Oxygen Analyzer Direct
Replacement Probe for
use with Existing Signal
Conditioning Electronics
http://www.processanalytic.com
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and
international standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them to ensure they continue to operate within their
normal specifications. The following instructions MUST be adhered to and integrated into your
safety program when installing, using, and maintaining Rosemount Analytical products. Failure to
follow the proper instructions may cause any one of the following situations to occur: Loss of life;
personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Rosemount Analytical repre-
sentative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and mainte-
nance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate In-
struction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program,
and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts
specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY.
Look-alike substitutions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except
when maintenance is being performed by qualified persons, to prevent electrical shock
and personal injury.
The information contained in this document is subject to change without notice.
Emerson Process Management
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
HIGHLIGHTS OF CHANGES
Effective June, 1996 Rev. 3
PageSummary
---General. Updated text and illustrations to reflect probe design
changes.
Page 1-15Added “Extended temperature by-pass arrangements” to Figure 1-15,
sheet 3 of 3.
Effective December, 1996 Rev. 3.1
PageSummary
Page P-2Added "Safety instructions for the wiring and installation of this
apparatus".
Page 1-4Corrected Figure 1-3 item listing.
Page 2-1Added one WARNING to read new safety instructions and another
WARNING regarding protective covers and grounds.
Page 3-1Added WARNING regarding protective covers and grounds.
Page 4-1Added WARNING regarding protective covers and grounds.
Page 4-6Added title block to Figure 4-6.
Page 5-1Added WARNING regarding protective covers and grounds.
Page 7-1Corrected cell replacement kit part numbers in Table 7-1.
Page 7-2Corrected cell replacement kit part number in Table 7-1.
Effective May, 1997 Rev. 3.2
PageSummary
Page P-2Added safety sheet.
Effective February, 1998 Rev. 3.3
PageSummary
Page 2-2Figure 2-1. Changed calibration gas tube dimensions.
The World Class 3000 Direct Replacement Probe can be used with any of the following
model electronics: 218, 225, 218A, ZA8C, and TC200. Read the appropriate setpoint adjustment in this manual to prevent damage to the probe.
The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of this particular probe.
Some sections may describe equipment not used in your configuration. The user should
become thoroughly familiar with the operation of this module before operating it. Read
this instruction manual completely.
The following definitions apply to WARNINGS, CAUTIONS, and NOTES found throughout this
publication.
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
CAUTION
!
PREFACE
DEFINITIONS
Highlights an operation or maintenance
procedure, practice, condition, statement, etc. If not strictly observed, could
result in injury, death, or long-term
health hazards of personnel.
NOTE
Highlights an essential operating procedure,
condition, or statement.
: EARTH (GROUND) TERMINAL
: PROTECTIVE CONDUCTOR TERMINAL
: RISK OF ELECTRICAL SHOCK
: WARNING: REFER TO INSTRUCTION BULLETIN
Highlights an operation or maintenance
procedure, practice, condition, statement, etc. If not strictly observed, could
result in damage to or destruction of
equipment, or loss of effectiveness.
NOTE TO USERS
The number in the lower right corner of each illustration in this publication is a manual illustration number. It is not a part number, and is not related to the illustration in any technical
manner.
Rosemount Analytical Inc. A Division of Emerson Process ManagementP-1
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
FOR THE WIRING AND INSTALLATION
The following safety instructions apply specifically to all EU member states. They should
be strictly adhered to in order to assure compliance with the Low Voltage Directive. NonEU states should also comply with the following unless superseded by local or National
Standards.
1. Adequate earth connections should be made to all earthing points, internal and external,
where provided.
2. After installation or troubleshooting, all safety covers and safety grounds must be replaced.
The integrity of all earth terminals must be maintained at all times.
3. Mains supply cords should comply with the requirements of IEC227 or IEC245.
World Class 3000
IMPORTANT
SAFETY INSTRUCTIONS
OF THIS APPARATUS
4. All wiring shall be suitable for use in an ambient temperature of greater than 75°C.
5. All cable glands used should be of such internal dimensions as to provide adequate cable
anchorage.
6. To ensure safe operation of this equipment, connection to the mains supply should only be
made through a circuit breaker which will disconnect all circuits carrying conductors during a
fault situation. The circuit breaker may also include a mechanically operated isolating switch.
If not, then another means of disconnecting the equipment from the supply must be provided
and clearly marked as such. Circuit breakers or switches must comply with a recognized
standard such as IEC947. All wiring must conform with any local standards.
7. Where equipment or covers are marked with the symbol to the right, hazardous voltages are likely to be present beneath. These covers should only be
removed when power is removed from the equipment — and then only by
trained service personnel.
8. Where equipment or covers are marked with the symbol to the right, there is a
danger from hot surfaces beneath. These covers should only be removed by
trained service personnel when power is removed from the equipment. Certain surfaces may remain hot to the touch.
9. Where equipment or covers are marked with the symbol to the right, refer to
the Operator Manual for instructions.
10. All graphical symbols used in this product are from one or more of the following standards: EN61010-1, IEC417, and ISO3864.
P-2Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
DESCRIPTION AND SPECIFICATIONS
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
SECTION 1
1-1 COMPONENT CHECKLIST
The Direct Replacement (DR) Probe package of
the World Class 3000 Oxygen Analyzer typically
includes the probe, the instruction bulletin, and
the mounting hardware. These three items are
shown in Figure 1-1. Record the Part Number,
Serial Number, and Order Number of your
system on the first page of this manual.
1-2 SYSTEM OVERVIEW
a. Scope
This Instruction Bulletin has been designed
to supply details needed to install, start up,
and troubleshoot the Rosemount World
Class 3000 Oxygen Analyzer. The Direct
Replacement Probe can be interfaced to a
number of different and earlier model
electronics packages. These electronic
packages are not covered in this manual.
For specific information concerning calibration and operation of the system, refer to
the Instruction Bulletin applicable to your
electronics package.
b. System Function
The Rosemount Oxygen Analyzer (Probe) is
designed to measure the net concentration
of oxygen in an industrial process; i.e., the
oxygen remaining after all fuels have been
oxidized. The probe is permanently positioned within an exhaust duct or stack. The
sampling system is in situ rather than
extractive.
2
1. Oxygen Analyzer Probe
2. Mounting Adapter Plate with
Mounting Hardware and
Gasket
3. Instruction Bulletin
3
1
19290001
Figure 1-1. Typical System Package
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
The equipment measures oxygen percentage by reading the voltage developed
across a heated electrochemical cell, which
consists of a small yttria-stabilized, zirconia
disc. Both sides of the disc are coated with
porous metal electrodes. When operated at
the proper temperature, the millivolt output
voltage of the sensor cell is given by the
following Nernst equation:
EMF = KT log
10(P1/P2
) + C
Where:
1. P
is the partial pressure of the oxygen
2
in the measured gas on one side of the
cell,
2. P
is the partial pressure of the oxygen
1
in the reference gas on the other side,
3. T is the absolute temperature,
4. C is the cell constant,
5. K is an arithmetic constant.
NOTE
For best results, use clean, dry,
instrument air (20.95% oxygen) as a
reference gas.
vapor. Therefore, it may be considered an
analysis on a "wet" basis. In comparison
with older methods, such as the Orsat
apparatus, which provides an analysis on a
"dry" gas basis, the "wet" analysis will, in
general, indicate a lower percentage of
oxygen. The difference will be proportional
to the water content of the sampled gas
stream.
c. System Configuration
The DR probe is similar to the Exchange
Probe (NE model) except that the DR uses
a 115 V heater like most older systems. The
DR does not require a separate power supply as does the NE. The DR probe (like the
NE) is available in five lengths. The length
of stack penetration depends on the size of
the stack or duct. The length options are
18 in., 3 ft, 6 ft, 9 ft, and 12 ft (457 mm, 0.9
m, 1.8 m, 2.7 m, and 3.7 m).
d. Features
1. Output voltage and sensitivity increase
as the oxygen concentration
decreases.
When the cell is at operating temperature,
oxygen ions will travel through the cell from
high to low oxygen partial pressures. The
resulting output voltage is about 50 mV per
decade. The voltage varies with the ratio of
oxygen partial pressures of the reference
gas over the sample gas. Thus, the output
signal increases as the sample oxygen level
decreases. This property gives the analyzer
high sensitivity at low oxygen levels.
Oxygen analyzer equipment measures net
oxygen concentration in the presence of all
the products of combustion, including water
2. In situ, non-sampling analyzer.
3. Field replaceable cell.
4. Analyzer constructed of rugged
316 LSS for all wetted parts.
5. Suitable for use in temperatures up to
1300°F (700°C).
6. The DR probe uses a 115 V heater
and does not require a separate heater
power supply.
1-2 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
e. Handling the Oxygen Analyzer
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
GASES
It is important that printed circuit
boards and integrated circuits are
handled only when adequate antistatic
precautions have been taken to prevent possible equipment damage.
The oxygen analyzer is designed for
industrial application. Treat each
component of the system with care to
avoid physical damage. The probe
contains components made from ceramics, which are susceptible to
shock when mishandled.
NOTE
Retain packaging in which the oxygen
analyzer arrived from the factory in
case any components are to be
shipped to another site. This packaging has been designed to protect the
product.
f. System Considerations
Prior to installation of your Rosemount
World Class 3000 Oxygen Analyzer, make
sure you have all of the components
necessary to make the system installation.
Ensure all components are properly
integrated to make the system functional.
STACK
CALIBRATION
INSTRUMENT
AIR SUPPLY
(REF. GAS)
GAS
PRESSURE
REGULATOR
FLOWMETER
EXISTING
ELECTRONICS
Figure 1-2. Typical System Installation
connecting test gas tanks to the oxygen
analyzer when the probe is to be calibrated.
NOTE
Ambient air is not recommended for
use as high test gas. An 8% O
balance in nitrogen is recommended
for high test gas.
DUCT
ADAPTER
OXYGEN
ANALYZER
(PROBE)
2
PLATE
VOLTAGE
19290002
LINE
Once you have verified you have all the
components, select mounting locations and
determine how each component will be
placed in terms of available power supply,
ambient temperatures, environmental considerations, convenience, and serviceability.
A typical installation is illustrated in Figure
1-2.
If test gas bottles will be hooked up permanently, an optional check valve is recommended next to the calibration fittings on
the probe head. This is to prevent breathing
of calibration gas line and subsequent flue
gas condensation and corrosion. The check
valve is in addition to the stop valve in the
test gas kit or the solenoid valve in the
multiprobe test gas sequencer units.
After selecting the probe mounting location,
provision should be made for a platform
1-3 PHYSICAL DESCRIPTION
where the probe can be easily serviced.
The Oxygen Analyzer Probe, Figure 1-3 conA source of instrument air is required at the
probe for reference gas use. Since the
sists of three component groups: probe exterior,
inner probe, and junction box, Figure 1-4.
probe is equipped with an in place calibration feature, provision should be made for
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
NOTE: NOT ALL PARTS SHOWN ARE AVAILABLE FOR
PURCHASE SEPARATELY. FOR LIST OF
AVAILABLE PARTS, SEE TABLE 7-1.
World Class 3000
1
26
20
18
23
21
19
16
14
20
12
29
10
17
11
13
15
9
28
3
2
4
30
27
25
24
11
22
10
5
6
NOTE: ITEM, CALIBRATION GAS TUBE,
FITS INTO HOLESWHEN PROBE IS
ASSEMBLED.
1. Heater, Strut, and Backplate11. Probe Junction Box Cover22. Gas Connection
Assembly12. Cover Gasket23. Seal Cap
1-4 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
PROBE
EXTERIOR
(SENSING CELL INSTALLED)
PROBE
INTERIOR
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
PROBE
JUNCTION
BOX
Figure 1-4. Main Probe Components
1-4 PROBE EXTERIOR
Primary probe exterior components include a
flange-mounted zirconium oxide cell, mounted
on a tube assembly and protected by a snubber
diffusion assembly.
a. Cell and Flange Assembly
The primary component in the cell and
flange assembly, Figure 1-5, is a
yttria-stabilized zirconium oxide cell. The
cell creates an electrical signal when the
oxygen level on one side is out of balance
with the oxygen level on the other side. This
signal is proportional to the difference in
oxygen levels.
b. Probe Tube Assembly
21240006
The tube assembly includes a flange which
mates with a stack-mounted flange (shown
attached to the probe flange in Figure 1-4).
Studs on the stack flange make installation
easy. There is also a tube to carry calibration gas from the junction box to the
process side of the cell during calibration.
PROBE TUBE
CORRUGATED
SEAL
CELL AND
FLANGE
ASSEMBLY
Four screws secure the cell and flange assembly, Figure 1-5, to the probe tube assembly. When in place, the cell is inside the
tube.
21240007
Figure 1-5. Cell and Tube Assemblies
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
Table 1-1. Specifications for Oxygen Analyzing Equipment
1 2
Probe lengths, nominal ...................................................................... 18 in. (457 mm), 3 ft (0.91 m),
6 ft (1.83 m), 9 ft (2.74 m), or 12 ft (3.66 m),
depending on duct dimensions
Temperature limits in process
measurement area ....................................................................... 50° to 1300°F (10° to 704°C)
Standard/current output ................................................................ 4-20 mA dc signal (factory set)
O
indication (Digital display
2
and analog output)........................................................................ 0.1% O
or ±3% of reading, whichever is greater
2
using Rosemount test gases
System speed of response ................................................................ less than 3 seconds (amplifier output)
Resolution sensitivity ......................................................................... 0.01% O
Calibration gas mixtures .................................................................... Rosemount Hagan Test Gas Kit Part No.
6296A27G01 contains 0.4% O
and 8% O
Nominal
2N2
Nominal
2N2
Calibration gas flow ........................................................................... 5 scfh (141.6 L/hr)
Power supply ..................................................................................... 115 ±10% Vac at 50/60 Hz
Power requirement ............................................................................ 200 VA
Ambient operating temperature (Junction Box)................................. 300°F (149°C) [160°F (71°C) max for YEW
replacement]
Approximate shipping weights:
18 in. (457 mm) package ....................................................... 55 lbs (24.97 kg)
3 ft (0.91 m) package ............................................................. 60 lbs (27.24 kg)
6 ft (1.83 m) package ............................................................. 65 lbs (29.51 kg)
9 ft (2.74 m) package ............................................................. 72 lbs (32.66 kg)
12 ft (3.66 m) package ........................................................... 78 lbs (35.38 kg)
1
All static performance characteristics are with operating variables constant.
2
Equipment ordered utilizing this document as reference will be supplied to the USA standard design.
Customers requiring the EEC standard design should request the EEC documentation and utilize its ordering
data. Temperatures over 1000°F (537°C) may affect the ease of field cell replaceability.
c. Snubber Diffusion Assembly
The snubber diffusion assembly protects
the cell from heavy particles and isolates
the cell from changes in temperature. The
snubber assembly threads onto the cell and
flange assembly. Pin spanner wrenches
An optional ceramic diffusor element and
vee deflector shown in Figure 1-6, is available. The ceramic diffusor assembly is also
available in a flame arresting version to
keep heat from the cell from igniting flue
gases.
(probe disassembly kit 3535B42G01) are
applied to holes in the diffusion element hub
to remove or install the snubber assembly.
1-6 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
is noted in the calibration data sheet supplied with each probe.
DIFFUSION
ELEMENT
HUB
PIN
WRENCH
VEE
DEFLECTOR
21240024
Figure 1-6. Optional Ceramic Diffusor and Vee
Deflector Assembly
Systems that use an abrasive shield require
a special diffusor assembly with a hub that
is grooved to accept two dust seal gaskets.
This special diffusor is available in both
snubber and ceramic versions. See Probe
Options, paragraph 1-8.
d. Cell – General
The components which make up the cell are
machined to close tolerances and assembled with care to provide accurate oxygen
measurements. Any replacement requires
attention to detail and care in assembly to
provide good results.
Every probe should be calibrated and
checked after repair or replacement of cell,
pad and wire, heater, and thermocouple, or
after disassembly of the probe.
There are two styles of the heater/strut
inner probe assembly. P/Ns
3D39441G06 through G10 identify the
44 V heater design. P/Ns 3D39441G01
through G05 and G11 through G15
identify the 115 V heater design. Both
styles look the same. When replacing
a strut, be sure the replacement has
the same part number as the original.
A stainless steel tag is attached to the
inner probe assembly to identify the
115V design. Operating a probe on the
wrong heater voltage can cause injury
or severe damage to equipment.
1-5 PROBE INTERIOR
The inner probe assembly, Figure 1-7, consists
of six main parts:
a. Ceramic support rod with four holes running
through the length. The holes serve as insulated paths for the cell signal wire and
thermocouple wires.
b. A heater that is helically wrapped on a
quartz support cylinder and insulated.
Failure to follow the instructions in
this manual could cause danger to
personnel and equipment. Read and
follow instructions in this manual
c. A chromel-alumel thermocouple which acts
as the sensing element for the temperature
controller. (Not visible in Figure 1-7; located
within ceramic support rod.)
carefully.
d. A platinum screen pad which forms electri-
The oxygen probe includes an inner electrode for the cell assembly. It consists of a
platinum pad and a platinum/inconel composite wire which produces the cell constant
offset voltage described in the Nernst equation.
With this pad and wire, the constant will be
between -10 and +15 mV. The cell constant
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-7
cal contact with the inner electrode of the
electrochemical cell. (Not visible in Figure
1-7; located at end of ceramic support rod.)
The pad is attached to an inconel/stainless
steel wire which carries the signal to the
terminal strip.
e. A V-strut assembly to give support to the
inner probe assembly.
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
HEATER
INSULATING
GASKET
V-STRUT
CERAMIC
SUPPORT
ROD
Figure 1-7. Inner Probe Assembly
f. A tube to carry reference gas to the cell.
Refer to Section 4, Maintenance and Service,
for repair procedures for probe components.
1-6 JUNCTION BOX
The junction box, Figure 1-8, is positioned at the
external end of the probe and contains a terminal strip for electrical connections and fittings for
reference and calibration gases. Fittings are for
0.250 inch stainless steel tubing on American
units, and 6 mm tubing on European units. The
calibration fitting has a seal cap which must remain in place except during calibration. A tubing
fitting is also supplied to be used with the
calibration gas supply during calibration.
If the calibration gas bottles will be permanently
hooked up to the probe, a manual block valve is
required at the probe (between the calibration
fitting and the gas line) to prevent condensation
of flue gas down the calibration gas line.
During operation and calibration, reference gas
is supplied through the reference gas fitting to
the reference side of the cell. This gives the
REFERENCE
GAS TUBE
TAG
21220004
PROBE
JUNCTION BOX
COVER
TERMINAL
STRIP
CALIBRATION
GAS FITTING
REFERENCE
GAS FITTING
21240029
Figure 1-8. Junction Box
system a known quantity of oxygen with which
to compare the oxygen level in the process gas.
Though ambient air can be used for this purpose, accuracy can only be assured if a
reference gas air set is used.
During calibration, two gases of different known
oxygen concentrations are injected one at a
time through the calibration gas fitting. Stainless steel tubing delivers this gas to the process
side of the cell. In a healthy cell, the difference
in oxygen pressure from the process side to the
reference side of the cell will cause a millivolt
output proportional to the difference in oxygen
levels. The electronics unit can use the two millivolt outputs caused by the two calibration
gases for either automatic or semi-automatic
calibration.
1-8 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
Do not attempt to remove a process
gas sample through either gas fitting.
Hot gases from the process would
damage gas hoses in the junction box.
1-7 CABLE ASSEMBLY
The Westinghouse and Rosemount systems
use a 7-conductor cable to connect the probe to
the electronics package. Standard length for this
cable is 20 ft (6 m), but lengths up to 150 ft
(45 m) are available. The seven conductors
include one shielded pair of wires for the cell
millivolt signal, one shielded pair of type K wires
for the thermocouple, and three individual
16-gauge wires for the heater and for ground.
The assembled conductors are wrapped by a
TM
Teflon
jacket and braided stainless steel
Instruction Manual
IB-106-300DR Rev. 3.3
TM
shield. The Teflon
ing is suitable for high temperature use. All
metal shields are isolated at the probe end and
connect by drain wires to ground at the
electronics.
1-8 PROBE OPTIONS
a. Abrasive Shield Assembly
The abrasive shield assembly, Figure 1-9, is
a stainless-steel tube that surrounds the
probe assembly. The shield protects the
probe against particle abrasion and corrosive condensations, provides a guide for
ease of insertion, and acts as a probe position support, especially for longer length
probes. The abrasive shield assembly uses
a modified diffusion and vee deflector assembly fitted with dual dust seal packing.
and stainless steel jacket-
February 2000
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-9
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
0.187
0.187
2
1
B
A
o
15
03.584
03.554
o
90
ON INSIDE BREAK
FOR SMOOTH
A
0.45 MIN
125
SKIN CUT FACE FOR 90
ROUNDED EDGE ON
BOTH ENDS OF
CHAMFER.
6.00
0.187
o
/
B
NOTES:
VIEW A
o
/
0.75 THRU 4 PLS.
EQ SP ON 4.75 B.C.
WELD ON BOTH SIDES WITH EXPANDING
CHILL BLOCK.
2
BEFORE WELDING, BUTT ITEM 2 WITH
ITEM 1 AS SHOWN.
o
/
Figure 1-9. Abrasive Shield Assembly
VIEW B
o
22.5
0.745
DIA ON A 7.50 DIA B.C. (REF)
0.755
P0009
1-10 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
P0010
Figure 1-10. Ceramic Diffusion/Dust Seal
Assembly
These modified diffusion and vee deflector
assemblies are available in standard, Figure
1-10, and flame arrestor version, Figure
1-11.
b. Ceramic Diffusion Assembly
The ceramic diffusion assembly, Figure
1-12, is the traditional design for the probe.
Used for over 25 years, the ceramic diffusion assembly provides a greater filter
surface area for the probe.
19280010
Figure 1-12. Ceramic Diffusion Assembly
c. Flame Arrestor Diffusion Assembly
Where a high concentration of unburned
fuel is present in the exhaust gases, a flame
arrestor diffusion assembly, Figure 1-11 and
Figure 1-13 is recommended.
The flame diffusion assembly includes a set
of baffles between the cell and the stack
gases. This keeps 1500°F (816°C) cell
temperatures from igniting unburned fuel in
the stack.
P0011
P0012
Figure 1-13. Flame Arrestor Diffusion Assembly
Figure 1-11. Flame Arrestor Diffusion/Dust Seal
Assembly
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-11
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
P0014
Figure 1-14. Snubber Diffusion/Dust Seal
Assembly
d. Snubber Diffusion/Dust Seal Assembly
The snubber diffusion/dust seal assembly,
Figure 1-14, is used in applications where
an abrasive shield is to be used with a
snubber type diffusion element. The dust
seal consists of two rings of packing to prevent abrasive dust from collecting inside the
abrasive shield.
World Class 3000
can be employed. The bypass system uses
an 18 inch (457 mm) or 3 foot (0.92 m)
probe mounted externally on the stack or
duct. The process or exhaust gases are directed out to the probe through a passive
sampling system using inconel tubes. Flue
gas flow induces the movement of gases
into, through, and out of the bypass unit.
The bypass arrangement does not require
the use of aspiration air and the gas which
flows past the probe is returned to the stack
or duct.
The bypass probe package is normally used
for process temperatures of 1300°F (704°C)
to 2000°F (1094°C). A higher temperature
version of the bypass provides for operation
at temperatures up to 2500°F (1372°C). In
this version the pick up tubes are made of a
special high-temperature alloy.
Overall dimensions and mounting details of
the American and European bypass systems are shown in Figure 1-15.
f. Probe Mounting Jacket Options
e. Bypass Probe Options
For processes where the flue gas exceeds
the maximum allowable temperature of
1300°F (704°C) a bypass sensor package
A probe mounting jacket option is available
to allow the probe to operate at temperatures of up to 2000°F (1095°C). A separate
instruction bulletin is available for this
option.
1-12 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
C).
o
A - A
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
19290005
OF FLOW
DIRECTION
TO 980
o
POINT OR
F (650
DEW
o
Y AIR REQUIRED. FITTING
TO 1800
ARDS DIRECTION ONLY.
o
NW
AX. CLEAN DR
TION GAS REQUIRED.
AX. CALIBRA
CALIBRATION AND PURGE GAS CONNECTION. BITE TYPE FITTING (PARKER CPI)
REFERENCE AIR SUPPLY CONNECTION BITE TYPE FITTING (PARKER CPI) FOR
1.
NOTES:
10 SCFH AT 32 PSIG M
0.250 O.D. TUBING. 2 SCFH AT 3 PSIG M
IS LOCATED ON FAR SIDE.
2.
TURE RANGE 1200
TURE DOES NOT GO BELOW
PERA
PERA
YZER IN A VERTICALLY DOW
C.
o
ITH ANAL
FLUE GAS OPERATING TEM
LAG TO ENSURE GAS TEM
EXCEED 500
INSTALL W
4.
5.
3.
A
GROUP NOTE
ARENTHESES.
AL CONDUCTIVITY K
ETERS IN P
TION. THERM
ILLIM
ITH M
O INCH THK INSULA
ENDED TW
M
ENSIONS ARE IN INCHES W
RECOM
EQUALS 0.5 FOR INSULATION.
DIM
7.
6.
01
A 3 FT (914.4) GAS TUBE PICK-UP
B 6 FT (1828.8) GAS TUBE PICK-UP
GASKET AND HARDWARE
06
FLANGE
CAL
GAS
VIEW
2
REF
GAS
1
4.026
(102.26)
04
03
02
I.D.
ELECT CABLE
C 9 FT (2743.2) GAS TUBE PICK-UP
BY CUSTOMER
OPTIONAL MOUNTING
ARRANGEMENT, 4 IN.
150# FLANGE SUPPLIED
05
HARDWARE
GASKET AND
06
30.62 (777.75) REF (4507C26G01)
62.50 (1587.5) REF (4507C26G03)
26.50 (673.1) REF (4507C26G01 AND G02)
37.00 (939.8) REF (4507C26G01)
73.00 (1854.2) REF (4507C26G02)
109.00 (2768.6) REF (4507C26G03)
6.0
(152.4) REF
6.50
(165.1) REF
DRAIN
STD 20 FT
(6.1 M) CABLE
A
4
6
3
FOR
INSERTION
FOR PROBE
27.31 (693.67)
AND REMOVAL
CLEARANCE REQ
3D3947G01
19.80 (502.92)
SQUARE TO GAS FLOW.
BE ORIENTED SO THAT VEE SHIELD IS
IF EQUIPPED WITH THE OPTIONAL CERAMIC
DIFFUSOR ASSEMBLY, PROBE ASSEMBLY MUST
Figure 1-15. Bypass Probe Option (Sheet 1 of 3)
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-13
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
o
o
POINT OR
DEW
o
ARDS DIRECTION ONLY.
o
NW
YZER IN A VERTICALLY DOW
o
o
ITH ANAL
ARENTHESES.
ITH MILLIMETERS IN P
GROUP NOTE
A 3 FT (914.4) GAS TUBE PICK-UP
ELECT
B 6 FT (1828.8) GAS TUBE PICK-UP
C 9 FT (2743.2) GAS TUBE PICK-UP
CAL
GAS
2
CABLE
OMER
16
15
14
REF
1
13
12
11
GAS
A - A
VIEW
04
03
02
OF FLOW
DIRECTION
I.D.
4.026
(102.26)
19290006
PICKUP (3D390004G07 OR G08)
26.5 (673.1) ON 3 FT (914.4) OR 6 FT (1828.8)
62.5 (1587.5) ON 9 FT (2743.2) PICKUP (3D390004G09)
37.0 (939.8) ON 3 FT (914.4) PICKUP (3D390004G07)
73.0 (1854.2) ON 6 FT (1828.8) PICKUP (3D390004G08)
109.0 (2768.6) ON 9 FT (2743.2) PICKUP (3D390004G09)
CALIBRATION AND PURGE GAS CONNECTION. BITE TYPE FITTING (PARKER CPI)
10 SCFH AT 32 PSIG (220.64 kPa GAUGE) MAX. CALIBRATION GAS REQUIRED.
LAG TO ENSURE GAS TEMPERATURE DOES NOT GO BELOW
REFERENCE AIR SUPPLY CONNECTION BITE TYPE FITTING (PARKER CPI) FOR
0.250 O.D. TUBING. 2 SCFH AT 3 PSIG (20.69 kPa GAUGE) MAX. CLEAN DRY AIR
REQUIRED. FITTING IS LOCATED ON FAR SIDE.
1.
NOTES:
EXCEED 932 F (500 C).
3.
2.
A
A
STD CABLE
20 FT (6.1 M)
RECOMMENDED 2.0 INCH (50.8) THK INSULATION. THERMAL CONDUCTIVITY K EQUAL
0.5 FOR INSULATION.
DIMENSIONS ARE IN INCHES W
FLUE GAS OPERATING TEMPERATURE RANGE 1200 TO 1800 F (650 TO 980 C).
INSTALL W
6.
5.
4.
7.
01
INSERTION
FOR PROBE
27.31 (693.67)
CLEARANCE REQ
FLANGE
AND REMOVAL
08
07
06
AND
GASKET
HARDWARE
4
Figure 1-15. Bypass Probe Option (Sheet 2 of 3)
ELD BY CUST
W
10
09
05
FOR
19.80
(502.92)
3D3947G01
PLATE
ELDED
W
TO STACK
6.50 (165.1) REF
DRAIN
3
6
ITH THE OPTIONAL CERAMIC
SQUARE TO GAS FLOW.
BE ORIENTED SO THAT VEE SHIELD IS
IF EQUIPPED W
DIFFUSOR ASSEMBLY, PROBE ASSEMBLY MUST
1-14 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
PARTS LIST
NOTEDEFINER
ITEM
01WORLD CLASS
02GAS PICK-UPDWG
03GAS PICK-UPDWG
04GAS PICK-UPDWG
05ANALYZER
06MTG HARDWARE DWG
07MTG HARDWARE DWG
08MTG HARDWARE DWG
09ANALYZER
10ANALYZER
11GAS PICK-UPDWG
12GAS PICK-UPDWG
13GAS PICK-UPDWG
14GAS PICK-UPDWG
15GAS PICK-UPDWG
16GAS PICK-UPDWG
PART NAME
3000
HOUSING
HOUSING
HOUSING
PARTS LIST UNITS: INCHESGROUP NOTEABCABCABC
SIZEREFERENCE
INFORMATION
DWG
DWG
DWG
DWG
PROBE 18 IN.
3 FT
6 FT
9 FT
FOR WORLD CLASS
3000 ANSI
FOR ANSI FLANGE
FOR JIS
FOR DIN
FOR JIS
FOR DIN
FOR JIS 3 FT
FOR JIS 6 FT
FOR JIS 9 FT
FOR DIN 3 FT
FOR DIN 6 FT
FOR DIN 9 FT
Instruction Manual
IB-106-300DR Rev. 3.3
FLANGE
STYLEANSIJISDIN
MAIL CODE
PART NUMBER
OR REF DWG
ORDER FROM
MATRIX
4507C26G01
4507C26G02
4507C26G03
3D39005G01
3535B58G01
3535B58G03
3535B58G05
3D39005G02
3D39005G03
4507C26G04
4507C26G05
4507C26G06
4507C26G07
4507C26G08
4507C26G09
G01 G02 G03 G04 G05 G06 G07 G08 G09
1
111
111
3D39004 GROUP
1
1
111
111
1
1
1
February 2000
111
111
1
1
1
Extended Temperature By-Pass Arrangements (2400°°°°F; 1300°°°°C)
GROUP
PART NO.
CODEDESCRIPTION
1U0571G013’ By-pass Package with ANSI bolt pattern.
1U0571G026’ By-pass Package with ANSI bolt pattern.
1U0571G039’ By-pass Package with ANSI bolt pattern.
1U0571G043’ By-pass Package with JIS bolt pattern.
1U0571G056’ By-pass Package with JIS bolt pattern.
1U0571G069’ By-pass Package with JIS bolt pattern.
1U0571G073’ By-pass Package with DIN bolt pattern.
1U0571G086’ By-pass Package with DIN bolt pattern.
1U0571G099’ By-pass Package with DIN bolt pattern.
Figure 1-15. Bypass Probe Option (Sheet 3 of 3)
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-15
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
1-16 Description and SpecificationsRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Before starting to install this equipment, read the "Safety instructions for
the wiring and installation of this apparatus" at the front of this Instruction
Bulletin. Failure to follow the safety
instructions could result in serious
injury or death.
Install all protective equipment covers
and safety ground leads after installation. Failure to install covers and
ground leads could result in serious
injury or death.
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
SECTION 2
INSTALLATION
b. Ensure that the area is clear of obstructions,
internal and external, that will interfere with
installation. Allow adequate clearance for
removal of probe (Figure 2-1).
The junction box temperature must
not exceed 300°°°°F (149°°°°C). If used with
the Yokogawa ZA8C electronics, junction box temperature is limited to
160°°°°F (71°°°°C). To meet these limits, the
user may need to make a heat shield
or provide cooling air to the junction
box. The temperature limit is due to
the thermocouple "cold reference"
device.
2-1 LOCATING THE OXYGEN ANALYZER
PROBE
The location of the probe in the stack or flue is
most important for maximum accuracy in the
oxygen analyzing process. The probe must be
positioned so that the gas it measures is representative of the process. Best results are normally obtained if the probe is positioned near
the center of the duct (40 to 60% insertion). A
point too near the edge or wall of the duct may
not provide a representative sample because of
the possibility of gas stratification. In addition,
the sensing point should be selected so that the
process gas temperature falls within a range of
50° to 1300°F (10° to 704°C). Figure 2-1 provides mechanical installation references.
a. Check the flue or stack for holes and air
leakage. The presence of this condition will
substantially affect the accuracy of the oxygen reading. Therefore, either make necessary repairs or install the probe upstream of
any leakage.
2-2 SERVICE REQUIRED
a. Power input: 115 ±10% Vac single phase,
50 to 60 Hz, 3 amp minimum. (See label.)
b. Compressed air: 10 psig (69 kPa gauge)
minimum, 225 psig (1551 kPa gauge)
maximum at 2 scfh (56.6 L/hr) maximum;
supplied by one of the following (less than
40 parts-per-million total hydrocarbons).
Regulator outlet pressure should be set at 5
psi (35 kPa).
1. Instrument air - clean, dry.
2. Bottled standard air with step-down
regulator.
3. Bottled compressed gas mixture
(20.95% oxygen in nitrogen).
4. Other equivalent clean, dry, oil-free air
supply.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
FURNISHED IN - XIT
4512C34
ADAPTER & ACCESSORY
TO AMBIENT
INSULATE IF EXPOSED
WEATHER CONDITIONS
4512C35
4512C36
2.27 (58)
DIA MAX
ROSEMOUNT
5.85 (148.6)
DIM "A"
WITH STANDARD
REF GAS
CAL GAS
7.58 (192)
SNUBBER
DIFFUSER
ELEC
DIM "B" REMOVAL ENVELOPE
1/4 IN. TUBE
6 MM TUBE
6 MM TUBE
ANSI
DIN
JIS
1/2"
CONN
CONDUIT
GAS
CAL
JIS
4512C18H01
6.10
(155)
1.88 (48)
0.59
GAS
REF
(15)
AT THE BOTTOM
BOTTOM VIEW
INSTALL WITH CONNECTIONS
5.12
(130)
THESE FLAT FACED FLANGES ARE MANUFACTURED
TO ANSI, DIN, AND JIS BOLT PATTERNS AND ARE NOT
DIMENSIONS ARE IN INCHES WITH MILLIMETERS IN
PARENTHESES.
PRESSURE RATED.
2.
NOTES: 1.
24610001
3.80 (96.5)
DIFFUSER
FOR PROBE
WITH CERAMIC
4.90 (124.5)
ADD TO DIM "A" FOR
ARRESTOR
DIFFUSER AND FLAME
PROBE WITH CERAMIC
DIN
7.28
4512C19H01
ANSI
6.00
4512C17H01
(185)
(153)
0.71
0.75
(18)
(20)
5.71
4.75
(145)
(121)
DIM "B"
DIM "A"
27.3 (694)
16 (406)
45.3 (1151)
34 (864)
81.3 (2065)
117.3 (2980)
70 (1778)
106 (2692)
153.3 (3894)
142 (3607)
3535B18H02
ANSI
0.062 THK GASKET
3635B48H01
3535B45H01
JIS
DIN
ADD TO DIM "A"
TABLE I MOUNTING FLANGE
3FT
6FT
9FT
12 FT
PROCESS FLOW MUST
BE IN THIS DIRECTION
WITH RESPECT TO
DEFLECTOR 3534848G01
FLANGE
DIA.
HOLE
DIA.
(4) HOLES
EQ SP ON BC
TABLE II INSTALLATION/REMOVAL
18 IN.
PROBE
Figure 2-1. Probe Installation (Sheet 1 of 5)
2-2 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
7.50
7.48
BOLT
CIRCLE
0.75
(8) HOLES
DIAMETER
FLANGE
9.00 (153)
DIAMETER
TABLE IV. FLANGE SIZE
*
ANSI
7.48
0.75
0.945
9.25 (235)
9.25 (235)
*
*
JIS
DIN
FLANGES ARE MANUFACTURED TO ANSI,
*
5.7
(145)
14.5
(369)
DIM "D" REMOVAL ENVELOPE
PRESSURE RATED.
FLAT FACED. THESE FLANGES ARE NOT
DIN, AND JIS BOLT PATTERNS AND ARE
7.00
(178)
SEE TABLE IV
FOR FLANGE
SIZES
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
21190008
REF AND
CAL GAS
CONNECTOR
ELECTRICAL
CONNECTOR
CAL GAS LINES
CHECK VALVE FOR
INSULATE IF
CONDITIONS
EXPOSED TO
AMBIENT WEATHER
31.1
(790)
DIM "E"
45.3
(1151)
DIM "D"
27
(686)
DIM "C"
NOMINAL MEASUREMENTS
TABLE III. REMOVAL / INSTALLATION
3 FT
67.1
81.3
63
6 FT
(1704)
(2065)
(1600)
103.1
(2619)
117.3
(2980)
99
(2515)
9 FT
139.1
(3533)
153.3
(3894)
(P/N 3535B58G04 - JIS)
135
12 FT
(3429)
DIM "C"
0.06 THK GASKET FURNISHED
DIM "E" (WITH FLAME ARRESTOR)
(P/N 3535B58G06 - DIN)
(P/N 3535B58G02 - ANSI)
IN HARDWARE PACKAGE
Figure 2-1. Probe Installation (Sheet 2 of 5)
3.6
NOMINAL
(P/N 4843B38G02)
SNUBBER DIFFUSION/
DUST SEAL ASSEMBLY
DIMENSIONS ARE IN INCHES WITH
MILLIMETERS IN PARENTHESES.
NOTE:
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
JIS
(P/N 3535B58G04)
9.25
(235)
4.92
(125)
(200)
7.894
(M-20 x 2.5)
World Class 3000
16860021
8 THREADED HOLES
EQUALLY SPACED ON
D DIA B.C.
ABRASIVE SHIELD
FLANGE O.D.
DIN
(P/N 3535B58G06)
ANSI
(P/N 3535B58G02)
IN.
TABLE VI. ADAPTOR PLATE DIMENSIONS FOR ABRASIVE SHIELD
(mm)
DIMENSIONS
JIS
(P/N 4512C35G01)
9.25
9.00
"A"
6.50
(235)
(229)
(165)
3.94
(100)
4.75
(121)
"B"
DIA
(M-12 x 1.75)
7.48
(M-16 x 2)
7.50
0.625-11
"C"
"D"
THREAD
(130)
5.118
(190)
(191)
DIA
NOTE: PART NUMBERS FOR ADAPTOR PLATES INCLUDE
ATTACHING HARDWARE.
o
22.5
OUTSIDE WALL SURFACE.
CROSSHATCHED AREA IN 4
CORNERS MAY BE USED TO
FIELD BOLTING OF PLATE TO
PROVIDE ADDITIONAL HOLES FOR
AND 12 FT ABRASIVE SHIELD
ADAPTOR PLATE FOR 3, 6, 9,
INSTALLATIONS. SEE SHEET 2.
C
A
B
A
4 STUDS,
LOCKWASHERS AND
NUTS EQUALLY
SPACED ON
DIN
ANSI
TABLE V. ADAPTOR PLATE DIMENSIONS FOR PROBE
IN.
DIMENSIONS
7.5
(191)
(M-16 x 2)
(P/N 4512C36G01)
6.00
(153)
0.625-11
(P/N 4512C34G01)
"B"
"A"
(mm)
THREAD
(145)
5.708
4.75
(121)
"C"
DIA
A
o
45
A
C
ATTACHING HARDWARE.
2.500 DIA
NOTE: PART NUMBERS FOR ADAPTOR PLATES INCLUDE
C DIA B.C.
B
ADAPTOR PLATE FOR
STD WORLD CLASS 3000
PROBE INSTALLATION.
SEE SHEET 1.
Figure 2-1. Probe Installation (Sheet 3 of 5)
2-4 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
INSTALLATION FOR METAL
WALL STACK OR DUCT
CONSTRUCTION
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
INSTALLATION FOR MASONRY
WALL STACK CONSTRUCTION
MTG HOLES
SHOWN ROTATED
o
OUT OF
22.5
TRUE POSITION
WELD OR BOLT MOUNTING
PLATE TO METAL WALL
OF STACK OR DUCT.
JOINT MUST BE AIRTIGHT.
0.50 (13)
3.75 (95)
MIN DIA HOLE
IN WALL
STACK OR DUCT
METAL WALL
BOLT ADAPTOR
PLATE TO OUTSIDE
WALL SURFACE
FIELD WELD
PIPE TO
ADAPTOR PLATE
MTG HOLES
SHOWN ROTATED
o
OUT OF
22.5
TRUE POSITION
JOINT MUST
BE AIRTIGHT
OUTSIDE WALL
SURFACE
NOTE: ALL MASONRY STACK WORK AND JOINTS EXCEPT
ADAPTOR PLATE NOT FURNISHED BY ROSEMOUNT.
ABRASIVE SHIELD MOUNTING
0.50 (13)
4.50 (114)
O.D. REF
PIPE 4.00 SCHED 40
PIPE SLEEVE (NOT
BY ROSEMOUNT)
LENGTH BY CUSTOMER
MASONRY
STACK WALL
WELD OR BOLT MOUNTING
PLATE TO METAL WALL
OF STACK OR DUCT.
JOINT MUST BE AIRTIGHT.
3.25 (82.5)
MIN DIA HOLE
IN WALL
STACK OR DUCT
METAL WALL
BOLT MOUNTING
PLATE TO OUTSIDE
WALL SURFACE
OUTSIDE WALL
PROBE MOUNTING
JOINT MUST
BE AIRTIGHT
SURFACE
NOTE: DIMENSIONS IN INCHES WITH
MILLIMETERS IN PARENTHESES.
FIELD WELD
PIPE TO
ADAPTOR PLATE
4.0 (102)
O.D. REF
PIPE 3.5 SCHED 40
PIPE SLEEVE (NOT
BY ROSEMOUNT)
LENGTH BY CUSTOMER
MASONRY
STACK WALL
P0038
Figure 2-1. Probe Installation (Sheet 4 of 5)
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
(BOTH BRACE CLAMP ASSEMBLIES ARE THE SAME.
INSTALLATION AND LOCATION OF CLAMP ASSEMBLIES
AND BRACE BARS TO BE DONE IN FIELD.)
ABRASIVE SHIELD
BY ROSEMOUNT
}
NOTE: BRACING IS FOR VERTICAL AND HORIZONTAL PROBE INSTALLATION.
EXTERNAL BRACING REQUIRED FOR 9 FT AND 12 FT
(2.75 M AND 3.66 M) PROBES AS SHOWN ABOVE.
Figure 2-1. Probe Installation (Sheet 5 of 5)
2-3 MECHANICAL INSTALLATION
a. Ensure that all components are available for
installation of the probe. If equipped with the
optional ceramic diffusor, check to ensure
that it is not damaged. Ensure that the system cable is the required length.
b. The probe may be installed intact as it is re-
ceived. It is recommended that you disassemble the adapter plate for each
installation.
36.00 (914)
P0039
NOTE
An abrasive shield is recommended
for high velocity particulate in the flue
stream (such as those in coal fired
boilers, kilns, and recovery boilers).
Vertical and horizontal brace clamps
are provided for 9 ft and 12 ft (2.75 m
and 3.66 m) probes to provide mechanical support of the probe. Refer to
Figure 2-1, sheet 5.
c. Weld or bolt adapter plate (Figure 2-1) onto
the duct. Existing stack penetration and
mounting hardware may be revised, or new
mounting flanges may be installed (see
Figure 2-1, sheets 3 and 4).
2-6 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
d. If using the optional ceramic diffusor ele-
ment, the vee deflector must be correctly
oriented. Before inserting the probe, check
the direction of gas flow in the duct. Orient
the vee deflector on the probe so that the
apex points upstream toward the flow (Figure 2-2). This may be done by loosening the
setscrews, and rotating the vee deflector to
the desired position. Retighten the
setscrews.
GAS FLOW
DIRECTION
VEE
APEX
FILTER
DEFLECTOR
DIFFUSION
ELEMENT
SETSCREW
VEE
DEFLECTOR
e. In horizontal installations, the probe cover
should be oriented so the system cable
drops vertically from the probe cover. In a
vertical installation, the system cable can be
oriented in any direction.
f. If the system has an abrasive shield, check
the diffusion element dust seal packings.
The joints in the two packings must be
staggered 180°. Also, make sure that the
packings are in the hub grooves as the
probe slides into the 15° forcing cone in the
abrasive shield.
g. Insert the probe through the opening in the
mounting flange and bolt the unit to the
flange When probe lengths selected are 9
or 12 ft (2.74 or 3.66 m), special brackets
are supplied to provide additional support
for the probe inside the flue or stack. See
Figure 2-1, sheet 5.
2-4 ELECTRICAL INSTALLATION
Figure 2-3 shows the junction box wiring for the
direct replacement probe including the 115 V
hookup. Consult the electronics manual for the
connections to power and the existing signal
conditioning electronics.
P0017
Figure 2-2. Orienting the Optional Vee Deflector
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
NOTE: FOR YOKOGAWA (YEW) ZA8C,
115V
HEATER
}
YE CHROMEL
RD ALUMEL
OR CELL +VE
BL CELL -VE
1234567
BL
OR
JUNCTION BOX
WIRING
Upon completing installation, make sure that the probe is turned on and operating prior to
firing up the combustion process. Damage can result from having a cold probe exposed to
the process gases.
YE
RD
BK
BK
GN
WHT
8
GN
WHT
Figure 2-3. Wiring the Direct Replacement Probe
NOTE
!
PROBE
REFER TO SECTION 3-6.
TEST GAS INPUT
AT 2-3 L/MIN
(4-7 SCFH)
CHECK
VALV E
REF. GAS SUPPLY
INPUT AIR AT
1 L/MIN (2 SCFH)
19290007
During outages, and if possible, leave all probes running to prevent condensation and premature aging from thermal cycling.
If the ducts will be washed down during outage, MAKE SURE to power down the probes and
remove them from the wash area.
2-8 InstallationRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
3
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
SECTION 3
STARTUP
3-1 OVERVIEW
This section covers the setup procedures for the
World Class 3000 Oxygen Analyzer. The DR
probe can be used with several electronics
packages including: Models 218, 218A, 225,
132 (analog and digital), TC200, and ZA8C.
Install all protective equipment covers
and safety ground leads after setup.
Failure to replace covers and ground
leads could result in serious injury or
death.
For Models 218, 225, and 132 (analog), refer to
paragraph 3-2. Additional information can be
found in IB-106-101.
For Model 218A, refer to paragraph 3-3. Additional information can be found in IB-106-101A.
For Model TC200, refer to paragraph 3-4. Additional information can be found in IB-107-020.
d. Attach voltmeter with positive (+) lead on
TP1 and negative (-) on either Pin 2 or 7.
NOTE
The voltage given above is for an ambient (machinery space) temperature
of 77°F (25°C). For each degree of ambient temperature above or below 77°F
(25°C), add or subtract 0.242 mV from
the nominal. Example: at 87°F (31°C),
the nominal voltage of -322.3 Mv
should be increased (made less negative) by 10 x 0.242 or 2.42 mV, making
the adjusted nominal -319.9 Mv.
e. Adjust potentiometer M110-1 to read -322.3
millivolts nominal.
f. Remove voltmeter leads.
g. Remove jumper wire.
TP1
M110-1
For Model 132 (digital), refer to paragraph 3-5.
Additional information can be found in IB-106106A.
For Yokogawa Model ZA8C, refer to paragraph
3-6.
3-2 MODELS 218, 225, AND 132 (ANALOG)
ELECTRONICS SETUP
Before beginning operation, it is important that
the probe heater setpoint of the existing electronics be changed to support the World Class
3000 probe. The setpoint adjustment procedure
required for Models 218, 225, and 132 analog
electronics is as follows:
a. Open electronics enclosure.
b. On temperature controller card, Figure 3-1,
connect jumper wire from TP3 to either Pin
2 or Pin 7.
c. Set voltmeter to read DC millivolts (MV).
TP3
PIN 2
(PSC)
Figure 3-1. Temperature Controller Card
23 567 1012
PIN 7 (PSC)
Calibration Points
P0019
Rosemount Analytical Inc. A Division of Emerson Process Management Startup 3-1
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
3-3 MODEL 218A ELECTRONICS SETUP
Before beginning operation, it is important that
the probe heater setpoint of the existing electronics be changed to support the World Class
3000 probe. To convert the Model 218A Digital
Electronics Package for use with the DR probe,
an EPROM change is necessary. Remove Main
PCB and check back of board to identify unit as
G02 or G04. The replacement EPROM needed
is as identified below:
G02G04
United States1M03192G011M02982G01
United Kingdom1M03192G021M02982G02
Germany1M03192G031M02982G03
France1M03192G041M02982G04
Italy1M03192G051M02982G05
NOTE
The replacement EPROM when using a
multiprobe averager unit is
1M02982G10.
To replace the EPROM, proceed as follows:
a. Shut off and lock out power to the electron-
ics package.
b. Open electronics enclosure.
c. On the Main PCB, Figure 3-2, locate and
remove old EPROM.
d. Replace with new EPROM.
e. Close electronics enclosure and power up
system.
EPROM
MAIN PCB
Figure 3-2. Main PCB (Model 218A) EPROM Replacement
3-2 StartupRosemount Analytical Inc. A Division of Emerson Process Management
P0020
World Class 3000
3
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
3-4 MODEL TC200 VERITRIM ELECTRONICS
SETUP
Before beginning operation, it is important that
the probe heater setpoint of the existing electronics be changed to support the World Class
3000 Probe. To convert the Model TC200 Electronics Package for use with the DR probe, an
EPROM change is necessary. The replacement
EPROM needed is part number 1M03154G02.
a. EPROM Replacement
To replace the EPROM, proceed as follows:
1. Shut off and lock out power to the
electronics package.
2. Open electronics enclosure.
3. On the main PCB, Figure 3-3, locate
and remove old EPROMs U11 and
U12.
b. Heater SetPoint Adjustment
EPROMS
U11 AND U12
4. Replace with new EPROMs (part number 1M03154G02) being careful to install U11 and U12 in their proper
locations.
5. Close electronics enclosure and power
up system.
The adjustment procedure required for the
Model TC200 Electronics Package is as
follows:
1. Open keylocked enclosure to access
membrane keyboard.
2. Put controller in PAR (parameter)
mode by depressing "LOCK" "▲"
"%O
" "INC" "ACK" in sequence.
2
3. Depress "ACK" pushbutton to clear
display.
DIS A/M
U12
U11
Figure 3-3. Main PCB (Model TC200) EPROM Replacement
P0021
Rosemount Analytical Inc. A Division of Emerson Process Management Startup 3-3
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
4. Press "NUM" pushbutton.
5. Using "INC", "DEC" buttons, display
parameter 125.
6. Press "VAL" button.
7. Using "INC", "DEC" buttons, change
parameter 125 value to 15.4.
8. Press "ENT" to save new value.
3-5 MODEL 132 DIGITAL
ELECTRONICS SETUP
Before beginning operation, it is important that
the probe heater setpoint of the existing electronics be changed to support the World Class
3000 Probe. To convert the Model 132 Digital
Electronics Package for use with the DR probe,
an EPROM change is necessary. The replacement EPROM needed is as identified below:
United StatesIM03222G01
United Kingdom
Germany
France
Italy
EPROM
MAIN PCB
To replace the EPROM, proceed as follows:
a. Shut off and lock out power to the
electronics package.
b. Open electronics enclosure.
c. On the Main PCB, Figure 3-4, locate and
remove old EPROM.
d. Replace with new EPROM.
e. Close electronics enclosure and power up
system.
3-6 THE YOKOGAWA ZA8C AND AV8C CON-
VERTER ELECTRONICS SETUP
The DR probe can be wired to work with the
Yokogawa® ZA8C and AV8C Converters.
Connect the cabling from the ZA8C or AV8C
terminal to the probe terminal in the junction box
as shown in Figure 3-5.
TX2
C34
M55
CR18
R8
P61
U7
C44
U2
C48
U1
C37
LM8
R38
03
TP1
R27
XT1
C26
C27
R12
C38
U6
R38
C29
C28
U4
C30
04
C29
R26
P25
P36
CR25
C13
D12
J1
P28
O18
LED1
08
R63
R62
R54
U9
1
R41
R48
RP1
C46
TP2
C24
D14
C21
C23
D13
C12
C14
D1
C16
D2
CR17
R48
C39
20
C31
U10
C43
U3
C42
R53
U13
C22
C13
C10
C41
R42
CR15
CR16
CR24
R36
CR13
CR14
CR12
CR11
R37
BR1
C30
R44
R49
R47
U14
R43
R38
R86
CR18
R68
09
P29
P22
R64
C86
C10
U12
R61
C16
U2
C32
TR1
TX1
C38
R56
CR19
C36
C11
R46
R48
P62
C7
C9
C8
U11
R6
C48
C4
C3 C8
C47
C1
C2
C3
R9
2001A
F2
2001A
CK1
F1
C7
I
R58
R57
R59
P68
U19
U17
R18
R16
R17
U16
R11
R18
CR3
CR4
U18
R4
R5
CR1
CR2
08
A
28
28
R16
28
C46
CR7
CR6
R19
CR8
CR5
28
28
LK2
28
R7
R6
R14
R23
R28
LK1
CR21
CR20
28
28
C17
R2
R1
R21
R13
R24
CR23
CR22
5A
F3
5A
F4
FIL1
R1
CR9
R81
ALARM
MTR-T/C CELL SKG STK-T/C ML-O/P
T83
C11
T82
OFF
N
L
HEATER
E
E
N
MAIN INPUT
L
T81
SW1
MOV1
MOV2
ON
P0022
Figure 3-4. Main PCB (Model 132) EPROM Replacement
3-4 StartupRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
3
ZA8C
AV8C
CONVERTER
OR
WORLD CLASS 3000
REPLACEMENT PROBE
WITH ADAPTER FLANGE
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
STACK WALL
TEST GAS
IF AVAILABLE
CELL
THERMOCOUPLE
COLD
JUNCTION
HEATER AND SIGNAL CABLE
REFERENCE AIR
IF AVAILABLE
SIGNAL
CABLE
ZA8C
OR
AV8C
TERMINAL
3+
}
4–
5+
}
6–
7+
}
8–
EXISTING
MOUNTING
COLD
JUNCTION
DEVICE
+–
GN OR YE RD BK BKGN
12345678
SHIELD
PROBE
PROBE
TERMINAL
32
ZA8C OR AV8C
NOTES: 1. HEATER TEMPERATURE SET TO 1380
GND
SHIELD
THE GREATER MASS OF THE WORLD CLASS DIRECT REPLACEMENT PROBE REQUIRES
2.
LONGER TIME TO HEAT UP. UPON STARTUP, THE YOKOGAWA ELECTRONICS MAY
INDICATE AN ERROR BECAUSE THE PROBE HAS NOT REACHED TEMPERATURE SETPOINT
IN THE NORMAL TIME. REMOVE POWER FROM THE YOKOGAWA ELECTRONICS OR PROBE
MODULE TO CLEAR THE ERROR, AND RESTORE POWER. THIS PROCEDURE MAY HAVE TO
BE REPEATED A COUPLE OF TIMES BEFORE PROBE OPERATING TEMPERATURE IS REACHED.
PROBE
HEATER
16
17
o
F (750oC)
624023
Figure 3-5. DR Probe Wired to the ZA8C or A V8C Converter
Rosemount Analytical Inc. A Division of Emerson Process Management Startup 3-5
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
3-6 StartupRosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
4
IB-106-300DR Rev. 3.3
World Class 3000
SECTION 4
MAINTENANCE AND SERVICE
NOTE
!
Upon completing installation, make sure that the probe is turned on and operating prior to
firing up the combustion process. Damage can result from having a cold probe exposed to
the process gases.
During outages, and if possible, leave all probes running to prevent condensation and premature aging from thermal cycling.
If the ducts will be washed down during outage, MAKE SURE to power down the probes and
remove them from the wash area.
February 2000
4-1 OVERVIEW
This section describes routine maintenance of
the oxygen analyzer probe. Spare parts referred
to are avail-able from Rosemount. Probe disassembly kit 3535B42G01 contains the required
spanner and hex wrenches. Refer to Section 7
for part numbers and ordering information.
Install all protective equipment covers
and safety ground leads after equipment repair or service. Failure to install covers and ground leads could
result in serious injury or death.
4-2 PROBE CALIBRATION
The oxygen analyzer system should be calibrated when commissioned. Under normal circumstances the probe will not require frequent
calibration. When calibration is required, follow
the procedure described in the Instruction Bulletin applicable to your electronics package.
The cell replacement kit contains a cell and
flange assembly, corrugated seal, setscrews,
socket head cap screws, and anti-seize compound. Items are carefully packaged to preserve
precise surface finishes. Do not remove items
from packaging until they are ready to be used.
Spanner wrenches and hex wrenches needed
for this procedure are part of an available special tools kit, Table 7-1.
Wear heat resistant gloves and clothing to remove probe from stack. Normal operating temperatures of diffusor
and vee deflector are approximately
600°°°° to 800°°°°F (316°°°° to 427°°°°C). They
could cause severe burns.
Disconnect and lock out power before
working on any electrical components.
There is voltage up to 115 Vac.
4-3 CELL REPLACEMENT
This paragraph covers oxygen sensing cell replacement. Do not attempt to replace the cell
until all other possibilities for poor performance
have been considered. If cell replacement is
needed, order cell replacement kit, Table 7-1.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-1
Do not remove cell unless it is certain
that replacement is needed. Removal
may damage cell and platinum pad. Go
through complete troubleshooting
procedure to make sure cell needs replacement before removing it.
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
BOMB TAIL
CONNECTOR
CELL EXTENSION
WIRE (ORANGE)
INCONEL
CELL WIRE
(CLEAR
SLEEVING)
CALIBRATION
GAS FITTING
CABLE
HEATER
WIRES
(BLACK
SLEEVING)
THERMOCOUPLE -
(RED ALUMEL)
(YELLOW CHROMEL)
REFERENCE
GAS FITTING
THERMOCOUPLE +
PROBE JUNCTION
BOX COVER
21220001
Figure 4-1. Cell Wiring Connection
a. Disconnect and lock out power to electron-
ics. Shut off and disconnect reference gas
and calibration gas supplies from junction
box, Figure 4-1. Wearing heat resistant
gloves and clothing, remove probe assembly from stack carefully and allow to cool to
room temperature. Do not attempt to work
on unit until it has cooled to a comfortable
working temperature.
b. If the probe uses the standard diffusion
element, use a spanner wrench to remove
the diffusion element.
c. If equipped with the optional ceramic diffu-
sor assembly, remove and discard
setscrews, Figure 4-2, and remove vee deflector. Use spanner wrenches from probe
disassembly kit, Table 7-1, to turn hub free
from retainer. Inspect diffusion element. If
damaged, replace element.
d. Loosen four socket head cap screws from
the cell and flange assembly and remove
the assembly and the corrugated seal. The
cell flange has a notch which may be used
to gently pry the flange away from the
probe. Note that the contact pad inside the
probe will sometimes fuse to the oxygen
sensing cell. If the cell is fused to the contact pad, push the cell assembly back into
the probe (against spring pressure), and
quickly twist the cell assembly. The cell and
contact pad should separate. If the contactpad stays fused to the cell, a new contact/thermocouple assembly must be installed. Disconnect the cell and the
thermocouple wires at the probe junction
box, and withdraw the cell with the wires still
attached (see paragraph 4-5).
4-2 Maintenance and ServiceRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
4
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
PIN
WRENCH
OPTIONAL CERAMIC
DIFFUSION ELEMENT
SETSCREW
FILLET
VEE
DEFLECTOR
21240026
CEMENT
PORT
RETAINER
HUB
CEMENT
Figure 4-2. Removal of Optional Ceramic Diffusor
and Vee Deflector
e. If contact assembly is damaged, replace
contact and thermocouple according to
paragraph 4-5, Replacement of Contact and
Thermocouple Assembly.
f. Remove and discard corrugated seal. Clean
mating faces of probe tube and retainer.
Remove burrs and raised surfaces with
block of wood and crocus cloth. Clean
threads on retainer and hub.
g. Rub a small amount of anti-seize on both
sides of corrugated seal.
h. Assemble cell and flange assembly, corru-
gated seal, and probe tube. Make sure the
calibration tube lines up with the calibration
gas passage in each component. Apply a
small amount of anti-seize compound to
screw threads and use screws to secure
assembly. Torque to 35 in-lbs (4 N·m).
i. Apply anti-seize compound to threads of
cell assembly, hub, and setscrews. Reinstall
hub on cell assembly. Using pin spanner
wrenches, torque to 10 ft-lbs (14 N·m). If
applicable, reinstall vee deflector, orienting
apex toward gas flow. Secure with
setscrews and anti-seize compound.
Torque to 25 in-lbs (2.8 N·m).
PROBE TUBE
(NOT INCLUDED
IN KIT)
CORRUGATED
SEAL
CELL AND
FLANGE
ASSEMBLY
SOCKET HEAD
CAP SCREWS
CALIBRATION GAS
PASSAGE
Figure 4-3. Cell Replacement Kit
j. On systems equipped with an abrasive
shield, install dust seal gaskets, with joints
180° apart.
k. Reinstall probe and gasket on stack flange.
If there is an abrasive shield in the stack,
make sure dust seal gaskets are in place as
they enter 15° reducing cone.
l. Turn power on to electronics and monitor
thermocouple output. It should stabilize at
29.3 ±0.2 mV. Set reference air flow at 2
scfh (56.6 L/hr). After probe stabilizes, calibrate probe per Instruction Bulletin applicable to your electronics package. If new
components have been installed, repeat
calibration after 24 hours of operation.
4-4 OPTIONAL CERAMIC DIFFUSION
ELEMENT REPLACEMENT
a. General
The optional ceramic diffusion element
protects the cell from particles in process
gases. It does not normally need to be replaced because the vee deflector protects it
from particulate erosion. In severe environments the filter may be broken or subject to
excessive erosion. Examine the diffusion
element whenever removing the probe for
any purpose. Replace if damaged.
21240009
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-3
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
Damage to the diffusion element may become apparent during calibration. Compare
probe response with previous response. A
broken diffusion element will cause a slower
response to calibration gas.
Hex wrenches needed to remove setscrews
and socket head screws in the following
procedure are available as part of a special
tool kit, Table 7-1.
Wear heat resistant gloves and clothing to remove probe from stack. Normal operating temperatures of diffusor
and vee deflector are approximately
600°°°° to 800°°°°F (300°°°° to 425°°°°C). These
high temperatures can cause severe
burns.
Disconnect and lock out power before
working on any electrical component.
There is voltage up to 115 Vac.
with M-6 x 6 stainless setscrews
coated with anti-seize compound.
3. On systems equipped with abrasive
shield, remove dual dust seal gaskets.
4. Use spanner wrenches from special
tool kit, Table 7-1, to turn hub free from
retainer.
5. Put hub in vise. Break out old diffusion
element with chisel along cement line
and 3/8 inch (9.5 mm) pin punch
through cement port.
6. Break out remaining diffusion element
by tapping lightly around hub with
hammer. Clean grooves with pointed
tool if necessary.
7. Replace diffusion element, using replacement kit listed in Table 7-1. This
consists of a diffusion element, cement, setscrews, anti-seize compound
and instructions.
Do not remove the cell unless it is certain that replacement is necessary.
Cell cannot be removed for inspection
without damaging it. Refer to paragraph 4-3, Cell Replacement.
b. Replacement Procedure.
1. Shut off power to electronics. Disconnect cable conductors and remove cable, Figure 4-1. Shut off and disconnect
reference gas and calibration gas supplies from probe head. Wearing heat
resistant gloves and clothing, carefully
remove probe assembly from stack
and allow to cool to room temperature.
Do not attempt to work on unit until it
has cooled to a comfortable working
temperature.
2. Loosen setscrews, Figure 4-2, using
hex wrench from special tools kit,
Table 7-1, and remove vee deflector.
Inspect setscrews. If damaged, replace
8. Test fit replacement element to be sure
seat is clean.
Do not get cement on diffusion element except where it touches the hub.
Any cement on ceramic element
blocks airflow through element. Wiping wet cement off of ceramic only
forces cement into pores.
9. Thoroughly mix cement and insert tip
of squeeze bottle into cement port. Tilt
bottle and squeeze while simultaneously turning diffusion element into
seat. Do not get any cement on upper
part of diffusion element. Ensure complete penetration of cement around
three grooves in hub. Cement should
extrude from opposite hole. Wipe excess material back into holes and wipe
top fillet of cement to form a uniform
fillet. (A Q-Tip is useful for this.) Clean
any excess cement from hub with
water.
4-4 Maintenance and ServiceRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
4
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
10. Allow filter to dry at room temperature
overnight or 1 to 2 hours at 200°F
(93°C).
11. Wipe a heavy layer of anti-seize compound onto the threads and mating
surfaces of the diffusion hub and retainer.
12. Assemble retainer and diffusion hub
with two pin spanner wrenches. Torque
to 10 ft-lbs (14 N·m).
13. On systems equipped with abrasive
shield, install dust seal gaskets with
joints 180° apart.
14. Reinstall vee deflector, orienting apex
toward gas flow. Apply anti-seize compound to setscrews, and tighten with
hex wrench.
15. Reinstall probe on stack flange.
16. Turn power on to electronics and
monitor thermocouple output. It should
stabilize at 29.3 ±0.2 mV. Calibrate
probe per Instruction Bulletin applicable to your electronics package.
4-5 REPLACEMENT OF CONTACT AND
THERMOCOUPLE ASSEMBLY
a. Disconnect and lock out power to electron-
ics. Using heat resistant gloves and clothing, remove junction box cover. Squeezing
tabs on hose clamps, remove hoses from
junction box, Figure 4-4. Remove four
screws in corners of the junction box. Pull
junction box and inner probe assembly free
from probe tube. Set on bench and allow to
cool to room temperature.
b. Disconnect cell extension wire (orange),
thermocouple wire (red alumel), and thermocouple wire (yellow chromel) by cutting
bomb tail connections from the terminal
strip, Figure 4-1.
c. Remove two screws, Figure 4-4, lockwash-
ers, and flat washers that connect junction
box to inner probe assembly. Pull heater,
V-strut and backplate assembly away from
junction box. Inspect all O-rings and insulating gasket; replace if worn or damaged.
PROBE JUNCTION BOX TO HEATER,
STRUT, AND BACKPLATE
ASSEMBLY SCREW
PROBE JUNCTION BOX
INNER PROBE ASSEMBLY
TO PROBE TUBE SCREWS
PROBE
JUNCTION
BOX
HOSE
CLAMP
Use heat resistant gloves and clothing
when removing junction box and inner
probe assembly. Do not attempt to
HOSE
work on these components until they
have cooled to room temperature.
Probe components can be as hot as
800°°°°F (427°°°°C), and can cause severe
burns.
Disconnect and lock out power before
21240027
working on any electrical components.
There is voltage up to 115 Vac.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-5
Figure 4-4. Junction Box Mechanical Connections
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
HEATER SCREWS
(NOT SHOWN)
HEATER
CERAMIC ROD
35900001
MOUNTING SCREW
(REAR VIEW)
CONTACT AND
THERMOCOUPLER
ASSEMBLY
SPRING
INSULATING
GASKET
CLIP
V-STRUT
Figure 4-5. Inner Probe Replacement (Heater,
V-Strut, and Backplate Assembly)
d. Use a pencil to mark locations of spring clip
on ceramic rod, Figure 4-5.
e. Pry or squeeze tabs on spring clips, and
pull contact and thermocouple assembly out
of probe assembly. Retain spring clips and
spring; replace if damaged.
Be very careful when handling contact
and thermocouple assembly. The ceramic rod in this assembly is fragile.
f. While very carefully handling new contact
and thermocouple assembly, lay old assembly next to new one. Transfer pencil
marks to new rod.
g. Note wire lengths of old assembly as an aid
for trimming new lengths in step (j). Trimming of wires will not always be necessary.
Throw away old contact and thermocouple
assembly.
h. Carefully guide new contact and thermo-
couple assembly through V-strut assembly
leaf spring (4, Figure 4-6), spring (9), spring
clip (10) (held open by squeezing tabs), and
tube supports (11, 13) until spring clip
reaches pencil mark.
i. Reinstall insulating gasket on backplate, re-
place two screws, O-rings, lockwashers and
flat washers connecting junction box to inner probe assembly.
Do not trim new wiring shorter than
existing (old) wiring. Excessive wire
trim will prevent connections from
being properly made and will require
new replacement kit.
j. Trim wires, if necessary, as noted in step
(g).
1
13
12
1. Heater Ceramic Rod
2. Contact and Thermocouple
Assembly
3. Strut
4. Leaf Spring
3
2
11
1011
5. Ring Lug
6. Butt Connector
7. Extension Wire
8. Backplate
4
5
6
9
8
7
10. Spring Clip Assembly
11. Common Tube Support
12. Heater
13. Short Tube Support
9. Spring
4
8
21240011
Figure 4-6. Heater, Strut, and Backplate Assembly (Inner Probe Assembly)
4-6 Maintenance and ServiceRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
4
k. Connect color coded wires to proper termi-
nals as shown in Figure 4-1. Rosemount
recommends connecting the thermocouple
wires directly to the terminal strip. This is
because the junction of different metals at
the wires and lugs and at the lugs and the
terminal strip could act as additional thermocouple junctions. This could produce a
voltage that would affect the thermocouple
output signal.
Do not bend wires closer than 1/4 in. (6.4
mm) from end of ceramic rod. Dress wires
so they do not touch sides of junction box.
l. Slide assembled junction box and inner
probe assembly into probe tube. To align
calibration gas tube with corresponding hole
in backplate (A, B, Figure 1-3), insert scriber
through hole in backplate and into calibration gas tube. Secure with screws. Reinstall
hoses and probe cover.
m. Power up system. Monitor thermocouple
output. It should stabilize at setpoint mV
±0.2 mV. Recalibrate probe per Instruction
Bulletin applicable to your electronics package.
4-6 REPLACEMENT OF HEATER, V-STRUT
AND BACKPLATE ASSEMBLY (Inner
ProbeAssembly; Includes Contact And
Thermocouple Assembly).
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
There are two styles of the heater/strut
inner probe assembly. P/Ns
3D39441G06 through G10 identify the
44 V heater design. P/Ns 3D39441G01
through G05 and G11 through G15
identify the 115 V heater design. Both
styles look the same. When replacing
a strut, be sure the replacement has
the same part number as the original.
A stainless steel tag is attached to the
inner probe assembly to identify the
115V design. Operating a probe on the
wrong heater voltage can cause injury
or severe damage to equipment.
NOTE
This replacement may be done without
removing the probe from the duct.
a. Disconnect and lock out power to electron-
ics. Using heat resistant gloves and clothing, remove probe cover. Squeeze tabs on
hose clamps and remove hoses from junction box, Figure 4-4. Remove four screws
and lockwashers (7 and 10, Figure 4-7) that
hold junction box and inner probe assembly
to probe tube. Pull junction box and inner
probe assembly free from probe tube. Set
on bench and allow to cool to room temperature.
b. Disconnect cell extension wire (orange),
thermocouple wire (red alumel), and ther-
Wear heat resistant gloves and clothing when remove probe head and inner probe assembly. Do not attempt to
work on these components until they
have cooled to room temperature.
Probe components can be as hot as
800°°°°F (427°°°°C). This can cause severe
burns.
Disconnect and lock out power before
working on any electrical components.
There is voltage up to 115 Vac.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-7
mocouple wire (yellow chromel) by cutting
bomb tail connections from the terminal
strip, Figure 4-1.
c. Remove two screws, lockwashers, and flat
washers that connect junction box to inner
probe assembly. Remove and discard inner
probe assembly (heater, V-strut, and backplate assembly). Replace with new inner
probe assembly. Reinstall screws, lockwashers and flat washers.
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
124
World Class 3000
678 9
5
11
3
10
21240012
1. Snubber Diffusion Element
2. Socket Hd Cap Screw
(0.25 in.-28 x 0.63 (16 mm)
nals as shown in Figure 4-1. Rosemount
recommends connecting the thermocouple
wires directly to the terminal strip. This is
because the junction of different metals at
the wires and lugs and at the lugs and the
terminal strip could act as additional thermocouple junctions. This could produce a
voltage that would affect the thermocouple
output signal.
Do not bend wires closer than 1/4 in. (6.4
mm) from end of ceramic rod. Dress wires
so they do not touch sides of junction box.
4.0 in. (102 mm) x 4.0 in.
x 0.12 in. (3 mm)
8-32 x 0.5 in. (12.7 mm)
e. Slide assembled junction box and inner
f. Power up system. Monitor thermocouple
9. Hose Clamp
10. Lockwasher, #8 Split
11. Heater Strut Assembly
probe assembly into probe tube. To align
calibration gas tube with corresponding hole
in backplate (A, B, Figure 1-3), insert aligning tool (included in probe disassembly kit,
P/N 3535B42G01) through hole in backplate and into calibration gas tube, while
sliding the heater strut into the probe tube.
Secure with screws. Reinstall hoses and
probe cover.
output. It should stabilize at set point ±0.2
mV. Recalibrate probe per Instruction
Bulletin applicable to your electronics
package.
4-8 Maintenance and ServiceRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
4
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
4-7 CALIBRATION AND REFERENCE GAS
LINES FOR HIGH TEMPERATURE CORROSIVE ENVIRONMENT OPERATION
A high temperature, corrosive environment kit is
available when the probe is exposed to these
types of operating conditions. The kit includes
stainless steel tubing and teflon fittings for inside the junction box. The kit part number is
4843B93G01.
a. Installation Procedure
Use heat resistant gloves and clothing
when removing junction box and inner
probe assembly. Do not attempt to
work on these components until they
have cooled to room temperature.
Probe components can be as hot as
800°°°°F (427°°°°C). This can cause severe
burns.
Disconnect and lock out power before
working on any electrical components.
There is voltage up to 115 Vac.
1. Disconnect and lock out power to digital electronics. Using heat resistant
gloves and clothing, remove probe
cover. Squeezing tabs on hose clamps,
remove hoses from junction box
(Figure 4-4).
2. First install the stainless steel tubing on
the fitting at the bottom of the junction
box. Install the other end of the stainless steel tube onto the tube going to
the probe (Figure 4-8).
NOTE
If abrasive conditions of high ash content and high velocity exist, an abrasive shield is recommended. To
balance out the wear on the shield, rotate the shield 90° every time the
probe is powered down for service.
TEFLON
SLEEVES
STAINLESS
STEEL
TUBING
21240028
Do not use sealant when installing the
stainless steel tubes. Gas samples
Figure 4-8. High Temperature – Corrosive
Environment Kit
may become contaminated.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-9
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
4-10 Maintenance and ServiceRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
5
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
SECTION 5
TROUBLESHOOTING
5-1 OVERVIEW
Troubleshooting for the oxygen analyzer system
is broken down to the main component level. In
addition to the information in this section,
instruction bulletins for individual models also
discuss troubleshooting.
Install all protective equipment covers
and safety ground leads after troubleshooting. Failure to replace covers
and ground leads could result in serious injury or death.
For Models 218, 225, and 132 (analog),
refer to IB-106-101.
For Model 218A, refer to IB-106-101A.
For Model TC200, refer to IB-107-020.
For Model 132 (digital), refer to IB-106-106A.
For Yokogawa (YEW) ZA8C Converter,
refer to original converter documentation.
5-2 PROBE TROUBLESHOOTING
a. Probe Faults
Listed below are the five symptoms of probe
failure.
1. The system does not respond to
changes in the oxygen concentration.
2. The system responds to oxygen
changes but does not give the correct
indication.
3. The system does not give an acceptable indication of the value of the oxygen test gas being applied during
calibration.
4. The probe takes a long time to return
to flue gas value after the calibration
gas is turned off.
5. The probe heater temperature is
unstable.
b. Table 5-1 provides a guide to fault finding
for the above symptoms.
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 5-1
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
SYMPTOMCHECKFAULTREMEDY
1. No response to oxygen
concentration change
when:
World Class 3000
Table 5-1. Fault Finding
Heater is cold and
TC mV output is less
than setpoint
Heater is hot and
T/C mV output is at set
point ±0.2 mV
2. System responds to
oxygen concentration
changes but does not
give correct indication
Thermocouple continuityThermocouple failureReplace thermocouple or
return probe to Rosemount.
Heater cold resistance
should be 67 ohm to
77 ohm
Triac O/P to heaterFailure of electronicsCheck electronics package.
Cell mV input to electronics
and cell mV at probe head
Heater failureReplace heater or return
probe to Rosemount.
Manual.
No cell mV at probe when
test gas applied
Probe cell mV OK but no
input to electronics
Cell mV satisfactory both at
junction box and input to
electronics - failure of electronics
Replace cell or return probe
to Rosemount.
Check out cable connection.
Check electronics package.
Good response, with
incorrect indication
3. Probe does not give
accurate indicationof
applied test gas
5-2 TroubleshootingRosemount Analytical Inc. A Division of Emerson Process Management
Recorder or remote
indicator
System calibrationCalibration errorRecalibrate system.
Probe mounting and condition of duct
Cell mV input to electronics Failure of electronicsCheck electronics package.
Test gas input portBlocked portClean port. If the flue gas is
Ceramic diffusion element
Blocked port
Calibration errorRecalibrate recorder or in-
dicator. Reference Recorder Instruction Manual.
Air ingress into ductStop air leaks or resite
probe.
condensing in the test gas
line, insulate the back of the
probe. Make sure that the
test gas line is capped between calibrations, or a
check valve is installed.
Diffusion element cracked,
broken, or missing
Replace diffusion element.
World Class 3000
5
SYMPTOMCHECKFAULTREMEDY
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
Table 5-1. Fault Finding (continued)
4. Probe takes a long time
to return to flue gas
value after calibration
gas is turned off
5. Probe heater
temperature unstable
Plugged diffusion elementPlugged diffusion elementChange diffusion element or
snubber diffusion element.
Proper voltage heater is
installed
Wrong heaterChange heater to proper
voltage.
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 5-3
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
5-4 TroubleshootingRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
6
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
SECTION 6
RETURN OF MATERIAL
6-1 If factory repair of defective equipment is re-
quired, proceed as follows:
a. Secure a return authorization number from
a Rosemount Analytical Sales Office or
Representative before returning the equipment. Equipment must be returned with
complete identification in accordance with
Rosemount instructions or it will not be accepted.
In no event will Rosemount be responsible
for equipment returned without proper
authorization and identification.
b. Carefully pack defective unit in a sturdy box
with sufficient shock absorbing material to
insure that no additional damage will occur
during shipping.
c. In a cover letter, describe completely:
1. The symptoms from which it was determined that the equipment is faulty.
2. The environment in which the equipment has been operating (housing,
weather, vibration, dust, etc.).
3. Site from which equipment was removed.
5. Complete shipping instructions for return of equipment.
6. Reference the return authorization
number.
d. Enclose a cover letter and purchase order
and ship the defective equipment according
to instructions provided in Rosemount
Return Authorization, prepaid, to:
Rosemount Analytical Inc.
RMR Department
1201 N. Main Street
Orrville, Ohio 44667
If warranty service is requested, the defective unit will be carefully inspected and
tested at the factory. If failure was due to
conditions listed in the standard Rosemount
warranty, the defective unit will be repaired
or replaced at Rosemount's option, and an
operating unit will be returned to the customer in accordance with shipping instructions furnished in the cover letter.
For equipment no longer under warranty,
the equipment will be repaired at the factory
and returned as directed by the purchase
order and shipping instructions.
4. Whether warranty or nonwarranty
service is requested.
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 6-1
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
6-2 Return of MaterialRosemount Analytical Inc. A Division of Emerson Process Management
4-6, 23D39003G01
4-6, 23D39003G02
1-93D39003G07
1-93D39003G08
1-94847B61G01Cell Replacement Kit, ANSI w/No lead wire
1-94847B61G02Cell Replacement Kit, ANSI 18 in. (45.6 cm)
4-34847B61G03Cell Replacement Kit, ANSI 3 ft (0.9 m)
4-34847B61G04Cell Replacement Kit, ANSI 6 ft (1.8 m)
4-34847B61G05Cell Replacement Kit, ANSI 9 ft (2.7 m)
4-34847B61G06Cell Replacement Kit, ANSI 12 ft (3.6 m)
1
1
1
1
1
1
Heater, V-Strut, and Backplate Assembly (115 V), 18 in. (45.6 cm)
Heater, V-Strut, and Backplate Assembly (115 V), 3 ft (0.9 m)
Heater, V-Strut, and Backplate Assembly (115 V), 6 ft (1.8 m)
Heater, V-Strut, and Backplate Assembly (115 V), 9 ft (2.7 m)
Heater, V-Strut, and Backplate Assembly (115 V), 12 ft (3.6 m)
Heater, V-Strut, and Backplate Assembly, 18 in. (45.6 cm)
(For use with YEW electronics)
1
Heater, V-Strut, and Backplate Assembly, 3 ft (0.9 m)
(For use with YEW electronics)
1
Heater, V-Strut, and Backplate Assembly, 6 ft (1.8 m)
(For use with YEW electronics)
1
Heater, V-Strut, and Backplate Assembly, 9 ft (2.7 m)
(For use with YEW electronics)
1
Heater, V-Strut, and Backplate Assembly, 12 ft (3.6 m)
(For use with YEW electronics)
2
2
2
2
2
2
Contact and Thermocouple Assembly, 18 in. (45.6 cm)
Contact and Thermocouple Assembly, 3 ft (0.9 m)
Contact and Thermocouple Assembly, 6 ft (1.8 m)
Contact and Thermocouple Assembly, 9 ft (2.7 m)
Contact and Thermocouple Assembly, 12 ft (3.6 m)
Contact and Thermocouple Assembly, 18 in. (45.6 cm)
(For use with YEW electronics)
2
Contact and Thermocouple Assembly, 3 ft (0.9 m)
(For use with YEW electronics)
2
Contact and Thermocouple Assembly, 6 ft (1.8 m)
(For use with YEW electronics)
2
Contact and Thermocouple Assembly, 9 ft (2.7 m)
(For use with YEW electronics)
2
Contact and Thermocouple Assembly, 12 ft (3.6 m)
(For use with YEW electronics)
3
3
3
3
Abrasive Shield Assembly, 3 ft (0.9 m)
Abrasive Shield Assembly, 6 ft (1.8 m)
Abrasive Shield Assembly, 9 ft (2.7 m)
Abrasive Shield Assembly, 12 ft (3.6 m)
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-1
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
Table 7-1. Replacement Parts for Probe (continued)
FIGURE and
INDEX No.PART NUMBERDESCRIPTION
4-34847B61G07Cell Replacement Kit, JIS, w/No lead wire
4-34847B61G08Cell Replacement Kit, JIS 18 in. (45.6 cm)
4-34847B61G09Cell Replacement Kit, JIS 3 ft (0.9 m)
4-34847B61G10Cell Replacement Kit, JIS 6 ft (1.8 m)
4-34847B61G11Cell Replacement Kit, JIS 9 ft (2.7 m)
4-34847B61G12Cell Replacement Kit, JIS 12 ft (3.6 m)
4-34847B61G13Cell Replacement Kit, DIN w/No lead wire
4-34847B61G14Cell Replacement Kit, DIN 18 in. (45.6 cm)
4-34847B61G15Cell Replacement Kit, DIN 3 ft (0.9 m)
4-34847B61G16Cell Replacement Kit, DIN 6 ft (1.8 m)
4-34847B61G17Cell Replacement Kit, DIN 9 ft (2.7 m)
4-34847B61G18Cell Replacement Kit, DIN 12 ft (3.6 m)
4-33535B42G01Probe Disassembly Kit
1-133535B18G01Ceramic Diffusor Assembly
1-113535B60G01
Ceramic Diffusor Dust Seal Hub Assembly
(For use with Abrasive Shield)
1-134841B03G02Stainless Steel Diffusor Assembly
1-123535B63G01Flame Arrestor Diffusor Dust Seal
1-143535B62G01Flame Arrestor Diffusor
4-23535B48G01
Vee Deflector Assembly
(For use with standard ceramic or dust seal type diffusors)
4-26292A74G02Diffusion Element Replacement Kit
2-11537B70G03Horizontal and Vertical Brace Clamp Assembly, 9 and 12 foot
(2.7 and 3.6 m) probe
4-84843B93G01High Temperature - Corrosive Environment Kit
Heater, V-strut, and backplate assembly includes contact and thermocouple assembly.
2
Contact and thermocouple assembly includes platinum pad and inconel wire.
3
Abrasive shield assembly includes accessories necessary for its use and a mounting plate and gasket.
World Class 3000
7-2 Replacement PartsRosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
8
This index is an alphabetized listing of parts, terms, and procedures having to do with the Hazardous Area Oxygen/Combustibles Transmitter. Every item listed in this index refers to a location
in the manual by one or more page numbers.
Rosemount Analytical Inc. A Division of Emerson Process Management Index 8-1
T
Thermocouple, 4-2, 4-5
V
Vee Deflector, 1-7, 2-8
V-Strut, 1-7, 4-6
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
W
Water Vapor, 1-2
Z
Zirconia Disc, 1-2
8-2 IndexRosemount Analytical Inc. A Division of Emerson Process Management
WARRANTY
Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from
defects in workmanship and material under normal use and service for a period of twelve (12)
months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid
junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and
material under normal use and service for a period of ninety (90) days from date of shipment by
Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or
material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the
goods, part(s) or consumables are returned to Seller's designated factory, transportation charges
prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in
the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in
effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day
warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such commercial unit when such goods, part(s) and
consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly
or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from
any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS,
SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED
WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD
WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF ACTION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC
GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER
AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDENTAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT
ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT.
IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER
NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS
COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO
BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED
HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond
Seller's direct control.
2461
3267/2-00
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
World Class 3000
World Class 3000 Probe
Pa rt n o. __ __ ____________
Se ria l n o._ ______________
Ord er no._______________
Emerson Process Management
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
E gas.csc@emersonprocess.com
ASIA - PACIFIC
Fisher-Rosemount
Singapore Private Ltd.
1 Pandan Crescent
Singapore 128461
Republic of Singapore
T 65-777-8211
F 65-777-0947