Rosemount 396R, 396RVP Instruction Manual

Page 1
Model 396R and 396RVP
Retractable pH/ORP Sensors
Instruction Manual
PN 5100396R
January 2001
Page 2
WARNING
Retractable sensors must not be inserted nor retracted when process pressures are in excess of 64 psig (542kPa), option 21 or 35 psig (343 kPa) option 25.
CAUTION
SENSOR/PROCESS
APPLICATION COMPATIBILITY
The wetted sensor materials may not be compatible with process com­position and operating conditions. Application compatibility is entirely the responsibility of the user.
DANGER
HAZARDOUS AREA INSTALLATION
Installations near flammable liquids or in haz­ardous area locations must be carefully eval­uated by qualified on site safety personnel. This sensor is not
Intrinsically Safe or
Explosion Proof. To secure and maintain an intrinsically safe
installation, the certified safety barrier, transmitter, and sensor combination must be used. The installation system must com­ply with the governing approval agency (FM, CSA or BASEEFA/CENELEC) haz­ardous area classification requirements. Consult your analyzer/transmitter instruction manual for details.
Proper installation, operation and servicing of this sensor in a Hazardous Area Instal­lation is entirely the responsibility of the user.
SS-SE
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures, and tests its products to meet many national and international stan­dards. Because these instruments are sophisticated tech­nical products, you must properly install, use, and main­tain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product. If this Instruction Manual is not the correct manual, telephone 1-800-654-7768 and the requested manual will be provided. Save this Instruction Manual for future reference.
• If you do not understand any of the instructions, contact your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper instal­lation, operation, and maintenance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qual­ified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance and place the safe operation of your process at risk. Look alike substitu­tions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protec­tive covers are in place, except while maintenance is being performed by qualified persons, to prevent elec­trical shock and personal injury.
Rosemount Analytical Inc. Uniloc Division
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 863-1181
© Rosemount Analytical Inc. 2001
Page 3
MODEL 396R pH/ORP TABLE OF CONTENTS
MODEL 396R AND 396RVP RETRACTABLE
pH/ORP SENSORS
TABLE OF CONTENTS
Section Title Page
1.0 DESCRIPTION AND SPECIFICATIONS............................................................. 1
1.1 Features and Applications................................................................................ 1
1.2 Performance and Physical Specifications ........................................................ 1
1.3 Ordering Information......................................................................................... 2
2.0 INSTALLATION.................................................................................................. 5
2.1 Unpacking and Inspection................................................................................ 5
2.2 Mechanical Installation .................................................................................... 5
3.0 WIRING MODEL 396R...................................................................................... 13
4.0 WIRING MODEL 396RVP.................................................................................. 22
5.0 START UP AND CALIBRATION......................................................................... 29
5.1 Start up.............................................................................................................. 29
5.2 pH Calibration................................................................................................... 29
5.3 ORP Calibration ................................................................................................ 30
6.0 MAINTENANCE................................................................................................. 31
6.1 Maintenance ..................................................................................................... 31
6.2 Sensor Removal................................................................................................ 31
6.3 pH Electrode Cleaning...................................................................................... 31
6.4 Platinum Electrode Cleaning............................................................................. 32
6.5 Automatic Temperature Compensation............................................................. 32
6.6 Sensor Tube Replacement................................................................................ 32
7.0 DIAGNOSTIC AND TROUBLESHOOTING ....................................................... 35
7.1 Diagnostics and Troubleshooting with Model 54/3081 pH/ORP Diagnostics... 35
7.2 Troubleshooting without Diagnostics................................................................ 36
8.0 RETURN OF MATERIAL.................................................................................... 37
LIST OF TABLES
Table No. Title Page
1-1
Commonly Used Accessories for Model 396R................................................. 3
1-2 Other Accessories for Model 396R................................................................... 3
1-3 Commonly Used Accessories for Model 396RVP............................................. 4
1-4 Other Accessories for Model 396RVP............................................................... 4
5-1 ORP of Saturated Quinhydrone Solutions......................................................... 30
6-1 Ro and R1 Values for Temperature Compensation Elements........................... 32
6-2 Temperature vs. Resistance of Auto T.C. Elements .......................................... 32
7-1 Troubleshooting with Diagnostics..................................................................... 35
7-2 Troubleshooting without Diagnostics................................................................ 36
i
Page 4
MODEL 396R pH/ORP TABLE OF CONTENTS
LIST OF FIGURES
Figure No. Title Page
2-1 Exploded View of Ball Valve Kit PN 23240-00 used with process connector
PN 23166-00 (or PN 23166-01)......................................................................... 6
2-2 Typical Mounting Configurations for Model 396R............................................. 7
2-3 Typical Mounting Configurations for Model 396RVP......................................... 7
2-4 Dimensional Drawing — Model 396R with Optional Ball Valve PN 23765-00 .. 8 2-5 Dimensional Drawing — Model 396R with Optional Ball Valve PN 23240-00 .. 9
2-6 Dimensional Warning Label for Hemi Bulb Sensors and Sensor Diagram....... 10
2-7 Dimensional Drawing — Model 396RVP with Optional 1-1/2 inch Ball Valve ..
PN 23240-00..................................................................................................... 11
2-8 Dimensional Drawing — Model 396RVP with Optional 1-1/4 inch Ball Valve ......
PN 23765-00..................................................................................................... 12
3-1 Cable Preparation Instructions for Model 396R................................................ 13
3-2 Wiring Details Model 396R-54 for use with Model 54/81/3081 pH/ORP........... 14
3-3 Wiring Details Model 396R-54 with Remote Junction Box & Pr eamp (PN 23555-00) 14 3-4 Wiring Details Model 396R-50 for use with Remote Junction Box (PN 23309-03). 15 3-5 Wiring Details Model 396R-54 for use with Remote Junction Box (PN 23309-04). 15
3-6 Wiring Details-Model 396R-54 for use with Model 1181 pH/ORP......................... 16
3-7 Wiring Details-Model 396R-54 for use with Models 1054A/B pH/ORP, 2054 pH, ..
and 2081 pH/ORP.............................................................................................. 16
3-8 Wiring Details-Model 396R-54 for use with Model SCL- (P/Q).............................. 17
3-9 Wiring Details-Model 396R-54 for use with Model 2700 Pr eamp J-Box (PN 23054-03) 17
3-10 Wiring Details-Model 396R-50/54-60 for use with Sensor Head J-Box ............ 18
3-11 Wiring Details-Model 396R-54-61 for use with Sensor Head J-Box ................. 18
3-12 Wiring Details-Model 396R-50 for use with J-Box (PN 23707-00) to................
Models 1181, 1050, 1060, 1030, 1023 pH Transmitters .................................. 19
3-13 Wiring Details-Model 396R-54 for use with J-Box (PN 23708-01) to ...............
Models 1054series, 2054, and 2081 pH Transmitters ...................................... 20
3-14 Wiring Model 396R-( )-54-62-(71) to Model 1055-10-22-32 ............................. 21
3-15 Wiring Model 396R-(02)-( )-54/55-(61) to Model 1055-10-22-32...................... 21
4-1 Wire Functions and Pin Connections for Model 396RVP.................................. 22
4-2 Wiring Model 396RVP to Model 81 ................................................................... 23
4-3 Wiring Model 396RVP to Model 1181 ............................................................... 23
4-4 Wiring Model 396RVP to Model 81 thru a Remote J-Box.................................. 23
4-5 Wiring Model 396R VP to Models 1181/1050/1060/1003/1023 thru Remote J-Box 23
4-6 Wiring Model 396RVP to Model 2081 ............................................................... 24
4-7 Wiring Model 396RVP to Models 3081 and 4081 ............................................. 24
4-8 Wiring Model 396RVP to Model 2081 thru a Remote J-Box.............................. 24
4-9 Wiring Model 396RVP to Models 3081 and 4081 thru a Remote J-Box............ 24
4-10 Wiring Model 396RVP to Model 1054............................................................... 25
4-11 Wiring Model 396RVP to Models 1054A/B and 2054........................................ 25
4-12 Wiring Model 396RVP to Model 1054 thru a Remote J-Box.............................. 25
4-13 Wiring Model 396RVP to Models 1054A/B and 2054 thru a Remote J-Box...... 25
4-14 Wiring Model 396RVP to Model 54................................................................... 26
4-15 Wiring Model 396RVP to Model 54 thru a Remote J-Box.................................. 26
4-16
Wiring Model 396RVP to Model 2700............................................................... 26
4-17 Wiring Model 396RVP to Model SCL-(P/Q)....................................................... 26
4-18 Wiring Model 396RVP to Model 1055pH/pH..................................................... 27
4-19 Wiring Model 396RVP to Model 1055pH/pH thru a Remote J-Box................... 27
4-20
Wiring Model 396RVP to Model 1055-10-22-32................................................ 28
6-1 Sensor Tube Replacement................................................................................ 34
6-2 Male Connector Tightening Diagram................................................................ 34
ii
Page 5
1.1 FEATURES AND APPLICATIONS The Model 396R and 396RVP Sensorsare specifical-
ly designed for improv ed life in harsh, dirty applications where a separate sample stream is difficult to provide and greater insertion depths are required. Model 396R is designed for use with a 1-1/4 in.or 1-1/2 in.ball valve for hot tap installation.The Model 396R is constructed of molded polypropylene housed in a titanium tube with EPDM seals to provide maximum chemical resistance.
Model 396R also features a titanium solution ground for advanced sensor diagnostics when used with the Model 54, 81, or 3081 pH/ORP Analyzer/Transmitter. Advanced sensor diagnostics provide preventative maintenance by notifying the operator for replacement and cleaning of an aged or fouled sensor for continu­ous optimum performance.
The sensor also features a shrouded tip for protection from breakage while allowing process to flow by the glass electrode for accurate and reliable pH measurement.
The Model 396R is available without an integral pre­amplifier only and 15 ft or 9.5 in. of integral high quali­ty 9 conductor cable.The preamplifier must be housed in a remote location or in a J-Box kit for attachment at the rear, cable end of the sensor (order separately) or integral to the Analyzer/Transmitter. The Model 396R is compatible with all Rosemount Analytical and vari­ous other manufacturers instruments.
Model 396RVP: Rosemount Analytical has recently released Model 396RVP.This model has identical per­formance and physical specifications to the Model 396R (see Section 1.2) with the following exception: the Model 396RVP has a Variopol (VP) connector on the back end of the sensor in place of a cable.
MODEL 396R pH/ORP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
MODEL 396R
Measured Range:
ORP: -1500 to 1500mV pH:GPLR A
CCUGLASS
Percent Linearity Over pH Range:
Wetted Materials: Polypropylene, EPDM, titanium, glass,
(platinum: ORP only)
Process Connections: 1-1/2 or 1-1/4 in. with ball valve,
1 in. without ball valve
Cable: Integral 15 ft or 9.5 in. 9 conductor cable except
option 60 (9.5 in. coaxial cable with BNC) Recommended Interconnect (PN 9200273)
Maximum Process Pressure and Temperature:
Hemi bulb: 150 psig (1136 kPa abs) at 212°F (100°C )
Flat bulb: 100 psig (790 kPa abs) at 212°F (100°C )
Maximum Pressure at Retraction or Insertion:
Code 21: 64 psig (542 kPa abs) Code 25: 35 psig (343 kPa abs)
Minimum Conductivity: 100 µS/cm Weight/Shipping Weight:
Sensor: Code 21: 2.0 lb/3.0 lb (.9 kg/1.40 kg)
Code 25: 3.0 lb/4.0 lb (1.40 kg/1.80 kg)
Ball Valve: PN 23240-00; 5 lb/7 lb (2.25 kg /3.20 kg)
PN 23634-00 8 lb/10 lb (3.65 kg/4.55 kg)
J-Box: 3 lb/4 lb (1.40 kg/1.80 kg)
MODEL 396RVP
Measured Range:
ORP: -1500 to 1500mV pH:0 to 14
Available pH glass types:GPLR hemi bulb or flat bulb Wetted Materials: Polypropylene, EPDM, titanium, glass,
(platinum: ORP only)
Process Connections: none, use 1-inch process connector
or ball valve kit (1-1/2 inch or 1-1/4 inch)
Temperature Range: 0 to 100C (32 to 212F) Pressure Range (hemi bulb): 100-1136 kPa abs (0-150 psig) Pressure Range (flat bulb): 100-790 kPa abs (0-100 psig) Maximum Pressure at Retraction or Insertion:
Code 21: 64 psig (542 kPa abs) Code 25: 35 psig (343 kPa abs)
Minimum Conductivity: 75 µS/cm, nominal Preamplifier options: remote Weight/Shipping Weight:
Sensor: Code 21: 2.0 lb/ 3.0 lb (.9 kg/1.40 kg)
Code 25: 3.0 lb/4.0 lb (1.40 kg/1.80 kg)
Ball Valve: PN 23240-00; 5 lb/7 lb (2.25 kg /3.20 kg)
Hemi Bulb Flat Bulb
1-2 pH 94% 93%
2-12 pH 99% 98%
12-13 pH 97% 95%
1
1.2 PERFORMANCE AND PHYSICAL SPECIFICATIONS
Page 6
2
MODEL 396R pH/ORP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.3 ORDERING INFORMATION
The Model 396R Sensor is housed in a titanium tube, with a patented polypropylene reference junction and tita- nium solution ground for use with a ball valve (order separately) for hot tap applications.The sensor is available with either a hemi or flat glass pH electrode and features a shrouded glass/platinum electrode and PT100 or 3K temperature compensation. The 396R is available with 9.5 in. or 15 ft of integral cable. The 396R sensor is not available with a preamp. Junction box kits with preamps must be ordered separately if the analyzer/transmitter does not have an integral preamp within 15 ft.of the probe. Process connector and ball valve assemblies m ust also be ordered separately.
MODEL
396R TUpH RETRACTABLE pH SENSOR
CODE MEASURING ELECTRODE TYPE (Required Selection)
10 GPLR hemi glass, General Purpose Low Resistivity (0-14 pH) 12 ORP 13 GPLR flat glass, General Purpose Low Resistivity (0-14 pH)
CODE SENSOR LENGTH
21 21 in. Titanium Tube 25 36 in. Titanium Tube
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For Models 1181 (3K TC) 54 For Models 1054, 1054A/B, 2054, 2081,54, 3081, 81, SCL-(P/Q), Solu Cube (PT 100 RTD)
CODE OPTIONAL OPTIONS
60 9.5 in. Cable with BNC (for use with Model 1181, 1054 series, 2054, 2081 Sensor Head J-Boxes) 61 9.5 in. Cable no BNC (Not Valid w/Option 50) (for use with Model 54, 81, 3081 Sensor Head J-Boxes)
396R - 10 - 21 - 54 EXAMPLE
MODEL 396RVP TUpH RETRACTABLE pH/ORP SENSOR
CODE MEASURING ELECTRODE TYPE (Required Selection)
10 Hemi bulb, General Purpose Low Resistivity (0-14 pH) 12 ORP 13 Flat, GPLR glass
CODE SENSOR LENGTH (Required Selection)
21 21 in. Titanium Tube 25 36 in. Titanium Tube
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For Models 1181 (3K TC) 54 For Models 1054, 1054A, 1054B, 2054, 2081,54, 3081, 4081, 81, SCL-(P/Q), 2700 (Pt 100 RTD)
The Model 396RVP ball valve retractable sensor features a gel-filled electrolyte solution with the large area, coating resistant TUpH polypropylene reference junction and a standard hemi or optional flat glass bulb. Model 396RVP is housed in a Titanium sensor tube and can be mounted directly into the process using a 1 in. MNPT threaded process connector and a ball valve assembly kit (both ordered separately). It is offered with the water­tight Variopol sensor-to-cable connector and uses the mating connector cable (ordered separately). Also available is a choice of temperature element, 3 K Balco or Pt 100 RTD. A remote preamplifier found in the analyzer/trans­mitter or in a junction box (ordered separately) must be used with this sensor for a reliable signal transmission.
396RVP- 10 - 21 - 54 EXAMPLE
Page 7
TABLE 1-2. OTHER ACCESSORIES FOR MODEL 396R
TABLE 1-1. COMMONLY USED ACCESSORIES FOR MODEL 396R
PART DESCRIPTION 22698-00 Preamplifier plug-in for J-box, for Model 1003, 22698-02 Preamplifier plug-in for J-box, for Models 1181/1050 22698-03 Preamplifier plug-in for J-box, for Models 1054A/B, 2054, 2081 23550-00 Remote Junction box with extension board 9550167 O-ring, 2-214, EPDM for process connector 9210012 Buffer solution, 4.01 pH, 16 oz 9210013 Buffer solution, 6.86 pH, 16 oz 9210014 Buffer solution, 9.18 pH, 16 oz 22743-01 Pt100 preamp for Model 1181 22744-01 3K Preamp for Model 1181 23557-00 Preamplifier for junction box for Models 54/3081
MODEL 396R pH/ORP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
3
For first time installations, Rosemount Analytical recommends using the following guide
1. Retractable Mounting
AA.. CChhoooossee oonnee ((rreeqquuiirreedd ffoorr aallll ffiirrsstt ttiimmee iinnssttaallllaattiioonnss))::
PN 23166-00, 1 in. x 1 in. NPT process connector, 316 SST PN 23166-01, 1 in. x 1 in. NPT process connector, Titanium
BB.. CChhoooossee oonnee::
PN 23240-00, 1-1/2 in. ball valve assembly, 316 SST PN 23765-00, 1-1/4 in. ball valve assembly, 316 SST
2. Junction Boxes (Optional; Choose either Sensor Head or Remote)
AA.. SSeennssoorr HHeeaadd JJuunnccttiioonn BBooxxeess ((uusseedd wwiitthh ooppttiioonnss -6600 oorr -6611 sseennssoorr)) - CChhoooossee oonnee::
PN 23709-00; includes preamplifier for Models 54, 81, 3081 PN 23708-01; includes preamplifier for Models 1054 series, 2054, 2081 PN 23707-00; includes preamplifier for Model 1181
BB.. RReemmoottee JJuunnccttiioonn BBooxxeess ((uusseedd wwiitthh ssttaannddaarrdd 1155 fftt.. ccaabbllee lleennggtthh sseennssoorr)) - CChhoooossee oonnee::
PN 23555-00; includes preamplifier for Models 54, 81, 3081 PN 23309-03; includes preamplifier for Model 1181 PN 23309-04; includes preamplifier for Models 1054 series, 2054, 2081 PN 23054-03; includes preamplifier for Solu Cube Model 2700
3. BNC Adapter - Choose one:
PN 9120516, BNC Adapter for use with remote junction boxes PN’s 23309-03 and 23309-04 Order option -60 (standard with BNC connector) for PN 23707-00 or 23708-01 sensor head junction boxes
4. Extension Cables - Choose one:
PN 23646-01, 11 conductor, shielded, prepped PN 9200273, 11 conductor, shielded, unprepped
Page 8
TABLE 1-4. OTHER ACCESSORIES FOR MODEL 396RVP
PART DESCRIPTION 22698-00 Preamplifier plug-in for junction box, for Model 1003, 22698-02 Preamplifier plug-in for junction box, for Models 1181, 1050 22698-03 Preamplifier plug-in for junction box, for Models 1054A/B, 2054, 2081 22743-01 Pt100 preamplifier for Model 1181 22744-01 3K Preamplifier for Model 1181 23557-00 Preamplifier for junction box for Models 54, 3081, 81, 4081 33046-00 Ferrule, 1 in., split 316SS 9310096 Nut, swage, 1 in. 316SST 9210012 Buffer solution, 4.01 pH, 16 oz 9210013 Buffer solution, 6.86 pH, 16 oz 9210014 Buffer solution, 9.18 pH, 16oz R508-160Z ORP solution, 460 mv ± 10 at 20°C 9550167 EPDM O-ring for Process Connector (PN 23166-00 or 23166-01) 12707-00 Jet Spray Cleaner
1. Variopol Cable (required for all first time installations)
Choose one: PN 23645-06, 15 ft cable with mating VP connector, prepped with BNC on analyzer end
PN 23645-07, 15 ft cable with mating VP connector, prepped without BNC on analyzer end*
2. Retractable Mounting
1A. Choose one (required for all first time installations, except as noted):
PN 23166-00 1 in. x 1 in. NPT process connector, 316 SST PN 23166-01 1 in. x 1 in. NPT process connector, Titanium
Choose one (optional process connector o-rings):
PN 9550220, Kalrez o-ring, 2-214 PN 9550099, Viton o-ring, 2-214
1B. Choose one:
PN 23240-00 1-1/2 in. ball valve assembly, 316 SST PN 23765-00 1-1/4 in. ball valve assembly, 316 SST (process connector not needed)
3. Remote Junction Boxes (Optional)
Choose one: PN 23555-00 includes preamplifier for Models 54, 81, 3081, 4081
PN 23309-03 and PN 22698-02 plug-in preamplifier for Model 1181Analyzer PN 23309-04 and PN 22698-03 plug-in preamplifier for Models 1054 series, 2054, 2081 Analyzers PN 23054-03 includes preamplifier for Solu Cube Model 2700
4. Extension cables
Choose one: PN 23646-01, 11 conductor, shielded, prepped
PN 9200273, 11 conductor, shielded, unprepped
FOR FIRST TIME 396RVP AND 398RVP INSTALLATIONS, ROSEMOUNT ANALYTICAL RECOMMENDS USING THE FOLLOWING GUIDE:
* Used for connections to Models 1181, 1054, 2081, 54, 81, 3081, 4081, and remote junction box PN 23555-00.
MODEL 396R pH/ORP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
4
TABLE 1-3. COMMONLY USED ACCESSORIES FOR MODEL 396RVP
Page 9
5
MODEL 396R pH/ORP SECTION 2.0
INSTALLATION
SECTION 2.0
INSTALLATION
2.1 UNPACKING AND INSPECTION. Inspect the out-
side of the carton for any damage. If damage is detected, contact the carrier immediately. Inspect the instrument and hardware. Make sure all items in the packing list are present and in good condition. Notify the factory if any part is missing.
NOTE
If the sensor is to be stored, the protective boot should be filled with either KCl elec­trolyte solution or pH 4.0 buffer solution and replaced on sensor tip until ready to use.
NOTE
Save the original packing cartons and materials as most carriers require proof of damage due to mishandling, etc. Also, if it is necessary to return the instrument to the factory, you must pack the instrument in the same manner as it was received. Refer to Section 6.0 for instructions.
WARNING
Glass electrode must be wetted at all times (in storage and in line) to maximize sensor life.
2.2 MECHANICAL INSTALLATION. The Model 396R Sensor may be installed through a weldalet or in a pipe tee or “Y”, as shown in Figure 2-1, when used with a ball valve. Insert the end of the sensor to a depth sufficient to ensure that the glass bulb is con­tinuously wetted by the process fluid. The Model 396R can also be inserted directly into the process without the use of a ball valve for applications not requiring continuous operation during sensor maintenance.
CAUTION
Allow sufficient room for safe retraction and insertion of the sensor. Personnel should have room for stable footing while performing removal or insertion of the sen­sor.
The sensor must be mounted within 10-90 degrees of the horizontal with the tip pointed downward, thus keeping air bubbles off of the pH sensitive glass bulb.
Bubbles settled on the glass bulb disrupt the electri­cal continuity between the pH sensitive glass and the silver/silver chloride measuring element.
If the retraction version is to be installed without a ball valve follow the installation procedure for insertion service (Section 2.2.2). Perform the following steps for sensor installation through a ball valve:
2.2.1 INSTALLATION THROUGH BALL VALVE.
1. Carefully remove the liquid filled rubber boot which protects the glass electrode and keeps the liquid junction wet during shipping and storage. Discard the liquid and boot. Make sure the lubri­cated O-ring is in place in the groove inside the male connector on the sensor body (Figure 4-1, item A).
CAUTION
Buffer solution, in the protective boot, may cause skin or eye irritation.
2. With the male connector on the sensor’s body, insert the sensor into the ball valve until it gently touches the closed valve. The molded electrode guard will protect the glass bulb from breakage.
3. Thread the male connector body tightly into the ball valve assembly. DO NOT tighten the hex nut on the male connector body; doing so would not allow the sensor to be inserted through the ball valve.
4. Pull back hard on the sensor assembly, as if try­ing to remove the sensor, to be certain that the sensor cannot come free of the ball valve assem­bly . The built-in r etraction stop will butt against the shoulder of the male connector if properly installed.
CAUTION
The sensor must be captured by the valve assembly and the male connector so that it cannot be blown free by process pres­sure if mishandled during insertion or retraction.
5. After confirming that the sensor assembly is prop­erly secured by the valve assembly, the valve may be opened and the sensor positioned into the process at the desired depth and orientation.
Page 10
6. While holding the sensor in position, tighten the hex nut of the male connector to firmly secure the sen­sor in place. When the hex nut is tightened, the Teflon ferrule inside the compression fitting clamps the sensor tube (see Figure 4-2).
NOTE
A stainless steel ferrule is available if the Teflon ferrule does not inadequately grip. When using the metallic ferrule, care must be taken to avoid over tightening and damaging the sensor tube. If the male connector leaks during insertion or retraction, replace the O­ring in the male connector.
2.2.2 INSTALLATION WITHOUT A BALL VALVE. The Model 396R Sensor may be installed through a weldalet or pipe tee or “Y” when used with a process connector (PN 23166-00 or 23166-01). The sensor should be installed within 80° of vertical, with the electrode facing down.
MODEL 396R pH/ORP SECTION 2.0
INSTALLATION
CAUTION
Over tightening the hex nut may damage the ferrule.
6
FIGURE 2-1. Exploded View of Ball Valve Kit PN 23240-00
used with process connector PN 23166-00 (or PN 23166-01)
Page 11
7
MODEL 396R pH/ORP SECTION 2.0
INSTALLATION
JUNCTION BOX IS OPTIONAL
FIGURE 2-2.Typical Mounting Configurations for Model 396R
FIGURE 2-3.Typical Mounting Configurations for Model 396RVP
Page 12
8
MODEL 396R pH/ORP SECTION 2.0
INSTALLATION
FIGURE 2-4. Dimensional Drawing — Model 396R with Optional Ball Valve PN 23765-00
Note: Add five (5) inches to dimension A if mounting a sensor head junction box onto the sensor.
INCH
MILLIMETER
Page 13
9
MODEL 396R pH/ORP SECTION 2.0
INSTALLATION
FIGURE 2-5. Dimensional Drawing — Model 396R with Optional Ball Valve PN 23240-00
Note: Add five (5) inches to dimension A if mounting a sensor head junction box onto the sensor.
DWG. NO. REV.
40396R05 A
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
Page 14
10
MODEL 396R pH/ORP SECTION 2.0
INSTALLATION
36”
36”
FIGURE 2-6. Dimensional Warning Label for Model 396R Hemi Bulb Sensors and Sensor Diagram
Note: Pressure rating for flat glass sensors is 100-790 kPa (0-100 psig).
Front Side 396R-21 Note: Retraction Pressure
4.920
2.25
Back Side 396R
Front Side 396R-25 Note: Retraction Pressure
A
MM / IN
548.64 / 21.60
916.94 / 36.10
OPTION
21 25
MILLIMETER
INCH
Page 15
11
MODEL 396R pH/ORP SECTION 2.0
INSTALLATION
FIGURE 2-7. Dimensional Drawing — Model 396RVP with Optional 1-1/2 inch Ball Valve PN 23240-00
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
Page 16
12
MODEL 396R pH/ORP SECTION 2.0
INSTALLATION
FIGURE 2-8. Dimensional Drawing — Model 396RVP with Optional 1-1/4 inch Ball Valve PN 23765-00
INCH
MILLIMETER
Page 17
13
MODEL 396R pH/ORP SECTION 3.0
WIRING MODEL 396R
SECTION 3.0
WIRING MODEL 396R
WIRING MODEL 396R.
Make electrical connections as shown on Figures 3-1 through 3-15 using the following guidelines. For wiring Model 396RVP, see Section 4.0.
1. Pa y particular attention to the analyzer or transmit­ter model number when following details on the wiring diagrams to ensure that the connections are made to the proper terminals.
2. Use Rosemount custom cable Part Number 9200273 for interconnect.
3. The maximum distance from the sensor to the analyzer is 15 ft without an integral preamplifier.
4. Signal cable should be run in a dedicated conduit and should be kept away from AC power lines.
NOTE
For maximum EMI/RFI protection when wiring from the sensor to the junction box, the outer braid of the sensor should be connected to the outer braided shield of the extension cable .The outer braid of the extension cable to the instru­ment must be terminated at earth ground or by using an appropriate metal cable gland fitting, that provides a secure connection to the instru­ment cable.
FIGURE 3-1. Cable Preparation Instructions (PN 900273)
WARNING:IF INNER BLACK CONDUCTIVE SHEATH IS IN CONTACT WITH THE EXPOSED LEADS, OR IS NOT PREPARED PROPERLY, IT MAY CAUSE AN ELECTRICAL SHORT.
DWG. NO. REV.
40396R24 A
Page 18
14
FIGURE 3-2.Wiring Details Model 396R-54 for use with Model 54 pH/ORP and 81/3081 pH/ORP
MODEL 396R pH/ORP SECTION 3.0
WIRING MODEL 396R
FIGURE 3-3.Wiring Details Model 396R-54 with Remote J-Box and Preamp (PN 23555-00)
DWG. NO. REV.
40396R08 A
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
40396R07 A
MODELS 54, 81, 3081
MODELS 54, 81, 3081
MODELS
54, 81, 3081
Page 19
15
MODEL 396R pH/ORP SECTION 3.0
WIRING MODEL 396R
DWG. NO. REV.
40396R011 A
FIGURE 3-4. Wiring Details - Model 396R-50 for use with Remote Junction Box (PN 23309-03)
1
PREP ORANGE WIRE FOR PREAMP WITH BNC ADAPTER (PN 9120516).
DWG. NO. REV.
40396R012 A
PREP ORANGE WIRE FOR PREAMP WITH BNC ADAPTER (PN 9120516).
1
DANGER:
Do not connect sensor
cable to power lines.
Serious injury may result.
DANGER:
Do not connect sensor
cable to power lines.
Serious injury may result.
FIGURE 3-5. Wiring Details - Model 396R-54 for use with Remote Junction Box (PN 23309-04)
Page 20
16
MODEL 396R pH/ORP SECTION 3.0
WIRING MODEL 396R
DWG. NO. REV.
40396R09 A
FIGURE 3-6. Wiring Details - Model 396R-50/54 for use with Model 1181 pH/ORP
1
PREP ORANGE WIRE FOR PREAMP WITH BNC ADAPTER (PN 9120516).
DWG. NO. REV.
40396R010 A
FIGURE 3-7. Wiring Details - Model 396R-54
for use with Models 1054A/B pH/ORP, 2054 pH, and 2081 pH/ORP
Page 21
17
MODEL 396R pH/ORP SECTION 3.0
WIRING MODEL 396R
DWG. NO. REV.
40396R13 A
DWG. NO. REV.
40396R14 A
FIGURE 3-8. Wiring Details - Model 396R-54 for use with Model SCL-(P/Q)
FIGURE 3-9. Wiring Details - Model 396R-54 for use with Model 2700 Preamp J-Box (PN 23054-03)
Page 22
18
MODEL 396R pH/ORP SECTION 3.0
WIRING MODEL 396R
FIGURE 3-11.Wiring Details - Model 396R-54-61 for use with
Sensor Head J-Box for use with Models 54, 81, 3081
FIGURE 3-10.Wiring Details - Model 396R-50/54-60 for use with Sensor Head
J-Boxes for Models 1181, 1054 Series, 2054, 2081
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
40396R21/22 B/C
DWG. NO. REV.
40396R06 C
Page 23
MILLIMETER
INCH
DANGER:
Do not connect sensor
cable to power lines.
Serious injury may result.
PREP ORANGE WIRE FOR PREAMP WITH BNC ADAPTER (PN 9120516).
1
MODEL 396R pH/ORP SECTION 3.0
WIRING MODEL 396R
FIGURE 3-12. Wiring Details-Model 396R-50 for use with J-Box (PN 23707-00)
to Models 1181, 1050, 1060, 1030, and 1023 pH Transmitters
DWG. NO. REV.
40396R21 A
19
Page 24
FIGURE 3-13. Wiring Details:
Model 396R-54 for use with J-Box (PN 23708-01) to Models 1054 Series, 2054, and 2081 pH Transmitters
MODEL 396R pH/ORP SECTION 3.0
WIRING MODEL 396R
20
DWG. NO. REV.
40396R22 B
PREPARE ORANGE WIRE FOR PREAMP, USE BNC ADAPTER PN 9120516
Page 25
FIGURE 3-14. Wiring Model 396R-( )-54-62-(71) to Model 1055-10-22-32
FIGURE 3-15. Wiring Model 396R-(02)-( )-54/55-(61) to Model 1055-10-22-32
MODEL 396R pH/ORP SECTION 3.0
WIRING MODEL 396R
21
Page 26
FIGURE 4-1. Wire Functions and Pin Connections for Model 396RVP
MODEL 396R pH/ORP SECTION 4.0
WIRING MODEL 396RVP
SECTION 4.0
WIRING MODEL 396RVP
WIRING MODEL 396RVP.
Make electrical connections as shown on Figures 4-1 through 4-20 using the following guidelines. For wiring Model 396R, see Section 3.0.
1. Pa y particular attention to the analyzer or transmit­ter model number when following details on the wiring diagrams to ensure that the connections are made to the proper terminals.
2. The Model 396RVP uses a mating VP cable.The cable part numbers are 23645-06 and 23645-07. See attached wiring sheet for wire functions of the cables and wiring diagrams to various analyzers.
3. The maximum distance from the sensor to the analyzer is 15 ft without an integral preamplifier.
4. Signal cable should be run in a dedicated conduit and should be kept away from AC power lines.
NOTE
For maximum EMI/RFI protection when wiring from the sensor to the junction box, the outer braid of the sensor should be connected to the outer braided shield of the extension cable .The outer braid of the extension cable to the instru­ment must be terminated at earth ground or by using an appropriate metal cable gland fitting, that provides a secure connection to the instru­ment cable.
Wire Functions for Variopol connector cable
PN 23645-06 (with BNC) PN 23645-07
22
Page 27
MODEL 396R pH/ORP SECTION 4.0
WIRING MODEL 396RVP
FIGURE 4-5. Model 1181, 1050/1060, & 1003/1023
Wiring through Remote Junction Box
FIGURE 4-2. Model 81 Wiring
FIGURE 4-4. Model 81 Wiring through Remote
Junction Box
FIGURE 4-3. Model 1181 Wiring
23
Page 28
MODEL 396R pH/ORP SECTION 4.0
WIRING MODEL 396RVP
FIGURE 4-6. Model 2081 Wiring
FIGURE 4-7. Model 3081 & 4081 Wiring
FIGURE 4-9. Model 3081 & 4081 Wiring through
Remote Junction Box
FIGURE 4-8. Model 2081 Wiring through Remote
Junction Box
24
Page 29
MODEL 396R pH/ORP SECTION 4.0
WIRING MODEL 396RVP
FIGURE 4-10. Model 1054 Wiring
FIGURE 4-12. Model 1054 Wiring through a Remote
Junction Box
FIGURE 4-11. Model 1054A/B & 2054 Wiring
FIGURE 4-13. Model 1054A/B & 2054 Wiring through
a Remote Junction Box
25
Page 30
MODEL 396R pH/ORP SECTION 4.0
WIRING MODEL 396RVP
FIGURE 4-14. Model 54 Wiring
FIGURE 4-15. Model 54 Wiring through Remote Junction Box
FIGURE 4-16. Model 2700 Wiring
FIGURE 4-17. Model SCL-(P/Q) Wiring
26
Page 31
MODEL 396R pH/ORP SECTION 4.0
WIRING MODEL 396RVP
FIGURE 4-18. Model 1055pH/pH Wiring
FIGURE 4-19. Model 1055pH/pH Wiring through Remote Junction Boxes
27
Page 32
FIGURE 4-20. Model 1055-10-22-32 Wiring
MODEL 396R pH/ORP SECTION 4.0
WIRING MODEL 396RVP
28
Page 33
29
5.1 START UP. To obtain best accuracy, the sensor
must be calibrated as a loop with the analyzer. Please refer to the respective analyzer/transmitter instruction manual for proper calibration procedures.
5.2 396R pH CALIBRATION
1. After a temporary connection is established between the sensor and the instrument, a buffer calibration may be performed.
2. Consult appropriate pH/ORP analyzer or transmitter instruction manual for specific calibration and standardization procedures or see below for recommended two point buffer calibration procedure.
5.2.1 Recommended two point buffer calibration
procedure:
Select two stable buffer solutions, preferably pH 4.0 and 10.0 (pH buffers other than pH 4.0 and pH 10.0 can be used as long as the pH values are at least two pH units apart).
Note: A pH 7 buffer solution reads a mV value of approx. zero, and pH buffers read approx. +/- 59.1 mV for each pH unit above or below pH 7. Check the pH buffer manufacturer specifications for millivolt values at various temperatures since it may affect the actual value of the buffer solution mV/pH value.
1. Immerse sensor in the first buffer solution. Allow sensor to adjust to the buffer temperature (to avoid errors due to temperature differences between the buffer solution and sensor temperature) and wait for reading to stabilize. Value of buffer can now be acknowledged by analyzer/transmitter
2. Once the first buffer has been acknowledged by the analyzer/transmitter, rinse the buffer solution off of the sensor with distilled or deionized water.
3. Repeat steps 1 and 2 using the second buffer solution.
4. Once the analyzer/transmitter has acknowledged both buffer solutions, a sensor slope (mV/pH) is established (the slope value can be found within the analyzer/transmitter).
5. The slope value should read about 59.1 mV/pH for a new sensor and will decrease over time to approximately 47 - 49 mV/pH. Once the slope reads below the 47-49 mV/pH range, a new sensor should be installed to maintain accurate readings.
5.2.2 Recommended pH Sensor Standardization
For maximum accuracy, the sensor can be standard­ized on-line or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sen­sor’ s slope but will simply adjust the analyzers r eading to match that of the known process pH.
1. While obtaining a process solution sample (it is recommended that the sample is taken close to the sensor), record the pH value that is shown on the analyzer/transmitter display.
2. Measure and record the pH of the process solution sample with a another temperature compensated, calibrated pH instrument. For best results, standardization should be performed at the process temperature.
3. Adjust the analyzer/transmitter value to the standardized value.
MODEL 396R pH/ORP SECTION 5.0
START UP AND CALIBRATION
SECTION 5.0
START UP AND CALIBRATION
Page 34
5.3 396R ORP CALIBRATION. An ORP loop is best calibrated using an ORP standard solution. Most industrial applications have a number of ORP reac­tions occurring in sequence or simultaneously. There can be several components that are oxidized or reduced by the reagents that are used. Theoretically, the ORP potential is absolute because it is the result of the oxidation-reduction equilibrium. However, the actual measured potential is dependent on many fac­tors, including the condition of the surface of the ORP platinum electrode. Therefore, the sensor should be
allowed 1-2 hours to become “conditioned” to the stream when first set-up or after being cleaned.
5.3.1 ORP Calibration Procedure CAUTION
The solution used during the following check is an acid and should be handled with care. Follow the directions of the acid manufacturer. Wear the proper protective equipment. Do not let the solution come in contact with skin or clothing. If contact with skin is made, immediately rinse with clean water.
1. Make a temporary electrical connection between
the sensor and the instrument.
2. Obtain a standard solution of saturated quinhy-
drone (PN R508-160Z). This can also be made
quite simply by adding a few crystals of quinhy­drone to either pH 4 or pH 7 buffer. Quinhydrone is only slightly soluble; therefore, only a few crys­tals will be required (refer to Section 6.3. for an alternate ORP standard solution).
3. Immerse the sensor in the standard solution. Allow 1-2 minutes for the ORP sensor to stabilize.
4. Adjust the standardize control of the instrument to the solution value shown in Table 5-1. The resulting potentials, measured with a clean platinum elec­trode and saturated KCl/AgCl reference electrode, should be within ±20 millivolts of the value shown in Table 5-1. Solution temperature must be noted to insure accurate interpretation of results. The ORP value of saturated quinhydrone solution is not sta­ble over long periods of time. Therefore, these stan­dards should be made up fresh each time they are used.
5. Remove the sensor from the buffer, rinse and install in the process.
TABLE 5-1
ORP of Saturated Quinhydrone Solution
pH 4 pH 7
TEMPERATURE
°C 20 25 30 20 25 30
Millivolt Potential 268 264 260 94 87 80
MODEL 396R pH/ORP SECTION 5.0
START UP AND CALIBRATION
30
Page 35
31
6.1 Maintenance. The Model 396R Sensor is a dis-
posal type sensor and therefore requires minimum maintenance. The sensor should be removed from the process periodically and checked in buffer solutions. If the sensor will not calibrate, refer to your analyzer/transmitters instruction manual for proper test procedures. If the sensor has failed, it should be discarded and replaced.
6.2 Sensor Removal. Please refer to the appropriate paragraph for instructions regarding removal of the sensor for periodic maintenance.
6.2.1 Retractable Version. WARNING
System pressure may cause the sensor to blow out with great force unless care is taken during removal. Make sure the following steps are adhered to.
A. Model 396R-21 (21” tube)
1. Be certain system pressure at the sensor is below
64 psig (542 kPa) before proceeding with the retraction. It is also recommended that the per­sonnel wear a face shield and have a stable foot­ing. Refer to Figure 6-1. Push in on the sensor end or the top of the J-box and slowly loosen the hex nut (B) of the process end male connector (A).
B. Model 396R-25 (36” tube)
2. Be certain that pressure at the sensor is below 35
psig (343 KPa) before proceeding with the retrac­tion. It is also recommended that the personnel wear a face shield and have a stable footing. Refer to Figure 6-1. Push in on the sensor end or the top of the J-box and slowly loosen the hex nut (B) of the process end male connector (A).
CAUTION
Do not remove nut at this time.
3. When the hex nut is loose enough, slowly ease
the sensor back completely until the retraction stop collar is reached.
CAUTION
Failure to withdraw the sensor completely may result in damage to the sensor when the valve is closed.
4. Close the ball valve slowly. If there is resistance, the valve may be hitting the sensor. Double check that the sensor has been retracted to the retrac­tion stop collar.
WARNING
Before removing the sensor from the ball valve, be absolutely certain that the ball valve is fully closed. Leakage from the male connector threads may indicate that the male connector is still under pressure. Leakage through a partially open valve could be hazardous, however with the ball valve closed, some residual process fluid may leak from the connector's pipe threads.
5. The Male Connector Body (A) may now be com­pletely unthreaded from the reducing coupling and the sensor removed for servicing.
CAUTION
If the male connector leaks during inser­tion or retraction, replace the O-ring (PN
9550099) in the male connector A.
6.3 pH Electrode Cleaning. If the electrode is coated
or dirty, it may be cleaned as follows:
1. Remove the sensor from process as instructed in Section 6.2.
2. Wipe the glass bulb with a soft, clean, lint free cloth or tissue. If this does not remove the dirt or coating, proceed to step 3. If the sensor appears to be clean, go to step 5.
3. Wash the glass bulb in a strong detergent solution and thoroughly rinse with tap water. If the bulb still appears to have a coating, proceed to step 4.
CAUTION
The solution used in the following step is an acid and should be handled with care. Follow the directions of the acid manufac­turer. Wear the proper protective equip­ment. Do not let the solution come in con­tact with skin or clothing. If contact with the skin is made, immediately rinse with clean water.
MODEL 396R pH/ORP SECTION 6.0
MAINTENANCE
SECTION 6.0
MAINTENANCE
Page 36
32
MODEL 396R pH/ORP SECTION 6.0
MAINTENANCE
4. Following the caution above, wash the glass bulb in dilute 5% hydrochloric acid solution and then rinse it thoroughly in tap water. Replace the sen­sor if it cannot be cleaned. If the glass bulb appears clean, proceed to step 5.
5. Buffer calibrate the sensor (Refer to Section 5.0). If the sensor appears to respond sluggishly to pH change, soaking it overnight in a weak acid solution (5% hydrochloric acid) may improve its response. Be sure to follow the CAUTION above and to rinse the sensor’s tip thoroughly with tap water. If the sen­sor will not calibrate, it must be replaced.
6.4 Cleaning Platinum Electrode. The electrode is
never exposed to these undesirable compounds. In the event poisoning is suspected, the electrode can be restored to normal operation by simply cleaning the plat­inum electrode with baking soda. Polish it by rubbing it with a damp paper towel and baking soda until a bright, shiny appearance is attained.
6.5 Automatic Temperature Compensator. The tem-
perature compensator element is temperature sensi­tive and can be checked with an ohmmeter. Resistance increases with temperature.
The 3K element will read 3000 ohms ± 1% at 25°C (77°F) and a Pt-100 will read 110 ohms. Resistance varies with temperature for a 3K and Pt-100 element and can be determined according to Table 6-2 or the following formula:
RT=Ro[l+R1(T-20)]
Where RT = Resistance T = Temperature in °C
Refer to Table 6-1 for Roand R1values:
TABLE 6-1
Roand R1VALUES FOR TEMPERATURE
COMPENSATION ELEMENTS
6.6 Sensor Tube Replacement When Used With A Sensor Head Junction Box. Replacement of the
retraction versions sensor tube assembly involves the removal and installation of two sets of male connectors: One at the process end of the sensor, and the other at the junction box end (See Figures 6-1, 6-2). Refer to Section 6.2 for proper removal of the sensor from process.
1. Remove sensor from process before proceeding. The junction box with attached male connector must be recovered from the old sensor for reuse. Unscrew the junction box cover and set aside. Disconnect electrical connections from printed circuit board inside junction box. Disconnect BNC connector to preamp. Unscrew hex nut (D) from male connector body (C). Separate junction box from used sensor. Set aside.
2. Pry off split ferrule from sensor and set aside for reuse. Remove hex nut (D) and set aside for reuse. Check that the internal O-ring is in place in the male connector body (C) attached to the junc­tion box.
3. Remove hex nut (B) from male connector body (A) at process end of sensor and set aside. Slide the Teflon ferrule and the male connector off sensor in the direction of junction box and set
NOTE
If stainless steel ferrule was used, male connector body (A) will have to be dis­carded with the sensor tube.
Resistance
Temperature °C (Ohms) ±1%
3K PT-100
0 2670 100.0 10 2802 103.8 20 2934 107.7 25 3000 109.6 30 3066 111.5 40 3198 115.4 50 3330 119.2 60 3462 123.1 70 3594 126.9 80 3726 130.8 90 3858 134.6
100 3990 138.5
Temperature
Compensation Element
R
o
R
1
3K 2934 .0045
PT-100 107.7 .00385
TABLE 6-2
TEMPERATURE vs RESISTANCE OF AUTO
T.C. ELEMENTS
Page 37
33
4. Discard used O-ring from male connector body (A). Coat new O-ring with a thin film of the O-ring lubricant provided. Position it in the machined O­ring groove in place of the discarded O-ring.
CAUTION
Make sure lubricant does not contact any part of the sensor tip particularly the glass bulb.
5. Cover the 1" MNPT pipe threads of the male con­nector body (A) with Teflon tape (not provided) to
protect them from galling during reinstallation.
6. Pass the wires from the new sensor through the process end male connector (A). Make sure that the beveled edge of the ferrule faces the process end of the sensor. Snug the hex nut (B) to keep it in place. Do not tighten down fully on the hex nut at this time.
7. Pass the wires from the new sensor through the hex nut (D), the split ferrule (from the old sensor), male connector body (C), O-ring, and through the junction box from the “neck” opening and out to the printed circuit board in the junction box. Butt the ferrule’s beveled edge and the sensor tube against the junction male connector (C). Screw the hex nut (D) by hand until the tube is “locked” into the male connector body. Make sure that the male connector body (C) is sufficiently tightened. The sensor will “click” into place by pulling the sensor tube away from the junction box, but will not move from side to side or pull clear of the male connector. If the sensor tube is correctly attached to the junction box, wrench tighten hex nut (D) on male connector body (C) (see Figure 6-1). Do not put the sensor tube in a vise or use a pipe wrench to tighten the hardware as these will damage the sensor. If sensor tube is not correctly attached to the junction box, loosen hex nut (D) and repeat.
8. Connect the sensor wires to the terminals on the printed circuit board in the junction box in the man­ner recommended on the junction box cover,and reattach the BNC connector to the preamp. Screw on the cover of the junction box aside. Discard sen­sor tube.
9. Insert the sensor in the process fitting. Stop it against the closed ball valve. Slide the process­end male connector down the sensor tube to mate with the process fitting. Tighten the male connector into the process fitting.
10. Pull back hard on the sensor assembly, as if trying to remove the sensor, to be certain that the sensor cannot come free from the valve assembly and male connector. The built-in retraction stop collar at the end of the sensor will butt against the shoulder of the male connector.
11. Open ball valve and position the sensor at the desired insertion depth and orientation. Using a crescent or open end wrench, tighten the hex nut (B) to secure the sensor in place. See Figure 6-2.
NOTE
A stainless steel ferrule is available if the Teflon ferrule does not adequately grip, be careful and avoid over tightening. This can damage the sensor tube.
CAUTION
If the male connector leaks during inser­tion or retraction, replace the O-Ring (PN
9550099) in the male connector body (A).
If the sensor is to be stored, the rubber boot should be filled with 7pH buffer solution and replaced on sensor tip until ready to use.
MODEL 396R pH/ORP SECTION 6.0
MAINTENANCE
Page 38
FIGURE 6-1. Example of Sensor Tube Replacement
FIGURE 6-2. Male Connector Tightening Diagram
MODEL 396R pH/ORP SECTION 6.0
MAINTENANCE
34
Page 39
35
MODEL 396R pH/ORP SECTION 7.0
DIAGNOSTICS AND TROUBLESHOOTING
SECTION 7.0
DIAGNOSTICS AND TROUBLESHOOTING
7.1 DIAGNOSTICS AND TROUBLESHOOTING WITH MODEL 54/81/3081 pH/ORP. The Model 54 and 54e
Analyzers and Models 81 and 3081 pH Transmitters automatically search for fault conditions that would cause an error in the measured pH value, as does the Model 1054A/B pH/ORP Analyzer to a lesser degree. Refer to the respective manual for a complete description of the analyzer’s fault conditions.
Table 7-1 lists the Analyzer/Transmitters diagnostic messages that indicate a possible sensor problem. A more complete description of the problem and a suggested remedy corresponding to each message is also listed.
DIAGNOSTIC MESSAGE 54 and 54e 81/3081 DESCRIPTION OF PROBLEM REMEDY
“Calibration Warning” 1. Aged glass. 1. Perform buffer calibration.
2. Sensor not immersed. 2. Be sure electrode measuring tip is in
CALIbrAtE process.
“Cracked glass failure” Broken or cracked glass. Replace Sensor.
6LASS FAIL
“High reference imped” 1. Liquid junction coated. 1. Clean sensor; replace if necessary.
2. Reference Cell gel depleted. 2. Replace sensor.
rEF FAIL or rEF WArn 3. Sensor not immersed. 3. Be sure electrode measuring tip is in
process.
“Input voltage high” pH input shorted or sensor. Check wiring. Replace sensor if “Input voltage low” miswired. necessary.
“Old glass warning” 1. Glass electrode worn out. 1. Replace sensor.
2. Sensor not immersed. 2. Be sure electrode measuring tip is in
6LaSS WArn
process.
“Reference offset err” Reference electrode poisoned. Replace sensor. (offline only)
Std Err “Ref voltage high” 1. Reference shorted or sensor Check wiring. Replace sensor if
“Ref voltage low” miswired. necessary.
2. Sensor not immersed.
“Sensor line open” 1.
Open wire between sensor and analyzer.
1. Check sensor wiring.
2. Interconnecting cable greater than 2. Relocate analyzer.
LInE FAIL 1000 ft.
“Sensor miswired” 1.
Open wire between sensor and analyzer.
1. Check wiring.
2. Bad preamplifier. 2. Replace preamplifier.
“Temp error high” 1. Open or shor ted RTD. 1. Replace sensor. “Temp error low” 2. Temperature out of range. 2. Check process temperature.
tEMP H I tEMP LO
TABLE 7-1 Troubleshooting with Diagnostics
Page 40
36
MODEL 396R pH/ORP SECTION 7.0
DIAGNOSTICS AND TROUBLESHOOTING
Problem Probable Cause Remedy
Meter reads off scale. (Display Defective preamplifier. Replace preamplifier (for code 02 reads overrange).
sensors). For code 01, replace sensor.
T.C. element shorted. Check T.C. element as instructed
in Section 6.5 and replace sensor if defective.
Sensor not in process or sample Make sure sensor is in process with stream is low. sufficient sample stream (refer to
Section 2.0 for installation details). Open glass electrode. Replace sensor. Reference element open - no contact.
Replace sensor.
Display reads between 3 and 6 pH Electrode cracked. Replace sensor. regardless of actual pH of solution or sample.
Meter or display indication swings T.C. element shorted. Check T.C. element as instructed or jumps widely in AUTO T.C. Mode. in Section 6.5 and replace
sensor if defective.
Span between buffers extremely T.C. element open. Check T.C. element as instructed short in AUTO T.C. Mode. in Section 6.5 and replace sensor
if defective.
Sluggish or slow meter indication Electrode coated. Clean sensor as instructed in for real changes in pH level. Sections 6.2 or 6.3. Replace
sensor if cracked. Electrode defective. Replace sensor.
Transmitter cannot be standardized. Electrode coated or cracked. Clean Sensor as instructed in
Sections 6.1 or 6.3 Replace
sensor if cracked. Defective preamplifier. Replace preamplifier.
Transmitter short spans between Aged glass electrode or high Replace sensor. two different buffer values. temperature exposure.
Electrode coated . Clean Sensor as instructed in
Sections 6.2 or 6.3. Replace
sensor if cracked.
TABLE 7-2.Troubleshooting without Diagnostics
7.2 TROUBLESHOOTING WITHOUT DIAGNOSTICS. Table 7-2 lists common problems, causes and remedies
typically encountered in process measurement.
Page 41
37
Model 396R pH/ORP SECTION 8.0
RETURN OF MATERIAL
SECTION 8.0
RETURN OF MATERIAL
8.1 GENERAL.
To expedite the repair and return of instruments, prop­er communication between the customer and the fac­tory is important. The “Return of Materials Request” form is provided for you (in this manual) to copy and use in case the situation arises. The accuracy and completeness of this form will affect the processing time of your materials. Call 1-949-863-1181 for a Return Materials Authorization (RMA) number.
8.2 WARRANTY REP AIR.
The following is the procedure for returning instru­ments still under warranty:
1. Call Rosemount Analytical for authorization.
2. To verify warranty, supply the factory sales order number or the original purchase order number. In the case of individual parts or sub-assemblies, the serial number on the unit must be supplied.
3. Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the materials in the same manner as they were received.
4. Send the package prepaid to: Rosemount Analytical Inc., Uniloc Division
Uniloc Division 2400 Barranca Parkway Irvine, CA 92606
Attn: Factory Repair
RMA No. ____________
Mark the package: Returned for Repair
Model No. ____
8.3 NON-WARRANTY REPAIR.
The following is the procedure for returning for repair instruments that are no longer under warranty:
1. Call Rosemount Analytical for authorization.
2. Supply the purchase order number, and make sure to provide the name and telephone number of the individual to be contacted should addition­al information be needed.
3. Do Steps 3 and 4 of Section 8.2.
NOTE
Consult the factory for additional informa­tion regarding service or repair.
Page 42
FROM: RETURN BILL TO:
_____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________
CUSTOMER/USER MUST SUBMIT MATERIAL SAFETY SHEET (MSDS) OR COMPLETE STREAM COMPOSITION, AND/OR LETTER CERTIFYING THE MATERIALS HAVE BEEN DISINFECTED AND/OR DETOXIFIED WHEN RETURNING ANY PRODUCT, SAMPLE OR MATERIAL THAT HAVE BEEN EXPOSED TO OR USED IN AN ENVIRONMENT OR PROCESS THAT CONTAINS A HAZARDOUS MATERIAL ANY OF THE ABOVE THAT IS SUBMITTED TO ROSEMOUNT ANALYTICAL WITHOUT THE MSDS WILL BE RETURNED TO SENDER C.O.D. FOR THE SAFETY AND HEALTH OF OUR EMPLOYEES. WE THANK YOU IN ADVANCE FOR COMPLIANCE TO THIS SUBJECT.
SENSOR OR CIRCUIT BOARD ONLY: (Please reference where from in MODEL / SER. NO. Column)
1. PART NO. _____________________________1. MODEL ____________________________________1. SER. NO. __________________
2. PART NO. _____________________________2. MODEL ____________________________________2. SER. NO. __________________
3. PART NO. _____________________________3. MODEL ____________________________________3. SER. NO. __________________
4. PART NO. _____________________________4. MODEL ____________________________________4. SER. NO. __________________
PLEASE CHECK ONE:
REPAIR AND CALIBRATE DEMO EQUIPMENT NO. ______________________________
EVALUATION OTHER (EXPLAIN) ___________________________________
REPLACEMENT REQUIRED? YES NO _______________________________________________________
DESCRIPTION OF MALFUNCTION: __________________________________________________________________________________________________________________ __________________________________________________________________________________________________________________ __________________________________________________________________________________________________________________
WARRANTY REPAIR REQUESTED:
YES-REFERENCE ORIGINAL ROSEMOUNT ANALYTICAL ORDER NO._______________________________________________
CUSTOMER PURCHASE ORDER NO. ________________________________________________________
NO-PROCEED WITH REPAIRS-INVOICE AGAINST P.O. NO._________________________________________________________
NO-CONTACT WITH ESTIMATE OF REPAIR CHARGES: LETTER ■_________________________________________________
PHONE ________________________________________________
NAME___________________________________________________________ PHONE______________________________________________
ADDRESS_______________________________________________________________________________________________________________
_____________________________________________________________________ ZIP______________________________________________
RETURN AUTHORITY FOR CREDIT ADJUSTMENT [Please check appropriate box(s)]
WRONG PART RECEIVED REPLACEMENT RECEIVED
DUPLICATE SHIPMENT REFERENCE ROSEMOUNT ANALYTICAL SALES ORDER NO._____________
RETURN FOR CREDIT RETURN AUTHORIZED BY: __________________________________________
WARRANTY DEFECT______________________________________________________________________________________________
_________________________________________________________________________________________________________________
24-6047
RETURN OF MATERIALS REQUEST
•IMPORTANT!
This form must be completed to ensure expedient factory service.
R E P A
I
R S
T A T U S
R E A S O N
F O R
R E
T U R N
C U S T O
M
E R
N O
T
I
C
E T
O
S E N D E R
Rosemount Analytical Inc. Uniloc Division
2400 Barranca Parkway Irvine, CA 92606 USA Tel: 1-800-854-8257 Fax: (949)-474-7250
Page 43
WARRANTY
Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in work­manship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, pH electrodes, membranes, liquid junctions, electrolyte, O-rings, etc. are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of ship­ment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and / or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, parts(s), or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty . This war­ranty shall be in effect for replacement or repaired goods, part(s) and consumables for the remaining portion of the peri­od of the twelve (12) month warranty in the case of goods and part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not oper­ate to condemn such commercial unit when such goods, parts(s) or consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage, directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STA T­ED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Rosemount Analytical Inc.
Uniloc Division
2400 Barranca Parkway
Irvine, CA 92606
The shipping container should be marked:
Return for Repair Model
_______________________________
The returned material should be accompanied by a letter of transmittal which should include the following information (make a copy of the "Return of Materials Request" found on the last page of the Manual and provide the following there­on):
1. Location type of service, and length of time of service of the device.
2. Description of the faulty operation of the device and the circumstances of the failure.
3. Name and telephone number of the person to contact if there are questions about the returned material.
4. Statement as to whether warranty or non-warranty service is requested.
5. Complete shipping instructions for return of the material.
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
Page 44
Rosemount Analytical Inc. Uniloc Division
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 863-1181 http://www.RAuniloc.com
© Rosemount Analytical Inc. 2001
Credit Cards for U.S. Purchases Only.
The right people, the right answers, right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.RAuniloc.com
UNILOC DIVISION
CUSTOMER SUPPORT CENTER
1-800-854-8257
Loading...