Rohde&Schwarz ESRP3, ESRP7, FSV-B9, ESRP-K53, ESRP-K56 Getting started

®
R&S
ESRP EMI Test Receiver Getting Started
1316469702 Version 07
This manual describes the following R&S®ESRP models with firmware version
3.36 and higher:
R&S®ESRP3 (1316.4500K03)
R&S®ESRP7 (1316.4500K07)
© 2021 Rohde & Schwarz GmbH & Co. KG Mühldorfstr. 15, 81671 München, Germany Phone: +49 89 41 29 - 0 Email: info@rohde-schwarz.com Internet: www.rohde-schwarz.com Subject to change – data without tolerance limits is not binding. R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG. Trade names are trademarks of the owners.
1316.4697.02 | Version 07 | R&S®ESRP
Throughout this manual, products from Rohde & Schwarz are indicated without the ® symbol , e.g. R&S®ESRP is indicated as R&S ESRP.
R&S®ESRP

Contents

1 Safety and Regulatory Information...................................... 9
1.1 Safety Instructions................................................................................9
1.2 Warning Messages in the Documentation........................................14
1.3 Korea certification class A.................................................................15
2 Documentation Overview....................................................16
2.1 Getting Started Manual.......................................................................16
2.2 User Manuals and Help...................................................................... 16
2.3 Service Manual....................................................................................16
Contents
2.4 Instrument Security Procedures....................................................... 17
2.5 Basic Safety Instructions................................................................... 17
2.6 Data Sheets and Brochures............................................................... 17
2.7 Release Notes and Open Source Acknowledgment (OSA).............17
2.8 Application Notes, Application Cards, White Papers, etc...............18
3 Conventions Used in the Documentation......................... 19
3.1 Typographical Conventions............................................................... 19
3.2 Conventions for Procedure Descriptions.........................................19
3.3 Notes on Screenshots........................................................................ 20
4 Instrument Tour................................................................... 21
4.1 The Front Panel...................................................................................21
4.1.1 Touchscreen Display.............................................................................22
4.1.2 Function Keys on the Front Panel.........................................................23
4.1.3 USB Interface........................................................................................26
4.1.4 Noise Source Control............................................................................26
4.1.5 RF Input 50.........................................................................................26
3Getting Started 1316.4697.02 ─ 07
R&S®ESRP
4.1.6 Probe Power......................................................................................... 27
4.1.7 AF Output..............................................................................................27
4.1.8 Tracking Generator Output (Optional)...................................................27
4.2 Rear Panel View.................................................................................. 28
4.2.1 AC Power Supply Connection and Main Power Switch........................ 29
4.2.2 LAN....................................................................................................... 29
4.2.3 VGA Interface....................................................................................... 29
4.2.4 Ext Trigger / Gate In..............................................................................29
4.2.5 Ref In.................................................................................................... 29
4.2.6 Ref Out..................................................................................................30
4.2.7 GPIB Interface...................................................................................... 30
Contents
4.2.8 Trigger Output.......................................................................................30
4.2.9 IF / Video...............................................................................................30
4.2.10 USB.......................................................................................................31
4.2.11 AUX Port............................................................................................... 31
4.2.12 OCXO option (R&S FSV-B4)................................................................ 31
4.2.13 Device ID.............................................................................................. 32
5 Preparing For Use................................................................33
5.1 Lifting and Carrying............................................................................33
5.2 Unpacking and Checking................................................................... 33
5.3 Choosing the Operating Site............................................................. 34
5.4 Setting up the R&S ESRP...................................................................34
5.5 Placing the R&S ESRP on a Bench Top............................................35
5.6 Mounting the R&S ESRP in a Rack................................................... 36
5.7 Connecting the AC Power..................................................................36
5.8 Connecting to a DC Power Source (Optional)..................................37
5.9 Switching the Instrument On and Off............................................... 38
4Getting Started 1316.4697.02 ─ 07
R&S®ESRP
5.10 Connecting to LAN............................................................................. 39
5.11 Connecting a Keyboard......................................................................41
5.12 Connecting an External Monitor........................................................41
5.13 Windows Operating System...............................................................42
5.14 Logging On..........................................................................................44
5.15 Checking the Supplied Options.........................................................46
5.16 Performing a Self Alignment and a Self Test....................................46
5.17 Considerations for Test Setup........................................................... 47
Contents
6 Basic Operations................................................................. 49
6.1 Information in the Diagram Area....................................................... 49
6.1.1 Channel Display....................................................................................50
6.1.2 Display of Hardware Settings................................................................51
6.1.3 Measurement Settings Information....................................................... 52
6.1.4 Diagram-specific and Trace Information............................................... 54
6.1.5 Instrument and Status Information........................................................56
6.2 Means of User Interaction.................................................................. 57
6.2.1 Toolbar.................................................................................................. 58
6.2.2 Touchscreen..........................................................................................59
6.2.3 On-screen Keyboard.............................................................................60
6.2.4 Keypad..................................................................................................61
6.2.5 Rotary Knob.......................................................................................... 62
6.2.6 Arrow Keys, UNDO/REDO Keys...........................................................62
6.2.7 Softkeys................................................................................................ 63
6.2.8 Context-sensitive Menus.......................................................................65
6.2.9 Dialog Boxes.........................................................................................66
6.3 Setting Parameters............................................................................. 67
6.3.1 Entering Numeric Parameters...............................................................67
5Getting Started 1316.4697.02 ─ 07
R&S®ESRP
6.3.2 Entering Alphanumeric Parameters...................................................... 68
6.3.3 Navigating in Dialog Boxes................................................................... 70
6.4 Changing the Display......................................................................... 73
6.4.1 Changing the Focus..............................................................................73
6.4.2 Switching between Split and Maximized Display.................................. 73
6.4.3 Soft and Mini Soft Front Panel Display................................................. 74
6.4.4 Enlarging the Display Area................................................................... 77
6.4.5 Displaying the Toolbar...........................................................................77
6.4.6 Zooming into the Diagram.....................................................................77
6.4.7 Adding a Title to the Diagram Header...................................................80
6.4.8 Removing the Diagram Footer..............................................................80
Contents
6.4.9 Selecting a Theme................................................................................ 80
6.4.10 Displaying and Setting the Date and Time............................................81
6.4.11 Changing the Display Update Rate.......................................................81
7 Basic Measurement Examples........................................... 83
7.1 Measuring a Sinusoidal Signal.......................................................... 83
7.1.1 Measuring the Level and Frequency Using Markers............................ 84
7.1.2 Measuring the Signal Frequency Using the Frequency Counter.......... 87
7.2 Measuring Harmonics of Sinusoidal Signals................................... 88
7.3 Measuring Signal Spectra with Multiple Signals............................. 92
7.3.1 Separating Signals by Selecting the Resolution Bandwidth................. 92
7.3.2 Measuring the Modulation Depth of an AM-Modulated Carrier (Span >
0)...........................................................................................................96
7.3.3 Measuring AM-Modulated Signals........................................................ 97
7.4 Measurements in Zero Span.............................................................. 99
7.4.1 Measuring the Power Characteristic of Burst Signals...........................99
7.4.2 Measuring the Signal-to-Noise Ratio of Burst Signals........................ 104
6Getting Started 1316.4697.02 ─ 07
R&S®ESRP
7.4.3 Measuring FM-Modulated Signals...................................................... 107
7.5 Storing and Loading Instrument Settings.......................................111
7.5.1 Storing an Instrument Configuration (without Traces).........................111
7.5.2 Storing Traces.....................................................................................112
7.5.3 Loading an Instrument Configuration (with Traces)............................ 112
7.5.4 Configuring Automatic Loading........................................................... 113
Contents
Index....................................................................................114
7Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Contents
8Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Safety and Regulatory Information
Safety Instructions

1 Safety and Regulatory Information

The product documentation helps you use the product safely and efficiently. Fol­low the instructions provided here and in the following chapters.
Intended use
The product is intended for the development, production and verification of elec­tronic components and devices in industrial, administrative, and laboratory envi­ronments. Use the product only for its designated purpose. Observe the operating conditions and performance limits stated in the data sheet.
Where do I find safety information?
Safety information is part of the product documentation. It warns you of potential dangers and gives instructions on how to prevent personal injury or damage caused by dangerous situations. Safety information is provided as follows:
In Chapter 1.1, "Safety Instructions", on page 9. The same information is provided in many languages as printed "Safety Instructions". The printed "Safety Instructions" are delivered with the product.
Throughout the documentation, safety instructions are provided when you need to take care during setup or operation.

1.1 Safety Instructions

Products from the Rohde & Schwarz group of companies are manufactured according to the highest technical standards. To use the products safely, follow the instructions provided here and in the product documentation. Keep the prod­uct documentation nearby and offer it to other users.
Use the product only for its intended use and within its performance limits. Inten­ded use and limits are described in the product documentation such as the data sheet, manuals and the printed "Safety Instructions". If you are unsure about the appropriate use, contact Rohde & Schwarz customer service.
Using the product requires specialists or specially trained personnel. These users also need sound knowledge of at least one of the languages in which the user interfaces and the product documentation are available.
9Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Never open the casing of the product. Only service personnel authorized by Rohde & Schwarz are allowed to repair the product. If any part of the product is damaged or broken, stop using the product. Contact Rohde & Schwarz customer service at http://www.customersupport.rohde-schwarz.com.
Lifting and carrying the product
The maximum weight of the product is provided in the data sheet. To move the product safely, you can use lifting or transporting equipment such as lift trucks and forklifts. Follow the instructions provided by the equipment manufacturer.
The product is heavy. Do not move or carry the product by yourself. A single per­son can only carry a maximum of 18 kg safely depending on age, gender and physical condition. Look up the maximum weight in the data sheet. Use the prod­uct handles to move or carry the product. Do not lift by the accessories mounted on the product. Accessories are not designed to carry the weight of the product.
Safety and Regulatory Information
Safety Instructions
To move the product safely, you can use lifting or transporting equipment such as lift trucks and forklifts. Follow the instructions provided by the equipment manu­facturer.
Choosing the operating site
Only use the product indoors. The product casing is not waterproof. Water that enters can electrically connect the casing with live parts, which can lead to elec­tric shock, serious personal injury or death if you touch the casing. If Rohde & Schwarz provides accessories designed for your product, e.g. a carrying bag, you can use the product outdoors.
Unless otherwise specified, you can operate the product up to an altitude of 2000 m above sea level. The product is suitable for pollution degree 2 environ­ments where nonconductive contamination can occur. For more information on environmental conditions such as ambient temperature and humidity, see the data sheet.
Setting up the product
Always place the product on a stable, flat and level surface with the bottom of the product facing down. If the product is designed for different positions, secure the product so that it cannot fall over.
If the product has foldable feet, always fold the feet completely in or out to ensure stability. The feet can collapse if they are not folded out completely or if the prod-
10Getting Started 1316.4697.02 ─ 07
R&S®ESRP
uct is moved without lifting it. The foldable feet are designed to carry the weight of the product, but not an extra load.
If stacking is possible, keep in mind that a stack of products can fall over and cause injury.
If you mount products in a rack, ensure that the rack has sufficient load capacity and stability. Observe the specifications of the rack manufacturer. Always install the products from the bottom shelf to the top shelf so that the rack stands securely. Secure the product so that it cannot fall off the rack.
Connecting to power
The product is an overvoltage category II product. Connect the product to a fixed installation used to supply energy-consuming equipment such as household appliances and similar loads. Keep in mind that electrically powered products have risks, such as electric shock, fire, personal injury or even death.
Safety and Regulatory Information
Safety Instructions
Take the following measures for your safety:
Before switching on the product, ensure that the voltage and frequency indica­ted on the product match the available power source. If the power adapter does not adjust automatically, set the correct value and check the rating of the fuse.
If a product has an exchangeable fuse, its type and characteristics are indica­ted next to the fuse holder. Before changing the fuse, switch off the instrument and disconnect it from the power source. How to change the fuse is described in the product documentation.
Only use the power cable delivered with the product. It complies with country­specific safety requirements. Only insert the plug into an outlet with protective conductor terminal.
Only use intact cables and route them carefully so that they cannot be dam­aged. Check the power cables regularly to ensure that they are undamaged. Also ensure that nobody can trip over loose cables.
If the product needs an external power supply, use the power supply that is delivered with the product or that is recommended in the product documenta­tion or a power supply that conforms to the country-specific regulations.
Only connect the product to a power source with a fuse protection of maxi­mum 20 A.
Ensure that you can disconnect the product from the power source at any time. Pull the power plug to disconnect the product. The power plug must be easily accessible. If the product is integrated into a system that does not meet
11Getting Started 1316.4697.02 ─ 07
R&S®ESRP
these requirements, provide an easily accessible circuit breaker at the system level.
Using laser products
Lasers are classified according to their potential risk. If exposure to the laser beam is possible, the product is labeled as shown in the table explaining the safety labels.
Class 1M lasers
Safe for the naked eye. If you look into the laser beam with optical instruments such as binoculars or eye loupes, you risk damaging your eyes.
Class 2 lasers
If you stare into the beam, you risk damaging your eyes.
Safety and Regulatory Information
Safety Instructions
Handling batteries safely
The product contains exchangeable or built-in lithium polymer or lithium ion cells or batteries. The use of the word battery in the following always means all types. Only the battery contents are potentially hazardous. As long as a battery is undamaged and the seals remain intact, there is no danger.
Impact, shock or heat can cause damage such as dents, punctures and other deformations. A damaged battery poses a risk of personal injury. Handle a dam­aged or leaking battery with extreme care. Immediately ventilate the area since the battery releases harmful gases. If you come into contact with the battery fluid, immediately remove all contaminated clothing. Irritation can occur if the battery fluid comes in contact with your skin or eyes. Immediately and thoroughly rinse your skin or eyes with water and seek medical aid.
For safe handling, follow these rules:
Do not short-circuit the battery.
Do not mechanically damage the battery. Do not open or disassemble the bat­tery.
Do not expose the battery to high temperatures such as open flames, hot sur­faces and sunlight.
Only use the battery with the designated Rohde & Schwarz product.
12Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Safety and Regulatory Information
Safety Instructions
Only use the appropriate Rohde & Schwarz charger to charge the batteries. If the batteries are improperly charged, there is a risk of explosion. For charging and discharging temperature ranges, see the product documentation.
Replace exchangeable batteries only with the same battery type.
Store the battery in the product or use the product packaging.
Dispose of exchangeable batteries separately from normal household waste as specified by the local waste disposal agency.
If you disregard these rules, you risk serious personal injury or even death due to explosion, fire or hazardous chemical substances. The product documentation provides further details.
If exchangeable batteries or products with built-in batteries are defective, contact the Rohde & Schwarz customer service. Rohde & Schwarz classifies the severity of the defect. When returning batteries or Rohde & Schwarz products containing batteries, use a carrier qualified to transport dangerous goods and notify the car­rier of this classification. Follow the carrier’s transport stipulations in line with IATA-DGR, IMDG-Code, ADR or RID.
Connecting headphones
Take the following measures to prevent hearing damage. Before using head­phones, check the volume and reduce it if necessary. If you monitor varying sig­nal levels, take off the headphones and wait until the signal has settled. Then adjust the volume.
Cleaning the product
Use a dry, lint-free cloth to clean the product. When cleaning, keep in mind that the casing is not waterproof. Do not use liquid cleaning agents.
Meaning of safety labels
Safety labels on the product warn against potential hazards.
Potential hazard Read the product documentation to avoid personal injury or product damage.
Heavy product Be careful when lifting, moving or carrying the product. Carrying the product requires
a sufficient number of persons or transport equipment.
13Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Electrical hazard Indicates live parts. Risk of electric shock, fire, personal injury or even death.
Hot surface Do not touch. Risk of skin burns. Risk of fire.
Protective conductor terminal Connect this terminal to a grounded external conductor or to protective ground. This
connection protects you against electric shock if an electric problem occurs.
Warning: laser beam The product contains a laser.
Avoid exposure to direct or reflected laser beam.
Safety and Regulatory Information
Warning Messages in the Documentation

1.2 Warning Messages in the Documentation

A warning message points out a risk or danger that you need to be aware of. The signal word indicates the severity of the safety hazard and how likely it will occur if you do not follow the safety precautions.
WARNING
Potentially hazardous situation. Could result in death or serious injury if not avoi­ded.
CAUTION
Potentially hazardous situation. Could result in minor or moderate injury if not avoided.
NOTICE
Potential risks of damage. Could result in damage to the supported product or to other property.
14Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Safety and Regulatory Information
Korea certification class A

1.3 Korea certification class A

이 기기는 업무용(A급) 전자파 적합기기로서 판매자 또는 사용자는 이 점을 주의하 시기 바라며, 가정외의 지역에서 사용하는 것을 목적으로 합니다.
15Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Documentation Overview
Service Manual

2 Documentation Overview

This section provides an overview of the R&S ESRP user documentation. Unless specified otherwise, you find the documents on the R&S ESRP product page at:
www.rohde-schwarz.com/manual/esrp

2.1 Getting Started Manual

Introduces the R&S ESRP and describes how to set up and start working with the product. Includes basic operations, typical measurement examples, and general information, e.g. safety instructions, etc.
A printed version is delivered with the instrument. A PDF version is available for download on the Internet.

2.2 User Manuals and Help

Contains the description of all instrument modes and functions. It also provides an introduction to remote control, a complete description of the remote control commands with programming examples, and information on maintenance, instru­ment interfaces and error messages. Includes the contents of the getting started manual.
The contents of the user manuals are available as help in the R&S ESRP. The help offers quick, context-sensitive access to the complete information.
All user manuals are also available for download or for immediate display on the Internet.

2.3 Service Manual

Describes the performance test for checking the rated specifications, module replacement and repair, firmware update, troubleshooting and fault elimination, and contains mechanical drawings and spare part lists.
16Getting Started 1316.4697.02 ─ 07
R&S®ESRP

Release Notes and Open Source Acknowledgment (OSA)

The service manual is available for download for registered users on the global Rohde & Schwarz information system (GLORIS):
https://gloris.rohde-schwarz.com).
Documentation Overview

2.4 Instrument Security Procedures

Deals with security issues when working with the R&S ESRP in secure areas. It is available for download on the Internet.

2.5 Basic Safety Instructions

Contains safety instructions, operating conditions and further important informa­tion. The printed document is delivered with the instrument.

2.6 Data Sheets and Brochures

The data sheet contains the technical specifications of the R&S ESRP. It also lists the options and their order numbers as well as optional accessories.
The brochure provides an overview of the instrument and deals with the specific characteristics.
www.rohde-schwarz.com/brochure-datasheet/esrp
2.7 Release Notes and Open Source Acknowledg­ment (OSA)
The release notes list new features, improvements and known issues of the cur­rent firmware version, and describe the firmware installation.
The open source acknowledgment document provides verbatim license texts of the used open source software.
17Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Application Notes, Application Cards, White Papers, etc.
www.rohde-schwarz.com/firmware/esrp
Documentation Overview

2.8 Application Notes, Application Cards, White Papers, etc.

These documents deal with special applications or background information on particular topics.
www.rohde-schwarz.com/application/esrp
18Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Conventions Used in the Documentation
Conventions for Procedure Descriptions

3 Conventions Used in the Documentation

3.1 Typographical Conventions

The following text markers are used throughout this documentation:
Convention Description
"Graphical user interface elements"
[Keys] Key and knob names are enclosed by square brackets.
Filenames, commands, program code
Input Input to be entered by the user is displayed in italics.
Links Links that you can click are displayed in blue font.
"References" References to other parts of the documentation are enclosed by
All names of graphical user interface elements on the screen, such as dialog boxes, menus, options, buttons, and softkeys are enclosed by quotation marks.
Filenames, commands, coding samples and screen output are distinguished by their font.
quotation marks.

3.2 Conventions for Procedure Descriptions

When operating the instrument, several alternative methods may be available to perform the same task. In this case, the procedure using the touchscreen is described. Any elements that can be activated by touching can also be clicked using an additionally connected mouse. The alternative procedure using the keys on the instrument or the on-screen keyboard is only described if it deviates from the standard operating procedures.
The term "select" may refer to any of the described methods, i.e. using a finger on the touchscreen, a mouse pointer in the display, or a key on the instrument or on a keyboard.
19Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Conventions Used in the Documentation
Notes on Screenshots

3.3 Notes on Screenshots

When describing the functions of the product, we use sample screenshots. These screenshots are meant to illustrate as many as possible of the provided functions and possible interdependencies between parameters. The shown values may not represent realistic usage scenarios.
The screenshots usually show a fully equipped product, that is: with all options installed. Thus, some functions shown in the screenshots may not be available in your particular product configuration.
20Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Instrument Tour
The Front Panel

4 Instrument Tour

On the instrument tour, you can learn about the different control elements and connectors on the front and back panel of the R&S ESRP.

4.1 The Front Panel

The front panel of the R&S ESRP is shown in Figure 4-1. Each element (function keys and connectors) is described in more detail in the subsequent sections.
Figure 4-1: Front panel of the R&S ESRP
1 = Function keys 2 = USB interface 3 = Power button 4 = Display 5 = Access to operating system and online keyboard 6 = Display options 7 = Navigation options for menus 8 = Measurement configuration 9 = Marker functions 10 = Measurement control 11 = Measurement start 12 = Data entry keys 13 = Rotary knob 14 = Navigation keys
21Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Instrument Tour
The Front Panel
15 = Undo / redo function 16 = AF output incl. volume control 17 = Noise source control 18 = Supply voltage for accessories 19 = Supply voltage for accessories 20 = Tracking generator output 21 = RF input 2 22 = RF input 1

4.1.1 Touchscreen Display

The touchscreen on the front panel of the R&S ESRP displays the measurement results. Additionally, the screen display provides status and setting information and allows you to switch between various measurement tasks. The screen is touch-sensitive, offering an alternative means of user interaction for quick and easy handling of the instrument.
Figure 4-2: Touchscreen elements
22Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Instrument Tour
The Front Panel
1 = Toolbar with standard application functions, e.g. print, save/open file etc. 2 = Tabs for individual measurement tasks 3 = Channel information bar for current measurement settings 4 = Diagram header with diagram-specific (trace) information 5 = Measurement results area 6 = Diagram footer with diagram-specific information, depending on measurement mode 7 = Error indicator 8 = Error message, if available 9 = Device status 10 = Progress bar for measurement 11 = Date and time display 12 = Softkeys for menu access
A touchscreen is a screen that is touch-sensitive, i.e. it reacts in a specified way when a particular element on the screen is tapped by a finger or a pointing device, for example. Any user interface elements that can be clicked on by a mouse pointer can also be tapped on the screen to trigger the same behavior, and vice versa.
Using the touchscreen, the following tasks (among others) can be performed by the tap of your finger:
Changing a setting
Changing the display
Moving a marker
Selecting a new evaluation method
Scrolling through a result list
Saving or printing results and settings

4.1.2 Function Keys on the Front Panel

A detailed description of the corresponding menus and the other function keys is provided in chapter 6 "Instrument Functions" of the Operating Manual.
Table 4-1: Function keys
Turns the instrument on and off.
Restores the default configuration of the R&S ESRP.
Provides functionality to save, restore and manage instrument settings and other files.
23Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Instrument Tour
The Front Panel
Provides functionality for general instrument configuration, for example:
to configure the display
to set the date and time
to establish a LAN connection
Provides functionality to configure printers and hardcopys.
Displays the online help.
Selects the operating mode or firmware application.
Opens the Windows "Start" menu
Turns the on-screen keyboard on and off. Repeatedly pressing the key changes the position of the keyboard (top or bottom of the display).
Selects alphanumeric characters.
Opens a dialog box to turn screen elements on or off.
Switches between maximized and split display of focus area.
Switches focus area between table and diagram.
Allows you to define and use softkeys to load custom configurations.
Opens the root menu of the current application.
FREQ (CHANNEL) Provides functionality to define frequency parameters, for example:
the center or receiver frequency
the frequency offset
(CHANNEL is for special applications)
24Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Instrument Tour
The Front Panel
SPAN Provides functionality to configure the frequency span.
AMPT (SCALE) Provides functionality to configure amplitude or level characteristics, for
example:
the reference level
the attenuation
the input impedance
the scale of the level axis
the preamplifier
AUTO SET Provides functionality to automatically define various parameters like the
level or frequency.
BW Provides functionality to define the resolution and video bandwidth.
SWEEP Provides functionality to configure the sweep, for example:
the measurement time
the number of measurement points
the measurement mode (single or continuous measurements)
TRACE Provides functionality to configure data acquisition and analyze mea-
sured data, for example:
the trace mode
the detector
TRIG Provides functionality to configure triggered and gated measurements,
for example:
MKR Provides functionality to activate and position absolute and relative mark-
ers (markers and delta markers).
PEAK SEARCH Performs a peak search for active markers.
If no marker is active, Marker 1 is activated and the peak search is per­formed for it.
MKR FUNC Provides additional analysis functions of the measurement markers, for
example:
the frequency counter
the noise measurement
the phase noise measurement
the AM/FM audio demodulator
MKR Provides functionality to position and control markers, for example:
to position the marker on the center frequency
to define the marker search area
to configure the peak excursion
25Getting Started 1316.4697.02 ─ 07
R&S®ESRP
MEAS Provides the measurement functions, for example:
the bargraph measurement (in Receiver mode)
the scan and final measurement (in Receiver mode)
the AF demodulation (in Receiver mode)
IF analysis (in Receiver mode with option R&S ESRP-K56)
the channel power and ACLR measurement (in Spectrum mode)
the occupied bandwidth measurement (in Spectrum mode)
the Spectrum Emission Mask (SEM) measurement (in spectrum mode)
the spurious emission measurement (in spectrum mode)
the signal statistics (in spectrum mode)
MEAS CONFIG Provides functionality to configure the measurement.
LINES Provides functionality to control display and limit lines.
INPUT/OUTPUT Provides functionality to configure inputs and outputs.
RUN SINGLE Starts a single measurement.
RUN CONT Starts a continuous measurement.
Instrument Tour
The Front Panel
UNDO Reverts the last operation.
REDO Repeats previously reverted operation.

4.1.3 USB Interface

The front panel provides two female USB connectors to connect devices like an external keyboard or mouse. You can also connect a memory stick to save and restore instrument settings and measurement data.

4.1.4 Noise Source Control

The noise source control female connector is used to provide the supply voltage for an external noise source, for example to measure the noise figure and gain of amplifiers and frequency converting DUTs.
Conventional noise sources require a voltage of +28 V in order to be switched on and 0 V to be switched off. The output supports a maximum load of 100 mA.

4.1.5 RF Input 50

The R&S ESRP provides an RF input for connection of a device under test (DUT) to the R&S ESRP. The DUT is connected to the RF Input via cable and an appro­riate connector (for example a male N connector).
26Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Instrument Tour
The Front Panel
The RF Input supports a frequency range from 9 kHz to f range from 0 dB to 70 dB.
With option R&S R&S ESRP-B29, the minimum frequency is extended to 10 Hz.
and an attenuation
max

4.1.6 Probe Power

The R&S ESRP provides two connectors to supply accessories like transducers or probes with power.
The probe power connector with five pins supports supply voltages of ±10 V and ground. The maximum permissible current is 200 mA. This probe power connec­tor is suitable, for example, for transducers from Rohde & Schwarz.
The probe power connector with three pins supports supply voltages from +15 V to -12.6 V and ground. The maximum permissible current is 150 mA. This probe power connector is suitable, for example, for high-impedance probes from Agi­lent.

4.1.7 AF Output

Headphones equipped with a miniature jack plug can be connected to the AF out­put female connector. The internal impedance is 10 Ω. The output voltage can be set by using the volume control to the right of the female connector. If a plug is connected, the internal loudspeaker is automatically switched off.
See also "Connecting headphones" on page 13.

4.1.8 Tracking Generator Output (Optional)

The R&S ESRP is equipped with an optional tracking generator and therefore provides a tracking generator output connector.
The output of the tracking generator is connected to the DUT via a cable equip­ped with a male N connector. The female connector is available only with the tracking generator option (R&S FSV-B9).
27Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Instrument Tour
Rear Panel View
Sensitive DUTs concerning matching
For DUTs with sensitive RF characteristics with regard to matching (VSWR) at the input, insert a 10 dB attenuator between the DUT and the tracking generator.

4.2 Rear Panel View

The rear panel of the R&S ESRP is shown in Figure 4-3. Each element is descri­bed in more detail in the subsequent sections.
Figure 4-3: R&S ESRP rear panel
1 = LAN interface 2 = Trigger output 3 = IF / Video connector 4 = USB interface 5 = AUX port 6 = External trigger / gate input 7 = VGA interface 8 = Reference in 9 = Reference out 10 = GPIB interface 11 = AC power supply and main switch
28Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Instrument Tour
Rear Panel View

4.2.1 AC Power Supply Connection and Main Power Switch

The AC power supply connector and main power switch are located in a unit on the rear panel of the instrument.
The main power switch works as follows.
Position "I": The instrument is supplied with power and in operation. The OCXO reference frequency is warmed up.
Position "O": The instrument is disconnected from the AC power supply.
For more information see Chapter 5.7, "Connecting the AC Power", on page 36.

4.2.2 LAN

The LAN interface can be used to connect the R&S ESRP to a local network for remote control, printouts and data transfer. The assignment of the RJ-45 connec­tor supports twisted-pair category 5 UTP/STP cables in a star configuration (UTP stands for unshielded twisted pair, and STP for shielded twisted pair).

4.2.3 VGA Interface

The female VGA connector is used to connect an external monitor. Instructions on how to connect an external monitor are provided in Chapter 5.12, "Connecting
an External Monitor", on page 41.

4.2.4 Ext Trigger / Gate In

The female connector for external trigger or gate input is used to control the mea­surement by means of an external signal. The voltage levels can range from 0.5 to 3.5 V. The default value is 1.4 V. The typical input impedance is 10 kΩ.

4.2.5 Ref In

As a reference signal, you can either use the internal reference, or connect an external one. The setup menu is used to switch between the internal and an external reference. The REF IN female connector is used as an input for a 1-20 MHz reference signal. The required input level is 0-10 dBm.
29Getting Started 1316.4697.02 ─ 07
R&S®ESRP
Instrument Tour
Rear Panel View

4.2.6 Ref Out

This connector can be used to provide an external reference signal (e.g. the OCXO) to other devices that are connected to this instrument. The REF OUT female connector can output a 10 MHz reference signal with an output level of 0 dBm.

4.2.7 GPIB Interface

The GPIB interface is in compliance with IEEE488 and SCPI. A computer for remote control can be connected via this interface. To set up the connection, a shielded cable is recommended.
For more information, refer to the user manual.

4.2.8 Trigger Output

The female BNC connector can be used to provide a signal to another device. The signal is TTL compatible (0 V / 5 V).
You can control the trigger output with the functionality provided in the "In-/ Output" menu ([INPUT/OUTPUT] key).
The trigger output also controls signals by the frequency mask trigger available in realtime mode (R&S ESR-K55).

4.2.9 IF / Video

The female BNC connector can be used for various outputs:
Intermediate frequency (IF) output of 32 MHz
Video output (1 V)
You can select between IF and video output with the functionality provided in the "In-/Output" menu ([INPUT/OUTPUT] key).
30Getting Started 1316.4697.02 ─ 07
Loading...