Reference Manual
0809-0100-3396
0
Rosemount™ 396/396VP/397/398/398VP
pH/ORP Sensors
Rev. AA
Page 2
hasgkas
Page 3
Essential Instructions
Read this page before proceeding!
merson designs, manufactures and tests its products to meet many national and international stan-
E dards. Because these sensors are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount products. Failure to follow the proper instructions may cause any one of the following situations to occur: loss of life; personal injury; property damage; damage to this sensor; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Emerson representative for
clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance
of the product.
Install your equipment as specified in the Installation Instructions of the appropriate
Reference Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update,
program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts
specified by Emerson. Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Third-party substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
DANGER
Hazardous Area InstallationN
This sensor is not Intrinsically Safe. or Explosion Proof. Installations near flammable liquids or in hazardous area locations must be carefully evaluated by qualified on site safety personnel.
To secure and maintain an intrinsically safe installation, an appropriate transmitter/safety barrier/sensor combi nation must be used. The installation system must be in accordance with the governing approval agency (FM, CSA or BASEEFA/CENELEC) hazardous area classification requirements. Consult your trans­mitter instruc tion manual for details.
Proper installation, operation and servicing of this sensor in a Hazardous Area Instal lation is entirely the responsibility of the user.
CAUTION
Sensor/Process Application Compatibility
The wetted sensor materials may not be compatible with process composition and operating conditions. Application compatibility is entirely the responsibility of the user.
Page 4
Page 5
Reference Manual Table of Contents
00809-0100-3396 February 2019
Contents
Section 1: Description and Specifications
1.1 TUpH Features and Applications ..........................................................................1
1.2 Specifications ......................................................................................................3
1.3 Product Certifications..........................................................................................4
1.4 Ordering Information...........................................................................................6
Section 2: Installation
2.1 First Time Installation ..........................................................................................9
2.2 Unpacking and Inspection .................................................................................10
2.3 Mounting ..........................................................................................................10
2.4 Electrical Installation..........................................................................................17
Section 3: Startup and Calibration
3.1 Sensor Preparation ............................................................................................33
3.2 Calibration Using Buffer Solution or Grab Sample...............................................33
Section 4: Maintenance
4.1 Electrode Cleaning.............................................................................................35
4.2 Automatic Temperature Compensator...............................................................36
Section 5: Troubleshooting ....................................................................................37
Section 6: Accessories
6.1 Accessories........................................................................................................39
EC Declaration of Conformity ...............................................................................41
Intrisicallly Safe Sensor Installation Drawing - FM ......................................47
Table of Contents i
Page 6
Table of Contents Reference Manual
February 2019 00809-0100-3396
ii Table of Contents
Page 7
Reference Manual Description and Specifications
00809-0100-3396 February 2019
Section 1: Description and Specifications
1.1 TUpH Features and Applications
The low maintenance and disposable Rosemount 396, 396VP, 397, 398, and 398VP TUpH sensors
offer long life and high performance in measuring pH or ORP in aqueous solutions in pipelines, open
tanks, or ponds.
These TUpH sensors feature a large area reference junction for maximum resistance to process
coatings, generally found in dirty, high solid applications. In addition, the secondary helical refer-
ence pathway serves as added protection from poisoning ions. The simplified construction,
designed with user convenience in mind, does not require electrolyte replenishment or any
component replacement. All models feature a large glass bulb for increased resistance to the
effects of aging for longer life.
1.1.1 Rosemount 396 and 396VP Sensor Features and Applications
Rosemount 396/396VP sensors are constructed of
polypropylene and stainless steel and are completely
sealed by EP (ethylene propylene) to eliminate process
intrusion.
Sensors are specially designed for improved life in
harsh, dirty applications such as lime slurry waste treat-
ment and paper machine headbox and pigment/dye
applications where large quantities of suspended solids
are present. Installation is easily achieved through a
wide variety of mounting configurations.
1.1.2 Rosemount 397 Sensor Features and Applications
The Rosemount 397 is housed in a highly chemical
resistant polypropylene body and completely sealed with
EP to eliminate process intrusion. The Rosemount 397
body is specifically designed for use with the Quik-Loc Kit
which consists of an adapter and coupler. The PEEK (poly
etheretherke tone) adapter enables the Rosemount 397
sensor to fit into a 1 in. MNPT Dixon coupler for quick and
easy removal without sensor cable twisting. The 316 stain
less steel DIxon coupler is sealed with EP an features
locking arms. The Quik-Loc Kit is not recommended for
use in processes with hazardous, corrosive, or strong
oxidizing chemicals due to a risk of spray and bodily hazards.
1.1.2 Rosemount 398 and 398VP Sensor Features and Applications
The chemical-resistant construction of Tefzel, titanium, and the TUpH reference junction make
the Rosemount 398 and 398VP the ideal sensors for measuring pH in harsh process liquids. Use
these sensors to measure pH in sour water strippers, in pulp bleaching towers that use chlorine
dioxide, and in process streams containing a variety of organic solvents.
Rosemount 398 and 398VP sensors use the highest quality materials to provide superior chem-
ical resistance. The sensors are housed in a titanium tube and features an optional 1 inch MNPT
process connector for insertion, submersion, or flow-through applications. The molded Tefzel
TUpH construction is offered with a choice of seals (Viton, EPDM, or Kalrez).
Description and Specifications 1
Page 8
Description and Specifications Reference Manual
February 2019 00809-0100-3396
Figure 1-1: Cross Section Diagram of the TUpH Reference Technology
All TUpH sensors are designed with a large area reference junction, helical reference pathway, and an AccuGlass pH glass bulb. This sensor technology ensures superior performance while only requiring minimal maintenance.
2 Description and Specifications
Page 9
Reference Manual Description and Specifications
00809-0100-3396 February 2019
1.2 Specifications
Table 1-1: Percent linearity over pH
396, 396VP, 397, 398, 398VP 396, 396VP, 397, 398, 398VP
pH Range GPHT Hemi GPLR Hemi
0-2 pH 94% 93%
2-12 pH 99% 98%
12-13 pH 97% 95%
13-14 pH 92% --
Table 1-2: Rosemount 396, 396VP, 397, 398 and 398VP Specifications
Rosemount 396
396VP
Measured Ranges 0 to 14 pH 0 to 14 pH
Available pH
ACCUGLASS Types
GPHT hemi bulb or
GPLR flat bulb
Rosemount 398
398VP
ORP: -1500 mV to 1500 mV
GPHT hemi bulb or
GPLR flat bulb GPHT hemi bulb
Rosemount 397 Quik-loc Kit
0 to 14 pH
_
_
316 SST,
Wetted Materials
Process Connection
Temperature Range 0-100 °C (32-212 °F) 0-100 °C (32-212 °F) 0-100 °C (32-212 °F)
Pressure Range-
Hemi bulb
Pressure Range-
Flat bulb
Minimum
Conductivity
Preamplifier Options Remote Remote Remote
Weight/Shipping
Weight
Polypropylene, EPDM, glass
None, use 1 in. NPT
process connector,
PN 23166-00 or 23166-01
(sold separately)
100-1136 kPa abs
(0-150 psig)
100-790 kPa abs
(0-100 psig)
75 μS/cm, nominal 75 μS/cm, nominal 75 μS/cm, nominal
0.45 kg/0.9 kg (1 lb/2 lb) 0.45 kg/0.9 kg (1 lb/2 lb) 0.45 kg/0.9 kg (1 lb/2 lb) 0.45 kg/0.9 kg
Titanium, Tefzel, glass,
choice of Kalrez, Viton, or
EPDM (platinum: ORP only)
None, use 1 in. NPT process connector,
PN 23166-00 or 23166-01
(sold separately)
100-1825 kPa abs
(0-250 psig)
100-790 kPa abs
(0-100 psig)
Polypropylene, EP, glass 316 SST, EP, PEEK
None, must use Quik­Lockit which includes 1 in. MNPT process connection
100-790 kPa abs
(0-100 psig)
_ _
1-in. MNPT
_
(1 lb/2 lb)
Description and Specifications 3
Page 10
Description and Specifications Reference Manual
February 2019 00809-0100-3396
1.3 Product Certifications
Please see online certificates for further details.
Rosemount 396/396VP:
IECEx
Sensors without preamp (pH and ORP) – Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Sensors with SMART preamp (pH only) – Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Sensors with standard preamp (396P only) – Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +80 °C) or
Ex ia IIC T5 Ga (-20 °C ≤ Ta ≤ +40 °C)
Per standards IEC60079-0 : 2011, IEC 60079-11 : 2011
ATEX
Sensors without preamp (pH and ORP) – II 1 G Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Sensors with SMART preamp (pH only) – II 1 G Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Sensors with standard preamp (396P only) – II 1 G Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +80 °C) or
II 1 G Ex ia IIC T5 Ga (-20 °C ≤ Ta ≤ +40 °C)
Per standards EN 60079-0: 2012+A11:2013, EN 60079-11:2012
FM
See online FM Certificate of Compliance for applicable sensor options:
Intrinsically Safe for use in Class I, II, and III, Division 1, Groups A, B, C, D, E, F, and G; Temperature
Class T6 Ta = -20 °C to +60 °C
Intrinsically Safe for use in Class I, Zone 0, AEx ia IIC T6 Ta = -20 °C to +60 °C
Nonincendive for use in Class I, Division 2, Groups A, B, C, and D; Temperature Class T6 Ta = -20 °C
to +60 °C
Suitable for use in Class II and III, Division 2, Groups E, F, and G; Temperature Class T6 Ta = -20 °C
to +60 °C Hazardous (Classified) Locations
IS/I,II,III/1/ABCDEFG/T6 Ta = 60 °C - 1400332; Entity; I/0/AEx ia IIC/T6 Ta = 60 °C - 1400332; Entity;
NI/I/2/ABCD/T6 Ta = 60 °C; S/II,III/2/EFG/T6 Ta = 60 °C
Per standards 3600:1998, 3610:2010, 3611:2004, 3810:2005
CSA
See online CSA Certificate of Compliance for applicable sensor options:
Sensors with preamp – Intrinsically Safe:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups
ABCD; Ambient temperature rating -20 °C to +60 °C; Ex ia IIC; T6
Sensors without preamp – Intrinsically Safe and Non-Incendive:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups
ABCD; Ex ia IIC; T6; Ambient temperature rating -20 °C to +60 °C: (Simple Apparatus)
Per standards C22.2 No. 0-10, C22.2 No. 0.4-M2004, C22.2 No. 94-M1991, C22.2 No. 142 –
M1987, C22.2 No 157 – M1992, CAN/CSA E60079-0:07, CAN/CSA E60079- 11:02, UL50 11th
Ed, UL508 17th Ed, UL913 7th Ed,, UL 60079-0: 2005, UL 60079-11: 2002
4 Description and Specifications
Page 11
Reference Manual Description and Specifications
00809-0100-3396 February 2019
osemount 397/398/398VP:
R
IECEx
Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Per standards IEC60079-0 : 2011, IEC 60079-11 : 2011
TEX
A
II 1 G Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Per standards EN 60079-0: 2012+A11:2013, EN 60079-11:2012
FM
Intrinsically Safe for use in Class I, II, and III, Division 1, Groups A, B, C, D, E, F, and G; Temperature
Class T6 Ta = -20 °C to +60 °C
Intrinsically Safe for use in Class I, Zone 0, AEx ia IIC T6 Ta = -20 °C to +60˚C
Nonincendive for use in Class I, Division 2, Groups A, B, C, and D; Temperature Class T6 Ta = -20 °C
to +60 °C
Suitable for use in Class II and III, Division 2, Groups E, F, and G; Temperature Class T6 Ta = -20 °C
to +60 °C Hazardous (Classified) Locations
IS/I,II,III/1/ABCDEFG/T6 Ta = 60 °C - 1400332; Entity; I/0/AEx ia IIC/T6 Ta = 60 °C - 1400332; Entity;
NI/I/2/ABCD/T6 Ta = 60 °C; S/II,III/2/EFG/T6 Ta = 60 °C; Entity Parameters
Per standards 3600:1998, 3610:2010, 3611:2004, 3810:2005
CSA
Intrinsically Safe and Non-Incendive:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2,
Groups ABCD; Ex ia IIC; T6; Ambient temperature rating -20°C to +60°C: (Simple Apparatus)
Per standards C22.2 No. 0-10, C22.2 No. 0.4-M2004, C22.2 No. 94-M1991, C22.2 No. 142 –
M1987, C22.2 No 157 – M1992, CAN/CSA E60079-0:07, CAN/CSA E60079- 11:02, UL50 11th
Ed, UL508 17th Ed, UL913 7th Ed, UL 60079-0: 2005, UL 60079-11: 2002
Description and Specifications 5
Page 12
Description and Specifications Reference Manual
February 2019 00809-0100-3396
1.4 Ordering Information
The Rosemount 396 pH sensor features either a standard hemi bulb or flat glass electrode. The
sensor is housed in a stainless steel body and must be used with a 1 inch MNPT threaded process
onnector (sold separately). The Rosemount 396 is not available with an integral preamplifier and
c
comes standard with a 15 ft integral cable.
Table 1-3: Rosemount 396 ordering information
Model Sensor type
396 pH/ORP Sensor
Transmitter/TC Compatibility
50 3K TC 54 Pt-100
Cable Option
_ No Selection 62 Cable without BNC
Electrode Option
_ No Selection 71 pH - GPHT Flat Glass
Typical Model Number: 396-54-62-71
(1)
(2)
(3)
1. For use with legacy transmitter models 1181, 1050, and 1003.
2. For wiring to Rosemount 1056, 1066, 1057, 56, and 5081 transmitters.
3. No selection will configure the sensor with a standard hemi bulb glass electrode.
The Rosemount 396VP sensor features a Variopol cable connection for use with a mating Variopol interconnecting cable (sold separately).
Table 1-4: Rosemount 396VP ordering information
Model Sensor type
396VP pH/ORP Sensor
Temperature Compensation
50 3K TC 54 Pt-100 55 Pt-100 (for SMART Preamp)
(1)
(2)
Electrode Option
_ No Selection 71 pH - GPHT Flat Glass
(3)
Preamplifier Option
_ No Selection 70 SMART Preamplifier
Typical Model Number: 396VP-55-71-70
1. For use with legacy transmitter models 1181, 1050, 1003.
2. Must be selected with Option 70 for SMART Preamplifier.
3. No selection will configure the sensor with a standard hemi bulb glass electrode.
4. Must be selected with Option 55.
(4)
6 Description and Specifications
Page 13
Reference Manual Description and Specifications
00809-0100-3396 February 2019
he Rosemount 398 sensor is housed in a titanium tube and made with a Tefzel reference junc-
T
tion. Sensors must be used with a 1 inch MNPT process connector (sold separately). These
sensors come with a standard 15 ft. integral cable. A preamplifier must be used if the sensor is
installed more than 15 ft. from the connected transmitter.
able 1-5: Rosemount 398 ordering information
T
Model Sensor type
398 pH/ORP Sensor
Measuring Electrode Type
10 pH - GPHT Glass 12 ORP 13 pH - GPLR Flat Glass
O-ring Material
30 EPDM 31 Viton 32 Kalrez
Transmitter/TC Compatibility
50 3K TC 54 Pt-100
Cable Option
_ No Selection 62 Cable without BNC
Typical Model Number: 398-10-30-54-62
(1)
(2)
1. For use with legacy transmitter model 1181. Not available with option 62. If selected with ORP, sensor will not come with TC.
2. For use with Rosemount transmitter models 1056, 1066, 1057, 56, and 5081. Only available with option 54.
The Rosemount 398VP sensor features a Variopol cable connection for use with a mating Variopol
interconnecting cable (sold separately).
Table 1-6: Rosemount 398VP ordering information
Model Sensor type
398VP pH/ORP Sensor
Measuring Electrode Type
10 pH - GPHT Glass 12 ORP 13 pH - GPLR Flat Glass
O-ring Material
30 EPDM 31 Viton 32 Kalrez
Transmitter/TC Compatibility
50 3K TC 54 Pt-100
Typical Model Number: 398VP-10-30-54
1. For use with legacy transmitter model 1181. If selected with ORP, sensor will not come with TC.
(1)
Description and Specifications 7
Page 14
Description and Specifications Reference Manual
February 2019 00809-0100-3396
The Rosemount 397 pH sensor is housed in a polypropylene body and is designed to be used with
the Quik-Loc Kit (sold separately).
Table 1-7: Rosemount 397 ordering information
Model Sensor type
397 pH/ORP Sensor
Preamplifier/Cable
02 Without Integral Preamplifier, 15 ft (4.6 m) Cable
Measuring Electrode Type
0
1 12 ORP
Transmitter/TC Compatibility
50 3K TC 54 Pt-100
Other Options
_ No Selection 62 Cable without BNC 64 Tefzel Body Material
Typical Model Number: 397-02-10-54-62
H - GPHT Glass
p
(1)
(2)
1. For use with legacy transmitter model 1181.
2. For use with Rosemount transmiter models 1056, 1066, 1057, 56, and 5081.
8 Description and Specifications
Page 15
Reference Manual Installation
00809-0100-3396 February 2019
Section 2: Installation
2.1 First Time Installation
For first time Rosemount 397 installations, using the following guide is recommended:
1. Quik-Loc Mounting (required for all first time installations)
Choose one: PN 23757-00, Quik-Loc Kit: for use in 1 in. tees; insertion depth 1.4 in. (35 mm)
2. Remote Junction Boxes (optional, recommended for sensor to analyzer distances of more than 15 ft)
Choose one: PN 23555-00 includes preamplifier
3. Extension cables (used with remote junction boxes)
Choose one: PN 23646-01, 11 conductor, shielded, prepped
PN 9200273, 11 conductor, shielded, unprepped
For first time Rosemount 396/396VP/398/398VP installations, using the following guide is recommended:
1. Process Connector Accessories (required for all first time installations with 1-inch process connection threads)
Choose one: PN 23166-00, 316 SST, 1 in. x 1 in. NPT process connector, with EPDM O-ring
PN 23166-01, Titanium, 1 in. x 1 in. NPT process connector, with EPDM O-ring
PN 9510066, Nylon, 1 in. x 1 in. NPT process connector (submersion only)
Choose one (optional process connector o-rings)
PN 9550220, Kalrez o-ring, 2-214
PN 9550099, Viton o-ring, 2-214
2. Variopol Cable (required for all first time installations) of Models 396VP and 398VP
Choose one: PN 24281-00, 15 ft cable with mating VP connector
PN 24281-06, 10 ft cable with mating VP connector
3. Mounting Accessories (optional)
Choose one: PN 915240-03 PVC flow through tee, 3/4 in. NPT process connection
PN 915240-04 PVC flow through tee, 1 in. NPT process connection
PN 915240-05 PVC flow through tee, 1 1/2 in. NPT process connection PN 11275-01 Sensor handrail mounting assembly PN 2002011 1-1/2 in. CPVC Tee with 1 in. FNPT connection PN 24091-00 Low Flow Cell
4. Remote Junction Boxes (optional, recommended for sensor to analyzer distances of more than 15 ft)
Choose one: PN 23555-00 includes preamplifier
5. Extension cables (used with remote junction boxes)
Choose one: PN 23646-01, 11 conductor, shielded, prepped
PN 9200273, 11 conductor, shielded, unprepped
Installation 9
Page 16
Installation Reference Manual
February 2019 00809-0100-3396
2.2 Unpacking and Inspection
Inspect the outside of the carton for any damage. If damage is detected, contact the carrier immediately. Inspect the instrument and hardware. Make sure all the items in the packing list are
resent and in good condition. Notify the factory if any part is missing. If the sensor appears to be
p in satisfactory condition, proceed to Section 2.2, Mounting.
Note: Save the original packing cartons and materials as most carriers require proof of damage due to mishandling, etc. Also, if it is necessary to return the instrument to the factory, you must pack the instrument in the same manner as it was received. Refer to Section 6 for return instructions. If the sensor is to be stored, the vinyl boot should be filled with pH buffer solution and replaced on sensor tip until ready to use.
WARNING
Glass electrode must be wetted at all times (in storage and in line) to maximize sensor life.
2.3 Mounting
Each sensor has been designed to be located in industrial process environments. Temperature and pressure limitations must not be exceeded at any time. A caution or warning label regarding this matter is attached to each sensor. For insertion, transfer the label as shown on label instructions. See Figure 2-2. For submersion applications, first note limits then remove and discard label.
Note: Before mounting the sensor, shake down the sensor to remove any air bubbles that may be present at the tip of the pH glass bulb. In most cases, the pH sensor can simply be installed as shipped and readings with an accuracy of ±0.6 pH may be obtained. To obtain greater accuracy or to verify proper operation, the sensor must be calibrated as a loop with its compatible analyzer or transmitter.
2.3.1 Flow Through and Insertion Mounting for Rosemount 396, 396VP, 398, and 398VP Sensors.
These sensors can be used with a 1 inch MNPT process connector at the front of the sensor for mounting into a 1-1/2 inch tee or the process. See Figure 2-1 for instal lation configurations.
2.3.2 Submersion Mounting for Rosemount 396, 396VP, 398, and 398VP Sensors.
These sensors also have a 1 inch MNPT process connector available for use on the back of the sensor. Tapered pipe threads in plastic fittings tend to loosen after installation. It is therefore recom mended that Teflon tape be used on the threads and that the tightness of the connection be checked frequently to assure that no loosening has occurred. To prevent rain water or condensa­tion from running into the sensor, a weatherproof junction box is recommended (see Figure 2-4). The sensor cable must be run through a protective conduit for isolation from electrical interfer­ence or physical abuse from the process. The sensor should be installed within 80° of vertical, with the electrode facing down. The sensor’s cable should not be run with power or control wiring.
10 Installation
Page 17
Reference Manual Installation
00809-0100-3396 February 2019
2.3.2 Quik-Loc Mounting for Rosemount 397
The Quik- Loc mounting is used with the Rosemount 397 sensor only.
WARNING
Once the Quik-Loc unit is installed the operator should wait for the process to cool to a safe temperature, use the pressure drain valve to relieve all process pressure and observe the pressure on the pressure gauge for proper removal of the sensor without spray or bodily injury. The Quik-Loc kit used with the Rosemount 397 sensor is not recommended for use with hazardous, corrosive, or strong oxidizing chemicals due to a risk of spray or bodily injury.
WARNING
It is recommended that a thermometer, drain valve to relieve pressure and pressure gauge be inserted near the Quik-Loc assembly (see Figure 2-5).
Wrap the pipe threads of the Twin-Kam Kamloc coupler with Teflon tape before placing it into the process pipe. The coupler can be connected to any 1 in. process connection and must be mounted within 80° of vertical, with the electrode facing down. Once the coupler is in place, the adapter should be positioned onto the back end of the sensor. (See Figure 2-5) Remove the parafilm wrap ping from the two 0-rings on the sensor, grease the o-rings with the lube provided, and feed the sensor cable through the adapter. Once the adapter is slipped over the sensor’s back end, the retaining ring (which is included with every Rosemount 397 sensor) should be installed on the black, grooved, back end of the sensor body. The retaining ring secures the sensor into the adapter. The adapter/sensor assembly is now ready to be inserted into the coupler. With both arms of the coupler in the released position, insert the adapter/sensor assembly into the coupler.
Note: The adapter can not be inserted completely or properly unless both arms are in the fully released position (see Figure 2-5).
Once the adapter has been properly placed in the coupler, both arms should be positioned in the locked position. A Sur- Loc™ spring arm has been provided on one arm so that the arm cannot be opened until the spring is released.
Note: The sensor may obstruct flow through smaller pipes.
Once the arms of the coupler have been locked in position, use tamper-evident safety wire on the metal rings of the Twin-Kam arms to prevent unauthorized and/or untrained personnel from using the Quik-Loc unit.
WARNING
It is the responsibility of each company using the Quik-Loc Kit/ 397 TUpH Sensor to train personnel of the injury risks associated with using a quick-release coupler that is placed in a hot or pressurized process.
The Quik-Loc unit should be used only within the pressure and temperature limits stated for the Rosemount 397 sensor in Section 1.2.
Installation 11
Page 18
Installation Reference Manual
February 2019 00809-0100-3396
igure 2-1: Recommended Flow Through and Insertion Installation for Rosemount 396,
F 396VP, 398, and 398VP Sensors
1-1/2 inch Pipe Tee (PN 2002011) with 1 inch threaded connections.
Figure 2-2: Recommended Flow-Through and Insertion Installation For Rosemount 397
12 Calibration and Maintenance
Page 19
Reference Manual Installation
00809-0100-3396 February 2019
Figure 2-3: Dimensional Drawing For Rosemount 396 and 398 Sensors
Calibration and Maintenance 13
Page 20
Installation Reference Manual
February 2019 00809-0100-3396
Figure 2-4: Rosemount 396VP and 398VP Sensor Dimensional Drawings (Shown with Process Connector)
Metal Process Connector PN 23166-xx
(xx = 00 for 316 SST and xx = 01 for titanium)
can be used for insertion or submersion mounting of Rosemount 396VP/398VP
sensors in 1-inch fittings.
14 Installation
The metal process connector gives the sensor various insertions depths, depending on where the user locates the compression fitting. Also thethreads can be switched to face the cable end of the sensor for connection to submersion pipes.
Page 21
Reference Manual Installation
00809-0100-3396 February 2019
Figure 2-5: Submersion Installations for Rosemount 396, 396VP, 398, and 398VP Sensors
Installation 15
Page 22
Installation Reference Manual
February 2019 00809-0100-3396
igure 2-6: Dimensional Drawings For Rosemount 397 in the Quik-Loc Unit
F
16 Installation
Page 23
Reference Manual Installation
00809-0100-3396 February 2019
2.4 Electrical Installation
Rosemount 396, 396VP, 397, 398, and 398VP sensors are for use with a remote preamplifier. Each sensor comes with either a special 15 ft low noise coax cable or a Variopol (VP) connector, which is used with a mating Variopol cable. The cable should be handled carefully, and kept dry and free of corrosive chemicals at all times. Extreme care should be used to prevent the cable from being twisted, damaged or scraped by rough, sharp edges or surfaces.
For additional wiring information on this product, please refer to Wiring Diagrams.
DANGER
DO NOT CONNECT SENSOR CABLE TO POWER LINES. SERIOUS INJURY MAY RESULT.
Note: Remove electrical tape or shrink sleeve from gray reference wire before connecting wire to
terminal.
Figure 2-7: Wiring for Rosemount 396/397/398 to Rosemount 5081-P
Installation 17
Page 24
Installation Reference Manual
February 2019 00809-0100-3396
Figure 2-8: Wiring for Rosemount 396-54, 397-54, and 398-54 (Pt-100 RTD). For Use With Junction Box (PN 23555-00) and Remote Preamplifier.
18 Installation
Page 25
Reference Manual Installation
00809-0100-3396 February 2019
igure 2-9: Wiring for Rosemount 396, 397, and 398. For Use With Model 1181 pH.
F
Installation 19
Page 26
Installation Reference Manual
February 2019 00809-0100-3396
Figure 2-10: Wiring for Rosemount 396-54 , 397-54, and 398-54. For Use With Rosemount 54,
4e, 81, 3081, 4081, and 5081.
5
20 Installation
Page 27
Reference Manual Installation
00809-0100-3396 February 2019
igure 2-11: Wiring for Rosemount 396VP and 398VP to Rosemount 3081, 4081, and 5081
F through a Remote Junction Box.
Installation 21
Page 28
Installation Reference Manual
February 2019 00809-0100-3396
Figure 2-12: Rosemount 396-54-62-xx Wiring to Rosemount 1056/1057/56 Transmitters
Figure 2-13: Rosemount 396-54-62-xx Wiring to Rosemount 1066 Transmitter
22 Installation
Page 29
Reference Manual Installation
00809-0100-3396 February 2019
Figure 2-14: Rosemount 396-54-62-xx Wiring to Rosemount 5081 Transmitter
Figure 2-15: Rosemount 396VP Wiring for Rosemount 1056/56 Transmitters
Installation 23
Page 30
Installation Reference Manual
February 2019 00809-0100-3396
Figure 2-16: Rosemount 396VP Wiring for Rosemount 1066 Transmitter
Figure 2-17: Rosemount 396VP Wiring for Rosemount 1057 Transmitter
24 Installation
Page 31
Reference Manual Installation
00809-0100-3396 February 2019
igure 2-18: Rosemount 396VP Wiring for Rosemount 5081 Transmitter
F
Figure 2-19: Rosemount 396VP-70 Wiring for Rosemount 1066 Transmitter
Installation 25
Page 32
Installation Reference Manual
February 2019 00809-0100-3396
Figure 2-20: Rosemount 396VP-70 Wiring for Rosemount 1056/1057/56 Transmitters
Figure 2-21: Rosemount 397-02-10-54-xx Wiring to Rosemount 1056/1057/56 Transmitter
26 Installation
Page 33
Reference Manual Installation
00809-0100-3396 February 2019
Figure 2-22: Rosemount 397-02-10-54-xx Wiring to Rosemount 1066 Transmitter
Figure 2-23: Rosemount 397-02-10-54-xx Wiring to Rosemount 5081 Transmitter
Installation 27
Page 34
Installation Reference Manual
February 2019 00809-0100-3396
Figure 2-24: Rosemount 398-xx-xx-xx-62 Wiring to Rosemount 1056/1057/56 Transmitters
Figure 2-25: Rosemount 398-xx-xx-xx-62 Wiring to Rosemount 1066 Transmitter
28 Installation
Page 35
Reference Manual Installation
00809-0100-3396 February 2019
Figure 2-26: Rosemount 398-xx-xx-xx-62 Wiring to Rosemount 5081 Transmitter
Figure 2-27: Rosemount 398VP Wiring to Rosemount 1056/56 Transmitters
Installation 29
Page 36
Installation Reference Manual
February 2019 00809-0100-3396
Figure 2-28: Rosemount 398VP Wiring to Rosemount 1057 Transmitter
Figure 2-29: Rosemount 398VP Wiring to Rosemount 5081 Transmitter
30 Installation
Page 37
Reference Manual Installation
00809-0100-3396 February 2019
Figure 2-30: Rosemount 398VP Wiring to Rosemount 1066 Transmitter
Installation 31
Page 38
Installation Reference Manual
February 2019 00809-0100-3396
32 Installation
Page 39
Reference Manual Startup and Calibration
00809-0100-3396 February 2019
Section 3: Startup and Calibration
3.1 Sensor Preparation
Shake down the sensor to remove any air bubbles that may be present at the tip of the pH glass bulb. In most cases, the pH sensor can simply be installed as shipped and readings with an accuracy of ±0.6 pH may be obtained. To obtain greater accuracy or to verify proper operation, the sensor must be calibrated as a loop with its compatible analyzer or transmitter.
3.2 Calibration Using Buffer Solution or Grab Sample
The pH sensor-transmitter loop may be calibrated by submersing the sensor in a buffer solution (standard solutions of known pH values) or in a process grab sample whose pH value maybe checked by a calibrated laboratory or portable pH meter.
Please refer to the transmitter Reference Manual for proper calibration procedures.
Startup and Calibration 33
Page 40
Startup and Calibration Reference Manual
February 2019 00809-0100-3396
34 Startup and Calibration
Page 41
Reference Manual Maintenance
00809-0100-3396 February 2019
Section 4: Maintenance
Rosemount 396, 396VP, 397, 398, and 398VP sensors are disposable type sensors and therefore require minimum maintenance. Every sensor should be kept clean and free of debris and sediment
t all times. The frequency of cleaning by wiping or brushing with a soft cloth or brush is
a determined by the nature of the solution being measured. The sensor should be removed from
the process periodically and checked in buffer solutions.
DANGER
SENSOR IS IN PRESSURIZED SYSTEM! May cause spray and bodily injury. Before removing sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level.
If the sensor will not calibrate, refer to your transmitter Reference Manual for proper test procedures. If it is determined that the sensor has failed, it should be discarded and replaced.
4.1 Electrode Cleaning
If the electrode is coated or dirty, clean as follows:
1. Remove the sensor from process.
2. Wipe the glass bulb with a soft, clean, lint free cloth or tissue. If this does not remove the dirt or coating, go to Step 3. (Detergents clean oil and grease; acids remove scale.)
3. Wash the glass bulb in a strong detergent solution and rinse it in clean water. If this does not clean the glass bulb, go to Step 4.
CAUTION
The solution used during the following check is an acid and should be handled with care. Follow the direc­tions of the acid manufacturer. Wear the proper protective equipment. Do not let the solution come in contact with skin or clothing. If contact with skin is made, immediately rinse with clean water.
4. Wash the glass bulb in a dilute 5% hydro chloric acid solution and rinse with clean water. Soaking the sensor overnight in the acid solution can improve cleaning action.
NOTICE
Erroneous pH results may occur immediately after acid soak, due to reference junction potential build-up. Replace the sensor if cleaning does not restore sensor operation.
Maintenance 35
Page 42
Maintenance Reference Manual
February 2019 00809-0100-3396
4.2 Automatic Temperature Compensator
The temperature compensator element is temp erature sensitive and can be checked with an ohmmeter. Resistance increases with temperature.
The 3K element will read 3000 ohms ±1% at 25°C (77°F) and a Pt-100 will read 110 ohms. Resist­ance varies with temperature for a 3K and Pt-100 element and can be determined according to Table 4- 2 or the following formula:
RT=R0[l+R1(T-20)]
Where RT= Resistance
T = Temperature in °C
Refer to Table 4-1 for R0and R1values
Table 4-1: R0and R1Values for Temperature Compensation Elements
Temperature Element R
3K 2934 0.0045
PT-100 107.7 0.00385
0
R
1
Table 4-2: Temperature vs Resistance of Automatic Temperature Compensator Elements
Temperature °C Resistance (Ohms) ±1
3K PT-100
0 2670 100.0
10 2802 103.8
20 2934 107.7
25 3000 109.6
30 3066 111.5
40 3198 115.4
50 3330 119.2
60 3462 123.1
70 3594 126.9
80 3726 130.8
90 3858 134.6
100 3990 138.5
36 Maintenance
Page 43
Reference Manual Troubleshooting
00809-0100-3396 February 2019
Section 5: Troubleshooting
able 5-1: Troubleshooting
T
Trouble Probable Cause Remedy
Meter reads off scale. (Display reads overrange.)
Display reads between 3 and 6 pH regardless of actual pH of solution or sample.
Meter or display indication swings or jumps widely in AUTO T.C. Mode.
Span between buffers extremely short in AUTO T.C. Mode.
Sluggish or slow meter indication for real changes in pH level.
T.C. element shorted. Check T.C. element as instructed in Section 4.2
and replace sensor if defective.
Sensor not in process or sample stream is low. Make sure sensor is in process with sufficient
sample stream (refer to Section 2.0 for installation details).
Open glass electrode. Replace sensor.
Reference element open - no contact. Replace sensor.
Electrode cracked. Replace sensor.
T.C. element shorted. Check T.C. element as instructed in Section 4.2
and replace sensor if defective.
T.C. element open. Check T.C. element as instructed in Section 4.2
and replace sensor if defective.
Electrode coated. Clean sensor as instructed in Sections 4.1
Replace sensor if cracked.
Electrode at end of life. Replace sensor.
Transmitter cannot be standardized.
Transmitter short spans between two different buffer values.
Electrode coated or cracked. Clean sensor as instructed in Sections 4.1
Replace sensor if cracked.
Electrode at end of life, due to old glass or high temperature exposure.
Coated glass. Clean sensor as instructed in Sections 4.1
Replace sensor.
Replace sensor if cracked.
Note: For any repair or warranty inquiries please contact our Customer Care group.
Troubleshooting 37
Page 44
Troubleshooting Reference Manual
February 2019 00809-0100-3396
38 Troubleshooting
Page 45
Reference Manual Accessories
00809-0100-3396 February 2019
Section 6: Accessories
6.1 Accessories
Table 6-1: Additional Accessories for Rosemount 396/396VP/398/398VP Sensors
Part Number Description
33046-00 Ferrule 1 in., split 316 SS
33211-00 Adapter retrofit for PN 915240-04
9310100 Ferrule, 1 in. Teflon
9310096 Nut, Swage, 1 in. stainless steel
Table 6-2: Additional Accessories for Rosemount 397 Sensor
Part Number Description
23753-00 PEEK replacement adapter for Quik-Loc kit (PN 23757-01)
23753-01 PEEK replacement adapter for Quik-Loc kit (PN 23757-00)
9160441 1 in. MNPT 316 stainless steel coupler
9160447 1 in. EP Gasket for coupler
2002011 1-1/2 in. CPVC Tee with 1 in. FNPT Connection
Table 6-3: Other Accessories
Part Number Description
9210012 Buffer solution, 4.01 pH, 16 oz.
9210013 Buffer solution, 6.86 pH, 16 oz.
9210014 Buffer solution, 9.18 pH, 16 oz.
2001492 Stainless Steel Tag, Specify Marking
Accessories 39
Page 46
Accessories Reference Manual
February 2019 00809-0100-3396
40 Accessories
Page 47
Reference Manual EC Declaration of Conformity
00809-0100-3396 February 2019
EC Declaration of Conformity
ote: Please see websitefor most recent Declaration.
N
EC Declaration of Conformity 41
Page 48
EC Declaration of Conformity Reference Manual
February 2019 00809-0100-3396
42 EC Declaration of Conformity
Page 49
Reference Manual EC Declaration of Conformity
00809-0100-3396 February 2019
EC Declaration of Conformity 43
Page 50
China RoHS Tables Reference Manual
February 2019 00809-0100-3396
44 China RoHS Table
Page 51
Reference Manual China RoHS Tables
00809-0100-3396 February 2019
China RoHS Tables 45
Page 52
China RoHS Tables Reference Manual
February 2019 00809-0100-3396
China RoHS Tables 46
Page 53
ANY FM APPROVED
ASSOCIATED APPARATUS
HAVING ENTITY PARAMETERS
NO
N-
H
AZARDOUS
(UNCLASSIFIED)
AREA
CLASS I, II, III, DIVISIO N 1, GROUPS A-G
T6 Ta = 60C
SENSOR
ENTITY PARAMETERS
Ui = 13.1U, Ii = 358mA
Pi = 698 m W
Ci = 0.967 F, Li = 0.1mH
1. NO REVISION TO THIS DRAW ING IS PERMITTED W ITHO UT
FM APPROVAL.
2. U
m ax
> U
t
; I
m ax
> I
t
; (C
i
OF ALL LOO PS + C CABLE) < C
a
;
(L
i
OF ALL LOOPS + L CABLE) < L
a
, P
m ax
OR P
i
> P
0.
3. SINGLE M ULTI-CHANNEL
IS BARRIER OR APPARATUS M UST BE FM APPROVED,
4. SINGLE M ULTI-CHANNEL
IS BARRIER OR APPARATUS M ANUFACTURE’S
CONTROL DRAW INGS M UST BE FOLLO W ED W HEN INSTALLING THE SYSTEM .
IS BARRIER OR EQUIPM ENT M AY BE INSTALLED W ITHIN THE HAZARDOUS
(CLASSIFIED) LOCATIO N FOR W HICH IT IS APPROVED
.
5. INSTALLATIO N M UST BE IN ACCORDANCE W ITH ARTICL
E 500 OF THE NEC
(ANSI/NFPA 70) AND ANSI/ISA RP 12.6.
W ARNING: SUBSTITUTION OF COM PONENTS M AY IM PA
IR INTRINSIC SAFETY.
6. pH & AM PEROM ETRIC SENSORS W ITHOUT PREAMPS ARE SIM PLE APPARATUS.
ZON E
ZONE 0
7. CONTROL EQUIPM ENT CONNECTED TO THE ASSOCIATED APPARATUS M UST
NOT USE OR GENERATE M ORE THAN 250V.
RESISTANCE BETW EEN INTRINSICALLY SAFE GROUND AND EARTH GROUND
M UST BE LESS THAN O R EQUAL TO 1 O HM .
ANY
FM APPROVED TRANSMITTER FOR
DIVISION 1 W ITH INTRINSICALLY SAFE
OUTPUT PARAMETERS. THIS
FM
APPROVED DEVICE M UST BE INSTALLED
PER ITS INSTALLATION DRAWING.
FM
APPROVED EQUIPM ENT (M AY BE
M ULTIPLE DEVICES, NUMBER IS LIM ITED
BY REQUIREMENTS TO M EET ALL OTHER
IS REQUIREMENTS FOR THE NETWORK)
W ITH EQUIVALENT HAZARDOUS AREA
APPROVALS.
H
AZARDOUS
(CLASSI
F
IED)
AREA
INTRISICALLY SA FE
9. pH/ORP SENSOR M O DELS THAT M AY CONTAIN THE PREAM PLIFIER:
3900/VP
3500/VP
3300HT/VP
3400HT/VP
396/VP
396R/VP
396P/VP
398R/VP
399/VP
389/VP
385/385+
10. W ARNING: TO PREVENT IGNITIO N O F FLAM M ABLE OR COM BUSTIBLE
ATMOSPHERES, DISCONNECT POW ER BEFORE SERVIC
ING.
11. THE ENTITY CONCEPT ALLOW S INTECONNECTION OF INTRINSICALLY SAFE
APPARATUS W ITH ASSOCIATED APPARATUS W HEN THE FOLLOW ING IS TRUE:
U
i
> U
o;
I
i
> I
o;
P
i
> P
o;
C
o
> C
i
+ C CABLE; L
o
> L
i
+ L CABLE.
12. COPY REVISIONS TO 1400332 TO pH/ORP SHIPPING M ANUALS.
13 Ci INCLUDES THE CAPACITANCE OF 500 FEET OF SENSOR CABLE.
13
DW G NO.
1400332
8.
Reference Manual FM Installation
00809-0100-3396 February 2019
Intrisicallly Safe Sensor Installation Drawing - FM
FM Installation 47
Page 54
00809-0100-3396
Rev. AA
February 2019
www.Emerson.com/RosemountLiquidAnalysis
Youtube.com/user/Rosemount
Twitter.com/Rosemount_News facebook.com/Rosemount
Emerson
8200 Market Blvd. Chanhassen, MN 55317, USA Tel +1 800 999 9307 Fax +1 952 949 7001
Liquid.CSC@Emerson.com
©2019 Emerson. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.
The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.
Loading...