The user documentation for the R&SFSV/FSVA is divided as follows:
●
Quick Start Guide
●
Operating Manuals for base unit and options
●
Service Manual
●
Online Help
●
Release Notes
Quick Start Guide
This manual is delivered with the instrument in printed form and in PDF format on the
CD. It provides the information needed to set up and start working with the instrument.
Basic operations and basic measurements are described. Also a brief introduction to
remote control is given. The manual includes general information (e.g. Safety Instructions) and the following chapters:
Chapter 1Introduction, General information
Chapter 2Front and Rear Panel
Chapter 3Preparing for Use
Chapter 4Firmware Update and Installation of Firmware Options
Chapter 5Basic Operations
Chapter 6Basic Measurement Examples
Chapter 7Brief Introduction to Remote Control
AppendixLAN Interface
Operating Manuals
The Operating Manuals are a supplement to the Quick Start Guide. Operating Manuals
are provided for the base unit and each additional (software) option.
The Operating Manual for the base unit provides basic information on operating the
R&S FSV/FSVA in general, and the "Spectrum" mode in particular. Furthermore, the
software options that enhance the basic functionality for various measurement modes
are described here. The set of measurement examples in the Quick Start Guide is
expanded by more advanced measurement examples. In addition to the brief introduction to remote control in the Quick Start Guide, a description of the basic analyzer commands and programming examples is given. Information on maintenance, instrument
interfaces and error messages is also provided.
In the individual option manuals, the specific instrument functions of the option are
described in detail. For additional information on default settings and parameters, refer
5Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
to the data sheets. Basic information on operating the R&S FSV/FSVA is not included
in the option manuals.
The following Operating Manuals are available for the R&S FSV/FSVA:
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
Preface
Documentation Overview
R&S FSV/FSVA base unit; in addition:
–R&S FSV-K9 Power Sensor Support
–R&S FSV-K14 Spectrogram Measurement
R&S FSV-K7 Analog Demodulation and R&S FSV-K7S FM Stereo Measurements
R&S FSV-K10 GSM/EDGE Measurement
R&S FSV-K30 Noise Figure Measurement
R&S FSV-K40 Phase Noise Measurement
R&S FSV-K70 Vector Signal Analysis Operating Manual
These manuals are available in PDF format on the CD delivered with the instrument.
Service Manual
This manual is available in PDF format on the CD delivered with the instrument. It
describes how to check compliance with rated specifications, instrument function,
repair, troubleshooting and fault elimination. It contains all information required for
repairing the R&S FSV/FSVA by replacing modules. The manual includes the following
chapters:
Chapter 1Performance Test
Chapter 2Adjustment
Chapter 3Repair
Chapter 4Software Update / Installing Options
Chapter 5Documents
Online Help
The online help contains context-specific help on operating the R&S FSV/FSVA and all
available options. It describes both manual and remote operation. The online help is
installed on the R&S FSV/FSVA by default, and is also available as an executable .chm file on the CD delivered with the instrument.
6Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
Release Notes
The release notes describe the installation of the firmware, new and modified functions, eliminated problems, and last minute changes to the documentation. The corresponding firmware version is indicated on the title page of the release notes. The current release notes are provided in the Internet.
Preface
Conventions Used in the Documentation
1.2Conventions Used in the Documentation
1.2.1Typographical Conventions
The following text markers are used throughout this documentation:
ConventionDescription
"Graphical user interface elements"
KEYSKey names are written in capital letters.
File names, commands,
program code
InputInput to be entered by the user is displayed in italics.
LinksLinks that you can click are displayed in blue font.
"References"References to other parts of the documentation are enclosed by quota-
All names of graphical user interface elements on the screen, such as
dialog boxes, menus, options, buttons, and softkeys are enclosed by
quotation marks.
File names, commands, coding samples and screen output are distinguished by their font.
tion marks.
1.2.2Conventions for Procedure Descriptions
When describing how to operate the instrument, several alternative methods may be
available to perform the same task. In this case, the procedure using the touchscreen
is described. Any elements that can be activated by touching can also be clicked using
an additionally connected mouse. The alternative procedure using the keys on the
instrument or the on-screen keyboard is only described if it deviates from the standard
operating procedures.
The term "select" may refer to any of the described methods, i.e. using a finger on the
touchscreen, a mouse pointer in the display, or a key on the instrument or on a keyboard.
1.2.3Notes on Screenshots
When describing the functions of the product, we use sample screenshots. These
screenshots are meant to illustrate as much as possible of the provided functions and
possible interdependencies between parameters.
7Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
The screenshots usually show a fully equipped product, that is: with all options installed. Thus, some functions shown in the screenshots may not be available in your particular product configuration.
Preface
How to Use the Help System
1.3How to Use the Help System
Calling context-sensitive and general help
► To display the general help dialog box, press the HELP key on the front panel.
The help dialog box "View" tab is displayed. A topic containing information about
the current menu or the currently opened dialog box and its function is displayed.
For standard Windows dialog boxes (e.g. File Properties, Print dialog etc.), no contextsensitive help is available.
► If the help is already displayed, press the softkey for which you want to display
help.
A topic containing information about the softkey and its function is displayed.
If a softkey opens a submenu and you press the softkey a second time, the submenu
of the softkey is displayed.
Contents of the help dialog box
The help dialog box contains four tabs:
●
"Contents" - contains a table of help contents
●
"View" - contains a specific help topic
●
"Index" - contains index entries to search for help topics
●
"Zoom" - contains zoom functions for the help display
To change between these tabs, press the tab on the touchscreen.
Navigating in the table of contents
●
To move through the displayed contents entries, use the UP ARROW and DOWN
ARROW keys. Entries that contain further entries are marked with a plus sign.
●
To display a help topic, press the ENTER key. The "View" tab with the corresponding help topic is displayed.
●
To change to the next tab, press the tab on the touchscreen.
Navigating in the help topics
●
To scroll through a page, use the rotary knob or the UP ARROW and DOWN
ARROW keys.
8Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
●
Searching for a topic
1. Change to the "Index" tab.
2. Enter the first characters of the topic you are interested in. The entries starting with
3. Change the focus by pressing the ENTER key.
4. Select the suitable keyword by using the UP ARROW or DOWN ARROW keys or
5. Press the ENTER key to display the help topic.
Changing the zoom
1. Change to the "Zoom" tab.
Preface
Notes for Users of R&S FSV 1307.9002Kxx Models
To jump to the linked topic, press the link text on the touchscreen.
these characters are displayed.
the rotary knob.
The "View" tab with the corresponding help topic is displayed.
2. Set the zoom using the rotary knob. Four settings are available: 1-4. The smallest
size is selected by number 1, the largest size is selected by number 4.
Closing the help window
► Press the ESC key or a function key on the front panel.
1.4Notes for Users of R&S FSV 1307.9002Kxx Models
Users of R&S FSV 1307.9002Kxx models should consider the following differences to
the description of the newer R&S FSV/FSVA 1321.3008Kxx models:
●
Functions that are based on the Windows7 operating system (e.g. printing or setting up networks) may have a slightly different appearance or require different settings on the Windows XP based models. For such functions, refer to the Windows
documentation or the documentation originally provided with the R&S FSV instrument.
●
The R&S FSV 1307.9002K03 model is restricted to a maximum frequency of
3 GHz, whereas the R&S FSV/FSVA1321.3008K04 model has a maximum frequency of 4 GHz.
●
The bandwidth extension option R&S FSV-B160 (1311.2015.xx) is not available for
the R&S FSV 1307.9002Kxx models. The maximum usable I/Q analysis bandwidth
for these models is 28 MHz, or with option R&S FSV-B70, 40 MHz.
9Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
2Introduction
Overview of Firmware Options R&S FSV-K84/-K85
This section contains all information required for operation of an R&S FSV/FSVA
equipped with Application Firmware R&S FSV-K84 or -K85. It covers operation via
menus and the remote control commands for the 1xEV-DO Analysis.
This part of the documentation consists of the following chapters:
●
●
●
●
●
●
Introduction
Chapter 3, "Measurement Examples for the 1xEV-DO BTS Analysis (K84)",
on page 11
Explains some basic 1xEV-DO base station tests.
Chapter 4, "Measurement Examples for the 1xEV-DO MS Analysis (K85)",
on page 20
Explains some basic 1xEV-DO mobile station tests.
Chapter 5, "Test Setup for Base Station and Mobile Station Tests", on page 29
Describes the measurement setup for base station and mobile station tests.
Chapter 6, "Instrument Functions of the 1xEV-DO Analysis", on page 31
Describes the instrument functions of 1xEV-DO Analysis.
Chapter 7, "Remote Commands of the 1xEV-DO Analysis", on page 199
Describes all remote control commands defined for the code domain measurement. An alphabetic list of all remote control commands and a table of softkeys
with the assignment of commands are provided at the end of this chapter.
Chapter 8, "Status Reporting System of the 1xEV-DO Analysis", on page 345
Contains device-specific error messages for R&S FSV-K84/-K85.
This part of the documentation includes only functions of the Application Firmware
R&S FSV-K84/-K85. For all other descriptions, please refer to the description of the
base unit.
10Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
3Measurement Examples for the 1xEV-DO
BTS Analysis (K84)
Measurement Examples for the 1xEV-DO BTS Analysis (K84)
Measuring the Signal Channel Power
3.1Measuring the Signal Channel Power
In the Power measurement, the total channel power of the 1xEV-DO signal is displayed. The measurement also displays spurious emissions like harmonics or intermodulation products that occur close to the carrier.
Test setup:
●
Connect the RF output of the signal generator to the RF input of the R&S FSV/
FSVA (coaxial cable with N connectors).
Signal generator settings:
Frequency: 878.49 MHz
Level: 0 dBm
Standard: 1xEV-DO BTS
Procedure:
1. Set the R&S FSV/FSVA to its default state.
a) Press the PRESET key.
2. Activate the "1xEV-DO BTS Analysis" mode.
a) Press the MODE key and activate the "1xEV-DO BTS Analysis" option.
3. Start the Power measurement
a) Press the MEAS key.
b) Press the "Power" softkey.
4. Set the center frequency.
a) Press the FREQ key and enter 878.49 MHz.
5. Set the reference level.
a) Press the AMPT key and enter 0 dBm.
On the screen, the spectrum of the signal and the corresponding power levels within
the 1.2288 MHz channel bandwidth are displayed. In the table below the diagram, the
numeric values of the channel bandwidth of the TX Channel and power level of the
analyzed signal are listed.
11Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
Measurement Examples for the 1xEV-DO BTS Analysis (K84)
Measuring the Spectrum Emission Mask
3.2Measuring the Spectrum Emission Mask
To detect spurious emissions such as harmonics or intermodulation products, the
R&S FSV/FSVA offers a spectrum emission mask measurement. The measurement
compares the power against the spectrum emission mask in the range from -4 MHz to
4 MHz around the carrier. The exact measurement settings like the filter that is used
depend on the Band Class parameter. For a list of supported bandclasses refer to the
"Bandclass"on page 124 softkey in the "Spectrum Emission Mask" menu.
Test setup:
●
Connect the RF output of the signal generator to the RF input of the R&S FSV/
FSVA (coaxial cable with N connectors).
Signal generator settings:
Frequency: 878.49 MHz
Level: 0 dBm
Standard: 1xEV-DO BTS
Procedure:
1. Set the R&S FSV/FSVA to its default state.
a) Press the PRESET key.
2. Select the "1xEV-DO BTS Analysis" mode.
a) Press the MODE key and select the "1xEV-DO BTS Analysis" option.
3. Start the measurement.
a) Press the MEAS key.
b) Press the "Spectrum Emission Mask" softkey.
4. Set the center frequency.
a) Press the FREQ key and enter 878.49 MHz.
5. Set the reference level.
a) Press the AMPT key and enter 0 dBm.
6. Select a bandclass
a) Press the "Bandclass" softkey and select BandClass 0: 800 MHz Cellular Band
from the list.
On the screen, the spectrum of the signal is displayed, including the limit line defined in
the standard. To understand where and about how much the measurement has failed,
the List Evaluation table shows the frequencies where spurious emissions occur.
12Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
Measurement Examples for the 1xEV-DO BTS Analysis (K84)
Measuring the Relative Code Domain Power and the Frequency Error
3.3Measuring the Relative Code Domain Power and the
Frequency Error
A Code Domain Power measurement analyses the signal over a single slot. It also
determines the power of all codes and channels.
The following examples show a Code Domain Power measurement on a test model
with 9 channels. In this measurement, changing some parameters one after the other
should demonstrate the resulting effects: values adapted to the measurement signal
are changed to non-adapted values.
In the following examples, adjusting the settings of the code domain measurements is
described using the dialog boxes. Alternatively, most of the settings can also be modified by using the corresponding hardkeys as in the base unit (e.g. the center frequency
can be either set in the "Frontend Settings" dialog box, or via the FREQ key).
Test setup:
●
Connect the RF output of the signal generator to the RF input of the R&S FSV/
FSVA.
●
Connect the reference input (EXT REF) on the rear panel of the R&S FSV/FSVA to
the reference output (REF) of the signal generator (coaxial cable with BNC connectors).
Signal generator settings:
Frequency: 878.49 MHz
Level: 0 dBm
Standard: 1xEV-DO BTS
Procedure:
1. Set the R&S FSV/FSVA to its default state.
a) Press the PRESET key.
2. Select the "1xEV-DO BTS Analysis" mode.
a) Press the MODE key and select "1xEV-DO BTS Analysis".
3. Select the Code Domain Analyzer.
a) Press the MEAS key
b) Press the "Code Domain Analyzer" softkey.
4. Start the measurement
a) In the "Code Domain Analyzer" menu, press the "Display Config" softkey.
b) Select the "Code Domain Power" measurement.
5. Set the center frequency and the reference level.
a) In the "Code Domain Analyzer" menu, press the "Frontend Settings" softkey.
13Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
In the two screens, the following results are displayed: screen A shows the power of
the code domain of the signal. The x-axis represents the individual channels (or
codes), while the y-axis shows the power of each channel.
In screen B the result summary is displayed. It shows the numeric results of the code
domain power measurement, including the frequency error.
By default, the R&S FSV-K84 displays two measurement screens. After a preset the
first (screen A) is always the Code Domain Power result display. The second (screen
B) is always the Result Display
For more information on the display concept refer to the "Display Config"on page 85
softkey.
Synchronization of the reference frequencies
Measurement Examples for the 1xEV-DO BTS Analysis (K84)
Measuring the Relative Code Domain Power and the Frequency Error
b) In the "Center Frequency" field enter 878.49 MHz.
c) In the "Ref Level" field enter 10 dBm.
d) Close the "Frontend Settings" dialog box.
The frequency error can be reduced by synchronizing the transmitter and the receiver
to the same reference frequency.
► Press the SETUP key.
a) Press the "Reference Int/Ext" softkey to switch to an external reference.
Screen A again shows the Code Domain Power measurement and screen B the result
summary. After the synchronization of the reference frequencies of the devices, the
frequency error should now be smaller than 10 Hz.
Behavior with deviating center frequency setting
A measurement can only be valid if the center frequency of the DUT and the analyzer
are balanced.
► On the signal generator, change the center frequency in steps of 0.1 kHz and
observe the analyzer screen.
Up to a frequency error of approximately 1.0 kHz, a Code Domain Power measurement on the R&S FSV/FSVA is still possible. A frequency error within this range
causes no apparent difference in the accuracy of the Code Domain Power measurement.
In case of a frequency error of more than 1.0 kHz, the probability of incorrect synchronization increases. This is indicated by the SYNC FAILED error message.
If the frequency error exceeds approximately 1.5 kHz, a Code Domain Power measurement cannot be performed. This is also indicated by the SYNC FAILED error
message.
Reset the center frequency of the signal generator to 878.49 MHz.
14Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
The center frequency of the DUT should not deviate by more than 1.0 kHz from that of
the R&S FSV/FSVA
Measurement Examples for the 1xEV-DO BTS Analysis (K84)
Measuring the Triggered Relative Code Domain Power
3.4Measuring the Triggered Relative Code Domain
Power
If the code domain power measurement is performed without external triggering, a section of the test signal is recorded at an arbitrary point of time and the firmware attempts
to detect the start of a slot. To detect this start, all possibilities of the PN sequence
location have to be tested in Free Run trigger mode. This requires computing time.
This computing time can be reduced by using an external (frame) trigger and entering
the correct PN offset. If the search range for the start of the power control group and
the PN offset are known then fewer possibilities have to be tested. This increases the
measurement speed.
Test setup:
●
Connect the RF output of the signal generator to the input of the R&SFSV/FSVA.
●
Connect the reference input (EXT REF) on the rear panel of the R&S FSV/FSVA to
the reference input of the signal generator (coaxial cable with BNC connectors).
●
Connect the external trigger input on the rear panel of the R&S FSV/FSVA (EXT
TRIGGER/GATE IN) to the external trigger output of the signal generator.
Signal generator settings (e.g. R&S SMU):
Frequency: 878.49 MHz
Level: 0 dBm
Standard: 1xEV-DO BTS
Procedure:
1. Set the R&S FSV/FSVA to its default state.
a) Press the PRESET key.
2. Select the "1xEV-DO BTS Analysis" Mode.
a) Press the MODE key and select 1xEV-DO BTS Analysis.
3. Select the Code Domain Analyzer.
a) Press the MEAS key
b) Press the "Code Domain Analyzer" softkey.
4. Start the measurement.
a) In the "Code Domain Analyzer" menu, press the "Display Config" softkey.
b) Select the tab for Screen A.
c) Select the "Code Domain Power" measurement.
15Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
5. Set the center frequency and the reference level.
In the two screens, the following results are displayed: by default, screen A shows the
code domain power of the signal. Compared to the measurement without an external
trigger (see previous example), the repetition rate of the measurement increases. In
screen B the result summary is displayed. In the row Trigger to Frame, the offset
between the trigger event and and the start of the slot is shown.
Measurement Examples for the 1xEV-DO BTS Analysis (K84)
Measuring the Triggered Relative Code Domain Power
a) In the "Code Domain Analyzer" menu, press the "Frontend Settings" softkey.
b) In the "Center Frequency" field enter 878.49 MHz.
c) In the "Ref Level" field enter 10 dBm.
d) Close the "Frontend Settings" dialog box.
3.4.1Adjusting the Trigger Offset
The delay between the trigger event and the start of the slot can be compensated for
by adjusting the trigger offset.
●
Set an external trigger source and the trigger offset.
–Open the IQ Capture dialog box.
–Set the "Trigger Source" option to "External".
–Set the "Trigger Offset" to 100µs to compensate analog delays of the trigger
event.
In the two screens, the following results are displayed: Screen A shows the the same
as above. In screen B the result summary is displayed. In the Trigger to Frame result,
the offset between the trigger event and the start of the slot has been adjusted.
3.4.2Behaviour With the Wrong PN Offset
The last adjustment is setting the PN (Pseudo Noise) offset correctly. The measurement can only be valid, if the PN offset on the analyzer is the same as that of the transmit signal.
●
Set a PN Offset.
–Open the "Result Settings" dialog box.
–In the "PN Offset" field enter 200.
Again, screen A shows the CDP measurement, screen B the result summary.
In the result summary, the Trigger to Frame result is not correct. Also, the error
message SYNC FAILED indicates that the synchronization has failed.
–In the "PN Offset" field enter 0.
After adjusting it, the PN offset on the R&S FSV/FSVA is the same as that of the signal. In the result summary the "Trigger to Frame" value is now shown correctly.
16Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
Measurement Examples for the 1xEV-DO BTS Analysis (K84)
Measuring the Composite EVM
3.5Measuring the Composite EVM
The Error Vector Magnitude (EVM) describes the quality of the measured signal compared to an ideal reference signal generated by the R&S FSV/FSVA. In the I-Q plane,
the error vector represents the ratio of the measured signal to the ideal signal on symbol level. The error vector is equal to the square root of the ratio of the measured signal to the reference signal. The result is given in %.
In the Composite EVM measurement the error is averaged over all channels (by
means of the root mean square) for a given slot. The measurement covers the entire
signal during the entire observation time. On screen the results are shown in a diagram, in which the x-axis represents the examined slots and the y-axis shows the EVM
values.
Test Setup:
●
Connect the RF output of the Signal Generator to the RF input of the R&S FSV/
FSVA. (coaxial cables with N connectors).
●
Connect the reference input (EXT REF IN/OUT) on the rear panel of the R&S FSV/
FSVA to the reference output (REF) on the signal generator (coaxial cable with
BNC connectors).
●
Connect external triggering of the analyzer (EXT TRIG GATE) to the signal generator’s trigger (TRIGOUT1 at PAR DATA).
Signal generator settings:
Frequency: 878.49 MHz
Level: 0 dBm
Standard: 1xEV-DO BTS
Procedure:
1. Set the R&S FSV/FSVA to its default state.
a) Press the PRESET key.
2. Select the "1xEV-DO BTS Analysis" Mode.
a) Press the MODE key and select 1xEV-DO BTS Analysis.
3. Select the Code Domain Analyzer.
a) Press the MEAS key
b) Press the "Code Domain Analyzer" softkey.
4. Start the measurement.
a) Press the "Display Config" softkey.
b) Select the tab for Screen A.
c) Select the "Composite EVM" measurement.
5. Set the center frequency and the reference level.
17Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
6. Set an external trigger source.
In the two screens, the following results are displayed: by default, Screen A shows the
diagram of the Composite EVM measurement result. In screen B the result summary is
displayed. It shows the numeric results of the Code Domain Power measurement,
including the values for the Composite EVM.
Measurement Examples for the 1xEV-DO BTS Analysis (K84)
Measuring the Peak Code Domain Error
a) Open the "Frontend Settings" dialog box.
b) In the "Center Frequency" field enter 878.49 MHz.
c) In the "Ref Level" field enter 10 dBm.
d) Close the "Frontend Settings" dialog box.
a) Open the "IQ Capture Settings" dialog box.
b) Set the "Trigger Source" option to "External".
3.6Measuring the Peak Code Domain Error
The Code Domain Error Power describes the quality of the measured signal compared
to an ideal reference signal generated by the R&S FSV/FSVA. In the I-Q plane, the
error vector represents the difference of the measured signal and the ideal signal. The
Code Domain Error is the difference in power on symbol level of the measured and the
reference signal projected to the class of of the base spreading factor. The unit of the
result is dB.
In the Peak Code Domain Error (PCDE) measurement, the maximum error value over
all channels is determined and displayed for a given slot. The measurement covers the
entire signal during the entire observation time. On screen the results are shown in a
diagram, in which the x-axis represents the slots and the y-axis shows the PCDE values.
A measurement of the RHO factor is shown in the second part of the example. RHO is
the normalized, correlated power between the measured and the ideal reference signal. The maximum value of RHO is 1. In that case the measured signal and the reference signal are identical. When measuring RHO, it is required that only the pilot channel is active.
Test setup:
●
Connect the RF output of the signal generator to the RF input of the R&S FSV/
FSVA (coaxial cable with N connectors).
●
Connect the reference input (EXT REF IN/OUT) on the rear panel of the R&S FSV/
FSVA to the reference output (REF) on the signal generator (coaxial cable with
BNC connectors).
●
Connect external triggering of the R&S FSV/FSVA (EXT TRIG GATE) to the signal
generator trigger (TRIGOUT1 at PAR DATA).
Signal generator settings:
Frequency: 878.49 MHz
18Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
Level: 0 dBm
Standard: 1xEV-DO BTS
Procedure:
1. Set the R&S FSV/FSVA to its default state.
2. Select the "1xEV-DO BTS Analysis" mode.
3. Select the Code Domain Analyzer.
4. Start the Peak Code Domain Error measurement.
Measurement Examples for the 1xEV-DO BTS Analysis (K84)
Measuring the Peak Code Domain Error
a) Press the PRESET key.
a) Press the MODE key and select the "1xEV-DO BTS Analysis" option.
a) Press the MEAS key.
b) Press the "Code Domain Analyzer" softkey.
a) Press the "Display Config" softkey
b) Select the tab for Screen A.
c) Select the "Peak Code Domain Error" softkey and start the measurement.
5. Set the center frequency and the reference level.
a) Open the "Frontend Settings" dialog box.
b) In the "Center Frequency" field enter 878.49 MHz.
c) In the "Ref Level" field enter 0 dBm.
d) Close the "Frontend Settings" dialog box.
6. Set an external trigger source.
a) Open the "IQ Capture Settings" dialog box.
b) Set the "Trigger Source" option to "External".
In the two screens, the following results are displayed: by default, screen A shows the
diagram of the Peak Code Domain Error. In screen B the result summary is displayed.
Displaying RHO
Make sure that all channels except the pilot channel (code 0.64) are OFF, so that only
the pilot channel is available in the measurement.
No specific measurement is required to get the value for RHO. The R&S FSV/FSVA
always calculates this value automatically regardless of the code domain measurement
performed. Besides the results of the code domain measurements, the numeric result
of the RHO measurement is shown in the result summary, by default shown in screen
B.
19Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
4Measurement Examples for the 1xEV-DO
MS Analysis (K85)
Measurement Examples for the 1xEV-DO MS Analysis (K85)
Measuring the Signal Channel Power
4.1Measuring the Signal Channel Power
In the Power measurement, the total channel power of the 1xEV-DO signal is displayed. The measurement also displays spurious emissions like harmonics or intermodulation products that occur close to the carrier.
Test setup:
●
Connect the RF output of the signal generator to the RF input of the R&S FSV/
FSVA (coaxial cable with N connectors).
Signal generator settings:
Frequency: 833.49 MHz
Level: 0 dBm
Standard: 1xEV-DO MS MS
Procedure:
1. Set the R&S FSV/FSVA to its default state.
a) Press the PRESET key.
2. Activate the "1xEV-DO MS Analysis" mode.
a) Press the MODE key and activate the "1xEV-DO MS Analysis" option.
3. Start the Power measurement
a) Press the MEAS key.
b) Press the "Power" softkey.
4. Set the center frequency.
a) Press the FREQ key and enter 833.49 MHz.
5. Set the reference level.
a) Press the AMPT key and enter 0 dBm.
On the screen, the spectrum of the signal and the corresponding power levels within
the 1.2288 MHz channel bandwidth are displayed. In the table below the diagram, the
numeric values of the channel bandwidth of the TX Channel and power level of the
analyzed signal are listed.
20Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
Measurement Examples for the 1xEV-DO MS Analysis (K85)
Measuring the Spectrum Emission Mask
4.2Measuring the Spectrum Emission Mask
To detect spurious emissions such as harmonics or intermodulation products, the
R&S FSV/FSVA offers a spectrum emission mask measurement. The measurement
compares the power against the spectrum emission mask in the range from -4 MHz to
4 MHz around the carrier. The exact measurement settings like the filter that is used
depend on the Band Class parameter. For a list of supported bandclasses refer to the
"Bandclass"on page 124 softkey in the "Spectrum Emission Mask" menu.
Test setup:
●
Connect the RF output of the signal generator to the RF input of the R&S FSV/
FSVA (coaxial cable with N connectors).
Signal generator settings:
Frequency: 833.49 MHz
Level: 0 dBm
Standard: 1xEV-DO MS
Procedure:
1. Set the R&S FSV/FSVA to its default state.
a) Press the PRESET key.
2. Select the "1xEV-DO MS Analysis" mode.
a) Press the MODE key and select the "1xEV-DO MS Analysis" option.
3. Start the measurement.
a) Press the MEAS key.
b) Press the "Spectrum Emission Mask" softkey.
4. Set the center frequency.
a) Press the FREQ key and enter 833.49 MHz.
5. Set the reference level.
a) Press the AMPT key and enter 0 dBm.
6. Select a bandclass
a) Press the "Bandclass" softkey and select BandClass 0: 800 MHz Cellular Band
from the list.
On the screen, the spectrum of the signal is displayed, including the limit line defined in
the standard. To understand where and about how much the measurement has failed,
the List Evaluation table shows the frequencies where spurious emissions occur.
21Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
Measurement Examples for the 1xEV-DO MS Analysis (K85)
Measuring the Relative Code Domain Power and the Frequency Error
4.3Measuring the Relative Code Domain Power and the
Frequency Error
A Code Domain Power measurement analyses the signal over a single slot. It also
determines the power of all codes and channels.
The following examples show a Code Domain Power measurement on a test model
with 9 channels. In this measurement, changing some parameters one after the other
should demonstrate the resulting effects: values adapted to the measurement signal
are changed to non-adapted values.
In the following examples, adjusting the settings of the code domain measurements is
described using the dialog boxes. Alternatively, most of the settings can also be modified by using the corresponding hardkeys as in the base unit (e.g. the center frequency
can be either set in the "Frontend Settings" dialog box, or via the FREQ key).
Test setup:
●
Connect the RF output of the signal generator to the RF input of the R&S FSV/
FSVA.
●
Connect the reference input (EXT REF) on the rear panel of the R&S FSV/FSVA to
the reference output (REF) of the signal generator (coaxial cable with BNC connectors).
Signal generator settings:
Frequency: 833.49 MHz
Level: 0 dBm
Standard: 1xEV-DO MS
Procedure:
1. Set the R&S FSV/FSVA to its default state.
a) Press the PRESET key.
2. Select the "1xEV-DO MS Analysis" mode.
a) Press the MODE key and select "1xEV-DO MS Analysis".
3. Select the Code Domain Analyzer.
a) Press the MEAS key
b) Press the "Code Domain Analyzer" softkey.
4. Start the measurement
a) In the "Code Domain Analyzer" menu, press the "Display Config" softkey.
b) Select the "Code Domain Power" measurement.
5. Set the center frequency and the reference level.
a) In the "Code Domain Analyzer" menu, press the "Frontend Settings" softkey.
22Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
In the two screens, the following results are displayed: screen A shows the power of
the code domain of the signal. The x-axis represents the individual channels (or
codes), while the y-axis shows the power of each channel.
In screen B the result summary is displayed. It shows the numeric results of the code
domain power measurement, including the frequency error.
Synchronization of the reference frequencies
The frequency error can be reduced by synchronizing the transmitter and the receiver
to the same reference frequency.
► Press the SETUP key.
Measurement Examples for the 1xEV-DO MS Analysis (K85)
Measuring the Relative Code Domain Power and the Frequency Error
b) In the "Center Frequency" field enter 833.49 MHz.
c) In the "Ref Level" field enter 10 dBm.
d) Close the "Frontend Settings" dialog box.
a) Press the "Reference Int/Ext" softkey to switch to an external reference.
Screen A again shows the Code Domain Power measurement and screen B the result
summary. After the synchronization of the reference frequencies of the devices, the
frequency error should now be smaller than 10 Hz.
Behavior with deviating center frequency setting
A measurement can only be valid if the center frequency of the DUT and the analyzer
are balanced.
► On the signal generator, change the center frequency in steps of 0.1 kHz and
observe the analyzer screen.
Up to a frequency error of approximately 4.0 kHz, a Code Domain Power measurement on the R&S FSV/FSVA is still possible. A frequency error within this range
causes no apparent difference in the accuracy of the Code Domain Power measurement.
In case of a frequency error of more than 4.3 kHz, the probability of incorrect synchronization increases. This is indicated by the SYNC FAILED error message.
If the frequency error exceeds approximately 4.3 kHz, a Code Domain Power measurement cannot be performed. This is also indicated by the SYNC FAILED error
message.
Reset the center frequency of the signal generator to 833.49 MHz.
The center frequency of the DUT should not deviate by more than 4.0 kHz from that of
the R&S FSV/FSVA.
23Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
Measurement Examples for the 1xEV-DO MS Analysis (K85)
Measuring the Triggered Relative Code Domain Power
4.4Measuring the Triggered Relative Code Domain
Power
If the code domain power measurement is performed without external triggering, a section of the test signal is recorded at an arbitrary point of time and the firmware attempts
to detect the start of a slot. To detect this start, all possibilities of the PN sequence
location have to be tested in Free Run trigger mode. This requires computing time.
This computing time can be reduced by using an external (frame) trigger and entering
the correct PN offset. If the search range for the start of the power control group and
the PN offset are known then fewer possibilities have to be tested. This increases the
measurement speed.
Test setup:
●
Connect the RF output of the signal generator to the input of the R&SFSV/FSVA.
●
Connect the reference input (EXT REF) on the rear panel of the R&S FSV/FSVA to
the reference input of the signal generator (coaxial cable with BNC connectors).
●
Connect the external trigger input on the rear panel of the R&S FSV/FSVA (EXT
TRIGGER/GATE IN) to the external trigger output of the signal generator.
Signal generator settings (e.g. R&S SMU):
Frequency: 833.49 MHz
Level: 0 dBm
Standard: 1xEV-0DO
Procedure:
1. Set the R&S FSV/FSVA to its default state.
a) Press the PRESET key.
2. Select the "1xEV-DO MS Analysis" Mode.
a) Press the MODE key and select 1xEV-DO MS Analysis.
3. Select the Code Domain Analyzer.
a) Press the MEAS key
b) Press the "Code Domain Analyzer" softkey.
4. Start the measurement.
a) In the "Code Domain Analyzer" menu, press the "Display Config" softkey.
b) Select the tab for Screen A.
c) Select the "Code Domain Power" measurement.
5. Set the center frequency and the reference level.
a) In the "Code Domain Analyzer" menu, press the "Frontend Settings" softkey.
b) In the "Center Frequency" field enter 833.49 MHz.
c) In the "Ref Level" field enter 10 dBm.
24Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
In the two screens, the following results are displayed: by default, screen A shows the
code domain power of the signal. Compared to the measurement without an external
trigger (see previous example), the repetition rate of the measurement increases. In
screen B the result summary is displayed. In the row Trigger to Frame, the offset
between the trigger event and and the start of the slot is shown.
Measurement Examples for the 1xEV-DO MS Analysis (K85)
Measuring the Composite EVM
d) Close the "Frontend Settings" dialog box.
4.4.1Adjusting the Trigger Offset
The delay between the trigger event and the start of the slot can be compensated for
by adjusting the trigger offset.
●
Set an external trigger source and the trigger offset.
–Open the IQ Capture dialog box.
–Set the "Trigger Source" option to "External".
–Set the "Trigger Offset" to 100µs to compensate analog delays of the trigger
event.
In the two screens, the following results are displayed: Screen A shows the the same
as above. In screen B the result summary is displayed. In the Trigger to Frame result,
the offset between the trigger event and the start of the slot has been adjusted.
4.5Measuring the Composite EVM
The Error Vector Magnitude (EVM) describes the quality of the measured signal compared to an ideal reference signal generated by the R&S FSV/FSVA. In the I-Q plane,
the error vector represents the ratio of the measured signal to the ideal signal on symbol level. The error vector is equal to the square root of the ratio of the measured signal to the reference signal. The result is given in %.
In the Composite EVM measurement the error is averaged over all channels (by
means of the root mean square) for a given slot. The measurement covers the entire
signal during the entire observation time. On screen the results are shown in a diagram, in which the x-axis represents the examined slots and the y-axis shows the EVM
values.
Test Setup:
●
Connect the RF output of the Signal Generator to the RF input of the R&S FSV/
FSVA. (coaxial cables with N connectors).
●
Connect the reference input (EXT REF IN/OUT) on the rear panel of the R&S FSV/
FSVA to the reference output (REF) on the signal generator (coaxial cable with
BNC connectors).
●
Connect external triggering of the analyzer (EXT TRIG GATE) to the signal generator’s trigger (TRIGOUT1 at PAR DATA).
25Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
Signal generator settings:
Frequency: 833.49 MHz
Level: 0 dBm
Standard: 1xEV-DO MS
Procedure:
1. Set the R&S FSV/FSVA to its default state.
2. Select the "1xEV-DO MS Analysis" Mode.
3. Select the Code Domain Analyzer.
Measurement Examples for the 1xEV-DO MS Analysis (K85)
Measuring the Peak Code Domain Error
a) Press the PRESET key.
a) Press the MODE key and select 1xEV-DO MS Analysis.
a) Press the MEAS key
b) Press the "Code Domain Analyzer" softkey.
4. Start the measurement.
a) Press the "Display Config" softkey.
b) Select the tab for Screen A.
c) Select the "Composite EVM" measurement.
5. Set the center frequency and the reference level.
a) Open the "Frontend Settings" dialog box.
b) In the "Center Frequency" field enter 833.49 MHz.
c) In the "Ref Level" field enter 10 dBm.
d) Close the "Frontend Settings" dialog box.
6. Set an external trigger source.
a) Open the "IQ Capture Settings" dialog box.
b) Set the "Trigger Source" option to "External".
In the two screens, the following results are displayed: by default, Screen A shows the
diagram of the Composite EVM measurement result. In screen B the result summary is
displayed. It shows the numeric results of the Code Domain Power measurement,
including the values for the Composite EVM.
4.6Measuring the Peak Code Domain Error
The Code Domain Error Power describes the quality of the measured signal compared
to an ideal reference signal generated by the R&S FSV/FSVA. In the I-Q plane, the
error vector represents the difference of the measured signal and the ideal signal. The
Code Domain Error is the difference in power on symbol level of the measured and the
reference signal projected to the class of of the base spreading factor. The unit of the
result is dB.
26Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
In the Peak Code Domain Error (PCDE) measurement, the maximum error value over
all channels is determined and displayed for a given slot. The measurement covers the
entire signal during the entire observation time. On screen the results are shown in a
diagram, in which the x-axis represents the slots and the y-axis shows the PCDE values.
A measurement of the RHO factor is shown in the second part of the example. RHO is
the normalized, correlated power between the measured and the ideal reference signal. The maximum value of RHO is 1. In that case the measured signal and the reference signal are identical. When measuring RHO, it is required that only the pilot channel is active.
Test setup:
●
●
●
Measurement Examples for the 1xEV-DO MS Analysis (K85)
Measuring the Peak Code Domain Error
Connect the RF output of the signal generator to the RF input of the R&S FSV/
FSVA (coaxial cable with N connectors).
Connect the reference input (EXT REF IN/OUT) on the rear panel of the R&S FSV/
FSVA to the reference output (REF) on the signal generator (coaxial cable with
BNC connectors).
Connect external triggering of the R&S FSV/FSVA (EXT TRIG GATE) to the signal
generator trigger (TRIGOUT1 at PAR DATA).
Signal generator settings:
Frequency: 833.49 MHz
Level: 0 dBm
Standard: 1xEV-DO MS
Procedure:
1. Set the R&S FSV/FSVA to its default state.
a) Press the PRESET key.
2. Select the "1xEV-DO MS Analysis" mode.
a) Press the MODE key and select the "1xEV-DO MS Analysis" option.
3. Select the Code Domain Analyzer.
a) Press the MEAS key.
b) Press the "Code Domain Analyzer" softkey.
4. Start the Peak Code Domain Error measurement.
a) Press the "Display Config" softkey
b) Select the tab for Screen A.
c) Select the "Peak Code Domain Error" softkey and start the measurement.
5. Set the center frequency and the reference level.
a) Open the "Frontend Settings" dialog box.
b) In the "Center Frequency" field enter 833.49 MHz.
c) In the "Ref Level" field enter 0 dBm.
27Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
6. Set an external trigger source.
In the two screens, the following results are displayed: by default, screen A shows the
diagram of the Peak Code Domain Error. In screen B the result summary is displayed.
Displaying RHO
Make sure that all channels except the pilot channel (code 0.64) are OFF, so that only
the pilot channel is available in the measurement.
No specific measurement is required to get the value for RHO. The R&S FSV/FSVA
always calculates this value automatically regardless of the code domain measurement
performed. Besides the results of the code domain measurements, the numeric result
of the RHO measurement is shown in the result summary, by default shown in screen
B.
Measurement Examples for the 1xEV-DO MS Analysis (K85)
Measuring the Peak Code Domain Error
d) Close the "Frontend Settings" dialog box.
a) Open the "IQ Capture Settings" dialog box.
b) Set the "Trigger Source" option to "External".
28Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
5Test Setup for Base Station and Mobile Sta-
tion Tests
This section describes the default settings of the R&S FSV/FSVA, if it is used as a
1xEV-DO base station or mobile station tester. Before starting the measurements, the
R&S FSV/FSVA has to be configured correctly and supplied with power as described
in the Quick Start Guide, "Preparing For Use". Furthermore, the application firmware of
the R&S FSV-K84 (base station tests) or -K85 (mobile station tests) must be enabled.
Installation and enabling of the application firmware are described in the Quick Start
Guide, chapter 3.
Risk of instrument damage during operation
An unsuitable operating site or test setup can cause damage to the instrument and to
connected devices. Ensure the following operating conditions before you switch on the
instrument:
●
●
●
●
●
●
Test Setup for Base Station and Mobile Station Tests
All fan openings are unobstructed and the airflow perforations are unimpeded. The
minimum distance from the wall is 10 cm.
The instrument is dry and shows no sign of condensation.
The instrument is positioned as described in the following sections.
The ambient temperature does not exceed the range specified in the data sheet.
Signal levels at the input connectors are all within the specified ranges.
Signal outputs are correctly connected and are not overloaded.
Connect the antenna output (or TX output) of the base/mobile station to the RF input of
the R&S FSV/FSVA. Use a power attenuator exhibiting suitable attenuation.
The following values for external attenuation are recommended to ensure that the RF
input of the analyzer is protected and the sensitivity of the unit is not reduced too
much:
Maximum PowerRecommended external attenuation
≥ 55 to 60 dBm35 to 40 dB
≥ 50 to 55 dBm30 to 35 dB
≥ 45 to 50 dBm25 to 30 dB
≥ 40 to 45 dBm20 to 25 dB
≥ 35 to 40 dBm15 to 20 dB
≥ 30 to 35 dBm10 to 15 dB
≥ 25 to 30 dBm0 to 10 dB
29Operating Manual 1176.7632.02 ─ 05
R&S® FSV-K84/-K85
Maximum PowerRecommended external attenuation
≥ 20 to 25 dBm0 to 5 dB
< 20 dBm0 dB
●
●
●
Presettings
1. Enter the external attenuation.
Test Setup for Base Station and Mobile Station Tests
For signal measurements at the output of two-port networks, connect the reference
frequency of the signal source to the rear reference input of the analyzer.
The R&S FSV/FSVA must be operated with an external frequency reference to
ensure that the error limits of the 1xEV-DO specification for frequency measurements on base stations/mobile stations are met. A rubidium frequency standard
can be used as a reference source for example.
If the base station or mobile station has a trigger output, connect the trigger output
of the base station/mobile station to the rear trigger input of the analyzer (EXT
TRIG GATE).
2. Enter the reference level.
3. Enter the center frequency.
4. Set the trigger.
5. If used, enable the external reference.
6. Select the standard and the desired measurement.
7. Set the PN offset.
30Operating Manual 1176.7632.02 ─ 05
Loading...
+ 329 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.