Rockwell Automation 1441-DYN25-MBAL User Manual

User Manual
Balancing Extension Module for the Dynamix 2500 Data Collector
Catalog Number
1441-DYN25-MBAL
Important User Information
IMPORTANT
Solid-state equipment has operational characteristics differing from those of electromechanical equipment. Safety Guidelines for the Application, Installation and Maintenance of Solid State Controls (publication SGI-1.1 your local Rockwell Automation sales office or online at http://www.rockwellautomation.com/literature/ important differences between solid-state equipment and hard-wired electromechanical devices. Because of this difference, and also because of the wide variety of uses for solid-state equipment, all persons responsible for applying this equipment must satisfy themselves that each intended application of this equipment is acceptable.
In no event will Rockwell Automation, Inc. be responsible or liable for indirect or consequential damages resulting from the use or application of this equipment.
The examples and diagrams in this manual are included solely for illustrative purposes. Because of the many variables and requirements associated with any particular installation, Rockwell Automation, Inc. cannot assume responsibility or liability for actual use based on the examples and diagrams.
No patent liability is assumed by Rockwell Automation, Inc. with respect to use of information, circuits, equipment, or software described in this manual.
Reproduction of the contents of this manual, in whole or in part, without written permission of Rockwell Automation, Inc., is prohibited.
Throughout this manual, when necessary, we use notes to make you aware of safety considerations.
WARNING: Identifies information about practices or circumstances that can cause an explosion in a hazardous environment, which may lead to personal injury or death, property damage, or economic loss.
) describes some
ATTENTION: Identifies information about practices or circumstances that can lead to personal injury or death, property damage, or economic loss. Attentions help you identify a hazard, avoid a hazard, and recognize the consequence
SHOCK HAZARD: Labels may be on or inside the equipment, for example, a drive or motor, to alert people that dangerous voltage may be present.
BURN HAZARD: Labels may be on or inside the equipment, for example, a drive or motor, to alert people that surfaces may reach dangerous temperatures.
Identifies information that is critical for successful application and understanding of the product.
Allen-Bradley, Rockwell Software, Rockwell Automation, Dynamix, Enpac, Emonitor, and TechConnect are trademarks of Rockwell Automation, Inc.
Trademarks not belonging to Rockwell Automation are property of their respective companies.

Table of Contents

Preface
Installing Optional Extension Modules
Balancing Extension Module
Index
Optional Extension Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Additional Resources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Chapter 1
Install Extension Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Uninstall Extension Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Manage Extension Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Extension Module Battery Status Indicators . . . . . . . . . . . . . . . . . . . . . . . . 13
Chapter 2
Balancing Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Plane Balancing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Static-couple Balancing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Set Up the Balancing Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Balance Setup Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Single Plane Balancing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Single Plane Balancing Procedure Overview. . . . . . . . . . . . . . . . . . . . . 26
Take the Initial Vibration Measurement. . . . . . . . . . . . . . . . . . . . . . . . 27
Add the Trial Weight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Enter the Trial Weight Manually . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Estimate the Trial Weight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Take a Trial Run Weight Measurement . . . . . . . . . . . . . . . . . . . . . . . . 32
Take a Correction Run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Perform a Trim Run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Add the Correction Weight and Take a Residual Measurement . . 41
Two Plane Balancing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Two Plane Balancing Procedure Overview. . . . . . . . . . . . . . . . . . . . . . 46
How to Set Up a Two-Plane Procedure . . . . . . . . . . . . . . . . . . . . . . . . 46
How to Perform Two-plane Balancing . . . . . . . . . . . . . . . . . . . . . . . . . 50
Move Around in a Balance Run. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Saving, Loading, and Reviewing Balance Measurements . . . . . . . . . . . . . 61
Save a Balance Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Load a Previously Saved Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Review Balancing Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
How to Delete Stored Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Rockwell Automation Publication 1441-UM004A-EN-P - May 2011 3
Table of Contents
Notes:
4 Rockwell Automation Publication 1441-UM004A-EN-P - May 2011

Preface

This manual describes the Balancing extension module for the Dynamix 2500 data collector. You install the extension module with the Balancing Secure Digital (SD) card.
Optional Extension Modules
See Installing Optional Extension Modules on page 7 instructions.
The balancing test determines the amount and location of the heavy spot on a rotating shaft so that you can balance it with an equal amount of weight in the opposite direction.
There are three types of measurements in the balancing process:
Initial Vibration
Tri al Wei gh t
Residual Measurements
These are the optional extension modules for the Dynamix 2500 data collector:
1441-DYN25-4C, 4-channel Activation The 4-channel activation lets you take 3- and 4-channel magnitude, time
waveform, spectra, and Offroute measurements.
1441-DYN25-MBMP Bump Test A bump test (or hammer test) determines the natural frequencies of a
machine or a structure.
1441-DYN25-MBAL Balancing
for installation
(1)
Balancing application resolves single-plane, two-plane, and static-couple balances with high precision.
1441-DYN25-MFRF Frequency Response Function The FRF test lets you determine the natural frequencies of a machine as
well as sophisticated information about the frequency response of the structure being tested.
1441-DYN25-MREC Time Recorder The Time Recorder test uses a the instrument as a data recorder for real-
time data acquisition and analysis.
1441-DYN25-MRUC Run Up Coast Down The RUCD test records and analyzes data from intermittent events and
transient vibration signals from non-steady state machines.
See Additional Resources on page 6
(1) This is an activation license for the Dynamix 2500 data collector.
Rockwell Automation Publication 1441-UM004A-EN-P - May 2011 5
for a listing of available publications.
Preface
Additional Resources
These documents contain additional information concerning related products from Rockwell Automation.
Resource Description
Dynamix 2500 Data Collector User Manual, publication 1441-UM001
Bump Test Extension Module for the Dynamix 2500 Data Collector User Manual, publication 1441-UM002
Frequency Response Function Extension Module for the Dynamix 2500 Data Collector User Manual, publication 1441-UM003
Time Recorder Extension Module for the Dynamix 2500 Data Collector User Manual, publication 1441-UM005
Run Up Coast Down Extension Module for the Dynamix 2500 Data Collector User Manual, publication 1441-UM006
Emonitor User’s Guide, publication
EMONTR-UM001
Dynamix 2500 Data Collector Kit Release Notes, publication 1441-RN001
Dynamix 2500 Data Collector Optional Extension Modules Release Notes, publication 1441-RN002
Industrial Automation Wiring and Grounding Guidelines, publication 1770-4.1
Product Certifications website, http://www.ab.com
Describes the Dynamix 2500 data collector which provides predictive maintenance by using noise and vibration analysis.
Describes how determine natural (or resonant) frequencies of a machine or structure.
Describes how to determine the natural frequencies of a machine or structure by using modal hammer.
Describes how to use the data collector as a data recorder for real-time data acquisition, post processing, and analysis.
Describes how to record and analyze data from intermittent events and transient vibration signals from non-steady state machines.
Describes data management for predictive maintenance services.
Provides information on the latest updates, for example, firmware, certifications, warnings, and hardware changes for the data collector.
Provides information on how to install the Optional Extension Modules on to the Dynamix 2500 data collector.
Provides general guidelines for installing a Rockwell Automation industrial system.
Provides declarations of conformity, certificates, and other certification details.
You can view or download publications at
http://www.rockwellautomation.com/literature
technical documentation, contact your local Allen-Bradley distributor or Rockwell Automation sales representative.
6 Rockwell Automation Publication 1441-UM004A-EN-P - May 2011
. To order paper copies of
Chapter
IMPORTANT
IMPORTANT
1

Installing Optional Extension Modules

The data collector uses the Extension Manager to install and uninstall extension modules. These extension modules are licensed and ordered separately from the basic entry-level product.
Topic Page
Install Extension Modules 7 Uninstall Extension Modules 9 Manage Extension Modules 12 Extension Module Battery Status Indicators 13
Install Extension Modules
The installation Secure Digital (SD) cards that you receive work with any Dynamix 2500 data collector. Once you have installed an extension module, the card is locked so that it can only be used with that instrument.
One installation SD card is required for each instrument that needs to be upgraded.
You can uninstall extension modules if required. When uninstalling an extension module, you have the option to free up the license so you can install the extension module on another instrument. This makes the extension module available to be transferred between units.
When ever you re-run the OS Loader software, you will re-load only the main OS firmware. The OS loader will backup licence files and data, but not the optional extension modules. Once you have updated the OS firmware, insatll the latest version of your optional extension modules.
See the Dynamix 2500 Data Collector User Manual, 1441-UM001 more information.
Follow these instructions to install an extension module.
1. Open the base cover at the bottom of the Dynamix 2500 data collector.
, for
2. Place the extension module SD card contact side-up into the unit until it is
firmly seated in place.
3. Close the base cover.
Rockwell Automation Publication 1441-UM004A-EN-P - May 2011 7
Chapter 3 Installing Optional Extension Modules
Extension Manager
4. Apply power to the data collector.
5. From the Main Menu, select Setup Utility and press Enter.
6. Press 0 (Shift) to display the second set of functions.
The Extension Manager function remains on the screen for about three seconds after releasing 0 (Shift).
7. Press F1 (Extn Mgr).
The Extension Manager screen appears showing the current extension module installations.
8 Rockwell Automation Publication 1441-UM004A-EN-P - May 2011
8. Press 0 (Shift) to display the Install Extension function.
9. Press F2 (Install) to install the new extension module.
Installing Optional Extension Modules Chapter 3
Extension Manager
When the installation is complete, a confirmation prompt appears.
10. Press F4 (OK).
The new extension module appears in the list.
Uninstall Extension Modules
11. Press F4 (Esc) to exit the Extension Manager screen.
Follow these instructions to uninstall an extension module.
1. Press 0 (Shift) from the Setup Utility screen to display the Extension
Manager function.
The Extension Manager function remains on the screen for about three seconds after releasing 0 (Shift).
2. Press F1 (Extn Mgr).
Rockwell Automation Publication 1441-UM004A-EN-P - May 2011 9
Chapter 3 Installing Optional Extension Modules
This screen lists the extension modules currently installed and the on the unit.
The Extension Manager screen appears.
3. Select the extension module you want to uninstall and press F3 (Select).
F3 (Select) toggles the selection on and off.
A checkmark appears next to that extension module.
4. Press F1 (Uninstall).
10 Rockwell Automation Publication 1441-UM004A-EN-P - May 2011
Installing Optional Extension Modules Chapter 3
IMPORTANT
A confirmation message appears.
5. Make sure your installation card is inserted into the instrument.
The extension module is uninstalled and the license on the card is released so that the card can be used to install the extension module on another Dynamix 2500 data collector.
If the installation card is not inserted in the instrument and no card, you are is found or the card does not have the extension module license for the unit, you are prompted to insert the correct installation card or continue without freeing the license.
6. Press F2 (Yes) to uninstall the extension module.
7. Press F4 (Esc) to exit the Extension Manager screen.
Rockwell Automation Publication 1441-UM004A-EN-P - May 2011 11
Chapter 3 Installing Optional Extension Modules
IMPORTANT
TIP
The extension modules that are hidden are shown in [square brackets].
Manage Extension Modules
The Dynamix 2500 data collector lets you to hide installed extension modules from the Main Menu. You may need to hide an advanced extension module icon from an inexperienced user, for example, RUCD and FRF.
Once an extension module is hidden, its icon is not represented on the Main Menu or displayed in the Dynamix 2500 data collector About screen.
Follow these instructions to hide or show an extension module icon on the Main Menu.
1. Press 0 (Shift) on the Setup Utility screen to display the Extension
Manager function.
The Extension Manager function should remain on the screen for approximately three seconds after releasing 0 (Shift).
2. Press F1 (Extn Mgr). The Extension Manager screen appears.
The Extension Manager displays a list of installed extensions module.
12 Rockwell Automation Publication 1441-UM004A-EN-P - May 2011
The F2 (Hide) toggles between Hide and Show depending if the selected extension module is hidden or not.
You have to exit and re-enter the Extension Manager after hiding an extension module to have the Show function appear.
Installing Optional Extension Modules Chapter 3
TIP
3. Select the extension module that you want to hide or show and press F2
(Show/Hide).
If you want to show or hide multiple extension modules simultaneously, select each extension module and press F3 (Select).
A checkbox appears next to each selected extension module. If you select multiple extension modules and some are hidden while others are shown, F2 (Show/Hide) reflects the status of the selected extension module.
4. Press F4 (Esc) to exit the Extension Manager.
Extension Module Battery Status Indicators
The battery status icons show the strength of the battery.
Table 1 - Extension Module Battery Icon Descriptions
Battery Icon Meaning
Battery status is good: >30% life remaining.
Battery status is low: >10% life remaining.
Battery status is very low: <10% life remaining.
Battery is charging.
Rockwell Automation Publication 1441-UM004A-EN-P - May 2011 13
Chapter 3 Installing Optional Extension Modules
Notes:
14 Rockwell Automation Publication 1441-UM004A-EN-P - May 2011
Chapter
2

Balancing Extension Module

The Balancing extension module is an optional module for the Dynamix 2500 data collector. You install the extension module with the Balancing Secure Digital (SD) storage card.
Topic Page
Balancing Measurements 16 Plane Balancing 16 Static-couple Balancing 18 Set Up the Balancing Parameters 20 Single Plane Balancing 26 Two Plane Balancing 45 Saving, Loading, and Reviewing Balance Measurements 61
Balancing application resolves single-plane, two-plane, and static-couple balances with high precision. You can use the Dynamix 2500 data collector coupled with a speed-measuring device such as a strobe light or laser tachometer.
The Dynamix 2500 data collector Balancing extension module provides a simple, direct method to balance your rotating machinery in one or two planes. You can use the internal laser tachometer in the data collector or an external laser tachometer, optical tachometer, or strobescope for measuring phase during the balance run.
Rockwell Automation Publication 1441-UM004A-EN-P - May 2011 15
Chapter 2 Balancing Extension Module
Balancing Measurements
Balancing is the technique for determining the amount and location of the heavy spot on a rotating shaft so that you can balance it with an equal amount of weight in the opposite direction. These methods go through the technique with stopping points where you start and stop the machine to perform weight addition or subtraction.
There are three types of measurements in the balancing process:
Initial Vibration
The initial vibration measurement is taken without any added weights on the machine. The initial vibration measurement establishes a reference of how the machine vibrates at each plane without any extra weight added. This vibration is what will be corrected by the correction weight.
Tri al Wei gh t
The trial weight measurements are taken with a single trial weight attached to the machine at one plane or the other. The trial weight measurements are used to determine how the machine is affected by the added weight. The ideal trial weight should produce either a 30 % change in amplitude or a 30
° change in phase.
Residual Measurements
Plane Balancing
The residual measurements are taken with the correction weight or the correction weight and trim weights attached to the machine. The correction weight should cancel out the initial unbalance. A residual vibration measurement is taken to measure the remaining unbalance. Trim weights are added to the machine to cancel out the vibration measured during a residual measurement.
Regardless of whether you are performing a single or two-plane balancing procedure, all balancing procedures progress through basic runs as you start (spin) and stop the rotor.
For two-plane procedures, you perform some of the runs twice, once with weights on correction plane one and again with weights on correction plane two. For simplicity, the following example describes a single-plane procedure.
These are the steps you need to take to take a single-plane balancing measurement.
1. Set up balancing equipment and the Balancing parameters.
With the machine stopped, set up the balancing equipment and mark your tachometer reference point on the rotor or shaft.
See Set Up the Balancing Parameters on page 20 measurement parameters for the balancing run sequence.
16 Rockwell Automation Publication 1441-UM004A-EN-P - May 2011
to configure balancing
Balancing Extension Module Chapter 2
2. Perform the initial measurement.
a. Start the machine. b. Take the initial measurement.
The initial measurement provides a starting place for balancing computations. It records the machine's initial imbalance, 1X vibration magnitude and phase angle. Later in the procedure, initial measurement data is automatically compared with trial (weight) measurement data to calibrate the machine's imbalance.
3. Attach a trial weight.
a. Stop the machine after collecting the initial measurement data. b. Add a trial weight.
A trial weight is a temporary weight attached to produce a change from the original unbalance readings (reference run readings). Input the trial weight amount and placement angle into the data collector.
4. Perform the trial measurement.
a. Start the machine again with the trial weight securely attached. b. Take trial measurement with the trial weight.
The trial measurement calibrates the machine's unbalance and allows the data collector to calculate the influence coefficient and permanent correction weight amount and placement. To allow for accurate balancing computations, the trial weight should pass the 30/30 rule.
The trial weight should do the following:
Increase or decrease the 1X vibration amplitude by at least 30 %
Change the phase angle by at least 30
°
Be a combination of A and B.
5. Attach the correction weight.
a. Stop the machine. b. Remove the trial weight c. Attach the specified correction weight at the specified angle.
6. Perform a correction weight measurement.
a. Start the machine again. b. Take correction weight measurement.
The correction weight run performs two functions:
It calculates the amount of residual imbalance with the correction weight in place to confirm that the machine is now balanced within tolerances.
If further balancing is necessary, it automatically applies the influence coefficient to calculate additional trim weights that can be added to further balance the machine.
Rockwell Automation Publication 1441-UM004A-EN-P - May 2011 17
Chapter 2 Balancing Extension Module
7. If additional trim weights are necessary to balance the machine within
tolerances, stop the machine and attach the specified trim weights. a. Stop the machine. b. Attach the specified trim weights. c. Take another measurement.
8. Perform a trim measurement.
a. Start the machine. b. Take a trim measurement to verify the machine is balanced within
tolerances. If not, you can take additional trim measurements.
Static-couple Balancing
With stiffer rotors, a two-plane procedure may reduce the amount of imbalance to target tolerances. However, with narrow or flexible rotors, more often a static-couple procedure works best.
This graphic shows a narrow rotor showing three heavy spots.
The static-couple procedure corrects the static unbalance with weights added to the center of gravity plane, and corrects the couple unbalance with weights added to the end planes.
In practice, a static-couple balancing procedure is almost identical to a two-plane balancing procedure, except that it provides three correction weight solutions instead of two. One correction weight for each plane, and an additional correction weight to correct the static unbalance.
Unless you are able to attach weights to the machine’s center of gravity plane, the static weight is typically split in half and applied to the end planes to correct the static unbalance.
In as such, on the data collector, the static-couple procedure is identical to the two-plane procedure except the Correction Weight and Trim Weight screens show three correction weights for the static-couple procedure:
One weight for end plane 1
One weight for end plane 2
18 Rockwell Automation Publication 1441-UM004A-EN-P - May 2011
Balancing Extension Module Chapter 2
Except for the Solution field, all static-couple balancing parameters are identical to two plane balancing settings.
If you want to compute three correction weights, select Static-Couple for the Solution parameter on the Balance - Setup screen.
See Balance Extension Module Parameters on page 22
for more information
about these parameters.
Except for attaching three correction and trim weights instead of two (or four if you split the static weight), all static-couple balancing run procedures are identical to the two-plane balancing procedures.
SeeTwo Plane Balancing on page 45
on how to perform a two- plane Procedure
section for details.
Rockwell Automation Publication 1441-UM004A-EN-P - May 2011 19
Chapter 2 Balancing Extension Module
Set Up the Balancing Parameters
All balancing measurement parameters are set up from Balance Setup screen.
Follow these steps to go to the Balancing extension module.
1. From the Main Menu, select Balancing and press Enter.
The Balance menu displays.
These are the Balance menu options:
Setup
Access the Balance - Setup screen where you can set up and perform a new balancing run.
Recall
Access the Balance - Load Setup screen where you can select a previously saved balancing run and perform a new run by using the saved settings.
Review
Review Data screen where you can review previously stored measurements or resume a previous balance measurement from the last completed run.
20 Rockwell Automation Publication 1441-UM004A-EN-P - May 2011
Balancing Extension Module Chapter 2
Balance Setup Screen
After setting up your balance measuring equipment and marking your tachometer reference point, the next step is to configure balancing measurement parameters for the balancing run sequence.
Follow these steps to setup the Balancing parameters.
1. From the Balance main menu, select Setup and press Enter.
The Balance Setup lets you configure the balance measurement parameters.
2. Select an option and press the Right arrow to open a list choices.
3. Use the arrow keys to select a parameter.
4. Press the Left arrow to save your selection.
Use the descriptions in Balance Extension Module Parameters
to help you
configure the parameters in the Balance Setup screen.
5. Save the setup or start taking the balancing measurement.
Rockwell Automation Publication 1441-UM004A-EN-P - May 2011 21
Loading...
+ 47 hidden pages