RTH100 Series Temperature & Humidity Recorder
Instruction Manual
Page 3
3
Alarm/Control Relays.
In recorder-controller versions of the RTH100 Series two single-pole changeover delayed-action relays are fitted for each pen.
These relays may be used either to activate an alarm should the recorder's pens move above or below the range of normal indication, or to connect the instrument to a closed-loop temperature or humidity control system.
For each pen, one relay is associated with the adjustable low contact of the mechanism (item 8 or 1 1 in figure 6) and a second relay
with the high contact (item 9 or 10). The status of the mechanism contacts is indicated by the light-emitting diodes (LEDs) mounted in the door of the instrument. The relays are designed to be failsafe and are energised when the actual value pen lies within the
contact limits of the desired value pointer (This is the 'normal' or 'non-alarm' condition; the mechanism contacts will be open and the
LEDs extinguished). It should be noted that the normally-open (NO) and normally-closed (NC) labelling, carried on the relay cover
plate and reproduced in figure 3, refers to the relay contact connections with power applied to the instrument and the pens within
their mechanism contact limits.
Each alarm/control relay is rated as follows:Contact arrangement : Single-pole changeover
Switched voltage max. : 30 Vdc / 277 Vac
Switched current max. : 10 A
Switched load max. : 150 W /1660 VA
When switching inductive loads it is recommended that an R-C protection network be wired in parallel with the relay contacts to
reduce arcing.
Audible Alarm.
Closure of the high or low mechanism contact on either pen will cause an audible alarm to sound. This alarm may be silenced in
one of two ways according to the setting of the mode switch which is located on the relay printed circuit board beneath the chart
plate.
Mode 1 : Permanent Silence.
In Mode 1 the alarm will be permanently silenced when the light-emitting diode (LED) built into the mute switch is illuminated.
Pressing the mute push-button will turn the LED on or off. When not silenced, the audible alarm will sound for as long as the pen
remains outside the contact limits.
Mode 2 : Temporary Silence.
In Mode 2 an audible alarm caused by either pen moving outside the contact limits may be silenced temporarily by pressing the
mute push-button. The mute switch LED will be illuminated to acknowledge that the alarm has been silenced. The alarm will only
be silenced while the pen remains outside contact limits: the temporary silence will be cancelled automatically when the pen returns
to within limits and the alarm will sound again should the pen subsequently move outside limits once more.
If an alarm caused by one of the two pens is temporarily silenced, the second pen will still cause the alarm to sound if it too also
moves outside its contact limits. Pressing the mute push button once more will silence the second alarm.
Installation
Ideally the site chosen should be free from dust, corrosive fumes, vibration and extremes of temperature.
Mounting the Controller
The controller may be surface or flush panel mounted.
For surface mounting, remove the screws holding the three brackets at the rear of the case, reverse and refit brackets with there
ends projecting from the edges of the case. The bracket with the keyhole slot must remain at the top of the case.
Drill three holes in the panel (or wall), of the size and at the centres given in Fig. 2 below.
Fit a screw of maximum diameter 6.35 mm (1/4") in the top hole. Locate the instrument on this screw and fit two similar screws
through the other two brackets to secure.
For flush panel mounting, make a cut out to the dimensions given in Fig. 2. Drill three holes and slot two to the cut-out edge as
shown.
Remove the nuts and washers from the studs projecting from behind the case flange, pass the instrument through the cut-out
and refit the washers and nuts to secure the instrument.