ROTRONIC CO2 Calibrator User Manual

Setup and adjustment CO2 transmitter
Required material:
CO
ROLINE USB to Serial Converter (Article No. 12.02.1089
Software UIP5 from SenseAir, link
CO2 CALIBRATOR (for CO
Transmitter (CF5/8-Series)
2
calibration only)
2
)
Installation of USB to Serial Converter Cable:
Insert installation disk and follow the instructions on the screen.
IMPORTANT: The cable must be configured from COM-Port 1 to 8 (Device Manager)
Hardware Setup:
CO
Connect CF5/8 Comm Cable with your computer (USB) and with the CO
Transmitter must be powered
2
Transmitter
2
Software (UIP5) setup
Start UIP5 , go to menu Help – Check for updates...
Software Manuals: go to Menu Help – Contents
Go to menu Meter – Connection configuration
Tick SA-Bus and select the right COM-Port and Save settings
Go to menu Meter – Connect to any
Check connection status on the bottom right (OK)
Device setup
Outputs
The most important settings are Outputs (see outputs tab)
Select block No. 3 (PRegulator) to adjust the CO2 scaling. (here 0...2000 ppm)
CO2 (calibration)
Select the CO2 tab to calibrate the sensor
Connect the CO2 calibrator tube with the sensor (see screenshot) and turn on the device (green light). Wait until to CO2 value is more or less stable.
Enter 25(ppm) as Target and push “Calibrate” in the Software. Device is now calibrated.
Loading...