DS2000 series digital oscilloscope consists of DS2202, DS2102 and DS2072.
Instrument Memory
This section contains inf ormation on the types of memory av ailable in y our instrument. It e xplains the
size of memory, how it is used, its location, volatility and the clearing procedure.
Instrument memory:
Memory
type and
size
SDRAM
32MB
NOR
FLASH
16MB
FRAM
2kB
DDR2
32MB*16B
*3
SRAM
512kB*36B
*2
Writable
during
normal
operation
Yes No Program Boot Main board
No Yes Firmware,
Yes Yes Settings Program Main board
Yes No Acquire
Yes No Display
Data
retained
when
powered
off
Purpose/
contents
Bin,
Internal
saving
memory
memory
Data
input
method
Firmware
update,
Internal
saving,
ref/setup
/pf
Acquire
system
Display
system
Location in
instrument
and
remarks
in CPU area
Main board
in CPU area
in CPU area
Main board
in acquire
area
Main board
in display
area
Sanitization
procedure
Cycle power
Security clear or
save files to C:
Default
Cycle power
Cycle power
Declassification Guide for Digital Oscilloscope
Page 4
RIGOL
2
Position of Instrument Memory on Main Board
Security Clear Procedure
To clear the memories, press StorageDisk.ManageSecurity ClearOK and the
system will be restored to the default settings.
Declassification Guide for Digital Oscilloscope
Page 5
RIGOL
3
DS4000 Series
DS4000 series digital oscilloscope consists of DS4054, DS4052, DS4034, DS4032, DS4024, DS4022,
DS4014 and DS4012.
Instrument Memory
This section contains inf ormation on the types of memory av ailable in y our instrument. It e xplains the
size of memory, how it is used, its location, volatility and the clearing procedure.
Instrument memory:
Memory
type and
size
SDRAM
32MB
NOR
FLASH
16MB
FRAM
2kB
DDR2
32MB*16B
*17
SRAM
512kB*36
B*8
Writable
during
normal
operation
Yes No Program Boot Main board
Yes Yes Firmware,
Yes Yes Settings Program M a i n board
Yes No Acquire
Yes No Display
Data
retained
when
powered
off
Purpose/
contents
Bin,
Internal
saving
memory
memory
Data
input
method
Firmware
update,
Internal
saving
ref/setup/
pf
Acquire
system
Display
system
Location in
instrument
and
remarks
in CPU area
Main board
in CPU area
in CPU area
Main board
in acquire
area
Main board
in display
area
Sanitization
procedure
Cycle power
Security clear
or Save files to
C:
Default
Cycle power
Cycle power
Declassification Guide for Digital Oscilloscope
Page 6
RIGOL
4
Position of Instrument Memory on Main Board
Security Clear Procedure
To clear the memories, press StorageDisk.ManageSecurity ClearOK and the
system will be restored to the default settings.
Declassification Guide for Digital Oscilloscope
Page 7
RIGOL
5
DS6000 Series
DS6000 series digital oscilloscope consis ts of DS6104, DS6102, DS6064 and DS60 62.
Instrument Memory
This section contains inf ormation on the types of memory av ailable in y our instrument. It explains the
size of memory, how it is used, its location, volatility and the clearing procedure.
Instrument memory:
Memory
type and
size
SDRAM
32MB
NAND
FLASH
1GB
NOR
FLASH
16MB
FRAM
2kB
DDR2
32MB*16
B*17
QDR2
22MB*18
B*8
Writable
during
normal
operation
Yes No Program Boot Main board
Yes Yes Internal
No Yes Firmware, Bin Firmware
Yes Yes Setting Program Main board
Yes No Acquire
Yes No Display
Data
retained
when
powered
off
Purpose/
contents
savings,
setup/wfm/ref
/pf/trc
memory
memory
Data
input
method
Internal
saving
update
Acquire
system
Display
System
Location in
instrument
and
remarks
in CPU area
Main board
in CPU area
Main board
in CPU area
in CPU area
Main board
in acquire
area
Main board
in draw area
Sanitization
procedure
Cycle power
Save files to
internal C:
No user data
Default
Cycle power
Cycle power
Declassification Guide for Digital Oscilloscope
Page 8
RIGOL
6
Position of Instrument Memory on Main Board
Security Clear Procedure
To clear the memories, press SAVEDisk.ManageSecurity ClearOK an d the
system will be restored to the default settings.
Declassification Guide for Digital Oscilloscope
Page 9
RIGOL
7
DS1000B Series
DS1000B series digital oscilloscope consists of DS1204B, DS1104B and DS1074B.
Instrument Memory
This section contains inf ormation on the types of memory av ailable in y our instrument. It explains the
size of memory, how it is used, its location, volatility and the clearing procedure.
DS1000CA digital oscilloscope consists of DS1302CA, DS1 202CA, DS1102CA and DS 1072CA.
Instrument Memory
This section contains inf ormation on the types of memory av ailable in y our instrument. It explains the
size of memory, how it is used, its location, volatility and the clearing procedure.
DS1000E series digital oscilloscope consists of DS1102E and DS1052E.
Instrument Memory
This section contains inf ormation on the types of memory av ailable in y our instrument. It explains the
size of memory, how it is used, its location, volatility and the clearing procedure.
DS1000D series digital oscilloscope c ons is ts of DS1 102D and DS1052D.
Instrument Memory
This section contains inf ormation on the types of memory av ailable in y our instrument. It explains the
size of memory, how it is used, its location, volatility and the clearing procedure.
Instrument memory:
Memory
type and
size
SDRAM
16MB
NOR
FLASH
8MB
FRAM
1kB
SRAM
512kB
SRAM
512kB
Writable
during
normal
operation
Yes No Program Boot Main boar d in
No Yes Firmware,
Yes Yes Settings Program Main board in
Yes No wfm points Acquire Main board in
Yes No LA points Acquire Main board in
Data
retained
when
powered
off
Purpose/
contents
Bin,
Internal
saving
Data
input
method
Firmware
update,
Internal
saving,
ref/setup
/pf/
record
/wfm
Location in
instrument
and remarks
CPU area
Main boar d in
CPU area
CPU area
acquire area
acquire area
Sanitization
procedure
Cycle power
Save files to
inter
Default
Cycle power
Cycle power