5
Access to Recorder
To open the door, unlock, and turn the latch anticlockwise. As the door is opened the pen lifter raises the pen (or
pens) away from the chart. Remove the chart (see page 6).
If the instrument is connected to the mains, switch off the mains supply before dismantling further.
For access to the terminal blocks remove the lower plastic plate from behind the chart by undoing the one
retaining screw.
For access to the relays and measuring systems, first remove the chart. Unhook the pens from their mounting
(see Fig. 3.). Remove the three screws from the outside edge of the chart plate. (The pen lifter is secured by one
of these screws and must also be removed). Lower the chart plate carefully. The chart motor is mounted on the
back of the chart plate and if the motor is an electric one it remains connected to the terminal block in the lower
part of the case.
When replacing the chart plate reverse the above procedure. The coil on the pen lifter should be against the side
of the case when the retaining screw is tightened.
Process Connections
To avoid confusion on instruments with more than one measuring system, setting pointers are labelled and
Bourdon tube capillary connections to the instrument are painted with the same colours as the inks used for the
corresponding pens.
Process connections – temperature measuring instruments:
Install the sensing element (bulb) of the instrument in the apparatus where the temperature is to be measured.
The circulation around this element should be good and it should not be too close to any heating or cooling coil or
other controlling medium. Fasten the capillary tubing at frequent intervals to rigid supports avoiding sharp bends
of less than 50mm. radius anywhere along its length. The tubing must not touch or run close to any hot apparatus
and should take the path least subject to temperature variation. If the temperature source is likely to move or
vibrate allow one or two coils of 100mm. diameter of the capillary tube to eliminate stress of the tubing.
If the atmosphere around the capillary tube is likely to be corrosive, paint the tube with anti-corrosive paint and
ensure this is maintained.
Process connections – pressure measuring instruments:
Connect the Bourdon tube sensing element to the measuring point on the apparatus using pressure tubing of the
same or similar material as the Bourdon tube and of sufficient pressure rating. Slope the run of the tubing so that
any condensate is drained away from the instrument and does not affect the accuracy of measurement. For
measurement of steam pressure or other hot vapours, form a condensate trap below the instrument.
Process connections – humidity measuring instruments (wet and dry bulb):
Install the sensing elements (bulbs) where the humidity is to be measured as described for temperature
measuring instruments. The wet bulb (red pen system) is kept moist by a fabric covering, forming a wick which
dips into a constant level water bath. Adistilled water supply tank should be connected to the inlet feed pipe with
a head of between 1 and 6 metres.
It is important to install the bulbs in a position where the air speed is 3.6 metres per second or more. Below this
speed inaccuracies of bulb depression may occur. The air should pass over the bulbs in such a direction that the
water evaporating from the wet bulb does not pass over the dry bulb. The water must be kept clean and free from
impurities. The temperature of the bath and all piping must not fall below 0
o
C. See also page 13.
Electrical Connections
Before making any connections to the instrument switch off the mains supply to be connected.
All electrical connections are made to terminal blocks in the lower part of the case behind the lower section of the
chart plate.
Insert the leads through the cable gland in the underside of the case (see Fig. 2.) and make the connections as
shown on the terminal labels or Wiring Diagram (page 14). The live line should be switched and fused with a 2
amp fuse. If the instrument is mounted on an insulated panel the case should be earthed.