Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated techni cal products, you
MUST properly install, use, and maintain them
normal specifications. The following instructions MUST be adhered to and integrated into your
safety program when installing, using, and maintaining Rosemount Analytical products. Failure to
follow the proper instructions may cause any one of the following situations to occur: Loss of life;
personal injury; property damage; damage to this instrument; and warranty invalidation.
to ensure they continue to operate within their
•Read all instructions
prior to installing, operating, and servicing the product.
•If you do not understand any of the instructions, contact your Rosemount Analytical representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of
the product.
•Install your equipment as specified in the Installation Instructions of the appropriate Instruc-
tion Manual and per applicable local and national codes. Connect all products to the proper
electrical and pressure sources.
•To ensure proper performance, use qualified personnel
to install, operate, update, program,
and maintain the product.
• When replacement parts are required, ensure that qualified people use replace ment parts
specified by Rosemount. Unauthorized parts and procedures can affect the product’ s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY
.
Look-alike substitutions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Ryton® is a registered trademark of Phillips Petroleum Co.
Teflon® is a registered trademark of E.I. duPont de Nemours and Co., Inc.
Emerson Process Management
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
e-mail: gas.csc@EmersonProcess.com
620434 Schematic Diagram, Isolated V/I Board
622228 Interconnect Diagram, Model 7003D Oxygen Monitor
622530 Schematic Diagram, Display Board
622538 Schematic Diagram, Power Supply Board
622617 Outline and Mounting Drawing, Model 7003D Oxygen Monitor
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Instruction Manual
748054-L
May 2003
Model 7003D
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 7003D
PREFACE
The purpose of this manual is to provide information concerning the components,
functions, installation and maintenance of the Model 7003D Oxygen Monitor.
Some sections may describe equipment not used in your configuration. The user should
become thoroughly familiar with the operation of this module before operating it. Read
this instruction manual completely.
DEFINITIONS
The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout
this publication.
DANGER .
748054-L
May 2003
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial
property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure,
condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Instruction Manual
748054-L
May 2003
Model 7003D
SAFETY SUMMARY
If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.
AUTHORIZED PERSONNEL
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property,
all personnel authorized to install, operate and service the this equipment should be thoroughly
familiar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can
cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a properly grounded
three-wire source of power. Electrical installation must be made in accordance with any applicable
national or local codes.
Alarm switching relay contacts wired to separate power source must be disconnected before servicing.
WARNING .
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this produc t.
Use only factory documented components for repair.
CAUTION .
HIGH PRESSURE GAS CYLINDERS
This module requires periodic use of pressurized gas. See General Precautions for Handling and
Storing High Pressure Gas Cylinders, page P-4.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 7003D
WARNING.
ENCLOSURE INTEGRITY
Unused cable conduit entries must be securely sealed by flameproof closures to prov ide enclosure
integrity in compliance with personnel safety and environmental protection requirements. The
plastic closures provided are for shipping protection only. When installing instrument, observe all
notes on drawing 622617 (in rear of this manual).
WARNING
ELECTRICAL SHOCK HAZARD
To avoid shock hazard and AC pickup, do not route potentiometric output or current outpu t cables
through the same conduit as power cable or alarm output cable.
WARNING
748054-L
May 2003
CORROSIVE MATERIAL
Concentrated nitric acid is used in rejuvenating the sensor cathode (Section 6-1b on page 6-3).
This material is highly corrosive.
CAUTION
CONDUIT GROUNDING
The non-metallic enclosure does not provide grounding between conduit connections. Use
grounding-type bushings and jumper wires.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Instruction Manual
748054-L
May 2003
Model 7003D
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed
Gases" published in 1981
Compressed Gas Association
1235 Jefferson Davis Highway
Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extreme s of
weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the
shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or
bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a
suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur whe n an empty cylinder is
attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125
never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc wel ding,
precautions must be taken to prevent striking an arc against the cylinder.
°
F (52°C). A flame should
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 7003D
DOCUMENTATION
The following Model 7003D instruction materials are available. Contact Customer Service Center or the
local representative to order.
748054 Operator Manual (this document)
COMPLIANCES
The Model 7003D Oxygen Monitor is designed to comply with applicable American standard s for protectio n
against electrical shock, mechanical and fire hazards in non-hazardous (ordinary) locations. The instrument must be installed in accordance with the provisions of the National Electrical Code (NEC),
ANSI/NFPA 70, and/or any applicable national or local code(s), and operated and maintained in the recommended manner.
748054-L
May 2003
The oxygen sensors and interconnecting cable used with the Model 7003D Oxygen Monitor are nonincendive in normal operation and comply with the requirements of Articles 501-3 (b)(1) c and 501-4 (b),
Exception, of the National Electrical Code, ANSI/NFPA 70-1987, for installations in Class I, Groups A, B, C,
D Division 2 classified locations.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Instruction Manual
748054-L
May 2003
Model 7003D
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
The Model 7003D Oxygen Monitor (Figure
1-1 below) automatically and continuously
measures the oxygen concentration in
gaseous samples. The determination is
based on measurement of the electrical current developed by an amperometric sensor
in contact with the sample.
The monitor provides direct readout of concentration in % by volume. Alarms and a
potentiometric output are provided as standard. The fullscale range of the alarms and
the potentiometric output are each independently selectable. Thus, the range of the
potentiometric output may be changed without the need to readjust alarm setpoints.
The Model 7003D is used with a sensor
which is housed in a submersion assembly
or flow assembly and is connected to the
monitor by a multi-conductor shielded cable.
Instruction Manual
748054-L
May 2003
SECTION 1
DESCRIPTION AND SPECIFICATIONS
1-1 OXYGEN MONITOR
The oxygen monitor conditions the sensor
output signal to provide direct readout of oxygen in % by volume. It also contains current-measuring circuitry, operating controls,
digital display, alarms, and signal outputs provisions.
The monitor is designed for panel mounting.
Accessory Pipe Mounting Kit permits the oxygen monitor to be mounted on a vertical or
horizontal pipe. Accessory Wall Mounting Kit
permits wall (surface) mounting. An optional
air purge kit is designed to meet requirements
for NFPA-496 Type Z air purge (see specifications) .
An accessory Isolated Current Output Board
provides a field-selectable 0 to 20 mA, or 4 to
20 mA isolated current output.
LOW
LOW
SET PT
%
SET PT
Model 7003D Oxygen Monitor
Figure 1-1. Model 7003D Oxygen Monitor
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
748054-L
May 2003
1-2 SENSORS
Rosemount Analytical offers rechargeable and
disposable oxygen sensors which can be
used with the Model 7003D. See Figure 1-2
below and Figure 1-3 below. These sensors
are supplied alone or in kits: Submersion, inline flow, and fast response.. Sensors are
available constructed of polypropylene or Ryton. See Section 2-4 on page 2-5 - Sensors,
Rechargeable, Section 2-5 on page 2-12 Sensors - Non-Rechargeable and Section 7-3
on page 7-2 - Sensors, Replacement Parts
for additional information.
Rechargeable - Fast-Response Flow
The fast-response flow assembly allows
minimum volume gas flow that permits mounting the sensor in a flowing gas stream. Sample is supplied at slightly above atmospheric
pressure, flows through the assembly and
discharges to atmospheric pressure. Internal
volume of the assembly is low to minimize
system response time.
Rechargeable - In-Line Flow
The in-line pressure compensated flow assembly permits mounting the sensor in a variable pressure gas stream at pressures up to
50 psig (345 kPa). The typical application is
in-line monitoring with the flow assembly connected directly into the process stream pipeline. An alternative application involves
discharge to atmospheric pressure where discharge rates are high.
Rechargeable - Submersion Assembly
The submersion assembly permits placing the
sensor at depth, in an open or closed vessel,
at a maximum pressure of 50 psig (345 kPa).
The submersion assembly provides a convenient means of mounting the sensor, and also
affords protection for the sensor cable connection a feature particularly desirable in high
humidity environments.
Non-Rechargeable - In-Line Flow
The in-line flow assembly permits mounting
the sensor in a flowing gas stream. Also included is a universal mounting bracket and a
Model 7003D
cable assembly for connecting the disposable
oxygen sensor to the monitor.
Non-Rechargeable - Submersion
The submersion assembly adapts the oxygen
sensor so that it may be inserted through the
wall of a vessel or pipe to monitor the oxygen
concentration existing in the vessel or pipe.
Figure 1-2. Rechargeable Sensor
Figure 1-3. Non-Rechargeable Sensor
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
1-3 SPECIFICATIONS
a. Performance
Operating Range........................... 0 - 19.99% to 0 to 25% oxygen by volume
Sample Temperature..................... 32°F to 110°F (0°C to 44°C)
Ambient Temperature.................... -20°F to 122°F (-29°C to 50°C) for instrument
Ambient Humidity .......................... Up to 95% relative humidity, non-condensing
System Linearity............................ For constant sample temperature after correction for sensor zero
b. Electrical
Power Requirements.....................107 to 127 VAC, 50/60 Hz, 0.2 A or 214 to 254 VAC, 50/60 Hz, 0.1
Potentiometric Output.................... Selectable: 0-25%, 0-10%, 0-5% or 0-1% oxygen fullscale
Voltage Output............................... Selectable: 0-10V, 0-5V, or 0-1V fullscale
Current Output (Option)................. Isolated 0-20mA or 4-20mA over same range as potentiometric
Instruction Manual
748054-L
May 2003
offset: ±1% of fullscale
Minimum load: 2K ohms
output. Minimum load: 600 ohms.
c. Alarm
Range............................................ Selectable: 0-25%, 0-10%, 0-5%, or 0-1% oxygen fullscale
Contacts ........................................ Two independently adjustable SPDT relay contact actuations
Deadband...................................... Adjustable from less than 2% to 20% of range at any setpoint
Repeatability.................................. ±0.1% of range
d. Physical
Mounting........................................Standard: Panel mount
Optional: Surface mount, pipe mount
Dimensions.................................... 10.4 x 8.9 x 8.3 inches (265 x 225 x 210 mm) HxWxD
Weight ........................................... 4.2 lbs (1.9 kg)
Enclosure....................................... NEMA-4X general purpose.
Optional air purge designed to NFPA-496 Type Z.
1
Sensor Cable................................. 1000 ft (305 m) Maximum length between instrument and sensor.
1
The optional air purge, when installed alon g with user supplied components, is designed to equip the instrument enclosure
with Type Z purge protection per Standard ANSI/NFPA 496-19 86. This reduces the classification within the enclosure from
Class I, Division 2 (normally non-hazardous ) to non-hazardous, thus permitting installation in a location classifie d Class I,
Groups A, B, C, D, Division 2. This method of protection is recognized in Article 500-1 of the National Electrical Code
(NEC)< ANSI/NFPA 70.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Stability.......................................... ±1% of fullscale at any given temperature per 24 hours, for an
Temperature Compensations........ 32°F to 110°F (0°C to 44°C) ±6% of reading
Response Time............................. 90% in 20 seconds for a step change, using an equilibrated sensor
Sample Pressure........................... 0 to 50 psig (0 to 345 kPa)
Model 7003D
equilibrated sensor
60°F to 90°F (15°C to 32°C) ±3% of reading
For any other 30°F (16°C) ±4% of reading
at 25°C
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
Instruction Manual
748054-L
May 2003
SECTION 2
INSTALLATION
2-1 UNPACKING AND FACILITY PREPARATION
Carefully examine the shipping carton and
contents for signs of damage. Immediately
notify the shipping carrier if the carton or contents is damaged. Retain the carton and
packing material until all components associated with the Model 7003D Oxygen Monitor
are operational.
Outline and mounting dimensions for the oxygen monitor are given in the outline and
mounting drawing 622617.
2-2 LOCATION AND MOUNTING
Mount the sensor in an environment within the
permissible range of 32°F to 110°F (0°C to
44°C) The sensor is supplied, as ordered, in a
kit that includes a submersion assembly or
flow chamber.
Sensor and amplifier are interconnected by a
multi-conductor shielded cable. It is supplied
in the standard 10 foot length, or in any length
specified by customer, up to 1000 feet (305
m). See Section 7.
The oxygen monitor is designed to meet
NEMA-4X requirements, and may be mounted
outdoors Permissible ambient temperature
range is -20°F to +122°F (-29°C to +50°C).
Panel mount the amplifier, or use the accessory wall mount kit or accessory pipe mount
kit as desired.
2-3 ELECTRICAL CONNECTIONS
WARNING
ELECTRICAL SHOCK HAZARD
Do not operate without doors and internal
circuit panel secure. Servicing requires
access to live parts which can cause death
or serious injury. Refer servicing to qualified personnel.
a. Line Power Connection
The Model 7003D provides
switch-selectable operation on either 107
VAC to 127 VAC or 214 VAC to 254 VAC,
50/60 Hz.
Electrical power is supplied to the monitor
via a customer-supplied three-conductor
cable, type SJT. minimum wire size 18
AWG. Route power cable through conduit and into appropriate opening in monitor enclosure. Connect power cable leads
to terminal strip TB5 on the power supply
board, as shown in Figure 2-1 on page 22 and drawing 622228.
Voltage Select
1. Open the monitor door.
2. Loosen the retainer screw which holds
the display board and pivot the display board to access the power supply board.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
3. Verify that the line voltage selector
switch S1 is set to indicate the nominal line voltage: either 115 or 230
(Figure 2-1 on page 2-2).
b. System Grounding Connection
A ground terminal (TB5-GND) is provided
on the power supply board. Refer to
Figure 2-1 on page 2-2. This terminal
must be connected, via the ground lead of
the three-conductor power cable to a
good earth ground.
Instruction Manual
A
A
A
748054-L
May 2003
Isolated V/I Board
(Option)
(see
Figure 3-2 on page 3-3)
Recorder
TB3
Sensor
TB2
R
1
R4
R3
R8
R9
POWER SUPPLY BOARD 622537
J1
R5 R6
U2
C2
U3
U6
C4
O
G
I
C5
CR2
1 2 3 4
I G O
TB1
30K
THERM
TB2
Current Output
TB3
C1
U1J1
CR
C3
U4
I
G
O
U5
3K/10K
THERM
SHLD
CATH
M
TB3
CR2
U5
O
G
I
U1
U3
O
G
I
O
O
I
G
G
I
U2U4
SHIELD
+VOLTAGE OUT
-VOLTAGE OUT
+CURRENT OUT
-CURRENT OUT
CR3
CR4
O
G
CR1
Model 7003D
C2C1
I
K2
F1
K1
TB4
NO
COM
LARM
NC
NO
COM
LARM
NC
TB5
C3
GND
NEUTRAL/L2
HOT/L1
T1
Voltage Select Switch
S1
S1
AC Power
TB5
c. Sensor Cable Connection
If a long cable is used, it should be routed
to the amplifier through appropriate conduit. Connect the amplifier end of the cable to terminal strip TB2 on the power
supply board, as shown in Figure 2-1
above and drawing 622228.
Connect the cable to the sensor when installation of the sensor is complete.
d. Output Cable Connection
ELECTRICAL SHOCK HAZARD
To avoid shock hazard and AC pickup,
do not route potentiometric output or
current output cables through the
same conduit as power cable or alarm
output cable.
If a recorder, controller, or other output
device is used, connect it via number 22
or number 24 AWG two-conductor
shielded cable. Route the output cable
Figure 2-1. Power Supply Board
through conduit to the oxygen monitor,
and into the case through the appropriate
opening shown in drawing 622617.
Potentiometric Output
1. Connect leads of shielded recorder
cable to VOLTOUT + and VOLTOUT
- terminals of TB3 on power supply
board (Figure 2-1 above). Connect
shield to SHD terminal.
2. Connect other end of output cable to
appropriate terminals of recorder or
other potentiometric device.
WARNING
a. For devices with spans of 0 to
1.0 to 5, or 0 to 10 volts connect
cable directly to input terminals
of the device, making sure polarity is correct.
b. For devices with intermediate
spans. i.e., between the specified values, connect cable to device via a suitable external
voltage divider. as shown in
Figure 2-2 on page 2-3.
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
r
Instruction Manual
748054-L
May 2003
Isolated Current Output Accessory
1. Verify that the current output board
(Isolated V/I Board) is properly
mounted on the power supply board
in connector J2. See Figure 2-1 on
page 2-2. If originally ordered with
the oxygen monitor, the board is fac-
must be connected in series. Refer to
Figure 2-2 below.
NOTE
Total resistance of all output devices
and associated interconnection cable
must not exceed 600 ohms.
tory installed.
4. Since neither the CUROUT + nor
2. Connect leads of shielded cable to
CURRENT OUT + and CURRENT
OUT - terminals of TB3 on power
supply board. Connect shield to SHD
terminal.
CUROUT - output terminal is
grounded, the current loop should be
grounded at some point within the circuit. The ground point should be
chosen to minimize noise or other undesirable interactions.
3. Connect other end of output cable to
input terminals of recorder or other
current-actuated device, making sure
polarity is correct. If two or more current-actuated devices are used, they
7003D
Monitor
Potentiometric
Recorder
Voltage Divide
(Customer Supplied)
Input
Terminals
(Verify pol arity
is correct )
Position of Recorder Outpu t
Selector Plu g
10 mV 1K Ohm
100 mV 10K Ohm
1 V 100K Ohm
5 V 2K Ohm
Min i mum Per missible Re si stance for
R1 + R2
Figure 2-2. Connections for Potentiometric Recorder with Non-Standard Span
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
748054-L
May 2003
Model 7003D
7003D
Monitor
Figure 2-3. Typical Example of Oxygen Monitor Connected in Series with Several Current Actuated
mA
+
-
+
-
+
-
+
-
Recorder
Controller
Remote
Indicator
Devices
e. Output Connections for Alarms
Alarm Output Connections
The alarm output provides two sets of relay contacts for actuation of alarm and/or
process-control functions. Leads from the
customer-supplied external alarm system
connect to terminal block TB4 on the
power supply board. See Figure 2-1 on
page 2-2 and drawing 622228.
If the alarm contacts are connected to any
device that produces radio frequency interference (RFI), it should be
arc-suppressed. Accessory Arc Suppressor is recommended. When possible, the
oxygen monitor should operate on a different AC power source, to minimize RFI.
Alarm Relay Characteristics
The HI ALARM and LO ALARM outputs
are provided by two identical single-pole
double-throw relays. Relay contacts are
rated at (resistive load):
3 A, 250 VAC
3 A, 30 VDC
Removal of AC power from the analyzer,
as in a power failure, de-energizes both
relays.
HI ALARM Relay
The HI ALARM relay coil is energized
when the display moves upscale through
the value that corresponds to the setpoint
plus deadband. This relay coil is
de-energized when display moves downscale through the value that corresponds
to setpoint minus deadband.
LO ALARM Relay
The LO ALARM relay coil is energized
when the display moves downscale
through the value that corresponds to
setpoint minus deadband. This relay coil
is de-energized when the display moves
upscale through the value that corresponds to setpoint plus deadband.
Alarm Reset
The HI ALARM and LO ALARM functions
both incorporate automatic reset. When
the meter reading goes beyond the preselected limits, the corresponding relay is
energized; when the meter reading returns within the acceptable range, the relay is automatically de-energized.
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
Instruction Manual
748054-L
May 2003
Fail Safe Applications
By appropriate connection to the double-throw relay contacts, it is possible to
obtain either a contact closure or a contact opening for an energized relay. The
de-energized relay then provides a contact opening or contact closure, respectively. It is important for failsafe
applications that the user understand
what circuit conditions are desired in
event of power failure and the resultant
relay de-energization. Relay contacts
should then be connected accordingly.
2-4 SENSORS – RECHARGEABLE
This section provides instructions on installation and gas connections of the sensors in the
three available mounting configurations. Refer to Section 2-3c on page 2-2 for electrical
connections to sensor. Refer to Section 7 for
part numbers of sensors, sensor kits, and accessories. The sensor is shipped assembled,
charged, and ready for use.
2-6B on page 2-7, is acceptable provided
the sample enters the lower port of the
chamber and discharges from the upper
port.
Installing Sensor
1. Refer to Figure 2-5 on page 2-6. Remove the sensor cap from the sensor.
Slide the stainless steel ring clamp
onto the threads for the sensor cap
until stops against sensor body. Replace the sensor cap. The ring clamp
is held between the cap and body of
the sensor.
2. Insert the o-ring and cap end of the
sensor into the flow chamber.
3. Slide the flow chamber nut down the
body of the sensor and tighten onto
flow chamber.
The o-ring between the bottom of the
chamber and the sensor cap now effects
a seal.
a. Installation of Sensor and Fast-
Response Flow Kit
The Fast-Response Flow Assembly is
used to mount the sensor in a gaseous
sample flow stream. It is used when the
sample is supplied at slightly above atmospheric pressure. Internal volume is
low to minimize system response time.
Mounting Flow Chamber
The flow chamber has two .2 diameter
clearance holes for mounting screws
(screws supplied by user). See Figure
2-4 on page 2-6.
Mount the flow chamber as shown in
Figure 2-6 on page 2-7. Preferably,
mount the flow chamber so that the sensor is vertical as shown in Figure 2-6A on
page 2-7. Horizontal mounting, Figure
Gas Connections
Sample inlet and outlet connections are
1/8 inch NPT (Figure 2-4 on page 2-6).
Normally, sample is supplied to the flow
chamber via a customer supplied needle
valve which provides flow adjustment.
Pressure and Flow
The flow chamber is designed for discharge at atmospheric pressure. Maximum recommended flow is 2 liters per
minute. Minimum acceptable flow rate
depends on required rate of system response, which is dependent on length of
the sample line, and in some cases, the
volume of gas sample that can be wasted.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
p
p
∅
748054-L
May 2003
Figure 2-4. Rechargeable Sensor with Fast Response Flow Assembly
Ring Clamp
Sample Inlet/Outlet
1/8 NPT
DIMENSIONS
Inch
[mm]
Flow Chamber
Sensor
Nut
O-Ring
Sensor Ca
Flow
Chamber
Ring
Clam
O-Ring
Cable to Oxygen Analyzer
.4
[9]
Nut
Sensor Body
1.5
[37]
3.0
[76]
Model 7003D
3.6
[9.1]
Mounting Holes (2)
∅.2 THRU
C'BORE ∅ .38 x .25 DP
1.5
[38]
Figure 2-5. Mounting Rechargeable Sensor in Fast Response Flow Assembly
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
Instruction Manual
748054-L
May 2003
A. Vertical Installation (Preferred)
Vent to
Atmosphere
Sample In
B. Horizontal Installation
Vent to
Atmosphere
Sample In
Figure 2-6. Typical Installation Orientation of Rechargeable Sensor and Fast Response Flow
Assembly
shown in Figure 2-8 on page 2-9. In this
b. Installation of Sensor and In-Line Flow
Kit
The In-Line Flow Assembly is used to
mount the sensor in a variable-pressure
gaseous or liquid sample stream, at pressures up to 50 psig (345 kPa).
orientation the sample enters the lower
port of the assembly and discharges to
the upper port. The horizontal arrangement ensures that during setup there will
be immediate discharge of any gas in the
flow chamber or any entrained gases in
the sample stream.
The typical application is in-line monitoring,
with the flow assembly connected directly
into the process-stream pipeline. An alternative application involves discharge to
For gaseous sample streams, the asse mbly may be mounted with the sensor horizontal, if desired.
atmospheric pressure where high discharge rates are desired.
Note that the sensor responds to partial
pressure of oxygen in the sample. If total
pressure changes, the oxygen responds
accordingly.
For liquid sample streams, it is recommended that the assembly be mounted so
the sensor is in the vertical position, as
Mounting Flow Chamber
Outline and mounting dimensions are
given in Figure 2-7 on page 2-8. The flow
chamber has two .22 diameter clearance
holes for mounting screws (screws supplied by user). The sample inlet and sample outlet are 1/2 FPT. Mount the flow
chamber as shown in Figure 2-8 on page
2-9.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
748054-L
May 2003
Installing Sensor
Model 7003D
3. Replace the adapter ring and clamp.
1. Remove the clamp and adapter ring
from the flow chamber.
Gas Connections
Sample inlet and sample outlet connec-
2. Insert the sensor (with gland) into the
tions are 1/2 FPT.
flow chamber.
Pressure and Flow
GAS STREAM LIQUID STREAM
MAXIMUM PRESSURE 50 psig (345 kPa) 50 psig (345 kPa)
MAXIMUM FLOW 2 cfm (60 L/min) 5 gal/min (6 L/min)
MINIMUM FLOW dependent on desired response time 1.5 gal/min (6 L/min)
Gland
Adapter Ring
Flow Chamber
Clamp
4.5
[127]
Sample Inlet/Outlet
1/2 NPT Female
.75
[19]
Mounting Holes
.219 [6] DIA
DIMENSIONS
Inch
[mm]
4.0
[102]
3.5
[89]
Figure 2-7. Rechargeable Sensor with Gland and In-Line Flow Assembly
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
Figure 2-8. Preferred Installation Orientation of Rechargeable Sensor and In-Line Flow
c. Installation of Sensor and Submersion
Kit
The Submersion Assembly permits placing
the sensor at depth in an open or closed
vessel. The Submersion Assembly provides a convenient means of mounting the
sensor, and also affords protection for the
sensor cable connection, a feature particularly desirable in high-humidity environments.
Sample In
Instruction Manual
748054-L
May 2003
To drain or process
stream
Assembly
Mounting in Gaseous Sample Stream
For gaseous samples, the submersion assembly may be oriented horizontally or
vertically.
Operating Pressure
Maximum permissible operating pressure
is 50 psig (345 kPa), equivalent to a water
depth of approximately 100 feet (approximately 30 m).
For liquid sample stream, it is recommended that the submersion assembly be
mounted in horizontal position, as shown
in Figure 2-10 on page 2-10. This arrangement prevents entrapment on the
membrane of gas bubbles which could
cause erroneous oxygen readings.
For installation made during plant construction, a swing-arm piping arrangement
installed in the side of the tank has proved
most satisfactory. The sensor cable is led
from the sensor through the supporting
pipe. A junction box located in a manhole
on the deck beside the aeration vessel
may be included optionally to facilitate pulling of the cable. Permanently installed
conduit carries the sensor cable to the
amplifier, which may be located in any
convenient building or protective enclosure. Figure 2-11 on page 2-11 and
Figure 2-12 on page 2-11 illustrate typical
installations.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-9
Liquid Sample Flow
For liquid sample streams, velocity of the
liquid past the sensor tip must be at least
1.5 feet per second (45 cm/sec) for flowindependent oxygen readout.
Mounting Chamber
The chamber is mounted to 3/4 inch pipe
(customer supplied), which the sensor cable is fed through. See Figure 2-9 on page
2-10and Figure 2-10 on page 2-10.
Installing Sensor
1. Remove the clamp and adapter ring
from the flow chamber.
2. Insert the sensor (with gland) into the
flow chamber.
3. Replace the adapter ring and clamp.
Instruction Manual
y
748054-L
May 2003
Sensor Cable
In a permanent installation, the sensor cable is normally routed through a customersupplied conduit, which screws into the 3/4
inch NPT connection on the submersion
assembly.
Submersion
Assembl
3.4
[86]
Chamber
Ring Clamp
Adapter Ring
2.0
[51]
1.9
[48]
Model 7003D
In applications where the sensor cable is
not housed in conduit, the cable connector
accessory (PN 856832) may be used.
This separately ordered item provides watertight sealing to prevent leakage around
the cable. The connector is constructed of
aluminum and includes a inner sealing
grommet and compression nut.
Aluminum Cable Connector
(Accessory)
2.9
[73]
3/4 - 14 NPT
7.4
[187]
Figure 2-9. Rechargeable Sensor with Submersion Assembly
Sensor Cable to Oxygen Monitor
Customer supplied 3/4 inch pipe
1. Preferred orientation in liquid sample stream.
2. Orientation in gaseous sample may be either horizontal or vertical.
Submersion Assembly horizontal
Figure 2-10. Typical Installation of Rechargeable Sensor and Submersion Assembly
Internal Sensor membrane vertical
2-10 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
Instruction Manual
748054-L
May 2003
Cable to Oxygen Moni-
Pipe and fittings to suit installation
Figure 2-11. Typical Permanent Installation of Rechargeable Sensor with Submersion Assembly
Pipe and fittings to suit installation
Manhole
Handrail
Diffuser
During Plant Construction
Cable to Oxygen Moni-
Chain to
Diffuser
Handrail
Submersion Assembly
Chain to Deck
Level
Capped pipe clamped to handrail
Union (loose to act as swivel)
Submersion Assembly
Figure 2-12. Typical Installation of Sensor/Submersion Assembly in an Existing Plant
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-11
Instruction Manual
748054-L
May 2003
2-5 SENSORS - NON-RECHARGEABLE
a. Conversion Of Oxygen Monitor From
Use With Rechargeable Sensor To
Use With Non-Rechargeable Sensor
The 7003D monitor is shipped from the
factory configured for use with a rechargeable sensor. To convert the instrument for
use with a non-rechargeable sensor, a
Non-Rechargeable Sensor Conversion Kit
must be installed. The Conversion Kit
changes the gain of the monitor to match
the output current available from the nonrechargeable sensor. Follow the instructions supplied in the kit.
b. Installation of Sensor and In-Line Flow
Assembly
This kit consists of a non-rechargeable
oxygen sensor designed for analysis of
gaseous oxygen samples, a flow chamber
and associated nut, a universal mounting
bracket, a connecting cable to connect the
monitor to the sensor, and appropriate
loose hardware (see Figure 2-13 on page
2-14). The PVC flow chamber is designed
for use with a non-rechargeable oxygen
sensor when a flowing gas stream with discharge of the effluent from the flow
chamber at atmospheric pressure is being
measured. This requirement means that
upstream sample pressure reduction must
be performed on the process sample from
a pressurized source before it is presented
to the flow chamber for analysis by the
sensor. Sample input flow rate should be
selected in the range of 50 to 100 cc/min
and care must be taken with downstream
pressure drops to prevent back pressurization of the sensor.
Mounting the Flow Chamber
Refer to Figure 2-15 on page 2-15. The
preferred mounting configuration of the
sensor is with the electrical connector at
the top of the flow chamber/sensor assembly. Operation in the horizontal plane
is also possible. To facilitate mounting the
flow chamber to a bench surface, wall or
pipe, employ the universal mounting
bracket included as part of the kit. The
Model 7003D
flow chamber has one port which should
be capped off with the fitting cap supplied
in the kit. The port to be capped is determined by the user, depending on the particular application.
Installing the Sensor
The sensor is inserted into the flow chamber to form a face seal on the front of the
sensor with the O-ring present in the bottom of the flow chamber. The nut is then
placed over the connector end of the sensor and tightened hand tight to form a seal
on the face of the sensor.
Gas Connections
Sample inlet and outlet connections are
1/8-inch NPT. Normally, sample is supplied at slightly above ambient pressure by
use of a customer supplied needle valve
which provides the necessary pressure
and flow adjustment to the desired 50 to
100 cc/min flow rate level.
c. Installation of Sensor with Submersion
Assembly
This kit consists of a non-rechargeable
oxygen sensor for gas phase oxygen
measurements, a head space submersion
assembly, and assorted required loose
hardware (see Figure 2-15 on page 2-15).
It is intended for use when the gas phase
sensor is to be inserted through a vessel
wall as in the monitoring of a process
vessel head space or when a large diameter process is being monitored directly by insertion of the sensor through
the pipe.
Mounting the Submersion Assembly
Refer to Figure 2-15 on page 2-15. The
sensor installs in the submersion assembly by placing the doughnut shaped thin
rubber gasket on the connector side of
flange of the oxygen sensor by gently
slipping it over the sensor body and then
connecting the cable to the cable sensor
using the mating connectors. The end of
the cable emerging from the threaded
portion of the submersion assembly
should then be gently pulled to seat the
2-12 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
Instruction Manual
748054-L
May 2003
sensor within the receiving cavity of the
submersion assembly. At this point, the
cap should be placed over the front end of
the submersion assembly and then
screwed down snugly. At this point the
blue silicone rubber plug has moved with
the cable and is now no longer seated in
the threaded portion of the submersion
assembly. Hold the cable securely and
move the plug back into the submersion
assembly until it is firmly seated.
The submersion assembly is now ready to
be connected to the 3/4-inch pipe in a
manner dictated by the local installation
requirements. See Figure 2-16 on page
2-16.
Sensor Cable
In permanent installations the sensor cable is normally routed through a customer
supplied conduit, which screws into the
3/4-inch NPT connection at the end of the
submersion assembly.
Operating Pressure
Maximum operating pressure is 50 psig
(345 kPa).
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-13
Figure 2-13. Dimensions and Components of Non-Rechargeable Sensor with In-Line Flow Kit
2-14 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
LIQUID GAS
To
Drain
or
Process
Stream
Preferred orientation of
In-Line Flow Assembly
To
Drain
or
Process
Stream
Sample
In
Instruction Manual
748054-L
May 2003
Sample
In
Preferred orientation of
In-Line Flow Assembly
Figure 2-14. Typical Installation of Non-Rechargeable Sensor with In-Line Flow Kit
Grommet/Plug
2.0
[51]
Adapter
1-11 1/2NPT
Cable
5.7
[146]
Nut
Gasket
1.0
[25]
2.3
[57]
Figure 2-15. Dimensions and Components of Non-Rechargeable Sensor with Submersion Kit
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-15
Instruction Manual
748054-L
May 2003
A. INSTALLATION DURING PLANT CONSTRUCTION
Cable to Oxygen Monitor
Pipe and fittings to suit
installation.
B. INSTALLATION IN EXISTING PLANT
Handrail
Pipe and fittings to suit
installation.
Manhole
Cable to Oxygen Monitor
Diffuser
Diffuser
Model 7003D
Chain to
Deck Level
Sensor/Submersion Assembly
Capped Pipe Clamped to
Union (loose to act as swivel)
Chain to
Deck Level
Sensor/Submersion Assembly
Figure 2-16. Typical Installation of Non-Rechargeable Sensor with Submersion Kit
2-16 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
SECTION 3
STARTUP AND CALIBRATION
Prior to startup and calibration, familiarization with
Figure 3-1 on page 3-3 is recommended. This figure
gives locations and brief descriptions of operating
controls and adjustments on the display board.
3-1 SYSTEM STARTUP
After completing the installation per Section 2,
proceed as follows:
1. On the Display Board (refer to diagram inside door and Figure 3-1 on page 3-3):
a. Select potentiometric output fullscale
range by placing output/alarm switch
S3 in ON position. Choices are 0 to
25%, 0 to 10%, 0 to 5%, and 0 to 1%
oxygen.
b. Select potentiometric output fullscale
voltage by placing output volts switch
S2 in ON position. Choices are 0 to 1,
0 to 5, and 0 to 10 volts, fullscale.
c. Set alarm switches S4 and S5 to OFF
position, to avoid activating either
alarm before setpoints and deadbands
have been adjusted (Section 3-3 on
page 3-2).
2. Turn on line power.
Instruction Manual
3-2 CALIBRATION
Sensor - Residual Current Zeroing
Each sensor has an individual residual current
that must be zeroed out by adjustment of R52
on the Display Board. The sensor is purged
with nitrogen gas until a stable reading is obtained. Adjust R52 for a zero reading. This
adjustment is required each time a new or recharged sensor is installed.
After the sensor has stabilized, the instrument
is ready for calibration. Perform the procedure
in Section 3-2a below or 3-2b on page 3-2, as
appropriate.
a. Calibration with Air
Calibration with dry air is the preferred
method, applicable in all situations. The
most convenient source of dry air, when
available, is a cylinder of compressed air.
It is important, however, to know the actual
oxygen content of the cylinder. Cylinder
compressed air is frequently a prepared
blend of O
labeled as such. U.S. Government standards require only that cylinders labeled
"air" contain between 19% and 23% oxygen.
and N2, and is not necessarily
2
748054-L
May 2003
3. Wait for a suitably stable reading on the
digital display. The time required by the
user for stabilization of the sensor to levels
of performance where the stability of the
sensor meets the requirement of a particular user is defined as the equilibration time,
that is, the time when the sensor is said to
be equilibrated. Normal operation of the
analyzer involves adjustments of only the
front-panel controls: ZERO and SPAN, and
RANGE Switch, if provided. See Figure
1-1 on page 1-1.
Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Calibration 3-1
To determine oxygen content, request an
analysis when purchasing the cylinder air.
Another way to determine oxygen content
is to pass ambient outdoor air through an
efficient drier on the way to the analyzer,
preferably using a diaphragm-type pump.
The concentration of oxygen will then be
very close to 21%.
1. On the Display Board, set RANGE
switch S7 at 25%, expose sensor to
suitable dry air at ambient atmospheric
pressure.
Instruction Manual
748054-L
May 2003
2. Change RANGE switch to CAL position, wait for display reading to stabilize. Then adjust CAL control
(disregard decimal point) to bring display reading to:
760 X % O
20.95
In most situations. ambient air may be
used in place of the recommended dry air.
Only in instances where temperature and
relative humidity are high will significant errors be introduced through the use of ambient air. At a near-sea-level barometric
pressure, errors are less than 2% for relative humidity under 40% and temperatures
below 95°F (35°C). At 60% relative humidity, ambient air temperatures as high
as 82°F (27.8°C) produce a maximum error of 2%. As a rule-of-thumb. dry air
should be used for calibration where ambient temperature exceeds 80°F (26.7°C)
and/or relative humidity exceeds 60%. For
calibration with ambient air, the procedure
is the same, except that the sensor is exposed to ambient air instead of dry air.
2 in sample
Model 7003D
tion. Choices are 0 to 25%. 0 to 10%, 0 to
5% and 0 to 1% oxygen. Refer to diagram
on inside of door.
2. The HI ALARM and LO ALARM setpoint
potentiometers are adjustable from 0% to
100% of the fullscale span. The potentiometers are graduated from 0 to 100.
Required potentiometer setting for either
setpoint adjustment is determined from the
equation:
Alarm range, 0 to 25%; desired Hi Alarm
setpoint, 20%; desired Lo Alarm setpoint,
15%
required Hi ALARM setting = x 100
= 80
required LO ALARM setting = x 100
= 60
20
25
15
25
b. Calibration with Span Gas
1. Set RANGE switch S7 to 25% or
19.99%, depending upon concentration of span gas, and expose sensor to
span gas, exhausting to ambient pressure.
2. Wait for reading to stabilize, then adjust
CAL control to bring display reading to
known concentration of span gas.
3-3 SELECTION OF ALARM RANGE,
SETPOINT, AND DEADBANDS
1. Select alarm fullscale range by placing one
slide of output/alarm switch S3 in ON posi-
3. Select the desired deadband. The HI
ALARM and LO ALARM deadband potentiometers are adjustable from 1% of
fullscale (counterclockwise limit) to 20%
of fullscale (clockwise limit). Deadband
is essentially symmetrical with respect
to the setpoint.
4. When setpoints and deadbands have
been selected, set alarm switches S4
and SS to the AUTO position, to activate
the alarms.
5. To test external alarm devices, temporarily set alarm switch S4 or S5 in the
ON position. The associated alarm is
then unconditionally on.
3-2 Startup and Calibration Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
A
LOW ALARM
INDICATOR LED
CR1
LOW ALARM
SETPOINT ADJUST
R43
LOW ALARM
DEADBAND ADJUST
R44
SENSOR CONVERSION
HEADER
J5
AUTO OFF ON
LOW ALARM SWITCH
S4
RANGE SWITCH S7
CAL CONTROL R14
OUTPUT VOLTAGE
SWITCH S2
OUTPUT/ALARM
SWITCH S3
19.99 25 CAL
S2
OFF ON
S3
OFF ON
E OUT
LARM
Instruction Manual
748054-L
May 2003
AMPLIFIER ZERO R46
DISPLAY SPAN R24
RESIDUAL ZERO CONTROL R52
HIGH ALARM
INDICATOR LED
CR2
HIGH ALARM
SETPOINT ADJUST
R40
HIGH ALARM
DEADBAND ADJUST
R41
AUTO OFF ON
HIGH ALARM
SWITCH S5
Figure 3-1. Display Board
Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Calibration 3-3
Instruction Manual
748054-L
May 2003
3-4 CURRENT OUTPUT RANGE
Select 0 to 20 or 4 to 20 milliamperes to be
minimum current for isolated current output (if
present).
1. On the Display Board (Figure 3-1 on
page 3-3), set RANGE switch S7 to CAL
position.
2. With power OFF, disconnect anode and
cathode leads from terminal strip TB2
on the Power Supply Board (Figure 2-1
on page 2-2). Secure the leads so that
they will not contact any board or component.
3. Turn power ON. On the Display Board
(Figure 3-1 on page 3-3), adjust residual
ZERO control R52 until potentiometric
output device indicates 0 volt.
4. On the Isolated Current Output Board,
adjust ZERO control R1 (Figure 3-2
below) so that current output device indicates the low range-limit desired for
ZERO
R1
SPAN
R2
R3
R8
R4
R
1
R
2
R5 R6
R9
CR2
U6
O
G
I
1 2 3 4
U3
U2
Model 7003D
the output range: 0 milliamperes or 4
milliamperes.
5. Turn power OFF. If sensor is rechargeable, connect a 10K ohm resistor between the anode and the cathode
terminals on the Power Supply Board
terminal strip TB2 (Figure 2-1 on page
2-2). If sensor is non-rechargeable,
connect a 320K ohm resistor across
TB2. This resistor permits the polarizing
voltage supply to provide a small current
that simulates the sensor output signal.
6. Turn power ON. On front panel controls, adjust CAL clockwise to produce
a fullscale reading on the potentiometric
output device.
7. On the Isolated Current Output Board
(Figure 3-2 below), adjust SPAN control
R2 for reading of 20 milliamps on current output device.
C1
U1J1
CR
C2
U4
C4
I G O
C3
I
G
O
U5
Figure 3-2. Isolated Current Output Board
3-4 Startup and Calibration Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
Instruction Manual
748054-L
May 2003
SECTION 4
OPERATION
4-1 ROUTINE OPERATION
After startup and calibration (Section 3) the
monitor will automatically and continuously indicate the oxygen concentration of the sample.
If potentiometric output or alarms are used on
the 0% to 1% oxygen range after calibration at
20.95% oxygen, an appropriate interval is
needed to permit the sensor to equilibrate to a
low concentration of oxygen. The time required for equilibration depends upon how
long the sensor was operated at the higher
oxygen level. Thus, a sensor which has been
in operation on air for 24 hours prior to calibration will typically show a reading of 0.05%
oxygen, or less within one hour after exposure
to pure nitrogen. This reading will decrease to
0.01% or so within the next few hours.
Inability to obtain a low-level oxygen reading
in a gas of known low-level oxygen content
usually indicates a leak in the sample system.
To check this possibility, increase sample flow
rate, block off the flow of gas into and out of
the flow chamber, and note the response of
the oxygen monitor. An increase in reading,
with time, indicates a leak in the system. If
there is no leak in the system, another possibility is that a rechargeable sensor should be
rejuvenated or that a disposable sensor
should be replaced. The procedure for sensor
rejuvenation, which is outlined in Section 6
must be followed carefully if satisfactory results are to be achieved.
4-2 RECOMMENDED CALIBRATION
FREQUENCY
For the first few days of operation the instrument should be calibrated daily to compensate for initial stabilization of the membrane in
the sensor.
After the sensor has stabilized, an instrument
in continuous operation should be calibrated
at least once a week until the appropriate
calibration interval is determined. A log of periodic calibrations helps establish desired calibration intervals for given applications.
4-3 FREQUENCY OF SENSOR RECHARGING
Nominal useful life of a sensor charge is approximately three to six months; after this
time, the sensor should be removed from the
installation and recharged. Refer to Section 61a on page 6-1. Physical damage to the
membrane is the most frequent cause of failure for service periods less than three
months. Proper handling and installation of
the sensor into the sample system are essential.
The service life of a disposable 0xygen sensor
is application-dependent and no rejuvenation
or recharging is possible. It is difficult to determine the prospective shelf life of the disposable oxygen sensor, since no time related
failure mechanisms are apparent at this time.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 4-1
Instruction Manual
748054-L
May 2003
Model 7003D
4-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
SECTION 5
THEORY
The Model 7003D Oxygen Monitor system consists
of an amperometric oxygen sensor and an amplifier
unit interconnected by a multi-conductor shielded
cable. The sensor responds to the partial pressure
of oxygen. The amplifier conditions the sensor signal, providing a readout of oxygen in percent by volume.
The following Sections detail the major principles of
operation. Section 5-2 on page 5-2 describes variables that influence the oxygen measurement.
5-1 ELECTROCHEMICAL THEORY
With the sensor placed in the process stream,
a voltage is applied across the cathode and
anode (see Figure 5-1 below). Oxygen in the
THERMISTOR
CAP
O-RINGS
MEMBRANE
Instruction Manual
748054-L
May 2003
process stream diffuses through the membrane and is reduced at the cathode. The reduction of oxygen results in a current flow
proportional to the partial pressure of oxygen
in the sample.
When oxygen is not present, no electrical current flows in the sensor. When oxygen is present, electrical current flows in the sensor
according to the characteristic curve for the
particular potential applied to the electrodes.
The magnitude of the current depends upon
the partial pressure of the dissolved oxygen in
the sample.
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 5-1
Instruction Manual
748054-L
May 2003
5-2 VARIABLES INFLUENCING OXYGEN
MEASUREMENT
Since the amperometric oxygen sensor responds to the partial pressure of oxygen, any
variable that affects that partial pressure
should be taken into account to ensure accurate measurement in the desired units. The
two basic variables involved are barometric
pressure and relative humidity.
Since dry air contains 20.95% oxygen by volume, regardless of barometric pressure, the
partial pressure of oxygen is directly proportional to the total barometric pressure, according to Dalton's law of partial pressures. Thus
for dry air, if the total barometric pressure is
known, the partial pressure of oxygen can be
computed. However, the procedure is valid
only for dry air. Humid air has the effect of reducing the partial pressure of oxygen and the
other gases in the air without affecting the total barometric pressure. Another way of expressing this relationship is by the following
equation:
Model 7003D
Thus, for constant barometric pressure, if the
humidity in the air is not zero, the partial pressure of oxygen is less than the value for dry
air. For most measurements taken below 80°F
(26.7°C) and below 60% RH, the effect of water vapor may be ignored.
At a barometric pressure of 760 mm Hg (101
kPa), the partial pressure of oxygen in dry air
is approximately 160 mm Hg (21.1 kPa).
To determine the partial pressure of oxygen in
air at various levels of humidity and barometric pressure. the partial pressure of water is
subtracted from the total barometric pressure:
this difference is then multiplied by 20.95%.
EXAMPLE
barometric pressure = 740 mm Hg (98.5
kPa)
partial pressure H
kPa)
0 = 20 mm Hg ( 2.7
2
P (atm) = P (gas)+ P (oxygen) + P (water)
where:
P (atm) = total barometric pressure
P (gas) = partial pressure of all gases
other than oxygen and water vapor
P (oxygen) = partial pressure of oxygen
P (water) = partial pressure of water vapor
partial pressure O
mm Hg = 151 mm Hg (20.1 kPa)
= (740-20) x 0.2095
2
5-2 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
ROUTINE SERVICE AND TROUBLESHOOTING
Most service and maintenance problems involve
the sensor. Failures within the amplifier unit are
less frequent. In system checkout, the recommended procedure is first to isolate the amplifier
from the sensor, then perform a few simple tests
to determine if the amplifier is performing satisfactorily. The sensor can be checked and then recharged, rejuvenated, or replaced, as necessary.
Instruction Manual
748054-L
May 2003
SECTION 6
the gold cathode should be rejuvenated as
described in Section 6-1b on page 6-3.
In event of "spiking," i.e.,
non-oxygen-related transient response, the
accessory Cell Separator Kit is recommended. Refer to Section 6-1c on page 6-
4.
6-1 RECHARGEABLE SENSORS
Most routine maintenance involves the sensor. Sensor maintenance consists of
periodic recharging and cleaning, or rejuvenating the sensor cathode. The usual indication that the sensor requires
rejuvenation and recharging is that, during
calibration. the correct upscale reading is
unobtainable by adjustment of the CAL
control. Normally, the inability to calibrate is
preceded by a gradual, day-to-day reduction in sensor output, with a resultant lower
instrument indication. The rate of reduction
increases with the increase in internal resistance of the sensor. Other indicators of
need for rejuvenation may be sluggish response or the presence of an appreciable
residual signal when the sensor is exposed
to a zero reference sample.
NOTE
If sensor is disassembled for inspection,
it must be recharged utilizing a new
membrane.
a. Recharging Sensor
The sensor must be removed from the
process installation and disconnected
from the sensor cable for recharging.
The recharging kit consist of electrolyte,
membranes, pressure-compensating
rubber diaphragms, O-rings for the
membrane retainer, O-rings for the
sensor body, and washers for the diaphragms.
Rechargeing Procedure
For this procedure refer to Figure 6-1
on page 6-2.
1. Unscrew knurled cap from end of
sensor body. Remove membrane
assembly, consisting of membrane
fixed between holder and retainer.
Empty all electrolyte from sensor.
Flush sensor with distilled or deionized water to remove all particulate
matter.
Normally, the sensor should be recharged
with fresh electrolyte at three-month intervals. However, the interval may be extended, depending upon the application in
which the sensor is used. In general, correcting a low output can be accomplished
by recharging with fresh electrolyte, as described in Section 6-1a below. If output still
remains low, or the other symptoms exist,
Rosemount Analytical Inc. A Division of Emerson Process Management Routine Service and Troubleshooting 6-1
2. Place a piece of adhesive tape
over the breather hole in the pressure compensation diaphragm port
(not the slotted fill plug).
3. Examine cathode for:
a. Staining or uneven coloration.
which indicates that the cath-
Instruction Manual
748054-L
May 2003
ode should be rejuvenated as
described in Section 6-1b on
page 6-3.
b. Any deposited material, typi-
cally white to gray. present in
or around the grooves in the
plastic surrounding the cathode. This must be removed to
ensure best operation. Most
of these deposits are water-soluble and may be removed by a water jet from a
squeeze bottle. Any insoluble
deposits in the annular and
channel grooves may be removed with a toothpick; however, care must be used to
avoid deforming the grooves.
4. Disassemble the membrane assembly. This consists of a plastic
membrane, a holder, a retainer,
and an O-ring. Remove retainer
from holder by placing finger into
center hole of holder and pressing
fingernail against inner edge of
CAP
MEMBRANE
RETAINER
CELL SEPARATOR
(Accessory)
HOLDER
O-RING
Model 7003D
retainer. Remove and discard the
old membrane.
5. Verify that O-ring is properly positioned in associated groove in
holder.
6. Holding a single membrane by the
edges only, place it across membrane holder and snap retainer in
place. Membrane is now fixed in
proper position, between holder
and retainer.
CAUTION
MEMBRANE CONTAMINATION
Never touch center area of membrane with fingers. Membranes are
easily contaminated with foreign
substances. Contaminated membranes cause drifting or erratic readings.
BREATHER
HOLE
WASHER
CATHODE
O-RING
DIAPHRAGM
DIAPHRAGM
SCREW
Figure 6-1. Rechargeable Sensor – Exploded View
6-2 Routine Service and Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
7. Using a sharp razor blade, carefully
trim away excess membrane around
edge of membrane assembly. Take
care that razor blade does not cut
into edges of membrane assembly.
8. Set sensor body on a flat surface,
with cathode facing upward. Verify
that O-ring in groove at end of sensor body is properly positioned
around cathode. Pour the electrolyte
over the cathode/central post assembly so that it runs down into the
sensor electrolyte well. Fill the well to
a level flush with the top of the sidewall.
Instruction Manual
748054-L
May 2003
electrolyte displaced from the electrolyte well into the side port may
now be removed by blotting with a
tissue.
10. Insert new rubber diaphragm into
side port, place new washer over
diaphragm, and secure with side port
screw. Do not overtighten screw.
11. Inspect sensor for possible leaks or
damage to membrane.
12. Remove the adhesive tape installed
in Step 2 from the pressure compensation port.
Put the membrane assembly directly
onto the cathode so that the face of
the holder (i.e., the part with the larger-diameter central hole) fits against
the O-ring in the end of the sensor
body.
The membrane is now in place. It will
spread any electrolyte remaining on
the cathode into a thin film that wets
the entire surface. The membrane
assembly should now be centered
over the cathode.
9. Taking care not to disturb the central
orientation of the membrane assembly, carefully place the cap on the
sensor body. Screw the cap on, fingertight only. Then lay the sensor on
its side, with the side port up. Remove side port screw, rubber pressure-compensating diaphragm, and
washer. Add electrolyte, if necessary, to bring the level into the side
port and then rock the sensor from
end to end to remove any air pockets. Add electrolyte, if necessary, to
bring the level even with the shoulder. With the side port still facing up,
tighten the cap further until it is snug,
and the membrane is stretched taut
across the cathode. Any excess
Sensor is ready for operation. Connect
cable. If sensor does not operate properly
refer to Section 6-3 on page 6-4.
If normal operation is not obtained with
the specified recharging procedure, perform rejuvenation, as detailed in Section
6-1b below.
b. Rejuvenating Cathode
If simple recharging does not correct
symptoms of low output, clean and/or rejuvenate the cathode as follows:
WARNING
CORROSIVE MATERIAL
Concentrated nitric acid is used in the
following procedure. This material is
highly corrosive. Avoid contact with
skin, eyes, clothing, and precision instrument parts. Use rubber gloves and
eye protection. If body contact occurs,
flush copiously with water and seek
medical attention.
1. Disassemble sensor. Remove cap
and membrane assembly. Discard the
used electrolyte. Flush the sensor
Rosemount Analytical Inc. A Division of Emerson Process Management Routine Service and Troubleshooting 6-3
Instruction Manual
748054-L
May 2003
with distilled or deionized water to
remove all particulate material.
2. Over a sink, use a cotton swab to
treat the cathode with concentrated
reagent grade nitric acid, obtainable
from a laboratory supply house. Submerge the tip of the swab in the nitric
acid. Press the swab against the side
of the nitric acid container to remove
excess acid. Swab the cathode area
lightly for five minutes.
3. Take care to confine the nitric acid to
the button area. Only a thin film of nitric acid should be present on the surface of the cathode during the
cleaning operation. Excessive application may result in the destruction of
the epoxy annulus surrounding the
button, with resultant sensor failure.
4. Rinse the button and sensor cavity
thoroughly with distilled or deionized
water. Then rinse the sensor with
electrolyte by pouring it over the cathode into the sensor cavity until it is
filled. Discard this electrolyte.
Model 7003D
The condition may be corrected by installing a separator disk which mechanically
separates the cathode and anode (see
Figure 6-1 on page 6-2). The kit contains
ten disks.
To install the separator in an oxygen sensor, remove the cap and membrane
holder. pour out the electrolyte, and position the hole in the separator over the
cathode post. Then gently push the separator down into the sensor body until it
rests on top of the anode. Fill the sensor
with fresh electrolyte and reassemble in
the conventional manner.
6-2 NON-RECHARGEABLE SENSORS
The non-rechargeable sensor has no user accessible internal parts. No sensor maintenance is possible other than gentle brushing
to remove deposits on the face of the sensor
to restore its sensitivity and response time.
If the performance of the sensor has degraded beyond acceptable limits, it must be
replaced.
5. Recharge the sensor in the normal
fashion.
If normal operation is not obtained with the
specified rejuvenation procedure, the sensor is depleted and must be replaced.
c. Cell Separator Kit
The cell separator is used to eliminate
"spiking," i.e., non-oxygen-related transient response on the analyzer output.
This condition is usually caused by reaction products which slough off the anode
and are transported to the cathode surface, where they are reduced. During reduction, the anode accepts electrons,
resulting in the undesired transient response.
6-3 TROUBLESHOOTING
The most frequent fault is a progressive development of insensitivity of the sensor. During calibration, the CAL control must be
turned farther and farther clockwise to set the
meter to the desired calibration value. Finally,
after several months of operation, rotating the
CAL control to its clockwise limit fails to bring
the meter to the desired calibration setting.
The cause of this characteristic change is the
gradual exhaustion or occlusion of the sensor.
Accordingly, the sensor must be maintained
as described in Section 6-1 and Section 6-2.
System problems can be isolated to the sensor or the amplifier by the following procedure:
1. Disconnect AC power from the instrument.
6-4 Routine Service and Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
2. Set alarm switches S4 and S5 (Figure 3-1
on page 3-3) to OFF position.
3. On TB2 (Figure 2-1 on page 2-2) make
following connections:
a. Disconnect all leads of sensor cable.
b. Connect a 10K ohm resistor across
the three l0K THMS terminals. This
resistor simulates the resistance of
the thermistor at 25°C.
c. If the sensor is rechargeable, connect
a 10 K ohm resistor across terminals
marked AN (for anode) and CATH (for
cathode). If the sensor is
non-rechargeable, connect a 320 K
ohm resistor across the terminals.
This resistor permits the polarizing
voltage supply to provide a small current that simulates the sensor output
signal.
4. Set RANGE switch to CAL (Figure 3-1 on
page 3-3).
5. Rotate CAL control throughout its range
to verify that the outputs of the instrument
respond from near zero to above fullscale.
If test yields correct results, amplifier is operational. The fault is in sensor or cable. Remove
10 K ohm resistors from TB2 and proceed to
tests given in Section 6-3a below or 6-3b on
page 7.
Instruction Manual
748054-L
May 2003
a. Checking Rechargeable Sensor and
Cable
For convenience, the sensor cable may
be regarded as part of the sensor. Accordingly, electrical checks are made at
the terminal ends of the cable, disconnected from the amplifier at TB2. Verification of electrical integrity of the sensor is
determined by making the following
checks with an ohmmeter:
1. Resistance between white and green
leads should be 10 ohms +1% at
25°C (approximately 30 ohms at 0°C,
and 4 ohms at 50°C). Readings from
white or green lead to any other lead
of the sensor cable should indicate
open circuit, or at least 100M ohms.
Readings less than this indicate a
shunt resistance path which produces
an error in the measurement.
2. The black and white lead is connected to the grounding shield. Readings from this lead to any other lead
of the sensor cable should be at least
100M ohms.
3. If checks in Steps I or 2 indicate trouble, or if other symptoms indicate the
sensor to be faulty, determine the
probable cause by reference to the
appropriate symptom(s) in Table 6-1
on page 6-6. This table is a compilation of the most common sensor problems.
If these tests do not yield correct results, contact a factory authorized service representative.
Rosemount Analytical Inc. A Division of Emerson Process Management Routine Service and Troubleshooting 6-5
Instruction Manual
748054-L
May 2003
Model 7003D
SYMPTOM PROBABLE CAUSE CORRECTIVE ACTION
Abnormally high oxygen readings
(unable to calibrate)
Abnormally low oxygen readings
(unable to calibrate)
Sensor noisy (motion sensitive)
Upscale reading with known oxygen-
"Contamination" may be the normal accumulation resulting from long-term operation, indicative that the standard cell-
rejuvenation procedure is required.
Rosemount Analytical Inc. A Division of Emerson Process Management Routine Service and Troubleshooting 6-7
Instruction Manual
748054-L
May 2003
Model 7003D
6-8 Routine Service and Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
SECTION 7
REPLACEMENT PARTS
WARNING
PARTS INTEGRITY
Tampering or unauthorized substitution of
components may adversely affect safety of
this product. Use only factory documented components for repair.
7-1 CIRCUIT BOARD REPLACEMENT POLICY
In most situations involving a malfunction of a
circuit board, it is more practical to replace the
Instruction Manual
748054-L
May 2003
board than to attempt isolation and replacement of the individual component, as the cost
of test and replacement will exceed the cost of
a rebuilt assembly. As standard policy, rebuilt
boards are available on an exchange basis.
Because of the exchange policy covering circuit boards, the following list does not include
individual electronic components. If circumstances necessitate replacement of an individual component, which can be identified by
inspection or from the schematic diagrams,
obtain the replacement component from a local source of supply.
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-3
Instruction Manual
748054-L
May 2003
Model 7003D
7-4 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7003D
8-1 RETURN OF MATERIAL
If factory repair of defective equipment is required, proceed as follows:
1. Secure a return authorization from a
Rosemount Analytical Inc. Sales Office or
Representative before returning the
equipment. Equipment must be returned
with complete identification in accordance
with Rosemount instructions or it will not
be accepted.
Rosemount CSC will provide the shipping
address for your instrument.
In no event will Rosemount be responsible for equipment returned without proper
authorization and identification.
2. Carefully pack the defective unit in a
sturdy box with sufficient shock absorbing
material to ensure no additional damage
occurs during shipping.
3. In a cover letter, describe completely:
•The symptoms that determined the
equipment is faulty.
•The environment in which the equip-
ment was operating (housing,
weather, vibration, dust, etc.).
•Site from where the equipment was
removed.
•Whether warranty or non-warranty
service is expected.
•Complete shipping instructions for the
return of the equipment.
4. Enclose a cover letter and purchase order
and ship the defective equipment according to instructions provided in the Rosemount Return Authorization, prepaid, to
the address provided by Rosemount CSC.
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
Instruction Manual
748054-L
May 2003
SECTION 8
RETURN OF MATERIAL
If warranty service is expected, the defective
unit will be carefully inspected and tested at
the factory. If the failure was due to the conditions listed in the standard Rosemount warranty, the defective unit will be repaired or
replaced at Rosemount’s option, and an operating unit will be returned to the customer in
accordance with the shipping instructions furnished in the cover letter.
For equipment no longer under warranty, the
equipment will be repaired at the factory and
returned as directed by the purchase order
and shipping instructions.
8-2 CUSTOMER SERVICE
For order administration, replacement Parts,
application assistance, on-site or factory repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
8-3 TRAINING
A comprehensive Factory Training Program of
operator and service classes is available. For
a copy of the Current Operator and Service Training Schedule contact the Technical Ser-
vices Department at:
Rosemount Analytical Inc.
Customer Service Center
1-800-433-6076
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 8-1
Instruction Manual
748054-L
May 2003
Model 7003D
8-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management
Model 7001D, 7002D, 7003D
CONVERTING RECHARGEABLE SENSOR TO
NON-RECHARGEABLE SENSOR
Instruction Sheet
748597-G
October 2000
To convert a Model 7001D, 7002D or 7003D
Oxygen Monitor which has been used with a
rechargeable oxygen sensor so that it can be
used with a non-rechargeable oxygen sensor,
it is necessary to modify the display board of
the instrument. This modification changes the
gain of the monitor to match the output current
available from the non-rechargeable oxygen
sensor. A header kit appropriate to each
model is provided for use in the modification.
To perform the modification to your
instrument, select the appropriate instruction
below.
7001D
1. Remove the existing header in J3 on the
display board. Mark it and save for
possible future re-conversion to use of a
rechargeable sensor.
2. Install the 14 pin header from the kit (P/N
623731) into J2 on the display board.
Refer to Figure 1. Verify that pin 1 on the
header is installed in the pin 1 socket on
the display board.
3. Install the 16 pin header from the kit into
J3 on the display board. Verify that pin 1
on the header is installed in the pin 1
socket on the display board.
7002D
1. Remove the existing header in J4 on the
display board. Mark it and save for
possible future re-conversion to use of a
rechargeable sensor.
2. Install the 14 pin header from the kit (P/N
623732) into J2 on the display board.
Refer to Figure 1. Verify that pin 1 on the
header is installed in the pin 1 socket on
the display board.
3. Install the 16 pin header from the kit into
J4 on the display board. Verify that pin 1
on the header is installed in the pin 1
socket on the display board.
Figure 1. Models 7001D and 7002D Display
7003D
1. Remove the existing header in J5 on the
2. Install the 14 pin header from the kit (P/N
Figure 2. Model 7003D Display Board
J5
Board
display board. Mark it and save for
possible future re-conversion to use of a
rechargeable sensor.
623733) into J5 on the display board.
Refer to Figure 2. Verify that pin 1 on the
header is installed in the pin 1 socket on
the display board.
J2
J3
J4
http://www.processanalytic.com
Electrolyte
PRODUCT:
Oxygen Sensor Electrolyte
PART NUMBER: 192580 (1 pint, 474 ml)
24 HOUR EMERGENCY TELEPHONE NUMBER:
CHEMTREC (800) 424-9300
SECTION I - GENERAL
Distributor: Rosemount Analytical Inc.
1201 N. Main St., Orrville, Ohio 44667-0901
330-682-9010
Chemical name and synonyms
Trade name and synonyms
Chemical family
Formula
CAS Number
SECTION II – HAZARDOUS INGREDIENTS
Hazardous mixtures of other liquids,
solids or gases
SECTION III – PHYSICAL DATA
Boiling point
Melting point
Vapor pressure
Vapor density (air=1)
Specific gravity (H2O=1)
% Volatile by volume
Evaporation rate (H2O=1)
Solubility in water
Appearance and odor
SECTION IV – FIRE AND EXPLOSION HAZARD DATA
Flash point
Extinguishing media
Special fire fighting procedures
Unusual fire and explosion hazards
SECTION V – REACTIVITY DATA
Stability
Conditions to avoid
Incompatibility (material to avoid)
Hazardous decomposition or byproducts
Hazardous polymerization
potassium chloride/water solution
electrolyte
metal halide salt solution
KCl + H
potassium chloride: 7447-40-7 water: 7732-18-5
none
100°C
NA
24 mm Hg
0.5
1
90+
1
water soluble
clear, odorless solution
NA
NA
NA
NA
stable
none
none
none
none
1
O (5% KCl, 95% H2O)
2
Material Safety Data Sheet
748595-J
July 1997
1
Is contained in the following sensors: 190404, 190408, 1904 09, 623246, 623370, 623371, 623372, 623373, 623374,
623375, 623740, 623741, 623742
http://www.processanalytic.com
Material Safety Data Sheet
748595-J
July 1997
SECTION VI – HEALTH HAZARD DATA
Threshold limit value
Routes of entry
Effects of overexposure
Emergency & first aid procedures
SECTION VII – SPILL OR LEAK PROCEDURE
Steps to be taken in case material is
released or spilled
Waste disposal method
SECTION VIII – SPECIAL PROTECTION INFORMATION
Respiratory protection
Ventilation
Protective gloves
Eye protection
Other protective equipment
SECTION IX – SPECIAL PRECAUTIONS
Precautions to be taken in handling
and storing
Other precautions
SECTION X – TRANSPORTATION
Must be compliance with federal, state and local regulations
NA
NA
NA
NA
NA
no special requirement
none
mechanical (general)
NA
safety glasses suggested
none
None
none
Electrolyte
NOTICE
WHILE ROSEMOUNT ANALYTICAL BELIEVES THE INFORMATION CONTAINED HEREIN IS VALID AND
ACCURATE, ROSEMOUNT ANALYTICAL MAKES NO WARRANTY OR REPRESENTATION AS TO ITS VALIDITY, ACCURACY, OR CURRENCY. ROSEMOUNT ANALYTICAL SHALL NOT BE LIABLE OR OTHERWISE RESPONSIBLE IN ANY WAY FOR USE OF EITHER THIS INFORMATION OR THE MATERIAL TO
WHICH IT APPLIES. DISPOSAL OF HAZARDOUS MATERIAL MAY BE SUBJECT TO FEDERAL, STATE,
OR LOCAL LAWS AND/OR REGULATIONS.
2 of 2 MSDS - Electrolyte
Rosemount Analytical Inc. A Division of Emerson Process Management
WARRANTY
Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from
defects in workmanship and material under normal use and service for a period of twelve (12)
months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid
junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by
Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or
material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the
goods, part(s) or consumables are returned to Seller's designated factory, transportation charges
prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in
the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in
effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day
warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such commercial unit when such goods, part(s) and
consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly
or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from
any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS,
SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED
WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy.
LAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD
WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF ACTION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC
GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER
AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDENTAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT
ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT.
IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER
NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS
COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO
BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED
HEREUNDER.
Force Majeure.
Seller's direct control.
Seller shall not be liable for failure to perform due to labor strikes or acts beyond
SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DE-
Instruction Manual
748054-L
May 2003
Model 7003D
EMERSON PROCESS MANAGEMENT
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
E gas.csc@emersonprocess.com
ASIA - PACIFIC
Fisher-Rosemount
Singapore Private Ltd.
1 Pandan Crescent
Singapore 128461
Republic of Singapore
T 65-777-8211
F 65-777-0947