Rosemount 3900, 3900VP Product Data Sheet

Rosemount 3900, 3900VP Product Data Sheet

Product Data Sheet

00813-0100-3900, Rev AB

October 2019

Rosemount3900/3900VP

General Purpose pH/ORP Sensors

High performance for your process pH/ORP needs

The Rosemount 3900 and 3900VP pH/ORP combination sensors are used for reliable pH or oxidation/ reduction potential measurements of aqueous solutions in pipelines, open tanks, or ponds. Its robust sensor design allows the Rosemount 3900 and 3900VP sensors to be used in a wide range of applications.

Rosemount 3900/3900VP

October 2019

Overview

A robust sensor design

Extended sensor life and protection against poisoning ions

 

using a double junction reference.

Enhanced performance and increased life with cracking

 

resistant glass.

Maximum chemical resistance provided by a rugged

 

polyphenylene sulfide body.

Operates in sub-zero temperatures down to 14 °F (-10 °C).

Built-ins solution ground for advanced diagnostics.

Versatile installation options

The sensor body features a one-piece construction with both front facing and rear facing process threads.

¾-in. and 1-in. MNPT process connections to meet a variety of application installation requirements.

Variopol (VP8) cable connection option, for quick cable-to-sensor release, eliminates cable twisting.

SMART preamplifier

Automatic recognition of pH sensors by Rosemount transmitters: 1066, 1057, 1056, and 56.

pH calibration data is stored, which allows sensors to be calibrated in advance for "plug and play" installations in the field.

Ordering information

The Rosemount 3900 and 3900VP General Purpose pH/ORP sensors feature a chemically resistant Ryton® plastic body, along with a built-in solution ground for advanced diagnostics and a Pt-100 resistance temperature device (RTD) for temperature compensation. These sensors are available with either an integral cable connection (Rosemount 3900) or Variopol (VP8) connector (Rosemount 3900VP). Variopol cables are sold separately (see Accessories).

Contents

 

Overview...........................................................................................................................................................................................

2

Ordering information........................................................................................................................................................................

2

Specifications....................................................................................................................................................................................

4

Dimensional and installation drawings..............................................................................................................................................

5

Accessories......................................................................................................................................................................................

16

2

Emerson.com/Rosemount

October 2019

Rosemount 3900/3900VP

Table 1: Rosemount 3900 Ordering Information

 

 

Option

Description

 

 

3900

General Purpose pH/ORP Sensor

 

 

Preamplifier option

 

 

 

01

SMART preamplifier, 32-ft. (10 m) cable (standard preamplifier if used with ORP)

 

 

02

No preamplifier, 15-ft. (5 m) cable

 

 

04

SMART preamplifier, 50-ft. (15 m) cable (standard preamplifier if used with ORP)

 

 

05

SMART preamplifier, 66-ft. (20 m) cable (standard preamplifier if used with ORP)

 

 

06

SMART preamplifier, 100-ft. (30 m) cable (standard preamplifier if used with ORP)

 

 

07

No preamplifier, 4-ft. (1 m) cable

 

 

08

No preamplifier, 10-ft. (3 m) cable

 

 

Measuring code

 

 

 

10

General purpose low resistivity (GPLR) pH glass (0-14 pH)

 

 

12

Platinum ORP

 

 

Calibration certificates - optional level

 

 

 

CC

Certificate of Calibration (no test data given)

 

 

LC

Loop Calibration Certificate (sensor and transmitter calibrated together, with test

 

data)

 

 

EC

Electronic Calibration Certificate (sensor calibrated against factory instrument, with

 

test data)

 

 

Note

The Rosemount 3900 pH/ORP sensor is housed in a Ryton plastic body with built-in solution ground for advanced diagnostics. Pt-100 is offered for a temperature compensator. The sensor is available with an integral cable. You can mount the sensor using a process connector, both ordered separately. You can also order junction box kits with preamplifiers separately if the transmitter does not have an integral preamplifier with 15 ft. (5 m) of the sensor.

Table 2: Rosemount 3900VP Ordering Information

Option

Description

 

 

3900VP

General Purpose pH/ORP Sensor with Variopol connector

 

 

Preamplifier option

 

 

 

01

SMART preamplifier (standard preamplifier if used with ORP)

 

 

02

No preamplifier

 

 

Measuring code

 

 

 

10

General purposel low resistivity (GPLR) pH glass (0-14 pH)

 

 

12

Platinum ORP

 

 

Calibration certificates - optional level

 

 

 

CC

Certificate of Calibration (no test data given)

 

 

LC

Loop Calibration Certificate (sensor and transmitter calibrated together, with test

 

data)

 

 

Rosemount 3900/3900VP

3

Rosemount 3900/3900VP

 

October 2019

Table 2: Rosemount 3900VP Ordering Information (continued)

 

 

 

Option

Description

 

 

 

EC

Electronic Calibration Certificate (sensor calibrated against factory instrument, with

 

test data)

 

 

 

Specifications

 

 

Table 3: Percent Linearity over pH Range

 

 

 

 

 

pH range

 

Linearity

 

 

 

0-7

 

97%

 

 

 

1-7

 

98%

 

 

 

4-7

 

98%

 

 

 

7-10

 

99%

 

 

 

7-12

 

97%

 

 

 

7-13

 

96%

 

 

 

7-14

 

95%

 

 

Table 4: Rosemount 3900/3900VP pH/ORP Sensor Specifications

 

 

 

Measurement range

 

 

 

 

pH

0 to 14

 

 

ORP

-1500 to +1500 mV

 

 

 

Operating temperature

 

 

 

14 to 212 °F (-10 to 100 °C), automatic temperature compensation from 14 to 212 °F (-10 to 100 °C)

 

 

 

Maximum pressure

 

 

 

 

 

100 psig (790 kPa [abs]) at 212 °F (100 °C)

 

 

CRN rating: 60 psig up to 212 °F (100 °C)

 

 

 

 

 

Materials of construction

 

 

 

 

Sensor body

Ryton®: polyphenylene sulfide (PPS)

O-ring

EPDM

 

 

pH electrode

Glass

 

 

ORP electrode

Glass, platinum

 

 

Solution ground

Stainless steel

 

 

Reference junction

PFTE (Teflon®)

Conductivity

 

 

 

Responds to changes in pH at a minimum conductivity of 0.1 µS/cm when used with the low flow panel. The sample flow rate

must be controlled to 2 gallons per hour (7.61 liters per hour).

 

 

 

Process connection

 

 

 

 

Front facing

¾-in. and 1-in. MNPT

 

 

 

4

Emerson.com/Rosemount

October 2019

Rosemount 3900/3900VP

Table 4: Rosemount 3900/3900VP pH/ORP Sensor Specifications (continued)

 

 

Rear facing

1-in. MNPT

 

 

Temperature compensation

 

 

 

Pt-100

 

 

 

Cable

 

 

 

3900

Various cable lengths (see Ordering information) with integral preamplifier; 15-ft.

 

(4.6 m) and less without preamplifier

 

 

3900VP

Use 24281-XX, 2.5-ft. (0.8 m) to 100 ft. (30.5 m). See Accessories.

 

 

Weight/shipping weight

 

 

 

1 lb./2 lb. (0.45 kg/0.9 kg)

 

 

 

Dimensional and installation drawings

Figure 1: Proper Sensor Installation Orientation

Install sensor within 80 degrees of vertical.

Rosemount 3900/3900VP

5

Rosemount 3900/3900VP

October 2019

Figure 2: Rosemount 3900 with Integral Cable Connection

A.1-in. MNPT

B.Wrench flats, 1.30 in. across

C.1-in. MNPT

D.¾-in. MNPT

6

Emerson.com/Rosemount

Loading...