Rosemount 3107, 3108 Quick Start Guide

00825-0400-4840, Rev AA
Rosemount Submersion Shield for the Rosemount 3107 and 3108
Quick Start Guide
May 2012
Quick Start Guide
A
B
C
D
E
This guide provides instructions for fitting a Rosemount Submersion Shield (part number 03107-7003-0001) which is an optional accessory for the Rosemount 3107 Level and 3108 Flow transmitters.
It does not provide installation instructions for the Rosemount 3107 Level and 3108 Flow transmitters (or other optional accessories). Refer to the 3107/3108 reference manual (document no. 00809-0100-4840) for detailed instructions.
Product data sheets, reference manuals, and quick-start guides are available electronically on www.emersonprocess.com.
Figure 1. Example application with a submersion shield fitted
May 2012
A. Rosemount 3107 Level or 3108 Flow Transmitter B. Rosemount Submersion Shield (part number 03107-7003-0001) C. Rosemount Head Verification Device (part number 03107-7003-0005) D. Conduit (not supplied) E. Rosemount Conduit Adaptor Boss,
1-in. BSP female to M20 x 1.5 female (part number 03107-7003-0003) or 1-in. NPT female to
1
/2-in. NPT female (part number 03107-7003-0004)
2
May 2012
D
C
B
A
Quick Start Guide

Rosemount 3107/3108 Submersion Shield

The Rosemount Submersion Shield (03107-7003-0001) is recommended for installations, e.g. wet wells, where a 3107/3108 transmitter may be at risk of being flooded.
It is easily fitted over the transducer cup, but requires the transmitter to be kept firmly vertical so that flood water causes an air pocket and keeps the transmitter face clear. The transmitter can then function normally when the flood water subsides.
Figure 1 on page 2 shows an example application with the submersion shield
already fitted to a Rosemount 3107/3108 transmitter, which is then mounted on a Rosemount Head Verification Device.
Figure 2. Features of the Rosemount Submersion Shield
A. PVC Submersion Shield (Medium Grey) B. Silicone O-ring (Red) C. Fluoroelastomer (FKM) O-ring (Dark Grey) D. Foam (PVC/NITRILE)
Failure to follow these installation guidelines could result in death or serious injury
Make sure only qualified personnel perform the installationUse the equipment only as specified in this manual. Failure to do so may impair the protection
Electrical shock could cause death or serious injury
Make sure the main power to the transmitter is off and the lines to any other external power
provided by the equipment
source are disconnected or not powered whilst installing the submersion shield
3