RKI Instruments Beacon 200 User Manual

Beacon 200 Gas Monitor
Operator’s Manual
Part Number: 71-0059RK Revision: D Released: 1/15/14
www.rkiinstruments.com
RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of sh ipm ent from RKI Instr uments, Inc. Any part s fo und defective within that period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items, which by their nature, are subject to deterioration or consumption in normal service, and which must be clean ed, rep aired, or replace d on a rout ine basis. Exa mples of such items are as follows:
a) Absorbent cartridges d) Batteries b) Pump diaphragms and valves e) Filter elements c) Fuses
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability , and we are not responsible for removal or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC., INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC., BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors, de alers, and r epresenta tives as appo inted by R KI Instrument s, Inc.
We do not assume indemn ification for any accident or damage caused by the operation of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
ii • Product Warranty
Table of Contents
Chapter 1: Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
About the Beacon 200 Gas Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
About this Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Chapter 2: Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
External Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Internal Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Chapter 3: Installation and Start Up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Mounting the Beacon 200 Gas Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Wiring the Beacon 200 Gas Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Starting Up the Beacon 200 Gas Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Chapter 4: Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Recorder Output Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Alarm Indications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Viewing & Resetting Min/Max Readings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Chapter 5: Channel Control and Setup Program . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Enable/Disable Channel(s) Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Configure Channel Settings Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
View System Information Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Table of Contents • iii
Chapter 6: Input Setup Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Setting Up a New Channel or Changing an Existing Channel . . . . . . . . . . . . . . 41
Chapter 7: Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Calibration Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Replacing the Fuses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Parts List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
iv • Table of Contents
Chapter 1: Introduction
Overview
This chapter briefly describes the Beacon 200 Gas Monitor. This chapter also describes the Beacon 200 Gas Monitor Operator’s Manual (this document). Table 1 at th e en d o f thi s ch apter lists the specifications for the Beacon 200.
About the Beacon 200 Gas Monitor
The Beacon 200 is a fixed-mounted, continuous-monitoring gas detection instrument. This gas monito r is capable of detecting gas at up to two locations. The display screen simultaneously displays the gas readings of the active channel or channel s. Both dir ect connect ( internal a mplifier type) detector heads and 4 - 20 mA transmitter (remote amplifier type) detector heads may be used with the Beacon 200.
The Beacon 200 includes audible and visual alarms that warn you of hazardous gas conditions. The alarm circuit includes two levels of gas alarms. The fail circuit alerts you to failures in the gas detector heads or Beacon 200.
Three instrument programs allow you to display and change channel and calibration settings and change channel types. They are the Channel Control & Setu p Program and the Ca libration P rogram, and the Inp ut Setup Program.
Overview • 5
About this Manual
The Beacon 200 Gas Monitor Operator’s Manual is organized as follows:
Chapters 1 through 6 describe components of the Beacon 200 and procedures to install, start up, operate, and maintain the Beacon 200.
The Beacon 200 Gas Monitor Operator’s Manual uses the following conventions for notes, cautions, and warnings.
NOTE: Describes additional or critical information.
CAUTION: Describes potential dam a ge to equ ipm en t.
WARNING: Describes potential danger that can result in injury or
death.
I Caution: refer to accompanying documentation
!
~ Vac (AC voltage)
Vdc (DC voltage)
6 • About this Manual
Specifications
Table 1 lists specifica t ion s for the Beaco n 200 Gas Mo ni to r.
Description Specification
Table 1: Beacon 200 Specifications
Input Power
Construction (housing) Fiberglass/polyester with lexan window (NEMA 4X) Dimensions 10.5 in. H x 8.5 in. W x 6.25 in. D
Weight 8 lbs. Environmental Conditions
Safety/Regulatory
115/220V ~ ±10%, 50/60Hz, 0.5/0.3A or 24 V +10% -8%, 0.6A VDC
(267 mm H x 216 mm W x 158 mm D)
For indoor or outdoor locations (Type 4X)
2000m max. altitude
-20°C to 50°C (-4°F to 122°F) max. ambient
Maximum humidity of 80% relative
Mains supply voltage fluctuations not exceeding ± 10% of nominal
DC supply voltage fluctuations not exceeding +10% -8% of nominal
Overvoltage Category II, Pollution Degree 2
1
C US
186718
Relays
Relay contacts rated for 10A @ 115/220V~ resistive or10A @ 30V resistive
SPDT Form C
User Controls Reset switch
Program buttons: ESCAPE, UP/YES, DOWN/NO, and ENTER
Standard Accessory Operator’s manual (this document)
*1The AC voltage input is factory set. The standard input is 115 V~.
CAUTION: Cleaning is not to be performed.
CAUTION: The operator of this instrument is advised that if the equipment
is used in a man ner not specif ied in thi s manua l, the pr otectio n provided by th e equipment may be impaired.
Specifications • 7
Chapter 2: Description
Overview
This chapter describes external and internal components of the Bea con 200 Gas Monitor.
External Description
This section describes the housing and all external components of the Beacon 200. For the purposes of this description, the housing door is considered the front of the monitor.
Housing
The Beacon 200’s fiberglass hou sing is weath er- and corros ion-resist ant. It is suitable for installation where general purpose equipment is in use. The housing door is hing ed on the lef t side and is secure d by two latches on the right side. The displa y screen and st atus lig hts are visible thr ough win dows in the housing door. Four mounting feet are attached to the back of the housing (one at each corner). The mounting feet allow you to install the housing to a vertical surface. Three conduit hubs on the bottom of the housing are for external wiring connections.
CAUTION: To avoid electrical inte rference, do not ro ute detect or head and
power wiring through the same conduit hub.
Reset Switch
The reset switch i s on the bottom of th e ho usi ng . I t i s in front of the conduit hubs. The reset switch serves three functions:
Resets the alarm circuits for “latch ed” alarms after an alarm 1 or alarm 2 condition passes.
You can set each channel for latched or self-resetting alarms in the Channel Control & Setup Program.
8 • Overview
Silences the bu zzer during an alarm 1 or al a rm 2 con dit io n. You canno t silence failure alarms.
Displays and resets the minimum and maximum gas concentration values.
Buzzer
The buzzer is on the botto m of the housi ng. It is on the far r ight. The buzze r sounds audible alarms to warn you of gas alarms and in strument failures.
Internal Description
This section describes the internal components of the Beacon 200.
Main PCB
Display PCB
(see Figure 2)
Power
Supply
DC Fuse
Alarm Relays
(see Figure 3)
Detector/Transmitter
Terminal Strip, Channel 1
Detector/Tra n s mitter
Terminal Strip, Channel 2
3/4" Conduit
Hub, 3X
Power Switch
AC Fuses
AC In Terminal Strip
Controller Terminal Strip
Alarm Terminal Strip
Reset Switch
Buzzer
Figure 1. Beacon 200 Gas Monitor Component Location
Internal Description • 9
Display Printed Circuit Board (PCB)
The display PCB is mou nted to the p ower supply mo unting plate which is in turn mounted to t he ma in PCB. The power su pply mo unting p late an d main PCB are described below. The display PCB includes the displ ay, the status lights, and the program buttons.
Display
The display simultaneously indicates the channel number, current gas reading, measuring unit, and target gas of all active channels.
The display also indicates messages, settings, and other data when you are operating the instrument programs.
ALARM 2 Light
ALARM 1 Light
Display Contrast Adjust Pot
Display Screen
Display Cable Connector
FAIL Light
J1
PILOT Light
Figure 2. Display Board Component Location
ESCAPE Button
UP/YES Button
DOWN/NO Button
ENTER Button
Button Repeater Buzzer
10 • Internal Description
Status Lights
The Beacon 200 includes four status lights that indicate the current status of the monitor. The status lights are to the left and right of the display (see Figure 2).
Pilot Light. The pilot light is on when the Beacon 200 is receiving incoming power.
Fail Light. The fail light turns on when the Beacon 200 is experiencing a fail condition. A fail condition can be caus ed by a failure within the Beacon 200 or detector hea d( s) w i r ed to the Be aco n 20 0. See Chapter 5, Maintenanc e, or the troubleshooting section in the applicable detector head manual to respond to a fail condition.
Alarm 1 Light. The alarm 1 light is on when the Beacon 200 is experiencing an alarm 1 gas condition.
Alarm 2 Light. The alarm 2 light is on when the Beacon 200 is experiencing an alarm 2 gas condition.
Program Buttons
The Beacon 200 includes four program buttons that allow you to enter the instrument programs, navigate through the programs, update instrument and channel settings, and save changes to the program settings. When a program button is pressed, a buzze r lo cate d on the disp la y PCB beeps. The program buttons are near the right edge of the display PCB (see Figure 2).
Table 2: Beacon 200 Program Button Functions
Button Function
ESCAPE Moves backward through the program menus
Cancels changes you make in the program menus
Enters the Channel Control and Setup program (press with the ENTER button)
UP/YES Accepts the displayed setting and proceeds to the next setting
Changes the displayed setting
Enters the Calibration program (press with the ENTER button)
DOWN/NO Allows you to update the displayed setting
Changes the displayed setting
ENTER Saves changes you make in the programs
Enters the Channel Control and Setup program (press with ESCAPE button)
Enters the Calibration program (press with the UP/YES button)
Enters the Input Setup Program (press while flipping power switch to ON position)
Internal Description • 11
Main PCB
The main PCB is mounted inside the housing. The power supply mounting plate is mounted to the main PCB with four standoffs and the display PCB is mounted to the power supply mounting plate with four standoffs. The main PCB includes the terminal strips, relays, fuses, and power switch.
Terminal Strips
The Beacon 200 includes four terminal strips for external wiring connections. See “Wiring the Beacon 200 Gas Monitor” for detailed wiring procedures.
Detector/Transmitter Terminal Strips. Two detector/transmitter terminal strip s are locate d near the bo ttom lef t corne r of the mai n circuit board (see Figure 1). These two 9-point terminal strips facilitate wiring connections to the detectors or transmitters. Although each terminal strip can accommodate several different detector head models, only one detector h ead at a time ma y be wi red per channel . The top term inal strip is for channel 1 and the bottom terminal strip is for channel 2.
Alarm Terminal Strip. The alarm terminal strip is located along the bottom edge of the main circuit board (see Figure 1). This 21-point terminal strips facilitates wiring connections to external alarm devices (horn, strobe, etc.). Termin als are provided for individual channel as well as common alarm relay contacts.
Controller T e rminal Str ip. The 10-point controller terminal strip is near the lower right edge of the main circuit board (see Figure 1). The controller terminal strip facilitates various interna l and external wiring connections. Table 3 lists the function of each terminal.
Table 3: Terminal Assignments for the Controller Terminal Strip
Terminal Connects to:
BAT -
BAT +
+ CH1 OUT + connection of 4 - 20 mA output, channel 1
- CH 1 OUT - connection of 4 - 20 mA output, channel 1 + CH2 OUT + connection of 4 - 20 mA output, channel 2
- CH2 OUT - connection of 4 - 20 mA output, channel 2
- connection from 24 VDC power source
+ connection from 24 VDC power source1 (or 24 V backup battery)
1
(or 24 V backup battery)
RESET (2) Reset switch (factory-wired) BUZ-/BUZ+ Internal buzzer (factory-wired)
1
If DC power is used as the primary power source, do not make wiring connections to the AC terminal strip.
12 • Internal Description
AC In T e rminal St rip. The AC in terminal strip is a 3-point terminal strip located above the controller terminal strip (see Figure 1). It facilitates wiring from the AC power source. Table 4 lists the function of each terminal.
T a ble 4: Terminal Assignments for the AC In Terminal Strip
Terminal Connects to:
LINE Hot wire from AC power source. NEUT Neutral wire from AC power source. GND Earth gro und
Relays
The Beacon 200 includes four channel relays (two per channel) and three common relays. Both sets of relays are single-pole, double-thr ow (SPDT) and are rated for 10 amps at 250 VAC (resistive).
NOTE: You can select normally energized (NE) or normally de-energized
(NDE) settings for ea ch channel in th e Channel Co ntrol and Setup program. This section describes the default setting : normally de­energized.
The alarm 1 and alarm 2 common relays are factory-set as NDE and the fail common relay is factory-set as NE. The alarm 1, alarm 2, and fail common relays’ NE/NDE settings are not user­adjustable.
Channel relays. The four channel relays are above the alarm terminal strip (see Figure 1). These relays are dedicated to specific channels and alarm levels.
For example, the channel 1, alarm 1 relay energizes if channel 1 recognizes an alarm 1 condition. Figure 3 below illustrates the allocation of the channel relays
Internal Description • 13
.
Channel 2, Alarm 2 Channel 2, Alarm 1 Channel 1, Alarm 2
Channel 1, Alarm 1
Figure 3. Beacon 200 Channel Relay Allocation
NOTE: The alarm 2 channel relays may be set to operate as individual
Common relays. The three common relay s, alarm 1, alarm 2, and fail, are to the lef t of th e controlle r terminal strip (see Figu re 1). These rela ys are common for both channels.
Common Alarm 1 Common Alarm 2
Common Fail
K1 K2 K3
K4
K5 K6 K7
channel fail relays. See the Configu re Channel Settings sect ion of Chapter 5 for instructions.
For example, the al arm 1 common rela y energizes if either channel 1 or channel 2 recognizes an alarm 1 condition.
Fuses
There are three fu ses that are used in t he Beacon 200 . T w o of them are AC fuses and one of them is a DC fuse.
AC Fuses. The two fuses located directly to the left of the controller terminal strip are the AC fuses. They cut off the incoming AC power in the event of a short circuit or other electric al fault which causes a high current draw in the Beacon 200. They are housed in vertical fuse holders and are held in the holder by a quarter turn cover. They are labelled as F2 (top fuse) and F3 (bottom fuse) on the PCB silk-screen and are rated at 3 A, 250 V.
DC Fuse. The fuse located to the left of the power switch is the DC fuse. It cuts off incoming DC power in the event of a short circuit or other electrical fault which causes a high current draw in the Beacon
200. It is also housed in a vertical fuse holder and is held in the holder by a quarter turn cover. It is labelled as F1 on the PCB silk-screen and is rated at 6A, 250 V.
14 • Internal Description
Power Switch
The power switch is located above the relays and in between the AC and DC fuses (see Figure 1). The power switch turns the incoming AC power source on and off at the Beacon 200. When the switch is up, the power switch is on.
NOTE: The DC power in put has no on/off switch and i s not af fected by the
position of this switch.
Power Supply
The power supply is mounted to the power supply mounting plate which is located behind the display PCB. The p ower supply mounting plate is mounted to the main PCB with fou r st a ndo ffs. The power supply receives the AC input from the external power source and converts it to a DC voltage that is usable by the Beacon 200 circuitry.
Internal Description • 15
Chapter 3: Installation and Start Up
Overview
This chapter describes procedures to mount the Beacon 200 Gas Monitor, make wiring connections to the monitor, and start up the monitor.
WARNING: Perform all installation and start-up procedures in a
“fresh air” environment (known to be free of combustible gas, toxic gas, and of normal oxygen content). The Beacon 200 is not in operation as a gas monitoring system until the start-up procedure is complete.
Mounting the Beacon 200 Gas Monitor
Perform the following procedure to install the Beacon 200 at the mounting site.
1. Select the mounting site. When you select the mounting site consider
the following factors:
Is an AC or DC power source available?
Is there enough room to open the housing door and make wiring
connections through the conduit hubs at the bottom of the housing?
Are the display screen and status lights visible?
2. Close and latch the housing door.
3. Position t he monitor on a vertical surface at e ye level (4 1/2 to 5 feet
from the floor).
4. The Beacon 200 is shipped with the mounting feet positioned behind
the housing. Loosen the screws that secure the feet to the housing, rotate the feet to their mounting position (as shown in Figure 4), then tighten the screws.
5. Insert 1/4 in. or 5/16 in. screws through the slot s in the mounting fe et at
16 • Overview
each corner of the housing to secure the housing to the mounting surface (see Figure 4).
Ø
.41
8.50
.38 x .50 slot, 4X
6.00
Mounting Feet, 4X
ALARM 2
FAIL
10.5010.94
PILOT
ALARM 1
BEACON 200
GAS MONITOR
RESET
.80
3/4 " Conduit Hub(3)
Figure 4. Beacon 200 Gas Monitor Outline and Mounting Dimensions
Mounting the Beacon 200 Gas Monitor • 17
AC In Terminal Strip
Wiring
Typical Detector/Transmitter
Terminal Strip Wiring
Only one detector or transmitter can be
wired to Ch1 or Ch2 at a time. See
detector head wiring diagram for
AMP/PREAM P
+S
+S
specific wiring.
CHANNEL 1
OXYGEN
+
RW G B
Green
White
- DC GROUND FB (4-20 mA)
+ 24 V
CHANNEL 2
OXYGENAMP/PREAMP
+
RW G B
LEL
Oxygen Detector
3-Wire 4-20 ma Transmitter
LEL
Controller Terminal
Strip Wiring
24 V
Channel 1 Recorder
1K Max Impedance
Channel 2 Recorder
1K Max Impedance
Reset Switch
(Factory Wired)
Buzzer
(Factory Wired)
Line (Hot) Neutral Ground
115V~
(220V~ Optional)
BAT BAT + +
4- 20mA Out
4- 20mA Out
CH1 OUT
+
CH2 OUT
RESET RESET
BUZ BUZ +
Typical Alarm Relay Terminal
Strip Wiring
CH1-A1 COM-A1 COM-FAILCOM-A2
NO NC C
CH2-A1CH1-A2
CH2-A2
NO NC C
NO NC CNO NC CNO NC C NO NC CNO NC C
ALARM DEVICE POWER
Black
Green
White
FB (4-20 mA) + 24 V
Red
2-Wire 4-20 ma Transmitter
LEL Detector
Alarm Devices
Contact Rating of 10 Amps at 115/220 V~ Resistive or 10A @ 30V Resistive for Each Set of Alarm Relay Contacts.
Figure 5. Beacon 200 Gas Monitor General External Wiring Diagram
18 • Mounting the Beacon 200 Gas Monitor
Wiring the Beacon 200 Gas Monitor
This section describes procedures to connect the AC power source, DC power source, external alarm(s), recorder, and detector head(s). See Figure 5 for a g eneral wiring diagram of all external wiri ng to the Beacon
200.
WARNING: Make all connections to the Beacon 200 before you plug
in or turn on the AC or DC power source. Before you make any wiring adjustments, al ways verify that all pow er sources are not live.
Connecting the AC Power Source
NOTE: If you are using DC power as the pr im a ry pow er source, go to the
next section, “Connecting the DC Power Source.”
The AC in terminal stri p will acce pt 24 - 14 AWG wire. When selecting wire to connect the AC power source to the Beacon 200, be sure to meet the local electric al code. Also be sure to use an appropriate circuit breaker in the AC line close to the Beacon 200 that meets the local electrical code.
Perform the following procedure to connect the AC power source to the Beacon 200.
WARNING: Verify that the power source is unplugged or turned off at
the power source end before you continue with this procedure.
1. Turn off or unplug all incoming power to the Beacon 200 at the power source end.
2. Open the housing door, then place the power switch in the OFF position.
3. Install an appropriately rate d cable bu shing or con duit to the ri ght-most conduit hub on the bottom of the Beacon 200 housing.
4. Locate the 3-point AC in terminal strip (see Figure 1). The terminals are labelled LINE, NEUT, and GND.
5. Guide the AC power cord or wires through the right-most conduit hub on the bottom of the Beacon 200 housing.
Wiring the Beacon 200 Gas Monit or • 19
CAUTION: Do not route power and det ector head wiring through the same
conduit hub. The p ower wiring may disr upt th e transmission of the detector head signal to the monitor.
6. Connect the AC wires to the AC power terminal strip as shown in Figure 6.
AC In Terminal Strip
Line (Hot)
Neutral
115 V~ ± 10%, 50/60 Hz
(220 V~Optional)
Ground
Figure 6. AC Power Wiring
Connecting the DC Power Source
WARNING: Verify that the power source is unplugged or turned off at
the power source end before you continue with this procedure.
DC power may be used as a primary power source. It may also be used as a backup power source with a 24 VDC battery if AC power is the primary power source. If DC power is the primary power source, DO NOT connect AC power.
1. Turn off or unplug all incoming power to the Beacon 200 at the power source end.
2. Open the housing door, then place the power switch in the OFF position.Locate the DC power terminals on the controller terminal strip (see Figure 1). They are labelled BAT - and BAT +
3. Install an appropriately ra ted cable bush ing or conduit to one of the unused conduit hubs on the bottom of the Beacon 200 housing.
4. Guide a DC power cord or wires through the selected conduit hub.
CAUTION: Do not route power and det ector head wiring through the same
conduit hub. The p ower wiring may disr upt th e transmission of the detector head signal to the monitor.
20 • Wiring the Beacon 20 0 Gas Monitor
CAUTION: When a batter y is used as backup power, the Beacon 200
trickle charges the battery. Do not use a non-rechargeable battery as backup power. Use RKI backup battery 49-8102RK or an appropriately rate d re cha rg ea bl e lead acid type batte ry.
5. Connect the DC wires to the controller terminal strip as sh own in Figure 7.
Controller Terminal Strip
24 VDC ± 2.5 VDC
+
BAT BAT +
Figure 7. DC Power Wiring
NOTE: If a 24 VDC battery is used as a backup power source, as long as
AC power is on t he Beacon 200 will keep a trickl e charge on the battery to maintain its charge. If AC power is interrupted, the Beacon 200 will operate from the DC backup battery unt il the battery voltage drops to 21.5 volts or less, or until AC power is restored. Battery recharge time will vary depending on how much the battery was drained.
Connecting External Alarms
Perform the following procedure to connect external alarm devices to the Beacon 200.
NOTE: The alarm termina l strip includes terminals for channel alarms
and common alarms. Channel alarms are activated by one particular channel. Common alarms are ac tivated by eith er of the two channels.The example used in this procedure describes connecting an external alarm devi ce to one of the channel alarm terminals: the channel 1, al arm 1 term inal s.
1. Turn off or unplug all incoming power to the Beacon 200 at the power source end.
2. Open the housing door, then place the power switch in the OFF position.Locate the alarm terminal strip (see Figure 1).
3. Install an appropriately rated cable bushing or conduit in an unused conduit hub on the bottom of the Beacon 200 housing.
Wiring the Beacon 200 Gas Monit or • 21
4. Guide the wiring of the external alarm device through the selected conduit hub on the bottom of the Beacon 200 housing.
CAUTION: Do not route the external alarm wiring and detector head
wiring through the same conduit hub. The external alarm wiring may disrupt the transmission of the detector signal to the Beacon 200.
5. Connect the leads from the external alarm device and power to the alarm terminals as shown in Figure 8.
External Alarm
Device
6. Repeat steps 3 and 4 for additional external alarm devices.
Connecting a Recorder
CH1-A1
NO NC C
+ (H)
-(N)
Figure 8. Typical External Alarm Wiring
Beacon 200 Alarm Terminal Strip
External Power
Source
Perform the following procedure to connect an analog recording device tot he Beacon 200. The output at the recorder output terminals is a 4 - 20 mA signal that is proportional to the detection range of the applicable detector head.
1. Turn off or unplug all incoming power to the Beacon 200 at the power source end.
2. Open the housing door, then place the power switch in the OFF position. Locate the recorder output terminals on the controller terminal strip (see Figure 1).
3. Install an appropriately rated cable bushing or conduit in an unused conduit hub on the bottom of the Beacon 200 housing.
22 • Wiring the Beacon 20 0 Gas Monitor
4. Guide the wiring from the recording device through the selected conduit hub on the Beacon 200.
5. Connect the wires form the recording device to the recorder output terminals as shown in Figure 9.
Recording Device #1, 1Kohm Max Impedance
Recording Device #2, 1Kohm Max Impedance
Connecting RKI Detector Heads
Perform the following procedure to connect an RKI detector head to the Beacon 200.
1. Turn off Power to the Beacon 200 at the power source.
2. Open the Beacon 200 door and place the power switch in the off position.
3. See the detector head instruction manual for instructions on how to connect wires to the detector head.
+
+
Figure 9. Recorder Output Wiring
+ +
CH1 OUT
CH2 OUT
4. Install an appropriately rated cable bushing or conduit in an unused conduit hub on the bottom of the Beacon 200 housing.
5. Route the wir es in conduit or shielded cable from the detector head through the selected conduit hub into the Beacon 200. See Table 5 below for wire size and distance guidelines.
Unshielded twisted pair cable in conduit or shielded twisted pair cable is recommended for all the direct connect detector heads. For the LEL detector, pair the R & B wires and the W & G wires. Shielded cable or wires in conduit are recommended for the 2-wire and 3-wire 4 - 20 mA transmitters.
6. Connect the wires from the det ector head to the appropriate det ector/ transmitter terminals. The top detector terminal strip is for channel 1 and the bottom one is for channel 2. See the detector head instruction manual for controller terminal connections.
Wiring the Beacon 200 Gas Monit or • 23
CAUTION: Do not route power and det ector head wiring through the same
conduit hub. The p ower wiring may disr upt th e transmission of the detector head’s to the Beacon 200.
.
T able 5: Wire Size Guidelines for RKI Detector Head Wiring
Number of
Detector Head Type
Direct Connect LEL 4 5 00 ft. 1,000 ft. 2,000 ft. Direct Connect Oxygen 2 500 ft. 1,000 ft. 2,000 ft. Direct Connect H2S 2 500 ft. 1,000 ft. 2,000 ft. Direct Connect CO 2 500 ft. 1,000 ft. 2,000 ft. Direct Connect Cl2 2 5 00 ft. 1,000 ft. 2,000 ft. Direct Connect SO2 2 500 ft. 1,000 ft. 2,000 ft. 2-Wire 4 - 20 mA Transmitter 2 2,500 ft. 5,000 ft. 8,000 ft. 3-Wire 4 - 20 mA Transmitter 3 2,500 ft. 5,000 ft. 8,000 ft.
Wires to
Controller
Max Distance
to Controller
w/18 Gauge
Wire
Max Distance
to Controller
w/16 Gauge
Wire
Max Distance
to Controller
w/14 Gauge
Wire
Connecting User-Supplied 4 - 20 mA Transmitters
The Beacon 200 ma y be used with a user s upplied 2-wire or 3-wire 4 - 20 mA transmitter which runs on 24 VDC. When this is done, the Beacon 200 is normally setup at RKI Instruments with the following channel parameters: unit of measure, item name, and full scale. For example, “PSI AIR” with a full scale of 10 PSI.
Perform the following procedure to connect a 4 - 20 mA transmitter, which you supply, to the Beacon 200.
1. Turn off power to the Beacon 200 at the power source.
2. Open the Beacon 200 door and turn off the power switch.
3. See the transmitter’s instruction manual for instructions on how to connect wires to the transmitter.
4. Install an appropriately rated cable bushing or conduit in an unused conduit hub on the bottom of the Beacon 200 housing.
5. Route the wires from the transmitter through the selected conduit hub into the Beacon 200.
24 • Wiring the Beacon 20 0 Gas Monitor
6. Connect the wires from the transmitter to the appropriate detector/ transmitter terminals. The top detector terminal strip is for channel 1 and the bottom one is for channel 2. See the transmitter instruction manual for contro ller terminal connections. Figure 10 below illustrates typical transmitter wiring connections.
CAUTION: Do not route power and transmitter wiring through the same
conduit hub. The p ower wiring may disr upt th e transmission of the transmitter’s signal to the Beacon 200.
3-Wire Connection
AMP/PREAMP
+ S
OXYGEN
+
-
(DC GROUND)
RW G B
FB (4-20 mA) + 24 VDC
Figure 10. Generic 4 to 20 mA Transmitter Output Wiring
LEL
3-Wire 4-20 ma Transmitter
2-Wire Connection
AMP/PREAMP OXYGEN
+ + S
FB (4-20 mA) + 24 VDC
RW G B
LEL
2-Wire 4-20 ma Transmitter
Wiring the Beacon 200 Gas Monit or • 25
Starting Up the Beacon 200 Gas Monitor
Perform the following procedure to place the Beacon 200 into normal operation.
1. Complete the mounting and wiring procedures describe d e ar l ie r in this chapter.
2. Complete all installation procedures described in the detector head or user supplied 4 - 20 mA transmitter instruction manual.
3. Verify that all wiring connections are correct and secure and that the Beacon 200’s power switch is in the OFF position.
4. Plug in or turn on the incoming power source (AC or DC) at the power source end.
5. Place the Beacon 200’s power switch in the ON position. RKI INSTRUMEN T S BE A CON 200 appears on the display for a few seconds, then WARMING UP appears for each active channel. The warm-up period will last for one minute.
NOTE: To prevent unwanted alarms during warm up, the alarm circuits
are not active while the WARMING UP message is displayed.
6. Verify that the PILOT light is on. If the PILOT light is not on, see the troubleshooting guide in Chapter 5, Maintenance.
7. Perform the start-up procedure for each detector head or user s upplied 4 - 20 mA transmitter as described in the detector head or transmitter instruction manual
26 • Starting Up the Beacon 200 Gas Monitor
Chapter 4: Operation
Overview
This chapter describes the Beacon 200 Gas Monitor in normal operation. This chapter also describes the Beacon 200 in alarm 1, alarm 2, and fail conditions and suggests response to these conditions.
Normal Operation
Normal operation is defined as follows:
The start-up procedure is complete.
The Beacon 20 0 is not indicating an alarm 1, alarm 2, or fail condition.
The Beacon 200 is not running the Channel Control & Setup or
Calibration Programs.
During normal operation, the Beacon 200 simultaneously displays the current gas reading, unit of measure, and target gas for each active channel.
:
1 2
:
0
20.9
% %
E
L
L
Oxy
CH4
g
e
n
The PILOT light is on during normal operation indicating that the Beacon 200 is receiving incoming power.
Overview • 27
Recorder Output Operation
The output at the recorder output terminals is a 4 - 20 mA signal for each active chann el that is proportional to the detection range of the channel. A channel that is set as CHANNEL NOT USED or CHANNEL DISABLED in the Channel Control & Setup Program (see Chapter 5) has an output of 0 mA.
There are several special circumstances where the recorder output will behave as follows:
When a channel is in WARMUP after the Beacon 200 is turned on, the
recorder output will be at 4 mA for all channel types except oxygen. For oxygen channels, the output will be 17.4 mA while in WARMUP.
If the Beacon 200 is being powered by a battery and is in low battery
alarm, the recorder output for eac h channel will be 0 mA.
When a channel is added or a channel type changed, the display will
indicate NEEDS CALIBRATION for that channel when the Beacon 200 is first turned on and will continue to indicate this until the channel is calibrated. In this situation, the recorder output will be at 3.2 mA u ntil the channel is calibrated.
If a channel goes into a fail condition, the recorder output will be 0 mA.
If you enter any of the instrument programs, such a s the Calibration
Program, the recorder output will hold at the value it was at when you entered the program was entered until you return to normal operation.
28 • Recorder Output Operation
Alarm Indications
This section describes the Beacon 200 in alarm 1, alarm 2, and fail conditions and suggests response to these conditions. Table 6 below lists the alarm indications for each condition.
NOTE: The Beacon 200 includ es alarm on and alar m off delay setti ngs for
each channel and level of gas alarm. The alarm indications described in this section operate according to the factory set delay settings. See the Configure Channel Settings Menu section of Chapter 5 for all the factory settings.
Table 6: Visual and Audible Alarm Indications
Condition Cause Visual Indication(s)
1
Alarm 1
Alarm 2
Fail
Low Battery
*
*2If DC power is used as primary or backup power source.
1
1
If the Beacon 200 is in both an alarm 1 and an alarm 2 condition, both alarm lights are on and the display
alternates between the gas reading and the ALARM-1 ALARM-2 message.
Increasing (decreasing for O2) gas reading at or above the alarm 1 setpoint
Increasing gas reading at or above the alarm 2 setpoint
Disconnected or misconnected detector wiring
Display reading below -10% of full scale or lower
Defective components
2
No AC power and DC power source (primary or backup) less than 21.5 volts.
ALARM 1 light is on
Gas reading flashes and alternates with ALARM-1 message
ALARM 2 light is on
Gas reading flashes and alternates with ALARM-2 message
FAIL light is on
FAIL message flashes in place of gas reading
FAIL light is on
SUPPLY VOLTAGE IS TOO
LOW LOW POWER STANDBY message and actual
voltage of incoming DC power
Pulsing tone
Pulsing tone
Steady tone
None
Audible
Indication
NOTE: You can select normally energized (NE) or normally de-energized
(NDE) channel relay settings in the Channel Control & Setup menu. The following sections describe the default setting for the channel relays which is NDE.
Common alarm 1 and alarm 2 relays are factory-set as NDE, and the common fail relay is factory set as NE. The common relays’ NE/NDE settings are not user-adjustable.
Alarm Indications • 2 9
Alarm 1 Condition
This section describes the audible and visual indications for an alarm 1 condition and sugge s t s re spo nse to an al ar m 1 condit ion .
Alarm 1 condition indications
When the gas reading of an active channel reaches the alarm 1 setpoint, the Beacon 200 senses an alarm 1 con dition. The Beacon 200 ale rts you to an alarm 1 condition as follows:
The ALARM 1 light turns on.
The gas reading in alarm 1 condition flashes and alternates with the ALARM-1 message.
The buzzer sounds a Pulsing tone.
The common al arm 1 relay energiz es .
The applicable alarm 1 channel relay energizes.
Responding to an alarm 1 condition
This section suggests response to an alarm 1 condition.
1. Follow your established procedu r e for a low level combustible or toxic
gas condition or a decreasing oxygen content condition.
2. Oxygen alarms are self-resetting and will automatically clear when the
oxygen rises above the alarm 1 setpoint.
3. Alarms for all other gas types are latching. After the gas reading falls
below the alarm 1 setpoint, press the reset switch to reset the alarm 1 circuit. Resetting the alarm 1 circuit silen c es the bu zzer, turns off the ALARM 1 light, resets the channel display, and de-energizes the common and channel alarm 1 relays.
NOTE: To silence the buzzer while in an alarm 1 condition, press the
reset switch. You cannot de-energize the alarm 1 relays until the gas reading
falls below (above for oxygen) the alarm 1 setpoint .
30 • Alarm Indications
Alarm 2 Condition
This section describes the audible and visual indications for an alarm 2 condition and sugge s t s re spo nse to an al ar m 2 condit ion .
Alarm 2 condition indications
When the gas reading of an active channel reaches the alarm 2 setpoint, the Beacon 200 senses an alarm 2 con dition. The Beacon 200 ale rts you to an alarm 2 condition as follows:
The ALARM 2 light turns on.
The gas reading in alarm 2 condition continues to flash and alternates with the ALARM-2 messages.
The buzzer sounds a Pulsing tone.
The common al arm 2 relay energiz es .
The applicable alarm 2 channel relay energizes.
Responding to an alarm 2 condition
This section suggests response to an alarm 2 condition.
1. Follow your establish ed proced ure fo r a high l evel combust ibl e or toxic
gas condition or an increasing oxygen content condition.
2. Oxygen alarms are self-resetting and will automatically clear when the
oxygen rises above the alarm 2 setpoint.
3. Alarms for all other gas types are latching. After the gas reading falls
below the alarm 2 setpoint, press the reset switch to reset the alarm circuit. Resetting the alarm circuit turns off the ALARM 2 light, and de­energizes the common and channel alarm 2 relays .
NOTE: To silence the buzzer while in an alarm 2 condition, press the
reset switch. You cannot de-energize the alarm 2 relays until the gas reading
falls below the alarm 2 setpoint.
Fail Condition
This section describes the audible and visu al indication s for a fail conditio n and suggests response to a fail condition.
Fail condition indications
The Beacon 200 senses a fail condition for any of the following:
The detector head wiring to the Beacon 200 is disconnected or
Alarm Indications • 3 1
incorrectly co nnected.
The detector head’s detector is disconnected or incorrectly connected.
The display reading is -10% of full scale or lower.
The Beacon 200 or detector head is malfunctioning.
When the Beacon 200 senses a fail condition, it alerts you as follows:
The FAIL light turns on.
The gas reading for the failing channel is replaced by the FAIL message.
The buzzer sounds a steady tone.
The common fail r elay de-energizes.
NOTE: If you elected to use the channel’s alarm 2 relay as an individual
fail relay in the Channel Control & Setup menu, the rel ay de­energizes in a fail condition. See the Chan nel Control & Setup section of Chapter 5 for a description of this setting.
Responding to a fail condition
This section suggests response to a fail condition.
1. Verify that the detector head wiring to the Beacon 200 is correctly and
securely connected.
2. Verify that the detector head’s detector is correctly and securely
connected.
3. See the troubleshooting guide in the detector head instruction manual.
Low Battery Condition
This section describes the audible and visual indications for a low battery condition and suggests respon se to a low battery condition. T his condition only applies when DC power is used as a primary or backup power source.
NOTE: When a 24 VDC battery is used as a backup power source, the
Beacon 200 ke eps t he battery charged by providing a trickle charge from the AC power source. If AC power is interrupted, the Beacon 200 will operate from the DC backup battery unt il the battery voltage drops to 21.5 volts or less, or until AC power is restored.
32 • Alarm Indications
Low battery condition indications
The Beacon 200 senses a low battery condition when:
AC power is disconnected, misconnected, or interrupted
AND
the DC power source is 21.5 volts or less
When the Beacon 200 senses a low battery condition, it alerts you as follows:
The FAIL light turns on.
The top display screen displays the SUPPLY VOLTAGE IS TOO LOW, LOW POWER STANDBY message and the actual voltage of incoming DC power.
Responding to a low battery condition
This section suggests response to a low battery condition.
If DC power is the primary power source:
1. For a temporary DC power source, disconnect primary DC power at
the Beacon 200, then connect a 24 VDC backup battery.
2. Determine and correct the cause of primary DC power loss. When the DC power source rises above 22.0 volts, the Beacon 200
begins the warm up process.
If DC power is the backup power source:
1. Replace or recharge the 24 VDC backup battery to resume backup
power capability.
2. Determine and correct the cause of primary AC power loss. When backup DC or primary AC power is restored, the Beacon 200
begins the warm up process. When AC power is restored, the Beacon 200 charges the backup battery until it is fully recharged. Charge time varies depending on the battery size and how much the battery was depleted. Once the battery is fully charged, the Beacon 200 reverts to a trickle charge to maintain the battery charge.
Alarm Indications • 3 3
Viewing & Resetting Min/Max Readings
The Reset switch may be used to view and reset the minimum and maximum gas readings for the active channel(s).
1. While the Beacon 200 is in norm al operat ion, press an d hold th e Reset
switch button for 3 seconds.
2. The display will indicate MIN / MAX Display Pres s RESET when
done viewing . . . for 5 seconds before displaying the minimum and maximum readings for the active chann el(s). The minimum reading is on the left and th e m axim um i s on the r ight si d e of the di splay for each channel.
3. Press and release the Reset switch button to exit the min/max screen.
The display will indicate To RESET Min/MAX values Press and HOLD RESET Button for 10 seconds and then return to normal operation.
To return to normal operation without resetting the minimum and
maximum readings, do not press the Reset switch button and allow the unit to return to normal operation.
To reset the minimum and maximum readings, before the unit
returns to normal operation press and hold the Reset switch button until the display indicates Min/Max Values Have Been Reset. Release the Reset switch button.The unit will then return to normal operation.
34 • Viewing & Resetting Min/Max Read ings
Chapter 5: Channel Control and Setup Program
Overview
The Channel Control & Setup Program allows viewing of and changes to instrument setup parameters. It is accessed using the program buttons. The Channel Control & Setup Program i ncludes thr ee m enus as de scribed in Table 7.
Table 7: Channel Control & Setup Program Menus
Menu Function
Enable/Disable Channel(s) Configures channels as enabled, disabled, or not used Configure Channel Settings Configures alarm settings, noise filter setting, and zero
suppression setting for each channel
View System Information Displays the firmware version number and the instrument
operating voltage
To enter the Channel Control & Setup Program, simultaneously press and hold the ESCAPE and ENTER buttons f or approximately 5 seconds.
The Channel Control & Setup Program menu includes a 5-minute time-out feature. If you do not press a button for 5 minutes, the Beacon 200 automatically returns to normal operation.
NOTE: If the Beacon 200 returns to normal operation because of a
program time-out, the active channels enter a warm-up period just as they do when the unit is first turned on.
If you are installing a new system, the channels have been setup at the factory for the ordered detector he ads. Use the Channel Control & Setup Program only if you want to disable or enable a channel, delete a channel, or change chan ne l setti n gs. If a channel is being added or a channel is being changed from one type to another, contact RKI Instruments, Inc. for additional documentation required to define the channel type.
Overview • 35
Enable/Disable Channel(s) Menu
1. From normal operation, simultaneously press and hold the ESCAPE
and ENTER button s for ap pr oxim ately 5 seconds to enter t he C h ann el Control & Setup Program. Release the buttons when the Control & Setup Program Proceed? [YES] or [NO] message appears on the display screen.
2. Press the UP/YES button to continue.
3. Press the UP/YES or DOWN/NO button until the 1) Enable/Disab le
Channel(s) message appears on the display screen, then press the ENTER button.
4. Use the UP/YES and DOWN/NO buttons to select the channel you
want to enable or disable, then press the ENTER button.
5. Press the DOWN/NO button. The CHANNEL USAGE setting displays
on the display screen.
6. Use the UP/YES and DOWN/NO buttons to display the setting you
want, then press the ENTER button to select the setting. The table below describes the three available settings.
Table 8: Beacon 200 Channel Usage Settings
Setting Description
CHANNEL ENABLED The Beacon 200 displays gas readings and initiates gas and channel failure alarms
when appropriate. Use this setting for normal operation when the channel has a detector head wired to it.
CHANNEL DISABLED The Beacon 200 displays DISABLED for the channel and the channel’s alarm circuit
is not active. Use this setting when the channel has a detector head wired to it, but gas readings and alarms are not required for the channel (fo r example if the detector hea d requires maintenance or is malfunctioning).
CHANNEL NOT USED The Beacon 200 leaves the channel blank on the display screen.
Use this setting when the channel does not have a detector head wired to it.
7. Press the ESCAPE button, then press the DOWN/NO button to return
to normal operation.
36 • Enable/Disable Channel(s) Menu
Configure Channel Settings Menu
This section describes how to view and change channel parameters for the installed gas channels.
1. Simultaneously press and hold the ESCAPE and ENTER buttons for
approximately 5 seconds to enter the Channel Control & Setup Program. Release the buttons when the Control & Setup Program Proceed? [YES] or [NO] message appears on the display screen.
2. Press the UP/YES button to continue.
3. Press the UP/YES or DOWN/NO button until the 2) Configure
Channel Setting(s) message appears on the display screen, then press the ENTER button.
4. Use the UP/YES and DOWN/NO buttons to select the channel for
which you want to set parameters, then press the ENTER button.
5. Press the UP/YES button until the parameter you want to set appears
on the display screen. The screen will display the current setting and ask if it is OK.
Table 9 lists the p aramete rs you can set for a channel . Table 9 also list s the factory set value for each parameter.
NOTE: Use the ESCAPE button to go back to a previously displayed
parameter.
6. If the setting is not OK and you want to change th e it, press the DOWN/
NO button. The parameter is now adjustable.
7. Use the UP/YES or DOWN/NO button to update the parameter, then
press the ENTER button to continue.
8. Repeat steps 5 through 7 to set any other channel paramet ers.
9. Press the UP/YES button until the following message appears on the
display screen.
Con f iguration for
CHANNEL 1
has nbee
eSav Set t i ngs
completed
[
Y
?
/N
]
Configure Channel Settings Menu • 37
10.Press the UP/YES button to save the configuration. The screen will
then return to the Channel Control & Setup menu.
11. Press ESCAPE to return to the screen which asks Control & Setup
Program Proceed? [YES] or [NO].
12.{Press the DOWN/NO button to return to normal operation.
Table 9: Channel Setting Parameters
Parameter
(Factory-Set Value)
ALARM-1 Level See the Beacon 200 Detector Head Specification Sheet for the detector head installed on this channel
ALARM-1 ON DELAY (1 sec)
ALARM-1 OFF DELAY (0 sec)
ALARM-1 (activation) (DECREASING for oxygen channels, INCREASING for all other channel types)
ALARM-1 Relay (action) (NORMALLY DE-ENERGIZED)
ALARM-1 Relay (reset) (SELF RESETTING for oxygen, LATCHING for all other channel types)
Description
The gas read ing at which the Beacon 200 initiates an alarm 1 condition for this channel.
The amount of time the Beacon 200 delays activation of the alarm 1 circuit once an alarm 1 condition is initiated.
The amount of time the Beacon 200 delays turning off the alarm 1 circuit once an alarm 1 condition passes.
Indicates if the alarm 1 circuit is activated by gas readings INCREASING or DECREASING to the ALARM-1 Level.
If set as NORMALL Y DE-ENERGIZED, t he cha nnel’ s a larm 1 re lay is de­energized in normal operat ion and energ ize s wh en an alarm 1 cond iti on is initiated. If set as NORMALLY ENERGIZED, the channel’s alarm 1 relay is energized in normal operation and de-energizes when an alarm 1 condition is initiated.
If set as LATCHING, you must press the RESET button to reset the alarm 1 circuit after the alarm 1 condition passes. If set as SELF RESETTING, the Beacon 200 automatically resets the alarm 1 circuit after the alarm 1 condition passes.
ALARM-2 Relay (used for) (ALARM-2 Condition)
ALARM-2 Level See the Beacon 200 Detector Head Specification Sheet for the detector head installed on this channel
ALARM-2 ON DELAY (2 min for oxygen, 1 sec for all other channel types)
ALARM-2 OFF DELAY (0 sec)
If set as ALARM-2 Condition, the channel’s alarm 2 relay activates when an alarm 2 condition is initiated for the channel. If set as FAIL Condition, the channel’s alarm 2 relay activates when a fail condition is initiated for the channel.
The gas read ing at which the Beacon 200 initiates an alarm 2 condition for this channel.
The amount of time the Beacon 200 delays activation of the alarm 2 circuit once an alarm 2 condition is initiated.
The amount of time the Beacon 200 delays turning off the alarm 2 circuit once an alarm 2 condition passes.
38 • Configure Channel Settings Menu
Table 9: Channel Setting Parameters (Continued)
Parameter
(Factory-Set Value)
ALARM-2 (activation) (INCREASING)
ALARM-2 Relay (action) (NORMALLY DE-ENERGIZED)
ALARM-2 Relay (reset) (SELF RESETTING for oxygen, LATCHING for all other channel types)
NOISE FILTER (3)
ZERO SUPPRESSION (0.0% of the detection range in t erms of the detection units for carbon dioxide channel types, 0.5% oxygen for oxygen channels, 2.0% of the detection range in terms of the detection units for all other channel types)
Description
Indicates if the alarm 2 circuit is activated by gas readings INCREASING or DECREASING to the ALARM-2 Level.
If set as NORMALL Y DE-ENERGIZED, t he cha nnel’ s a larm 2 re lay is de­energized in normal operat ion and energ ize s wh en an alarm 2 cond iti on is initiated. If set as NORMALLY ENERGIZED, the channel’s alarm 2 relay is energized in normal operation and de-energizes when an alarm 2 condition is initiated.
If set as LATCHING, you must press the RESET button to reset the alarm 2 circuit after the alarm 2 condition passes. If set as SELF RESETTING, the Beacon 200 automatically resets the alarm 2 circuit after the alarm 2 condition passes.
The noise filter feature helps “smooth out” jumpy or noisy signals from the detector head. You can set the noise filter from 1 to 8. A setting of 8 produces the greatest amount of smoothing but also responds slowest to changes in the response reading. A setting of 1 responds fastest to changes in the response reading but produces the least amount of smoothing.
The zero suppression feature helps prevent “jumpy” readings near the fresh air reading. For example, if the zero suppression setting on a %LEL channel is 2.0% and the full scale is 100% LEL, the Beacon 200 will display a reading of 0% LEL for gas readings from -2% LEL to 2% LEL.
View System Information Menu
The View System Information Menu consists of only one display screen which indicates the version number of the firmware that is running the instrument and the system voltage. The system voltage is the voltage that is directly running the instrument’s circuitry. When the unit is running from AC power, this voltage is normally 24.0 volts ± 0.2 volts. When the un it is running from DC power, this voltage reading can be used as an indication of the DC voltage powering the unit.
Configure Channel Settings Menu • 39
Chapter 6: Input Setup Program
Overview
This chapter describes how to use the Input Setup Program to add a channel or change the cha nn el ty pe of an in stalled channel on the Bea c on
200. The Input Setup Program allows you to define the type of detector
head, the units and gas type, and the full scale for that channel. To enter the Input Setup Program, the Beacon 200 must first be off. While
the Beacon 200 is off, press and hold the ENTER button, then turn on the Beacon 200 with the ON/OFF switch.
The Input Setup Program m enu include s a 5-minute time-out feature. If you do not press a button for 5 minutes, the Beacon 200 automatically begins normal operation.
NOTE: If the Beacon 200 enters normal operation because of a program
time-out, the active channels enter a warm-up period just as they do when the unit is first turned on.
40 • Overview
Setting Up a New Channel or Changing an Existing Channel
1. While the Beacon 200 is off, press and hold the ENTER button, then
turn on the Beacon 200 with the ON/OFF switch.
2. The Beacon 200 will beep repeatedly while you are holding down the
ENTER button and then the screen will show INPUT SETUP PROGRAM on the top line.
3. Press the ENTER button to continue.
If you press the ESCAPE button, the unit will start-up and enter its warm-up period.
4. Press the UP/YES or DOWN/NO button until the channel you wish to
add or change appears on the display screen, then press the ENTER button. The detector head type screen appears.
5. The display screen indicates what type of detector head is currently
selected for that channel and asks if it is OK. Table 1 below briefly describes each type.
Table 10: Beacon 200 Detector Head Types
Detector Head Type Description
4 - 20 mA Amp [+S(-)] A 4 - 20 mA detector head is connected to the Beacon 200
with 2 or 3 wires, depending on the detector head model, using the AMP/PREAMP +, S, and - terminals on a detector head terminal strip. All calibration adjustments are made at the detector head.
PreAmp [+S] A PreAmp detector head is connected to the Beacon 200 with
2 wires using the AMP/PREAMP + and S terminals on a detector head terminal strip. All calibration adjustments are made at the Beacon 200.
O2 Direct [+-] An O2 Direct detector head is an oxygen detector head in
which the oxygen detector is wired to the Beacon 200 with 2 wires using the OXYGEN + and - terminals on a detector head terminal strip. All calibration adjustments are made at the Beacon 200.
LEL Direct [RWGB] An LEL Direct detector head is a combustible gas detector
head in which the combustible gas detector is wired to the Beacon 200 with 4 wires using the LEL R, W, G, and B terminals on a detector head terminal strip. All calibration adjustments are made at the Beacon 200.
If the detector head type is not correct, press the DOWN/NO button and continue with step 6.
Setting Up a New Channel or Changing an Existing Channel • 41
If the detector head type is correct, press the UP/YES button to proceed to t he units and gas type screen and skip to step 7.
NOTE: See the detector head operator’s manual and the Beacon 200
Detector Head Specification sheet for the detector head to determine the detector head type.
6. Use the UP/YES and DOWN/NO buttons to scroll through the list of
detector head types until the correct one is displayed. Press the ENTER button to accept the type.The units and gas type screen appears.
7. If the units and gas type are not correct, press the DOWN/NO button
and continue with step 8. If the units and gas type are correct, press the UP/YES button to
proceed to the full scale screen and skip to step 9.
8. Use the UP/YES and DOWN/NO buttons to scroll through a list of units
and gas type choices.
NOTE: See the Beacon 200 Detector Head Specification Sheet for the
detector head to determine t he correct units and gas type.
When the desired units and gas type appears, press the ENTER button to accept the choice. The full scale screen appears.
One of the choices is User Will Specify. If the desired units and gas type setting is not in the list, this setting will allow you to enter a 10 character units and gas type setting. With this choice displayed, press the ENTER button and a screen will appear which allows you to input the characters. Use the UP/YES and DOWN/NO buttons to select a character and then press ENTER to accept it and continue with the next character. When all characters have been entered, the full scale screen will appear.
9. If the full scale setting is not correct, press the DOWN/NO button and
continue with step 10. If the full scale setting is correct, press the UP/YES button to proceed
to the save screen and skip to step 11.
10.Use the UP/YES and DOWN/NO buttons to scroll through a list of full
scale settings.
NOTE: See the Beacon 200 Detector Head Specification Sheet for the
detector head to determine the correct full scale setting.
42 • Setting Up a New Channel or Changing an Existing Channel
When the desired full scale setting appears, press the ENTER button to accept the setting. The save screen appears.
One of the choices is User W ill S pecify. If the desired full scale setting is not in the list, this setting will allow you to enter a full scale setting. With this choice displayed, press the ENTER button and a screen will appear which prompts you to choose how many decimal places you want in the full scale setting. You can select up to 3 decimal places, then press the EN TER button to proceed t o the select f ull scale scr een. Use the UP/YES and DOWN/NO buttons to select a full scale setting, then press ENTER to accept it. The save screen appears.
11. To save the settings, press the UP/YES button and the settings will be
saved. The display will return to the first screen of the Input Setup Program.
If you do not want to save the settings, press the DOWN/NO button. The display will return to the first screen of the Input Setup Program.
12. Press the ENTER button to perform additional setup or the ESCAPE
button to exit the Input Setup Program and begin the Beacon 200’s warm-up sequence.
NOTE: Once the Beacon 200 channels are configured correctly, see the
detector head operator’s manual(s) for a complete description of detector head installation procedures.
Setting Up a New Channel or Changing an Existing Channel • 43
Chapter 7: Maintenance
Overview
This chapter describes use of the Calibration Program and corrective maintenance procedures for the Beacon 200. It includes a troubleshooting guide for problem s you may encoun ter with the B eacon 200 . Procedures to replace compon ents of the Beacon 200 are at the end of this chapter.
Calibration Program
The Calibration Pr ogram is used to calibrate the Beacon 200’s active channel(s). Since the Beacon 200 can support both di rect connect (int ernal amp) and 4-20 mA transmitter (remote amp) detector heads, w hen calibrating the active channel(s) there are three possible detector head combinations:
Direct connect detector head(s) only. If one or two direct connect detector heads are active, then all
calibration adjustments are made at the Beacon 200 after calibration gas is applied at the detector (s).
4-20 mA transmitter detector head(s) only. If one or two 4-20 mA transmitter detector heads are active, then all
calibration adjustments are made at the detector head(s) after calibration gas is applied to the detector(s).
A direct connect and a 4-20 mA transmitter detector head. If one direct connect and one 4-20 mA transmitter detector head are
installed, then calibration adjustments must be made at the Beacon 200 for the direct connect detector head after applying gas to its detector, and at the detector head for the 4-20 mA transmitter detector head.
Calibration Program Flow
Figures 11, 12, and 13 below illustrate the general flow of the Calibration Program for the three possible detector head combinations. See the next section, Entering the Calibration Program, for instructions to enter the Calibration Program. In general, the program screens provide instructions to guide you through the program. At any point in the calibration program, the ESCAPE key may be used to either return to the previous screen or
44 • Overview
abort a process.
Normal
Operation
ESCAPE
Calibration
Program Enter
ENTER/ESCAPE
ENTER
Calibration
Timeout
Selection
Calibrate
Channel 1 Y/N?
Calibrate
Channel 2 Y/N?
(If Install e d )
Channel 2 Y/N?
(If Install e d )
Press Enter to
Adjust Span
Normal
Operation
Air Adjust Channel 2
(If Install e d )
Select Cal. Gas
Concentration
for Channel 2
Span
Air Adjust Channel 1
Select Cal. Gas
Concentration
for Channel 1
Figure 11. Direct Connect Detector Head(s) Only
ESCAPE
Calibration
Program Enter
ENTER/ESCAPE
ENTER
Calibration
Timeout
Selection
Span
Channel 1 Y/N?
Apply Gas to
Ch. 1 & Ch. 2
Detectors
Calibrate
Channel 1 Y/N?
Calibrate
Channel 2 Y/N?
(If Installed)
Calibrate at
Detector
Head(s)
Press Enter When Done
Figure 12. 4 - 20 mA Transmitter Detector Head(s) Only
Calibration Program • 45
Normal
Operation
Calibrate
Channel 1 at
Detector Head
Calibrate
Channel 1 Y/N?
Transmitter
YesNoYes
Direct Connect
Air Adjust
Channel 1
ESCAPE
Calibration
Program Enter
ENTER/ESCAPE
ENTER
Calibration
Timeout
Selection
Calibrate
Channel 2 at
Detector Head
Channel 2 Y/N?
TransmitterTransmitter
Calibrate
Direct Connect
No
Yes
Air Adjust
Channel 2
Press Enter When Done
Select Cal. G as
Concentration
for Channel 1
Apply Gas to
Ch. 1 Detector
Press Enter to
Adjust Span
Press Enter
When Done
Select Cal. Gas
Concentration for Channel 2
Apply Gas to
Ch. 2 Detector
Press Enter to
Adjust Span
Figure 13. 4 - 20 mA & Direct Connect Detector Head(s)
46 • Calibration Program
Entering the Calibration Program
1. Assemble the calibration kit(s). See the instruction manual for each
detector head for procedures specific to that detector head.
2. Simultaneously press and hold the ENTER and UP/YES buttons for
approximately 5 secon ds to enter the Calibratio n Program. Release the buttons when the CALIBRATIO N PROGRAM... message displ ays and asks if you want to continue or return to normal operation.
NOTE: While in the Calibration Program, the alarm status of the Beacon
200 will be locked in the state i t was when the Calibration Program was entered.
3. Press the ENTER button to continue and dis play the Calibr ation Time-
out setting.
The Beacon 200 will rem ain in the Calibration Program for the amo unt of time indicated by the Calibration Time-out setting or until you exit the program. If necessary, adjust the setting using the UP/YES and DOWN/ NO buttons. Make sure you have allotted enough time to perform the calibration pr ocedure. Consider the type of detector head(s) installe d and the distance from the Beacon 200 when determining the time required.
NOTE: If you do not accept a Time-out setting and stay at this screen or
press ESCAPE and do not press control buttons again, then the Beacon 200 will exit the Calibration Pr ogram and begin its warm­up period 5 minutes after the last button was pushed. This warm­up period functi ons as if the unit were just powered up.
4. Press the ENTER button to accep t th e Time-out setting, start the Time-
out period and continue in the Calibration Program.
NOTE: The following describes calibration with a typical channel
allocation which may be different from yours. This example allocation has two active channels with both types of detector heads installed. Channel 1 is a 4-20 mA detector head and channel 2 is a direct connect detector head. See the detector head instruction manuals for calibration information and procedures specific to the detector heads.
Calibration Program • 47
Calibrating a 4-20 mA detector head
1. The display asks if you want to calibrate channel 1 (in this example a 4-
20 mA detector head). Press the UP/YES button to continue with calibrating channel 1.
If you press the DOWN/NO button, the display will skip channel 1 and ask if you want to calibrate channel 2.
2. If you pressed the UP/YES button, since channel 1 is a 4 - 20 mA
detector head in this example, the unit will display the following message for a few second s be fore continuing: Reminder. Calibration Must Be Done at the Detector Head. Then the display will alternate between the current gas reading for channel 1 and the message CALIBRATE AT HEAD above the time remaining in the calibration Time-out.
3. Verify that the detector head is in a fresh-air environment. (If
necessary, use a zero-emission air cylinder, also known as zero air, to introduce a fresh-air sample whe n adjusting the zero reading below.)
4. Adjust the detector head’s zero (fresh air reading for oxygen) reading.
See the detector head’s instruction manual for instructions on how to adjust the zero reading (fresh air reading for oxygen).
5. Apply calibration gas to the detector head’s detector and adjust the
detector head’s span reading (zero reading for oxygen). See the detector head’s i nstruction man ual for instr uctions on how to a djust the span (zero reading for oxygen) reading.
6. Press the ENTER button to indicate that you are done with calibrating
channel 1 and are ready to continue.
Calibrating a Direct Connect Detector Head
1. The display asks if you want to calibrate channel 2 (in this example a
direct connect detector head). Press the UP /YES button to continue with calibrating channel 2.
If you press the DOWN/NO button, the display will skip channel 2 and return to the first calibration program screen which asks if you want to continue or escape from the program.
2. If you pressed the UP/YES button, since channel 2 is a direct connect
detector head in this example, the unit will display the following message for a few seconds before continuing: Expose Dectector(s) To Fresh Air. . . When Done Press ENTER. Then the display will alternate between the gas reading and the message FRESH AIR ENTER to ACCEPT ESCAPE to ABORT above the time remaining in the calibration Time-out.
48 • Calibration Program
3. If the detector is in a fresh air environment, press the ENTER button.
The unit will adjust the zero reading an d display the message Fresh Air Adjust Passed for: Channel 2 before contin uing.
If you suspect the detector area is not a fres h air environment, apply zero air to the detector before pressing the ENTER button. See the detector head instru ction ma nual for inst ructions to apply zero a ir to th e detector. The Beacon 200 will freeze t he display reading for channel 2 at the lowest level reached while applying zero air (highest level for an oxygen channel). A fter applyi ng zero air for the required am ount of time , usually two minutes, remove the zero air source from the detector. Press the ENTER button at the Beacon 200. The unit will adjust the zero reading and display the message Fresh Air Adjust Passed for: Channel 2 before continuing.
4. The display asks if you want to perform a span (zero for an oxygen
channel) operation on channel 2 by applying gas. Press the UP/YES button to continue with adjusting the span on channel 2.
If you press the DOWN/NO button, the unit will return to the first calibration program screen which asks if you want to con tinue or escape from th e Calibration Program.
5. If you pressed the UP/YES button, the display will prompt you for the
span gas (zero gas for an oxygen channel) concentration that will be used. Adjust the displayed concentration up or down as needed using the UP/YES and DOWN/NO buttons so that it matches the concentration in the calibration cylinder.
6. Press the ENTER button to accept the calibration gas concentration
and continue. The unit will display the message Expose Dectector(s) To Gas. . . When Done Press ENTER for a few seconds. It will then alternate between the gas reading for channel 2 and the message APPLYING GAS ENTER to ACCEPT ESCAPE to ABORT above the time remaining in the calibration time-out.
7. Apply the calibration gas to the detector. See the detector head
instruction manual for instructions on how to apply gas to the detector.
When the calibration gas is applied, the Beacon 200 will freeze the display gas reading at the highest level reached whil e the gas was applied (lowest for an oxygen channel).
8. After applying calibration gas for the required amount of time, usually
two minutes, remove the gas from the detector.
Calibration Program • 49
9. Press the ENTER button at the Beacon 200 to proceed with the
calibration adjustment.
If the Beacon 200 is able to succes sfully make the calibration adjustment, it will display the message Cal Passed for: Channel 2. It will then return to the first calibration program screen which asks if you want to continue or escape from the program. Press ESCAPE to return to normal operation.
If the Beacon 200 is not able to make the calibration ad justment, it will display the message Not Enough Response! CAL Failed for: Channel 2, then return to the first calibration program screen which asks if you want to continue or escape from the program. Press ENTER to start the calibration process again or ESCAPE to return to normal operation.
NOTE: Be sure to allow enough time between the application of
calibration gas and returning to normal operation for the gas reading to return to normal levels to avoid false alarms. The alarm circuits are active when the Beacon 200 returns to normal operation.
Replacing the Fuses
The Beacon 200 includes three fuses: one DC and two AC fuses. To replace other components of the Beacon 200, contact RKI Instruments, Inc., for further information.
1. Turn off or unplug all incoming power to the Beacon 200 at the power
source end.
2. Open the housing door of the Beacon 200, then place the power switch
in the OFF position.
3. Locate the vertical fuse holders on eith er side of the power switch. The DC fuse which is l abelled F1 on the P CB silk-scree n is on the le ft of
the power switch. The AC fuses are labelled F2 and F3 and are on the right of the power switch.
4. Use a flat-blade screwdriver to rotate the applicable fuse holder 1/4
turn counterclockwise. The fuse holder releases from the socket.
50 • Replacing the Fuses
5. Remove the fuse holder from the socket, the n remove the fuse from the
fuse holder.
CAUTION: Verify that the r eplaceme nt fuse is the same t ype and rating a s
the fuse you are replacing.
6. Install the appropriate replacement fuse in the fuse holder, then place
the fuse holder in the socket.
7. Push the fuse holder into the socket, then turn the holder 1/4 turn
clockwise to secure it in the socket.
8. Plug in or turn on all incoming power to the Beacon 200 at the power
source end.
9. Place the Beacon 200’s power switch in the ON position, then verify
that the PILOT light is on.
10.Close and secure the housing door.
Preventive Maintenance
Preventive maintenance of the Beacon 200 consists of daily, monthly, and quarterly procedur es to ensure that the dete ctor head is op erating p roperly. See the detector head instruction manual(s) for preventive maintenance procedures that apply to the detector head(s)) supplied with your Beacon
200.
Preventive Maintenance • 5 1
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended act i on for problems you may encounter with th e Beacon
200.
NOTE: This troubleshooting guide describes controller problems only.
See the detector head instruction manual(s) for trouble shooting procedures that apply to the detector head(s) supplied with your Beacon 200.
Table 11: Troubleshooting the Beacon 200
Condition Symptom(s) Probable Causes Recommended Action
No Power
Frequent or Suspect Alarms
Flickering Display
The PILOT light is off.
The display screens are blank.
The Beacon 200 alerts you to frequent or suspect alarms.
The detector head’s fresh air readings remain on zero (20.9 for oxygen).
The display readings flicker often.
The power wiring is disconnected or misconnected.
One or both AC fuses or the DC fuse is blown.
The display cable is disconnected or misconnected.
The Beacon 200 is experiencing false readings due to RFI or EMI.
The detector head wiring is disconnected, misconnected, or intermittent.
The Beacon 200 is experiencing false readings due to RFI or EMI.
The noise filter setting is too low.
The zero suppression setting is too low.
The display screen is malfunctioning.
1. Verify that the wiring to the power source is correct and secure.
2. At the Beacon 200, verify that the wiring to the AC terminals is correct and secure.
3. Check the continuity of the applicable (AC or DC) fuse.
4. Verify that the display cable is connected.
The display (ribbon) cable plugs into connectors on the top edge of the display board (labeled J1) and on the top edge of the main circuit board (also labelled J1).
5. If the power difficulties continue, contact RKI for further instruction.
1. Verify that the detector head wiring is properly shielded (see the detector head instruction manual.)
2. Verify that the detector head wiring is correct and secure.
3. Verify that power and detector head wiring is routed through separate conduit hubs on the bottom of the Beacon 200 housing.
4. Increase the alarm on delay setting in the Channel Control & Setup Program.
5. If the frequent or suspect alarm difficulties continue, contact RKI for further instruction.
1. Verify that the detector head wiring is properly shielded (see the detector head instruction manual.)
2. Verify that the detector head wiring is correct and secure.
3. Verify that power and detector head wiring is routed through separate conduit hubs on the bottom of the Beacon 200 housing.
4. Increase the noise filter setting in the Channel Control & Setup Program.
5. Increase the zero suppression setting in the Channel Control & Setup Program
6. If the display difficulties continue, contact RKI for further instruction.
52 • Troubleshooting
Table 11: Troubleshooting the Beacon 200 (Continued)
Condition Symptom(s) Probable Causes Recommended Action
Buzzer not Working
Reset Switch not Working
Parts List
The buzzer does not sound an audible alarm during alarm conditions.
The buzzer sounds weak or broken.
The buzzer does not silence when you press the reset switch.
The applicable alarm circuit does not reset when you press the reset switch after an alarm condition passes.
The buzzer is disconnected or misconnected.
The buzzer is malfunctioning.
The reset switch is disconnected or misconnected.
The reset switch is malfunctioning.
1. Verify that the wiring to the BUZ- and BUZ+ terminals of the controller terminal strip is correct and secure.
2. If the buzzer difficulties continue, contact RKI for further instruction.
1. Verify that the wiring to the RESET (2) terminals of the controller terminal strip is correct and secure.
2. If the reset switch difficulties continue, contact RKI for further instruction.
This appendix list s the part n umbers and descriptio ns for replacemen t parts and accessories offered for the Beacon 200 Gas Monitor.
Table 12: Parts List, Beacon 200 Gas Monitor
Part No. Description
18-0107RK Conduit Hub (3/4 in.) 43-0440RK Reset Switch 43-4165RK AC Fuse, 1/4” x 1 1/4”, fast acting, 3A, 250 V 43-4170RK DC Fuse, 1/4” x 1 1/4”, fast acting, 6A 52-1016RK Buzzer 71-0059RK Beacon 200 Gas Monitor Operator’s Manual (this document)
Parts List • 5 3
Loading...