.
2
[Example]
For an internal CO
density of 5.0% (setting) and a measured value of 4.5%.
2
1. Press the calibration key (CAL) once.
2. Press the calibration key (CAL) for about 5 seconds again.
3. The “3” on the digital temperature indicator blinks, and the digital CO
density indicator goes out.
2
4. Press the calibration key (CAL) once again.
5. The “0” on the digital CO
left digit on the digital CO
6. Set the present correct CO
density indicator blinks, and the digital temperature indicator goes out. The
2
density indicator blinks, and the temperature indicator goes out.
2
density (4.5%) with the digit shift key ( ) and numerical value shift key
2
( ), then press the enter key (ENT).
7. The incubator automatically reverts to the display mode.
Note:
It is important to accurately measure the temperature inside the incubator when calibrating the
temperature. Particularly, the temperature gauge used must have an accuracy of JIS 0.5 Class or better.
The temperature must be measured at several points.
The temperature setting must not change by more than ±1.0
o
C during calibration. If these values are
exceeded, an error buzzer will sound, the input data will be ignored, and the incubator will return to the
o
display mode. If it is necessary to change the temperature by more than 1.0
C, perform calibration in
several stages over a period of time.
- 81 -
32
Page 34
TROUBLESHOOTING
If the incubator does not seem to be working properly, check the following items before calling for service.
Malfunction Check/Remedy
The incubator does not
operate at all.
The key operation is disable • The key lock function is set in ON mode.
If the alarm function and the
buzzer operates
If the chamber temperature
is not equal to the set
temperature
If the CO
gas density does
2
not coincide with the set
value
If the chamber humidity does
not rise
If the CO
consumption is
2
too much
If normal cultivation cannot
be done and chamber gas
density is suspect
If it takes much time to
recover the CO
CO
gas is not being
2
gas density
2
injected.
• The incubator is not plugged correctly into a power outlet.
• The circuit breaker at the power source is active or a power
failure has occurred.
[At the beginning of operation]
• The chamber temperature is not equal to the set value.
• The chamber CO
a. The secondary pressure of the pressure regulator is not
equal to the set value (0.03MPaG, 0.3kgf/cm
b. The tube is not connected securely between the pressure
regulator and the incubator.
[During operation]
• The upper limit alarm temperature is not set at least 1
than the set chamber temperature.
• The set temperature value was changed, or the door was left
open for a long period. Or a low temperature load was placed
inside the incubator. In this case, if the incubator is left as it is,
the alarm will eventually clear itself.
• The gas tube has slipped off or the gas leaks.
• The set value of the gas density was changed.
• The gas cylinder is empty. Check the primary pressure of the
CO
cylinder once a week. (The primary pressure of less than
2
3.8MPaG (38kgf/cm
Replace the cylinder soon.)
• The ambient temperature must always be at least 5
the set temperature.
• The outer door was closed while the inner door was left open.
• The secondary pressure is not set to 0.03MPaG (0.3kgf/cm
4.3psiG).
• The gas tube is clogged or chinked.
• The humidifying pan is not filled with sterile distilled water.
(Always use sterile distilled water.)
• The humidifying pan is not filled with sterile distilled water.
(Always use sterile distilled water.)
• The door is opened frequently.
• There is any gas leakage at the connection or pin hole on the
tube. It is recommended to replace the tube once a year.
• The gasket of the inner door is not completely sealed.
• The access port at the upper left corner is opened.
• The environment around the incubator is not normal. The
source of the contaminated gas is nearby.
• The incubator is installed in an enclosed space.
• HEPA filter is provided in the gas piping. If it takes much time
to recover the gas density even though the gas pressure is
normal, it may be that dust on the HEPA filter prevents the gas
flow. Contact a Sanyo sales representative or agent.
• The CO
2
method. CO
concentration in the chamber approaches the set value.
Injections may be stopped for periods of approximately 15
seconds, but that is not an error.
density is not equal to the set value.
2
2
G, 4.3psiG).
o
C higher
2
G) means a little gas in the cylinder.
o
C less than
2
G,
control method for the incubator is the ON-OFF
gas is intermittently injected as the gas
2
33
- 82 -
Page 35
Note: If the problem still has not been solved after trying the above checks and countermeasures, or for
any problems not covered here, contact a Sanyo sales representative or agent.
DISPOSING OF THE CO2 INCUBATOR
When disposing of the CO2 incubator, contact a Sanyo sales representative or agent.
WARNING
The incubator must be dismantled and disposed of by qualified personnel only. If the incubator is
left where outsiders enter, it may result unexpected accident (for example, children to become locked
inside).
Before disposing a CO
extent possible by the user.
incubator with biohazardous danger, decontaminate the CO2 incubator to the
2
- 83 -
34
Page 36
Note:
This symbol mark and recycle system are applied only to EU countries
and not applied to the countries in the other area of the world.
Waste Electrical and Electronic Equipment (WEEE) Directive-2002/96/EC
(English)
Your SANYO product is designed and manufactured with high quality materials and components which
can be recycled and reused.
This symbol means that electrical and electronic equipment, at their end-of-life, should be disposed of
separately from your household waste.
Please dispose of this equipment at your local community waste collection/recycling centre.
In the European Union there are separate collection systems for used electrical and electronic products.
Please help us to conserve the environment we live in!
(German)
Ihr SANYO Produkt wurde entworfen und hergestellt mit qualitativ hochwertigen Materialien und
Komponenten, die recycelt und wiederverwendet werden können.
Dieses Symbol bedeutet, daß elektrische und elektronische Geräte am Ende ihrer Nutzungsdauer von
Hausmüll getrennt entsorgt werden sollen.
Bitte entsorgen Sie dieses Gerät bei Ihrer örtlichen kommunalen Sammelstelle oder im Recycling Centre.
In der Europäischen Union gibt es unterschiedliche Sammelsysteme für Elektrik- und Elektronikgeräte.
Helfen Sie uns bitte, die Umwelt zu erhalten, in der wir leben!
35
- 84 -
Page 37
(French)
Votre produit Sanyo est conçu et fabriqué avec des matèriels et des composants de qualité supérieure qui
peuvent être recyclés et réutilisés.
Ce symbole signifie que les équipements électriques et électroniques en fin de vie doivent être éliminés
séparément des ordures ménagères.
Nous vous prions donc de confier cet équipement à votre centre local de collecte/recyclage.
Dans l’Union Européenne, il existe des systèmes sélectifs de collecte pour les produits électriques et
électroniques usagés.
Aidez-nous à conserver l’environnement dans lequel nous vivons !
Les machines ou appareils électriques et électroniques contiennent fréquemment des matières qui, si
elles sont traitées ou éliminées de manière inappropriée, peuvent s’avérer potentiellement dangereuses
pour la santé humaine et pour l’environnement.
Cependant, ces matières sont nécessaires au bon fonctionnement de votre appareil ou de votre machine.
Pour cette raison, il vous est demandé de ne pas vous débarrasser de votre appareil ou machine usagé
avec vos ordures ménagères.
(Spanish)
Los productos SANYO están diseñados y fabricados con materiales y componentes de alta calidad, que
pueden ser reciclados y reutilizados.
Este símbolo significa que el equipo eléctrico y electrónico, al final de su ciclo de vida, no se debe
desechar con el resto de residuos domésticos.
Por favor, deposite su viejo “televisor” en el punto de recogida de residuos o contacte con su
administración local.
En la Unión Europea existen sistemas de recogida específicos para residuos de aparatos eléctricos y
electrónicos.
Por favor, ayúdenos a conservar el medio ambiente!
- 85 -
36
Page 38
(Portuguese)
O seu produto SANYO foi concebido e produzido com materiais e componentes de alta qualidade que
podem ser reciclados e reutilizados.
Este símbolo significa que o equipamento eléctrico e electrónico no final da sua vida útil deverá ser
descartado separadamente do seu lixo doméstico.
Por favor, entregue este equipamento no seu ponto local de recolha/reciclagem.
Na União Europeia existem sistemas de recolha separados para produtos eléctricos e electrónicos
usados.
Por favor, ajude-nos a conservar o ambiente em que vivemos!
(Italian)
Il vostro prodotto SANYO è stato costruito da materiali e componenti di alta qualità, che sono riutilizzabili o
riciclabili.
Prodotti elettrici ed elettronici portando questo simbolo alla fine dell’uso devono essere smaltiti
separatamente dai rifiuti casalinghi.
Vi preghiamo di smaltire questo apparecchio al deposito comunale.
Nell’Unione Europea esistono sistemi di raccolta differenziata per prodotti elettrici ed elettronici.
Aiutateci a conservare l’ambiente in cui viviamo!
37
- 86 -
Page 39
(Dutch)
Sanyo producten zijn ontwikkeld en gefabriceerd uit eerste kwaliteit materialen, de onderdelen kunnen
worden gerecycled en weer worden gebruikt.
Het symbool betekent dat de elektrische en elektronische onderdelen wanneer deze vernietigd gaan
worden , dit separaat gebeurt van het normale huisafval.
Zorg ervoor dat het verwijderen van de apparatuur bij de lokaal erkende instanties gaat gebeuren.
In de Europese Unie wordt de gebruikte elektrische en elektronische apparatuur bij de daarvoor wettelijke
instanties aangeboden.
Alstublieft help allen mee om het milieu te beschermen.
(Swedish)
Din SANYO produkt är designad och tillverkad av material och komponenter med hög kvalitet som kan
återvinnas och återanvändas.
Denna symbol betyder att elektriska och elektroniska produkter, efter slutanvändande, skall sorteras och
lämnas separat från Ditt hushållsavfall.
Vänligen, lämna denna produkt hos Din lokala mottagningstation för avfall/återvinningsstation.
Inom den Europeiska Unionen finns det separata återvinningssystem för begagnade elektriska och
elektroniska produkter.
Vänligen, hjälp oss att bevara miljön vi lever i!
- 87 -
38
Page 40
AUTOMATIC CO2 CYLINDER CHANGEOVER
An automatic CO
system switches the gas supply line when one CO
cylinder changeover system (MCO-21GC) is available as an optional accessory. This
2
gas cylinder is empty.
2
Note: The installation of the MCO-21GC must be performed only by qualified service personnel.
Install the MCO-21GC and then perform the following procedure
1. Connect the CO
gas pipes to connecting ports for CO2 gas pipes A and B. Connecting ports for CO
2
gas pipes are located on the left back of the incubator. (Refer to page 8).
Note: Only port A can be used when the MCO-21GC is not installed. There will be 2 ports, ports A and
B, after the MCO-21GC is installed.
2. Connect a CO
for information on connecting the CO
to the CO
CO
2
gas pipe connected to port B.)
2
gas cylinder equipped with a gas pressure regulator to each gas pipe. Refer to page 14
2
gas cylinders. (“Gas cylinder A” is the CO2 gas cylinder connected
2
gas pipe connected to port A, and “gas cylinder B” is the CO2 gas cylinder connected to the
3. Open the valve of both gas cylinders.
4. Check that the CO
gas supply line indicator on the control panel is lit by pressing the CO2 gas supply
2
line switching key.
2
5. Select a CO
6. The MCO-21GC detects that there is no more CO
chamber does not increase for a while even after opening the CO
gas supply line (A or B).
2
gas in a cylinder when the CO2density in the
2
gas valve in the incubator. It switches
2
the gas supply line when it detects that there is no more gas. At that time, the following indications of
switching the gas cylinder are provided.
gas supply line indicator
CO
Digital temperature indicator
Buzzer
2
Empty line The other line
E01 is displayed alternately with the
temperature on the digital
Intermittent tone Blinks Lights
temperature indicator.
●To silence the buzzer, press the alarm buzzer stop key (BUZZER). The alarm condition is cleared and
normal operation continues.
7. When gas cylinder A is switched to B, remove gas cylinder A and replace it with new one.
Note: Be careful when handling the empty CO
gas cylinder. Some gas may still be left in the cylinder.
2
8. When gas cylinder B becomes empty, the line is switched to gas cylinder A again.
The MCO-21GC detects that there is no more CO
does not increase for a while even after opening the CO
gas in a cylinder when the CO2density in the chamber
2
gas valve in the incubator. The supply line may
2
also be switched for other reasons even though there is gas left in the cylinder: blocking or restricting of
gas tube, reduction of CO
gas pressure, or improper opening of CO2 gas cylinder. Always check the gas
2
quantity in the cylinder before disconnecting it.
39
- 88 -
Page 41
STACKING INCUBATORS
Use the following procedure to stack incubators. This work is potentially dangerous, so contact a Sanyo
sales representative or agent.
CAUTION
x Select a floor that is strong enough to support the stacked incubators.
x Never stack three or more incubators. Doing so is dangerous.
1. Take out two stacking plates A, two stacking plates B, four protective stickers, and four screws from
accessory bag.
Note:
When stacking two incubators, use two stacking plates (A and B) for each one. Only one pair of
stacking plates is provided per incubator, so take out the stacking plates (A and B) from both incubators.
Similarly only 2 protective stickers are provided per incubator
2. Unplug both of the incubators that are to be stacked.
3. Remove the caps at the two places in the front of the top surface of the incubator that is to be on the
bottom, and attach the protective stickers to the four corners on the top surface of the incubator that is to
be on the bottom. (See Fig. A. on the page 41)
4. Secure stacking plates A with the screws that are provided.
5. Remove the front panel (4 screws) from the incubator that is to be on the top.
6. Place the upper incubator on top of the lower one, being careful not to pinch your fingers.
7. Make the lower incubator level by adjusting the leveling feet, and then level the upper incubator in the
same way.
8. Secure stacking plates A on the top surface of lower incubator to the upper one with the accessory
screws.
9. Remove from the back of the lower incubator the two hooks (4 screws) and remove from the upper
incubator the 2 screws on the lower rear panel.
10. Use the 6 screws that were removed in step 9 to secure stacking plates B to the rear panels of the
upper and lower incubators.
11. Replace the front panel (4 screws) to the upper incubator.
12. Use the hooks on the back of the upper incubator to prevent the incubators from falling over.
- 89 -
40
Page 42
Front panel
Front side
Stacking plate A
Stacking plate B
Protective sticker
Hook
Fig.A
41
- 90 -
Page 43
SPECIFICATIONS
Name CO2 incubator
Model MCO-18AC MCO-18AC(UV)
External dimensions W620 x D710 x H900 (mm)
Internal dimensions W490 x D523x H665 (mm)
Interior volume 170 L
Exterior Painted steel
Interior Stainless steel containing copper
Outer door Painted steel
Inner door Tempered glass
Tray 3 trays made of stainless steel containing copper
Access port Inner diameter; 30 mm, On the back side
Insulation Rigid polyurethane foamed-in place
Heating system DHA system (heater jacket + air jacket system)
Heater 282 W
Humidifying system Natural evaporation with humidifying pan
Temperature controller PID control system
Temperature display Digital display
CO
controller ON-OFF control system/TC sensor
2
CO
density display Digital display
2
Air circulation Fan assisted
Air filter
UV lamp
Water level sensor
Alarm
Automatic set temperature alarm, Automatic set CO
Upper limit temperature alarm, various gas/sensor/heater alarms
Remote alarm contact Allowable contact capacity: DC 30 V, 2 A
CO
inlet connection 4 to 6 mm diameter tube can be connected
2
inlet pressure 0.03MPaG (0.3kgf/cm2G, 4.3psiG)
CO
2
Accessories 3 trays, 3 sets of tray support, 1 gas tube, 1 humidifying pan,
Stacking plate A and B, 2 protective stickers, 2 tube bands
Weight 92 kg 93 kg
Optional accessory
Note: Design or specifications will be subject to change without notice.
UV system kit (MCO-18UVS3) Standard equipment
UV replacement kit (MCO-20UV), 4-20mA Interface(MCO-420MA, USA only)
Extra tray (MCO-47ST) including 2 tray supports, Half tray (MCO-25ST)
cylinder changeover system (MCO-21GC), Roller base (MCO-18RB)
SANYO Data acquisition software (MTR-5000)
- 91 -
42
Page 44
PERFORMANCE
Temperature control range
Temperature distribution
Temperature variation
CO
control range
2
variation
CO
2
Ambient temperature +5
o
± 0.25
C (ambient temperature; 25oC, setting; 37oC, CO2 5%, no load)
o
± 0.1
C (ambient temperature; 25oC, setting; 37oC, CO2 5%, no load)
± 0.15% (ambient temperature; 25
Chamber humidity
Heat emission
o
C to 35oC, Humidity: equal or less than 80% R.H.
Applicable environment
condition
Temperature: 5
(The designed performance may not be obtained when
the ambient temperature is equal or less than 15
Noise level
Power consumption
Rated voltage, frequency
Amperage
Max. 300 W Max. 310 W
110 to 120 V, 60 Hz 220 V, 60 Hz 220 to 240 V, 50 Hz
Max. 2.6 A Max. 1.4 A Max. 1.4 A
Note: The unit with CE mark complies with EC directives.
It is based on our measuring method.
o
C to 50oC (ambient temperature; 5oC to 35oC)
0 to 20%
o
C, setting; 37oC, CO2 5%, no load)
95 ± 5%RH
Max. 1116 kJ/h
o
C)
24 dB (A scale)
43
- 92 -
Page 45
CAUTION
Please fill in this form before servicing.
Hand over this form to the service engineer to keep for his and your safety.
Safety check sheet
1. Incubator contents :
Risk of infection:
Risk of toxicity:
Risk from radioactive sources:
(List all potentially hazardous materials that have been stored in this unit.)
Notes :
2.
Contamination of the unit
Unit interior
No contamination
Decontaminated
Contaminated
Others:
□Yes □No
□Yes □No
□Yes □No
□Yes □No
□Yes □No
□Yes □No
□Yes □No
□Yes □No
□Yes □No
□Yes □No
□Yes □No
□Yes □No
□Yes □No
□Yes □No
□Yes □No
□Yes □No
3. Instructions for safe repair/maintenance of the unit
a) The unit is safe to work on
b) There is some danger (see below)
Procedure to be adhered to in order to reduce safety risk indicated in b) below.
Date :
Signature :
Address, Division :
Telephone :
Product name :
CO
incubator
2
Please decontaminate the unit yourself before calling the service engineer.
Model :
MCO-18AC
MCO-18AC(UV)
- 93 -
□Yes □No
□Yes □No
□Yes □No
□Yes □No
Serial number : Date of Installation :
44
Page 46
SANYO Electric Co., Ltd
Printed in Japan
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.