Rohde & Schwarz FSVA3000, FSV3000 Getting Started

®
R&S R&S
FSVA3000/
®
FSV3000 Signal and Spectrum Analyzer
Getting Started
(=NÞ×2)
1330807302
Getting Started
This manual applies to the following R&S®FSV3000 and R&S®FSVA3000 models with firmware version 1.20 and higher:
R&S®FSV3004 (1330.5000K04) / R&S®FSVA3004 (1330.5000K05)
R&S®FSV3007 (1330.5000K07) / R&S®FSVA3007 (1330.5000K08)
R&S®FSV3013 (1330.5000K13) / R&S®FSVA3013 (1330.5000K14)
R&S®FSV3030 (1330.5000K30) / R&S®FSVA3030 (1330.5000K31)
R&S®FSV3044 (1330.5000K43) / R&S®FSVA3044 (1330.5000K44)
In addition to the base unit, the following options are described:
R&S®FSV3-B3, Audio Demodulator (1330.3765.02)
R&S®FSV3-B4, OCXO frequency reference (1330.3794.02)
R&S®FSV3-B5, Additional Interfaces (1330.3820.02)
R&S®FSV3-B6, 10Gbit/s LAN interface (1330.3913.02)
R&S®FSV3-B10, External generator control (1330.3859.02)
R&S®FSV3-B11, YIG preselector bypass (1330.3865.02)
R&S®FSV3-B24, Preamplifier (1330.4049.XX)
R&S®FSV3-B25, Electronic attenuator (1330.4078.02)
R&S®FSV3-B40, 40 MHz analysis bandwidth extension (1330.4103.02)
R&S®FSV3-B114, enhanced computing power (1330.4910.02)
R&S®FSV3-B200, 200 MHz analysis bandwidth extension (1330.4132.02)
R&S®FSV3-B400, 400 MHz analysis bandwidth extension (1330.7154.02) / R&S®FSV3-U400 (1330.7183.02)
R&S®FSV3-K3, Power Sensor support (1346.3676.02)
© 2019 Rohde & Schwarz GmbH & Co. KG Mühldorfstr. 15, 81671 München, Germany Phone: +49 89 41 29 - 0 Fax: +49 89 41 29 12 164 Email: info@rohde-schwarz.com Internet: www.rohde-schwarz.com Subject to change – Data without tolerance limits is not binding. R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG. Trade names are trademarks of the owners.
1330.8073.02 | Version 03 | R&S®FSVA3000/R&S®FSV3000
The following abbreviations are used throughout this manual: R&S®FSVA3000 is abbreviated as R&S FSVA3000. R&S®FSV3000 is abbreviated as R&S FSV3000. R&S®FSV/A refers to both the R&S FSV3000 and the R&S FSVA3000.
1

Safety Instructions

Risk of injury and instrument damage
The instrument must be used in an appropriate manner to prevent electric shock, fire,
Keep the "Basic Safety Instructions" and the product documentation
Riesgo de lesiones y daños en el instrumento
El instrumento se debe usar de manera adecuada para p descargas eléctricas, incendios, lesiones o daños materiales.
especificaciones técnicas pueden contener condiciones adicionales
Instrucciones de seguridad Sicherheitshinweise Consignes de sécurité
personal injury or instrument damage.
Do not open the instrument casing. Read and observe the "Basic Sa fety Instructions" delivered as
printed brochure with the instrument.
Read and observe the safety instructions in the following sections.
Note that the data sheet may specify additional operating conditions.
in a safe place and pass them on to the subsequent users.
No abrir la carcasa del instrumento. Lea y cumpla las "Instrucciones de seguridad elementales"
suministradas con el instrumento como folleto impreso.
Lea y cumpla las instrucciones de seguridad incluidas en las
siguientes secciones. Se debe tener en cuenta que las para su uso.
Guarde bien las instrucciones de seguridad elementales, así como
la documentación del producto, y entréguelas a usuarios posteriores.
revenir
1171.1307.42 - 05
2
Gefahr von Verletzungen und Schäden am Ger ät
Betreiben Sie das Gerät immer ordnungsgemäß, um elektrischen Schlag, Brand, Verletzungen von Personen oder Geräteschäden zu verhindern.
Risque de blessures et d'endommagement de l'appareil
L'ap les électrocutions, incendies, dommages corporels et matériels.
suivantes. Il ne faut pas oublier que la fiche technique peut indiquer
Öffnen Sie das Gerätegehäuse nicht. Lesen und beachten Sie die "Grundlegenden Sicherheitshinweise",
die als gedruckte Broschüre dem Gerät beiliegen.
Lesen und beachten Sie die Sicherheitshinweise in den folgenden
Abschnitten; möglicherweise enthält das Datenblatt weitere Hinweise zu speziellen Betriebsbedingungen.
Bewahren Sie die "Grundlegenden Sicherheitshinweise" und die
Produktdokumentation gut auf und geben Sie diese an weitere Benutzer des Produkts weiter.
pareil doit être utilisé conformément aux prescriptions afin d'éviter
N'ouvrez pas le boîtier de l'appareil. Lisez et respectez les "consignes de sécurité fondamentales"
fournies avec l’app ar eil sous forme de brochure imprimée.
Lisez et respectez les instructions de sécurité dans les sections
des conditions d’exploitation supplémentaires.
Gardez les consignes de sécurité fondamentales et la
documentation produit dans un lieu sûr et transmettez ces documents aux autres utilisateurs.
1171.1307.42 - 05
R&S®FSVA3000/R&S®FSV3000

Contents

1 Preface....................................................................................5
1.1 Key Features......................................................................................... 5
1.2 Conventions Used in the Documentation...........................................6
2 Safety Information................................................................. 8
3 Documentation Overview......................................................9
3.1 Getting Started Manual.........................................................................9
3.2 User Manuals and Help........................................................................ 9
3.3 Service Manual....................................................................................10
Contents
3.4 Instrument Security Procedures....................................................... 10
3.5 Basic Safety Instructions................................................................... 10
3.6 Data Sheets and Brochures............................................................... 10
3.7 Release Notes and Open Source Acknowledgment (OSA).............11
3.8 Application Notes, Application Cards, White Papers, etc...............11
4 Preparing for Use.................................................................12
4.1 Putting into Operation........................................................................ 12
4.2 Windows Operating System...............................................................20
4.3 Connecting USB Devices................................................................... 25
4.4 Connecting an External Monitor........................................................27
4.5 Setting Up a Network (LAN) Connection.......................................... 28
4.6 Configuring the Initial Instrument Settings...................................... 35
4.7 Protecting Data Using the Secure User Mode..................................35
5 Instrument Tour................................................................... 39
5.1 Front Panel View................................................................................. 39
5.2 Rear Panel View.................................................................................. 50
3Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Contents
6 Trying Out the Instrument...................................................56
6.1 Measuring a Basic Signal...................................................................56
6.2 Displaying a Spectrogram..................................................................58
6.3 Activating Additional Measurement Channels.................................60
6.4 Performing Sequential Measurements............................................. 65
6.5 Setting and Moving a Marker............................................................. 66
6.6 Displaying a Marker Peak List........................................................... 68
6.7 Zooming into the Display................................................................... 69
6.8 Zooming into the Display Permanently............................................ 73
6.9 Saving Settings................................................................................... 76
6.10 Printing and Saving Results.............................................................. 78
7 Operating the Instrument....................................................80
7.1 Understanding the Display Information............................................81
7.2 Accessing the Functionality.............................................................. 90
7.3 Entering Data.......................................................................................96
7.4 Touchscreen Gestures..................................................................... 100
7.5 Displaying Results............................................................................104
7.6 Getting Help.......................................................................................113
7.7 Remote Control................................................................................. 114
8 Contacting Customer Support..........................................117
Index....................................................................................118
4Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Key Features
Preface

1 Preface

1.1 Key Features

The R&S FSV3000 Signal and Spectrum Analyzer and the R&S FSVA3000 Sig­nal and Spectrum Analyzer set standards in RF performance and usability. Key features are:
Wide analysis bandwidth (up to 200 MHz for the FSV3000 and up to 400 MHz for the FSVA3000)
Very high measurement speed
Low phase noise
Capacitive touchscreen with multi touch operation
Excellent dynamic range
Smart wizards for fast setup of complex measurement tasks
SCPI recorder
Virtual one box instrument with a Rohde & Schwarz signal generator
Multiple measurement applications can be run and displayed in parallel
Optional measurement applications for demodulation of analog and digital modulated signals
For a detailed specification, refer to the data sheet.
Due to these features, the R&S FSV3000 and the R&S FSVA3000 are ideal for various measurement tasks, for instance:
Functional check and repair in service and maintenance due to the excellent usability
Measurements on 5G NR and other wireless standards with the optional mea­surement applications for cellular and WLAN standards
Production of components due to the high measurement speed
Analysis of wideband modulated signals, short events or the linearization of amplifiers due to the wide analysis bandwidth
Measuring oscillators due to the low phase noise
5Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Conventions Used in the Documentation
Identifying and analyzing spurious emissions due to the large spurious-free dynamic range and low DANL

1.2 Conventions Used in the Documentation

1.2.1 Typographical Conventions

The following text markers are used throughout this documentation:
Convention Description
Preface
"Graphical user interface elements"
[Keys] Key and knob names are enclosed by square brackets.
Filenames, commands, program code
Input Input to be entered by the user is displayed in italics.
Links Links that you can click are displayed in blue font.
"References" References to other parts of the documentation are enclosed by
All names of graphical user interface elements on the screen, such as dialog boxes, menus, options, buttons, and softkeys are enclosed by quotation marks.
Filenames, commands, coding samples and screen output are distinguished by their font.
quotation marks.

1.2.2 Conventions for Procedure Descriptions

When operating the instrument, several alternative methods may be available to perform the same task. In this case, the procedure using the touchscreen is described. Any elements that can be activated by touching can also be clicked using an additionally connected mouse. The alternative procedure using the keys on the instrument or the on-screen keyboard is only described if it deviates from the standard operating procedures.
The term "select" may refer to any of the described methods, i.e. using a finger on the touchscreen, a mouse pointer in the display, or a key on the instrument or on a keyboard.
6Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Conventions Used in the Documentation
Preface

1.2.3 Notes on Screenshots

When describing the functions of the product, we use sample screenshots. These screenshots are meant to illustrate as many as possible of the provided functions and possible interdependencies between parameters. The shown values may not represent realistic usage scenarios.
The screenshots usually show a fully equipped product, that is: with all options installed. Thus, some functions shown in the screenshots may not be available in your particular product configuration.
7Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Safety Information

2 Safety Information

The product documentation helps you use the R&S FSV/A safely and efficiently. Follow the instructions provided here and in the printed "Basic Safety Instruc­tions". Keep the product documentation nearby and offer it to other users.
Intended use
The R&S FSV/A is intended for the development, production and verification of electronic components and devices in industrial, administrative, and laboratory environments. Use the R&S FSV/A only for its designated purpose. Observe the operating conditions and performance limits stated in the data sheet.
Where do I find safety information?
Safety information is part of the product documentation. It warns you about the potential dangers and gives instructions how to prevent personal injuries or dam­age caused by dangerous situations. Safety information is provided as follows:
The printed "Basic Safety Instructions" provide safety information in many lan­guages and are delivered with the R&S FSV/A.
Throughout the documentation, safety instructions are provided when you need to take care during setup or operation.
8Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Documentation Overview
User Manuals and Help

3 Documentation Overview

This section provides an overview of the R&S FSV/A user documentation. Unless specified otherwise, you find the documents on the R&S FSV/A product page at:
www.rohde-schwarz.com/product/FSVA3000.html/
www.rohde-schwarz.com/product/FSV3000.html

3.1 Getting Started Manual

Introduces the R&S FSV/A and describes how to set up and start working with the product. Includes basic operations, typical measurement examples, and general information, e.g. safety instructions, etc.
A printed version is delivered with the instrument. A PDF version is available for download on the Internet.

3.2 User Manuals and Help

Separate user manuals are provided for the base unit and the firmware applica­tions:
Base unit manual Contains the description of all instrument modes and functions. It also pro­vides an introduction to remote control, a complete description of the remote control commands with programming examples, and information on mainte­nance, instrument interfaces and error messages. Includes the contents of the getting started manual.
Firmware application manual Contains the description of the specific functions of a firmware application, including remote control commands. Basic information on operating the R&S FSV/A is not included.
The contents of the user manuals are available as help in the R&S FSV/A. The help offers quick, context-sensitive access to the complete information for the base unit and the firmware applications.
9Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
All user manuals are also available for download or for immediate display on the Internet.
Documentation Overview
Data Sheets and Brochures

3.3 Service Manual

Describes the performance test for checking the rated specifications, module replacement and repair, firmware update, troubleshooting and fault elimination, and contains mechanical drawings and spare part lists.
The service manual is available for registered users on the global Rohde & Schwarz information system (GLORIS):
https://gloris.rohde-schwarz.com

3.4 Instrument Security Procedures

Deals with security issues when working with the R&S FSV/A in secure areas. It is available for download on the Internet.

3.5 Basic Safety Instructions

Contains safety instructions, operating conditions and further important informa­tion. The printed document is delivered with the instrument.

3.6 Data Sheets and Brochures

The data sheet contains the technical specifications of the R&S FSV/A. It also lists the firmware applications and their order numbers, and optional accessories.
The brochure provides an overview of the instrument and deals with the specific characteristics.
See www.rohde-schwarz.com/brochure-datasheet/FSV3000 /
www.rohde-schwarz.com/brochure-datasheet/FSVA3000
10Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Application Notes, Application Cards, White Papers, etc.
Documentation Overview
3.7 Release Notes and Open Source Acknowledg­ment (OSA)
The release notes list new features, improvements and known issues of the cur­rent firmware version, and describe the firmware installation.
The open-source acknowledgment document provides verbatim license texts of the used open source software.
See www.rohde-schwarz.com/firmware/FSV3000 /
www.rohde-schwarz.com/firmware/FSVA3000

3.8 Application Notes, Application Cards, White Papers, etc.

These documents deal with special applications or background information on particular topics.
See www.rohde-schwarz.com/application/FSV3000 /
www.rohde-schwarz.com/application/FSVA3000
11Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Preparing for Use
Putting into Operation

4 Preparing for Use

Putting into Operation..................................................................................... 12
Windows Operating System............................................................................20
Connecting USB Devices................................................................................25
Connecting an External Monitor......................................................................27
Setting Up a Network (LAN) Connection.........................................................28
Configuring the Initial Instrument Settings...................................................... 35
Protecting Data Using the Secure User Mode................................................ 35

4.1 Putting into Operation

This section describes the basic steps to be taken when setting up the R&S FSV/A for the first time.
Risk of injury due to disregarding safety information
Observe the information on appropriate operating conditions provided in the data sheet to prevent personal injury or damage to the instrument. Read and observe the basic safety instructions provided with the instrument, in addition to the safety instructions in the following sections. In particular:
Do not open the instrument casing.
Risk of instrument damage due to inappropriate operating conditions
Specific operating conditions are required to ensure accurate measure­ments and to avoid damage to the instrument. Observe the information on appropriate operating conditions provided in the basic safety instructions and the instrument's data sheet.
12Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Instrument damage caused by electrostatic discharge
Electrostatic discharge (ESD) can damage the electronic components of the instrument and the device under test (DUT). Electrostatic discharge is most likely to occur when you connect or disconnect a DUT or test fixture to the instrument's test ports. To prevent electrostatic discharge, use a wrist strap and cord and connect yourself to the ground, or use a conductive floor mat and heel strap combination.
Risk of instrument damage due to inappropriate operating conditions
An unsuitable operating site or test setup can damage the instrument and connected devices. Before switching on the instrument, observe the infor­mation on appropriate operating conditions provided in the data sheet. In particular, ensure the following:
Preparing for Use
Putting into Operation
All fan openings are unobstructed and the airflow perforations are unim­peded. A minimum distance of 10 cm to other objects is recommended.
The instrument is dry and shows no sign of condensation.
The instrument is positioned as described in the following sections.
The ambient temperature does not exceed the range specified in the data sheet.
Signal levels at the input connectors are all within the specified ranges.
Signal outputs are connected correctly and are not overloaded.
EMI impact on measurement results
Electromagnetic interference (EMI) may affect the measurement results. To suppress generated electromagnetic interference (EMI):
Use suitable shielded cables of high quality. For example, use double­shielded RF and LAN cables.
Always terminate open cable ends.
Note the EMC classification in the data sheet.
13Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Unpacking and Checking the Instrument........................................................ 14
Accessory List.................................................................................................14
Placing or Mounting the Instrument................................................................ 15
Connecting the AC Power...............................................................................17
Switching the Instrument On and Off.............................................................. 17
Performing a Self-Alignment and a Selftest.................................................... 18
Checking the Supplied Options.......................................................................19
Preparing for Use
Putting into Operation

4.1.1 Unpacking and Checking the Instrument

Check the equipment for completeness using the delivery note and the accessory lists for the various items. Check the instrument for any damage. If there is dam­age, immediately contact the carrier who delivered the instrument. Make sure not to discard the box and packing material.
Packing material
Retain the original packing material. If the instrument needs to be transpor­ted or shipped later, you can use the material to protect the control ele­ments and connectors.
Risk of injury during transportation
The carrying handles at the front and side of the casing are designed to lift or carry the instrument. Do not apply excessive force to the handles. If a handle is ripped off, the falling instrument can cause injury.
Be aware of the weight of the instrument when lifting it. Observe the infor­mation on transporting heavy instruments in the basic safety instructions provided with the instrument.

4.1.2 Accessory List

The instrument comes with the following accessories:
Power cable
Printed Getting Started manual
14Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Preparing for Use
Putting into Operation

4.1.3 Placing or Mounting the Instrument

The R&S FSV/A is designed for use under laboratory conditions, either on a bench top or in a rack.
Bench Top Operation
If the R&S FSV/A is operated on a bench top, the surface should be flat. The instrument can be used in horizontal position, standing on its feet, or with the sup­port feet on the bottom extended.
Risk of injury if feet are folded out
The feet can fold in if they are not folded out completely or if the instrument is shifted. Collapsing feet can cause injury or damage the instrument.
Fold the feet completely in or out to ensure stability of the instrument. Never shift the instrument when the feet are folded out.
When the feet are folded out, do not work under the instrument or place anything underneath.
The feet can break if they are overloaded. The overall load on the fol­ded-out feet must not exceed 500 N.
15Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Risk of injury when stacking instruments
A stack of instruments can tilt over and cause injury if not stacked correctly. Furthermore, the instruments at the bottom of the stack can be damaged due to the load imposed by the instruments on top.
Observe the following instructions when stacking instruments:
Never stack more than three instruments. If you need to stack more than three instruments, install them in a rack.
The overall load imposed on the lowest instrument must not exceed 500 N.
It is best if all instruments have the same dimensions (width and length). If you need to stack smaller instruments on the top, the overall load imposed on the lowest instrument must not exceed 250 N.
Preparing for Use
Putting into Operation
If the instruments have foldable feet, fold them in completely.
Rackmounting
The R&S FSV/A can be installed in a rack using a rack adapter kit (order no. see data sheet). The installation instructions are part of the adapter kit.
Risk of instrument damage due to insufficient airflow in a rack
If you mount several instruments in a rack, you need an efficient ventilation concept to ensure that the instruments do not overheat. Insufficient airflow for a longer period can disturb the operation and even cause damage.
16Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Preparing for Use
Putting into Operation

4.1.4 Connecting the AC Power

In the standard version, the R&S FSV/A is equipped with an AC power supply connector. The R&S FSV/A can be used with different AC power voltages and adapts itself automatically to it. Refer to the datasheet for the requirements of voltage and frequency. The AC power connector is located on the rear panel of the instrument.
For details on the connector, refer to Chapter 5.2.1, "AC Power Supply Connec-
tion and Main Power Switch", on page 51.
► Connect the R&S FSV/A to the AC power supply using the supplied power
cable. Since the instrument complies with the specifications for safety class EN61010, only connect it to an outlet that has a ground contact.

4.1.5 Switching the Instrument On and Off

Switching on the instrument
► Press the AC power switch on the rear panel to position "I".
The instrument is supplied with AC power. After booting, the instrument is ready for operation. The [POWER] key lights up green. If the [POWER] key lights up orange, the instrument is in standby mode.
Switching off the instrument
1. Press the [POWER] key on the front panel.
The R&S FSV/A switches to standby mode.
2. Set the AC power switch on the rear panel to position "O", or disconnect the
instrument from the AC power supply.
The R&S FSV/A changes into off mode.
17Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Putting into Operation
If the instrument temperature exceeds the limit specified in the data sheet, the R&S FSV/A automatically shuts down to protect the instrument from damage.
Risk of losing data
If you switch off the running instrument using the rear panel switch or by disconnecting the power cord, the instrument loses its current settings. Fur­thermore, program data can be lost.
Press the Power key first to shut down the application properly.

4.1.6 Performing a Self-Alignment and a Selftest

Preparing for Use
During instrument start, the installed hardware is checked against the cur­rent firmware version to ensure that the hardware is supported. If not, an error message is displayed ("WRONG_FW") and you are asked to update the firmware. Until the firmware version is updated, self-alignment fails and the instrument's performance is not as specified.
(For details refer to the R&S FSV/A User Manual).
When temperature changes occur, you may have to perform a self-alignment to align the data to a reference source. See the data sheet for details.
Operating temperature
Before performing this functional test, make sure that the instrument has reached its operating temperature (for details, refer to the data sheet).
A message in the status bar ("Instrument warming up...") indicates that the oper­ating temperature has not yet been reached.
Performing a self-alignment
1. Press the [SETUP] key.
2. Press the "Alignment" softkey.
18Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
3. Select the "Start Self-Alignment" button in the "Alignment" dialog box.
Once the system correction values have been calculated successfully, a mes­sage is displayed.
To display the alignment results again later
Press the [SETUP] key.
Press the "Alignment" softkey.
Performing a selftest
You do not have to repeat the selftest every time the instrument is switched on. It is only necessary when instrument malfunction is suspected.
1. Press the [SETUP] key.
Preparing for Use
Putting into Operation
2. Select the "Service and Support" softkey.
3. Switch to the "Selftest" tab in the "Service" dialog box.
4. Select the "Start Selftest" button.
Once the instrument modules have been checked successfully, a message is displayed.

4.1.7 Checking the Supplied Options

The instrument can be equipped with both hardware and firmware options. To check whether the installed options correspond to the options indicated on the delivery note, proceed as follows.
1. Press the [SETUP] key.
2. Press the "System Config" softkey.
3. Switch to the "Versions + Options" tab in the "System Configuration" dialog
box. A list with hardware and firmware information is displayed.
4. Check the availability of the hardware options as indicated in the delivery
note.
19Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Windows Operating System
Preparing for Use

4.2 Windows Operating System

The instrument contains the Windows 10 operating system which has been con­figured according to the instrument's features and needs. Changes in the system setup are only required when peripherals like keyboard or a printer are installed or if the network configuration does not comply with the default settings. After the R&S FSV/A is started, the operating system boots and the instrument firmware is started automatically.
To ensure that the instrument software functions properly, certain rules must be adhered to concerning the operating system.
Tested software
The drivers and programs used on the instrument under Windows 10 are adapted to the instrument. Only install update software released by Rohde & Schwarz to modify existing instrument software.
You can install additional software on the instrument, however, additional soft­ware can impair instrument function. Thus, run only programs that Rohde & Schwarz has tested for compatibility with the instrument software.
The following program packages have been tested:
R&S Power Viewer Plus - virtual power meter for displaying results of the power sensor R&S NRPxx (install only this component!)
Symantec Endpoint Security – virus-protection software
FileShredder - for reliable deletion of files on the hard disk

4.2.1 Virus Protection

Take appropriate steps to protect your instruments from infection. Use strong fire­wall settings and scan any removable storage device used with a Rohde & Schwarz instrument regularly. It is also recommended that you install anti-virus software on the instrument. Rohde & Schwarz does NOT recommend running anti-virus software in the background ("on-access" mode) on Windows­based instruments, due to potentially degrading instrument performance. How­ever, Rohde & Schwarz does recommend running it during non-critical hours.
20Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Windows Operating System
For details and recommendations, see the following Rohde & Schwarz white paper:
1EF96: Malware Protection Windows 10
Preparing for Use

4.2.2 Service Packs and Updates

Microsoft regularly creates security updates and other patches to protect Win­dows-based operating systems. These are released through the Microsoft Update website and associated update server. Instruments using Windows, especially those that connect to a network, should be updated regularly.
For details and recommendations, see the following Rohde & Schwarz white paper:
1EF96: Malware Protection Windows 10

4.2.3 Login

Windows 10 requires that users identify themselves by entering a user name and password in a login window. By default, the R&S FSV/A provides two user accounts:
"Instrument": a standard user account with limited access
"Administrator": an administrator account with unrestricted access to the computer/domain
Some administrative tasks require administrator rights (e.g. the configuration of a LAN network). Refer to the description of the basic instrument setup ([SETUP] menu) to find out which functions are affected.
21Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Secure user mode
If the secure user mode option (R&S FSV/A-K33) is installed, an additional account is provided: the "SecureUser".
The "SecureUser" is a standard user account with limited functionality. In particular, administrative tasks such as LAN configuration or general instru­ment settings are not available. Furthermore, for a "SecureUser", data that the R&S FSV/A normally stores on the solid-state drive is redirected to vola­tile memory instead. You can access data that is stored in volatile memory during the current instrument session. However, when the instrument’s power is removed, all data in volatile memory is erased.
For details, see Chapter 4.7, "Protecting Data Using the Secure User
Mode", on page 35.
Automatic login
Preparing for Use
Windows Operating System
For the standard instrument account, an automatic login function is active by default. If activated, login is carried out automatically in the background when the R&S FSV/A is started, without having to enter a password. This function is active until an administrator explicitly deactivates it or changes the password.
For information on how to deactivate or reactivate the automatic login, refer to
Chapter 4.2.3.1, "Automatic Login Function", on page 23.
Passwords
For all default user accounts, the initial password is 894129. Note that this pass­word is very weak, and it is recommended that you change the password for both users after initial login. An administrator can change the password in Windows 10 for any user at any time via "Start > Settings > Account > SignIn Options > Pass­word > Change".
Changing the password and use of auto-login function
Note that when you change the default passwords, the default auto-login function no longer works! Reactivate it manually as described in "Reactivat-
ing the automatic login function" on page 24.
22Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Windows Operating System
4.2.3.1 Automatic Login Function
When shipped, the instrument automatically logs on the default "Instrument" user to Windows 10 using the default password.
Switching users when using the automatic login function
Which user account is used is defined during login. If automatic login is active, the login window is not displayed. However, you can also switch the user account to be used when the automatic login function is active.
1. Select the "Windows" icon in the toolbar to access the operating system of the
R&S FSV/A (see also Chapter 4.2.4, "Accessing the Start Menu", on page 24).
Preparing for Use
2. Press [CTRL + ALT + DEL], then select "Sign out".
The "Login" dialog box is displayed, in which you can enter the different user account name and password.
Deactivating the automatic login function
To deactivate the automatic login function, perform the following steps:
1. Select the "Windows" icon in the toolbar to access the operating system of the
R&S FSV/A (see also Chapter 4.2.4, "Accessing the Start Menu", on page 24).
2. In the "Start" menu, select "Run".
The "Run" dialog box is displayed.
3. Enter the command C:\R_S\INSTR\USER\NO_AUTOLOGIN.REG.
4. Press the [ENTER] key to confirm.
This command deactivates the automatic login function. The next time you switch on the instrument, the operating system prompts you to enter your user name and password before it starts the firmware.
23Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Windows Operating System
Adapting the automatic login function to a new password
If you change the "Instrument" user's password, which is used during automatic login, this function no longer works. Adapt the settings for the command that acti­vates the auto login function first.
1. Open the C:\R_S\INSTR\USER\NO_AUTOLOGIN.REG file in any text editor
(e.g. Notepad).
2. In the line "DefaultPassword"="894129", replace the default password
(894129) by the new password for automatic login.
3. Save the changes to the file.
Reactivating the automatic login function
1. Select the "Windows" icon in the toolbar to access the operating system of the
R&S FSV/A (see also Chapter 4.2.4, "Accessing the Start Menu", on page 24).
Preparing for Use
2. In the "Start" menu, select "Run".
The "Run" dialog box is displayed.
3. Enter the command C:\R_S\INSTR\USER\AUTOLOGIN.REG.
4. Press the [ENTER] key to confirm.
This command reactivates automatic login function. It is active the next time the instrument reboots.

4.2.4 Accessing the Start Menu

The Windows "Start" menu provides access to the Windows 10 functionality and installed programs.
To open the "Start" menu:
► Select the "Windows" icon in the toolbar, or press the "Windows" key or the
[CTRL + ESC] key combination on the (external) keyboard.
24Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Connecting USB Devices
All necessary system settings can be defined in the "Start > Settings" menu.
(For required settings refer to the Windows 10 documentation and to the hard­ware description).
Preparing for Use

4.2.5 Accessing the Windows Taskbar

The Windows taskbar also provides quick access to commonly used programs, for example Paint or WordPad. IECWIN, the auxiliary remote control tool provided free of charge and installed by Rohde & Schwarz, is also available from the task­bar.
For details on the IECWIN tool, see the "Network and Remote Control" chapter of the R&S FSV/A user manual.
To open the taskbar, select the "Windows" icon on the R&S FSV/A toolbar, or press the "Windows" key or the [CTRL + ESC] key combination on your (external) keyboard.

4.3 Connecting USB Devices

The USB interfaces of the R&S FSV/A allow you to connect USB devices, such as a mouse, directly to the instrument. Increase the number of possible connec­tions using USB hubs. Due to the large number of available USB devices, there is almost no limit to the expansions that are possible with the R&S FSV/A.
All USB devices can be connected to or disconnected from the instrument during operation.
To connect a USB device
1. Connect the device to the USB interface of the R&S FSV/A.
The operating system automatically searches for a suitable device driver.
2. If Windows does not find a suitable driver:
a) If the driver software is on a CD, connect a USB CD-ROM drive to the
instrument.
b) Specify a directory that contains the driver software.
25Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Connecting USB Devices
To disconnect a USB device
► Remove the device from the USB interface of the R&S FSV/A.
Windows immediately detects the change in hardware configuration and deac­tivates the corresponding driver.
Preparing for Use

4.3.1 Connecting a Keyboard

The keyboard is detected automatically when it is connected. The default input language is English – US.
However, you can also connect foreign language keyboards; currently the follow­ing languages are supported for the R&S FSV/A:
German
Swiss
French
Russian
To configure the keyboard language
1. To access the Windows operating system, press the Windows key on the
external keyboard.
2. Select "Start > Settings > Time & language > Region & language > Add a lan-
guage" .

4.3.2 Connecting a Printer

When printing a file, the instrument checks whether a printer is connected and turned on and whether the appropriate printer driver is installed. If necessary, printer driver installation is initiated. You only have to install a printer driver once.
You can load updated and improved driver versions or new drivers from an instal­lation disk, USB memory stick or another external storage medium. If the instru­ment is integrated in a network, you can also install driver data stored in a net­work directory.
26Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Connecting an External Monitor
Preparing for Use

4.4 Connecting an External Monitor

You can connect an external monitor (or projector) to the "DVI" or "display port" connector on the rear panel of the R&S FSV/A (see also Chapter 5.2.2, "Display
Port and DVI", on page 51).
Screen resolution and format
The touchscreen of the R&S FSV/A is calibrated for a 16:10 format. If you connect a monitor or projector using a different format (e.g. 4:3), the calibra­tion is not correct and the screen does not react to your touch actions prop­erly.
The touchscreen has a screen resolution of 1280x800 pixels. Most external monitors have a higher screen resolution. If the screen resolution of the monitor is set higher than the instrument's resolution, the application win­dow uses an area of 1280x800 pixels on the monitor display. For full screen display, adjust the monitor's screen resolution.
The R&S FSV/A supports a minimum resolution of 1280x768 pixels.
1. Connect the external monitor to the R&S FSV/A.
2. Press the [SETUP] key.
3. Press the "Display" softkey.
4. Select the "Configure Monitor" tab in the "Display" dialog box.
The standard Windows "Screen Resolution" dialog box is displayed.
27Getting Started 1330.8073.02 ─ 03
R&S®FSVA3000/R&S®FSV3000
Preparing for Use
Setting Up a Network (LAN) Connection
5. If necessary, change the screen resolution. Consider the information in the
note above.
6. Select the instrument to be used for display:
"Display 1": internal monitor only
"Display 2": external monitor only
"Duplicate": both internal and external monitor
7. Tap "Apply" to try out the settings before they are accepted permanently, then
you can easily return to the previous settings, if necessary.
8. Select "OK" if the settings are suitable.

4.5 Setting Up a Network (LAN) Connection

A LAN connection is the prerequisite for all network operations. The LAN connec­tion settings can be configured directly in the Windows operating system.
The R&S FSV/A is equipped with a network interface and can be connected to an Ethernet LAN (local area network). Provided the network administrator has assigned you the appropriate rights and adapted the Windows firewall configura­tion, you can use the interface, for example:
28Getting Started 1330.8073.02 ─ 03