The Rosemount system described in this document is the property of Emerson.
No part of the hardware, software, or documentation may be reproduced, transmitted, transcribed, stored in a retrieval system,
or translated into any language or computer language in any form or by any means without prior written permission of Emerson.
While great efforts have been made to ensure the accuracy and clarity of this document, Emerson assumes no responsibility
resulting from any omissions in this document or from misuse of the information obtained herein. The information in this
document has been carefully checked and is believed to be entirely reliable with all the necessary information included. Emerson
does not assume any liability arising out of the application or any use of any product or circuit described herein; neither does it
convey license under its patent rights or the rights of others.
WARNING
All individuals who have or will have responsibility for using, maintaining, or servicing the product must read this manual
thoroughly.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could
be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access
by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
NOTICE
The source and detector are not field-repairable due to the meticulous alignment and calibration of the sensors and the respective
circuits.
Do not attempt to modify or repair the internal circuits or change their settings, as this will impair the system's performance and
void the Emerson product warranty.
The Rosemount™ 936 WinHost is communication software for the Rosemount 936
detector. The program displays information (such as status, gain, address, signals, and
LEL) and makes it possible to change the detector’s configuration.
1.1Software overview
The Rosemount 936 WinHost software makes it possible to:
• Communicate with Rosemount 936 gas detectors
• Read status and parameters from the detector
• Record relevant detector data to a log file (S/N.txt)
• Calibrate the detector
1.2Minimal requirements
Following requirements are the minimum for operating this software:
• Pentium® 3 GHz
• Windows XP,7,8, or 10
• 2 GB of RAM
• 10 GB hard disk free space
• Isolated RS-485 Interface Card to be defined as COM1, COM2, COM3, or COM4; or
RS-232/RS-485 converter to connect to standard COM port
1.3Standard
EIA 485:
•
Electrical characteristics of enhanced Voltage Digital Interface Circuits.
Rosemount 9365
Product overviewReference Manual
April 202200809-0200-4036
6Emerson.com/Rosemount
Reference ManualInstallation
00809-0200-4036April 2022
2Installation
2.1Installing WinHost software
To load your computer with the Rosemount 935 WinHost configuration and diagnostic
software:
Procedure
1. Copy the Rosemount 936 Winhost installation files into the correct drive.
2. Start the Rosemount 936 WinHost software installation by running the setup.exe
file.
3. Follow the installation instructions.
4. Connect the detector unit to the RS-485 communications port (see Connecting the
detector to the computer).
5. Start the Rosemount 936 WinHost software with the specified COM port number as
a parameter (see Establish the COM port).
Rosemount 9367
InstallationReference Manual
April 202200809-0200-4036
8Emerson.com/Rosemount
Reference ManualGetting started
00809-0200-4036April 2022
3Getting started
3.1Connecting the detector to the computer
Before you can perform any configuration or diagnostic operation on a detector, you must
connect the computer to the detector using the harness cable provided.
3.1.1Connect the computer to a detector
Procedure
1. Connect one end of the USB cable to the computer USB port.
2. Connect the other end of the USB cable to the USB serial (RS-485) adapter.
3. Connect the serial port of the adapter to the harness cable.
3.1.2
3.1.3
3.1.4
Connect the detector to the harness cable
Procedure
1. Connect one side of the cable to detector terminal 5 for RS-485 (+).
2. Connect the other side of the cable to detector terminal 6 for RS-485 (-)
Connect a socket D-Type on the other side of the cable
Procedure
1. Connect RS-485 (+) to pin 2.
2. Connect RS-485 (-) to pin 1.
3. Connect RTN to pin 5.
Perform USB adapter setup
Procedure
1. Unscrew the cover of the USB adapter.
2. There are two options for setting up the jumpers:
Figure 3-1: USB adapter setup option 1
Rosemount 9369
Getting startedReference Manual
April 202200809-0200-4036
Figure 3-2: USB adapter setup option 2
A. Dip switch
B. On
C. Off
3. Close the USB adapter cover.
4. Connect the cable.
CAUTION
If using a different adapter than the one recommended, check that the D-connector
adapter wiring is similar to the wiring above (if not, adjust the cable wiring to fit the
desired adapter).
3.2Establish the COM port
Before using the software, you need to establish the number of the COM port.
To view the COM port used by the adapter:
Procedure
Select Start→Settings→Control→Panel→System→Hardware→Device Manager.
The COM port number is displayed. This is the COM port number you will use.
10Emerson.com/Rosemount
Reference ManualGetting started
00809-0200-4036April 2022
Figure 3-3: COM port number
A. Assigned COM port
Rosemount 93611
Getting startedReference Manual
April 202200809-0200-4036
3.3Run WinHost
Procedure
1. Open the Winhost software application.
The opening window will appear
After a few seconds, the opening window disappears and the communication setup
dialog box appears:The communication setup dialog box allows the user to select
the communication port number.
Figure 3-4: Communication setup dialog box
2. From the drop-down menu, select the assigned communication port number (see
Establish the COM port).
3. Click OK.
The main window appears.
12Emerson.com/Rosemount
Reference ManualOperating WinHost
00809-0200-4036April 2022
4Operating WinHost
4.1Main window
The main window monitors the detector.
Note
For Windows XP, in the case of poor communication, press F12 and wait until good
communication is achieved.
The main window is divided into the following areas:
1. Device information
2. Signal information
3. Device status
4. Left menu
4.1.1
Device information components
Table 4-1 describes the composition of the device information area in the main window.
Table 4-1: Device information components
ElementDescription
ModelThe detector’s model.
Serial numberThe detector’s serial number. Each detector has
a unique serial number.
4.1.2Signal information components
Table 4-2 describes the composition of the signal information area in the main window.
Table 4-2: Signal information components
ElementDescription
Signal 1Voltage signal of Sensor 1 after electronic
amplification (depends on gain).
ReferenceVoltage signal of reference after electronic
amplification.
Signal 2Voltage signal of Sensor 2 after electronic
amplification (depends on gain).
NQ Ratio 1Ratio1/QO1 (Normalized ratio 1). Equals 1 when
no gas is present and goes down when gas is
introduced.
Rosemount 93613
Operating WinHostReference Manual
April 202200809-0200-4036
Table 4-2: Signal information components (continued)
ElementDescription
NQ Ratio 2Ratio2/QO2 (Normalized ratio 2) Equals 2 when
no gas is present and goes down when gas is
introduced.
4.1.3Device status components
Table 4-3 describes the composition of the device status components in the main window.
Table 4-3: Device status components
ElementDescription
StatusThe detector’s current operational status.
GainElectronic amplification.
Up and down arrows
4.1.4
AddressThe address currently viewed by the software. (You can change the
current address by using the up and down arrows).
Indicates the status of the communication between the detector and the
computer.
Change addressChanges address to that selected in dropdown menu
Note
Do not click the address finding button when more than one detector is connected.
Toolbar
Table 4-4: Toolbar Buttons
ButtonButton nameDescription
New addressOpens a dialog box that enables you to set a new
address location for the detector.
AlignAligns the detector with the source.
14Emerson.com/Rosemount
Reference ManualOperating WinHost
00809-0200-4036April 2022
Table 4-4: Toolbar Buttons (continued)
ButtonButton nameDescription
SetupOpens a dialog box that enables you to configure the
detector.
MaintOpens the Maintenance window.
Primary micro software Displays the version and details of the primary micro
software.
Secondary Micro
Software Version
MasterSeeks the address of the connected detector (from 1
StopCloses the application.
4.1.5Detector status
The WinHost software displays the status in two fields: a letter field and a number field.
The detector can have the following statuses:
Characters
BBBN0NormalNormal operation during
Displays the version and details of the secondary micro
software.
to 247).
GroupDescription
interference
OMN0Maintenance call for low signal
or ratio
PPP74FaultParameters fault
VVV83Low voltage
Rosemount 93615
Operating WinHostReference Manual
April 202200809-0200-4036
CharactersGroupDescription
FFF34, 35, 36High noise
OON0Obscuration, saturation, or low
signal
QQN0Wrong alignment
DDD0Disconnection
AAN0AlarmAlarm
WWN0WarningWarning
XXX0AlignmentAlign
SSY0Startup/standby
GGG0Zero calibration
16Emerson.com/Rosemount
Reference ManualOperating WinHost
00809-0200-4036April 2022
4.2Setup table window
The setup table window allows you to view and program the various detector functions.
Figure 4-1: Setup Table Window
To configure the heater state body:
Procedure
1. From the main window, click Setup.
The Setup table window appears.
2. Define the parameters as required.
Rosemount 93617
Operating WinHostReference Manual
April 202200809-0200-4036
See Setup table parameters for all available parameters.
3. To define the new setup configuration, click F3-Burn.
4.2.1Setup table parameters
Table 4-5: Gas Type
RangeDescription
ShortFor short range
Medium LongFor medium/long range
Table 4-6: Heater Mode
ParameterDescription
OFFHeater always in off mode
AUTOThe heater functions automatically according to
the following conditions:
• If the ambient temperature level is below
the Heater ON at (C) parameter plus 10 °C,
the heater turns on.
• If the ambient temperature is higher than
the sum of Heater ON at (C) plus 20 °C, the
heater turns off.
ONHeater always in on mode
Table 4-7: Heater ON
Temperature (˚C)
0
5
10
15
20
25
30
4.3Set a new address
In this dialog box, you can set a new address for the detector. The valid detector addresses
are in range of 1 to 247.
To change the detector’s address:
Procedure
1. From the main window, click Address.
18Emerson.com/Rosemount
Reference ManualOperating WinHost
00809-0200-4036April 2022
The New address dialog appears.
2. In the Set new address field, type the new address.
3. Click F3-Burn.
The new address is set.
Figure 4-2: New Address Dialog Box
4.4Maintenance window
The Maintenance window allows the user to see the detector’s various realtime data
channels. In this window, you can also check the signals/noise values and the alignment of
the detector.
To view the detector’s various real-time channels:
Procedure
From the main window, click Maint.
The Maintenance window appears:
Rosemount 93619
Operating WinHostReference Manual
April 202200809-0200-4036
Figure 4-3: Maintenance window
The Maintenance window contains the following elements:
Parameter
GainElectronic amplification.
Signal 1Voltage signal of Sensor 1 after electronic amplification (depends on
ReferenceVoltage signal of reference after electronic amplification.
Signal 2Voltage signal of Sensor 2 after electronic amplification (depends on
Ratio 1Ratio between Sensor 1 and reference.
Ratio 2Ratio between Sensor 2 and reference.
NQratio 1Ratio1 / QO1 (Normalized ratio1). Equals 1 when no gas is present and
NQratio 2Ratio1 / QO2 (Normalized ratio2) Equals 1 when no gas is present and
LEL/PPM x mLower explosion level by meter.
TemperatureThe temperature inside the detector in °C.
VoltageDetector voltage.
Ref/NoiseReference/Noise
Sig1/NoiseSignal1/Noise
Description
gain).
gain).
goes down when gas is introduced.
goes down when gas is introduced.
Sig2/NoiseSignal2/Noise
StatusCurrent status
Esc - ExitExits the window.
20Emerson.com/Rosemount
Reference ManualOperating WinHost
00809-0200-4036April 2022
ParameterDescription
F9 - Check S/NChecks signals to noise.
F10 - Normal ModeCancels the S/N check and returns to normal operation.
AlignAligns the detector with the source.
GoodCommunications status.
4.5Check signals-to-noise
You can check signals-to-noise from the Maintenance window
Procedure
From the Maintenance window, click F9 - Check S/N.
The signals/noise values appear on the Ref/Noise, S1/Noise, and S2/Noise fields.
Note
You can revert to normal operation at any time by clicking F10 -Normal Mode.
4.6View micro software version
Procedure
1. To view the primary micro software version, from the main window, click Version.
A field appears, displaying the software version:
Figure 4-4: Primary micro version
2. To view the secondary micro software version, from the main window, click SeC
ver.
A field appears, displaying the software version:
Figure 4-5: Secondary micro version
4.7Log detector event
The detector does not have an internal log. In order to record events for diagnostic or
other purposes, the detector must be connected to a computer with the WinHost
software.
Rosemount 93621
Operating WinHostReference Manual
April 202200809-0200-4036
When you start logging, set the log file period in seconds. A line is subsequently written to
the log whenever that number of seconds passes (for example, every 60 seconds) and
whenever there is a change in the detector's status.
Each line in the log notes the following information:
• Detector temperature
• Detector gain
• Detector status
• Date and time
• Detector data
The filename of the recording is the serial number of the device, followed by .txt suffix. The
default location where the file is saved: C:\Users\Username\AppData\Local\VirtualStore
\Program Files (x86)\T88860A
To log detector events:
Procedure
1. 1 From the main window, click File→Start Logging .
The log record dialog box appears:
2. In the text field, enter the log file period (in seconds).
3. Click OK.
Logging begins. A line is written to the log (S/N.txt) every time the log file period is
over and any time there is a change in the detector’s status.
22Emerson.com/Rosemount
Reference ManualOperating WinHost
00809-0200-4036April 2022
Figure 4-6: Log File Record Message
4.8View log file
To view the log file:
Procedure
1. From the main window, click File→View log.
The log file viewer dialog box appears:
2. In the text field, enter the detector’s serial number, and click OK.
The Log file viewer window appears:
Rosemount 93623
Operating WinHostReference Manual
April 202200809-0200-4036
Figure 4-7: Log file viewer
4.9View WinHost software version
The About window displays the WinHost software version information.
Procedure
From the main window drop-down menu, click About.
24Emerson.com/Rosemount
Reference ManualMaintaining detector
00809-0200-4036April 2022
5Maintaining detector
The Rosemount 936 system requires only simple periodic maintenance to provide
satisfactory service and achieve maximum performance.
Use standard tools and equipment to maintain the detector and source. Recording the
periodic test results in the Maintenance log is recommended.
To maintain your detector
• Clean the optical surfaces of the detector and source once a month.
• Perform alignment and calibration once every six months or after each time the units
are moved or opened.
To clean the optical surfaces:
Procedure
1. Turn off the detector.
2. In places where dust or dirt have accumulated on the optical surface, clean the
surface with a small soft bristle brush.
3. Wash the surfaces thoroughly with water and a mild non-abrasive detergent.
4. Thoroughly rinse the glass surface with clean water, ensuring no residue is left
behind.
5. Dry the glass with a clean dry soft cloth.
6. Enter the following information into the Maintenance log: date and name of person
and company who performed the maintenance service.
Emerson Terms and Conditions of Sale are available upon
request. The Emerson logo is a trademark and service mark of
Emerson Electric Co. Rosemount is a mark of one of the
Emerson family of companies. All other marks are the property
of their respective owners.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.