Rigol DS1000CA series, DS1302CA, DS1202CA, DS1102CA, DS1072CA User Manual

Page 1
RIGOL
User’s Guide
DS1000CA Series
Digital Oscilloscopes
Aug. 2011
RIGOL Technologies, Inc.
Page 2
Page 3
RIGOL
User‟s Guide for DS1000CA Series
I
Guaranty and Declaration
Copyright
© 2009 RIGOL Technologies, Inc. All Rights Reserved.
Trademark Information RIGOL is registered trademark of RIGOL Technologies, Inc.
Publication Number
UGA03113-1110
Notices
RIGOL products are protected by patent law in and outside of P.R.C.. RIGOL Technologies, Inc. reserves the right to modify or change parts of or all
the specifications and pricing policies at company‟s sole decision.
Information in this publication replaces all previously corresponding material. RIGOL shall not be liable for losses caused by either incidental or
consequential in connection with the furnishing, use or performance of this manual as well as any information contained.
Any part of this document is forbidden to copy or photocopy or rearrange
without prior written approval of RIGOL.
Product Certification RIGOL guarantees this product conforms to the standards of national and industrial.
Meanwhile, the related standards conform to other ISO will get further. At present, DS1000CA has passed CE, cTUVus and GOST certification.
Contact Us
If you have any problem or requirement when using our products, please contact RIGOL Technologies, Inc. or your local distributors, or visit: www.rigol.com
Page 4
RIGOL
User‟s Guide for DS1000CA Series
II
Safety Notices
Review the following safety precautions carefully before operating the instrument to avoid any personal injuries or damages to the instrument and any products connected to it. To avoid potential hazards use the
instrument as specified by this user‟s
guide only.
The instrument should be serviced by qualified personnel only.
To Avoid Fire or Personal Injury.
Use Proper Power Cord.
Use the power cord designed for the instrument and authorized in your country only.
Ground The Instrument.
The oscilloscope is grounded through the grounding conductor of the power cord. To avoid electric shock the instrument grounding conductor(s) must be grounded properly, before making connections to the input or output terminals of the instrument.
Connect The Probe.
The probes‟ ground terminals are at the
same voltage level of the instrument ground. Do not connect the ground terminals to a high voltage.
Observe All Terminal Ratings.
To avoid fire or shock hazard, observe all ratings and marks on the instrument.
Follow the user‟s guide for further
ratings information before making connections to the instrument.
Do Not Operate Without Covers.
Do not operate the instrument with covers or panels removed.
Use Proper Fuse.
Use the fuse of the type, voltage and current ratings as specified for the instrument.
Avoid Circuit or Wire Exposure.
Do not touch exposed connections and components when power is on.
Do Not Operate With Suspected Failures.
If suspected damage occurs with the instrument, have it inspected by qualified service personnel before further operations.
Provide Proper Ventilation.
Refer to the installation instructions for proper ventilation of the instrument.
Do Not Operate in Wet Conditions
In order to avoid short circuiting to the
Page 5
RIGOL
User‟s Guide for DS1000CA Series
III
interior of the device or electric shock, please do not operate in a humid environment.
Do Not Operate in an Explosive Atmosphere
In order to avoid damages to the device or personal injuries, it is important to operate the device away from an explosive atmosphere.
Keep Product Surfaces Clean and Dry
To avoid the influence of dust and/or moisture in air, please keep the surface of device clean and dry.
The disturbance test of all the models meet the limit values of A in the standard of EN 61326:
1997+A1+A2+A3, but can’t meet
the limit values of B.
Measurement Category
The DS1000CA series Digital Oscilloscope is intended to be used for measurements in Measurement Category I.
Measurement Category
Definitions
Measurement Category I is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS derived circuits. In the latter case, transient stresses are variable; for that reason, the transient withstand capability of the equipment is made known to the user.
WARNING
IEC Measurement Category I. The input terminals may be connected to circuit terminal in IEC Category I installations for voltages up to 300 VAC. To avoid the danger of electric shock, do not
connect the inputs to circuit‟s voltages
above 300 VAC.
Transient overvoltage is also present on circuits that are isolated from mains. The DS1000CA series Digital Oscilloscopes is designed to safely withstand occasional transient overvoltage up to 500 Vpk. Do not use this equipment to measure circuits where transient overvoltage could exceed this level.
Page 6
RIGOL
User‟s Guide for DS1000CA Series
IV
Safety Terms and Symbols
Terms in this Manual. These terms may appear in this manual:
WARNING
Warning statements indicate the conditions or practices that could result in injury or loss of life.
CAUTION
Caution statements indicate the conditions or practices that could result in
damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury or hazard that may immediately happen. WARNING indicates a potential injury or hazard that may immediately happen. CAUTION indicates that a potential damage to the instrument or other property
might occur.
Symbols on the Product. These symbols may appear on the product:
!
Hazardous Voltage
Refer to Instructions
Protective Earth Terminal
Test Grounding
Chassis Ground
!
!
Page 7
RIGOL
User‟s Guide for DS1000CA Series
V
Instrument at a Glance
Four types of DS1000CA series digital oscilloscopes will be introduced in this book: DS1072CA, DS1102CA, DS1202CA, DS1302CA
The instrument could help users to achieve high-quality measurement by excellent performance, powerful functions and easy operation.
DS1000CA is designed for 2GSa/s maximum real-time sample rate, 50GSa/s maximum equivalent sample rate, 300MHz maximum bandwidth and 20000wfms/s waveform capture, which are used to capture rapidly changing waveforms fast and accurately. Besides, abundant trigger, math and acquire functions enable it easy to capture and analyze waveforms, also, AUTO button is available to display a signal automatically. Whats more, design of panels, LCD, interface and buttons are friendly and clear, making operation more comfortable.
Main Features:
Dual Channel, Bandwidth:
300MHz (DS1302CA) 200MHz (DS1202CA) 100MHz (DS1102CA 70MHz (DS1072CA)
Color TFT LCD, 320× 234 pixels resolution. Plug and play USB storage. Adjustable waveform intensity, more effective waveform viewing. One-touch automatic setup, ease of use (AUTO). Supports waveforms, setups, bitmap and CSV format storage and reappearance
of waveforms and setups.
Delayed Scan Function, easy to give attention to both details and overview of a
waveform.
20 Automatic measurements. Automatic cursor tracking measurements. Waveform recorder, record and replay dynamic waveforms. User selectable fast offset calibration. Built-in FFT function, Frequency Counter.
Page 8
RIGOL
User‟s Guide for DS1000CA Series
VI
Digital filters, includes LPF, HPF, BPF, BRF. Pass/Fail Function, optically isolated Pass/Fail output. Add, Subtract and Multiply Mathematic Functions. Advanced trigger types include: Edge, Video, Pulse width, Slope, Alternative. Adjustable trigger sensitivity. Multiple Language User Interface. Pop-up menu makes it easy to read and easy to use. Built-in Chinese and English help system. Easy-to-use file system supports Chinese & English characters file name input.
Page 9
RIGOL
User‟s Guide for DS1000CA Series
VII
Structure of This Document
Chpter 1 Quick Start
To introduce the panls and menus to help user to realize the base operations.
Chapter 2 Operating Your Oscilloscope
More details to help users to understand the functions and measurements of instrument deeply.
Chapter 3 Application Examples
To show the measurement functions directly by application examples.
Chapter 4 Troubleshooting
Provide you troubleshooting to solve the problems during operating.
Chpter 5 Specifications
List common specifications and characteristics of DS1000CA series oscilloscope.
Chapter 6 Appendix
Information about accessories, warranties, services and supports and the like.
Page 10
RIGOL
User‟s Guide for DS1000CA Series
VIII
Content
Guaranty and Declaration ....................................................................... I
Safety Notices ........................................................................................ II
Instrument at a Glance ............................................................................. V
Chapter 1 Quick Start ........................................................................ 1-1
General Inspect .................................................................................... 1-2
Panels and User Interface ...................................................................... 1-3
Function Check ..................................................................................... 1-7
To Compensate Probes .......................................................................... 1-9
To Display a Signal Automatically ......................................................... 1-10
To Understand the Vertical System ....................................................... 1-11
To Understand the Horizontal System ................................................... 1-13
To Understand the Trigger System ....................................................... 1-15
Chapter 2 Operating Your Oscilloscope .............................................. 2-1
To Set Up the Vertical System ................................................................ 2-2
Channel Settings ............................................................................ 2-4
Math functions ............................................................................. 2-14
REF function ................................................................................ 2-18
To Set Up the Horizontal System .......................................................... 2-26
Delayed Scan ............................................................................... 2-28
X-Y Format .................................................................................. 2-29
To Set Up the Trigger System .............................................................. 2-30
Edge Trigger ................................................................................ 2-31
Pulse Width Trigger ...................................................................... 2-32
Video Trigger ............................................................................... 2-33
Slope Trigger ............................................................................... 2-36
Alternative Trigger ........................................................................ 2-38
Trigger Settings ............................................................................ 2-41
Trigger Holdoff ............................................................................. 2-42
To Set Up the Sampling System ........................................................... 2-46
To Set Up the Display System .............................................................. 2-50
To Store and Recall Waveforms or Setups ............................................. 2-53
To Set Up the Utility ............................................................................ 2-60
I/O Setup .................................................................................... 2-62
Page 11
RIGOL
User‟s Guide for DS1000CA Series
IX
Language ..................................................................................... 2-63
Pass/Fail ...................................................................................... 2-64
Waveform Recorder ...................................................................... 2-68
Print Set ...................................................................................... 2-72
Self-Calibration ............................................................................. 2-73
Service ........................................................................................ 2-74
Preference ................................................................................... 2-75
To Measure Automatically .................................................................... 2-77
Automatic measurement of voltage parameters ............................... 2-81
Automatic measurement of time parameters .................................... 2-82
To Measure with Cursors ...................................................................... 2-83
Manual Mode ................................................................................ 2-84
Track Mode .................................................................................. 2-86
Auto Mode ................................................................................... 2-88
To Use Run Control Buttons ................................................................. 2-89
Chapter 3 Application & Examples ..................................................... 3-1
Example 1: Taking Simple Measurements ................................................ 3-1
Example 2: View a Signal Delay Caused by a Circuit.................................. 3-2
Example 3: Capture a Single-Shot Signal ................................................. 3-3
Example 4: To Reduce the Random Noise on a Signal ............................... 3-4
Example 5: Making Cursor Measurements ................................................ 3-6
Example 6: The application of the X-Y operation ...................................... 3-8
Example 7: Triggering on a Video Signal ................................................ 3-10
Example 8: FFT Cursor Measurement .................................................... 3-12
Example 9: Pass/Fail Test .................................................................... 3-13
Chapter 4 Troubleshooting ................................................................ 4-1
Chapter 5 Specifications .................................................................... 5-1
Technical Specifications ................................................................... 5-2
General Specifications ..................................................................... 5-6
Chapter 6 Appendix ........................................................................... 6-1
Appendix A: DS1000CA series Accessories ............................................... 6-1
Appendix B: Warranty ............................................................................ 6-2
Appendix C: Maintenance ....................................................................... 6-3
Appendix D: Any Comment or Question? ................................................. 6-4
Index ...................................................................................................... 1
Page 12
Page 13
RIGOL
User‟s Guide for DS1000CA Series
1-1
Chapter 1 Quick Start
This chapter covers the following topics:
General Inspect Panels and User Interface Function Check To Compensate Probes To Display a Signal Automatically To Understand the Vertical System To Understand the Horizontal System To Understand the Trigger System
Page 14
RIGOL
User‟s Guide for DS1000CA Series
1-2
General Inspect
After receiving a new DS1000CA series oscilloscope, please inspect the instrument as follows:
1. Inspect the shipping container for damage.
Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has passed both electrical and mechanical test.
The consigner or carrier shall be liable for the damage to instrument resulting from shipment, without free maintenance or replacement by RIGOL.
2. Inspect the instrument.
In case of any damage, or defect, or failure, notify your RIGOL Sales Representative.
3. Check the accessories.
Accessories supplied with the instrument are listed in "Appendix A: DS1000CA series Accessories" in this guide. If the contents are incomplete or damaged, please notify your RIGOL Sales Representative.
Page 15
RIGOL
User‟s Guide for DS1000CA Series
1-3
Panels and User Interface
The first thing to do with a new oscilloscope is to know its panels. This chapter helps to be familiar with the layout of the knobs and keys and how to use them. Read the chapter carefully before further operations.
The Front Panel
The front panel is divided into function menu, “run control, “trigger control”, horizontal control and “vertical control” areas.
Thereinto, five grey menu buttons which lie beside LCD screen (defined No.1 to No.5) are used to switch current menu options. Other function buttons are used to enable the access to many measurement features associated with advanced functions, mathematics, reference or to run control features. In addition, multi-function knob is available to adjust waveform brightness or switch menus, but others are applied to adjust offset, volts/div or s/div.
Figure 1-1
Figure 1-1
Front panel of DS1000CA series oscilloscope
Power Switch Turn On/Off Multi-function Function Run Control
LCD Menu Knob Menu
USB Host Menu Buttons Analog Signal EXT TRIG
Input Channels Channel
Probe Compensator
Trigger Control
Horizontal Control
Vertical Control
Page 16
RIGOL
User‟s Guide for DS1000CA Series
1-4
Interfaces:
USB Host: It is used to connect DS1000CA with external USB devices when the
oscilloscope is regarded as Host Device”. For example, connect a USB flash device.
Signal Input: Two channels are provided: CH1 and CH2. EXT TRIG Input: It is applied to input external trigger signal.
Page 17
RIGOL
User‟s Guide for DS1000CA Series
1-5
The Rear Panel
Figure 1-2
Rear panel of DS1000CA series oscilloscope
Interfaces:
Pass/Fail: It is used to output the pass/fail test result. RS232: To connect the oscilloscope with external devices by serial port. USB Device: It is used to connect DS1000CA with external USB devices when the
oscilloscope is regarded as a Slave Device”. For example, connect with PC for remote control.
Notation definitions in this User’s Guide:
Throughout this manual, notation symbols of buttons and knobs are the same of those on front-panel.
A box around the name of the key denotes MENU function buttons on
front-panel, such as Measure.
( ) denotes the multi-function knob .  denotes the two POSITION knobs. denotes the two SCALE knobs. denotes the LEVEL knob. The name with a drop shadow denotes the menu operating key, such as
Waveforms soft key in Storage menu.
Use “ to divide operation steps. Such as: Storage Storage, indicates press
Storage button on the front panel, then press “Storage” menu.
Pass/Fail RS232 USB Device
Page 18
RIGOL
User‟s Guide for DS1000CA Series
1-6
Users Interface
Figure 1-3 Users Interface
Channel 2
Menu
Location of Waveform Trigger Point Run Status Window in Memory in Memory
CH1 Volts/Div CH2 Volts/Div Timebase Trigger Offset
Channel 1
Status Bar
Trigger Point in Waveform Window
Page 19
RIGOL
User‟s Guide for DS1000CA Series
1-7
Function Check
Perform this quick functional check to verify that the instrument is operating correctly.
1. Turn on the instrument
Use the power cord designed for the oscilloscope to connect the instrument and AC power which delivers 100 to 240 VAC
RMS
, 45Hz to 440Hz. Press the power switch to turn on the instruments, waiting until the display shows the waveform window.
2. Recall factory setting
Press Storage Storage Factory Load, to recall factory setting.
Figure 1-4
Turn on the instrument
WARNNING
To avoid electric shock, be sure the oscilloscope is properly grounded.
3. Input a signal to a channel of the oscilloscope
Push BNC plug of probe into CH1 BNC connector, then twist the BNC cable to
right to lock the probe in place. Switch on the probe to 10X.
!
Power Switch Storage Button
Page 20
RIGOL
User‟s Guide for DS1000CA Series
1-8
Figure 1-5 Probe connection Figure 1-6 Probe adjustment
Press CH1 Probe → 10X to set probe attenuation of the oscilloscope, so as
to make sure the measurement result is correct. (Default probe attenuation scale is 1X.)
Attach the probe tip to the Probe compensator connector and the reference
lead to the ground pin. Press AUTO, within a few seconds, a square wave will display (approximately 1 kHz 3 V peak- to- peak).
Push the OFF button or push the CH1 button again to turn off Channel 1.
Push the CH2 button to turn on channel 2, repeat steps above.
Probe compensator connector
Ground pin
Page 21
RIGOL
User‟s Guide for DS1000CA Series
1-9
To Compensate Probes
Perform this adjustment to match the characteristics of the probe and the channel input. This should be performed whenever attaching a probe to any input channel the first time.
1. From CH1 menu, set the Probe attenuation to 10X (press CH1 Probe 10X).
Set the switch to 10X on the probe and connect it to CH1 of the oscilloscope. When using the probe hook-tip, inserting the tip onto the probe firmly to ensure a proper connection.
2. Attach the probe tip to the Probe compensator connector and the reference lead
to the ground pin, Select CH1, and then press AUTO.
3. Check the shape of the displayed waveform.
Over compensated Correctly Compensated Under Compensated
Figure 1-9
Figure 1-7 Waveform compensation
4. If necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe
for the flattest square wave being displayed on the oscilloscope.
5. Repeat if necessary.
WARNNING
To avoid electric shock while using the probe, be sure the perfection of the insulated cable, and do not touch the metallic portions of the probe head while it is connected with a voltage source.
!
Page 22
RIGOL
User‟s Guide for DS1000CA Series
1-10
To Display a Signal Automatically
The oscilloscope has an automated feature to display the input signal best-fit. The input signal should be 50Hz or higher and a duty cycle is greater than 1%. Please take the following steps:
1. Connect a signal to the oscilloscope input channel.
2. Press AUTO.
The oscilloscope will set up VERTICAL, HORIZONTAL and TRIGGER controls automatically. You also could adjust the controls manually to get the best results if necessary.
Page 23
RIGOL
User‟s Guide for DS1000CA Series
1-11
To Understand the Vertical System
See figure 1-8 as below, the VERTICAL Control Area contains a series of buttons and knobs, which are not only used to set CH1, CH2, math and reference waveform functions, but also to adjust vertical position and Volts/div”.
Figure 1-8 Vertical Control
1. Press CH1, CH2, MATH, REF, then the corresponding menu, sign, waveform and
volts/div state will appear on the screen. To turn off channels, please press OFF.
2. Set vertical position of waveform display by knob.
Turning the knob, a voltage value is displayed for a short time indicating its value with respect to the ground reference located at the center of the screen. Also notice that the ground symbol on the left side of the display moves in conjunction with the knob.
Measurement hints
If the channel is DC coupled, measuring the DC component of the signal by simply noting its distance from the ground symbol. If the channel is AC coupled, the DC component of the signal is blocked, allow you to use greater sensitivity to display the AC component of the signal.
Page 24
RIGOL
User‟s Guide for DS1000CA Series
1-12
Vertical offset back to 0 shortcut key
Turn the knob to change the vertical display position of channel and press the knob to set the vertical display position back to 0 as a shortcut key, this is especially helpful when the trace position is far out of the screen and want it to get back to the screen center immediately.
3. Set vertical Volts/div by knob.
Rotate to change Volts/div”, the change will be displayed in real time at the bottom of interface. Besides, the Volts/div contains two forms: Fine and Coarse.
Coarse/Fine Shortcut key
The Coarse/Fine vertical control can be set by simply pressing the vertical
knob.
Page 25
RIGOL
User‟s Guide for DS1000CA Series
1-13
To Understand the Horizontal System
See figure 1-9 as below, the HORIZONTAL Control Area contains a button and two knobs, which are not only used to set up horizontal parameters and timebase, but also to adjust horizontal position of waveforms.
Figure 1-9 Horizontal Control
1. The horizontal knob adjusts signal horizontal position on
waveform window.
Turning the horizontal knob, the waveform will move to left or right and a sign at the left-bottom of the screen will vary to show the horizontal position of the waveform.
Horizontal offset back to 0 shortcut key
Press the knob to set the horizontal offset to 0 as a shortcut key, this is especially helpful when the trigger point is far out of the screen and want it to get back to the screen center immediately.
2. Press MENU button to display Time menu, setting delayed state, time base
mode and trigger offset.
Page 26
RIGOL
User‟s Guide for DS1000CA Series
1-14
Term Explanations
Trigger offset: Denotes the real position of the trigger point relative to the midpoint of the memory. In this setting, the trigger position will be changed horizontally when you turning the knob.
3. Set horizontal s/div by knob.
The horizontal knob changes the sweep speed in a 1-2-5 step sequence, and displays the value in the status bar. The time base ranges of the DS1000CA series is from 2ns/div* to 50s/div.
NOTE*: The speed of horizontal scan varies by different models.
Delayed Scan Shortcut key
To press the knob in the horizontal control area on the front-panel is another way to enter or exit Delayed Scan mode.
Page 27
RIGOL
User‟s Guide for DS1000CA Series
1-15
To Understand the Trigger System
See figure 1-10 below, the TRIGGER Control Area contains a knob and three buttons, which are used to set trigger level, trigger parameters and force trigger.
Figure 1-10
Trigger Control
1. Set trigger level by knob.
A nacarat trigger line and trigger sign will move up and down when you turn the
knob, and they will disappear about 5 seconds after stop turning knob. The trigger level value will change on the screen at the same time you turn the trigger line.
2. Press MENU button in the Trigger control. A soft button menu appears on the
display showing the trigger setting choices as shown in Figure 1-11.
Trigger Level to 0 Shortcut Key
Turn the knob to change the trigger level value and press the
knob to set trigger level back to 0 as a shortcut key.
Page 28
RIGOL
User‟s Guide for DS1000CA Series
1-16
· Press NO.1 menu button to choose Edge.
· Press NO.2 menu button to choose Source as CH1.
· Press NO.3 menu button to choose Slope as .
· Press NO.4 menu button to choose Sweep as Auto.
· Press NO.5 menu button to enter secondary menu of Set Up.
Figure 1-11
NOTE: The trigger mode, slope and source change in conjunction with the status bar on the top-right of the screen.
3. Press 50%
The 50% button sets the trigger level to the center of the signal.
4. Press FORCE
Starts an acquisition regardless of an adequate trigger signal, usually used in “Normal” or “Single” trigger mode. This button has no effect if the acquisition is already stopped.
Local/Remote control
When the instrument is in remote mode, press FORCE to switch to local mode.
Page 29
RIGOL
User‟s Guide for DS1000CA Series
2-1
Chapter 2 Operating Your Oscilloscope
By now, a user should understand the VERTICAL, HORIZONTAL and TRIGGER control systems and knows how to determine the system setup from the status bar of a DS1000CA-series digital oscilloscope. This chapter will go through all groups of front-panel buttons, knobs and menus; and further the knowledge of the operation by hints in this guide. It is recommended to perform all of the following exercises to get the most of the powerful measurement capabilities of the oscilloscope. If you haven‟t know the operations above, please refers to chapter 1.
This chapter covers the following topics:
To Set Up the Vertical System ( CH1, CH2, MATH, REF, OFF, Vertical
, Vertical )
To Set Up the Horizontal System ( MENU, Horizontal , Horizontal
)
To Set Up the Trigger System ( , MENU, 50%, FORCE) To Set Up the Sampling System (Acquire) To Set Up the Display System (Display) To Store and Recall Waveforms or Setups (Storage) To Set Up the Utility (Utility) To Measure Automatically (Measure) To Measure with Cursors (Cursor) To Use Run Control Buttons (AUTO, RUN/STOP)
Page 30
RIGOL
User‟s Guide for DS1000CA Series
2-2
To Set Up the Vertical System
Two analog channels are provided by DS1000CA: CH1 and CH2. Press the corresponding button on the front panel to turn the channels on/off. The backlight indicates the channel is currently active. Press the button again to turn the channel off. When channel is currently selected, press OFF will turn it off and the backlight goes off.
Table 2-1 The Channels menu
Channel Mode
Settings
Status Indicator
Channel 1 (CH1)
ON Selected OFF
CH1 (yellow letter)
CH1 (black letter) No indicator
Channel 2 (CH2)
ON Selected OFF
CH2 (blue letter)
CH2 (black letter) No indicator
MATH
ON Selected OFF
MATH (purple letter)
MATH(black letter) No indicator
All functions applied will base on operating the instrument with channels, so MATH and REF can be regarded as relatively isolated channels.
Page 31
RIGOL
User‟s Guide for DS1000CA Series
2-3
Knobs Explanations
Use the vertical controls to display signal waveforms by adjusting the vertical
and , and setting the input parameters.
1. Using vertical knob
The vertical control changes the vertical position of signal waveforms in all channels (including MATH and REF). The resolution changes according to the vertical level set. Pressing this knob will clear the channel offset to zero.
2. Using vertical knob
The vertical control changes the vertical sensitivity of signal waveforms in all channels (including MATH and REF). If the Volts/Div is set to
“Coarse”, the waveform scales in a 1-2-5 step sequence from 2 mV to 10 V. If the Volts/Div is set to “Fine”, it scales to small steps between the coarse settings.
3. Channels would be adjustable by the vertical and only
when they are selected.
4. During the vertical positioning, a position message is displayed on the left
bottom of the screen, in the same color as the corresponding channel. The unit is V (Volts).For instance: POS: 32.4mV Colours of words and channel waveform are the same, the unit is V.
Page 32
RIGOL
User‟s Guide for DS1000CA Series
2-4
Channel Settings
Each channel of DS1000CA has an operation menu and it will pop up after pressing CH1 or CH2 button. Take CH1 as an example, the settings of all items in the menu are shown in the table below.
Figure 2-1 Table 2-2 The Channel menu (Page 1/2)
Menu
Settings
Comments
Coupling
AC
DC
GND
Block the DC component of the input Signal. Pass both AC and DC components of the input signal Disconnect the input signal.
BW Limit
ON
OFF
Limit the channel bandwidth to 20MHz to reduce display noise. Get full bandwidth.
Probe
1X 5X 10X 50X 100X 500X 1000X
Set this to match the probe attenuation factor to make the vertical scale readout correct.
Digital filter
Setup digital filter (See table 2-5).
1/2
Go to the next menu page (The followings are the same, no more explanation).
Page 33
RIGOL
User‟s Guide for DS1000CA Series
2-5
Figure 2-2 Table 2-3 The Channel menu (Page 2/2)
Menu
Settings
Comments
2/2
Back to the previous menu page (The followings are the same, no more explanation).
Volts/Div
Coarse Fine
Define a 1-2-5 sequence. To change the resolution to small steps between the coarse settings.
Invert
ON OFF
Turn on the invert function. Restore to original display of the waveform.
Unit
V/ A/ W/ U
Set “V”, “A”, “W” or “U” as the unit of
vertical channel.
Input
50Ω* 1MΩ
Press this key to select the input resistance to 50Ω or 1MΩ.
NOTE*: This parameter is for DS1202CA and DS1302CA only.
Page 34
RIGOL
User‟s Guide for DS1000CA Series
2-6
1. Channel coupling
To use CH 1 as an example, input a sine wave signal with DC shift.
Press CH1CouplingAC to set “AC” coupling. It will pass AC component blocks the DC component of the input signal.
Figure 2-3 AC coupling setting
Press CH1CouplingDC, to set “DC” coupling. It will pass both AC and DC components of the input signal.
Figure 2-4 DC coupling setting
DC Coupling Setup
DC Coupling Status Symbol
AC Coupling Setup
AC Coupling Status Symbol
Page 35
RIGOL
User‟s Guide for DS1000CA Series
2-7
Press CH1CouplingGND, to set “GND” coupling, it disconnects the input signal.
Figure 2-5 GND coupling setting
GND Coupling
GND Coupling Status Symbol
Page 36
RIGOL
User‟s Guide for DS1000CA Series
2-8
2. Set up the channel bandwidth limit
Take CH1 for an example, input a signal containing high frequency component.
Press CH1BW LimitOFF, to set up bandwidth limit to “OFF” status. The oscilloscope is set to full bandwidth and passing the high frequency component in the signal.
Figure 2-6 Turn off the BW limit
Press CH1BW LimitON, to set up bandwidth limit to “ON” status. It will reject the frequency component higher than 20MHz.
Figure 2-7 Turn on the BW limit
Turn Off the BW Limit
Mark of BW Limit
20M BW Limit
Page 37
RIGOL
User‟s Guide for DS1000CA Series
2-9
3. Probe Attenuation Setting
The oscilloscope allows selecting the attenuation factor for the probe. The attenuation factor changes the vertical scaling of the oscilloscope so that the measurement results reflect the actual voltage levels at the probe tip. To change (or check) the probe attenuation setting, press the CH1 or CH2 button (according to which channel in using). Toggle the Probe soft button to match the attenuation factor of the probe. This setting remains in effect until changed again.
Figure 2-8 show an example for using a 1000:1 probe and its attenuation factor.
Figure 2-8 Setting probe attenuation
Table 2-4 Probe setting
Probe attenuation factors
Corresponding settings
1:1
1X
5:1
5X
10:1
10X
50:1
50X
100:1
100X
500:1
500X
1000:1
1000X
Vertical Volts/div
Probe Attenuation
Page 38
RIGOL
User‟s Guide for DS1000CA Series
2-10
4. Digital Filter
Press CH1Digital filter, display the digital filter menu.
Figure 2-9 Turn off digital filter
Figure 2-10 Turn on digital filter
Turn off Digital Filter
Mark of Digital Filter
Turn on Digital Filter
Page 39
RIGOL
User‟s Guide for DS1000CA Series
2-11
Turn ( ) knob to set high and low limit of frequency.
Figure 2-11 Table 2-5 The Filter menu
Menu
Settings
Comments
Digital Filter
Off On
Turn off the digital filter. Turn on the digital filter.
Filter Type
Setup as LPF (Low Pass Filter). Setup as HPF (High Pass Filter). Setup as BPF (Band Pass Filter). Setup as BRF (Band Reject Filter).
Upper limit
<frequency upper limit>
Turn ( ) knob to set high limit.
Lower limit
<frequency lower limit>
Turn ( ) knob to set low limit.
Back to higher level menu (The followings are the same, no more explanation).
Page 40
RIGOL
User‟s Guide for DS1000CA Series
2-12
5. Volts/Div settings
The Volts/Div control has Coarse or Fine configuration. The Vertical Sensitivity is 1mv/div ~10V/div.
Coarse: It is the default setting of Volts/Div in a 1-2-5 step sequence from 1mV/div, 2mV/div, 5mV/div, 10mV/div to10 V/div.
Fine: This setting changes the vertical scale to small steps between the coarse settings. It will be helpful to adjust the waveform in smooth steps.
Figure 2-12 Fine configurations
Coarse/Fine Shortcut key
To change Coarse/Fine setting, not only by menu but also by pressing vertical knob.
Fine Adjustment
Fine Adjustment Data
Page 41
RIGOL
User‟s Guide for DS1000CA Series
2-13
6. To invert a waveform
Invert turns the displayed waveform 180 degrees, as respect to the ground level. When the oscilloscope is triggered on the inverted signal, the trigger is also inverted. Figure 2-13 and 2-14 show the changes before after the inversion respectively.
Figure 2-13
The waveform before inversion
Figure 2-14
The waveform after inversion
Invert Off
Invert On
Page 42
RIGOL
User‟s Guide for DS1000CA Series
2-14
Math functions
The mathematic functions include “Add”, “Subtract”, “Multiply” and “FFT” for Channel 1 and Channel 2. The mathematic result can be measured by the grid and the cursor.
Press MATH button to enter “Math” menu, meanwhile, sign of math scale will appear under the screen. See figure below, waveform of channel1 is marked yellow, waveform of channel2 is marked blue, and waveform of channel1 adding channel2 is marked purple.
Figure 2-15 The Math function
Figure 2-16 Table 2-6 The Math menu
Menu
Settings
Comments
Operation
A+B A-B A× B FFT
Add source A and source B. Subtract source B from source A. Multiply source B by source A. Fast Fourier Transform.
Source A
CH1 CH2
Define CH1 or CH2 as source A.
Source B
CH1 CH2
Define CH1 or CH2 as source B.
Invert
ON OFF
Invert the MATH waveform. Restore to original waveform display.
MATH scale
Page 43
RIGOL
User‟s Guide for DS1000CA Series
2-15
1. FFT Analyse
The FFT (Fast Fourier Transform) process converts a time-domain signal into its frequency components mathematically. FFT waveforms are useful in the following applications:
Measuring harmonic content and distortion in systems Characterizing noise in DC power supplies Analyzing vibration
Press MATHOperateFFT to enter FFT setting menu.
Figure 2-17 Table 2-7 The FFT menu
Menu
Settings
Comments
Operate
A+B A-B A× B FFT
Add source A to source B. Subtract source B from source A. Multiply source B by source A. Fast Fourier Transform.
Source
CH1 CH2
Define CH1 or CH2 as FFT source.
Window
Rectangle Hanning Hamming Blackman
Select window for FFT. Display
Split Full screen
Display FFT waveform on half screen. Display FFT waveform on full screen.
Scale
V
RMS
dBV
RMS
Set “V
RMS
” as vertical unit.
Set “dBV
RMS
” as vertical unit.
Key points for FFT
1. Signals that have a DC component or offset can cause incorrect FFT
waveform component magnitude values. To minimize the DC component, choose AC Coupling on the source signal.
2. To reduce random noise and aliases components in repetitive or
single-shot events, set the oscilloscope acquisition mode to Average.
3. To display FFT waveforms with a large dynamic range, use the dBV
RMS
scale. The dBV
RMS
scale displays component magnitudes using a log scale.
Page 44
RIGOL
User‟s Guide for DS1000CA Series
2-16
2. Selecting an FFT Window
Oscilloscope record the FFT change for a limit time assume that YT waveform repeat ceaselessly. If cycle is integer, YT waveform voltage is the same at starting and ending point, and waveform is consecutive. Contrary, If cycle is non-integer, the voltage is different at starting and ending point, leading to the high frequency instantaneous intermit, which is called “leak”. To avoid leak, multiply a window function to force the value is 0 at starting and ending point.
DS1000CA provides four FFT functions with different feature, which are need to be selected according to measuring waveform.
Table 2-8 FFT Windows
Window
Features
Best for measuring
Rectangle
Best frequency resolution worst magnitude resolution. This is essentially the same as no window.
Transients or bursts, the signal levels before and after the event are nearly equal. Equal-amplitude sine waves with fixed frequencies. Broadband random noise with a relatively slow varying spectrum.
Hanning
Better frequency, poorer magnitude accuracy than Rectangular.
Sine, periodic, and narrow-band random noise.
Hamming
Hamming has slightly better frequency resolution than Hanning.
Transients or bursts where the signal levels before and after the events are significantly different.
Blackman
Best magnitude, worst frequency resolution.
Single frequency waveforms, to Find higher order harmonics.
Page 45
RIGOL
User‟s Guide for DS1000CA Series
2-17
Term Explanations
FFT Resolution: The quotient between sampling rate and number of FFT
points. With a fixed FFT points, the lower sampling rate results in better resolution.
Nyquist Frequency
The highest frequency that any real-time digitizing oscilloscope can acquire
without aliasing. It‟s normally half of the sample rate. This frequency is called
the Nyquist frequency. Frequency above the Nyquist frequency will be under sampled, causing a situation known as aliasing.
Page 46
RIGOL
User‟s Guide for DS1000CA Series
2-18
REF function
Reference Waveforms are saved waveforms to be selected for display. The reference function will be available after saving the selected waveform to non-volatile memory.
Press REF button to display the Reference waveform menu.
Figure 2-18 Table 2-9 REF menu when using internal memory
Menu
Settings
Comments
Source
CH1 CH2 MATH/FFT
Select channel 1 as REF channel.
Select channel 2 as REF channel.
Select MATH/FFT as REF channel.
Location
Internal External
Select memory location in scope. Select memory location out scope.
Save
Save REF waveform
Imp./Exp.
Go to import/export menu (see table 2-11).
Reset
Reset REF waveform.
Figure 2-19 Table 2-10 REF menu when using external memory
Menu
Settings
Comments
Source
CH1 CH2 MATH/FFT
Select channel 1 as REF channel.
Select channel 2 as REF channel.
Select Math/FFT as REF channel.
Location
Internal External
Select memory location in scope. Select memory location out scope.
Save
Save REF waveform to outer memory location.
Import
Go to import menu (see table 2-15).
Reset
Reset REF waveform.
Page 47
RIGOL
User‟s Guide for DS1000CA Series
2-19
1. Import and Export
Switch in external memory and select Internal location. Then, press REF Imp. /Exp. and go to the following interface.
Figure 2-20 Import of export the figure
Figure 2-21 Table 2-11
(a) (a) The Imp. /Exp. Menu (Files)
Note: When the explorer is set to Path, Import and Delete File are grayed out
and disabled. When the explorer is set to Directories, the menu and comments are as shown in the table below.
Menu
Settings
Comments
Explorer
Path Directories Files
Set the explorer to Path, directories or files.
Export
Export the REF file from internal memory to export memory (see table 2-12).
Import
Import the REF file to internal memory.
Delete File
Delete file.
Page 48
RIGOL
User‟s Guide for DS1000CA Series
2-20
(b) (b) The Imp. /Exp. Menu (Directories)
Menu
Settings
Comments
Explorer
Path Directories Files
Set the explorer to Path, directories or files.
New Folder
Create a new folder under the current path.
Del Folder
Delete the folder currently selected.
Page 49
RIGOL
User‟s Guide for DS1000CA Series
2-21
2. Export
Switch in external memory and select Internal location. Then, press REFImp. /Exp. Export and go to the following interface.
Figure 2-22 Figure export
Figure 2-23 Table 2-12 The Export menu
Menu
Settings
Comments
Move the cursor up.
Move the cursor down.
To delete chosen letter.
Save
Execute the operation.
Page 50
RIGOL
User‟s Guide for DS1000CA Series
2-22
3. Save to External memory
Switch in external memory and select External location. Then, press REFSave and go to the following interface.
Figure 2-24 Save the figure
Figure 2-25 Table 2-13 The Save menu
Menu
Settings
Comments
Explorer
Path Directories Files
Switch among Path, Directories and Files.
New File (Folder)
Set up new file in Path and Files. Set up new folder in Directories.
Delete File (Folder)
Delete file (folder).
Page 51
RIGOL
User‟s Guide for DS1000CA Series
2-23
4. New File (or New Folder)
Switch in external memory and select External location. Then, press REFSaveNew File (or New Folder) and go to the following interface.
Figure 2-26 Input interface
Figure 2-27 Table 2-14 The New File menu
Menu
Settings
Comments
Move the cursor up.
Move the cursor down.
To delete chosen letter.
Save
Execute the operation.
Character Window
Switch Capital On/Off
Switch Chinese/English
Spell Key In
File Name Key In
Page 52
RIGOL
User‟s Guide for DS1000CA Series
2-24
5. Import
Switch in external memory and select External location. Then, press REFImport and go to the following interface.
Figure 2-28 Figure import
Figure 2-29 Table 2-15 The Import menu
Menu
Settings
Comments
Explorer
Path Directories Files
Switch among Path, Directories and Files.
Import
Import the REF file into internal memory.
Page 53
RIGOL
User‟s Guide for DS1000CA Series
2-25
6. Display a Reference Waveform
Figure 2-30 Reference waveform display
Operation Explanations:
1. Push REF button to show the reference waveform menu.
2. Press soft button 1 to select the reference channel: CH1, CH2, MATH/FFT.
3. Turn vertical and vertical to adjust the REF waveform
to a suitable position.
4. Press soft button No.2 to select the save location of REF waveform.
5. Press soft button No.3 to save the waveform as REF.
NOTE: The Reference function is not available in X-Y mode.
Page 54
RIGOL
User‟s Guide for DS1000CA Series
2-26
To Set Up the Horizontal System
Press MENU button to enter the horizontal setting system.
Figure 2-31 Table 2-16 The Horizontal menu
Menu
Settings
Comments
Delayed
ON OFF
Turn on Delayed Scan mode. Turn off the Delayed Scan mode.
Time Base
Y-T
X-Y
Roll
[1]
Y-axis indicates vertical voltage, X-axis indicates horizontal time. Y-axis indicates voltage of CH2, X-axis indicates voltage of CH1. The waveform display updates from right to left.
Trig-offset Reset
Reset
Adjust to the center.
NOTE:
[1] No trigger or horizontal offset control of waveforms is available
during Roll Mode, and it‟s only available when set to 500ms/div or slower.
Term Explanations:
Slow Scan Mode: This mode is available when the horizontal time base is
set to 50ms or slower. In this mode, the oscilloscope acquires sufficient data for the left part to the trigger point, then wait for trigger, when trigger occurs, it continue to draw the rest part from the trigger point to the end of the right side. When choosing this mode to view low frequency signals, it is recommended that the channel coupling should be set as DC.
Page 55
RIGOL
User‟s Guide for DS1000CA Series
2-27
The Horizontal Knobs
The horizontal controls change the horizontal scale and position of waveforms. The horizontal center of the screen is the time reference for waveforms. Changing the horizontal scale causes the waveform to expand or contract with respect to the screen center. Horizontal position changes the displayed waveform position, relative to the trigger point.
1. Horizontal knob is used to adjust the horizontal position of all
channel (include Math) waveforms. The resolution of this control varies with the time base. Pressing this button clears trigger offset and moves the trigger point to the horizontal center of the screen.
2. Horizontal is used to select the horizontal time/div (scale factor) for
the main or the Delayed Scan time base. When Delayed Scan is enabled, it changes the width of the window zone by changing the Delayed Scan time base.
Page 56
RIGOL
User‟s Guide for DS1000CA Series
2-28
Delayed Scan
The Delayed Scan is a magnified portion of the main waveform window. Use Delayed Scan to locate and horizontally expand part of the main waveform window for a more detailed (higher horizontal resolution) analysis of signal. The Delayed Scan time base setting cannot be set slower than the Main time base setting.
Press MENUDelayedON, or press horizontal knob to turn on delayed scan function. See the waveform below.
Figure 2-32 Delayed Scan windows
The screen splits into two parts as above: The upper half displays the main waveform window and the lower half displays an expanded portion of the main waveform window. This expanded portion of the main window is called the Delayed Scan window. Two blocks shaded at the upper half; the un-shaded portion is expanded in the lower half. The horizontal and
knobs control the size and position of the Delayed Scan. The value at bottom of the screen is the main time base and the value on the center bottom means the Delayed Scan time.
Waveform to be Horizontally Expanded
Expanded Waveform in Horizontal Main Timebase
Time Base of Delayed Scan
Page 57
RIGOL
User‟s Guide for DS1000CA Series
2-29
X-Y Format
This format is useful for studying phase relationships between two signals. CH1 in the horizontal axis(X) and CH2 in the vertical axis(Y), the oscilloscope uses a none-trigger acquisition mode, data is displayed as dots.
Press MENUTime BaseX-Y to enter X-Y mode.
Figure 2-33 X-Y display format
The oscilloscope could capture waveform with arbitrary sample rate in Y-T mode (2GSa/s maximum for single channel, 1GSa/s maximum for double channels). While, in X-Y mode, the sample rate is up to 500MSa/s. Generally, the sample rate is more lower, the Lissajous graphics is more better.
The following modes or functions will not work in X-Y format.
Cursor Measurements Pass/Fail Test REF or MATH Operations Delayed Scan Mode Vector Display Mode Horizontal knob Trigger Controls
Page 58
RIGOL
User‟s Guide for DS1000CA Series
2-30
To Set Up the Trigger System
The trigger determines when the oscilloscope starts to acquire data and display a waveform. When a trigger is set up properly, it can convert unstable displays or blank screens into meaningful waveforms.
When the oscilloscope starts to acquire a waveform, it collects enough data so that it can draw the waveform to the left of the trigger point. The oscilloscope continues to acquire data while waiting for the trigger condition to occur. After it detects a trigger, the oscilloscope continues to acquire enough data so that it can draw the waveform to the right of the trigger point.
Press MENU button to enter trigger system setting interface.
Figure 2-34 Trigger controls
DS1000CA provides five trigger modes: Edge, Pulse, Slope, Video and Alternative.
Edge: An edge trigger occurs when the trigger input passes through a specified voltage level in the specified slope direction.
Pulse: Use this trigger type to catch pulses with certain pulse width. Video: Use video trigger on fields or lines for standard video signals. Slope: The oscilloscope begins to trigger according to the signal rising or falling
speed. Alternative: Trigger on non-synchronized signals.
Page 59
RIGOL
User‟s Guide for DS1000CA Series
2-31
Edge Trigger
An edge trigger determines whether the oscilloscope finds the trigger point on the rising or the falling edge of a signal. Select Edge trigger Mode to trigger on Rising
edge, falling edge or rising & falling edge.
Press MENUModeEdge to enter the following menu.
Figure 2-35 Table 2-17 The Trigger menu
Menu
Settings
Comments
Source
CH1 CH2 EXT EXT/5
AC Line
Select CH1 as trigger signal. Select CH2 as trigger signal. Select EXT TRIG as trigger signal. Select attenuated EXT TRIG/5 as trigger signal. Select power line as trigger signal.
Slope
Trigger on rising edge. Trigger on falling edge. Trigger on both ring & falling edge.
Sweep
Auto Normal Single
Acquire waveform even no trigger occurred. Acquire waveform when trigger occurred. When trigger occurs, acquire one waveform then stop.
Set up
To go to Set Up menu, see table 2-32
Page 60
RIGOL
User‟s Guide for DS1000CA Series
2-32
Pulse Width Trigger
Pulse trigger occurs according to the width of pulse. The abnormal signals can be
detected through setting up the pulse width condition.
Press MENUModePulse to enter the following menu.
Figure 2-36 Table 2-18 The Trigger menu (Page 1/2)
Menu
Settings
Comments
Source
CH1 CH2 EXT
EXT/5
Selects CH1 as trigger signal. Select CH2 as trigger signal. Select EXT TRIG as trigger signal. Select attenuated EXT TRIG/5 as trigger signal.
When
Set Pulse width as “+pulse width less than” Set Pulse width as +pulse width more than” Set Pulse width as “+pulse width equal to” Set Pulse width as -pulse width less than” Set Pulse width as -pulse width more than” Set Pulse width as -pulse width equal to
Settings
<Width>
Set required pulse width.
Figure 2-37 Table 2-19 The Trigger menu (Page 2/2)
Menu
Settings
Comments
Sweep
Auto Normal Single
Acquire waveform even no trigger occurred. Acquire waveform when trigger occurred. When trigger occurs, acquire one waveform and then stop.
Set Up
To go to Set Up menu, see table 2-32.
NOTE: The adjustable range of Pulse width is 20ns to 10s. When the condition is met, it will trigger and acquire the waveform.
Page 61
RIGOL
User‟s Guide for DS1000CA Series
2-33
Video Trigger
Choose video trigger to trigger on fields or lines of NTSC, PAL, or SECAM standard video signals. Trigger coupling preset to DC.
Press MENUModeVideo to enter the following menu.
Figure 2-38 Table 2-20 The Video Trigger menu (Page 1/2)
Menu
Settings
Comments
Source
CH1 CH2 EXT EXT/5
Selects CH1 as trigger source. Select CH2 as trigger source. Select EXT TRIG as trigger source. Select EXT TRIG/5 as trigger source.
Polarity
Normal polarity
Inverted polarity
Trigger on the negative edge of the sync pulse
Trigger on the positive edge of the sync pulse
Sync
All Lines Line Num Odd field Even field
Trigger on all lines. Trigger on an specified line. Select to trigger on odd field. Select to trigger on even field.
Note*: normal polarity is applicable to signal with negative horizontal sync pulse;
inverted polarity is applicable to signal with positive horizontal sync pulse.
Page 62
RIGOL
User‟s Guide for DS1000CA Series
2-34
Figure 2-39 Table 2-21 The Video Trigger menu (Page 2/2, when Sync is set as the
specified line)
Menu
Settings
Comments
Line Num
<line number>
Specify the line on which the oscilloscope
triggers. The range of the line number available is from 1 to 525 (NTSC) or from 1 to 625 (PAL/SECAM).
Standard
PAL/SECM NTSC
Select Video standard.
Sweep
Auto
Normal
Single
Force the oscilloscope to trigger in the absence of trigger condition. Lets oscilloscope to trigger in the suitable trigger condition. Lets oscilloscope to trigger one time in the suitable trigger condition, and then stop.
Set Up
To go to set up menu, see table 2-33.
Page 63
RIGOL
User‟s Guide for DS1000CA Series
2-35
Figure 2-40 Table 2-22 The Video Trigger menu (When the Sync is set as All lines,
Odd field and Even field)
Select Line Synchronization: Select Filed Synchronization:
Figure 2-41 Figure 2-42
Video Trigger: Line Synchronization Video Trigger: Field Synchronization
Menu
Settings
Comments
Standard
PAL/SECAM NTSC
Select Video standard.
Sweep
Auto
Normal
Single
Force the oscilloscope to trigger in the absence of trigger condition. Lets oscilloscope to trigger in the suitable trigger condition. Lets oscilloscope to trigger one time in the suitable trigger condition, then stop.
Set Up
To go to set up menu, see table 2-33.
Page 64
RIGOL
User‟s Guide for DS1000CA Series
2-36
Slope Trigger
Slope trigger sets the oscilloscope as the positive/negative slope trigger within the specified time.
Press MENUModeSlope to enter the following menu.
Figure 2-43 Table 2-23 The Slope Trigger menu (Page 1/2)
Menu
Settings
Comments
Source
CH1 CH2 EXT EXT/5
Set channel 1 as trigger source. Set channel 2 as trigger source. Set EXT. channel as trigger source. Set EXT/5 as trigger source.
When
Set slope condition as +slope more than
Set slope condition as +slope less than
Set slope condition as +slope equal to
Set slope condition as -slope more than
Set slope condition as -slope less than
Set slope condition as -slope equal to
Time <Time Set>
Set slope time.
Figure 2-44 Table 2-24 The Slope Trigger menu (Page2/2)
Menu
Settings
Comments
Vertical
Select the level that can be adjusted by
.
Sweep
Auto
Normal
Single
Force the oscilloscope to trigger in the
absence of trigger condition.
Lets oscilloscope to trigger in the suitable
trigger condition.
Lets oscilloscope to trigger one time in the
suitable trigger condition, then stop.
Set Up
To go to set up menu. See Table 2-32.
Page 65
RIGOL
User‟s Guide for DS1000CA Series
2-37
NOTE: Slope time can be set from 20ns to 10s. When a signal meets the trigger condition, scope will execute the acquisition. You can adjust LEVEL A/ LEVEL B or both simultaneous by turning the knob.
Page 66
RIGOL
User‟s Guide for DS1000CA Series
2-38
Alternative Trigger
When alternative trigger is on, the trigger sources come from two vertical channels. This mode can be used to observe two non-related signals. You can choose two different trigger modes for the two vertical channels. The options are as follows: Edge, Pulse, Slope and video. The info of the trigger level of the two channels will be displayed on the upper-right of the screen.
Press MENUModeAlternative to enter the following menu.
Figure 2-45 Table 2-25 The Alternative menu (Trigger Type: Edge)
Menu
Settings
Comments
Select
CH1
CH2
Set trigger mode for Channel 1.
Set trigger mode for Channel 2.
Type
Edge
Set Edge Trigger as the trigger type.
Slope
Trigger on rising edge.
Trigger on falling edge.
Trigger on both ring & falling edge.
Set Up
To go to set up menu. See Table 2-32.
Figure 2-46 Table 2-26 The Alternative menu (Trigger Type: Pulse Page 1/2)
Menu
Settings
Comments
Select
CH1
CH2
Set trigger mode for Channel 1.
Set trigger mode for Channel 2.
Type
Pulse
Set Pulse Trigger for the channel.
When
Set Pulse width as “+pulse width less than” Set Pulse width as +pulse width more than” Set Pulse width as “+pulse width equal to” Set Pulse width as -pulse width less than” Set Pulse width as -pulse width more than” Set Pulse width as -pulse width equal to
Page 67
RIGOL
User‟s Guide for DS1000CA Series
2-39
Figure 2-47 Table 2-27 The Alternative menu (Trigger Type: Pulse Page 2/2)
Men
u
Settings
Comments
Setting
<pulse width>
Set the width of the pulse.
Set Up
To go to set up menu. See Table 2-32.
Figure 2-48 Table 2-28 The Alternative menu (Trigger Type: Slope Page 1/2)
Menu
Settings
Comments
Select
CH1 CH2
Set trigger mode for Channel 1.
Set trigger mode for Channel 2.
Type
Slope
Set Slope Trigger for the vertical channel.
When
Set slope condition as +slope more than
Set slope condition as +slope less than
Set slope condition as +slope equal to
Set slope condition as -slope more than
Set slope condition as -slope less than
Set slope condition as -slope equal to
Figure 2-49 Table 2-29 The Alternative menu (Trigger Type: Slope Page 2/2)
Menu
Settings
Comments
Time
<Time Set >
Set slope time.
Vertical
Select the level to be adjusted by
.
Set Up
To go to set up menu. See Table 2-32
and table 2-33.
Page 68
RIGOL
User‟s Guide for DS1000CA Series
2-40
Figure 2-50 Table 2-30 The Alternative menu (Trigger Type: Video Page 1/2)
Menu
Settings
Comments
Select
CH1 CH2
Set trigger mode for CH1.
Set trigger mode for CH2.
Type
Video
Video Trigger for the channel.
Polarity
Normal polarity Inverted polarity
Trigger on the negative edge
of the sync pulse
Trigger on the positive edge of
the sync pulse
Note*: normal polarity is applicable to signal with negative horizontal sync pulse;
inverted polarity is applicable to signal with positive horizontal sync pulse.
Figure 2-51 Table 2-31 The Alternative menu (Trigger Mode: Video Page 2/2)
Menu
Settings
Comments
Sync
ALL lines
Line Num
Odd field
Even field
Trigger on all lines.
Trigger on an specified line.
Select to trigger on odd field or even
field.
Line Num
<Lines Set >
Select the specified line number for
sync.
Standard
PAL/SECM
NTSC
Select Video standard.
Set Up
To go to set up menu. See Table
2-33.
Page 69
RIGOL
User‟s Guide for DS1000CA Series
2-41
Trigger Settings
Set up different trigger settings according to different trigger modes. When in the mode of Edge and Pulse, only Holdoff is adjustable. When source is non-digital channel and in slope trigger, only trigger coupling, trigger sensitivity and Holdoff can be set. For video trigger, Sensitivity and Holdoff can be set.
Figure 2-52 Table 2-32 The Trigger Set Up menu (Settings for trigger coupling,
trigger sensitivity and holdoff)
Figure 2-53 Table 2-33 The Trigger Set Up menu (Settings for sensitivity and holdoff)
Menu
Settings
Comments
Sensitivity
<Sensitivity Setting>
Set trigger sensitivity.
Holdoff
<Holdoff Setting>
Set time slot before another
trigger event.
Holdoff
Reset
Reset Holdoff time to 500ns.
Menu
Settings
Comments
Coupling
DC
AC
HF Reject
LF Reject
Allow all signals pass.
Block DC signals.
Reject high frequency signals.
Reject DC and low frequency
signals.
Sensitivity
<Sensitivity Setting>
Set trigger sensitivity and the
range is from 0.10div to
1.00div.
Holdoff
<Holdoff Setting>
Set time slot before another
trigger event and the range is
from 500ns to 1.50s.
Holdoff
Reset
Reset Holdoff time to 500ns.
Page 70
RIGOL
User‟s Guide for DS1000CA Series
2-42
Trigger Holdoff
Use trigger Holdoff to stabilize a complex waveform, such as a pulse range. Holdoff time is the oscilloscope‟s waiting period before starting a new trigger. During Holdoff, oscilloscope will not trigger until Holdoff ends. For instance: To trigger on the first pulse on a group of them, users can set the holdoff time to Pulse cluster width.
Figure 2-54 Trigger Holdoff
Operation Explanations:
1 Press the trigger MENU button to display the Menu. 2 Press Set Up key to display trigger set up menu. 3 Turn the multi function knob ( ) to change Holdoff time until waveform is
stable.
4 Pushing Trigger Hold off reset to reset the Holdoff time to its default value.
Time of Holdoff
Trigger Position
Page 71
RIGOL
User‟s Guide for DS1000CA Series
2-43
Trigger Terms Explanation
1. Trigger Source:
Trigger occurs from several sources: Input channels (CH1 and CH2), AC Line, Ext, Ext/5.
Input Channels:
It is the most commonly used trigger source. The channel works when selected as a trigger source whatever displayed.
Ext Trig:
The instrument can be triggered from a third source while acquiring data from CH1 and CH2. For example, to trigger from an external clock or with a signal from another part of the test circuit. The Ext, Ext/5 trigger sources use an external trigger signal connected to the EXT TRIG connector. Ext uses the signal directly; it has a trigger level range of +1.6 V to -1.6 V. The EXT/5 trigger source attenuates the signal by 5X, which extends the trigger level range to +8 V to -8 V allowing the oscilloscope to trigger on a larger signal.
AC Line:
AC power can be used to display signals related to the power line frequency, such as lighting equipment and power supply devices. The oscilloscope gets triggered on its AC power input, an AC trigger signal is not required.
2. Sweep Mode:
The sweep mode determines how the oscilloscope behaves in the absence of a trigger event. The oscilloscope provides three trigger modes: Auto, Normal, and Single.
Auto:
This sweep mode allows the oscilloscope to acquire waveforms even when it does not detect a trigger condition. If no trigger condition occurs while the oscilloscope is waiting for a specific period (as determined by the time-base setting), it will force itself to trigger.
When forcing invalid triggers, the oscilloscope cannot synchronize the waveform, and the waveform seems to roll across the display. If valid triggers
Page 72
RIGOL
User‟s Guide for DS1000CA Series
2-44
occur, the display becomes stable on the screen. Any factor results in the un-stability of waveforms can be detected by Auto Trigger, such as the output of Power supply. NOTE: When horizontal control is set under 50 ms/div, Auto mode allows the oscilloscope not to capture trigger signal.
Normal:
The Normal mode allows the oscilloscope to acquire a waveform only when it is triggered. If no trigger occurs, the oscilloscope keeps waiting, and the previous waveform, if any, will remain on the display.
Single:
In Single mode, after pressing the RUN/STOP key, the oscilloscope waits for trigger. While the trigger occurs, the oscilloscope acquires one waveform then stop.
3. Coupling:
Trigger coupling determines which signal component passing to the trigger circuit. Coupling types include AC, DC, LF Reject and HF Reject.
AC: AC coupling blocks DC components. DC: DC coupling passes both AC and DC components. LF Reject: LF Reject coupling blocks DC component, and attenuates all signal
with a frequency lower than 8 kHz.
HF Reject: HF Reject coupling attenuates all signals with a frequency higher
than 150 kHz.
4. Pre-trigger/delayed trigger:
The data collected before and after trigger. The trigger position is typically set at the horizontal center of the screen. In the full-screen display the 6div data of pre-trigger and delayed trigger can be surveyed. More data (14div) of pre-trigger and 1s delayed trigger can be surveyed by adjusting the horizontal knob. This feature is very useful to study the events that led up to the trigger point.
Everything to the right of the trigger point is called post-trigger information. The delay range (pre-trigger and post-trigger information) depends on the sweep speed selected.
Page 73
RIGOL
User‟s Guide for DS1000CA Series
2-45
5. Adjustable trigger sensitivity:
To avoid the influence of noise from the physical world, and get the stable trigger, the trigger circuit has adopted Stickiness. In DS1000CA series, the stickiness is adjustable from 0.1div-1.0div, which means when it sets to 1.0div, the trigger circuit will not affect any signal with peak-peak amplitude less than 1.0div, so as to avoid the influence of the noise.
Page 74
RIGOL
User‟s Guide for DS1000CA Series
2-46
To Set Up the Sampling System
Figure 2-55 shows the menu button for acquire system on the front panel.
Figure 2-55 Acquire button
Press Acquire button, the interface menu shows as follows:
Figure 2-56 Table 2-34 The Acquire menu
Menu
Settings
Comments
Acquisition
Normal
Average
Peak Detect
Normal Acquisition mode.
Average Acquisition mode.
Peak Detect Acquisition mode.
Averages
[1]
2 to 256
Step by 2 N times the power. Set
average times from 2 to 256.
Sampling
Real-Time
Equ-Time
Real-time sampling mode.
Equivalent sampling mode.
Sinx/x
[2]
ON
OFF
Set the insert mode is Sinx or
Line.
Sa Rate
[3]
Display sampling rate.
NOTES:
[1] This function appears when acquisition as Average”. [2] This function appears when sampling as Real-Time”. [3] Sample rate of CH1 and CH2 will be displayed respectively in alternative trigger
mode.
Acquire Setup Button
Page 75
RIGOL
User‟s Guide for DS1000CA Series
2-47
The waveform displayed on the screen will change in conjunction with the setting of Acquire menu.
To reduce the displayed random noise, select the Average Acquisition. This
mode would make the screen refresh slower.
To Avoid signal aliasing, select Peak Detect Acquisition. Select Real-time acquisition to observe the single-shot or pulse signals. Select Equ-Time to observe high frequency repetitive signals.
Figure 2-57
Signal that contains noise, and without average sampling
Figure 2-58
Display signal after average sampling
Page 76
RIGOL
User‟s Guide for DS1000CA Series
2-48
Figure 2-59
Adopt Normal acquisition
Figure 2-60
Adopt peak detect acquisition, signal contains envelope
Key points
Stop Acquisition: When the scope is acquiring waveforms, the
waveforms is in a live status; when acquisition is stopped, frozen waveform
will be displayed, the position and scale can still be adjusted by vertical
control and horizontal control.
Page 77
RIGOL
User‟s Guide for DS1000CA Series
2-49
Term Explanations
Real-time Sampling: DS1000CA has Real-time sampling rate up to
2GSa/s. At the time base 20ns or faster, the oscilloscopes use the sine(x)/x
interpolation to expand the horizontal time base.
Equivalent sampling: Known as Repetitive sampling to get up to 20ps of
horizontal resolution (equivalent 50Gsa/s). This mode is good for observing
repetitive signals, and it is not recommended for single-shot or pulse.
Normal Acquisition: Oscilloscope acquires signal by equal time interval.
Average Acquisition: Apply averaging to your signal to remove
uncorrelated noise and improve measurement accuracy. Reduces random
or uncorrelated noise in the signal display. The averaged waveform is a
running average over a specified number of acquisitions from 2 to 256.
Peak Detect: Peak Detect mode captures the maximum and minimum
values of a signal. Finds highest and lowest record points over many
acquisitions.
Page 78
RIGOL
User‟s Guide for DS1000CA Series
2-50
To Set Up the Display System
Figure 2-61 shows the menu button for the display system on the front panel.
Figure 2-61 Display button
Press the Display button, the interface menu for settings shows as follows.
Figure 2-62 Table 2-35 The Display menu (Page 1/3)
Menu
Setting
Comments
Type
Vectors
Dots
Display waveforms as vectors.
Display waveforms as dots.
Clear
Clear all existing waveforms from
screen.
Persist
OFF
Infinite
Turn off the persistence function.
The sample points remain displayed
until turn the persistence “OFF”.
Intensity
<percentage>
Set up waveform intensity
(0%~100%).
Key points
Display type: Display type includes Vector and Dot. In vectors type,
oscilloscope connects dots through digital interpolation including both linearity
and sin(x)/x. Sin(x)/x interpolation is suitable for Real-time sampling and will
be more effective at 50ns or faster time base.
Display Setup Button
Page 79
RIGOL
User‟s Guide for DS1000CA Series
2-51
Figure 2-63 Table 2-36 The Display menu (Page 2/3)
Menu
Settings
Comments
Function
Normal
Intensity
The waveform is displayed at the
same intensity.
The intensity of waveforms is
connecting to probability.
Grid
Display grids and coordinates on the
screen.
Turn off the grids.
Turn off the grids and coordinates.
Menu
Display
1s
2s
5s
10s
20s
Infinite
Set the time before menu fades away.
The menu will be hidden after the set
time of last button pressing.
Figure 2-64 Table 2-37 The Display menu (Page 3/3)
Menu
Settings
Comments
Brightness
<percentage>
Set up grid brightness.
Screen
Normal
Inverted
Set to normal mode.
Set to invert color display mode.
Term Explanations
Refresh rate: It is an important performance of digital oscilloscopes. It means
the number of display refreshing per second and it will affect the ability to
observe signal.
Page 80
RIGOL
User‟s Guide for DS1000CA Series
2-52
Adjusting waveform intensity
Default setup of multi-function knob ( ) is adjusting waveform intensity.
Page 81
RIGOL
User‟s Guide for DS1000CA Series
2-53
To Store and Recall Waveforms or Setups
Figure 2-65 shows the menu button for the storage system on the front panel.
Figure 2-65 Storage button
Press the Storage button, the interface menu for settings shows as follows.
Figure 2-66 Table 2-38 The Storage menu (Waveform or Setups storage)
Menu
Settings
Comments
Storage
[1]
Waveform
[2]
Setups
Bit map
CSV
Factory
Store or recall waveform.
Store or recall instrument setups.
Create or delete bit map files.
Create or delete CSV files.
Recall factory setups.
Internal
Go to menu for internal memory
operation (see Table 2-42).
External
Go to menu for external memory
operation (see Table 2-43).
Disk Mana.
Go to disk manage menu (see
Table 2-44).
NOTES:
[1] The waveform and setups files in internal memory can be saved, loaded and
deleted. While, the waveform, setups, bit map and CSV files in external memory can be created and deleted. To restore the factory settings, please select Factory”.
[2] Both waveforms and status settings could be saved in Waveform Storage”.
Storage Setup Button
Page 82
RIGOL
User‟s Guide for DS1000CA Series
2-54
Figure 2-67 Table 2-39 The Storage menu (Bit Map storage)
Menu
Settings
Comments
Storage
Bit map
Create or delete bit map files.
Para Save
On
Off
Save the current oscilloscope
settings in different format with the
same file name.
External
Go to menu for external memory
operation (see Table 2-43).
Disk Mana.
Go to disk manage menu (see Table
2-44).
Figure 2-68 Table 2-40 The Storage menu (CSV storage)
Menu
Settings
Comments
Storage
CSV
Create or delete CSV files.
Data
Depth
Displayed
Maximum
Save currently displayed waveform
data to CSV file.
Save the whole waveform data in
memory to CSV file.
Para Save
On
Off
Save the current oscilloscope
settings in different format with the
same file name.
External
Go to menu for external memory
operation (see Table 2-43).
Disk Mana.
Go to disk manage menu (see Table
2-44).
Figure 2-69 Table 2-41 The Storage menu (Factory storage)
Menu
Settings
Comments
Storage
Factory
Recall factory setups.
Load
Recall factory setups or files.
Disk Mana.
Go to disk manage menu (see Table
2-44).
Page 83
RIGOL
User‟s Guide for DS1000CA Series
2-55
Internal Memory
Press StorageInternal to go to the following menu.
Figure 2-70 Table 2-42 The Internal Memory menu
Menu
Settings
Comments
Internal
Int_00 (S)
Int_01 (S)
.
.
Int_09 (N)
Set up the location of files in
internal memory among Int_00 to
Int_09. Thereinto, S in brackets
indicates the location has storage
file, N indicates has not.
Load
Recall waveform files and setup
files from the internal memory
location.
Save
Save waveform files and setup
files to the internal memory
location.
Delete File
Delete the files
NOTE: To ensure the setups being saved properly, only after the settings are changed for more than 5 seconds can the user turn off the instrument. The oscilloscope can store 10 settings permanently and can restore at anytime.
Page 84
RIGOL
User‟s Guide for DS1000CA Series
2-56
External Memory
Switch in external memory, press Storage External to go to the following menu.
Figure 2-71 File system
Figure 2-72 Table 2-43 The External Memory menu
Menu
Settings
Comments
Explorer
Path
Directory
File
Switch among Path, Directory and File.
New File
(Folder)
To create new file or folder.
Delete
File
Delete file (Folder).
Load
Recall waveform and setup from USB
storage device.
Current Directory
Current Path
Files
Back to Previous
Directory
File Information
Page 85
RIGOL
User‟s Guide for DS1000CA Series
2-57
Disk Management
Press StorageDisk Mana. to go to the following menu.
Figure 2-73 Table 2-44 The Disk Management menu (Page 1/2)
Menu
Settings
Comments
Explorer
Path
Directory
File
Switch among Path, Directory and
File.
New folder
To create new folder (same as new
files).
Delete File
Delete file.
Load
Recall waveform, setup, recorded
waveform, Pass/Fail file.
Figure 2-74 Table 2-45 The Disk Management menu (Page 2/2)
Menu
Settings
Comments
Rename
To rename a file (see Table 2-46).
Disk info
Display disk information.
Page 86
RIGOL
User‟s Guide for DS1000CA Series
2-58
1. Disk Information
Press Storage Disk Mana Diskinfo to go to the disk information interface.
Figure 2-75 Disk information interface
2. Rename
Press StorageDisk Mana.Rename to go to the following interface.
Figure 2-76
Rename the file
Page 87
RIGOL
User‟s Guide for DS1000CA Series
2-59
Figure 2-77 Table 2-46 The Rename menu
Menu
Settings
Comments
To move the cursor up.
To move the cursor down.
To delete chosen letter.
OK
Rename the file.
Key point
Factory: The oscilloscope has default settings and can be recalled at
anytime by user.
Memory location: Specify the memory location to save/recall the
waveforms and setups.
Load: Recall saved waveforms, setups and default settings.
Save: Save waveforms and setups.
Page 88
RIGOL
User‟s Guide for DS1000CA Series
2-60
To Set Up the Utility
Figure 2-78 shows the menu button for the Utility on the front panel.
Figure 2-78 Utility button
Press the Utility button to show the menu of the settings in the Utility system.
Figure 2-79 Table 2-47 The Utility menu (Page 1/3)
Menu
Setting
Comments
I/O setting
Setup I/O configuration.
Sound
(ON)
(OFF)
Turn beeper sound on/off.
Counter
OFF
ON
Turn off Frequency Counter.
Turn on Frequency Counter.
Language
简体中文
繁体中文
한국어
日本語
English
Deutsch
Français
Português
Русский
Espaňol
Select languages.
System Function Setup Button
Page 89
RIGOL
User‟s Guide for DS1000CA Series
2-61
Figure 2-80 Table 2-48 The Utility menu (Page 2/3)
Menu
Settings
Comments
Pass/Fail
Setup Pass/Fail test.
Record
Setup Waveform Recorder.
Print set
Setup printing.
Figure 2-81 Table 2-49 The Utility menu (3/3)
Menu
Settings
Comments
Self-Cal
Execute Self-calibration.
Service
Go to service menu.
Preference
Go to preference menu.
Term Explanations:
Self-Cal: Oscilloscope will calibrate parameter of vertical system (CH1,
CH2, and Ext), horizontal system and trigger system.
Page 90
RIGOL
User‟s Guide for DS1000CA Series
2-62
I/O Setup
Press Utility I/O setting to go to the following menu.
Figure 2-82 Table 2-50 The I/O Setup menu
Menu
Display
Comments
RS-232
Baud
300
.
.
38400
Set RS-232 baud rate as 300, 2400,
4800, 9600, 19200 or 38400.
GPIB#
0 to 30
Set the GPIB address.
Page 91
RIGOL
User‟s Guide for DS1000CA Series
2-63
Language
The DS1000CA series oscilloscopes have multi-language user menu, choose as your desire.
Press UtilityLanguage to select the language.
Figure 2-83 Language interface
Page 92
RIGOL
User‟s Guide for DS1000CA Series
2-64
Pass/Fail
The Pass/Fail function monitors changes of signals and output pass or fail signals by comparing the input signal is within the pre-defined mask. The testing results not only can be displayed on screen or output by isolated pass/fail port but also can be alarmed if turn on system sound.
Press UtilityPass/Fail to go to the following menu.
Figure 2-84 Table 2-51 The Pass/Fail menu (Page 1/2)
Menu
Setting
Comments
Enable Test
ON
OFF
Turn on Pass/Fail test.
Turn off Pass/Fail test.
Source
CH1
CH2
Select Pass/Fail test on CH1.
Select Pass/Fail test on CH2.
Operate
(RUN)
(STOP)
Pass/Fail test stopped, press to run.
Pass/Fail test running, press to stop.
Msg display
ON
OFF
Turn on Pass/Fail information display.
Turn off Pass/Fail information display.
Figure 2-85 Table 2-52 The Pass/Fail menu (Page 2/2)
Menu
Settings
Comments
Output
Fail
Fail +
[1]
Pass
Pass +
[1]
Output when Fail condition detected.
Output and beep when Fail condition
detected.
Output when Pass condition detected.
Output and beep when Pass condition
detected.
Stop
on Output
ON
OFF
Stop test when output occur.
Continue test when output occur.
Mask
Go to mask setting menu.
Note: [1] Only available when sound is enabled.
Page 93
RIGOL
User‟s Guide for DS1000CA Series
2-65
1. Mask Setting
Press UtilityPass/FailMask Setting to go to the following menu.
Figure 2-86 Table 2-53 The Mask setting menu (Page 1/2)
Menu
Settings
Comments
X Mask
<x div>
Set horizontal clearance to the
waveform (0.04div-4.00div).
Y Mask
<y div>
Set vertical clearance to the waveform
(0.04div-4.00div).
Create
Mask
Create a test mask according to the
above clearance.
Location
Internal
External
Set the memory location of the mask
files.
Figure 2-87 Table 2-54 The Mask setting menu (Page 2/2)
When the save as internal memory
Menu
Settings
Comments
Save
Store created test mask into internal
memory.
Load
Recall mask setting file from internal
memory.
Imp./Exp.
Go to import/export menu (same as
REF import/export menu.).
Page 94
RIGOL
User‟s Guide for DS1000CA Series
2-66
Figure 2-88 Table 2-55 The Mask setting menu (Page 2/2)
When the save as external memory
Menu
Settings
Comments
Save
Go to save menu (same as REF save
menu).
Load
Go to load menu (see Table 2-56).
Import
Go to import menu. (The same as REF
import menu).
2. Mask Setting
Press UtilityPass/FailMask SettingLoad to go to the following menu.
Figure 2-89 Table 2-56 The Load menu
Menu
Settings
Comments
Explorer
Path
Directory
File
Switch among Path, Directory and
File.
Load
Recall the specified file.
NOTE: Pass/Fail function is unavailable in X-Y mode.
Page 95
RIGOL
User‟s Guide for DS1000CA Series
2-67
3. Pass/Fail connection
The Pass/Fail connection has an optically isolated output. It needs to connect to another circuit to fulfill the function.
Before connecting to an external circuit, make sure the maximum voltage/current does not exceed the internal photo MOS relay‟s rating, 400V/100mA. DS1000CA series adopts the technique of Optical Isolation, and the output device has no polarity and can be used in any circuit within the ratings.
Figure 2-90
Schematic Diagram of Pass/Fail output
Page 96
RIGOL
User‟s Guide for DS1000CA Series
2-68
Waveform Recorder
Waveform recorder records input waveforms from CH1 and CH2, with a maximum record length of 1000 frames. This performance can also be activated by the Pass/Fail test output, which makes this function especially useful to capture abnormal signals in long term without keeping an eye watching it. This function contains four modes: OFF, Record, Play back and Storage. More details about the function please refer to the following introduction.
1. OFF: It will close all record functions.
2. Record: To record waveform by selected time interval till the last frame.
Press UtilityRecordModeRecord to go to the following menu
Figure 2-91 Table 2-57 The Record menu (Page 1/2)
Menu
Settings
Comments
Mode
OFF
Record
Play back
Storage
Turn off all recorder functions.
Select record mode.
Select play back mode.
Select storage mode.
Source
CH1
CH2
P/F-OUT
Select record source channel.
End Frame
<1-1000>
Set number of record frames.
Operate
(Run)
(Stop)
Record stopped, press to Start
recording.
Press to stop recording.
Page 97
RIGOL
User‟s Guide for DS1000CA Series
2-69
Figure 2-92 Table 2-58 The Record menu (Page 2/2)
Menu
Settings
Comments
Interval
<1.00ms-1000s>
Set time interval between
record frames.
3. Play back: Play back the recorded waveforms.
Figure 2-93 Table 2-59 The Play Back menu (Page 1/2)
Menu
Settings
Comments
Operate
(Run)
(Stop)
Play stopped, press to Start
playback.
Press to stop playing.
Play mode
Set repeat play mode.
Set single time play mode.
Interval
<1.00ms-20s>
Set up interval.
Figure 2-94 Table 2-60 The Play Back menu (Page 2/2)
Menu
Settings
Comments
Start Frame
<1-1000>
Set start frame.
Current
Frame
<1-1000>
Select current frame to be played.
End Frame
<1-1000>
Set End frame.
Page 98
RIGOL
User‟s Guide for DS1000CA Series
2-70
Key Points
The RUN/STOP button can also replay or continue the waveform display.
4. Storage: Store recorded waveforms to non-volatile memory according to the
setup frames.
Figure 2-95 Table 2-61 The Storage menu (Page 1/2)
Menu
Settings
Comments
Start Frame
<1-1000>
Set first frame to be saved.
End Frame
<1-1000>
Set last frame to be saved.
Location
Internal
external
Set up Store location.
Figure 2-96 Table 2-62 The Storage menu when the save to Internal memory
(Page 2/2)
Menu
Settings
Comments
Save
Save recorded waveform to internal
memory location.
Load
Recall recorded waveform from
internal memory location.
Imp./Exp.
Go to import/export menu (Same as
REF import/export menu).
Page 99
RIGOL
User‟s Guide for DS1000CA Series
2-71
Figure 2-97 Table 2-63 The Storage menu when the save as external memory
(Page 2/2)
Menu
Settings
Comments
Save
Go to save menu (same as REF save
menu).
Load
Go to load menu (see Table 2-56).
Import
Go to import menu. (Same as REF
import menu).
Page 100
RIGOL
User‟s Guide for DS1000CA Series
2-72
Print Set
DS1000CA series oscilloscopes support external printers.
Press Utility Print set to go to the following menu.
Figure 2-98 Table 2-64 The Print Set menu
Menu
Settings
Comments
Print
Execute print function.
Inverted
ON
OFF
Invert the color for print.
Print original color.
Palette
Gray scale
Color
Set up print color.
Loading...