Samson TROVIS SAFE 3730 Series, TROVIS SAFE 3730-6, 3730 Series, 3730-6 Operating Instructions Manual

Page 1
Series 3730 TROVIS SAFE 3730-6 Electropneumatic
Positioner EXPERTplus Valve Diagnostics
Operating Instructions
EB 8389-1S EN
Firmware version 1.1x Edition May 2015
Page 2
DANGER!
Hazardous situations which, if not avoided, will result in death or serious injury
WARNING!
Hazardous situations which, if not avoided, could result in death or seri­ous injury
2 EB 8389-1S EN
NOTICE
Property damage message or mal­function
Note:
Additional information
Tip:
Recommended action
Page 3
Contents
1 Description ....................................................................................................7
1.1 General ........................................................................................................7
1.2 Start-up .........................................................................................................9
1.3 Diagnostic functions .....................................................................................10
1.3.1 Type of application.......................................................................................11
1.3.2 Assessment ..................................................................................................11
2 Monitoring ..................................................................................................13
2.1 Status messages ...........................................................................................13
2.1.1 Resetting status messages .............................................................................13
2.1.2 Condensed state ..........................................................................................17
2.2 Logging .......................................................................................................18
3 Data logger ................................................................................................21
3.1 Permanent data logging ...............................................................................21
3.2 Triggered data logging ................................................................................. 22
3.2.1 Triggered by condensed state ........................................................................22
3.2.2 Triggered by set point, valve position, set point deviation or signal pressure ......23
3.2.3 Triggered by binary input .............................................................................24
3.2.4 Triggered by internal solenoid valve/forced venting ........................................25
3.2.5 Triggered by set point or internal solenoid valve/forced venting .......................25
4 Valve signature ...........................................................................................27
4.1 Reference graphs .........................................................................................28
4.1.1 Analysis and monitoring ...............................................................................28
4.2 Valve signature, signal pressure(x) .................................................................29
4.2.1 Analysis and monitoring ...............................................................................30
4.3 Course of supply pressure .............................................................................31
4.3.1 Analysis and monitoring ...............................................................................32
4.4 Valve signature, friction(x) ............................................................................33
4.4.1 Analysis and monitoring ...............................................................................34
4.5 Resetting single status messages ....................................................................34
5 On/off valve ...............................................................................................37
5.1 Diagnostics for on/off valve ..........................................................................38
5.2 Analysis and monitoring ...............................................................................39
5.3 Resetting single status messages ....................................................................40
EB 8389-1S EN 3
Page 4
Contents
6 Valve position x histogram ..........................................................................41
6.1 Analysis and monitoring ...............................................................................42
6.2 Resetting single status messages ....................................................................42
7 Set point deviation e histogram ....................................................................44
7.1 Analysis and monitoring ...............................................................................45
7.2 Resetting single status messages ....................................................................46
8 Cycle counter histogram ..............................................................................49
8.1 Analysis and monitoring ...............................................................................50
8.2 Resetting single status messages ....................................................................50
9 Leakage sensor ...........................................................................................53
9.1 Start-up of the leakage sensor .......................................................................54
9.1.1 Manufacturer reference ................................................................................54
9.1.2 Process reference .........................................................................................57
9.2 Short-term monitoring ..................................................................................62
9.2.3 Resetting single status messages ....................................................................63
9.3 Long-term monitoring ...................................................................................63
9.3.1 Resetting single status messages ....................................................................63
9.4 Sound level(x) ..............................................................................................64
9.4.2 Resetting single status messages ....................................................................64
10 Course of end position .................................................................................67
10.1 Analysis and monitoring ...............................................................................68
10.2 Resetting single status messages ....................................................................68
11 Valve dead band ......................................................................................... 69
11.1 Resetting single status messages ....................................................................70
12 Partial stroke test (PST) ................................................................................73
12.1 Start triggered by the set point ......................................................................78
12.2 Start triggered by the binary input.................................................................78
12.3 Analysis and monitoring ...............................................................................79
12.4 Resetting single status messages ....................................................................80
13 Full stroke test (FST) .....................................................................................83
13.1 Analysis and monitoring ...............................................................................87
4 EB 8389-1S EN
Page 5
Contents
13.2 Resetting single status messages ....................................................................87
14 SIL operator test ..........................................................................................88
14.1 Analysis and monitoring ...............................................................................90
15 Binary input ................................................................................................ 91
®
16 Dynamic HART
variables ............................................................................93
17 Appendix ....................................................................................................95
17.1 Code list ......................................................................................................95
17.2 Error messages and recommended corrective action .....................................102
17.3 Diagnostic data points saved in a non-volatile memory ................................. 107
17.4 Resetting parameters ..................................................................................108
EB 8389-1S EN 5
Page 6
Overview
Throttling service
On/off service
No conguration required for monitoring
Travel histogram
u Section 6
Set point deviation histogram e
u Section 7
Cycle counter histogram
u Section 8
Course of end position
u Section 10
Note:
− Tests highlighted in red border require an initialization with valve signature
Tests highlighted in gray can optimize the proper functioning of safety equipment according to IEC61508 and IEC61511, provided these tests are performed regularly.
Conguration required for diagnosis
Data logger
u Section 3
Valve signature
u Section 4
Packing
u Section 8
Leakage sensor
u Section 9
Valve dead band
u Section 11
Full stroke test (FST)
u Section 13
No conguration required for monitoring
Travel histogram
u Section 6
Set point deviation histogram e
u Section 7
Cycle counter histogram
u Section 8
Course of end position
u Section 10
Conguration required for diagnosis
Data logger
u Section 3
On/off diagnosis
u Section 5
Packing
u Section 8
Leakage sensor
u Section 9
Valve dead band
u Section 11
Partial stroke test (PST)
u Section 12
Full stroke test (FST)
u Section 13
6 EB 8389-1S EN
Page 7
Description
1 Description
1.1 General
These instructions supplement the standard Mounting and Operating Instructions for the
TROVIS SAFE3730-6 Positioner (uEB8384-6S EN).
EXPERTplus is a diagnostic rmware inte-
grated into the positioner which allows the predictive, status-oriented maintenance of valves with pneumatic actuators.
EXPERTplus records the valve condition while the process is running (in automatic mode) and generates messages on the required maintenance work. In addition, numerous tests can be performed in manual mode to pinpoint emerging faults.
The diagnostic functions of EXPERTplus are completely integrated into the positioner. Di­agnostic data are compiled, saved and ana-
lyzed in the positioner itself. Classied status
messages on the state of the valve are gener­ated from the analysis.
All the parameter settings that are changed must also be downloaded onto the posi­tioner to allow them to become effective.
Local operation
Some parameters can be changed at the po­sitioner as well as over the operator interface. T
he positioner code of these parameters are written in parentheses. Refer to the Mounting and Operating Instructions of the TROVIS
SAFE3730-6 Positioner (EB8384-6SEN) for
a list of all parameters that can be changed at the positioner.
The operation described in the following sections illustrates operation using TROVIS­VIEW4. The default settings of the position­er and TROVIS-VIEW4 are written in square brackets [ ]. Settings highlighted in gray refer to operation using TROVIS­VIEW4.
Operation using TROVIS-VIEW 4/DD/DTM/
eDD
EXPERTplus allows the parameters to be viewed or changed using the TROVIS-
VIEW4 software or DD/DTM/eDD. TROVIS-VIEW4 · SAMSON operator inter-
face used to congure various SAMSON de-
vices
DTM · Device type manager to describe the device and communication properties
DD/eDD · Device description/enhanced device description
EB 8389-1S EN 7
Page 8
DescriptionDescription
Initialization mode
â â â
MAX NOM MAN/MAN2
â
Initialization including valve
signature
ß à
Yes
START-UP
â â
Zero
Valve signature (D1)
Code 81 Code 81
ß Not successful
ß Successful
Automatic change to
AUTO operating mode
à
No
Throttling or on/off service
Later recording of
reference graphs
â
Operating mode = MAN
Valve signature (D1)
All reference values are overwritten!
ß Not successful
â
Operating mode = AUTO
ß
8 EB 8389-1S EN
Page 9
Description
1.2 Start-up
The positioner must be initialized to use the full scope of the valve diagnostics. During ini­tialization the positioner adapts itself optimal­ly to the friction conditions and the signal pr
essure required by the control valve. The positioner can be initialized using one of the following initialization modes: maximum
range (MAX), nominal range (NOM), manual setting 1 (MAN) and manual setting 2 (MAN2).
Maximum range (MAX)
Initialization mode for simple start-up of valves with two clearly dened mechani­cal end positions, e.g. three-way valves
Nominal range (NOM)
Initialization mode for all globe valves
Manual setting 1 (MAN)
Initialization mode for globe valves re-
quiring OPEN position to be entered
manually
Manual setting 2 (MAN2)
Initialization mode for globe valves re-
quiring the end positions (OPEN and
CLOSED positions) to be entered manu­ally
The type of application, pressure limit and the start-up parameters required for the se­lected initialization must be entered to initial­ize the positioner.
Note:
Positioner start-up is described in detail in the associated Mounting and Oper­ating Instructions (
uEB8384-6S EN
).
The monitoring of friction, supply pressure, leakage, actuator springs, course of end po­sition and zero point shift requires additional reference tests of the valve signature to be performed. Refer to Section 4. The reference tests required for the diagnostic functions cannot be performed if the positioner has been initialized in the substitute calibration
(SUB) mode. By setting 'Initialization includ­ing valve signature' = Yes, the positioner re-
cords the reference data automatically after initialization.
Start-up
− Type of application (Code 49 - h0): Control valve, [On/off valve]
− Initialization mode (Code 6): [Maximum
range(MAX)], Nominal range (NOM), Manual setting 1 (MAN) or Manual setting 2 (MAN2)
− Initialization including valve signature (Code 48 - h0): [Yes]
− Pin position (Code 4): [Off], 17, 25, 35, 50, 70, 100, 200 mm, 90°
− Pressure limit (Code 16): 1.4 to [7.0] bar
During positioner initialization, the 'Kp level' and 'Tv level' are optimally set. If the posi-
tioner tends to overshoot impermissibly due to other disturbances, the proportional-ac­tion coefcient (Kp level) and derivative-ac­tion time (Tv level) can be adapted accord­ingly. Increment the derivative-action time until the desired behavior is reached. When the maximum value of 4 is reached for the derivative-action time, the proportional-ac-
tion coefcient can be reduced in steps.
EB 8389-1S EN 9
Page 10
Description
NOTICE
Kp level changes affect the set point deviation.
After changing the proportional-action coef-
cient 'Kp level', we recommend to recali­brate the ne lter in the positioner. Use the 'Fine-tuning after Kp change' function for
this.
Start-up > Control parameters
Proportional-action coefcient Kp level (Code 17): 0 to 17, [7]
− Derivative-action time Tv level (Code 18): Off, 1 to 4, [2]
Fine-tuning after Kp change
1.3 Diagnostic functions
There are two different types of diagnostic functions:
1. Monitoring functions
Data are compiled, saved and analyzed by the positioner while the process is running without disrupting the process. The positioner follows the set point to po-
sition the valve. A classied status alarm
or fault alarm is generated if the posi­tioner detects an event.
2. Dynamic tests
Similar to the monitoring function, data are compiled, saved and analyzed by the positioner. However, in this case, the valve position is not determined by the set point, but by the settings of the test. The dynamic tests can only be started when the conditions in the plant allow it (e.g. plant shutdown or service work in the workshop). For reasons of safety, the dynamic tests, except for partial stroke testing, can only be performed in the
MAN operating mode.
A dynamic test is stopped and the posi­tioner changes to the fail-safe position when the electrical signal falls below a certain level or when the solenoid valve is triggered or the forced venting func­tion is activated.
10 EB 8389-1S EN
Page 11
1.3.1 Type of application
Different diagnostic functions are available depending on the type of application select­ed in EXPERTplus. The types of application 'Control valve' and 'On/off valve' are available. Depending on the type of applica­tion selected, the positioner behaves differ­ently in the automatic mode (AUTO):
Control valve
The positioner uses the set point to posi­tion the valve. The valve position (current position) ap­pears in % on the display.
On/off valve
Discrete analysis of the set point The valve position (current position) in % and O/C (Open/Close) appear in alter­nating sequence on the display. See
uSection 5.
Description
1.3.2 Assessment
Table 1 shows the diagnostic functions and their statements on the condition of the valve depending on the type of application.
EB 8389-1S EN 11
Page 12
Description
Description
Table 1: Diagnostic functions and test analysis
Control
Diagnostic functions
valve
Monitoring
Data logger Depending on trigger status
Valve signature
On/off valve Breakaway time
Valve position histogram
Set point deviation histogram
Cycle counter histogram
Leakage sensor − Seat leakage u 9 on page 53 Course of end
position
Dynamic tests
Valve dead band − Dead band u 11 on page 69 Partial stroke test
(PST)
Full stroke test (FST) − Overshooting
• Full scope of functions
U Function is performed, but not analyzed
– Function is not performed
On/off
valve Assessment See section
selected
U
− Change in friction
− Supply pressure
− Defective actuator springs
− Pneumatic leakage
− Transit time
− Valve end position
U
− Change in manipulated variable range
− Manipulated variable range
− Manipulated variable range limitation
− Seat leakage
− Positioner-valve linkage
− Average set point deviation
− Packing leakage
− Dynamic stress factor
− Course of end position
− Zero shift
− Overshooting
− Dead time
− T86
− Settling time
− Dead time
− T86
− Settling time
u 3 on page 21
u 4 on page 27
u 5 on page 37
u 6 on page 41
u 7 on page 44
u 8 on page 49
u 10 on page 67
u 12 on page 73
u
13 on page 83
12 EB 8389-1S EN
Page 13
Monitoring
2 Monitoring
2.1 Status messages
The valve diagnostics integrated into the po-
sitioner generates classied status messages.
Messages generated from the analysis of the
diagnosis can be classied according to the
possible causes. See u Section 4 to 13.
The following classications are possible:
No message
If an event is classied as “No message”,
this event does not have any affect on the condensed state.
Function check
Test or calibration procedures are per­formed in the positioner. The positioner is temporarily unable to perform its control task as long as the procedure is taking place.
− Maintenance request/maintenance re- quired
The positioner still performs its control task (with restrictions). A maintenance demand or above average wear has been determined. The wear tolerance will soon be exhausted or is reducing at a faster rate than expected. Maintenance is necessary in the medium term.
Out of specication
The positioner is running outside the
specied operating conditions.
Failure
The positioner cannot perform its control task due to a functional fault in the posi­tioner itself or in one of its peripherals or
an initialization has not yet been suc­cessfully completed.
You can view these messages in TROVIS-
VIEW4 in the Diagnostics folder (> Monitor- ing) and its subfolders. The 'Positioner sta­tus', 'Valve status', 'Actuator status' and 'Valve position status' messages provide a
condensed state of the status messages of each subfolder.
2.1.1 Resetting status messages
When a status message is generated, you
should rst locate the source of the fault and
take action to remedy it. For recommended action concerning the sta-
tus messages see u Section 17.2. Status messages can be reset individually or
using the reset function. Table 2 on page 15 contains an overview on how the di­agnosis can be reset. Resetting is performed in the Diagnostics folder (> Service/mainte­nance > Reset).
If you want to keep measured data and the analysis after resetting the positioner, it is possible to upload them onto a computer.
Resetting single status messages
− Status messages represented by a code
in the positioner can be conrmed at the
positioner itself. Select the error code
and conrm it by pushing the rotary
pushbutton. See the standard instructions of the positioner u EB 8384-6S EN.
− On resetting histograms and diagrams, the data for short-term monitoring are also reset.
EB 8389-1S EN 13
Page 14
Monitoring
− Resetting measured data does not cause the diagnostic parameters and reference value to be reset as well.
− The positioner does not need to be re-ini- tialized after resetting.
Reset measured diagnostic data
Code 36 - Diag
− Parameters are reset as described in u Section 17.4.
− Resets diagnostics assessment.
− Reference values remain unchanged.
Status classication and data logs re- main saved.
− The positioner does not need to be re-ini- tialized after resetting.
− If the diagnostic data are to be reset at regular intervals, enter the time in "De­sired time until ‘Reset measured diagnos­tic data" (Code 48 - h3). The setting "00:00:00" causes the resetting at regu­lar intervals to be deactivated.
Reset start-up parameters
Code 36 - Std
− Parameters are reset as described in
u Section 17.4.
− Resets diagnostics assessment.
− Reference values are deleted.
Status classication remains unchanged.
− Data logs are reset.
− The positioner must be re-initialized after resetting.
Reset to default settings
Code 36 - DS
− Parameters are reset as described in
u Section 17.4.
− Resets diagnostics assessment.
− Reference values are deleted.
Status classication and data logs are deleted.
− The positioner must be re-initialized after resetting.
Note:
Before mounting the positioner on a new control valve, perform a reset by selecting the command 'Reset to de-
fault settings' (Code 36 - DS) and re-
initialize the positioner.
14 EB 8389-1S EN
Page 15
Monitoring
Monitoring
Table 2: Resetting the diagnosis
Resetting
single status
messages
Reference graphs
Data logger Conguration and measured values NO YES YES YES
Valve signature Signal pressure(x) Measured values YES YES YES YES
On/off valve Conguration YES NO YES YES
Valve position x histogram Measured values YES YES YES YES
Set point deviation e histogram Measured values YES YES YES YES
Cycle counter histogram Measured values YES YES YES YES
Leakage sensor Short-term monitoring 'Sensitivity sound level'
Course of lower end position Measured values YES YES YES YES
Valve dead band Conguration and
Partial stroke test (PST) Conguration and
Valve signature NO NO YES YES
Leakage sensor Manufacturer reference YES NO YES YES
Process reference YES NO YES YES
Course of supply pressure Conguration NO NO YES YES
Measured values YES YES YES YES
Friction(x) Measured values YES YES YES YES
Measured values YES YES YES YES
Short-term monitoring 'Sampling time' and
measured values
Short-term monitoring 'Sampling time' and
measured values
Short-term monitoring Measured values YES YES YES YES
and measured values
Long-term monitoring Measured values YES YES YES YES
Sound level(x) Measured values YES YES YES YES
Reference value YES NO YES YES
measured values
measured values
* Except for 'Sampling time' parameter
YES YES YES YES
YES YES YES YES
YES YES YES YES
YES YES YES YES
YES NO* YES YES
Code 36
Diag Std DS
EB 8389-1S EN 15
Page 16
Monitoring
Resetting
single status
messages
Full stroke test (FST) Conguration and
measured values
* Except for 'Sampling time' and 'Max. test duration' parameters
Alarm settings NO See uSec.17.4
Status classication according to NAMUR107 NO NO NO YES
All logged messages, see u Section 2.2 YES NO YES YES
Operating hours counter NO NO NO NO
Device in operation NO NO YES YES
Device switched on since initialization NO NO YES YES
Device in operation since initialization NO NO YES YES
YES NO* YES YES
Code 36
Diag Std DS
16 EB 8389-1S EN
Page 17
Monitoring
2.1.2 Condensed state
To provide a better overview on the condi­tion of the positioner, all status messages are summarized in a condensed state which is
made up from a summary of all classied
messages in the positioner. The status mes­sage with the highest priority determines which condensed state is set.
The condensed state appears in TROVIS-
VIEW4 on the right-hand side of the info
bar and in the Measured process values folder. See u Table 3 for a description of the icons and their meaning.
Additionally, the condensed state can be used to trigger the data logger. See uSec­tion 3.2.1.
Note:
The condensed state is marked by until the positioner data have been up­loaded.
The condensed state can be read in the posi­tioner display in Code 48 - d6. See u Table
3.
Condensed state at the fault alarm output
The condensed state also be read out at the fault alarm output if one of the following conditions occurs:
1. Condensed state 'Failure' is activated.
2. Condensed state 'Function check' is acti­vated and the fault alarm output is acti­vated.
3. Condensed state 'Maintenance required'
or 'Out of specication' is activated and
the fault alarm output is activated.
Table 3: Condensed state reading
Status message TROVIS-VIEW 4/DTM Positioner Priority
Failure
Function check
Out of specication
Maintenance request/ maintenance required
No message, OK
EB 8389-1S EN 17
red
orange
yellow
blue
green
Text e.g. TESTING, TUNE or
TEST
blinking
Page 18
Monitoring
Device settings > Alarm settings
Fault alarm at condensed state 'Function
2.
check' (Code 32): Yes
3. − Error message in case of 'Maintenance re- quired' and 'Out of specication' con-
densed states (Code 33): Yes
2.2 Logging
The last 30 generated messages are saved in the positioner with a time-stamp (logged by the operating hours counter) and with details on how long the message exists.
You can view these messages in TROVIS­VIEW4 in the Diagnostics folder (> Monitor- ing > Logging).
Logging starts automatically 15 minutes after initialization. It does not need to be activated by the user.
Messages are logged when:
Their status classication is not 'No mes-
sage'.
− Their recording in the logging is activat- ed.
The 'Internal solenoid valve/forced vent- ing/supply pressure' message is only ad­ditionally logged when the time entered
in 'Min. interval for new logging of inter­nal solenoid valve' has elapsed between the generation of two 'Internal solenoid valve/forced venting/supply pressure'
messages.
Note:
Logging can only be deactivated when
the associated status classication is set to 'No message'.
Device settings > Alarm settings > Status classication > Logging
Supply pressure: [Yes], No
Change in friction: [Yes], No
Seat leakage: [Yes], No
Packing leakage: [Yes], No
Pneumatic leakage: [Yes], No
Defective actuator springs: [Yes], No
− Manipulated variable range limitation: [Yes],
No
Course of end position: [Yes], No
Positioner-valve linkage: [Yes], No
Manipulated variable range: [Yes], No
− Change in manipulated variable range: [Yes],
No
Partial stroke test (PST) [Yes], No
Full stroke test (FST): [Yes], No
On/off valve: [Yes], No
Code 50-58, 61, 63, 76, 81: [Yes], No
Binary input: [Yes], No
Data logger: [Yes], No
− Internal solenoid valve/forced venting/supply
pressure: [Yes], No
− Min. interval for new logging of internal solenoid valve: 0 to 5000 s, [300 s]
18 EB 8389-1S EN
Page 19
The following messages are not logged if they are generated due to a hardware error:
− Packing leakage
− Pneumatic leakage
− Manipulated variable range limitation
− Course of end position
− Positioner-valve linkage
− Manipulated variable range
− Change in manipulated variable range
In this case, only the original hardware error is logged:
− x > range (Code 50)
− Internal solenoid valve/forced venting/ supply pressure (Code 54)
Transit time not reached (Code55)
− Inconsistent data memory (Code 59)
− Internal device error (Code 60)
− x signal (Code 62)
− i/p converter (Code 64)
− Hardware (Code 65)
Monitoring
EB 8389-1S EN 19
Page 20
Monitoring
Data logger
â â
Permanent Triggered
â â
Sampling time
â
Start
START-UP
â
100 measured val-
(w, x, e, p
Function
1)
â â â â â
Con-
densed
state
â â â
Trigger condition
ues
) â
out
Triggered by
, x, e
w or p
out
Trigger value
Trigger band
Trigger condition
Trigger lead time
Sampling time
Start
Solenoid
BE
valve/ forced
venting
w or solenoid
valve/forced
venting
Trigger value
Trigger band
Trigger condition
â
1)
â
Trigger event
â
Logged by operating hours
counter
â â
PROCESS
Measured values (w, x, e, pout) before trigger
event
(Number = Trigger lead/Sampling time)
Measured values (w, x, e, pout) after trigger event
+
(Number = 100 – Trigger lead/Sampling time)
â
Total 100 measured values (w, x, e, pout)
1)
The data logger is started by the software, e.g. TROVIS-VIEW 4
20 EB 8389-1S EN
Page 21
Data logger
3 Data logger
The data logger records the measured vari­ables (valve position x, set point w, set point deviation e and signal pressure p corded data are plotted against time in a graph.
Note:
The data logger is interrupted and must be reactivated when one of the following events occurs:
Failure of power supply of the posi­tioner
Failure of power supply of the exter­nal solenoid valve
). The re-
out
3.1 Permanent data logging
The measured variables are logged at the
rate dened in 'Sampling time' and saved in
a circular buffer, which holds 100 data points per measured variable at one time.
Note:
You can read the measured data
logged over the past 24 hours from
the 'Data logger' graph when the Di­agnostics folder (> Data logger) is left
open over this period.
Fig. 1: Diagnostics > Data logger
EB 8389-1S EN 21
Page 22
Data logger
Dening parameters
1. Select 'Permanent' (Function).
2. Enter sampling time.
3. Start data logger.
The 'Test information' status indicates 'Test active'.
Diagnostics > Data logger
1.
− Function: [Permanent]
2. − Sampling time: 0.2 to 3600.0 s, [1.0 s]
3. − Start data logger
Note:
Right-click 'Cancel data logger' and select 'Execute' to stop the data logger ('Test information' = 'Test not active').
3.2 Triggered data logging
Measured values are saved in a circular buf-
fer after the event dened in 'Triggered by'
has occurred (see u Section 3.2.1 to 3.2.5). The event that has triggered data logging is recorded. Data logging is terminated after 100 measured values per measured variable have been saved in the circular buffer. The
'Sampling time' determines the time between recordings. A 'Trigger lead time' greater
than 0 also leads to the measuring values before the triggering event for the time se­lected being included in the 100 measured
values per measured variable. The 'Trigger lead time' may include the value 100 x 'Sampling time' at the maximum.
Dening parameters
1. Select 'Triggered' (Function).
2. Select the triggering event.
3. Enter sampling time.
4. Start data logger.
The 'Test information' status indicates 'Test active'.
When the data logging is nished, the Progress bar indicates 'Memory full, data recording completed'.
Note:
Right-click 'Cancel data logger' and select 'Execute' to stop the data logger ('Test information' = 'Test not active').
3.2.1 Triggered by condensed state
The measured values are included in the trig­gered event when the condensed state de­ned in 'Triggered by condensed state' aris­es.
Note:
If 'Function check' is selected as the condensed state ('Triggered by con­densed state'), data are logged when a dynamic test starts. Data from each test start are logged.
Diagnostics > Data logger
1.
− Function: Triggered.
2. − Sampling time: 0.2 to 3600.0 s, [1.0 s]
3. − Triggered by: Condensed state
22 EB 8389-1S EN
Page 23
Data logger
rigger band
gg
w, x, y, e
ger band
w, x, y, e
rigger band
rigger band
w, x, y, e
− Trigger lead time:
0.0 s to 100 x 'Sampling time', [20.0 s]
Triggered by condensed state: No message, Function check [Maintenance required], Maintenance demanded, Out of specica­tion, Failure
4. − Start data logger
3.2.2 Triggered by set point, valve position, set point deviation or signal pres­sure
The measured values are included in the trig­gered data logging when the conditions for the selected measured variable (set point w, valve position x, set point deviation e or sig­nal pressure p
'Trigger band' and 'Trigger condition' are
met.
'Trigger condition' = Decreasing signal/ lower band value undercut
) dened in 'Trigger value',
out
'Trigger condition' = Increasing signal/ upper band value exceeded
Upper band limit = Trigger value Trigger value
Trigger band
+ ½ Trig
The conditions for starting a trigger event are met when the value falls below the limit
('Trigger value' – ½ 'Trigger band').
'Trigger condition' = Band exit
Upper band limit = Trigger value + ½ T
Trigger value
Lower band limit = Trigger value - ½ T
Trigger band
Trigger value Lower band limit = Trigger value – ½ T
er band
Tri
The conditions for starting a trigger event are met when the value falls below the limit
('Trigger value' – ½ 'Trigger band') or ex­ceeds the limit ('Trigger value' + ½ 'Trigger band').
The conditions for starting a trigger event are met when the value falls below the limit
This function is only active when 'Trigger band' ≠ 0.
('Trigger value' – ½ 'Trigger band').
EB 8389-1S EN 23
Page 24
Data logger
ger band
w, x, y, e
'Trigger condition' = Band entry
Upper band limit = Trigger value Trigger value Lower band limit = Trigger value - ½ Trigger band Trigger band
+ ½ Trig
The conditions for starting a trigger event are met when the value exceeds the limit
('Trigger value' – ½ 'Trigger band') or falls below the limit ('Trigger value' + ½ 'Trigger band').
This function is only active when 'Trigger band' ≠ 0.
Diagnostics > Data logger
1.
− Function: Triggered
2. − Triggered by: Valve position, set point devi- ation, signal pressure or set point
− Trigger value:
0.0 to 100.0 %, [99.0 %] (set point, valve position, set point deviation)
0.0 to 7000.0 mbar, [99.0 mbar] (signal pressure)
− Trigger band:
0.0 to 100.0 %, [99.0 %] (set point, valve position, set point deviation)
0.0 to 10000.0 1/s, [99.0 1/s] (drive sig­nal)
− Trigger lead time:
0.0 s to 100 x 'Sampling time', [20.0 s]
− Trigger condition: [Decreasing signal/lower band value undercut], Increasing signal/up-
per band value exceeded, Band exit, Band
entry
3. − Sampling time: 0.2 to 3600.0 s, [1.0 s]
4. − Start data logger
3.2.3 Triggered by binary in­put
This setting is only active when a binary in­put is installed in the positioner. If this is not
the case, the Progress bar indicates 'Cannot start data logger – trigger cannot be set'.
The measured values are included in the trig­gered data logging when the state of the bi­nary input changes.
Diagnostics > Data logger
1.
− Function: Triggered
2. − Triggered by: Binary input
− Trigger lead time:
0.0 s to 100 x 'Sampling time', [20.0 s]
3. − Sampling time: 0.2 to 3600.0 s, [1.0 s]
4. − Start data logger
24 EB 8389-1S EN
Page 25
Data logger
3.2.4 Triggered by internal so­lenoid valve/forced venting
This setting is only active when an internal solenoid valve/forced venting is installed in the positioner. See 'Internal solenoid valve/
forced venting' reading. If this is not the case, the Progress bar indicates 'Cannot start data logger – trigger cannot be set'.
The measured values are included in the trig­gered data logging when the solenoid valve is triggered or the forced venting is activat­ed.
Diagnostics > Data logger
1.
− Function: Triggered
2. − Triggered by: Internal solenoid valve/
forced venting
− Trigger lead time:
0.0 s to 100 x 'Sampling time', [20.0 s]
3. − Sampling time: 0.2 to 3600.0 s, [1.0 s]
4. − Start data logger
tions dened in 'Triggered by internal sole­noid valve/forced venting“ or 'Triggered by set point' are met.
Diagnostics > Data logger
1.
− Function: Triggered
− Triggered by: Set point or internal solenoid
valve/forced venting
Trigger value: 0.0 to 100.0 %, [99.0%]
− Trigger band: 0.0 to 100.0 %, [99.0 %]
− Trigger lead time:
0.0 s to 100 x 'Sampling time', [1.0 s]
− Trigger condition: [Decreasing signal/lower band value undercut], Increasing signal/up-
per band value exceeded, Band exit, Band
entry
3. − Sampling time: 0.2 to 3600.0 s, [20.0 s]
4. − Start data logger
3.2.5 Triggered by set point or internal solenoid valve/ forced venting
This setting is only active when an internal solenoid valve/forced venting is installed in
the positioner. See 'Internal solenoid valve/ forced venting' reading. If this is not the case, the Progress bar indicates 'Cannot start data logger – trigger cannot be set'.
The measured values are included in the trig­gered data logging when one of the condi-
EB 8389-1S EN 25
Page 26
Data logger
Valve signature
Requirements:
– Actuator: Single-acting
– Type of application: Control valve
Only applies to rst start-up after reset with Code 36 –
DS or Std.
â
Initialization including valve signature
â à
Start initialization
â
Valve signature recording
â
START-UP
Valve signature, signal pressure(x) Course of supply pressure Valve signature, friction(x)
â â
Alarm settings Recording threshold
Alar
â
m settings
ATC ATO and ATC
Zero limit,
lag time
Lower limit Upper limit
Status classication (NE 107) Status classication (NE 107) Status classication (NE 107)
Zero
x0 > (ZP
+ Zero limit)
Init
x0 < (– Zero limit)
Spring pre-load
-
ing
Reduced spring pre-loading
essure Change in friction
Supply pr
Not
available
Too low
Perm.
Too high
limit exceeded
â â â â â â
Analysis (NE 107)/logged by
operating hours counter
PROCESS
Zero
Defective actuator
springs
Analysis (NE 107)/logged by
operating hours counter
Analysis (NE 107)/logged by
operating hours counter
Supply pressure status Change in friction
Min. supply pressure
Max. supply pressure
Dwell time (supply pressure
< 1.0bar)
l time (supply pressure
Dwel
> 7.0bar)
Valve signature
canceled
Not successful
â
Lower in total range Lower/higher in mid-
position Lower/higher near
max. OPEN position
Lower/higher near CLOSED position
â
R
eset 'Pressure sensor monitoring values'
26 EB 8389-1S EN
Page 27
Valve signature
4 Valve signature
The valve signature plots the signal pressure p
as a function of valve position x.
out
All diagnostic functions dependent on the signal pressure are based on the valve sig­nature, e.g. to detect pneumatic leakage or to reveal an excessively high or low supply pressure.
Note:
Additionally, the pneumatics are mon­itored using the 'Leakage limit' param eter, which is adjustable between 0 and 100% in the Device settings fold- er (> Alarm settings). The positioner generates the 'Pneumatic leakage' message whenever the leakage ex
ceeds this limit value. 'No message' is
assigned by default.
Requirements
1. A single-acting actuator is mounted on the valve.
2. The valve operates as a control valve.
Device settings > Actuator
1.
− Principle of operation (Code 48 - d11):
Single-acting
Start-up
2.
− Type of application (Code 49 - h0):
Control valve
-
To perform monitoring while the process is
running, the reference graph must rst be
plotted.
-
Fig. 2: Start-up > Reference graphs > Valve signature
EB 8389-1S EN 27
Page 28
Valve signature
4.1 Reference graphs
During the plotting of the reference graph (Fig. 2), the valve is moved very slowly from the end position at which no pressure is ap­plied to the actuator to the position at which the maximum air signal is applied to the ac­tuator. The valve is then moved back again to its end position. During which, the posi­tioner switches to open-loop control (control without feedback).
The sensitivity of reference graph plotting de-
nes at which speed the valve is moved. The
time span how long the reference graph is to be plotted depends on the sensitivity select-
ed. Select 'High' for valves with small bench
ranges and/or high friction. After the reference graph has been plotted,
the recorded data points for signal pressure p
and valve position x are converted into
out
xed points.
Note:
The valve positions cannot be predict­ed in open-loop control (control with­out feedback). The reference graph may differ for the same valve after each logging.
Dening parameters
The reference graph is automatically plotted
after initialization when 'Initialization includ­ing valve signature' = Yes.
Start-up
− Initialization including valve signature (Code 48 - h0): [Yes]
The reference graph can also be plotted sep­arately outside initialization.
1. Switch to manual mode.
2. Select 'Sensitivity' from the drop-down list.
3. Start test.
The 'Test information' status indicates 'Test active'. 'D1' and 'TEST' are indicat-
ed in alternating sequence on the posi­tioner display.
'Function check'
is activated as the
condensed state.
Start-up
1.
− Enter operating mode (Code 0): Manual
2.Start-up > Reference graphs > Valve
signature
− Sensitivity: Low, [Medium], High
3. − Start test
4.1.1 Analysis and monitoring
The positioner records the data for lling and
venting the actuator. It then determines the characteristic values listed below:
'Mean hysteresis': Average hysteresis
(average signal pressure change Δp
relation to the spring range)
'Min. hysteresis': Lowest possible hystere- sis (minimum signal pressure change
Δp
in relation to the spring range)
out
'Max. hysteresis': Highest possible hys- teresis (maximum signal pressure change
Δp
in relation to the spring range)
out
in
out
28 EB 8389-1S EN
Page 29
Valve signature
'Detected lower spring range value': Sig- nal pressure p
lled with the minimum amount of air
'Detected upper spring range value': Sig- nal pressure p
lled with the maximum amount of air
The 'Valve signature canceled' message is
generated if the test is canceled.
Start-up > Initialization result
− Valve signature canceled (Code 81): , , ,
The 'Test status' reading indicates the reason
why the test was canceled:
No supply pressure: The supply pressure
was under 500mbar during the test.
− Valve moved too quickly. Recommended action: Increase sensitivity
− Position at p air): (1) Even though the pressure was in-
creased in steps from 0bar to the maxi­mum amount of air for lling the actua-
tor, the valve was not able to reach the maximum position in closed-loop opera­tion. (2) The minimum valve position without tight-closing is below 96%. Recommended action: Increase supply pressure.
− Position at p (1) Even though the pressure was re­duced in steps from the maximum amount of air for lling the actuator, the valve was not able to reach the minimum
when the actuator is
out
when the actuator is
out
not reached (lling with
max
not reached (venting):
min
position in closed-loop operation. The valve might be jammed. (2) The minimum valve position without
tight-closing is above 4%.
− Time-out
− Double-acting actuator
− Internal error
− Current too low
− Internal solenoid valve/forced venting triggered
SUB initialization
4.2 Valve signature, signal pressure(x)
The positioner records the data for lling and
venting the actuator during closed-loop op­eration. The recorded data can be directly compared with the reference graph. Addi­tionally, the minimum, maximum and current supply pressure are shown in the graph.
Data are recorded in the background re­gardless of the operating mode selected if a reference graph has already been plotted. Data logging does not need to be activated.
Note:
The positioner determines the mini­mum and maximum supply pressure while plotting the reference graph. The current supply pressure may fall below the minimum supply pressure and ex­ceed the maximum supply pressure.
Dening parameters
1. Record reference graph. u Section 4.2.
EB 8389-1S EN 29
Page 30
Valve signature
2. Set the limit for zero monitoring in con­trol valves with closed position ATC (air to close).
3. Select classication for status messages. See u Section 4.2.1.
Device settings > Alarm settings
2.
− Zero limit (Code 48 - d5):
0.0 to 100.0 %, [5.0 %]
− Lag time: 1 to 9999 s, [30 s]
3.Device settings > Alarm settings > Status
classication > Positioner
− Zero (Code 58):
Device settings > Alarm settings > Status classication > Actuator
Defective actuator springs
− Reduced spring pre-loading:
, [ ], ,
[ ], , ,
4.2.1 Analysis and monitoring
If the monitoring line has a smaller gradient than the reference line, this pinpoints to a re­duced compression of the actuator springs.
Diagnostics > Monitoring > Actuator
− Defective actuator springs:
Valves with ATC (air to close) closed posi­tion
In valves with ATC closed position, the posi­tioner generates a zero error when the valve position in the tight-closing position is larger than the sum of the initialization zero points
and 'Zero limit' or smaller than the negative 'Zero limit' after the adjusted 'Lag time' has
elapsed.
Diagnostics > Monitoring > Positioner
− Zero (Code 58):
, , ,
, , ,
Fig. 3: Diagnostics > Monitoring/tests > Valve signature > Valve signature, signal pressure(x)
30 EB 8389-1S EN
Page 31
Valve signature
4.3 Course of supply pressure
During control-loop operation, ve values of
the supply pressure are recorded and saved in a circular buffer. A new supply pressure is recorded and saved in the circular buffer when it deviates from the last value plotted in
the graph by the amount entered in 'Record­ing threshold'.
Data are recorded in the background re­gardless of the operating mode selected. Da­ta logging does not need to be activated. To
monitor the limits ('Lower limit' and 'Upper limit'), the corresponding limit must be acti­vated rst. The limits are automatically deter-
mined while the valve signature is being
plotted. Alternatively, user-dened limits can
be entered.
Dening parameters
1. Dene 'Recording threshold'.
2. Enter limits for monitoring. See u Sec­tion 4.3.1.
3. Select classication for status messages. See u Section 4.3.1.
1.Diagnostics > Monitoring/tests > Valve
signature > Course of supply pressure
− Recording threshold:
0.10 to 14.00 bar, [0.50 bar]
Device settings > Alarm settings
2.
Activate lower limit: [Yes], No
− Lower limit: [0.0] to 7.0 bar
Activate upper limit: Yes, [No]
− Upper limit: [0.0] to 7.0 bar
Fig. 4: Diagnostics > Monitoring/tests > Valve signature > Course of supply pressure
EB 8389-1S EN 31
Page 32
Valve signature
3.Device settings > Alarm settings > Status
classication > Supply pressure
Permissible limit exceeded: , , , [ ]
− Too high: , [ ], ,
− Fluctuates: [ ], , ,
− Too low: , [ ], ,
Not available: , , , [ ]
4.3.1 Analysis and monitoring
The positioner generates the 'Supply pres­sure status' message with the dened status classication.
− 'Too high' if the supply pressure exceeds the 'Upper limit'.
− 'Not available' if the supply pressure falls below 0.1bar.
− 'Too low' if the supply pressure exceeds the 'Lower limit'.
− 'Permissible limit exceeded' if the supply pressure exceeds 7.0bar.
A uctuating supply pressure is recognized
by the positioner whenever the supply pres-
sure continuously falls below the 'Lower limit' and rises above the 'Upper limit'. In such cases, the positioner generates the 'Supply pressure status' message with the dened status classication.
− Min. supply pressure
− Time stamp of min. supply pressure
− Max. supply pressure
− Time stamp of max. supply pressure
− Dwell time (supply pressure < 1.0 bar)
− Dwell time (supply pressure > 7.0 bar)
Diagnostics > Monitoring > Actuator
− Supply pressure status:
− Supply pressure (Code 48 - d7)
, , ,
32 EB 8389-1S EN
Page 33
Valve signature
4.4 Valve signature, friction(x)
The positioner calculates the friction during closed-loop operation and compares it with the friction determined when the reference graph was plotted.
Data are recorded automatically 15 minutes after initialization (regardless of the operat­ing mode selected) if a reference graph has already been plotted. Data logging does not need to be activated.
Dening parameters
1. Record the reference graph. See u Sec­tion 4.1.
2. Select classication for status messages. See uSection 4.2.1.
2.Device settings > Alarm settings > Status
classication > Valve
Change in friction
− Higher in total range: [
− Lower in total range: [ ], , ,
− Higher in mid-position: [ ], , ,
− Lower in mid-position: [ ], , ,
Higher near max. OPEN position: [ ], , ,
Lower near max. OPEN position: [ ], , ,
− Higher near CLOSED position: [ ], , ,
− Lower near CLOSED position: [ ], , ,
], , ,
Fig. 5: Diagnostics > Monitoring/tests > Valve signature > Valve signature, friction(x)
EB 8389-1S EN 33
Page 34
Valve signature
4.4.1 Analysis and monitoring
The positioner generates the 'Change in fric­tion' message with the selected status classi-
cation. The friction for the total range of the valve, the mid valve position and for the ranges near to the end positions are com­pared.
The positioner calculates the friction during closed-loop operation from the actuator ll­ing and venting graphs at the point where a directional change in valve travel takes place. The positioner converts the friction da­ta into xed points close to the point of di­rectional change and compares them to the reference friction.
If the friction at a xed point increases to
more than double of the reference friction, the friction is regarded to be higher.
If the friction at a xed point drops to less
than half of the reference friction, the friction is regarded to be lower.
Note:
To ensure that sufcient data points
are available for calculating the fric­tion, the valve must not be moved too quickly.
4.5 Resetting single status mes­sages
All messages generated by the valve signa­ture can be reset together by selecting and
executing "Reset 'Pressure sensor monitoring
values". These messages include:
− Change in friction
− Supply pressure status
− Defective actuator springs
− Zero
At the same time, the supply pressure data (supply pressure, min. supply pressure and max. supply pressure) are reset as well.
Diagnostics > Service/maintenance > Reset
Reset 'Pressure sensor monitoring values'
Diagnostics > Monitoring > Valve
− Change in friction:
34 EB 8389-1S EN
, , ,
Page 35
Page 36
Valve signature
On/off valve
START-UP
Prerequisite:
– Type of application: On/off valve
â
Operating point, Limit fail-safe position, Limit operating point
â
Travel time assessment limit
Valve end position limit
Status classication (NE 107)
Breakaway time increasing/decreasing
Transit time increasing/decreasing
Final travel/angle value not reached
Final travel/angle value increasing/
easing
decr
â â
On/off diagnosis Discrete set point analysis PST
Dene parameters for partial stroke
test, see u Section 12
â
First analysis after dening parameters =
Breakaway time, transit
time, valve end position (increasing/decreasing)
logged by operating hours counter
PROCESS
Reset measured 'On/off valve' values
Reset all 'On/off valve' parameters
Reference
â â
Analysis (NE 107)/
On/off valve status
â
Analysis, see
u Section 12.3
36 EB 8389-1S EN
Page 37
On/off valve
Operating point limit
Fail-safe action limit
x [%]
Operating point limit
Fail-safe action limit
x [%]
Lower range value (PST) Upper range value (PST)
x [%]
5 On/off valve
The travel range of open/close (on/off) is dened by the fail-safe position and the op­erating point. As a result, the following pa­rameters to determine the working range and set point range are not analyzed and cannot be changed.
− Lower travel/angle range value
− Upper travel/angle range value
− Lower travel/angle limit
− Upper limit for travel/angle
− Lower set point range value
− Upper set point range value
The discrete analysis of the reference vari­able is performed in automatic mode.
If the set point ( ) is below 'Operating point limit' when the automatic mode starts, the valve (
the set point increases and exceeds 'Operat­ing point limit', the valve moves to the 'Op­erating point'. The valve moves back to the fail-safe position (0% in the example) if the set point then falls below 'Fail-safe action limit'.
) moves to the fail-safe position.
w [%]
Operating point
If
If the set point ( ) is above 'Operating point limit' when the automatic mode starts, the valve (
) moves to the operating
point. The valve moves back to the fail-safe
position (0% in the example) if the set point then falls below 'Fail-safe action limit'.
w [%]
Operating point
50
25
Fail-safe position
t <6s
t [s]
Starting the partial stroke test (PST)
A partial stroke test is started when the set point ( ) moves in the range between 25
and 50% of the travel range and remains
there for longer than six seconds. See
uSection 12.1.
w [%]
Operating point =
Operating point limit
Stop PSTStart PST
50
25
>6s
t
t [s]
Fail-safe position
50
25
t > 6 s
Fail-safe action limit
Fail-safe position
t [s]
EB 8389-1S EN 37
Page 38
On/off valve
The PST diagnostic parameter 'Lower range value' must be within the dened position ± 'Tolerance limit' for the partial stroke test to start.
After the partial stroke test is completed, the valve moves back to its last position (fail-safe position or operating point).
Canceling the partial stroke test (PST)
The partial stroke test is canceled whenever the set point (
'Fail-safe action limit' and 'Operating point limit', falling below 'Operating point limit'.
After the partial stroke test is canceled, the valve moves back to its last position (fail-safe position or operating point).
Dening parameters
Note:
The parameters can only be dened in TROVIS-VIEW4 after the 'Type of ap-
plication' has been set to 'On/off valve'.
1. Select the type of application.
2. Dene parameters for on/off valve.
3. Dene parameters for partial stroke test (PST). See u Section 12.
Start-up
1.
− Type of application (Code 49 - h0): On/off
valve
2.Device settings > Positioner > Transfer
characteristic on/off
− Operating point (Code 49 - h1): 0.0 to
100%, [100 %]
) leaves the range between
− Fail-safe action limit (Code 49 - h2): 0.0 to
20.0 %, [12.5 %]
− Operating point limit (Code 49 - h5):
55.0 to 100.0 %, [75.0 %]
3.Diagnostics > Monitoring/tests > Dynamic
tests > Partial stroke test (PST)
See u Section 12
5.1 Diagnostics for on/off valve
The diagnostics for on/off valve provide statements on the valve end positions, transit times (increasing/decreasing) and break­away times (increasing/decreasing).
The data are recorded automatically for the diagnostics for on/off valve in automatic mode. Data logging does not need to be ac­tivated.
The positioner compares the current break­away time, transit time and valve position with the values recorded during the reference
measurement (rst analysis) while the plant is
running.
Dening parameters
Note:
The parameters can only be dened in TROVIS-VIEW4 after the 'Type of ap-
plication' has been set to 'On/off valve'.
1. Enter limits for monitoring. See u
5.2.
2. Select classication for status messages.
Section
38 EB 8389-1S EN
Page 39
On/off valve
(decreasing)
(increasing)
x [%]
t [s]
Device settings > Alarm settings
1.
− Travel time assessment limit (Code 49 - h7): [0.6] to 30.0 s
− Valve end position limit (Code 49 - h8): [0.3] to 100.0 %
2.Device settings > Alarm settings > Status
classication > On/off valve
− Increasing breakaway time: [
− Decreasing breakaway time: [ ], , ,
− Transit time increasing: [ ], , ,
− Transit time (decreasing): [ ], , ,
− Final travel/angle value not reached: [ ], , ,
Increasing nal travel/angle value: [ ], , ,
Decreasing nal travel/angle value: [ ], , ,
5.2 Analysis and monitoring
The analysis pinpoints to a fault when at least one of the following conditions is met while the valve is moving:
The current 'Increasing breakaway time' differs from the reference value by the
amount entered in 'Travel time assess­ment limit'.
The current 'Decreasing breakaway time' differs from the reference value by the
amount entered in 'Travel time assess­ment limit'.
The current 'Transit time increasing' dif- fers from the reference value by the
amount entered in 'Travel time assess­ment limit'.
], , ,
The current 'Transit time (decreasing)' dif- fers from the reference value by the
amount entered in 'Travel time assess­ment limit'.
The current 'Transit time increasing' dif- fers from the reference value by the
amount entered in 'Valve end position limit'.
The current 'Transit time (decreasing)' dif- fers from the reference value by the
amount entered in 'Valve end position limit'.
− The valve end position is not reached.
w [%]
Valve end position
(increasing)
Valve end position
(decreasing)
Breakaway time
(decreasing)
Transit time
Breakaway time
(increasing)
Transit time
Fig. 6: Analysis of diagnostics for on/off valve
If one of these conditions is met, the position-
er generates an 'On/off valve status' mes­sage according the selected status classica-
tion.
Diagnostics > Monitoring
− On/off valve status (Code 85):
, , ,
EB 8389-1S EN 39
Page 40
On/off valve
5.3 Resetting single status mes­sages
The message and the limit values are reset by selecting and executing "Reset measured 'On/off valve' values".
The positioner saves the reference analysis and two further test analyses. The analysis of the oldest test is deleted when another test is performed.
Diagnostics > Service/maintenance > Reset
Reset measured 'On/off valve' values
40 EB 8389-1S EN
Page 41
Valve position x histogram
6 Valve position x histogram
The valve position histogram is a statistical analysis of the plotted valve positions. The histogram provides information about where the valve mainly spends the majority of its time during its service life and whether it shows a recent trend concerning changes in its operating range.
Data are recorded automatically 15 minutes after initialization (regardless of the operat­ing mode selected). Data logging does not need to be activated.
The positioner records the valve position ev­ery second and assigns the data into pre­dened valve position classes. The distribu­tion showing how often the sound level oc­curred within a valve position class is shown in a bar graph.
'Mean value': Average class assignment
of valve positions throughout 'Monitoring duration'.
'Number of measurement points': Total number of values recorded during the
'Monitoring duration'.
'Monitoring duration'
The measured data are saved every 24 hours in the positioner.
Short-term monitoring
In order to be able to recognize any short­term changes in valve position, the position­er records the valve positions according to
the adjusted 'Sampling time' and analyses
the last 100 measured values.
'Mean time': Contains the average class assignment of valve positions for the last 100 measured values.
'Adjusted monitoring duration': 100 x
'Sampling time'
The positioner saves the valve positions in a circular buffer, which holds 100 measured values at one time.
Fig. 7: Diagnostics > Monitoring/tests > Histograms > Valve position x histogram
EB 8389-1S EN 41
Page 42
Valve position x histogram
Note:
Changing 'Sampling time' causes all
existing measured values to be delet­ed from the circular buffer.
Dening parameters
1. Set the 'Sampling time' for the short-term monitoring.
2. Select classication for status messages. See u Section 6.1.
Diagnostics > Monitoring/tests > Histograms > Valve position x histogram > Short-term monitoring
1.
− Sampling time: Adjustable as required, [00:14:24 d.h:min:s]
2.Device settings > Alarm settings > Status
classication > Valve position
Manipulated variable range
− Mainly near CLOSED position:
], , ,
[
Mainly near max. OPEN position: [ ], , ,
− Mainly in CLOSED position: [ ], , ,
Mainly near max. OPEN position: [ ], , ,
Change in manipulated variable range:
− Operating range shifted to CLOSED position: [ ], , ,
Operating range shifted to max. OPEN position: [ ], , ,
− Short-term change: [ ], , ,
6.1 Analysis and monitoring
Analysis of the histogram for control valves starts one hour after the monitoring period begins. No analysis is performed for on/off valves.
If the control valve mainly works during the monitoring duration near or in one of the end positions, the positioner generates the
'Manipulated variable range' message with the selected status classication.
For analysis of the short-term monitoring, a complete set of data (100 measured values) is required. The analysis is only active when the sampling time setting is greater or equal to one minute.
The positioner generates the 'Change in ma­nipulated variable range' message with the selected status classication whenever a
trend showing a change in the operating range is found from the analysis of the histo­gram and the short-term monitoring.
Diagnostics > Monitoring > Valve position
− Manipulated variable range:
− Change in manipulated variable range: , , ,
, , ,
6.2 Resetting single status mes­sages
The 'Manipulated variable range' and the 'Change in manipulated variable range'
messages can be reset by selecting and exe­cuting the command "Reset 'Reset valve posi­tion x histogram". This command resets all diagnostic parameters and measured data
42 EB 8389-1S EN
Page 43
of the histogram and the short-term monitor­ing.
By selecting and executing the command "Reset 'Short-term valve position x histo-
gram", the diagnostic parameters and mea­sured data in the Short-term monitoring folder are reset.
Diagnostics > Service/maintenance > Reset
Reset 'Valve position x histogram'
Reset 'Short-term valve position x histogram'
Valve position x histogram
EB 8389-1S EN 43
Page 44
Set point deviation e histogram
7 Set point deviation e histo-
gram
The set point deviation histogram contains a statistical analysis of any set point deviations recorded. It provides information on to which extent a set point deviation has occurred during the valve service life and whether faults may occur due to a restricted operat­ing range or due to seat leakage.
Data are recorded automatically 15 minutes after initialization (regardless of the operat­ing mode selected). Data logging does not need to be activated.
The positioner records the set point deviation every second and assigns the data into pre-
dened classes. The distribution showing
how often the set point deviation remained within a class is shown in a bar graph.
'Mean value': Average class assignment
of set point deviation throughout 'Moni­toring duration'.
'Number of measurement points': Total number of values recorded during the
'Monitoring duration'.
'Monitoring duration'
The measured data are saved every 24 hours in the positioner.
Short-term monitoring
In order to be able to recognize any short­term changes in set point deviation, the posi­tioner records the set point deviation accord-
ing to the adjusted 'Sampling time' and
analyses the last 100 measured values.
'Mean time': Contains the average class assignment of set point deviations for the last 100 measured values.
'Adjusted monitoring duration': 100 x
'Sampling time'
The positioner saves the set point deviations in a circular buffer, which holds 100 mea­sured values at one time.
Fig. 8: Diagnostics > Monitoring/tests > Histograms > Set point deviation e histogram
44 EB 8389-1S EN
Page 45
Set point deviation e histogram
Note:
Changing 'Sampling time' causes all
existing measured values to be delet­ed from the circular buffer.
Dening parameters
1. Set the 'Sampling time' for the short-term monitoring.
2. Select classication for status messages. See u Section 7.1.
Diagnostics > Monitoring/tests > Histograms > Set point deviation e histogram > Short­term monitoring
1.
− Sampling time: Adjustable as required, [00:14:24 d.h:min:s]
2.Device settings > Alarm settings > Status
classication > Valve
Seat leakage
− May exist: [
Device settings > Alarm settings > Status classication > Valve position
Manipulated variable range limitation
− Lower: [
− Upper: [ ], , ,
No change possible: [ ], , ,
Positioner-valve linkage
− Travel transmission not optimal: [ ], , ,
], , ,
], , ,
7.1 Analysis and monitoring
Ideally, the set point deviation should be
nearly 0%. Set point deviations greater than 1% follow-
ing in quick succession pinpoint to a limita­tion of the upper operating range. In this
case, the positioner generates the 'Manipu­lated variable range limitation' and 'Posi­tioner-valve linkage' messages with the se­lected status classications.
Set point deviations smaller than 1% follow-
ing in quick succession pinpoint to a limita­tion of the lower operating range or to seat
leakage. The positioner generates the 'Ma­nipulated variable range limitation', 'Posi­tioner-valve linkage' and 'Seat leakage' messages with the selected status classica-
tions. If almost all set point deviations during the
short-term monitoring are greater than 1% or smaller than –1%, this may indicated that
the actuator or valve stem is jammed. The
positioner generates the 'Manipulated vari­able range limitation' and 'Positioner-valve linkage' messages with the selected status classications.
Diagnostics > Monitoring > Valve
− Seat leakage:
Diagnostics > Monitoring > Valve position
− Manipulated variable range limitation: , ,
− Positioner-valve linkage: , , ,
, , ,
,
EB 8389-1S EN 45
Page 46
Set point deviation e histogram
7.2 Resetting single status mes­sages
The 'Seat leakage', 'Manipulated variable range limitation' and 'Positioner-valve link­age' messages can be reset by selecting and executing the command "Reset 'Set point de­viation e histogram" or "Reset 'Short-term set
point deviation e histogram".
By selecting and executing the "Reset 'Set
point deviation e histogram" command, all diagnostic parameters and measured data of the histogram and the short-term monitor­ing are reset.
By selecting and executing the command "Reset 'Short-term set point deviation e histo-
gram", the diagnostic parameters and mea­sured data in the Short-term monitoring folder are reset.
Diagnostics > Service/maintenance > Reset
Reset 'Set point deviation histogram e'
Reset 'Short-term set point deviation e histo-
gram'
46 EB 8389-1S EN
Page 47
Page 48
Set point deviation e histogram
t [s]
x [%]
t [s]
x [%]
Cycle counter histogram
Stem seal
â
â â â â
Self-adjusting Adjustable Other, -/- Bellows seal
â
Max. cycle count
START-UP
PROCESS
â â â
â â
Status classication (NE 107)
External leakage possibly to be
expected soon
â â
Determining the cycle spans
1
3
2
4
â â
Dynamic stress factor Dynamic stress factor
Dynamic stress factor > 90 %
â â
Analysis (NE 107)/
logged by operating hours counter
acking leakage
P
Status classication (NE 107)
External leakage possibly to be
expected soon
Determining the cycle height
2
1
4
3
Dynamic stress factor > 90 %
Analysis (NE 107)/
logged by operating hours counter
Packing leakage
â
Reset 'Cycle counter histogram'
48 EB 8389-1S EN
Page 49
Cycle counter histogram
8 Cycle counter histogram
The cycle counter histogram provides a statis­tical analysis of the cycles. As a result, the cycle c
ounter also provides information on the dynamic stress of a bellows seal and/or packing.
Data are recorded automatically 15 minutes after initialization (regardless of the operat­ing mode selected). Data logging does not need to be activated.
The positioner records the number of cycle spans when the stem seal setting is set to
'Live-loaded', 'Adjustable', 'Other' or '-/-'.
A valve cycle span starts at the point where the valve stroke changes direction until the point where it changes direction again. The valve stroke between these two changes in direction is the cycle span.
The positioner records the cycle height when
the 'Stem seal’ setting is set to 'Bellows seal'.
The cycle height is the travel performed be­tween two changes in direction.
The cycle spans or cycle heights are as­signed to classes. The distribution showing how often the cycle span or height occurred within a class is shown in a bar graph.
'Mean value': Average class assignment
of cycle height calculated from the 'No. of cycles'
'No. of cycles'
The measured data are saved every 24 hours in the positioner.
Dening parameters
1. Select the type of stem seal.
(*When 'Other' is selected as the stem seal, the additional parameter 'Max. cy­cle count' must also be set).
Device settings > Alarm settings
1.
− Stem seal: [-/-], Live-loaded, Adjustable,
Bellows seal, Other
* Max. cycle count:
1 to 1000000000, [1000000]
Fig. 9: Diagnostics > Monitoring/tests > Histograms > Cycle counter histogram
EB 8389-1S EN 49
Page 50
Cycle counter histogram
Short-term monitoring
To recognize short-term changes in the cycle spans or cycle height, the positioner analyz­es the last 100 cycle heights or cycle spans.
The positioner saves the cycle heights or cy­cle spans in a circular buffer, which holds 100 measured values at one time.
'Mean time': Contains the average class assignment for the last 100 measured values.
Dening parameters
1. Select classication for status messages.
1.Device settings > Alarm settings > Status
classication > Valve
Packing leakage
− Possibly to be expected soon: [ ,
], ,
8.1 Analysis and monitoring
The load on the bellows and/or packing can be read from the 'Dynamic stress factor' pa­rameter. The value is determined from the cy­cle spans or cycle heights and takes into ac­count the type of packing used in the valve.
A 'Packing leakage' message is generated with the selected status classication whenev-
er:
− The number of measured cycle spans ex- ceeds 450000 when 'Live-loaded' is se­lected as the stem seal.
− The number of measured cycle spans ex- ceeds 180000 when 'Adjustable' is se­lected as the stem seal.
− The number of measured cycle spans ex-
ceeds 90% of the 'Max. cycle count' when 'Other' is selected as the stem seal.
− The number of measured cycle heights
exceeds 180000 when 'Bellows seal' is
selected as the stem seal.
Measured process values
− Dynamic stress factor
Diagnostics > Monitoring > Valve
− Packing leakage:
, , ,
8.2 Resetting single status mes­sages
The 'Packing leakage' message is reset by
selecting and executing the command "Reset
'Cycle counter histogram" or "Reset 'Short-
term cycle counter histogram".
By selecting and executing the "Reset 'Cycle
counter histogram" command, all measured data of the histogram and the short-term
monitoring as well as the 'Dynamic stress factor' are reset.
By selecting and executing the command "Reset 'Short-term cycle counter histogram",
the measured data in the Short-term moni­toring folder are reset.
Diagnostics > Service/maintenance > Reset
Reset 'Cycle counter histogram'
Reset 'Short-term cycle counter histogram'
50 EB 8389-1S EN
Page 51
Page 52
Cycle counter histogram
x [%]
t [s]
Leakage sensor
Requirements
– Positioner accessories identication: Leakage sensor – Activate CLOSED end position: On
â
START-UP
Manufacturer reference
with specied standard conditions
Level (x) diagram
ß
Phase 3
â â
Reset 'Leakage sensor
manufacturer reference'
Settling time before sound level
measurement
à
Set points, sensitivity sound level
Start
ß Not successful
Phase 2
Alarm settings
Response time, preset alarm limits
Status classication (NE 107)
Alarm limit 2
exceeded
Process reference
with process conditions
User-dened
á
â
Sensitivity sound
â
level lower
á
Cancel
â
â á
ß Successful
Successful
Phase 1
Reset 'Leakage sensor process
reference'
â â
ß
Alarm limits 1,
2, 3
Recommended:
10 dB, 15 dB, 25 dB
Alarm limit 3
exceeded
Determining the mean sound levels while the valve is tightly shut
Cyclic analysis of sensor level, e.g. over dynamic HART® variables
0
PROCESS
Mean sound level exceeds
alarm limit 2
â â
Mean sound level exceeds alarm limit 3
Analysis (NE 107)/logged by operating hours counter
Seat leakage
â
Reset 'Leakage sensor process reference'
52 EB 8389-1S EN
Page 53
Leakage sensor
9 Leakage sensor
By upgrading the positioner with a leakage
sensor, it is possible to detect seat leakage when the valve is in the closed position. To achieve this, the leakage sensor measures
the sound pressure level (dB) while the valve
is tightly shut and compares the current
sound pressure level with predened alarm
limits. The positioner generates a message if the current sound pressure level exceeds one of the alarm limits.
Requirements for using the seat leakage sensor:
1. A leakage sensor is mounted to the valve. Refer to the standard mounting and operating instructions of the posi­tioner uEB8384-6 EN
Alarm limit 3
2. The leakage sensor option has been se­lected.
3. The tight-closing function has been acti­vated.
4. The leakage sensor has been put into op­eration. See u Section 9.1.
2.Start-up > Reference graphs > Leakage
sensor
Positioner accessories identication:
Leakage sensor
Device settings > Positioner > Transfer characteristic or Transfer characteristic on/ off
3.
− Activate CLOSED end position (Code 14):
On
− CLOSED end position (Code 14):
0.0 to 49.9 %, [1.0 %]
Alarm limit 2
Fig. 10: Start-up > Reference graphs > Leakage sensor > Manufacturer reference
Alarm limit 1
EB 8389-1S EN 53
Page 54
Leakage sensor
9.1 Start-up of the leakage sensor
To be able to use the full scope of functions, the response of the leakage sensor to stan­dardized conditions and to the prevailing process conditions must be measured. Fur­thermore, the limit to activate the alarm must be entered.
9.1.1 Manufacturer reference
The manufacturer reference (u Fig. 10) mea­sures the response of the leakage sensor. We recommend performing the manufacturer reference test. On request, it can also be
performed by SAMSON and must not be
performed again. In this case, the standard conditions are:
• Medium = air
• Input pressure = 4 bar
• Output pressure = atmosphere
Default values of alarm limits are A2 = 15
dB and A3 = 25 dB. If the leakage sensor has been tted later onto the valve, the alarm limits must be manually congured or adjusted by performing a manufacturer or process reference test before the leakage sensor can be used. See u Section 9.1.2.
While the manufacturer reference test is run­ning, the parameters listed below are auto­matically deactivated:
− Activate CLOSED end position
− Activate ramp function
Phase 1: The valve moves to eleven dened
set points one after the other. After reaching
a set point and after the 'Settling time before sound level measurement' has elapsed, the
leakage sensor measures the sound pressure level.
If the difference between two neighboring set
points is larger or equal to the adjusted 'Sen­sitivity sound level', the valve does not move
to the next set points. Instead, Phase 2 starts.
Phase 1 successful: The sound pressure level
exceeded the adjusted 'Sensitivity sound level' (10dB) between set points 4 and 5. Phase 2
starts.
If the 'Sensitivity sound level' is not reached
after the valve has moved to all eleven set points, the test is canceled. The canceled test
is logged with a time stamp. The 'Test infor­mation' reading indicates 'Test failed: sound level change too small'.
Phase 1 not successful: The difference in sound pressure level between two neighboring set points
is smaller than the 'Sensitivity sound level' (10dB). The reference test is canceled.
54 EB 8389-1S EN
Page 55
Phase 2: A 0.30 % band is placed around
the last set point that the valve moved to. One third of this band lies in front of the set point that the valve moved to and two thirds of the band lies behind it. The band itself is
subdivided into eleven xed points. Each xed point is located at a distance of 0.03%
to the next point. The valve moves to the new
xed points one after the other. After reach­ing a xed point and after the 'Settling time before sound level measurement' has
elapsed, the leakage sensor measures the sound pressure level. The manufacturer reference test is successful
when the difference between the rst and last newly dened points is larger or equal to the adjusted 'Sensitivity sound level'.
Leakage sensor
Phase 2 not successful: The difference in sound
pressure level between the rst and last newly dened points is lower than 'Sensitivity sound level' (10 dB). Phase 3 starts.
Phase 3: The valve moves to the set points valid for phase 1 one after the other. This is plotted in a sound level vs. travel graph. The graph shows where the point of activation is
and to which value the 'Sensitivity sound lev­el' must be reduced to allow the test to be
completed successfully.
Dening parameters
Note:
The parameters can only be dened in TROVIS-VIEW4 after the 'Positioner
Phase 2 successful: The 'Sensitivity sound level' (10 dB) is reached between the rst and last newly dened points. The manufacturer test is
successfully completed.
If the 'Sensitivity sound level' is not reached
after the valve has moved to all eleven newly
dened points, then the change in sound
pressure level is too low. In this case, phase 3 starts.
EB 8389-1S EN 55
accessories identication' has been set
to 'Leakage sensor'.
1. Switch to manual mode.
2. Dene the parameters for the manufac­turer reference test. Refer to Note con- cerning editing set points.
3. Start manufacturer reference test. The start of the reference test is docu­mented in the Time stamp. 'D8' and 'TEST' are indicated in alternat­ing sequence on the positioner display.
Page 56
Leakage sensor
Start-up
1.
− Enter operating mode (Code 0): Manual
2.Start-up > Reference graphs > Leakage
sensor > Manufacturer reference
− Settling time before sound level measure- ment: 1 to 255 s, [5 s]
Sensitivity sound level: 3 to 255 dB, [10 dB]
− Edit set points: 0.00 to 100.00 % [1: 0.00 %; 2: 0.10 %; 3: 0.20 %; 4: 0.30%; 5: 0.40 %; 6: 0.50 %; 7: 0.60 %; 8: 0.70 %; 9: 0.80 %; 10: 0.90 %,
11: 1.00%]
3. − Start manufacturer reference
Note:
By selecting and executing 'Cancel manufacturer reference' or by press­ing the rotary pushbutton, the manu­facturer reference test is canceled (Test information = 'Test canceled manual­ly'). After canceling the manufacturer reference test, the positioner remains in the manual mode.
In TROVIS-VIEW4 the test information and progress ag of the manufacturer reference
test are displayed. When the manufacturer reference test has been successfully complet­ed, the Test information reading indicates
'Test completed successfully'.
Note concerning editing set points
− The entered set points must continuously
increase from 'Set point 1' to 'Set point 11'.
− The valve moves to the set points in steps
of 0.1%. Set points must be rounded up
to two decimal places.
User-dened settings can be saved for other functions (e.g. process reference
test) in a le.
9.1.1.1 Assessment
While the manufacturer reference test is running, the positioner determines three alarm limits. The relation between Valve
position x [%] and Sound level [dB] is shown
in TROVIS-VIEW 4:
− Relation 1: Valve position and sound lev-
el at 0% position
− Relation 2: Valve position and sound lev- el at the point where the curve in the
'Leakage sensor manufacturer reference'
graph starts to rise monotonously
− Relation 3: Valve position and sound lev- el at the last measurement
9.1.1.2 Resetting single status messages
The short-term monitoring (parameters, mea­sured data and analysis) can be reset by se­lecting and executing the command "Reset
'leakage sensor manufacturer reference".
If the test is restarted and a test analysis has already been performed, the analysis of the old manufacturer reference test is overwrit­ten.
Diagnostics > Service/maintenance > Reset
Reset 'Leakage sensor manufacturer reference'
56 EB 8389-1S EN
Page 57
Leakage sensor
9.1.2 Process reference
The process reference test (Fig. 11) measures the response of the leakage sensor to pro­cess conditions. Process medium, inlet and outlet pressures as well as the process envi-
ronment can have an effect on the sensor's
response. The alarm limits are determined from the measured data.
The process reference test is performed and analyzed in the same manner as the manu­facturer reference test described in u Section
9.1.1. This test must be performed after the valve has been installed and the plant has been commissioned.
If it is not possible to perform the process ref-
erence test, the user-dened alarm limits can
be entered. See u Section 9.1.2.2.
Alarm limit 3
While the process reference test is running, the parameters listed below are automatical­ly deactivated:
− Activate CLOSED end position
− Activate ramp function
Phase 1: The valve moves to eleven dened
set points one after the other. After reaching
a set point and after the 'Settling time before sound level measurement' has elapsed, the
leakage sensor measures the sound pressure level.
If the difference between two neighboring set
points is larger or equal to the adjusted 'Sen­sitivity sound level', the valve does not move
to the next set points. Instead, Phase 2 starts.
Alarm limit 2
Alarm limit 1
Fig. 11: Start-up > Reference graphs > Leakage sensor > Process reference
EB 8389-1S EN 57
Page 58
Leakage sensor
Phase 1 successful: The sound pressure level
exceeded the adjusted 'Sensitivity sound level' (10 dB) between set points 4 and 5. Phase 2
starts.
xed points one after the other. After reach­ing a xed point and after the 'Settling time before sound level measurement' has
elapsed, the leakage sensor measures the sound pressure level.
If the 'Sensitivity sound level' is not reached
after the valve has moved to all eleven set points, the test is canceled. The canceled test
is logged with a time stamp and the 'Test in­formation' reading indicates 'Test failed: sound level change too small'.
Phase 1 not successful: The difference in sound pressure level between two neighboring set points
is smaller than the 'Sensitivity sound level' (10dB). The process reference test is canceled.
Phase 2: A 0.30 % band is placed around the last set point that the valve moved to. One third of this band lies in front of the set point that the valve moved to and two thirds of the band lies behind it. The band itself is
subdivided into eleven xed points. Each xed point is located at a distance of 0.03%
to the next point. The valve moves to the new
Phase 2 successful: The 'Sensitivity sound level' (10 dB) is reached between the rst and last newly dened points. The process reference test is
successfully completed.
The process reference test is successful when
the difference between the rst and last new­ly dened points is larger or equal to the ad­justed 'Sensitivity sound level'.
If the 'Sensitivity sound level' is not reached
after the valve has moved to all eleven newly
dened points, then the change in sound
pressure level is too low. In this case, phase 3 starts.
Phase 2 not successful: The difference in sound
pressure level between the rst and last newly dened points is lower than 'Sensitivity sound level' (10 dB). Phase 3 starts.
58 EB 8389-1S EN
Page 59
Leakage sensor
Phase 3: The valve moves to the set points valid for phase 1 one after the other. This is plotted in a sound level vs. travel graph. The graph shows where the point of activation is
and to which value the 'Sensitivity sound lev­el' must be reduced to allow the test to be
completed successfully.
Dening parameters
Note:
The parameters can only be dened in TROVIS-VIEW4 after the 'Positioner accessories identication' has been set
to 'Leakage sensor'.
1. Switch to manual mode (Code0 = MAN).
2. Dene the parameters for the process reference test. Refer to Note concerning editing set points.
3. Start process reference test. The start of the reference test is docu­mented in the Time stamp. 'D9' and 'TEST' are indicated in alternat­ing sequence on the positioner display.
Start-up
1.
− Enter operating mode (Code 0): Manual
2.Start-up > Reference graphs > Leakage
sensor > Process reference
− Settling time before sound level measure- ment: 1 to 255 s, [5 s]
Sensitivity sound level: 3 to 255 dB, [10 dB]
− Edit set points: 0.00 to 100.00 %
[1: 0.00 %; 2: 0.10 %; 3: 0.20 %; 4: 0.03%;
5: 0.04 %; 6: 0.05 %; 7: 0.06 %; 8: 0.07 %; 9: 0.08 %; 10: 0.09 %, 11: 1.00 %]
3. − Start process reference
Note:
By selecting and executing 'Cancel ref­erence reference' or by pressing the rotary pushbutton, the process refer­ence test is canceled (Test information = 'Test canceled manually'). After the test has been canceled, the positioner remains in manual mode.
In TROVIS-VIEW4 the test information and progress ag of the test are displayed. When
the process reference test has been success­fully completed, the Test information reading
indicates 'Test completed successfully'.
Note concerning editing set points
− The entered set points must continuously
increase from 'Set point 1' to 'Set point 11'.
− The valve moves to the set points in steps
of 0.1%. Set points must be rounded up
to two decimal places.
User-dened settings can be saved for other functions (e.g. another process ref-
erence test) in a le.
9.1.2.1 Assessment
While the process reference test is running, the positioner determines three alarm limits. The relation between Valve position x [%]
and Sound level [dB] is shown in TROVIS­VIEW4:
− Relation 1: Valve position and sound lev-
el at 0% position
EB 8389-1S EN 59
Page 60
Leakage sensor
33
x
[%]
t [s]
− Relation 2: Valve position and sound lev- el at the point where the curve in the
'Leakage detection process reference'
graph starts to rise monotonously
− Relation 3: Valve position and sound lev- el at the last measurement
9.1.2.2 Alarm settings
After connecting the leakage sensor and performing the manufacturer and process reference tests, the positioner is able to pin­point any seat leakage. To do this, it records the sound level in closed-loop operation while the valve is tightly shut. The seat leak­age monitoring is performed automatically while the process is running.
The alarm limits detected in the manufacturer reference test, process reference test or user-
dened settings can be selected. When user­dened limits are entered, the alarm limits
must rise continuously from 'Alarm limit 1' to 'Alarm limit 3'.
During closed-loop operation, the mean sound levels while the valve is tightly shut are compared to the alarm limits. Which mean sound level is to be used for comparison can
be selected in 'Response time':
Very quick: The mean value calculated from the current sound level and from the last four sound levels measured while the valve is tightly shut is used for monitoring (Fig. 12: Tight-closing event 31 and sound level E to I).
Quick: The mean value calculated from all sound levels measured while the valve is tightly shut is used for monitoring (Fig. 12: Tight-closing event 31 and sound level A to I).
4 … 29
[dB]
031
H
G
D
F
A
B
C
I
E
t [s]
1
0
Fig. 12:
Tight-closing process and calculated mean values for analysis of the process reference test
2
60 EB 8389-1S EN
Page 61
Leakage sensor
Slow: The mean value calculated from the last 30 sound levels measured during short-term monitoring (see uSection 9.2) is used for monitoring (Fig. 12: Tight­closing event 2 to Tight-closing event 31 with all sound levels).
Very slow: The mean value calculated from all the sound levels measured dur­ing long-term monitoring (see u Section
9.3) is used for monitoring (Fig. 12: Tight-closing event 1 to Tight-closing event 31 with all sound levels).
The 'No alarm triggering' setting deactivates
the alarm function.
Dening parameters
1. Dene alarm parameters.
2. Select classication for status messages. See u Section 9.1.2.3.
1.Start-up > Reference graphs > Leakage
sensor > Process reference
Response time: [No alarm triggering], Very quick, Quick, Slow, Very slow
1)
− Preset alarm limits
reference], Process reference, User-dened
2.Device settings > Alarm settings > Status
classication > Valve
− Alarm limit 2 exceeded:
], , ,
[
− Alarm limit 3 exceeded: [ ], , ,
1)
On selecting the alarm limit 'User-dened', we
recommend checking the correct setting of the alarm limits using an operating time of one to three months based on the measured data in
'Sound level monitoring' diagram (see
u Section 9.4).
: [Manufacturer
9.1.2.3 Monitoring
If the determined mean sound level exceeds
the 'Alarm limit 2', the positioner generates a 'Seat leakage' message with status classi­cation selected for 'Alarm limit 2 exceeded'.
If the determined mean sound level exceeds
the 'Alarm limit 3', the positioner generates a 'Seat leakage' message with status classi­cation selected for 'Alarm limit 3 exceeded'.
Diagnostics > Monitoring > Valve
− Seat leakage:
, , ,
9.1.2.4 Reset
The process reference test (diagnostic param­eters, measured data and analysis) and the 'Seat leakage' message can be reset by se­lecting and executing the command "Reset
'leakage sensor process reference".
If the test is restarted and a test analysis has already been performed, the analysis of the old process reference test is overwritten.
Diagnostics > Service/maintenance > Reset
Reset 'Leakage sensor process reference'
EB 8389-1S EN 61
Page 62
Leakage sensor
9.2 Short-term monitoring
Short-term monitoring provides an insight in­to short-term changes in the sound level while the valve is tightly shut.
Data are recorded in the background re­gardless of the operating mode selected. Da­ta logging does not need to be activated.
The leakage sensor records the sound level when the valve leaves the tightly shut posi­tion or whenever the sound level changes by
2 dB. A mean value is calculated from the
recorded sound level and last four recorded sound levels. If this value deviates from the previous mean
value by 'Sensitivity sound level', the new
mean value is saved in the short-term moni­toring.
The last 'Mean value' recorded in the short-
term monitoring is indicated. The positioner saves the mean values of the
sound level and valve travel in a circular buf­fer, which holds 30 measured values at one time together with a time stamp. The saved values can be read in the Measured data
assessment folder.
Dening parameters
1. Enter the sensitivity sound level.
1.Diagnostics > Monitoring/tests > Leakage
sensor > Short-term monitoring
Sensitivity sound level: 3 to 255 dB, [3 dB]
Fig. 13: Start-up > Reference graphs > Leakage sensor > Short-term monitoring
62 EB 8389-1S EN
Page 63
Leakage sensor
9.2.3 Resetting single status messages
The short-term monitoring (diagnostic param­eters, measured data and analysis) can be re­set by selecting and executing the command
"R
eset 'Short-term leakage sensor monitor-
ing". The data in the Measured data assess­ment folder are reset as well.
Diagnostics > Service/maintenance > Reset
Reset 'Short-term leakage sensor monitoring'
9.3 Long-term monitoring
To obtain a sound level trend over a long pe­riod of time, the long-term monitoring con­tains all the mean values saved in the short­term monitoring since the last reset:
'Mean value': The average sound level
calculated from 'No. of measuring points'
'No. of measuring points'
Data are recorded in the background re­gardless of the operating mode selected. Da­ta logging does not need to be activated.
9.3.1 Resetting single status messages
The measured data of the long-term monitor­ing are reset by selecting and executing the
command "Reset 'Leakage detection - Long-
term monitoring".
Diagnostics > Service/maintenance > Reset
Reset 'Leakage sensor - Long-term monitoring'
EB 8389-1S EN 63
Page 64
Leakage sensor
9.4 Sound level(x)
The sound level monitoring is shown in a his­togram. The distribution of recorded sound
levels within xed classes of valve position x
is revealed. The leakage sensor records the sound level
every second and assigns the data into pre­dened valve positions classes. The distribu­tion showing how often the sound level oc­curred within a valve position class is shown in a bar graph.
Data are recorded in the background re­gardless of the operating mode selected. Da­ta logging does not need to be activated.
9.4.2 Resetting single status messages
The measured data of the sound level moni­toring are reset by selecting and executing
the command "Reset 'Leakage sensor sound
level(x)".
Diagnostics > Service/maintenance > Reset
Reset 'Leakage sensor sound level(x)'
Fig. 14: Start-up > Reference graphs > Leakage sensor > Level(x)
64 EB 8389-1S EN
Page 65
Page 66
Leakage sensor
x [%]
t [s]
Course of end position
Prerequisite
– Activate CLOSED end position: On
â
START-UP
PROCESS
à
à
irst zero point =
F
Reference
Zero calibration
Recording threshold
â
Alarm settings (NE 107)
Course of end position
0
Valve signature complete?
u Section 4.1
ß NO
YES
â
à
ß
Zero shifted monotonically increasing/decreasing, alternating
Subsequent zero calibration over
Code 6 - ZP
â
Status classication (NE 107)/logged by operating hours counter
Course of end position
â â
á
Reset 'Reference values for course of
lower end position'
66 EB 8389-1S EN
Reset 'Course of lower end position'
Page 67
Course of end position
10 Course of end position
The course of end position is used to detect an alternating zero point or a creeping zero point shift due to seat and plug wear or dirt between the seat and plug.
Data are recorded in the background re­gardless of the operating mode selected if the tight-closing function (Code 14) is active. Data logging does not need to be activated.
The course of the end position records the valve position x and the signal pressure p together with the time stamp by the operat­ing hours counter when the valve moves to the lower end position. The new recorded valve position is compared to the last saved
zero point. If it differs by the 'Recording
out
threshold' from the last value, the data of the
new zero point are saved. A graph of the recorded valve positions at
the lower end position is plotted over time. The positioner saves the valve positions in a
circular buffer, which holds 30 measured values at one time. The recorded measured data are listed in the Lower end position folder.
Dening parameters
1. Activate tight-closing function.
2. Dene 'Recording threshold'.
3. Select classication for status messages.
Fig. 15: Diagnostics > Monitoring/tests > Course of end position
EB 8389-1S EN 67
Page 68
Course of end position
Device settings > Positioner > Transfer characteristic or Transfer characteristic on/ off
1.
− Activate CLOSED end position (Code 14):
On
− CLOSED end position (Code 14):
0.0 to 49.9 %, [1.0 %]
Device settings > Alarm settings
2.
− Recording threshold:
0.10 to 5.00 %, [0.25 %]
3.Device settings > Alarm settings > Status
classication > Valve position
Course of end position
− Monotonically increasing: [
− Monotonically decreasing: [ ], , ,
− Alternating: [ ], , ,
], , ,
10.1 Analysis and monitoring
A reference zero point must be recorded to analyze the course of the end position. This is recorded during the valve signature refer­ence test (see u Section 4.1). In case a refer-
ence test has not been performed, the rst
zero point that the valve moved to serves as the reference value. The reference value is represented by a straight line in the course of end position graph.
not exceed the 'Zero limit' (Code 48 -
d5).
If the analysis of course of the end position pinpoints to a fault, the positioner generates
the 'Course of end position' message with the selected status classication.
Diagnostics > Monitoring > Valve position
− Course of end position:
, , ,
10.2 Resetting single status mes-
sages
The 'Course of end position' message and
the measured data for the course of the end position can be reset by selecting and exe-
cuting "Reset 'Course of lower end position".
If only the reference zero point is to be reset,
select and execute "Reset 'Reference values
for course of lower end position".
Diagnostics > Service/maintenance > Reset
Reset 'Course of lower end position'
Reset 'Reference values for course of lower end
position'
Note:
If the reference value has been reset by selecting and executing "Reset 'Reference values for course of lower end position" (see
u Section 2.1.1
),
the rst zero point that the valve
moves to after the reset serves as the new reference value, provided it does
68 EB 8389-1S EN
Page 69
Valve dead band
11 Valve dead band
The difference in set point w that causes a minimal change in the valve position x is
termed ‘dead band’.
The valve dead band is affected by the fric­tion hysteresis and the elastic processes in the valve stem packing.
The test is started in the manual mode.
The positioner species the set point w in a dened test range ('Lower range value' and 'Stop') in small steps and records the re-
sponse of the valve position x after waiting a
'Waiting time after step change'. The step
height is determined automatically from the
number of measured values ('No. until re­versing') and the dened test range. The as-
cendent and descendent are plotted within
the test range. The response of the valve po-
sition x to the change in set point (∆w) is
plotted in a graph. The dead band is analyzed in the positioner
when a step height is smaller than 0.2%.
'Mean dead band': Average change in set point that causes a minimal change in the valve position.
'Min. dead band': Minimum change in set point that causes a minimal change in the valve position.
'Max. dead band': Maximum change in set point that causes a minimal change in the valve position.
Fig. 16: Diagnostics > Monitoring/tests > Dynamic tests > Valve dead band
EB 8389-1S EN 69
Page 70
Valve dead band
Dening parameters
1. Switch to manual mode (Code0 = MAN).
2. Select test parameters.
3. Start test.
The 'Test information' status indicates 'Test active'. 'D3' and 'TEST' are indicat-
ed in alternating sequence on the posi­tioner display.
'Function check'
condensed state.
Diagnostics > Service/maintenance > Operating mode
1.
− Enter operating mode (Code 0): Manual
2.Diagnostics > Monitoring/tests > Dynamic
tests > Valve dead band
− Lower range value: 0.0 to 100.0 %,
[50.0%]
− Stop: 0.0 to 100.0 % [52.0 %]
− Waiting time after step change:
0.1 to 25.0 s, [1.0 s]
No. until reversing: 1 to [50]
3. − Start test
is activated as the
11.1 Resetting single status mes­sages
The diagnostic parameters and measured data of the last test are reset by selecting and
executing the command "Reset 'Valve dead
band". If the test is restarted and a test analysis has
already been performed, the analysis of the old test is overwritten.
Diagnostics > Service/maintenance > Reset
Reset 'Valve dead band'
Note:
Cancel the test by right-clicking 'Can­cel test' and selecting 'Execute' or by pressing the rotary pushbutton at the positioner. After the test has been can­celed, the positioner remains in man­ual mode.
In TROVIS-VIEW4 the test information and progress ag of the test are displayed. The 'Test information' status indicates 'Test not active' after the test is nished.
70 EB 8389-1S EN
Page 71
Page 72
Valve dead band
x [%]
t [s]
x [%]
x [%]
t [s]
Partial stroke test (PST)
Step response settings: Lower range value, Stop, Tolerance limit, Settling time
before starting test, Waiting time after step change, Sampling time, Number of
100
90
Max. test
duration
Max. break-
away time
START-UP
not moved
before
reaching the
max.
breakaway
time.
The valve has
before 'Perm.
step end' has
â
not reached
The valve has
Max. test
duration
exceeded
â
Other cancelation causes: solenoid valve trig-
gered/forced venting active; Supply pressure too
low; Too much friction; Step start too large;
Change in set point; Current too low
â â â â â â â
TEST/PROCESS
step responses
Activate ramp function
â
â â
No Yes
Ramp time (increasing)
Ramp time (decreasing)
â
elation conditions
Canc
erm. time until
P
step end
x monitoring value
Status classication (NE 107)
PST/FST
â
Observe start conditions
Table 4
(see
)
â
Test interval
â â
Test start
â â
the 'Stop'
position
time until
elapsed.
â
Immediately
â
T start
PS
â â
Time-con
trolled*
First PST after dening parameters =
Reference
â
Start further PST
Analysis (NE 107)/logged by operating hours counter
PST/FST
â
Reset 'PST parameters'
â
100
Δp out moni
90
-
toring value
The signal
-
The valve position has fallen below the monitoring value.
â
Other cancelation causes: solenoid valve trig­gered/forced venting active; Supply pressure too low; Too much friction; Step start too large; Change in set point; Current too low
pressure change has exceeded or fallen below the reference value.
â
PST tolerance
band
The valve position deviation exceeds the PST tolerance band.
â
* Note! The positioner is write-protected when tests are performed according to a schedule (local operation and operating software).
72 EB 8389-1S EN
Page 73
Partial stroke test (PST)
12 Partial stroke test (PST)
The partial stroke test (PST) is particularly suitable for the status-oriented detection of malfunctions in pneumatic shut-off valves. As a result, the probability of failure on demand (PFD) can be reduced and it may be possible to extend maintenance intervals.
A shut-off valve normally in its end position can be prevented from seizing up or getting jammed. The breakaway torque must be overcome when the plug starts to move out of its open position. This breakaway torque depends on the seal, deposits on the seat and plug, the process medium and friction in the valve trim. After the breakaway torque has been overcome, it can be assumed that the valve is able to completely close. Record­ing the test results additionally allows an analysis of the dynamic control response.
The partial stroke test can be performed once (test immediately started) or, with an on/off valve in automatic mode, regularly (time-controlled), provided the start condi­tions are met (see u Table 4):
− A control valve is in the manual mode.
− An on/off valve is in the manual or auto-
matic mode. In automatic mode, the test is only started when the Set point w is
greater than the 'Fail-safe action limit' (Code49 - h2).
The following listed parameters are activated while the partial stroke test is being per­formed:
− Characteristic selection (Code 20): Linear
Required transit time OPEN (Code 21):
Variable
− Required transit time CLOSED (Code 22):
Variable
Fig. 17: Diagnostics > Monitoring/tests > Dynamic tests > Partial stroke test (PST)
EB 8389-1S EN 73
Page 74
Partial stroke test (PST)
Operating point –
Start PST
t [s]
During the partial stroke test, the valve moves from its current operating point to a
dened 'Stop' value and back to the initial
position again. The change in travel can be performed ei-
ther in steps or in a ramp function (Fig. 18). For the test in a ramp function, additionally the ramp times for the increasing and de-
creasing ramps need to be dened.
To perform a partial stroke test, the 'Lower range value' must be near the current oper­ating point ± 'Tolerance limit'.
After being activated, the test does not start
until the 'Settling time before starting test' has elapsed. Based on the operating point, the valve moves until reaching the 'Stop'
x [%]
Operating point + Tolerance limit
Lower range value
Tolerance limit
Operating point
valve position. The valve remains in this po-
sition for the time dened by the 'Waiting time after step change' before performing a
second step change in the opposite direction
from 'Stop' to the operating position. The 'Sampling time' denes the time interval be-
tween which the measured values are re­corded during the test.
Test cancelation conditions
Various cancelation conditions provide addi­tional protection against the valve slamming shut or moving past the end position. The positioner cancels the partial stroke test when one of the following cancelation condi-
tions is fullled:
Start PST
Stop
t
decreasing
t
increasing
Settling time before
starting test
Waiting time after
‘Lower range value’ – ‘Stop’
‘Lower range value’ – ‘Stop’
step change
100
100
Settling time before
starting test
x ‘Ramp time (decreasing)’
x ‘Ramp time (increasing)’
Waiting time after
t
decreasing
step change
t
Fig. 18: Course of the partial stroke test with step response (left) and ramp function (right)
74 EB 8389-1S EN
increasing
Page 75
Partial stroke test (PST)
Cancelation conditions, time-out
'Max. test duration': The test is canceled when the maximum permissible test du­ration is reached.
'Max. breakaway time': The test is can- celed when the valve position has
reached less than 10% of the PST target travel after the dened time has elapsed.
This cancelation condition only becomes effective after it has been activated ("Ac-
tivate 'Max. breakaway time" = [Yes]).
− 'Perm. time until step end': The test is canceled when the time entered in 'Set­tling time before starting test' has
elapsed before the valve has reached its 'Stop' position. This cancelation condi­tion only becomes effective after it has been activated ("Activate 'Perm. time un­til step end" = [Yes]).
Cancelation condition - valve position x (to check the valve position)
'x monitoring value': The test is canceled as soon as the valve position falls below the adjusted value. This cancelation con­dition only becomes effective after it has
been activated ('Activate x monitoring' =
Yes).
Cancelation condition, signal pressure Δp
out
'∆p out monitoring value': The test is canceled if the change in signal pressure exceeds or falls below the reference val­ue.
The reference value is made up of 'Δp out reference value' and 'Δp out moni­toring value'. This cancelation condition
only becomes effective after it has been
activated ('Activate Δp out monitoring' =
Yes).
The positioner determines the '∆p out reference value' from both signal pres-
sures issued at the start and end of the step. It only applies to adjusted step change and ramp values.
Cancelation condition, tolerance band
'PST tolerance band': The test is canceled as soon as the deviation of the valve po­sition (in relation to the step end value (Stop)) exceeds the adjusted PST toler­ance band. This cancelation condition only becomes effective after it has been
activated ('Activate PST tolerance band monitoring' = Yes).
Note:
The partial stroke test must be per­formed with deactivated cancelation conditions for valves with double­acting actuator and pneumatic booster as well as for valves that
have been initialized using the SUB
mode (substitute calibration).
Excessive overshooting may occur in
valves tted with boosters. In this
case, the cancelation conditions 'x
monitoring value' and 'PST toler
ance band' must be increased ac­cordingly.
Additionally, the partial stroke test is can­celed when one of the following events aris­es:
-
EB 8389-1S EN 75
Page 76
Partial stroke test (PST)
− Cancel internal solenoid valve/forced venting: The test was canceled by the ac­tivation of the solenoid valve/forced venting function.
− Canceled by control loop error: A control loop error has occurred.
− Set point start difference too high: During
the step change, 'Lower range value' is outside the range (operating point ± 'Tol­erance limit').
− Set point change: Time-controlled start of the test. Due to a set point change before
the step change, 'Lower range value' is outside the range (operating point + 'Tol­erance limit').
− Current too low
− Supply pressure too low
− Electric current change: The partial stroke test of a control valve is canceled when
the change in electric current ≥ 'Tolerance band' (Code 19) arises.
The partial stroke test of an on/off valve is canceled when the valve moves from the operating position to the fail-safe po­sition or from the fail-safe position to the operating position due to a change in electric current.
Note:
The 'Measured data storage out of memory' reading (Failure) is generat-
ed when the 'Sampling time' is too
low. After recording 100 measured values per variable, logging is stopped, but the test continues until it is completed.
After the partial stroke test is canceled, the
'PST status' reading indicates 'Not success­ful'. The reason for cancelation is marked by the 'Failure' message in the Measured data
assessment folder (> Current test).
Dening parameters
1. Dene parameters for the partial stroke test. See also u Note concerning setting the PST diagnostic parameters. (The de­fault settings partly depend on the clos­ing position ATO/ATC.)
2. Dene parameters for cancelation condi­tions.
3. Select classication for status messages.
4. Start the partial stroke test.
The 'Test information' status indicates 'Test active'. 'D4' and 'TEST' are indicat-
ed in alternating sequence on the posi-
tioner display. 'Function check'
is ac-
tivated as the condensed state.
1. Diagnostics > Monitoring/tests > Dynamic
tests > Partial stroke test (PST)
− Lower range value (Code 49 - d2): 0.0 to
100.0 %, [ATO: 100.0 %; ATC: 0.0 %]
− Stop (Code 49 - d3): 0.0 to 100.0 %
[ATO:90.0 %; ATC: 10 %]
− Tolerance limit: 0.1 to 10.0 %, [2.0 %]
− Activate ramp function (Code 49 - d4):
[Yes], No
− Settling time before starting test
(Code 49 - d7): 1 to 240s, [2 s]
− Waiting time after step change (Code 49 - d8):
1.0 to 240.0 s, [2.0 s]
− Sampling time (Code 49 - d9):
0.2 to 250.0s, [0.8s}
76 EB 8389-1S EN
Page 77
Partial stroke test (PST)
Only when the ramp function is activated:
− Ramp time (decreasing) (Code 49 - d5): 0 to 9999 s, [600 s]
− Ramp time (increasing) (Code 49 - d6): 0 to 9999 s, [60 s]
2. − Max. test duration (Code 49 - E7): 30 to 25000 s, [90 s]
Activate 'Max. breakaway time' [Yes], No
− Max. breakaway time: 0.0 to 25000 s,
[30.0s]
Activate 'Perm. time until step end' [Yes], No
− Perm. time until step end:
0.0 to 25000 s, [70.0 s]
− Activate x monitoring (Code 49 - E0): [Yes],
No
− x monitoring value (Code 49 - E1): –10.0 to 110.0 %, [ATO: 0.0 %; ATC: 85 %]
Activate Δp out monitoring (Code49 - A8):
[Yes], No
− Δp out monitoring value (Code 49 - A9):
0.00 to 7.00bar, [1.00 bar]
− Activate PST tolerance band monitor-
ing(Code 49 - E5): Yes, [No]
− PST tolerance band (Code 49 - E6):
0.1 to 100.0 %, [5.0 %]
3.Device settings > Alarm settings > Status
classication > PST/FST
− x cancelation: [
Δp out cancelation: [ ], , ,
− Tolerance band exceeded: [ ], , ,
− Max. test duration exceeded:
], , ,
[
− Test canceled manually: [ ], , ,
− Measured data memory full: [ ], , ,
], , ,
− Cancel internal solenoid valve/forced venting: [ ], , ,
− Canceled by control loop error: [ ], , ,
− Set point start difference too high: [ ], , ,
− Set point change: [ ], , ,
− Current too low: [ ], , ,
− Max. breakaway time exceeded: [ ], , ,
− Perm. time until step end exceeded: [ ], , ,
− Canceled by supply pressure: [ ], , ,
4.Diagnostics > Monitoring/tests > Dynamic
tests > Partial stroke test (PST)
Either:
− Enter test start (Code 49 - A2) =
[Immediately]
− Start test
Or:
(only when 'Type of application' = On/off
valve)
− Enter test interval (Code 49 - A3): [1 h] to 2345 d
− Enter test start (Code 49 - A2) = Time-
controlled
NOTICE! The positioner is write-protected when tests are performed according to a schedule (local operation and operating software).
Code 0 reading: "O/C" and "PST" in
alternating sequence
Code 3 reading: "PST" blinks.
EB 8389-1S EN 77
Page 78
Partial stroke test (PST)
Note:
Cancel the test by right-clicking 'Can­cel test' and selecting 'Execute' or by pressing the rotary pushbutton at the positioner. After the test has been can­celed, the positioner remains in select-
ed mode. The 'PST status' reading in dicates 'Not successful'.
In TROVIS-VIEW4 the test information and progress ag of the test are displayed. The 'Test information' status indicates 'Test not active' after the test is nished.
Note concerning setting the PST diagnostic parameters
− We recommend only to start the partial stroke test when the valve is in the end position. In on/off valves, the start value must be the same as the operating point.
The 'Ramp time (increasing)' must be greater than the corresponding value for
'Min. transit time CLOSED' (Code 41)
determined during initialization.
The 'Ramp time (decreasing)' must be greater than the corresponding value for 'Min. transit time OPEN' (Code 40) de­termined during initialization.
The 'Sampling time' must not be lower
than the indicated 'Recommended min. sampling time'. The 'Recommended min. sampling time' is calculated from the 'Ex­pected duration of test'.
12.1 Start triggered by the set point
The partial stroke test of on/off valves is trig­gered when the set point w moves in the
range between 25 and 50% of the travel
and remains there for longer than six sec-
-
onds. See u Section 5 and chart u page
72.
The PST diagnostic parameter 'Lower range value' must be within the dened position ± 'Tolerance limit' for the partial stroke test to start.
The test and its cancelation are described in section 12, while the test assessment is de­scribed in u Section 12.3.
12.2 Start triggered by the binary input
If the positioner is tted with the optional bi­nary input, the partial stroke test can be started by the binary input when the condi­tions to start the partial stroke test are met:
− A control valve is in the manual mode.
− An on/off valve is in the manual or auto- matic mode. In automatic mode, the test
is only started when the 'Safety set point' is greater than the 'Fail-safe action limit'
(Code 49 - h2).
The test and its cancelation are described in section 12, while the test assessment is de­scribed in u Section 12.3.
It is important to make sure that the diag­nostic parameter 'Lower range value' of the
78 EB 8389-1S EN
Page 79
Partial stroke test (PST)
partial stroke test is within the range of the 'Safety set point' ± 'Tolerance limit'.
Dening parameters
1. Select 'Binary input' in 'Positioner acces­sories identication'.
2. Congure the binary input.
3. Select classication for status message.
Device settings > Positioner > Options
1.
Positioner accessories identication: Binary
input
2.Device settings > Positioner > Options >
Conguration, binary input
− Binary input conguration: [For oating contact (switch function)], For non-oating
contact (0-24 V)
− Select function: Start PST
Binary input control: Activate function: switch open, [Activate function: switch closed]
Safety set point: 0.0 to 100.0 %, [50.0%]
3. Binary input classication: ], , ,
[
Note:
Further details on optional binary in­put u Section 15.
12.3 Analysis and monitoring
The analysis of the last three partial stroke tests are saved with a time stamp in the PST measured data assessment folder. A graph of the last partial stroke test is shown in the
Partial stroke test (PST) folder.
Test completed successfully
When a partial stroke test has been complet­ed successfully, the analyzed parameters are displayed separately for the increasing and decreasing characteristics.
Analysis of measured data (step response test):
'Overshooting' (relative to the step height) [%]
'Dead time' [s]
'T86' [s]
'Settling time' [s]
Analysis of measured data (ramp test):
'Overshooting' (relative to the step height) [%]
The results of the rst partial stroke test are
used as the reference measurement.
Note:
Changes in the diagnostic parame­ters listed below affect the test. The results of the next following partial stroke test is used as the new refer­ence measurement:
− 'Lower range value'
'Stop'
− 'Activate ramp function'
− 'Ramp time (increasing)'
− 'Ramp time (decreasing)'
'Waiting time after step change'
− Test not completed
If the test was not completed, the reason for cancelation is indicated in the corresponding reading by the 'Failure' message. The posi-
EB 8389-1S EN 79
Page 80
Partial stroke test (PST)
tioner generates a 'PST/FST' message with the selected status classication.
Diagnostics > Monitoring
− PST/FST status (Code 84):
Note:
The ‘No test available’ status remains
active until a partial stroke is complet­ed successfully.
, , ,
12.4 Resetting single status mes­sages
The diagnostic parameters and measured data analysis of the partial stroke test are re­set by selecting and executing the command
“Reset ‘PST parameters”.
The positioner saves the measured data analysis of the last three partial stroke tests. The analysis of the penultimate test is deleted when another test is performed.
Diagnostics > Service/maintenance > Reset
Reset 'PST parameters'
80 EB 8389-1S EN
Page 81
Table 4: Start options for the partial stroke test (PST)
Operating mode
Type of application
PST start
Rotary pushbut-
by
ton or operating
software
See page 73 See page 73 See section12.1 See section12.2
Manual Automatic Automatic Automatic
Test interval
(time-controlled) Set pointw Binary input
Partial stroke test (PST)
Automatic mode AUTO
Manual mode
Control valve
MAN
Automatic mode AUTO
Manual mode
On/off valve
MAN
Triggering event
Not possible Not possible Not possible
Possible Not possible Possible
Possible Possible Possible Possible
Possible Possible Not possible Possible
Code 49 - A0 =
YES or
'Start test'
parameter
activated
'Test interval'
elapsed
Set point is six
seconds long in
the range
between 25 and
50% of travel
Binary input
changes to ‘Switch
open’ or ‘Switch
closed’ (see ‘Binary
input control’
parameter
EB 8389-1S EN 81
Page 82
Partial stroke test (PST)
x [%]
t [s]
x [%]
t [s]
Full stroke test (FST)
100
START-UP
0
Solenoid valve
triggered/forced
venting active;
Control loop error; Set point start difference
too high; Current
too low; Supply
pressure too low
Max. test duration
exceeded
â â â â â
Manual mode (TEST)
Prerequisite
– Operating mode: Manual
â
Tolerance limit, Settling time before starting test,
Waiting time after step change, Sampling time
ivate ramp function
Act
â
â â
1)
No
Yes
â
Ramp time (increasing)
Ramp time (decreasing)
â
elation condition
Canc
Max. test duration Max. breakaway time
Status classication
PST/FST
â
Start FST
The valve has not moved before reaching the max. breakaway time.
Analysis (NE 107)/logged by operating hours counter
PST/FST
â
Reset 'FST parameters'
100
0
Perm. time until
CLOSED position
The valve has not reached the
'Stop'
position before
'Perm. time
until CLOSED
position' has
elapsed.
1)
Recommended
82 EB 8389-1S EN
Page 83
Full stroke test (FST)
13 Full stroke test (FST)
The dynamic valve performance can be eval­uated by performing this test.
The full stroke test is started in the manual mode.
The following listed parameters are activated while the full stroke test is being performed:
− Characteristic selection (Code 20): Linear
Required transit time OPEN (Code 21): Variable
− Required transit time CLOSED (Code 22): Variable
During the full stroke test, the valve moves through its entire working range.
The rst step ends in the fail-safe position. As
a result, the second step starts from the fail­safe position.
The change in travel can be performed either in steps or in a ramp function (Fig. 20). For the test in a ramp function, additionally the times for the increasing and decreasing
ramps need to be dened.
After being activated, the test does not start
until the 'Settling time before starting test'
has elapsed. This ensures that the valve has reached the start position.
Starting from the start position, the valve moves to the fail-safe position. The valve re-
mains in this position for the time dened by the 'Waiting time after step change' before
performing a second step change in the op­posite direction from the fail-safe position to
the start position of the rst step. After the 'Waiting time after step change' has
elapsed, the valve moves back to its operat­ing point.
Fig. 19: Diagnostics > Monitoring/tests > Dynamic tests > Full stroke test (FST)
EB 8389-1S EN 83
Page 84
Full stroke test (FST)
Operating point +
Operating point –
t [s]
x [%]
The 'Tolerance limit' parameter denes the
permitted valve positions for the start value and the end value for the step.
The 'Sampling time' denes the time interval
between which the measured values are re­corded during the test.
Test cancelation conditions
Various cancelation conditions provide addi­tional protection against the valve slamming shut or moving past the end position. The positioner cancels the full stroke test when one of the following cancelation conditions is
fullled:
Tolerance limit
Tolerance limit
100
Operating point
'Max. test duration': The test is canceled when the maximum permissible test du­ration is reached.
'Max. breakaway time': The test is can-
celed when the time entered in 'Settling time before starting test' has elapsed and
the valve has still not changed its posi­tion. This cancelation condition only be­comes effective after it has been activat-
ed ("Activate 'Max. breakaway time" =
[Yes]).
t
t
decreasing
increasing
Settling time before
starting test
‘Lower range value’ – ‘Stop’
100
‘Step start’ – ‘Step end’
100
Waiting time after
step change
Waiting time after
step change
x ‘Ramp time (decreasing)’
x ‘Ramp time (increasing)’
Settling time before
starting test
Waiting time after
t
decreasing
step change
t
increasing
Fig. 20: Course of the full stroke test with step response (left) and ramp function (right), fail-close
84 EB 8389-1S EN
Waiting time after
step change
Page 85
Full stroke test (FST)
'Perm. time until CLOSED position': The test is canceled when the valve does not reach the CLOSED position before the adjusted time has elapsed. This cancel­ation condition only becomes effective after it has been activated ("Activate 'Perm. time until step end" = [Yes]).
Additionally, the full stroke test is canceled when one of the following events arises:
− Cancel internal solenoid valve/forced venting: The test was canceled by the ac­tivation of the solenoid valve/forced venting function.
− Canceled by control loop error: A control loop error has occurred.
− Set point start difference too high: During
the step change, 'Lower range value' is outside the range (operating point ± 'Tol­erance limit').
− Current too low
− Supply pressure too low
Note:
The 'Measured data storage out of memory' reading (Failure) is generat-
ed when the 'Sampling time' is too
low. After recording 100 measured values per variable, logging is stopped, but the test continues until it is completed.
After the full stroke test is canceled, the 'FST status' reading indicates 'Not successful'. The
reason for cancelation is marked by the
'Failure' message in the Measured data as­sessment folder (> Current test).
Dening parameters
1. Switch to manual mode (Code0 =MAN).
2. Dene parameters for full stroke test. See also u Note concerning setting the FST diagnostic parameters.
3. Dene parameters for cancelation condi­tions.
4. Select classication for status messages.
5. Start full stroke test.
The 'Test information' status indicates 'Test active'. 'D6' and 'TEST' are indicat-
ed in alternating sequence on the posi-
tioner display. 'Function check'
is ac-
tivated as the condensed state.
Diagnostics > Service/maintenance > Operating mode
− Enter operating mode (Code 0): Manual
2.Diagnostics > Monitoring/tests > Dynamic
tests > Full stroke test (FST)
− Tolerance limit: 0.1 to 10.0 %, [2.0 %]
Activate ramp function: [Yes], No
− Settling time before starting test:
[1] to 240s
− Waiting time after step change:
[2.0] to 25.0s
Sampling time: 0.2 to 250.0 s, [1.4s]
Only when the ramp function is activated:
− Ramp time (increasing): 0 to 9999 s, [60 s]
− Ramp time (decreasing): 0 to 9999 s, [60 s]
3. − Max. test duration: 30 to 25000 s, [140 s]
Activate 'Max. breakaway time': Yes, [No]
− Max. breakaway time: 0.0 to 25000 s,
[30.0s]
EB 8389-1S EN 85
Page 86
Full stroke test (FST)
− Activate 'Perm. time until CLOSED position exceeded': Yes, [No]
− Perm. time until CLOSED position exceeded:
0.0 to 25000 s, [70.0 s]
Device settings > Alarm settings > Status classication > PST/FST
− Max. test duration exceeded:
], , ,
[
− Test canceled manually: [ ], , ,
− Measured data memory full: [ ], , ,
− Cancel internal solenoid valve/forced venting: [ ], , ,
− Canceled by control loop error: [ ], , ,
− Set point start difference too high: [ ], , ,
− Set point change: [ ], , ,
− Current too low: [ ], , ,
− Max. breakaway time exceeded: [ ], , ,
− Perm. time until CLOSED position exceeded: [ ], , ,
− Canceled by supply pressure: [ ], , ,
The 'Ramp time (decreasing)' must be greater than the corresponding value for 'Min. transit time OPEN' (Code 40) de­termined during initialization.
The 'Sampling time' must not be lower
than the indicated 'Recommended min. sampling time'. The 'Recommended min. sampling time' is calculated from the 'Ex­pected duration of test'.
4.Diagnostics > Monitoring/tests > Dynamic
tests > Full stroke test (FST)
Start test
Note concerning setting the FST diagnostic parameters
The 'Ramp time (increasing)' must be greater than the corresponding value for
'Min. transit time CLOSED' (Code 41)
determined during initialization.
86 EB 8389-1S EN
Page 87
Full stroke test (FST)
13.1 Analysis and monitoring
The analysis of the last three full stroke tests are saved with a time stamp in the FST mea-
sured data assessment folder.
Test completed successfully
When a full stroke test has been completed successfully, the analyzed parameters are displayed separately for the increasing and decreasing characteristics.
Analysis of measured data (step response test):
'Overshooting' (relative to the step height) [%]
'Dead time' [s]
'T86' [s]
'Settling time' [s]
Analysis of measured data (ramp test):
'Overshooting' (relative to the step height) [%]
The results of the rst full stroke test are used
as the reference measurement.
Note:
Changes in the diagnostic parame­ters listed below affect the test. The results of the next following full stroke test is used as the new reference measurement:
− 'Activate ramp function'
− 'Ramp time (increasing)'
− 'Ramp time (decreasing)'
'Waiting time after step change'
Test not completed
If the test was not completed, the reason for cancelation is indicated in the corresponding
reading by the 'Failure' message. The posi­tioner generates a 'PST/FST' message with the selected status classication.
Diagnostics > Monitoring
− PST/FST status (Code 84):
Note:
The 'No test available' status remains
active until a full stroke is completed successfully.
, , ,
13.2 Resetting single status mes­sages
The diagnostic parameters of the full stroke test are reset by selecting and executing the
command 'Reset FST parameters'. The mea­sured data analysis and the 'PST/FST status'
message cannot be reset. The positioner saves the measured data
analysis of the last three full stroke tests. The analysis of the penultimate test is deleted when another test is performed.
Diagnostics > Service/maintenance > Reset
Reset 'FST parameters'
EB 8389-1S EN 87
Page 88
SIL operator test
G
+81 -82 +11 -12
+83
-84 +51
-52
A2A3 A1
+41 -42
+31 -32
A
+31 -32
G
+31 -32
14 SIL operator test
The SIL operator ensures that the safety func­tion of the positioner is working. The safety function is based on the shutdown of the i/p converter (6, Fig. 22). This causes the pneu­matic actuator to be vented and the valve to move to its fail-safe position.
Monitoring of the input signal
The i/p converter is switched off when the input signal of the positioner at terminals
+11/ –12 falls below 3.8mA or 4.4mA depend-
ing on the positioner version (a signal range
of 4 to 20mA is required). See Fig. 21.
Optional Optional Optional Optional
Monitoring the voltage supply (version with forced venting and solenoid valve)
The i/p converter and the solenoid valve (when installed) are shut down whenever the
voltage at terminals +81/–82 falls below 12V (an input voltage of 24V DC is re-
quired). See Fig. 21. The SIL operator test checks the emergency
shutdown by the integrated safety function (SIL). It can be started in the manual or auto­matic mode of control valves.
The positioner generates a permanent mes­sage in the event that an error occurs during the operator test. All control properties of the positioner remain kept. Only the emergency shutdown function of the positioner is no lon­ger guaranteed.
A2A3 A1
G
G
+81 -82 +11 -12
24 V DC
Forced
venting/
solenoid valve
(optional)
mA
control
signal
-84 +51
+83
Switching amplier according to EN 60947-5-6
ault
A3 F
alarm
output
-52
Limit switches
A2
Softwar
e
+41 -42
A1
Software
optionally
ive
induct
+31 -32
A
Two-wire transmitter Supply unit
only for
optional
ansmitter
tr
+31 -32
Leakage
sensor
+31 -32
Binary input
Fig. 21: Electrical connections
88 EB 8389-1S EN
Page 89
BE
21
w
24 V DC
Serial
20
Interface
FSK
4
>12 V
1)
x
16
&
17 18
12 25
SIL operator test
SIL operator test
22
19
13
15
14
14
2
9
A3
A2
A1
11
GG
%
mm
5
6
mm
%
S
w
PD
x
3
10
24
Q
1
y
x
8
23
7
1 Control valve 14 Software limit switches A1/A2 2 Travel sensor 15 Störmeldeausgang A3 3 PD controller 16 LCD 4 A/D converter 17* Actuation of solenoid valve 5 Microcontroller 18* Electrical insulation 6 i/p converter 19 D/A converter 7 Air capacity booster 20 Communication interface
®
8 Pressure regulator 21 HART
connection
9 Flow regulator 22 Binary input BE*
10 Volume restriction 23 Pressure sensor for supply air ps
11* Inductive limit switch 24 Pressure sensor for signal pressure p
out
12* Solenoid valve 25* Forced venting 13* Analog position transmitter or
binary input * Options
Fig. 22: Functional diagram
EB 8389-1S EN 89
Page 90
SIL operator test
NOTICE
The SIL operator test causes the valve
to automatically move. It must only be started when the plant allows it.
Dening parameters
1. Set Type of application to Control valve.
2. Apply an external set point:
4.0 to 4.5 mA for 3.8 mA emergency shutdown
4.8 to 5.5 mA for 4.4 mA emergency shutdown
3. Start SIL operator test. S001 to S030 appear on the positioner display one after the other.
Start-up
1.
− Type of application (Code 49 - h0): Control
valve
3.Diagnostics > Service/maintenance > SIL
operator test
− Start SIL operator test
The emergency shutdown function of the po­sitioner is no longer guaranteed. The posi­tioner needs to be returned immediately to
SAMSON.
Diagnostics > Service/maintenance > SIL operator test
− SIL test:
,
14.1 Analysis and monitoring
SIL operator test completed successfully
OK appears on the positioner display.
SIL operator test not completed
Possible error code E001 to E030 appears on the positioner display. The positioner generates the permanent error code 86 (Fail-
ure - this status classication cannot be
changed).
90 EB 8389-1S EN
Page 91
Binary input
15 Binary input
TROVIS SAFE 3730-6xxxxx3x00x0x00
The positioner has an optional binary input to activate various functions:
− [Transmit switching state] The switching state of the binary input is logged.
− Activate local write protection
After the rst initialization, a local write
protection can be activated. While the binary input is active, no settings can be changed at the positioner. The positioner cannot be restarted. Enabling congura­tion over Code 3 is not active.
− Start PST The positioner starts a single partial stroke test. The test is performed using the settings in Code 49 - d2 to Code 49
- d9. See u Section 12.
− Move valve to safety set point An on/off valve moves to the predeter­mined safety set point when the position­er is in automatic mode. This function is not performed if the positioner is in the manual mode or fail-safe position mode.
Switch AUTO/MAN The positioner changes from the auto­matic mode to the manual mode or vice versa. This function is not performed if the positioner is in the fail-safe position mode.
− Start data logger Activation of the binary input causes the data logger to start. See u Section 3.
− Reset diagnostics Active monitoring and dynamic tests are stopped and the diagnostic data is reset once.
Note:
The optional binary input can only be
congured using the TROVIS-VIEW4
software and using the DD parame­ters. In the default setting, the switch state is logged with a closed switch.
Dening parameters
Note:
− The parameters can only be dened in TROVIS-VIEW4 after the 'Posi­tioner accessories identication' has
been set to 'Binary input'.
The 'Safety set point' can only be set
when the 'Type of application' is set to 'On/off valve'.
1. Select 'Binary input' in 'Positioner acces­sories identication'.
2. Congure the binary input.
Device settings > Positioner > Options
1.
Positioner accessories identication: Binary input
EB 8389-1S EN 91
Page 92
Binary input
2.Device settings > Positioner > Options >
Conguration, binary input
− Binary input conguration: [For oating contact (switch function)], For non-oating
contact (0-24 V)
− Select function: [Transmit switching state], Activate local write protection, Start PST, Move valve to safety set point, Switch
AUTO/MAN, Start data logger, Reset
diagnostics
Binary input control: Activate function: switch open, [Activate function: switch closed]
− Safety set point: 0.0 to 100.0 %, [50.0 %] (only when Type of application = On/off valve)
Binary input classication: [
], , ,
92 EB 8389-1S EN
Page 93
Dynamic HART® variables
16 Dynamic HART® variables
The HART® specication denes four dynam­ic variables consisting of a value and an en­gineering unit. These variables can be as­signed to device parameters as required. The universal HART namic variables out of the device. This al­lows manufacturer-specic parameters to al­so be transferred using a universal com­mand.
In the TROVIS SAFE3730-6 Positioner, the
dynamic variables can be assigned in the
Device settings folder (> Positioner > HART communication) as listed in uTable 5:
Device settings > Positioner > HART communication
− Primary variable assignment: [Set point], Direction of action set point, Set point after
transit time specication, Valve position, Set
point deviation e, Absolute total valve travel,
Binary input status, Internal solenoid valve/
forced venting status, Condensed state, Temperature, Leakage sensor sound level, Ambient pressure, Signal pressure p out, Supply pressure, Flow rate, Differential pressure, All active errors
− Secondary variable assignment: Set point, Direction of action set point, Set point after
transit time specication, [Valve position], Set
point deviation e, Absolute total valve travel,
Binary input status, Internal solenoid valve/
forced venting status, Condensed state, Temperature, Leakage sensor sound level, Ambient pressure, Signal pressure p out, Supply pressure, Flow rate, Differential pressure, All active errors
®
command 3 reads the dy-
− Tertiary variable assignment: Set point, Direction of action set point, Set point after
transit time specication, Valve position, [Set
point deviation e], Absolute total valve travel,
Binary input status, Internal solenoid valve/
forced venting status, Condensed state, Temperature, Leakage sensor sound level, Ambient pressure, Signal pressure p out, Supply pressure, Flow rate, Differential pressure, All active errors
− Quaternary variable assignment: Set point, Direction of action set point, Set point after
transit time specication, Valve position, Set
point deviation e, [Absolute total valve travel],
Binary input status, Internal solenoid valve/
forced venting status, Condensed state, Temperature, Leakage sensor sound level, Ambient pressure, Signal pressure p out, Supply pressure, Flow rate, Differential pressure, All active errors
EB 8389-1S EN 93
Page 94
Dynamic HART® variables
Table 5: Dynamic HART® variables assignment
Variable Meaning Unit
Set point Set point % Direction of action set point Direction of action set point % Set point after transit time
specication
Valve position Valve position % Set point deviation e Set point deviation e % Absolute total valve travel Absolute total valve travel
Binary input status 0 = Not active
Internal solenoid valve/forced venting status
Condensed state 0 = No message
Temperature Temperature °C Leakage sensor sound level Leakage sensor sound level dB Ambient pressure Ambient pressure mbar Signal pressure p out Signal pressure p Supply pressure Supply pressure bar Flow rate Flow rat m³/h Differential pressure Differential pressure bar All active errors 0 = No errors
Set point after transit time specication %
1 = Active 255 = –/–
0 = De-energized 1 = Energized
2 = Not installed
1 = Maintenance required 2 = Maintenance request 3 = Failure
4 = Out of specication
7 = Function check
out
1 = Control loop 2 = Zero 4 = w too low 8 = PST/FST status 16 = On/off error activated 32 = SIL test 64 = Set point outside range 128 = Total valve travel exceeded 256 = Operating mode not AUTO
bar
94 EB 8389-1S EN
Page 95
Appendix
Appendix
17 Appendix
17.1 Code list
Code
Parameter – Readings/
no.
values [default setting]
Note: Codes with marked with an asterisk (*) must be enabled with Code3 prior to conguration.
48* Diagnostic parameters
Description
d0 Temperature
Read only
d1 Min. temperature
Read only
d2 Max. temperature
Read only
d3 No. of zero calibra-
tions
Read only
d4 No. of initializations
Read only
d5* Zero limit
0.0 to 100.0%, [5.0%] of the nominal
range, ESC
d6 Condensed state
Read only
Current operating temperature [°C] inside the positioner (accuracy
±3%)
Lowest temperature [°C] recorded inside the positioner since start­ing the operating hours counter
Highest temperature [°C] recorded inside the positioner since start­ing the operating hours counter
The number of zero calibrations performed since last initialization
Number of initializations performed since last start with default
values
Valve position limit relating to nominal range
'Zero point' message (Code58) triggered depending on selected status classication if limit is exceeded.
Note: The 'Zero point' message (Code58) has the default status classication 'Maintenance required'. This classication can only be changed in the operator software (e.g. TROVIS-VIEW).
Summary of all status messages classied according to NAMUR Recommendation NE107 OK: No message
C: Maintenance required CR: Maintenance demanded
S: Out of specication B: Failure
I: Function check
EB 8389-1S EN 95
Page 96
Appendix
Appendix
Code
Parameter – Readings/
no.
values [default setting]
Description
Note: Codes with marked with an asterisk (*) must be enabled with Code3 prior to conguration.
48* d7 Supply pressure p
Current supply pressure [bar]
s
Read only
d8 Signal pressure p
out
Read only
d9 Flow rate
Read only
d10 Differential pressure
Read only
d11*
Direction of action (ac­tuator)
[-/-], SA, DA, ELSE, ESC
Diagnostic parameters h
h0* Initialization including
valve signature
No, [YES], ESC
h1, h2
Current signal pressure [bar]
Current valve ow rate
Note: – – – – appears on the display when the ow rate calcula- tion is not active or has failed.
Current differential pressure [bar]
Indicates the actuator's direction of action
SA: Single-acting DA: Double-acting ELSE: Other
Initialization with [YES] or without [No] plotting the valve signature
The valve signature involves the signal pressure p
being plotted
out
in relation to the valve position. Regardless of the adjusted settings, the valve signature is recorded
automatically when initializing a positioner that has not yet been
initialized (e.g. after initialization was reset (Code36 - Std and Code36 - DS)), after each further initialization if the settings for 'Initialization mode', 'Pin position', 'Direction of action', 'Pressure limit', 'Proportional-action coefcient Kp level' or 'Derivative-action time Tv level' were changed, and when the switch position (ATO/
ATC) was changed.
Note: The valve signature is required to perform diagnostic func­tions.
Unassigned
96 EB 8389-1S EN
Page 97
Appendix
Code
Parameter – Readings/
no.
values [default setting]
Note: Codes with marked with an asterisk (*) must be enabled with Code3 prior to conguration.
48* h3* Desired time until 'Re-
set measured diagnos­tic data'
[0] to 365 days, ESC
h4 Remaining time until
'Reset measured diag­nostic data'
Read only
49* Partial stroke test (PST)
A Partial stroke test (PST)
A0* Start test
[No], YES, ESC
A1 Time until next test
Read only
Description
Time interval for scheduled resetting of measured diagnostic data
Remaining time (time and unit of time displayed in alternating se­quence) until the next scheduled reset of the diagnosis data
Starts partial stroke test (test D4). The valve moves through the test range (Lower range value
(Code49 - d2) to Upper range value (Code49 - d3)) and back
again in a ramp or in steps. Recorded variables: time, set point, valve position, set point devia-
tion, control signal.
Remaining time (time and unit of time displayed in alternating se­quence) until the the next time-controlled test starts
A2* Test start
AUTO, [MAN], ESC
The write protection is active in the AUTO set­ting (local operation and operation over software locked).
A3* Test interval Time interval (time and unit of time displayed in alternating se-
A4 Unassigned
A5 Recommended min.
sampling time
Read only
Activates (AUTO) or deactivates (MAN) the time-controlled partial
stroke test.
quence) between time-controlled tests
Sampling time [s] used to record entire step response in the graph of the partial stroke test
Page 98
Appendix
Appendix
Code
Parameter – Readings/
no.
values [default setting]
Note: Codes with marked with an asterisk (*) must be enabled with Code3 prior to conguration.
49* A6 Unassigned
A7 Δp_out reference val-
ue
Read only
A8* Activate Δp_out moni-
toring
[NO], YES, ESC
A9* Δp_out monitoring
value
0.00 to 7.00bar, [1.00bar,] ESC
d Step parameters for the partial stroke test (PST)
d0, d1
d2* Lower range value
0.0 to [100.0]%, ESC
d3* Upper range value
0.0 to 100.0%, [90.0%], ESC
d4* Activation of ramp
function
No, [YES]
d5* Ramp time (decreas-
ing)
0 to 9999s, (600s],
ESC
Description
The valve moves to the lower range value (Code49 - d2) and up­per range value (Code49 - d3) with a certain signal pressure. These signal pressures are used to form the Δp_out reference value
[bar].
Note: The reference value only applies to the adjusted step and ramp values.
Activates Δp_out monitoring (YES) or deactivates it (No)
Test canceled if signal pressure change exceeds or falls below the
reference value. The reference value is made up of Δp_out refer­ence value Δp_out reference value (Code49 - A7) and the Δp_out
monitoring value.
Unassigned
Start value of the test range
Note: To perform a partial stroke test, the 'Lower range value' must be near the current operating point ± 'Tolerance limit'.
The Tolerance limit is 2.0% by default. It can be changed in the operator software, e.g. TROVIS-VIEW.
End value of the test range
YES: The valve is moved through the test range within the ad-
No:
The time required by the valve to move through the range from
CLOSED to OPEN position. The time to move from Lower range value (Code49 - d2) to Upper
range value (Code49 - d3) is calculated:
|Upper range value – Lower range value| / 100 x Ramp time (in­creasing)
justed ramp time. The valve is moved in steps through the test range (step
response).
98 EB 8389-1S EN
Page 99
Appendix
Appendix
Code
Parameter – Readings/
no.
values [default setting]
Note: Codes with marked with an asterisk (*) must be enabled with Code3 prior to conguration.
49* d6* Ramp time (increas-
ing)
0 to 9999s, [60s],
ESC
d7* Settling time before
starting test
1 to 240s, [2s], ESC
Description
The time required by the valve to move through the range from
OPEN to CLOSED position. The time to move from Upper range value (Code49 - d3) to Lower
range value (Code49 - d2) is calculated:
|Lower range value – Upper range value| / 100 x Ramp time (decreasing)
Settling time between reaching the upper range value (Code49 -
d3) and the valve moving through the test range in the reverse di­rection
d8* Waiting time after
step change
1.0 to 240.0s, [2.0s],
ESC
d9* Sampling time
0.2 to 250.0 s, [0.8s],
ESC
E Cancelation conditions of the partial stroke test (PST)
E0* Activate x monitoring
No, [YES]
E1* x monitoring value
–10.0 to 110.0%, [85.0%] of total travel,
ESC
E2, E3, E4 Unassigned
E5* Activate PST tolerance
band monitoring
[No], YES
E6* PST tolerance band
0.1 to 100.0%, [5.0%], ESC
Waiting time between step change from Lower range value (Code49 - d2) to Upper range value (Code49 - d3) and vice ver­sa
Interval for measured data recording
Activates x monitoring (YES) or deactivates it (No)
The test is canceled when the valve position – falls below the adjusted value (step end < step start) – exceeds the adjusted value (step end > step start)
Activates PST tolerance band monitoring (YES) or deactivates it
(No)
The test is canceled when the valve position's deviation (relating to
step end) exceeds the adjusted value.
EB 8389-1S EN 99
Page 100
Appendix
Appendix
Code
Parameter – Readings/
no.
values [default setting]
Note: Codes with marked with an asterisk (*) must be enabled with Code3 prior to conguration.
49* E7* Max. test duration
30 to 25000s, [90 s],
ESC
E8
E9* Reset 'PST parame-
ters'
F Partial stroke test (PST) information · Read only
F0 No test available
F1
F2 x cancelation The test was canceled.
F3 Δp_out cancelation The test was canceled.
F4 Tolerance band ex-
ceeded
F5 Max. test duration ex-
ceeded
F6 Test canceled manual-
ly
F7 Measured data mem-
ory full
F8 Cancel internal sole-
noid valve/forced venting
Description
Test canceled when max. test duration is reached.
Unassigned
Resets partial stroke test parameters.
Unassigned
The valve position fell below the x monitoring value (Code49 -
E1).
The signal pressure change Δp_out exceeded or fell below the ref­erence value.
The valve position deviation exceeds the PST tolerance band
(Code49 - E6).
The test was canceled.
The Max. test duration (Code49 - E7) was reached.
The Sampling time (Code49 - d9) is too low. After recording 100
measured values per variable, logging is stopped, but the test con­tinues until it is completed.
The test was canceled. The internal solenoid valve has been energized/the forced venting
has been activated.
100 EB 8389-1S EN
Loading...