Rosemount World Class 3000 O2 Analyzer DR Direct Replacement Probe-Rev 3.3 Manuals & Guides

Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
Oxygen Analyzer Direct Replacement Probe for use with Existing Signal Conditioning Electronics
http://www.processanalytic.com

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical repre-
sentative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and mainte-
nance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate In- struction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s per­formance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
HIGHLIGHTS OF CHANGES
Effective June, 1996 Rev. 3
Page Summary
--- General. Updated text and illustrations to reflect probe design changes.
Page 1-15 Added “Extended temperature by-pass arrangements” to Figure 1-15,
sheet 3 of 3.
Effective December, 1996 Rev. 3.1
Page Summary
Page P-2 Added "Safety instructions for the wiring and installation of this
apparatus".
Page 1-4 Corrected Figure 1-3 item listing.
Page 2-1 Added one WARNING to read new safety instructions and another
WARNING regarding protective covers and grounds.
Page 3-1 Added WARNING regarding protective covers and grounds.
Page 4-1 Added WARNING regarding protective covers and grounds.
Page 4-6 Added title block to Figure 4-6.
Page 5-1 Added WARNING regarding protective covers and grounds.
Page 7-1 Corrected cell replacement kit part numbers in Table 7-1.
Page 7-2 Corrected cell replacement kit part number in Table 7-1.
Effective May, 1997 Rev. 3.2
Page Summary
Page P-2 Added safety sheet.
Effective February, 1998 Rev. 3.3
Page Summary
Page 2-2 Figure 2-1. Changed calibration gas tube dimensions.
World Class 3000
PREFACE.........................................................................................................................P1
Definitions ......................................................................................................................... P1
Safety Instructions ........................................................................................................... P2
1-0 DESCRIPTION AND SPECIFICATIONS........................................................................ 1-1
1-1 Component Checklist ..................................................................................................... 1-1
1-2 System Overview............................................................................................................ 1-1
1-3 Physical Description ....................................................................................................... 1-3
1-4 Probe Exterior................................................................................................................. 1-5
1-5 Probe Interior .................................................................................................................. 1-7
1-6 Junction Box .................................................................................................................... 1-8
1-7 Cable Assembly...............................................................................................................1-9
1-8 Probe Options................................................................................................................. 1-9
2-0 INSTALLATION .............................................................................................................. 2-1
2-1 Locating The Oxygen Analyzer Probe.......................................................................... 2-1
2-2 Service Required .............................................................................................................2-1
2-3 Mechanical Installation ................................................................................................... 2-6
2-4 Electrical Installation ........................................................................................................ 2-7
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000

TABLE OF CONTENTS

3-0 STARTUP........................................................................................................................ 3-1
3-1 Overview.......................................................................................................................... 3-1
3-2 Models 218, 225, And 132 (Analog) Electronics Setup............................................. 3-1
3-3 Model 218A Electronics Setup ..................................................................................... 3-2
3-4 Model TC200 Veritrim Electronics Setup ..................................................................... 3-3
3-5 Model 132 Digital Electronics Setup............................................................................ 3-4
3-6 The Yokogawa ZA8C And AV8C Converter Electronics Setup ................................ 3-4
4-0 MAINTENANCE AND SERVICE .................................................................................. 4-1
4-1 Overview.......................................................................................................................... 4-1
4-2 Probe Calibration ............................................................................................................. 4-1
4-3 Cell Replacement ........................................................................................................... 4-1
4-4 Optional Ceramic Diffusion Element Replacement ..................................................... 4-3
4-5 Replacement Of Contact And Thermocouple Assembly ............................................. 4-5
4-6 Replacement Of Heater, V-Strut And Backplate Assembly ..................................... 4-7
4-7 Calibration And Reference Gas Lines For High Temperature -
Corrosive Environment Operation ................................................................................... 4-9
5-0 TROUBLESHOOTING .................................................................................................... 5-1
5-1 Overview.......................................................................................................................... 5-1
5-2 Probe Troubleshooting ................................................................................................... 5-1
6-0 RETURN OF MATERIAL .............................................................................................. 6-1
7-0 REPLACEMENT PARTS ................................................................................................ 7-1
8-0 INDEX.............................................................................................................................. 8-1
Rosemount Analytical Inc. A Division of Emerson Process Management i
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
Figure 1-1. Typical System Package ....................................................................................... 1-1
Figure 1-2. Typical System Installation .................................................................................... 1-3
Figure 1-3. Oxygen Analyzer Probe - Exploded View.............................................................. 1-4
Figure 1-4. Main Probe Components ....................................................................................... 1-5
Figure 1-5. Cell and Tube Assemblies ..................................................................................... 1-5
Figure 1-6. Optional Ceramic Diffusor and Vee Deflector Assembly....................................... 1-7
Figure 1-7. Inner Probe Assembly............................................................................................ 1-8
Figure 1-8. Junction Box .......................................................................................................... 1-8
Figure 1-9. Abrasive Shield Assembly ................................................................................... 1-10
Figure 1-10. Ceramic Diffusion/Dust Seal Assembly ............................................................... 1-11
Figure 1-11. Flame Arrestor Diffusion/Dust Seal Assembly..................................................... 1-11
Figure 1-12. Ceramic Diffusion Assembly................................................................................ 1-11
Figure 1-13. Flame Arrestor Diffusion Assembly ..................................................................... 1-11
Figure 1-14. Snubber Diffusion/Dust Seal Assembly............................................................... 1-12
Figure 1-15. Bypass Probe Option (Sheet 1 of 3) .................................................................... 1-13
Figure 2-1. Probe Installation (Sheet 1 of 5) ............................................................................ 2-2
Figure 2-2. Orienting the Optional Vee Deflector ..................................................................... 2-7
Figure 2-3. Wiring the Direct Replacement Probe ................................................................... 2-8
Figure 3-1. Temperature Controller Card Calibration Points ................................................... 3-1
Figure 3-2. Main PCB (Model 218A) EPROM Replacement ................................................... 3-2
Figure 3-3. Main PCB (Model TC200) EPROM Replacement ................................................. 3-3
Figure 3-4. Main PCB (Model 132) EPROM Replacement...................................................... 3-4
Figure 3-5. DR Probe Wired to the ZA8C or A V8C Converter ................................................ 3-5
Figure 4-1. Cell Wiring Connection .......................................................................................... 4-2
Figure 4-2. Removal of Optional Ceramic Diffusor and Vee Deflector .................................... 4-3
Figure 4-3. Cell Replacement Kit ............................................................................................. 4-3
Figure 4-4. Junction Box Mechanical Connections.................................................................. 4-5
Figure 4-5. Inner Probe Replacement (Heater, V-Strut, and Backplate Assembly)................ 4-6
Figure 4-6. Heater, Strut, and Backplate Assembly (Inner Probe Assembly).......................... 4-6
Figure 4-7. Oxygen Analyzer Probe, Cross-Sectional View .................................................... 4-8
Figure 4-8. High Temperature – Corrosive Environment Kit .................................................... 4-9
World Class 3000

LIST OF ILLUSTRATIONS

LIST OF TABLES

Table 1-1. Specifications for Oxygen Analyzing Equipment
Table 5-1. Fault Finding .......................................................................................................... 5-2
Table 7-1. Replacement Parts for Probe ................................................................................ 7-1
ii Rosemount Analytical Inc. A Division of Emerson Process Management
1 2
............................................... 1-6
World Class 3000
The World Class 3000 Direct Replacement Probe can be used with any of the following model electronics: 218, 225, 218A, ZA8C, and TC200. Read the appropriate setpoint adjust­ment in this manual to prevent damage to the probe.
The purpose of this manual is to provide information concerning the components, func­tions, installation and maintenance of this particular probe.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.
The following definitions apply to WARNINGS, CAUTIONS, and NOTES found throughout this publication.
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
CAUTION
!

PREFACE

DEFINITIONS
Highlights an operation or maintenance procedure, practice, condition, state­ment, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
NOTE
Highlights an essential operating procedure, condition, or statement.
: EARTH (GROUND) TERMINAL
: PROTECTIVE CONDUCTOR TERMINAL
: RISK OF ELECTRICAL SHOCK
: WARNING: REFER TO INSTRUCTION BULLETIN
Highlights an operation or maintenance procedure, practice, condition, state­ment, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE TO USERS
The number in the lower right corner of each illustration in this publication is a manual illus­tration number. It is not a part number, and is not related to the illustration in any technical manner.
Rosemount Analytical Inc. A Division of Emerson Process Management P-1
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
FOR THE WIRING AND INSTALLATION
The following safety instructions apply specifically to all EU member states. They should be strictly adhered to in order to assure compliance with the Low Voltage Directive. Non­EU states should also comply with the following unless superseded by local or National Standards.
1. Adequate earth connections should be made to all earthing points, internal and external, where provided.
2. After installation or troubleshooting, all safety covers and safety grounds must be replaced. The integrity of all earth terminals must be maintained at all times.
3. Mains supply cords should comply with the requirements of IEC227 or IEC245.
World Class 3000
IMPORTANT
SAFETY INSTRUCTIONS
OF THIS APPARATUS
4. All wiring shall be suitable for use in an ambient temperature of greater than 75°C.
5. All cable glands used should be of such internal dimensions as to provide adequate cable anchorage.
6. To ensure safe operation of this equipment, connection to the mains supply should only be made through a circuit breaker which will disconnect all circuits carrying conductors during a fault situation. The circuit breaker may also include a mechanically operated isolating switch. If not, then another means of disconnecting the equipment from the supply must be provided and clearly marked as such. Circuit breakers or switches must comply with a recognized standard such as IEC947. All wiring must conform with any local standards.
7. Where equipment or covers are marked with the symbol to the right, hazard­ous voltages are likely to be present beneath. These covers should only be removed when power is removed from the equipment — and then only by trained service personnel.
8. Where equipment or covers are marked with the symbol to the right, there is a danger from hot surfaces beneath. These covers should only be removed by trained service personnel when power is removed from the equipment. Cer­tain surfaces may remain hot to the touch.
9. Where equipment or covers are marked with the symbol to the right, refer to the Operator Manual for instructions.
10. All graphical symbols used in this product are from one or more of the follow­ing standards: EN61010-1, IEC417, and ISO3864.
P-2 Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
DESCRIPTION AND SPECIFICATIONS
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
SECTION 1

1-1 COMPONENT CHECKLIST

The Direct Replacement (DR) Probe package of the World Class 3000 Oxygen Analyzer typically includes the probe, the instruction bulletin, and the mounting hardware. These three items are shown in Figure 1-1. Record the Part Number, Serial Number, and Order Number of your system on the first page of this manual.

1-2 SYSTEM OVERVIEW

a. Scope
This Instruction Bulletin has been designed to supply details needed to install, start up, and troubleshoot the Rosemount World Class 3000 Oxygen Analyzer. The Direct
Replacement Probe can be interfaced to a number of different and earlier model electronics packages. These electronic packages are not covered in this manual. For specific information concerning calibra­tion and operation of the system, refer to the Instruction Bulletin applicable to your electronics package.
b. System Function
The Rosemount Oxygen Analyzer (Probe) is designed to measure the net concentration of oxygen in an industrial process; i.e., the oxygen remaining after all fuels have been oxidized. The probe is permanently posi­tioned within an exhaust duct or stack. The sampling system is in situ rather than extractive.
2
1. Oxygen Analyzer Probe
2. Mounting Adapter Plate with Mounting Hardware and Gasket
3. Instruction Bulletin
3
1
19290001
Figure 1-1. Typical System Package
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
The equipment measures oxygen percent­age by reading the voltage developed across a heated electrochemical cell, which consists of a small yttria-stabilized, zirconia disc. Both sides of the disc are coated with porous metal electrodes. When operated at the proper temperature, the millivolt output voltage of the sensor cell is given by the following Nernst equation:
EMF = KT log
10(P1/P2
) + C
Where:
1. P
is the partial pressure of the oxygen
2
in the measured gas on one side of the cell,
2. P
is the partial pressure of the oxygen
1
in the reference gas on the other side,
3. T is the absolute temperature,
4. C is the cell constant,
5. K is an arithmetic constant.
NOTE
For best results, use clean, dry, instrument air (20.95% oxygen) as a reference gas.
vapor. Therefore, it may be considered an analysis on a "wet" basis. In comparison with older methods, such as the Orsat apparatus, which provides an analysis on a "dry" gas basis, the "wet" analysis will, in general, indicate a lower percentage of oxygen. The difference will be proportional to the water content of the sampled gas stream.
c. System Configuration
The DR probe is similar to the Exchange Probe (NE model) except that the DR uses a 115 V heater like most older systems. The DR does not require a separate power sup­ply as does the NE. The DR probe (like the NE) is available in five lengths. The length of stack penetration depends on the size of the stack or duct. The length options are 18 in., 3 ft, 6 ft, 9 ft, and 12 ft (457 mm, 0.9 m, 1.8 m, 2.7 m, and 3.7 m).
d. Features
1. Output voltage and sensitivity increase as the oxygen concentration decreases.
When the cell is at operating temperature, oxygen ions will travel through the cell from high to low oxygen partial pressures. The resulting output voltage is about 50 mV per decade. The voltage varies with the ratio of oxygen partial pressures of the reference gas over the sample gas. Thus, the output signal increases as the sample oxygen level decreases. This property gives the analyzer high sensitivity at low oxygen levels.
Oxygen analyzer equipment measures net oxygen concentration in the presence of all the products of combustion, including water
2. In situ, non-sampling analyzer.
3. Field replaceable cell.
4. Analyzer constructed of rugged 316 LSS for all wetted parts.
5. Suitable for use in temperatures up to 1300°F (700°C).
6. The DR probe uses a 115 V heater and does not require a separate heater power supply.
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
e. Handling the Oxygen Analyzer
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
GASES
It is important that printed circuit boards and integrated circuits are handled only when adequate antistatic precautions have been taken to pre­vent possible equipment damage.
The oxygen analyzer is designed for industrial application. Treat each component of the system with care to avoid physical damage. The probe contains components made from ce­ramics, which are susceptible to shock when mishandled.
NOTE
Retain packaging in which the oxygen analyzer arrived from the factory in case any components are to be shipped to another site. This packag­ing has been designed to protect the product.
f. System Considerations
Prior to installation of your Rosemount World Class 3000 Oxygen Analyzer, make sure you have all of the components necessary to make the system installation. Ensure all components are properly integrated to make the system functional.
STACK
CALIBRATION
INSTRUMENT
AIR SUPPLY
(REF. GAS)
GAS
PRESSURE
REGULATOR
FLOWMETER
EXISTING
ELECTRONICS
Figure 1-2. Typical System Installation
connecting test gas tanks to the oxygen analyzer when the probe is to be calibrated.
NOTE
Ambient air is not recommended for use as high test gas. An 8% O balance in nitrogen is recommended for high test gas.
DUCT
ADAPTER
OXYGEN
ANALYZER
(PROBE)
2
PLATE
VOLTAGE
19290002
LINE
Once you have verified you have all the components, select mounting locations and determine how each component will be placed in terms of available power supply, ambient temperatures, environmental con­siderations, convenience, and serviceability. A typical installation is illustrated in Figure 1-2.
If test gas bottles will be hooked up perma­nently, an optional check valve is recom­mended next to the calibration fittings on the probe head. This is to prevent breathing of calibration gas line and subsequent flue gas condensation and corrosion. The check valve is in addition to the stop valve in the test gas kit or the solenoid valve in the
multiprobe test gas sequencer units. After selecting the probe mounting location, provision should be made for a platform

1-3 PHYSICAL DESCRIPTION

where the probe can be easily serviced.
The Oxygen Analyzer Probe, Figure 1-3 con­A source of instrument air is required at the probe for reference gas use. Since the
sists of three component groups: probe exterior,
inner probe, and junction box, Figure 1-4. probe is equipped with an in place calibra­tion feature, provision should be made for
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
NOTE: NOT ALL PARTS SHOWN ARE AVAILABLE FOR
PURCHASE SEPARATELY. FOR LIST OF AVAILABLE PARTS, SEE TABLE 7-1.
World Class 3000
1
26
20
18
23
21
19
16
14
20
12
29
10
17
11
13
15
9
28
3
2
4
30
27
25
24
11
22
10
5
6
NOTE: ITEM , CALIBRATION GAS TUBE,
FITS INTO HOLES WHEN PROBE IS
ASSEMBLED.
1. Heater, Strut, and Backplate 11. Probe Junction Box Cover 22. Gas Connection Assembly 12. Cover Gasket 23. Seal Cap
2. Diffusion Assembly 13. Insulating Sheet 24. Label
3. Retainer Screw 14. O-Ring 25. Probe Junction Box
4. Cell and Flange 15. Terminal Block Screw 26. Ground Wires
5. Corrugated Seal 16. Terminal Block 27. Insulating Gasket
6. Probe Tube Assembly 17. Terminal Block Marker 28. Washer
7. Cover Screw 18. Terminal Block Mounting Plate 29. Screw
8. Washer 19. Probe Junction Box Screws 30. Tag
9. Cover Chain Screw 20. Hose Clamp
10. Cover Chain 21. Hose
35900002
8
7
Figure 1-3. Oxygen Analyzer Probe - Exploded View
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
PROBE EXTERIOR (SENSING CELL INSTALLED)
PROBE INTERIOR
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
PROBE
JUNCTION
BOX
Figure 1-4. Main Probe Components

1-4 PROBE EXTERIOR

Primary probe exterior components include a flange-mounted zirconium oxide cell, mounted on a tube assembly and protected by a snubber diffusion assembly.
a. Cell and Flange Assembly
The primary component in the cell and flange assembly, Figure 1-5, is a yttria-stabilized zirconium oxide cell. The cell creates an electrical signal when the oxygen level on one side is out of balance with the oxygen level on the other side. This signal is proportional to the difference in oxygen levels.
b. Probe Tube Assembly
21240006
The tube assembly includes a flange which mates with a stack-mounted flange (shown attached to the probe flange in Figure 1-4). Studs on the stack flange make installation easy. There is also a tube to carry calibra­tion gas from the junction box to the process side of the cell during calibration.
PROBE TUBE
CORRUGATED
SEAL
CELL AND
FLANGE
ASSEMBLY
Four screws secure the cell and flange as­sembly, Figure 1-5, to the probe tube as­sembly. When in place, the cell is inside the tube.
21240007
Figure 1-5. Cell and Tube Assemblies
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
Table 1-1. Specifications for Oxygen Analyzing Equipment
1 2
Probe lengths, nominal ...................................................................... 18 in. (457 mm), 3 ft (0.91 m),
6 ft (1.83 m), 9 ft (2.74 m), or 12 ft (3.66 m), depending on duct dimensions
Temperature limits in process
measurement area ....................................................................... 50° to 1300°F (10° to 704°C)
Standard/current output ................................................................ 4-20 mA dc signal (factory set)
O
indication (Digital display
2
and analog output)........................................................................ 0.1% O
or ±3% of reading, whichever is greater
2
using Rosemount test gases
System speed of response ................................................................ less than 3 seconds (amplifier output)
Resolution sensitivity ......................................................................... 0.01% O
transmitted signal
2
Probe reference air flow .................................................................... 2 scfh (56.6 L/hr) clean, dry, instrument
quality air (20.95% O
), regulated to
2
5 psi (34 kPa)
Calibration gas mixtures .................................................................... Rosemount Hagan Test Gas Kit Part No.
6296A27G01 contains 0.4% O
and 8% O
Nominal
2N2
Nominal
2N2
Calibration gas flow ........................................................................... 5 scfh (141.6 L/hr)
Power supply ..................................................................................... 115 ±10% Vac at 50/60 Hz
Power requirement ............................................................................ 200 VA
Ambient operating temperature (Junction Box)................................. 300°F (149°C) [160°F (71°C) max for YEW
replacement] Approximate shipping weights:
18 in. (457 mm) package ....................................................... 55 lbs (24.97 kg)
3 ft (0.91 m) package ............................................................. 60 lbs (27.24 kg)
6 ft (1.83 m) package ............................................................. 65 lbs (29.51 kg)
9 ft (2.74 m) package ............................................................. 72 lbs (32.66 kg)
12 ft (3.66 m) package ........................................................... 78 lbs (35.38 kg)
1
All static performance characteristics are with operating variables constant.
2
Equipment ordered utilizing this document as reference will be supplied to the USA standard design. Customers requiring the EEC standard design should request the EEC documentation and utilize its ordering data. Temperatures over 1000°F (537°C) may affect the ease of field cell replaceability.
c. Snubber Diffusion Assembly
The snubber diffusion assembly protects the cell from heavy particles and isolates the cell from changes in temperature. The snubber assembly threads onto the cell and flange assembly. Pin spanner wrenches
An optional ceramic diffusor element and vee deflector shown in Figure 1-6, is avail­able. The ceramic diffusor assembly is also available in a flame arresting version to keep heat from the cell from igniting flue gases.
(probe disassembly kit 3535B42G01) are applied to holes in the diffusion element hub to remove or install the snubber assembly.
1-6 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
is noted in the calibration data sheet sup­plied with each probe.
DIFFUSION
ELEMENT
HUB
PIN
WRENCH
VEE
DEFLECTOR
21240024
Figure 1-6. Optional Ceramic Diffusor and Vee
Deflector Assembly
Systems that use an abrasive shield require a special diffusor assembly with a hub that is grooved to accept two dust seal gaskets. This special diffusor is available in both snubber and ceramic versions. See Probe Options, paragraph 1-8.
d. Cell – General
The components which make up the cell are machined to close tolerances and assem­bled with care to provide accurate oxygen measurements. Any replacement requires attention to detail and care in assembly to provide good results.
Every probe should be calibrated and checked after repair or replacement of cell, pad and wire, heater, and thermocouple, or after disassembly of the probe.
There are two styles of the heater/strut inner probe assembly. P/Ns 3D39441G06 through G10 identify the 44 V heater design. P/Ns 3D39441G01 through G05 and G11 through G15 identify the 115 V heater design. Both styles look the same. When replacing a strut, be sure the replacement has the same part number as the original. A stainless steel tag is attached to the inner probe assembly to identify the 115V design. Operating a probe on the wrong heater voltage can cause injury or severe damage to equipment.

1-5 PROBE INTERIOR

The inner probe assembly, Figure 1-7, consists of six main parts:
a. Ceramic support rod with four holes running
through the length. The holes serve as in­sulated paths for the cell signal wire and thermocouple wires.
b. A heater that is helically wrapped on a
quartz support cylinder and insulated.
Failure to follow the instructions in this manual could cause danger to personnel and equipment. Read and follow instructions in this manual
c. A chromel-alumel thermocouple which acts
as the sensing element for the temperature controller. (Not visible in Figure 1-7; located within ceramic support rod.)
carefully.
d. A platinum screen pad which forms electri-
The oxygen probe includes an inner elec­trode for the cell assembly. It consists of a platinum pad and a platinum/inconel com­posite wire which produces the cell constant offset voltage described in the Nernst equa­tion.
With this pad and wire, the constant will be between -10 and +15 mV. The cell constant
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-7
cal contact with the inner electrode of the electrochemical cell. (Not visible in Figure 1-7; located at end of ceramic support rod.) The pad is attached to an inconel/stainless steel wire which carries the signal to the terminal strip.
e. A V-strut assembly to give support to the
inner probe assembly.
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
HEATER
INSULATING
GASKET
V-STRUT
CERAMIC SUPPORT
ROD
Figure 1-7. Inner Probe Assembly
f. A tube to carry reference gas to the cell.
Refer to Section 4, Maintenance and Service, for repair procedures for probe components.
1-6 JUNCTION BOX
The junction box, Figure 1-8, is positioned at the external end of the probe and contains a termi­nal strip for electrical connections and fittings for reference and calibration gases. Fittings are for
0.250 inch stainless steel tubing on American units, and 6 mm tubing on European units. The calibration fitting has a seal cap which must re­main in place except during calibration. A tubing fitting is also supplied to be used with the calibration gas supply during calibration.
If the calibration gas bottles will be permanently hooked up to the probe, a manual block valve is required at the probe (between the calibration fitting and the gas line) to prevent condensation of flue gas down the calibration gas line.
During operation and calibration, reference gas is supplied through the reference gas fitting to the reference side of the cell. This gives the
REFERENCE
GAS TUBE
TAG
21220004
PROBE

JUNCTION BOX

COVER
TERMINAL
STRIP
CALIBRATION
GAS FITTING
REFERENCE GAS FITTING
21240029
Figure 1-8. Junction Box
system a known quantity of oxygen with which to compare the oxygen level in the process gas. Though ambient air can be used for this pur­pose, accuracy can only be assured if a reference gas air set is used.
During calibration, two gases of different known oxygen concentrations are injected one at a time through the calibration gas fitting. Stain­less steel tubing delivers this gas to the process side of the cell. In a healthy cell, the difference in oxygen pressure from the process side to the reference side of the cell will cause a millivolt output proportional to the difference in oxygen levels. The electronics unit can use the two mil­livolt outputs caused by the two calibration gases for either automatic or semi-automatic calibration.
1-8 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
Do not attempt to remove a process gas sample through either gas fitting. Hot gases from the process would damage gas hoses in the junction box.
1-7 CABLE ASSEMBLY
The Westinghouse and Rosemount systems use a 7-conductor cable to connect the probe to the electronics package. Standard length for this cable is 20 ft (6 m), but lengths up to 150 ft (45 m) are available. The seven conductors include one shielded pair of wires for the cell millivolt signal, one shielded pair of type K wires for the thermocouple, and three individual 16-gauge wires for the heater and for ground. The assembled conductors are wrapped by a
TM
Teflon
jacket and braided stainless steel
Instruction Manual
IB-106-300DR Rev. 3.3
TM
shield. The Teflon ing is suitable for high temperature use. All metal shields are isolated at the probe end and connect by drain wires to ground at the electronics.

1-8 PROBE OPTIONS

a. Abrasive Shield Assembly
The abrasive shield assembly, Figure 1-9, is a stainless-steel tube that surrounds the probe assembly. The shield protects the probe against particle abrasion and corro­sive condensations, provides a guide for ease of insertion, and acts as a probe posi­tion support, especially for longer length probes. The abrasive shield assembly uses a modified diffusion and vee deflector as­sembly fitted with dual dust seal packing.
and stainless steel jacket-
February 2000
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-9
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
0.187
0.187
2
1
B
A
o
15
03.584
03.554
o
90
ON INSIDE BREAK FOR SMOOTH
A
0.45 MIN
125
SKIN CUT FACE FOR 90
ROUNDED EDGE ON BOTH ENDS OF CHAMFER.
6.00
0.187
o
/
B
NOTES:
VIEW A
o
/
0.75 THRU 4 PLS.
EQ SP ON 4.75 B.C.
WELD ON BOTH SIDES WITH EXPANDING CHILL BLOCK.
2
BEFORE WELDING, BUTT ITEM 2 WITH ITEM 1 AS SHOWN.
o
/
Figure 1-9. Abrasive Shield Assembly
VIEW B
o
22.5
0.745 DIA ON A 7.50 DIA B.C. (REF)
0.755
P0009
1-10 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
P0010
Figure 1-10. Ceramic Diffusion/Dust Seal
Assembly
These modified diffusion and vee deflector assemblies are available in standard, Figure 1-10, and flame arrestor version, Figure 1-11.
b. Ceramic Diffusion Assembly
The ceramic diffusion assembly, Figure 1-12, is the traditional design for the probe. Used for over 25 years, the ceramic diffu­sion assembly provides a greater filter surface area for the probe.
19280010
Figure 1-12. Ceramic Diffusion Assembly
c. Flame Arrestor Diffusion Assembly
Where a high concentration of unburned fuel is present in the exhaust gases, a flame arrestor diffusion assembly, Figure 1-11 and Figure 1-13 is recommended.
The flame diffusion assembly includes a set of baffles between the cell and the stack gases. This keeps 1500°F (816°C) cell temperatures from igniting unburned fuel in the stack.
P0011
P0012
Figure 1-13. Flame Arrestor Diffusion Assembly
Figure 1-11. Flame Arrestor Diffusion/Dust Seal
Assembly
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-11
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
P0014
Figure 1-14. Snubber Diffusion/Dust Seal
Assembly
d. Snubber Diffusion/Dust Seal Assembly
The snubber diffusion/dust seal assembly, Figure 1-14, is used in applications where an abrasive shield is to be used with a snubber type diffusion element. The dust seal consists of two rings of packing to pre­vent abrasive dust from collecting inside the abrasive shield.
World Class 3000
can be employed. The bypass system uses an 18 inch (457 mm) or 3 foot (0.92 m) probe mounted externally on the stack or duct. The process or exhaust gases are di­rected out to the probe through a passive sampling system using inconel tubes. Flue gas flow induces the movement of gases into, through, and out of the bypass unit. The bypass arrangement does not require the use of aspiration air and the gas which flows past the probe is returned to the stack or duct.
The bypass probe package is normally used for process temperatures of 1300°F (704°C) to 2000°F (1094°C). A higher temperature version of the bypass provides for operation at temperatures up to 2500°F (1372°C). In this version the pick up tubes are made of a special high-temperature alloy.
Overall dimensions and mounting details of the American and European bypass sys­tems are shown in Figure 1-15.
f. Probe Mounting Jacket Options
e. Bypass Probe Options
For processes where the flue gas exceeds the maximum allowable temperature of 1300°F (704°C) a bypass sensor package
A probe mounting jacket option is available to allow the probe to operate at tempera­tures of up to 2000°F (1095°C). A separate instruction bulletin is available for this option.
1-12 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
C).
o
A - A
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
19290005
OF FLOW
DIRECTION
TO 980
o
POINT OR
F (650
DEW
o
Y AIR REQUIRED. FITTING
TO 1800
ARDS DIRECTION ONLY.
o
NW
AX. CLEAN DR
TION GAS REQUIRED.
AX. CALIBRA
CALIBRATION AND PURGE GAS CONNECTION. BITE TYPE FITTING (PARKER CPI)
REFERENCE AIR SUPPLY CONNECTION BITE TYPE FITTING (PARKER CPI) FOR
1.
NOTES:
10 SCFH AT 32 PSIG M
0.250 O.D. TUBING. 2 SCFH AT 3 PSIG M
IS LOCATED ON FAR SIDE.
2.
TURE RANGE 1200
TURE DOES NOT GO BELOW
PERA
PERA
YZER IN A VERTICALLY DOW
C.
o
ITH ANAL
FLUE GAS OPERATING TEM
LAG TO ENSURE GAS TEM
EXCEED 500
INSTALL W
4.
5.
3.
A
GROUP NOTE
ARENTHESES.
AL CONDUCTIVITY K
ETERS IN P
TION. THERM
ILLIM
ITH M
O INCH THK INSULA
ENDED TW M
ENSIONS ARE IN INCHES W
RECOM
EQUALS 0.5 FOR INSULATION.
DIM
7.
6.
01
A 3 FT (914.4) GAS TUBE PICK-UP
B 6 FT (1828.8) GAS TUBE PICK-UP
GASKET AND HARDWARE
06
FLANGE
CAL
GAS
VIEW
2
REF
GAS
1
4.026 (102.26)
04 03 02
I.D.
ELECT CABLE
C 9 FT (2743.2) GAS TUBE PICK-UP
BY CUSTOMER
OPTIONAL MOUNTING
ARRANGEMENT, 4 IN.
150# FLANGE SUPPLIED
05
HARDWARE
GASKET AND
06
30.62 (777.75) REF (4507C26G01)
62.50 (1587.5) REF (4507C26G03)
26.50 (673.1) REF (4507C26G01 AND G02)
37.00 (939.8) REF (4507C26G01)
73.00 (1854.2) REF (4507C26G02)
109.00 (2768.6) REF (4507C26G03)
6.0
(152.4) REF
6.50
(165.1) REF
DRAIN
STD 20 FT
(6.1 M) CABLE
A
4
6
3
FOR
INSERTION
FOR PROBE
27.31 (693.67)
AND REMOVAL
CLEARANCE REQ
3D3947G01
19.80 (502.92) SQUARE TO GAS FLOW.
BE ORIENTED SO THAT VEE SHIELD IS
IF EQUIPPED WITH THE OPTIONAL CERAMIC
DIFFUSOR ASSEMBLY, PROBE ASSEMBLY MUST
Figure 1-15. Bypass Probe Option (Sheet 1 of 3)
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-13
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
o
o
POINT OR
DEW
o
ARDS DIRECTION ONLY.
o
NW
YZER IN A VERTICALLY DOW
o
o
ITH ANAL
ARENTHESES.
ITH MILLIMETERS IN P
GROUP NOTE
A 3 FT (914.4) GAS TUBE PICK-UP
ELECT
B 6 FT (1828.8) GAS TUBE PICK-UP
C 9 FT (2743.2) GAS TUBE PICK-UP
CAL
GAS
2
CABLE
OMER
16
15
14
REF
1
13
12
11
GAS
A - A
VIEW
04
03
02
OF FLOW
DIRECTION
I.D.
4.026 (102.26)
19290006
PICKUP (3D390004G07 OR G08)
26.5 (673.1) ON 3 FT (914.4) OR 6 FT (1828.8)
62.5 (1587.5) ON 9 FT (2743.2) PICKUP (3D390004G09)
37.0 (939.8) ON 3 FT (914.4) PICKUP (3D390004G07)
73.0 (1854.2) ON 6 FT (1828.8) PICKUP (3D390004G08)
109.0 (2768.6) ON 9 FT (2743.2) PICKUP (3D390004G09)
CALIBRATION AND PURGE GAS CONNECTION. BITE TYPE FITTING (PARKER CPI)
10 SCFH AT 32 PSIG (220.64 kPa GAUGE) MAX. CALIBRATION GAS REQUIRED.
LAG TO ENSURE GAS TEMPERATURE DOES NOT GO BELOW
REFERENCE AIR SUPPLY CONNECTION BITE TYPE FITTING (PARKER CPI) FOR
0.250 O.D. TUBING. 2 SCFH AT 3 PSIG (20.69 kPa GAUGE) MAX. CLEAN DRY AIR
REQUIRED. FITTING IS LOCATED ON FAR SIDE.
1.
NOTES:
EXCEED 932 F (500 C).
3.
2.
A
A
STD CABLE
20 FT (6.1 M)
RECOMMENDED 2.0 INCH (50.8) THK INSULATION. THERMAL CONDUCTIVITY K EQUAL
0.5 FOR INSULATION.
DIMENSIONS ARE IN INCHES W
FLUE GAS OPERATING TEMPERATURE RANGE 1200 TO 1800 F (650 TO 980 C).
INSTALL W
6.
5.
4.
7.
01
INSERTION
FOR PROBE
27.31 (693.67) CLEARANCE REQ
FLANGE
AND REMOVAL
08
07
06
AND
GASKET
HARDWARE
4
Figure 1-15. Bypass Probe Option (Sheet 2 of 3)
ELD BY CUST W
10
09
05
FOR
19.80 (502.92)
3D3947G01
PLATE
ELDED W
TO STACK
6.50 (165.1) REF
DRAIN
3
6
ITH THE OPTIONAL CERAMIC
SQUARE TO GAS FLOW.
BE ORIENTED SO THAT VEE SHIELD IS
IF EQUIPPED W
DIFFUSOR ASSEMBLY, PROBE ASSEMBLY MUST
1-14 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
PARTS LIST
NOTE DEFINER
ITEM
01 WORLD CLASS
02 GAS PICK-UP DWG
03 GAS PICK-UP DWG
04 GAS PICK-UP DWG
05 ANALYZER
06 MTG HARDWARE DWG
07 MTG HARDWARE DWG
08 MTG HARDWARE DWG
09 ANALYZER
10 ANALYZER
11 GAS PICK-UP DWG
12 GAS PICK-UP DWG
13 GAS PICK-UP DWG
14 GAS PICK-UP DWG
15 GAS PICK-UP DWG
16 GAS PICK-UP DWG
PART NAME
3000
HOUSING
HOUSING
HOUSING
PARTS LIST UNITS: INCHES GROUP NOTE A B C A B C A B C
SIZE­REFERENCE INFORMATION
DWG
DWG
DWG
DWG
PROBE 18 IN.
3 FT
6 FT
9 FT
FOR WORLD CLASS
3000 ANSI
FOR ANSI FLANGE
FOR JIS
FOR DIN
FOR JIS
FOR DIN
FOR JIS 3 FT
FOR JIS 6 FT
FOR JIS 9 FT
FOR DIN 3 FT
FOR DIN 6 FT
FOR DIN 9 FT
Instruction Manual
IB-106-300DR Rev. 3.3
FLANGE
STYLE ANSI JIS DIN
MAIL CODE
PART NUMBER
OR REF DWG
ORDER FROM
MATRIX
4507C26G01
4507C26G02
4507C26G03
3D39005G01
3535B58G01
3535B58G03
3535B58G05
3D39005G02
3D39005G03
4507C26G04
4507C26G05
4507C26G06
4507C26G07
4507C26G08
4507C26G09
G01 G02 G03 G04 G05 G06 G07 G08 G09
1
111
111
3D39004 GROUP
1
1
111
111
1
1
1
February 2000
111
111
1
1
1
Extended Temperature By-Pass Arrangements (2400°°°°F; 1300°°°°C)
GROUP
PART NO.
CODE DESCRIPTION
1U0571 G01 3’ By-pass Package with ANSI bolt pattern.
1U0571 G02 6’ By-pass Package with ANSI bolt pattern.
1U0571 G03 9’ By-pass Package with ANSI bolt pattern.
1U0571 G04 3’ By-pass Package with JIS bolt pattern.
1U0571 G05 6’ By-pass Package with JIS bolt pattern.
1U0571 G06 9’ By-pass Package with JIS bolt pattern.
1U0571 G07 3’ By-pass Package with DIN bolt pattern.
1U0571 G08 6’ By-pass Package with DIN bolt pattern.
1U0571 G09 9’ By-pass Package with DIN bolt pattern.
Figure 1-15. Bypass Probe Option (Sheet 3 of 3)
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-15
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
1-16 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Before starting to install this equip­ment, read the "Safety instructions for the wiring and installation of this ap­paratus" at the front of this Instruction Bulletin. Failure to follow the safety instructions could result in serious injury or death.
Install all protective equipment covers and safety ground leads after installa­tion. Failure to install covers and ground leads could result in serious injury or death.
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
SECTION 2
INSTALLATION
b. Ensure that the area is clear of obstructions,
internal and external, that will interfere with installation. Allow adequate clearance for removal of probe (Figure 2-1).
The junction box temperature must not exceed 300°°°°F (149°°°°C). If used with the Yokogawa ZA8C electronics, junc­tion box temperature is limited to 160°°°°F (71°°°°C). To meet these limits, the user may need to make a heat shield or provide cooling air to the junction box. The temperature limit is due to the thermocouple "cold reference" device.
2-1 LOCATING THE OXYGEN ANALYZER
PROBE
The location of the probe in the stack or flue is most important for maximum accuracy in the oxygen analyzing process. The probe must be positioned so that the gas it measures is repre­sentative of the process. Best results are nor­mally obtained if the probe is positioned near the center of the duct (40 to 60% insertion). A point too near the edge or wall of the duct may not provide a representative sample because of the possibility of gas stratification. In addition, the sensing point should be selected so that the process gas temperature falls within a range of 50° to 1300°F (10° to 704°C). Figure 2-1 pro­vides mechanical installation references.
a. Check the flue or stack for holes and air
leakage. The presence of this condition will substantially affect the accuracy of the oxy­gen reading. Therefore, either make neces­sary repairs or install the probe upstream of any leakage.
2-2 SERVICE REQUIRED
a. Power input: 115 ±10% Vac single phase,
50 to 60 Hz, 3 amp minimum. (See label.)
b. Compressed air: 10 psig (69 kPa gauge)
minimum, 225 psig (1551 kPa gauge) maximum at 2 scfh (56.6 L/hr) maximum; supplied by one of the following (less than 40 parts-per-million total hydrocarbons). Regulator outlet pressure should be set at 5 psi (35 kPa).
1. Instrument air - clean, dry.
2. Bottled standard air with step-down regulator.
3. Bottled compressed gas mixture (20.95% oxygen in nitrogen).
4. Other equivalent clean, dry, oil-free air supply.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
FURNISHED IN - XIT
4512C34
ADAPTER & ACCESSORY
TO AMBIENT
INSULATE IF EXPOSED
WEATHER CONDITIONS
4512C35
4512C36
2.27 (58)
DIA MAX
ROSEMOUNT
5.85 (148.6)
DIM "A"
WITH STANDARD
REF GAS
CAL GAS
7.58 (192)
SNUBBER
DIFFUSER
ELEC
DIM "B" REMOVAL ENVELOPE
1/4 IN. TUBE
6 MM TUBE
6 MM TUBE
ANSI
DIN
JIS
1/2"
CONN
CONDUIT
GAS CAL
JIS
4512C18H01
6.10
(155)
1.88 (48)
0.59
GAS REF
(15)
AT THE BOTTOM
BOTTOM VIEW
INSTALL WITH CONNECTIONS
5.12
(130)
THESE FLAT FACED FLANGES ARE MANUFACTURED
TO ANSI, DIN, AND JIS BOLT PATTERNS AND ARE NOT
DIMENSIONS ARE IN INCHES WITH MILLIMETERS IN
PARENTHESES.
PRESSURE RATED.
2.
NOTES: 1.
24610001
3.80 (96.5)
DIFFUSER
FOR PROBE
WITH CERAMIC
4.90 (124.5)
ADD TO DIM "A" FOR
ARRESTOR
DIFFUSER AND FLAME
PROBE WITH CERAMIC
DIN
7.28
4512C19H01
ANSI
6.00
4512C17H01
(185)
(153)
0.71
0.75
(18)
(20)
5.71
4.75
(145)
(121)
DIM "B"
DIM "A"
27.3 (694)
16 (406)
45.3 (1151)
34 (864)
81.3 (2065)
117.3 (2980)
70 (1778)
106 (2692)
153.3 (3894)
142 (3607)
3535B18H02
ANSI
0.062 THK GASKET
3635B48H01
3535B45H01
JIS
DIN
ADD TO DIM "A"
TABLE I MOUNTING FLANGE
3FT
6FT
9FT
12 FT
PROCESS FLOW MUST
BE IN THIS DIRECTION
WITH RESPECT TO
DEFLECTOR 3534848G01
FLANGE
DIA.
HOLE
DIA.
(4) HOLES
EQ SP ON BC
TABLE II INSTALLATION/REMOVAL
18 IN.
PROBE
Figure 2-1. Probe Installation (Sheet 1 of 5)
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
7.50
7.48
BOLT
CIRCLE
0.75
(8) HOLES
DIAMETER
FLANGE
9.00 (153)
DIAMETER
TABLE IV. FLANGE SIZE
*
ANSI
7.48
0.75
0.945
9.25 (235)
9.25 (235)
*
*
JIS
DIN
FLANGES ARE MANUFACTURED TO ANSI,
*
5.7 (145)
14.5
(369)
DIM "D" REMOVAL ENVELOPE
PRESSURE RATED.
FLAT FACED. THESE FLANGES ARE NOT
DIN, AND JIS BOLT PATTERNS AND ARE
7.00
(178)
SEE TABLE IV
FOR FLANGE
SIZES
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
21190008
REF AND
CAL GAS
CONNECTOR
ELECTRICAL
CONNECTOR
CAL GAS LINES
CHECK VALVE FOR
INSULATE IF
CONDITIONS
EXPOSED TO
AMBIENT WEATHER
31.1
(790)
DIM "E"
45.3 (1151)
DIM "D"
27
(686)
DIM "C"
NOMINAL MEASUREMENTS
TABLE III. REMOVAL / INSTALLATION
3 FT
67.1
81.3
63
6 FT
(1704)
(2065)
(1600)
103.1
(2619)
117.3
(2980)
99
(2515)
9 FT
139.1
(3533)
153.3
(3894)
(P/N 3535B58G04 - JIS)
135
12 FT
(3429)
DIM "C"
0.06 THK GASKET FURNISHED
DIM "E" (WITH FLAME ARRESTOR)
(P/N 3535B58G06 - DIN)
(P/N 3535B58G02 - ANSI)
IN HARDWARE PACKAGE
Figure 2-1. Probe Installation (Sheet 2 of 5)
3.6
NOMINAL
(P/N 4843B38G02)
SNUBBER DIFFUSION/
DUST SEAL ASSEMBLY
DIMENSIONS ARE IN INCHES WITH
MILLIMETERS IN PARENTHESES.
NOTE:
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
JIS
(P/N 3535B58G04)
9.25
(235)
4.92
(125)
(200)
7.894
(M-20 x 2.5)
World Class 3000
16860021
8 THREADED HOLES
EQUALLY SPACED ON
D DIA B.C.
ABRASIVE SHIELD
FLANGE O.D.
DIN
(P/N 3535B58G06)
ANSI
(P/N 3535B58G02)
IN.
TABLE VI. ADAPTOR PLATE DIMENSIONS FOR ABRASIVE SHIELD
(mm)
DIMENSIONS
JIS
(P/N 4512C35G01)
9.25
9.00
"A"
6.50
(235)
(229)
(165)
3.94
(100)
4.75
(121)
"B"
DIA
(M-12 x 1.75)
7.48
(M-16 x 2)
7.50
0.625-11
"C"
"D"
THREAD
(130)
5.118
(190)
(191)
DIA
NOTE: PART NUMBERS FOR ADAPTOR PLATES INCLUDE
ATTACHING HARDWARE.
o
22.5
OUTSIDE WALL SURFACE.
CROSSHATCHED AREA IN 4
CORNERS MAY BE USED TO
FIELD BOLTING OF PLATE TO
PROVIDE ADDITIONAL HOLES FOR
AND 12 FT ABRASIVE SHIELD
ADAPTOR PLATE FOR 3, 6, 9,
INSTALLATIONS. SEE SHEET 2.
C
A
B
A
4 STUDS,
LOCKWASHERS AND
NUTS EQUALLY
SPACED ON
DIN
ANSI
TABLE V. ADAPTOR PLATE DIMENSIONS FOR PROBE
IN.
DIMENSIONS
7.5 (191)
(M-16 x 2)
(P/N 4512C36G01)
6.00
(153)
0.625-11
(P/N 4512C34G01)
"B"
"A"
(mm)
THREAD
(145)
5.708
4.75
(121)
"C"
DIA
A
o
45
A
C
ATTACHING HARDWARE.
2.500 DIA
NOTE: PART NUMBERS FOR ADAPTOR PLATES INCLUDE
C DIA B.C.
B
ADAPTOR PLATE FOR
STD WORLD CLASS 3000
PROBE INSTALLATION.
SEE SHEET 1.
Figure 2-1. Probe Installation (Sheet 3 of 5)
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
INSTALLATION FOR METAL
WALL STACK OR DUCT
CONSTRUCTION
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
INSTALLATION FOR MASONRY WALL STACK CONSTRUCTION
MTG HOLES
SHOWN ROTATED
o
OUT OF
22.5
TRUE POSITION
WELD OR BOLT MOUNTING
PLATE TO METAL WALL
OF STACK OR DUCT.
JOINT MUST BE AIRTIGHT.
0.50 (13)
3.75 (95) MIN DIA HOLE
IN WALL
STACK OR DUCT METAL WALL
BOLT ADAPTOR
PLATE TO OUTSIDE
WALL SURFACE
FIELD WELD
PIPE TO
ADAPTOR PLATE
MTG HOLES
SHOWN ROTATED
o
OUT OF
22.5
TRUE POSITION
JOINT MUST
BE AIRTIGHT
OUTSIDE WALL
SURFACE
NOTE: ALL MASONRY STACK WORK AND JOINTS EXCEPT
ADAPTOR PLATE NOT FURNISHED BY ROSEMOUNT.
ABRASIVE SHIELD MOUNTING
0.50 (13)
4.50 (114) O.D. REF
PIPE 4.00 SCHED 40 PIPE SLEEVE (NOT BY ROSEMOUNT) LENGTH BY CUSTOMER
MASONRY STACK WALL
WELD OR BOLT MOUNTING
PLATE TO METAL WALL
OF STACK OR DUCT.
JOINT MUST BE AIRTIGHT.
3.25 (82.5)
MIN DIA HOLE IN WALL
STACK OR DUCT METAL WALL
BOLT MOUNTING
PLATE TO OUTSIDE
WALL SURFACE
OUTSIDE WALL
PROBE MOUNTING
JOINT MUST
BE AIRTIGHT
SURFACE
NOTE: DIMENSIONS IN INCHES WITH
MILLIMETERS IN PARENTHESES.
FIELD WELD PIPE TO ADAPTOR PLATE
4.0 (102) O.D. REF
PIPE 3.5 SCHED 40 PIPE SLEEVE (NOT BY ROSEMOUNT) LENGTH BY CUSTOMER
MASONRY STACK WALL
P0038
Figure 2-1. Probe Installation (Sheet 4 of 5)
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
o
60 MAX.
o
30 MIN.
4.12
(105)
4.12
(105)
BRACE BARS (NOT BY ROSEMOUNT)
2.00 (51)
1.00 (25)
2 HOLES - 0.625 (16) DIA. FOR
0.50 (12) DIA. BOLT
1.00 (25) MAX.
0.375 (10)
NOTE: DIMENSIONS IN INCHES WITH
5.62
(143)
5.62
(143)
MILLIMETERS IN PARETHESES.
VERTICAL BRACE CLAMP ASSY. HORIZONTAL BRACE CLAMP ASSY.
(BOTH BRACE CLAMP ASSEMBLIES ARE THE SAME. INSTALLATION AND LOCATION OF CLAMP ASSEMBLIES AND BRACE BARS TO BE DONE IN FIELD.)
ABRASIVE SHIELD
BY ROSEMOUNT
}
NOTE: BRACING IS FOR VERTICAL AND HORIZONTAL PROBE INSTALLATION.
EXTERNAL BRACING REQUIRED FOR 9 FT AND 12 FT (2.75 M AND 3.66 M) PROBES AS SHOWN ABOVE.
Figure 2-1. Probe Installation (Sheet 5 of 5)
2-3 MECHANICAL INSTALLATION
a. Ensure that all components are available for
installation of the probe. If equipped with the optional ceramic diffusor, check to ensure that it is not damaged. Ensure that the sys­tem cable is the required length.
b. The probe may be installed intact as it is re-
ceived. It is recommended that you disas­semble the adapter plate for each installation.
36.00 (914)
P0039
NOTE
An abrasive shield is recommended for high velocity particulate in the flue stream (such as those in coal fired boilers, kilns, and recovery boilers). Vertical and horizontal brace clamps are provided for 9 ft and 12 ft (2.75 m and 3.66 m) probes to provide me­chanical support of the probe. Refer to Figure 2-1, sheet 5.
c. Weld or bolt adapter plate (Figure 2-1) onto
the duct. Existing stack penetration and mounting hardware may be revised, or new mounting flanges may be installed (see Figure 2-1, sheets 3 and 4).
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
d. If using the optional ceramic diffusor ele-
ment, the vee deflector must be correctly oriented. Before inserting the probe, check the direction of gas flow in the duct. Orient the vee deflector on the probe so that the apex points upstream toward the flow (Fig­ure 2-2). This may be done by loosening the setscrews, and rotating the vee deflector to the desired position. Retighten the setscrews.
GAS FLOW DIRECTION
VEE
APEX
FILTER
DEFLECTOR
DIFFUSION
ELEMENT
SETSCREW
VEE
DEFLECTOR
e. In horizontal installations, the probe cover
should be oriented so the system cable drops vertically from the probe cover. In a vertical installation, the system cable can be oriented in any direction.
f. If the system has an abrasive shield, check
the diffusion element dust seal packings. The joints in the two packings must be staggered 180°. Also, make sure that the packings are in the hub grooves as the probe slides into the 15° forcing cone in the abrasive shield.
g. Insert the probe through the opening in the
mounting flange and bolt the unit to the flange When probe lengths selected are 9 or 12 ft (2.74 or 3.66 m), special brackets are supplied to provide additional support for the probe inside the flue or stack. See Figure 2-1, sheet 5.
2-4 ELECTRICAL INSTALLATION
Figure 2-3 shows the junction box wiring for the direct replacement probe including the 115 V hookup. Consult the electronics manual for the connections to power and the existing signal conditioning electronics.
P0017
Figure 2-2. Orienting the Optional Vee Deflector
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
NOTE: FOR YOKOGAWA (YEW) ZA8C,
115V
HEATER
}
YE CHROMEL
RD ALUMEL
OR CELL +VE
BL CELL -VE
1234567
BL
OR
JUNCTION BOX
WIRING
Upon completing installation, make sure that the probe is turned on and operating prior to firing up the combustion process. Damage can result from having a cold probe exposed to the process gases.
YE
RD
BK
BK
GN
WHT
8
GN
WHT
Figure 2-3. Wiring the Direct Replacement Probe
NOTE
!
PROBE
REFER TO SECTION 3-6.
TEST GAS INPUT
AT 2-3 L/MIN
(4-7 SCFH)
CHECK
VALV E
REF. GAS SUPPLY
INPUT AIR AT
1 L/MIN (2 SCFH)
19290007
During outages, and if possible, leave all probes running to prevent condensation and prema­ture aging from thermal cycling.
If the ducts will be washed down during outage, MAKE SURE to power down the probes and remove them from the wash area.
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
3
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
SECTION 3
STARTUP
3-1 OVERVIEW
This section covers the setup procedures for the World Class 3000 Oxygen Analyzer. The DR probe can be used with several electronics packages including: Models 218, 218A, 225, 132 (analog and digital), TC200, and ZA8C.
Install all protective equipment covers and safety ground leads after setup. Failure to replace covers and ground leads could result in serious injury or death.
For Models 218, 225, and 132 (analog), refer to paragraph 3-2. Additional information can be found in IB-106-101.
For Model 218A, refer to paragraph 3-3. Addi­tional information can be found in IB-106-101A.
For Model TC200, refer to paragraph 3-4. Addi­tional information can be found in IB-107-020.
d. Attach voltmeter with positive (+) lead on
TP1 and negative (-) on either Pin 2 or 7.
NOTE
The voltage given above is for an am­bient (machinery space) temperature of 77°F (25°C). For each degree of am­bient temperature above or below 77°F (25°C), add or subtract 0.242 mV from the nominal. Example: at 87°F (31°C), the nominal voltage of -322.3 Mv should be increased (made less nega­tive) by 10 x 0.242 or 2.42 mV, making the adjusted nominal -319.9 Mv.
e. Adjust potentiometer M110-1 to read -322.3
millivolts nominal.
f. Remove voltmeter leads.
g. Remove jumper wire.
TP1
M110-1
For Model 132 (digital), refer to paragraph 3-5. Additional information can be found in IB-106­106A.
For Yokogawa Model ZA8C, refer to paragraph 3-6.
3-2 MODELS 218, 225, AND 132 (ANALOG)
ELECTRONICS SETUP
Before beginning operation, it is important that the probe heater setpoint of the existing elec­tronics be changed to support the World Class 3000 probe. The setpoint adjustment procedure required for Models 218, 225, and 132 analog electronics is as follows:
a. Open electronics enclosure.
b. On temperature controller card, Figure 3-1,
connect jumper wire from TP3 to either Pin 2 or Pin 7.
c. Set voltmeter to read DC millivolts (MV).
TP3
PIN 2 (PSC)
Figure 3-1. Temperature Controller Card
23 567 1012
PIN 7 (PSC)
Calibration Points
P0019
Rosemount Analytical Inc. A Division of Emerson Process Management Startup 3-1
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000

3-3 MODEL 218A ELECTRONICS SETUP

Before beginning operation, it is important that the probe heater setpoint of the existing elec­tronics be changed to support the World Class 3000 probe. To convert the Model 218A Digital Electronics Package for use with the DR probe, an EPROM change is necessary. Remove Main PCB and check back of board to identify unit as G02 or G04. The replacement EPROM needed is as identified below:
G02 G04
United States 1M03192G01 1M02982G01 United Kingdom 1M03192G02 1M02982G02 Germany 1M03192G03 1M02982G03 France 1M03192G04 1M02982G04 Italy 1M03192G05 1M02982G05
NOTE
The replacement EPROM when using a multiprobe averager unit is 1M02982G10.
To replace the EPROM, proceed as follows:
a. Shut off and lock out power to the electron-
ics package.
b. Open electronics enclosure.
c. On the Main PCB, Figure 3-2, locate and
remove old EPROM.
d. Replace with new EPROM.
e. Close electronics enclosure and power up
system.
EPROM
MAIN PCB
Figure 3-2. Main PCB (Model 218A) EPROM Replacement
3-2 Startup Rosemount Analytical Inc. A Division of Emerson Process Management
P0020
World Class 3000
3
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
3-4 MODEL TC200 VERITRIM ELECTRONICS
SETUP
Before beginning operation, it is important that the probe heater setpoint of the existing elec­tronics be changed to support the World Class 3000 Probe. To convert the Model TC200 Elec­tronics Package for use with the DR probe, an EPROM change is necessary. The replacement EPROM needed is part number 1M03154G02.
a. EPROM Replacement
To replace the EPROM, proceed as follows:
1. Shut off and lock out power to the electronics package.
2. Open electronics enclosure.
3. On the main PCB, Figure 3-3, locate and remove old EPROMs U11 and U12.
b. Heater SetPoint Adjustment
EPROMS
U11 AND U12
4. Replace with new EPROMs (part num­ber 1M03154G02) being careful to in­stall U11 and U12 in their proper locations.
5. Close electronics enclosure and power up system.
The adjustment procedure required for the Model TC200 Electronics Package is as follows:
1. Open keylocked enclosure to access membrane keyboard.
2. Put controller in PAR (parameter) mode by depressing "LOCK" "▲"
"%O
" "INC" "ACK" in sequence.
2
3. Depress "ACK" pushbutton to clear display.
DIS A/M
U12
U11
Figure 3-3. Main PCB (Model TC200) EPROM Replacement
P0021
Rosemount Analytical Inc. A Division of Emerson Process Management Startup 3-3
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
4. Press "NUM" pushbutton.
5. Using "INC", "DEC" buttons, display parameter 125.
6. Press "VAL" button.
7. Using "INC", "DEC" buttons, change parameter 125 value to 15.4.
8. Press "ENT" to save new value.
3-5 MODEL 132 DIGITAL
ELECTRONICS SETUP
Before beginning operation, it is important that the probe heater setpoint of the existing elec­tronics be changed to support the World Class 3000 Probe. To convert the Model 132 Digital Electronics Package for use with the DR probe, an EPROM change is necessary. The replace­ment EPROM needed is as identified below:
United States IM03222G01 United Kingdom Germany France Italy
EPROM
MAIN PCB
To replace the EPROM, proceed as follows:
a. Shut off and lock out power to the
electronics package.
b. Open electronics enclosure.
c. On the Main PCB, Figure 3-4, locate and
remove old EPROM.
d. Replace with new EPROM.
e. Close electronics enclosure and power up
system.
3-6 THE YOKOGAWA ZA8C AND AV8C CON-
VERTER ELECTRONICS SETUP
The DR probe can be wired to work with the Yokogawa® ZA8C and AV8C Converters.
Connect the cabling from the ZA8C or AV8C terminal to the probe terminal in the junction box as shown in Figure 3-5.
TX2
C34
M55
CR18
R8
P61
U7
C44
U2
C48
U1
C37
LM8
R38
03
TP1
R27
XT1
C26
C27
R12
C38
U6
R38
C29
C28
U4
C30
04
C29
R26
P25
P36
CR25
C13
D12
J1
P28
O18
LED1
08
R63
R62
R54
U9
1
R41
R48
RP1
C46
TP2
C24
D14
C21
C23
D13
C12
C14
D1
C16
D2
CR17
R48
C39
20
C31
U10
C43
U3
C42
R53
U13
C22
C13
C10
C41
R42
CR15
CR16
CR24
R36
CR13
CR14
CR12
CR11
R37
BR1
C30
R44
R49
R47
U14
R43
R38
R86
CR18
R68
09
P29
P22
R64
C86
C10
U12
R61
C16
U2
C32
TR1
TX1
C38
R56
CR19
C36
C11
R46
R48
P62
C7
C9
C8
U11
R6
C48
C4
C3 C8
C47
C1
C2
C3
R9
2001A
F2
2001A
CK1
F1
C7
I
R58
R57
R59
P68
U19
U17
R18
R16
R17
U16
R11
R18
CR3
CR4
U18
R4
R5
CR1
CR2
08
A
28
28
R16
28
C46
CR7
CR6
R19
CR8
CR5
28
28
LK2
28
R7
R6
R14
R23
R28
LK1
CR21
CR20
28
28
C17
R2
R1
R21
R13
R24
CR23
CR22
5A
F3
5A
F4
FIL1
R1
CR9
R81
ALARM
MTR-T/C CELL SKG STK-T/C ML-O/P
T83
C11
T82
OFF
N
L
HEATER
E
E
N
MAIN INPUT
L
T81
SW1
MOV1
MOV2
ON
P0022
Figure 3-4. Main PCB (Model 132) EPROM Replacement
3-4 Startup Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
3
ZA8C
AV8C
CONVERTER
OR
WORLD CLASS 3000
REPLACEMENT PROBE
WITH ADAPTER FLANGE
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
STACK WALL
TEST GAS
IF AVAILABLE
CELL
THERMOCOUPLE
COLD
JUNCTION
HEATER AND SIGNAL CABLE
REFERENCE AIR
IF AVAILABLE
SIGNAL
CABLE
ZA8C
OR
AV8C
TERMINAL
3+
}
4–
5+
}
6–
7+
}
8–
EXISTING
MOUNTING
COLD
JUNCTION
DEVICE
+–
GN OR YE RD BK BK GN
12345678
SHIELD
PROBE
PROBE TERMINAL
32
ZA8C OR AV8C
NOTES: 1. HEATER TEMPERATURE SET TO 1380
GND SHIELD
THE GREATER MASS OF THE WORLD CLASS DIRECT REPLACEMENT PROBE REQUIRES
2. LONGER TIME TO HEAT UP. UPON STARTUP, THE YOKOGAWA ELECTRONICS MAY INDICATE AN ERROR BECAUSE THE PROBE HAS NOT REACHED TEMPERATURE SETPOINT
IN THE NORMAL TIME. REMOVE POWER FROM THE YOKOGAWA ELECTRONICS OR PROBE MODULE TO CLEAR THE ERROR, AND RESTORE POWER. THIS PROCEDURE MAY HAVE TO BE REPEATED A COUPLE OF TIMES BEFORE PROBE OPERATING TEMPERATURE IS REACHED.
PROBE
HEATER
16
17
o
F (750oC)
624023
Figure 3-5. DR Probe Wired to the ZA8C or A V8C Converter
Rosemount Analytical Inc. A Division of Emerson Process Management Startup 3-5
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
3-6 Startup Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
4
IB-106-300DR Rev. 3.3
World Class 3000
SECTION 4
MAINTENANCE AND SERVICE
NOTE
!
Upon completing installation, make sure that the probe is turned on and operating prior to firing up the combustion process. Damage can result from having a cold probe exposed to the process gases.
During outages, and if possible, leave all probes running to prevent condensation and prema­ture aging from thermal cycling.
If the ducts will be washed down during outage, MAKE SURE to power down the probes and remove them from the wash area.
February 2000
4-1 OVERVIEW
This section describes routine maintenance of the oxygen analyzer probe. Spare parts referred to are avail-able from Rosemount. Probe disas­sembly kit 3535B42G01 contains the required spanner and hex wrenches. Refer to Section 7 for part numbers and ordering information.
Install all protective equipment covers and safety ground leads after equip­ment repair or service. Failure to in­stall covers and ground leads could result in serious injury or death.
4-2 PROBE CALIBRATION
The oxygen analyzer system should be cali­brated when commissioned. Under normal cir­cumstances the probe will not require frequent calibration. When calibration is required, follow the procedure described in the Instruction Bulle­tin applicable to your electronics package.
The cell replacement kit contains a cell and flange assembly, corrugated seal, setscrews, socket head cap screws, and anti-seize com­pound. Items are carefully packaged to preserve precise surface finishes. Do not remove items from packaging until they are ready to be used. Spanner wrenches and hex wrenches needed for this procedure are part of an available spe­cial tools kit, Table 7-1.
Wear heat resistant gloves and cloth­ing to remove probe from stack. Nor­mal operating temperatures of diffusor and vee deflector are approximately 600°°°° to 800°°°°F (316°°°° to 427°°°°C). They could cause severe burns.
Disconnect and lock out power before working on any electrical components. There is voltage up to 115 Vac.

4-3 CELL REPLACEMENT

This paragraph covers oxygen sensing cell re­placement. Do not attempt to replace the cell until all other possibilities for poor performance have been considered. If cell replacement is needed, order cell replacement kit, Table 7-1.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-1
Do not remove cell unless it is certain that replacement is needed. Removal may damage cell and platinum pad. Go through complete troubleshooting procedure to make sure cell needs re­placement before removing it.
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
BOMB TAIL
CONNECTOR
CELL EXTENSION
WIRE (ORANGE)
INCONEL
CELL WIRE
(CLEAR
SLEEVING)
CALIBRATION
GAS FITTING
CABLE
HEATER
WIRES
(BLACK
SLEEVING)
THERMOCOUPLE -
(RED ALUMEL)
(YELLOW CHROMEL)
REFERENCE GAS FITTING
THERMOCOUPLE +
PROBE JUNCTION
BOX COVER
21220001
Figure 4-1. Cell Wiring Connection
a. Disconnect and lock out power to electron-
ics. Shut off and disconnect reference gas and calibration gas supplies from junction box, Figure 4-1. Wearing heat resistant gloves and clothing, remove probe assem­bly from stack carefully and allow to cool to room temperature. Do not attempt to work on unit until it has cooled to a comfortable working temperature.
b. If the probe uses the standard diffusion
element, use a spanner wrench to remove the diffusion element.
c. If equipped with the optional ceramic diffu-
sor assembly, remove and discard setscrews, Figure 4-2, and remove vee de­flector. Use spanner wrenches from probe disassembly kit, Table 7-1, to turn hub free from retainer. Inspect diffusion element. If damaged, replace element.
d. Loosen four socket head cap screws from
the cell and flange assembly and remove the assembly and the corrugated seal. The cell flange has a notch which may be used to gently pry the flange away from the probe. Note that the contact pad inside the probe will sometimes fuse to the oxygen sensing cell. If the cell is fused to the con­tact pad, push the cell assembly back into the probe (against spring pressure), and quickly twist the cell assembly. The cell and contact pad should separate. If the contact pad stays fused to the cell, a new con­tact/thermocouple assembly must be in­stalled. Disconnect the cell and the thermocouple wires at the probe junction box, and withdraw the cell with the wires still attached (see paragraph 4-5).
4-2 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
4
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
PIN
WRENCH
OPTIONAL CERAMIC
DIFFUSION ELEMENT
SETSCREW
FILLET
VEE
DEFLECTOR
21240026
CEMENT
PORT
RETAINER
HUB
CEMENT
Figure 4-2. Removal of Optional Ceramic Diffusor
and Vee Deflector
e. If contact assembly is damaged, replace
contact and thermocouple according to paragraph 4-5, Replacement of Contact and Thermocouple Assembly.
f. Remove and discard corrugated seal. Clean
mating faces of probe tube and retainer. Remove burrs and raised surfaces with block of wood and crocus cloth. Clean threads on retainer and hub.
g. Rub a small amount of anti-seize on both
sides of corrugated seal.
h. Assemble cell and flange assembly, corru-
gated seal, and probe tube. Make sure the calibration tube lines up with the calibration gas passage in each component. Apply a small amount of anti-seize compound to screw threads and use screws to secure assembly. Torque to 35 in-lbs (4 N·m).
i. Apply anti-seize compound to threads of
cell assembly, hub, and setscrews. Reinstall hub on cell assembly. Using pin spanner wrenches, torque to 10 ft-lbs (14 N·m). If applicable, reinstall vee deflector, orienting apex toward gas flow. Secure with setscrews and anti-seize compound. Torque to 25 in-lbs (2.8 N·m).
PROBE TUBE
(NOT INCLUDED
IN KIT)
CORRUGATED
SEAL
CELL AND
FLANGE
ASSEMBLY
SOCKET HEAD
CAP SCREWS
CALIBRATION GAS
PASSAGE
Figure 4-3. Cell Replacement Kit
j. On systems equipped with an abrasive
shield, install dust seal gaskets, with joints 180° apart.
k. Reinstall probe and gasket on stack flange.
If there is an abrasive shield in the stack, make sure dust seal gaskets are in place as they enter 15° reducing cone.
l. Turn power on to electronics and monitor
thermocouple output. It should stabilize at
29.3 ±0.2 mV. Set reference air flow at 2 scfh (56.6 L/hr). After probe stabilizes, cali­brate probe per Instruction Bulletin applica­ble to your electronics package. If new components have been installed, repeat calibration after 24 hours of operation.
4-4 OPTIONAL CERAMIC DIFFUSION
ELEMENT REPLACEMENT
a. General
The optional ceramic diffusion element protects the cell from particles in process gases. It does not normally need to be re­placed because the vee deflector protects it from particulate erosion. In severe environ­ments the filter may be broken or subject to excessive erosion. Examine the diffusion element whenever removing the probe for any purpose. Replace if damaged.
21240009
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-3
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
Damage to the diffusion element may be­come apparent during calibration. Compare probe response with previous response. A broken diffusion element will cause a slower response to calibration gas.
Hex wrenches needed to remove setscrews and socket head screws in the following procedure are available as part of a special tool kit, Table 7-1.
Wear heat resistant gloves and cloth­ing to remove probe from stack. Nor­mal operating temperatures of diffusor and vee deflector are approximately 600°°°° to 800°°°°F (300°°°° to 425°°°°C). These high temperatures can cause severe burns.
Disconnect and lock out power before working on any electrical component. There is voltage up to 115 Vac.
with M-6 x 6 stainless setscrews coated with anti-seize compound.
3. On systems equipped with abrasive shield, remove dual dust seal gaskets.
4. Use spanner wrenches from special tool kit, Table 7-1, to turn hub free from retainer.
5. Put hub in vise. Break out old diffusion element with chisel along cement line and 3/8 inch (9.5 mm) pin punch through cement port.
6. Break out remaining diffusion element by tapping lightly around hub with hammer. Clean grooves with pointed tool if necessary.
7. Replace diffusion element, using re­placement kit listed in Table 7-1. This consists of a diffusion element, ce­ment, setscrews, anti-seize compound and instructions.
Do not remove the cell unless it is cer­tain that replacement is necessary. Cell cannot be removed for inspection without damaging it. Refer to para­graph 4-3, Cell Replacement.
b. Replacement Procedure.
1. Shut off power to electronics. Discon­nect cable conductors and remove ca­ble, Figure 4-1. Shut off and disconnect reference gas and calibration gas sup­plies from probe head. Wearing heat resistant gloves and clothing, carefully remove probe assembly from stack and allow to cool to room temperature. Do not attempt to work on unit until it has cooled to a comfortable working temperature.
2. Loosen setscrews, Figure 4-2, using hex wrench from special tools kit, Table 7-1, and remove vee deflector. Inspect setscrews. If damaged, replace
8. Test fit replacement element to be sure seat is clean.
Do not get cement on diffusion ele­ment except where it touches the hub. Any cement on ceramic element blocks airflow through element. Wip­ing wet cement off of ceramic only forces cement into pores.
9. Thoroughly mix cement and insert tip of squeeze bottle into cement port. Tilt bottle and squeeze while simultane­ously turning diffusion element into seat. Do not get any cement on upper part of diffusion element. Ensure com­plete penetration of cement around three grooves in hub. Cement should extrude from opposite hole. Wipe ex­cess material back into holes and wipe top fillet of cement to form a uniform fillet. (A Q-Tip is useful for this.) Clean any excess cement from hub with water.
4-4 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
4
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
10. Allow filter to dry at room temperature overnight or 1 to 2 hours at 200°F (93°C).
11. Wipe a heavy layer of anti-seize com­pound onto the threads and mating surfaces of the diffusion hub and re­tainer.
12. Assemble retainer and diffusion hub with two pin spanner wrenches. Torque to 10 ft-lbs (14 N·m).
13. On systems equipped with abrasive shield, install dust seal gaskets with joints 180° apart.
14. Reinstall vee deflector, orienting apex toward gas flow. Apply anti-seize com­pound to setscrews, and tighten with hex wrench.
15. Reinstall probe on stack flange.
16. Turn power on to electronics and monitor thermocouple output. It should stabilize at 29.3 ±0.2 mV. Calibrate probe per Instruction Bulletin applica­ble to your electronics package.
4-5 REPLACEMENT OF CONTACT AND
THERMOCOUPLE ASSEMBLY
a. Disconnect and lock out power to electron-
ics. Using heat resistant gloves and cloth­ing, remove junction box cover. Squeezing tabs on hose clamps, remove hoses from junction box, Figure 4-4. Remove four screws in corners of the junction box. Pull junction box and inner probe assembly free from probe tube. Set on bench and allow to cool to room temperature.
b. Disconnect cell extension wire (orange),
thermocouple wire (red alumel), and ther­mocouple wire (yellow chromel) by cutting bomb tail connections from the terminal strip, Figure 4-1.
c. Remove two screws, Figure 4-4, lockwash-
ers, and flat washers that connect junction box to inner probe assembly. Pull heater, V-strut and backplate assembly away from junction box. Inspect all O-rings and insu­lating gasket; replace if worn or damaged.
PROBE JUNCTION BOX TO HEATER,
STRUT, AND BACKPLATE
ASSEMBLY SCREW
PROBE JUNCTION BOX
INNER PROBE ASSEMBLY
TO PROBE TUBE SCREWS
PROBE
JUNCTION
BOX
HOSE
CLAMP
Use heat resistant gloves and clothing when removing junction box and inner probe assembly. Do not attempt to
HOSE
work on these components until they have cooled to room temperature. Probe components can be as hot as 800°°°°F (427°°°°C), and can cause severe burns.
Disconnect and lock out power before
21240027
working on any electrical components. There is voltage up to 115 Vac.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-5
Figure 4-4. Junction Box Mechanical Connections
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
HEATER SCREWS
(NOT SHOWN)
HEATER
CERAMIC ROD
35900001
MOUNTING SCREW
(REAR VIEW)
CONTACT AND
THERMOCOUPLER
ASSEMBLY
SPRING
INSULATING
GASKET
CLIP
V-STRUT
Figure 4-5. Inner Probe Replacement (Heater,
V-Strut, and Backplate Assembly)
d. Use a pencil to mark locations of spring clip
on ceramic rod, Figure 4-5.
e. Pry or squeeze tabs on spring clips, and
pull contact and thermocouple assembly out of probe assembly. Retain spring clips and spring; replace if damaged.
Be very careful when handling contact and thermocouple assembly. The ce­ramic rod in this assembly is fragile.
f. While very carefully handling new contact
and thermocouple assembly, lay old as­sembly next to new one. Transfer pencil marks to new rod.
g. Note wire lengths of old assembly as an aid
for trimming new lengths in step (j). Trim­ming of wires will not always be necessary. Throw away old contact and thermocouple assembly.
h. Carefully guide new contact and thermo-
couple assembly through V-strut assembly leaf spring (4, Figure 4-6), spring (9), spring clip (10) (held open by squeezing tabs), and tube supports (11, 13) until spring clip reaches pencil mark.
i. Reinstall insulating gasket on backplate, re-
place two screws, O-rings, lockwashers and flat washers connecting junction box to in­ner probe assembly.
Do not trim new wiring shorter than existing (old) wiring. Excessive wire trim will prevent connections from being properly made and will require new replacement kit.
j. Trim wires, if necessary, as noted in step
(g).
1
13
12
1. Heater Ceramic Rod
2. Contact and Thermocouple Assembly
3. Strut
4. Leaf Spring
3
2
11
1011
5. Ring Lug
6. Butt Connector
7. Extension Wire
8. Backplate
4
5
6
9
8
7
10. Spring Clip Assembly
11. Common Tube Support
12. Heater
13. Short Tube Support
9. Spring
4
8
21240011
Figure 4-6. Heater, Strut, and Backplate Assembly (Inner Probe Assembly)
4-6 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
4
k. Connect color coded wires to proper termi-
nals as shown in Figure 4-1. Rosemount recommends connecting the thermocouple wires directly to the terminal strip. This is because the junction of different metals at the wires and lugs and at the lugs and the terminal strip could act as additional ther­mocouple junctions. This could produce a voltage that would affect the thermocouple output signal.
Do not bend wires closer than 1/4 in. (6.4 mm) from end of ceramic rod. Dress wires so they do not touch sides of junction box.
l. Slide assembled junction box and inner
probe assembly into probe tube. To align calibration gas tube with corresponding hole in backplate (A, B, Figure 1-3), insert scriber through hole in backplate and into calibra­tion gas tube. Secure with screws. Reinstall hoses and probe cover.
m. Power up system. Monitor thermocouple
output. It should stabilize at setpoint mV ±0.2 mV. Recalibrate probe per Instruction Bulletin applicable to your electronics pack­age.
4-6 REPLACEMENT OF HEATER, V-STRUT
AND BACKPLATE ASSEMBLY (Inner ProbeAssembly; Includes Contact And Thermocouple Assembly).
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
There are two styles of the heater/strut inner probe assembly. P/Ns 3D39441G06 through G10 identify the 44 V heater design. P/Ns 3D39441G01 through G05 and G11 through G15 identify the 115 V heater design. Both styles look the same. When replacing a strut, be sure the replacement has the same part number as the original. A stainless steel tag is attached to the inner probe assembly to identify the 115V design. Operating a probe on the wrong heater voltage can cause injury or severe damage to equipment.
NOTE
This replacement may be done without removing the probe from the duct.
a. Disconnect and lock out power to electron-
ics. Using heat resistant gloves and cloth­ing, remove probe cover. Squeeze tabs on hose clamps and remove hoses from junc­tion box, Figure 4-4. Remove four screws and lockwashers (7 and 10, Figure 4-7) that hold junction box and inner probe assembly to probe tube. Pull junction box and inner probe assembly free from probe tube. Set on bench and allow to cool to room tem­perature.
b. Disconnect cell extension wire (orange),
thermocouple wire (red alumel), and ther-
Wear heat resistant gloves and cloth­ing when remove probe head and in­ner probe assembly. Do not attempt to work on these components until they have cooled to room temperature. Probe components can be as hot as 800°°°°F (427°°°°C). This can cause severe burns.
Disconnect and lock out power before working on any electrical components. There is voltage up to 115 Vac.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-7
mocouple wire (yellow chromel) by cutting bomb tail connections from the terminal strip, Figure 4-1.
c. Remove two screws, lockwashers, and flat
washers that connect junction box to inner probe assembly. Remove and discard inner probe assembly (heater, V-strut, and back­plate assembly). Replace with new inner probe assembly. Reinstall screws, lock­washers and flat washers.
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
124
World Class 3000
678 9
5
11
3
10
21240012
1. Snubber Diffusion Element
2. Socket Hd Cap Screw (0.25 in.-28 x 0.63 (16 mm)
3. Cell and Flange assembly
4. Corrugated Seal
5. Probe Tube Assembly
6. Gasket
7. Fillister Hd Screw
8. Cover Head Assembly
Figure 4-7. Oxygen Analyzer Probe, Cross-Sectional View
d. Connect color coded wires to proper termi-
nals as shown in Figure 4-1. Rosemount recommends connecting the thermocouple wires directly to the terminal strip. This is because the junction of different metals at the wires and lugs and at the lugs and the terminal strip could act as additional ther­mocouple junctions. This could produce a voltage that would affect the thermocouple output signal.
Do not bend wires closer than 1/4 in. (6.4 mm) from end of ceramic rod. Dress wires so they do not touch sides of junction box.
4.0 in. (102 mm) x 4.0 in. x 0.12 in. (3 mm)
8-32 x 0.5 in. (12.7 mm)
e. Slide assembled junction box and inner
f. Power up system. Monitor thermocouple
9. Hose Clamp
10. Lockwasher, #8 Split
11. Heater Strut Assembly
probe assembly into probe tube. To align calibration gas tube with corresponding hole in backplate (A, B, Figure 1-3), insert align­ing tool (included in probe disassembly kit, P/N 3535B42G01) through hole in back­plate and into calibration gas tube, while sliding the heater strut into the probe tube. Secure with screws. Reinstall hoses and probe cover.
output. It should stabilize at set point ±0.2 mV. Recalibrate probe per Instruction Bulletin applicable to your electronics package.
4-8 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
4
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
4-7 CALIBRATION AND REFERENCE GAS
LINES FOR HIGH TEMPERATURE ­CORROSIVE ENVIRONMENT OPERATION
A high temperature, corrosive environment kit is available when the probe is exposed to these types of operating conditions. The kit includes stainless steel tubing and teflon fittings for in­side the junction box. The kit part number is 4843B93G01.
a. Installation Procedure
Use heat resistant gloves and clothing when removing junction box and inner probe assembly. Do not attempt to work on these components until they have cooled to room temperature. Probe components can be as hot as 800°°°°F (427°°°°C). This can cause severe burns.
Disconnect and lock out power before working on any electrical components. There is voltage up to 115 Vac.
1. Disconnect and lock out power to digi­tal electronics. Using heat resistant gloves and clothing, remove probe cover. Squeezing tabs on hose clamps, remove hoses from junction box (Figure 4-4).
2. First install the stainless steel tubing on the fitting at the bottom of the junction box. Install the other end of the stain­less steel tube onto the tube going to the probe (Figure 4-8).
NOTE
If abrasive conditions of high ash con­tent and high velocity exist, an abra­sive shield is recommended. To balance out the wear on the shield, ro­tate the shield 90° every time the probe is powered down for service.
TEFLON
SLEEVES
STAINLESS
STEEL
TUBING
21240028
Do not use sealant when installing the stainless steel tubes. Gas samples
Figure 4-8. High Temperature – Corrosive
Environment Kit
may become contaminated.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-9
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
4-10 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
5
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
SECTION 5
TROUBLESHOOTING
5-1 OVERVIEW
Troubleshooting for the oxygen analyzer system is broken down to the main component level. In addition to the information in this section, instruction bulletins for individual models also discuss troubleshooting.
Install all protective equipment covers and safety ground leads after trouble­shooting. Failure to replace covers and ground leads could result in seri­ous injury or death.
For Models 218, 225, and 132 (analog), refer to IB-106-101.
For Model 218A, refer to IB-106-101A.
For Model TC200, refer to IB-107-020.
For Model 132 (digital), refer to IB-106-106A.
For Yokogawa (YEW) ZA8C Converter, refer to original converter documentation.

5-2 PROBE TROUBLESHOOTING

a. Probe Faults
Listed below are the five symptoms of probe failure.
1. The system does not respond to changes in the oxygen concentration.
2. The system responds to oxygen changes but does not give the correct indication.
3. The system does not give an accept­able indication of the value of the oxy­gen test gas being applied during calibration.
4. The probe takes a long time to return to flue gas value after the calibration gas is turned off.
5. The probe heater temperature is unstable.
b. Table 5-1 provides a guide to fault finding
for the above symptoms.
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 5-1
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
SYMPTOM CHECK FAULT REMEDY
1. No response to oxygen concentration change when:
World Class 3000
Table 5-1. Fault Finding
Heater is cold and TC mV output is less than setpoint
Heater is hot and T/C mV output is at set point ±0.2 mV
2. System responds to oxygen concentration changes but does not give correct indication
Thermocouple continuity Thermocouple failure Replace thermocouple or
return probe to Rosemount.
Heater cold resistance should be 67 ohm to 77 ohm
Triac O/P to heater Failure of electronics Check electronics package.
Recorder chart Recorder failure See Recorder Instruction
Cell mV input to electronics and cell mV at probe head
Heater failure Replace heater or return
probe to Rosemount.
Manual.
No cell mV at probe when test gas applied
Probe cell mV OK but no input to electronics
Cell mV satisfactory both at junction box and input to electronics - failure of elec­tronics
Replace cell or return probe to Rosemount.
Check out cable connec­tion.
Check electronics package.
Good response, with incorrect indication
3. Probe does not give accurate indicationof applied test gas
5-2 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Recorder or remote indicator
System calibration Calibration error Recalibrate system.
Probe mounting and condi­tion of duct
Cell mV input to electronics Failure of electronics Check electronics package.
Test gas input port Blocked port Clean port. If the flue gas is
Ceramic diffusion element Blocked port
Calibration error Recalibrate recorder or in-
dicator. Reference Re­corder Instruction Manual.
Air ingress into duct Stop air leaks or resite
probe.
condensing in the test gas line, insulate the back of the probe. Make sure that the test gas line is capped be­tween calibrations, or a check valve is installed.
Diffusion element cracked, broken, or missing
Replace diffusion element.
World Class 3000
5
SYMPTOM CHECK FAULT REMEDY
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
Table 5-1. Fault Finding (continued)
4. Probe takes a long time to return to flue gas value after calibration gas is turned off
5. Probe heater temperature unstable
Plugged diffusion element Plugged diffusion element Change diffusion element or
snubber diffusion element.
Proper voltage heater is installed
Wrong heater Change heater to proper
voltage.
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 5-3
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
5-4 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
6
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
SECTION 6
RETURN OF MATERIAL
6-1 If factory repair of defective equipment is re-
quired, proceed as follows:
a. Secure a return authorization number from
a Rosemount Analytical Sales Office or Representative before returning the equip­ment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be ac­cepted.
In no event will Rosemount be responsible for equipment returned without proper authorization and identification.
b. Carefully pack defective unit in a sturdy box
with sufficient shock absorbing material to insure that no additional damage will occur during shipping.
c. In a cover letter, describe completely:
1. The symptoms from which it was de­termined that the equipment is faulty.
2. The environment in which the equip­ment has been operating (housing, weather, vibration, dust, etc.).
3. Site from which equipment was re­moved.
5. Complete shipping instructions for re­turn of equipment.
6. Reference the return authorization number.
d. Enclose a cover letter and purchase order
and ship the defective equipment according to instructions provided in Rosemount Return Authorization, prepaid, to:
Rosemount Analytical Inc. RMR Department 1201 N. Main Street Orrville, Ohio 44667
If warranty service is requested, the defec­tive unit will be carefully inspected and tested at the factory. If failure was due to conditions listed in the standard Rosemount warranty, the defective unit will be repaired or replaced at Rosemount's option, and an operating unit will be returned to the cus­tomer in accordance with shipping instruc­tions furnished in the cover letter.
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.
4. Whether warranty or nonwarranty service is requested.
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 6-1
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
6-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
7
SECTION 7
REPLACEMENT PARTS
Table 7-1. Replacement Parts for Probe
FIGURE and
INDEX No. PART NUMBER DESCRIPTION
4-5 3D39441G01 4-5 3D39441G02 4-5 3D39441G03 4-5 3D39441G04 4-5 3D39441G05 4-5 3D39441G11
4-5 3D39441G12
4-5 3D39441G13
4-5 3D39441G14
4-5 3D39441G15
4-6, 2 3534B56G04 4-6, 2 3534B56G05 4-6, 2 3534B56G06 4-6, 2 3534B56G07 4-6, 2 3534B56G08 4-6, 2 4513C61G01
4-6, 2 4513C61G02
4-6, 2 4513C61G03
4-6, 2 4513C61G04
4-6, 2 4513C61G05
4-6, 2 3D39003G01 4-6, 2 3D39003G02 1-9 3D39003G07 1-9 3D39003G08 1-9 4847B61G01 Cell Replacement Kit, ANSI w/No lead wire 1-9 4847B61G02 Cell Replacement Kit, ANSI 18 in. (45.6 cm) 4-3 4847B61G03 Cell Replacement Kit, ANSI 3 ft (0.9 m) 4-3 4847B61G04 Cell Replacement Kit, ANSI 6 ft (1.8 m) 4-3 4847B61G05 Cell Replacement Kit, ANSI 9 ft (2.7 m) 4-3 4847B61G06 Cell Replacement Kit, ANSI 12 ft (3.6 m)
1
1
1
1
1
1
Heater, V-Strut, and Backplate Assembly (115 V), 18 in. (45.6 cm) Heater, V-Strut, and Backplate Assembly (115 V), 3 ft (0.9 m) Heater, V-Strut, and Backplate Assembly (115 V), 6 ft (1.8 m) Heater, V-Strut, and Backplate Assembly (115 V), 9 ft (2.7 m) Heater, V-Strut, and Backplate Assembly (115 V), 12 ft (3.6 m) Heater, V-Strut, and Backplate Assembly, 18 in. (45.6 cm)
(For use with YEW electronics)
1
Heater, V-Strut, and Backplate Assembly, 3 ft (0.9 m) (For use with YEW electronics)
1
Heater, V-Strut, and Backplate Assembly, 6 ft (1.8 m) (For use with YEW electronics)
1
Heater, V-Strut, and Backplate Assembly, 9 ft (2.7 m) (For use with YEW electronics)
1
Heater, V-Strut, and Backplate Assembly, 12 ft (3.6 m) (For use with YEW electronics)
2
2
2
2
2
2
Contact and Thermocouple Assembly, 18 in. (45.6 cm) Contact and Thermocouple Assembly, 3 ft (0.9 m) Contact and Thermocouple Assembly, 6 ft (1.8 m) Contact and Thermocouple Assembly, 9 ft (2.7 m) Contact and Thermocouple Assembly, 12 ft (3.6 m) Contact and Thermocouple Assembly, 18 in. (45.6 cm)
(For use with YEW electronics)
2
Contact and Thermocouple Assembly, 3 ft (0.9 m) (For use with YEW electronics)
2
Contact and Thermocouple Assembly, 6 ft (1.8 m) (For use with YEW electronics)
2
Contact and Thermocouple Assembly, 9 ft (2.7 m) (For use with YEW electronics)
2
Contact and Thermocouple Assembly, 12 ft (3.6 m) (For use with YEW electronics)
3
3
3
3
Abrasive Shield Assembly, 3 ft (0.9 m) Abrasive Shield Assembly, 6 ft (1.8 m) Abrasive Shield Assembly, 9 ft (2.7 m) Abrasive Shield Assembly, 12 ft (3.6 m)
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-1
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
Table 7-1. Replacement Parts for Probe (continued)
FIGURE and
INDEX No. PART NUMBER DESCRIPTION
4-3 4847B61G07 Cell Replacement Kit, JIS, w/No lead wire 4-3 4847B61G08 Cell Replacement Kit, JIS 18 in. (45.6 cm) 4-3 4847B61G09 Cell Replacement Kit, JIS 3 ft (0.9 m) 4-3 4847B61G10 Cell Replacement Kit, JIS 6 ft (1.8 m) 4-3 4847B61G11 Cell Replacement Kit, JIS 9 ft (2.7 m) 4-3 4847B61G12 Cell Replacement Kit, JIS 12 ft (3.6 m) 4-3 4847B61G13 Cell Replacement Kit, DIN w/No lead wire 4-3 4847B61G14 Cell Replacement Kit, DIN 18 in. (45.6 cm) 4-3 4847B61G15 Cell Replacement Kit, DIN 3 ft (0.9 m) 4-3 4847B61G16 Cell Replacement Kit, DIN 6 ft (1.8 m) 4-3 4847B61G17 Cell Replacement Kit, DIN 9 ft (2.7 m) 4-3 4847B61G18 Cell Replacement Kit, DIN 12 ft (3.6 m) 4-3 3535B42G01 Probe Disassembly Kit 1-13 3535B18G01 Ceramic Diffusor Assembly 1-11 3535B60G01
Ceramic Diffusor Dust Seal Hub Assembly
(For use with Abrasive Shield) 1-13 4841B03G02 Stainless Steel Diffusor Assembly 1-12 3535B63G01 Flame Arrestor Diffusor Dust Seal 1-14 3535B62G01 Flame Arrestor Diffusor 4-2 3535B48G01
Vee Deflector Assembly
(For use with standard ceramic or dust seal type diffusors) 4-2 6292A74G02 Diffusion Element Replacement Kit 2-1 1537B70G03 Horizontal and Vertical Brace Clamp Assembly, 9 and 12 foot
(2.7 and 3.6 m) probe 4-8 4843B93G01 High Temperature - Corrosive Environment Kit
4843B37G01 Snubber Diffusion Assembly
1-10 4843B38G01 Dust Seal/Snubber Diffusion Assembly
1
Heater, V-strut, and backplate assembly includes contact and thermocouple assembly.
2
Contact and thermocouple assembly includes platinum pad and inconel wire.
3
Abrasive shield assembly includes accessories necessary for its use and a mounting plate and gasket.
World Class 3000
7-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
8
This index is an alphabetized listing of parts, terms, and procedures having to do with the Haz­ardous Area Oxygen/Combustibles Transmitter. Every item listed in this index refers to a location in the manual by one or more page numbers.
Instruction Manual
IB-106-300DR Rev. 3.3
February 2000
SECTION 8
INDEX
A
Abrasive Shield, 1-7, 1-9, 2-8 Absolute Temperature, 1-2 Adapter Plate, 2-2 Ambient Temperature, 1-6 Arithmetic Constant, 1-2
B
Bypass Probe Option, 1-12
C
Calibration, 4-1 Cell Constant, 1-2 Check Valve, 1-3
D
Diffusion Element, 1-6, 1-12, 4-3 Diffusor Element, 1-6
E
Electrochemical Cell, 1-2 Electronic Packages, 1-1, 3-1, 4-1 EPROM, 3-2, 3-3, 3-4
J
Junction Box, 1-3, 1-8
M
Model 218A Digital Electronic Package, 3-2 Model 132 Digital Electronics Package, 3-4 Model 225 Electronics Package, 3-1 Model TC200 VERITRIM Electronics Package, 3-3 Models 218 Electronics Package, 3-1 Mounting Adapter Plate, 1-1
N
Nernst Equation, 1-2
O
Orsat Apparatus, 1-2 Oxygen Analyzer, 1-1
P
Probe Heater Setpoint, 3-3
R
Reference Gas, 1-3
F
Flame Arrestor, 1-11
G
Gas Stratification, 2-1
S
Sampling System, 1-1 Sensing Point, 2-1 Sensor Cell, 1-2, 1-4 Snubber, 1-12 Solenoid Valve, 1-3
H
Heater, 1-8, 3-1, 4-7, 5-2 Hub Grooves, 2-8
I
In Situ, 1-1, 1-2 Instrument Air, 1-2, 2-1
Rosemount Analytical Inc. A Division of Emerson Process Management Index 8-1
T
Thermocouple, 4-2, 4-5
V
Vee Deflector, 1-7, 2-8 V-Strut, 1-7, 4-6
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
W
Water Vapor, 1-2
Z
Zirconia Disc, 1-2
8-2 Index Rosemount Analytical Inc. A Division of Emerson Process Management

WARRANTY

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the com­mercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WAR­RANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DE­LAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WAR­RANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF AC­TION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDEN­TAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVE­NUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond Seller's direct control.
2461
3267/2-00
Instruction Manual
IB-106-300DR Rev. 3.3 February 2000
World Class 3000
World Class 3000 Probe
Pa rt n o. __ __ ____________ Se ria l n o._ ______________ Ord er no._______________
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com
ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd.
1 Pandan Crescent Singapore 128461 Republic of Singapore T 65-777-8211 F 65-777-0947
http://www.processanalytic.com
© Rosemount Analytical Inc. 2001
Fisher-Rosemount GmbH & Co.
Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209
EUROPE, MIDDLE EAST, AFRICA Fisher-Rosemount Ltd.
Heath Place Bognor Regis West Sussex PO22 9SH England T 44-1243-863121 F 44-1243-845354
LATIN AMERICA Fisher - Rosemount
Av. das Americas 3333 sala 1004 Rio de Janeiro, RJ Brazil 22631-003 T 55-21-2431-1882
Loading...