R&S®ist eingetragenes Warenzeichen der Fa. Rohde & Schwarz GmbH & Co. KG.
Eigennamen sind Warenzeichen der jeweiligen Eigentümer.
®
is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
R&S
Proper names are trademarks of the respective owners.
Printed in the Federal
Republic of Germany
Rohde & Schwarz GmbH & Co. KG
Mühldorfstraße 15
D-81671 München
www.rohde-schwarz.com
3030.3059.72-02.00
Page 3
Brand Names UMS100/120/120 System Manual
Dear Customer,
The following abbreviations are used throughout this manual:
®
MS100/120 is abbreviated with UMS100/120.
&S
U
R
®
R&S
UMS100/120G is abbreviated with UMS100/120G.
®
UMS100/120HF is abbreviated with UMS100/120HF.
R&S
®
UMS100/120SHF is abbreviated with UMS100/120SHF.
R&S
®
ARGUS-UMS is abbreviated with ARGUS-UMS.
R&S
®
ARGUS is abbreviated with ARGUS.
R&S
®
is a registered trademark of Rohde & Schwarz GmbH & Co. KG and its subsidiaries.
R&S
CAUTION
This document is only applicable for ARGUS-UMS software of version 1.1 or higher.
(The controlled UMS100/120 therefore require firmware of version 1.1 or higher.) Older SW
versions have a different MMI and are not covered by this manual.
A3030.3059.72-02.00
Page 4
Page 5
Contents UMS100/120/120 System Manual
Contents
Technical information ................................................................................................................................
Figure 3-1: Service Tool Firmware Update ............................................................................................. 64
Figure 3-2: Service Tool - Maintenance .................................................................................................. 65
3030.3059.72-02.00 6
Page 9
Operation preparation UMS100/120/120 System Manual
1 Operation preparation
1.1 Foreword
This manual describes the interconnection of one or more UMS100/120 with a control PC to a highcapacity, extensively automatic radio monitoring network.
In addition, steps are described required for setting-up measurements which run automatically on the
UMS100/120.
The system operator is then only required if a UMS100/120 detects a deviation and has triggered an
alarm on the control PC in the main office. Manual tests must then be performed interactively to obtain
further information. This procedure is also outlined as an example.
Figure 1-1: Main UMS100/120 connections to the central control PC (star network)
73030.3059.72-02.00
Page 10
UMS100/120/120 System Manual Operation preparation
1.2 Installation/preparation
The UMS100/120 components and control PC are also referred to as network elements in the following,
in accordance with ARGUS nomenclature. Before the actual configuration of these network elements
(UMS100/120 and control PC) can begin, the type of connection must be determined. Therefore it is
strongly recommended to enter the names, IP addresses, modem telephone numbers and other access
parameters such as user names and passwords, etc into a corresponding table. Only then can the
configuration of the network elements take place smoothly and enable communication.
Allocating IP addresses is essential for the network elements to be able to communicate clearly in the
network.
The definition of IP addresses in the private address range will not be discussed here in any detail.
There are explanations in the corresponding literature (e.g. using the Internet at Wikipedia:
http://de.wikipedia.org/wiki/Hauptseite or according to RFC 1918, go directly to
http://www.ietf.org/rfc/rfc1918.txt).
If more than one UMS100/120 is operating in the network, the configuration must take place before it is
connected to the network (LAN/WAN), as all UMS100/120 are set with a factory default address
(192.168.0.31).
One such configuration table is attached in appendix A which is filled in with several stations and the
accompanying parameters as an example.
This or the UMS100/120 to be connected is/are to be installed and connected to the power supply in
accordance with the "Mechanical Installation Manual". In the case of a LAN/WAN connection, the
UMS100/120 must be configured before with regards to the address.
The main office must be accessible at all times and supplied with the appropriate equipment for set-up
and operating a standard PC with keyboard, mouse and a printer.
Depending on the type of UMS100/120 connection, suitable telecommunication lines must be available.
This includes LAN/WAN access and/or analogue telephone lines for at least one, but usually two
modems. GSM module as modem can also be used alternatively. In this way the main office also has a
"wireless" connection.
A standard PC is suitable as a control PC equipped with a Pentium IV processor, 512 MB memory, at
least 40 GB hard drive, CD or DVD-RW drive and an adequate number of RS232 ports (for external
modems) or PCI card slots (for internal modems). At least one free USB connector must be available
for the USB hardlock.
3030.3059.72-02.00 8
Page 11
Operation preparation UMS100/120/120 System Manual
Figure 1-2: UMS100/120 connection to the main office (via LAN/WAN (red) or mobile network (blue))
When using a LAN/WAN connection via e.g. SHDSL connections, that are available in many countries
now as point-to-point connections should be completely end-to-end tested by the telephone network
operator, including modems.
1.3 Web-based Service Tool in UMS100/120
To carry out the minimal service work when up and running and for the initial configuration, a user
interface can be used which is web-based and can be started on each UMS100/120. A web browser
such as MS Internet Explorer from version 5.0 (better 6.0) can be used or e.g. Mozilla Firefox 1.4 on
any control or service PC. The control or service PC is connected to the system that has to to be
configured via a LAN/WAN or mobile network connection. Once the IP address has been entered the
configuration tool starts up. The user interface of the configuration tool is only available in English.
Furthermore, a service PC does not require any other special programs.
93030.3059.72-02.00
Page 12
UMS100/120/120 System Manual Operation preparation
1.4 UMS100/120 configuration for LAN/WAN
connections
CAUTION
The configuration must take place before installing the device within the LAN as they all
have the same IP address (192.168.0.31) from the factory by default. Without previous
configuration, the devices would not be able to communicate individually in the network and
there would be an IP address conflict.
The configuration is either done in a central delivery store before they are distributed to the site or it is
done by a technician on site after the mechanical installation using a service notebook.
A standard PC with a LAN interface and a web browser (e.g. MS Internet Explorer, see chapter 1.3) is
required for the configuration.
Connect the PC to the UMS100/120 in one of the two following ways:
Connect the RJ45 socket on the PC to the corresponding socket on the bottom of the UMS100/120
•
to be configured directly using a cross wired LAN cable.
Connect the RJ45 socket on the PC to the corresponding socket on the bottom of the UMS100/120
•
to be configured using an uncrossed cable by interconnecting a simple LAN switch or hub.
Then start the UMS100/120 by connecting it to the power supply.
Now start up the PC and set the network card to an IP address within the address range 192.168.0.xxx,
Subnet. 255.255.255.0, e.g. 192.168.0.11. The 192.168.0.31 IP address may not be used as this is the
UMS100/120 address ( IP address conflict).
Then start up the web browser.
To open the service tool in UMS100/120, enter the following address in the browser window.
CAUTION
As this is a matter of a secure connection it is essential to enter "https:…..".
https://192.168.0.31
The following security message appears:
3030.3059.72-02.00 10
Page 13
Operation preparation UMS100/120/120 System Manual
Figure 1-3: Service Tool, security certificate
As the communication between the control PC and UMS100/120 is processed in SSL (encrypted /
secure sockets layer), the UMS100/120 provides a certificate. A warning always appears to inform that
the name does not match the name of the site (see Figure 1-3). Acknowledge this certificate by clicking
on <Yes>.
UMS120 incorporates the modular version of the UMS100 complete system. Users therefore can
connect to the UMS120 a large range of monitoring antennas from Rohde & Schwarz (transducers are
interpreted) as well as from other suppliers (transducers are not interpreted) and configure these within
the service tool.
The displayed pages of the service tool are slightly different to UMS100 having an additional menu
entry for antenna configuration with UMS120 (refer to Figure 1-7).
11 3030.3059.72-02.00
Page 14
UMS100/120/120 System Manual Operation preparation
Then the following Login window appears:
Figure 1-4: Service Tool UMS100, Login window
Figure 1-5: Service Tool UMS120, Login window
3030.3059.72-02.00 12
Page 15
Operation preparation UMS100/120/120 System Manual
To log-in, enter the default user name and the corresponding password:
<Customer>
<21kd&woe>
In the following logged on status, the Home Page of the Service Tool appears with general information
and for currently installed versions of Firmware components.
CAUTION
For reasons of security, a timer runs which triggers an automatic logout after 30 minutes. In
this case the Login window appears again automatically.
Now the actual configuration can begin.
Figure 1-6: Service Tool UMS120 Home Page
For a better distinction between different UMS100/120 within the network it is possible to give the unit
an individual name under Measurement Unit.
Enter the name in the appropriate input field and click on <Set> to permanently take over the name.
13 3030.3059.72-02.00
Page 16
UMS100/120/120 System Manual Operation preparation
As the first page Date and Time is now opened. Check or correct the settings.
The same time zone must be set for all the UMS100/120s otherwise results between the
•
UMS100/120s cannot be correlated or automated measurements (AMMs) cannot start because
start time has passed already.
The time must be the actual time in this time zone and not UTC time. Check the data again and
•
click on <Submit> to send.
In a later operation the UMS100/120s synchronize their system time with the control PC.
Figure 1-7: Service Tool, date and time entries incl. time zone
Now enter the parameters from the configuration table in the corresponding masks of the Service Tool.
3030.3059.72-02.00 14
Page 17
Operation preparation UMS100/120/120 System Manual
On the Web Login Account page, the default login parameters can be replaced with customer-specific
ntries.
e
Overwrite <Customer> at Current User Name.
•
nter <21kd&woe> at Current Password.
E
•
Enter a new password at New Password.
•
Repeat the new entry in the sections underneath.
•
Then check the data again and click on <Submit> to send.
•
CAUTION
Make a detailed note of the new access data and store it in a safe place. If you forget your
details or they get lost, access to the UMS100/120 may no longer be possible.
If this is the case, you will be obliged to pay for the factory customer services.
For UMS100 with Ser. Numbers above 100075 and for UMS120 it is possible to perform a
factory preset by using the internal Preset Button (for further details please refer to the
UMS100/120 installation manual).
Figure 1-8: Service Tool, changing the Login parameters
15 3030.3059.72-02.00
Page 18
UMS100/120/120 System Manual Operation preparation
Now enter the details for the Control PC. Go to the relevant page.
For LAN/WAN connections, enter the IP Address for the LAN interface.
•
For GSM connections, enter the address as it is specified for the incoming dial-up networking.
•
In the Control Unit field, always enter the "ARGUS Name" from the standard ARGUS installation.
•
For ARGUS-UMS - this is always "ARGUS-UMS".
The Service to be selected is always “ARGUS_MT1” if the UMS100/120 is controlled by ARGUS-
•
UMS software, for standard ARGUS this may be different.
When all the fields are filled in and the Connection Type <LAN> or <Dial-up Connection> has
•
been selected, send the data by clicking on <Submit>.
Figure 1-9: Service Tool, details via the control PC in use
The settings are necessary to route the alarm messages back to the (correct) control PC within the
network as well as to send back the results from automatic measurements that have been started by
ORM (order report module).
In case a GSM/CDMA mobile phone link is used for the connection, further settings as e.g. phone No.
etc. are necessary (please refer to next mask).
3030.3059.72-02.00 16
Page 19
Operation preparation UMS100/120/120 System Manual
Now open the Communication window and enter the following parameters:
The parameters for the LAN/WAN connection to be configured
•
The parameters for the dial-up connection via e.g. GSM or CDMA (UMS120)
•
It is recommended that also these details are entered in case a wireless connection is necessary in the
future. But they can also be entered at a later date.
CAUTION
If you click on <Submit> once you have entered a new IP Address, Subnet Mask and
Default Gateway (only necessary with a router), the data is sent and a connection cannot
be kept open because the UMS100/120 is now only available via the new address
(https://<new address>).
If you want to open the service tool again, put the IP address as well as the Subnet mask of the service
or control PC into the new address range of the UMS100/120. Then connect to the Service Tool using
the new IP address.
Figure 1-10: Service Tool, setting the communication port
17 3030.3059.72-02.00
Page 20
UMS100/120/120 System Manual Operation preparation
Options Management
Type depending different options for UMS100 and UMS120 are available.
Installed options are “removed” by activating of “remove” and clicking on <submit>.
In case new options are to be installed, the following instructions should be followed:
Put the options-CD into the drive of the “Control-PC”.
•
After selection of <Browse …> mark the CD-drive and the options file in there.
•
Activate “install”.
•
And finally click on <Submit>.
•
Figure 1-11: Options for UMS100
CAUTION
The configured options in UMS100 not only extend the selectable frequency range but also
set the appropriate transducer tables for the supplied antennas (HF or SHF). In case that
antennas from other suppliers are connected to the UMS100, the displayed field strength
levels in the ARGUS control SW will be not correct.
3030.3059.72-02.00 18
Page 21
Operation preparation UMS100/120/120 System Manual
Figure 1-12: Options for UMS120
CAUTION
Options installed in UMS120 only control the settable frequency range.
Antenna Configuration (only UMS120)
Two different antennas can be connected to the UMS120 out of the portfolio of R&S monitoring
antennas (or from other suppliers) having corresponding or also slightly overlapping frequency ranges.
After selection in the service tool the corresponding transducer tables and connection cable types and
lengths are considered in the ARGUS control SW. Through this correct field strength levels can be
displayed.
In case of overlapping frequency ranges of the antennas the Transition Frequency can be calculated
from the tool (mid of overlapping range) or can be set manually by entering the value manually into the
appropriate input field if this is preferred.
Further the used cable types and lengths and – by use of active antennas – the necessary DC-feeds
can be selected. Clicking the <Update> button shows the antenna parameters.
After clicking on <Submit> the selected settings are activated on the UMS120.
19 3030.3059.72-02.00
Page 22
UMS100/120/120 System Manual Operation preparation
Figure 1-13: Antenna configuration for UMS120
CAUTION
In case that a combination between “no antenna” and “unknown antenna dBm” or
“unknown antenna dBMV” is selected, the ARGUS control SW only displays voltage levels
in dBm or dBMV.
A combination of “unknown antenna” and known (R&S) antenna nevertheless only displays
voltage levels. (Transducer tables from other antenna suppliers can not be entered.)
Now the UMS100/120 has been set so that it can be installed on the network and connected to the
ARGUS-UMS (or ARGUS) control program.
Finally open the page Logout and click on <Logout>. Wait until the login window appears and close the
browser afterwards.
Another part of the manual explains the other functions of the Service Tool.
3030.3059.72-02.00 20
Page 23
Operation preparation UMS100/120/120 System Manual
1.5 UMS100/120 configuration for dial-up
connections using a mobile network
CAUTION
Before the UMS100/120s can be connected via a mobile phone network to, for example,
GSM technology, "SIM cards" must be obtained from that network operator which provides
sufficient coverage at the planned UMS100/120 locations. In addition, UMS100/120
operator must apply for a special number for the Data Call function with the "SIM card".
Before insertion of the “SIM card” into the UMS100/120 the “PIN entry” must be disabled.
This can be done by placing the card into a standard mobile phone.
Insert the SIM card prior to starting the UMS100/120. The UMS100/120 must be disconnected from the
power supply. For further information please refer to the UMS100/120 User Manual.
It is now possible to carry out full configuration via the GSM connection which is available once the SIM
card has been inserted and the power supply is connected up. But further preparation is necessary on
the control PC for setting up dial-up networking which will be described in due course.
Due to the high speed, we recommend that the UMS100/120 configurations are done using the LAN
interface, as previously described.
The procedure is the same as the one already described (Figure 1-3 to Figure 1-9). In addition, you
must also fill in the parameters for Incoming Connection and Dialup Connection for the reverse
direction in the last entry field (Figure 1-10) and click on <Submit> each time to send it to the
UMS100/120.
1.6 Configuring the control PC for a LAN/WAN
connection
When all the UMS100/120 are connected by LAN to the control PC in the main office, make sure the
network card address is set as follows:
The address must be located in the same address range as all the UMS100/120.
•
The Subnet Mask must also correspond.
•
If a router is being used in the network to connect different addresses, set its address as Default
•
Gateway.
21 3030.3059.72-02.00
Page 24
UMS100/120/120 System Manual Operation preparation
1.7 Configuring the control PC for a GSM connection
Diagrams of the configuration window and the corresponding descriptions are based on Microsoft
Windows XP professional operating system. The configuration in MS Windows 2000 varies slightly in
different places. However, basically the specific settings can be made in the same way.
For the following configuration it is assumed that at least one, but preferably two modems are set up on
the control PC and connected to the phone lines.
For two or more modems, you can use one for the outgoing connections to the UMS100/120 if the
operator is actively connected. Any other modems can then be available to receive UMS100/120
generated and automatically transmitted alarms for incoming connections.
Please note the terminology regarding the connection set-ups here:
Dial-up networking to the
control PC
Direction of initiation UMS100/120 Service Tool
Outgoing connection
Incoming connection
Incoming
Outgoing
1.7.1 Outgoing connections (to the UMS100/120)
Install a separate outgoing dial-up networking using the Assistant for New Connections for each
UMS100/120 connected via GSM.
Open the Network Connections folder using the MS Windows desktop (start button or system
•
control).
Select Create New Connection with the mouse at Network Tasks. The Assistant opens.
•
a) Click on <Next> in the first window and open the second window.
b) Select Setting up a connection with the network at the office and click on <Next> to open
the third window.
c) In the third window, select Dial-up Networking and click on <Next> to open the next window.
d) (available only if more than one modem is installed, else continue with step e)
From the proposed (installed) modems in this window, select the one that will be used for the
outgoing connection and click <Next> to open the next window.
e) In this window, assign a name for the connection, such as the station name. Click on <Next> to
open the next window.
f) In this window, enter the telephone number for data connections which is assigned to the SIM
card in the UMS100/120. Click on <Next> to open the next window.
g) In this window select “… no smart card” and click on <Next> to open the next window.
h) In the following window click on <Complete> to finalize the definition. A connection to the
desktop is not necessary and is not selected.
3030.3059.72-02.00 22
Page 25
Operation preparation UMS100/120/120 System Manual
The information for
. Then you can always see whether
i) Then the Setting up connection with “name” window opens automatically.
the User Name and Password for the Login Parameter defined in the configuration table for
the relevant UMS100/120 is now entered.
Click on <Save User Name and Password for...> and <All users for this computer>.
j) Click on <Select> to start a connection set up to test the connection.
If everything has been configured correctly, a connection is set up after a short while and the user name
and password are verified. The connection has been made and can be used. The configuration of the
UMS100/120 can also be done remotely using the Service Tool and the web browser, as described in
Section 1.4.
Select disconnect.
•
There are some details left to configure for the connection. Select the dial-up network connection
•
just defined in the Network Connections folder. Open the Properties window by clicking it with the
right mouse button:
a) At the General tab, deselect All devices dial the same number and select the Show symbol
for a connection in the information section of the task bar
the connection has been made.
b) Keep the standard settings on the Options tab. At Idle time after hanging up, select 5 min or
10 min to avoid blockings and high telephone costs by forgetting to hang up.
c) The Security, Network and Extended tabs remain unchanged.
d) Close the window with <OK>.
Then set up and test all other dial-up networking for the remaining UMS100/120 in the same way.
•
1.7.2 Incoming connections (from the UMS100/120)
Before the incoming connection can be set up, a user account must be available or has to be
generated on the control PC where the UMS100/120 can register on the system.
Do not
Instead, go to Control Panel Administrative Tools Computer Management System Tools
and mark the Local User folder within Local Users and Groups.
Select the New User window from the Action command in the menu.
carry out this action via Windows Start System Control User Account.
23 3030.3059.72-02.00
Page 26
UMS100/120/120 System Manual Operation preparation
Figure 1-14: User administration
Fill in the mask with the information from the configuration table (entries, Dialup Connection) as you
can see in the following Figure 1-15 and click on <Create>.
Figure 1-15: Creating a user for dial-up
Please close the following empty mask with <Close>.
For reasons of security, you must specify group membership for the new user “name”:
Double-click on the new entry of the user created and select the Member of tab. Delete the
•
membership to the user class User.
The “name” Properties window should be empty after that.
3030.3059.72-02.00 24
Page 27
Operation preparation UMS100/120/120 System Manual
Figure 1-16: User properties, Member of .. window
In the following the membership Power Users is exclusively entered.
Click on <Add ..>. In the following Select Groups window click on <Advanced ..>.
•
Figure 1-17: Group selection
In the expanded Select Groups window finally click on <Find Now>. Select Power Users and
•
close the window with <OK>.
25 3030.3059.72-02.00
Page 28
UMS100/120/120 System Manual Operation preparation
Figure 1-18: Group selection expanded
Close the Select Users window with <OK> also.
•
The “name” Properties window now only has the entry “Power Users”.
Figure 1-19: “name” Properties with entry
Close “name” Properties window and the Computer Management window with <OK>.
3030.3059.72-02.00 26
Page 29
Operation preparation UMS100/120/120 System Manual
Creation of incoming telephone connection(s)
For automatic dial-in from the UMS100/120, create a connection using the Assistant for new
connections in the following way:
Click on <Next> to close the first window. The second window appears.
•
In the second window, select Setting up another connection. Click on <Next> to open the third
•
window.
In the third window, select Allow incoming connections. Click on <Next> to open the next
•
window.
The following window Device for incoming connections shows you all the available modems. You
•
may select any modem except the modem already configured for the outgoing connection. Click on
<Next> to open the next window.
In the next Incoming VPN connection window, you must select VPN connections not permitted.
•
Click on <Next> to open the next window.
In the User Permissions window, only select the previously defined user and then click on
•
Properties.
Figure 1-20: Dial-in user permissions
The Properties of ... window opens. Check, and if necessary set the Call-back tab to activate Call-
•
back not permitted. Then click on <OK> to close the window. Click on <Next> to open the next
window.
In the Network Software window, configure the first entry in the table Internet Protocol (TCP/IP)
•
by double-clicking on it or by clicking on <Properties>.
Enter the following settings:
- The first IP address corresponds to the address in the configuration table.
27 3030.3059.72-02.00
Page 30
UMS100/120/120 System Manual Operation preparation
- The last digit in the second IP address should be one digit higher than in the first.
A total of two IP addresses must be available, as shown in Figure 1-21.
-
Figure 1-21: Assigning an IP address for dial-up networking
Transfer the other settings from Figure 1-21. Click on <OK> to close the window. In the Assistant
•
window click on <Next>.
In the following window close the Assistant by clicking on <Complete>.
•
3030.3059.72-02.00 28
Page 31
Operation preparation UMS100/120/120 System Manual
1.8 Installing ARGUS-UMS for communicating with
UMS100/120
When ARGUS-UMS is started for the first time after installation, there must be at least one User Group
and one user in the User Class of Administrator because only this type of user has the rights to make
he necessary configurations. This procedure is largely self-explanatory. For more detailed information
t
see also the ARGUS-UMS Reference Manual.
Once the user name has been entered at <Insert>, the standard registration window opens. Enter the
defined user name there. You will then be asked to enter a new password and to repeat the entry. Then
the program starts and the standard control window appears.
Before UMS100/120 operation can be carried out by the control PC, some entries and assignments
must still be made in System View tab in Configuration Control Unit Configuration command in
the menu.
Now a new entry for a UMS100/120 can be generated:
In the Input Field area under Measurement PC overwrite the localhost entry with the IP address
•
of the new UMS100/120
In case of dial-in connections select the name of the previously installed Dial-up Networking
•
connection to the relevant UMS100/120.
In the field below enter the name of the UMS100/120 as it was defined in the configuration table.
•
Always enter the ARGUS_MT1 service which is automatically assigned to the instance.
•
In the Transfer Rate box below select the line speed according to the type of connection:
•
- for LAN/WAN usually the transfer rate is > 1.5 Mbit/s
- for mobile phone links select the appropriate rate (e.g. 9.600 kbit/s for GSM).
Finally click on <Insert>. The entry is saved and can be used.
•
Further UMS100/120 can be entered in the same way.
The complete System View Configuration can be saved following the menu Configuration Save System View .. and can be retrieved later.
29 3030.3059.72-02.00
Page 32
UMS100/120/120 System Manual Operation preparation
Figure 1-22: System View with some UMS100/120
CAUTION
The <Modify> and <Delete> buttons are enabled. Be careful not to click on them.
The procedure is now complete for configuring a UMS100/120. Click on <OK> to close the window.
If you want to set up more UMS100/120, begin with the intermediate step each time and carry out the
configuration as already described.
CAUTION
If at any point you want to make some changes to the input field, firstly click on <Modify>.
Otherwise the changes will be discarded and not saved.
3030.3059.72-02.00 30
Page 33
Operating with ARGUS-UMS UMS100/120/120 System Manual
2 Operating with ARGUS-UMS
2.1 Interface overview
Once you have started the program by clicking the icon on the desktop, the registration window appears
where you can enter your user name and password. Then the standard control window opens which
an be subsequently displayed in the following sizes. All additional windows opened from this window
c
can be placed on the desktop. They are not restricted to the size of the standard window.
You can view information in the help text on the buttons and input fields by simply gliding the mouse
over them. Details on possible value ranges are outlined in the ARGUS-UMS manual respectively in the
Online Help and can be accessed using the <F1> button. The button functions are also displayed in
the menu.
Any other functions and settings used less frequently can only be accessed via the menu. In particular,
the Configuration command.
Figure 2-1: ARGUS-UMS window
31 3030.3059.72-02.00
Page 34
UMS100/120/120 System Manual Operating with ARGUS-UMS
2.2 Connecting to a UMS100/120
There can be several connections to UMS100/120 in a LAN/WAN network. For dial-up networking, this
is restricted to the number of previous modems for outgoing connections.
However, you can only activate one connection from ARGUS. An active connection is necessary to
perform direct measurements (DMMs) or define automatic measurements (AMMs). If you need to
activate another connection it will be declined if there are any control windows still open. Otherwise an
activated connection will be disconnected.
Therefore the Select window contains the buttons <Connect> as well as <Connect and Activate>.
Depending on the type of connection, setting up the UMS100/120 connection takes up to one minute.
During this time a corresponding window appears. For dial-up networking the RAS States (2 4 5 10-> DUN connected) are displayed via the Connection Set Up Status.
If a connection to a UMS100/120 is set up for the first time, the configuration is not available on the
control PC. Therefore, a window appears while the system is setting up the connection and informs you
of a "mismatch" in the configuration. Then the configuration is automatically loaded. The reception
procedure can take up to one minute for slower connections. Once the transfer is successfully
completed, press <OK> to quit.
Figure 2-2: Connection set up with configuration transfer
The configuration between the control PC and the UMS100/120 is now coordinated.
3030.3059.72-02.00 32
Page 35
Operating with ARGUS-UMS UMS100/120/120 System Manual
Updating it again will only be necessary if
the UMS100/120 was modified by changes made to the enabled frequency range and/or the
•
antenna configuration when options/antenna configurations from the Service Tool were loaded or
deleted.
the data area was "emptied" with the Clean Up Data Area maintenance function from the Service
•
Tool in the UMS100/120.
The connected UMS100/120s are listed in the corresponding table in the Select window. The activated
connection is displayed in red.
Figure 2-3: Select UMS100/120-window with activated connection
33 3030.3059.72-02.00
Page 36
UMS100/120/120 System Manual Operating with ARGUS-UMS
This window is to be closed when the connection set up is complete. The ARGUS-UMS standard
indow displays the connection as follows:
w
Figure 2-4: ARGUS-UMS connected to a UMS100/120
The header informs about the name of the connected UMS100/120. The bolt of lightening on a yellow
background shows the active remote connection.
The input fields for the geographical coordinates and height can be filled in or changed. These
parameters are transferred to the specific UMS100/120 before the connection is terminated. If this
information has been entered correctly, the position of the UMS100/120 can be charted on a map using
an optional program from Rohde & Schwarz. Click on the central box to activate the transfer to the
program.
Now the Measurement mode is open and can be configured.
3030.3059.72-02.00 34
Page 37
Operating with ARGUS-UMS UMS100/120/120 System Manual
2.3 Manual operation in "Direct Measurement Mode"
DMM
The window of the manual operation, also called "Direct Measurement Mode" DMM is directly opened
after closing Select UMS window or using the menu command of the same name under Measurement
r directly using <F4>.
o
If frequency extensions are activated, it is shown next to the UMS100/120 name, e.g. UMS100
100.000 KHz – 6.000000000 GHz..
The menu item View Antenna opens the ANTENNA window for more information.
Figure 2-5: Antenna information
The DMM control window for UMS100/120 offers three types of measurements:
Fixed Frequency Mode - FFM: single frequency measurement, level, frequency, offset
•
Scan with start and stop frequency input and increments for a quick overview.
•
Frequency List Scan - FLS: single frequency measurement for preset frequency lists
•
For all of these measurements a detector (FM, AM) and listening to demodulated audio signals on the
control PC can be selected.
All levels will be displayed with UMS100 in dBMV/m (field strength) and with UMS120 depending on the
antenna configuration also in dBm/dBMV (voltage).
Depending on the bandwidth of the UMS100/120 connection, you can select a suitable Audio Transfer
Rate (in kbit/s)with more or less audio signal compression.
CAUTION
In case of GSM connections only the two lowest Audio Transfer Rates may be used.
Higher rates also block the transfer of measurement values leading to "timeouts" in the
communication and disturbances to the connection.
35 3030.3059.72-02.00
Page 38
UMS100/120/120 System Manual Operating with ARGUS-UMS
2.3.1 Gain Control
The available settings for Gain Control allow an attenuation of the input signal up to -30 dB as well as
an amplification of up to +25 dB to adapt the UMS100/120 to various measurement activities.
For an automatic adoption (AGC) it is possible to select
Fast (1 ms)
•
Medium (10 ms, only in FFM) and
•
Slow (100 ms, only in FFM)
•
2.3.2 Overload Indicator
For all manual measurements an overload indicator is available.
In case that the antenna input level reaches a critical value, where the pre-amplifier could produce
intermodulation or more critical, the analogue-digital-converter could be overdriven, the indicator color is
changing from green as shown above to red.
CAUTION
Performing a wide frequency scan (see next chapter) beginning with 100 kHz, step width of
20 MHz and wide IF bandwidth of 1 MHz the following effect can be seen:
In the area of the start frequency (100 KHz) the frequency of the local oscillator is within the
IF bandwidth of 1 MHz so that as a result a very strong level of about 130 dBMV/m is
displayed. Even with a selected attenuation of -30 dB the overload indicator will be still on.
This is a normal effect that can be observed with every receiver with corresponding
settings.
3030.3059.72-02.00 36
Page 39
Operating with ARGUS-UMS UMS100/120/120 System Manual
2.3.3 Getting an overview of the spectrum using Scan
For a quick look at the frequency spectrum, such as after an alarm, use the <Scan> function. Select the
start and stop frequency according to the areas of interest. The increments (step width) are oriented to
various criteria.
he smaller the increment, the slower is the scan. Alternatively, in the display there are small
T
increments for a higher resolution. In this way, signals can be detected that are very close to one
another. However, this can be restricted by the resolution of the screen you are using. Generally, the
ratio between the selected IF bandwidth and the increment is important. However, you can overlook this
ratio, as the Detector is generally in a max. peak position. Each signal is then displayed, even within
larger increments. The strongest signal determines the level value of the increment.
Here you are asked to develop suitable parameters for the signals which are of particular interest to
you.
Some examples for various parametrizing, for a better visualization of the time sequence with switched
on 2D waterfall diagram (for further information please refer to chapter 2.3.4):
The highest possible speed for a large frequency range (Figure 2-6)
•
The highest possible resolution for small frequency ranges, such as in the channel grid used - here:
•
analogue trunked radio, downlink - or GSM CHH downlink and the highest resolution (100 Hz) for
modulation analysis (Figure 2-7)
Figure 2-6: Example of a fast scan for a wide frequency range
37 3030.3059.72-02.00
Page 40
UMS100/120/120 System Manual Operating with ARGUS-UMS
Figure 2-7: Example: Scan for high resolution
If an interesting signal (peak) is found in the spectrum, depending on the step width, another scan may
be necessary over a more limited frequency range with smaller steps.
2.3.4 Waterfall Diagram in Scan Mode
The examples in chapter 2.3.3 show also the 2D Waterfall Diagram. Here the signals were shown in
the time sequence. Thus clear information is available about the duration and how often a signal is
transmitted.
The relevant level however can only be classified according to the displayed color.
3030.3059.72-02.00 38
Page 41
Operating with ARGUS-UMS UMS100/120/120 System Manual
2.3.5 Max Hold in Scan Mode
The red enveloping line shows the strongest signals that have been appeared since Max Hold has
een switched on. Here also signals with a short period are displayed with their peak level as it is in
b
voice radio transmissions. Also signals with fluctuating levels are recognized with their max. peaks. The
peaks can be selected with the marker where exact marker frequency and level is shown in the line
above the scan graphic.
Figure 2-8: Scan with Max Hold
2.3.6 Listen Mode in Scan
Before starting the scan, a suitable Demodulation (AM/FM), IF Bandwidth, Audio On, Audio Transf.
Rate and Volume should have been selected.
During a running scan a double click on the signal peak stops the scan, sets the marker on that position
and the audio modulation of the signal can be heard. The following Figure 2-9 shows the listening at
439,25 MHz. If the scan shall be restarted, the listen symbol (clef) has to be clicked once. To stop click
on the listen symbol again.
If the Listen Mode is activated the marker can also be moved with the mouse (or the cursor buttons
left and right) to other peaks.
39 3030.3059.72-02.00
Page 42
UMS100/120/120 System Manual Operating with ARGUS-UMS
Figure 2-9: Listen Mode in Scan
2.3.7 Bandwidth Measurement in Scan
Before starting the Bandwidth Measurement the Start and Stop Frequency should be set so that the
range is at least about two times wider than the expected bandwidth (test-scan). The Bandwidth Measurement can only be selected with button <B> if the scan is stopped.
There are two possibilities:
x dB down (default is 26 dB)
•
Percent below peak (ß measurement) (default 1 %)
•
Please insure before starting the measurement that a sufficient measurement dynamic range is
available (e.g. S 30 dB).The measurement is always performed with Max Peak. Also the step width
should be made small enough to get a fine resolution.
After starting the scan the center frequency marker is manually set with the mouse or by entering
•
the frequency in the corresponding input field.
The measurement window is set in that way that one of the yellow lines is moved with the mouse
•
where the other follows symmetrically around the center. Another method would be to enter the
value in the corresponding input field.
Depending on the selected number of scans (1, 10, 100) and sufficient measurement dynamic the
•
differential markers appear on the screen (small yellow circles at the left and right flanges of the
signal of interest) and the corresponding calculated bandwidth is displayed in the line above the
graphic.
The following picture shows the ß-method with 1 % setting.
3030.3059.72-02.00 40
Page 43
Operating with ARGUS-UMS UMS100/120/120 System Manual
Figure 2-10: Bandwidth Measurement
2.3.8 Scan with handover to (FFM)
In order to prepare a following Fixed Frequency Measurement the following options can be selected:
Set Marker Frequency
•
(with switchover to FFM the marker frequency will be the measured frequency)
Transfer Settings to Single Mode
•
The settings for IF Bandwidth, Meas. Time, etc. will be transferred to FFM)
If now the scan is started and the marker is positioned to a frequency of interest (using the mouse or
cursor left and right buttons) the settings are transferred if switched to FFM.
41 3030.3059.72-02.00
Page 44
UMS100/120/120 System Manual Operating with ARGUS-UMS
Figure 2-11: Settings for transferring configurations and marker on frequency of interest
2.3.9 Fixed Frequency Measurement
By clicking on the FFM symbol the window is switched to FFM mode. (Measurement of Level,
frequency and offset can be performed).
Directly after clicking the <Start> button the Level is displayed and the audio modulation can be heard if
the relevant demodulator is selected and Audio is switched on.
If the measurement is not running, the recording can be switched on by clicking the “Disk” symbol and
defining a File Name. Audio recording can be selected additionally. Start the measurement and stop it
again later. Now the recording can be switched off again.
Click on
to activate the frequency measurement. In addition the demodulator is switched off.
3030.3059.72-02.00 42
Page 45
Operating with ARGUS-UMS UMS100/120/120 System Manual
Figure 2-12: Frequency measurement in FFM
CAUTION
System based, only carriers or amplitude modulated carriers can be measured exactly. For
frequency modulated carriers there is no center frequency displayed but a frequency within
the bandwidth of the signal. Depending on modulation deviation of the FM signal the Offset
fluctuates, i.e. the difference between to the set frequency can be up to some kHz.
The above example (Figure 2-12) shows the frequency of a narrow-band FM signal where a
measurement is realistic for such kind of transmitters.
The accuracy can be evaluated by watching the differences over a period of some seconds. If the
differences are only some Hz the result can be trusted.
The measurement window (here 100.000 kHz) should not be made too large. Otherwise, another
signals with a higher level also located in this window, might be displayed instead.
The next figure shows an example of measuring an AM transmitter. The frequency is exactly coupled
with a normal frequency source.
43 3030.3059.72-02.00
Page 46
UMS100/120/120 System Manual Operating with ARGUS-UMS
Figure 2-13: Frequency and offset measurement for AM carrier
3030.3059.72-02.00 44
Page 47
Operating with ARGUS-UMS UMS100/120/120 System Manual
2.4 Definition of automatic measurements (AMM)
To a large extent automatic measurements will be the main kind of operation, as in a network of
different UMS100/120, so the operator is released from routine activities an can concentrate on other
activities.
During a short online session with the UMS100/120 the required settings are defined. Then the
connection can be disconnected and the UMS100/120 carry out the measurements “offline”
continuously.
When AMM is selected for the first time the following window appears:
Figure 2-14: Preconditions for AMM
To start the automatic measurement, the Range Definition, Measurement Definition and Timing Definition have to be defined. Each of these definitions can be available several times and be assigned
for a specific automatic measurement with additional adaptations. During operation, a large set of
different definitions are then available for the operator's various measurement requirements.
Note:
Before an automatic measurement is defined and started, first run a test with specific settings in the
manual measurement mode. There is a range of combinations which exclude one another. In addition,
there are limits for measurement times, bandwidths and automatic settings resulting in different
parameters.
If automatic measurements are started with parameters that do not match, you possibly will not receive
an error message straight away but only later in the log file.
2.4.1 Memory required for different measurements and
audio recordings
The following table gives the user a roughly overview of the memory required for different
measurements. From these durations an AMM planning could be derived.
Type of Measurement in AMM (Each takes 1 hour) Memory required in %
of the max. space
available
Scan 118 MHz – 135 MHz, increment = 25 kHz, meas. time = 1 ms, IF
bandwidth = 10 kHz, preamplifier = one, level measurement, continuous
measurement
FFM at 123.9 MHz, meas. time = 5 ms, IF bandwidth = 10 kHz, preamplifier = one, level measurement, continuous measurement
FFM at 123.9 MHz, meas. time = 5 ms, IF bandwidth = 10 kHz, preamplifier = one, level measurement, for values exceeding the limit
S 55 dBMV/m a second measurement is carried out with "high" quality
2.1
(921,993 data records)
0.11
(8,822 data records)
0.43
(443/649 data records)
45 3030.3059.72-02.00
Page 48
UMS100/120/120 System Manual Operating with ARGUS-UMS
Type of Measurement in AMM (Each takes 1 hour) Memory required in %
of the max. space
available
audio recording
FFM at 105.7 MHz, meas. time = 20 ms, IF bandwidth = 100 kHz, preamplifier = one, level measurement with continuous measurement "low"
quality audio recording
FFM at 105.7 MHz, meas. time = 20 ms, IF bandwidth = 100 kHz, preamplifier = one, level measurement with continuous measurement audio
recording with "SpeexQ1" quality
1.65
(5,000 data records)
0.43
(4,970 data records)
2.4.2 Creating automatic measurements with AMM Wizard
By use of the AMM wizard an automatic measurement can easily be set up in 5 steps. This procedure is
recommended especially for unskilled beginners.
Begin by accessing the AMM menu, go to Extras AMM Wizard... and open the first mask.
Figure 2-15: Accessing AMM Wizard
3030.3059.72-02.00 46
Page 49
Operating with ARGUS-UMS UMS100/120/120 System Manual
Step 1:
The first step (Step 1) defines the timing. To keep things flexible, almost all variants of any and
repeated measurements are possible.
In the following example, two periods on Tuesday at 9 o'clock for 6 hours are defined.
•
However, for continuous monitoring activities with UMS100/120, select Measurement is always
•
performed, that is "round the clock" 365 days of the year (first selection).
The last point is of interest if measurement data is recorded over a longer time period and for
•
example, measurements are carried out only every 15 minutes so as to save memory space.
Note:
The total available memory space is not the issue here. As a rule, the memory is sufficient for
measurements taking several weeks. More importantly, the recorded measurement values must be
transferred so they can be analysed by the operator. The transmission can take a long time, particularly
for large amounts of data when using a slow GSM connection. Please refer to the notes in the previous
chapter.
Figure 2-16: AMM Wizard Step 1, Timing Definition
Once you have made all the settings, click on <Next> to close the window and the Range Definition
window (Step 2) appears.
47 3030.3059.72-02.00
Page 50
UMS100/120/120 System Manual Operating with ARGUS-UMS
Step 2:
In case individual frequencies (based on a Frequency List
•
necessary list. The lists must have been previously made. For details on frequency lists please refer
to chapter 2.4.5.
To continuously search a frequency range, select Scan and specify the range by entering the Start
•
and Stop Frequency. If "narrow band" signals i.e. with an IF bandwidth corresponding to the type
signal shall be monitored, the Step Width is usually adjusted to the channel spacing of the radio
service in that frequency range.
Do not follow this rule in case that high scan speeds over large frequency ranges should be
•
achieved. Here increments of up to 20 MHz with IF setting of 1 MHz should be selected.
Even for an IF bandwidth of max. 1 MHz no signal will be lost, also not with narrow band carriers.
If a fixed frequency needs to be checked, select FFM and enter the frequency below.
•
) have to be checked, select the
Figure 2-17: AMM Wizard Step 2, Range Definition
Once you have made all the settings, click on <Next> to close the window and the Measurement Definition (Step 3) window appears.
3030.3059.72-02.00 48
Page 51
Operating with ARGUS-UMS UMS100/120/120 System Manual
Step 3:
In Step 3 the measurement settings are selected.
If stronger signals are expected, select -20 dB or even -30 dB at Gain Control.
•
For very weak signals select the +25 dB setting.
In case of frequently changing signal levels the Automatic Gain Control (Fast, i.e. with FFM also Medium/Slow) may be the preferred choice.
Keep the measurement period for scans as short as possible, select ≤ 1 ms for fast scans.
•
Demodulation is not important during a scan.
•
For fixed frequency measurements also Frequency and Offset can be selected or - alternatively
•
Audio. In this case the demodulated audio signal will be directly recorded on the UMS100/120. For
this purpose a demodulator must be switched on which is not possible at this point. Please refer to
chapter 2.4.3.
If there is an active connection to the UMS100/120 during AMM, the continuous measurements can
•
then be shown in the form of a graph. You can select the graphic type here.
In the last but one configuration window, set when and if measurement values are to be saved as well
as if and how an alarm is to be triggered automatically at the Control PC. in the center. The <Alarm>
button opens a window for this purpose. Here the essential things are now defined for automatic
operation.
If measurement results are to be saved on the UMS100/120 the Measurement Time preset must
•
be taken into account. A continuously running measurement creates huge amounts of data and the
memory runs at full capacity. Besides fast scans, this can occur particularly when there are
continuous audio recordings (see Section 2.4.1).
What may be more important is that the large amounts of recorded data must also be “moved” from
UMS100/120 to the Control PC using an extremely slow interface. For example, a limited audio
recording with RF level takes at least approx. one minute to transfer, even with a fast LAN
connection. Using a GSM connection a transfer would take approx. 6 hours.
49 3030.3059.72-02.00
Page 52
UMS100/120/120 System Manual Operating with ARGUS-UMS
If a continuous measurement with automatic alarm is to be performed, the Measurement Result
•
during an Alarm or Beginand End of an Alarm settings are sufficient. There are various options
for configuring the alarm criteria in the Alarm window.
One option is to enter the highest limit value as a fixed line via the scan range. This variant runs
•
ell if there is no other carrier apart from the transmissions expected. If the level is exceeded by a
w
signal, the alarm criterion is fulfilled. The actions selected at Alarm Actions are carried out, such
as starting a second measurement which e.g. measures the alarm frequency exactly.
The Message function and a Signal Tone are only released on the Control PC if there is an active
•
connection.
In case SMS is selected and a mobile phone number with information text has been entered, the
•
UMS100/120 sends a short message to this mobile phone if an alarm occurs. The only condition is
that a SIM card has been inserted. Please follow the instructions in chapter 1.5.
The SMS alarming is send only once after an initialization of the alarm. New SMS will only be sent
after restart of the AMM or after the first new alarm following midnight (00:00 h).
Another option is to measure against a predefined Limit Line. To do this, a spectrum is saved
•
during a DMM as the actual status with any Offset of ± x dB. If a new signal is detected, it is
recognized by an Overshoot measurement. If a carrier fails, it is recognized by an Undershoot
measurement. In Section 2.4.4 the setting for the Limit Line is described.
3030.3059.72-02.00 50
Page 53
Operating with ARGUS-UMS UMS100/120/120 System Manual
Figure 2-19: AMM Wizard result analysis and alarm function
Once all the settings have been carried out, click on <Next> to close the window and the next window
(Step 5) appears.
Step 5:
In Step 5 proposals are made for the names of the three files to be created which can be accepted or
overwritten. Click on <Finish> to end the definition.
The following question if the settings should be stored as presets can be answered at discretion.
2.4.3 Other AMM settings
To obtain other settings the AMM window can be enlarged by clicking on <Settings>.
Forward Messages is switched on in the Messages area for the automatic alarming.
•
At the Station input field, the name Control Unit is always shown in ARGUS-UMS.
•
The Delay Time determines a delay after the first alarm has been detected before the UMS100/120
•
connects to the control PC automatically. This delay is very important particularly with a GSM
connection. The value should not be set at too short intervals so as to avoid frequent connections. If
continuous alarms are expected, do not set the time at less than one minute intervals. Otherwise,
there is a danger that AMM can not be stopped due to a blocked outgoing connection coming up
against a frequently occupied UMS100/120 because this is permanently sending alarms.
The setting Summarize Messages controls if alarms that are displayed in the General Log File
•
have only information about date and time of alarm or detailed information about the kind of alarm
with e.g. frequency, level etc..
51 3030.3059.72-02.00
Page 54
UMS100/120/120 System Manual Operating with ARGUS-UMS
Figure 2-20: AMM window, other functions
3030.3059.72-02.00 52
Page 55
Operating with ARGUS-UMS UMS100/120/120 System Manual
Clicking on <Timing Definition> opens the corresponding window presenting a table with various
•
Measurement Definitions also e.g. made by the settings in Wizard. These can be extended and
more specified here.
Several AMM at different times can be defined; using different frequency ranges etc. and inserting
•
hem into the table.
t
Figure 2-21: AMM Timing Definition window
The subsequent Measurement Definition window displays a setting for simultaneous level
•
measurement and audio recording.
For recording audio only when a transmission is active, set a Squelch level.
•
Figure 2-22: AMM Measurement Definition window
53 3030.3059.72-02.00
Page 56
UMS100/120/120 System Manual Operating with ARGUS-UMS
In the Range Definition window all the available types of measurement parameters are listed. Click
•
on <Range Definition> to open this window from Measurement Definition window.
For creation of a new measurement the parameter has to be selected and inserted by clicking of
•
<Insert>.
The icons under System Path are a special feature. Click on them to open the appropriate device
•
drive. Select the settings here which otherwise cannot be set or changed within the other AMM
windows, such as the settings for the Audio Transfer Rate in the Sound drive.
Always click on modify before closing a window if changes have been made before. After closing a
•
window, a question appears if changes should be saved. Always accept this question.
Figure 2-23: AMM Range Definition window
In the Measurement Definition window click on <Further Settings ..> to open the Further Settings
•
window and to go to the settings for Limit Monitoring. In the relevant input fields Limit Lines for
Overshoot and Undershoot as well as Alarm Actions can be set.
The defined frequency range for the measurement (start and stop frequency) must correspond to
the frequency range of the Limit Line or fall within this frequency range.
During an active connection an additional graph can be displayed on the control PC which you can
•
select in the Graphic field. You can display measurements against a Limit Line in a diagram.
Here also the setting can be done for SMS alarming.
•
3030.3059.72-02.00 54
Page 57
Operating with ARGUS-UMS UMS100/120/120 System Manual
Figure 2-24: AMM Further Settings … window
55 3030.3059.72-02.00
Page 58
UMS100/120/120 System Manual Operating with ARGUS-UMS
After all the AMM settings are completed click <Start> to begin AMM.
The AMM window under Status shows the measurements performed. (Meas. Channel, see
•
Figure 2-25) and the accompanying parameters at Active Task.
You can also view a selected Diagram and the General Log File which opens when the first alarm
•
arrives.
Figure 2-25: AMM window with continuous measurement and the General Log File
Once the AMM has started, the connection to the UMS100/120 can be interrupted in the Select
window. Alarms are still transferred and displayed in the General Log File.
The General Log File shows the latest alarms on top whereas elder alarms are moving to the bottom.
The operator can acknowledge an alarm by a double click in that line if he has checked it. So the color
is changing from red to black.
In order to directly connect to an alarming UMS100/120 the operator can right click on this entry and
select the corresponding function in the context mouse menu.
In this menu additionally entries can also be selected and deleted.
3030.3059.72-02.00 56
Page 59
Operating with ARGUS-UMS UMS100/120/120 System Manual
2.4.4 Creating limit lines
In the actual measurement mode, a scan is started for the required frequency range and then
•
stopped when the display result is satisfactory. This shall form the “as it is” status of the spectrum.
In the scan window click the button <Create Limit Line> (disk symbol with green scan curve on top).
•
The Navigator window opens for entering a name. Click <New> to close the window.
Figure 2-26: Creating a Limit Line from the Scan graph
Enter an Offset for the measured spectrum in the Create Limit Line window that follows.
Set the appropriate Offset. Positive or negative offset depends on whether the alarm goes off when
•
the Limit Line is later overshoot or undershoot in the automatic measurement.
A positive offset of 5 dB for example, pushes the limit up. Slight level fluctuations do not trigger the
•
alarm with the Overshoot setting; however it will be triggered if there is a new, stronger and
unknown signal.
If a signal is too weak and is under the limit with the Undershoot setting or disappears completely,
•
the alarm will then be triggered.
In case that previously Max Hold was switched on also the Max Hold line can be saved as limit line.
•
If Max Hold was running some time before saving, smaller fluctuations in signals are recorded and
as an effect out of this, for Overshoots the Offset can be set to smaller values.
Figure 2-27: Offset setting for the Limit Line
Click <OK> to quit.
The Limit Line can now be selected during the definition of AMM.
57 3030.3059.72-02.00
Page 60
UMS100/120/120 System Manual Operating with ARGUS-UMS
2.4.5 Creating frequency lists
Sometimes it makes more sense not to scan a wide frequency range but just some individual
frequencies. These are put into a frequency list.
In Navigator, select File Type Frequency List.
•
Enter a new „name“.
•
Click on <OK> to close the window. Now a blank frequency list opens.
•
Enter the frequencies you want to scan. After closing the list, it directly can be used for measuring.
•
Figure 2-28: Creating a frequency list
Instead of entering manually, you can also import a frequency list created in Microsoft Excel. Go to File Import FrequencyList.
Another possibility exists by takeover certain frequencies out of the scan graphic in DMM.
Firstly open an existing or new frequency list in the Navigator. Now the Scan Mode is started in DMM.
If the spectrum has been measured, certain frequencies can be selected with the marker. By clicking on
the symbol the frequency is taken over in to the Frequency List.
3030.3059.72-02.00 58
Page 61
Operating with ARGUS-UMS UMS100/120/120 System Manual
Figure 2-29: Take over frequencies into the frequency list
59 3030.3059.72-02.00
Page 62
UMS100/120/120 System Manual Operating with ARGUS-UMS
2.5 Analyzing alarms and measurement results
In conclusion with an alarm warning or after an AMM has been stopped the measurement results have
to be evaluated.
Before, it may have happened that under certain circumstances with special measurements including
audio recording the available disk space came to a minimum of less than 5 %. In this situation the
following alarm appears in the General Log File:
Figure 2-30: Disk capacity shortage on UMS100/120
Now the following tasks should be done:
Make a connection to the relevant UMS100/120 if there is not one already.
•
Open the Navigator and select the Remote Measurement Files tab. All the measurement results
•
and log files are displayed here that are saved in the remote UMS100/120.
Mark the files by clicking on them. Several or all of them can be selected simultaneously. Selection
•
of single files allows displaying the file size. Go to File Attributes to view the relevant information.
More detailed information is available on Measurement Results Files by clicking <File Info>
(please refer to Figure 2-32).
3030.3059.72-02.00 60
Page 63
Operating with ARGUS-UMS UMS100/120/120 System Manual
Figure 2-31: Measurement result files on UMS100/120
Figure 2-32: Corresponding File Info
61 3030.3059.72-02.00
Page 64
UMS100/120/120 System Manual Operating with ARGUS-UMS
Click on <Move> to transfer the marked files to the control PC.
•
Figure 2-33: File transfer to the Control PC
Once the transfer is complete, the files are deleted on the UMS100/120, thus making more memory
available.
Should you inadvertently copy very large files (particularly audio) which cannot be transferred within an
acceptable time due to a slower GSM connection, click on <Delete> to erase the files directly.
After transferring, the files are located on the control PC under System Files. You can open the files as
text or as a graph and export them to another program for processing, if required. Click <Replay..> to
playback audio recordings. You can also save the files as Backup Files.
Figure 2-34: Analyzing and processing measurement data
3030.3059.72-02.00 62
Page 65
Operating with ARGUS-UMS UMS100/120/120 System Manual
2.5.1 Audio playback
Audio records can be replayed by clicking on the button <Replay> in the Navigator. By clicking on the <Preview>
button the graphic “Replay Level XY” is switched on. Click on “A” to listen to the audio.
The following recording was made by use of the squelch function, so that recordings of walky-talkies are reduced to
the pure audio.
Figure 2-35: Audio playback
2.6 ORM_ADC interface to other SW user interfaces
Some monitoring system users already have a customized user interface geared to their specific
purposes which it is essential to keep. For these situations there is a fitting solution by using O
eport Module with ARGUS Data Channel. This is available on the control PC for the remote control of
R
all the UMS100/120 connected up in the network. Simple commands are transferred in XML files to an
input folder in the control PC via LAN and then processed by the system. The results are then put into
an output folder ready for collection. You can forward live measurement results to the higher level
system by “UDP data stream” and then the data is displayed in the external application.
A detailed description of the ORM_ADC interface is provided in a special document.
rder
63 3030.3059.72-02.00
Page 66
UMS100/120/120 System Manual Maintenance
3 Maintenance
3.1 Firmware update
Sometimes it is necessary to update the Firmware in the UMS100/120. Open the function in the
Service Tool which is started by the web browser and entering the specific IP address.
The Firmware is either provided by Rohde & Schwarz Service on a CD or in some cases it can be
downloaded from the following Internet address:
http://www.ums.rohde-schwarz.com
CAUTION
In many cases, it is necessary to update the version of the control software on the control
PC before a Firmware update on the UMS100/120 is performed; otherwise operation is no
longer possible.
Before the Firmware update, clarify with each UMS100/120 whether it is compatible with
the relevant ARGUS-UMS version or whether a Service Pack must be obtained.
The update must be performed via a LAN/WAN interface as the entire Firmware is always transferred
(currently approx. 23.5 MB). Updating a UMS100/120 via GSM is actually not possible, because this
involves a downloading period of almost eight hours but the Service Tool is already shut down after 30
minutes for safety reasons.
Figure 3-1: Service Tool Firmware Update
3030.3059.72-02.00 64
Page 67
Maintenance UMS100/120/120 System Manual
3.2 Cleaning the hard drive and restarting
UMS100/120
In rare cases, the memory needed for the measurement data is not properly released. The
UMS100/120 then is sending an alarm message via the general Log File informing the operator that
nly a small percentage of the memory is available. If moving or deletion of measurement files is not
o
successful and the message still appears, it might be necessary to perform a memory cleanup.
Open the Service Tool. Go to Maintenance and click on <Execute> that comes after Cleanup Data
Area.
CAUTION
If Clean up Data Area has been executed, all saved measurement results on the
UMS100/120 are lost.
If the problem has not been rectified after that, a restart is recommended.
To restart, click <Execute> that comes after Restart UMS100/120.
•
Confirm the following question. The UMS100/120 is then not available for a several minutes while it
•
shuts down and restarts.
CAUTION
If a connection is made again from ARGUS-UMS, you may receive a warning that there is a
Mismatch in the configuration of the control and measurement unit. This is normal and you
can receive the configuration by clicking on <Receive…>
Figure 3-2: Service Tool - Maintenance
65 3030.3059.72-02.00
Page 68
UMS100/120/120 System Manual Maintenance
3.3 System support
In case of serious problems coming up during operation of UMS100/120 please contact the following
address via mail:
ums.support@rohde-schwarz.com
or by phone under:
(+49) 89 4129 – 12194
Further information may be available on the following Web Site:
www.ums.rohde-schwarz.com
For a comprehensive service or for hardware faults (the UMS100/120G can no longer be contacted or
the fans do not function despite the fuses being fully functional) please contact an authorized Rohde &
Schwarz service centre or the Rohde & Schwarz central service office.
3030.3059.72-02.00 66
Page 69
Appendix A UMS100 System Ma nual
Appendix A
System/network configuration table
As an example, the following table describes the configuration of a small monitoring network.
The default settings for all UMS100 are shown in the first line on a gray background. The tables describes a mixed connection of both UMS100s via LAN/WAN and via GSM mobile network. The configuration that is not required is in gray. It
is advisable to enter these settings beforehand. Otherwise, a service done locally on the LAN interface of a UMS100, connected by GSM, must still use the default IP address even though the login parameter is not default. This could be
confusing for service personnel.
Furthermore, it makes sense for any WAN connections that might be insecure to also have a parallel GSM connection (i.e. using a GSM SIM card). Therefore you have the option to reconfigure the backup connection (GMS) on the control
PC at any time.
Appendix 3030.3059.72-01.00
Page 70
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.