RIGOL is a registered trademark of RIGOL (SUZHOU) TECHNOLOGIES INC.
Publication Number
QGA27100-1110
Notices
RIGOL products are covered by P.R.C. and foreign patents, issued and pending.
RIGOL reserves the right to modify or change parts of or all the specifications
and pricing policies at the company’s sole decision.
Information in this publication replaces all previously released materials.
Information in this publication is subject to change without notice.
RIGOL shall not be liable for either incidental or consequential losses in
connection with the furnishing, use, or performance of this manual, as well as
any information contained.
Any part of this document is forbidden to be copied, photocopied, or rearranged
without prior written approval of RIGOL.
Product Certification
RIGOL guarantees that this product conforms to the national and industrial
standards in China as well as the ISO9001:2015 standard and the ISO14001:2015
standard. Other international standard conformance certifications are in progress.
Contact Us
If you have any problem or requirement when using our products or this manual,
please contact RIGOL.
E-mail: service@rigol.com
Website: www.rigol.com
DS1000Z-E Quick Guide
RIGOL
1. Only the exclusive power cord
designed for the instrument and
authorized for use within the local
country could be used.
2. Ensure that the instrument is
safely grounded.
3. Connect the probe correctly.
4. Observe all terminal ratings.
5. Use proper overvoltage protection.
6. Do not operate without covers.
7. Do not insert objects into the air
outlet.
8. Use the proper fuse.
9. Avoid circuit or wire exposure.
10. Do not operate the instrument with
suspected failures.
11. Provide adequate ventilation.
12. Do not operate in wet conditions.
13. Do not operate in an explosive
atmosphere.
14. Keep instrument surfaces clean and
dry.
15. Prevent electrostatic impact.
16. Handle with caution.
WARNING
Indicates a potentially hazardous situation or practice which, if not
avoided, will result in serious injury or death.
CAUTION
Indicates a potentially hazardous situation or practice which, if not
avoided, could result in damage to the product or loss of important data.
DANGER
It calls attention to an operation, if not correctly performed, could
result in injury or hazard immediately.
WARNING
It calls attention to an operation, if not correctly performed, could
result in potential injury or hazard.
CAUTION
It calls attention to an operation, if not correctly performed, could
result in damage to the product or other devices connected to the
product.
Hazardous
Voltage
Safety
Warning
Protective
Earth
Terminal
Chassis
Ground
Test
Ground
Safety Requirement
General Safety Summary
Safety Notices and Symbols
Safety Notices in this Manual:
Safety Terms on the Product:
Safety Symbols on the Product:
II DS1000Z-E Quick Guide
RIGOL
III
WARNING
This oscilloscope can only be used for measurements within its specified
measurement categories.
WARNING
Inadequate ventilation may cause temperature increase which would
damage the instrument. So please keep the instrument well ventilated
during operation and inspect the intake and fan regularly.
Measurement Category
Measurement Category
DS1000Z-E series digital oscilloscopes can make measurements in Measurement
Category I.
Measurement Category Definitions
Measurement category I is for measurements performed on circuits not directly
connected to MAINS. Examples are measurements on circuits not derived from
MAINS, and specially protected (internal) MAINS derived circuits. In the latter case,
transient stresses are variable; for that reason, the transient withstand capability of
the equipment is made known to the user.
Measurement category II is for measurements performed on circuits directly
connected to the low voltage installation. Examples are measurements on household
appliances, portable tools and similar equipment.
Measurement category III is for measurements performed in the building installation.
Examples are measurements on distribution boards, circuit-breakers, wiring,
including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed
installation, and equipment for industrial use and some other equipment, for
example. Stationary motors with permanent connection to the fixed installation.
Measurement category IV is for measurements performed at the source of the
low-voltage installation. Examples are electricity meters and measurements on
primary overcurrent protection devices and ripple control units.
Ventilation Requirement
This oscilloscope uses fan to force cooling. Please make sure that the air intake and
exhaust areas are free from obstructions and have free air. When using the
oscilloscope in a bench-top or rack setting, provide at least 10 cm clearance beside,
above and behind the instrument for adequate ventilation.
DS1000Z-E Quick Guide
RIGOL
WARNING
To avoid short circuit inside the instrument or electric shock, please do
not operate in humid environment.
WARNING
Make sure that no overvoltage (such as that caused by thunderbolt) can
reach the product, or else the operator might expose to danger of electric
shock.
Working Environment
Temperature
Operating: 0℃ to +50℃
Non-operating: -40℃ to +70℃
Humidity
0℃ to +30℃: ≤95% relative humidity
+30℃ to +40℃: ≤75% relative humidity
+40℃ to +50℃: ≤45% relative humility
Altitude
Operating: below 3 km
Non-operating: below 15 km
Installation (Overvoltage) Category
This product is powered by mains conforming to installation (overvoltage) category
II.
Installation (Overvoltage) Category Definitions
Installation (overvoltage) category I refers to signal level which is applicable to
equipment measurement terminals connected to the source circuit. In these
terminals, precautions are done to limit the transient voltage to the corresponding
low level.
Installation (overvoltage) category II refers to the local power distribution level
which is applicable to equipment connected to the AC line(AC power).
Pollution Degree
Degree 2
Pollution Degree Definitions
Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The
pollution has no influence. For example: a clean room or air-conditioned office
environment.
IV DS1000Z-E Quick Guide
RIGOL
V
CAUTION
To avoid damage to the instrument, do not expose it to caustic liquids.
WARNING
To avoid short-circuit resulting from moisture or personal injuries, ensure
that the instrument is completely dry before connecting it to the power
supply.
Pollution degree 2: Normally only dry, non-conductive pollution occurs. Occasionally
a temporary conductivity caused by condensation may occur. For example: general
indoor environment.
Pollution degree 3: Conductive pollution occurs, or dry, non-conductive pollution
occurs which becomes conductive due to condensation which is expected. For
example: Sheltered outdoor environment.
Pollution degree 4: Pollution that generates persistent conductivity through
conductive dust, rain, or snow. For example: outdoor locations.
Safety Class
Class 1 – Grounded Product
Care and Cleaning
Care
Do not store or leave the instrument where it may be exposed to direct sunlight for
long periods of time.
Cleaning
Clean the instrument regularly according to its operating conditions.
1. Disconnect the instrument from all power sources.
2. Clean the external surfaces of the instrument with a soft cloth dampened with
mild detergent or water. Avoid having any water or other objects into the
chassis via the heat dissipation hole. When cleaning the LCD, take care to avoid
scarifying it.
DS1000Z-E Quick Guide
RIGOL
Environmental Considerations
The following symbol indicates that this product complies with the WEEE Directive
2002/96/EC.
Product End-of-Life Handling
The equipment may contain substances that could be harmful to the environment or
human health. To avoid the release of such substances into the environment and
avoid harm to human health, we recommend you to recycle this product
appropriately to ensure that most materials are reused or recycled properly. Please
contact your local authorities for disposal or recycling information.
Document Overview
This manual is used to guide users to quickly get familiar with the front panel, rear
panel, user interface and basic operation method of DS1000Z-E series digital
oscilloscope. You can download the newest version of the manual from RIGOL
official website (www.rigol.com).
Format Conventions in this Manual:
1. Key
The front panel keys are denoted by the format of "Key Name (Bold) + Text Box".
For example, Utility denotes the "Utility" key.
2. Menu
The menu softkeys are denoted by the format of "Menu Word (Bold) +
Character Shading". For example, System denotes the "System" menu under
Utility.
3. Operation Step
The next step of operation is denoted by an arrow "". For example, Utility
System denotes that first press Utility on the front panel and then press
System.
VI DS1000Z-E Quick Guide
RIGOL
VII
Label
Knob
HORIZONTAL SCALE
Horizontal Scale Knob
HORIZONTAL POSITION
Horizontal Position Knob
VERTICAL SCALE
Vertical Scale Knob
VERTICAL POSITION
Vertical Position Knob
TRIGGER LEVEL
Trigger Level Knob
Model
Analog Bandwidth
Number of Analog Channels
DS1202Z-E
200 MHz
2
4. Knob
Content Conventions in this Manual:
DS1000Z-E series includes the following models. Unless otherwise noted, this
manual takes DS1202Z-E as an example to illustrate the front panel, rear panel, user
interface and basic operation methods of DS1000Z-E series.
DS1000Z-E Quick Guide
RIGOL
Contents
Guaranty and Declaration ................................................................................................ I
Safety Requirement ....................................................................................................... II
General Safety Summary .............................................................................................. II
Safety Notices and Symbols .......................................................................................... II
Measurement Category ............................................................................................... III
Ventilation Requirement .............................................................................................. III
Working Environment .................................................................................................. IV
Care and Cleaning ........................................................................................................V
Environmental Considerations ...................................................................................... VI
Document Overview ....................................................................................................... VI
If the packaging has been damaged, do not dispose the damaged packaging or
cushioning materials until the shipment has been checked for completeness and
has passed both electrical and mechanical tests.
The consigner or carrier shall be liable for the damage to the instrument
resulting from shipment. RIGOL would not be responsible for free
maintenance/rework or replacement of the instrument.
2. Inspect the instrument
In case of any mechanical damage, missing parts, or failure in passing the
electrical and mechanical tests, contact your RIGOL sales representative.
3. Check the accessories
Please check the accessories according to the packing lists. If the accessories
are damaged or incomplete, please contact your RIGOL sales representative.
DS1000Z-E Quick Guide
RIGOL
Appearance and Dimensions
Figure 1 Front View Unit: mm
Figure 2 Top View Unit: mm
2 DS1000Z-E Quick Guide
RIGOL
3
Power Cord Connector
To Prepare for Operation
To Adjust the Supporting Legs
Adjust the supporting legs properly to use them as stands to tilt the oscilloscope
upwards for stable placement of the oscilloscope as well as better operation and
observation.
Figure 3 To Adjust the Supporting Legs
To Connect to AC Power Supply
This oscilloscope can accept 100-240 V, 45-440 Hz AC power supply. Please use the
power cord supplied with the accessories to connect the oscilloscope to the power
source as shown in the figure below.
DS1000Z-E Quick Guide
Figure 4 To Connect to AC Power Supply
RIGOL
Turn-on Checkout
When the oscilloscope is connected to power, press the Power key at the
lower-left corner of the front panel to start the oscilloscope. During the start-up
process, the oscilloscope performs a series of self-test items. After the self-test, the
welcome screen is displayed. The instrument is installed with the trial versions of the
options before leaving factory and the remaining time is about 2,000 minutes. The
"Installed Options" dialog box will be displayed if your instrument has currently
installed the trial versions of options. From this dialog box you can view the name,
detail, version and remaining trial time of the option currently installed.
To Connect the Probe
RIGOL provides passive probe for DS1000Z-E series. For the model of the probes,
please refer to
probes, please refer to corresponding Probe User’s Guide.
Connect the Probe:
1. Connect the BNC terminal of the probe to an analog channel input of the
oscilloscope on the front panel.
2. Connect the ground alligator clip of the probe to the circuit ground terminal and
then connect the probe tip to the circuit point under test.
DS1000Z-E Series Datasheet
. For detailed technical information of the
Figure 5 To Connect the Passive Probe
After you connect the passive probe, check the probe function and probe
compensation adjustment before making measurements. For detailed procedures,
refer to "Function Inspection" and "Probe Compensation" sections in this
manual.
4 DS1000Z-E Quick Guide
RIGOL
5
WARNING
To avoid electric shock during the use of probe, please make sure that the
insulated wire of the probe is in good condition. Do not touch the metallic
part of the probe when the probe is connected to high voltage source.
Tip
The signal output from the probe compensation connector can only be used for
probe compensation adjustment and cannot be used for calibration.
Compensation Signal Output Terminal
Ground Terminal
Function Inspection
1. Press StorageDefault to restore the oscilloscope to its default
configuration.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" as shown
in the figure below.
3. Use the probe to connect the input terminal of CH1 and the "Compensation
Signal Output Terminal" of the oscilloscope.
Figure 6 To Use the Compensation Signal
4. Set the attenuation on the probe to 10X. Then press AUTO.
5. Observe the waveform on the display. In normal condition, the display should be
a square waveform as shown in the figure below:
Figure 7 Square Waveform
6. Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the figure above, please perform "Probe Compensation".
DS1000Z-E Quick Guide
RIGOL
Over compensated Perfectly compensated Under compensated
Probe Compensation
When the probes are used for the first time, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorly
compensated probes may cause measurement inaccuracy or errors. The probe
compensation procedures are as follows:
1. Perform Step 1, 2, 3 and 4 specified in "Function Inspection".
2. Check the displayed waveforms and compare them with the waveforms in the
following figure.
Figure 8 Probe Compensation
3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment
hole on the probe until the displayed waveform is displayed as "Perfectly
compensated" in the figure above.
6 DS1000Z-E Quick Guide
RIGOL
7
No.
Description
No.
Description
1
Measurement Menu Softkeys
10
Help/Print
2
LCD
11
Power Key
3
Function Menu Softkeys
12
USB Host Interface
4
Multifunction Knob
13
Analog Channel Input
5
Common Operation Keys
14
VERTICAL Control Area
6
CLEAR
15
HORIZONTAL Control Area
7
AUTO
16
External Trigger Input
8
RUN/STOP
17
TRIGGER Control Area
9
SINGLE
18
Probe Compensation Signal
Output Terminal/Ground Terminal
3 4 5 6 7 8 9 10
1 2
11 12 13 14 15 16 17 18
Front Panel Overview
Figure 9 Front Panel Overview
Table 1 Front Panel Description
DS1000Z-E Quick Guide
RIGOL
1 2 3
4 5 6 7
Rear Panel Overview
Figure 10 Rear Panel Overview
1.Handle
Pull up the handle vertically for easy carrying of the instrument. When you do
not need the handle, press it down.
2. LAN
Connect the instrument to the network via this interface for remote control. This
oscilloscope conforms to the LXI CORE 2011 DEVICE class instrument standards
and can quickly build test system with other instruments.
3. USB DEVICE
You can connect the oscilloscope to a PictBridge printer or PC via this interface.
When a PC is connected, users can send SCPI commands using the PC software
or control the oscilloscope via user-defined programming. When a printer is
connected, users can print the waveforms displayed on the screen using the
printer.
4. Trigger Out and Pass/Fail
Trigger Out:
The oscilloscope can output a signal that can reflect the current capture
rate of the oscilloscope at each trigger via this interface. Connect the signal
to a waveform display device to measure the frequency of the signal. The
measurement result is equal to the current capture rate.
8 DS1000Z-E Quick Guide
RIGOL
9
Pass/Fail:
The instrument will output a negative pulse via this connector when a failed
waveform is detected during the pass/fail test. The instrument continuously
outputs a low level via this connector when no failed waveform is detected.
5. Lock Hole
You can lock the instrument to a fixed location by using the security lock (please
purchase it yourself) via the lock hole.
6. Fuse
If a new fuse is required, please use the specified fuse (250V, T2A). The
operation method is as follows:
a) Turn off the instrument, disconnect the power, and then remove the power
cord.
b) Insert a small straight screwdriver into the slot at the power cord connector
and pry out the fuse holder gently.
c) Take out the fuse and replace it with a specified fuse. Then, install the fuse
holder.
7. AC Power Cord Connector
AC power input terminal. The power requirements of this oscilloscope are
100-240 V, 45-440 Hz. Use the power cord provided with the accessories to
connect the instrument to AC power. Then, you can press the Power key on the
front panel to start the instrument.
DS1000Z-E Quick Guide
RIGOL
CH1, CH2: analog channel setting keys.
The 2 channels are marked by different
colors which are also used to mark both the
corresponding waveforms on the screen
and the channel input connectors. Press
any key to open the corresponding channel
menu and press again to turn off the
channel.
MATH: press MATHMath, you can
open the math operation menu under which
A+B, A-B, A×B, A/B, FFT, A&&B, A||B,
A^B, !A, Intg, Diff, Sqrt, Lg, Ln, Exp and
Abs are provided. You can also press MATH
to open the decoding menu and set the
decoding options.
REF: press this key to enable the reference waveform function to compare the
waveform actually measured with the reference waveform.
Vertical POSITION: modify the vertical position of the current channel
waveform. Turn clockwise to increase the position and turn counterclockwise to
decrease. During the modification, the waveform would move up and down and the
position message (e.g. ) at the lower-left corner of the screen would
change accordingly. Press down this knob to quickly reset the vertical position to
zero.
VERTICAL SCALE: modify the vertical scale of the current channel. Turn
clockwise to decrease the scale and turn counterclockwise to increase. During the
modification, the amplitude of the waveform would enlarge or reduce and the scale
information (e.g.) at the lower side of the screen would change
accordingly. Press down this knob to quickly switch the vertical scale adjustment
modes between "Coarse" and "Fine".
Tip
How to set the vertical scale and vertical position of each channel?
The 2 channels of DS1000Z-E use the same VERTICAL POSITION and
VERTICAL SCALE knobs. If you want to set the vertical scale and vertical
position of a certain channel, please press CH1 or CH2 at first to select the
desired channel. Then rotate the VERTICAL POSITION and VERTICAL
SCALE knobs to set the values.
Front Panel Function Overview
VERTICAL
10 DS1000Z-E Quick Guide
RIGOL
11
HORIZONTAL POSITION: modify the horizontal
position. The trigger point would move left or right relative to
the center of the screen when you rotate the knob. During the
modification, waveforms of all the channels would move left or
right and the horizontal position message (e.g.
) at the upper-right corner of the screen
would change accordingly. Press down this knob to quickly
reset the horizontal position (or the delayed sweep position).
MENU: press this key to open the horizontal control menu
where you can turn on or off the delayed sweep function and
switch between different time base modes.
HORIZONTAL SCALE: modify the horizontal time base. Turn clockwise to
reduce the time base and turn counterclockwise to increase the time base. During
the modification, waveforms of all the channels will be displayed in expanded or
compressed mode and the time base message (e.g. ) at the upper side of
the screen would change accordingly. Press down this knob to quickly switch to
delayed sweep state.
MODE: press this key to switch the trigger mode to Auto,
Normal or Single and the corresponding state backlight of
the current trigger mode would be illuminated.
TRIGGER LEVEL: modify the trigger level. Turn clockwise
to increase the level and turn counterclockwise to reduce the
level. During the modification, the trigger level line would
move up and down and the value in the trigger level message
box (e.g. ) at the lower-left corner of the
screen would change accordingly. Press down the knob to
quickly reset the trigger level to zero point.
MENU: press this key to open the trigger operation menu.
This oscilloscope provides various trigger types; for the details,
please refer to the introduction in
DS1000Z-E User Guide.
FORCE: press this key to generate a trigger signal forcibly.
HORIZONTAL
TRIGGER
DS1000Z-E Quick Guide
RIGOL
Press this key to clear all the waveforms on the screen. If the
oscilloscope is in "RUN" state, new waveforms will still be
displayed.
Press this key to enable the waveform auto setting function.
The oscilloscope will automatically adjust the vertical scale,
horizontal time base and trigger mode according to the input
signal to realize optimum waveform display.
Press this key to "RUN" or "STOP" waveform sampling. In the
"RUN" state, the key is illuminated in yellow. In the "STOP"
state, the key is illuminated in red.
Press this key to set the trigger mode to "Single". In single
trigger mode, press FORCE to generate a trigger signal
immediately.
CLEAR
AUTO
Note: Waveform auto setting function requires that the frequency of sine is no lower
than 41 Hz; the duty cycle should be greater than 1% and the amplitude must be at
least 20 mVpp for square. Otherwise, the Waveform auto setting function may be
invalid and the quick parameter measurement function displayed in the menu will
also be unavailable.
RUN/STOP
SINGLE
12 DS1000Z-E Quick Guide
RIGOL
13
Adjust waveform brightness:
In non-menu-operation mode, turn this knob to adjust the
brightness of waveform display. The adjustable range is from
0% to 100%. Turn clockwise to increase the brightness and
counterclockwise to reduce. Press down this knob to reset the
brightness to 60%.
You can also press Display Intensity and use the knob to
adjust the waveform brightness.
Multifunctional:
In menu operation, the backlight of the knob goes on. Press any menu softkey and
rotate the knob to select the submenus under this menu and then press down the
knob to select the current submenu. It can also be used to modify parameters
(please refer to the introduction in "Parameter Setting Method") and input
filename.
Multifunction Knob
Function Menus
Measure: press this key to open the measurement setting menu. You can set the
measurement source, turn on or off the frequency counter, all measure, statistic
function and etc. Press MENU at the left of the screen to switch the measurement
menus of 37 waveform parameters. Then, press down the corresponding menu
softkey to quickly realize one-key measurement and the measurement result will be
displayed at the bottom of the screen.
Acquire: press this key to enter the sample setting menu to set the acquisition
mode, Sin(x)/x and memory depth of the oscilloscope.
Storage: press this key to enter file store and recall interface. The storable file
types include picture, traces, waveforms, setups, CSV and parameters. Internal and
external storage as well as disk management are also supported.
Cursor: press this key to enter the cursor measurement menu. The oscilloscope
provides four cursor modes: manual, track, auto and XY. Note that XY cursor mode is
only available when the horizontal time base is set to XY.
DS1000Z-E Quick Guide
RIGOL
Press this key to print the screen or save the screen to a USB
storage device.
― If a PictBridge printer is connected currently and the
printer is in idle state, pressing this key will execute the
print operation.
― If no printer is connected but a USB storage device is
inserted, pressing this key can save the screen to the USB
storage device in the specified format.
― If both a printer and a USB storage device are connected
at the same time, the printer enjoys higher priority when
pressing this key.
Display: press this key to enter display setting menu to set the display type,
persistence time, wave intensity, grid type and grid brightness.
Utility: press this key to enter the system function setting menu to set the
system-related functions or parameters, such as the I/O, sound and language.
Besides, some advanced functions (such as the pass/fail test, waveform record, etc.)
are also supported.
Print
14 DS1000Z-E Quick Guide
RIGOL
15
11 12 13 14 15 16
1 2 3 4 5 6 7 8 9 10
User Interface
DS1000Z-E provides 7.0 inch WVGA (800*480) TFT LCD.
1. Auto Measurement Items
Provide 20 horizontal (HORIZONTAL) and 17 vertical (VERTICAL) measurement
parameters. Press the softkey at the left of the screen to activate the
corresponding measurement item. Press MENU continuously to switch between
the horizontal and vertical parameters.
2. Status
Available states include RUN, STOP, T’D (triggered), WAIT and AUTO.
3. Horizontal Time Base
Represent the time per grid on the horizontal axis on the screen.
Use HORIZONTAL SCALE to modify this parameter. The range
available is from 5 ns to 50 s.
4. Sample Rate/Memory Depth
Display the current sample rate and memory depth of the oscilloscope.
The sample rate and memory depth will change in accordance with the
Figure 11 User Interface
DS1000Z-E Quick Guide
RIGOL
waveform on the screen
waveform in the memory
horizontal time base.
5. Waveform Memory
Provide the schematic diagram of the memory position of the waveform
currently on the screen.
6. Trigger Position
Display the trigger position of the waveform in the waveform memory and on
the screen.
7. Horizontal Position
Use HORIZONTAL POSITION to modify this parameter. Press down the
knob to automatically set the parameter to zero.
8. Trigger Type
Display the currently selected trigger type and trigger condition setting.
Different labels are displayed when different trigger types are selected.
For example: represents triggering on the rising edge in “Edge” trigger.
9. Trigger Source
Display the trigger source currently selected (CH1, CH2,AC Line, or EXT).
Different labels are displayed when different trigger sources are selected and
the color of the trigger parameter area will change accordingly.
For example: denotes that CH1 is selected as the trigger source.
10. Trigger Level
When the trigger source is set to analog channel, you need to set proper
trigger level.
at the right of the screen is the trigger level label and the trigger level
value is displayed at the upper-right corner of the screen.
When using TRIGGER LEVEL to modify the trigger level, the trigger
level value will change with the up and down of .
Note: In slope trigger, runt trigger and windows trigger, there are two
trigger level labels ( and ).
11. CH1 Vertical Scale
Display the voltage value per grid of CH1 waveform vertically.
Use VERTICAL SCALE to modify this parameter.
The following labels will be displayed according to the current channel
setting: channel coupling (e.g. ) and bandwidth limit (e.g. ).
16 DS1000Z-E Quick Guide
RIGOL
17
Denote that you can rotate the multifunction knob to modify
parameters. The backlight of turns on during the parameter
modification.
Denote that you can rotate to select the desired item, the item
currently selected is displayed in blue, press down the knob to enter the
corresponding menu. The backlight of is constant on after menus
with this symbol are selected.
Denote that press to input desired paramrter values directly using
the pop-up numeric keyboard. The backlight of is constant on after
menus with this symbol are selected.
Denote that the current menu has several options.
Denote that the current menu has a lower level menu.
Press this key to return to the previous menu.
The number of the dots denotes that the number of the pages the
current menu has.
12. Analog Channel Label/Waveform
Different channels are marked with different colors and the colors of the channel
label and waveform are the same.
13. CH2 Vertical Scale
Display the voltage value per grid of CH2 waveform vertically.
Use VERTICAL SCALE to modify this parameter.
The following labels will be displayed according to the current channel
setting: channel coupling (e.g. ) and bandwidth limit (e.g. ).
14. Message Box
Display prompt messages.
15. Notification Area
Display sound icon and USB disk icon.
Sound Icon: Press Utility Sound to enable or disable the sound. When
the sound is enabled, will be displayed; when the sound is disabled,
will be displayed.
USB Disk Icon: when a USB disk is detected, will be displayed.
16. Operation Menu
Press any softkey to activate the corresponding menu.
The following symbols might be displayed in the menu:
DS1000Z-E Quick Guide
RIGOL
Security Lock Hole
Parameter Setting Method
This oscilloscope provides two methods to set the basic parameters.
Method 1:
When is displayed on the menu, you can rotate to set the desired value
directly.
Method 2:
When is displayed, press down and the numeric keyboard is displayed, as
shown below. Rotate the knob to select the desired value, and then press down the
knob to input the value. After all the value input, rotate the knob to select the desired
unit, press down the knob to finish your input.
Figure 12 Numeric Keyboard
To Use the Security Lock
If needed, you can use the security lock (please buy it yourself) to lock the
oscilloscope to a fixed location. The method is as follows, align the lock with the lock
hole and plug it into the lock hole vertically, turn the key clockwise to lock the
oscilloscope and then pull the key out.
Figure 13 To Use the Security Lock
Note: Please do not insert other articles into the security lock hole to avoid
damaging the instrument.
18 DS1000Z-E Quick Guide
RIGOL
19
Help Options Help Display Area
To Use the Built-in Help System
The help system of this oscilloscope provides instructions for all the function keys
(including menu keys) on the front panel. Press Help to open the help interface and
press again to close the interface. The help interface mainly consists of two parts.
The left is "Help Options" and the right is "Help Display Area".
Figure 14 Help Information
You can press the button (except the Power key , the knob and the menu
Page Up/Down key /) on the front panel directly to get the corresponding help
information in the "Help Display Area".
DS1000Z-E Quick Guide
RIGOL
Troubleshooting
The commonly encountered failures and their solutions are listed below. When you
encounter those failures, please solve them following the corresponding steps. If the
problem remains still, please contact RIGOL and provide your device information
(acquisition method: Utility SystemSystem Info).
1. The screen is still dark (no display) after you powered on the
instrument.
(1) Check whether the power switch is really on.
(2) Check whether the power is correctly connected.
(3) Check whether the fuse is burned out. If the fuse needs to be changed,
please use the specified fuse.
(4) Restart the instrument after finishing the above inspections.
(5) If it still persists, please contact RIGOL.
2. The signal is sampled but no waveform of the signal is displayed.
(1) Check whether the probe is correctly connected to the item under test.
(2) Check whether signals are generated from the item under test (you can
connect the probe compensation signal to the problematic channel to locate
the problem).
(3) Resample the signal.
3. The voltage amplitude measured is greater or smaller than the actual
value (note that this failure usually only occurs when probe is used).
Check whether the probe ratio of the channel complies with the attenuation ratio
of the probe.
4. Waveforms are found to be displayed but not stable.
(1) Check the trigger signal source: press MENU (in the trigger control area
(TRIGGER) on the front panel) Source to check whether the trigger
source selected complies with the signal channel actually used.
(2) Check the trigger type: general signals should use "Edge" trigger and video
signal should use "Video" trigger. Only when the proper trigger type is used,
can the waveform be displayed stably.
(3) Check the trigger level: adjust the trigger level to the middle of the signal.
(4) Change the trigger holdoff setting.
5. No display is found after pressing RUN/STOP.
Check whether the trigger mode in the trigger control area (TRIGGER) on the
front panel is "Normal" or "Single" and whether the trigger level exceeds the
waveform range. If yes, set the trigger level to the middle or press MODE to set
the trigger mode to "Auto".
Note: Pressing AUTO could automatically finish the above setting.
20 DS1000Z-E Quick Guide
RIGOL
21
6. The display of waveform is ladder-like.
(1) The horizontal time base might be too low. Increase the horizontal time
base to increase the horizontal resolution and improve the display.
(2) If the display type is "Vectors", the lines between the sample points may
cause ladder-like display. Press DisplayType to set the display type to
"Dots" to solve the problem.
7.I Fail to connect PC or PictBridge printer through USB.
(1) Press UtilityIO SettingUSB Device to check whether the current
setting matches the device currently connected.
(2) Check whether the USB cable is correctly connected to the instrument and
PC.
(3) Check whether the USB cable is in good condition. If needed, restart the
oscilloscope.
8. The USB storage device cannot be recognized.
(1) Check whether the USB storage device can work normally.
(2) Check whether the USB storage device being used is flash type. This
oscilloscope does not support USB3.0 and hardware type USB storage
device.
(3) Check whether the capacity of the USB storage device is too large. It is
recommended that the capacity of the USB storage device being used with
this oscilloscope should not exceed 8 GBytes.
(4) Restart the instrument and then insert the USB storage device to check it.
(5) If the USB storage device still cannot be used normally, please contact
RIGOL.
DS1000Z-E Quick Guide