(1) 检查 U 盘是否可以正常工作。
(2) 确认使用的为 Flash 型 U 盘,本仪器不支持硬盘型 U 盘。
(3) 确认使用的 U 盘容量是否过大,本示波器推荐使用不超过 8 GBytes 的 U 盘。
(4) 重新启动仪器后,再插入 U 盘进行检查。
(5) 如果仍然无法正常使用 U 盘,请与 RIGOL 联系。
RIGOL is a registered trademark of RIGOL Technologies, Inc.
Publicatio n Number
QGA15103-1110
Notices
RIGOL products are pr o t ec t ed by patent law in and outside of P.R.C.
RIGOL reserves the right to modify or change parts of or all the specifications and pricing
policies at company’s sole decision.
Information in this publication replaces all previously corresponding materials.
RIGOL shall not be liable for losses caused by either incidental or consequential in
connection with the furnishing , use or performa nce of this manual as well as any inf ormation
contained.
Any part of this document is forbidden to be copied or photocopied or rearranged without
prior written approval of RIGOL.
Product Certification
RIGOL guarantees this product conforms to the national and industrial standards in China as well
as the ISO9001:2008 standard and the ISO14001:2004 standard. Other international standard
conformance certification is in progress.
Contact Us
If you have any pro blem o r requ irement when usin g our produc ts or this manual , ple ase c ont act
RIGOL.
E-mail: service@rigol.com
Website: www.rigol.com
MSO4000/DS4000 Quick Guide
RIGOL
II
Safety Requirement
General Safety Summary
Please review the following safety precautions carefully before putting the instrument into
operation so as to avoid any personal injuries o r damages to the instrum ent and any product
connected to it. To prevent po tential hazards, please use t he instrument only specifi ed by this
manual.
Use Proper Power Cord.
Only the power cord designed for the instrument and authorized by local country could be used.
Ground The Instrument.
The instrument is grounded through the Protective Earth lead of the power cord. To avoid electric
shock, it is essential to connect the earth terminal of power cord to the Protective Earth terminal
before any inputs or outputs.
Connect the Probe Correctly.
Do not connect the ground lead to high voltage since it has the isobaric electric potential as ground.
Observe All Terminal Ratings.
To avoid f ire o r sh ock ha zard, obser ve al l ratings and m arkers on t he in strum ent and check your
manual for more information about r a t i ngs be fore connecting.
Use Proper Overvoltage Protection.
Make sure that no overvoltage (such as that caused by a thunderstorm) can reach the product, or
else the operator might expose to danger of electrical shock.
Do Not Operate Without Covers.
Do not operate the instrument with cover s or panels removed.
Avoid Circuit or Wire Exposure.
Do not touch exposed junctions and components when the unit is powered.
Do Not Operate With Suspected Failures.
If you suspect damage occurs to the instrument, have it inspected by qualified service personnel
before further operations. Any maintenance, adjustment or replacement especially to circuits or
accessories must be performed by RIGOL authorized personnel.
Keep Well Ventilation.
Inadequate ventilation may cause increasing of temperature or damages to the device. So please
keep well ventilated and inspect the intake and fan regularly .
Do Not Operate in Wet Conditions.
In order to avoid sho rt circuiting to the in terior of the device o r electric shock, please do not
MSO4000/DS4000 Quick Guide
RIGOL
III
operate in a humid environment.
Do Not Operate in an Explosive Atmosphere.
In order to avoid damages to the device or personal injuries, it is important to operate the device
away from an explosive atmosphere.
Keep Product Surfaces Clean and Dry.
To avoid the influence of dust and/or moisture in air, please keep the surface of device clean a nd
dry.
Electrostatic Prevention.
Operate in an electrostatic dischar ge protective ar ea environm ent to avoid d amages indu ced by
static discharges. Always ground both the internal and external conductors of the cable to release
static before connecting.
Handling Safety.
Please handle w ith care during transpo rtation to avoid damages t o buttons, knob inter faces and
other parts on the panels.
Safety Terms and Symbols
Terms in this Manual. The following terms may appear in this manual:
WARNING
Warning statements indicate the conditions or practices that could result in injuries or
loss of life.
CAUTION
Caution statements indicate the conditions or practices that could result in damage to
this product or loss of data.
Terms on the Product. The following terms may appear on the product:
DANGER indicates an injury or hazard may immediately happen.
WARNING indicates an injury or hazard may be accessible potentially.
CAUTION indicates a potential damage to the instrument or other property might occur.
MSO4000/DS4000 Quick Guide
RIGOL
IV
Symbols on the Product. The following symbols may appear on the product:
Hazardous
Voltage
Safety Warning
Protective
Earth
Terminal
Chassis
Ground
Test
Ground
Measurement Category
Measurement Category
MSO4000/DS4000 series digital oscilloscopes can make measurements in Measurement Category I.
WARNING
This oscilloscope can only be used for measurements within its specified measurement
categories.
Measurement Category Definitions
Measurement category I is for measurements performed on circuits not directly connected to
MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected
(internal) MAINS derived circuits. In the latter case, transient stresses are variable; for that reason,
the transient withstand capability of the equipment is made known to the user.
Measurement category II is for measurements performed on circuits directly connected to the low
voltage installation. Examples are measurements on household appliances, portable tools and
similar equipment.
Measurement category III is for measurements performed in the building installation. Examples are
measurements on distribution boards, circuit-breakers, wiring, including cables, bus-bars, junction
boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some
other equipment, for example. Stationary motors with permanent connection to the fixed
installation.
Measurement category IV is for measurements performed at the source of the low-voltage
installation. Examples are electricity meters and measurements on primary overcurrent protection
devices and ripple control units.
MSO4000/DS4000 Quick Guide
RIGOL
V
Ventilation Requirement
This oscilloscope uses fan to force cooling. Please make sure that the air intake and exhaust areas
are free from obst ruct ions and hav e free air. When using the osc illosc ope in a bench-top setting,
provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation.
WARNING
Inadequate ventilation may cause temperature increase which would damage the
instrument. So please keep the instrument well ventilated during operation and inspect
the intake and fan regularly.
Working Environment
Temperature
Operating: 0℃ to +50℃
Non-operating: -40℃ to +70℃
Humidity
0℃ to +30℃:≤95% relative humidity
+30℃ to +40℃:≤75% relative humidity
+40℃ to +50℃:≤45% relative humility
WARNING
To avoid short circuit inside the instrument or electric shock, please do not operate in
humid environment.
Altitude
Operating: less than 3 km
Non-operating: less than 15 km
Installation (overvo lta ge) Cat ego ry
This product is powered by MAINS conforming to installation (overvoltage) category II.
MSO4000/DS4000 Quick Guide
RIGOL
VI
WARNING
Make sure that no overvoltage (such as that caused by thunderbolt) can reach the
product, or else the operator might expose to danger of electric shock.
Installation (overvoltage) Category Definitions
Installation (overvoltage) category I refers to signal level which is applicable to equipment
measurement terminals connected to the source circuit. In these terminals, precautions are done
to limit the transient voltage to the corresponding low level.
Installation (overvoltage) category II refers to the local power distribution level which is applicable
to equipment connected to the AC line(AC power).
Pollution Degree
Degree 2
Pollution Degree Definitions
Pollution deg ree 1 : No pol l ut i on or only dry, non-conductive pollution occurs. The pollution has no
influence. For example: a clean room or air-conditioned office environment.
Pollution degree 2: Normally only dry, non-conductive pollution occurs. Occasionally a temporary
conductivity caused by condensation may occur. For example: general indoor environment.
Pollution degree 3: Conductive pollution occurs, or dry, non-conductive pollution occurs which
becomes conductive due to condensation which is expected. For example: Sheltered outdoor
environment.
Pollution degree 4: Pollution that generates persist e nt conductivi ty t hroug h cond ucti ve dust, rain,
or snow. For example: outdoor locations.
Safety Class
Class 1 – Grounded Product
General Care and Cleaning
General Care:
Do not store or leave the instrument in where the instrument will be exposed to direct sunlight for
long periods of time.
MSO4000/DS4000 Quick Guide
RIGOL
VII
Cleaning:
Clean the instrument regularly according to its operating conditions. To clean the exterior surface,
perform the following steps:
1. Disconnect the instrument from all power sources.
2. Clean the loose dust o n the outside of the in strument with a lint- free cloth (with a mild
detergent or water). When cleaning the LCD, take care to avoid scarifying it.
CAUTION
To avoid damages to the instrument, do not expose them to liquids which have
causticity.
WARNING
T o a void injury res ulting from short circuit, make sure the instrument is completely dry
before reconnecting to a power source.
Environmental Consideratio ns
The following symbol indicates that this product complies with the requirements of European Union
according to Directives 2002/96/EC on waste electrical and electronic equipment (WEEE).
Product End-of-Life Handling
The equipment may contain substances that could be harmful to the environment or human health.
In order to avoid release of such substances into the environment and harm to human health, we
encourage you to recycle this product in an appropriate system that will ensure that most of the
materials are reused or recycled appropriately. Please contact your local authorities for disposal or
recycling information.
MSO4000/DS4000 Quick Guide
RIGOL
VIII
Contents
Guaranty and Declaration ................................................................................................ I
Safety Requirement ....................................................................................................... II
General Safety Summary .............................................................................................. II
Safety Terms and Symbols ........................................................................................... III
Measurement Category ............................................................................................... IV
Keep the damaged shipping container or cushioning material until the contents
of the shipment have been checked for completeness and the instrument has
passed both ele ctrical and mechanical tests.
The consigner or carrier shall be liable for the damage to instrumentresulting
from shipment. RIGOL would not be responsible for free maintenance/rework
or replacement of the unit.
2. Inspect the instrument
In case of any damage, or defect, or failure, notify your RIGOL sales
representative.
3. Check the accessories
Please check the accessories according to the packing lists. If the accessories
are incomplete or damaged, pleas e contact your RIGOL sales representative.
RIGOL
MSO4000/DS4000 Quick Guide
RIGOL
2
Appearance and Dimensions
Figure 1 Front View Unit: mm
Figure 2 Side View Unit: mm
MSO4000/DS4000 Quick Guide
RIGOL
3
To Prepare for Operation
To Remove the Cover
Before using the oscilloscope, remove the front panel cover by grasping the
transverse grab on each side and pull them in the arrow directions as shown in the
figure below.
Figure 3 To Rem ove th e Cover
To Adjust the Supporting Legs
Adjust the suppo rting legs properly to us e them as stands to tilt the oscilloscope
upwards, thus to place the oscilloscope stably and to operate an d observe the display
better. You can u nf old or f old the supp orting le gs in the ar row dire ctions as sh own in
the fi g ure below.
Figure 4 To Adjust the Supporting Legs
MSO4000/DS4000 Quick Guide
RIGOL
4
AC Power Input Terminal
To Connect to AC Power Supply
This oscilloscope accepts two kinds of AC power supplies: 100-127V, 45-440Hz and
100-240V, 45-65Hz. Please use the power cord supplied with the accessories to
connect the oscilloscope to the power source. Turn on the power switch under t he
power plug; at this point, the oscilloscope is energized and the power key
the lower-left corner of the front panel is in breath state.
at
Figure 5 To Connect to AC Power Suppl y
Power-on Inspection
When the oscilloscope is energized, press the power key
corner of the front panel to start the oscilloscope. During the start-up process, the
oscilloscope performs a series of self-tests and you can hear the sound of relay
switching. After the self-test is finishe d, the welcome screen is displayed and yo u can
view the Option Type, Option name, Option Edition and Left time of the option
currently installed in the “Installed Options” pop-up dialog box on the screen. When
the instrument is shipped, a t rial version of the optio n is provided and the left time i s
about 2000 minutes. Press UtilitySystemSelfTestInfo to view the self-test
results.
at the low er -left
MSO4000/DS4000 Quick Guide
RIGOL
5
To Connect the Probe
RIGOL provides passive and active probes for the MSO4000/DS4000 series
oscilloscopes. For detailed technical information of the probes, please refer to the
corresponding Probe User’s Guide. The following are the probes recommended for
this oscilloscope.
Model Description
RP3500A 500 MHz, passive probe, standard, auto detection
RP7150 1.5 GHz, active probe, optional, auto detection
Connect the Probe:
1. Connect the BNC terminal of the probe to a channel BNC connector of the
oscilloscope at the front panel.
2. Connect the probe tip to the circuit point under test and connect the ground
alligator clip of the probe to the circuit ground terminal.
Figure 6 To Connect the Probe
Function Inspection
1. Press Default to restore the oscilloscope to its default configuration.
2. Connect the ground alligator clip of the pr obe to the “Ground Terminal” under the
“Compensation Signal Output Terminal”.
3. Connect the input terminal of CH1 of the oscilloscope with the “Compensation
Signal Output Terminal” of the probe using the probe.
MSO4000/DS4000 Quick Guide
RIGOL
6
o avoid electric shock during the use of probe, please make sure that
Tip
probe compensation adjustment and can not be used for calibration.
Compensation Signal Output T erminal
Ground Terminal
Figure 7 To Use the Compensation Signal
4. Press the AUTO key.
5. Observe the wavef orm o n the displa y. In normal condition, the dis play s hould be
a square waveform as shown in the figure below:
Figure 8 Square Waveform
6. Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the f igure above, please perform “Probe
Compensation” in the next section.
WARNING
T
the insulated wire of the probe is in good condition and do not touch
the metallic part of the probe when the probe is connected to high
voltage source.
The signal output from the probe compensation connector can only be used for
MSO4000/DS4000 Quick Guide
RIGOL
7
Over compensated Perfectly compensated Under c ompensated
Probe Compensation
When the probes are used for the first time, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorl y
compensate d probes may cause mea s ure m ent i n accurac y and error. The probe
compensation procedure is as follows:
1. Perform s tep 1, 2, 3 and 4 of “Function Inspection” in the previous section.
2. Check the displayed waveforms and compare them with the following figures.
Figure 9 Probe Compensation
3. Use a nonmetallic driver to adjust the variable capacitor on the probe until the
displayed waveform is as the “Perfectly compensated” in the figure above.
To Connect the Logic Probe
RIGOL provides logic probe for the MSO4000 series digital oscilloscope. For detailed
technical information of the logic probe, please refer to the corresponding Probe
User’s Guide.
Connect the logic probe single head to the digital channel input terminal [LOGIC D0-D15] at the front panel of the MSO4000 series digital oscilloscope. Note:
connect the logic probe adaptor supplies with the accessories to the corresponding
branch header (called channel g roup) before connecting the logic probe to the devi ce
under test.
Figure 10 To Connect the Logic Probe
MSO4000/DS4000 Quick Guide
RIGOL
8
Front Panel Overview
1 2
13 14 15 16 17 18 19 20 21 22
Figure 11 Front Panel Overview
Table 1 Front Panel Description
3 4 5
12
11
6 7 8 9 10
No. Description No. Description
1 Menu 12 Record
2 LCD 13 Power Key
3 Knob 14 USB HOST
4 Navigation Knob 15 Digital Channel Input Terminal
5 HORIZONTAL 16 Function Setting Menu Softkeys
6 CLEAR 17 VERTICAL
7 AUTO 18 Analog Channel Input Terminals
8 RUN/STOP 19 Function Menu Keys
9 SINGLE 20 TRIGGER
10 Default&Print 21 EXT TRIG Input Terminal
11 LOGIC ANALYZER 22 Probe Compensation Signal Output
Terminal/Ground Terminal
MSO4000/DS4000 Quick Guide
9
Rear Panel Overview
1. Trig Out/Calibratio n
Various kinds of signals can be output from this connector (press Utility
AuxOutput to select the desired outpu t type).
1) TrigOut: the oscilloscope outputs a signal which can reflect the current capture rate of the oscilloscope each time a trigger is generated.
2) Fast: the oscilloscope outputs a fast edge signal which can be used in the self-calibration of the oscilloscope .
3) GND: the oscilloscope outputs a ground level.
4) PassFail: the oscilloscope outputs a pulse when failed waveform is
detected and the pulse can be transmitted to o ther control system for easy
view of the test result.
2. Reference Clock
Provide more precise sample clock signal for the oscilloscope and it can also
synchronize two or more oscilloscope clocks.
3. Video Output
Through this interface, the oscilloscope can be conne cted to external m onitors to
get clearer w aveform dis play. Note that the display of the oscilloscope is still
valid.
1 2 3 4 5 6 7 8
Figure 12 Rear Panel Overview
RIGOL
MSO4000/DS4000 Quick Guide
RIGOL
10
4. LAN
Through this interface, the oscilloscope can be connected to the network for
remote cont rol. As t he osci lloscope conforms to the LXI-C instrument standard, a
test system can be built quickly an d accessed t hrough we b page.
5. USB DEVICE
Through this interface, the oscill osc ope can be c onnected to the printer or PC to
print data waveform or control the oscilloscope through PC software.
6. USB HOST
Through this interface, the oscilloscope can be connec ted to the printer to print
wavef orm data or be connected to a USB storage device to store waveform files.
GPIB interface communication can be realized by using the USB-GPIB interface
converter provided by RIGOL.
Note that there is also a USB HOST interface at the front panel.
7. Lock Hole
Use th e security lock (please buy it yourself) to lock the oscilloscope to a fixed
location.
8. AC Input/Switch
AC power input terminal. The power supply require ments of this oscilloscope are
100-127 V, 45-440 Hz and 100-240 V, 45-65 Hz; CAT II. Please connect the
oscilloscope to AC power using the power cord prov i de d with the accessories.
Then, turn on the AC power switch to energize the oscillosc ope. Press the power
key at the front panel to start the oscilloscope.
MSO4000/DS4000 Quick Guide
RIGOL
11
Front Panel Function Overview
VERTICAL
MSO4000/DS4000 provi des independent vertical control systems for the four analog
input channels.
CH1, CH2, CH3, CH4: analog input channels. The four channels are marked
by different colors which are also used to mark both the corresponding
waveforms on the screen and the channel input connectors. Press any key to
open the corresponding channel menu and press again to turn off the channel.
AC: illuminated when AC channel coupling mode is enabled.
50: illuminated when the channel input impedance is 50 Ω.
BW: illuminated when bandwidth limit is enabled.
MATH: press this key to open the math operation menu under which add,
subtract, multiply, divide operations, FFT operation, logic operation and
advanced operation are p r o vided .
REF: press this key to enable t h e re fer e n c e wav efo r m fu nction t o c o mp ar e t h e
waveform actually tested with the reference waveform.
VERTICAL POSITION: vertical position knobs. During the modification,
the position label at the lower left corner of the sc reen would change a ccordingly.
Turn clockwise to increase the position and counterclockwise to decrease. Press
down the knob to quickly reset the vertical position to zero.
VERTICAL SCALE: vertical scale knobs. During the modification, the scale
label at the bottom of the screen would change accordingly. Turn clockwise to
decrease the scale with a step of 5-2-1 and turn counterclockwise to increase
MSO4000/DS4000 Quick Guide
RIGOL
12
with a step of 1-2-5. Press down the knob to switch the vertical scale adjustment
mode between “Coarse” and “Fine”.
Decode1 and Decode2: decoding function keys. Press the corresponding key
to open the decoding function menu. MSO4000/DS4000 supports parallel
decoding and p rotocol decoding.
HORIZONTAL
MENU: press this key to open the horizont al control menu with which users can
turn on or off the delayed sweep function, switch between different time base
modes, switch between “Coarse” and “Fine” adjustments of scale as well as
modify the horizontal reference setting.
HORIZONTAL SCALE: modify the horizontal time base. Turn clockwise to
reduce the time base and turn counterclockwise to increase the time base.
During the modification, waveforms of all the channels will be displayed in
expanded or compressed mode and the time base message (e.g.
at the upper side of the screen would change accordingly. Press down the knob
to quickly switch to delayed sweep state.
HORIZONTAL
would move left or right relative to the center of the screen when you turn the
knob. During the modification, waveforms of all the channels would move left or
right and the trigger position message (e.g.
upper-right corne r of the screen w ould change accor dingly . P ress down the knob
to quickly reset the trigger position (or the delayed sweep position)
POSITION: modify the trigger position. The trigger point
) at the
)
MSO4000/DS4000 Quick Guide
13
TRIGGER
MODE: press this key to switch the trigger mode to Auto,
RIGOL
Normal or Single and the state backlight of the current
trigger mode would be illuminated.
TRIGGER LEVEL: modify the trigger level. Turn clockwise
to increase the level and turn counterclockwise to reduce the
level. During the modification, the trigger level line moves up
and down and the value in the trigger level message box (e.g.
) at the lower-left corner of the screen would
MENU: press this key to turn on t he trig ger oper atio n menu. This oscillos cope
provides various trigger types.
FORCE: in Normal and Single trigger modes, press this key to generate a
trigger signal forcefully.
50%: press this key to set the trigger level to the vertical midpoint of the
trigger signal amplitude.
change accordingly. Press down the knob to quickly reset the
trigger level to zero point.
CLEAR
Press this key to clear all the waveforms on the screen. If the
oscilloscope is in “Run” state, waveforms will still be displayed.
MSO4000/DS4000 Quick Guide
RIGOL
14
ote that auto set ting requires t hat the f requency of the signal unde r test shou ld be
AUTO
Press this key to enable the waveform auto setting function. The
oscilloscope will automatically adjust the vertical scale, horizontal
time base and trigger mode according to the input signal to realize
optimum waveform dis play . Besides, quick parameter measurement
function is also provided.
N
no lower than 50 Hz, the duty cycle be greater than 1% and the amplitude be at
least 20 mVpp. If the parameters exceed theses limits, “Auto failed” would be
displayed after pressing this key and the quick parameter measurement menu
might not be displayed.
RUN/STOP
Press this key to set the state of the oscilloscope to “Run” or “Stop”.
In “Run” state, the key is illuminated in yellow and red in “Stop”
state.
SINGLE
Press this key to set the trigger mode to “Single”. In single trigger
mode, the oscilloscope triggers when the trigger condition is met
and then stops.
MSO4000/DS4000 Quick Guide
RIGOL
15
be used to quickly l ocate t he w a vef orm fr a me (“C urr ent
Knob
This knob can be used to adjust the value of the parameters, adjust the waveform
brightness and select the desired menu (the backlight turns on during the operation).
During parameter input, turn the knob clockwise to increase the parameter and
count erclockwise to decrease.
When the menu blanks off, rotate the knob to adjust the waveform brightness.
The adjustable range is from 0% to 100%. Turn the knob clockwise to increase
the waveform brightness and counterclockwise to decrease. Press down the
knob to reset the waveform brightness to 50%. Users can also press Display WaveIntensity and then rotate the knob to adjust the waveform brightness.
During menu operation, after pressing a menu softkey, rotate the knob to switch
to t he desired su b-menu under that menu and then press down the knob to
select this sub-menu. You can also use it to modify parameter and input
filename.
Navigation K n ob
This knob provides quick Adjust/Locate fun ction for
numerical parameters with relatively larger settable
ranges. Turn t h e knob clockwise (counter-clockwise) to
increase (reduce) the value. Note that the inner knob is
used for fine adjustment and the outer knob for coarse
MSO4000/DS4000 Quick Guide
adjustment. The more the outer knob is rotated, the
faster the value will cha nge. F or example, this knob can
Frame” men u) to be playe d back in wa vefor m play back.
Similar menus include trigger holdoff, pulse width
setting, slope time etc.
RIGOL
16
Press this key to restore the oscilloscope to its default
Default
Print
configuration.
Press this key to execute print function or save the screen
in the USB storage device. If the oscilloscope is currently
connected to a printer and the printer is in i dle state, press
this key to execute print function. If no printer but a USB
storage device is currently connected, press this key to
save the screen to the USB storage device in “.bmp” format
(if the current storage type is picture, the screen will be
stored in the USB storage device in picture format). When
printer and USB storage device are connected at the same
time, the printer enjoys higher priority.
Function Keys
Measure: press this key to enter the measurement setting menu to set the
measurement-related parameters.
Acquire: press this key to enter the sample setting menu to set t he acquisition
mode, the memory depth and the antialiasing function of the oscilloscope.
Storage: press this key to enter the file store and recall interface. The storable
file types include traces, waveforms, setups, picture and CSV. This instrument
supports internal and external storage as well as disk management.
Cursor: press this key to enter the cursor measurement menu. The oscilloscope
MSO4000/DS4000 Quick Guide
RIGOL
17
Stop Play/Pause Record
provides thre e cursor modes: manual, track and auto.
Display: press this key to enter the display setting menu to set the dis play type,
persistence time, wave intensity, grid type, grid brightness and menu display
time o f t he waveform.
Utility: press this key to enter the system function setting menu to set the
system-related functions or parameters, such as I/O setting, sound and
language. The oscilloscope also supports some advanced functions such as
pass/fail test and print setting.
Record: press this key to enter the waveform record menu. The oscilloscope
provides wavef orm record function and supports rec ord constant ope n mode as
well as playback and analysis of recorded waveform.
Help: press this key to open the help interface. For detailed information, pleas e
refer to the introduction in “To Use the Built-in Help Syste m ”.
Record
Record: press this key to start recording the waveform. The backlight is
illuminated in red. Besides, the backlight will also be illuminated when record
constant on (open) mode is enabled.
Play/Pause: in stop or pause state, press this key to play back the wavefo rm
and press again to pausethe play. The backlight is illuminated in yellow.
Stop: press this key to stop the waveform in record or p layback mode. The
backlight is illuminated in orange.
MSO4000/DS4000 Quick Guide
RIGOL
18
LA
LOGIC ANALYZER
LA_: this knob can be used to reposition the digital
Note: this function is only available for MSO4000 series digital oscilloscope.
: press this key to enter the logic analyzer setting
interface. You can enable or disable groups of channels or
a single channel, change the display size and the logic
threshold of the digital channels and group the 16 digital
channels and display them as a bus. You also can set a
label for each digital channel.
channels. Press this knob continu ously to swit ch the digital
channels currently enabled by the number of the digital
channel (from small to large) and the channel selected is
highlighted in red. Rotate the knob to move the digital
waveform to desired position, press this knob again to fix
this channel and select the next channel.
MSO4000/DS4000 Quick Guide
RIGOL
19
13 14 15 16 17 18 19 20
User Interface
MSO4000/DS4000 oscilloscope pro vi des 9 inches, WVGA (800*480) 160,000 color
TFT LCD. What is worth mentioning is that the 14-grid ultra-wide screen makes you
view “longer” waveform.
1 2 3 4 5 6 7 8 9 10 11 12
Figure 13 User Interfa ce
1. Measurement Menu
Provide 12 horizontal (HORIZO NTAL) and 12 vertical (VERTICAL)
measurement parameters. Press the softkey at the left of the screen to
activate corresponding measurement item.
2. Analog Channel Label/Waveform
Different channels are marked by different colors and the color of the
waveform complies with the color of the channel.
3. Digital Channel Label/Waveform
MSO4000/DS4000 Quick Guide
RIGOL
20
The logic high levels of the digital waveforms are displayed in blue, the logic
low levels are displayed in green which complies with the color of the digital
channel label and the edges are displayed in white. Both the label and the
waveform of the digital channel currently selected are displayed in red.
Note: this function is only available for MSO4000 series digital oscilloscope.
4. Status
Available states include RUN, STOP, T’D (triggered), WAIT and AUTO.
5. Horizontal Time Base
Represent t he time p er grid on the horizontal axis on the screen.
Use HORIZONTAL
SCALE
to modify this parameter. The range
available is from 1.000 ns to 1000 s.
6. Sample Rate/Memo ry Depth
Display the current sample rate and memory depth of the oscilloscope.
Use HORIZONTAL
SCALE to modify this parameter.
7. Waveform Memory
Provide the schematic diagram of the memory position of the waveform
currently on the screen.
waveform on the
screen
8. Trigger Position
Display the trigger position of the waveform in the waveform memo ry and
on the screen.
9. Horizontal Position
Use HORIZONTAL
POSITION to modify this parameter. Press down
the knob to automatically set the parameter to zero.
10. Trigger Type
Display the currently selected trigger type and trigger condition setting.
Different labels are displayed when different trigger types are selected.
For example:
represents triggering on the rising edge in “Edge” trigger.
11. Trigger Source
Display the trigger source (CH1-CH4, EXT, EXT/5,AC L ine or D 0-D15)
currently selected. Different labels are displayed when different trigger
sources are sel e cted an d t he colo r o f t he t ri gger pa r a met er a rea w ill cha n ge
accordingly.
For example:
MSO4000/DS4000 Quick Guide
denotes that CH1 is selecte d as the trigger so urce.
21
12. Trigger Level/threshold
The trigger level label
will be displayed at the right of the screen
when the trigger source is selected CH1 to CH4 and the trigger level
value is displayed at the upper-right corner of the screen. Note: in runt
trigger and slope trigger, there are two trigger level labels (
).
When Using TRIGGER
trigger level value will change with the up and down of
There is no trigger level label
LEVEL to modify the trigger level, the
.
displayed when the trigger source is
selected D0 to D15 and the trigger threshold is displayed at the
upper-right corner of the screen (only available for MSO4000 series
digital oscilloscope.).
13. CH1 Vertical Scale
Display the voltage value per grid of CH1 wa veform vertically.
Use VIRTICAL
SCALE to modify this par ameter.
The following labels will be provided according to the current channel
setting: channel coupling (e.g.
in AC coupling), ban dwidth limit (e .g .
when bandwidth limit is enabled) and input impedance (e.g.
when the input impedance is 50Ω).
14. CH2 Vertical Scale
Displaythe voltage value per grid of CH2 waveform vertically.
Use VIRTICAL
SCALE to modify this par ameter.
The following labels will be provided according to the current channel
setting: channel coupling (e.g.
in AC coupling), ban dwidth limit (e .g .
when bandwidth limit is enabled) and input impedance (e.g.
when the input impedance is 50Ω).
15. CH3 Vertical Scale
Display the voltage value per grid of CH3 waveform vertically.
Use VIRTICAL
SCALE to modify this par ameter.
The following labels will be provided according to the current channel
setting: channel coupling (e.g.
in AC coupling), bandwidth limit (e.g.
when bandwidth limit is enabled) and input impedance (e.g.
when the input impedance is 50Ω).
16. CH4 Vertical Scale
Display the voltage value per grid of CH4 waveform vertically.
Use VIRTICAL
SCALE to modify this parameter.
The following labels will be provided according to the current channel
RIGOL
and
MSO4000/DS4000 Quick Guide
RIGOL
22
setting: channel coupling (e.g. in AC coupling), bandwidth limit (e.g.
when bandwidth limit is enabled) and input impedance (e.g.
when the input impedance is 50Ω).
17. Message Box
Display prompt messages.
18. Digital Channel Status Area
Display the current state of the 16 digital channels. It is D0 to D15 from right
to left. The digital channels cur rently ena bled are displayed in green and the
digital channel currently selected is highlighted in red. Any digital channel
that is turned off will be grayed out in the Digital Channel Status Area.
Note: this function is only available for MSO4000 series digital oscilloscope.
19. Notification Area
Display system time, sound icon and USB storage device icon.
System Time: displayed in “hh:mm (hour:minute)” format. When
printing or storing the waveform, the output file will contain this time
message. Press UtilitySystemSyste m Time to set the time in
the following format:
yyyy-mm-dd hh-mm-ss (year-month-dat e hour-minute-second)
Sound Icon: when sound is enabled,
is displayed. Press Utility
Sound to turn the sound on or off.
USB Storage Device Icon: when the oscilloscope detects a USB storage
device,
is displayed.
20. Operation Menu
Press any softkey to activate the corresponding menu. The following
symbols might be displayed in the menu:
Denote that
items. The backlight of
at the front panel can be used to select parameter
turns on when parameter selection is
valid.
Denote that
backlight of
Denote that you can use the nevigation knob to quickly adjust/locate
can be used to mod ify paramete r value. The
turns on when parameter input is valid.
parameters.
Denote that users can rotate and then press down
to select the
desired paramet er.
Denote that the current menu has several options.
Denote that the current menu has a lower level menu.
MSO4000/DS4000 Quick Guide
RIGOL
23
Press this key to return to the previous menu.
Note that the following direction keys might appear in the grid at the
lower-left corner of the menu bar:
Denote that you can open the next page menu.
Denote that you can open the previous page menu.
To Use the Security Lock
If needed, you can use the security lock (please buy it yourself) to lock the
oscilloscope to a fixed location. The method is as follows, align the lock wit h the lo ck
hole and plug it into the lock hole vertically, turn the key clockwise to lock the
oscilloscope and then pull the key out.
Security Lock Hole
Figure 14 To Use the Security Loc k
Note: do not insert other articles into the security lock hole to avoid damaging the
instrument.
MSO4000/DS4000 Quick Guide
RIGOL
24
To Use the Built-in Help System
The help system of this oscilloscope provides instructions for all the function keys
(including menu keys) at t he front panel. Press Help to open the help interface and
press ag ain to clos e the inter face. The help interface mainly consists of two parts.
The left is “Help Options” and you can use “Button” or “Index” mode for selection.
The right is “Help Display Area”.
Help Options Help Display Area
Figure 15 Help Information
Button:
Default mode. In this mode, you can press the button (ex cep t th e pow er key
the knob
and menu page up/ down key /) or rotate the knob at the front
panel directly to get the corresponding help information of that button in the “Help
Display Area”.
Use
to select “To I nd e x” and then press down the knob to switch to Index
mode.
Index:
In this mode, use
to select th e item that needs to get help (for example: “Band
Width”). The item currently selected is displayed with brown shading. Press down
the knob to get the corresponding help information in the “Help Display Area”.
Use
to select “To B ut t o n ” a nd then pre ss down t he knob to switch t o Button
mode.
MSO4000/DS4000 Quick Guide
,
RIGOL
25
Troubleshooting
The commonly encountered failures and their solutions are listed below. When you
encounter those problems, please solve them following the corresponding steps. If
the problem remains still, please contact RIGOL and provide yo ur device
information (acquisition method: Utility System System Info).
1. The screen is still dark (no display) after power on:
(1) Check whether the power is correctly conne cted.
(2) Check whether the power switch is really on.
(3) Restart the instrument after finishing the above inspections.
(4) If it still does not work correctly, please contact RIGOL.
2. The signal is sampled but no waveform of the signal is displayed:
(1) Check whether the probe is correctly connected to the signal connecting
wire.
(2) Check whether the signal connecting wi re is correctly connected to the BNC
(namely channel connector).
(3) Check whether the probe is correctly connected to the item under test.
(4) Check whether there are signals generated from the item under test (you
can connect the probe compensation output signal to the problematic
channel to determine which has problems, the channel or the item under
test).
(5) Resample the signal.
3. The tested voltage amplitude is greater or lower than the actual value
(note that this problem usually occurs when probe is used):
Check whether the attenuation coefficient of the channel complies with the
attenuation ratio of the probe.
4. There is waveform display but not stable:
(1) Check the trigger signal source: check whether the Source in the trigger
menu complies with the signal channel actually used.
(2) Check the trigger t y pe: ge neral sig nals sh ould use “ Edge” trigger and video
signal should use “ Video” trigger. Only when the proper trigger type is used,
can the waveform be displayed stably.
MSO4000/DS4000 Quick Guide
RIGOL
26
(3) Try to change the Coupling to “HF Reject” or “LF Reject” to filter out the
high-frequency or low-frequency noise that disturbs the trigger.
(4) Change the trigger holdoff setting.
5. No display after pressing RUN/STOP:
Check whether the MODE at the trigger panel (TRIGGER) is “Normal” or
“Single” and whether the trigger level exceeds the waveform range. If yes, set
the trigger level to the middle or set the MODE to “Auto”.
Note that using the AUTO button could automatically finish the above setting.
6. The display of waveform is ladder-like:
(1) The horizontal ti me base might be too low. Increase the horizontal time
base to increase the horizontal resolution and improve the display.
(2) If the display Type is “Vectors”, the lines between the sample points may
cause ladder-like display. Set Type to “Dots” to solve the problem.
7. Fail to connect PC or printer through USB:
Check the IO Se tting menu in Utility t o make sure whet her the setting in USB
Device matches the currently connected device. If needed, restart the oscilloscope.
8. The USB storage device can not be recognized:
(1) Check whether the USB storage device can work normally.
(2) Make sure that the USB storage device used is flash storage type. This
oscilloscope does not support hardware storage type.
(3) Make sure whether the capacity of the USB storage device is too large. It is
recommended that the capacity of the USB storage device being used with
this oscilloscope is no larger than 8 GBytes.
(4) Restart the instrument and then insert the USB storage device.
(5) If the USB storage device still can not be used normally, please contact
RIGOL.
MSO4000/DS4000 Quick Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.