The firmware of the instrument makes use of several valuable open source software packages. For information, see the "Open Source Acknowledgement" on the user documentation CD-ROM (included in delivery).
Rohde & Schwarz would like to thank the open source community for their valuable contribution to embedded computing.
The R&S FSW Signal and Spectrum Analyzer sets standards in RF performance
and usability. Outstanding key features are:
Unmatched phase noise
●
Excellent dynamic range
●
Up to 160 MHz analysis bandwidth
●
High sensitivity even at low frequencies
●
High measurement rates and fast sweep times with sweep rates up to 1000
●
sweeps per second
Multiple measurement applications can be run and displayed in parallel
●
Easy and intuitive to operate via the large touchscreen user interface and opti-
●
mized user guidance
●
Integrated support of R&S®NRP-Zxx power sensors
For a detailed specification refer to the data sheet.
Due to these features the R&S FSW is ideal for various measurement tasks, for
instance:
Measuring oscillators for radar and communications applications due to the
●
low phase noise
Identifying and analyzing spurious emissions due to the large spurious-free
●
dynamic range and low DANL
Measuring harmonics due to integrated highpass filters
●
Measuring wide-band modulated or frequency-agile signals due to the large
●
bandwidth
Detecting errors caused by interaction between signals by measuring multiple
●
standards simultaneously
5Getting Started 1312.9420.02 ─ 15
Page 6
R&S®FSW
Preface
Documentation Overview
1.2Documentation Overview
The user documentation for the R&S FSW consists of the following parts:
Printed Getting Started manual
●
Online Help system on the instrument
●
Documentation CD-ROM with:
●
–Getting Started
–User Manuals for base unit and firmware applications
–Service Manual
–Release Notes
–Data sheet and product brochures
Online Help
The Online Help is embedded in the instrument's firmware. It offers quick, context-sensitive access to the complete information needed for operation and programming. Online help is available using the
icon on the toolbar of the
R&S FSW.
Web Help
The web help provides online access to the complete information on operating
the R&S FSW and all available options, without downloading. The content of the
web help corresponds to the user manuals for the latest product version. The web
help is available from the R&S FSW product page at http://www.rohde-
schwarz.com/product/FSW.html > Downloads > Web Help.
Getting Started
This manual is delivered with the instrument in printed form and in PDF format on
the CD. It provides the information needed to set up and start working with the
instrument. Basic operations and handling are described. Safety information is
also included.
The Getting Started manual in various languages is also available for download
from the Rohde & Schwarz website, on the R&S FSW product page at http://
www2.rohde-schwarz.com/product/FSW.html.
6Getting Started 1312.9420.02 ─ 15
Page 7
R&S®FSW
Documentation Overview
User Manuals
User manuals are provided for the base unit and each additional (firmware) application.
The user manuals are available in PDF format - in printable form - on the Documentation CD-ROM delivered with the instrument. In the user manuals, all instrument functions are described in detail. Furthermore, they provide a complete
description of the remote control commands with programming examples.
The user manual for the base unit provides basic information on operating the
R&S FSW in general, and the Spectrum application in particular. Furthermore, the
software functions that enhance the basic functionality for various applications are
described here. An introduction to remote control is provided, as well as information on maintenance, instrument interfaces and troubleshooting.
In the individual application manuals, the specific instrument functions of the
application are described in detail. For additional information on default settings
and parameters, refer to the data sheets. Basic information on operating the
R&S FSW is not included in the application manuals.
Preface
All user manuals are also available for download from the Rohde & Schwarz website, on the R&S FSW product page at http://www2.rohde-schwarz.com/product/
FSW.html.
Service Manual
This manual is available in PDF format on the Documentation CD-ROM delivered
with the instrument. It describes how to check compliance with rated specifications, instrument function, repair, troubleshooting and fault elimination. It contains
all information required for repairing the R&S FSW by replacing modules.
Release Notes
The release notes describe the installation of the firmware, new and modified
functions, eliminated problems, and last minute changes to the documentation.
The corresponding firmware version is indicated on the title page of the release
notes.
The most recent release notes are also available for download from the Rohde &
Schwarz website, on the R&S FSW product page at http://www2.rohde-
The following text markers are used throughout this documentation:
ConventionDescription
"Graphical user interface
elements"
KEYSKey names are written in capital letters.
File names, commands,
program code
InputInput to be entered by the user is displayed in italics.
LinksLinks that you can click are displayed in blue font.
"References"References to other parts of the documentation are enclosed by
All names of graphical user interface elements on the screen,
such as dialog boxes, menus, options, buttons, and softkeys are
enclosed by quotation marks.
File names, commands, coding samples and screen output are
distinguished by their font.
quotation marks.
1.3.2Conventions for Procedure Descriptions
When describing how to operate the instrument, several alternative methods may
be available to perform the same task. In this case, the procedure using the
touchscreen is described. Any elements that can be activated by touching can
also be clicked using an additionally connected mouse. The alternative procedure
using the keys on the instrument or the on-screen keyboard is only described if it
deviates from the standard operating procedures.
The term "select" may refer to any of the described methods, i.e. using a finger on
the touchscreen, a mouse pointer in the display, or a key on the instrument or on
a keyboard.
8Getting Started 1312.9420.02 ─ 15
Page 9
R&S®FSW
Preparing for Use
Putting into Operation
2Preparing for Use
●Putting into Operation....................................................................................... 9
●Connecting USB Devices................................................................................21
●Connecting an External Monitor......................................................................23
●Setting Up a Network (LAN) Connection.........................................................25
●Configuring the Initial Instrument Settings...................................................... 30
●Protecting Data Using the Secure User Mode................................................ 31
2.1Putting into Operation
This section describes the basic steps to be taken when setting up the R&S FSW
for the first time.
Risk of injury and instrument damage
The instrument must be used in an appropriate manner to prevent electric
shock, fire, personal injury, or damage.
Do not open the instrument casing.
●
Read and observe the "Basic Safety Instructions" at the beginning of
●
this manual or on the documentation CD-ROM, in addition to the safety
instructions in the following sections. Notice that the data sheet may
specify additional operating conditions.
Risk of instrument damage
Note that the general safety instructions also contain information on operating conditions that will prevent damage to the instrument. The instrument's
data sheet may contain additional operating conditions.
9Getting Started 1312.9420.02 ─ 15
Page 10
R&S®FSW
Risk of electrostatic discharge (ESD)
Electrostatic discharge (ESD) can cause damage to the electronic components of the instrument and the device under test (DUT). ESD is most likely
to occur when you connect or disconnect a DUT or test fixture to the instrument's test ports. To prevent ESD, use a wrist strap and cord and connect
yourself to the ground, or use a conductive floor mat and heel strap combination.
For details, refer to the safety instructions delivered in electronic format on
the documentation CD-ROM.
Risk of instrument damage during operation
Preparing for Use
Putting into Operation
An unsuitable operating site or test setup can cause damage to the instrument and to connected devices. Ensure the following operating conditions
before you switch on the instrument:
All fan openings are unobstructed and the airflow perforations are unim-
●
peded. The minimum distance from the wall is 10 cm.
The instrument is dry and shows no sign of condensation.
●
The instrument is positioned as described in the following sections.
●
The ambient temperature does not exceed the range specified in the
●
data sheet.
Signal levels at the input connectors are all within the specified ranges.
●
Signal outputs are correctly connected and are not overloaded.
●
EMI impact on measurement results
Electromagnetic interference (EMI) may affect the measurement results.
To suppress generated electromagnetic interference (EMI):
Use suitable shielded cables of high quality. For example, use double-
●
shielded RF and LAN cables.
Always terminate open cable ends.
●
Note the EMC classification in the data sheet.
●
10Getting Started 1312.9420.02 ─ 15
Page 11
R&S®FSW
Preparing for Use
Putting into Operation
●Unpacking and Checking the Instrument........................................................ 11
●Placing or Mounting the Instrument................................................................ 12
●Connecting the AC Power...............................................................................15
●Switching the Instrument On and Off.............................................................. 15
●Performing a Self-Alignment and a Selftest.................................................... 16
●Checking the Supplied Options.......................................................................17
2.1.1Unpacking and Checking the Instrument
Check the equipment for completeness using the delivery note and the accessory
lists for the various items. Check the instrument for any damage. If there is damage, immediately contact the carrier who delivered the instrument. Make sure not
to discard the box and packing material.
Packing material
Retain the original packing material. If the instrument needs to be transported or shipped at a later date, you can use the material to protect the control
elements and connectors.
11Getting Started 1312.9420.02 ─ 15
Page 12
R&S®FSW
Preparing for Use
Putting into Operation
Risk of damage during transportation and shipment
Insufficient protection against mechanical and electrostatic effects during
transportation and shipment can damage the instrument.
Always make sure that sufficient mechanical and electrostatic protection
●
is provided.
When shipping an instrument, the original packaging should be used. If
●
you do not have the original packaging, use sufficient padding to prevent the instrument from moving around inside the box. Pack the instrument in antistatic wrap to protect it from electrostatic charging.
Secure the instrument to prevent any movement and other mechanical
●
effects during transportation.
The carrying handles at the front and side of the casing are designed to lift
or carry the instrument. Do not apply an excessive external force to the
handles.
Observe the information on transporting heavy instruments in the basic
safety instructions included at the front of the printed manual and on the
supplied CD ROM.
2.1.2Accessory List
The instrument comes with the following accessories:
Power cable
●
Printed Getting Started manual
●
"R&S FSW User Documentation" CD-ROM
●
2.1.3Placing or Mounting the Instrument
The R&S FSW is designed for use under laboratory conditions, either on a bench
top or in a rack.
Bench Top Operation
If the R&S FSW is operated on a bench top, the surface should be flat. The
instrument can be used in horizontal position, standing on its feet, or with the support feet on the bottom extended.
12Getting Started 1312.9420.02 ─ 15
Page 13
R&S®FSW
Preparing for Use
Putting into Operation
Risk of injury if feet are folded out
The feet may fold in if they are not folded out completely or if the instrument
is shifted. This may cause damage or injury.
Fold the feet completely in or completely out to ensure stability of the
●
instrument. Never shift the instrument when the feet are folded out.
When the feet are folded out, do not work under the instrument or place
●
anything underneath.
The feet can break if they are overloaded. The overall load on the fol-
●
ded-out feet must not exceed 500 N.
F < 500 N
13Getting Started 1312.9420.02 ─ 15
Page 14
R&S®FSW
Risk of injury and instrument damage if stacking instruments
A stack of instruments may tilt over and cause injury. Furthermore, the
instruments at the bottom of the stack may be damaged due to the load
imposed by the instruments on top.
Observe the following instructions when stacking instruments:
Never stack more than three instruments. If you need to stack more
●
than three instruments, install them in a rack.
The overall load imposed on the lowest instrument must not exceed
●
500 N.
All instruments must have the same dimensions (width and length).
●
If the instruments have foldable feet, fold them in completely.
●
Preparing for Use
Putting into Operation
Rackmounting
The R&S FSW can be installed in a rack using a rack adapter kit (order no. see
data sheet). The installation instructions are part of the adapter kit.
Risk of instrument damage in a rack
An insufficient airflow can cause the instrument to overheat, which may disturb the operation and even cause damage.
Make sure that all fan openings are unobstructed, that the airflow perforations are unimpeded, and that the minimum distance from the wall is 10 cm.
14Getting Started 1312.9420.02 ─ 15
Page 15
R&S®FSW
Preparing for Use
Putting into Operation
2.1.4Connecting the AC Power
In the standard version, the R&S FSW is equipped with an AC power supply connector. The R&S FSW can be used with different AC power voltages and adapts
itself automatically to it. Refer to the datasheet for the requirements of voltage
and frequency. The AC power connector is located on the rear panel of the instrument.
For details on the connector refer to chapter 3.2.2, "AC Power Supply Connection
and Main Power Switch", on page 51.
►
Connect the R&S FSW to the AC power supply using the supplied power
cable.
Since the instrument is assembled in line with the specifications for safety
class EN61010, it may only be connected to an outlet that has a ground contact.
2.1.5Switching the Instrument On and Off
Switching the instrument on
► Press the AC power switch on the rear panel to position "I".
The instrument is supplied with AC power. After booting, the instrument is
ready for operation. A green LED above the POWER key indicates this.
An orange LED indicates the instrument is in standby mode
Warm-up time for OCXO
When the instrument is switched on, the OCXO requires an extended
warm-up time (see data sheet). No warm-up time is required when starting
from standby mode.
Switching the instrument off
1. Press the POWER key on the front panel.
15Getting Started 1312.9420.02 ─ 15
Page 16
R&S®FSW
Preparing for Use
Putting into Operation
The R&S FSW switches to standby mode.
2. Set the AC power switch on the rear panel to position "O", or disconnect the
instrument from the AC power supply.
The R&S FSW changes into off mode.
Risk of losing data
If you switch off the running instrument using the rear panel switch or by
disconnecting the power cord, the instrument loses its current settings. Furthermore, program data may be lost.
Press the Power key first to shut down the application properly.
2.1.6Performing a Self-Alignment and a Selftest
During instrument start, the installed hardware is checked against the current firmware version to ensure the hardware is supported. If not, an error
message is displayed ("WRONG_FW") and you are asked to update the
firmware. Until the firmware version is updated, self-alignment fails.
(For details refer to the R&S FSW User Manual).
Furthermore, it may be necessary to align the data to a reference source by performing a self-alignment when strong temperature changes occur.
Operating temperature
Before performing this functional test, make sure that the instrument has
reached its operating temperature (for details, refer to the data sheet).
A message in the status bar ("Instrument warming up...") indicates that the operating temperature has not yet been reached.
Performing a self-alignment
1. Press the SETUP key.
2. Press the "Alignment" softkey.
16Getting Started 1312.9420.02 ─ 15
Page 17
R&S®FSW
Preparing for Use
Putting into Operation
3. Select the "Start Self-Alignment" button in the "Alignment" dialog box.
Once the system correction values have been calculated successfully, a message is displayed.
To display the alignment results again later
Press the SETUP key.
●
Press the "Alignment" softkey.
●
Performing a selftest
The selftest does not need to be repeated every time the instrument is switched
on. It is only necessary when instrument malfunction is suspected.
1. Press the SETUP key.
2. Press the "Service" softkey.
3. Switch to the "Selftest" tab in the "Service" dialog box.
4. Select the "Start Selftest" button.
Once the instrument modules have been checked successfully, a message is
displayed.
2.1.7Checking the Supplied Options
The instrument may be equipped with both hardware and firmware options. In
order to check whether the installed options correspond to the options indicated
on the delivery note, proceed as follows.
1. Press the SETUP key.
2. Press the "System Config" softkey.
3. Switch to the "Versions + Options" tab in the "System Configuration" dialog
box.
A list with hardware and firmware information is displayed.
4. Check the availability of the hardware options as indicated in the delivery
note.
17Getting Started 1312.9420.02 ─ 15
Page 18
R&S®FSW
Preparing for Use
Windows Operating System
2.2Windows Operating System
The instrument contains the Windows 7 operating system which has been configured according to the instrument's features and needs. Changes in the system
setup are only required when peripherals like keyboard or a printer are installed
or if the network configuration does not comply with the default settings. After the
R&S FSW is started, the operating system boots and the instrument firmware is
started automatically.
To ensure that the instrument software functions properly, certain rules must be
adhered to concerning the operating system.
Risk of causing instrument unusability
The instrument is equipped with the Windows 7 operating system. Additional software can therefore be installed on the instrument. The use and
installation of additional software may impair instrument function. Thus, run
only programs that Rohde & Schwarz has tested for compatibility with the
instrument software.
The drivers and programs used on the instrument under Windows 7 have
been adapted to the instrument. Existing instrument software must always
be modified using only update software released by Rohde & Schwarz.
The following program packages have been tested:
R&S Power Viewer Plus - virtual power meter for displaying results of the
●
Power Sensor R&S NRP-Zxx (install only this component!)
Symantec Endpoint Security – Virus-protection software
●
FileShredder - for reliable deletion of files on the hard disk
●
2.2.1Virus Protection
Users must take appropriate steps to protect their instruments from infection.
Beside the use of strong firewall settings and regularly scanning any removable
storage device used with a Rohde & Schwarz instrument, it is also recommended
that anti-virus software be installed on the instrument. While Rohde & Schwarz
does NOT recommend running anti-virus software in the background ("on-
18Getting Started 1312.9420.02 ─ 15
Page 19
R&S®FSW
Preparing for Use
Windows Operating System
access" mode) on Windows-based instruments, due to potentially degrading
instrument performance, it does recommend running it during non-critical hours.
For details and recommendations, see the Rohde & Schwarz White Paper "Malware Protection" available at http://www.rohde-schwarz.com/appnote/1EF73.
2.2.2Service Packs and Updates
Microsoft regularly creates security updates and other patches to protect Windows-based operating systems. These are released through the Microsoft Update
website and associated update server. Instruments using Windows, especially
those that connect to a network, should be updated regularly.
For more details and information on configuring automatic updates see the R&S
White Paper "Malware Protection" (available at http://www.rohde-schwarz.com/
appnote/1EF73).
2.2.3Login
Windows 7 requires that users identify themselves by entering a user name and
password in a login window. By default, the R&S FSW provides two user
accounts: an administrator account ("Instrument") with unrestricted access to the
computer/domain, and a standard user account ("NormalUser") with limited
access. In both cases the initial password is 894129. Note that this password is
very weak, and it is recommended that you change the password for both users
after initial login. You can change the password in Windows 7 for any user at any
time via "Start > Control Panel > User Accounts".
Secure User Mode
If the secure user mode option (R&S FSW-K33) is installed, an additional
account is provided: the "SecureUser", also with the password 894129.
When the secure user mode is activated the first time after installation, the
user is prompted to change the passwords for all user accounts in order to
improve system security.
For details see chapter 2.7, "Protecting Data Using the Secure User Mode",
on page 31.
For the administrator account, an automatic login function is active by default. If
activated, login is carried out automatically for the administrator (with full access)
19Getting Started 1312.9420.02 ─ 15
Page 20
R&S®FSW
Preparing for Use
Windows Operating System
in the background when the R&S FSW is started, without having to enter a password.
Some administrative tasks require administrator rights (e.g. the configuration of a
LAN network). Refer to the description of the basic instrument setup (SETUP
menu) to find out which functions are affected.
2.2.3.1The Automatic Login Function
Deactivating the automatic login function
When shipped, the instrument is already configured to automatically log on under
Windows 7. To deactivate the automatic login function, perform the following
steps:
1.
Select the "Windows" icon in the toolbar to access the operating system of the
R&S FSW (see also chapter 2.2.4, "Accessing the Start Menu",
on page 21).
2. In the "Start" menu, select "Run".
The "Run" dialog box is displayed.
3. Enter the command C:\R_S\INSTR\USER\NO_AUTOLOGIN.REG.
4. Press the ENTER key to confirm.
The automatic login function is deactivated. The next time you switch on the
instrument, you are prompted to enter your user name and password before
the firmware is started.
Reactivating the automatic login function
1. In the "Start" menu, select "Run".
The "Run" dialog box is displayed.
2. Enter the command C:\R_S\INSTR\USER\AUTOLOGIN.REG.
3. Press the ENTER key to confirm.
The automatic login function is reactivated. It will be applied the next time the
instrument is switched on.
20Getting Started 1312.9420.02 ─ 15
Page 21
R&S®FSW
Preparing for Use
Connecting USB Devices
Switching users when using the automatic login function
Which user account is used is defined during login. However, you can also switch
the user account to be used when the automatic login function is active.
► In the "Start" menu, select the arrow next to the "Shut down" button and then
"Log off".
The "Login" dialog box is displayed, in which you can enter the different user
account name and password.
2.2.4Accessing the Start Menu
The Windows "Start" menu provides access to the Windows 7
functionality and installed programs. To open the "Start" menu,
select the "Windows" icon on the R&S FSW toolbar, or press the "Windows" key
or the CTRL + ESC key combination on the (external) keyboard.
All necessary system settings can be defined in the "Start > Control Panel" menu
(for required settings refer to the Windows 7 documentation and to the hardware
description).
2.2.5Accessing the Windows Task Bar
The Windows task bar also provides quick access to commonly used programs,
for example Paint or Wordpad, but also IECWIN, the auxiliary remote control tool
provided free of charge and installed by Rohde & Schwarz.
For details on the IECWIN tool, see the "Network and Remote Control"
chapter of the R&S FSW User Manual.
To open the task bar, select the "Windows" icon on the R&S FSW
toolbar, or press the "Windows" key or the CTRL + ESC key combination on your (external) keyboard.
2.3Connecting USB Devices
The USB interfaces on the front and (optionally) rear panels of the R&S FSW
allow you to connect USB devices directly to the instrument. This number can be
21Getting Started 1312.9420.02 ─ 15
Page 22
R&S®FSW
Connecting USB Devices
increased as necessary by using USB hubs. Due to the large number of available
USB devices, there is almost no limit to the expansions that are possible with the
R&S FSW.
The following list shows various USB devices that can be useful:
Memory stick for easy transfer of data to/from a computer (e.g. firmware
●
updates)
CD-ROM drives for easy installation of firmware applications
●
Keyboard or mouse to simplify the entry of data, comments, file names, etc.
●
Printer for printing out measurement results
●
Power sensors, e.g. of the NRP Zxy family
●
Installing USB devices is easy under Windows, because all USB devices are
plug&play. After a device is connected to the USB interface, automatically
searches for a suitable device driver.
Preparing for Use
If Windows does not find a suitable driver, it will prompt you to specify a directory
that contains the driver software. If the driver software is on a CD, connect a USB
CD-ROM drive to the instrument before proceeding.
When a USB device is subsequently disconnected from the R&S FSW, Windows
immediately detects the change in hardware configuration and deactivates the
corresponding driver.
All USB devices can be connected to or disconnected from the instrument during
operation.
Connecting a memory stick or CD-ROM drive
If installation of a memory stick or CD-ROM drive is successful, Windows informs
you that the device is ready to use. The device is made available as a new drive
and is displayed in Windows Explorer. The name of the drive depends on the
manufacturer.
Connecting a keyboard
The keyboard is detected automatically when it is connected. The default input
language is English – US.
However, you can also connect foreign language keyboards; currently the following languages are supported for the R&S FSW:
German
●
Swiss
●
22Getting Started 1312.9420.02 ─ 15
Page 23
R&S®FSW
Preparing for Use
Connecting an External Monitor
French
●
Russian
●
Select "Start > Control Panel > Clock, Language, and Region > Change keyboards or other input methods" to configure the keyboard properties. To access
the Windows operating system, press the Windows key on the external keyboard.
Connecting a mouse
The mouse is detected automatically when it is connected.
Select "Start > Control Panel > Devices and Printers > Mouse" to configure the
mouse properties. To access the Windows operating system, press the Windows
key on the external keyboard.
Connecting a printer
When printing a file, the instrument checks whether a printer is connected and
turned on and whether the appropriate printer driver is installed. If required,
printer driver installation is initiated. A printer driver needs to be installed only
once.
To install a printer, select "Start > Control Panel > Devices and Printers > Add a
printer". To access the Windows operating system, press the Windows key on the
external keyboard.
You can load updated and improved driver versions or new drivers from an installation disk, USB memory stick or another external storage medium. If the instrument is integrated in a network, you can also install driver data stored in a network directory.
Select "Start > Control Panel > Devices and Printers > Device Manager > Update
Device drivers" to install the driver.
2.4Connecting an External Monitor
You can connect an external monitor (or projector) to the DVI or DISPLAY PORT
connector on the instrument's rear panel (see also chapter 3.2.3, "DISPLAY
PORT and DVI", on page 51).
23Getting Started 1312.9420.02 ─ 15
Page 24
R&S®FSW
Screen resolution and format
The touchscreen of the R&S FSW is calibrated for a 16:10 format. If you
connect a monitor or projector using a different format (e.g. 4:3), the calibration will not be correct and the screen will not react to your touch actions
properly.
The touchscreen has a screen resolution of 1280x800 pixels. Most external
monitors have a higher screen resolution. If the screen resolution of the
monitor is set higher than the instrument's resolution, the application window uses an area of 1280x800 pixels on the monitor display. For full screen
display, adjust the monitor's screen resolution.
The R&S FSW supports a minimum resolution of 1280x768 pixels.
1.
Connect the external monitor to the R&S FSW.
2. Press the SETUP key.
Preparing for Use
Connecting an External Monitor
3. Press the "Display" softkey.
4. Select the "Configure Monitor" tab in the "Display" dialog box.
The standard Windows "Screen Resolution" dialog box is displayed.
5. If necessary, change the screen resolution to be used. Consider the informa-
tion in the note above.
24Getting Started 1312.9420.02 ─ 15
Page 25
R&S®FSW
Preparing for Use
Setting Up a Network (LAN) Connection
6. Select the instrument to be used for display:
●"Display 1": internal monitor only
●"Display 2": external monitor only
●"Duplicate": both internal and external monitor
7. Tap "Apply" to try out the settings before they are accepted permanently, then
you can easily return to the previous settings, if necessary.
8. Select "OK" if the settings are suitable.
2.5Setting Up a Network (LAN) Connection
The R&S FSW is equipped with a network interface and can be connected to an
Ethernet LAN (local area network). Provided the appropriate rights have been
assigned by the network administrator and the Windows firewall configuration is
adapted accordingly, the interface can be used, for example:
To transfer data between a controlling device and the test device, e.g. in order
●
to run a remote control program.
See chapter "Remote Control" in the R&S FSW User Manual.
To access or control the measurement from a remote computer using the
●
"Remote Desktop" application (or a similar tool)
To connect external network devices (e.g. printers)
●
To transfer data from a remote computer and back, e.g. using network folders
●
This section describes how to configure the LAN interface. It includes the following topics:
chapter 2.5.1, "Connecting the Instrument to the Network", on page 26
●
chapter 2.5.2, "Assigning the IP Address", on page 27
●
Note that only user accounts with administrator rights can configure LAN networks.
LXI
The R&S FSW complies with LXI Class C. LXI gives you direct access to
the LAN settings described below.
For further information on the LXI interface refer to the User Manual.
25Getting Started 1312.9420.02 ─ 15
Page 26
R&S®FSW
Preparing for Use
Setting Up a Network (LAN) Connection
2.5.1Connecting the Instrument to the Network
There are two methods to establish a LAN connection to the instrument:
A non-dedicated network (Ethernet) connection from the instrument to an
●
existing network made with an ordinary RJ-45 network cable. The instrument
is assigned an IP address and can coexist with a computer and with other
hosts on the same network.
A dedicated network connection (Point-to-point connection) between the
●
instrument and a single computer made with a (crossover) RJ-45 network
cable. The computer must be equipped with a network adapter and is directly
connected to the instrument. The use of hubs, switches, or gateways is not
required, however, data transfer is still performed using the TCP/IP protocol.
An IP address has to be assigned to the instrument and the computer, see
chapter 2.5.2, "Assigning the IP Address", on page 27.
Note: As the R&S FSW uses a 1 GBit LAN, a crossover cable is not neces-
sary (due to Auto-MDI(X) functionality).
Risk of network failure
Before connecting the instrument to the network or configuring the network,
consult your network administrator. Errors may affect the entire network.
► To establish a non-dedicated network connection, connect a commercial
RJ-45 cable to one of the LAN ports.
To establish a dedicated connection, connect a (crossover) RJ-45 cable
between the instrument and a single PC.
If the instrument is connected to the LAN, Windows automatically detects the network connection and activates the required drivers.
The network card can be operated with a 1 GBit Ethernet IEEE 802.3u interface.
26Getting Started 1312.9420.02 ─ 15
Page 27
R&S®FSW
Preparing for Use
Setting Up a Network (LAN) Connection
2.5.2Assigning the IP Address
Depending on the network capacities, the TCP/IP address information for the
instrument can be obtained in different ways.
If the network supports dynamic TCP/IP configuration using the Dynamic Host
●
Configuration Protocol (DHCP), all address information can be assigned automatically.
If the network does not support DHCP, or if the instrument is set to use alter-
●
nate TCP/IP configuration, the addresses must be set manually.
By default, the instrument is configured to use dynamic TCP/IP configuration and
obtain all address information automatically. This means that it is safe to establish
a physical connection to the LAN without any previous instrument configuration.
Risk of network errors
Connection errors can affect the entire network. If your network does not
support DHCP, or if you choose to disable dynamic TCP/IP configuration,
you must assign valid address information before connecting the instrument
to the LAN. Contact your network administrator to obtain a valid IP address.
Assigning the IP address on the instrument
1. Press the SETUP key.
2. Press the "Network + Remote" softkey.
3. Select the "Network" tab.
4. In the "Network + Remote" dialog, toggle the "DHCP On/Off" setting to the
required mode.
If DHCP is "Off", you must enter the IP address manually, as described in the
following steps.
Note: When DHCP is changed from "On" to "Off", the previously set IP
address and subnet mask are retrieved.
If DHCP is "On", the IP address of the DHCP server is obtained automatically.
The configuration is saved, and you are prompted to restart the instrument.
You can skip the remaining steps.
Note: When a DHCP server is used, a new IP address may be assigned each
time the instrument is restarted. This address must first be determined on the
instrument itself. Thus, when using a DHCP server, it is recommended that
27Getting Started 1312.9420.02 ─ 15
Page 28
R&S®FSW
Setting Up a Network (LAN) Connection
you use the permanent computer name, which determines the address via the
DNS server (see "Using a DNS server to determine the IP address"
on page 28 and chapter 2.5.3, "Using computer names", on page 29).
5. Enter the "IP Address", for example 192.0.2.0. The IP address consists of four
number blocks separated by dots. Every block contains 3 numbers in maximum.
6. Enter the "Subnet Mask", for example 255.255.255.0. The subnet mask con-
sists of four number blocks separated by dots. Every block contains 3 numbers in maximum.
7. Select "Configure Network".
If you have entered an invalid IP address or subnet mask, the message "out of
range" is displayed in the status line. If the settings are correct, the configuration is saved, and you are prompted to restart the instrument.
8. Confirm the displayed message ("Yes" button) to restart the instrument.
Preparing for Use
Using a DNS server to determine the IP address
If a DNS server is configured on the R&S FSW, the server can determine the current IP address for the connection using the permanent computer name.
1. Obtain the name of your DNS domain and the IP addresses of the DNS and
WINS servers on your network.
2. Press the SETUP key and then the "Network + Remote" softkey.
3. In the "Network" tab, select the "Open Dialog 'Network Configuration" button.
4. Double-tap the "Local Area Connection" icon.
5. In the "Local Area Connection Status" dialog box, select the "Properties" but-
ton.
The items used by the LAN connection are displayed.
6. Tap the entry named "Internet Protocol Version 4 (TCP/IPv4)" to highlight it.
28Getting Started 1312.9420.02 ─ 15
Page 29
R&S®FSW
Preparing for Use
Setting Up a Network (LAN) Connection
7. Select the "Properties" button.
8. On the "General" tab, select "Use the following DNS server addresses" and
enter your own DNS addresses.
For more information refer to the Windows operating system Help.
2.5.3Using computer names
In a LAN that uses a DNS server (Domain Name System server), each PC or
instrument connected in the LAN can be accessed via an unambiguous computer
name instead of the IP address. The DNS server translates the host name to the
IP address. This is especially useful when a DHCP server is used, as a new IP
address may be assigned each time the instrument is restarted.
Each instrument is delivered with an assigned computer name, but this name can
be changed.
To change the computer name
1. Press the SETUP key and then the "Network + Remote" softkey.
The current "Computer Name" is displayed in the "Network" tab.
29Getting Started 1312.9420.02 ─ 15
Page 30
R&S®FSW
Configuring the Initial Instrument Settings
2. Enter the new computer name and close the dialog box.
Preparing for Use
2.5.4Changing the Windows Firewall Settings
A firewall protects an instrument by preventing unauthorized users from gaining
access to it through a network. Rohde & Schwarz highly recommends the use of
the firewall on your instrument. Rohde & Schwarz instruments are shipped with
the Windows firewall enabled and preconfigured in such a way that all ports and
connections for remote control are enabled. For more details on firewall configuration see the Rohde & Schwarz White Paper "Malware Protection" (available at
http://www2.rohde-schwarz.com/file_13784/1EF73_0E.pdf) and the Windows
Help system.
Note that changing firewall settings requires administrator rights.
2.6Configuring the Initial Instrument Settings
This section describes how to setup the R&S FSW initially. For further basic
instrument settings see the R&S FSW User Manual.
●Setting the Date and Time.............................................................................. 30
2.6.1Setting the Date and Time
Users with administrator rights can set the date and time for the internal real time
clock as follows:
Opening the Date and Time Properties dialog box
1. Press the SETUP key.
2. Press the "Display" softkey.
3. Select the "General" tab in the "Display" dialog box.
4. Press the "Set Date and Time" button to open the standard Windows "Date
and Time Properties" dialog box.
30Getting Started 1312.9420.02 ─ 15
Page 31
R&S®FSW
Preparing for Use
Protecting Data Using the Secure User Mode
5. If necessary, toggle the "Date and Time Format" between German (DE) and
US.
After you have made a change and closed the dialog box, the new date and
time is also adopted by the instrument.
2.7Protecting Data Using the Secure User Mode
During normal operation, the R&S FSW uses a solid-state drive to store its operating system, instrument firmware, instrument self-alignment data, and any user
data created during operation. If necessary, the solid-state drive can be removed
from the R&S FSW and locked in a secure place to protect any classified data it
may contain.
Redirecting storage to volatile memory
Alternatively, to avoid storing any sensitive data on the R&S FSW permanently,
the secure user mode was introduced (option R&S FSW-K33). In secure user
mode the instrument’s solid-state drive is write-protected so that no information
can be written to memory permanently. Data that the R&S FSW normally stores
on the solid-state drive is redirected to volatile memory instead, which remains
available only until the instrument is switched off. This data includes:
Windows 7 operating system files
●
Firmware shutdown files containing information on last instrument state
●
Self-alignment data
●
General instrument settings such as the IP address
●
Measurement settings
●
User data created during operation
●
●Any data created by other applications installed on the R&SFSW, for example
text editors (Notepad), the Clipboard, drawing tools etc.
Data that is stored in volatile memory can be accessed by the user just as in normal operation; however, when the instrument’s power is removed, all data in this
memory is cleared. Thus, in secure user mode, the instrument always starts in a
defined, fixed state when switched on.
In order to store data such as measurement results permanently, it must be
stored to an external storage device, such as a memory stick.
31Getting Started 1312.9420.02 ─ 15
Page 32
R&S®FSW
Protecting Data Using the Secure User Mode
Limited storage space
The volatile memory used to store data in secure user mode is restricted to
256 MB. Thus, a "Memory full" error may occur although the hard disk indicates that storage space is still available.
Storing required data permanently
Any data that is to be available for subsequent sessions with the R&S FSW, such
as predefined instrument settings, transducer factors or self-alignment data, must
be stored on the instrument permanently, prior to activating the secure usermode.
Self-alignment data
Note that self-alignment data becomes invalid with time and due to temperature changes. Therefore, to achieve optimal accuracy, it may be preferable
to perform a new self-alignment at the start of each new session on the
R&S FSW.
Preparing for Use
Restricted operation
Due to the fact that permanent storage is not possible, the following functions are
not available in secure user mode:
Firmware update
●
Activating a new option key
●
Activating and deactivating secure user mode
Only a user with administrator rights can activate the secure user mode. Once
activated, a restart is required. The special user "SecureUser" is then logged on
to the R&S FSW automatically (using the automatic login function, see chap-
ter 2.2.3.1, "The Automatic Login Function", on page 20). While the secure user
mode is active, a message is displayed in the status bar at the bottom of the
screen.
32Getting Started 1312.9420.02 ─ 15
Page 33
R&S®FSW
Protecting Data Using the Secure User Mode
Secure Passwords
By default, the initial password for both the administrator account ("Instrument") and the "SecureUser" account is "894129". When the secure user
mode is activated the first time after installation, you are prompted to
change the passwords for all user accounts in order to improve system
security. Although it is possible to continue without changing the passwords, it is strongly recommended that you do so.
You can change the password in Windows 7 for any user at any time via
"Start > Control Panel > User Accounts".
In order to deactivate the secure user mode, the "SecureUser" must log off and
the "Instrument" user (administrator) must log on.
Switching users when using the automatic login function
Preparing for Use
In the "Start" menu, select the arrow next to the "Shut down" button and
then "Log off".
The "Login" dialog box is displayed, in which you can enter the different
user account name and password.
The secure user mode setting and automatic login is automatically deactivated
when the "Instrument" user logs on. The "SecureUser" is no longer available.
The secure user mode setting is available in the general system configuration settings (SETUP key > "System Configuration" softkey > "Config" tab > "Secure
User Mode": "ON", see the R&S FSW User Manual).
Remote control
Initially after installation of the R&S FSW-K33 option, secure user mode must be
enabled manually once before remote control is possible. This is necessary to
prompt for a change of passwords.
33Getting Started 1312.9420.02 ─ 15
Page 34
R&S®FSW
Instrument Tour
Front Panel View
3Instrument Tour
3.1Front Panel View
This chapter describes the front panel, including all function keys and connectors.
Fig. 3-1: Front panel view
1 = Touchscreen
2 = POWER key
3 = SYSTEM keys
4 = Headphones connector and volume control
5 = USB connectors
6 = POWER SENSOR connector
7 = PROBE connector
8 = NOISE SOURCE CONTROL
9 = Function keys
10 = Keypad
11 = Navigation controls
12 =
(Analog) Baseband Input 50Ω connectors for I/Q signal or Rohde&Schwarz active probes
(optional)
13 =
(Analog) Baseband Input 50Ω connectors for inverse part of differential I/Q signal (optional)
14 = TRIGGER INPUT/OUTPUT connectors
15 = EXT MIXER connector (optional)
16 =
RF Input 50 Ω connector
34Getting Started 1312.9420.02 ─ 15
Page 35
R&S®FSW
Instrument Tour
Front Panel View
Instrument damage caused by cleaning agents
Cleaning agents contain substances that may damage the instrument, for
example cleaning agents that contain a solvent may damage the front panel
labeling, plastic parts, or the display.
Never use cleaning agents such as solvents (thinners, acetone, etc), acids,
bases, or other substances.
The outside of the instrument can be cleaned sufficiently using a soft, lintfree dust cloth.
3.1.1Touchscreen
All measurement results are displayed on the screen on the front panel. Additionally, the screen display provides status and setting information and allows you to
switch between various measurement tasks. The screen is touch-sensitive, offering an alternative means of user interaction for quick and easy handling of the
instrument.
Risk of touchscreen damage during operation
The touchscreen may be damaged by inappropriate tools or excessive
force.
Observe the following instructions when operating or cleaning the touchscreen:
Never touch the screen with ball point pens or other pointed objects with
●
sharp edges.
It is recommended that you operate the touchscreen by finger only.
●
As an alternative, you may use a stylus pen with a smooth soft tip.
Never apply excessive force to the screen. Touch it gently.
●
Never scratch the screen surface, for example with a finger nail.
●
Never rub it strongly, for example with a dust cloth.
35Getting Started 1312.9420.02 ─ 15
Page 36
R&S®FSW
Instrument Tour
Front Panel View
1
2
3
6
7
4
5
8
9
Fig. 3-2: Touchscreen elements
1 = Toolbar with standard application functions, e.g. print, save/open file etc.
2 = Tabs for individual measurement channels
3 = Channel bar for firmware and measurement settings
4 = Input field for measurement setting
5 = Softkeys for function access
6 = Window title bar with diagram-specific (trace) information
7 = Measurement results area
8 = Diagram footer with diagram-specific information, depending on application
9 = Instrument status bar with error messages, progress bar and date/time display
A touchscreen is a screen that is touch-sensitive, i.e. it reacts in a specified way
when a particular element on the screen is tapped by a finger or a pointing
device, for example. Any user interface elements that can be clicked on by a
mouse pointer can also be tapped on the screen to trigger the same behavior,
and vice versa. Using the touchscreen, the following tasks (among others) can be
performed by the tap of your finger (see also chapter 4, "Trying Out the Instru-
ment", on page 58):
Changing a setting
●
Changing the display
●
36Getting Started 1312.9420.02 ─ 15
Page 37
R&S®FSW
Instrument Tour
Front Panel View
Moving a marker
●
Zooming into a diagram
●
Selecting a new evaluation method
●
Scrolling through a result list
●
Saving or printing results and settings
●
To imitate a right-click by mouse using the touchscreen, for example to open a
context-sensitive menu for a specific item, press the screen for about 1 second.
3.1.1.1Toolbar
Standard functions can be performed via the icons in the toolbar at the top of the
screen.
You can hide the toolbar display, e.g. when using remote control, in order to
enlarge the display area for the measurement results ("Setup > Display >
Displayed Items"). See the R&S FSW User Manual for details.
The following functions are available:
Table 3-1: Standard Application Functions in the Toolbar
IconDescription
Windows: displays the Windows "Start" menu and task bar
Open: opens a file from the instrument ("Save/Recall" menu)
Store: stores data on the instrument ("Save/Recall" menu)
Print: defines print settings ("Print" menu)
Undo: reverts last operation
Redo: repeats previously reverted operation
Selection mode: the cursor can be used to select (and move) markers in a zoomed
display
37Getting Started 1312.9420.02 ─ 15
Page 38
R&S®FSW
Instrument Tour
Front Panel View
IconDescription
Zoom mode: displays a dotted rectangle in the diagram that can be expanded to
define the zoom area
Multiple zoom mode: multiple zoom areas can be defined for the same diagram
Zoom off: displays the diagram in its original size
SmartGrid: activates "SmartGrid" mode to configure the screen layout
Sequencer: opens the "Sequencer" menu to perform consecutive measurements
Help (+ Select): allows you to select an object for which context-specific help is displayed
ALAnalysis line (MSRA mode only): opens a dialog box to display and position an analy-
sis line in MSRA measurements (see R&S FSW MSRA User Manual)
Help: displays context-sensitive help topic for currently selected element
Print immediately: prints the current display (screenshot) as configured
In "SmartGrid" mode only:
Exit "SmartGrid" mode
3.1.1.2Softkeys
Softkeys are virtual keys provided by the software. Thus, more functions can be
provided than those that can be accessed directly via the function keys on the
instrument. Softkeys are dynamic, i.e. depending on the selected function key, a
different list of softkeys is displayed on the right side of the screen.
A list of softkeys for a certain function key is also called a menu. Softkeys can
either perform a specific function or open a dialog box.
The "More" softkey indicates that the menu contains more softkeys than can be
displayed at once on the screen. When pressed, it displays the next set of softkeys.
38Getting Started 1312.9420.02 ─ 15
Page 39
R&S®FSW
Instrument Tour
Front Panel View
Recognizing the softkey status by color
ColorMeaning
orangeassociated dialog box is open
blueassociated function is active; for toggle keys: currently active state
grayinstrument function is temporarily not available due to a specific setting or
missing option
You can hide the softkey display, e.g. when using remote control, in order
to enlarge the display area for the measurement results ("Setup > Display >
Displayed Items"). See the User Manual for details.
3.1.1.3Context-sensitive Menus
Markers and traces in the display, as well as the information in the channel bar,
have context-sensitive menus. If you right-click on a marker or trace in the display, or information in the channel bar (or tap it for about 1 second), a menu is
displayed which contains the same functions as the corresponding softkey. This
is useful, for example, when the softkey display is hidden.
39Getting Started 1312.9420.02 ─ 15
Page 40
R&S®FSW
Instrument Tour
Front Panel View
3.1.1.4On-screen Keyboard
The on-screen keyboard is an additional means of interacting with the instrument
without having to connect an external keyboard.
The on-screen keyboard display can be switched on and off as desired using the
"On-Screen Keyboard" function key beneath the screen.
When you press this key, the display switches between the following options:
Keyboard displayed at the top of the screen
●
Keyboard displayed at the bottom of the screen
●
No keyboard displayed
●
You can use the TAB key on the on-screen keyboard to move the focus
from one field to another in dialog boxes.
3.1.2POWER Key
The POWER key is located on the lower left corner of the front panel. It starts up
and shuts down the instrument.
See also chapter 2.1.5, "Switching the Instrument On and Off", on page 15.
3.1.3SYSTEM Keys
SYSTEM keys set the instrument to a predefined state, change basic settings,
and provide print and display functions.
40Getting Started 1312.9420.02 ─ 15
Page 41
R&S®FSW
Instrument Tour
Front Panel View
A detailed description of the corresponding functions is provided in the User Manual.
Reference frequency (external/internal), noise source
●
Date, time, display configuration
●
LAN interface
●
Self-alignment
●
Firmware update and enabling of options
●
Information about instrument configuration incl. firmware ver-
●
sion and system error messages
Service support functions (self test etc.)
●
Switches between the on-screen keyboard display:
at the top of the screen
●
at the bottom of the screen
●
off
●
Switches between maximized and split display of focus area.
Moves focus area from one active window to the next.
3.1.4PHONES and VOLUME
Headphones equipped with a miniature jack plug can be connected to the
PHONES female connector. The internal impedance is 10 Ω. The output voltage
can be set by using the VOLUME control to the right of the female connector. If a
plug is connected, the internal loudspeaker is automatically switched off.
To use the PHONES connector, in the "Output" dialog box (INPUT/OUTPUT key),
select "Video Output" = "Video". The output voltage (volume) is 1V.
41Getting Started 1312.9420.02 ─ 15
Page 42
R&S®FSW
Instrument Tour
Front Panel View
Risk of hearing damage
To protect your hearing, make sure that the volume setting is not too high
before putting on the headphones.
3.1.5USB
The front panel provides three female USB connectors (USB-A) to connect devices like a keyboard (recommended: R&S PSL-Z2, order number 1157.6870.03)
or a mouse (recommended: R&S PSL-Z10, order number 1157.7060.03). In addition, a memory stick can be connected to store and reload instrument settings
and measurement data.
The rear panel provides further USB connectors, including a male (USB-B)
connector. See chapter 3.2.5, "USB", on page 52.
All USB connectors support standard 2.0.
3.1.6POWER SENSOR
The LEMOSA female connector is used to connect power sensors of the R&S
NRP-Zxy family. For a detailed list of supported sensors see the data sheet.
For details on configuring and using power sensors, see the User Manual.
3.1.7PROBE
The R&S FSW provides a connector for supply voltages of +15 V to -12 V and
ground for active probes and preamplifiers. A maximum current of 140 mA is
available. This connector is suitable as power supply for high-impedance probes.
3.1.8NOISE SOURCE CONTROL
The noise source control female connector is used to provide the supply voltage
for an external noise source, e.g., to measure the noise figure and gain of amplifiers and frequency converting devices.
42Getting Started 1312.9420.02 ─ 15
Page 43
R&S®FSW
Instrument Tour
Front Panel View
Conventional noise sources require a voltage of +28 V in order to be switched on
and 0 V to be switched off. The output supports a maximum load of 100 mA.
3.1.9Function Keys
Function keys provide access to the most common measurement settings and
functions.
A detailed description of the corresponding functions is provided in the User Manual.
Table 3-3: Function keys
Function keyAssigned functions
Basic measurement settings
FREQSets the center frequency as well as the start and stop frequen-
cies for the frequency range under consideration. This key is also
used to set the frequency offset and the signal track function.
SPANSets the frequency span to be analyzed.
AMPTSets the reference level, the displayed dynamic range, the RF
attenuation and the unit for the level display.
Sets the level offset and the input impedance.
Activates the preamplifier (option RF Preamplifier, R&S FSW-
B24).
AUTO SETEnables automatic settings for level, frequency or sweep type
mode.
BWSets the resolution bandwidth and the video bandwidth.
SWEEPSets the sweep time and the number of measurement points.
Selects continuous measurement or single measurement.
TRACEConfigures the measured data acquisition and the analysis of the
measurement data.
TRIGSets the trigger mode, the trigger threshold, the trigger delay, and
the gate configuration in the case of gated sweep.
Marker functions
MKRSets and positions the absolute and relative measurement mark-
ers (markers and delta markers).
PEAK SEARCHPerforms a peak search for active marker. If no marker is active,
normal marker 1 is activated and the peak search is performed
for it.
43Getting Started 1312.9420.02 ─ 15
Page 44
R&S®FSW
Instrument Tour
Front Panel View
Function keyAssigned functions
MKR FUNCProvides additional analysis functions of the measurement mark-
ers:
Frequency counter (Sig Count)
Fixed reference point for relative measurement markers (Ref
Fixed)
Noise marker (Noise Meas)
Phase noise (Phase Noise)
n dB down function
AM/FM audio demodulation
Peak list
MKR->Used for search functions of the measurement markers (maxi-
mum/minimum of the trace).
Assigns the marker frequency to the center frequency, and the
marker level to the reference level.
Restricts the search area (Search Limits) and characterizes the
maximum points and minimum points (Peak Excursion).
Measurement and evaluation functions
MEASProvides the measurement functions.
Measurement of multicarrier adjacent channel power (Ch Power
ACLR)
Mask)
Spurious emissions (Spurious Emissions)
Measurement of time domain power (Time Domain Power)
Signal statistics: amplitude probability distribution (APD) and
cumulative complementary distribution function (CCDF)
Third-order intercept point (TOI)
AM modulation depth (AM Mod Depth)
MEAS CONFIGUsed to define measurement configuration.
LINESConfigures display lines and limit lines.
INPUT/OUTPUTDisplays softkeys for input/output functions.
Measurement start functions
RUN SINGLEStarts a single new measurement (Single Sweep Mode).
RUN CONTStarts a continuous measurement (Continuous Sweep Mode).
Function execution (in navigation controls area)
44Getting Started 1312.9420.02 ─ 15
Page 45
R&S®FSW
Function keyAssigned functions
UNDOReverts last operation
REDORepeats previously reverted operation.
Instrument Tour
Front Panel View
3.1.10Keypad
The keypad is used to enter alphanumeric parameters, including the corresponding units (see also chapter 5.4, "Entering Data", on page 90). It contains the following keys:
Table 3-4: Keys on the keypad
Type of keyDescription
Alphanumeric keysEnter numbers and (special) characters in edit dialog boxes.
Decimal pointInserts a decimal point "." at the cursor position.
Sign keyChanges the sign of a numeric parameter. In the case of an
alphanumeric parameter, inserts a "-" at the cursor position.
Unit keys (GHz/-dBm MHz/
dBm, kHz/dB and Hz/dB)
ESC keyCloses all kinds of dialog boxes, if the edit mode is not active.
Adds the selected unit to the entered numeric value and complete the entry.
In the case of level entries (e.g. in dB) or dimensionless values,
all units have the value "1" as multiplying factor. Thus, they have
the same function as an ENTER key.
Quits the edit mode, if the edit mode is active. In dialog boxes
that contain a "Cancel" button it activates that button.
For "Edit" dialog boxes the following mechanism is used:
If data entry has been started, it retains the original value
●
and closes the dialog box.
If data entry has not been started or has been completed, it
●
closes the dialog box.
BACKSPACE keyIf an alphanumeric entry has already been started, this key dele-
tes the character to the left of the cursor.
ENTER key
Concludes the entry of dimensionless entries. The new value
●
is accepted.
With other entries, this key can be used instead of the
●
"Hz/dB" unit key.
In a dialog box, selects the default or focused element.
●
45Getting Started 1312.9420.02 ─ 15
Page 46
R&S®FSW
Instrument Tour
Front Panel View
3.1.11Navigation Controls
The navigation controls include a rotary knob, navigation keys, and UNDO /
REDO keys. They allow you to navigate within the display or within dialog boxes.
3.1.11.1 Rotary Knob
The rotary knob has several functions:
For numeric entries: increments (clockwise direction) or decrements (counter-
●
clockwise direction) the instrument parameter at a defined step width
In lists: toggles between entries
●
For markers, limit lines, and other graphical elements on the screen: moves
●
their position
For active scroll bars: moves the scroll bar vertically
●
For dialog boxes: Acts like the ENTER key when pressed
●
3.1.11.2 Navigation Keys
The navigation keys can be used alternatively to the rotary knob to navigate
through dialog boxes, diagrams or tables.
Arrow Up/Arrow Down Keys
The <arrow up> or <arrow down> keys do the following:
For numeric entries: increments (Arrow Up) or decrements (Arrow Down) the
●
instrument parameter at a defined step width
In a list: scrolls forward and backward through the list entries
●
In a table: moves the selection bar vertically
●
In windows or dialog boxes with a vertical scroll bar: moves the scroll bar
●
46Getting Started 1312.9420.02 ─ 15
Page 47
R&S®FSW
Instrument Tour
Front Panel View
Arrow Left/Arrow Right Keys
The <arrow left> or <arrow right> keys do the following:
In an alphanumeric edit dialog box, move the cursor.
●
In a list, scroll forward and backward through the list entries.
●
In a table, move the selection bar horizontally.
●
In windows or dialog boxes with horizontal scroll bar, move the scroll bar.
●
3.1.12UNDO/REDO keys
The UNDO key reverts the previously performed action, i.e. the status before
●
the previous action is retrieved.
The undo function is useful, for example, if you are performing a zero span
measurement with several markers and a limit line defined and accidentally
select a different measurement. In this case, many settings would be lost.
However, if you press UNDO immediately afterwards, the previous status is
retrieved, i.e. the zero span measurement and all settings.
The REDO key repeats the previously reverted action, i.e. the most recently
●
performed action is repeated.
The UNDO function is not available after a PRESET or "RECALL" operation. When these functions are used, the history of previous actions is
deleted.
The UNDO/REDO functions are not available for some applications; see
Release Notes for details.
The R&S FSW Analog Baseband Interface option (R&S FSW-B71) provides four
BASEBAND INPUT BNC connectors on the front panel of the instrument for analog I and Q signals.
47Getting Started 1312.9420.02 ─ 15
Page 48
R&S®FSW
The upper BNC connectors BASEBAND INPUT I and BASEBAND INPUT Q are
used to input single-ended signals, the positive signal input for differential signals,
as well as input from active Rohde & Schwarz probes (R&S RT-ZSxx and differential probes RT-ZDxx).
The lower BNC connectors Ī and Ǭ are used to input the negative signal for differential signals.
Instrument Tour
Front Panel View
For complex signal input (I+jQ), always use two identical cables for the I
and Q connectors (same length, same type, same manufacturer). Otherwise, time delay or gain imbalance may occur between the different cables,
which cannot be calibrated.
All connectors have a fixed impedance of 50 Ω and may receive a maximum input
level of 4 Vpp each.
Risk of instrument damage
Do not overload the BASEBAND INPUT connectors. An input voltage of 4 V
must never be exceeded. Noncompliance will destroy the Analog Baseband
Interface components.
Input via the Analog Baseband Interface can be enabled in the I/Q Analyzer, the
Analog Demodulation application, or in one of the optional applications that process I/Q data (where available).
For more information on the Analog Baseband Interface (R&S FSW-B71) see the
R&S FSW I/Q Analyzer and I/Q Input User Manual.
48Getting Started 1312.9420.02 ─ 15
Page 49
R&S®FSW
Instrument Tour
Front Panel View
3.1.14TRIGGER INPUT / OUTPUT
The female TRIGGER INPUT connector for external trigger or gate input is used
to control the measurement by means of an external signal. The voltage levels
can range from 0.5 to 3.5 V. The default value is 1.4 V. The typical input impedance is 10 kΩ.
The female BNC TRIGGER INPUT / OUTPUT connector can be used to receive
a second external signal or to provide a signal to another device. The signal is
TTL compatible (0 V / 5 V). You can control the connector usage in the "Trigger"
settings (TRIG key).
The rear panel provides a third TRIGGER INPUT / OUTPUT connector, see
chapter 3.2.10, "TRIGGER 3 INPUT/ OUTPUT", on page 54.
3.1.15EXT MIXER Connector (optional)
External mixers can be connected at the EXT MIXER LO OUT/IF IN and IF IN
female connectors to increase the available frequency range. This connector is
optional and only available with R&S FSW-B21.
If no external mixers are connected to the R&S FSW, cover the two front connectors LO OUT / IF IN and IF IN with the supplied SMA caps.
3.1.16
RF INPUT 50 Ω
A device under test (DUT) can be connected to the R&S FSW to provide RF input
which is then analyzed. The DUT is connected to the instrument's RF INPUT via
a cable equipped with an appropriate connector.
Risk of instrument damage
Do not overload the input. For maximum allowed values, see the data
sheet. For AC-coupling, a DC input voltage of 50 V must never be exceeded. For DC-coupling, DC voltage must not be applied at the input. In both
cases, noncompliance will destroy the input mixers.
49Getting Started 1312.9420.02 ─ 15
Page 50
R&S®FSW
Instrument Tour
Rear Panel View
3.2Rear Panel View
This figure shows the rear panel view of the R&S FSW. The individual elements
are described in more detail in the subsequent sections.
1
2
36
8
16
17
14
4
5
7
9
10
11
1213
15
18
Fig. 3-3: Rear panel view
1 = Removable system hard drive
2 = AC Power Supply Connection and Main Power Switch
3 = DISPLAY PORT for external display
4 = DVI connector for external display
5 = LAN connector
6 = USB (DEVICE) connectors
7 =
Bandwidth Extension 160 MHz/ 320 MHz/ 500 MHz with IF WIDE OUTPUT connector
(option B160/ B320/B500)
8 = IF/VIDEO/DEMOD connector
9 = DIGITAL BASEBAND INPUT/OUTPUT connectors (option B17)
10 = TRIGGER 3 INPUT/OUTPUT connector
11 = SYNC TRIGGER OUTPUT/INPUT
12 = AUX PORT
13 = GPIB interface
14 = Analog baseband interface (option B71)
15 = External generator control (option B10)
16 =
IF OUT (2GHz) connector, for models R&S FSW43/50/67 only
17 = OCXO external reference (option B4)
18 = REF INPUT/OUTPUT connectors
50Getting Started 1312.9420.02 ─ 15
Page 51
R&S®FSW
Instrument Tour
Rear Panel View
3.2.1Removable System Hard Drive
The removable system hard drive contains all measurement data from the
R&S FSW, allowing you to store the data securely in an external location.
3.2.2AC Power Supply Connection and Main Power Switch
An AC power supply connector and main power switch are located in a unit on
the rear panel of the instrument.
Main power switch function:
Position 1: The instrument is in operation.
Position O: The entire instrument is disconnected from the AC power supply.
For details refer to chapter 2.1.4, "Connecting the AC Power", on page 15.
3.2.3DISPLAY PORT and DVI
You can connect an external monitor or other display device to the R&S FSW to
provide an enlarged display. Two different types of connectors are provided for
this purpose:
DISPLAY PORT
●
DVI (Digital visual interface)
●
For details see chapter 2.4, "Connecting an External Monitor", on page 23.
3.2.4LAN
The LAN interface can be used to connect the R&S FSW to a local network for
remote control, printouts or data transfer. The assignment of the RJ-45 connector
supports twisted-pair category 5 UTP/STP cables in a star configuration (UTP
stands for unshielded twisted pair, and STP for shielded twisted pair).
For details see chapter 2.5, "Setting Up a Network (LAN) Connection",
on page 25.
51Getting Started 1312.9420.02 ─ 15
Page 52
R&S®FSW
Instrument Tour
Rear Panel View
3.2.5USB
The rear panel provides four additional female USB (USB-A) connectors to connect devices like a keyboard, a mouse or a memory stick (see also chapter 3.1.5,
"USB", on page 42).
Furthermore, a male USB DEVICE connector (USB-B) is provided, for example
for a printer.
All USB connectors support standard 2.0.
3.2.6
Bandwidth Extension 160 MHz/ 320 MHz/ 500 MHz with IF
WIDE OUTPUT Connector (R&S FSW-B160/-B320/-B500/U160/-U320)
The signal analysis bandwidth of the R&S FSW can be extended by a hardware
option (R&S FSW-B160/-B320/-B500 or R&S FSW-U160/-U320). The bandwidth
extension allows for a linear bandwidth up to a maximum of 160 MHz (with option
B160/U160), 320 MHz (with option B320/U320), or 500 MHz (with option B500),
with an output sample rate of up to 10 GHz. While the extension can be activated
or deactivated manually in the R&S FSW base unit (I/Q Analyzer application), it is
activated automatically in some applications that also support I/Q data analysis.
See the application-specific documentation for details.
The bandwidth extension option R&S FSW-B320/-U320 requires a reference board revision 3.14 or higher.
The bandwidth extension option R&S FSW-B500 requires a reference
board (1312.8075.06) revision 4.06 or higher and a motherboard
1313.4180.02 or 1313.7698.02.
Together with the bandwidth extension an additional IF output connector is provided (IF WIDE OUTPUT). As opposed to the default IF/VIDEO/DEMOD OUTPUT
connector, the IF output frequency of the optional connector cannot be defined
manually, but is determined automatically depending on the center frequency. For
details on the used frequencies see the data sheet. The IF WIDE OUTPUT connector is used automatically when the bandwidth extension is activated (i.e. for
bandwidths > 80 MHz).
52Getting Started 1312.9420.02 ─ 15
Page 53
R&S®FSW
Instrument Tour
Rear Panel View
3.2.7IF/VIDEO/DEMOD OUTPUT
The female BNC connector can be used for various outputs:
Intermediate frequency (IF) output of approximately 20 MHz
●
Video output (1V)
●
Which output is provided is defined in the software (INPUT/OUTPUT key).
For details see the User Manual.
3.2.8
IF OUTPUT (2 GHz)
The female SMA connector is only available for instrument models
R&S FSW43/50/67 and can be used to provide intermediate frequency (IF) output
of approximately 2 GHz at a frequency of 2 GHz.
Output is activated in the software (INPUT/OUTPUT key).
For details see the User Manual.
3.2.9
Digital Baseband Interface (R&S FSW-B17) and R&S EXIQ-BOX
The R&S FSW Digital Baseband Interface option (R&S FSW-B17) provides an
online digital I/Q data interface on the rear panel of the instrument for input and
output. The digital input and output can be enabled in the base unit or in one of
the applications (where available).
Optionally, an R&S EX-IQ-BOX can be connected to the Digital Baseband Interface to convert signal properties and the transmission protocol of the R&S FSW
into user-defined or standardized signal formats and vice versa.
Since the Digital I/Q input and the Analog Baseband input use the same
digital signal path, both cannot be used simultaneously. When one is activated, established connections for the other are disconnected. When the second input is deactivated, connections to the first are re-established. This
may cause a short delay in data transfer after switching the input source.
53Getting Started 1312.9420.02 ─ 15
Page 54
R&S®FSW
Instrument Tour
Rear Panel View
For more information on the Digital Baseband Interface (R&S FSW-B17) and typical applications see the R&S FSW I/Q Analyzer User Manual.
3.2.10TRIGGER 3 INPUT/ OUTPUT
The additional female BNC TRIGGER INPUT / OUTPUT connector can be used
to receive a third external signal or to provide a signal to another device. The signal is TTL compatible (0 V / 5 V). You can control the connector usage in the
"Trigger" settings (TRIG key).
3.2.11SYNC TRIGGER OUTPUT/INPUT
The SYNC TRIGGER OUTPUT/INPUT connectors allow you to synchronize several devices (e.g. two R&S FSWs) with respect to the trigger signal, but also the
reference frequency. A 100 MHz signal can be output as a trigger or reference
signal to another device, and an external trigger or reference signal can be
received at the input connector by the R&S FSW.
3.2.12AUX PORT
The 9 pole SUB-D male connector provides control signals for controlling external
devices. The voltage levels are of the TTL type (max. 5 V).
PinSignalDescription
1+5 V / max. 250 mASupply voltage for external circuits
2 to 7I/OControl lines for user ports (see
User manual)
8GNDGround
9READY FOR TRIGGERSignal indicating that the instru-
ment is ready to receive a trigger
signal (Low active = 0 V)
54Getting Started 1312.9420.02 ─ 15
Page 55
R&S®FSW
Instrument Tour
Rear Panel View
Short-circuit hazard
Always observe the designated pin assignment. A short-circuit may damage
the instrument.
3.2.13GPIB Interface
The GPIB interface is in compliance with IEEE488 and SCPI. A computer for
remote control can be connected via this interface. To set up the connection, a
shielded cable is recommended. For more details refer to "Setting Up Remote
Control" in the User Manual.
3.2.14
External Generator Control Option (R&S FSW-B10)
The external generator control option provides an additional GPIB connector and
AUX port.
The GPIB connector can be used to connect the external generator to the
R&S FSW.
The AUX port is required for TTL synchronization, if supported by the generator. It
is identical to the standard AUX port (see chapter 3.2.12, "AUX PORT",
on page 54).
For details on connecting an external generator see the "External Generator Control" section of the R&S FSW User Manual.
3.2.15
OCXO Option (R&S FSW-B4)
This option generates a 10 MHz reference signal with a very precise frequency. If
installed, and if no external signal is used, this signal is used as an internal refer-
55Getting Started 1312.9420.02 ─ 15
Page 56
R&S®FSW
Instrument Tour
Rear Panel View
ence. It can also be used to synchronize other connected devices via the REF
OUTPUT 10 MHz connector.
Warm-up time for OCXO
When the instrument is switched on, the OCXO requires an extended
warm-up time (see data sheet). No warm-up time is required when starting
from standby mode.
3.2.16REF INPUT / REF OUTPUT
The REF INPUT connectors are used to provide an external reference signal to
the R&S FSW.
The REF OUTPUT connectors can be used to provide an external reference signal (or the optional OCXO reference signal) from the R&S FSW to other devices
that are connected to this instrument.
Various connectors are provided for different reference signals:
ConnectorReference signalUsage
REF INPUT1...20 MHz
0...10 dBm
REF OUTPUT1...20 MHz
0...10 dBm
REF OUTPUT10 MHz
10 dBm
To provide an external reference signal on the
R&S FSW.
To provide the same external reference signal
received by the REF INPUT 1...20 MHZ connector to another device, when available.
To provide the internal reference signal from the
R&S FSW to another device continuously.
Also used to provide OCXO reference signal to
another device.
REF INPUT100 MHz
0...10 dBm
REF OUTPUT100 MHz
6 dBm
REF OUTPUT640 MHz
16 dBm
To provide an external reference signal on the
R&S FSW.
To provide a 100 MHz reference signal from the
R&S FSW to another device.
To provide a 640 MHz reference signal from the
R&S FSW to another device.
56Getting Started 1312.9420.02 ─ 15
Page 57
R&S®FSW
SYNC TRIGGER
The SYNC TRIGGER connector can also be used to synchronize the reference frequency on several devices (see chapter 3.2.11, "SYNC TRIGGER
OUTPUT/INPUT", on page 54).
Instrument Tour
Rear Panel View
57Getting Started 1312.9420.02 ─ 15
Page 58
R&S®FSW
Trying Out the Instrument
Measuring a Basic Signal
4Trying Out the Instrument
This chapter introduces the most important functions and settings of the
R&S FSW step by step. The complete description of the functionality and its
usage is given in the R&S FSW User Manual. Basic instrument operation is
described in chapter 5, "Operating the Instrument", on page 78.
Prerequisites
The instrument is set up, connected to the mains system, and started up as
●
described in chapter 2, "Preparing for Use", on page 9.
For these first measurements, you use the internal calibration signal, so you do
not need any additional signal source or instruments. Try out the following:
●Measuring a Basic Signal................................................................................58
●Displaying a Spectrogram............................................................................... 61
●Printing and Saving Results............................................................................ 77
4.1Measuring a Basic Signal
We will start out by measuring a simple sinus wave, using the internal calibration
signal as the input.
To display the internal 64 MHz calibration signal
1. Press the PRESET key to start out in a defined instrument configuration.
2. Press the SETUP key on the front panel.
3. Tap the "Service" softkey.
4. Tap the "Calibration Signal" tab.
58Getting Started 1312.9420.02 ─ 15
Page 59
R&S®FSW
Trying Out the Instrument
Measuring a Basic Signal
5. Tap the "Calibration Frequency RF" option. Leave the frequency at the default
64 MHz, with a narrowband spectrum.
The calibration signal is now sent to the RF input of the R&S FSW. By default,
a continuous frequency sweep is performed, so that the spectrum of the calibration signal is now displayed in the standard level versus frequency diagram.
Fig. 4-1: Calibration signal as RF input
Instrument warmup time
Note that the instrument requires an initial warmup time after switching it
on. A message in the status bar ("Instrument warming up...") indicates that
the operating temperature has not yet been reached. Wait till this message
is no longer displayed before you start a measurement.
To optimize the display
In order to optimize the display for the calibration signal, we will adjust the main
measurement settings.
1. Set the center frequency to the calibration frequency:
59Getting Started 1312.9420.02 ─ 15
Page 60
R&S®FSW
Trying Out the Instrument
Measuring a Basic Signal
a) Tap the "Overview" softkey to display the configuration "Overview".
b) Tap the "Frequency" button.
c) In the "Center" field, enter 64 on the number pad on the front panel.
d) Press the "MHz" key next to the number pad.
2.
Reduce the span to 20 MHz:
a)
In the "Span" field of the "Frequency" dialog box, enter 20
MHz.
b) Close the "Frequency" dialog box.
3.
Set the reference level to -25 dBm:
a) In the configuration "Overview", tap the "Amplitude" button.
b)
In the "Value" field of the "Amplitude" dialog box, enter -25 dBm.
The display of the calibration signal is now improved. The maximum at the
center frequency (=calibration frequency) of 64 MHz becomes visible.
Fig. 4-2: Calibration signal with optimized display settings
60Getting Started 1312.9420.02 ─ 15
Page 61
R&S®FSW
Trying Out the Instrument
Displaying a Spectrogram
4.2Displaying a Spectrogram
In addition to the standard "level versus frequency" spectrum display, the
R&S FSW also provides a spectrogram display of the measured data. A spectrogram shows how the spectral density of a signal varies over time. The x-axis
shows the frequency, the y-axis shows the time. A third dimension, the power
level, is indicated by different colors. Thus you can see how the strength of the
signal varies over time for different frequencies.
1. Tap the "Overview" softkey to display the general configuration dialog box.
2. Tap the "Display Config" button.
The SmartGrid mode is activated, and the evaluation bar with the available
evaluation methods is displayed.
3.
Drag the "Spectrogram" icon from the evaluation bar to the diagram area. The
blue area indicates that the new diagram would replace the previous spectrum
display. Since we do not want to replace the spectrum, drag the icon to the
lower half of the display to add an additional window instead.
Fig. 4-3: Adding a Spectrogram to the display
Drop the icon.
61Getting Started 1312.9420.02 ─ 15
Page 62
R&S®FSW
Trying Out the Instrument
Activating Additional Measurement Channels
4. Close the SmartGrid mode by tapping the "Close" icon at the top right corner
of the toolbar.
You see the spectrogram compared to the standard spectrum display. Since
the calibration signal does not change over time, the color of the frequency
levels does not change over time, i.e. vertically. The legend at the top of the
spectrogram window describes the power levels the colors represent.
Fig. 4-4: Spectrogram of the calibration signal
4.3Activating Additional Measurement Channels
The R&S FSW features multiple measurement channels, i.e. you can define several measurement configurations in parallel and then switch between the channels automatically to perform the measurements sequentially. We will demonstrate this feature by activating additional measurement channels for a different
frequency range, a zero span measurement, and an I/Q analysis.
62Getting Started 1312.9420.02 ─ 15
Page 63
R&S®FSW
Trying Out the Instrument
Activating Additional Measurement Channels
To activate additional measurement channels
1. Press the MODE key on the front panel.
2. On the "New Channel" tab of the "Signal + Spectrum Mode" dialog box, tap
the "Spectrum" button.
Fig. 4-5: Adding a new measurement channel
3. Change the frequency range for this spectrum display:
In the "Frequency" dialog box, set the center frequency to 500 MHz and the
span to 1 GHz.
63Getting Started 1312.9420.02 ─ 15
Page 64
R&S®FSW
Trying Out the Instrument
Activating Additional Measurement Channels
Fig. 4-6: Frequency spectrum of the calibration signal with a larger span
4. Repeat the previous steps to activate a third Spectrum window.
Change the frequency range for this spectrum display:
In the "Frequency" dialog box, set the center frequency to 64 MHz and tap
"Zero Span".
As the calibration signal does not vary over time, the level versus time diagram displays a straight line.
64Getting Started 1312.9420.02 ─ 15
Page 65
R&S®FSW
Trying Out the Instrument
Activating Additional Measurement Channels
Fig. 4-7: Time domain display of the calibration signal
5. Create a new channel for I/Q analysis:
a) Press the MODE key.
b) Tap the "IQ Analyzer" button to activate a channel for the I/Q Analyzer
application.
c) Tap the "Display Config" softkey to activate the SmartGrid mode.
65Getting Started 1312.9420.02 ─ 15
Page 66
R&S®FSW
Trying Out the Instrument
Activating Additional Measurement Channels
d) Drag the "Real/Imag (I/Q)" icon from the evaluation bar to the SmartGrid.
Fig. 4-8: Inserting a Real/Imag diagram for I/Q analysis
e) Close the SmartGrid mode.
The "IQ Analyzer" channel displays the real and imaginary signal parts in sep-
arate windows.
To display the MultiView tab
An overview of all active channels is provided in the "MultiView" tab. This tab is
always displayed and cannot be closed.
► Tap the "MultiView" tab.
66Getting Started 1312.9420.02 ─ 15
Page 67
R&S®FSW
Trying Out the Instrument
Performing Sequential Measurements
Fig. 4-9: The "MultiView" tab
4.4Performing Sequential Measurements
Although only one measurement can be performed at any one time, the measurements configured in the active channels can be performed sequentially, that
means: one after the other, automatically, either once or continuously.
1.
Tap the "Sequencer" icon in the toolbar.
2. Toggle the "Sequencer" softkey in the "Sequencer" menu to "On".
A continuous sequence is started, i.e. each channel measurement is performed one after the other until the Sequencer is stopped.
67Getting Started 1312.9420.02 ─ 15
Page 68
R&S®FSW
Trying Out the Instrument
Setting and Moving a Marker
Fig. 4-10: "MultiView" tab with active Sequencer
In figure 4-10, the "Spectrum 2" measurement is currently active (indicated by
the "channel active" icon in the tab label).
3. Stop the Sequencer by tapping the "Sequencer" softkey again.
4.5Setting and Moving a Marker
Markers are useful to determine the position of particular effects in the trace. The
most common use is to determine a peak, which is the default setting when you
activate a marker. We will set a marker on the peak in our first Spectrum measurement.
1. In the "MultiView" tab, double-tap the "Spectrum" window (frequency sweep
with spectrogram display) to return to the "Spectrum" channel.
2. Tap the spectrum display to set the focus on that window.
3.
68Getting Started 1312.9420.02 ─ 15
Page 69
R&S®FSW
Press the "Split/Maximize" key on the front panel to maximize the spectrum
window, as we currently do not need the spectrogram display.
4. Press the "RUN SINGLE" key on the front panel to perform a single sweep so
we have a fixed trace to set a marker on.
5. Press the MKR key on the front panel to display the "Marker" menu.
Marker 1 is activated and automatically set to the maximum of trace 1. The
marker position and value is indicated in the diagram area as M1[1].
Trying Out the Instrument
Setting and Moving a Marker
6. Now you can move the marker by tapping and dragging it to a different posi-
tion. The current position is indicated by a dotted blue line. Notice how the
position and value change in the marker area of the diagram.
69Getting Started 1312.9420.02 ─ 15
Page 70
R&S®FSW
Trying Out the Instrument
Displaying a Marker Peak List
4.6Displaying a Marker Peak List
The marker peak list determines the frequencies and levels of peaks in the spectrum automatically. We will display a marker peak list for the Spectrum 2 channel.
1. Tap the "Spectrum 2" tab.
2. Press the "RUN SINGLE" key on the front panel to perform a single sweep for
which we will determine the peaks.
3.
Tap the "SmartGrid" icon in the toolbar to activate SmartGrid mode.
4. Drag the "Marker Peak List" icon from the evaluation bar to the lower half of
the display to add a new window for the peak list.
5. Close the SmartGrid mode.
6. To obtain a more conclusive peak list that does not contain noise peaks, for
example, define a threshold that is higher than the noise floor:
a) Press the MKR key on the front panel.
70Getting Started 1312.9420.02 ─ 15
Page 71
R&S®FSW
Trying Out the Instrument
Zooming into the Display
b) Tap the "Marker Config" softkey in the "Marker" menu.
c) Tap the "Search" tab in the "Marker" dialog box.
d)
In the "Threshold" field, enter -68 dBm.
e) Tap the "State" box for "Threshold" to activate its use.
Only peaks that are larger than -68 dBm will be included in the peak list.
The marker peak list displays the determined peaks that are above the
defined threshold.
Fig. 4-11: Marker Peak List
4.7Zooming into the Display
In order to analyze the areas around the peak levels in more detail, we will zoom
into the top 3 peaks.
1.
Tap the "Multiple Zoom" icon in the toolbar.
The icon is highlighted orange to indicate that multiple zoom mode is active.
71Getting Started 1312.9420.02 ─ 15
Page 72
R&S®FSW
Trying Out the Instrument
Zooming into the Display
2. Tap the diagram near the first peak and drag your finger to the opposite cor-
ner of the zoom area. A white rectangle is displayed from the point where you
tapped to the current position.
Fig. 4-12: Defining the zoom area
When you remove your finger, the zoom area is enlarged in a second
(sub-)window.
Fig. 4-13: Zoomed display around a peak
Zoom mode is automatically disabled, and the default selection mode is
restored, indicated by the highlighted "Selection mode" icon.
72Getting Started 1312.9420.02 ─ 15
Page 73
R&S®FSW
Trying Out the Instrument
Zooming into the Display
3. In figure 4-13 the enlarged peak is represented by a very thick trace. This is
due to the insufficient number of sweep points. The missing sweep points for
the zoomed display are interpolated, which provides poor results. To optimize
the results, we will increase the number of sweep points from the default 1001
to 32001.
a) Press the SWEEP key on the front panel.
b) Tap the "Sweep Config" softkey in the "Sweep" menu.
c) In the "Sweep Points" field, enter 32001.
d) Press the RUN SINGLE key on the front panel to perform a new sweep
with the increased number of sweep points.
Fig. 4-14: Zoomed peak with increased number of sweep points
Note that the trace becomes much more precise.
4.
Tap the "Multiple Zoom" icon in the toolbar again and define a zoom area
around markers M4, M5 and M6.
73Getting Started 1312.9420.02 ─ 15
Page 74
R&S®FSW
Trying Out the Instrument
Zooming into the Display
Fig. 4-15: Multiple zoom windows
5. Tap the "Multiple Zoom" icon in the toolbar again and define a zoom area
around marker M8.
6. To increase the size of the third zoom window, drag the "splitter" between the
windows to the left or right or up or down.
74Getting Started 1312.9420.02 ─ 15
Page 75
R&S®FSW
Trying Out the Instrument
Saving Settings
Fig. 4-16: Enlarged zoom window
4.8Saving Settings
In order to restore the results of our measurements at a later time, we will store
the instrument settings to a file.
To save the instrument settings to a file
1.
Tap the "Save" icon in the toolbar.
2.
Press the keyboard key on the front panel to display the online keyboard, as
you will have to enter text in the next step.
3. In the "Save" dialog box, tap the "File Name" field and enter MyMultiView-
Setup using the keyboard.
75Getting Started 1312.9420.02 ─ 15
Page 76
R&S®FSW
Trying Out the Instrument
Saving Settings
Keep the default "File Type" setting "Instrument with all Channels" to store the
configuration of all channels.
Fig. 4-17: Saving the instrument settings to a file
4. Tap the "Save" button.
The file MyMultiViewSetup.dfl is stored in the default directory C:/R_S/instr/user.
To load stored instrument settings
You can restore the settings to the instrument at any time using the settings file.
1. Press the PRESET button to restore the default instrument settings so you
can check that the stored user settings are actually restored afterwards.
2.
Tap the "Load" icon in the toolbar.
3. In the "Load" dialog box, select the MyMultiViewSetup.dfl file in the
default directory C:/R_S/instr/user.
76Getting Started 1312.9420.02 ─ 15
Page 77
R&S®FSW
Trying Out the Instrument
Printing and Saving Results
4. Tap the "Load" button.
All instrument settings are restored and the display should resemble fig-
ure 4-16, which shows the instrument display right before the settings were
stored.
4.9Printing and Saving Results
Finally, after a successful measurement, we will document our results. First we
will export the numeric trace data, then we will create a screenshot of the graphical display.
To export the trace data
1. Press the TRACE key on the front panel.
2. Tap the "Trace Config" softkey.
3. Tap the "Trace Export" tab.
4. Tap the "Export Trace to ASCII File" button.
5. Enter the file name MyMultiViewResults.
The trace data is stored to MyMultiViewResults.DAT
To create a screenshot of the display
1.
Tap the "Print immediately" icon in the toolbar.
A screenshot of the current display is created. Note that the colors on the
screen are inverted in the screenshot to improve printout results.
2. In the "Save Hardcopy as Portable Network Graphics (PNG)" dialog box,
enter a file name, e.g. MyMultiViewDisplay.
The screenshot is stored to MyMultiViewDisplay.png.
77Getting Started 1312.9420.02 ─ 15
Page 78
R&S®FSW
Operating the Instrument
Understanding the Display Information
5Operating the Instrument
This chapter provides an overview on how to work with the R&S FSW. It
describes what kind of information is displayed in the diagram area, how to operate the R&S FSW via the front panel keys and other interaction methods, and
how to use the Online Help.
●Understanding the Display Information........................................................... 78
●Accessing the Functionality.............................................................................89
●Changing the Focus........................................................................................ 89
The following figure shows a measurement diagram in Spectrum mode. All different information areas are labeled. They are explained in more detail in the following sections.
78Getting Started 1312.9420.02 ─ 15
Page 79
R&S®FSW
Operating the Instrument
Understanding the Display Information
1
23
4
5
6
1= Channel bar for firmware and measurement settings
2+3 = Window title bar with diagram-specific (trace) information
4= Diagram area with marker information
5= Diagram footer with diagram-specific information, depending on measurement application
6= Instrument status bar with error messages, progress bar and date/time display
Hiding elements in the display
You can hide some of the elements in the display, for example the status
bar or channel bar, in order to enlarge the display area for the measurement results ("Setup > Display > Displayed Items").
Using the R&S FSW you can handle several different measurement tasks (channels) at the same time (although they can only be performed asynchronously).
For each channel, a separate tab is displayed on the screen. In order to switch
from one channel display to another, simply select the corresponding tab.
Alternatively, if many tabs are displayed, select the tab selection list icon at the
right end of the channel bar and select the channel you want to switch to from the
list.
MultiView tab
An additional tab labelled "MultiView" provides an overview of all active channels
at a glance. In the "MultiView" tab, each individual window contains its own channel bar with an additional button. Tap this button to switch to the corresponding
channel display quickly.
80Getting Started 1312.9420.02 ─ 15
Page 81
R&S®FSW
Operating the Instrument
Understanding the Display Information
Icons in the channel bar
The icon on the tab label indicates that the displayed trace no longer matches
the current instrument settings. This may be the case, for example, if a trace is
frozen and the instrument settings are changed. As soon as a new measurement
is performed, the icon disappears.
The icon indicates that an error or warning is available for that measurement
channel. This is particularly useful if the MultiView tab is displayed.
An orange "IQ" (in MSRA mode only) indicates that the results displayed in the
MSRA application(s) no longer match the data captured by the MSRA Master.
The "IQ" disappears after the results in the application(s) are refreshed.
The icon indicates the currently active channel during an automatic measurement sequence (Sequencer functionality).
Channel-specific settings
Beneath the channel name, information on channel-specific settings for the measurement are displayed in the channel bar. Channel information varies depending on the active application.
In the Spectrum application, the R&S FSW shows the following settings:
Table 5-1: Channel settings displayed in the channel bar in the Spectrum application
Ref LevelReference level
m.+el.AttMechanical and electronic RF attenuation that has been set.
Ref OffsetReference level offset
SWTSweep time that has been set.
If the sweep time does not correspond to the value for automatic
coupling, a bullet is displayed in front of the field. The color of the
bullet turns red if the sweep time is set below the value for automatic coupling. In addition, the UNCAL flag is shown. In this case,
the sweep time must be increased.
Meas TimeMeasurement time, calculated from analysis bandwidth and num-
ber of samples (for statistics measurements)
RBWResolution bandwidth that has been set.
If the bandwidth does not correspond to the value for automatic
coupling, a green bullet appears in front of the field.
VBWVideo bandwidth that has been set.
If the bandwidth does not correspond to the value for automatic
coupling, a green bullet is displayed in front of the field.
A bullet next to the setting indicates that user-defined settings are used, not automatic settings. A green bullet indicates this setting is valid and the measurement
is correct. A red bullet indicates an invalid setting that does not provide useful
results.
Common settings
In addition to the channel-specific settings, the channel bar above the diagram
also displays information on instrument settings that affect the measurement
results even though this is not immediately apparent from the display of the measured values. This information is displayed in gray font and only when applicable
for the current measurement, as opposed to the channel-specific settings that are
always displayed.
The following types of information may be displayed, if applicable.
Table 5-2: Common settings displayed in the channel bar
SGLThe sweep is set to single sweep mode.
Sweep Count The current signal count for measurement tasks that involve a specific number
of subsequent sweeps
(see "Sweep Count" setting in "Sweep settings" in the User Manual)
82Getting Started 1312.9420.02 ─ 15
Page 83
R&S®FSW
TRGTrigger source
(for details see "Trigger settings" in the User Manual)
●BBP: Baseband power (with Digital Baseband Interface R&S FSW-B17
only)
EXT: External
●
●GP_0: General purpose bit (with Digital Baseband Interface R&SFSW-B17
only)
IFP: IF power (+trigger bandwidth)
●
PSE: Power sensor
●
RFP: RF power
●
SQL: Squelch
●
TIM: Time
●
VID: Video
●
Operating the Instrument
Understanding the Display Information
6dB/RRC/
CHN
PAThe preamplifier is activated.
YIG BypassThe YIG filter is deactivated.
GATThe frequency sweep is controlled via the TRIGGER INPUT connector.
TDFA transducer factor is activated.
75 ΩThe input impedance of the instrument is set to 75 Ω.
FRQ
DC/ACAn external DC or AC calibration signal is in use.
ExtMix
<band>
<NOR | APR>
Ext. Gen
Filter type for sweep bandwidth
(see "Bandwidth settings" in the User Manual)
A frequency offset ≠ 0 Hz is set.
An external mixer is activated for input (requires option R&S FSW-B21); the
used band is also indicated
An external generator is being controlled by the R&S FSW (requires option
R&S FSW-B10).
NOR: the measurements are normalized with the results of the external generator calibration
APR (approximation): the measurements are normalized with the results of the
external generator calibration; however, the measurement settings have been
changed since calibration
If neither label is displayed, no calibration has been performed yet or normalization is not active.
For details see the "External Generator Control" section in the R&S FSW User
Manual.
LVLA level offset is applied to the external generator signal (only if external genera-
tor control is active).
83Getting Started 1312.9420.02 ─ 15
Page 84
R&S®FSW
Operating the Instrument
Understanding the Display Information
5.1.2Window Title Bar
Each channel in the R&S FSW display may contain several windows. Each window can display either a graph or a table as a result of the channel measurement.
Which type of result evaluation is displayed in which window is defined in the display configuration (see chapter 5.5, "Displaying Results", on page 92). The window's title bar indicates which type of evaluation is displayed.
Trace Information in Window Title Bar
Information on the displayed traces is indicated in the window title bar.
Trace color
Trace no.
Color of trace display in diagram
Trace number (1 to 6)
Detector
Trace Mode
APAUTOPEAK detector
PkMAX PEAK detector
MiMIN PEAK detector
SaSAMPLE detector
AvAVERAGE detector
RmRMS detector
QPQUASIPEAK detector
ClrwCLEAR/WRITE
MaxMAX HOLD
MinMIN HOLD
AvgAVERAGE (Lin/Log/Pwr)
ViewVIEW
Selected detector:
Sweep mode:
Norm/NCor
Correction data is not used.
84Getting Started 1312.9420.02 ─ 15
Page 85
R&S®FSW
Operating the Instrument
Understanding the Display Information
5.1.3Marker Information
Marker information is provided either in the diagram grid or in a separate marker
table, depending on the configuration.
Marker information in diagram grid
Within the diagram grid, the x-axis and y-axis positions of the last 2 markers or
delta markers that were set are displayed, if available, as well as their index. The
value in the square brackets after the index indicates the trace to which the
marker is assigned. (Example: M2[1] defines marker 2 on trace 1.) For more than
2 markers, a separate marker table is displayed beneath the diagram by default.
Marker information in marker table
In addition to the marker information displayed within the diagram grid, a separate
marker table may be displayed beneath the diagram. This table provides the following information for all active markers:
TypeMarker type: N (normal), D (delta), T (temporary, internal), PWR
(power sensor)
RefReference (for delta markers)
TrcTrace to which the marker is assigned
X-valuex-value of the marker
Y-valuey-value of the marker
FuncActivated marker or measurement function
Func .ResultResult of the active marker or measurement function
The functions are indicated with the following abbreviations:
FXDFixed reference marker
PHNoisePhase noise measurement
CNTSignal count
TRKSignal tracking
NOIseNoise measurement
MDepthAM modulation depth
TOIThird order intercept measurement
85Getting Started 1312.9420.02 ─ 15
Page 86
R&S®FSW
Operating the Instrument
Understanding the Display Information
5.1.4Frequency and Span Information in Diagram Footer
The information in the diagram footer (beneath the diagram) depends on the current application.
For Spectrum mode, it contains the following information:
LabelInformation
CFCenter frequency
SpanFrequency span (frequency domain display)
ms/Time per division (time domain display)
PtsNumber of sweep points or (rounded) number of currently displayed points
in zoom mode
5.1.5Instrument and Status Information
Global instrument settings and functions (such as LXI configuration mode), the
instrument status and any irregularities are indicated in the status bar beneath the
diagram.
In the MultiView tab the status bar always displays the information for the currently selected measurement.
The following information is displayed:
Instrument status
The instrument is configured for operation with an external reference.
The optional Digital Baseband Interface (R&S FSW-B17) is being used for
digital input
For details on the Digital Baseband Interface (R&S FSW-B17) see the
R&S FSW I/Q Analyzer User Manual.
The optional Digital Baseband Interface (R&S FSW-B17) is being used to
provide digital output
For details on the Digital Baseband Interface (R&S FSW-B17) see the
R&S FSW I/Q Analyzer User Manual.
Progress
The progress of the current operation is displayed in the status bar.
86Getting Started 1312.9420.02 ─ 15
Page 87
R&S®FSW
Operating the Instrument
Understanding the Display Information
In the MultiView tab, the progress bar indicates the status of the currently
selected measurement, not the measurement currently being performed by
a Sequencer, for example.
Date and time
The date and time settings of the instrument are displayed in the status bar.
5.1.6Error Information
If errors or irregularities are detected, a keyword and an error message, if available, are displayed in the status bar.
Depending on the type of message, the status message is indicated in varying
colors.
Table 5-3: Status bar information - color coding
ColorTypeDescription
redFatalA serious error occurred in the application; regular operation is no
longer possible.
redErrorAn error occurred during a measurement, e.g. due to missing data
or wrong settings, so that the measurement cannot be completed
correctly.
orangeWarningAn irregular situation occurred during measurement, e.g. the set-
tings no longer match the displayed results, or the connection to an
external device was interrupted temporarily.
grayInformationInformation on the status of individual processing steps.
grayMessageAn event or state has occurred that may lead to an error during fur-
ther operation.
greenNo errorsNo messages displayed.
87Getting Started 1312.9420.02 ─ 15
Page 88
R&S®FSW
If any error information is available for a measurement channel, the icon
is displayed next to the channel name. This is particularly useful when the
MultiView tab is displayed, as the status bar in the MultiView tab always displays the information for the currently selected measurement only.
The following keywords are used:
Operating the Instrument
Understanding the Display Information
DATA ERR
FIFO OVLD
IF OVLDOverload of the IF signal path after the input mixer.
INPUT OVLDThe signal level at the RF input connector exceeds the maximum.
LOUNLError in the instrument's frequency processing hardware was detected.
NO REFInstrument was set to an external reference but no signal was detected on
For Digital Baseband Interface (R&S FSW-B17) only:
Error in digital I/Q input data
For details on the Digital Baseband Interface (R&S FSW-B17) see the
R&S FSW I/Q Analyzer User Manual.
For Digital Baseband Interface (R&S FSW-B17) only:
Input sample rate from connected instrument is too high
For details on the Digital Baseband Interface (R&S FSW-B17) see the
R&S FSW I/Q Analyzer User Manual.
Increase the reference level.
●
The RF input is disconnected from the input mixer to protect the device. In
order to re-enable measurement, decrease the level at the RF input connector and reconnect the RF input to the mixer input.
(See "RF Input Protection" in the R&S FSW User Manual).
the reference input.
OVEN
OVLDOverload of the input signal path after the input mixer (only when RF INPUT
PLL UNLOCK
RF OVLDOverload of the input mixer.
OCXO reference frequency (option R&S FSW-B4) has not yet reached its
operating temperature. The message usually disappears a few minutes
after power has been switched on.
path is NOT used, e.g. for input from the Digital Baseband Interface
(R&S FSW-B17) or the Analog Baseband Interface (R&S FSW-B71)).
Reduce the input level.
●
For Digital Baseband Interface (R&S FSW-B17) only:
Error in digital I/Q input data
For details on the Digital Baseband Interface (R&S FSW-B17) see the
R&S FSW I/Q Analyzer User Manual.
Increase the RF attenuation (for RF input).
●
Reduce the input level (for digital input)
●
88Getting Started 1312.9420.02 ─ 15
Page 89
R&S®FSW
Operating the Instrument
Accessing the Functionality
UNCALOne of the following conditions applies:
Correction data has been switched off.
●
No correction values are available, for example after a firmware update.
●
Record the correction data by performing a self alignment
●
(For details refer to chapter 2.1.6, "Performing a Self-Alignment and a
Selftest", on page 16).
WRONG_FWThe firmware version is out-of-date and does not support the currently
installed hardware. Until the firmware version is updated, this error message is displayed and self-alignment fails.
(For details refer to the R&S FSW User Manual).
5.2Accessing the Functionality
All tasks necessary to operate the instrument can be performed using this user
interface. Apart from instrument specific keys, all other keys that correspond to an
external keyboard (e.g. arrow keys, ENTER key) operate conform to Microsoft.
For most tasks, there are at least 2 alternative methods to perform them:
Using the touchscreen
●
Using other elements provided by the front panel, e.g. the keypad, rotary
●
knob, or arrow and position keys.
The measurement and instrument functions and settings can be accessed by
selecting one of the following elements:
System and function keys on the front panel of the instrument
●
Softkeys on the touchscreen
●
Context-sensitive menus for specific elements on the touchscreen
●
Icons on the tool bar in the touchscreen
●
Displayed setting on the touchscreen
●
5.3Changing the Focus
Any selected function is always performed on the currently focused element in the
display, e.g. a dialog field, diagram, or table row. Which element is focused is
indicated by a blue frame (diagram, window, table) or is otherwise highlighted
(softkey, marker etc.). Moving the focus is most easily done by tapping on the ele-
89Getting Started 1312.9420.02 ─ 15
Page 90
R&S®FSW
Operating the Instrument
Entering Data
ment on the touchscreen. Alternatively, use the "Tab" key on the on-screen keyboard or the rotary knob to move the focus from one element to the next on the
display.
To move the focus between any displayed diagrams or tables in a
window, press the "Change focus" key on the front panel. The
focus moves from the diagram to the first table to the next table etc. and then
back to the diagram, within the same window.
In fullscreen mode, where a single window is displayed in full size on the screen,
this key switches the focus (and the display) from one active window to the next.
5.4Entering Data
Data can be entered in dialog boxes using one of the following methods:
Using the touchscreen, via the online keyboard
●
Using other elements provided by the front panel, e.g. the keypad, rotary
●
knob, or navigation keys
The rotary knob acts like the ENTER key when it is pressed.
Using a connected external keyboard
●
Particularities in Windows dialog boxes
In some cases, e.g. if you want to install a printer, original Windows dialog
boxes are used. In these dialog boxes, the rotary knob and function keys do
not work. Use the touchscreen instead.
5.4.1Entering Numeric Parameters
If a field requires numeric input, the keypad provides only numbers.
1. Enter the parameter value using the keypad, or change the currently used
parameter value by using the rotary knob (small steps) or the UP or DOWN
keys (large steps).
2. After entering the numeric value via keypad, press the corresponding unit key.
The unit is added to the entry.
90Getting Started 1312.9420.02 ─ 15
Page 91
R&S®FSW
Operating the Instrument
Entering Data
3. If the parameter does not require a unit, confirm the entered value by pressing
the ENTER key or any of the unit keys.
The editing line is highlighted in order to confirm the entry.
5.4.2Entering Alphanumeric Parameters
If a field requires alphanumeric input, you can use the on-screen keyboard to
enter numbers and (special) characters (see chapter 3.1.1.4, "On-screen Key-
board", on page 40).
Alternatively, you can use the keypad. Every alphanumeric key represents several characters and one number. The decimal point key (.) represents special
characters, and the sign key (-) toggles between capital and small letters. For the
assignment refer to table 5-4.
Entering numbers and (special) characters via the keypad
1. Press the key once to enter the first possible value.
2. All characters available via this key are displayed.
3. To choose another value provided by this key, press the key again, until your
desired value is displayed.
4. With every key stroke the next possible value of this key is displayed. If all
possible values have been displayed, the series starts with the first value
again. For information on the series refer to table 5-4.
5. To change from capital to small letters and vice versa, press the sign key (-).
6. When you have chosen the desired value, wait for 2 seconds (to use the
same key again), or start the next entry by pressing another key.
Entering a blank
► Press the "Space" bar, or press the "0" key and wait 2 seconds.
Correcting an entry:
1. Using the arrow keys, move the cursor to the right of the entry you want to
delete.
2. Press the BACKSPACE key.
91Getting Started 1312.9420.02 ─ 15
Page 92
R&S®FSW
The entry to the left of the cursor is deleted.
3. Enter your correction.
Completing the entry
► Press the ENTER key or the rotary knob.
Aborting the entry
► Press the ESC key.
The dialog box is closed without changing the settings.
Table 5-4: Keys for alphanumeric parameters
Operating the Instrument
Displaying Results
Key name
(upper inscription)
77 µ Ω ° € ¥ $ ¢
8A B C 8 Ä ÆÅ Ç
9D E F 9 É
4G H I 4
5J K L 5
6M N O 6 Ň Ö
1P Q R S 1
2T U V 2 Ü
3W X Y Z 3
0<blank> 0 – @ + / \ < > = % &
.. * : _ , ; " ' ? ( ) #
–<toggles between capital and small letters>
Series of (special) characters and number provided
5.5Displaying Results
The R&S FSW provides several instrument applications for different analysis
tasks and different types of signals, e.g. 3G FDD, I/Q analysis or basic spectrum
analysis. For each application, a new measurement channel is created and displayed in a separate tab on the screen.
92Getting Started 1312.9420.02 ─ 15
Page 93
R&S®FSW
Operating the Instrument
Displaying Results
The results of a measurement channel can be evaluated in many different ways,
both graphically and numerically. For each evaluation method the results are displayed in a separate window in the tab.
The R&S FSW allows you to configure the display to suit your specific requirements and optimize analysis.
5.5.1Activating and Deactivating Channels
When you activate an application, a new measurement channel is created which
determines the measurement settings for that application. The same application
can be activated with different measurement settings by creating several channels for the same application. Whenever you switch channels, the corresponding
measurement settings are restored. Each channel is displayed in a separate tab
on the screen.
An additional tab ("MultiView") provides an overview of all currently active channels at once.
Only one measurement can be performed at any time, namely the one in the currently active channel. However, in order to perform the configured measurements
consecutively, a Sequencer function is provided.
93Getting Started 1312.9420.02 ─ 15
Page 94
R&S®FSW
Operating the Instrument
Displaying Results
To start a new channel
1. Select the MODE key.
2. In the "Mode" dialog box, select the required application on the "New Chan-
nel" tab.
A new tab is displayed for the new channel.
To change the application in an active channel
1. Select the tab of the channel you want to change.
2. Select the MODE key.
3. In the "Mode" dialog box, select the new application to be displayed on the
"Replace Current Channel" tab.
The selected application is displayed in the current channel.
To close a measurement channel
►
Select the "Close" icon on the tab of the measurement channel.
The tab is closed, any running measurements are aborted, and all results for
that channel are deleted.
5.5.2Laying out the Result Display with the SmartGrid
Measurement results can be evaluated in many different ways, for example
graphically, as summary tables, statistical evaluations etc. Each type of evaluation is displayed in a separate window in the channel tab. Up to 16 individual windows can be displayed per measurement channel (i.e. per tab). To arrange the
diagrams and tables on the screen, the Rohde & Schwarz SmartGrid function
helps you find the target position simply and quickly.
(For details on evaluation methods see the User Manual.)
Principally, the layout of the windows on the screen is based on an underlying
grid, the SmartGrid. However, the SmartGrid is dynamic and flexible, allowing for
many different layout possibilities. The SmartGrid functionality provides the following basic features:
Windows can be arranged in columns or in rows, or in a combination of both.
●
Windows can be arranged in up to four rows and four columns.
●
94Getting Started 1312.9420.02 ─ 15
Page 95
R&S®FSW
Operating the Instrument
Displaying Results
Windows are moved simply by dragging them to a new position on the screen,
●
possibly changing the layout of the other windows, as well.
All evaluation methods available for the currently selected measurement are
●
displayed as icons in the evaluation bar. If the evaluation bar contains more
icons than can be displayed at once on the screen, it can be scrolled vertically. The same evaluation method can be displayed in multiple windows
simultaneously.
New windows are added by dragging an evaluation icon from the evaluation
●
bar to the screen. The position of each new window depends on where you
drop the evaluation icon in relation to the existing windows.
All display configuration actions are only possible in SmartGrid mode. When
●
SmartGrid mode is activated, the evaluation bar replaces the current softkey
menu display. When the SmartGrid mode is deactivated again, the previous
softkey menu display is restored.
●Background Information: The SmartGrid Principle..........................................95
●How to Activate SmartGrid Mode....................................................................97
●How to Add a New Result Window................................................................. 98
●How to Close a Result Window.......................................................................98
●How to Arrange the Result Windows.............................................................. 98
5.5.2.1Background Information: The SmartGrid Principle
SmartGrid display
During any positioning action, the underlying SmartGrid is displayed. Different
colors and frames indicate the possible new positions. The position in the SmartGrid where you drop the window determines its position on the screen.
95Getting Started 1312.9420.02 ─ 15
Page 96
R&S®FSW
Operating the Instrument
Displaying Results
Fig. 5-1: Moving a window in SmartGrid mode
The brown area indicates the possible "drop area" for the window, i.e. the area in
which the window can be placed. A blue area indicates the (approximate) layout
of the window as it would be if the icon were dropped at the current position. The
frames indicate the possible destinations of the new window with respect to the
existing windows: above/below, right/left or replacement (as illustrated in fig-
ure 5-2). If an existing window would be replaced, the drop area is highlighted in a
darker color shade.
Positioning the window
The screen can be divided into up to four rows. Each row can be split into up to
four columns, where each row can have a different number of columns. However,
rows always span the entire width of the screen and may not be interrupted by a
column. A single row is available as the drop area for the window in the SmartGrid. The row can be split into columns, or a new row can be inserted above or
below the existing row (if the maximum of 4 has not yet been reached).
96Getting Started 1312.9420.02 ─ 15
Page 97
R&S®FSW
A
Operating the Instrument
Displaying Results
1
B
C
Fig. 5-2: SmartGrid window positions
1 = Insert row above or below the existing row
2 = Create a new column in the existing row
3 = Replace a window in the existing row
22233
1
SmartGrid functions
Once the evaluation icon has been dropped, icons in each window provide delete
and move functions.
The "Move" icon allows you to move the position of the window,
possibly changing the size and position of the other displayed win-
dows.
The "Delete" icon allows you to close the window, enlarging the
display of the remaining windows.
5.5.2.2How to Activate SmartGrid Mode
All display configuration actions are only possible in SmartGrid mode. In SmartGrid mode the evaluation bar replaces the current softkey menu display. When
the SmartGrid mode is deactivated again, the previous softkey menu display is
restored.
► To activate SmartGrid mode, do one of the following:
●
Select the "SmartGrid" icon from the toolbar.
●Select the "Display Config" button in the configuration "Overview".
●Select the "Display Config" softkey from the MEAS CONFIG menu.
The SmartGrid functions and the evaluation bar are displayed.
97Getting Started 1312.9420.02 ─ 15
Page 98
R&S®FSW
To close the SmartGrid mode and restore the previous softkey
menu select the "Close" icon in the right-hand corner of the toolbar, or press any key on the front panel.
Operating the Instrument
Displaying Results
5.5.2.3How to Add a New Result Window
Each type of evaluation is displayed in a separate window. Up to 16 individual
windows can be displayed per measurement channel (i.e. per tab).
1. Activate SmartGrid mode.
All evaluation methods available for the currently selected measurement are
displayed as icons in the evaluation bar.
2. Select the icon for the required evaluation method from the evaluation bar.
If the evaluation bar contains more icons than can be displayed at once on the
screen, it can be scrolled vertically. Touch the evaluation bar between the
icons and move it up or down until the required icon appears.
3. Drag the required icon from the evaluation bar to the SmartGrid, which is dis-
played in the diagram area, and drop it at the required position. (See chap-
ter 5.5.2.5, "How to Arrange the Result Windows", on page 98 for more
information on positioning the window).
5.5.2.4How to Close a Result Window
► To close a window, activate SmartGrid mode and select the "Delete" icon for
the window.
5.5.2.5How to Arrange the Result Windows
1. Select an icon from the evaluation bar or the "Move" icon for an existing eval-
uation window.
98Getting Started 1312.9420.02 ─ 15
Page 99
R&S®FSW
Operating the Instrument
Displaying Results
2. Drag the evaluation over the SmartGrid.
A blue area shows where the window will be placed.
3. Move the window until a suitable area is indicated in blue.
4. Drop the window in the target area.
The windows are rearranged to the selected layout, and "Delete" and "Move"
icons are displayed in each window.
5. To close a window, select the corresponding "Delete" icon.
5.5.3Changing the Size of Windows
Each channel tab may contain several windows to evaluate the measurement
results using different methods. A "splitter" allows you to change the size of
neighboring windows.
99Getting Started 1312.9420.02 ─ 15
Page 100
R&S®FSW
Operating the Instrument
Displaying Results
The splitters are not available in SmartGrid mode.
► To change the size of two neighboring windows, drag the splitter between the
windows in either direction.
5.5.4Switching Between a Split and Maximized Window Dis-
play
To get an overview of the results, displaying several windows at the same time
may be helpful. However, the individual windows may become rather small. In
this case it is useful to maximize an individual window to the entire screen temporarily in order to analyze the results in more detail.
To switch between a split and a maximized display without having to close and
re-open windows, press the SPLIT/MAXIMIZE key on the front panel. In maximized display, the currently focused window is maximized. In split display, all
active windows are displayed.
100Getting Started 1312.9420.02 ─ 15
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.