− Type of protection against electric shocks: Class I
− Degree of protection against electric shocks (when the patient is in
physical contact): Type BF or type CF applied part
− Degree of protection against the ingress of harmful liquids: General
equipment
− Degree of safety of use in the presence of flammable anesthetic agent
mixed with air, oxygen, or nitrous oxide: Not suitable for use near
flammable anesthetic agent mixed with air, oxygen, or nitrous oxide
− Mode of operation: Continuous operation
■Safety standards the device conforms to
− Medical electrical equipment, part 1: General requirements for basic safety
and essential performance IEC 60601-1:2005
− Medical electrical equipment, part 1-2: General requirements for basic
safety and essential performance - Collateral standard: Electromagnetic
compatibility - Requirements and tests IEC 60601-1-2:2007
− Medical electrical equipment, part 1-6: General requirements for basic
safety and essential performance - Collateral standard: Usability IEC 60601-1-6:2010
− Medical electrical equipment, part 2-37: Particular requirements for the
basic safety and essential performance of ultrasonic medical diagnostic
and monitoring equipment IEC 60601-2-37:2007
− Medical electrical equipment, part 1: General requirements for safety IEC
60601-1:1988, A1:1991, A2:1995
− Medical electrical equipment, part 1-1: General requirements for safety -
Collateral standards: General requirements for medical electrical systems
IEC 60601-1-1:2000
− Medical electrical equipment, part 1-2: General requirements for safety -
Collateral standards: Electromagnetic compatibility - Requirements and
tests IEC 60601-1-2:2001, A1:2004
− Medical electrical equipment, part 1-4: General requirements for safety -
Collateral standards: Programmable electrical medical systems IEC 60601-1-4:1996, A1:1999
− Medical electrical equipment, part 2-37: Particular requirements for the
basic safety and essential performance of ultrasonic medical diagnostic
and monitoring equipment IEC 60601-2-37:2001, A1:2004, A2:2005
− Medical devices - Application of risk management to medical devices ISO
14971:2007
− Medical electrical equipment, part 1: General requirements for safety UL
60601-1:2003
− Medical electrical equipment - part 1: General requirements for safety
CAN/CSA C22.2 No. 601.1-M90:1990, R2003, R2005
− Biological evaluation of medical devices – part 1: evaluation and testing
ISO 10993-1: 2009
− Standard means for reporting the acoustic output of medical diagnostic
ultrasonic equipment IEC 61157:2007
■
Statements
This is the CSA symbol used in Canada and the
United States.
This mark certifies that the product conforms to
applicable EEC standards and has been certified by
the European certification agency.
This mark certifies that the product conforms to
applicable EEC standards.
GMP symbol represents the good manufacturing
practice and quality standards in accordance with
the Korean Quality Standards.
Precautions for Use
Be sure to read this operation manual thoroughly, to familiarize yourself with the
operation of the product and the relevant safety information, before attempting to
use the product.
■ Keep this manual near the product and refer to it when using the product.
■ Please familiarize yourself with the safety precautions in 'Chapter 1. Safety' and
'Chapter 8. Maintenance' in particular.
■ This operation manual does not contain clinical opinions or diagnoses. Also,
please consult the reference for each study area before evaluating the
measurement result of an application.
■ This product is an ultrasound diagnostic system, and cannot be used with your
personal computer. If you use this product in such an environment, we cannot
be held responsible for any resulting problems.
■ This product must be used by a person who possesses clinical pathology
training and/or certification; use by unqualified persons is prohibited.
■ The manufacturer is not responsible for any damage to this product caused by
user carelessness and/or neglect.
■ The contents of this operation manual may be changed without notice.
■ Products that are not manufactured by Samsung Medison are indicated with the
trademarks of their respective owners.
■ The following terms are used to highlight safety precautions that the user must
be aware of:
DANGER
WARNING
CAUTION
NOTE
Disregarding this instruction may result in death, serious injury, or
other dangerous situations.
Follow these instructions to prevent a serious accident or damage
to property.
Follow these instructions to prevent a minor accident or damage
to property.
The accompanying information covers an installation, operation, or
maintenance procedure that requires careful attention from the
user, but has little chance of leading directly to a dangerous
situation.
Revision History
The revision history of this manual is as follows.
VERSION DATE NOTE
V3.00.00-00 2015.10.30
Initial Release
System Upgrades and Manual Set Updates
Samsung Medison Ultrasound is committed to innovation and continued
improvement. Upgrades may be announced that consist of hardware or software
improvements. Updated manuals will accompany those system upgrades.
Verify that Check if this version of the manual is correct for the system version. If
not, please contact the Customer Service Department.
If You Need Assistance
If you need any assistance with the equipment, like the service manual, please
contact the Samsung Medison Customer Service Department or one of their
worldwide customer service representatives, immediately.
Image orientation (left/right and up/down, rotation)
White on black/black on white
Zoom
Trackball operation of multiple cursors
2D mode: Linear measurements and area measurements using elliptical
approximation or trace
M mode: Continuous readout of distance, time, and slope rate
Doppler mode: Velocity and trace
Auxiliary
User Interface
Pressure Limits
Humidity Limits
Temperature Limits
USB Video Printer
USB to RS-232 Serial Cable
Foot Switch(IPX8)
USB Flash Memory Media
USB HDD
Monitor
English, German, French, Spanish, Italian, Russian, Chinese
Operating: 700 – 1,060hPa
Storage: 700 – 1,060hPa
Operating: 30 – 75%
Storage & Shipping: 20 – 90%
Operating: 10 – 35OC
Storage & Shipping: -25 – 60
O
C
Chapter 1. Introduction 1 - 5
Product Configuration
This Product consists of monitor, control panel, console, peripheral devices and probes.
① Monitor
② Monitor arm
③ DVD drive
④ Speaker
⑤ Control panel
⑥ Probe holder
⑦ Keyboard
Lift
USBport
[Figure 1.1 Front of the product]
CW probe port
Probe port
Air filter
Brake
Wheels
Touch panel
Gel Warmer
(Option)
1 - 6 WS80A Service Manual
① Handle
② Storage compartments
③ Ventilation
④ Rear panel
⑤ Cable hook
⑥ Power terminal
⑦ ID Label
[Figure 1.2 Back of the product]
Chapter 1. Introduction 1 - 7
The Monitor
Ultrasound images and other information are displayed on the color LCD monitor.
▐ Screen Layout
The screen displays ultrasound images, operation menus and a variety of other information.
The main areas of the screen are ① Title Area, ② Preset Change and EZ Exam Area, ③
Image Area, ④ Thumbnail Area, ⑤ User Information Area, and ⑥ User Defined Key Area,
as shown below.
[Figure 1.3 Screen Layout]
① Title Area
Displays patient name, hospital name, application, frame rate and depth, probe information,
acoustic output information, and date and time.
② Preset Change and Ez Exam+ Area
Displays Preset Change. You can quickly change the preset of a probe. The Ez Exam+
menu will also appear if being used.
You can set up the Ez Exam+ at Ez Exam+ Setup; please refer to ‘Chapter 3.
Utilities’ for information on Ez Exam+ Setup.
NOTE
1 - 8 WS80A Service Manual
③ Image Area
Displays ultrasound images, TGC, image information, annotation, and measurement
information are also displayed.
④ Thumbnail Area
Images saved by selecting Save are displayed as thumbnails. If saving Single screens, up
to 5 images are shown in a list; for Quad screens, up to 16 images are displayed. Clicking
with the pointer will enlarge the selected thumbnail in the Image area.
⑤ User Information Area
Information that is useful to the user, such as current system status, image information,
selectable items, etc., is displayed.
⑥ User Defined Key Area
Settings for User Defined Keys, including the positions of Set and Exit, are displayed.
You can change the setting of each button in Setup > User Defined Key.
NOTE
TIP Principles of Operation of the Diagnostic Ultrasound System
Medical ultrasound images are created when the computer's digital memory
converts the high-frequency wave signals that are transmitted and received by the
probe.
As ultrasound waves propagate through the human body, they generate reflected
signals whenever they encounter a change in density. For example, reflected
signals are generated when signals pass from fatty tissues to muscle tissues.
Reflected signals return to the probe where they are converted into electronic
signals. The reflected signals are amplified and processed by analog and digital
circuits that have filters for various frequencies and response time options. Then
they are again converted into high-frequency electronic signals, and saved as a
series of digital image signals. The monitor displays the image signals stored on
the storage device in real time.
The entire process of transmitting, receiving, and processing signals is controlled
by the computer.
For information on User Key Setup, please refer to 'Chapter 3.
Utilities'.
The Control Panel
The system can be controlled by using the control panel.
Chapter 1. Introduction 1 - 9
[Figure 1.4 Control Panel]
The control panel consists of a keyboard, soft menus, buttons, dials, dial-buttons, a slider, and a
trackball.
The dial-button can be used both as a dial and a button.
▐ Functions of the Control Panel
The followings describe each control in the control panel and show how to use them.
Controls with multiple functions are described in detail in the following chapters of Chapter 3
in this manual.
Button Turns the system on/off.
On/Off
1 - 10 WS80A Service Manual
2D
M / x
PD / y
PW / z
C / Ref. Slice Dial-button
3D
4D
CW / TB
Dial-button
Dial-button
Dial-button
Dial-button
Button
Button
Dial-button
Button: Starts 2D mode.
Dial: Adjusts the 2D gain.
Button: Starts or ends M mode.
Dial: Adjusts M gain. Also, turning this dial-button when in 3D
View rotates the image along the x-axis.
Button: Starts or ends Power Doppler mode.
Dial: Adjusts the PD gain. Also, turning this dial-button, when in
3D View rotates the image along the y-axis.
Button: Starts or ends PW Spectral Doppler mode.
Dial: Adjusts the PW gain. Also, turning this dial-button when in
3D View rotates the image along the z-axis.
Button: Starts or ends Color Doppler mode.
Dial: Adjusts the C gain. Moves the reference slice horizontally in
3D View.
Starts or ends 3D mode.
Starts or ends 4D mode.
Button: Starts or ends CW Spectral Doppler mode.
Dial: Adjusts the CW gain. Adjusts top and bottom margins of
ROI in 3D View-MPR. TB is an abbreviation for “Top-Bottom”.
Angle / LR
Depth
Focus
Zoom
Q Scan
Freeze
Save
Button: Adjusts the angle of the sample volume in Spectral
Doppler mode. It is also used to adjust the BodyMarker’s probe
Dial-button
Switch
Switch
Dial-button You can magnify an image.
Button
Button
Button
cursor or indicator angle.
Adjusts left and right margins of ROI in 3D View-MPR. LR is
Dial:
an abbreviation for “Left-Right”.
Adjusts the scanning depth of the image.
Changes location and number of focus on the target location you
wish to study.
Press this button to turn the Quick Scan function on. The ‘Q
Scan’ mark will appear at the top of an image.
Pauses/resumes scanning.
Saves an image or a report displayed on the screen to the
database.
U1, 2, 4
P 1~2
Button
Button
Chapter 1. Introduction 1 - 11
Stands for ‘User Key.’ This button allows users to select a
function to apply to the button. The function of each button can
be set in Setup > User Defined Key. The selected settings will be
displayed in the User Defined Key area of the monitor.
Stands for ‘Peripheral Key.’ This button allows users to select a
function to apply to the button. The function of each button can
be set in Setup > User Defined Key. The selected settings will be
displayed in the User Defined Key area of the monitor.
BodyMarker
Text
EZ Exam
Set / Exit
Button
Button
Dial-button Use the EZ Exam and Preset Change features.
Button
Button
Button
Button
Allows the user to enter a BodyMarker over an image.
Allows the user to place text on an image.
In this mode, only the image is displayed on the screen.
Compares two independent images.
Compares four independent images.
This button is used to assign user-defined functions. The
function of each button can be set in Utility > Setup > User
Defined Key.
Set: Selects an item or value using the trackball or changes the
function of the trackball.
Exit: Exits the function currently being used and returns to the
previous state.
Pointer
Clear
Change
Calculator
Caliper
Trackball
Button
Button
Button This is used to change the current trackball function.
Button
Button
Trackball
When this is pressed, an arrow marker appears to point to parts
of the displayed image.
Deletes text, indicator, BodyMarker, and measurement result,
etc. displayed on an image.
Starts measurements by application.
Starts to measure distance, circumference, area, and volume.
Moves the cursor on the screen and scrolls through Cine
images.
1 - 12 WS80A Service Manual
Keyboard
The keyboard is used to type in text.
[Figure 1.5 Keyboard]
▐ Touch Panel
These control tools are located on both sides of the touch screen. Available buttons are as
follows:
[Figure 1.6 Touch Panel]
Chapter 1. Introduction 1 - 13
Patient
Displays the Patient Information screen, which is used for selecting a patient
ID from the list or entering new patient information.
End Exam
Probe
Report
SonoView Runs SonoView, which is the image filing program.
U3
Utility The Utility Menu appears on the touch screen.
Finishes the exam of the currently selected patient and resets the related
data.
Displays the Probe Selection screen to select or change the probe and
application.
Displays the Report screen that shows the measurement results of the
current application and other information.
Stands for User Key; functions can be assigned to these buttons as desired.
The function of each button can be set in Setup > User Defined Key. The
settings are displayed in the User Defined Key area in the monitor.
S-Flow Initiates or terminates the S-Flow (Directional Power Doppler) Mode.
ADVR Initiates recording feature.
TGC
The TGC screen will be displayed on the touch screen. TGC stands for Time
Gain Compensation.
1 - 14 WS80A Service Manual
▐ Touch Screen
[Figure 1.7 Touch Screen]
The touch screen is an operating tool that can be touched by the user to input data. The
functions that are available in the current mode are shown in the form of buttons or a
dial-button.
■Touch screen display
① Information Area: Shows the title of the touch screen currently displayed.
② Tab Area: Shows diagnostic modes and utilities under different tabs. The touch screen
can be changed by pressing one of the tabs.
③ Menu Area: The menu items that are available in the current input mode are shown in
the form of buttons. The user can access the desired menu item by pressing the
corresponding button. The menu currently in use is shown in blue.
④ Soft Menu Area: The soft menu items that are available in the current input mode are
shown.The menus in use are shown with blue borders. Press or rotate the dial-buttons
right below each menu.
⑤ Quick Preset: With predefined diagnosis mode and presets of probes frequently used by
the user, this function provides quick and easy access to frequently used probe in each
diagnosis mode
For further details about setting up Quick Preset, please refer to ‘Setup >
General > Quick Preset > Quick Preset Setup’ in ‘Chapter 3 Utilities’.
NOTE
Chapter 1. Introduction 1 - 15
※Tip! When there are two Soft Menus
When there are two menus available – upper and lower, both menus can be adjusted with the corresponding dial-button. Or tap the button for the menu you want to use on the touch screen and then use the dial-button.
▐ Adjusting the Control Panel
Do not apply excessive force to the control panel.
CAUTION
Adjust right/left
Hold the control panel handle and move it carefully to the right or left.
Adjust up/down
Use the handle at the back of the product when moving it.
Press the lever on the control panel handle and move it carefully up or down.
1 - 16 WS80A Service Manual
① ② ③ ④ ⑤
⑥
⑦ ⑧
The Console
The console consists of two parts – the inner and outer units. The inside of the console contains
ultrasound imaging components. On the exterior of the console are various connectors, probe
holders, storage compartments, handles, and wheels, etc.
▐ Rear Panel
Various peripheral devices including monitors are connected via the rear panel at the back of
the system.
① Trig port (In/Out): Not used.
② Microphone port (Input): Connect a microphone to
this port.
③ VHS port (Output): Outputs composite image to the
monitor.
④ Audio port (Output): Used to output audio signals.
⑤ S-VHS port (Output): Outputs S-VHS image to the
monitor.
⑥ DVI port (Output): Outputs the digital signal (DVI
Full HD) and analog signal (DVI R.G.B) to the
monitor.
[Figure 1.9 Rear panel]
⑦ USB port: Used to connect to USB peripheral
devices..
⑧ Network port: Used to connect to a network. You
can transfer patient information to another server
via the DICOM network.
Loading...
+ 236 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.