Rosemount 3051T Data sheet

Product Data Sheet
00813-0100-4001, Rev TC
Rosemount™ 3051 Pressure Transmitter
March 2018
With the Rosemount 3051 Pressure Transmitter, you’ll gain more control over your plant. You’ll be able to reduce product variation and complexity, as well as your total cost of ownership by leveraging one device across a number of pressure, level, and flow applications. You’ll have access to information you can use to diagnose, correct, and even prevent issues. And with unparalleled reliability and experience, the Rosemount 3051 is the industry standard that will help increase efficiency and safety so your operations can remain globally competitive.
Rosemount 3051

Setting the standard for pressure measurement

Proven best-in-class performance, reliability and safety

More than seven million installed
Reference accuracy 0.04 percent of span
Installed total performance of 0.14 percent of span
10-year stability of 0.2 percent of URL
SIL2/3 certified (IEC 61508)

Maximize installation and application flexibility with the coplanar platform

Improve reliability and performance with integrated DP
Flowmeters, DP Level solutions and integral manifolds
Easy installation with all solutions fully assembled, leak-tested
and calibrated
Meet your application needs with an unsurpassed offering
March 2018

Advanced functionality

Power advisory diagnostics
Detect on-scale failures caused by electrical loop issues before they impact your process
operation
This capability is safety certified for your most critical applications
Local operator interface (LOI)
Straightforward menus and built-in configuration buttons allow you commission the device in
less than a minute
Configure in hazardous-area locations without removing the transmitter cover using external
buttons
Contents
Rosemount 3051C Coplanar Pressure Transmitter . . . 4
Rosemount 3051T In-Line Pressure Transmitter . . . . .11
Rosemount 3051L Level Transmitter . . . . . . . . . . . . . . 37
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Rosemount 3051CF Flowmeter selection guide . . . . .17
Rosemount 3051CFA Annubar Flowmeter . . . . . . . . .18
Rosemount 3051CFC Compact Flowmeter . . . . . . . . .26
Rosemount 3051CFP Integral Orifice Flowmeter . . . .31
2
Product certifications
Pipe I.D. range codes
. . . . . . . . . . . . . . . . . . . . . . . . . . 56
. . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Dimensional drawings . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Emerson.com/Rosemount
March 2018
Rosemount 3051
Industry leading capabilities extended to IEC 62591
®
(WirelessHART
Cost effectively implement wireless on the industry's most proven platform
Optimize safety with the industry's only intrinsically safe power module
Eliminate wiring design and construction complexities to lower costs by 40 to 60 percent
Quickly deploy new pressure, level, and flow measurements in 70 percent less time
)

Innovative, integrated DP Flowmeters

Fully assembled, configured, and leak tested for out-of-the-box installation
Reduce straight pipe requirements, lower permanent pressure loss, and achieve accurate
measurement in small line sizes
Up to 1.65 percent volumetric flow accuracy at 8:1 turndown

Proven, reliable, and innovative DP Level Technologies

Connect to virtually any process with a comprehensive offering of process connections, fill
fluids, direct mount, or capillary connections and materials
Quantify and optimize total system performance with QZ option
Operate at higher temperature and in vacuum applications
Optimize level measurement with cost efficient Rosemount Tuned-System
Assemblies

Instrument manifolds – quality, convenient, and easy

Designed and engineered for optimal performance with Rosemount transmitters
Save installation time and money with factory assembly
Offers a variety of styles, materials, and configurations
Emerson.com/Rosemount
3
Rosemount 3051
March 2018

Rosemount 3051C Coplanar Pressure Transmitter

Rosemount 3051C Coplanar™ Pressure Transmitters are the industry standard for differential, gage, and absolute pressure measurement. The coplanar platform enables seamless integration with manifolds, flow and level solutions. Capabilities include:
Power advisory can proactively detect degraded electrical loop integrity issues
(option code DA0)
LOI with straightforward menus and built-in configuration buttons (option code M4)
Safety Certification (option code QT)
Additional information:
Specifications: page 44 Certifications: page 56 Dimensional drawings: page 67
See Specifications and options for more details on each configuration. Specification and selection of product materials, options, or components must be made by the purchaser of the equipment. See page 53 for more information on material selection.
Table 1. Rosemount 3051C Coplanar Pressure Transmitters Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
(1)
Model
3051C Coplanar pressure transmitter
Transmitter t ype
Measurement type
D Differential
G Gage
(2)
A
Absolute
Pressure range
Differential (Rosemount 3051CD)
1
2
3
4
5
(3)
0
–25 to 25 inH2O (–62,16 to 62,16 mbar)
–250 to 250 inH2O (–621,60 to 621,60 mbar)
–1000 to 1000 inH2O (–2,48 to 2,48 bar)
–300 to 300 psi (–20,68 to 20,68 bar)
–2000 to 2000 psi (–137,89 to 137,89 bar)
–3 to 3 inH2O (-7,46 to 7,46 mbar)
Gage (Rosemount 3051CG)
–25 to 25 inH2O (–62,16 to 62,16 mbar)
–250 to 250 inH2O (–621,60 to 621,60 mbar)
–393 to 1000 inH2O (–0,97 to 2,48 bar)
–14.2 to 300 psi (–0,97 to 20,68 bar)
–14.2 to 2000 psi (–0,97 to 137,89 bar)
N/A N/A
Absolute (Rosemount 3051CA)
0 to 30 psia (0 to 2,06 bar)
0 to 150 psia (0 to 10,34 bar)
0 to 800 psia (0 to 55,15 bar)
0 to 4000 psia (0 to 275,79 bar)
N/A
Transmitter output
(4)
A
F FOUNDATION™ Fieldbus Protocol
(5)
W
(6)
X
(7)
M
4–20 mA with digital signal based on HART® Protocol
PROFIBUS® PA Protocol
Wireless (requires wireless options and engineered polymer housing)
Low-power, 1–5 Vdc with digital signal based on HART Protocol
4
Emerson.com/Rosemount
March 2018
Rosemount 3051
Table 1. Rosemount 3051C Coplanar Pressure Transmitters Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Materials of construction
Process flange type Flange material Drain/vent
2 Coplanar Stainless steel (SST) SST
(8)
3
Coplanar Cast C-276 Alloy C-276
4 Coplanar Alloy 400 Alloy 400/K-500
5 Coplanar Plated carbon steel (CS) SST
(8)
7
(8)
8
Coplanar SST Alloy C-276
Coplanar Plated CS Alloy C-276
0 Alternate process connection
Isolating diaphragm
(8)
2
(8)
3
(9)
4
(9)
5
(9)
6
(9)
7
316L SST
Alloy C-276
Alloy 400
Tantalum (available on Rosemount 3051CD and CG, ranges 2–5 only; not available on Rosemount 3051CA)
Gold-plated alloy 400 (use in combination with O-ring option code B)
Gold-plated 316 SST
O-ring
A Glass-filled PTFE
B Graphite-filled PTFE
Sensor fill fluid
1 Silicone
(9)
2
Inert (differential and gage only)
Housing material Conduit entry size
A Aluminum
B Aluminum M20 x 1.5
E Aluminum, ultra low copper
F Aluminum, ultra low copper M20 x 1.5
J SST
K SST M20 x 1.5
(10)
P
(11)
D
(11)
M
Engineered polymer No conduit entries
Aluminum G1/2
SST G1/2
1
/2–14 NPT
1
/2–14 NPT
1
/2–14 NPT
Wireless options (requires wireless output code X and engineered polymer housing code P)
Wireless transmit rate, operating frequency, and protocol
WA3 User Configurable Transmit Rate, 2.4GHz WirelessHART Protocol
Antenna and SmartPower
WP5 Internal antenna, compatible with Green Power Module (I.S. Power Module sold separately)
Emerson.com/Rosemount
5
Rosemount 3051
March 2018
Table 1. Rosemount 3051C Coplanar Pressure Transmitters Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Options (include with selected model number)
Extended product warranty
WR3 3-year limited warranty
WR5 5-year limited warranty
Plantweb™ control functionality
A01 FOUNDATION Fieldbus control function block suite
Plantweb diagnostic functionality
(13)
DA0
(12)
D01
Alternate flange
Power Advisory HART Diagnostic
FOUNDATION Fieldbus Diagnostics Suite
(14)
H2 Traditional flange, 316 SST, SST drain/vent
(8)
H3
Traditional flange, alloy C, alloy C-276 drain/vent
H4 Traditional flange, cast alloy 400, alloy 400/K-500 drain/vent
(8)
H7
Traditional flange, 316 SST, alloy C-276 drain/vent
HJ DIN-compliant traditional flange,SST,7/16-in. adapter/manifold bolting
FA Level flange, SST, 2-in., ANSI class 150, vertical mount 316 SST drain/vent
FB Level flange, SST, 2-in., ANSI Class 300, vertical mount 316 SST drain/vent
FC Level flange, SST, 3-in., ANSI Class 150, vertical mount 316 SST drain/vent
FD Level flange, SST, 3-in., ANSI Class 300, vertical mount 316 SST drain/vent
FP DIN level flange, SST, DN 50, PN 40, vertical mount 316 SST drain/vent
FQ DIN level flange, SST, DN 80, PN 40, vertical mount 316 SST drain/vent
(15)
HK
DIN compliant traditional flange, SST, 10 mm adapter/manifold bolting 316 SST
HL DIN compliant traditional flange, SST, 12 mm adapter/manifold bolting 316 SST
Manifold assembly
(16)
S5 Assemble to Rosemount 305 Integral Manifold
S6 Assemble to Rosemount 304 Manifold or Connection System
Integral mount primary element
S3 Assemble to Rosemount 405 Compact Orifice Plate
(17)
S4
Seal assemblies
(18)
S1
(19)
S2
Mounting bracket
B4 Coplanar flange bracket, all SST, 2-in. pipe and panel
B1 Traditional flange bracket, CS, 2-in. pipe
B2 Traditional flange bracket, CS, panel
B3 Traditional flange flat bracket, CS, 2-in. pipe
B7 Traditional flange bracket, B1 with SST bolts
B8 Traditional flange bracket, B2 with SST bolts
B9 Traditional flange bracket, B3 with SST bolts
Assemble to Rosemount Annubar™ or Rosemount 1195 Integral Orifice
(16)
Assemble to one Rosemount 1199 seal
Assemble to two Rosemount 1199 seals
(20)
(12)
(15)(16)
6
Emerson.com/Rosemount
March 2018
Rosemount 3051
Table 1. Rosemount 3051C Coplanar Pressure Transmitters Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
BA Traditional flange bracket, B1, all SST
BC Traditional flange bracket, B3, all SST
Product certifications
E8 ATEX Flameproof and Dust Certification
(21)
I1
IA ATEX FISCO Intrinsic Safety; for FOUNDATION Fieldbus or PROFIBUS PA Protocol only
N1 ATEX Type n Certification and Dust
K8 ATEX Flameproof, Intrinsic Safety, Type n, Dust (combination of E8, I1, and N1)
(22)
E4
E5 FM Explosion-proof, Dust Ignition-Proof
(23)
I5
IE FM FISCO Intrinsically Safe; for FOUNDATION Fieldbus or PROFIBUS PA Protocol only
K5 FM Explosion-proof, Dust Ignition-Proof, Intrinsically Safe, and Division 2
C6 CSA Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2
(10)
I6
K6 CSA and ATEX Explosion-proof, Intrinsically Safe, and Division 2 (combination of C6, E8, and I1)
E7 IECEx Flameproof, Dust Ignition-proof
I7 IECEx Intrinsic Safety
N7 IECEx Type n Certification
K7 IECEx Flame-proof, Dust Ignition-proof, Intrinsic Safety, and Type n (combination of I7, N7, and E7)
E2 INMETRO Flameproof
I2 INMETRO Intrinsic Safety
IB INMETRO FISCO intrinsically safe; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only
K2 INMETRO Flameproof, Intrinsic Safety
E3 China Flameproof
I3 China Intrinsic Safety
N3 China Type n
EM Technical Regulations Customs Union (EAC) Flameproof
IM Technical Regulations Customs Union (EAC) Intrinsic Safety
KM Technical Regulations Customs Union (EAC) Flameproof and Intrinsic Safety
KB FM and CSA Explosion-proof, Dust Ignition Proof, Intrinsically Safe, and Division 2 (combination of K5 and C6)
KD FM, CSA, and ATEX Explosion-proof, Intrinsically Safe (combination of K5, C6, I1, and E8)
Drinking water approval
DW NSF drinking water approval
Shipboard approvals
SBS American Bureau of Shipping
(25)
SBV
SDN Det Norske Veritas
(25)
SLL
Custody transfer
C5
ATEX Intrinsic Safety and Dust
TIIS Flame-proof
FM Intrinsically Safe, Nonincendive
CSA Intrinsic Safety
(24)
(9)
Bureau Veritas (BV)
Lloyds Register (LR)
(13)
Measurement Canada Accuracy Approval (limited availability depending on transmitter type and range;
contact an Emerson
representative)
Emerson.com/Rosemount
7
Rosemount 3051
March 2018
Table 1. Rosemount 3051C Coplanar Pressure Transmitters Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Bolting material
L4 Austenitic 316 SST bolts
L5 ASTM A 193, grade B7M bolts
L6 Alloy K-500 bolts
Display and interface options
(26)
M4
LCD display with LOI
M5 LCD display
Calibration certificate
Q4 Calibration Certificate
(27)
QG
Calibration Certificate and GOST Verification Certificate
QP Calibration certification and tamper evident seal
Material traceability certification
Q8 Material Traceability Certification per EN 10204 3.1
Quality certification for safety
QS Prior-use certificate of FMEDA data
QT Safety certified to IEC 61508 with certificate of FMEDA
(13)
Configuration buttons
(13)
D4
(28)
DZ
Transient protection
Analog zero and span
Digital zero trim
(9)(29)
T1 Transient protection terminal block
Software configuration
(28)
Custom Software Configuration
C1
(For wired, see the Rosemount 3051 Configuration Data Sheet For wireless, see the Rosemount 3051 Wireless Configuration Data Sheet.
.
)
Low power output
C2 0.8–3.2 Vdc output with digital signal based on HART Protocol (available with output code M only)
Gage pressure calibration
C3 Gage calibration (Rosemount 3051ca4 only)
Alarm levels
C4 Analog output levels compliant with NAMUR recommendation NE 43, alarm high
CN Analog output levels compliant with NAMUR recommendation NE 43, alarm low
CR
CS
CT Rosemount standard low alarm
(13)
Custom alarm and saturation signal levels, high alarm (requires C1 and Rosemount 3051 Configuration Data Sheet
Custom alarm and saturation signal levels, low alarm (requires C1 and Rosemount 3051 Configuration Data Sheet
)
)
Pressure testing
P1 Hydrostatic testing with certificate
8
Emerson.com/Rosemount
March 2018
Rosemount 3051
Table 1. Rosemount 3051C Coplanar Pressure Transmitters Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Cleaning process area
P2 Cleaning for special service
P3 Cleaning for <1 PPM chlorine/fluorine
Flange adapters
DF
(30)
1
/2–14 NPT flange adapter(s)
Vent/drain valves
D7 Coplanar flange without drain/vent ports
Conduit plug
(9)(31)
DO 316 SST conduit plug
RC1/4 RC1/2 process connection
(32)
D9 RC 1/4 flange with RC 1/2 flange adapter - SST
Max static line pressure
P9 4500 psig (310,26 bar) static pressure limit (Rosemount 3051CD ranges 2–5 only)
Ground screw
V5 External ground screw assembly
(9)(33)
Surface finish
Q16 Surface finish certification for sanitary remote seals
Toolkit total system performance reports
QZ Remote seal system performance calculation report
Conduit electrical connector
(9)
GE M12, 4-pin, male connector (eurofast®)
GM A size Mini, 4-pin, male connector (minifast®)
NACE® certificate
(34)
Q15 Certificate of Compliance to NACE MR0175/ISO 15156 for wetted materials
Q25 Certificate of Compliance to NACE MR0103 for wetted materials
Cold temperature
(35)
BR5 –58 °F (–50 °C) cold temperature
BR6 –76 °F (–60 °C) cold temperature
HART Revision configuration (requires HART output code A)
(4)
HR5 Configured for HART Revision 5
HR7 Configured for HART Revision 7
Typical model number: 3051CD 2 A 2 2 A 1 A B4
1. Select configuration buttons (option code D4 or DZ) or LOI (option code M4) if local configuration buttons are required.
2. If ordered with Wireless output code X, only range 1–4, 316L SST diaphragm material (code 2), silicone fill fluid (code 1) and wireless housing (code P) are available.
3. Rosemount 3051CD0 is only available with output code A and X. For output code A, only process flange code 0 (alternate flange H2, H7, HJ, or HK), isolating diaphragm code 2, O ring code A and bolting option L4 are available. For output code X, only process flange code 0 (alternate flange H2), isolating diaphragm code 2, O ring code A and bolting option L4 are available.
Emerson.com/Rosemount
9
Rosemount 3051
4. O pti on HR 5 co nfig ure s the HAR T out put to HA RT Re vis ion 5 . Op tion HR7 conf igu res t he H ART outp ut to HAR T Rev isio n 7. The devi ce can be field configured to H ART Revision 5 or 7 if desired. HART Revision 5 is the default HART output.
5. For local addressing and configuration, M4 LOI is required.
6. Available approvals are FM Intrinsically Safe, (option code I5), CSA Intrinsically Safe (option code I6), ATEX Intrinsic Safety (option code I1), IECEx Intrinsic Safety (option code I7) and EAC Intrinsic Safety (option code IM).
7. Only available with C6, E2, E5, I5, K5, KB, and E8 product certifications. Not available with GE, GM, SBS, DA0, M4, D4, DZ, QT, HR5, HR7, CR, CS, and CT.
8. Materials of Construction comply with recommendations per NACE MR0175/ISO 15156 for sour oil field production environments. Environmental limits apply to certain materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining environments.
9. Not available with wireless output (code X).
10. Only available with wireless output (code X).
11. Not available with Product certifications options E8, K8, E5, K5, C6, K6, E7, K7, E2, K2, E3, KB, and KD.
12. Only valid with F
13. Only available with HART 4–20 mA output (code A).
14. Requires 0 code in materials of construction for alternate process connection.
15. Not valid with option code P9 for 4500 psi static pressure.
16. “Assemble-to” items are specified separately and require a completed model number.
17. Process flange limited to coplanar (option codes 2, 3, 5, 7, 8) or traditional (option codes H2, H3, H7).
18. Not valid with option code D9 for RC
19. Not valid for option codes DF and D9 for adapters.
20. Panel mounting bolts are not supplied.
21. Dust approval not applicable to output code X. See “Certificate: AG-0226; AG-0454; AG-0477” on page 62 for wireless approvals.
22. Only available with output codes A - 4–20mA HART, F - F
23. Nonincendive certification not provided with Wireless output (code X).
24. Not available with Alloy C-276 isolator (code 3), tantalum isolator (code 5), all cast C-276 flanges, all plated carbon steel (CS) flanges, all DIN flanges, all Level flanges, assemble-to manifolds (codes S5 and S6), assemble-to seals (codes S1 and S2), assemble-to primary elements (codes S3 and S4), surface finish certification (code Q16), and remote seal system report (code QZ).
25. Only available with product certifications E7, E8, I1, I7, IA, K7, K8, KD, N1, and N7
26. Not available with F
27. Contact an Emerson representative for availability.
28. Only available with HART 4–20 mA output (code A) and wireless output (code X)
29. The T1 option is not needed with FISCO Product Certifications; transient protection is included in the FISCO product certification codes IA, IB, and IE.
30. Not valid with alternate process connection options S3, S4, S5, and S6.
31. Transmitter is shipped with a 316 SST conduit plug (uninstalled) in place of standard CS conduit plug.
32. Not available with alternate process connection; DIN flanges and level flanges.
33. The V5 option is not needed with the T1 option; external ground screw assembly is included with the T1 option.
34. NACE compliant wetted materials are identified by Footnote 8.
35. Only available on pressure ranges 1–5, with silicone sensor fill fluid and SST or C-276 isolating diaphragms.
OUNDATION Fieldbus output (code F).
1
/2 adapters.
OUNDATION Fieldbus, and W - PROFIBUS PA. Also only available with G
OUNDATION Fieldbus (code F), wireless (code X), or low power (code M).
1
/2 housing thread types.
March 2018
10
Emerson.com/Rosemount
March 2018
Rosemount 3051

Rosemount 3051T In-Line Pressure Transmitter

Rosemount 3051T In-Line Pressure Transmitters are the industry standard for gage and absolute pressure measurement. The in-line, compact design allows the transmitter to be connected directly to a process for quick, easy and cost effective installation. Capabilities include:
Power advisory can proactively detect degraded electrical loop integrity issues
(option code DA0)
LOI with straightforward menus and built-in configuration buttons (option code M4)
Safety Certification (option code QT)
Additional information:
Specifications: page 44 Certifications: page 56 Dimensional drawings: page 67
See “Specifications” on page 44 and options for more details on each configuration. Specification and selection of product materials, options, or components must be made by the purchaser of the equipment. See page 53 for more information on material selection.
Table 2. Rosemount 3051T In-Line Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
(1)
Model
Transmitter type
3051T In-line pressure transmitter
Pressure type
G Gage
(2)
A
Absolute
Pressure range
Gage (Rosemount 3051TG)
1 –14.7 to 30 psi (–1,01 to 2,06 bar) 0 to 30 psia (0 to 2,06 bar)
2 –14.7 to 150 psi (–1,01 to 10,34 bar) 0 to 150 psia (0 to 10,34 bar)
3 –14.7 to 800 psi (–1,01 to 55,15 bar) 0 to 800 psia (0 to 55,15 bar)
4 –14.7 to 4000 psi (–1,01 to 275,79 bar) 0 to 4000 psia (0 to 275,79 bar)
5 –14.7 to 10000 psi (–1,01 to 689,47 bar) 0 to 10000 psia (0 to 689,47 bar)
(4)
6
–14.7 to 20000 psi (–1,01 to 1378,95 bar) 0 to 20000 psia (0 to 1378,95 bar)
(3)
Absolute (Rosemount 3051TA)
Transmitter outpu t
(5)
A
F FOUNDATION Fieldbus Protocol
(6)
W
(7)
X
(8)
M
4–20 mA with digital signal based on HART Protocol
PROFIBUS PA Protocol
Wireless (requires wireless options and engineered polymer housing)
Low-power 1–5 Vdc with digital signal based on HART Protocol
Emerson.com/Rosemount
11
Rosemount 3051
March 2018
Table 2. Rosemount 3051T In-Line Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Process connection style
2B
(9)
2C
(10)
2F
(11)
61
Isolating diaphragm
2 316L SST 316L SST
3 Alloy C-276 Alloy C-276
7 Gold-plated 316 SST 316L SST
1
/2–14 NPT female (range 1–5 only)
G1/2 A DIN 16288 male (range 1–4 only)
Coned and threaded, compatible with autoclave type F-250-C (range 5–6 only)
Non-threaded Instrument flange (range 1–4 only)
(12)
Process connection wetted parts material
Sensor fill fluid
1 Silicone
(11)
2
Inert
Housing material Conduit entry size
A Aluminum
B Aluminum M20 x 1.5
E Aluminum, ultra low copper
F Aluminum, ultra low copper M20 x 1.5
J SST
K SST M20 x 1.5
(13)
P
(14)
D
(14)
M
Engineered polymer No conduit entries
Aluminum G1/2
SST G1/2
1
/2–14 NPT
1
/2–14 NPT
1
/2–14 NPT
Wireless options (requires wireless output code X and engineered polymer housing code P)
Wireless transmit rate, operating frequency, and protocol
WA3 User configurable transmit rate, 2.4GHz WirelessHART Protocol
Antenna and SmartPower
WP5 Internal antenna, compatible with Green Power Module (I.S. Power Module sold separately)
Options (include with selected model number)
Extended product warranty
WR3 3-year limited warranty
WR5 5-year limited warranty
12
Emerson.com/Rosemount
March 2018
Rosemount 3051
Table 2. Rosemount 3051T In-Line Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Plantweb control functionality
(15)
A01 FOUNDATION Fieldbus control function block suite
Plantweb diagnostic functionality
(24)
DA0
(15)
D01
Integral assembly
S5 Assemble to Rosemount 306 Integral Manifold
Diaphragm seal assemblies
S1 Assemble to one Rosemount 1199 seal
Mounting bracket
B4 Bracket for 2-in. pipe or panel mounting, all SST
Power Advisory HART diagnostic
FOUNDATION Fieldbus diagnostics suite
(16)
(16)
(17)
Product certifications
E8 ATEX Flameproof and Dust Certification
(18)
I1
IA ATEX Intrinsic Safety for FISCO; for F
N1 ATEX Type n Certification and Dust
K8 ATEX Flame-proof, Intrinsic Safety, Type n, Dust (combination of E8, I1, and N1)
(19)
E4
E5 FM Explosion-proof, Dust Ignition-proof
(20)
I5
IE FM FISCO Intrinsically Safe; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only
K5 FM Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2
C6 CSA Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2
(13)
I6
K6 CSA and ATEX Explosion-proof, Intrinsically Safe, and Division 2 (combination of C6, E8, and I1)
E7 IECEx Flameproof, Dust Ignition-proof
I7 IECEx Intrinsic Safety
N7 IECEx Type n Certification
ATEX Intrinsic Safety and Dust
OUNDATION Fieldbus or PROFIBUS PA Protocols only
TIIS Flameproof
FM Intrinsically Safe, Nonincendive
CSA Intrinsic Safety
K7 IECEx Flameproof, Dust Ignition-proof, Intrinsic Safety, and Type n (combination of I7, N7, and E7)
E2 INMETRO Flameproof
I2 INMETRO Intrinsic Safety
IB INMETRO FISCO intrinsically safe; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only
K2 INMETRO Flameproof, Intrinsic Safety
E3 China Flameproof
Emerson.com/Rosemount
13
Rosemount 3051
March 2018
Table 2. Rosemount 3051T In-Line Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
I3 China Intrinsic Safety
N3 China Type n
EM Technical Regulations Customs Union (EAC) Flameproof
IM Technical Regulations Customs Union (EAC) Intrinsic Safety
KM Technical Regulations Customs Union (EAC) Flameproof and Intrinsic Safety
KB FM and CSA Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2 (combination of K5 and C6)
KD FM, CSA, and ATEX Explosion-proof, Intrinsically Safe (combination of K5, C6, I1, and E8)
Drinking water approval
(21)
DW NSF drinking water approval
Shipboard approvals
(11)
SBS American Bureau of Shipping
(22)
SBV
Bureau Veritas (BV)
SDN Det Norske Veritas
(22)
SLL
Lloyds Register (LR)
Custody transfer
C5
Measurement Canada Accuracy Approval (Limited availability depending on transmitter type and range. Contact an Emerson representative.)
Calibration certification
Q4 Calibration Certificate
(23)
QG
Calibration Certificate and GOST Verification Certificate
QP Calibration Certification and tamper evident seal
Material traceability certification
Q8 Material Traceability Certification per EN 10204 3.1
Quality certification for safety
(24)
QS Prior-use certificate of FMEDA Data
QT Safety certified to IEC 61508 with certificate of FMEDA
Configuration buttons
(24)
D4
DZ
(25)
Analog zero and span
Digital zero trim
Display and interface options
(26)
M4
LCD display with LOI
M5 LCD display
Wireless SST sensor module
(13)
WSM Wireless SST sensor module
14
Emerson.com/Rosemount
March 2018
Rosemount 3051
Table 2. Rosemount 3051T In-Line Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Conduit plug
(11)(27)
DO 316 SST conduit plug
Transient terminal block
(11)(28)
T1 Transient protection terminal block
Software configuration
(25)
Custom Software Configuration
C1
(For wired, see the Rosemount 3051 Configuration Data Sheet. For wireless, see the Rosemount 3051 Wireless Configuration Data Sheet.
)
Low power output
C2 0.8–3.2 Vdc output with digital signal based on HART Protocol (available with output code M only)
Alarm levels
C4 Analog output levels compliant with NAMUR recommendation NE 43, alarm high
CN Analog output levels compliant with NAMUR recommendation NE 43, low alarm
CR
CS
(24)
Custom alarm and saturation signal levels, high alarm (requires C1 and Rosemount 3051 Configuration Data Sheet
Custom alarm and saturation signal levels, low alarm (requires C1 and Rosemount 3051 Configuration Data Sheet)
)
CT Rosemount standard low alarm
Pressure testing
P1 Hydrostatic testing with certificate
Cleaning process area
(29)
P2 Cleaning for special service
P3 Cleaning for <1 PPM chlorine/fluorine
Ground screw
(11)(30)
V5 External ground screw assembly
Surface finish
Q16 Surface finish certification for sanitary remote seals
Toolkit total system performance reports
QZ Remote seal system performance calculation report
Conduit electrical connector
GE M12, 4-pin, male connector (eurofast)
GM A size mini, 4-pin, male connector (minifast)
NACE certificate
(31)
(11)
Q15 Certificate of Compliance to NACE MR0175/ISO15156 for wetted materials
Q25 Certificate of Compliance to NACE MR0103 for wetted materials
Emerson.com/Rosemount
15
Rosemount 3051
March 2018
Table 2. Rosemount 3051T In-Line Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Cold temperature
BR5 –50 °F (–58 °C) cold temperature
BR6 –76 °F (–60 °C) cold temperature
HART Revision configuration (requires HART output code A)
(5)
HR5 Configured for HART Revision 5
HR7 Configured for HART Revision 7
Typical model number: 3051T G 5 F 2A 2 1 A B4
1. Select configuration buttons (option code D4 or DZ) or LOI (option code M4) if local configuration buttons are required.
2. Wireless output (code X) only available in absolute measurement type (code A) in range 1–5 with (code P). Wireless output and range 6 is only available with coned and threaded process connection (code 2F) and polymer housing.
3. Rosemount 3051TG lower range limit varies with atmospheric pressure.
4. Not available with PROFIBUS PA or low power 1–5 Vdc transmitter output (option code W or M), inert sensor fill fluid (option code 2), NSW drinking water approval (option code DW), or assemble to manifolds (option code S5).
5. O pti on HR 5 co nfig ure s the HAR T out put to H ART Re vis ion 5. O ptio n HR 7 con figu res the HART out put t o HA RT Re visi on 7 . Th e dev ice can be field configured to HART Revision 5 or 7 if desired. HART Revision 5 is the default HART output.
6. For local addressing and configuration, M4 LOI is required.
7. Requires wireless options and engineered polymer housing. Available approvals are FM Intrinsically Safe, (option code I5), CSA Intrinsically Safe (option code I6), ATEX Intrinsic Safety (option code I1), IECEx Intrinsic Safety (option code I7), and EAC Intrinsic Safety (option code IM).
8. Only available with C6, E2, E5, I5, K5, KB, and E8 product certifications. Not available with GE, GM, SBS, DA0, M4, D4, DZ, QT, HR5, HR7, CR, CS, and CT.
9. Wireless output (code X) only available in G fluid (code 1) and housing (code P).
10. Not available with wireless output for range 5.
11. Not available with wireless (output code X).
12. Materials of construction comply with recommendations per NACE MR0175/ISO 15156 for sour oil field production environments. Environmental limits apply to certain materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining environments.
13. Only available with wireless (output code X).
14. Not available with Product certifications options E8, K8, E5, K5, C6, K6, E7, K7, E2, K2, E3, KB, and KD.
15. Only valid with F
16. “Assemble-to” items are specified separately and require a completed model number.
17. Panel mounting bolts are not supplied.
18. Dust approval not applicable to output code X. See “Certificate: AG-0226; AG-0454; AG-0477” on page 62 for wireless approvals.
19. Only available with output codes A - 4–20mA HART, F - F
20. Nonincendive certification not provided with wireless output (code X).
21. Not available with Alloy C-276 isolator (option code 3), assemble-to manifolds (option code S5), assemble-to seals (option code S1), surface finish certification (option code Q16), and remote seal system report (option code QZ).
22. Only available with product certifications E7, E8, I1, I7, IA, K7, K8, KD, N1, and N7.
23. Contact an Emerson representative for availability.
24. Only available with HART 4–20 mA output (code A).
25. Only available with HART 4–20 mA output (code A) and wireless output (code X).
26. Not available with F
27. Transmitter is shipped with 316 SST conduit plug (uninstalled) in place of standard CS conduit plug.
28. The T1 option is not needed with FISCO Product Certifications; transient protection is included in the FISCO product certification codes IA, IB, and IE.
29. Not valid with alternate process connection S5.
30. The V5 option is not needed with T1 option; external ground screw assembly is included with the T1 option.
31. NACE compliant wetted materials are identified by Footnote 11.
OUNDATION Fieldbus output code F.
OUNDATION Fieldbus (code F) and wireless output (code X) or low power (code M).
1
/2 A DIN 16288 male process connection (code 2C) with range 1–4, 316 SST isolating diaphragm (code 2), silicone fill
OUNDATION Fieldbus, and W - PROFIBUS PA. Also only available with G
1
/2–14 NPT process connection (code 2B), and polymer housing
1
/2 housing thread types.
16
Emerson.com/Rosemount
March 2018
Rosemount 3051

Rosemount 3051CF Flowmeter selection guide

Rosemount 3051CF Flowmeters combine the proven Rosemount 3051 Pressure Transmitter and the latest primary element technologies. All flowmeters are fully assembled, calibrated, configured, and leak tested for out-of-the-box installation and are available with wired or wireless capabilities to meet all of your application needs.
Rosemount 3051CFA Annubar Flowmeter
Rosemount Annubar technology minimizes permanent pressure loss while delivering best in class accuracy.
Lowest material costs for large line sizes
Flo-tap enables installation without process shutdown
Realize up to 96 percent less permanent pressure loss compared to
traditional orifice plate installations
Rosemount 3051CFC Compact Conditioning Flowmeter
Rosemount Compact Conditioning technologies provide unprecedented performance with minimal straight-run requirements. Solutions include conditioning orifice plate or Rosemount Annubar primary elements.
Conditioning Orifice requires only two pipe diameters upstream and
downstream
Eliminate swirl and regular profiles resulting in more stable and
accurate flow measurement
Savings up to 55 percent when compared to a traditional orifice plate
installation can be realized
Emerson.com/Rosemount
Rosemount 3051CFP Integral Orifice Flowmeter
Rosemount Integral Orifice Flowmeters deliver highly accurate small-bore flow measurement capability with minimal installation and maintenance requirements.
Best performance for small line sizes
(15 to 40 mm)
Precision honed pipe section and tight machining tolerances deliver
higher installed performance
Reduces uncertainty by up to five percent compared to traditional
orifice plate installation
1
/2- to 11/2-in.
17
Rosemount 3051
March 2018

Rosemount 3051CFA Annubar Flowmeter

The Rosemount 3051CFA Annubar Flowmeter utilizes the T-shaped sensor design that delivers best in class accuracy and performance while meeting the needs of diverse process applications, whether it is high accuracy for precision control or high strength for severe flow applications. Main capabilities include:
Up to 1.8 percent of flow rate accuracy
Available in 2-to 96-in. (50 to 2400 mm) line
Fully assembled and leak tested for out-of-the-box installation
Power advisory can proactively detect degraded electrical loop integrity issues
(option code DA0)
LOI with straightforward menus and built-in configuration buttons
(option code M4)
Additional information:
Specifications: page 44 Certifications: page 56 Dimensional drawings: page 67
See “Specifications” on page 44 and options for more details on each configuration. Specification and selection of product materials, options, or components must be made by the purchaser of the equipment. See page 53 for more information on material selection.
Table 3. Rosemount 3051CFA Annubar Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
(1)
Model
3051CFA Rosemount Annubar Flowmeter
Product description
Measurement type
D Differential Pressure
Fluid type
L Liquid
G Gas
S Steam
Line size
020 2-in. (50 mm)
025 21/2-in. (63.5 mm)
030 3-in. (80 mm)
035 31/2-in. (89 mm)
040 4-in. (100 mm)
050 5-in. (125 mm)
060 6-in. (150 mm)
070 7-in. (175 mm)
080 8-in. (200 mm)
100 10-in. (250 mm)
120 12-in. (300 mm)
140 14-in. (350 mm)
160 16-in. (400 mm)
180 18-in. (450 mm)
200 20-in. (500 mm)
18
Emerson.com/Rosemount
March 2018
Table 3. Rosemount 3051CFA Annubar Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
240 24-in. (600 mm)
300 30-in. (750 mm)
360 36-in. (900 mm)
420 42-in. (1066 mm)
480 48-in. (1210 mm)
600 60-in. (1520 mm)
720 72-in. (1820 mm)
780 78-in (1950 mm)
840 84-in. (2100 mm)
900 90-in. (2250 mm)
960 96-in (2400 mm)
Rosemount 3051
Pipe I.D. range
C Range C from the Pipe I.D. range codes table
D Range D from the Pipe I.D. range codes table
A Range A from the Pipe I.D. range codes table
B Range B from the Pipe I.D. range codes table
E Range E from the Pipe I.D. range codes table
Z Non-standard Pipe I.D. range codes or line sizes greater than 12-in.
Pipe material/mounting assembly material
C CS (A105)
S 316 SST
0 No mounting (customer supplied)
G Chrome-moly grade F-11
N Chrome-moly grade F-22
J Chrome-moly grade F-91
Piping orientation
H Horizontal piping
D Vertical piping with downwards flow
U Vertical piping with upwards flow
Annubar type
P Pak-lok
F Flanged with opposite side support
L Flange-lok
G Gear-drive flo-tap
M Manual flo-tap
Sensor material
S 316 SST
H Alloy C-276
Sensor size
1 Sensor size 1 — line sizes 2- to 8-in. (50 to 200 mm)
2 Sensor size 2 — line sizes 6- to 96-in. (150 to 2400 mm)
3 Sensor size 3 — line sizes greater than 12-in. (300 mm)
Emerson.com/Rosemount
19
Rosemount 3051
March 2018
Table 3. Rosemount 3051CFA Annubar Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Mounting type
T1 Compression or threaded connection
A1 Class 150 RF ANSI
A3 Class 300 RF ANSI
A6 Class 600RF ANSI
D1 DN PN16 flange
D3 DN PN40 flange
D6 DN PN100 flange
(2)
A9
(2)
AF
(2)
AT
R1 Class 150 RTJ flange
R3 Class 300 RTJ flange
R6 Class 600RTJ flange
(2)
R9
(2)
RF
(2)
RT
Class 900 RF ANSI
Class 1500 RF ANSI
Class 2500 RF ANSI
Class 900 RTJ flange
Class 1500 RTJ flange
Class 2500 RTJ flange
Opposite side support or packing gland
0 No opposite side support or packing gland (required for pak-lok and flange-lok models)
Opposite side support – required for flanged models
C NPT threaded opposite support assembly – extended tip
D Welded opposite support assembly – extended tip
Packing gland – required for flo-tap models
Packing gland material Rod material Packing material
(3)
J
(3)
K
(3)
L
(3)
N
SST packing gland/cage nipple CS PTFE
SST packing gland/cage nipple SST PTFE
SST packing gland/cage nipple CS Graphite
SST packing gland/cage nipple SST Graphite
R Alloy C-276 packing gland/cage nipple SST Graphite
Isolation valve for flo-tap models
0 Not applicable or customer supplied
1 Gate valve, CS
2 Gate valve, SST
5 Ball valve, CS
6 Ball valve, SST
Temperature measurement
T Integral RTD – not available with flanged model greater than Class 600
0 No temperature sensor
R Remote Thermowell and RTD
20
Emerson.com/Rosemount
March 2018
Rosemount 3051
Table 3. Rosemount 3051CFA Annubar Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Transmitter connection platform
3 Direct-mount, Integral 3-valve manifold– not available with flanged model greater than Class 600
5 Direct -mount, 5-valve manifold – not available with flanged model greater than Class 600
7 Remote-mount NPT connections (1/2-in. NPT)
6 Direct-mount, high temperature 5-valve Manifold – not available with flanged model greater than Class 600
8 Remote-mount SW connections (1/2-in.)
Differential pressure range
1 0 to 25 in H2O (0 to 62,16 mbar)
2 0 to 250 in H2O (0 to 621,60 mbar)
3 0 to 1000 in H2O (0 to 2,48 bar)
Transmitter outpu t
(4)
A
F FOUNDATION Fieldbus Protocol
(5)
W
(6)
X
(7)
M
4–20 mA with digital signal based on HART Protocol
PROFIBUS PA Protocol
Wireless (requires wireless options and engineered polymer housing)
Low-power 1–5 Vdc with digital signal based on HART Protocol
Transmitter housing material Conduit entry size
A Aluminum
B Aluminum M20 x 1.5
J SST
E Aluminum, ultra copper low
F Aluminum, ultra copper low M20 x 1.5
K SST M20 x 1.5
(8)
P
(9)
D
(9)
M
Engineered polymer No conduit entries
Aluminum G1/2
SST G1/2
1
/2–14 NPT
1
/2–14 NPT
1
/2–14 NPT
Transmitter performance class
1 1.8 percent flow rate accuracy, 8:1 flow turndown, 5-yr. stability
Wireless options (requires wireless output code X and Engineered Polymer Housing Code P)
Wireless transmit rate, operating frequency, and protocol
WA3 User configurable transmit rate, 2.4GHz WirelessHART Protocol
Antenna and SmartPower
WP5 Internal antenna, compatible with Green Power Module (I.S. Power Module sold separately)
Options (include with selected model number)
Extended product warranty
WR3 3-year limited warranty
WR5 5-year limited warranty
Emerson.com/Rosemount
21
Rosemount 3051
Table 3. Rosemount 3051CFA Annubar Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Pressure testing
P1 Hydrostatic testing with certificate
PX Extended hydrostatic testing
(10)
March 2018
Special cleaning
P2 Cleaning for special services
PA Cleaning per ASTM G93 Level D (section 11.4)
Material testing
V1 Dye penetrant exam
Material examination
V2 Radiographic Examination
Flow calibration
W1 Flow calibration (Average K)
Special inspection
QC1 Visual and dimensional inspection with certificate
QC7 Inspection & performance certificate
Surface finish
RL Surface finish for low pipe Reynolds number in gas and steam
RH Surface finish for high pipe Reynolds number in liquid
Material traceability certification
Q8 Material Traceability Certification per EN 10474:2004 3.1
Code conformance
J2 ANSI/ASME B31.1
J3 ANSI/ASME B31.3
Materials conformance
J5 NACE MR-0175 / ISO 15156
(12)
(13)
(11)
Country certification
J6 European Pressure Directive (PED)
J1 Canadian Registration
Installed in flanged pipe spool section
H3 Class 150 flanged connection with Rosemount standard length and schedule
H4 Class 300 flanged connection with Rosemount standard length and schedule
H5 Class 600 flanged connection with Rosemount standard length and schedule
Instrument connections for remote mount options
G2 Needle valves, SST
G6 OS&Y gate valve, SST
G1 Needle valves, CS
G3 Needle valves, alloy C-276
G5 OS&Y gate valve, CS
G7 OS&Y gate valve, alloy C-276
22
Emerson.com/Rosemount
March 2018
Rosemount 3051
Table 3. Rosemount 3051CFA Annubar Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Special shipment
Y1 Mounting hardware shipped separately
Special dimensions
VM Variable mounting
VT Variable tip
VS Variable length spool section
Plantweb control functionality
(14)
A01 FOUNDATION Fieldbus control function block suite
Plantweb diagnostic functionality
(15)
DA0
D01
(14)
Power Advisory HART diagnostic
FOUNDATION Fieldbus diagnostics suite
Product certifications
E8 ATEX Flameproof, Dust
(16)
I1
IA ATEX FISCO Intrinsic Safety; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only
N1 ATEX Type n and Dust
K8 ATEX Flameproof, Intrinsic Safety, Type n, Dust (combination of E8, I1 and N1)
E5 FM Explosion-proof, Dust Ignition-proof
(17)
I5
IE FM FISCO Intrinsically Safe; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only
K5 FM Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2 (combination of E5 and I5)
C6 CSA Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2
(8)
I6
K6 CSA and ATEX Explosion-proof, Intrinsically Safe, and Division 2 (combination of C6, E8, and I1)
E7 IECEx Flameproof, Dust Ignition-proof
I7 IECEx Intrinsic Safety
N7 IECEx Type n
K7 IECEx Flameproof, Dust Ignition-proof, Intrinsic Safety, and Type n (combination of I7, N7, and E7)
E2 INMETRO Flameproof
I2 INMETRO Intrinsic Safety
IB INMETRO FISCO intrinsically safe; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only
K2 INMETRO Flameproof, Intrinsic Safety
E3 China Flameproof
I3 China Intrinsic Safety
KB FM and CSA Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2 (combination of K5 and C6)
KD CSA, FM, and ATEX Explosion-proof, Intrinsically Safe (combination of K5, C6, I1, and E8)
ATEX Intrinsic Safety and Dust
FM Intrinsically Safe, Nonincendive
CSA Intrinsically Safe
Sensor fill fluid and O-ring options
(18)
L1
Inert sensor fill fluid (silicone fill fluid is standard)
L2 Graphite-filled (PTFE) O-ring
(18)
LA
Inert sensor fill fluid and graphite-filled (PTFE) O-ring
Emerson.com/Rosemount
23
Rosemount 3051
March 2018
Table 3. Rosemount 3051CFA Annubar Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Shipboard approvals
(18)
SBS American Bureau of Shipping
(19)
SLL
Lloyds Register (LR)
Display and interface options
(20)
M4
LCD display with LOI
M5 LCD display
Transmitter calibration certification
Q4 Calibration certificate for transmitter
Quality certification for safety
(15)
QS Prior-use certificate of FMEDA data
QT Safety certified to IEC 61508 with certificate of FMEDA
Transient protection
(18)(21)
T1 Transient terminal block
Manifold for remote mount option
F2 3-valve manifold, SST
F6 5-valve manifold, SST
F1 3-valve manifold, CS
F3 3-valve manifold, alloy C-276
F5 5-valve manifold, CS
F7 5-valve manifold, alloy C-276
Lower power output
C2 0.8–3.2 Vdc output with digital signal based on HART Protocol (available with output code M only)
Alarm levels
(15)
C4 NAMUR alarm and saturation levels, high alarm
CN NAMUR alarm and saturation levels, low alarm
CR Custom alarm and saturation signal levels, high alarm
CS Custom alarm and saturation signal levels, low alarm
CT Rosemount standard low alarm
Configuration buttons
(15)
D4
(22)
DZ
Ground screw
Analog zero and span
Digital zero trim
(18)(23)
V5 External ground screw assembly
HART Revision configuration (requires HART Protocol output code A)
(4)
HR5 Configured for HART Revision 5
HR7 Configured for HART Revision 7
Typical model number: 3051CFA D L 060 D C H P S 2 T1 0 0 0 3 2 A A 1
1. Select configuration buttons (option code D4 or DZ) or LOI (option code M4) if local configuration buttons are required.
24
Emerson.com/Rosemount
March 2018
2. Available in remote mount applications only.
3. The cage nipple is constructed of Rosemount 304 SST.
4. O pti on HR 5 co nfig ure s the HAR T out put to HA RT Re vis ion 5. Op tio n HR7 con figu res the H ART outp ut t o HA RT Rev isi on 7 . The dev ice can be field configured to HART Revision 5 or 7 if desired. HART Revision 5 is the default HART output.
5. For local addressing and configuration, M4 LOI is required.
6. Requires wireless options and engineered polymer housing. Available approvals are FM Intrinsically Safe, (option code I5), CSA Intrinsically Safe (option code I6), ATEX Intrinsic Safety (option code I1), and IECEx Intrinsic Safety (option code I7).
7. Only available with C6, E2, E5, I5, K5, KB, and E8 approval. Not available with GE, GM, SBS, DA0, M4, D4, DZ, QT, HR5, HR7, CR, CS, and CT.
8. Only available with wireless (code X).
9. Not available with Product certifications options E8, K8, E5, K5, C6, K6, E7, K7, E2, K2, E3, KB, and KD.
10. Applies to assembled flowmeter only, mounting not tested.
11. Instrument connections for remote mount options and isolation valves for flo-tap models are not included in the Material Traceability Certification.
12. Not available with transmitter connection platform 6.
13. Materials of Construction comply with metallurgical requirements within NACE MR0175/ISO for sour oil field production environments. Environmental limits apply to certain materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining environments.
14. Only valid with F
15. Only available with 4–20 mA HART (code A).
16. Dust approval not applicable to output code X. See “Certificate: AG-0226; AG-0454; AG-0477” on page 62 for wireless approvals.
17. Nonincendive certification not provided with wireless output (code X).
18. Not available with wireless output (code X).
19. Only available with Product cer tifications E7, E8, I1, I7, IA, K7, K8, KD, N1, and N7
20. Not available with F
21. The T1 option is not needed with FISCO Product Certifications, transient protection is included with the FISCO Product Certification codes IA, IB, and IE.
22. Only available with 4–20 mA HART (output code A) and wireless (output code X).
23. The V5 option is not needed with the T1 option; external ground screw assembly is included with the T1 option.
OUNDATION Fieldbus (code F).
OUNDATION Fieldbus (output code F) or wireless (output code X) or low power (output code M).
Rosemount 3051
Emerson.com/Rosemount
25
Rosemount 3051
March 2018

Rosemount 3051CFC Compact Flowmeter

Rosemount 3051CFC Compact Flowmeters provide a quick, reliable installation between existing raised face flanges. Depending on your application needs, you can reduce energy loss with the Compact Annubar or minimize straight run requirements with the conditioning orifice.
Up to 1.8 percent of flow rate accuracy
Available in
Fully assembled and leak tested for out-of-the-box installation
Power advisory can proactively detect degraded electrical loop integrity issues.
(option code DA0)
LOI with straightforward menus and built-in configuration buttons
(option code M4)
Additional information:
Specifications: page 44 Certifications: page 56 Dimensional drawings: page 67
See “Specifications” on page 44 and options for more details on each configuration. Specification and selection of product materials, options, or components must be made by the purchaser of the equipment. See page 53 for more information on material selection.
Table 4. Rosemount 3051CFC Compact Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
(1)
Model
Product description
3051CFC Compact Flowmeter
Measurement type
D Differential Pressure
Primary element technology
A Annubar averaging pitot tube C Conditioning orifice plate P Orifice plate
Material type
S 316 SST
Line size
(2)
005
(2)
010
(2)
015 020 2-in. (50 mm) 030 3-in. (80 mm) 040 4-in. (100 mm) 060 6-in. (150 mm) 080 8-in. (200 mm)
(3)
100
(3)
120
1
/2-in. (15 mm) 1-in. (25 mm) 11/2-in. (40 mm)
10-in. (250 mm) 12-in. (300 mm)
Primary element type
N000 Rosemount Annubar sensor size 1 N040 0.40 beta ratio N050 0.50 beta ratio
(4)
N065
0.65 beta ratio
1
/2- to 12-in. (15 to 300 mm) line sizes
26
Emerson.com/Rosemount
March 2018
Rosemount 3051
Table 4. Rosemount 3051CFC Compact Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Temperature measurement
0 No temperature sensor R Remote thermowell and RTD
Transmitter connection platform
3 Direct-mount 7 Remote-mount, NPT connections
Differential pressure range
1 0 to 25 inH2O (0 to 62,16 mbar) 2 0 to 250 inH2O (0 to 621,60 mbar) 3 0 to 1000 inH2O (0 to 2,48 bar)
Transmitter outpu t
(5)
A
4–20 mA with digital signal based on HART Protocol
F FOUNDATION Fieldbus Protocol
(6)
W X M
(7)
(8)
PROFIBUS PA Protocol Wireless (requires wireless options and engineered polymer housing) Low-power 1–5 Vdc with digital signal basedon HART Protocol
Transmitter housing material Conduit entry size
A Aluminum B Aluminum M20 x 1.5 E Aluminum, ultra low copper F Aluminum, ultra low copper M20 x 1.5 J SST K SST M20 x 1.5
(9)
P
(10)
D
(10)
M
Engineered polymer No conduit entries Aluminum G1/2 SST G1/2
1
/2–14 NPT
1
/2–14 NPT
1
/2–14 NPT
Transmitter per formance class
1 Up to ±1.65% flow rate accuracy, 8:1 flow turndown, 5-year stability
Wireless options (requires wireless output code X and engineered polymer housing code P)
Wireless transmit rate, operating frequency, and protocol
WA3 User configurable transmit rate, 2.4 GHz WirelessHART Protocol
Antenna and SmartPower
WP5 Internal antenna, compatible with Green Power Module (I.S. Power Module sold separately)
Options (include with selected model number)
Extended product warranty
WR3 3-year limited warranty WR5 5-year limited warranty
Installation accessories
AB ANSI alignment ring (Class 150) (only required for 10- and 12-in. (250 and 300mm) line sizes) AC ANSI alignment ring (Class 300) (only required for 10- and 12-in. (250 and 300mm) line sizes) AD ANSI alignment ring (Class 600) (only required for 10- and 12-in. (250 and 300mm) line sizes) DG DIN alignment ring (PN16)
Emerson.com/Rosemount
27
Rosemount 3051
March 2018
Table 4. Rosemount 3051CFC Compact Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
DH DIN alignment ring (PN40) DJ DIN alignment ring (PN100) JB JIS alignment ring (10K) JR JIS alignment ring (20K) JS JIS alignment ring (40K)
Remote adapters
FE Flange adapters 316 SST (1/2-in NPT)
High temperature application
HT Graphite valve packing (T
Flow calibration
(11)
WC Flow calibration, 3 pt, conditioning orifice option C (all pipe schedules)
(12)
WD
Flow calibration, 10 pt, conditioning option C (all schedules), Rosemount Annubar option A (schedule 40)
= 850 °F)
max
Pressure testing
P1 Hydrostatic testing with certificate
Special cleaning
(13)
P2
Cleaning for special services
PA Cleaning per ASTM G93 level D (section 11.4)
Special inspection
QC1 Visual & dimensional inspection with certificate QC7 Inspection and performance certificate
Transmitter calibration certification
Q4 Calibration certificate for transmitter
Quality certification for safety
(14)
QS Prior-use certificate of FMEDA data QT Safety certified to IEC 61508 with certificate of FMEDA
Material traceability certification
Q8 Material Traceability Certification per EN 10204:2004 3.1
Code conformance
J2 ANSI/ASME B31.1 J3 ANSI/ASME B31.3 J4 ANSI/ASME B31.8
Materials conformance
(15)
J5 NACE MR-0175/ISO 15156
Country certification
J1 Canadian Registration
Product certifications
E8 ATEX Fla meproof, Dust
(16)
I1 IA ATEX FISCO Intrinsic Safety; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only N1 ATEX Type n and Dust K8 ATEX Flameproof, Intrinsic Safety, Type n, Dust (combination of E8, I1, and N1) E5 FM Explosion-proof, Dust Ignition-proof
(17)
I5
ATEX Intrinsic Safety and Dust
FM Intrinsically Safe, Nonincendive
28
Emerson.com/Rosemount
March 2018
Rosemount 3051
Table 4. Rosemount 3051CFC Compact Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
IE FM FISCO Intrinsically Safe; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only K5 FM Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2 (combination of E5 and I5) C6 CSA Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2
(9)
I6
CSA Intrinsically Safe
K6 CSA and ATEX Explosion-proof, Intrinsically Safe, and Division 2 (combination of C6, E8, and I1) E7 IECEx Flameproof, Dust Ignition-proof I7 IECEx Intrinsic Safety N7 IECEx Type n K7 IECEx Flameproof, Dust Ignition-proof, Intrinsic Safety, and Type n (combination of I7, N7, and E7) E2 INMETRO Flameproof I2 INMETRO Intrinsic Safety IB INMETRO FISCO intrinsically safe; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only K2 INMETRO Flameproof, Intrinsic Safety E3 China Flameproof I3 China Intrinsic Safety KB FM and CSA Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2 (combination of K5 and C6) KD CSA, FM, and ATEX Explosion-proof, Intrinsically Safe (combination of K5, C6, I1, and E8)
Sensor fill fluid and O-ring options
(18)
L1
Inert sensor fill fluid
L2 Graphite-filled (PTFE) O-ring
(18)
LA
Shipboard approvals
Inert sensor fill fluid and graphite-filled (PTFE) O-ring
(18)
SBS American Bureau of Shipping
(19)
SLL
Lloyds Register (LR)
Display and interface options
(20)
M4
LCD display with LOI
M5 LCD display
Transient protection
(18)(21)
T1 Transient terminal block
Manifold for remote mount option
F2 3-valve manifold, SST F6 5-Valve Manifold, SST
Plantweb control functionality
(22)
A01 FOUNDATION Fieldbus Control Function Block Suite
Plantweb diagnostic functionality
(14)
DA0 D01
(22)
Power Advisory HART Diagnostic FOUNDATION Fieldbus Diagnostic Suite
Low power output
C2 0.8–3.2 Vdc output with digital signal based on HART Protocol (available with output code M only)
Alarm levels
C4 NAMUR alarm and saturation levels, high alarm CN NAMUR alarm and saturation levels, low alarm CR Custom alarm and saturation signal levels, high alarm CS Custom alarm and saturation signal levels, low alarm CT Rosemount standard low alarm
Emerson.com/Rosemount
(14)
29
Rosemount 3051
March 2018
Table 4. Rosemount 3051CFC Compact Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Ground screw
(18)(23)
V5 External ground screw assembly
Configuration buttons
(14)
D4
(24)
DZ
HART Revision Configuration (requires HART Protocol output code A)
Analog zero and span Digital zero trim
(5)
HR5 Configured for HART Revision 5 HR7 Configured for HART Revision 7
Typical model number: 3051CFC D C S 060 N 065 0 3 2 A A 1 WC E5 M5
1. Select configuration buttons (option code D4 or DZ) or LOI (option code M4) if local configuration buttons are required.
2. Available wi th primary element tech nology P only.
3. 10- and 12-in. (250 and 300 mm) line sizes not available with Primary Element Technology A.
4. For 2-in. (50 mm) line sizes the primary element type is 0.6 for primary element technology (code C).
5. O pti on HR 5 co nfig ure s the HAR T out put to HA RT Re vis ion 5 . Op tion HR7 conf igu res t he H ART o utp ut to HAR T Rev isio n 7. The devi ce can be field configured to HART Revision 5 or 7 if desired. HART Revision 5 is the default HART output.
6. For local addressing and configuration, M4 LOI is required.
7. Requires wireless options and engineered polymer housing. Available approvals are FM Intrinsically Safe, (option code I5), CSA Intrinsically Safe (option code I6), ATEX Intrinsic Safety (option code I1), and IECEx Intrinsic Safety (option code I7).
8. Only available with C6, E2, E5, I5, K5, KB and E8 approval. Not available with GE, GM, SBS, DA0, M4, D4, DZ, QT, HR5, HR7, CR, CS, and CT.
9. Only available with Wireless output code X.
10. Not available with product certifications options E8, K8, E5, K5, C6, K6, E7, K7, E2, K2, E3, KB, and KD.
11. Available with primary element technology C only.
12. For Annubar option A, consult factory for pipe schedules other than schedule 40.
13. Available with primary element technology C or P only.
14. Only available with HART 4–20 mA output code A.
15. Materials of construction comply with metallurgical requirements within NACE MR0175/ISO for sour oil field production environments. Environmental limits apply to certain materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining environments.
16. Dust approval not applicable to output code X. See “Certificate: AG-0226; AG-0454; AG-0477” on page 62 for wireless approvals
17. Nonincendive certification not provided with wireless option (code X).
18. Not available with Wireless output code X.
19. Only available with product certifications E7, E8, I1, I7, IA, K7, K8, KD, N1, and N7
20. Not available with output code F - F
21. The T1 option is not needed with FISCO Product Certifications, transient protection is included with the FISCO Product Certification code IA, IB, and IE.
22. Only valid with F
23. The V5 option is not needed with the T1 option; external ground screw assembly is included with the T1 option.
24. Only available with 4–20 mA HART output code A and wireless output code X.
OUNDATION Fieldbus (output code F).
OUNDATION Fieldbus or Wireless output code X or low power (output code M).
30
Emerson.com/Rosemount
March 2018
Rosemount 3051

Rosemount 3051CFP Integral Orifice Flowmeter

Rosemount 3051CFP Integral Orifice Flowmeters enable highly accurate flow measurement in small line sizes. Integral Orifice utilize precision honed pipe section for increased accuracy and self-centering plate design to prevent alignment errors that magnify measurement inaccuracies in small line sizes.
Up to 1.75 percent of flow rate accuracy
Available in
Fully assembled and leak tested for out-of-the-box installation
Power advisory can proactively detect degraded electrical loop integrity issues.
(option code DA0)
LOI with straightforward menus and built-in configuration buttons
(option code M4)
Additional information:
Specifications: page 44 Certifications: page 56 Dimensional drawings: page 67
See “Specifications” on page 44 and options for more details on each configuration. Specification and selection of product materials, options, or components must be made by the purchaser of the equipment. See page 53 for more information on material selection.
Table 5. Rosemount 3051CFP Integral Orifice Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
(1)
Model
Product description
3051CFP Integral Orifice Flowmeter
Measurement type
D Differential Pressure
Body material
S 316 SST
Line size
005 010 1-in. (25 mm) 015 11/2-in. (40 mm)
1
/2-in. (15 mm)
Process connection
T1 NPT female body (not available with remote thermowell and RTD)
(2)
S1 P1 Pipe ends: NPT threaded P2 Pipe ends: beveled D1 Pipe ends: Flanged, DIN PN16, slip-on D2 Pipe ends: Flanged, DIN PN40, slip-on D3 Pipe ends: Flanged, DIN PN100, slip-on W1 Pipe ends: Flanged, RF, ANSI Class 150, weld-neck W3 Pipe ends: Flanged, RF, ANSI Class 300, weld-neck W6 Pipe ends: Flanged, RF, ANSI Class 600, weld-neck A1 Pipe ends: Flanged, RF, ANSI Class 150, slip-on A3 Pipe ends: Flanged, RF, ANSI Class 300, slip-on A6 Pipe ends: Flanged, RF, ANSI Class 600, slip-on R1 Pipe ends: Flanged, RTJ, ANSI Class 150, slip-on R3 Pipe ends: Flanged, RTJ, ANSI Class 300, slip-on R6 Pipe ends: Flanged, RTJ, ANSI Class 600, slip-on
Socket weld body (not available with remote thermowell and RTD)
1
/2- to 11/2-in. (15 to 40 mm) line sizes
Emerson.com/Rosemount
31
Rosemount 3051
March 2018
Table 5. Rosemount 3051CFP Integral Orifice Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Orifice plate material
S 316 SST H Alloy C-276 M Alloy 400
Bore size option
0066 0.066-in. (1.68 mm) for 1/2-in. pipe 0109 0.109-in. (2.77 mm) for 1/2-in. pipe 0160 0.160-in. (4.06 mm) for 1/2-in. pipe 0196 0.196-in. (4.98 mm) for 1/2-in. pipe 0260 0.260-in. (6.60 mm) for 1/2-in. pipe 0340 0.340-in. (8.64 mm) for 1/2-in. pipe 0150 0.150-in. (3.81 mm) for 1-in. pipe 0250 0.250-in. (6.35 mm) for 1-in. pipe 0345 0.345-in. (8.76 mm) for 1-in. Pipe 0500 0.500-in. (12.70 mm) for 1-in. pipe 0630 0.630-in. (16.00 mm) for 1-in. pipe 0800 0.800-in. (20.32 mm) for 1-in. pipe 0295 0.295-in. (7.49 mm) for 11/2-in. pipe 0376 0.376-in. (9.55 mm) for 11/2-in. pipe 0512 0.512-in. (13.00 mm) for 11/2-in. pipe 0748 0.748-in. (19.00 mm) for 11/2-in. pipe 1022 1.022-in. (25.96 mm) for 11/2-in. pipe 1184 1.184-in. (30.07 mm) for 11/2-in. pipe 0010 0.010-in. (0.25 mm) for 1/2-in. pipe 0014 0.014-in. (0.36 mm) for 1/2-in. pipe 0020 0.020-in. (0.51 mm) for 1/2-in. pipe 0034 0.034-in. (0.86 mm) for 1/2-in. pipe
Transmitter connection platform
D3 Direct-mount, 3-valve manifold, SST D5 Direct-mount, 5-valve manifold, SST R3 Remote-mount, 3-valve manifold, SST R5 Remote-mount, 5-valve manifold, SST D4 Direct-mount, 3-valve manifold, alloy C-276 D6 Direct-mount, 5-valve manifold, alloy C-276 D7 Direct-mount, high temperature, 5-valve manifold, SST R4 Remote-mount, 3-valve manifold, alloy C-276 R6 Remote-mount, 5-valve manifold, alloy C-276
Differential pressure ranges
1 0 to 25 inH2O (0 to 62,16 mbar) 2 0 to 250 inH2O (0 to 621,60 mbar) 3 0 to 1000 in H2O (0 to 2,48 bar)
Transmitter outpu t
(3)
A F FOUNDATION Fieldbus Protocol
(4)
W
(5)
X
(6)
M
32
4–20 mA with digital signal based on HART Protocol
PROFIBUS PA Protocol Wireless Low-power 1–5 Vdc with digital signal based on HART Protocol
Emerson.com/Rosemount
March 2018
Rosemount 3051
Table 5. Rosemount 3051CFP Integral Orifice Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Transmitter housing material Conduit entry size
A Aluminum B Aluminum M20 x 1.5 E Aluminum, ultra low copper F Aluminum, ultra low copper M20 x 1.5 J SST K SST M20 x 1.5
(7)
P
(8)
D
(8)
M
Engineered polymer No conduit entries Aluminum G1/2 SST G1/2
1
/2–14 NPT
1
/2–14 NPT
1
/2–14 NPT
Transmitter per formance class
1 up to ±1.8% flow rate accuracy, 8:1 flow turndown, 5-year stability
Wireless options (requires Wireless output code X and engineered polymer housing code P)
Wireless transmit rate, operating frequency, and protocol
WA3 User configurable transmit rate, 2.4GHz WirelessHART Protocol
Antenna and SmartPower
WP5 Internal antenna, compatible with Green Power Module (I.S. Power Module sold separately)
Options (include with selected model number)
Extended product warranty
WR3 3-year limited warranty WR5 5-year limited warranty
Transmitter body/bolt material
GT High temperature (850 °F/454 °C)
Temperature sensor
RT Thermowell and RTD
Optional connection
G1 DIN 19213 transmitter connection
Pressure testing
P1 Hydrostatic testing with certificate
Special cleaning
P2 Cleaning for special services PA Cleaning per ASTM G93 level D (section 11.4)
Material testing
V1 Dye penetrant exam
Material examination
V2 Radiographic examination
Flow calibration
WD Discharge coefficient verification
Special inspection
QC1 Visual and dimensional inspection with certificate QC7 Inspection and performance certificate
(9)
(10)
(11)
Emerson.com/Rosemount
33
Rosemount 3051
March 2018
Table 5. Rosemount 3051CFP Integral Orifice Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Material traceability certification
Q8 Material traceability certification per EN 10204:2004 3.1
Code conformance
(12)
J2 ANSI/ASME B31.1 J3 ANSI/ASME B31.3 J4 ANSI/ASME B31.8
Materials conformance
(13)
J5 NACE MR-0175/ISO 15156
Country certification
J6 European Pressure Directive (PED) J1 Canadian Registration
Transmitter calibration certification
Q4 Calibration Certificate for Transmitter
Quality certification for safety
(14)
QS Prior-use certificate of FMEDA data QT Safety certified to IEC 61508 with certificate of FMEDA
Product certifications
E8 ATEX Flameproof, Dust
(15)
I1
ATEX Intrinsic Safety and Dust IA ATEX FISCO Intrinsic Safety; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only N1 ATEX Type n and Dust K8 ATEX Flameproof, Intrinsic Safety, Type n, Dust (combination of E8, I1, and N1) E5 FM Explosion-proof, Dust Ignition-proof
(16)
I5
FM Intrinsically Safe, Nonincendive IE FM FISCO Intrinsically Safe; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only K5 FM Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2 (combination of E5 and I5) C6 CSA Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2
(7)
I6
CSA Intrinsically Safe K6 CSA and ATEX Explosion-proof, Intrinsically Safe, and Division 2 (combination of C6, E8, and I1) E7 IECEx Flameproof, Dust Ignition-proof I7 IECEx Intrinsic Safety N7 IECEx Type n K7 IECEx Flameproof, Dust Ignition-proof, Intrinsic Safety, and Type n (combination of I7, N7 and E7) E2 INMETRO Flameproof I2 INMETRO Intrinsic Safety IB INMETRO FISCO intrinsically safe; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only K2 INMETRO Flameproof, Intrinsic Safety E3 China Flameproof I3 China Intrinsic Safety KB FM and CSA Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2 (combination of K5 and C6) KD CSA, FM, and ATEX Explosion-proof, Intrinsically Safe (combination of K5, C6, I1, and E8)
Sensor fill fluid and O-ring options
(17)
L1 L2 Graphite-filled (PTFE) O-ring
(17)
LA
Inert sensor fill fluid
Inert sensor fill fluid and graphite-filled (PTFE) O-ring
34
Emerson.com/Rosemount
March 2018
Rosemount 3051
Table 5. Rosemount 3051CFP Integral Orifice Flowmeter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Shipboard approvals
(17)
SBS American Bureau of Shipping
(18)
SLL
Lloyds Register (LR)
Display and interface options
(19)
M4
LCD display with LOI M5 LCD display
Transient protection
(17)(20)
T1 Transient terminal block
Plantweb control functionality
(21)
A01 FOUNDATION Fieldbus control function block suite
Plantweb diagnostic functionality
(14)
DA0 D01
(21)
Power Advisory HART diagnostic
FOUNDATION Fieldbus diagnostic suite
Low power output
C2 0.8–3.2 Vdc output with digital signal based on HART Protocol (available with output code M only)
Alarm levels
(14)
C4 NAMUR alarm and saturation levels, high alarm CN NAMUR alarm and saturation levels, low alarm CR Custom alarm and saturation signal levels, high alarm CS Custom alarm and saturation signal levels, low alarm CT Rosemount standard low alarm
Ground screw
(17)(22)
V5 External ground screw assembly
Configuration buttons
(14)
D4
(23)
DZ
Hart Revision configuration (requires hart protocol output code A)
Analog zero and span
Digital zero trim
(3)
HR5 Configured for HART Revision 5 HR7 Configured for HART Revision 7
Typical model number: 3051CFP D S 010 W1 S 0500 D3 2 A A 1 E5 M5
1. Select configuration buttons (option code D4 or DZ) or LOI (option code M4) if local configuration buttons are required.
2. To improve pipe perpendicularity for gasket sealing, socket diameter is smaller than standard pipe O.D.
3. Option HR5 configures the HART output to HART Revision 5. Option HR7 configures the HART output to HART Revision 7. The device can be field configured to HART Revision 5 or 7 if desired. HART Revision 5 is the default HART output.
4. For local addressing and configuration, M4 (LOI) is required.
5. Requires wireless options and engineered polymer housing. Available approvals are FM Intrinsically Safe, (option code I5), CSA Intrinsically Safe (option code I6), ATEX Intrinsic Safety (option code I1), and IECEx Intrinsic Safety (option code I7).
6. Only available with C6, E2, E5, I5, K5, KB, and E8 approval. Not available with GE, GM, SBS, DA0, M4, D4, DZ, QT, HR5, HR7, CR, CS, and CT.
7. Only available with wireless output code X.
8. Not available with product certifications options E8, K8, E5, K5, C6, K6, E7, K7, E2, K2, E3, KB, and KD.
9. Thermowell Material is the same as the body material.
10. Does not apply to process connection codes T1 and S1.
11. Not available for bore sizes 0010, 0014, 0020, 0034, 0066, or 0109.
12. Not available with DIN Process Connection codes D1, D2, or D3.
Emerson.com/Rosemount
35
Rosemount 3051
13. Materials of Construction comply with metallurgical requirements within NACE MR0175/ISO for sour oil field production environments. Environmental limits apply to certain materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining environments.
14. Only available with HART 4–20 mA output ( code A).
15. Dust approval not applicable to output code X. See “Certificate: AG-0226; AG-0454; AG-0477” on page 62 for wireless approvals.
16. Nonincendive certification not provided with Wireless output (code X).
17. Not available with wireless output code X.
18. Only available with product certifications E7, E8, I1, I7, IA, K7, K8, KD, N1, and N7.
19. Not available with F
20. The T1 option is not needed with FISCO Product Certifications, transient protection is included with the FISCO Product Certification code IA, IB, and IE.
21. Only valid with F
22. The V5 option is not needed with the T1 option; external ground screw assembly is included with the T1 option.
23. Only available with 4–20 mA (code A) and wireless (code X).
OUNDATION Fieldbus (code F) or wireless output (code X) or low power (code M).
OUNDATION Fieldbus output code F.
March 2018
36
Emerson.com/Rosemount
March 2018
Rosemount 3051

Rosemount 3051L Level Transmitter

The Rosemount 3051L Level Transmitter combines the performance and capabilities of Rosemount 3051 Transmitters with the reliability and quality of a direct mount seal in one model number. 3051L Level Transmitters offer a variety of process connections, configurations, and fill fluid types to meet a breadth of level applications. Capabilities of a Rosemount 3051L Level Transmitter include:
Quantify and optimize total system performance (option code QZ)
Tuned-System Assembly (option code S1)
Power advisory can proactively detect degraded electrical loop integrity issues
(option code DA0)
LOI with straightforward menus and built-in configuration buttons (option code M4)
Additional information:
Specifications: page 44 Certifications: page 56 Dimensional drawings: page 67
See Specifications and options for more details on each configuration. Specification and selection of product materials, options, or components must be made by the purchaser of the equipment. See page 53 for more information on material selection.
Table 6. Rosemount 3051L Level Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
(1)
Model
Transmitter type
3051L Level Transmitter
Pressure range
2 –250 to 250 inH2O (–621,60 to 621,60 mbar)
3 –1000 to 1000 inH2O (–2,48 to 2,48 bar)
4 –300 to 300 psi (–20,68 to 20,68 bar)
Transmitter outpu t
(2)
A
4–20 mA with digital signal based on HART Protocol
F FOUNDATION Fieldbus Protocol
(3)
W
X
M
(4)
(5)
PROFIBUS PA Protocol
Wireless (requires wireless options and engineered polymer housing)
Low-power 1–5 Vdc with digital signal based on HART Protocol
Process connection size, material, extension length (high side)
Code Process connection size Material Extension length
(6)
G0
(6)
H0
J0 2-in./DN 50 Ta n t a lu m Flush mount only
(6)
A0
(6)
A2
(6)
A4
(6)
A6
(6)
B0
2-in./DN 50/A 316L SST Flush mount only
2-in./DN 50 Alloy C-276 Flush mount only
3-in./DN 80 316L SST Flush mount
3-in./DN 80 316L SST 2-in./50 mm
3-in./DN 80 316L SST 4-in./100 mm
3-in./DN 80 316L SST 6-in./150 mm
4-in./DN 100 316L SST Flush mount
Emerson.com/Rosemount
37
Rosemount 3051
March 2018
Table 6. Rosemount 3051L Level Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
(6)
B2
B4
B6
C0
C2
C4
C6
D0
D2
D4
D6
(6)
(6)
(6)
(6)
(6)
(6)
(6)
(6)
(6)
(6)
4-in./DN 100 316L SST 2-in./50 mm
4-in./DN 100 316L SST 4-in./100 mm
4-in./DN 100 316L SST 6-in./150 mm
3-in./DN 80 Alloy C-276 Flush mount
3-in./DN 80 Alloy C-276 2-in./50 mm
3-in./DN 80 Alloy C-276 4-in./100 mm
3-in./DN 80 Alloy C-276 6-in./150 mm
4-in./DN 100 Alloy C-276 Flush Mount
4-in./DN 100 Alloy C-276 2-in./50 mm
4-in./DN 100 Alloy C-276 4-in./100 mm
4-in./DN 100 Alloy C-276 6-in./150 mm
E0 3-in./DN 80 Ta nt a l u m Flush mount only
F0 4-in./DN 100 Ta n t al u m Flush mount only
Mounting flange size, rating, material (high side)
Size Rating Material
M 2-in. ANSI/ASME B16.5 Class 150 CS
A 3-in. ANSI/ASME B16.5 Class 150 CS
B 4-in. ANSI/ASME B16.5 Class 150 CS
N 2-in. ANSI/ASME B16.5 Class 300 CS
C 3-in. ANSI/ASME B16.5 Class 300 CS
D 4-in. ANSI/ASME B16.5 Class 300 CS
P 2-in. ANSI/ASME B16.5 Class 600 CS
E 3-in. ANSI/ASME B16.5 Class 600 CS
(6)
X
(6)
F
(6)
G
(6)
Y
(6)
H
(6)
J
(6)
Z
(6)
L
Q DN 50 PN 10-40 per EN 1092-1 CS
R DN 80 PN 40 per EN 1092-1 CS
S DN 100 PN 40 per EN 1092-1 CS
V DN 100 PN 10/16 per EN 1092-1 CS
(6)
K
(6)
T
(6)
U
2-in. ANSI/ASME B16.5 Class 150 316 SST
3-in. ANSI/ASME B16.5 Class 150 316 SST
4-in. ANSI/ASME B16.5 Class 150 316 SST
2-in. ANSI/ASME B16.5 Class 300 316 SST
3-in. ANSI/ASME B16.5 Class 300 316 SST
4-in. ANSI/ASME B16.5 Class 300 316 SST
2-in. ANSI/ASME B16.5 Class 600 316 SST
3-in. ANSI/ASME B16.5 Class 600 316 SST
DN 50 PN 10-40 per EN 1092-1 316 SST
DN 80 PN 40 per EN 1092-1 316 SST
DN 100 PN 40 per EN 1092-1 316 SST
Mounting flange size, rating, material (high side)
Size Rating Material
(6)
W
7
(6)
DN 100 PN 10/16 per EN 1092-1 316 SST
4 in. ANSI/ASME B16.5 Class 600 316 SST
38
Emerson.com/Rosemount
March 2018
Rosemount 3051
Table 6. Rosemount 3051L Level Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
1 N/A 10K per JIS B2238 CS
2 N/A 20K per JIS B2238 CS
3 N/A 40K per JIS B2238 CS
(6)
4
(6)
5
(6)
6
N/A 10K per JIS B2238 316 SST
N/A 20K per JIS B2238 316 SST
N/A 40K per JIS B2238 316 SST
Seal fill fluid (high side) Specific gravity Temperature limits (ambient temperature of 70° F [21° C])
D Silicone 200 0.93 –49 to 401 °F (–45 to 205 °C)
Silicone 200 for
F
Vacuum
0.93
Applications
Dow Corning® 704
L
Diffusion Pump
1.07 32 to 401 °F (0 to 205 °C)
Fluid
D.C. Silicone 704
C
for Vacuum
1.07
Applications
R Silicone 705 1.09 68 to 401 °F (20 to 205 °C)
Silicone 705 for
V
Vacuum
1.09
Applications
A SYLTHERM™ XLT 0.85 –102 to 293 °F (–75 to 145 °C)
H Inert (halocarbon) 1.85 –49 to 320 °F (–45 to 160 °C)
G Glycerin and water 1.13 5 to 203 °F (–15 to 95 °C)
N Neobee® M-20 0.92 5 to 401 °F (–15 to 205 °C)
P
Propylene glycol and water
1.02 5 to 203 F (–15 to 95 °C)
For use in vacuum applications below 14.7 psia (1 bar-a), refer to vapor pressure curves in Rosemount DP Level Fill Fluid Specification
Technical Note
.
For use in vacuum applications below 14.7 psia (1 bar-a), refer to vapor pressure curves in Rosemount DP Level Fill Fluid Specification
Technical Note.
For use in vacuum applications below 14.7 psia (1 bar-a), refer to vapor pressure curves in Rosemount DP Level Fill Fluid Specification
Technical Note.
Low pressure side
Configuration
(6)
11
21 Differential SST 316L SST Silicone
(6)
22
(7)
2A
(6)(7)
2B
(6)
31
Gage SST 316L SST Silicone
Differential SST Alloy C-276 Silicone
Differential SST 316L SST Inert (halocarbon)
Differential SST Alloy C-276 Inert (halocarbon)
Tu ne d- S ys te m Assembly with Remote Seal
Flange adapter
Diaphragm material Sensor fill fluid
None 316L SST
Silicone (requires option code S1)
O-ring
A Glass-filled PTFE
Housing material Conduit entry size
A Aluminum
B Aluminum M20 x 1.5
E Aluminum, ultra low copper
F Aluminum, ultra low copper M20 x 1.5
1
/2–14 NPT
1
/2–14 NPT
Emerson.com/Rosemount
39
Rosemount 3051
March 2018
Table 6. Rosemount 3051L Level Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
J SST
1
/2–14 NPT
K SST M20 x 1.5
(8)
P
(9)
D
(9)
M
Engineered polymer No conduit entries
Aluminum G1/2
SST G1/2
Wireless options (requires wireless output code X and engineered polymer housing code P)
Wireless transmit rate, operating frequency, and protocol
WA3 User configurable transmit rate, 2.4GHz WirelessHART Protocol
Antenna and SmartPower
WP5 Internal antenna, compatible with Green Power Module (I.S. Power Module sold separately)
Options (include with selected model number)
Extended product warranty
WR3 3-year limited warranty
WR5 5-year limited warranty
Plantweb control functionality
A01 FOUNDATION Fieldbus Control Function Block Suite
(10)
Plantweb diagnostic functionality
(18)
DA0
(10)
D01
Seal assemblies
Power advisory HART diagnostic
FOUNDATION Fieldbus diagnostics suite
(11)
S1 Assembled to One Rosemount 1199 Seal
Product certifications
E8 ATEX Flameproof and Dust Certification
(12)
I1
IA ATEX FISCO Intrinsic Safety; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only
N1 ATEX Type n Certification and Dust
K8 ATEX Flameproof, Intrinsic Safety, Type n, Dust (combination of E8, I1, and N1)
(13)
E4
E5 FM Explosion-proof, Dust Ignition-proof
(14)
I5
IE FM FISCO Intrinsically Safe; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only
K5 FM Explosion-proof, Dust Ignition-Proof, Intrinsically Safe, and Division 2
C6 CSA Explosion-proof, Dust Ignition-proof, Intrinsically Safe, and Division 2
(8)
I6
K6 CSA and ATEX Explosion-proof, Intrinsically Safe, and Division 2 (combination of C6, E8, and I1)
E7 IECEx Flameproof, Dust Ignition-proof
I7 IECEx Intrinsic Safety
N7 IECEx Type n Certification
K7 IECEx Flameproof, Dust Ignition-proof, Intrinsic Safety, and Type n (combination of I7, N7, and E7)
E2 INMETRO Flameproof
ATEX Intrinsic Safety and Dust
TIIS Flameproof
FM Intrinsically Safe, Nonincendive
CSA Intrinsic Safety
40
Emerson.com/Rosemount
March 2018
Rosemount 3051
Table 6. Rosemount 3051L Level Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
I2 INMETRO Intrinsic Safety
IB INMETRO FISCO intrinsically safe; for FOUNDATION Fieldbus or PROFIBUS PA Protocols only
K2 INMETRO Flameproof, Intrinsic Safety
E3 China Flameproof
I3 China Intrinsic Safety
N3 China Type n
EM Technical Regulations Customs Union (EAC) Flameproof
IM Technical Regulations Customs Union (EAC) Intrinsic Safety
KM Technical Regulations Customs Union (EAC) Flameproof and Intrinsic Safety
KB FM and CSA Explosion-proof, Dust Ignition Proof, Intrinsically Safe, and Division 2 (combination of K5 and C6)
KD FM, CSA, and ATEX Explosion-proof, Intrinsically Safe (combination of K5, C6, I1, and E8)
Shipboard approvals
(7)
SBS American Bureau of Shipping
(15)
SBV
Bureau Veritas (BV)
SDN Det Norske Veritas
(15)
SLL
Lloyds Register (LR)
Bolting material
L4 Austenitic 316 SST bolts
L5 ASTM A 193, grade B7M bolts
L6 Alloy K-500 Bolts
L8 ASTM A 193 Class 2, grade B8M bolts
Display and interface options
(16)
M4
LCD display with LOI
M5 LCD display
Calibration certification
Q4 Calibration Certificate
QP Calibration Certificate and tamper evident seal
(17)
QG
Calibration Certificate and GOST Verification Certificate
Material traceability certification
Q8 Material Traceability Certification per EN 10204 3.1
Quality certification for safety
(18)
QS Prior-use certificate of FMEDA data
QT Safety certified to IEC 61508 with certificate of FMEDA
Toolkit total system performance reports
QZ Seal system performance calculation report
Conduit electrical connector
GE M12, 4-pin, male connector (eurofast)
GM A size Mini, 4-pin, male connector (minifast)
(7)
Configuration buttons
(18)
D4
DZ
(19)
Analog zero and span
Digital zero trim
Emerson.com/Rosemount
41
Rosemount 3051
March 2018
Table 6. Rosemount 3051L Level Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Transient protection
(7)(20)
T1 Transient protection
Software configuration
(19)
Custom Software Configuration
C1
(For wired, see the Rosemount 3051 Configuration Data Sheet. For wireless, see the Rosemount 3051 Wireless Configuration Data Sheet.
)
Low power output
C2 0.8–3.2 Vdc output with digital signal based on HART Protocol (available with output code M only)
Alarm levels
C4 NAMUR alarm and saturation levels, high alarm
CN NAMUR alarm and saturation levels, low alarm
CR
CS
CT Rosemount Standard low alarm
(18)
Custom alarm and saturation signal levels, high alarm (requires C1 and Rosemount 3051 Configuration Data Sheet
Custom alarm and saturation signal levels, low alarm (requires C1 and Rosemount 3051 Configuration Data Sheet
)
)
Conduit plug
DO 316 SST Conduit Plug
Ground screw
(7)(21)
V5 External ground screw assembly
Lower housing flushing connection options
Ring material Number Size (NPT)
F1 316 SST 1
F2 316 SST 2
F3 Alloy C-276 1
F4 Alloy C-276 2
F7 316 SST 1
F8 316 SST 2
F9 Alloy C-276 1
F0 Alloy C-276 2
1
/4–18 NPT
1
/4–18 NPT
1
/4–18 NPT
1
/4–18 NPT
1
/2–14 NPT
1
/2–14 NPT
1
/2–14 NPT
1
/2–14 NPT
Lower housing intermediate gasket material
S0 No gasket for lower housing
(22)
SY
NACE certificate
Q15 Certificate of compliance to NACE MR0175/ISO 15156 for wetted materials
Q25 Certificate of compliance to NACE MR0103 for wetted materials
HART Revision configuration
HR5 Configured for HART Revision 5
HR7 Configured for HART Revision 7
Thermo-Tork® TN-9000
(23)
(2)
(requires HART Protocol output code A)
Typical model number: 3051L 2 A A0 D 21 A A F1
42
Emerson.com/Rosemount
March 2018
1. Select configuration buttons (option code D4 or DZ) or LOI (option code M4) if local configuration buttons are required.
2. O ption HR5 co nfig ures t he HA RT out put to HART R evisi on 5. Optio n HR7 confi gures the H ART ou tput t o HART Revis ion 7 . The d evice can be field configured to HART Revision 5 or 7 if desired. HART Revision 5 is the default HART output.
3. Option code M4 - LCD display with LOI required for local addressing and configuration.
4. Requires wireless options and engineered polymer housing. Available approvals are FM Intrinsically Safe, (option code I5), CSA Intrinsically Safe (option code I6), ATEX Intrinsic Safety (option code I1), IECEx Intrinsic Safety (option code I7) and EAC Intrinsic Safety (option code IM).
5. Only available with C6, E2, E5, I5, K5, KB, and E8 approval. Not available with GE, GM, SBS, DA0, M4, D4, DZ, QT, HR5, HR7, CR, CS, and CT.
6. Materials of Construction comply with metallurgical requirements highlighted within NACE MR0175/ISO 15156 for sour oil field production environments. Environmental limits apply to certain materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining environments.
7. Not available with wireless output code X.
8. Only available with wireless output code X.
9. Not available with product certifications options E8, K8, E5, K5, C6, K6, E7, K7, E2, K2, E3, KB, and KD.
10. Only valid with F
11. “Assemble-to” items are specified separately and require a completed model number.
12. Dust approval not applicable to output code X. See “Certificate: AG-0226; AG-0454; AG-0477” on page 62 for wireless approvals.
13. Only available with output codes A - 4–20mA HART, F - F
14. Nonincendive certification not provided with Wireless output code X.
15. Only available with product certifications E7, E8, I1, I7, IA, K7, K8, KD, N1, and N7.
16. Not available with F
17. Contact an Emerson representative for availability.
18. Only available with HART 4–20 mA output code A).
19. Only available with 4–20 mA HART (output code A) and Wireless (output code X).
20. The T1 option is not needed with FISCO Product Certifications; transient protection is included in the FISCO product certification codes IA, IB, and IE.
21. The V5 option is not needed with the T1 option; external ground screw assembly is included with the T1 option.
22. Gasket provided when lower housing is ordered.
23. NACE compliant wetted materials are identified by Footnote 6.
OUNDATION Fieldbus output code F.
OUNDATION Fieldbus, and W - PROFIBUS PA. Also only available with G
OUNDATION Fieldbus (output code F) or wireless (output code X) or low power (output code M).
Rosemount 3051
1
/2 housing thread types.
Emerson.com/Rosemount
43
Rosemount 3051
0.015 0.005+
URL
Span
--------------


% of Span±
0.015 0.005+
URL
Span
--------------


% of Span±
0.025 0.005+
URL
Span
--------------


% of Span±
0.0075
URL
Span
--------------


% of Span±
0.0075
URL
Span
--------------


% of Span±
0.0075
URL
Span
--------------


% of Span±
0.025 0.005+
URL
Span
--------------


% of Span±
March 2018

Specifications

Performance specifications

This product data sheet covers HART, WirelessHART, FOUNDATION Fieldbus, and PROFIBUS PA unless specified.
Conformance to specification (±3σ [Sigma])
Technology leadership, advanced manufacturing techniques, and statistical process control ensure specification conformance to at least ±3σ.
Reference accuracy
Stated reference accuracy equations include terminal based linearity, hysteresis, and repeatability. For wireless, FOUNDATION Fieldbus and PROFIBUS PA devices, use calibrated range in place of span.
Models Rosemount 3051 and WirelessHART Protocol
Rosemount 3051C
(1)
Range 5 ±0.065% of span
For spans less than 10:1, accuracy =
Ranges 2–4 ± 0.04% of span
Range 0 (CD) ± 0.10% of span
Rosemount 3051CA
Ranges 1–4 ± 0.04% of span
Rosemount 3051T
Ranges 1–4 ± 0.04% of span
Range 5–6 ± 0.075% of span
Rosemount 3051L
Ranges 2-4 ± 0.075% of span
For spans less than 10:1
Range 1 ± 0.10% of span
For spans less than 15:1, accuracy =
For spans less than 2:1, accuracy = ± 0.05% of URL
For spans less than 10:1, accuracy =
(4)
For spans less than 10:1, accuracy =
For spans less than 10:1, accuracy =
For spans less than 10:1, accuracy =
(2)
(3)
, accuracy =
(2)
(2)
1. For Rosemount 3051C and 3051T with 1199 assemble to code S1, use Rosemount 3051L specification.
2. For output code W and M, ±0.065% span.
3. For output code F, for span less than 5:1.
4. For Rosemount 3051C and 3051T with 1199 assemble to code S1, use Rosemount 3051L specification.
44
Emerson.com/Rosemount
March 2018
β
β
β
βββββ
Rosemount 3051
Flow performance - flow reference accuracy
(1)
Rosemount 3051CFA Annubar Flowmeter
Ranges 2–3 ±1.80% of flow rate at 8:1 flow turndown
Rosemount 3051CFC_A Compact Annubar Flowmeter – Rosemount Annubar option A
Ranges 2–3
Uncalibrated ±2.10% of flow rate at 8:1 flow turndown Calibrated ±1.80% of flow rate at 8:1 flow turndown
Rosemount 3051CFC_C Compact Orifice Flowmeter – conditioning option C
Ranges 2–3
Rosemount 3051CFC_P Compact Orifice Flowmeter – orifice type option P
Ranges 2–3
= 0.4 ±1.75% of flow rate at 8:1 flow turndown = 0.50, 0.65 ±1.95% of flow rate at 8:1 flow turndown
(2)
= 0.4 ±2.00% of flow rate at 8:1 flow turndown = 0.65 ±2.00% of flow rate at 8:1 flow turndown
Rosemount 3051CFP Integral Orifice Flowmeter
<0.1 ±3.00% of flow rate at 8:1 flow turndown
Ranges 2–3
1. Accuracy over range of use is always application dependent. Range 1 flowmeters may experience an additional uncertainty up to 0.9 percent. Consult your Emerson representative for exact specifications.
2. Applicable to 2- to 12-in. line sizes. For smaller line sizes, see the Rosemount DP Flowmeters and Primary Elements Product Data Sheet
0.1< <0.2 ±1.95% of flow rate at 8:1 flow turndown
0.2< <0.6 ±1.75% of flow rate at 8:1 flow turndown
0.6< <0.8 ±2.15% of flow rate at 8:1 flow turndown
.
Tot al performance
Total performance is based on combined errors of reference accuracy, ambient temperature effect, and static pressure effect at normal operating conditions (70 percent of span typical reading, 740 psi (51,02 bar) line pressure).
For ±50 °F (28 °C) temperature changes; 0–100% relative humidity, from 1:1 to 5:1 rangedown
Models Total performance
Rosemount 3051C
Ranges 2–5 ± 0.14% of span
Rosemount 3051T
Ranges 1–4
Rosemount 3051L
Ranges 2–4
1. For output code W, F and M, total performance is ±0.15% of span.
± 0.14% of span
Use Instrument Toolkit™ or the QZ option to quantify the total performance of a remote seal assembly under operating conditions.
(1)
Emerson.com/Rosemount
45
Rosemount 3051
T
c
T
d
Td = Dead time T
c
= Time constant
Pressure released
Response time = Td+T
c
63.2% of total step change
0%
100%
36.8%
Tra nsm itter outpu t vs. T ime
Time
Long term stability
Models Long term stability
March 2018
Rosemount 3051C
Ranges 2–5
Rosemount 3051CD, 3051CG Low/Draft Range
Ranges 0–1
Rosemount 3051CA Low Range
Rosemount 3051T
Ranges 1–4
Rosemount 3051L
Ranges 2–3 Ranges 4–5
Dynamic performance
4–20 mA HART
Tot al response time (Td + Tc)
Rosemount 3051C Ranges 2–5 Range 1 Range 0 Rosemount 3051T Rosemount 3051L
(2)
100 ms 255 ms 700 ms 100 ms See Instrument Toolkit.
Range 1
:
±0.2% of URL for 10 years ±50 °F (28 °C) temperature changes, and up to 1000 psi (68,95 bar) line pressure.
±0.2% of URL for 1 year
±0.2% of URL for 10 years ±50 °F (28 °C) temperature changes, and up to 1000 psi (68,95 bar) line pressure.
±0.2% of URL for 10 years ±50 °F (28 °C) temperature changes, and up to 1000 psi (68,95 bar) line pressure.
±0.1% of URL for 1 year ±0.2% of URL for 1 year
(1)
FOUNDATION Fieldbus and PROFIBUS PA Protocols
(3)
Typical HART transmitter response time
152 ms 307 ms N/A 152 ms See Instrument Toolkit.
Dead time (Td) Update rate
1. Dead time and update rate apply to all models and ranges; analog output only.
2. Nominal total response time at 75 °F (24 °C) reference conditions.
3. Transducer block response time, Analog Input block execution time not included.
4. Does not apply to wireless (output code X). See “Wireless (output code X)” on page 51 for wireless update
Line pressure effect per 1000 psi (68,95 bar)
For line pressures above 2000 psi (137,90 bar) and Ranges 4–5, see the following documents.
For HART, see the Rosemount 3051 Reference Manual. For WirelessHART, see the Rosemount 3051 Wireless Reference Manual. For F
Rosemount 3051CD, 3051CF Zero error
46
45 ms (nominal) 97 ms
(4)
rate.
OUNDATION Fieldbus, see the Rosemount 3051 Reference Manual. For PROFIBUS PA, see the Rosemount 3051 Reference Manual.
22 times per second 22 times per second
Models Line pressure effect
Ranges 2–3 ±0.05% of URL/1000 psi (68,95 bar) for line pressures from 0 to 2000 psi (0 to 137,90 bar)
Range 1 ±0.25% of URL/1000 psi (68,95 bar) for line pressures from 0 to 2000 psi (0 to 137,90 bar) Range 0 ±0.125% of URL/100 psi (6,89 bar) for line pressures from 0 to 750 psi (0 to 51,71 bar)
Span error
Ranges 2–3 ±0.1% of reading/1000 psi (68,95 bar)
Range 1 ±0.4% of reading/1000 psi (68,95 bar) Range 0 ±0.15% of reading/100 psi (6,895 bar)
Emerson.com/Rosemount
March 2018
Ambient temperature effect per 50 °F (28 °C)
Models Ambient temperature effect
Rosemount 3051
Rosemount 3051C
Ranges 2–5
Range 1 ±(0.1% URL + 0.25% span) from 1:1 to 30:1
Range 0 ±(0.25% URL + 0.05% span) from 1:1 to 30:1
Rosemount 3051CA
Ranges 1–4
Rosemount 3051T
Range 2–4
Range 1 ±(0.025% URL + 0.125% span) from 1:1 to 10:1
Range 5–6 ±(0.1% URL + 0.15% span) from 1:1 to 5:1
Rosemount 3051L See instrument toolkit software.
±(0.0125% URL + 0.0625% span) from 1:1 to 5:1 ±(0.025% URL + 0.125% span) from 5:1 to 150:1
±(0.14% URL + 0.15% span) from 30:1 to 50:1
±(0.025% URL + 0.125% span) from 1:1 to 30:1 ±(0.035% URL + 0.125% span) from 30:1 to 150:1
±(0.025% URL + 0.125% span) from 1:1 to 30:1 ±(0.035% URL + 0.125% span) from 30:1 to 150:1
±(0.05% URL + 0.125% span) from 10:1 to 100:1
Mounting position effects
Models Mounting position effects
Rosemount 3051C Zero shifts up to ±1.25 inH2O (3,11 mbar), which can be calibrated out. No span effect.
Rosemount 3051CA, 3051T Zero shifts up to ±2.5 inH2O (6,22 mbar), which can be calibrated out. No span effect.
With liquid level diaphragm in vertical plane, zero shift of up to ±1 inH2O (2,49 mbar). With
Rosemount 3051L
diaphragm in horizontal plane, zero shift of up to ±5 inH on extended units. All zero shifts can be calibrated out. No span effect.
O (12,43 mbar) plus extension length
2
Vibration effect
Less than ±0.1% of URL when tested per the requirements of IEC60770-1: 1999 field or pipeline with high vibration level (10–60 Hz 0.21 mm displacement peak amplitude/60–2000 Hz 3g).
Power supply effect
Less than ±0.005% of calibrated span per volt change
Electromagnetic compatibility (EMC)
Meets all relevant requirements of EN61326-1:2006 and Namur NE-21.
1. NAMUR NE-21 does not apply to wireless output code X.
(1)
Transient protection (option code T1)
Tested in accordance with IEEE C62.41.2-2002, location category B
6 kV crest (0.5 μs - 100 kHz)3 kA crest (8 x 20 μs)6 kV crest (1.2 x 50 μs)
Emerson.com/Rosemount
47
Rosemount 3051

Functional specifications

Range and sensor limits
Table 7. Rosemount 3051CD, 3051CG, 3051CF, and 3051L Range and Sensor Limits
Minimum span Range and sensor limits
Lower (LRL)
(1)
Rosemount
Range
0
1
2
3
4
5
3051CD, 3051CG,
3051CF,
(2)
3051L
0.10 inH2O
(0,24 mbar)
0.50 inH2O
(1,24 mbar)
1.67 inH2O
(4,15 mbar)
6.67 inH2O
(16,58 mbar)
2.00 psi
(137,89 mbar)
13.33 psi
(919,01 mbar)
Upper
(URL)
3.00 inH2O
(7,45 mbar)
25.00 inH2O
(62,16 mbar)
250.00 inH2O
(621,60 mbar)
1000.00 inH2O (2,48 bar)
300.00 psi
(20,68 bar)
2000.00 psi
(137,89 bar)
Rosemount 3051CD
differential,
3051CF Flowmeters
–3.00 inH2O
(–7,45 mbar)
–25.00 inH2O
(–62,16 mbar)
–250.00 inH2O
(–621,60 mbar) –1000.00 inH2O
(–2,48 bar)
–300.00 psi
(–20,68 bar)
– 2000.00 psi (–137,89 bar)
Rosemount
3051CG gage
N/A N/A N/A
–25.00 inH2O
(–62,16 mbar)
–250.00 inH2O
(–621,60 mbar)
0.50 psia
(34,47 mbar)
0.50 psia
(34,47 mbar)
0.50 psia
(34,47 mbar)
Rosemount
(3)
differential
–250.00 inH2O
(–621,60 mbar)
–1000.00 inH2O
(–2,48 bar)
–300.00 psi
(–20,68 bar)
March 2018
3051L
N/A N/A
N/A N/A
Rosemount
3051L gage
–250.00 inH2O
(–621,60 mbar)
0.50 psia
(34,47 mbar)
0.50 psia
(34,47 mbar)
(3)
1. Range 0 only available with Rosemount 3051CD. Range 1 only available with Rosemount 3051CD, 3051CG, or 3051CF. inH2O referenced at 68 °F.
2. For outputs options W and M, minimum span are: Range 1 - 0.50 inH 4 - 3.00 psi (0,21 bar), range 5 - 20.00 psi (1,38 bar).
3. Assumes atmospheric pressure of 14.7 psig.
Table 8. Rosemount 3051CA and 3051T Range and Sensor Limits
Rosemount 3051CA
Minimum span
Range
1
2
3
4
5 N/A N/A N/A 5
6 N/A N/A N/A 6
0.30 psi
(20,68 mbar)
1.00 psi
(68,94 mbar)
5.33 psi
(367,49 mbar)
26.67 psi
(1,83 bar)
(1)
Range and sensor limits Minimum span
Upper
(URL)
30 psia
(2,06 bar)
150 psia
(10,34 bar)
800 psia
(55,15 bar)
4000 psia
(275,79 bar)
Lower
(LRL)
0 psia
(0 bar)
0 psia
(0 bar)
0 psia
(0 bar)
0 psia
(0 bar)
O (1,24 mbar), range 2 - 2.50 inH2O (6,21 mbar), range 3 - 10.00 inH2O (24,86 mbar), range
2
Rosemount 3051T
(1)
Range and sensor limits
Range
1
2
3
(367,49 mbar)
4
Upper
0.30 psi
(20,68 mbar)
1.00 psi
(68,94 mbar)
5.33 psi
26.67 psi
(1,83 bar)
2000.00 psi
(137,89 bar)
4000.00 psi
(275,79 bar)
(URL)
30.00 psi
(2,06 bar)
150.00 psi
(10,34 bar)
800.00 psi
(55,15 bar)
4000.00 psi
(275,79 bar)
10000.00 psi (689,47 bar)
20000.00 psi
(1378,95 bar)
Lower
(LRL) (absolute)
0 psia
(0 bar)
0 psia
(0 bar)
0 psia
(0 bar)
0 psia
(0 bar)
0 psia
(0 bar)
0 psia
(0 bar)
(2)
Lower
(LRL) (gage)
–14.70 psig
(–1,01 bar)
–14.70 psig
(–1,01 bar)
–14.70 psig
(–1,01 bar)
–14.70 psig
(–1,01 bar)
–14.70 psig
(–1,01 bar)
–14.70 psig
(–1,01 bar)
1. For output options W and M, minimum span are: range 2 – 1.50 psi (0,10 bar), range 3 – 8.00 psi (0,55 bar), range 4 – 40.00 psi (2,75 bar), range 5 for Rosemount 3051T – 2000.00 psi (137,89 bar).
2. Assumes atmospheric pressure of 14.7 psig.
48
Emerson.com/Rosemount
March 2018
Communication requires a minimum loop resistance of 250 ohms.
1. For CSA approval, power supply must not exceed 42.4 V.
Max. loop resistance = 43.5 (power supply voltage – 10.5)
Voltage (Vdc )
Load (Ωs)
Operating
region
1387
1000
500
0
10.5 20 30
42.4
(1)
Rosemount 3051
Service
Liquid, gas, and vapor applications.
4–20 mA HART (output code A)
Power supply
External power supply required. Standard transmitter (4–20 mA) operates on 10.5–42.4 Vdc with no load.
Load limitations
Maximum loop resistance is determined by the voltage level of the external power supply described by:
Indication
Optional 2-line LOI/LCD display
Optional configuration buttons
Configuration buttons need to be specified:
Digital zero trim (option code DZ) changes digital value of the
transmitter and is used for performing a sensor zero trim.
Analog zero span (option code D4) changes analog value and
can be used to rerange the transmitter with an applied pressure.
Output
graphical, task-based interface that provides single-click access to critical process/device information and descriptive graphical troubleshooting.
LOI
The LOI utilizes a 2-button menu with internal and external configuration buttons. Internal buttons are always configured for LOI. External buttons can be configured for either LOI (option code M4), Analog Zero and Span (option code D4) or Digital Zero Trim (option code DZ). See Rosemount 3051 Reference
Manual for LOI configuration menu.
FOUNDATION Fieldbus Protocol (output code F)
Power supply
External power supply required; transmitters operate on 9.0 to
32.0 Vdc transmitter terminal voltage. FISCO transmitters operate on 9.0 to 17.5 Vdc.
Current draw
17.5 mA for all configurations (including LCD display option)
Indication
Optional 2-line LCD display
OUNDATION Fieldbus block execution times
F
Block Execution time
Resource N/A
Sensor and SPM Transducer N/A
LCD Display N/A
Analog Input 1, 2 20 milliseconds
PID 25 milliseconds
Input Selector 20 milliseconds
Arithmetic 20 milliseconds
Signal Characterizer 20 milliseconds
Integrator 20 milliseconds
Output Splitter 20 milliseconds
Control Selector 20 milliseconds
2-wire 4–20mA, user selectable for linear or square root output. Digital process variable superimposed on 4–20 mA signal, available to any host that conforms to HART Protocol. The Rosemount 3051 comes with Selectable HART Revisions. Digital communications based on HART Revision 5 (default) or Revision 7 (option code HR7) can be selected. The HART revision can be switched in the field using any HART based configuration tool or the optional LOI (M4).
Power advisory diagnostics
Power advisory diagnostics pro-actively detect and notify you of degraded electrical loop integrity before it can affect your process operation. Example loop problems that can be detected include water in the terminal compartment, corrosion of terminals, improper grounding, and unstable power supplies. The device dashboard presents the diagnostics in a
Emerson.com/Rosemount
OUNDATION Fieldbus parameters
F
Links 25 (max.)
Virtual communications relationships (VCR) 20 (max.)
F
OUNDATION Fieldbus function blocks (option A01)
Resource block
The resource block contains diagnostic, hardware, and electronics information. There are no linkable inputs or outputs to the resource block.
Sensor transducer block
The sensor transducer block contains sensor information and the ability to calibrate the pressure sensor or recall factory calibration.
49
Rosemount 3051
March 2018
LCD transducer block
The LCD display transducer block is used to configure the LCD display meter.
Analog input block
The analog input (AI) function block processes the measurements from the sensor and makes them available to other function blocks. The output value from the AI block is in engineering units and contains a status indicating the quality of the measurement. The AI Block is widely used for scaling functionality.
Input selector block
The input selector (ISEL) function block can be used to select the first good, hot backup, maximum, minimum, or average of as many as eight input values and place it at the output. The block supports signal status propagation.
Integrator block
The integrator (INT) function block integrates one or two variables over time. The block compares the integrated or accumulated value to pre-trip and trip limits and generates discrete output signals when the limits are reached. The INT function block is used as a totalizer. This block will accept up to two inputs, has six options how to totalize the inputs, and two trip outputs.
Arithmetic block
The arithmetic (ARTH) function block provides the ability to configure a range extension function for a primary input. It can also be used to compute nine different arithmetic functions including flow with partial density compensation, electronic remote seals, hydrostatic tank gaging, ratio control, and others.
Signal characterizer block
The signal characterizer (SGCR) function block characterizes or approximates any function that defines an input/output relationship. The function is defined by configuring as many as 20 X,Y coordinates. The block interpolates an output value for a given input value using the curve defined by the configured coordinates. Two separate analog input signals can be processed simultaneously to give two corresponding separate output values using the same defined curve.
PID block
The PID function block combines all of the necessary logic to perform proportional/integral/derivative (PID) control. The block supports mode control, signal scaling and limiting, feed forward control, override tracking, alarm limit detection, and signal status propagation.
Control selector block
The control selector function block selects one of two or three inputs to be the output. The inputs are normally connected to the outputs of PID or other function blocks. One of the inputs would be considered normal and the other two overrides.
Output splitter block
The output splitter function block provides the capability to drive two control outputs from a single input. It takes the output of one PID or other control block to control two valves or other actuators.
Backup Link Active Scheduler (LAS)
The transmitter can function as a Link Active Scheduler if the current link master device fails or is removed from the segment.
OUNDATION Fieldbus Diagnostics Suite
F (option code D01)
The Rosemount 3051C FOUNDATION Fieldbus Diagnostics Suite features SPM technology to detect changes in the process, process equipment, or installation conditions (such as plugged impulse lines) of the transmitter. This is done by modeling the process noise signature (using the statistical values of mean and standard deviation) under normal conditions and then comparing the baseline values to current values over time. If a significant change in the current values is detected, the transmitter can generate an alert.
PROFIBUS PA Protocol (output code W)
Profile version
3.02
Power supply
External power supply required; transmitters operate on 9.0 to
32.0 Vdc transmitter terminal voltage. FISCO transmitters operate on 9.0 to 17.5 Vdc.
Current draw
17.5 mA for all configurations (including LCD display option)
Output update rate
Four times per second
Standard function blocks
Analog input (AI block)
The AI function block processes the measurements and makes them available to the host device. The output value from the AI block is in engineering units and contains a status indicating the quality of the measurement.
Physical block
The physical block defines the physical resources of the device including type of memory, hardware, electronics and diagnostic information.
Transducer block
Contains actual sensor measurement data including the sensor diagnostics and the ability to trim the pressure sensor or recall factory defaults.
50
Indication
Optional 2-line LCD display
Emerson.com/Rosemount
March 2018
Rosemount 3051
LOI
The LOI utilizes a 2-button menu with external configuration buttons.
Wireless (output code X)
Output
IEC 62591 (WirelessHART), 2.4 GHz DSSS
Wireless radio (internal antenna, WP5 option)
Frequency: 2.400 – 2.485 GHz
Channels: 15
Modulation: IEEE 802.15.4 compliant DSSS
Transmission: Maximum of 10 dBm EIRP
Local display
The optional 3-line, 7-digit LCD display can display user-selectable information such as primary variable in engineering units, scaled variable, percent of range, sensor module temperature, and electronics temperature. The display updates based on the wireless update rate.
Digital zero trim
Digital zero trim (option DZ) is an offset adjustment to compensate for mounting position effects, up to 5% of URL.
Power module
Field replaceable, keyed connection eliminates the risk of incorrect installation, Intrinsically Safe Lithium-thionyl chloride Power Module with PBT/PC enclosure. 10-year life at one minute update rate.
1. Reference conditions are 70 °F (21 °C), and routing data for three additional
(1)
network devices.
Note
Continuous exposure to ambient temperature limits of –40 °F or 185 °F (–40 °C or 85 °C) may reduce specified life by less than 20 percent.
Low power output
1–5 Vdc HART low power (output code M)
Output
3-wire 1–5 Vdc (option code C2) user-selectable output. Also user selectable for linear or square root output configuration. Digital process variable superimposed on voltage signal, available to any host conforming to the HART Protocol. Low-power transmitter operates on 6–12 Vdc with no load.
Update rate
User selectable 1 sec. to 60 min.
Wireless sensor module for in-line transmitters
The Rosemount 3051 Wireless Transmitter requires the engineered polymer housing to be selected. The standard sensor module will come with aluminum material. If SST is required, the option WSM must be selected.
Emerson.com/Rosemount
51
Rosemount 3051
Ω
March 2018
Power consumption
3.0 mA, 18–36 mW
Minimum load impedance
100 k (V
out
wiring)
Indication
Optional 5-digit LCD display
Overpressure limits
Rosemount 3051CD/CG/CF
Range 0: 750 psi (51,71 bar)
Range 1: 2000 psig (137,90 bar)
Ranges 2–5: 3626 psig (250,00 bar)
4500 psig (310,26 bar) for option code P9
Rosemount 3051CA
Range 1: 750 psia (51,71 bar)
Range 2: 1500 psia (103,42 bar)
Range 3: 1600 psia (110,32 bar)
Range 4: 6000 psia (413,69 bar)
Rosemount 3051TG/TA
Range 1: 750 psi (51,71 bar)
Range 2: 1500 psi (103,42 bar)
Range 3: 1600 psi (110,32 bar)
Range 4: 6000 psi (413,69 bar)
Range 5: 15000 psi (1034,21 bar)
Range 6: 24000 psi (1654,74 bar)
For Rosemount 3051L or level flange option codes FA, FB, FC, FD, FP, and FQ, limit is 0 psia to the flange rating or sensor rating, whichever is lower.
Table 9. Rosemount 3051L and Level Flange Rating Limits
Standard Typ e CS rating SST rating
ANSI/ASME Class 150 285 psig 275 psig ANSI/ASME Class 300 740 psig 720 psig ANSI/ASME Class 600 1480 psig 1440 psig
At 100 °F (38 °C), the rating decreases
with increasing temperature, per ANSI/ASME B16.5.
DIN PN 10–40 40 bar 40 bar DIN PN 10/16 16 bar 16 bar DIN PN 25/40 40 bar 40 bar
At 248 °F (120 °C), the rating decreases
with increasing temperature, per DIN 2401.
Static pressure limit
Burst pressure limits
Rosemount 3051C, 3051CF Coplanar or Traditional process flange
10081 psig (695,06 bar)
Rosemount 3051T In-Line
Ranges 1–4: 11016 psi (759,53 bar)
Range 5: 26016 psig (1793,74 bar)
Range 6: 46092 psi (3177,93 bar)
Failure mode alarm
HART 4–20 mA (output option code A)
If self-diagnostics detect a sensor or microprocessor failure, the analog signal is driven either high or low to alert the user. High or low failure mode is user-selectable with a jumper/switch on the transmitter. The values to which the transmitter drives its output in failure mode depend on whether it is configured to standard, NAMUR-compliant, or custom levels (see alarm configuration below). The values for each are as follows:
High alarm Low alarm
Default 21.75 mA 3.75 mA
NAMUR compliant
Custom levels
1. Analog output levels are compliant with NAMUR recommendation NE 43, see option codes C4 or C5.
2. Low alarm must be 0.1 mA less than low saturation and high alarm must be 0.1 mA greater than high saturation.
Output code M
If self-diagnostics detect a gross transmitter failure, the analog signal will be driven either below 0.94 V or above 5.4 V to alert the user (below 0.75 V or above 4.4 V for Option C2). High or low alarm signal is user-selectable by internal jumper.
Output code F, W, and X
If self-diagnostics detect a gross transmitter failure, that information gets passed as an alert and a status along with the process variable.
(1)
22.5 mA 3.6 mA
(2)
20.2 – 23.0 mA 3.4 – 3.8 mA
Temperature limits
Ambient
(1)
–40 to 185 °F (–40 to 85 °C)
With LCD display
(2)(3)
: –40 to 176 °F (–40 to 80 °C)
Rosemount 3051CD only
Operates within specifications between static line pressures of
0.5 psia and 3626 psig (4500 psig [310, 26 bar] for option code P9). Range 0: 0.5 psia and 750 psig (0,03 bar and 51,71 bar)
Range 1: 0.5 psia and 2000 psig (0,03 bar and 137, 90 bar)
52
1. For BR5 option code ambient temperature limits are –58 to 185 °F (–50 to 85 °C) and for BR6 option code ambient temperature limits are –76 to 185 °F (–60 to 85 °C).
2. For the output code M and W, LCD display may not be readable and LCD display updates will be slower at temperatures below –22 °F (–30 °C).
3. Wireless LCD display may not be readable and LCD display updates will be slower at temperature below –4
°F (–20 °C).
Emerson.com/Rosemount
March 2018
Rosemount 3051
Storage
1. If storage temperature is above 85 °C, perform a sensor trim prior to instal-
(1)
–76 to 230 °F (–60 to 110 °C)
With LCD display: –76 to 185 °F (–60 to 85 °C)
With wireless output: –40 to 185 °F (–40 to 85 °C)
lation.
Process
At atmospheric pressures and above. See Table 10 .
Table 10. Rosemount 3051 Process Temperature Limits
Rosemount 3051CD, 3051CG, 3051CF, 3051CA
(1)(4)
(1)
–40 to 185 °F (–40 to 85 °C)
(5)(6)
Silicone fill sensor
with Coplanar flange –40 to 250 °F (–40 to 121 °C)
with Traditional flange –40 to 300 °F (–40 to 149 °C)
with Level flange –40 to 300 °F (–40 to 149 °C)
with 305 Integral Manifold –40 to 300 °F (–40 to 149 °C)
Inert fill sensor
Rosemount 3051T (process fill fluid)
Silicone fill sensor
Inert fill sensor
(1)
(1)
–40 to 250 °F (–40 to 121 °C)
–22 to 250 °F (–30 to 121 °C)
(2)
(2)(3)
(2)
(2)
(2)
(2)
Humidity limits
0–100% relative humidity
Turn-on time
Performance within specifications less than 2.0 seconds (20.0 seconds for PROFIBUS PA and F after power is applied to the transmitter.
1. Does not apply to wireless option code X.
OUNDATION Fieldbus Protocols)
(1)
Volumetric displacement
Less than 0.005-in3 (0,08 cm3)
Damping
4–20 mA HART Protocol
Analog output response to a step input change is user-enterable from 0.0 to 60 seconds for one time constant. This software damping is in addition to sensor module response time.
OUNDATION Fieldbus Protocol
F
Transducer block: User configurable
AI Block: User configurable
PROFIBUS PA Protocol
AI block only: User configurable
Rosemount 3051L low-side temperature limits
Silicone fill sensor
Inert fill sensor
(1)
(1)
–40 to 250 °F (–40 to 121 °C)
–40 to 185 °F (–40 to 85 °C)
(2)
(5)
Rosemount 3051L high-side temperature limits
(process fill fluid)
SYLTHERM XLT –102 to 293 °F (–75 to 145 °C)
D.C. Silicone 704 32 to 401 °F (0 to 205 °C)
D.C. Silicone 200 –49 to 401 °F (–45 to 205 °C)
Inert –49 to 320 °F (–45 to 160 °C)
Glycerin and water 5 to 203 °F (–15 to 95 °C)
Neobee M-20 5 to 401 °F (–15 to 205 °C)
Propylene glycol and Water
1. Process temperatures above 185 °F (85 °C) require derating the ambient limits by a 1.5:1 ratio.
2. 220 °F (104 °C) limit in vacuum service; 130 °F (54 °C) for pressures below
0.5 psia.
3. Rosemount 3051CD0 process temperature limits are –40 to 212 °F (–40 to 100 °C).
4. Inert fill with traditional flange on Range 0: limits are 32 to 185 °F (0 to 85 °C).
5. 160 °F (71 °C) limit in vacuum service.
6. Not available for Rosemount 3051CA.
5 to 203 °F (–15 to 95 °C)

Physical specifications

Material selection
Emerson provides a variety of Rosemount products with various product options and configurations including materials of construction that can be expected to perform well in a wide range of applications. The Rosemount product information presented is intended as a guide for the purchaser to make an appropriate selection for the application. It is the purchaser’s sole responsibility to make a careful analysis of all process parameters (such as all chemical components, temperature, pressure, flow rate, abrasives, contaminants, etc.), when specifying product materials, options, and components for the particular application. Emerson is not in a position to evaluate or guarantee the compatibility of the process fluid or other process parameters with the product options, configuration, or materials of construction selected.
Electrical connections
1
/2–14 NPT, G1/2, and M20 x 1.5 conduit. The polymer housing code P) has no conduit entries. HART interface connections fixed to terminal block for output code A and to 701P Power Module for output code X.
Process connections
Rosemount 3051C
•1/4–18 NPT on 21/8-in. centers
1
/2–14 NPT on 2-, 21/8-, or 21/4-in. centers
Emerson.com/Rosemount
53
Rosemount 3051
March 2018
Rosemount 3051L
High pressure side: 2-, 3-, or 4-in., ASME B 16.5 (ANSI) Class 150,
300, or 600 flange; 50, 80, or 100 mm, PN 40 or 10/16 flange
Low pressure side:
1
/4–18 NPT on flange 1/2–14 NPT on adapter
Rosemount 3051T
•1/2–14 NPT female
1
/2 A DIN 16288 Male (range 1–4 only)
G
Autoclave type F-250-C (pressure relieved
1
/4 OD high pressure tube 60° cone; available for range 5–6
transmitters only).
9
/16–18 gland thread;
Rosemount 3051CF
For Rosemount 3051CFA, see Rosemount 485 Annubar Product Data Sheet.
For Rosemount 3051CFC, see Rosemount 405 Compact Orifice
Plate Product Data Sheet
For Rosemount 3051CFP, see Rosemount 1195 Integral Orifice Product Data Sheet
.
.
Process-wetted parts
Drain/vent valves
316 SST, alloy C-276, or alloy 400 material (Alloy 400 not available with Rosemount 3051L)
Process flanges and adapters
Plated CS
SST: CF-8M (Cast 316 SST) per ASTM A743
Cast C-276: CW-12MW per ASTM A494
Cast Alloy 400: M-30C per ASTM A494
Wetted O-rings
Glass-filled PTFE or graphite-filled PTFE
Process isolating diaphragms
Model
Isolating diaphragm material
3051CD
3051CG
3051T
316L SST (UNS S31603) • Alloy C-276 (UNS N10276) • Alloy 400 (UNS N04400) N/A • Tantalum (UNS R05440) N/A N/A Gold-plated Alloy 400 N/A • Gold-plated 316L SST
Rosemount 3051L process wetted parts
Flanged process connection (transmitter high side)
Process diaphragms, including process gasket surface
316L SST, Alloy C-276, or Tantalum
Extension
CF-3M (Cast version of 316L SST, material per ASTM-A743), or Alloy C-276. Fits schedule 40 and 80 pipe.
Mounting flange
Zinc-cobalt plated CS or SST
Reference process connection (transmitter low side)
Isolating diaphragms
316L SST or Alloy C-276
Reference flange and adapter
CF-8M (cast version of 316 SST, material per ASTM-A743)
Non-wetted parts
Electronics housing
Low-copper aluminum or CF-8M (cast version of 316 SST) Enclosure type 4X, IP 65, IP 66, IP 68 Housing material code P: PBT/PC with NEMA® 4X and IP66/67/68
Coplanar sensor module housing
SST: CF-3M (Cast 316L SST)
Bolts
Plated CS per ASTM A449, Type 1
Austenitic 316 SST per ASTM F593
ASTM A193, Grade B7M alloy steel
Alloy K-500
Sensor module fill fluid
Coplanar: Silicone or Inert Halocarbon
In-line: Silicone or Fluorinert
FC-43
Process fill fluid (Rosemount 3051L only)
SYLTHERM XLT, D.C. Silicone 704, D.C. Silicone 200, inert, glycerin and water, Neobee M-20, or propylene glycol and water
Paint
Polyurethane
Cover O-rings
Buna-N
3051CA
Silicone (for wireless option code X)
Power module
Field replaceable, keyed connection eliminates the risk of incorrect installation, Intrinsically Safe Lithium-thionyl chloride Power Module with PBT enclosure.
Shipping weights
Table 11. Transmitter Weights without Options
Tra ns mitte r
3051C 6.0 (2,7) 3.9 (1,8) 3051T 3.0 (1,4) 1.9 (0,86) 3051L Tabl e 12 Ta ble 1 2
1. Transmitter weights include the sensor module and housing only (aluminum for Rosemount 3051 and polymer for wireless).
3051 In lb. (kg) Wireless In lb. (kg)
(1)
54
Emerson.com/Rosemount
March 2018
Table 12. Rosemount 3051L Weights without Options
Flange
2-in., Class 150
3-in., Class 150
4-in., Class 150
2-in., Class 300
3-in., Class 300
4-in., Class 300
2-in., Class 600
3-in., Class 600
DN 50/ PN 40
DN 80/ PN 40
DN 100/ PN 10/16
DN 100/ PN 40
Flush
lb. (kg)
12.5 (5,7) N/A N/A N/A
17.5 (7,9) 19.5 (8,8) 20.5 (9,3) 21.5 (9,7)
23.5 (10,7) 26.5 (12,0) 28.5 (12,9) 30.5 (13,8)
17.5 (7,9) N/A N/A N/A
22.5 (10,2) 24.5 (11,1) 25.5 (11,6) 26.5 (12,0)
32.5 (14,7) 35.5 (16,1) 37.5 (17,0) 39.5 (17,9)
15.3 (6,9) N/A N/A N/A
25.2 (11,4) 27.2 (12,3) 28.2 (12,8) 29.2 (13,2)
13.8 (6,2) N/A N/A N/A
19.5 (8,8) 21.5 (9,7) 22.5 (10,2) 23.5 (10,6)
17.8 (8,1) 19.8 (9,0) 20.8 (9,5) 21.8 (9,9)
23.2 (10,5) 25.2 (11,5) 26.2 (11,9) 27.2 (12,3)
2-in. Ext.
lb. (kg)
4-in. Ext.
lb. (kg)
6-in. Ext.
lb. (kg)
Rosemount 3051
Table 13. Transmitter Option Weights
Code Option
J, K, L, M SST housing (T) 3.9 (1,8) J, K, L, M SST housing (C, L, H, P) 3.1 (1,4) M4/M5 LCD display for wired transmitter 0.5 (0,2) M5 LCD display for wireless output 0.1 (0,04)
B4
B1, B2, B3
B7, B8, B9
BA, BC SST bracket for traditional flange 2.3 (1,0) H2 Traditional flange 2.4 (1,1) H3 Traditional flange 2.7 (1,2) H4 Traditional flange 2.6 (1,2) H7 Traditional flange 2.5 (1,1) FC Level flange—3-in., Class 150 10.8 (4,9) FD Level flange—3-in., Class 300 14.3 (6,5) FA Level flange—2-in., Class 150 10.7 (4,8) FB Level flange—2-in., Class 300 14.0 (6,3) FP DIN level flange, SST, DN 50, PN 40 8.3 (3,8) FQ DIN level flange, SST, DN 80, PN 40 13.7 (6,2) WSM SST sensor module 1.0 (0,45)
SST mounting bracket for coplanar flange
Mounting bracket for traditional flange
Mounting bracket for traditional flange
Power Module (701PGNKF) 0.4 (0,18)
Add
lb. (kg)
1.0 (0,5)
2.3 (1,0)
2.3 (1,0)
Emerson.com/Rosemount
55

Rosemount 3051

Product certifications

March 2018
Rosemount 3051
Rev 1.13
European Directive Information
A copy of the EU Declaration of Conformity can be found at the end of the Quick Start Guide. The most recent revision of the EU Declaration of Conformity can be found at
Emerson.com/Rosemount
Ordinary Location Certification
As standard, the transmitter has been examined and tested to determine that the design meets the basic electrical, mechanical, and fire protection requirements by a nationally recognized test laboratory (NRTL) as accredited by the Federal Occupational Safety and Health Administration (OSHA).
North America
E5 USA Explosionproof (XP) and Dust-Ignitionproof (DIP)
Range 1–5
Certificate:FM16US0121
Standards: FM Class 3600 - 2011, FM Class 3615 - 2006, FM
Class 3810 - 2005, ANSI/NEMA 250 - 2003
Markings: XP CL I, DIV 1, GP B, C, D; DIP CL II, DIV 1, GP E, F,
G; CL III; T5(–50 °C ≤ T Sealed; Type 4X
Range 6
Certificate: 1053834
Standards: ANSI/ISA 12.27.01-2003,
CSA Std. C22.2 No. 30 -M1986, CSA Std. C22.2 No.142-M1987, CSA Std. C22.2 No. 213 - M1987
Markings: XP Class I, Division 1, Groups B, C and D, T5,
(–50 °C ≤ T 1, Group IIB+H2, T5; DIP Class II and Class III, Division 1, Groups E, F and G, T5, (–50 °C ≤ T Single Seal (See drawing 03031-1053)
I5 FM Intrinsic Safety (IS) and Nonincendive (NI)
Range 1–5
Certificate:FM16US0120X
Standards: FM Class 3600 - 2011, FM Class 3610 - 2010,
FM Class 3611 - 2004, FM Class 3810 - 2005, ANSI/NEMA 250 - 2008
Markings: IS CL I, DIV 1, GP A, B, C, D; CL II, DIV 1, GP E, F,
G; Class III; DIV 1 when connected per Rosemount drawing 03031-1019; NI CL 1, DIV 2, GP A, B, C, D; T4(–50 °C ≤ Ta +70 °C) [HART], T5(–50 °C T4(–50 °C ≤ Ta +60 °C) [Fieldbus/PROFIBUS]; Typ e 4x
.
+85 °C); Factory
a
+85 °C) Suitable for Class I, Zone
a
+85 °C);Type 4X; Factory Sealed;
a
T
+40 °C) [HART];
a
Special Conditions for Safe Use (X):
1. The Rosemount 3051 Transmitter housing contains aluminum and is considered a potential risk of ignition by impact or friction. Care must be taken into account during installation and use to prevent impact and friction.
2. The Rosemount 3051 Transmitter with the transient terminal block (option code T1) will not pass the 500 Vrms dielectric strength test and this must be taken into account during installation.
Range 6
Certificate: 1053834
Standards: ANSI/ISA 12.27.01-2003, CSA Std. C22.2
No.142-M1987, CSA Std. C22.2. No.157-92
Markings: IS Class I, II, III, Division 1 Groups A, B, C, D, E, F,
and G when connected in accordance with Rosemount drawing 03031-1024, Suitable for Class I, Zone 0 Group IIC; Class I, Division 2, Groups A, B, C and D; NIFW; Suitable for Class I Zone 2, Group IIC; HART: T4 (–60 °C ≤ T T5 (–60 °C ≤ T Fieldbus/PROFIBUS: T4 (–60 °C ≤ Ta 60 °C) Type 4X; Factory Sealed; Single Seal (See drawing 03031-1053)
IE USA FISCO
Range 1–5
Certificate:FM16US0120X
Standards: FM Class 3600 - 2011, FM Class 3610 - 2010,
FM Class 3611 - 2004, FM Class 3810 - 2005
Markings: IS CL I, DIV 1, GP A, B, C, D when connected per
Rosemount drawing 03031-1019 (–50 °C ≤ T
Special Conditions for Safe Use (X):
1. The Rosemount 3051 Transmitter housing contains aluminum and is considered a potential risk of ignition by impact or friction. Care must be taken into account during installation and use to prevent impact and friction.
2. The Rosemount 3051 Transmitter with the transient terminal block (option code T1) will not pass the 500 Vrms dielectric strength test and this must be taken into account during installation.
Range 6
Certificate: 1053834
Standards: ANSI/ISA 12.27.01-2003, CSA Std. C22.2
No.142-M1987, CSA Std. C22.2. No.157-92
Markings: IS Class I, Division 1 Groups A, B, C, D,
T4 (–60 °C ≤ T accordance with Rosemount drawing 03031-1024, Suitable for Class I, Zone 0 Group IIC; Type 4X; Factory Sealed; Single Seal (See drawing 03031-1053)
+60 °C); Type 4x
a
70 °C);
a
40 °C)
a
≤ +60 °C) when connected in
a
56
Emerson.com/Rosemount
March 2018
Rosemount 3051
C6 Canada Explosionproof, Dust-Ignitionproof, Intrinsic Safety
and Nonincendive
Certificate: 1053834
Standards: ANSI/ISA 12.27.01-2003,
CSA Std. C22.2 No. 30 -M1986, CSA Std. C22.2 No.142-M1987, CSA Std. C22.2. No.157-92, CSA Std. C22.2 No. 213 - M1987
Markings: Explosionproof for Class I, Division 1, Groups B,
C and D; Suitable for Class I, Zone 1, Group IIB+H2, T5 (–50 °C ≤ Ta 85 °C); Dust-Ignitionproof Class II, III, Division 1, Groups E, F, G, T5 (–50 °C ≤ T
85 °C); Class III
a
Division 1; Intrinsically Safe Class I, Division 1 Groups A, B, C, D when connected in accordance with Rosemount drawing 03031-1024, Temperature Code T4; Suitable for Class I, Zone 0; Class I Division 2 Groups A, B, C and D, T5 (–50 °C ≤ T
85 °C); Suitable for
a
Class I Zone 2, Group IIC; Type 4X; Factory Sealed; Single Seal (See drawing 03031-1053)
E6 Canada Explosionproof, Dust-Ignitionproof and Division 2
Certificate: 1053834
Standards: ANSI/ISA 12.27.01-2003,
CSA Std. C22.2 No. 30 -M1986, CSA Std. C22.2 No.142-M1987, CSA Std. C22.2 No. 213 - M1987
Markings: Explosionproof Class I, Division 1, Groups B, C
and D; Suitable for Class I, Zone 1, Group IIB+H2, T5(–50°C ≤ Ta 85°C); Dust-Ignitionproof for Class II and Class III, Division 1, Groups E, F and G; T5 (–50 °C ≤ T
85 °C); Class I, Division 2,
a
Groups A, B, C and D; T5 (–50 °C ≤ Ta 85 °C); Suitable for Class I Zone 2, Group IIC; Type 4X; Factory Sealed; Single Seal (See drawing 03031-1053)
Special Conditions for Safe Use (X):
1. This device contains a thin wall diaphragm. Installation, maintenance and use shall take into account the environmental conditions to which the diaphragm will be subjected. The manufacturer's instructions for installation and maintenance shall be followed in detail to assure safety during its expected lifetime.
2. Flameproof joints are not intended for repair.
3. Non-standard paint options may cause risk from electrostatic discharge. Avoid installations that could cause electrostatic build-up on painted surfaces, and only clean the painted surfaces with a damp cloth. If paint is ordered through a special option code, contact the manufacturer for more information.
4. Some variants of the equipment have reduced markings on the nameplate. Refer to the Certificate for full equipment marking.
I1 ATEX Intrinsic Safety and Dust
Certificate: BAS97ATEX1089X; Baseefa11ATEX0275X
Standards: EN60079-0:2012, EN60079-11:2012,
EN60079-31:2009
Markings: HART: Ex II 1 G Ex ia IIC T5/T4 Ga,
T5(–60 °C ≤ T
+40 °C),
a
T4(–60 °C ≤ Ta +70 °C) Fieldbus/PROFIBUS: Ex II 1 G Ex ia IIC Ga T4(–60°C ≤ T DUST: Ex II 1 D Ex ta IIIC T95 °C T
+60°C)
a
500
105 °C Da (–20 °C ≤ Ta +85 °C)
Table 15. Input Parameters
Parameter HART Fieldbus/PROFIBUS
Voltage U
Current I
Power P
i
i
i
30 V 30 V
200 mA 300 mA
0.9 W 1.3 W
Europe
E8 ATEX Fla meproof and Dust
Certificate: KEMA00ATEX2013X; Baseefa11ATEX0275X
Standards: EN60079-0:2012 + A11:2013,
EN60079-1:2014, EN60079-26:2015, EN60079-31:2009
Markings: II 1/2 G Ex db IIC T6…T4 Ga/Gb T6
(–60 °C ≤ T T4/T5(–60 °C ≤ T II 1 D Ex ta IIIC T95 °C T (–20 °C ≤ T
Table 14. Process Temperature
Temperature class Process temperature
T6 –60 to +65 °C
T5 –60 to +80 °C
T4 –60 to +120 °C
Emerson.com/Rosemount
+70 °C),
a
+80 °C);
a
+85 °C)
a
105 °C Da
500
Capacitance C
Inductance L
0.012 μF 0 μF
i
i
0 mH 0 mH
Special Conditions for Safe Use (X):
1. The apparatus is not capable of withstanding the 500 V insulation test required by clause 6.3.12 of EN60079-11:2012. This must be taken into account when installing the apparatus.
2. The enclosure may be made of aluminum alloy and given a protective polyurethane paint finish; however care should be taken to protect it from impact or abrasion if located in Zone 0.
3. Some variants of the equipment have reduced markings on the nameplate. Refer to the Certificate for full equipment marking.
57
Rosemount 3051
March 2018
IA ATEX FISCO
Certificate:BAS97ATEX1089X
Standards: EN60079-0:2012, EN60079-11:2009
Markings: Ex II 1 G Ex ia IIC T4 Ga (–60 °C ≤ T
+60 °C)
a
Table 16. Input Parameters
Parameters FISCO
Voltage U
Current I
Power P
i
i
i
Capacitance C
Inductance L
i
i
17.5 V
380 mA
5.32 W
<5 nF
<10 μH
Special Conditions for Safe Use (X):
1. The apparatus is not capable of withstanding the 500 V insulation test required by clause 6.3.12 of EN60079-11:2012. This must be taken into account when installing the apparatus.
2. The enclosure may be made of aluminum alloy and given a protective polyurethane paint finish; however care should be taken to protect it from impact or abrasion if located in Zone 0.
N1 ATEX Type n and Dust
Certificate:BAS00ATEX3105X; Baseefa11ATEX0275X
Standards: EN60079-0:2012, EN60079-15:2010,
EN60079-31:2009
Markings: Ex II 3 G Ex nA IIC T5 Gc (–40 °C ≤ T
Ex II 1 D Ex ta IIIC T95 °C T
500
105 °C
+70 °C);
a
Da (–20 °C ≤ Ta +85 °C)
Special Conditions for Safe Use (X):
1. This apparatus is not capable of withstanding the 500 V insulation test that is required by clause 6.8.1 of EN60079-15. This must be taken into account when installing the apparatus.
2. Some variants of the equipment have reduced markings on the nameplate. Refer to the Certificate for full equipment marking.
International
E7 IECEx Flameproof and Dust
Certificate: IECEx KEM 09.0034X; IECEx BAS 10.0034X
Standards: IEC60079-0:2011, IEC60079-1:2014-06,
IEC60079-26:2014-10, IEC60079-31:2013
Markings: Ex db IIC T6…T4 Ga/Gb
T6(–60 °C ≤ T T4/T5(–60 °C ≤ T Ex ta IIIC T95 °C T (–20 °C ≤ T
+70 °C),
a
a
+85 °C)
a
+80 °C);
105 °C Da
500
Table 17. Process Temperature
Temperature class Process temperature
T6 –60 °C to +70 °C
T5
–60 °C to +80 °C
T4
Special Conditions for Safe Use (X):
1. This device contains a thin wall diaphragm. Installation, maintenance and use shall take into account the environmental conditions to which the diaphragm will be subjected. The manufacturer's instructions for installation and maintenance shall be followed in detail to assure safety during its expected lifetime.
2. Flameproof joints are not intended for repair.
3. Non-standard paint options may cause risk from electrostatic discharge. Avoid installations that could cause electrostatic build-up on painted surfaces, and only clean the painted surfaces with a damp cloth. If paint is ordered through a special option code, contact the manufacturer for more information.
4. Some variants of the equipment have reduced markings on the nameplate. Refer to the Certificate for full equipment marking.
I7 IECEx Intrinsic Safety
Certificate: IECEx BAS 09.0076X
Standards: IEC60079-0:2011, IEC60079-11:2011
Markings: HART: Ex ia IIC T5/T4 Ga,
T5(–60 °C ≤ T
+40 °C),
a
T4(–60 °C ≤ Ta +70 °C) Fieldbus/PROFIBUS: Ex ia IIC T4(–60 °C ≤ T
+60 °C)
a
Table 18. Input Parameters
Parameter HART Fieldbus/PROFIBUS
Voltage U
Current I
Power P
i
i
i
Capacitance C
Inductance L
i
i
30 V 30 V
200 mA 300 mA
0.9 W 1.3 W
0.012 μF 0 μF
0 mH 0 mH
Special Conditions for Safe Use (X):
1. If the apparatus is fitted with an optional 90V transient suppressor, it is not capable of withstanding the 500 V insulation test required by clause 6.3.12 of IEC60079-11. This must be taken into account when installing the apparatus.
58
Emerson.com/Rosemount
March 2018
Rosemount 3051
2. The enclosure may be made of aluminum alloy and given a protective polyurethane paint finish; however, care should be taken to protect it from impact or abrasion if located in Zone 0.
IECEx Mining (Special A0259)
Certificate:IECEx TSA 14.0001X
Standards: IEC60079-0:2011, IEC60079-11:2011
Markings: Ex ia I Ma (–60 °C ≤ T
Parameter HART
Voltage U
Current I
Power P
i
i
Capacitance C
Inductance L
i
i
30 V 30 V 17.5 V
200 mA 300 mA 380 mA
0.9 W 1.3 W 5.32 W
0.012 μF 0 μF <5 nF
i
0 mH 0 mH <10 μH
+70 °C)
a
Fieldbus/
PROFIBUS
FISCO
Special Conditions for Safe Use (X):
1. If the apparatus is fitted with an optional 90V transient suppressor, it is not capable of withstanding the 500 V insulation test required by IEC60079-11. This must be taken into account when installing the apparatus.
2. It is a condition of safe use that the above input parameters shall be taken into account during installation.
3. It is a condition of manufacture that only the apparatus fitted with housing, covers and sensor module housing made out of SST are used in Group I applications.
IG IECEx FISCO
Certificate: IECEx BAS 09.0076X
Standards: IEC60079-0:2011, IEC60079-11:2011
Markings: Ex ia IIC T4 Ga (–40 °C ≤ T
+70 °C)
a
Table 19. Input Parameters
Parameter Fieldbus/PROFIBUS
Voltage U
Current I
Power P
i
i
i
Capacitance C
Inductance L
i
i
17.5 V
380 mA
5.32 W
5 nF
10 μH
Special Conditions for Safe Use (X):
1. If the apparatus is fitted with an optional 90V transient suppressor, it is not capable of withstanding the 500V insulation test required by clause 6.3.12 of IEC 60079-11. This must be taken into account when installing the apparatus.
2. The enclosure may be made of aluminum alloy and given a protective polyurethane paint finish; however, care should be taken to protect it from impact or abrasion if located in Zone 0.
N7 IECEx Type n
Certificate: IECEx BAS 09.0077X
Standards: IEC60079-0:2011, IEC60079-15:2010
Markings: Ex nA IIC T5 Gc (–40 °C ≤ T
+70 °C)
a
Special Conditions for Safe Use (X):
1. The apparatus is not capable of withstanding the 500 V insulation test required by IEC60079-15. This must be taken into account when installing the apparatus.
Brazil
E2 INMETRO Flameproof
Certificate: UL-BR 13.0643X
Standards: ABNT NBR IEC60079-0:2008 + Errata 1:2011,
ABNT NBR IEC60079-1:2009 + Errata 1:2011, ABNT NBR IEC60079-26:2008 + Errata 1:2008
Markings: Ex db IIC T6…T4 Ga/Gb,
T6(–60 °C ≤ T T4/T5(–60 °C ≤ T
Special Conditions for Safe Use (X):
1. This device contains a thin wall diaphragm less than 1 mm thickness that forms a boundary between zone 0 (process connection) and zone 1 (all other parts of the equipment). The model code and datasheet are to be consulted for details of the diaphragm material. Installation, maintenance and use shall take into account the environmental conditions to which the diaphragm will be subjected. The manufacturer's instructions for installation and maintenance shall be followed in detail to assure safety during its expected lifetime.
2. Flameproof joints are not intended for repair.
3. Non-standard paint options may cause risk from electrostatic discharge. Avoid installations that could cause electrostatic build-up on painted surfaces, and only clean the painted surfaces with a damp cloth. If paint is ordered through a special option code, contact the manufacturer for more information.
I2 INMETRO Intrinsic Safety
Certificate: UL-BR 13.0584X
Standards: ABNT NBR IEC60079-0:2008 + Errata 1:2011,
ABNT NBR IEC60079-11:2009
+70 °C),
a
+80 °C)
a
Emerson.com/Rosemount
59
Rosemount 3051
March 2018
Markings: HART: Ex ia IIC T5/T4 Ga,
T5(–60 °C ≤ Ta +40 °C), T4(–60 °C ≤ Ta +70 °C) Fieldbus/PROFIBUS: Ex ia IIC T4 Ga
le
Table 20. Input Parameters
(–60 °C ≤ T
+60 °C)
a
Parameter HART Fieldbus/PROFIBUS
Voltage U
Current I
Power P
i
i
i
Capacitance C
Inductance L
i
i
30 V 30 V
200 mA 300 mA
0.9 W 1.3 W
0.012 μF 0 μF
0 mH 0 mH
Special Conditions for Safe Use (X):
1. If the equipment is fitted with an optional 90V transient suppressor, it is not capable of withstanding the 500 V insulation test required by ABNT NBR IRC 60079-11. This must be taken into account when installing the equipment.
2. The enclosure may be made of aluminum alloy and given protective polyurethane paint finish; however, care should be taken to protect it from impact or abrasion if equipment requires EPL Ga.
IB INMETRO FISCO
Certificate: UL-BR 13.0584X
Standards: ABNT NBR IEC60079-0:2008 + Errata 1:2011,
ABNT NBR IEC60079-11:2009
Markings: Ex ia IIC T4 Ga (-60 °C ≤ T
Table 21. Input Parameters
+60 °C)
a
Parameters FISCO
Voltage U
Current I
Power P
i
i
i
Capacitance C
Inductance L
i
i
17.5 V
380 mA
5.32 W
<5 nF
<10 μH
Special Conditions for Safe Use (X):
1. If the equipment is fitted with an optional 90V transient suppressor, it is not capable of withstanding the 500 V insulation test required by ABNT NBR IEC 60079-11. This must be taken into account when installing the equipment.
2. The enclosure may be made of aluminum alloy and given protective polyurethane paint finish; however, care should be taken to protect it from impact or abrasion if equipment requires EPL Ga.
China
E3 China Flameproof
Certificate: GYJ14.1041X; GYJ15.1368X [Flow Meters]
Standards: GB12476-2000; GB3836.1-2010,
GB3836.2-2010, GB3836.20-2010
Markings: Ex d IIC T6/T5 Ga/Gb, T6(–50 °C ≤ T
T5(–50 °C ≤ T
+80 °C)
a
Special Conditions for Safe Use (X):
1. The relation between ambient temperature arrange and
temperature class is as follows:
T
a
–50 °C
–50 °C
+80 °C T5
~
+65 °C T6
~
When used in a combustible dust environment, the
maximum ambient temperature is 80 °C.
2. The earth connection facility in the enclosure should be connected reliably.
3. Cable entry certified by notified body with type of protection Ex d IIC in accordance with GB3836.1-2000 and GB3836.2-2000, should be applied when installed in a hazardous location. When used in combustible dust environment, cable entry in accordance with IP66 or higher level should be applied.
4. Obey the warning “Keep tight when the circuit is alive.”
5. End users are not permitted to change any internal components.
6. During installation, use and maintenance of this product, observe the following standards: GB3836.13-1997, GB3836.15-2000, GB3836.16-2006, GB50257-1996, GB12476.2-2006, GB15577-2007.
I3 China Intrinsic Safety
Certificate: GYJ13.1362X; GYJ15.1367X [Flow Meters]
Standards: GB3836.1-2010, GB3836.4-2010,
GB3836.20-2010, GB12476.1-2000
Markings: Ex ia IIC Ga T4/T5
Special Conditions for Safe Use (X):
1. Symbol “X” is used to denote specific conditions of use: a. If the apparatus is fitted with an optional 90 V
transient suppressor, it is not capable of withstanding the 500 V insulation test for one minute. This must be taken into account when installing the apparatus.
b. The enclosure may be made of aluminum alloy and
given a protective polyurethane paint finish; however, care should be taken to protect it from impact or abrasion if located in Zone 0.
Temperature class
+65 °C),
a
60
Emerson.com/Rosemount
March 2018
Rosemount 3051
2. The relation between T code and ambient temperature range is:
Model T code Temperature range
HART
Fieldbus/PROFIBUS/
FISCO
T5 –60 °C ≤ T
–60 °C ≤ T
T4
–60 °C ≤ Ta +60 °C
+40 °C
a
+70 °C
a
3. Intrinsically Safe parameters:
Parameter HART
Voltage U
Current I
Power P
Capacitance C
Inductance L
i
i
i
200 mA 300 mA 380 mA
0.012 μF 0 μF <5 nF
i
i
30 V 30 V 17.5 V
0.9 W 1.3 W 5.32 W
0 mH 0 mH <10 μH
Fieldbus/
PROFIBUS
FISCO
Note
FISCO parameters apply to both Group IIC and IIB.
7. End users are not permitted to change any intern components but to settle the problem in conjunction with the manufacturer to avoid damage to the product.
8. During installation, use and maintenance of this product, observe the following standards: GB3836.13-1997, GB3836.15-2000, GB3836.16-2006, GB50257-1996, GB12476.2-2006, GB15577-2007.
N3 China Type n
Certificate: GYJ15.1105X
Standards: GB3836.1-2010, GB3836.8-2003
Markings: Ex nA nL IIC T5 Gc (–40 °C ≤ T
+70 °C)
a
Special Condition for Safe Use (X):
1. Symbol “X” is used to denote specific conditions of use: The apparatus is not capable of withstanding the 500 V test to earth for one minute. The must be taken into consideration during installation.
Japan
E4 Japan Flameproof
Certificate: TC20577, TC20578, TC20583, TC20584
[HART]; TC20579, TC20580, TC20581, TC20582 [Fieldbus]
Markings: Ex d IIC T5
[For Flow Meters] When Rosemount 644 Temperature Transmitter is used, the Rosemount 644 Transmitter should be used with Ex-certified associated apparatus to establish explosion protection system that can be used in explosive gas atmospheres. Wiring and terminals should comply with the instruction manual of both Rosemount 644 Transmitter and associated apparatus. The cables between Rosemount 644 Transmitter and associated apparatus should be shielded cables (the cables must have insulated shield). The shielded cable has to be grounded reliably in a non-hazardous area.
4. Transmitters comply with the requirements for FISCO field devices specified in IEC60079-27:2008. For the connection of an intrinsically safe circuit in accordance with FISCO Model, FISCO parameters are listed in the table above.
5. The product should be used with Ex-certified associated apparatus to establish explosion protection system that can be used in explosive gas atmospheres. Wiring and terminals should comply with the instruction manual of the product and associated apparatus.
6. The cables between this product and associated apparatus should be shielded cables (the cables must have insulated shield). The shielded cable has to be grounded reliably in a non-hazardous area.
Republic of Korea
EP Republic of Korea Flameproof
Certificate:11-KB4BO-0188X [Mfg Singapore]
Markings: Ex d IIC T6…T4
IP Republic of Korea Intrinsic Safety
Certificate:13-KB4BO-0203X [HART – Mfg USA],
13-KB4BO-0204X [Fieldbus – Mfg USA], 10-KB4BO-0138X [HART – Mfg Singapore], 13-KB4BO-0206X [Fieldbus – Mfg Singapore]
Markings: Ex ia IIC T5/T4 (HART)
Ex ia IIC T4 (Fieldbus)
Technical Regulations Customs Union (EAC)
EM EAC Flameproof
Markings: Ga/Gb Ex d IIC T4... T6 X,
T4/T5(–60 °C ≤ T T6(–60 °C ≤ T
Special Condition for Safe Use (X):
1. See certificate for special conditions.
+80 °C),
a
+70 °C)
a
Emerson.com/Rosemount
61
Rosemount 3051
March 2018
IM EAC Intrinsically Safe
Markings: HART: 0Ex ia IIC T4/T5 Ga X,
T4(–60 °C ≤ T T5(–60 °C ≤ T
+70 °C),
a
+40 °C)
a
Fieldbus/PROFIBUS: 0Ex ia IIC T4 Ga X (–60 °C ≤ Ta +60 °C)
Special Condition for Safe Use (X):
1. See certificate for special conditions.
Combinations
K2 Combination of E2 and I2 K5 Combination of E5 and I5 K6 Combination of C6, E8, and I1 K7 Combination of E7, I7, and N7 K8 Combination of E8, I1, and N1 KB Combination of E5, I5, and C6 KD Combination of E8, I1, E5, I5, and C6 KM Combination of EM and IM KP Combination of EP and IP
Conduit plugs and adapters
IECEx Flameproof and Increased Safety
Certificate: IECEx FMG 13.0032X
Standards: IEC60079-0:2011, IEC60079-1:2007,
IEC60079-7:2006-2007
Markings: Ex de IIC Gb
ATEX Flameproof and Increased Safety
Certificate: FM13ATEX0076X
Standards: EN60079-0:2012, EN60079-1:2007,
IEC60079-7:2007
Markings: Ex II 2 G Ex de IIC Gb
Table 22. Conduit Plug Thread Sizes
Thread Identification mark
M20 x 1.5 M20
1
/2 –14 NPT
1
/2 NPT
Special Conditions for Safe Use (X):
1. When the thread adapter or blanking plug is used with an enclosure in type of protection increased safety “e” the entry thread shall be suitably sealed in order to maintain the ingress protection rating (IP) of the enclosure.
2. The blanking plug shall not be used with an adapter.
3. Blanking Plug and Threaded Adapter shall be either NPT or Metric thread forms. G
1
/2 thread forms are only acceptable
for existing (legacy) equipment installations.
Additional Certifications
SBS American Bureau of Shipping (ABS) Type Approval
Certificate: 09-HS446883A-5-PDA
Intended Use: Marine & Offshore Applications -
Measurement of either gauge or absolute pressure for liquid, gas and vapor.
SBV Bureau Veritas (BV) Type Approval
Certificate: 23155
Requirements: Bureau Veritas Rules for the Classification of
Steel Ships
Application: Class notations: AUT-UMS, AUT-CCS,
AUT-PORT and AUT-IMS; Pressure transmitter type 3051 cannot be installed on diesel engines
SDN Det Norske Veritas (DNV) Type Approval
Certificate: TAA000004F
Intended Use: DNV GL Rules for Classification - Ships and
offshore units
Application:
Location Classes
Tem pe ra tu re D
Humidity B
Vibration A
EMC B
Enclosure D
Table 23. Thread Adapter Thread Sizes
Male Thread Identification mark
M20 x 1.5 – 6H M20
1
/2–14 NPT
3
/4 –14 NPT
1
/2 –14 NPT
3
/4 –14 NPT
Female Thread Identification mark
M20 x 1.5 – 6H M20
1
/2–14 NPT
1
G
/2 G1/2
1
/2–14 NPT
62
SLL Lloyds Register (LR) Type Approval
Certificate:11/60002
Application: Environmental categories ENV1, ENV2, ENV3,
and ENV5
C5 Custody Transfer - Measurement Canada Accuracy
Approval
Certificate: AG-0226; AG-0454; AG-0477
Emerson.com/Rosemount
March 2018
Rosemount 3051

Rosemount 3051 Wireless

Rev 1.4
European Directive Information
A copy of the EU Declaration of Conformity can be found at the end of the Quick Start Guide. The most recent revision of the EU Declaration of Conformity can be found at
Emerson.com/Rosemount
Telecommunication compliance
All wireless devices require certification to ensure that they adhere to regulations regarding the use of the RF spectrum. Nearly every country requires this type of product certification. Emerson is working with governmental agencies around the world to supply fully compliant products and remove the risk of violating country directives or laws governing wireless device usage.
FCC and IC
This device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: This device may not cause harmful interference. This device must accept any interference received, including interference that may cause undesired operation. This device must be installed to ensure a minimum antenna separation distance of 20 cm from all persons.
Ordinary Location Certification
As standard, the transmitter has been examined and tested to determine that the design meets the basic electrical, mechanical, and fire protection requirements by a nationally recognized test laboratory (NRTL) as accredited by the Federal Occupational Safety and Health Administration (OSHA).
Installing in North America
The US National Electrical Code® (NEC) and the Canadian Electrical Code (CEC) permit the use of Division marked equipment in Zones and Zone marked equipment in Divisions. The markings must be suitable for the area classification, gas, and temperature class. This information is clearly defined in the respective codes.
USA
I5 U.S.A. Intrinsically Safe (IS)
Range 1–5 Certificate: FM 3046325 Standards: FM Class 3600 - 2011, FM Class 3610 - 2010, FM
Class 3810 - 2005, ANSI/ISA 60079-0 - 2009, ANSI/ISA 60079-11 - 2009, NEMA ANSI/IEC 60529
.
®
250 - 2003,
Markings: IS CL I, DIV 1, GP A, B, C, D T4; CL 1, Zone 0 AEx ia
IIC T4; T4(–40 °C ≤ Ta +70 °C) when installed per Rosemount drawing 03031-1062; Type 4X/IP66/IP68
Special Conditions for Safe Use (X):
1. The Rosemount 3051 Wireless Pressure Transmitter shall only be used with the 701PGNKF Rosemount SmartPower Battery Pack.
2. The inline pressure sensor may contain more than 10% aluminum and is considered a potential risk of ignition by impact or friction. Care must be taken into account during installation and used to prevent impact and friction.
3. The surface resistivity of the transmitter housing is greater than one gigaohm. To avoid electrostatic charge build-up, it must not be rubbed or cleaned with solvents or a dry cloth.
Range 6 Certificate: CSA 2526009 Standards: FM Class 3600 - 2011, FM Class 3610 - 2010, FM
Class 3810 - 2005, ANSI/ISA 60079-0 - 2009, ANSI/ISA 60079-11 - 2009, UL 61010-1 (3rd edition), UL50E (1st Edition)
Markings: IS CL I, DIV 1, GP A, B, C, D T4; CL 1, Zone 0 AEx ia
IIC T4; T4(-40 °C ≤ T per Rosemount drawing 03031-1063; Type 4X/IP66/IP68
+70 °C) when installed
a
Canada
I6 Canada Intrinsically Safe
Certificate: CSA 2526009 Standards: CAN/CSA C22.2 No. 0-M91, CAN/CSA C22.2
No.94-M91, CSA Std C22.2 No. 142-M1987, CSA Std C22.2 No. 157-92, CSA Std C22.2 No. 60529:05
Markings: Intrinsically Safe for Class I, Division 1, Groups A,
B, C, D, T4 when installed per Rosemount drawing 03031-1063; Type 4X/IP66/IP68
Europe
I1 ATEX Intrinsic Safety
Certificate: Baseefa12ATEX0228X Standards:EN 60079-0: 2012, EN 60079-11: 2012 Markings: Ex II 1 G Ex ia IIC T4 Ga, T4(–40 °C ≤ T
IP66/IP68
+70 °C)
a
Emerson.com/Rosemount
63
Rosemount 3051
March 2018
Special Conditions for Safe Use (X):
1. The plastic enclosure may constitute a potential electrostatic ignition risk and must not be rubbed or cleaned with a dry cloth.
2. The Model 701PGNKF Power Module may be replaced in a hazardous area. The power module has a surface resistivity greater than 1 GΩ and must be properly installed in the wireless device enclosure. Care must be taken during transportation to and from the point of installation to prevent electrostatic charge build-up.
International
I7 IECEx Intrinsic Safety
Certificate: IECEx BAS 12.0124X Standards: IEC 60079-0: 2011, IEC 60079-11: 2011 Markings: Ex ia IIC T4 Ga, T4(–40 °C ≤ T
IP66/IP68
Special Conditions for Safe Use (X):
1. The plastic enclosure may constitute a potential electrostatic ignition risk and must not be rubbed or cleaned with a dry cloth.
2. The Model 701PGNKF Power Module may be replaced in a hazardous area. The power module has a surface resistivity greater than 1 GΩ and must be properly installed in the wireless device enclosure. Care must be taken during transportation to and from the point of installation to prevent electrostatic charge build-up.
+70 °C)
a
Special Condition for Safe Use (X):
1. See certificate for special conditions.
EAC - Belarus, Kazakhstan, Russia
IM Technical Regulation Customs Union (EAC) Intrinsic Safety
Certificate: TU RU C-US.AA87.B.00534 Markings: 0Ex ia IIC T4 Ga X; (–40 °C ≤ T
Special Condition for Safe Use (X):
1. See certificate for special conditions.
+70 °C)
a
Korea
IP Korea Intrinsic Safety
Certificate: 13-KB4BO-0295X Markings: Ex ia IIC T4 (–40 °C ≤ T
Special Condition for Safe Use (X):
1. See certificate for special conditions.
+70 °C)
a
Brazil
I2 INMETRO Intrinsic Safety
Certificate: UL-BR 13.0534X Standards: ABNT NBR IEC 60079-0:2008 + Errata 1:2011,
ABNT NBR IEC 60079-11:2009
Markings:Ex ia IIC T4 IP66 Ga, T4(–40 °C ≤ T
Special Condition for Safe Use (X):
1. See certificate for special conditions.
+70 °C)
a
China
I3 China Intrinsic Safety
Certificate: GYJ13.1362X, GYJ15.1367X [Flow Meters] Standards: GB3836.1-2010, GB3836.4-2010,
GB3836.20-2010
Markings: Ex ia IIC T4 Ga, T4(–40
Special Condition for Safe Use (X):
1. See certificate for special conditions.
+70 °C)
~
Japan
I4 TIIS Intrinsic Safety
Certificate: TC22022X (Rosemount 3051C/L), TC22023X
(Rosemount 3051T), TC22024X (Rosemount 3051CFx)
Markings: Ex ia IIC T4 Ga, T4(–20
+60 °C)
~
64
Emerson.com/Rosemount
March 2018
Rosemount 3051

Pipe I.D. range codes

For pipes with an inner diameter (I.D.) range/pipe wall thickness not found in this table or with a line size greater than 12-in. (300 mm), choose option code Z and specify the exact pipe dimensions (I.D. and pipe wall thickness) on the Configuration Data
Sheet. The Emerson sizing program will determine this code, based on the application piping.
Nominal Max. O.D.
2-in.
(50 mm)
21/2-in.
(63.5 mm)
3-in.
(80 mm)
N/A
31/2-in.
(89 mm)
4-in.
(100 mm)
5-in.
(125 mm)
Sensor
(150 mm)
size 1
6-in.
6.93-in. (176.02
Line size
2.625-in.
(66.68 mm)
3.188-in.
(80.98 mm)
3.75-in.
(95.25 mm)
4.25-in.
(107.95 mm)
5.032-in.
(127.81 mm)
6.094-in.
(154.79 mm)
mm)
Option
code
020
025
030
035
040
050
060
I.D. range
1.784 to 1.841-in.
(45.31 to 46.76 mm)
1.842 to 1.938-in.
(46.79 to 49.23 mm)
1.939 to 2.067-in.
(49.25 to 52.50 mm)
2.068 to 2.206-in.
(52.53 to 56.03 mm)
2.207 to 2.322-in.
(56.06 to 58.98 mm)
2.323 to 2.469-in.
(59.00 to 62.71 mm)
2.470 to 2.598-in.
(62.74 to 65.99 mm)
2.599 to 2.647-in.
(66.01 to 67.23 mm)
2.648 to 2.751-in.
(67.26 to 69.88 mm)
2.752 to 2.899-in.
(69.90 to 73.63 mm)
2.900 to 3.068-in.
(73.66 to 77.93 mm)
3.069 to 3.228-in.
(77.95 to 81.99 mm)
3.229 to 3.333-in.
(82.02 to 84.66 mm)
3.334 to 3.548-in.
(84.68 to 90.12 mm)
3.549 to 3.734-in.
(90.14 to 94.84 mm)
3.735 to 3.825-in.
(94.87 to 97.16 mm)
3.826 to 4.026-in.
(97.18 to 102.26 mm)
4.027 to 4.237-in.
(102.29 to 107.62 mm)
4.238 to 4.437-in.
(107.65 to 112.70 mm)
4.438 to 4.571-in.
(112.73 to 116.10 mm)
4.572 to 4.812-in.
(116.13 to 122.22 mm)
4.813 to 5.047-in.
(122.25 to 128.19 mm)
5.048 to 5.249-in.
(128.22 to 133.32 mm)
5.250 to 5.472-in.
(133.35 to 138.99 mm)
5.473 to 5.760-in.
(139.01 to 146.30 mm)
5.761 to 6.065-in.
(146.33 to 154.05 mm)
6.066 to 6.383-in.
(154.08 to 162.13 mm)
Pipe wall thickness
ANSI pipes Non-ANSI pipes
0.065 to 0.488-in. (1.7 to 12.4 mm)
0.065 to 0.449-in.
0.065 to 0.545-in. (1.7 to 13.8 mm)
0.083 to 0.563-in. (2.1 to 14.3 mm)
0.083 to 0.563-in. (2.1 to 14.3 mm)
0.120 to 0.600-in. (3.0 to 15.2 mm)
0.120 to 0.600-in. (3.0 to 15.2 mm)
0.134 to 0.614-in. (3.4 to 15.6 mm)
0.134 to 0.614-in. (3.4 to 15.6 mm)
(1.7 to 11.4 mm)
0.065 to 0.417-in. (1.7 to 10.6 mm)
0.065 to 0.407-in. (1.7 to 10.3 mm)
0.083 to 0.448-in. (2.1 to 11.4 mm)
0.083 to 0.417-in. (2.1 to 10.6 mm)
0.083 to 0.435-in. (2.1 to 11.0 mm)
0.083 to 0.515-in. (2.1 to 13.1 mm)
0.083 to 0.460-in. (2.1 to 11.7 mm)
0.083 to 0.416-in. (2.1 to 10.6 mm)
0.083 to 0.395-in. (2.1 to 10.0 mm)
0.083 to 0.404-in (2.1 to 10.3 mm)
0.120 to 0.496-in. (3.0 to 12.6 mm)
0.120 to 0.386-in.
(3.0 to 9.8 mm)
0.120 to 0.415-in. (3.0 to 10.5 mm)
0.120 to 0.510-in. (3.0 to 13.0 mm)
0.120 to 0.400-in. (3.0 to 10.2 mm)
0.120 to 0.390-in.
(3.0 to 9.9 mm)
0.120 to 0.401-in. (3.0 to 10.2 mm)
0.134 to 0.481-in. (3.4 to 12.2 mm)
0.134 to 0.374-in.
(3.4 to 9.5 mm)
0.134 to 0.380-in.
(3.4 to 9.7 mm)
0.134 to 0.413-in. (3.4 to 10.5 mm)
0.134 to 0.3919-in. (3.4 to 9.9 mm)
0.134 to 0.327-in. (3.4 to 8.3 mm)
0.134 to 0.31-in. (3.4 to 7.9 mm)
0.134 to 0.297-in. (3.4 to 7.5 mm)
I.D.
range
code
A
B
C
D
B
C
D
E
A
B
C
D
B
C
D
B
C
D
E
A
B
C
D
A
B
C
D
Emerson.com/Rosemount
65
Rosemount 3051
March 2018
Sensor
Sensor
Sensor
Sensor
Sensor
size 2
size 1
size2
size 1
size 2
N/A
6-in.
(150 mm)
7-in.
(180 mm)
7-in.
(180 mm)
8-in.
(200 mm)
8-in.
(200 mm)
10-in.
(250 mm)
12-in.
(300 mm)
6.93-in.
(176.02 mm)
7.93-in.
(201.42 mm)
7.93-in.
(201.42 mm)
9.688-in.
(246.08 mm)
9.688-in.
(246.08 mm)
11.75-in.
(298.45 mm)
13.0375-in.
(331.15 mm)
060
070
070
080
080
100
120
5.250 to 5.472-in.
(133.35 to 139.99 mm)
5.473 to 5.760-in.
(139.01 to 146.30 mm)
5.761 to 6.065-in.
(146.33 to 154.05 mm)
6.066 to 6.383-in.
(154.08 to 162.13 mm)
6.384 to 6.624-in.
(162.15 to 168.25 mm)
6.625 to 7.023-in.
(168.28 to 178.38 mm)
7.024 to 7.392-in.
(178.41 to 187.76 mm)
6.384 to 6.624-in.
(162.15 to 168.25 mm)
6.625 to 7.023-in.
(168.28 to 178.38 mm)
7.024 to 7.392-in.
(178.41 to 187.76 mm)
7.393 to 7.624-in.
(187.78 to 193.65 mm)
7.625 to 7.981-in.
(193.68 to 202.72 mm)
7.982 to 8.400-in.
(202.74 to 213.36 mm)
8.401 to 8.766-in.
(213.39 to 222.66 mm)
7.393 to 7.624-in.
(187.78 to 193.65 mm)
7.625 to 7.981-in.
(193.68 to 202.72 mm)
7.982 to 8.400-in.
(202.74 to 213.36 mm)
8.401 to 8.766-in.
(213.39 to 222.66 mm)
8.767 to 9.172-in.
(222.68 to 232.97 mm)
9.173 to 9.561-in.
(232.99 to 242.85 mm)
9.562 to 10.020-in.
(242.87 to 254.51 mm)
10.021 to 10.546-in.
(254.53 to 267.87 mm)
10.547 to 10.999-in.
(267.89 to 279.37 mm)
11.000 to 11.373-in.
(279.40 to 288.87 mm)
11.374 to 11.938-in.
(288.90 to 303.23 mm)
11.939 to 12.250-in.
(303.25 to 311.15 mm)
0.134 to 1.354-in. (3.4 to 34.4 mm)
0.134 to 0.614-in. (3.4 to 15.6 mm)
0.134 to 1.354-in. (3.4 to 34.4 mm)
0.250 to 0.73-in. (6.4 to 18.5 mm)
0.250 to 1.47-in. (6.4 to 37.3 mm)
0.250 to 1.470-in. (6.4 to 37.3 mm)
0.250 to 1.470-in. (6.4 to 37.3 mm)
0.134 to 1.132-in. (3.4 to 28.7 mm)
0.134 to 1.067-in. (3.4 to 27.1 mm)
0.134 to 1.05-in. (3.4 to 26.7 mm)
0.134 to 1.037-in. (3.4 to 26.3 mm)
0.134 to 0.374-in.
(3.4 to 9.5 mm)
0.134 to 0.216-in.
(3.4 to 5.5 mm)
0.134 to 0.246-in.
(3.4 to 6.2 mm)
0.134 to 1.114-in. (3.4 to 28.3 mm)
0.134 to 0.956-in. (3.4 to 24.3 mm)
0.134 to 0.986-in. (3.4 to 25.0 mm)
0.250 to 0.499-in. (6.4 to 12.6 mm)
0.250 to 0.374-in.
(6.4 to 9.5 mm)
0.250 to 0.312-in.
(6.4 to 7.9 mm)
0.250 to 0.364-in.
(6.4 to 9.2 mm)
0.250 to 1.239-in. (6.4 to 31.4 mm)
0.250 to 1.114-in. (6.4 to 28.3 mm)
0.250 to 1.052-in. (6.4 to 26.7 mm)
0.250 to 1.104-in. (6.4 to 28.0 mm)
0.250 to 1.065-in. (6.4 to 27.1 mm)
0.250 to 1.082-in. (6.4 to 27.5 mm)
0.250 to 1.012-in. (6.4 to 25.7 mm)
0.250 to 0.945-in. (6.4 to 24.0 mm)
0.250 to 1.018-in. (6.4 to 25.9 mm)
0.250 to 1.097-in. (6.4 to 27.9 mm)
0.250 to 0.906-in. (6.4 to 23.0 mm)
0.250 to 1.159-in. (6.4 to 29.4 mm)
A
B
C
D
B
C
D
B
C
D
B
C
D
E
B
C
D
E
A
B
C
D
E
B
C
D
66
Emerson.com/Rosemount
March 2018
A
B
E
H
K
L
I
M
J
N
O
P
Q
C
G
D
F
Rosemount 3051

Dimensional drawings

Figure 1. Rosemount 3051C Exploded View
(1)
A. Cover B. Cover O-ring C. Terminal block D. Electronics housing E. Configuration buttons cover F. Local configuration buttons
G. Electronics board H. Name plate I. Housing rotation set screw (180° maximum rotation
without further disassembly) J. Sensor module K. Coplanar flange
L. Drain/vent valve M. Flange adapters N. Process O-ring O. Flange adapter O-ring P. Flange alignment screw (not pressure retaining) Q. Flange bolts
1. This section contains dimensional drawings for output codes A, F, and X. For output codes W and M, visit
Emerson.com/Rosemount
Emerson.com/Rosemount /Documentation-and-Dra wings
.
67
Rosemount 3051
4.09
(104)
7.12
(181)
4.29
(109)
6.55
(166)
Figure 2. Rosemount 3051C Coplanar Flange
Dimensions are in inches (millimeters).
March 2018
Figure 3. Rosemount 3051 Wireless Housing with Coplanar Flange
4.196
(106.6)
Dimensions are in inches (millimeters).
5.49
(139)
7.41
(188)
68
Emerson.com/Rosemount
March 2018
6.75
(171)
5.50 (140) Max open
p
Rosemount 3051
Figure 4. Rosemount 3051C Coplanar Flange with Rosemount 305RC3 3-Valve Coplanar Integral Manifold
9.20 (234) Max O
en
Dimensions are in inches (millimeters).
Emerson.com/Rosemount
69
Rosemount 3051
2.81 (71)
4.73
(120)
6.20
(157)
A
2.81 (71)
B
3.35 (85)
6.25
(159)
March 2018
Figure 5. Coplanar Flange Mounting Configurations with Optional Bracket (B4) for 2-in. Pipe or Panel Mounting
Panel
mounting
Pipe
mounting
A. 5/16–18 bolts (not supplied)
3
B.
/8–16 bolts
Dimensions are in inches (millimeters).
C. 2-in. U-bolt
3
D.
/8–16 bolts
C
D
3.35 (85)
3.50 (89)
70
Emerson.com/Rosemount
March 2018
(54)
6.28
(159)
A
3.63 (92)
0.91
(23)
3.50 (89)
1.10 (28)
(
)
Figure 6. Rosemount 3051C Coplanar with Traditional Flange
4.09
(104)
A
1.16 (29)
A. Flange adapters (optional) Dimensions are in inches (millimeters).
3.40 (86)
1.10 (28)
Rosemount 3051
1.63 (41)
2.13
Figure 7. Rosemount 3051C Coplanar with Rosemount 305RT3 3-Valve Traditional Integral Manifold
2.13
2.52 (64)
1
A.
/2–14 NPT flange adapter (optional)
Dimensions are in inches (millimeters).
6.38 162
(54)
Emerson.com/Rosemount
71
Rosemount 3051
Figure 8. Traditional Flange Mounting Configurations with Optional Brackets for 2-in. Pipe or Panel Mounting
Panel mounting bracket (option B2/B8) Pipe mounting bracket (option B1/B7/BA)
March 2018
9.27 (235)
2.62 (67)
2.62 (67)
A
6.65 (169)
0.93 (24)
6.65 (169)
A. 5/16-18 bolts (not supplied) Dimensions are in inches (millimeters).
2.81 (71)
4.12 (105)
Pipe mounting bracket (option B3/B9/BC)
4.85 (123)
1.10 (28)
3.40 (86)
1.75 (44)
72
Emerson.com/Rosemount
March 2018
4.09
(104)
R1-4: 7.22 (183)
R5: 7.32 (186)
A
4.20
(107)
Figure 9. Rosemount 3051T
Dimensions are in inches (millimeters).
Figure 10. Rosemount 3051T Wireless Housing
Rosemount 3051
4.29
(109)
5.49
(139)
R1-4: 7.56 (192)
R5: 7.66 (195)
A. U-Bolt bracket Dimensions are in inches (millimeters).
Emerson.com/Rosemount
73
Rosemount 3051
4.13
(105)
4.88
(124)
6.25 (159) max open
Figure 11. Rosemount 3051T In-Line Range 6
In-line
7.86 (199,5)
A
A. Autoclave type F-250-C
Figure 12. Rosemount 3051T with Rosemount 306 2-Valve Integral Manifold
Wireless in-line
8.13 (206,6)
March 2018
Dimensions are in inches (millimeters).
74
Emerson.com/Rosemount
March 2018
3.49 (89)
6.15
(156)
A
B
D
C
Figure 13. Rosemount 3051T Typical Mounting Configurations with Optional Mounting Bracket
Pipe mounting Panel mounting
6.15
(156)
Dimensions are in inches (millimeters).
2.81 (71)
4.78
(121)
Rosemount 3051
Figure 14. Rosemount 3051CFA Pak-Lok Annubar Flowmeter
Front view Side view Top view
(1)
Emerson.com/Rosemount
75
Rosemount 3051
B
A
C
A
B
D
Table 24. Rosemount 3051CFA Pak-Lok Annubar Flowmeter Dimensional Data (Maximum Dimensions)
Sensor size A B C D
1 8.50 (215.9) 15.60 (396.9) 9.00 (228.6) 6.00 (152.4)
2 11.00 (279.4) 18.10 (460.4) 9.00 (228.6) 6.00 (152.4)
3 12.00 (304.8) 19.10 (485.8) 9.00 (228.6) 6.00 (152.4)
Dimensions are in inches (millimeters).
Figure 15. Rosemount 3051CFC Compact Orifice Flowmeter
Orifice plate side view Orifice plate front view Orifice plate top view
D
March 2018
Compact orifice plate
(Primary element type code P)
C
Conditioning orifice plate
(Primary element type code C)
Primary element typ e
A B Tran sm itte r h eig ht C D
Type P and C 5.62 (143) Transmitter height + A 6.27 (159)
Dimensions are in inches (millimeters).
7.75 (197) - closed
8.25 (210) - open
6.00 (152) - closed
6.25 (159) - open
1. The Rosemount Annubar pak-lok model is available up to Class 600 ANSI (1440 psig at 100 °F [99 bar at 38 °C]).
76
Emerson.com/Rosemount
March 2018
8.80
(223.46)
A
10.30
(261.81)
5.30
(134.51)
6.30
(160.55)
K
Downstream
J
Upstream
Figure 16. Rosemount 3051CFP Integral Orifice Flowmeter
Side view
Bottom view Front view
Rosemount 3051
A. B.D. (bore diameter) Dimensions are in inches (millimeters).
Dimension
J (Beveled/threaded pipe ends) 12.54 (318.4) 20.24 (514.0) 28.44 (722.4) J (RF slip-on, RTJ slip-on, RF-DIN slip on) 12.62 (320.4) 20.32 (516.0) 28.52 (724.4) J (RF Class 150, weld neck) 14.37 (364.9) 22.37 (568.1) 30.82 (782.9) J (RF Class 300, weld neck) 14.56 (369.8) 22.63 (574.7) 31.06 (789.0) J (RF Class 600, weld neck) 14.81 (376.0) 22.88 (581.0) 31.38 (797.1) K (Beveled/threaded pipe ends) 5.74 (145.7) 8.75 (222.2) 11.91 (302.6) K (RF slip-on, RTJ slip-on, RF-DIN slip on) K (RF Class 150, weld neck) 7.57 (192.3) 10.88 (276.3) 14.29 (363.1) K (RF Class 300, weld neck) 7.76 (197.1) 11.14 (282.9) 14.53 (369.2) K (RF Class 600, weld neck) 8.01 (203.4) 11.39 (289.2) 14.85 (377.2) B.D. (Bore diameter) 0.664 (16.87) 1.097 (27.86) 1.567 (39.80)
Dimensions are in inches (millimeters).
1. Downstream length shown here includes plate thickness of 0.162-in. (4.11 mm).
Emerson.com/Rosemount
(1)
Line size
1
/2-in. (15 mm) 1-in. (25 mm) 11/2-in. (40 mm)
5.82 (147.8) 8.83 (224.2) 11.99 (304.6)
77
Rosemount 3051
4.09
(104)
A
B
H
E
Figure 17. Rosemount 3051L
2-in. flange
configuration
(flush mount only)
3- and 4-in. flange configuration Diaphragm assembly and mounting flange
March 2018
D
C
A
J
B
Optional flushing connection ring (lower housing)
F
4.09
(104)
5.13
(130)
4.29
(109)
6.80
(173)
7.10
(181)
8.20
(209)
G
D
I
A–H. See Table 25 on page 79. I. Flushing connection Dimensions are in inches (millimeters).
78
J. 2-, 4-., or 6-in. extension
(only available with 3-. and 4-in., DN80, and DN100 flange configurations)
Emerson.com/Rosemount
March 2018
Table 25. Rosemount 3051L Dimensional Specifications
Rosemount 3051
(1)
Class
ASME B16.5 (ANSI) 150
ASME B16.5 (ANSI) 300
ASME B16.5 (ANSI) 600
DIN 2501 PN 10–40
DIN 2501 PN 25/40
DIN 2501 PN 10/16
Dimensions are in inches (millimeters).
1. Tolerances are 0.040 (1.02), - 0.020 (0.51).
Pipe
size
2 (51) 0.69 (18) 4.75 (121) 6.0 (152) 4 0.75 (19) N/A 3.6 (92)
3 (76) 0.88 (22) 6.0 (152) 7.5 (191) 4 0.75 (19) 2.58 (66) 5.0 (127)
4 (102) 0.88 (22) 7.5 (191) 9.0 (229) 8 0.75 (19) 3.5 (89) 6.2 (158)
2 (51) 0.82 (21) 5.0 (127) 6.5 (165) 8 0.75 (19) N/A 3.6 (92)
3 (76) 1.06 (27) 6.62 (168) 8.25 (210) 8 0.88 (22) 2.58 (66) 5.0 (127)
4 (102) 1.19 (30) 7.88 (200) 10.0 (254) 8 0.88 (22) 3.5 (89) 6.2 (158)
2 (51) 1.00 (25) 5.0 (127) 6.5 (165) 8 0.75 (19) N/A 3.6 (92)
3 (76) 1.25 (32) 6.62 (168) 8.25 (210) 8 0.88 (22) 2.58 (66) 5.0 (127)
DN 50 20 mm 125 mm 165 mm 4 18 mm N/A 4.0 (102)
DN 80 24 mm 160 mm 200 mm 8 18 mm 66 mm 5.4 (138)
DN 100 24 mm 190 mm 235 mm 8 22 mm 89 mm 6.2 (158)
DN 100 20 mm 180 mm 220 mm 8 18 mm 89 mm 6.2 (158)
Flange
thickness
A
Bolt circle
diameter
H
Outside
diameter
E
No. of
bolts
Bolt hole diameter
Extension
diameter
C
(1)
surface D
O.D.
gasket
(1)
Class
ASME B16.5 (ANSI) 150
ASME B16.5 (ANSI) 300
ASME B16.5 (ANSI) 600
DIN 2501 PN 10–40 DN 50 2.40 (61) 0.97 (25) 1.31 (33) 5.65 (143)
DIN 2501 PN 25/40
DIN 2501 PN 10/16 DN 100 3.60 (91) 0.97 (25) 1.31 (33) 5.65 (143)
1. Tolerances are 0.040 (1.02), - 0.020 (0.51).
Pipe
size
2 (51) 2.12 (54) 0.97 (25) 1.31 (33) 5.65 (143)
3 (76) 3.60 (91) 0.97 (25) 1.31 (33) 5.65 (143)
4 (102) 3.60 (91) 0.97 (25) 1.31 (33) 5.65 (143)
2 (51) 2.12 (54) 0.97 (25) 1.31 (33) 5.65 (143)
3 (76) 3.60 (91) 0.97 (25) 1.31 (33) 5.65 (143)
4 (102) 3.60 (91) 0.97 (25) 1.31 (33) 5.65 (143)
2 (51) 2.12 (54) 0.97 (25) 1.31 (33) 7.65 (194)
3 (76) 3.60 (91) 0.97 (25) 1.31 (33) 7.65 (194)
DN 80 3.60 (91) 0.97 (25) 1.31 (33) 5.65 (143)
DN 100 3.60 (91) 0.97 (25) 1.31 (33) 5.65 (143)
Process
side G
Lower housing F
1
/4-in. NPT1/2 -in. NPT
B
Emerson.com/Rosemount
79
Rosemount 3051
March 2018

Options

Standard configuration

Unless otherwise specified, transmitter is shipped as follows:
Engineering units
Differential/Gage:
Absolute/ Rosemount 3051TA/ Rosemount 3051TG:
(1)
4 mA
:
(1)
20 mA
: Upper range limit
Output:
External buttons:
Flange type:
Flange material:
O-ring material:
Drain/vent:
LCD Display:
(1)
Alarm
: High
Software tag:
Damping:
1. Not applicable to FOUNDATION Fieldbus, PROFIBUS PA, or wireless.
2. For Fieldbus Protocols, default damping is 1 second.
inH2O (Range 0, 1, 2, and 3)
psi (all ranges)
0 (engineering units above)
Linear
None
Specified model code option
Specified model code option
Specified model code option
Specified model code option
None
(Blank)
0.4 seconds
(2)
Refer to the Rosemount 3051 Configuration Data Sheet
for
Rosemount 3051 HART Protocol.
For Wireless, refer to the Rosemount 3051 Wireless
(1)
Configuration Data Sheet
.

Tagging (three options available)

Standard SST hardware tag is wired to the transmitter. Tag
character height is 0.125-in. (3,18 mm), 56 characters maximum.
Tag may be permanently stamped on transmitter nameplate
upon request, 56 characters maximum.
Tag may be stored in transmitter memory. Character limit is
dependent on protocol.
HART Revision 5: 8 characters
HART Revision 7 and wireless: 32 characters
OUNDATION Fieldbus: 32 characters
F
PROFIBUS PA: 32 characters
Commissioning tag
A temporary commissioning tag is attached to all transmitters. The tag indicates the device ID and allows an area for writing the location.
Optional Rosemount 304, 305 or 306 Integral Manifolds
Factory assembled to 3051C and 3051T transmitters. Refer to the following Product Data Sheet for Rosemount 304, 305, and 306 for additional information.
Other seals
Refer to Rosemount DP Level Transmitters and 1199 Diaphragm Seal System Product Data Sheet for additional information.
(2)

Custom configuration

(1)
If option code C1 is ordered, the customer may specify the following data in addition to the standard configuration parameters.
Output information
Transmitter information
LCD display configuration
Hardware selectable information
Signal selection
Wireless information
Scaled variable
1. Not applicable to FOUNDATION Fieldbus or PROFIBUS PA Protocols. 2. Only applicable to FOUNDATION Fieldbus Protocol.
80
Emerson.com/Rosemount
March 2018
Rosemount 3051
Output information
Output range points must be the same unit of measure. Available units of measure include:
Pressure
atm inH2O at 4 °C
(2)
mbar mmH2O kg/cm
bar mmHg Pa
inH20
mmH2O at 4 °C
(2)
inHg ftH20 MPa
(1)(2)
hPa
mHg
at 0 °C
1. Field configurable only, not available for factor y calibration or custom configuration (option code C1 “Software configuration”).
2. Not available with low power (output code M) or PROFIBUS PA Protocol (output option code W).
(1)(2)
inH2O at 60 °F
psf
(2)
(1)(2)
g/cm
kPa
kg/m
ftH2O
at 4 °C
2
(2)
(1)(2)
(1)(2)
2
2
psi
torr
cmH2O
(1)(2)
at 4 °C
mH2O
(1)(2)
at 4 °C
ftH2O
(1)(2)
at 60 °F
cmHg
(1)(2)
at 0 °C
Display and interface options
M4 Digital display with LOI
Available for 4–20 mA HART and PROFIBUS PA
M5 Digital display
2-Line, 5-Digit LCD display for low power output
2-Line, 8-Digit LCD display for 4–20 mA HART, F
Fieldbus, and PROFIBUS PA Protocols
3-Line, 7-digit LCD display for wireless
Direct reading of digital data for higher accuracy
Displays user-defined flow, level, volume, or pressure units
Displays diagnostic messages for local troubleshooting
90° rotation capability for easy viewing
OUNDATION
Configuration buttons
Rosemount 3051 will ship with no buttons unless option D4 (analog zero and span), DZ (digital zero), or M4 (LOI) for local configuration buttons are specified. The Rosemount 3051 Wireless Transmitter is available with a Digital zero button installed with or without the LCD display digital display.
Transient protection (option code T1)
Tested in accordance with IEEE C62.41.2-2002, location category B
6 kV crest (0.5 μs–100 kHz) 3 kA crest (8 x 20 μs) 6 kV crest (1.2 x 50 μs)
Bolts for flanges and adapters
Options permit bolts for flanges and adapters to be obtained
in various materials
Standard material is plated CS per ASTM A449,
type 1
L4 austenitic 316 SST bolts
L5 ASTM A 193, grade B7M bolts
L6 alloy k-500 bolts
Conduit plug
DO 316 SST conduit plug
Single 316 SST conduit plug replaces CS plug
Rosemount 3051C Coplanar Flange and 3051T bracket option
B4 Bracket for 2-in. pipe or panel mounting
For use with the standard coplanar flange configuration
Bracket for mounting of transmitter on 2-in. pipe or panel
SST construction with SST bolts
Rosemount 3051C Traditional Flange bracket options
B1 Bracket for 2-in. pipe mounting
• For use with the traditional flange option
• Bracket for mounting on 2-in. pipe
• CS construction with CS bolts
• Coated with polyurethane paint
B2 Bracket for panel mounting
• For use with the traditional flange option
• Bracket for mounting transmitter on wall or panel
• CS construction with CS bolts
• Coated with polyurethane paint
B3 Flat Bracket for 2-in. pipe mounting
• For use with the traditional flange option
• Bracket for vertical mounting of transmitter on 2-in. pipe
• CS construction with CS bolts
• Coated with polyurethane paint
B7 B1 Bracket with SST bolts
Same bracket as the B1 option with series 300 SST bolts
B8 B2 Bracket with SST bolts
Same bracket as the B2 option with series 300 SST bolts
B9 B3 Bracket with SST bolts
Same bracket as the B3 option with series 300 SST bolts
BA SST B1 bracket with SST bolts
B1 bracket in SST with series 300 SST bolts
Emerson.com/Rosemount
BC SST B3 Bracket with SST bolts
B3 bracket in SST with series 300 SST bolts
81
Rosemount 3051
00813-0100-4001, Rev TC
Global Headquarters
Emerson Automation Solutions
6021 Innovation Blvd. Shakopee, MN 55379, USA
+1 800 999 9307 or +1 952 906 8888 +1 952 949 7001 RFQ.RMD-RCC@Emerson.com
North America Regional Office
Emerson Automation Solutions
8200 Market Blvd. Chanhassen, MN 55317, USA
+1 800 999 9307 or +1 952 906 8888 +1 952 949 7001 RMT-NA.RCCRFQ@Emerson.com
Latin America Regional Office
Emerson Automation Solutions
1300 Concord Terrace, Suite 400 Sunrise, FL 33323, USA
+1 954 846 5030 +1 954 846 5121 RFQ.RMD-RCC@Emerson.com
Product Data Sheet
March 2018
Europe Regional Office
Emerson Automation Solutions Europe GmbH
Neuhofstrasse 19a P.O. Box 1046 CH 6340 Baar Switzerland
+41 (0) 41 768 6111 +41 (0) 41 768 6300 RFQ.RMD-RCC@Emerson.com
Asia Pacific Regional Office
Emerson Automation Solutions
1 Pandan Crescent Singapore 128461
+65 6777 8211 +65 6777 0947 Enquiries@AP.Emerson.com
Middle East and Africa Regional Office
Emerson Automation Solutions
Emerson FZE P.O. Box 17033 Jebel Ali Free Zone - South 2 Dubai, United Arab Emirates
+971 4 8118100 +971 4 8865465 RFQ.RMTMEA@Emerson.com
Linkedin.com/company/Emerson-Automation-Solutions
Twitter.com/Rosemount_News
Facebook.com/Rosemount
Youtube.com/user/RosemountMeasurement
Google.com/+RosemountMeasurement
Emerson Terms and Conditions of Sale are available upon request. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.
© 2018 Emerson. All rights reserved.
Loading...