Rigol RP1003C, RP1004C, RP1005C User Manual

Page 1
RIGOL
User’s Guide
RP1003C/RP1004C/RP1005C
Current Probe
Jan. 2013
Page 2
Page 3
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide I
Guaranty and Declaration
Copyright
© 2013 RIGOL Technologies, Inc. All Rights Reserved.
Trademark Information
RIGOL is a registered trademark of RIGOL Technologies, Inc.
Publication Number
UGE19101-1110
Notices
RIGOL products are protected by patent law in and outside of
P.R.C.
RIGOL reserves the right to modify or change parts of or all the
specifications and pricing policies at company’s sole decision.
Information in this publication replaces all previously
corresponding material.
RIGOL shall not be liable for losses caused by either incidental or
consequential in connection with the fur ni shing, use or performance of this manual as well as any information contained.
Any part of this document is forbidden to be copied or photocopied
or rearranged without prior written approval of RIGOL.
Product Certification
RIGOL guarantees this pro duct conforms to t he national and indus trial standards in China as well as the ISO9001:2008 standard and the ISO14001:2004 standard. Other international standard conformance certification is in progress.
Contact Us
If you have any problem or requirement when using our products, please contact RIGOL Technologies, Inc. or your local distributors, or visit: www.rigol.com
.
Page 4
RIGOL
II RP1003C/RP1004C/RP1005C User’s Guide
General Safety Summary
Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injuries or damages to the instrume nt and any prod uct connected to it. To prevent potential hazards, please use the instrument only specified by this manual.
Ground The Instrument.
The instrument is grounded through the Protective Earth lead of the power cord. To avoid electric shock, it is essential to connect the earth terminal of power cord to the Protective Earth terminal before any inputs or outputs.
Observe All Terminal Ratings.
To avoid fire or shock hazard, observe all ratings and markers on the instrument and check your manual for more information about ratings before connecting.
Do Not Operate Without Covers.
Do not operate the instrument with covers or panels removed.
Avoid Circuit or Wire Exposure.
Do not touch exposed junct ions and components when the unit is powered.
Do Not Operate With Suspected Failures.
If you suspect damage occurs to the instrument , have it inspected by qualified service personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.
Page 5
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide III
Keep Well Ventilation.
Inadequate ventilation may cause increasing of temperature or damages to the device. So please keep well ventilated and inspect the intake and fan regularly.
Do Not Operate in Wet Conditions.
In order to avoid short circuiting to the interior of the device or electric shock, please do not operate in a humid environment.
Do Not Operate in an Explosive Atmosphere.
In order to avoid damages to the device or personal injuries, it is important to operate the device away from an explosive atmosphere.
Keep Product Surfaces Clean and Dry.
To avoid the influence of dust and/or moisture in air, please kee p t h e surface of device clean and dry.
Electrostatic Prevention.
Operate in an electrostatic discharge protective area environment to avoid damages induced by static discharges. Always ground both the internal and external conductors of the cable to release static before connecting.
Page 6
RIGOL
IV RP1003C/RP1004C/RP1005C User’s Guide
Safety Terms and Symbols
Terms in this Manual. These terms may appear in this manual:
WARNING
Warning statements indicate the conditions or practices that could result in injury or loss of life.
CAUTION
Caution statements indicate the conditions or practices that could result in damage to this product or other property.
Terms on the Product. These terms may appear on the Product:
DANGER
indicates an injury or hazard may immediately happen.
WARNING indicates an injury or hazard may be accessible potentially. CAUTION indicates a potential damage to the instrument or other
property might occur.
Symbols on the Product. These symbols may appear on the product:
Double
Insulation
Safety
Warning
Protective
Earth Terminal
Chassis
Ground
Test
Ground
Page 7
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide V
Precautions
To avoid short circuits and potentially life-threatening hazards,
never connect the current probe to a circuit that operates at greater than the maximum rated voltage or over bare conductors.
When conductors being measured carry in excess of the safe
voltage level and not more than the maximum rated voltage, to prevent short circuits and electric shock while the core section is open, make sure that conductors to be measured are insulated with material conforming to the requirements of measurement category, working voltage and pollution degree.
Isolate the output terminal of the probe from other terminals on
the measuring instrument using basic insulation conforming to the requirements of measurement category, working voltage and pollution degree.
The core and shield case of the current sensor are not insulated.
Please do not use the current sensor on bare conductors.
Be careful to avoid damaging the insulation surface while taking
measurements.
Please use the power supply provided with this product or other
power supply conforming to the specifications of this product. Make sure that the power supply has a protective earthing with double-insulation construction.
Make sure that the m easuri ng instrume nt connected to the o utput
terminal of the current probe is equipped with a protective earthing with double-insulation construction.
Core
Page 8
RIGOL
VI RP1003C/RP1004C/RP1005C User’s Guide
If the measuring instrument connected to the output terminal
(BNC) of the current probe is equipped with any other measurement terminals, take the following precautions to ensure that the other instrument does not form a bridge between the probe and any hazardous live of a part. a) Isolate the terminal of the probe from other terminals on the
measuring instrument using basic insulation conforming to the requirements of measurement category, working voltage and pollution degree.
b) If basic insulation requirement cannot be met between the
terminal to which the current probe is connected and other terminals on the measuring instrument, make sure that the voltage input to the measurement terminal does not exceed the safe voltage level.
When using a measuring instrument t hat do es no t pr ovide
isolation between its input terminals and chassis or other input terminals, please do not connect the ground terminal to any non-ground potential. Otherwise, short-c ircuit current will flow through the power adaptor or current probe from the ground terminal, which could cause an electrical accident or damage.
RP1003C and RP1004C:
×
Page 9
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide VII
RP1005C:
To avoid electric shock, do not allow the device get wet and
do not take measurement with wet hands.
To avoid electric shock when measuring live lines, wear
appropriate protective gear, such as insulated gloves, boots and a safety helmet.
The device is not designed to be entirely water-proof or
dust-proof. To avoid damage, do not use it in a wet or dusty environment.
The current sensor is a precision assembly including a molded
component, a ferrite core and a Hall effect element. It may be damaged if subjected to sudden changes in ambient temperature, or mechanical strain or shock.
The mating surfaces of the current sensor are precisely
ground and should be treated wi th car e. If t hese su rfaces ar e scratched, the performance may be impaired.
Measurements are degraded by dirt on the mating surfaces
of the current sensor, so keep the surfaces clean by gently wiping with a soft cloth.
To avoid damaging the current probe cable, do not bend or
pull the cable.
Keep the current sensor closed, except when clamping it
onto the conductor to be measured. The mating su rf ace o f the core ca n b e scra tch ed when it is op en .
Avoid stepping on or pinc hing the ca ble , whic h co uld dama ge
×
Page 10
RIGOL
VIII RP1003C/RP1004C/RP1005C User’s Guide
the cable insulation.
Keep the cables well away from heat sources, as bare
conductors could be exposed if the insulation melts.
Correct measurement may be impossible in the presence of
strong magnetic fields, such as near transformers and high-current conductors, or in the presence of strong electromagnetic fields such as near radio transmitters.
Do not store or use the device where it could be exposed to
direct sunlight, high temperature, humidity or condensation. Under such conditions, the device might be damaged and insulation may deteriorate so that it no longer meets specifications.
Page 11
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide IX
General Inspection
1. Inspect the shipping container for damage
Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the probe has passed both electrical and mechanical tests.
The consigner or carrier shall be liable for the damage to probe resultin g from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the unit.
2. Inspect the probe
In case of any damage, or defect, or failu re, notify your RIGOL sales representative.
Page 12
RIGOL
X RP1003C/RP1004C/RP1005C User’s Guide
Contents
Guaranty and Declaration .......................................................... I
General Safety Summary ......................................................... II
Safety Terms and Symbols ....................................................... IV
Precautions ............................................................................... V
General Inspection .................................................................. IX
Current Probe Overview ........................................................... 1
To Use the Current Probe .......................................................... 7
Measurement Preparations ......................................................... 7
Demagnetizing .......................................................................... 8
Zero Adjustment ...................................................................... 10
Measurement Steps ................................................................. 10
Precautions during Measurements ............................................. 12
Specifications .......................................................................... 16
Technical Parameters ............................................................... 16
General Specifications .............................................................. 19
Appendix ................................................................................. 21
Appendix 1 Amplitude Characteristics ........................................ 21
Appendix 2 Relation between Max Input Current and Frequency . 23
Appendix 3 Input Impedance (Typical) ...................................... 25
Page 13
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide 1
Current Probe Overview
RP1003C, RP1004C and RP1005C current probes can detect the current flowing through the conductor and convert it to voltage that can be displayed on and measured by the measuring instrument.
Main Features:
Highly accurat e c urrent detection Easy current measurement Br oadband frequency characteristics (RP1003C: DC to 50MHz;
RP1004C: DC to 100MHz; RP1005C: DC to 10MHz)
Compact design, permits measurement of low current levels Easy protect function at excessive input Unique thin film Hall effect element
Page 14
RIGOL
2 RP1003C/RP1004C/RP1005C User’s Guide
RP1003C Parts Overview
Figure 1 RP1003C Parts
Note: For more information of parts 1 to 7 in the figure above, refer to the introductions on page 4.
Current Direction
Indication
On/Off Indication
Terminator
Power Supply
Cable
Sensor Cable
5
2
7
4
3
1
6
Gain
Page 15
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide 3
RP1004C Parts Overview
Figure 2 RP1004C Parts
Note: For more information of parts 1 to 7 in the figure above, refer to the introductions on page 4.
Gain
Current Direction Indication
On/Off Indication
Terminator
Power Supply
Cable
Sensor Cable
5
2
7
4
3
1
6
Page 16
RIGOL
4 RP1003C/RP1004C/RP1005C User’s Guide
RP1005C Parts Overview
Figure 3 RP1005C Parts
1 Slide Switch
It is used to open and lock the current sensor. You are recommended to lock the current sensor when measuring the conductor to be measured to avoid danger. For RP1003C and RP1004C, there are OPEN, FREE and LOCK indications on one side o f the slide switch. The on/off status o f the current sensor is related to the position of the slide switch. When the slide switch is at the OPEN position, the current
sensor is open and at this point, the conductor to be measured can be connected to the current sensor;
When the slide switch is at the FREE position, the current
sensor is closed but not locked;
When the slide switch is at the LOCK position, the current
Current Direction
Indication
On/Off Indication
Terminator
Power Supply Cable
Sensor Cable
2
7
4
3
1
6
Gain
Page 17
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide 5
sensor is locked and at this point, the UNLOCK indication is covered (cannot be seen).
For RP1005C, there are LOCK and UNLOCK indications on the slide
switch. The current sensor is locked when the LOCK indication is displayed on the slide switch (the UNLOCK indication disappears).
2 Current Sensor
It is used to clamp the co nductor to be measured and carries out the actual current measurement. It is a precision assembly including a molded component, a ferrite core and a Hall effect element. It may be damaged if subjected to sudden changes in ambient temperature, or mechanical strain or shock, and therefore great care should be taken in handling it.
3 Demagnetizing Switch
It is used to demagnetizing the core if it has been magnetized by switching the power on and off, or by an excessive input. Always carry out demagnetizing before measurement. The demagnetizing process takes about one second. During demagnetizing, the current probe outputs a demagnetizing waveform.
4 Zero Adjustment Dial
Use the zero adjustment dial to correct for the effect of a DC
voltage offset or temperature drift on t he dev ice. W hen b egin nin g measurement, after demagnetizing always carry out zero adjustment to adjust the baseline to zero.
5 Coarse Adjustment Trimmer (only applicable to RP1003C
and RP1004C)
When adjustment is not possible within the range of zero adjustment dial, use a nonconductive screwdriver to (such as
Page 18
RIGOL
6 RP1003C/RP1004C/RP1005C User’s Guide
ceramic screwdriver) to adjust the baseline to the adjustable range of the zero adjustment dial via the coarse adjustment trimmer; then, use the zero adjustment dial to adjust the baseline to zero.
6 Output Connector
This connector can be directly connected to the BNC input terminal of the measuring instrument (such as oscilloscope or recorder), or be connected to other measuring instrument via BNC-to-Banana plug adaptor or similar adaptor. At this point, the current waveform of the conductor under measurement is output at a constant gain. Note: a) For RP1003C and RP1004C, 0.1V/A on the output connector i s
the gain when the input impedance of the measuring instrument is 1MΩ; the corresponding gain is 1/2×0.1V/A=0.05V/A when the input impedance of the measuring instrument is 50Ω.
a) For RP1005C, 0.01V/A on the output connector is the gain
when the input impedance of the measuring instrument is 1MΩ; the corresponding gain is 1/2×0.01V/A=0. 005V/A when the input impedance is 50Ω.
7 Power Plug
Connect the power plug to the corresponding receptacle of the power adaptor to supply power to the current probe terminator. Note: To avoid electric shock, do not touch the portion beyond the protective barrier during use.
Page 19
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide 7
To Use the Current Probe
Before using the current probe to measure the current of the device under measurement, make measurement preparations and perform demagnetizing and zero adjustment. Then, make measurements according to the method described in Measurement Steps. Note: You are recommended to read the Precautions during Measurements before making measurements to avoid unnecessary loss.
Measurement Preparations
1. Connect the power plug of the current probe to the power adaptor.
2. Turn off the power adaptor and connect it to the power supply. Note: The voltage of the power supply used should match the rated supply voltage of the power adaptor.
3. Turn on the power adaptor and check whether the powe r indicator lights.
Power Indicator
Page 20
RIGOL
8 RP1003C/RP1004C/RP1005C User’s Guide
Demagnetizing
The core can be magnetized by switching the power on and off, or by an excessive input. Therefore, perform demagnetizing before making measurements.
1. With the input of the measuring instrument at ground, set the coupling of the measuring instrument to GND and adjust the baseline to the zero position.
2. Set the coupling of the measuring instrument to DC.
3. Connect the output connector of the current probe to the BNC input terminal of the measuring instrument. Turn the collar of the output connector as show n in the figur e below and check w hethe r it is locked securely.
Note: When disconnecting the output connector and the measuring instrument, be sure to release the lock by rotating the collar in the opposite direction of the arrow as shown in the figure above before pulling off the connector. Forcibly pulling the connector without releasing the lock or pulling on the cable can damage the terminator.
4. Make sure the current sensor is locked (for RP1003C and RP1004C,
Collar
Page 21
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide 9
the slide switch should be at the LOCK position; for RP1005C, LOCK should be displayed on the s li de switch a nd UNLOCK should disappear.
5. Turn on the power adaptor and press the demagnetizing switch on the terminator. The demagnetizing process takes about one second. During the demagnetizing, the current probe outputs a demagnetizing waveform.
Note: Do not demagne tize the curr ent prob e when it is cl amping a
conductor to be measured. Demagnetizing causes current to flo w into the conductor, which may damage the parts in the circuit to be measured.
RP1003C and RP1004C:
RP1005C:
×
×
Demagnetizing
Demagnetizing
Page 22
RIGOL
10 RP1003C/RP1004C/RP1005C User’ s Guide
Zero Adjustment
As the voltage offset and temperature drift might affect the position of the baseline of the current probe, perform zero adjustment after performing demagnetizing when beginning measurements.
1. Turn the zero adjustment dial on the terminator to adjust the baseline to the zero position.
2. When adjustment is not possible within the range of zero adjustment dial, use a nonconductive screwdriver to (such as ceramic screwdriver) to adjust the baseline to the adjustable range of the zero adjustment dial via the coarse adjustment trimmer; then, use the zero adjustment dial to adjust the baseline to the zero position (only applicable to RP1003C and RP1004C).
Measurement Steps
Make measurements after the above operations are finished.
1. Open the current se n s or by pushing the slide switch in the direction of the arrow as shown in the figure below (for RP1003C and RP1004C, the slide switch sho uld be at the OPEN position; for RP1005C, UNLOCK should be displayed on the slide switch and LOCK should disappear).
2. Align the current sensor so that the current direction indication corresponds to the direction of current flowing through the
Slide Switch
Page 23
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide 11
conductor to be measured. Clamp the conductor to be measured so that the conductor is in the center of the sensor aperture.
3. Lock the current sensor by pushing the slide switch in the direction of the arrow as shown in the figure below (for RP1003C and RP1004C, the slide switch should be at the LOCK position; for RP1005C, you need to first press the current probe to close the current sensor and then push the slide switch until LOCK is displayed and UNLOCK disappears).
4. At this point, you can view the voltage amplitude of the waveform measured by the measuring instrument and calculate the actual current amplitude (current amplitude=voltage amplitude measured/gain of the current probe). For example, when the voltage amplitude measured is 2V and the gain of the current probe is 0.1V/A, the actual current am plitude is 2V/(0.1V/A)=20A.
Note: The current sensitivity can be derived from the volta g e sensitivity of the measuring instrument. The current sensitivity equals the ratio of the voltage sensitivity of the measuring instrument to the gain of the current probe. For example, when the voltage sensitivity of the measuring instrument is 0.001V/div
Slide Switch
I
Current Direction Indication
Page 24
RIGOL
12 RP1003C/RP1004C/RP1005C User’ s Guide
and the gain of the curren t probe i s 0.1V/A, the current sensitivity is (0.001V/div)/(0.1V/A)=0.01A/div.
Precautions during Measurements
1. The curre nt consumption of the current probe depends on the current to be measured. Make sure that the total current consumption of the current probes do not exceed the rated output current of the power adaptor when multiple current probes are connected to the same power adaptor. The figure below is the relation curve between the output current and current consumption.
f=50 Hz
AC
0
100
200
300
400
500
600
0
10
20 30
40
Current: A
Current Consumption: mA
DC
Note: The current consumption is the a lgebraic sum of the positive and negative current consumption.
2. The maximum continuous input range is bas ed on the heat that is internally generated during the meas ureme nt. Note that the inp ut current should not exceed this range; otherwise, the current probe might be damaged.
Page 25
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide 13
3. The linear response range of the input current varies according to the frequency of the current being measured. For the variation relation curve between them, refer to Appendix 2 Relation between Max Input Current and Frequency.
4. If excess current is input, generated heat activates a built-in safety function that blocks normal output. If this happens, disconnecti ng the current sensor and the conductor under measurement or reduce the input current to zero. Wait until the current sensor has had sufficient time to cool before resuming operation. Note: Even if the input current does not exceed the maximum continuous input range, continuous input for an extended period of time may result in activation of the safety circ uit to prevent damage resulting from heating of the current sensor.
5. At high ambient temperatures, the built-in safety circuit may activate at current input levels below the maximum continuous input range.
6. Continuous input of current exceeding the maximum continuous input range or repeated safety circuit activation will degrade the performance of the safety circuit, possibly resulting in damage to the device.
7. Do not place any unclamped conductor with an electric current of a frequency of 10kHz or higher near t he current senso r (as shown in the figures on the next page). Current flowing in the conductor nearby may heat up the sensor and cause its temperature to rise, leading to damage to the current sensor.
Page 26
RIGOL
14 RP1003C/RP1004C/RP1005C User’ s Guide
RP1003C and RP1004C:
RP1005C:
8. Be sure to use the sl id e switch to open the current sen s or . For RP1003C and RP1004C, please do not press the upper core in the direction of the arrow as shown in the figure below to avoid damaging the internal structure of the slide switch.
9. Immediatel y aft er po w er ing on , the device may b e subject to an appreciable DC drift due to the effect of self-heating. To counteract this, allow the current probe to warm up for more tha n 30 minutes before carrying out measurements.
10. Under certain circumstances, oscillation may occur when the probe is connected to the power adaptor when the power supply is on. This does not indicate a malfunction. Oscillation can be stopped and operation restored to normal by opening and closing the current sensor.
11. When the frequency of the input signal is relatively high, the common mode voltage range corresponding to the signal measured by the current probe reduces. At this point, connect the current probe to the low-voltage side of the circuit, as shown in the figures on the next page.
Upper Core
Page 27
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide 15
RP1003C and RP1004C:
RP1005C:
Low
Signal
Source
Load
High
×
Load
×
Low
High
Signal Source
Page 28
RIGOL
16 RP1003C/RP1004C/RP1005C User’ s Guide
Specifications
The following parameters are applicable when the instrument has been working for more than 30 minutes under the specif ie d op erating temperature (10℃ to 30) and humidity (0%RH t o 80%RH).
Technical Parameters
RP1003C:
Bandwidth
DC to 50MHz (-3dB), refer to Appendix 1 Amplitude Characteristics (RP1003C)
Rise Time
≤7ns
Linear Response Range of Input Current
30Arms, refer to Appendix 2 Relation between Max Input Current and Frequency (RP1003C)
Maximum Peak
Current
50A peak, non-continuous
Gain 0.1V/A
Gain Accuracy
±1.0%rdg±1mV, ≤30A ±2.0%rdg, 30A to 50A peak non-continuous (DC and 45Hz to 66Hz, within the linear res ponse range of the input current)
Noise
2.5mArms (for 20MHz band measuring instrument)
Input Impedance
Refer to
Appendix 3 Input Impedance
(Typical) (RP1003C)
Gain Accuracy Drift
±2% (0 to 40 temperature, input of
Page 29
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide 17
50Hz, 30A)
Rated Power 5.6W Supply Voltage +12V±0.5V Maximum Rated Voltage
300V, CAT I (insulated conductor)
Gain Accuracy Period
1 year (opening/closing up to 10,000 times)
Effect of External Magnetic Fields
≤20mA (DC and 60Hz, magnetic field of 400A/m)
RP1004C:
Bandwidth
DC to 100MHz (-3dB), refer to Appendix 1 Amplitude Characteristics (RP1004C)
Rise Time ≤3.5ns Linear Response
Range of Input Current
30Arms, refer to Appendix 2 Relation between Max Input Current and Frequency (RP1004C)
Maximum Peak Current
50A peak, non-continuous
Gain 0.1V/A
Gain Accuracy
±1.0%rdg±1mV, 30A ±2.0%rdg, 30A to 50A peak (DC and 45Hz to 66Hz, within the linear response range of the input current)
Noise
≤2.5mArms (for 20MHz band measuring instrument)
Input Impedance
Refer to Appendix 3 Input Impedanc e (Typical) (RP1004C)
Gain Accuracy Drift
±2% (0 to 40 temperature, input of
Page 30
RIGOL
18 RP1003C/RP1004C/RP1005C User’ s Guide
50Hz, 30A)
Rated Power 5.3W Supply Voltage +12V±0.5V Maximum Rated Voltage
300V, CATI (insulated conductor)
Gain Accuracy Period
1 year (opening/closing up to 10,000 times)
Effect of External Magnetic Fields
5mA (DC and 60Hz, magnetic field of 400A/m)
RP1005C:
Bandwidth
DC to 10MHz (-3dB), refer to
Appendix 1
Amplitude Characteristics (RP1005C)
Rise Time
≤35ns
Linear Response
Range of Input Current
150A, refer to
Appendix 2 Relation
between Max Input Current and Frequency
(RP1005C)
Maximum Peak Current
300A peak, non-continuous 500A peak, pulse width≤30μs
Gain
0.01V/A
Gain Accuracy
±1.0%rdg±1mV, 150A ±2.0%rdg, 150A to 300A peak (DC and 45Hz to 66Hz)
Noise
≤25mArms (for 20MHz band measuring instrument)
Input Impedance
Refer to
Appendix 3 Input Impedance
(Typical) (RP1005C)
Gain Accuracy Drift
±2% (0 to 40 temperature, input of 55Hz, 150A)
Rated Power 5.5W
Page 31
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide 19
Supply Voltage +12V±1V Maximum Rated Voltage
600V: CATII; 300V: CATIII (insulated conductor)
Gain Accuracy Period
1 year (opening/closing up to 10,000 times)
Effect of External
Magnetic Fields
150mA (DC and 60Hz, magnetic field of 400A/m)
General Specifications
Probe Dimensions
Current Sensor
RP1003C: approx.
175mmx18mmx32mm RP1004C: approx. 175mmx18mmx32mm RP1005C: approx.
176mmx69mmx34mm
Terminator
RP1003C: approx.
6.9mmx2.5mmx1.6mm RP1004C: approx.
6.9mmx2.5mmx1.6mm RP1005C: approx.
6.9mmx2.5mmx1.6mm
Maximum
Dimensions of Conductor to be
Measured
RP1003C: approx. 5mm
RP1004C: approx. 5mm RP1005C: approx. 20mm
Cable Length
Current
Sensor
Cable
RP1003C: approx. 150cm
RP1004C: approx. 150cm
RP1005C: approx. 200cm
Power
Supply
Cable
RP1003C: approx. 100cm
RP1004C: approx. 100cm
RP1005C: approx. 100cm
Weight
RP1003C: approx. 230g
RP1004C: approx. 240g RP1005C: approx. 500g
Working
0℃
to +4
0℃
, 0 to 80%RH
Page 32
RIGOL
20 RP1003C/RP1004C/RP1005C User’ s Guide
Temperature and Humidity
Storage
Temperature
and Humidity
-1
0℃
to +5
0℃
, 0 to 80%RH
Location for Use
Indoor, altitude up to 2000m
Electromagnetic Compatibility
EN61326
Measurement Category
RP1003C: CAT I (anticipated transient voltage
1500V) RP1004C: CAT I (anticipated transient voltage 1500V) RP1005C: CAT II, CAT III (anticipated transient
voltage 4000V)
Pollution Degree
Degree 2
Safety
EN61010
Page 33
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide 21
Appendix
Appendix 1 Amplitude-frequency Characteristics
RP1003C:
1
100
1k
1M
100M
-60
-50
-40
-
30
-20
-10
Frequency:Hz
Amplitude dB 0dB=1V
RP1004C:
1 100
10k 1
M
100M
-10
-
20
-30
-40
-50
-60 Frequency: Hz
Amplitude dB0dB=1V
Page 34
RIGOL
22 RP1003C/RP1004C/RP1005C User’ s Guide
RP1005C:
10k 100k 1M 10M 100M1k Frequency: Hz
Amplitude dB0dB=1V
-30
-40
-80
-60
-50
-70
Page 35
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide 23
Appendix 2 Relation between Max Input Current and Frequency
RP1003C:
10 100 1k 10k 100k 1M
10M
100M
35 30
25 20
15 10
5 0
Frequency: Hz
Max Input Current: A
RP1004C:
10 100 1k 10k 100k 1M
10M
100M
35 30
25 20
15 10
5 0
Frequency: Hz
Max Input Current: A
Page 36
RIGOL
24 RP1003C/RP1004C/RP1005C User’ s Guide
RP1005C:
1
k 10k 1M 10M
10
100k
100
0
50
100
150
Max Input Current: Arms
200
Frequency: Hz
Page 37
RIGOL
RP1003C/RP1004C/RP1005C User’s Guide 25
Appendix 3 Input Impedance (Typical)
RP1003C:
1k
10k 1M 10M 100
M100k
100
1
m
10m
100m
1
Input Impedance: Ω
10
Frequency: Hz
RP1004C:
100
1k
10k
1M 10M
100M
100k
1m
10m
100m
1
10
Frequency: Hz
Input Impedance: Ω
Page 38
RIGOL
26 RP1003C/RP1004C/RP1005C User’ s Guide
RP1005C:
10k
1M
10
M
100k
0.001
0
.01
0.
1
1
Input Impedance: Ω
Frequency: Hz
Loading...