Rohde & Schwarz HMC8015 User Manual

¸HMC8015
Power Analyzer User Manual
*5800532402*
5800532402
User Manual
Version 01
Test & Measurement
General Information Regarding the CE Marking
General Information Regarding the CE Marking
General Information Regarding the CE Marking
All ROHDE & SCHWARZ measuring instruments comply with the regulations of the EMC directive. The ROHDE & SCHWARZ conformity evaluation is based on applicable professional and product standards. In cases where multiple limit values can be used, ROHDE & SCHWARZ will apply the stricter standard of the two. With regard to interference emission, the limit values for businesses and commercial operations including the small business sector are applied (Class 1B). For the susceptibility to interference emissions, the applicable limit values for the industrial sec­tor are used. The signal and data cables that must be con­nected to the device during operation will necessarily affect the compliance with the acceptance limits to some extent. However, the cables used for this purpose will vary, depend­ing on the application. For operation of the device, the follow­ing information and conditions should therefore be observed with regard to interference emission and susceptibility:
1. Data Cables
All connections between measuring instruments or their interfaces and external devices (printers, computers, etc.) must always be implemented with adequately shielded
cables. Unless the user manual species a shorter maximum
cable length, the data cable (input/output, signal/control) may not exceed a total length of 3 m, and may not be located outdoors. If the device interface allows the connection of more than one cable, only one may be connected at a time. Data cables in general should always be double-shielded cable. For the IEEE bus cable, the double-shielded cable HZ72 is recommended and can be ordered from ROHDE & SCHWARZ.
2. Signal Cables
Signal cables used for signal transmission between the measurement point and the measuring device should gener­ally be kept as short as possible. Unless a shorter maximum
cable length is specied, the signal cables (input/output,
signal/control) may not exceed a total length of 1 m, and may not be located outdoors. All signal cables must be shielded (RG58/U coaxial cable). Care must be taken to achieve a proper ground connection. For signal generators, double­shielded coaxial cables (RG223/U, RG214/U) must always be used.
General Information
3. Impact on the Equipment
In the presence of strong high-frequency electric or
magnetic elds, unwanted signal components may enter
the equipment, despite careful setup of the connected cables and lines. With ROHDE & SCHWARZ measuring instruments, this will neither be destructive nor lead to the shutdown of the equipment. As a result of such external conditions, the displayed values and readings may show minor deviations with respect to the applicable instrument
specications in individual cases.
Regarding the CE Marking

Content

General Information Regarding the CE Marking .... 2
1 Important Information ...................4
1.1 Symbols ...................................4
1.2 Unpacking ..................................4
1.3 Instrument Setup ............................4
1.4 Safet y .....................................4
1.5 Intended Use ...............................4
1.6 Ambient Conditions ..........................5
1.7 Warranty and Repair ..........................5
1.8 Maintenance ................................5
1.9 CAT II ......................................6
1.10 Supply Voltage ..............................6
1.11 Limit Values .................................6
1.12 Batteries and Accumulators/Cells ...............6
1.13 Product Disposal .............................7
2 Instrument Controls and Indicators .........8
3 Options/Upgrade Vouchers ............... 9
4 Connecting a Device under Test .......... 10
4.1 Connecting a Device under Test via Adapter ......10
4.2 Connecting a Device under Test Directly .........11
4.3 Sensor Menu ...............................12
5 Setting Parameters ..................... 13
5.1 Using the Instrument for the First Time ..........13
5.2 Screen Display Description ...................13
5.3 Multifunction Softkeys .......................14
5.4 Conguring the Operating Mode from a List of
Measuring Parameter Options (“Cells”) ..........14
5.5 Readout Display ............................14
5.6 Selecting the Measuring Range ................15
6 Instrument Functions ................... 16
6.1 Measurement Parameters ....................16
6.2 Display Modes (VIEW Menu) ..................16
6.3 Crest Factor ................................19
6.4 Energy Counter (Integrator Values) .............19
6.5 Invert Function .............................20
6.6 Acquisition Mode ...........................20
6.7 Peak Hold .................................21
6.8 Limits .....................................21
6.9 Analog / Digital Inputs and Outputs .............21
7 Data Logging .........................23
8 Documenting, Saving and Retrieving ....... 25
8.1 Device Settings .............................25
9 General Settings ....................... 26
9.1 Updates (Instrument Firmware) ................26
9.2 Interface Setup .............................26
9.3 General Settings (Misc). ......................26
10 Remot Operation ......................28
10.1 USB VCP ..................................28
10.2 USB TMC .................................28
10.3 Ethernet ...................................30
10.4 IEEE 488.2 / GPIB ...........................32
11 Technical Data ........................33
12 Appendix ............................34
12.1 List of gures ..............................34
12. 2 Glos sar y ..................................34
Content
Content

Important Information

Important Information
1 Important
Information

1.1 Symbols

must follow the instructions and warnings contained in this manual carefully. The enclosure, chassis and all measuring terminals are connected to protective ground lead. The device complies with protection class II.
Disabling or separating the intended safety ground
connection, either inside or outside the device, is not permitted!
(1) (2) (3) (4)
(5) (6) (7)
Symbol 1: CAUTION: General hazard – Observe product documentation Symbol 2: Danger of electric shock Symbol 3: Earth (ground) Symbol 4: Protective conductor terminal Symbol 5: ON (supply voltage) Symbol 6: OFF (supply voltage) Symbol 7: Ground terminal

1.2 Un packing

When unpacking, check the package contents for com­pleteness (measuring instrument, power cord, product CD, any optional accessories). After you have nished unpack­ing, check the equipment for any visible shipping-related damage or other mechanical problems. For any damage occured during transport and shipping, contact the carrier and supplier immediately. Do not use the equipment if it is damaged.

1.3 Instrument Setup

As shown in the picture below, the instrument may be set up at a slight angle by using small extension legs that unfold from its feet. Ensure that the extension legs are completely unfolded to ensure secure footing.
Fig. 1.1: Operating positions.
Position the instrument so the power switch is not ob­structed at any time.

1.4 Safety

This instrument is designed in accordance with EN 61010-1 (VDE 0411 Part 1), Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, and left the factory in perfect condition. It thus also meets the requirements of European Standard EN 61010-1 and the international standard IEC 61010-1. To maintain this condition and ensure safe operation, the user
All exposed metal parts have been tested against the AC power leads with a test voltage of 2200 VDC. The instru-
ment complies with overvoltage category CAT II (600V). To
ensure safe operation, the instrument may only be oper­ated from a properly grounded power outlet. Plug in the power cord before connecting any signal circuits. Never use the product if the power cable is damaged. Check the power cable on a regular basis to ensure that it is in proper operating condition. Take appropriate safety measures and carefully lay the power cable to ensure that the cable will not be damaged and that no one can be hurt by, for exam­ple, tripping over the cable or suffering an electric shock. Any time there is reason to believe that safe operation is no longer possible, the instrument must be taken out of service and secured against inadvertent use.
Safe operation is not possible:
❙ If the instrument shows visible damage ❙ If the instrument is no longer working ❙ Following long-term storage under unfavorable conditions
(e.g. outdoors or in damp rooms)
❙ After severe transport stress (e.g. in packaging that does
not meet the minimum conditions for shipping by post, rail or truck)
Prior to switching on the product, always ensure that the nominal voltage set on the product matches the nominal voltage of the mains supply. If a different voltage is to be set, the mains fuse of the product may have to be changed accordingly.

1.5 Intended Use

This instrument is only intended to be used by persons who are familiar with the risks associated with measuring electrical quantities. The instrument may only be oper­ated with a properly grounded power outlet. Never use a “cheater plug” or other means to defeat or disconnect the protective ground lead. The power plug must be con­nected before any signal circuits are connected. The prod­uct may be operated only under the operating conditions
and in the positions specied by the manufacturer, without
the product's ventilation being obstructed. If the manufac-
turer's specications are not observed, this can result in electric shock, re and/or serious personal injury or death.
Applicable local or national safety regulations and rules for the prevention of accidents must be observed in all work performed.
Important Information
Important Information
The instrument is intended for use in the following areas:
❙ Industrial ❙ Residential ❙ Business and commercial ❙ Small business
The measuring device is to be used only with authentic ROHDE & SCHWARZ test accessories, signal cables and power cords. Never
use power cords with insufcient ratings. Before starting each
measurement, the signal cables should be checked for damage and replaced if necessary. Damaged or worn components may damage the equipment or cause injury.
The instrument may only be used indoors. Before each measurement, check the instrument for proper operation using a known signal source or sample.
To disconnect from the power supply, the rear uncooled IEC plug must be removed.

1.6 Ambient Conditions

The permissible operating temperature is +5 °C to +40 °C (connector contamination level 2). The maximum permis­sible relative humidity (non-condensing) is 80 %. During storage or transport, the permissible temperature range is –25 °C to +60 °C. If condensation has formed during transport or storage, the equipment should be allowed to acclimatize to ambient levels for about two hours before use. The instrument is intended for use in clean, dry areas. It may not be operated in dusty or damp conditions, an explosion hazard area, or in the vicinity of aggressive chemicals. The instrument can be positioned as needed,
but sufcient air circulation must always be ensured. For
prolonged operation, a horizontal or slightly angled operat­ing position (extension legs) is preferable.
When installing one or more R&S®HMC8015 units in a 19­inch rack, make sure that there is enough space around
its periphery to ensure sufcient cooling (see gure).
Recommended minimum clearance: 1 height unit (HU)
The instrument may be operated at an altitude of up to
2000 m. The specications and tolerances apply after
a warm-up time of at least 60 minutes, at an ambient temperature of 23 °C (± 2 °C). All values stated without tolerances are typical values for an average unit.
The ventilation openings must not be covered!

1.7 Warranty and Repair

ROHDE & SCHWARZ equipment is subject to strict quality
control. Before leaving the production line, each device
is subjected to a 10-hour “burn-in test”. This is followed
by an extensive functional test and quality inspection, in which all operating modes and compliance with speci-
cations are checked. These tests are carried out using
equipment that is calibrated traceable to national stan­dards. The statutory warranty provisions of the country in which the ROHDE & SCHWARZ product was purchased are applicable. In the event of any complaints, please contact the dealer from whom you purchased the ROHDE & SCHWARZ product.
All adjustments, parts replacements, maintenance and repairs must be carried out by authorized ROHDE & SCHWARZ specialists. Only original parts may be used for replacing safety-relevant parts (e.g. mains switches, mains transformers, fuses). After each replacement of safety-relevant parts, a safety check (visual inspection, PG conductor test, insulation resistance test, leakage current measurement, functional testing) must be performed. This helps ensure the continued safety of the product.
The product may only be opened by authorized,
specially trained personnel. Before any work is
performed on the product or before the product is opened, it must be disconnected from the supply voltage. Otherwise, personnel will be exposed to the risk of an electric shock.

1.8 Maintenance

The outside of the instrument should be cleaned regularly with a soft, lint-free dust cloth.
The display may only be cleaned with water or a suit­able glass cleaner (not with alcohol or solvents), followed by wiping with a clean, dry, lint-free cloth. Never allow cleaning liquid to enter into the equipment. Other cleaning products may dissolve the labels or mar the plastic surface and painted or varnished surfaces.
Before cleaning the instrument, make sure that it is turned off
and disconnected from any power supply (e.g. AC power grid or battery).
No part of the equipment should be cleaned with chemical sol­vents such as alcohol, acetone or paint thinners!
Important Information
Important Information
1.9 CAT II
This instrument is intended for measurements on cir­cuits that are or are not connected directly to the normal (low-voltage) AC power grid. The device complies with measurement category CAT II; the input voltage must not exceed 600 V
in a CAT II application.
RMS
The following information relates exclusively to user safety. Other aspects, e.g. the maximum permissible
input voltage, are specied in the technical data sheets
and must also be observed. Indirect measurements on the power grid may be performed by means of appropri­ate transducers (e.g. current probes) that meet or exceed the requirements of the protection class applicable to the measurement task. When measuring, the transducer’s
measurement category – as specied by the manufacturer
– must be observed.

Measurement categories

The measurement categories relate to transients super­imposed on the AC line voltage. Transients are short, very fast (steep) voltage and current changes that may occur either periodically or non-periodically. The amplitude of possible transients increases as the distance to the source of the low-voltage network decreases.
Fig. 1.2: Rear panel with supply voltage selector and connections.
The measuring circuit is equipped with a current overow
fuse against excess current that cannot be serviced or replaced by the user.

1.11 Limit Values

The R&S HMC8015 is equipped with an overload protec­tion circuit. The overload protection circuit is used to prevent damage to the unit, and to protect against possible electrical shock. The limit values of the unit must not be
exceeded. The front panel of the R&S HMC8015 species
the protection limit values in order to ensure safe opera­tion of the unit. These protection limits must be strictly observed:
Measurement category IV: Measurements at the
source of the low-voltage network (e.g. on electricity meters).
Measurement category III: Measurements in a building
network (e.g. distribution panels, circuit breakers, permanently installed electrical outlets, stationary motors, etc.).
Measurement category II: Measurements on circuits
directly connected to the low-voltage power grid (e.g. household appliances, power tools, etc.)
0 (Devices with no measurement category rating): Other
circuits that are not directly connected to the power grid.

1.10 Supply Voltage

Before turning on this device, please check that the avail­able supply voltage (115 V or 230 V ±10 %, 50 Hz to 60 Hz)
matches the specications on the supply voltage selector
switch of the device. If this is not the case, the supply volt-
age must be adjusted using the switch. The mains voltage
selector switch is located on the rear panel of the unit.
The unit has a main power fuse, which is located on the rear panel of the unit for uncooled IEC connectors. Depending on the supply voltage setting, a fuse of the cor­rect type must be installed.
❙ At the 115 V setting: F630L250V ❙ At the 230 V setting: F400L250V
If the equipment must be left unattended for a long period of time, it must be switched off at the power switch for safety reasons.
Fig. 1.3: Connections on the front panel.
Voltage input: 600 V
RMS
Current input: 20 A Max. voltage between COM and IN: 5 V
Peak
Connection to power supply:
115 V / 230 V, selectable with the voltage selector on the rear panel of the unit
Frequency: 50 -60 Hz Power consumption: 35 W maximum, 15 W typical

1.12 Batteries and Accumulators/Cells

If the information regarding batteries and accumulators/cells
is ignored or not sufciently observed, this may lead to explo­sion, re and serious personal injury or death. The handling of
batteries and accumulators based on alkaline electrolytes must conform to EN 62133.
1. Cells must not be taken apart, opened or crushed.
2. Cells or batteries must not be exposed to heat or re.
Storage in direct sunlight must be avoided. Keep cells and batteries clean and dry. Clean soiled connectors using a dry, clean cloth.
3. Cells or batteries must not be short-circuited. Cells or batteries must not be stored loose in a box or drawer where they can short-circuit each other, or where they can be short-circuited by other conductive materials. Cells and batteries must not be removed from their original packaging until they are ready to be used.
4. Keep cells and batteries out of the reach of children. If a cell or battery has been swallowed, seek immediate medical care.
5. Cells and batteries must not be subjected to any me­chanical impacts that are stronger than permitted.
6. If a cell develops a leak, the uid must not come into
contact with the skin or eyes. If contact occurs, wash the affected area with plenty of water and seek medical aid.
7. Cells or batteries that are not properly replaced or charged pose a risk of explosion. To ensure that the product remains safe to use, always replace cells or batteries with the matching type.
Important Information
Important Information
8. Cells and batteries must be recycled and kept sepa­rate from regular household waste. Accumulators or batteries that contain either lead, mercury or cadmium are hazardous waste. Observe the national regulations regarding waste disposal and recycling.

1.13 Product Disposal

Fig. 1.4: Product labeling in accordance
with EN 50419.
The German Electrical and Electronic Equipment Act (Elek­troG) implements the following EC directives:
❙ 2002/96/EC (WEEE) for electrical and electronic
equipment waste and
❙ 2002/95/EC on the Restriction of Use of Hazardous
Substances in electrical and electronic equipment (RoHS directive)
At the end of the lifecycle of this product, it may not be disposed of with normal household waste. Moreover, disposal via the municipal collection points for electrical equipment waste is also not permissible. Rohde & Schwarz GmbH & Co. KG fully assumes the manufacturer’s obliga­tions to take back and dispose as set forth by the ElektroG, to ensure environmentally sound disposal or recycling of the materials used.
Please contact your local service representative in order to dispose of the product.

Instrument Controls and Indicators

Instrument Controls and Indicators
2 Instrument Con-
trols and Indicators
¸HMC8015 front panel
1
Display – 3.5" TFT color display with QVGA resolution
(320 pixels x 240 pixels)
2
Interactive softkeys – Direct access to all relevant func-
tions
3
Function Keys – Operates as a numeric keypad using
SHIFT function U – Increases voltage range manually I – Increases current range manually VIEW – Switches between graphical and numeric dis-
play U – Decreases voltage range manually I – Decreases current range manually ACQ – Data acquisition menu for recording data START / STOP – Starts/stops the energy counter RESET – Resets peak values / energy counter MEAS – Logging function / energy counter / limit val-
ues HOLD – Stops the update of measurement values NULL – Sets the current display values as zero refer-
ence values EXT – Sensor menu and additional functions
4
SAVE/RECALL – Saves and recalls/uploads device set-
tings
5
SETUP – Gives access to general device settings
6
HELP – Integrated help screen
7
SHIFT – Activates the numeric keypad
8
Universal rotary control knob with arrow keys – Sets
the parameters cells (editing keys)
V jack – Input for voltage measurement
9
COM jack – Shared measuring input for voltage and
10
current measurement
A jack – Input for current measurement
11
IN/COM jacks – Sensor input for external shunts or cur-
12
rent probes
USB port – USB host port for USB ash drives when
13
saving parameters
POWER – ON/OFF for standby mode
14
¸HMC8015 rear panel
15
Fig. 2.2: ¸HMC8015 rear panel.
15
Digital / Analog IN/OUT – Digital / analog inputs and
19181716
20 21 2422
23
outputs for Pass/Fail sorting and limit values
16
IEEE-488 (GPIB) interface (optional) – Installation avail-
able at factory only
17
Ethernet (LAN) interface
18
USB interface
19
Ground terminal
20
Uncooled IEC plug
21
Fuse
22
Power switch
Supply voltage selector switch (115V or 230V)
23
24
Kensington lock
1
14
Fig. 2.1: ¸HMC8015 front panel.
13
2
3
12
4
5 6 7
11
8
10
9
3 Options/Upgrade
Vouchers
The R&S®HMC8015 has available options that increase the application scope of the instrument. Options and upgrade vouchers can be combined. The following options (or up­grade vouchers) are available for the R&S®HMC8015 power analyzer:

Options/Upgrade Vouchers

Options/Upgrade Vouchers
Description ¸HMC8 015
Advanced Analysis ¸HOC151 ¸HVC151
Advanced IO ¸HOC152 ¸HVC152
Table 3.1: Overview of ¸HMC8015 options/upgrade vouchers
options
1)
¸HMC8 015
upgrade vouchers
Fig. 3.1: Upgrade menu.
2)
The R&S®HOC151 Advanced Analysis option or R&S®HOC152 Advanced IO function can both be pur­chased from the factory with an R&S®HMC8015. In con­trast, the R&S®HV151 Advanced Analysis upgrade voucher
1) Only when ordered in combination with a ¸HMC8015 base unit
2) Subsequent activation of ¸HMC8015 options via upgrade voucher
Function Description
P Active power (W)
S Apparent power (VA)
Q Reactive power (var)
PF Power factor lambda (λ)
PHI Phase shift (φ)
FU Frequency value for the voltage (Hz)
FI Frequency value for the current (Hz)
FPLL
URMS True root mean square (RMS) voltage (U
UAVG Average voltage
IRMS True root mean square (RMS) current (I
IAVG Average current
UTHD Total harmonic distortion U
ITHD Total harmonic distortion I
WHM, WHP, WH, AHM, AHP, AH Energy counter (integrator values)
Logging Data recording (comma-separated values CSV)
USB, Ethernet (GPIB optional – HMC8015- G)
UPPeak Maximum voltage (U
UMPeak Minimum voltage (U
IPPeak Maximum current (I
IMPeak Minimum current (I
PPPeak Maximum power (P
PMPeak Minimum power (P
HARMONICS mode Bar graph of 50 harmonics
WAVEFORM mode
TRENDCHART mode Current or voltage displayed as signal prole
INRUSH mode Triggered display of a signal prole (single shot)
Sensor input Input jack for current probe / external shunt
DIN / AIN
Limit value / PASS-FAIL Displays limit values
Table 3.2: Feature overview: options/upgrade vouchers
Frequency and phase locked loop recording fre­quency (Hz)
Remote interfaces
)
+pk
)
–pk
)
+pk
)
–pk
)
+pk
)
–pk
Displays the waveform (for one cycle of voltage, cur­rent or power data)
Digital / analog inputs and outputs (Bayonet Neill – Concelman BNC)
or R&S®HV152 Advanced IO upgrade voucher enable a subsequent upgrade via a license key. The installed op­tions or upgrade vouchers can be looked up under “Device
R&S®HMC8015
)
RMS
)
RMS
base unit
¸HOC151 ¸HVC151
¸HOC152 ¸HVC152
Options/Upgrade Vouchers
Options/Upgrade Vouchers
Information” in the SETUP menu. Table 3.2 shows an over­view of the most important optional features. For com-
plete specications, please visit the ROHDE & SCHWARZ
homepage.
To generate a license key, go to http://voucherrs.hameg. com and enter the voucher code. The license key (SERI-
ALNUMBER.hlk) is an ASCII le and can be opened using a le editor. The key is listed in the le in plain text. Enable
the option in the instrument by either reading the key in automatically or entering it manually. The fastest and easi­est method is to read the key in automatically from a USB
ash drive. To do this, save the license le to a USB ash drive and then insert the ash drive into the USB port on
the instrument front panel. Press the SETUP button and then softkey LICENSE to open the menu for reading in options and upgrade vouchers. Press softkey READ KEYS FROM FILE to open the File Manager. Use the rotary con-
trol knob to select the license le, then press softkey LOAD
to load the key. The license key is loaded and the option is available after the instrument has been restarted.
Alternatively, the license key can be entered manually. To do this, press softkey INPUT KEY MANUALLY. This opens an input window where the license key can be typed in
manually. Once the key has been entered, conrm by
pressing softkey ACCEPT. Restart the instrument to enable the option.
4 Connecting a
Device under Test

4.1 Connecting a Device under Test via Adapter

Fig. 4.1: Connecting HZC815 DE version to ¸HMC8015.
A device under test (DUT) can easily and securely be connected to the ¸HMC8015 using the optional HZC815 socket adapter. In this case, the cables enclosed with the HZC815 (length = 25 cm) are connected to the ¸HMC8015 socket on the front panel. The connection of the socket adapter to the ¸HMC8015 front panel is determined by the color of the HZC815 accessory cables. The cables on the socket adapter represent an extension of contacts for a standard AC outlet with grounded safety sockets and are used for measuring a DUT. Power is sup­plied to the HZC815 socket adapter through the uncooled IEC plug on the top of the adapter and a power cord.
Fig. 3.2: Manually entering the license key.
There are 3 different types of HZC815 available ex factory:
❙ HZC815-DE: DE version; EU socket adapter (230V) ❙ HZC815-UK: UK version; UK socket adapter (230V) ❙ HZC815-US: US version; US socket adapter (115V)
Fig. 4.2: HZC815 DE, UK and US version plug adapters.
The HZC815 standard AC grounded safety outlet system is only approved for a 10 A continuous load. A load of 20 A would destroy the adapter.
10
10
On the sides of the HZC815 adapter, special faceplates hide mounting screws. The socket adapter can be perma­nently mounted with these mounting screws if so desired.

4.2.1 Internal Jack Wiring

Options/Upgrade Vouchers
Options/Upgrade Vouchers
Fig. 4.3: HZC815 DE version plug adapter with mounting screws.

4.2 Connecting a Device under Test Directly

The ¸HMC8015 can measure voltages up to 600 V and currents up to 20 A
. In these cases, a device under
RMS
RMS
test (DUT) can be directly connected to the power analyzer
using the 4 mm safety jacks on the front panel.
Before the start of each measurement, the signal cables to be used
must be checked for damage and wear.
Voltages are measured using the red V jack; currents are measured using the blue A jack. The black COM jack acts
here as the shared measuring input for voltage and current measurement.
If the maximum values for current and voltage are ex­ceeded, the ¸HMC8015 power analyzer is automati­cally switched off.
If the safety warning “Overcurrent Protection” appears on the screen, then more than 2 A were supplied in the low current range. The unit switches itself off in order to protect the internal shunt. In this case, switch to the automatic range selection (AUTO) or a higher range (> 500 mA) manually.
Fig. 4.4: Internal wiring of A, V and COM jacks.
The 4 mm A, V and COM safety jacks on the front panel
are not electrically isolated and separate from each other, and therefore offer several methods for measuring a DUT:
❙ Voltage-based measurement ❙ Current-based measurement ❙ Measurements via external shunt / current probe

4.2.2 Voltage-Based Measurement

Fig. 4.5: Voltage-based measurement.
Voltage-based measurement is typically used with high currents, since the voltage drop at the shunt increases as the current increases. The voltage U is correctly measured at the load.
Fig. 4.6: Voltage-based measurement, wiring example.

4.2.3 Current-Based Measurement

Current-based measurement is preferred with low currents
and high voltages, because it prevents current owing through the voltage measuring jack from going through
the shunt, which would falsify the reading. With this mea­surement type, the current reading must be inverted (IN-
11
11
Options/Upgrade Vouchers
Options/Upgrade Vouchers
Fig. 4.7: Current-based measurement.
VERT) using the ACQ menu and softkey menu CURRENT. In addition, the I and COM cables need to be reversed (see Fig. 4.8).

4.3 Sensor Menu

The Sensor conguration menu is opened via the EXT
menu and the softkey menu SENSOR. Press softkey ACTIVATE to activate (ON) or deactivate (OFF) the sensor measurement. When the sensor measurement is activated (ON), the automatic current measurement (AUTO) is deac­tivated (if it was activated previously).
Fig. 4.10: Sensor menu.
Fig. 4.8: Current-based measurement, wiring example.
4.2.4 Measurements with an External Shunt or
Clamp-on Current Probe
The 4 mm safety jacks IN and COM are used for making
measurements with an external shunt or clamp-on current probe. Safety plugs connect the shunt or clamp-on probe
directly to the jacks. Accessories such as HZC50 or HZC51
can be used for the clamp-on probe.
The softkey ACTIVATE activates the sensor inputs IN and COM on the instrument front panel (hardware switch).
Use the softkey RANGE and the rotary control knob to select the maximum measuring range for the measured
voltage. The available settings are 100mV, 1V and 4V. As a rule, the 100mV range is appropriate for measurements on shunts, while 1V and 4V are used for measurements
with a current probe. This setting is referenced to the peak value.
Example for shunt:
Set RESISTANCE to 1Ω. Set RANGE to 4V.
Convert as follows:
1V = 1A 4V divided by CF3 = 1.33A. Use the cell conguration to display this power range as
IRANGE on the screen.
Fig. 4.9: HZC51 Clamp-on current probe.
The connection of a clamp-on probe through a BNC adapter is not allowed, since depending on the polarity of the volt-
age, the measuring voltage may be present across the BNC
plug coupling.
12
12
Fig. 4.11: Example of a sensor.
Use softkey MODE and the rotary control knob to select the sensor mode. The available modes are SHUNT (to connect an external shunt) and CLAMP (to connect a current probe, e.g. HZC50 / HZC51). The Sensor menu
changes dynamically to match the selected sensor mode. In sensor mode SHUNT, softkey RESISTANCE is used to enter the value of the externally connected shunt in ohms. Select the value using the rotary control knob or press the SHIFT button to input the value numerically. If you entered the value numerically, press the appropriate unit on the
softkey menu to conrm the value. Once the shunt value
is entered, all measured values (current, power and values derived from these) can be calculated correctly and dis­played. Manual calculations are no longer necessary. The data from the external shunt is converted to the correct measured values internally.
Setting Parameters
Setting Parameters

5 Setting Parameters

5.1 Using the Instrument for the First Time

Pay particular attention to the safety instructions indicated
above when using the instrument for the rst time!
After connecting the uncooled IEC power cable and operating the power switch on the rear panel, the unit can be switched on by pressing the POWER button on the front panel. A help screen appears automatically. If you do not want to see the help screen, it can be hidden with the HIDE function. With the HIDE FOREVER softkey, the help screen feature will be disabled and no longer automatically displayed after restarting. After doing this, the help func­tion can only be opened using the HELP button.
Whenever the power analyzer is switched on, it is in the same mode as when it was last shut down. All device settings are stored in nonvolatile memory (NVM) and retrieved when restarting.
Fig. 4.12: Sensor mode.
In sensor mode CLAMP, softkey RATIO and the rotary con-
trol knob are used to dene a conversion factor. This factor denes how many millivolts the sensor transmits when a current of 1A is applied. Once the factor is entered, all
measured values (current, power and values derived from these) can be calculated correctly and displayed. Manual calculations are no longer necessary. Available factors are
1mV/A, 10mV/A and a user-dened value (USER). The conversion factors 1mV/A and 10mV/A are appropriate for
the current probes HZO50 and HZO51, for example. If you are using a current probe from a different manufacturer, use the rotary control knob to select the factor or use the SHIFT button to input the factor numerically.

5.2 Screen Display Description

1 2 3 4 6
7
8
9
1
Voltage measuring range
2
Current measuring range
3
Operating mode
4
Time displayed as function of the operating mode
5
Logging enabled (yellow) / disabled (white)
6
Type of interface: GPIB / TMC / VCP / LAN
7
Cell function
8
Measuring parameter cell with reading
9
Measuring range and modulation shown as bar
5
graphs (URange / IRange function)
10
Softkey menu 1 to 4 used to congure the meas-
uring parameter cell / display toggles between six and 10 measurement values
10
Fig. 5.1: Screen display description.
13
13
Setting Parameters
Setting Parameters
Fig. 5.2: Help function.

5.3 Multifunction Softkeys

The menu eld that appears on the screen can be manipu­lated using the softkeys on the right side of the screen. After a menu is activated by using the function keys on the front panel, control and conguration of the menu func­tions is done by using the softkeys. If a device function is
not available within a specic setting (e.g. if an option or
upgrade voucher is not installed), then the corresponding softkey is disabled and the caption is grayed out. Close a menu or jump back one menu level using the softkey at the very bottom. In addition, the Quick Menu and the
softkeys 1 to 4 can be used to congure the display of the
measuring parameter cells. If the softkey of the already
active conguration page is pressed again, it is possible
to toggle between a display of six and 10 parameter cells.
Each cell can be fully customized. Up to four conguration
menus are available.
5.4 ConguringtheOperatingModefromaListof
Measuring Parameter Options (“Cells”)
The individual functions and operating modes of the power meter can be selected using the buttons on the front panel. To select a device function, the corresponding function key is pressed. The white LED inside the key is lit as an indication that the function is active. The editing of a cell is activated by pressing on the rotary control knob
or one of the arrow keys. The rst cell on the top left is
highlighted in blue. The relevant parameter cell can then be selected using the rotary control knob or the arrow keys. Once a parameter cell is selected, the parameter list
can be opened by pressing the rotary control knob again. When a measuring parameter has been selected, it can be
conrmed by pressing the rotary control knob, or after 20
seconds without input, the unit will automatically return to the previous screen (see Section 9.3.5 Key Fallback Time).

5.5 Readout Display

The R&S®HMC8015 power analyzer has a TFT color dis­play. The screen is divided into two sections. The rst sec­tion is the status bar, in which the currently active modules and interface information are displayed. Table 3.2 presents a general overview of the power meter’s split screen with all possible function displays and their descriptions.
Fig. 5.4: Readout display with six active cells.
The eld of cells is variable and dependent on the oper­ating mode. In NUMERIC mode (set by default), typical numeric measurement values, such as power, voltage, cur­rent etc., are displayed. By contrast, in HARMONICS mode (HVC151 option required), the harmonics of the signals are displayed as bar graphs. Each mode has its own menu
structures, which can be congured by using the softkeys.
By default, the display consists of either six or 10 cells with
four different conguration pages.
Fig. 5.3: Selecting the measurement function.
14
14
Fig. 5.5: Readout display with 10 active cells.
With the HOLD function, the current measurement results are paused and shown on the display. Therefore when the HOLD button is pressed (button is lit), the particular measurement values are frozen on the display. The data recording is then running in the background. If the HOLD button is pressed again (LED turned off), the function is deactivated.
Setting Parameters
Setting Parameters
Using the NULL button, the current readings are taken as reference values and any deviation from them is shown on the display. If the NULL function is activated, the LED inside the button is lit. To deactivate the NULL function, press the button again (LED turned off).
If the device shows only dashes “---” on the display, the ¸HMC8015 could not determine a valid frequency. This can occur, for example, when the AC function was selected and the signal amplitude is too small or when the measuring range has been set too high.

5.6 Selecting the Measuring Range

The individual functions and operating modes of the R&S®HMC8015 unit can be selected using the buttons on the front panel. Basic functions, such as voltage or current measuring range, can be selected through the correspond­ing function keys. Using the U, I, U and I keys, the measuring range can be increased or decreased manually. U or I switches to the next higher, U or I to the next lower voltage or current measuring range.
In automatic range selection mode, this leads to a switch to the next higher range if the signal readings exceed
120% of the current measuring range. The unit switches to the next lower range if 0.6s have elapsed since the last range change and the RMS reading is within 95% of the
next smaller measuring range.
Moreover, by using the ACQ menu, the automatic measur­ing range selection (AUTORANGE) for current or voltage can be activated (ON) or deactivated (OFF) in the VOLTAGE or CURRENT softkey menu. The manual range selection is automatically disabled when using AUTORANGE.
In the case of manual range selection, if the applied value is too large, the display will show the message “OL” (overload).
Fig. 5.6: Example of measuring range display.
Push long the U , I , U and I keys to activate the automatic measurement range mode.
Using the URANGE or IRANGE measurement parameters, the modulation of the selected measuring range can be graphically displayed (bar graph) on the screen. If the mea­suring range is manually adjusted with the U, I, U and I keys, the bar graph is changed accordingly. A red measuring range bar indicates that the measuring range of the incoming signal is too small.
Fig. 5.7: Selecting the measurement function.
If, at crest factor 3, the instrument is in a current measuring range less than 200 mA and automatic current range selection is activated, then the data acquisition hardware automatically switches to the highest current measuring range for any large overcurrent (e.g. 2 A).
15
15

Instrument Functions

Instrument Functions
6 Instrument
Functions

6.1 Measurement Parameters

Fig. 6.1: R&S®HMC8015 readout display.
In NUMERIC mode, the instrument screen provides an overview of the measurement values. The numeric values displayed on the instrument screen are instantaneous val-
Fig. 6.2: Option / upgrade voucher info box.
ues that are constantly updated. If an instrument function is grayed out in the menu, the purchase of an option/up­grade voucher is required in order to unlock this feature. If the softkey of the grayed-out function is pressed, then an info box appears on the screen with the required option/ upgrade voucher (see Chapter 3).

6.2 Display Modes (VIEW Menu)

R&S®HMC8015 base unit
P Active power (W)
S Apparent power (VA)
Q Reactive power (var)
PF Power factor lambda (λ)
PHI Phase shift (φ)
FU Frequency value for the voltage (Hz)
FI Frequency value for the current (Hz)
FPLL Frequency and phase locked loop recording frequency (Hz)
URMS True root mean square (RMS) voltage (U
UAVG Average voltage
IRMS True root mean square (RMS) current (I
IAVG Average current
UTHD Total harmonic distortion U
ITHD Total harmonic distortion I
Wh+ Positive watt hours
Wh– Negative watt hours
Wh Sum of positive and negative watt hours
Ah+ Positive ampere hours
Ah– Negative ampere hours
Ah Sum of positive and negative ampere hours
R&S®HMC8015 base unit + HOC151 option /HVC151 voucher
UPPeak Maximum voltage (U
UMPeak Minimum voltage (U
IPPeak Maximum current (I
IMPeak Minimum current (I
PPPeak Maximum power (P
PMPeak Minimum power (P
R&S®HMC8015 base unit + HOC152 option /HVC152 voucher
DIN Digital input value (digital IN)
AIN Analog input value (analog IN)
Limit N Limit display
Table 6.1: Overview of measurement functions
–pk
–pk
+pk
+pk
–pk
+pk
)
)
)
)
)
)
RMS
RMS
)
)
Fig. 6.3: VIEW menu.
The R&S®HMC8015 power analyzer includes ve differ­ent display modes for the display of measurement results; these can be enabled by activating options/upgrade vouchers. The display modes can be selected using the VIEW menu. The following display modes are available:
Numeric mode: (default display)
Numeric display, toggles between six and ten measurement parameters presented as cells with four
different conguration pages; this provides a quick overview of the individual, congurable measurement
value cells.
Harmonics mode:
Bar graph of the rst 50 harmonics; this represents the
density of voltage / current as a percentage, absolute value or as a list. HARMONICS mode is only available in
conjunction with the HOC151 option or the HVC151
upgrade voucher.
Waveform mode:
Displays the waveform (for one cycle of voltage, current or power data). WAVEFORM mode is only available in
conjunction with the HOC151 option or the HVC151
upgrade voucher.
16
16
Instrument Functions
Instrument Functions
❙ Trendchart mode:
Current or voltage displayed as a signal prole (similar to
the roll mode in an oscilloscope). TRENDCHART mode is
only available in conjunction with the HOC151 option or
the HVC151 upgrade voucher.
Inrush mode:
Triggered display of a signal prole (single shot); INRUSH mode is only available in conjunction with the HOC151
option or the HVC151 upgrade voucher.

6.2.1 Harmonics Mode

Fig. 6.4: Logarithmic readout display in Harmonics mode.
FFT is used to calculate the RMS value and phase angle of
the harmonics. This includes a calculation of the rst 50 harmonics (100kHz max.). Instead of the total RMS value,
the display shows the voltage level or the current level of the individual harmonics selected using the cursor. The cursor, which appears in the form of a small triangle under the indicator bar, is moved using the rotary control knob. The voltage levels (softkey SOURCE U – Quick Menu U) are displayed as blue bars and the current levels (softkey SOURCE I – Quick Menu I) as blue bars. Using SCALING (Quick Menu A or %), it is possible to select a scaling fac­tor for the signal in the Y-axis with reference to the funda­mental (ABS or %). The measurement diagram is displayed logarithmically on the screen so that even the smallest signal details are visible.
key NUMBER (Quick Menu #) and the rotary control knob can be used to display the number of harmonics included in the FFT calculation. The softkey SUBSET (Quick Menu E
or O) can be used to dene what harmonics are to be used
in the calculation (EVEN, ODD or all). The voltage U or current I or both can be selected as the source for the FFT calculation (softkey SOURCE).
The FFT is calculated continuously. Settings made using softkeys affect only the display of values on the screen.
Pressing the bottommost softkey closes the HAR­MONICS menu and opens a Quick Menu while the mode remains active. This permits convenient control of the harmonics settings via the Quick Menu. Pressing the VIEW button and then selecting display mode NUMERIC closes Harmonics mode and returns to the default display mode.

6.2.2 Waveform Mode

The WAVEFORM display mode can be used to display the voltage, current and power simultaneously as a trace on the screen (hardware-based triggering). Exactly one signal period is displayed in the measurement diagram and
automatically scaled to the measuring range nal value.
Softkeys U, I and P are used to display or hide the traces for voltage (blue), current (red) and power (green). In addi­tion, softkey R (Readouts) can be used to display statistical values, including U
RMS
, I
, FPLL (= trigger frequency)
RMS
and P (power).
Fig. 6.5: Tabular display in Harmonics mode.
The softkey VIEW SELECT (Quick Menu B or T) can be used to switch the harmonics display between a bar graph (BAR) or a table (TAB). Unlike the bar graph, the tabular display includes both the magnitude and phase. The soft-
Fig. 6.6: Quick Menu in Waveform mode.
Pressing the bottommost softkey closes the WAVE­FORM menu and opens a Quick Menu while the mode remains active. This permits convenient control of the set­tings via the Quick Menu. Pressing the VIEW button and then selecting display mode NUMERIC closes Waveform mode and returns to the default display mode.

6.2.3 Trendchart Mode

TRENDCHART display mode is used to view two selected values over time. Softkeys SRC 1 (Source 1 – Quick Menu
1) and SRC 2 (Source 2 – Quick Menu 2) are used to select the values (sources). The following measurement functions are available as sources:
17
17
Instrument Functions
Instrument Functions
Fig. 6.7: Trendchart mode. Fig. 6.8: Inrush mode.
Acquisition of the trend is restarted any time the mea­surement function is changed. The trace scaling is set automatically on the screen and cannot be changed. The softkey TIME BASE can be used to select the temporal
resolution per scale division. The available values are 1s/ Div, 10s/Div, 1m/Div or 10m/Div. The existing trace data
is deleted any time the temporal resolution is changed. The softkey CLEAR deletes the trace data from the screen and starts a new acquisition. VISIBLE is used to activate Source 1 and Source 2 (yellow) or to deactivate them (white).
Function Description
P Active power (W)
S Apparent power (VA)
Q Reactive power (var)
PF Power factor lambda (λ)
PHI Phase shift (φ)
FU Frequency value for the voltage (Hz)
FI Frequency value for the current (Hz)
FPLL
URMS True root mean square (RMS) voltage (U
UAVG Average voltage
IRMS True root mean square (RMS) current (I
IAVG Average current
UTHD Total harmonic distortion U
ITHD Total harmonic distortion I
DIN Digital input value (digital IN)
AIN Analog input value (analog IN)
Table 6.2: Overview of SOURCE measurement functions in Trendchart mode
Frequency and phase locked loop recording fre­quency (Hz)
)
RMS
)
RMS
Pressing the bottommost softkey closes the TREND­CHART menu and opens a Quick Menu while the mode remains active. This permits convenient control of the set­tings via the Quick Menu. Pressing the VIEW button and then selecting display mode NUMERIC closes Trendchart mode and returns to the default display mode.

6.2.4 Inrush Mode

INRUSH display mode uses an internal hardware buffer to trigger acquisition of a rare event and to output it to the display (similar to the single shot function found on oscil­loscopes). This might be used to acquire inrush currents
on loads, for example. After an event occurs, the current and voltage trends are acquired. The trend display starts at the right-hand side of the screen. The softkeys VOLTAGE (Quick Menu U) and CURRENT (Quick Menu I) are used to activate (ON) and deactivate (OFF) the voltage and current traces, respectively. Under SETTINGS, the trigger is set to one of these two values, or it can be selected manu­ally. A threshold (softkey LEVEL) can also be set, as well as the offset between the rising and falling edges (softkey SLOPE). At the full sampling rate, the acquisition runs for
16ms (softkey ACQ. TIME). The sampling rate can also be
varied to permit acquisition of signal sequences of over a minute (see Table 6.3).
Sampling rate Acquisition time
500.00 ksample 16ms
250.00 ksample 32ms
125.00 ksample 65ms
62.50 ksample 130ms
31.25 ksample 260ms
15.63 ksample 520ms
7.81 ksample 1s
3.91 ksample 2s
1.95 ksample 4 s
976.56 sample 8s
488.28 sample 16s
244.14 sample 33s
122.07 sample 67s
Table 6.3: Overview of sampling rates
The trace display does not update dynamically. Instead, it must be updated manually by pressing softkey ACTIVATE (ON) (Quick Menu A). The trace is updated each time the button is pressed (single shot). The Inrush display cannot be zoomed. A maximum of 8000 measurement values each for current and voltage can be acquired.
Pressing the bottommost softkey closes the INRUSH menu and opens a Quick Menu while the mode remains active. This permits convenient control of the settings via the Quick Menu. Pressing the VIEW button and then selecting display mode NUMERIC closes Inrush mode and returns to the default display mode.
18
18
Instrument Functions
Instrument Functions

6.3 Crest Factor

The crest factor (also known as peak factor) describes the factor by which a signal’s amplitude (peak value) is greater than the rms-calculated effective value. This is important for the measurement of pulse-shaped sizes. For pure si-
nusoidal wave types, the following ratio applies: √2=1.414.
The accuracy of the calculated effective value remains the same as long as the peak value stays within the peak range. A crest factor of 1.8 or higher indicates a strong distortion from the waveform. To evaluate the waveform distortion of non-linear electric loads, the Waveform or TRENDCHART mode can be used.
Example:
If a battery is charged with 4AH and discharged with 3AH, the display will show AHP = 4, AH– = –3 and the
total AH = 1 (i.e. 4 + (–3) = 1).
The Integrator menu can be opened by pressing the soft­key INTEGRATOR in the MEAS menu. The relevant integra­tor mode is then selected using the softkey MODE and the rotary control knob.

6.4.1 Manual Integrator Mode (Default)

Fig. 6.9: Sample conguration for crest factor 6.
From the ACQ menu, the crest factor can be varied be­tween 3 and 6 by using the softkey CREST FACT in the VOLTAGE or the CURRENT softkey menu. When changing the crest factor, the measuring range is adjusted automati­cally. A crest factor of 5 or higher is used e.g. for sinusoidal waveforms with spikes (see example in Fig. 6.4) and is also required for standby measurements according to IEC
62018.

6.4 Energy Counter (Integrator Values)

The power analyzer makes it possible to display integra­tor values (WHP, WHM, AHP, AHM) on the screen. These values are also activated via the measuring parameter cell. In this way, the energy use of electronic devices under test (DUT) can be recorded where power consumption is not constant. WHP / AHP represent the power / current in the positive direction and WHM / AHM in the negative direc­tion. WH / AH correspond to the sum of the watt-hour/ ampere-hour positive and negative portions, respectively.
Fig. 6.11: MANUAL integrator menu.
The manual integrator mode is pre-selected by default. Using the START softkey, or directly using the START/ STOP button on the front panel of the instrument, the energy counter can be started as soon as it has been activated (ON) using the ACTIVATE softkey. Pressing either the softkey START or the START/STOP button on the front panel once more will stop the energy counter. The RESET softkey or the RESET button on the front panel will reset the integrator values.

6.4.2 Span Integrator Mode

A start time (START TIME) and an integration time span (DURATION) can be set using the SPAN integrator mode. Once the energy counter has been activated (ON) via the ACTIVATE softkey, the counter will automatically start at the preset start time and then automatically stop after the integration time span (DURATION) has elapsed. The time span for the integration can be set using the rotary control knob or, by using the SET TO CURR. softkey, the current
system time can be applied and adjusted accordingly.
Fig. 6.10: Readout display with integrator output values. Fig. 6.12: Span integration mode selection menu.
19
19
Instrument Functions
Instrument Functions
Initiating the start separately by using the START/STOP button on the front panel is not required. The RESET button on the front panel will reset the integrator values.
The maximum duration for running the energy counter is 96 hours (DURATION).
Fig. 6.13: Setting the integration duration.
Fig. 6.14: Inversion.
blue A jack. In order to record and display the current cor­rectly with this type of wiring, the current direction must be inverted in the ACQ menu using the CURRENT softkey menu and the INVERT softkey. When the INVERT func­tion is activated, it is displayed as a green box around the measuring range displayed in the status bar.

6.4.3 Duration Integrator Mode

The DURATION integrator mode denes how long the
energy counter is active. Using the START softkey, or directly using the START/STOP button on the front panel of the instrument, the energy counter can be started as soon as it has been activated (ON) using the ACTIVATE softkey. After the integration time (DURATION) has elapsed, the energy counter is automatically stopped. The time display in the status bar is replaced with the integration time. The time span for the integration can be set using the rotary control knob or, by using the SET TO CURR. softkey, the current system time can be applied and adjusted accord­ingly. If the integration time is expired, a new integrator measurement can only be started via START button, if the integrator values were reset with RESET softkey.

6.4.4 External Integrator Mode

(only with the HOC152 option/HVC152 voucher)
Using the external integrator mode (EXTERN), the energy counter can be started and stopped by an external pulse at the digital inputs on the rear panel. Once activated (ON) using the ACTIVATE softkey, the energy counter is automatically started by a “high” signal and automatically stopped by a “low” signal. Initiating the start / stop sepa­rately by using the START/STOP button on the front panel is not required.

6.5 Invert Function

The power analyzer also records the sign of the mea­sured current with AC, for the correct calculation of the exact input or output of power and energy, for example. Depending on the external wiring (e.g. with the HZC815 adapter) the voltage or the current is measured inverted. For current-based measurements, the wires going to the
jacks A and COM on the front panel must be swapped
(see Section 4.2.3). When using the HZC815 measuring adapter, the blue cable must therefore be inserted into the
black COM jack, and the black cable is inserted into the

6.6 Acquisition Mode

Fig. 6.15: Selecting the data acquisition mode.
The acquisition of current and voltage is performed simul­taneously and so can be compared with a two-channel oscilloscope (16 bit). The sampling frequency is 500 ksample/s. The values displayed on the screen are updated 10 times per second. In the ACQ menu, the data acquisi­tion mode can be selected by using the MODE softkey menu, followed by the MODE softkey. The following set­tings can be selected:
AUTO: Automatic selection of the data acquisition mode
(default); automatic detection of an incoming signal.
DC: For measuring pure DC loads; in contrast to the AC
mode, there is no synchronization with the signal period, instead synchronization is over a certain period of time (error is displayed if AC signal is applied).
AC: Average value of a voltage / current period
(synchronization with the signal period).
EXTERN: Specifying a signal period via the external BNC
connector DIGITAL IN (only available in conjunction with
option HOC152 / upgrade voucher HVC152); e.g. used
with a fuzzy signal which must be ltered externally.
20
20
Instrument Functions
Instrument Functions

6.6.1 Frequency Filter

The power analyzer R&S®HMC8015 features the following lter settings which can be adjusted in the ACQ menu us­ing the softkey menu MODE (for AUTO and AC only):
Frequencylter:
Reliable acquisition of the zero crossings for the precise determination of the signal period.
Digitallter:
Automatic, adaptive averaging of the lter setting for
several periods to increase accuracy; ideal for highly
uctuating values (values become more stable).
BWL:
Classic low-pass lter (1 kHz) as bandwidth limit (BWL), in order to produce a clean, HF-adjusted signal.
Fig. 6.16: Filters arranged as a chain.

6.6.2 Frequency Source

Use the phase locked loop source frequency synchroniza­tion (PLL SRC) softkey to set the reference variable for the fundamental frequency (VOLTAGE/CURRENT). The funda­mental frequency of each period is the reference for the measured harmonics of the reading. Voltage (VOLTAGE) is selected as the reference variable by default. The CUR­RENT setting makes sense for signals with a very small or an irregular current trace. Using the EXTERN function, the frequency can be set via the digital input located on the rear panel (only available in conjunction with the HOC152 option or the HVC152 upgrade voucher).
Function Description
P Active power (W)
S Apparent power (VA)
Q Reactive power (var)
PF Power factor lambda (λ)
PHI Phase shift (φ)
FU Frequency value for the voltage (Hz)
FI Frequency value for the current (Hz)
FPLL
URMS True root mean square (RMS) voltage (U
UAVG Average voltage
IRMS True root mean square (RMS) current (I
IAVG Average current
UTHD Total harmonic distortion U
ITHD Total harmonic distortion I
Table 6.4: Overview of LIMIT measurement functions
Frequency and phase locked loop recording fre­quency (Hz)
)
RMS
)
RMS
dene maximum and minimum limit values. Similar to the
measuring range modulation, a bar is used to display the limit values. This two-part bar indicates the modulation of
the limit from –100% (minimum modulation) to +100%
(maximum modulation). It is also possible to link the limit function with the ANALOG OUT or DIGITAL OUT (EXT menu). See Section 6.9.

6.7 Peak Hold

The PEAK HOLD displays maximum peak values, ensuring a stable readout display for loads. The sampling frequency is 500 ksample/s. The PEAK HOLD function is activated (ON) or deactivated (OFF) in the ACQ menu by selecting softkey menu PEAK HOLD and softkey ACTIVATE.
If “OL” (Overload) appears on the display, the peak values must be reset on the display using softkey RESET

6.8 Limits

In the MEAS menu, the softkey menu LIMIT can be used
to dene up to six limit values. Once up to six limit val­ues (LIM1 to LIM6) are selected in the cell conguration,
they can be activated (ON) or deactivated (OFF) using the softkey ACTIVATE. The softkey SOURCE and the rotary control knob can be used to select the measurement func­tion for the source. Table 6.4 shows the possible measure­ment functions. The softkeys HIGH and LOW are used to
If the measuring instrument displays only dashes “-----” on the display, either the limit function is not activated or an error oc-
curred (e.g. limit value LOW ≥ limit value HIGH).
Fig. 6.17: LIMIT menu.

6.9 Analog / Digital Inputs and Outputs

On the R&S®HMC8015 rear panel, there are four BNC connectors that can be used as analog or digital inputs and outputs (Digital / Analog IN/OUT). The connectors are available for limit values and Pass/Fail sorting, for example.
Fig. 6.18: BNC connectors Digital / Analog IN/OUT.
21
21
Instrument Functions
Instrument Functions

6.9.1 Analog IN

The ANALOG IN can be activated (ON) or deactivated (OFF) in the EXT menu by selecting softkey menu ANA­LOG IN and softkey ACTIVATE. In addition, the AIN
function in the cell conguration can be used to display
the measured value of the analog input signal on the screen.
Fig. 6.19: ANALOG OUT menu.

6.9.2 Analog OUT

The ANALOG OUT can be activated (ON) or deactivated (OFF) in the EXT menu by selecting softkey menu ANA­LOG OUT and softkey ACTIVATE. The external mode is selected by using softkey MODE and the rotary control knob. The following modes can be selected:
Voltage:
Minimum modulation at the output corresponds to 0V
and maximum modulation corresponds to the measuring range.
Current:
Minimum modulation at the output corresponds to 0V
and maximum modulation corresponds to the measuring range.
Power:
Minimum modulation at the output corresponds to 0W
and maximum modulation corresponds to the maximum current and voltage measuring ranges.
Limit:
The limit module is connected to the analog output. The output modulation is represented as a percentage %.
–100% corresponds to minimum modulation at the output (≙ –5V) and +100% corresponds to maximum modulation (≙ +5V). The limit value (LIM1 to LIM6) is
selected using softkey LIMIT NO and the rotary control knob.
Frequency:
The power analyzer measures the frequency at the DIGITAL IN.
PWM:
The power analyzer measures the relationship between High and Low.
State:
The power analyzer measures the High and Low levels.
Fig. 6.20: DIGITAL IN menu.
The softkey INVERT is used to invert the signal applied at DIGITAL IN. This makes sense in STATE mode, for ex­ample.

6.9.4 Digital OUT

The DIGITAL OUT can be activated (ON) or deactivated (OFF) in the EXT menu by selecting softkey menu DIGITAL OUT and softkey ACTIVATE. The external mode is selected by using softkey MODE and the rotary control knob. The following modes can be selected:
Limit:
The limit module is connected to the digital output. The High level at the output corresponds to logical True and the Low level corresponds to False.
FPLL:
The output is controlled by an FPGA and always outputs the frequency of the selected source.

6.9.3 Digital IN

The DIGITAL IN can be activated (ON) or deactivated (OFF) in the EXT menu by selecting softkey menu DIGITAL IN and softkey ACTIVATE. In addition, the DIN function in the
cell conguration can be used to display the measured
value of the digital input signal on the screen. The differ­ent modes can be selected using softkey MODE and the rotary control knob:
22
22

7 Data Logging

Data Logging
Data Logging
function should start. Using the DURATION softkey menu,
the recording duration for the readings is dened.
The maximum logging duration is 96 hours (DURATION).
Using the MEAS button and the LOGGING softkey menu, the data acquisition can be started and various settings can be selected. Using the ACTIVATE softkey, data acqui­sition and storage can be activated (ON) or deactivated (OFF). Using the STORAGE softkey menu, the type of stor-
age device can be selected (Internal / USB ash drive), as well as the le name (File Name) and the le format (CSV
/ TXT). Using the LOG PAGE softkey, the active measur­ing parameter cells menu (Quick Menu 1 to 4 on the main screen) can be selected. If LOG PAGE 1 is selected, only
the readings from the rst measuring parameter cell menu
(Quick Menu 1 on the main screen) are recorded. If LOG PAGE 2 is selected, only the readings from the second measuring parameter cell menu (Quick Menu 2 on the main screen) are recorded, and so on.
Using the INTERVAL softkey and the rotary control knob, a measuring interval can be set. The measuring interval is the time between readings. If, for example, the INTERVAL function is set to 5, then a reading is taken every 5 s. The MODE softkey menu offers various logging modes. Select
the UNLIMITED function if you intend to make an innitely
long data record. The limiting factor here is the size of the internal memory (512 kbyte max.) or that of the connected
USB ash drive (4 Gbyte max., FAT/FAT32 formatting). If
the COUNT function is active, the number of recorded events can be set with the COUNT softkey and the rotary control knob. If, for example, an interval of 5 s and a count of 5 are set, then 5 readings will be recorded at intervals of 5 s. If the DURATION function is activated, the record­ing duration for the readings can be adjusted with the DU­RATION softkey and the rotary control knob. If the SPAN
function is activated, a logging time span can be dened. Use the softkey menu START to dene when the logging
External USB hard drives (or USB extensions) are not supported. Only USB ash drives with FAT/FAT32 formatting (4Gbyte max.)
are recognized by the R&S®HMC8015.
Using the SPAN function, a start time (START TIME) and a logging time span (DURATION) can be set. Once the logging function has been activated (ON) via the ACTIVATE softkey, the data acquisition will automatically start at the preset start time and then automatically stop after the log­ging time span (DURATION) has elapsed. The time span for the logging can be set using the rotary control knob or, by using the SET TO CURR. TIME softkey, the current
system time can be applied and adjusted accordingly.
The log function is active if the color of the LOG display in the
status bar has changed to green, or inactive if it is white.
Fig. 7.2: Setting the integration duration.
Using the CSV softkey menu, the structure of the CSV le can be congured. This is used to dene the decimal separa-
tor (DEC.SEP.) as well as the line separator (FIELD DELIM.).
Data saved in a CSV le can be opened with Microsoft
Excel, for example, and displayed as graphical charts (see example in Fig. 7.1).
Gaps in the logging may occur under the following circum­stances:
❙ High load caused by SCPI commands over the interface ❙ Use of a “slow” USB ash drive ❙ Sector in le system is sized too large
Fig. 7.1: Graphical log le analysis in Excel.
23
23
Data Logging
Data Logging

7.1.1 Example of Data Format in a Log File

#Device;HMC8015 #Device Name;PM #Format;LOG #Date;2000 -01-01 #Version;00.021-02.430-03.730 #Serial No.;012345678 #Mode;Count #Logging Interval[s];1.000
#Specied Logging Count;1800 #Specied Logging Time[s];-----
#Start Time;04:49:33 #Stop Time;04:49:37 #Actual Count;4
URMS[V];IRMS[A];P[W];FU[Hz];EMPTY;EMPTY;S[VA];Q[var ];LAMBDA[];UTHD[%];Timestamp
231. 27E+00;45.0E-03;6.63E+00;50.0E+00;nan;nan;10.38 E+ 00;7.98E+00;639E-03;2E+00;04:49:33:000
231.38E+00;45.0E-03;6.64E+00;50.0E+00;nan;nan;10.38E +00;7.98E+00;639E-03;2E+00;04:49:34:000
231.35E+00;45.0E-03;6.63E+00;50.0E+00;nan;nan;10.38E +00;7.98E+00;639E-03;2E+00;04:49:35:000
231.38E+0 0;44.9E- 03;6.63E+00;50.0E+00;nan;nan;10.38E +00;7.99E+00;639E-03;2E+00;04:49:36:000
the time stamp can be adjusted by rst selecting the time
stamp values and then clicking the right mouse button for
“Format Cells”. Then, using the “User Dened” function, the formatting can be congured manually with the key-
board in the “type” section as follows:
hh:mm:ss,000
After conrming with the OK button, the user-dened
custom formatting is saved, and the time stamp values are displayed in the selected format. By doing this, even the smallest millisecond values can be read easily.
Fig. 7.4: Excel Import Wizard (Text Import Wizard).
If the CSV le contains “NaN” (not a number), then either the cor­responding measuring parameter cell was empty (EMPTY), or the reading could not be displayed due to the chosen settings.

7.1.2 Data Formatting in Excel/Import Wizard

If a CSV log le is opened with Excel, the measurement
data is displayed correctly. However, we recommend that you change the formatting of the time stamp (Timestamp)
manually. After opening the CSV le, the formatting of
It is also possible to use the Import Wizard in Excel. After
starting Excel, rst select the relevant CSV le by clicking on the tab “Data” > “From Text”. After selecting the le and conrming with the “Import” button, the Import Wiz-
ard is started. This Text Import Wizard can also be used
to format the CSV le as needed. For example, the “Start import on line” function can be congured in a way that
only the readings are imported without the header (e.g. 18, see Fig. 7.4).
Fig. 7.3: User-dened customization of time stamp formatting in Excel.
24
24
8 Documenting,
Saving and Retrieving

Documenting,Saving and Retrieving

Documenting,Saving and Retrieving

8.2 Screenshots

The most important method of storing data for docu­mentation purposes is the screenshot. A screenshot is an
image le that shows the current screen contents at the moment it is captured to a le. However, screenshots can only be saved to a USB ash drive. When a USB ash drive
is connected, directories can also be changed, created or
deleted. The choice of the target directory is conrmed
with ACCEPT.
The R&S®HMC8015 power analyzer allows the user to save all screen displays and user settings. Internal memory is used to store device settings. This type of data can also
be saved to any USB ash drive connected to the unit.
However, screen captures can only be saved to a USB
ash drive. The main menu for the saving and recalling/
uploading of functions is opened by pressing the SAVE/ RECALL button.

8.1 Device Settings

Using the DEVICE SETTINGS softkey menu, the currently active device settings can be saved, and previously saved settings can be retrieved/uploaded.
Pressing the SAVE softkey will open the Save to Memory menu. Using the STORAGE softkey, a possible location (INTERNAL or FRONT) is selected for the device settings to be stored. After selecting the desired storage location and pressing the ACCEPT softkey, the File System Man-
ager is opened. The le name (FILE NAME) can be adjusted to the currently active setting or changed entirely
(the default name is SET). Use the COMMENT softkey to add a comment that will appear in the footer of the File
Manager if the le has been selected. Device settings are
stored in HDS format (binary). The format cannot be changed. The settings are saved using the SAVE com­mand.
The name of the le (FILE NAME) can be adjusted to the currently active setting or changed entirely (the default name is SCR).
The selected le format for the graphic le then deter­mines the color depth and the type of compression. There is no difference in the quality of these formats with respect
to the instrument graphics. The following le formats can
be selected from the softkey menu FORMAT:
❙ BMP = Windows Bitmap Format ❙ PNG = Portable Network Graphic
Using the COLOR MODE softkey and the rotary control knob, the user can choose between GRAYS CALE and COLOR. In GR AYSC ALE, the colors are converted to grayscale, while saving in COLOR saves the image as displayed on the screen. Pressing the softkey SAVE triggers an immediate saving of the current screen to the preset location and using the preset name and format.
A screenshot will be saved (depending on the menu setting) on
a connected USB ash drive by holding down the SAVE/RECALL
button.
Device settings from an old rmware version cannot be loaded with a new rmware version.
To upload les containing saved settings, the LOAD soft­key menu is opened by pressing the corresponding soft­key. This will open the File Manager, in which the desired
le can be selected using the rotary control knob. Once the storage location and the settings le are selected, the le is uploaded by pressing the LOAD softkey. To remove settings les that are no longer needed, the corresponding le is selected with the rotary control knob and removed
using the REMOVE FILE softkey. When a USB ash drive is connected, directories can be changed or deleted as well.
The menu item DEFAULT SET TINGS provides the ability to upload and return to the factory default settings.
25
25
General Settings
General Settings

9 General Settings

Important basic settings such as general or interface set­tings can be set using the SETUP button. Use the button to go back one level.

9.1 Updates (Instrument Firmware)

The rmware is compressed into a ZIP le. When the ZIP le is downloaded, it is decompressed into the root directory of a USB ash drive. Once the USB ash drive
is connected to the USB port of the instrument, press the SETUP button. Using the UPDATE softkey, the update
menu opens and the currently installed rmware version is
displayed together with its version number, date and build information.
When the FIRMWARE softkey is pressed, a search for the
corresponding le is started on the USB ash drive, and all information of the new rmware to be installed from the ash drive is displayed. If the rmware on the device is
already identical to the most current version, then the
version number from the USB ash drive is displayed in
red, otherwise the version number is displayed in green. In the latter case, the update should be started by press­ing the EXECUTE softkey.
Fig. 9.2: Setup menu.

9.3 General Settings (Misc)

9.3.1 Device Info

Using this softkey, it is possible to access device informa­tion, such as serial number, software version etc. In addition, the remaining available internal memory is displayed.
Fig. 9.1: Firmware update menu.

9.2 Interface Setup

Using the INTERFACE softkey menu, settings for the fol­lowing can be created:
❙ VCP (virtual COM port) ❙ USB TMC ❙ Ethernet (IP address, subnet mask etc.) and ❙ IEEE-488 GPIB interfaces (GPIB address)
The interface to be used for communications is selected with the corresponding softkey. The necessary interface
parameters are congured under the softkey menu item
for parameters. See Chapter 10 for more information about the available interfaces.
Fig. 9.3: Instrument information.

9.3.2 Date & Time

Using the DATE & TIME softkey, it is possible to set the time of day and the date, which are then used to create a date and time stamp on printer output or in saved data­sets. The date and time can always be reset by the user.
The settings for date and time may be congured using
the rotary control knob. The corresponding softkey menu item is active if it is highlighted in yellow. The date or time parameters can be transferred by using the SAVE com­mand.

9.3.3 Sound

The power analyzer provides the option for issuing a signal in the event of failure (or even as a test). This option can be activated (ON) or deactivated (OFF) using the ERROR BEEP or CTRL BEEP softkeys. The corresponding softkey menu item is active if it is highlighted in yellow.
26
26

9.3.4 Display

Using the DISPLAY softkey menu and the BACKLIGHT softkey, the screen intensity (i.e. brightness and contrast)
can be varied between 10% to 100% using the rotary
control knob. Using the CONTRAST and BRIGHTNESS softkeys, screen contrast and brightness can each be set between 10% and 100%. The GRID INTENS. sofkey is used to set the intensity of the display grid for graphical
displays. This menu item is only available in conjunction
with the HOC151 option or the HVC151 upgrade voucher. The corresponding softkey menu item is active if it is high­lighted in yellow.

9.3.5 Key Brightness (KEY)

With the KEY BRIGHT softkey, the button brightness can
be varied from 10% to 100% using the rotary control
knob. You can use the FALLBACK softkey to set the key fallback time to 5 s, 10 s or 20 s. Once the fallback time is set, the settings window will be closed automatically after the set time. In addition, the automatic return to the previ­ous level can also be deactivated (OFF). The corresponding softkey menu item is active if it is highlighted in yellow.
General Settings
General Settings

9.3.6 Device Name

The device name can be edited in this menu. A keypad is opened by pressing the DEVICE NAME softkey. Characters can be selected using the rotary control knob. The respec-
tive character is then conrmed by using the ENTER button (SHIFT). The entered device name is nally con­rmed by using the ACCEPT softkey.
Fig. 9.4: Entering the instrument name.

9.3.7 CSV

Using the CSV softkey menu, the structure of the CSV
le can be congured. This is used to dene the decimal
separator (DEC.SEP.) as well as the line separator (FIELD DELIM.).
27
27
Remote Operation
Remote Operation

10 Remote Operation

The USB VCP driver can only be installed on a PC if the following
prerequisites are met:
The R&S®HMC8015 comes with standard Ethernet and USB interfaces. In addition to a LAN interface, the R&S®HMC8015 has a
To enable communications, the selected interface and the associ­ated settings in the device must exactly match those on the PC.
USB device port. For this interface, the user can choose whether the device is to be addressed via a virtual COM port (VCP) or via the USB TMC class. A GPIB interface is available as option from the factory (R&S®HMC8015-G). The GPIB interface has a dedicated slot on the rear panel.
Fig. 10.1: Rear panel.

10.1 USB VCP

The available USB VCP drivers have been fully tested and are
available for Windows XP™, VISTA™, Windows 7™, Windows 8™
and Windows 10™ (32 bit and 64 bit).
When using the classic Virtual COM Port (VCP) variant, the user can, after installing the corresponding Windows driver for a terminal program, communicate with the R&S®HMC unit using SCPI commands. The latest USB VCP driver can be downloaded free of charge from the ROHDE&SCHWARZ website <www.rohde-schwarz. com> in the downloads section, and decompressed into an appropriate directory. If a driver for the R&S®HMC com­pact series has not yet been installed on the computer, the operating system responds with the message “Found New Hardware” once the connection between the device and the PC has been established. In addition, the “Found
1. R&S®HMC8015 with USB VCP interface enabled
2. A PC with the Windows XP™, VISTA™, Windows 7™, Win-
dows 8™ or Windows 10™ (32 bit or 64 bit) operating system
3. Administrator privileges are required for installing the driver. If an error message is displayed citing a write error, this is generally caused by not having the required access creden­tials for the installation of the driver. In this case, contact your IT department to obtain the necessary privileges.
New Hardware Wizard” is displayed. If this occurs ,the USB driver must be installed. For more information about USB VCP driver installation, see the installation instruc-
tions in the driver le.
In addition, the free HMExplorer software can be used. This Windows application offers the ability to create screenshots for the R&S®HMC8015 in addition to having a terminal program function.

10.2 USB TMC

The modern alternative to a Virtual COM Port (VCP) for external control is using the USB TMC procedure. TMC stands for “Test & Measurement Class” which means that once Virtual Instrument Software Architecture (VISA) drivers have been installed, the connected device can be detected without specialized Windows drivers, and used directly in the corresponding environments. The structure of the TMC procedure is modeled on the GPIB interface. Therefore, the great advantage of the USB TMC is that, by querying special registers, it can be determined whether commands were correctly processed and have been completed. By contrast, when communicating using VCP, testing and polling mechanisms in the control software are needed for this purpose, which may lead to a considerable workload on the device interface. This problem is solved when using the USB TMC by using the TMC status regis­ters, similar to a GPIB interface hardware-based solution through corresponding control lines.
Communications via USB TMC are not supported by the HMEx­plorer software.
10.2.1 USBTMCConguration
The R&S®HMC8015 power analyzer assumes a generic USB device driver when it is operated in USB TMC mode. The USB Test & Measurement Class (USB TMC) is a pro­tocol that enables GPIB-like communications using USB interfaces and represents a unique device class within the
USB specication. The USB TMC protocol supports ser­vice queries, triggers and other GPIB-specic instructions.
The driver is included in the NI Virtual Instrument Software Architecture (VISA) package and can be downloaded from <http://www.ni.com/downloads/ni-drivers/>.
Fig. 10.2: USB VCP settings.
28
28
First, the NI-VISA driver must be installed on your Win­dows system. To do this, download the latest version of the NI-VISA driver package. Decompress the driver
package you just downloaded and follow the installation
instructions.
Here is an example for NI-VISA version 5.4.1:
Remote Operation
Remote Operation
Fig. 10.5: Installing NI-VISA application locally.
After successfully installing the NI-VISA driver, you can now switch your R&S®HMC8015 to the USB TMC inter­face. To do so, please access the SETUP menu of your R&S®HMC8015 and select “Interface”.
Fig. 10.3: NI-VISA 5.4.1.
Start the installation with “Next” and follow the installation instructions.
Fig. 10.4: NI-VISA installation guide.
For this step, select “NI-VISA xxx All applications in­stalled locally”.
Fig. 10.6: Setup menu.
Then, using the softkey, switch to “USB TMC”.
Fig. 10.7: Interface menu.
The top of the screen now shows “TMC” as the selected interface type.
29
29
Remote Operation
Remote Operation
Fig. 10.8: TMC display on the screen.
Finally, establish the connection between the power supply unit and your Windows PC with a USB interface cable (type A – B). When used for the rst time, the operat­ing system reports back with the message “Found New
Hardware”. After successful conguration, the window
“Device Driver Installation” appears with the message “USB Test and Measurement Device (IVI), use now enabled”.
Fig. 10.9: Device driver installation.
Open the Windows Device Manager. The device is now listed under “USB Test and Measurement Devices > USB Test and Measurement Device (IVI)”.

10.3.1 Internet Protocol (IP) Networks

In order for two or more network elements to communi­cate with each other through a network (e.g. measuring instruments, hosts/PCs, etc.), a number of basic connec­tions must be observed, to facilitate the correct and un­disturbed execution of the data transmission in networks. Each network element within a network must be assigned an IP address in order for them to exchange data. IP ad­dresses are represented (in the IP version 4 or IPv4) in the form of four decimal numbers separated by periods (e.g.
192.168.15.1). Each decimal number represents a binary number of 8 bits (or one byte). IP addresses are divided into public and private address ranges. Public IP addresses are routed through the Internet and are provided by an Internet Service Provider (ISP). Network elements which have a public IP address can be reached via the Internet as well as exchange data over the Internet directly. Private IP addresses are not routed through the Internet and are re­served for private networks. Network elements which have a private IP address cannot be reached via the Internet and cannot exchange data over the Internet directly.
In order for network elements with private IP addresses to exchange data over the Internet, a router that performs IP network address translation (NAT) must connect them to the Internet. Using this router, which has a private IP address (the LAN IP address) and a public IP address (the WAN IP address), the network elements are connected to the Internet and can exchange data. In cases where network elements only ever exchange data through a local area network (without connection to the Internet), it is best to use private IP addresses. Choose a private IP address for the device and also a private IP address for the host (PC) used to control the device. If you later connect your private network through a router to the Internet, you can keep the previously used private IP addresses for use on your local network.
Fig. 10.10: Display in the device manager.

10.3 Ethernet

For direct connection to a host (PC) or indirect connec­tion using a switch, a double-shielded network cable (e.g. CAT.5, CAT.5e, CAT.5+, CAT.6 or CAT.7) is required, with RJ-45 connectors on both ends. Either a straight or a crossover cable may be used for the network cable.
30
30
As with any IP address range, the rst IP address denotes
the network and the last IP address is used as a broadcast IP address, therefore these two IP addresses must be sub­tracted from the “number of possible host addresses” (see Table 10.1: Private IP address ranges).
In addition to the classication of IP addresses into public
and private address ranges, IP addresses are further divided into classes (Class: A, B, C, D, E). The private IP ad­dress ranges described above would fall within categories
A, B, and C. The classication of IP addresses is important
for the allocation of public IP address ranges, and depends mainly on the size of a local area network (i.e. maximum number of hosts on the network), when it is connecting to the Internet (see Table 10.2: Classes of IP addresses).
IP addresses can be allocated as permanent (static) or variable (dynamic). If permanent IP addresses are assigned in a network, an IP address must be set manually for each network element. If variable IP addresses are automatically (dynamically) assigned to the connected network ele­ments, the task of allocating the IP addresses then requires
Remote Operation
Remote Operation
a Dynamic Host Conguration Protocol (DHCP) server.
When using a DHCP server, a range of IP addresses can be preset for the automatic allocation of the IP addresses. DHCP servers are typically already integrated into routers. If however a network element (device) is directly con­nected via a network cable to a host (PC), no IP addresses can be allocated automatically to the measuring device and the host (PC), because in this case, there is no avail­able network with a DHCP server. They must therefore be
specied manually on the device and the host (PC).
With the help of subnet masks, IP addresses are broken up into the network portion and the host portion, simi-
lar to a phone number being broken up into area prex
(international country code and local area code) and phone number (subscriber number). Subnet masks have the same form as IP addresses. They are represented in the form of four decimal numbers separated by periods (e.g.
255.255.255.0). As with the IP addresses, each decimal number represents a binary number of 8 bits (or one byte). The subnet mask manages the separation of an IP ad­dress into the network portion and the host portion (e.g. the IP address 192.168.10.10 is split up by the subnet mask
255.255.255.0 into a network part 192.168.10.0 and a host part 0.0.0.10). The allocation is achieved by the conver­sion of both IP address and subnet mask into binary form, followed by a bitwise logical AND operation between IP address and subnet mask. The result is the network por­tion of the IP address.
elements (e.g. measuring devices, hosts/PCs etc.) can be simultaneously operated in such a network when using this subnet mask.
Very often, a network will have a default gateway. In most local networks, this gateway is identical to the router used for the Internet (DSL router, ISDN router etc). Using this (gateway) router, a connection can be made to another network. This makes it possible to reach network elements which are not in the same (local area) network, or network elements from the local network can exchange data with network elements from other networks. In order to enable network-wide data exchange, the IP address of the default gateway must also be set. In local networks, the rst IP ad­dress within a network is generally used for this (gateway) router. Routers which are used as a gateway in a local area network often have an IP address with a “1” as the last digit of the IP address (e.g. 192.168.10.1).

10.3.2 Ethernet Settings

PC and device must be in the same network, otherwise a connec­tion is not possible.
The network interface card comes with a USB inter-
face and an Ethernet interface. The conguration of
the required parameters is managed directly by the R&S®HMC8015 after “Ethernet” has been selected as the interface, and the PARAMETER softkey has been pressed.
By contrast, the host portion of the IP address is formed by a bitwise logical NAND operation between IP address and subnet mask. By using subnet masks to achieve a variable (dynamic) allocation of IP addresses into network portion and host portion, the user can specify IP address ranges for large as well as small networks individually. This allows the user to operate both large and small IP networks and, if necessary, connect them through a router to the Internet. In smaller local area networks, the subnet mask of
255.255.255.0 is generally used. In this case, the network
portion (the rst three of the four decimal numbers) and the host part (the last decimal number) can be identied
without much mathematical effort, and up to 254 network
Fig. 10.11: Ethernet settings.
Address range IPv4 subnet mask(s) CIDR notation Number of available host addresses
10.0.0.0 to 10.255.255.255 255.0.0.0 10.0.0.0/8 224 − 2 = 16 777 214
172.16.0.0 to 172.31.255.255 255.240.0.0 172.16.0.0/12 220 − 2 = 1 048 574
192.168.0.0 to 192.168.255.255 255.255.0.0
Table 10.1: Private IP address ranges
Classes Address range
A 0.0.0.1 to 127.255.255.255 8 bit 24 bit 126 16.777.214
B 128.0.0.1 to 191.255.255.255 16 bit 16 bit 16.384 65.534
C 192.0.0.1 to 223.255.255.255 24 bit 8 bit 2.097.151 254
D 224.0.0.1 to 239.255.255.255 Reserved for multicast applications
E 240.0.0.1 to 255.255.255.255 Reserved for special applications
Table 10.2: IP address classes
255.255.255.0
Network por­tion
192.168.0.0/16
192.168.0.0/24
Host por­tion
216 − 2 = 65 534
8
2
− 2 = 254
Max. number of net­works
Max. number of hosts per network
31
31
Remote Operation
Remote Operation
A complete parameter conguration may be performed, including the allocation of a xed IP address. Alternatively,
a dynamic assignment of IP addresses is possible through activation of the DHCP function. Contact your IT manager for more information regarding the correct settings.
Once the device has an IP address, it can be reached with a web browser using this IP, since the Ethernet interface has an embedded web server. When the IP address is en­tered in the browser address bar (http://xxx.xxx.xxx.xxx), a window will appear with details on the device type and serial number.
If DHCP is used and the R&S®HMC8015 cannot obtain an IP address (e.g. if no Ethernet cable is plugged in or the network does not support DHCP), it takes up to three minutes for a “time out” message to be returned, thereby clearing the interface for
another conguration.

10.3.3 LXI

LAN eXtensions for Instrumentation (LAN LXI) is a device platform for measuring instruments and test equipment that is based on standard Ethernet technology. LXI as a LAN-based successor to GPIB is intended to combine the advantages of Ethernet with the simplicity and ease of use of GPIB.
The R&S®HMC8015 is LXI certied and complies with
Version 1.4 (LXI Core 2011). Advanced features are not sup­ported. The LXI certication central components are the In­terchangeable Virtual Instrument (IVI) device drivers. These IVI.net drivers are based on Microsoft’s .NET Framework
4. LabView and LabWindows/CVI drivers that were created based on LabWindows/CVI 2012 are also available.
❙ Displaying the Ethernet settings (Settings)
Fig. 10.13: Ethernet settings
❙ Password assignment (Security)
Fig. 10.14: Password setting
❙ SCPI Terminal

10.3.4 Web Server

The Ethernet interface comes with a web server that can be used with any web browser (e.g. Internet Explorer). The web server supports the following features:
❙ Displaying the device data (Device information)
Fig. 10.12: Display of the device information
Fig. 10.15: SCPI terminal

10.4 IEEE 488.2 / GPIB

In addition to the GPIB features already made available by USB TMC, the R&S®HMC8015 is also available with an option that includes a GPIB interface (R&S®HMC8015-G). This solution may be attractive for those customers who already have an existing GPIB environment. In this way, a new R&S®HMC8015 model can replace any existing outdated equipment with minimal conversion effort. The
conguration of the required parameters is managed by
the R&S®HMC8015 once “IEEE488” has been selected as the interface, and the PARAMETER softkey has been pressed.
The optional IEEE 488 (GPIB) interface can only be installed at
the factory, since the unit must be opened and the warranty seal would have to be broken.
32
32

11 Technical Data

Power Analyzer R&S®HMC8015
Range configuration
CF3 CF6 Peak
voltage 5 V
15 V 30 V 60 V 150 V 300 V 600 V
2.5 V
7.5 V 15 V 30 V 75 V 150 V 300 V
±15 V ±45 V ±90 V ±180 V ±450 V ±900 V ±1800 V
current (500mΩ) 5 mA
10 mA 20 mA 50 mA 100 mA 200 mA
2.5 mA 5 mA 10 mA 25 mA 50 mA 100 mA
±15 mA ±30 mA ±60 mA ±150 mA ±300 mA ±600 mA
current (10mΩ) 0.5 A
1 A 2 A 5 A 10 A 20 A
0.25 A
0.5 A 1 A
2.5 A 5 A 10 A
±1.5 A ±3 A ±6 A ±15 A ±30 A ±60 A
sensor 0.033 V
0.33 V
1.33 V
0.0165 V
0.165 V
0.665 V
±0.1 V ±1 V
±4 V Measurement accuracy (± reading in % ± peak range in %)
Frequency Voltage Current / Sensor Active Power
DC f < 45 Hz 45 Hz < f < 66 Hz 66 Hz < f < 1 kHz 1 kHz < f < 10 kHz 10 kHz < f < 100 kHz
0.05 + 0.05
0.05 + 0.05
0.05 + 0.05
0.05 + 0.1 (0.1 + 0.02 *F) + 0.1 (0.1 + 0.04 *F) + 0.1
0.05 + 0.05
0.05 + 0.05
0.05 + 0.05
0.05 + 0.1 (0.1 + 0.03 *F) + 0.1 (0.1 + 0.04 *F) + 0.2
0.05 + 0.05
0.075 + 0.075
0.05 + 0.05
0.075 + 0.075
(0.1 + 0.07 *F) + 0.1
(0.1 + 0.07 *F) + 0.1 voltage, current: F = frequency in kHz
sensor input: F = frequency in kHz *2 Additional errors
Power factor < 1 ±(0.2 + 0.2 *F)%, only for active power
Common mode error ±0.01 % of peak range
Technical Data
All specifications valid for a temperature range between 20 °C and 30 °C at 80% relative humidity after 60 minutes warm-up
BasicSpecications
Measurement method simultaneous voltage and current sampling
Analog bandwidth DC to 100 kHz
Frequency accuracy 0.1 % of reading
A/D converter resolution 16 bit (voltage), 16 bit (current)
Basic accuracy 0.05 % of reading
Display resolution 5 digits
Sampling frequency 500 kHz
Filters
Analog signal filter Digital filter Frequency filter
1 kHz automatic adaptiv filter 500 Hz (independent from signal filters)
Additional rear panel inputs / outputs (BNC)
Analog input ±10 V
p
Analog input accuracy 0.5 % of reading
Analog output ±5 V
p
Digital input
Low level High level
0 V to 2 V 3 V to 24 V
Digital output 5 V TTL (up to 100 mA source/sink)
Voltage input impedance 2 MΩ
PLL synchronisation sources U, I, external
GeneralSpecications
Display
Type 8.9 cm (3.5“) TFT (Farbe)
Resolution 320 x 240 Pixel (QVGA)
Power supply 100 V
ac
to 115 Vac / 230 Vac @ 50-60 Hz
Power consumption 35 W max, 15 W typ.
Operating temperature 5° C to 40° C
Storage temperature -25° C to 60° C
Standards CAN/CSA-C22.2 NO. 61010-1, UL 61010-1,
DIN EN 61010-1, DIN EN 61326-1, DIN EN 55011
Common mode voltage CAT II, 600 V
rms
Dimensions 222 x 88 x 280 mm
Weight ca 3.250 kg
Warm-up time 60 minutes
Specifications apply to sine wave as input, PF = 1, voltage to earth = 0V, analog signal filter deactivated, digital filter activited and are valid above 1% of range.
Technical Data
Technical Data
33
33
Appendix
Appendix

12 Appendix

12.1 Listofgures
Fig. 1.1: Operating positions. ......................4
Fig. 1.2: Rear panel with supply voltage selector and
connections. ............................6
Fig. 1.3: Connections on the front panel. .............6
Fig. 1.4: Product labeling in accordance with EN 50419. 7
Fig. 2.1: ¸HMC8015 front panel. ................8
Fig. 2.2: ¸HMC8015 rear panel. .................8
Fig. 3.1: Upgrade menu. ..........................9
Fig. 3.2: Manually entering the license key. ..........10
Fig. 4.1: Connecting HZC815 DE version to
¸HMC8015. .........................10
Fig. 4.2: HZC815 DE, UK and US version plug adapters. 10 Fig. 4.3: HZC815 DE version plug adapter with mounting
screws. ...............................11
Fig. 4.4: Internal wiring of A, V and COM jacks. ......11
Fig. 4.5: Voltage-based measurement. .............11
Fig. 4.6: Voltage-based measurement, wiring example. 11
Fig. 4.7: Current-based measurement. ..............12
Fig. 4.8: Current-based measurement, wiring example. 12
Fig. 4.9: HZC51 Clamp-on current probe. ...........12
Fig. 4.10: Sensor menu. ..........................12
Fig. 4.11: Example of a sensor. .....................12
Fig. 4.12: Sensor mode. ..........................13
Fig. 5.1: Screen display description. ................13
Fig. 5.2: Help function. ..........................14
Fig. 5.3: Selecting the measurement function. .......14
Fig. 5.4: Readout display with six active cells. ........14
Fig. 5.5: Readout display with 10 active cells. ........14
Fig. 5.6: Example of measuring range display.. . . . . . . . 15
Fig. 5.7: Selecting the measurement function. .......15
Fig. 6.1: R&S®HMC8015 readout display. ............16
Fig. 6.2: Option / upgrade voucher info box. .........16
Fig. 6.3: VIEW menu. ...........................16
Fig. 6.4: Logarithmic readout display in Harmonics
mode. ................................17
Fig. 6.5: Tabular display in Harmonics mode. ........17
Fig. 6.6: Quick Menu in Waveform mode. ...........17
Fig. 6.7: Trendchart mode. .......................18
Fig. 6.8: Inrush mode. ...........................18
Fig. 6.9: Sample conguration for crest factor 6. ......19
Fig. 6.10: Readout display with integrator output values. 19
Fig. 6.11: MANUAL integrator menu. ................19
Fig. 6.12: Span integration mode selection menu. .....19
Fig. 6.13: Setting the integration duration. ...........20
Fig. 6.14: Inversion. ..............................20
Fig. 6.15: Selecting the data acquisition mode. ........20
Fig. 6.16: Filters arranged as a chain. ................21
Fig. 6.17: Limit menu. ............................21
Fig. 6.18: BNC connectors Digital / Analog IN/OUT. ....21
Fig. 6.19: ANALOG OUT menu. ....................22
Fig. 6.20: DIGITAL IN menu. .......................22
Fig. 7.1: Graphical log le analysis in Excel. ..........23
Fig. 7.2: Setting the integration duration. ...........23
Fig. 7.3: User-dened customization of time stamp for-
matting in Excel. ........................24
Fig. 7.4: Excel Import Wizard (Text Import Wizard). ...24
Fig. 9.1: Firmware update menu. ..................26
Fig. 9.2: Setup menu. ...........................26
Fig. 9.3: Instrument information. ..................26
Fig. 9.4: Entering the instrument name. .............27
Fig. 10.1: Rear panel. ............................28
Fig. 10.2: USB VCP settings. .......................28
Fig. 10.3: NI-VISA 5.4.1. ..........................29
Fig. 10.4: NI-VISA installation guide. ................29
Fig. 10.5: Installing NI-VISA application locally. ........29
Fig. 10.6: Setup menu. ...........................29
Fig. 10.7: Interface menu. .........................29
Fig. 10.8: TMC display on the screen. ...............30
Fig. 10.9: Device driver installation. .................30
Fig. 10.10: Display in the device manager. .............30
Fig. 10.11: Ethernet settings. .......................31
Fig. 10.12: Display of the device information ...........32
Fig. 10.13: Ethernet settings ........................32
Fig. 10.14: Password setting ........................32
Fig. 10.15: SCPI terminal ...........................32

12.2 Glossary

A Active power: 9, 16, 18, 21 Analog In: 13 Apparent power: 9, 16, 18, 21 Average current: 9, 16, 18, 21 Average voltage: 9, 16, 18, 21
D Data acquisition: 23 DC voltage: 4 Device settings: 25
E Energy counter: 9
F File manager: 10 Frequency value for the current: 9, 16, 18, 21 Frequency value for the voltage: 9, 16, 18, 21
G GPIB: 32
H HARMONICS mode: 9, 14 HDS format: 25 Help: 5
I INRUSH mode: 9
K Key fallback time: 27
34
34
L License key: 9, 10 Limits: 6 Logging: 8, 9, 13, 23, 24 M Maintenance: 5 Measurement category: 5, 6 Measuring interval: 23
P Phase shift: 9, 16, 18, 21 Portable Network Graphic: 25 Power factor lambda: 9, 16, 18, 21 Product disposal: 7
R Reactive power: 9, 16, 18, 21 Repair: 5
S Safety class: 4, 5 Screenshots: 25
Appendix
Appendix
T Total harmonic distortion: 9, 16, 18, 21 TRENDCHART mode: 9
U USB: 8
W Warranty: 5 WAVEFORM mode: 9 Windows Bitmap Format: 25
35
35
Appendix
Appendix
36
36
Appendix
Appendix
37
37
Appendix
Appendix
38
38
Appendix
Appendix
39
39
© 2015 Rohde & Schwarz GmbH & Co. KG
Mühldorfstr. 15, 81671 München, Germany
Phone: +49 89 41 29 - 0
Fax: +49 89 41 29 12 164
Email: info@rohde-schwarz.com
Internet: www.rohde-schwarz.com
Customer Support: www.customersupport.rohde-schwarz.com
Service www.service.rohde-schwarz.com
Subject to change – Data without tolerance limits is not binding.
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
Trade names are trademarks of the owners.
5800.5324.02 │ Version 01 │R&S®HMC8015
The following abbreviations are used throughout this manual: R&S®HMC8015 is abbreviated as R&S HMC8015.
Loading...