Rosemount 3300 and 5300 Series-Mounting a Centering Disc Manual Supplement

Rosemount 3300 and 5300 Series
Mounting Instruction Centering Disc
The mounting of a centering disc on a rigid single lead probe requires one hole (8 mm probe) or two holes (13 mm probe) at certain distances from the end of the lead. The drilling fixture included in your shipment should be used to make the hole(s) according to the figure below:
Figure 1 Drilling fixture to make hole in probe
After the hole has been drilled, the centering disc should be attached to the probe according to figures below:
Rigid Single Lead probe (8 mm)
Figure 2 Mounting procedure for centering disc on rigid single lead probe (8 mm; 4A)
NOTE that the washer should not be mounted if the thickness of the plate is 4 mm or if the disc material is C-276, Duplex 2205 or Alloy 400.
Art No. 9270 001-248 Rev. D
Rosemount 3300 and 5300 Series
Mounting Instruction Centering Disc
Rigid Single Lead probe (13 mm)
Figure 3 Mounting procedure for centering disc on rigid single lead probe (13 mm; 4B)
NOTE that the washer should not be mounted if the thickness of the plate is 4 mm or if the disc material is C-276, Duplex 2205 or Alloy 400.

Mounting a Centeri ng Disc on Flexible Lead Probes

The procedure for m ounting a centering disc on a flexible lead is described by figure below:
Figure 4 Mounting procedure for centering disc on flexi b le lead probe
Art No. 9270 001-248 Rev. D