RIGOL is a registered trademark of RIGOL TECHNOLOGIES, INC.
Publication Number
UGD07112-1110
Software Version
DSA815:00.01.18
DSA832/DSA875/DSA832E:00.01.05
Software upgrade might c hange or add prod uct feat ures. Please a cqui re the la test
version of the manual from RIGOL website or contact RIGOL to upgrade the
software.
Notices
RIGOL products are covered by P.R.C. and foreign patents, issued and
pending.
RIGOL reserves the right to modify or change parts of or all the
specifications and pricing policies at the company’s sole decision.
Information in this publication replaces all previously released materials.
Information in this publication is subject to change without notice.
RIGOL shall not be liable for either incidental or consequential losses in
connection with the furnishing, use, or performance of this man ual, as well as
any information contained.
Any part of this document is forbidden to be copied, ph otocopied, or
rearranged without prior written approval of RIGOL.
Product Certification
RIGOL guarantees that this product conforms to the national and industrial
standards in China as wel l as the ISO9 001:2 008 stan dard an d the IS O1400 1: 200 4
standard. Other international standard conformance certifications are in progress.
Contact Us
If you have any problem or requirement when using our products or this manual,
please contact RIGOL.
E-mail: service@rigol.com
Website: www.rigol.com
User’s Guide for DSA800/E Series
RIGOL
II
Safety Requirement
General Safety Summary
Please review the following safety precautions carefully before putting the
instrument into operation so as to avoid any personal injury or damage to the
instrument and any product connected to it. To prevent potential hazards, please
follow the instructions specified in this manual to use the instrument properly.
Use Proper Power Cord.
Only the exclusive power cord designed for the instrument and authorized for use
within the local country could be used.
Ground the Instrument.
The instrument is grounded through the Protective Earth lead of the power cord.
To avoid electric shock, connect the earth terminal of the power cord to the
Protective Earth terminal before connecting any input or output terminals.
Connect the Probe Correctly.
If a probe is used, the probe ground lead must be connected to earth ground. Do
not connect the ground lead to high voltage. Improper way of connection could
result in dangerous voltages being present on the connectors, controls or other
surfaces of the oscilloscope and probes, which will cause potential hazards for
operators.
Observe All Terminal Ratings.
To avoid fire or shock hazard, observe all ratings and ma rkers on the instrument
and check your manual for more information about ratings before connecting the
instrument.
Use Proper Overvoltage Protection.
Ensure that no overvoltage (such as that caused by a bolt of lightning) can reach
the product. Otherwise, the operator might be exposed to the dange r of an el ectri c
shock.
User’s Guide for DSA800/E Series
RIGOL
III
Do Not Operate Without Covers.
Do not operate the instrument with covers or panels removed.
Do Not Insert Objects Into the Air Outlet.
Do not insert objects into the air outlet, as doing so may cause damage to the
instrument.
Use Proper Fuse.
Please use the specified fuses.
Avoid Circuit or Wire Exposure.
Do not touch exposed junctions and components when the unit is powered on.
Do Not Operate With Suspected Failures.
If you suspect that any damage may occur to the instrument, have it inspected by
RIGOL authorized personnel before further operations. Any maintenance,
adjustment or replacement especially to ci rcuits or accessories must be perf ormed by
RIGOL authorized personnel.
Provide Adequate Ventilation.
Inadequate ventilation may cause an increase of t emperature in the instrument,
which would cause damage to the instrument. So please keep the inst ru ment well
ventilated and inspect the air outlet and the fan regularly.
Do Not Operate in Wet Conditions.
To avoid short circuit inside the instrument or electric shock, never operate the
instrument in a humid environment.
Do Not Operate in an Explosive Atmosphere.
To avoid personal injuries or damage to the instrument, never operat e the
instrument in an explo sive atmospher e.
Keep Instrument Surfaces Clean and Dry.
To avoid dust or moisture from affecting the performance of the instrument, keep
the surfaces of the instrument clean and dry.
User’s Guide for DSA800/E Series
RIGOL
IV
Prevent Electrostatic Impact.
Operate the instrument in an electrostatic discharge protective environment to
avoid damage induced by static discharges. Always ground both the internal and
external conductors of cables to release static before making connections.
Use the Battery Properly.
Do not expose the battery (if available) to high temperature or fire.
Keep it out of the reach of children. Improper change of a battery (lithium battery)
may cause an explosion. Use the RIGOL specified battery only.
Handle with Caution.
Please handle with care during transportation to avoid damage t o keys, knobs,
interfaces, and other parts on the panels.
User’s Guide for DSA800/E Series
RIGOL
V
CAUTION
DANGER
Hazardous
Safety
Protective
Terminal
Chassis
Test
Safety Notices and Symbols
Safety Notices in this Manua l:
WARNING
Indicates a potentially hazardous situation or practice which, if not
avoided, will result in serious injury or death.
Indicates a potentially hazardous situation or practice which, if not
avoided, could result in damage to the product or loss of important data.
Safety Terms on the Product:
It calls attention to an operation, if not cor rectly performed, could
result in injury or hazard immediately.
WARNING It calls attention to an operation, if not correctly pe rformed, could
result in
CAUTIONIt calls attention to an operation, if not correctly pe rformed, could
result in
product.
Safety Symbols on the Product:
potential injury or hazard.
damage to the product or other devices connected to the
Voltage
Warning
Earth
Ground
Ground
User’s Guide for DSA800/E Series
RIGOL
VI
Allgemeine Sicherheits Informationen
Überprüfen Sie diefolgenden Sicherheitshinweise
sorgfältigumPersonenschädenoderSchäden am Gerätundan damit verbundenen
weiteren Gerätenzu vermeiden. Zur V ermeidung von Gefahren, nutzen Sie bit te das
Gerät nur so, wiein diesem Handbuchangegeben.
Um Feuer oder Verletzungen zu vermeiden, verwenden Sie ein
ordnungsgemäßes Netzkabel.
Verwenden Sie für dieses Gerät nur das für ihr Land zugelassene und genehmigte
Netzkabel.
Erden des Gerätes.
Das Gerät ist durch den Schutzleiter im Netzkabel geerdet. Um Gefahren durch
elektrischen Schlag zu vermeiden, ist es unerlässlich, die Erdung durchzuführen.
Erst dann dürfen weitere Ein- oder Ausgänge verbunden werden.
Anschluss einesTastkopfes.
Die Erdungsklemmen der Sonden sindauf dem gleichen Spannungspegel des
Instruments geerdet. SchließenSie die Erdungsklemmen an keine hohe Spannung
an.
Beachten Sie alle Anschlüsse.
Zur Vermei dung v on F eue r o der St romschlag , bea chten Sie alle Beme rkungen un d
Markierungen auf dem Instrument. Befolgen Sie die Bedienungsanleitung für
weitere Informationen, bevor Sie weitere Anschlüsse an das Instrument legen.
Verwenden Sie einen geeigneten Überspannungsschutz.
Stellen Sie sicher, daß kein erlei Überspannung (wie z .B. durch Gewit ter verursacht )
das Gerät erreichen kann. Andernfallsb e stehtf ür den Anwender die
GefahreinesStromschlages.
Nicht ohne Abdeckung einschalten.
Betreiben Sie das Gerät nicht mit entfernten Gehäuse-Abdeckungen.
User’s Guide for DSA800/E Series
RIGOL
VII
Betreiben Sie das Gerät nicht geöffnet.
Der Betrieb mit offenen oder entfernten Gehäuseteilen ist nicht zulässig. Nichts in
entsprechende Öffnungen stecken (Lüfter z.B.)
Passende Sicherung verwenden.
Setzen Sie nur die spezifikationsgemäßen Sicherungen ein.
Vermeiden Sie ungeschützte Verbindungen.
Berühren Sie keine unisolierten Verbindungen oder Baugruppen, während das
Gerät in Betrieb ist.
Betreiben Sie das Gerät n ic h t i m Fe hlerfall.
Wenn Sie am Gerät einen Defekt vermuten, sorgen Sie dafür, bevor Sie das Gerät
wieder betreiben, dass eine Untersuchung durch RIGOL autorisiertem Personal
durchgeführt wird. Jedwede Wartung, Einstellarbeiten oder Austausch von Teilen
am Gerät, sowie am Zubehör dürfen nur von RIGOL autorisiertem Personal
durchgeführt werden.
Belüftung sicherstellen.
Unzureichende Belüftung kann zu Temperaturanstiegen und s o mit zu thermischen
Schäden am Gerät führen. Stellen Sie deswegen die Belüftung sicher und
kontrollieren regelmäßig Lüfter und Belüftungsöffnungen.
Nicht in feuc h te r Um g ebung betreiben.
Zur Vermeidung von Kurzschluß im Geräteinneren und Stromschlag betreiben Sie
das Gerät bitte niemals in feuchter Umgebung.
Nicht in explosiver Atmosphäre betreiben.
Zur Ve rm e idung von Personen- und Sachschä den ist es unumgä ngli ch, das Ger ät
ausschließlich fernab jedweder explosiven At mosphäre zu betreiben.
Geräteoberflächen sauber und trocken halten.
Um den Einfluß v on Staub und Fe uchtigkeit aus der Luft ausz uschließen, halte n Sie
bitte die Geräteoberflächen sauber und trocken.
Schutz gegen elektrostatische Entladung (ESD).
Sorgen Sie für eine elektrostatisch geschützte Umgebung, um somit Schäden und
User’s Guide for DSA800/E Series
RIGOL
VIII
Funktionsstörungen durch ESD zu vermeiden. Erden Sie vor dem Anschluß immer
Innen- und Außenleiter der Verbindungsleitung, um statische Aufla dun g zu
entladen.
Die richtige Verwendung desAkku.
Wenneine Batterieverwendet wird, vermeiden Sie hohe Temperaturen bzw. Feuer
ausgesetzt werden. Bewahren Sie es außerhalbder Reichweitevon Kindern auf.
Unsachgemäße Änderun g derBatterie (Anmerkung: Lithium-Batterie) kann zu einer
Explosion führen. VerwendenSie nur von RIGOL angegebenenAkkus.
Sicherer Transport.
Transportieren Sie das Gerät sorgfältig (Verpackung!), um Schäden an
Bedienelementen, Anschlüssen und anderen Teilen zu vermeiden.
User’s Guide for DSA800/E Series
RIGOL
IX
CAUTION
DANGER
Sicherheits Begriffe und Symbole
Begriffe in diesem Guide:
WARNING
Die Kennzeichnung WARNING beschreibt Gefahrenquellen die leibliche
Schäden oder den Tod von Personen zur Folge haben können.
Die Kennzeichnung Caution (Vor sicht) beschreibt Gefahrenquellen die
Schäden am Gerät hervorrufen können.
Begriffe auf dem Produkt:
weist auf eine Verletzung oder Gefährdung hin, die sof ort
geschehen kann.
WARNINGweist auf eine V erletzung oder Gefährdung hin , die möglicherweise
nicht sofort geschehen.
CAUTIONweist auf eine Verletzung oder Gefährdung hin und bedeutet, dass
eine mögliche Beschädigung des Instruments oder anderer
Gegenstände auftreten kann.
Symbole auf dem Produkt:
Gefährliche
Spannung
SicherheitsHinweis
Schutz-erde Gehäusemasse Erde
User’s Guide for DSA800/E Series
RIGOL
X
WARNING
Care and Cleaning
Care
Do not store or leave the instrument where it may be exposed to direct sunlight for
long periods of time.
Cleaning
Clean the instrument regularly according to its operating conditions.
1. Disconnect the instrument from all power sources.
2. Clean the external su rfaces of the instrument with a soft cloth dampened with
mild detergent or water. When cleaning the LCD, take care to avoid scarifying
it.
CAUTION
To avoid damage to the instrument, do not expose it to caustic liquids.
To avoid short-circuit resulting from moisture or personal injuries, ensure
that the instrument is completely dry before connecting it to the power
supply.
User’s Guide for DSA800/E Series
RIGOL
XI
Environmental Consideratio ns
The following symbol indicates that this product complies with the WEEE Directive
2002/96/EC.
Product End-of-Life Handling
The equipment may contain substances that could be harmf ul to the envi ronm ent or
human health. To avoid the release of such substances into the environment and
avoid harm to human health, we recommend you to recycle this product
appropriately to ensure that most materials are reused or recycled properly. Please
contact your local authorities for disposal or recycling information.
User’s Guide for DSA800/E Series
RIGOL
XII
User’s Guide for DSA800/E Series
RIGOL
XIII
DSA800/E Series Overview
DSA800/E series spectrum analyzers which are small, light and cost-effective, are
portable spectrum analyzers designed for starters. Configured with
easy-to-operate numeric keyboard, high-resolution color LCD display and various
remote communication interfaces, they can be widely used in various fields, such
as education, company research and development as well as industrial
manufacture.
8 inch (800×480 pixels) high-definition display with clear, vivid, and
easy-to-use graphical interface
Various interfaces such as LAN, USB Host, USB Device and GPIB (option)
User’s Guide for DSA800/E Series
RIGOL
XIV
Document Overview
Topics in this manual:
Chapter 1 Quick Start
This chapter introduces the front/rear panel and user interface as well as
announcements during first use of the analyzer.
Chapter 2 Front Panel Oper a tio n
This chapter gives detailed function des criptions of the front panel keys with their
associated menu keys.
Chapter 3 Remote Control
This chapter shows how to control the analyzer in remote mode.
Chapter 4 Troubleshooting&Message
This chapter lists the troubleshooting information and messages that may appear
during the use of the analyzer.
Chapter 5 Specifications
This chapter lists the specific ations and general specifications of the analyzer.
Chapter 6 Appendix
This chapter lists the options and accessories that can be ordered along with your
analyzer as well as the service and support information.
User’s Guide for DSA800/E Series
RIGOL
XV
Model
Frequency Range
Tracking Generator
Format conventions in this manual:
1. Keys:
The keys at the front panel are usually denoted by the format of “Key name
(Bold) +textbox”. For example, FREQ denotes the FREQ key.
2. Menu keys:
The menu softkeys are usually denoted by the format of “Menu word (Bold)
+character shading”. For example, Center Freq denotes the center
frequency menu item under the FREQ functi on key.
3. Connectors:
The connectors at the front or rear panel are usually denoted b y the for mat of
“Connector name (Bold) +square brackets (Bold)”, such as [GEN OUTPUT 50Ω].
4. Operation steps:
“” represents the next step of operation. For example, FREQ Center
Freq indicates pressing FREQ at the front panel and then pressing the menu softkey Center Freq.
Content conventions in this manual:
The DSA800/E series spectrum analyzer includes the following eight models. The
illustrations in this manual are based on DSA875.
DSA815 9 kHz to 1.5 GHz None
DSA832 9 kHz to 3.2 GHz None
DSA875 9 kHz to 7.5 GHz None
DSA832E 9 kHz to 3.2 GHz None
DSA815-TG 9 kHz to 1.5 GHz 1.5 GHz
DSA832-TG 9 kHz to 3.2 GHz 3.2 GHz
DSA875-TG 9 kHz to 7.5 GHz 7.5 GHz
DSA832E-TG 9 kHz to 3.2 GHz 3.2 GHz
User manuals provided with this product:
User’s Guide, Quick Guide, Programming Guide, Data sheet etc. For the desired
manual, please download it from www.rigol.com.
User’s Guide for DSA800/E Series
RIGOL Contents
XVI
Contents
Guaranty and Declaration ......................................................................... I
Safety Requirement ................................................................................. II
General Safety Summary........................................................................... II
Safety Notices and Symbols ...................................................................... V
Allgemeine Sicherheits Informationen ....................................................... VI
Sicherheits Begriffe und Sy mbole .............................................................. IX
Care and Cleaning .................................................................................... X
Environmental Considerations ................................................................... XI
DSA800/E Series Overview ................................................................. XIII
Document Overview ............................................................................. XIV
Index ............................................................................................. 1
User’s Guide for DSA800/E Series
Chapter 1 Quick Start RIGOL
1-1
Chapter 1 Quick Start
This chapter guides users to quickly get familia r with the a ppear a nce, dimensio ns,
front/ rear panel and the u ser interface, as well as announcements during first use
of DSA800/E series spectrum analyzer.
Subjects in this chapter:
General Inspection
Appearance and Dim ensions
To Prepare for Use
Front Panel
Rear Panel
User Interface
Menu Operation
Parameter Setting
To Input Filename
To Loc k the Keyboard
To Use the Built-in Help
To Use the Security Lock
To Replace the Fuse
User’s Guide for DSA800/E Series
RIGOL Chapter 1 Quick Start
1-2
General Inspection
1. Inspect the packaging
If the packaging has been damaged, do not dispose the damaged packaging
or cushioning materials until the shipment has been checked for completeness
and has passed both electric al and mechanical tests.
The consigner or carrier shall be liable for the damage to the instrument
resulting from shipment. RIGOL would not be responsible for free
maintenance/rework or replacement of the instrument.
2. Inspect the instrument
In case of any mechanical damage, m issing parts, or failure in passing the
electrical and mechanical tests, contact your RIGOL sales representative.
3. Check the accessories
Please check t he a cce ss or i es according to the packing lists. If the accessories
are damaged or incomplete, please contact your RIGOL sales
representative.
User’s Guide for DSA800/E Series
Chapter 1 Quick Start RIGOL
1-3
Appearance and Dimensions
Figure 1-1 Front View Unit: mm
Figure 1-2 Side View Unit: mm
User’s Guide for DSA800/E Series
RIGOL Chapter 1 Quick Start
1-4
To Prepare for Use
To Adjust the Supporting Legs
Users can unfold the supporting legs to use them as stands to tilt the instrument
upwards for easier operation and observation. Users can also fold the supporting
legs when the instrument is not in use for easier storage or shipment.
To unfold the supporting legs
Figure 1-3 To Adjust the Supporting Legs
To fold the supporting legs
User’s Guide for DSA800/E Series
Chapter 1 Quick Start RIGOL
1-5
CAUTION
To Connect to Power
Please use the power cord provided with the accessories to connect the spectrum
analyzer to AC power source as shown in the figure below. The AC power supply
specif ication of this spectrum analyzer is 100-240 V, 45-440 Hz. The power
consumption of the instrument cannot exceed 50 W. When the spectrum analyzer
is connected to AC power supply via the powe r c ord, the instrument select the
correct voltage range automatically and users do not need to select the voltage
range manually.
Figure 1-4 Power Cord Connection
Make sure that the instrument is properly grounded to avoid electric
shock.
User’s Guide for DSA800/E Series
RIGOL Chapter 1 Quick Start
1-6
Power-on Inspection
After connecting the instrument to power so urce correctly, press
panel to start the spectrum analyzer. Following the start-up screen which shows
the start-up initialization process information, the sweep curve is displayed.
at the front
Self-calibration
After the instrument starts, execute self-calibration.
Press System Calibrate Cal Now and the instrument will perform
self-calibration using the internal calibration source.
To Set the System Language
DSA800/E series spectrum analyzer supports mul t iple syste m languages. You can
press System Language to switch the system language.
User’s Guide for DSA800/E Series
Chapter 1 Quick Start RIGOL
1-7
1 2 3
14
13
12
11 10 9 8 7
Front Panel
The front panel of DSA800/E is as shown in the figure below.
Figure 1-5 Front Panel
Table 1-1 Front Panel Descri ption
4
5
6
NO. Description NO. Description
1 LCD 8 Tracking generator output*
2 Menu softkeys/menu control keys 9 Earphone jack
3 Function key area 10 USB Host
4 Knob 11 Power switch
5 Direction keys 12 Help
6 Numeric Keyboard 13 Print
7 RF input 14
Note: *This function is only available for DSA815-TG/DSA832-TG/DSA875-TG/DSA832E-TG.
User’s Guide for DSA800/E Series
Preset
RIGOL Chapter 1 Quick Start
1-8
Front Panel Function K eys
Figure 1-6 Function Key Area
Table 1-2 Function keys description
Key Description
FREQ
Set the center, start and stop frequency, etc; enable the
signal tracking function.
SPAN Set the frequency span of the sweep.
Set the reference level, RF attenuator, scal e and the unit of
Y-axis, etc.
AMPT
Set the reference level offset, maximum mixing level and
input impedence.
Execute auto scale and auto range as well as turn on the RF
preamplifier.
Set the resolution bandwi dth (RBW), video bandwidth (VBW)
BW/Det
and V/R ratio.
Set the detector and filter types.
Sweep/Trig Set the sweep and trigger parameters.
Trace/P/F
Set the parameters related to trace.
Configure the Pass/Fail test.
TG Set the tracking generator*.
Meas Select and control the measurement function**.
Meas Setup Set the par ameters for the selected measurement fu nction**.
User’s Guide for DSA800/E Series
Chapter 1 Quick Start RIGOL
1-9
User Key
Preset
Demod Set the demodulation function.
Marker
Marker—>
Marker Fctn
Peak
Read the amplitude, frequency and sweep time of a certain
point on the trace via marke r.
Set other system parameters on the basis of the current
marker value.
Special functions of the marker such as noise m a r ker, N dB
bandwidth measurement and frequency counter.
Open the peak search menu and search for peaks
immediately.
System Set the system parameters.
Print Setup Set the print parameters.
Storage Provide f ile storage and recall functions.
Auto Search for signals automatically within full frequency range.
User-defined shortcut key.
Restore the system to factory settings or user-defined state.
Print or save the current screen.
Help Turn on the built-in help.
Note:
*This function is only available for DSA815-TG/DSA832-TG/DSA875-TG/DSA832E-TG.
**This function is only available for DSA800/E with the corresponding option.
User’s Guide for DSA800/E Series
RIGOL Chapter 1 Quick Start
1-10
Front Panel Key Backlight
The on/off state and the color of the backlights of some keys at the front panel
indicate the working state of the spectrum analyzer. The states are as listed below.
1. Power Switch
Flash on and off alternatively, in breathing state: indicate the unit is in
stand-by state.
Constant on: indicate the instrument is in n ormal operating state.
2.TG*
When the TG function is enabled, the backlight of TG turns on and turns off
when the function is disabled.
3. Auto
When Auto is pressed, the backlight turns on. The instrument starts sweeping
within the full frequency range, searches for the signal with the maximum
amplitude and moves it to the center of the screen. Then the backlight tu rns off .
4.Meas**
The backlight of Meas turns on when VSWR or any of the advanced
measurements is enabled and stays on until all measurement functions are
disabled.
Note:
*This function is only available for DSA815-TG/DSA832-TG/DSA875-TG/DSA832E-TG.
**This function is only applicable to DSA800/E installed with the corresponding option.
User’s Guide for DSA800/E Series