Troubleshooting
6-4 Samsung Electronics
6-4 50” HD D1.0 Logic Main Board T/S
If the PDP unit and the logic board operates properly, the operation LED of Figure 1 would blink at about
1 second interval.
If the unit is out of order, check the status of the operation LED through eye-inspection first.
If the behavior of the operation LED is different from that of normal state, you have to replace the board.
To check the trouble on the board, complete the following logic board test procedures described below.
Required testing equipment : - Oscilloscope (digital 400 MHz with more than 3 channels)
- Multi-meter
Other equipment : - DC power supply (5V: 1EA)
- Sub-PCB ASS'Y for JIG: 1 EA
① First, perform eye-inspection and short-circuit inspection for the power stage of the target logic
board. Then, perform the following tests on the board in sequence if no problem has been found.
② Replace the (Serial EEPROM) on the logic board with the Test EERPOM.
Change the clock setting of the logic board to internal.
(Refer to the configuration procedures in the next page)
If the Test EEPROM is not available, perform the tests by installing key scan and set the address
number 9F to 00.
➂ Implement 5V to the LD1, and check if the LED on the left-top of the board blinks at about one
second intervals.
➃ If the LED blinks too fast or the LED is not turned on, it means that the LOGIC board is not operating
properly.
➄ If no problem is found in the above step, connect sub-PCB to check the logic output. Then measure
the output waveform of the sub-PCB, and compare it with the appended waveform in normal state.
Record the test result (OK or NG).
➅ Check drive Y s/w , drive X s/w , and address signals in that order.
⑦ Set probe 1 of the oscilloscope to the trigger signal, and connect it to TP105 on the logic board.
➇ Set the oscilloscope to 2ms/div. After adjusting probe 2 to 5V/div, check the output signal.
⑨ After T/S, turn off the power supply, and disconnect the connector.
➉ Record the test result on the test sheet.