Rosemount 396P, 396PVP Instruction Manual

Page 1
Models 396P and 396PVP
Combination pH/ORP Sensor
Instruction Manual
PN 5100396P
February 2001
MODEL 396P
MODEL 396PVP with mating cable
Page 2
CAUTION
SENSOR/PROCESS
APPLICATION COMPATIBILITY
The wetted sensor materials may not be compatible with process com­position and operating conditions. Application compatibility is entirely the responsibility of the user.
DANGER
HAZARDOUS AREA INSTALLATION
This sensor is not Intrinsically Safe. or Explosion Proof. Installations near flammable liquids or in hazardous area locations must be carefully evaluated by qualified on site safety personnel.
To secure and maintain an intrinsically safe installation, an appropriate transmitter/ safety barrier/sensor combination must be used. The installation system must be in accordance with the governing approval agency (FM, CSA or BASEEFA/CENELEC) hazardous area classification requir ements. Consult your analyzer/transmitter instruction manual for details.
Proper installation, operation and servicing of this sensor in a Hazardous Area Instal­lation is entirely the responsibility of the user.
SS-SE
August 1995
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures, and tests its products to meet many national and international stan­dards. Because these instruments are sophisticated tech­nical products, you must properly install, use, and main­tain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product. If this Instruction Manual is not the correct manual, telephone 1-800-654-7768 and the requested manual will be provided. Save this Instruction Manual for future reference.
• If you do not understand any of the instructions, contact your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper instal­lation, operation, and maintenance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qual­ified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance and place the safe operation of your process at risk. Look alike substitu­tions may result in fire, electrical hazards, or improper operation.
Rosemount Analytical Inc. Uniloc Division
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 863-1181 http://www.RAuniloc.com
Page 3
i
MODEL 396P and 396PVP TABLE OF CONTENTS
MODEL 396P and 396PVP
COMBINATION pH/ORP SENSOR
TABLE OF CONTENTS
Section Title Page
11..00 DDEESSCCRRIIPPTTIIOONN AANNDD SSPPEECCIIFFIICCAATTIIOONNSS ...................................................................................................................................... 11
1.1 Features and Applications....................................................................................... 1
1.2 Physical Specifications............................................................................................ 1
1.3 Ordering Information................................................................................................ 2
22..00 IINNSSTTAALLLLAATTIIOONN................................................................................................................................................................................................................ 44
2.1 Unpacking and Inspection ...................................................................................... 4
2.2 Mounting.................................................................................................................. 4
33..00 WWIIRRIINNGG MMOODDEELL 339966PP-0011................................................................................................................................................................................ 99
44..00 WWIIRRIINNGG MMOODDEELLSS 339966PP-0022 AANNDD 339966PPVVPP................................................................................................................................ 1155
55..00 SSTTAARRTT UUPP AANNDD CCAALLIIBBRRAATTIIOONN.............................................................................................................................................................. 2222
5.1 Model 396P pH........................................................................................................ 22
5.2 Model 396P ORP ..................................................................................................... 22
66..00
MMAAIINNTTEENNAANNCCEE .............................................................................................................................................................................................................. 2233
6.1 Automatic Temperature Compensator..................................................................... 23
6.2 Model 396P pH........................................................................................................ 23
6.3 Model 396P ORP ..................................................................................................... 24
77..00 DDIIAAGGNNOOSSTTIICCSS AANNDD TTRROOUUBBLLEESSHHOOOOTTIINNGG ........................................................................................................................ 2255
7.1 Diagnostics & Troubleshooting With the Models 54, 3081, 81, and 4081.............. 25
7.2 Troubleshooting Without Advanced Diagnostics..................................................... 26
88..00 RREETTUURRNN OOFF MMAATTEERRIIAALL...................................................................................................................................................................................... 2288
8.1 General.................................................................................................................... 28
8.2 Warranty Repair....................................................................................................... 28
8.3 Non Warranty Repair................................................................................................ 28
LIST OF TABLES
Number Title Page
1-1 Recommended Accessories for First Time Installations.......................................... 3
3-1 Wiring Matrix Guide - Model 396P........................................................................... 9
4-1 Wiring Matrix Guide - Model 396PVP....................................................................... 15
5-1 ORP of Saturated Quinhydrone Solution (In Millivolts)............................................. 22
6-1 Ro and R1 Values for Temperature Compensation Elements .................................. 23
6-2 Temperature vs Resistance of Auto T.C. Elements .................................................. 23
7-1 Troubleshooting With Advanced Diagnostics.......................................................... 25
7-2 Troubleshooting Without Advanced Diagnostics..................................................... 26
7-3
Model 396P Replacement Parts and Accessories................................................... 27
Page 4
MODEL 396P and 396PVP TABLE OF CONTENTS
LIST OF FIGURES
Number Title Page
2-1 Dimensional Drawing .......................................................................................................................................................... 5
2-2 Flow-Through Tee with Adapter (PN 915240-xx)................................................................................................................. 5
2-3 Flow-Through and Insertion Installations ........................................................................................................................... 6
2-4 Model 396P/396PVP with Insertion Mounting Adapter (PN 23242-02)............................................................................... 6
2-5 Submersion Installations .................................................................................................................................................... 7
2-6 Low Flow Cell PN 23728-00 ................................................................................................................................................ 8
2-7 Jet Spray Cleaner PN 12707-00.......................................................................................................................................... 8
3-1 Wire Functions for Model 396P with & without Preamplifier................................................................................................ 10
3-2 Wiring Details Model 396P-01-50 for use with & without Junction Box (PN 22719-02) for Models 1181, 1050, & 1060.... 11
3-3 Wiring Details Model 396P-01-54 for use with & sithout Junction Box (PN 22719-02) for Models 1054/A/B, 2054, 2081
and Extension Cable (PN 9200254).................................................................................................................................... 12
3-4 Wiring Details Model 396P-01-55 for use with Junction Box (PN 23550-00) for Models 54, 3081, 4081, and 81 ............. 13
3-5 Wiring Details Model 396P-01-55 for use with Models 54, 81, 3081, and 4081 ................................................................. 14
3-6 Wiring Model 396P-01)-( )-55 Sensor to Model 1055-10-22-32 Analyzer........................................................................... 14
4-1 Wire Functions for Mating Variopol Cable used with Model 396PVP.................................................................................. 16
4-2 Connector Pins and their Functions.................................................................................................................................... 16
4-3 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Model 81....... 16
4-4 Wiring Details for Models 396PVP or 396P-02-50 with Mating Variopol Cable (PN 23645-07) for use with Model 1181... 16 4-5 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-07) for use with Model 2081... 16 4-6 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Models 3081
and 4081............................................................................................................................................................................. 16
4-7 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Remote .........
Junction Box (PN 23555-00) to Model 81 ........................................................................................................................... 17
4-8 Wiring Details for Models 396PVP or 396P-02-50 with Mating Variopol Cable (PN 23645-07) for use with Remote .........
Junction Box (PN 23309-03) to Model 1181 ....................................................................................................................... 17
4-9 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Model 54...... 17
4-10 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Remote .........
Junction Box (PN 23555-00) to Model 54 ........................................................................................................................... 17
4-11 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-06) for use with Remote .........
Junction Box (PN 23309-03) to Model 2081 ....................................................................................................................... 18
4-12 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Remote .........
Junction Box (PN 23555-00) to Models 3081 and 4081..................................................................................................... 18
4-13 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-07) for use with Model 2700... 18
4-14 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-07) for use with ......................
Model SCL-(P/Q)................................................................................................................................................................. 18
4-15 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Model 1055... 19
4-16 Wiring Model 396P-(02)-( )-54/55-(61) Sensor to Model 1055-01-10-22-32 Analyzer........................................................ 19
4-17 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Remote .........
Junction Box (PN 23557-00) to Model 1055 ....................................................................................................................... 19
4-18 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-07) for use with Model 1054... 20
4-19 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-06) for use with Models..........
1054A/B and 2054 .............................................................................................................................................................. 20
4-20 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-06) for use with Remote .........
Junction Box (PN 23557-00) to Model 1054 ....................................................................................................................... 20
4-21 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-06) for use with Remote .........
Junction Box (PN 23309-04) to Models 1054A/B and 2054 ............................................................................................... 20
4-22 Wiring Details Model 396P-02-50 for use with Junction Box (PN 23309-03), Remote Preamplifier, and Extension ..........
Cable (PN 9200254)............................................................................................................................................................ 21
4-23
Wiring Details Model 396P-02-54 for use with Junction Box (PN 23309-04), Remote Preamplifier (PN 22698-03), and ..
Extension Cable (PN 9200254)
........................................................................................................................................... 21
ii
Page 5
construction, designed with user convenience in mind, does not require electrolyte (KCl) replenishment or any high maintenance troubleshooting procedures.
The Model 396P is offered with or without an optional integral preamplifier to convert the high impedance pH or ORP mV signal into a stable, noise-free signal with transmission capability of up to three miles. A remote preamplifier must be used with Model 396PVP.
Installation is easily achieved through the wide variety of mounting configurations. The Models 396P and 396PVP feature 1 in.(MNPT) front and rear facing con­nections for insertion, submersion or flow through pH and ORP applications.
Model 396PVP is offered with a watertight sensor-to­cable connector which eliminates re-wiring and cable twisting when replacing sensors.The Variopol VP mul­tiple pin connector is an integral part of each sensor model and uses a mating VP cable. Once the cable is installed and wired to the analyzer, sensors are easily replaced without replacing the cable, and, if the replacement sensor is the same as its predecessor, without rewiring the analyzer. Also the cable can be disconnected from the sensor before removal from the process which eliminates cable twisting.
U.S.Patent No. 5,152,882 Foreign Patent Pending U.S.Patent No. 6,054,031, Foreign Patents Pending.
1
AccuGlass is a trademark of Rosemount Analytical.
1
MODEL 396P and 396PVP SECTION 1.0
DESCRIPTIONS AND SPECIFICATIONS
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.1 FEATURES AND APPLICATIONS
The Rosemount Analytical Model 396P and 396PVP Sensors measure the pH or the ORP of aqueous solu­tions in pipelines, open tanks, or ponds. It is designed for harsh, dirty applications such as sourwater waste treatment and scrubbers, where a high performance, low maintenance, disposable sensor is required.
The combination electrode features a patented, large area polypropylene reference junction for resistance to process coatings, generally found in dirty, high solid applications. The patented helical reference pathway provides added resistance for poisoning applications. In addition, it is available in two configurations: glass/platinum electrode completely recessed within the reference junction for added life in abrasive or rough applications and the glass/platinum electrode partially exposed for added life and accuracy in vis­cous and low flow applications.
Both Models feature a titanium solution ground con­structed in an annular design around the pH/ORP elec­trode. The solution ground provides glass and refer­ence sensor diagnostics for preventative maintenance when used with the Model 54, 1055, 3081, 4081, and 81 pH/ORP transmitters. In addition Models 396P and 396PVP can be used with most non-diagnostic Rosemount Analytical and other manufacturer’s instru­ments.
Each sensor is housed in a molded polypropylene body with EPDM seals, making it virtually indestructi­ble and chemically resistant. Complete encapsulation eliminates leakage or high humidity problems tradition­ally found in other pH/ORP designs. The simplified
Measurements and Ranges *pH: 0-14
ORP: -1500 to 1500 mv
Available pH ACCUGLASS Types GPLR hemi or flat glass Wetted Materials Titanium, Polypropylene, EPDM, glass; platinum (ORP only) Process Connection 1 in. MNPT front and rear facing threads Temperature Range 0-100°C (32-212°F) Pressure Range-Hemi bulb 100-1136 kPa [abs] (0-150 psig) Pressure Range-Flat bulb 100-790 kPa [abs] (0-100 psig) Minimum Conductivity 75 µS/cm, nominal; 100 µS/cm Integral Cable 396P: Code 01 - 25 ft; Code 02 - 15 ft coaxial
396PVP: none - must use mating VP cable
Preamplifier Options 396P: Remote or Integral
396PVP: Remote only
Weight/Shipping W eight 0.45 kg/0.9 kg (1 lb/2 lb)
PERCENT LINEARITY*
pH range GPLR hemi bulb GPLR flat bulb
0 - 2 pH 94%
2 - 12 pH 97% 98% 12 - 13 pH 98% 95% 13 - 14 pH 98%
1.2 SPECIFICATIONS
Page 6
2
MODEL 396P and 396PVP SECTION 1.0
DESCRIPTIONS AND SPECIFICATIONS
1.3 ORDERING INFORMATION
The Model 396P Sensor is housed in a molded reinforced polypropylene
body with 1 in. MNPT threads suitable for inser tion, submersion or flow through installation. The sensor includes a general purpose pH electrode or a platinum ORP electrode, a patented reference junction and a solution ground. The Model 396P comes standard with a recessed electrode; an optional slotted tip is also available. In addition, the 396P features an optional integral hermetically sealed preamplifier and 15 ft or 25 ft cable lengths. Automatic temperature compensation, Pt 100 or 3K Balco, is standard with the Model 396P.
CODE PREAMPLIFIER/CABLE (Required Selection)
01 With integral preamplifier, 25 ft cable 02 Without integral preamplifier, 15 ft cable
CODE MEASURING ELECTRODE TYPE (Required Selection)
10 GPLR hemi bulb, General Purpose Low Resistivity (0-14 pH) 12 ORP 13 GPLR flat bulb, General Purpose Low Resistivity (2 - 13 pH)
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For Models 1181 (3K TC) 54 For Models 1054, 1054A, 2054, 2081;2700-Code 02 only (Pt 100 RTD) 55 For Models 54, 3081, 81, 4081, 1055 (PT-100 RTD)
396P - 01 - 10 - 55 EXAMPLE
NOTE:The Model 396P is also compatible with Models 1003 (option 02-50 only) and SCL-P/Q and (option 02-54 only).
CODE OPTIONAL SELECTION
41 Slotted Tip (not available on flat bulb sensors)
1. Mounting Accessories (optional)
Choose one: PN 915240-03, PVC flow through tee, 3/4 in. NPT process connection
PN 915240-04, PVC flow through tee, 1 in. NPT process connection PN 915240-05, PVC flow through tee, 1-1/2 in. NPT process connection PN 23728-00, acrylic low flow cell PN 2002011, 1-1/2 in. CPVC tee with 1-in. FNPT connection PN 11275-01, sensor handrail assembly
2. Junction Boxes (optional)
Remote Junction Boxes (used with option -02 sensors, for sensor to analyzer distances of more than 15 ft)
Choose one: PN 23555-00 includes preamplifier for Models 54, 81, 3081, 4081
PN 23309-03 and PN 22698-02 plug-in preamplifier for Model 1181Analyzer PN 23309-04 and PN 22698-03 plug-in preamplifier for Models 1054 series, 2054, 2081 Analyzers
PN 23054-03 includes preamplifier for Solu Cube Model 2700 Choose one: PN 9120516 BNC Adapter for use with remote junction boxes PNs 23309-03, 23309-04 Remote Junction Box (used with option -01 sensors) Choose one: PN 23550-00 cable extension board
3. Extension cables (used with remote junction boxes)
Choose one: PN 23646-01, 11 conductor, shielded, prepped
PN 9200273, 11 conductor, shielded, unprepped
TABLE 1-1. RECOMMENDED ACCESSORIES FOR FIRST TIME INSTALLATIONS
The Model 396P insertion/submersion sensor with integral cable is offered with or without a built-in preamplifier
MODEL
396P TUpH INSERTION/SUBMERSION POLYPROPYLENE pH/ORP SENSOR
Page 7
3
MODEL 396P and 396PVP SECTION 1.0
DESCRIPTIONS AND SPECIFICATIONS
OOppttiioonnaall SSllootttteedd TTiipp
is available on all
hemi bulb sensors,
ordered as option -41
FFllaatt TTiipp
is avail-
able with flat glass
bulb sensors
Examples of all sensing tip offerings
SShhrroouuddeedd TTiipp
is
standard on all
hemi bulb sensors
VVaarriiooppooll ccoonnnneeccttoorr
shown with mating variopol cable receptacle
MODEL 396PVP TUpH INSERTION/SUBMERSION POLYPROPYLENE pH/ORP SENSOR
CODE MEASURING ELECTRODE TYPE (Required Selection)
10 GPLR hemi bulb, General Purpose Low Resistivity (0-14 pH) 12 ORP 13 GPLR flat bulb, General Purpose Low Resistivity (2 - 13 pH)
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For Models 1181 (3K TC) 54 For Models 1054, 1054A, 2054, 2081;2700-Code 02 only (Pt 100 RTD) 55 For Models 54, 3081, 81, 4081 (PT-100 RTD)
CODE OPTIONAL SELECTION
41 Slotted Tip (not available on flat bulb sensors)
PN MATING VP CONNECTOR CABLE
((RReeqquuiirreedd ffoorr aallll nneeww iinnssttaallllaattiioonnss))
23645-06 15 ft.cable with mating VP connector, prepped with BNC on analyzer end 23645-07 15 ft.cable with mating VP connector, prepped without BNC on analyzer end*
The Model 396PVP Sensor has similar features to the Model 396P. However, the Model 396PVP is offered with the new Variopol (VP) connector and uses a mating VP cable (purchased separately). A remote preamplifier must be used with this sensor.
A Variopol cable is required for all new installations. See below for cable selection.
The Model 396PVP insertion/ submersion sen­sor with the VP (Variopol) connector
OTHER ACCESSORIES FOR MODELS 396P AND 396PVP
PART DESCRIPTION 22698-00 Preamplifier plug-in for junction box, for Model 1003, 22698-02 Preamplifier plug-in for junction box, for Models 1181, 1050 22698-03 Preamplifier plug-in for junction box, for Models 1054A/B, 2054, 2081 22743-01 Pt100 preamplifier for Model 1181 22744-01 3K Preamplifier for Model 1181 23557-00 Preamplifier for junction box for Models 54, 3081, 81, 4081 9210012 Buffer solution, 4.01 pH, 16 oz 9210013 Buffer solution, 6.86 pH, 16 oz 9210014 Buffer solution, 9.18 pH, 16oz R508-160Z ORP solution, 460 mv ± 10 at 20°C 12707-00 Jet Spray Cleaner
*For use with connections to Models 1181, 1054/A/B, 2081, 54, 81, 3081, 4081, and remote junction box PN 23555-00.
Page 8
MODEL 396P and 396PVP SECTION 2.0
INSTALLATION
SECTION 2.0
INSTALLATION
2. Do not install the sensor on the horizontal. The sensor must be 10° off the horizontal to ensure accuracy.
3. Do not install the sensor upside down.
4. Air bubbles may become trapped in the sensor end between the glass bulb and the sensor body. This problem is most commonly encountered in areas of low flow or during calibration. Shake the probe while immersed in solution to remove bub­bles. This problem can be avoided by ordering the sensor with the slotted tip (option -41).
In most cases, the pH sensor can simply be installed as shipped and readings with an accuracy of ± 0.6 pH may be obtained. To obtain greater accuracy or to ver­ify proper operation, the sensor must be calibrated as a loop with its compatible analyzer or transmitter.
2.2.1 Flow Through and Insertion Mounting.
Model 396P and 396PVP Sensors have a 1-inch MNPT process connection at the front of the sensor for mounting into a 1-1/2 inch tee or the process pipes. See Figure 2-1 through Figure 2-7 for installation con­figurations.
NOTE
LARGE PIPE WRENCHES MUST NOT BE USED TO TIGHTEN THE SENSOR INTO A FLANGE OR OTHER TYPE OF MOUNTING.
2.2.2 Submersion Mounting. Model 396P and
396PVP Sensors also have a 1 inch MNPT process connection at the back of the sensor. Utilizing a stan­dard 1 inch union, the sensor may be mounted to a 1 inch SCH 80 CPVC or PVDF standpipe (see Figure 2­x). Tapered pipe threads in plastic tend to loosen after installation. It is therefore recommended that Teflon
1
tape be used on the threads and that the tightness of the connection be checked frequently to assure that no loosening has occurred. To prevent rain water or con­densation from running into the sensor, a weatherproof junction box is recommended (see Figure 2-x). The sensor cable must be run through a protective conduit for isolation from electrical interference or physical abuse from the process. The sensor should be installed within 80° of vertical, with the electrode facing down. The sensor’ s cable should not be run with power or control wiring.
1
Teflon is a registered trademark of E.I. du Pont de Nemours & Co.
2.1 UNPACKING AND INSPECTION.
Inspect the outside of the carton for any damage. If dam­age is detected, contact the carrier immediately. Inspect the hardware. Make sure all the items in the packing list are present and in good condition. Notify the factory if any part is missing. If the sensor appears to be in satis­factory condition, proceed to Section 2.2, Mounting.
NOTE
Save the original packing cartons and materi­als as most carriers require proof of damage due to mishandling, etc. Also, if it is neces­sary to return the sensor to the factory, you must pack the sensor in the same manner as it was received. Refer to Section 6.0 for return instructions. If the sensor is to be stored, the vinyl boot should be filled with pH buffer solu­tion and replaced on sensor tip until ready to use.
CAUTION
Buffer solution, in the vinyl boot, may cause skin or eye irritation.
WARNING
Glass electrode must be wetted at all times (in storage and in line) to maximize sensor life.
2.2 MOUNTING.
The sensor has been designed to be located in indus­trial process environments. Temperature and pr essur e limitations must not be exceeded at any time. A cau­tion label regarding this matter is attached to the sen­sor. Please do not remove the label. See Figure 2-1.
CAUTION
Internal electrolyte fill solution may cause skin or eye irritation.
Mounting Guidelines:
1. Shake the sensor in a downward motion to remove any air bubbles that may be present inside the tip of the pH glass.
4
Page 9
5
MODEL 396P and 396PVP SECTION 2.0
INSTALLATION
FIGURE 2-1. Dimensional Drawing
DWG. NO. REV.
40396P01 D
MILLIMETER
INCH
FIGURE 2-2. Flow-Through Tee with Adapter (PN 915240-xx*)
MILLIMETER
INCH
xx* Process Connection
Threads
03 3/4 inch 04 1 inch 05 1-1/2 inch
SENSOR CABLE (OR VP CONNECTOR - NOT SHOWN)
SENSOR CABLE
(OR VP CONNECTOR
- NOT SHOWN)
Page 10
MODEL 396P and 396PVP SECTION 2.0
INSTALLATION
FIGURE 2-4. Model 396P with Insertion Mounting Adapter (PN 23242-02).
Not for use with Model 396PVP.
Mounting adapter allows for sensor removal without twisting or
disconnecting interconnecting cable for ease of maintenance.
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
40396P02 A
FIGURE 2-3. Flow-Through and Insertion Installations
1-1/2” PIPE “Y”
1-1/2” X 1” Reducing Bushing
1-1/2” X 1”
Reducing
Bushing
1-1/2” X 1”
Reducing
Bushing
1-1/2” Pipe Tee PN 2002011
1-1/2” Pipe Tee PN 2002011
PIPE “Y”
INSTALLATION
SHOWN
STRAIGHT
FLOW
SHOWN
FLOW
ANGLE
FLOW
SHOWN
200.0
7.875
121.92
4.80
85.85
3.375
NECK, UNION FITTING (316 SST) 2-5/8 IN. WRENCH OPENING
INSERTION MOUNTING ADAPTER PN 23242-02 (INCLUDES: PEEK ADAPTER, 304 SST UNION FITTING)
1-1/2 IN. MNPT
2.531.8 ACME THREAD (TYP)
CABLE (OR VP CONNECTOR - NOT SHOWN)
PEEK ADAPTER 1 IN FNPT X 3/4 IN.FNPT (REVERSIBLE)
NUT, HEX UNION 2 IN. 3 IN.WRENCH OPENING (304 SST)
6
Page 11
MODEL 396P and 396PVP SECTION 2.0
INSTALLATION
FIGURE 2-5. Submersion Installations
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
40396P04 A
DWG. NO. REV.
40396P03 A
Handrail Mounting Assembly PN 11275-01
7
Page 12
8
MODEL 396P and 396PVP SECTION 2.0
INSTALLATION
FIGURE 2-6. Low Flow Cell PN 23728-00
MILLIMETER
INCH
FIGURE 2-7. Jet Spray Cleaner PN 12707-00
SENSOR CABLE
(OR VP CONNECTOR
- NOT SHOWN)
Page 13
9
Figures in this section provide the guidelines for wiring the 396P-01 sensor to various Analyzer/Transmitter instruments.
To determine which wiring guideline to use, locate the model number of the sensor to be installed.
1. If the cable needs to be extended, use a high quality eleven conductor double shielded instru­ment cable available from Rosemount Analytical. Refer to Table 3-1 for the appropriate junction box to use and the corresponding wiring details.
NOTE
If the cable is too long, loop up the excess cable. If the cable has to be shortened, cut and terminate each conductor neatly and
make sure that the overall (outermost) drain wire is not shorted out with either of the two inner drain wires (shields).
2. Signal cable should be run in a dedicated conduit (preferably an earth grounded metallic conduit) and should be kept away from AC power lines. For your convenience, a wire nut kit is furnished (in a plastic bag wrapped around the cable).
NOTE
For maximum EMI/RFI protection when wiring from the sensor to the junction box, the outer braid of the sensor should be connected to the outer braided shield of the extension cable. The outer braid of the extension cable to the instrument must be terminated at earth ground or by using an appropriate metal cable gland fitting that provides a secure connection to the instru­ment cable.
Preamplifier Location Analyzer/Transmitter
Sensor Integral to Integral to Remote Interconnect 1054, 1054A, 54, 81, 1055 1181 1050 1003 SCL or Refer to
Model Sensor Analyzer Junction (unprepped 1054B, 2054 3081, 1060 1023 SoluCube Figure
Options
Box (PN) PN) 2081 4081 #
396P-01-()-50 ✓✓ 3-2 396P-01-()-50 Ext. box 9200254 ✓✓ 3-2
(22719-02)
396P-01-()-54 ✓✓ 3-3 396P-01-()-54 Ext. box 9200254 3-3
(22719-02)
396P-01-()-55 ✓✓3-5 396P-01-()-55 Ext. box 9200273 3-4
(23550-00)
396P-01-()-55 9200273 3-6
or 23646-01
TABLE 3-1. Wiring Matrix Guide
MODEL 396P and 396PVP SECTION 3.0
WIRING - MODEL 396P
SECTION 3.0
WIRING — MODEL 396P-01
Page 14
10
MODEL 396P and 396PVP SECTION 3.0
WIRING - MODEL 396P
Wiring. The Model 396P has an optional built-in pre-
amplifier and is offered with a shielded cable. The cable should be handled carefully and kept dry and free of corrosive chemicals at all times. Extreme care should be used to prevent it from being twisted, dam­aged or scraped by rough, sharp edges or surfaces. Please refer to Figures 3-1 through 3-6 for wiring Model 396P-01.
FIGURE 3-1.Wire Functions for Model 396P with and without Preamplifier
DANGER
DO NOT CONNECT SENSOR CABLE TO POWER LINES. SERIOUS INJURY MAY RESULT.
Model 396P Option 01 (with preamplifier) cable
Model 396P Option 02 (without preamplifier) cable
Page 15
11
MODEL 396P and 396PVP SECTION 3.0
WIRING - MODEL 396P
FIGURE 3-2. Wiring Details Model 396P-01-50 for use with and without Junction Box
(PN 22719-02) for Models 1181, 1050, and 1060.
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
40396P15 B
Page 16
12
FIGURE 3-3. Wiring Details Model 396P-01-54 for use with and without Junction Box
(PN 22719-02) for Models 1054, 1054A, 1054B, 2054, 2081, and Extension Cable (9200254).
MODEL 396P and 396PVP SECTION 3.0
WIRING - MODEL 396P
DWG. NO. REV.
40396P13 D
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
Page 17
13
MODEL 396P and 396PVP SECTION 3.0
WIRING - MODEL 396P
FIGURE 3-4. Wiring Details Model 396P-01-55 for use with Junction Box
(PN 23550-00) for Models 54, 3081, 4081, and 81.
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
40396P08 B
Page 18
14
MODEL 396P and 396PVP SECTION 3.0
WIRING - MODEL 396P
FIGURE 3-5.Wiring Details Model 396P-01-55 for use with Models 54, 81, 3081, and 4081
FIGURE 3-6.Wiring Model 396P-01-( )--55 Sensor to Model 1055-10-22-32 Analyzer
Page 19
Figures 4-1 thru 4-22 provide the guidelines for wiring the sensor to various Analyzer/T ransmitter instruments.
To determine which wiring guideline to use, locate the model number of the sensor to be installed.
1. If the cable needs to be extended, use a high qual­ity eleven conductor double shielded instrument cable available from Rosemount Analytical. Refer to Table 4-1 to refer to the appropriate junction box to use and the corresponding wiring details.
NOTE
If the cable is too long, loop up the excess cable. If the cable has to be shortened, cut and terminate each conductor neatly and
make sure that the overall (outermost) drain wire is not shorted out with either of the two inner drain wires (shields).
2. Signal cable should be run in a dedicated conduit (preferably an earth grounded metallic conduit) and should be kept away from AC power lines. For your convenience, a wire nut kit is furnished (in a plastic bag wrapped around the cable).
NOTE
For maximum EMI/RFI protection when wiring from the sensor to the junction box, the outer braid of the sensor should be connected to the outer braided shield of the extension cable. The outer braid of the extension cable to the instrument must be terminated at earth ground or by using an appropriate metal cable gland fitting that provides a secure connection to the instrument cable.
TABLE 4-1. Wiring Matrix Guide
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
SECTION 4.0
WIRING — MODELS 396P-02 AND 396PVP
15
Preamplifier Location Analyzer/Transmitter
Sensor Integral to Integral to Remote Interconnect 1054, 1054A, 54, 81, 1055 1181 1050 1003 SCL or Refer to
Model Sensor Analyzer Junction (unprepped 1054B, 2054 3081, 1060 1023 SoluCube Figure
Options
Box (PN) PN) 2081 4081 #
396P-02-()-55 ✓✓4-3, 4-6,
396PVP-55 4-9
396P-02-()-50 ✓✓4-4
396PVP-50
396P-02-()-50 23309-03 9200254 ✓✓✓ 4-22
396PVP-50
396P-02-()-50 23309-03 9200273 or ✓✓✓ 4-8
396PVP-50 23646-01 396P-02-()-54 23309-04 9200254 4-23 396P-02-()-54 23309-04 9200273 or 4-20
396PVP-54 23646-01 396P-02-()-54 ✓✓ 4-5, 4-18
396PVP-54 4-19 396P-02-()-55 23555-00 9200273 or 4-7, 4-10
396PVP-55 23646-01 4-12 396P-02-()-54
661-898695
4-14
396PVP-54
4-13
396P-02-54 23309-04 9200273 or
4-11
396PVP-54 23646-01
4-21
396P-02-55
✓✓4-15
396PVP-55 4-16
396P-02-55
9200273 or
4-17
396PVP-55 23646-01
Page 20
16
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
FIGURE 4-1.Wire Functions
for Mating Variopol Cable used
with Model 396PVP
FIGURE 4-2. Connector Pins
and Their Functions
FIGURE 4-3. Wiring Details for Models 396PVP
or 396P-02-55 with Mating Variopol Cable
(PN 23645-07) for use with Model 81
FIGURE 4-4. Wiring Details for Models 396PVP
or 396P-02-50 with Mating Variopol Cable
(PN 23645-07) for use with Model 1181
FIGURE 4-5. Wiring Details for Models 396PVP
or 396P-02-54 with Mating Variopol Cable
(PN 23645-07) for use with Model 2081
FIGURE 4-6. Wiring Details for Models 396PVP
or 396P-02-55 with Mating Variopol Cable
(PN 23645-07) for use with Models 3081 and 4081
Page 21
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
FIGURE 4-8. Wiring Details for Models 396PVP
or 396P-02-50 with Mating Variopol Cable
(PN 23645-06) for use with Remote Junction Box
(PN 23309-03) to Model 1181
FIGURE 4-7. Wiring Details for Models 396PVP
or 396P-02-55 with Mating Variopol Cable
(PN 23645-07) for use with Remote Junction Box
(PN 23555-00) to Model 81
FIGURE 4-9. Wiring Details for
Models 396PVP or 396P-02-55 with
Mating Variopol Cable (PN 23645-07)
for use with Model 54
FIGURE 4-10. Wiring Details for Models 396PVP or 396P-02-55
with Mating Variopol Cable (PN 23645-07) for use with Remote
Junction Box (PN 23555-00) to Model 54
17
Page 22
18
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
FIGURE 4-12. Wiring Details for Models 396PVP or 396P­02-55 with Mating Variopol Cable (PN 23645-07) for use with Remote Junction Box (PN 23555-00) to Models 3081 & 4081
FIGURE 4-11. Wiring Details for Models 396PVP or 396P-
02-54 with Mating Variopol Cable (PN 23645-06) for use
with Remote Junction Box (PN 23309-04) to Model 2081
FIGURE 4-13. Wiring Details for Models 396PVP or
396P-02-54 with Mating Variopol Cable (PN 23645-07)
for use with Model 2700
FIGURE 4-14. Wiring Details for Models 396PVP or
396P-02-54 with Mating Variopol Cable (PN 23645-07)
for use with Model SCL-(P/Q)
Page 23
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
FIGURE 4-15. Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Model 1055
FIGURE 4-16. Wiring Model 396P-(02)-( )--54/55-(61) Sensor to Model 1055-01-10-22-32 Analyzer
FIGURE 4-17. Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07)
for use with Remote Junction Box (PN 23557-00) to Model 1055
NOTE:This wiring diagram can also be used for wiring a Model 396P-01.The wire colors and functions for
the extension cable (PN 9200273) are the same as the 396P-01 wire colors and functions.
19
Page 24
20
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
FIGURE 4-18. Wiring Details for Models 396PVP
or 396P-02-54 with Mating Variopol Cable
(PN 23645-07) for use with Model 1054
FIGURE 4-19. Wiring Details for Models 396PVP
or 396P-02-54 with Mating Variopol Cable
(PN 23645-06) for use with Models 1054A/B & 2054
FIGURE 4-21. Wiring Details for Models 396PVP or
396P-02-54 with Mating Variopol Cable (PN 23645-06)
for use with Remote Junction Box (PN 23309-04) to
Models 1054A/B and 2054
FIGURE 4-20. Wiring Details for Models 396PVP or 396P-
02-54 with Mating Variopol Cable (PN 23645-06) for use
with Remote Junction Box (PN 23309-04) to Model 1054
Page 25
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
FIGURE 4-22. Wiring Details Model 396P-02-50 for use with Junction Box
(PN 23309-03) and
Remote Preamplifier, Extension Cable
(PN 9200254).
DWG. NO. REV.
40396P25 A
FIGURE 4-23. Wiring Details Model 396P-02-54 for use with Junction Box (PN 23309-04) and
Remote Preamplifier (PN 22698-03), Extension Cable (9200254).
DWG. NO. REV.
40396P24 A
21
Page 26
22
1. While obtaining a process solution sample (it is recommended that the sample is taken close to the sensor), record the pH value that is shown on the analyzer/transmitter display.
2. Measure and record the pH of the process solution sample with another temperature compensated, calibrated pH instrument. For best results, standardization should be performed at the process temperature.
3. Adjust the analyzer/transmitter value to the standardized value.
5.2 MODEL 396P and 396PVP ORP SENSORS
5.2.1 SENSOR PREPARATION. Most industrial applications have
a number of ORP reactions occurring in sequence or simultaneous­ly. There can be several components that are oxidized or reduced by the reagents that are used. Theoretically, the ORP potential is absolute because it is the result of the oxidation-reduction equilibri­um. However, the actual measured potential is dependent on many factors, including the condition of the surface of the ORP platinum electrode. Therefore, the sensor should be allowed 1-2 hours to
become “conditioned” to the stream when first set-up or after being cleaned.
5.2.2 ORP CALIBRATION
1. Make a temporar y electrical connection between the sensor and the instrument.
2. Obtain an ORP standard solution (PN R508-160Z), or a standard solution can also be made quite simply by adding a few crystals of quinhydrone to either pH 4 or pH 7 buffer. Quinhydrone is only slight­ly soluble therefore a few crystals will be required. (Refer to Section
4.3. for an alternate ORP standard solution).
3. Immerse the sensor in the standard solution. Allow 1-2 minutes for the ORP sensor to stabilize.
4. Adjust the standardize control of the instrument to the solution value shown in Table 5-1 (below) or on the label of the standard solution. The resulting potentials, measured with a clean platinum electrode and saturated KCl/AgCl reference electrode, should be within ±20 millivolts of the value. Solution temperature must be noted to ensure accurate interpretation of results. The ORP value of saturated quin­hydrone solution is not stable over long periods of time. Therefore, these standards should be made up fresh each time they are used.
5. Remove the sensor from the buffer, rinse and install in the pr ocess.
MODEL 396P and 396PVP SECTION 5.0
START UP AND CALIBRATION
SECTION 5.0
START UP AND CALIBRATION
5.1 MODELS 396P and 396PVP pH SENSORS
5.1.1 SENSOR PREPARATION. Shake down the sensor
to remove any air bubbles that may be present at the tip of the pH glass bulb. In most cases, the pH sensor can simply be installed as shipped and readings with an accuracy of ±
0.6 pH may be obtained. To obtain greater accuracy or to verify proper operation, the sensor must be calibrated as a loop with its compatible analyzer or transmitter.
5.1.2 pH CALIBRATION. After a temporar y connection is
established between the sensor and the instrument, a buffer calibration may be performed. Consult appropriate pH/ORP analyzer or transmitter instruction manual for specific calibra­tion and standardization procedures, or see below for recom­mended two-point buffer calibration procedure.
Recommended tw
o-point buffer calibration procedure:
Select two stable buffer solutions, preferably pH 4.0 and 10.0 (pH buffers other than pH 4.0 and pH 10.0 can be used as long as the pH values are at least two pH units apart).
NOTE
A pH 7.0 buffer solution reads a mV value of approxi­mately zero, and pH b uffers read appro ximately 59.1 mV for each pH unit above or below pH 7.0. Check the pH buffer manufacturer specifications for millivolt values at various temperatures since it may aff ect the actual value of the buffer solution mV/pH value.
1. Immerse sensor in the first buffer solution. Allow sensor to adjust to the buffer temperature (to avoid errors due to temperature differences between the buffer solution and sensor temperature) and wait for reading to stabilize. Value of buffer can now be acknowledged by analyzer/transmitter.
2. Once the first buffer has been acknowledged by the ana­lyzer/transmitter, rinse the buffer solution off of the sensor with distilled or deionized water.
3. Repeat steps 1 and 2 using the second buffer solution.
4. Once the analyzer/transmitter has acknowledged both buffer solutions, a sensor slope (mV/pH) is established (the slope value can be found within the analyzer/ transmitter).
5. The slope value should read about 59.1 mV/pH for a new sensor and will decrease over time to approximately 47­49 mV/pH. Once the slope reads below the 47-49 mV/pH range, a new sensor should be installed to maintain accurate readings.
Recommended pH Sensor Standar
dization:
For maximum accuracy, the sensor can be standardized online or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sensor’s slope but will simply adjust the analyzer’s reading to match that of the known process pH.
pH 4 Solution pH 7 Solution
Temp °°C 20 25 30 20 25 30 Millivolt Potential 268 264 260 94 87 80
TABLE 5-1. ORP of Saturated Quinhydrone
Solution (In Millivolts)
CAUTION
The solution used during the following check is an acid and should be handled with care. Follow the directions of the acid manufacturer. Wear the proper protective equipment. Do not let the solution come in contact with skin or clothing. If con­tact with skin is made, immediately rinse with clean water.
Page 27
23
MODEL 396P and 396PVP SECTION 6.0
MAINTENANCE
The Model 396P and 396PVP Sensors require mini­mum maintenance. The sensor should be kept clean and free of debris and sediment at all times. The fre­quency of cleaning by wiping or brushing with a soft cloth or brush is determined by the nature of the solu­tion being measured. The sensor should be removed from the process periodically and checked in buffer solutions.
WARNING
BEFORE REMOVING THE SENSOR, be absolutely certain that the process pres­sure is reduced to 0 psig and the process temperature is lowered to a safe level!
If the sensor will not calibrate, refer to your analyzer/ transmitter instruction manual for proper test proce­dures. If it is determined that the sensor has failed, it should be discarded and replaced.
6.1 AUTOMATIC TEMPERATURE COMPENSATOR.
The temperature compensator element is temp­erature sensitive and can be checked with an ohm­meter. Resistance increases with temperature.
The 3K element will read 3000 ohms ± 1% at 25°C (77°F) and a Pt100 will read 110 ohms. Resistance varies with temperature for a 3K and Pt100 element and can be determined according to Table 6-2 or the following formula:
RT=Ro[l+R1(T-20)]
Where RT = Resistance T = Temperature in °C
Refer to Table 6-1 for Roand R1values
6.2 MODEL 396P & 396PVP pH SENSORS
4.2.1 ELECTRODE CLEANING. If the electrode is
coated or dirty, clean as follows:
1. Remove the sensor from process.
2. Wipe the glass bulb with a soft, clean, lint free cloth or tissue. If this does not remove the dirt or coating, go to Step 3. (Detergents clean oil and grease; acids remove scale.)
3. Wash the glass bulb in a mild detergent solution and rinse it in clean water. If this does not clean the glass bulb, go to Step 4.
SECTION 6.0
MAINTENANCE
CAUTION
The solution used during the following check is an acid and should be handled with care. Follow the directions of the acid manufacturer. Wear the proper protective equipment. Do not let the solution come in contact with skin or clothing. If contact with skin is made, immediately rinse with clean water.
4. Wash the glass bulb in a dilute 5% hydrochloric acid solution and rinse with clean water. Soaking the sensor overnight in the acid solution can improve cleaning action.
NOTE
Erroneous pH results may result immedi­ately after acid soak, due to reference junction potential build-up.
Replace the sensor if cleaning does not restore sensor operation.
TABLE 6-1
Roand R1VALUES FOR TEMPERATURE
COMPENSATION ELEMENTS
Resistance
Temperature °C (Ohms) ±1%
3K Pt100
0 2670 100.0 10 2802 103.8 20 2934 107.7 25 3000 109.6 30 3066 111.5 40 3198 115.4 50 3330 119.2 60 3462 123.1 70 3594 126.9 80 3726 130.8 90 3858 134.6
100 3990 138.5
Temperature
Compensation Element
R
o
R
1
3K 2934 .0045
Pt100 107.7 .00385
TABLE 6-2
TEMPERATURE vs RESISTANCE OF AUTO
T.C. ELEMENTS
Page 28
24
MODEL 396P and 396PVP SECTION 6.0
MAINTENANCE
6.3 MODEL 396P and 396PVP ORP
4.3.1 Platinum Electrode Check. The platinum elec-
trode may be checked as follows: There are three types of standard solutions which may be used to check the ORP electrode/transmitter system.
Type 1: A prepared ORP standard solution (PN R508-160Z)
Type 2: Another type of commonly used ORP stan­dard solution is the saturated quinhydrone solution. Refer to Section 5.2.
CAUTION
The solution used during the following check is an acid and should be handled with care. Follow the directions of the acid manufacturer. Wear the proper protective equipment. If contact with skin of clothing is made, immediately rinse with plenty of clean water.
Type 3: A third ORP standard solution can be pre­pared from the following recipe: Dissolve 39.2 grams of reagent grade ferrous ammonium sulfate, Fe(NH4)
2
(SO4)2• 6H2O and 48.2 grams of reagent grade fer­ric ammonium sulfate, FeNH4(SO4)2• 12H2O, in approximately 700 milliliters of water (distilled water is preferred, but tap water is acceptable). Slowly and carefully add 56.2 milliliters of concentrated sulfuric acid. Add sufficient water to bring the total solution volume up to 1000 ml. This standard ORP solution, although not as simple to prepare as the quinhydrone recipe, is much more stable, and will maintain its milli­volt value for approximately one year when stored in glass containers. This solution (ferric/ferrous ammoni­um sulfate) will produce a nominal ORP of 476 +20 mV at 25°C when used with a saturated KCl/AgCl refer­ence electrode and platinum measuring electrode. Some tolerance in mV values is to be expected due to the rather large liquid reference junction potentials which can arise when measuring this strongly acidic and concentrated solution. However, if the measuring electrodes are kept clean and in good operating con­dition, consistently repeatable calibrations can be car­ried out using this standard solution.
6.3.2 Cleaning Platinum Electrode. The electrode can be restored to normal operation by simply cleaning the platinum electrode with baking soda. Polish it by rub­bing it with a damp paper towel and baking soda until a bright, shiny appearance is attained.
Page 29
25
MODEL 396P and 396PVP SECTION 7.0
DIAGNOSTICS AND TROUBLESHOOTING
SECTION 7.0
DIAGNOSTICS AND TROUBLESHOOTING
7.1 MODEL 54/81/3081/4081 DIAGNOSTICS AND TROUBLESHOOTING
The Model 54 Analyzer and Models 81 and 3081 Transmitters automatically search for fault conditions that would cause an error in the measured pH value, as does the Model 1054B pH/ORP Analyzer to a lesser degree. Refer to the applicable Instruction Manual for a complete description of the analyzer’s fault conditions.
Table 7-1, below , lists the Model 54’s, 4081’s, 3081’s, and 81’s diagnostic messages that indicate a possible sen­sor problem. A more complete description of the problem and a suggested remedy corresponding to each mes­sage is also listed.
DIAGNOSTIC MESSAGE 54 3081/81/4081 DESCRIPTION OF PROBLEM REMEDY
“Calibration Warning” 1. Aged glass. 1. Perform buffer calibration.
2. Sensor not immersed. 2. Be sure electrode measuring tip is in
CALIbrAtE process.
“Cracked glass failure” Broken or cracked glass. Replace sensor.
6LASS fAIL
“High reference impede” 1. Liquid junction coated. 1. Clean sensor; replace if necessary.
2. Reference Cell gel depleted. 2. Replace sensor.
rEF fAIL or rEF WjArn 3. Sensor not immersed. 3. Be sure electrode measuring tip is in
process.
“Input voltage high” pH input shor ted or sensor Check wiring. Replace sensor if “Input voltage low” miswired. necessary.
“Old glass warning” 1. Glass electrode worn out. 1. Replace sensor.
2. Sensor not immersed. 2. Be sure electrode measuring tip is in
6LaSS WjArn
process.
“Reference offset err” Reference electrode poisoned. Replace sensor. (offline only)
Std Err
“Ref voltage high” 1. Reference shorted or sensor Check wiring and installation. Replace “Ref voltage low” miswired. sensor if necessary.
2. Sensor not immersed
“Sensor line open” 1.
Open wire between sensor and analyzer.
1. Check sensor wiring.
2. Interconnecting cable greater than 2. Relocate analyzer.
LInE FAIL 1000 ft. “Sensor miswired” 1.
Open wire between sensor and analyzer.
1. Check wiring.
2. Bad preamplifier. 2. Replace preamplifier. (Code 02 only)
“Temp error high” 1. Open or shorted RTD. 1. Replace sensor. “Temp error low” 2. Temperature out of range. 2. Check process temperature.
tEMP H I tEMP LO
TABLE 7-1. Troubleshooting with Advanced Diagnostics
Page 30
26
7.2 TROUBLESHOOTING WITHOUT ADVANCED DIAGNOSTICS.
Table 7-2, below, lists common problems, causes and remedies typically encountered in process measurement.
Problem Probable Cause Remedy
Meter reads off scale. (Display Defective preamplifier Replace preamplifier (for code 02 reads overrange).
sensors). For code 01, replace sensor.
T.C. element shorted Check T.C. element as instructed
in Section 6.1 and replace sensor if defective.
Sensor not in process. Sample Make sure sensor is in process with stream is low or air bubbles are sufficient sample stream (refer to present. Section 2.0 for installation details).
Open glass electrode Replace sensor. Reference element open - no contact
Replace sensor.
Display reads between 3 and 6 pH Electrode cracked Replace sensor. regardless of actual pH of solution or sample.
Meter or display indication swings T.C. element shorted Check T.C. element as instructed or jumps widely in AUTO T.C. Mode. in Section 6.1 and replace
sensor if defective.
Span between buffers extremely T.C. element open Check T.C. element as instructed short in AUTO T.C. Mode. in Section 6.1 and replace sensor
if defective.
Sluggish or slow meter indication Electrode coated Clean sensor as instructed in for real changes in pH level. Sections 6.2 or Section 6.3.2.
Replace sensor if cracked.
Electrode defective Replace sensor.
Transmitter cannot be standardized. Electrode coated or cracked Clean Sensor as instructed in
Sections 6.2 or Section 6.3.2 Replace sensor if cracked.
Defective preamplifier Replace preamplifier.
Transmitter short spans between Aged glass electrode or high Replace sensor. two different buffer values. temperature exposure
Electrode coated Clean Sensor as instructed in
Section 6.2 or Section 6.3.2.
Replace sensor if cracked. Air bubbles trapped in sensor end Shake the sensor in solution. See between glass bulb and sensor body Section 2.0 for mounting guidelines.
TABLE 7-2.Troubleshooting without Advanced Diagnostics
MODEL 396P and 396PVP SECTION 7.0
TROUBLESHOOTING
Page 31
27
MODEL 396P and 396PVP SECTION 7.0
TROUBLESHOOTING
TABLE 7-3. Model 396P and 396PVP pH/ORP Replacement Parts and Accessories
PN DESCRIPTION QUANTITY
11275-01 Sensor Handrail Mounting Assembly 2002011 Flow Cell, CPVC, 1 inch FNPT 23242-02 Mounting Adapter, Insertion, 1¼-inch MNPT (304 S.S.) X 1” FNPT (PEEK) 23309-03 Junction Box, for remote preamplifier Code-50 23309-04 Junction Box, for remote preamplifier Code-54 23646-01 Cable, Extension (Prepped) for Models 54, 81, 3081, and 4081 23555-00 Junction Box with preamplifier, Models 54, 81, 3081, and 4081 compatible 23557-00 Preamplifier, remote for Junction Box, Models 54, 81, 3081, and 4081 compatible 22698-00 Preamplifier, Plug-in, Model 1003 compatible (for Code 02-50) 1 22698-02 Preamplifier, Plug-in, Models 1181 and 1050 compatible (for Code 02-50) 1 22698-03 Preamplifier, Plug-in, Models 1054, 1054A, 1054B, 2054, and 2081 compatible (for Code 02-54) 1 22719-02 Junction Box, w/o Preamplifier 33081-00 Adapter Insert, PEEK, 1 X 3/4-inch, for 23242-02 7901631 Shroud, PVC 9200254 Cable, 4 conductor, 22 AWG, shielded pair, for 1054/A/B, 2054, and 1181 9200273 Cable, Extension (Unprepped) for Models 54, 81, 3081, and 4081 23645-06 15 ft (4.6 m) cable with mating VP connector with BNC on transmitter end 23645-07 15 ft (4.6 m) cable with mating VP connector with bare wires on transmitter end 9210012 Buffer Solution, 4.01pH, 16 oz 4 9210013 Buffer Solution, 6.86pH, 16 oz 4 9210014 Buffer Solution, 9.18pH, 16 oz 4 9322014 Union, KYNAR
1
9330022 Union, CPVC 9120516 BNC Adapter 915240-04 Tee, Flow-through, 2” PVC, 1” NPT 9550175 O-ring for Mounting Adapter (23242-02) R508-160Z ORP Standard Solution, 460mV ±10 at 20°C 23550-00 Junction Box with Extension Board, Models 54, 81, 3081, and 4081 compatible 661-898695 Cable 5 Conductor (for Model 2700 only)
Page 32
28
MODEL 396P and 396PVP SECTION 8.0
RETURN OF MATERIAL
SECTION 8.0
RETURN OF MATERIAL
8.1 GENERAL.
To expedite the repair and return of instruments, proper communication between the customer and the factory is important. Before returning a product for repair, call 1-949-863-1181 for a R etur n Materials Authorization (RMA) number.
8.2 W ARRANTY REP AIR.
The following is the procedure for returning instru­ments still under warranty:
1. Call Rosemount Analytical for authorization.
2. T o verify warranty, supply the factory sales order number or the original purchase order number. In the case of individual parts or sub-assem­blies, the serial number on the unit must be sup­plied.
3. Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the materials in the same man­ner as they were received.
4. Send the package prepaid to: Rosemount Analytical Inc., Uniloc Division
Uniloc Division 2400 Barranca Parkway Irvine, CA 92606
Attn: Factory Repair
RMA No. ____________
Mark the package: Returned for Repair
Model No. ____
8.3 NON-WARRANTY REPAIR.
The following is the procedure for returning for repair instruments that are no longer under warran­ty:
1. Call Rosemount Analytical for authorization.
2. Supply the purchase order number, and make sure to provide the name and telephone number of the individual to be contacted should addi­tional information be needed.
3. Do Steps 3 and 4 of Section 8.2.
NOTE
Consult the factory for additional infor­mation regarding service or repair.
Page 33
FROM: RETURN BILL TO:
_____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________
CUSTOMER/USER MUST SUBMIT MATERIAL SAFETY SHEET (MSDS) OR COMPLETE STREAM COMPOSITION, AND/OR LETTER CERTIFYING THE MATERIALS HAVE BEEN DISINFECTED AND/OR DETOXIFIED WHEN RETURNING ANY PRODUCT, SAMPLE OR MATERIAL THAT HAVE BEEN EXPOSED TO OR USED IN AN ENVIRONMENT OR PROCESS THAT CONTAINS A HAZARDOUS MATERIAL ANY OF THE ABOVE THA T IS SUBMITTED TO ROSEMOUNT ANALYTICAL WITHOUT THE MSDS WILL BE RETURNED TO SENDER C.O.D. FOR THE SAFETY AND HEALTH OF OUR EMPLOYEES. WE THANK YOU IN ADVANCE FOR COMPLIANCE TO THIS SUBJECT.
SENSOR OR CIRCUIT BOARD ONLY: (Please reference where from in MODEL / SER. NO. Column)
1. PART NO. _____________________________1. MODEL ____________________________________1. SER. NO.__________________
2. PART NO. _____________________________2. MODEL ____________________________________2. SER. NO.__________________
3. PART NO. _____________________________3. MODEL ____________________________________3. SER. NO.__________________
4. PART NO. _____________________________4. MODEL ____________________________________4. SER. NO.__________________
PLEASE CHECK ONE:
REPAIR AND CALIBRATE DEMO EQUIPMENT NO. ______________________________
EVALUATION OTHER (EXPLAIN) ___________________________________
REPLACEMENT REQUIRED? YES NO _______________________________________________________
DESCRIPTION OF MALFUNCTION: __________________________________________________________________________________________________________________ __________________________________________________________________________________________________________________ __________________________________________________________________________________________________________________
WARRANTY REPAIR REQUESTED:
YES-REFERENCE ORIGINAL ROSEMOUNT ANALYTICAL ORDER NO. _______________________________________________
CUSTOMER PURCHASE ORDER NO. ________________________________________________________
NO-PROCEED WITH REPAIRS-INVOICE AGAINST P.O. NO. _________________________________________________________
NO-CONTACT WITH ESTIMATE OF REPAIR CHARGES: LETTER ■_________________________________________________
PHONE ________________________________________________
NAME___________________________________________________________ PHONE______________________________________________ ADDRESS_______________________________________________________________________________________________________________
_____________________________________________________________________ ZIP______________________________________________
RETURN AUTHORITY FOR CREDIT ADJUSTMENT [Please check appropriate box(s)]
WRONG PART RECEIVED REPLACEMENT RECEIVED
DUPLICATE SHIPMENT REFERENCE ROSEMOUNT ANALYTICAL SALES ORDER NO. _____________
RETURN FOR CREDIT RETURN AUTHORIZED BY: __________________________________________
WARRANTY DEFECT______________________________________________________________________________________________
_________________________________________________________________________________________________________________
24-6047
RETURN OF MATERIALS REQUEST
•IMPORTANT!
This form must be completed to ensure expedient factory service.
R E P A
I
R S
T A T U S
R E A S O N
F O R
R E T U R N
C U S T
O
M
E R
N O
T
I
C
E T
O
S E N D E R
Rosemount Analytical Inc. Uniloc Division
2400 Barranca Parkway Irvine, CA 92606 USA Tel: 1-800-854-8257 Fax: (949)-474-7250
Page 34
U.S. Field Service
Field Watch Response Center 1-800-654-7768
International
ARGENTINA
Emerson Argentina Buenos Aires 54 1 780311, 795-9338
AUSTRALIA
Fisher-Rosemount Pty. Ltd. Baysw ater, Victoria (61)(39)721-0200
AUSTRIA
Fisher-Rosemount - AG Neudorf (43)(2236)607
BELGIUM
Fisher-Rosemount N.V./S.A. Diegem (32)(2)716-77-11
BRAZIL
Fisher-Rosemount do Brasil Industria e Comercio Ltda. Sorocada (55)(11)228 3788
CANADA
Rosemount Instruments Ltd. Calgary (800) 268-1151
CHINA
Fisher-Rosemount Pudong, Shanghai (86)(215)899-4415
CROATIA
Fisher-Rosemount Croatia Zagreb (385)(1)305 071
CZECH REPUBLIC
Fisher-Rosemount Praha (420)(2)81002666-7
DENMARK
Fisher-Rosemount A/S Alleroed (45)48-17-03-33
ENGLAND
Fisher-Rosemount Limited Bognor Regis (44)(1243)863-121
FINLAND
Oy Valment-Rosemount AG ESPOO (358)(9)549-541
FRANCE
Fisher-Rosemount S.A. Lyon (33)(7)215-98-00
GERMANY
Fisher-Rosemount GmbH & Co. D-63549 Hasselroth (49)(6055)884-0
HUNGARY
Fisher-Rosemount Budapest (36)(1)4624001
INDIA
Fisher-Rosemount (India) Ltd. New Bombay (91)(22)762-9191/9200
ITALY
Fisher-Rosemount Italia s.r.l. Milan (11)(39)(39)27021
JAPAN
Fisher-Rosemount Japan Co., Ltd. Tokyo (81)(3)5403-8529
KOREA
Fisher-Rosemount Korea Ltd. Seoul (82)(2)515-2321
MIDDLE EAST
Fisher-Rosemount Middle-East Dubai, United Arab Emirates (971)(4)835235
MALAYSIA
Rosemount Instruments SDN BHD Kuala Lumpur (60)(3)244-5688
MEXICO
Fisher-Rosemount S.A. de C.V. Tlalnepantla (011)(525)728-0885
NEW ZEALAND
Fisher-Rosemount Auckland (64)(9)444-1646
NORWAY
Fisher-Rosemount A/S Porsgrunn (47)35-55-5600
POLAND
Fisher-Rosemount Warszawa (48)(22)857 3856/ 647 3865
PORTUGAL
Fisher-Rosemount Portugal Instrumentos Limitada Alfragide (351)(1)472 88 50
PUERTO RICO
Fisher-Rosemount Inc. Guaynabo (787)782-9955, 9956, or 9957
ROMANIA
Fisher-Rosemount Bucharest 40 1230 4149/50/59
RUSSIA
Fisher-Rosemount Moscow 7095 245 86 86
SINGAPORE
Fisher-Rosemount Singapore Pte, Ltd (65)777-8211
SLOVAK REPUBLIC
Fisher-Rosemount scr Bratislava 42 17 761 973
SPAIN
Fisher-Rosemount S.A. Madrid (34)(1)358-91-41
SWEDEN
Fisher-Rosemount AB Karlstad (46)(54)19-0090
SWITZERLAND
Fisher-Rosemount AG Baar (41)(41)768 62 60
THE NETHERLANDS
Fisher-Rosemount BV Rijswijk (31)(70)413-6542
TURKEY
Fisher-Rosemount Proses Kontrol Ltd.Sri Istanbul (90)(216)49 240 42-46
A Worldwide Network of Sales and Service
The Rosemount customer sales and service organization comprises a network of fully equipped support centers strategical­ly located throughout the world. From many of these locations, the Rosemount Group provides support, distribution of fin­ished products, repair facilities, and training for our customers.
Immediate, Reliable Analytical Support
Now there’s a way to quickly get the right answers for your liquid analytical instrumentation questions: the Analytical Customer Support Center.
Our staff of trained professionals are ready to provide the information you need. If you are placing an order, verifying delivery, requesting application information, or just want to con­tact a Rosemount Analytical representative, a toll-free call to 1-800-854-8257 will provide you with the right people, the right answers, right now.
The right people, the right answers, right now.
Page 35
WARRANTY
Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in work­manship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, pH electrodes, membranes, liquid junctions, electrolyte, O-rings, etc. are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of ship­ment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and / or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, parts(s), or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty . This war­ranty shall be in effect for replacement or repaired goods, part(s) and consumables for the remaining portion of the peri­od of the twelve (12) month warranty in the case of goods and part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not oper­ate to condemn such commercial unit when such goods, parts(s) or consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage, directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STAT­ED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Rosemount Analytical Inc.
Uniloc Division
2400 Barranca Parkway
Irvine, CA 92606
The shipping container should be marked:
Return for Repair Model
_______________________________
The returned material should be accompanied by a letter of transmittal which should include the following information (make a copy of the "Return of Materials Request" found on the last page of the Manual and provide the following there­on):
1. Location type of service, and length of time of service of the device.
2. Description of the faulty operation of the device and the circumstances of the failure.
3. Name and telephone number of the person to contact if there are questions about the returned material.
4. Statement as to whether warranty or non-warranty service is requested.
5. Complete shipping instructions for return of the material.
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
Page 36
Rosemount Analytical Inc. Uniloc Division
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 863-1181 http://www.RAuniloc.com
© Rosemount Analytical Inc. 2001
Credit Cards for U.S. Purchases Only.
The right people, the right answers, right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.RAuniloc.com
UNILOC DIVISION
CUSTOMER SUPPORT CENTER
1-800-854-8257
Loading...