Rigol DS1104Z-S Plus, DS1000Z Series, DS1074Z-S Plus, DS1054, DS1104Z Plus Quick Manual

...
Page 1
RIGOL
Quick Guide
DS1000Z Series Digital
Oscilloscope
Apr. 2019
Page 2
Page 3
RIGOL
I

Guaranty and Declaration

Copyright
© 2013 RIGOL (SUZHOU) TECHNOLOGIES INC. All Rights Reserved .
Trademark Information
RIGOL is a registered trademark of RIGOL (SUZHOU) TECHNOLOGIES INC.
Publication Number
QGA19108-1110
Notices
RIGOL produ cts are cove red by P.R.C. and f oreign pa tents, issue d and pendin g. RIGOL reserves the right to modify or change parts of or all the specifications
and pricing policies at the company’s s ole de cision .
Information in this publication replaces all previously released materials. Information in this publication is subject to change without notice. RIGOL shall not be liable for either incidental or consequential losses in
connection with the furnishing, use, or performance of this manual, as well as any information contained.
Any part of this document is f orbi dden to be c opie d, ph otoc opie d, o r rearranged
without prior written approval of RIGOL.
Product Certification
RIGOL guarantees that this product conforms to the national and industrial standards in China as well as the ISO9001:2015 standard and the ISO14001:2015 standard. Other international standard conformance certifications are in progress.
Contact Us
If you have any problem or requirement when using our products or this manual, please contact RIGOL. E-mail: service@rigol.com Website: www.rigol.com
DS1000Z Quick Guide
Page 4
RIGOL
1. Only the exclusive power cord
8. Use the proper fuse.
WARNING
avoided, will result in serious injury or death.
CAUTION
avoided, could result in damage to the product or lo ss of i mportant data.
DANGER
It calls attention to an operation, if not correctly performed, could result in injury or hazard immediately.
WARNING
It calls attention to an operation, if not correctly performed, could result in potential injury or hazard.
CAUTION
It calls attention to an operation , if not correctly performed, coul d product.
Hazardous
Safety
Protective Terminal
Chassis
Test

Safety Requirement

General Safety Summary

designed for the instrument and authorized for use within the local country could be used.
2. Ensure that the instrument is safely grounded.
3. Connect the probe correctly.
4. Observe all terminal ratings.
5. Use proper o vervolt age prot ection.
6. Do not operate without covers.
7. Do not insert objects into the air outlet.
9. Avoid circuit or wire exposure.
10. Do not operate the instrument with suspected failures.
11. Provide adequate ventilation.
12. Do not operate in wet conditions.
13. Do not operate in an explos ive atmosphere.
14. Keep instrument surfaces clean and dry.
15. Prevent electrostatic impact.
16. Handle with caution.

Safety Notices and Symbols

Safety Notic e s in this Manua l:
Indicates a potentially hazardous situation or practice which, if not
Indicates a potentially hazardous situation or practice which, if not
Safety Terms on the Product:
result in damage to the product or other devices connected to the
Safety Symbols on the Product:
Voltage
II DS1000Z Quick Guide
Warning
Earth
Ground
Ground
Page 5
RIGOL
III
WARNING
measurement categories.
WARNING
during operation and inspect the intake and fan regularly.

Measurement Category

Measurement Category
DS1000Z series digital oscilloscopes can make measurements in Measurement Category I.
This oscilloscope can only be used for measurements within its specified
Measurement Category Definitions
Measurement category I is for measurements performed on circuits not directly connected to MAINS. Examples are measureme nts on circuits n ot de ri ved from MAINS, and specially protected (internal) MAINS derived circuits. In the latter case, transient stresses are variable; for that reason, the transient withstand capability of the equipment is made known to the user.
Measurement category II is for measurements performed on circuits directly connected to the low volta ge install ation. Exa mples are measure ments on househol d appliances, portable tools and similar equipment.
Measurement category III is fo r measure ments perf ormed in the b uilding inst allation. Examples are measurements on distribution boards, circuit-breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the f ixed installation, and equipment for industrial use and some other equipment, for example. Stationary motors with permanent connection to the fixed installation.
Measuremen t c ategory IV is for measurements performed at the source of the low-voltage installation. Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.

Ventilation Requirement

This oscilloscope uses fan to force cooling. Please make sure that the air intake and exhaust areas are free from obstructions and have free air. When using the oscilloscope in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation.
Inadequate ventilation may cause temperature increase which would damage the instrument. So please keep the instrument well ventilated
DS1000Z Quick Guide
Page 6
RIGOL
WARNING
not operate in humid environment.
WARNING
reach the p roduct, or else the ope rat or mi ght expo se to danger of elect ric shock.

Working Environment

Temperature
Operating: 0 to +50 Non-operating: -40℃ to +70
Humidity
0 to + 30: ≤95% relative humidity +30 to +40: ≤75% relative humidity +40 to +50: ≤45% relative humility
To avoid short circuit inside the instrument or electric shock, please do
Altitude
Operating: below 3 km Non-operating: below 15 km
Installation (Overvoltage) Category
This product is powered by mains conforming to installation (overvoltage) category II.
Make sure that no over volta ge (such as that caused by thunderbolt) can
Installation (Overvoltage) Cate g o ry Definitions
Installation (overvoltage) category I refers to signal level which is applicable to equipment measurement terminals connected to the source circuit. In these terminals, precautions are done to limit the transient voltage to the corresponding low level.
Installation (overvoltage) category II refers to the local power distribution level which is applicable to equipment connected to the AC line (AC power) .
Pollution Degree
Degree 2
Pollution Degree Definitions
Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. For example: a clean room or air-conditioned office environment.
IV DS1000Z Quick Guide
Page 7
RIGOL
V
CAUTION
WARNING
supply.
Pollution degree 2: Normally only dry, non-conductive pollution occurs. Occasionally a temporary conductivity caused by condensation may occur. For example: general indoor environment.
Pollution degree 3: Conductive pollution occurs, or dry, non-con du ctiv e pollutio n occurs which becomes conductive due to condensation which is expected. For example: Sheltered outdoor environment.
Pollution degree 4: Pollution that generates persistent conductivity through conductive dust, rain, or snow. For example: outdoor locations.
Safety Class
Class 1 – Grounded Product

Care and Cleaning

Care
Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time.
Cleaning
Clean the instrument regularly according to its operating conditions.
1. Disconnect the instrument from all power sources.
2. Clean the external surfaces of the instrument with a soft cloth dampened with mild detergent or water. Avoid having any water or other objects into the chassis via the heat dissi pation hole . When cleaning the LCD , take care to avoid scarifying it.
To avoid damage to the instrument, do not expose it to caustic liquids.
To avoid short-circuit resulting from moisture o r pe rsonal injuries, ensure that the instrument is completely dry before connecting it to the power
DS1000Z Quick Guide
Page 8
RIGOL

Environmental Consideratio ns

The following symbol indicates that this product complies with the WEEE Directive 2002/96/EC.
Product End-of-Life Handling
The equipment may contain substances that could be harmful to the envi ronm ent or human health. To avoid the release of such substances into the environment and avoid harm to human health, we recommend you to rec ycle this product appropriately to ensure that most materials are reused or recycled properly. Please contact your local authorities for disposal or recycling information.

Document Overview

This manual is used to guide users to quickly get familiar with the front panel, rear panel, user interface and basic operation method of DS1000Z series digital oscilloscope. You can download the newest version of the manual from RIGOL official website (www.rigol.com).
Format Conventions in this Manual:
1. Key
The fr ont panel ke ys are den oted by the f ormat of "Key Name (Bold) + T ext Box". For ex ample, Utility denotes the "Utility" key.
2. Menu
The menu softkeys are denoted by the format of "Menu Word (Bold) + Character Shading". For example , System denotes the "System" menu under
Utility.
3. Operation Step
The next step of ope ration is denote d b y an a rrow "". For example, Utility System denotes that first press Utility on the front panel and then press System.
VI DS1000Z Quick Guide
Page 9
RIGOL
VII
HORIZONTAL SCALE
VERTICAL SCALE
Number of
Channels
Number of
Channels
Number of
Channels
DS1104Z-S Plus
100 MHz
4 2 16
[1]
DS1074Z-S Plus
70 MHz
4 2 16
[1]
DS1104Z Plus
100 MHz
4
--
16
[1]
DS1074Z Plus
70 MHz
4
--
16
[1]
DS1054Z
50 MHz
4
--
--
4. Knob
Label Knob
Horizontal Scale Knob
HORIZONTAL POSITION
Horizontal Position Knob Vertical Sca le Knob
VERTICAL POSITION TRIGGER LEVEL
Vertical Positi on Knob Trigger Level Knob
Content Con v entions in th is Man ual:
DS1000Z series includes the following models. Unless otherwise noted, this manual takes DS1104Z-S Plus as an example to illustrate the front panel, rear panel, user interface and basic operatio n m e t h ods of DS1000Z series.
Model
[1]
Note
RPL1116 and the model label).
: Required to upgrade through the MSO upgrade option (includes the logic analyzer probe
Analog Bandwidth
Analog
Source
Digital
DS1000Z Quick Guide
Page 10
RIGOL
Contents
Guaranty and Declaration ......................................................................... I
Safety Requirement ................................................................................. II
General Safety Summary........................................................................... II
Safety Not ices and Symbol s ...................................................................... II
Measurement Category ............................................................................ III
Ventilation Requirement ........................................................................... III
Working Enviro nment .............................................................................. IV
Care and Cleaning .................................................................................... V
Environmental Considerations ................................................................... VI
Document Overview ............................................................................... VI
Quick Start ............................................................................................... 1
General Inspection .................................................................................... 1
Appearance and Dimensions ...................................................................... 2
To Prepare for Operation ............................................................................ 3
To Adjust the Supporting Legs ............................................................. 3
To Connect to AC Power Supply ........................................................... 3
Turn-on Checkout ............................................................................... 4
To Connect the Probe .......................................................................... 4
Function Inspection ............................................................................ 6
Probe Compensation ........................................................................... 7
Panel Overview ................................................................................ 8
Front
Rear Panel Overview ................................................................................. 9
Front Panel Function Overview ................................................................. 11
VERTICAL ........................................................................................ 11
Logic Analyzer .................................................................................. 12
Signal Source ................................................................................... 12
HORIZONTAL ................................................................................... 13
TRIGGER ......................................................................................... 13
CLEAR ............................................................................................. 14
AUTO .............................................................................................. 14
RUN/STOP ....................................................................................... 14
SINGLE ............................................................................................ 14
Multifunction Knob ............................................................................ 15
Function Menus ................................................................................ 15
Print ................................................................................................ 16
User Interface ......................................................................................... 17
Parameter Setting Method ........................................................................ 21
To Use the Security Lock .......................................................................... 21
To Use the Built-in Help System ................................................................ 22
Troubleshooting ..................................................................................... 23
VIII DS1000Z Quick Guide
Page 11
RIGOL
1

Quick Start

General Inspection

1. Inspect the packaging
If the packagi ng has been da m aged, do not dispose t h e da maged packaging or cushioning materials u ntil the shipment has been checked for completeness and has passed both electrical and mechanical tests.
The consigner or carrier shall be liable for the damage to the instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacem ent of the instrument.
2. Inspect the instrument
In case of any mechanical damage, missing parts, or failure in passing the electrical and mechanical tests, contact your RIGOL sales representative.
3. Check the accessories
Please check the accessories according to the packi ng l is ts . If t he accessories are damaged or incomplete, please contact your RIGOL sales representative.
DS1000Z Quick Guide
Page 12
RIGOL

Appearance and Dimensions

Figure 1 Front View Unit: mm
Figure 2 Top View Unit: mm
2 DS1000Z Quick Guide
Page 13
RIGOL
3
Power Cord Connector

To Prepare for Operation

To Adjust the Supporting Legs
Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement of the oscilloscope as well as better operation and observation.
Figure 3 To Adjust the Supporting Legs
To Connect to AC Power Supply
This oscilloscope can accept 100-240 V, 45-440 Hz AC power supply. Please use the power cord supplied with the accessories to connect the oscilloscope to the power source as shown in the figure below.
DS1000Z Quick Guide
Figure 4 To Connect to AC Power Supply
Page 14
RIGOL
Turn-on Checkout
When the oscilloscope is connected to power, press the Power key lower-left corner of the fr ont panel to start the oscilloscope. During the start-up process, the oscilloscope performs a series of self-test items. After the self-test, the welcome screen is displayed. The instrument is insta lled with the trial v ersions of the options before leaving factory and the remaining time is about 2,000 minutes. The "Installed Options" dialog box will be displayed if your instrument has currently installed the trial versions of options. From this dialog box you can view the name, detail, version and remaining trial time of the option currently installed.
at the
To Connect the Probe
RIGOL provides passive probe and logic probe for DS1000Z series. For the model of the probes, please refer to information of the probes, please refer to corresponding Probe User’s Gui de .
Connect the Probe:
1. Connect the BNC terminal of the probe to an analog channel input of the oscilloscope on the front panel.
2. Connect the ground alligator clip of the probe to the circuit ground terminal and then connect the probe tip to the circuit point under test.
DS1000Z Series Datasheet
. For detailed technical
Figure 5 To Connect the Passive Probe
After you connect the passive probe, check the probe function and probe compensation adjustment before making measurements. For detailed procedures, refer to "Function Inspection" and "Probe Compensation" sections in this manual.
4 DS1000Z Quick Guide
Page 15
RIGOL
5
Digital Channel Input Terminal
Connect the Logic Probe:
1. Connect the single-wire terminal of the logic probe to the digital channel input terminal on the front panel of the oscilloscope in the correct direction.
2. Connect the other terminal of the logic probe to the signal under test. RIGOL provides DS1 000Z Plus with the RPL1116 logic probe option. To realize convenient and flexible detection of signals, two connection methods for the signal under test are a vailable. For the details, please refer to the
Probe User Guide
.
RPL1116 Logic
Figure 6 To Connect the Logic Probe
Note: The digital channel input terminal does not support hot plugging. Please do
not insert or pull out the logic probe when the instrument is in power-on state.
DS1000Z Quick Guide
Page 16
RIGOL
WARNING
electric shock during the use of probe, please make sure that t he
part of the probe when the probe is connected to high voltage source.
Tip
probe compensation adjustment and cannot be used fo r cali bration.
Compensation Signal Output Terminal
Function Inspection
1. Press Storage Default to restore the oscilloscope to its default configuration.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" as shown in the figure below.
3. Use the probe to connect the input terminal of CH1 and the "Compensation Signal Output Terminal" of the oscilloscope.
Ground Terminal
Figure 7 To Use the Compensation Signal
4. Set the attenuation on the probe to 10X. Then press AUTO.
5. Observe the wavef orm on the dis play. In normal condition, the display should be a square waveform as shown in the figure below:
Figure 8 Square Waveform
6. Use the same method to test the other channels. If the square waveforms actually shown do not match that in the figure above, please perform "Probe Compensation".
To avoid insulated wire of the probe is in good c ondition. Do not touch the met allic
The signal output from the probe compensation connector can only be used for
6 DS1000Z Quick Guide
Page 17
RIGOL
7
Over compensated Perfectly compensated Under com pen sated
Probe Compensation
When the probes are used for the first time, you should compensate the probes to match the input channels of the oscilloscope. Non-compe ns a ted or poorly compensate d probes may cause measurement inaccuracy or errors. The probe compensation procedures are as follows:
1. Perform Step 1, 2, 3 and 4 specified in "Function Inspection".
2. Check the displayed waveforms and compare them with the waveforms in the following figure.
Figure 9 Probe Compensation
3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment hole on the probe until the displayed waveform is displayed as "Perfectly compensated" in the figure above.
DS1000Z Quick Guide
Page 18
RIGOL
No.
Description
No.
Description
1
Measurement Menu Sof tkeys
11
Power Key
2
LCD
12
USB Host Interface
3
Function Menu Softkeys
13
Digital Channel Input Interface
[1]
4
Multifunction Knob
14
Analog Channel Input Interface
5
Common Operation Keys
15
Logic Analysis Control Key
[1]
6
CLEAR
16
Signal Source
[2]
7
AUTO
17
VERTICAL Control
8
RUN/STOP
18
HORIZONTAL Control
9
SINGLE
19
TRIGGER Control
Probe Compensation Signal Output Terminal/Ground Terminal
3 4 5 6 7 8 9 10
1 2
11 12 13 14 15 16 17 18 19 20

Front Panel Overview

Figure 10 Front Panel Ove rview
Table 1 Front Panel Description
10 Help/Print 20
[1]
Note Note
: Only applicable to DS1000Z Plus.
[2]
: Only applicable to digital oscilloscopes with source channels.
8 DS1000Z Quick Guide
Page 19
RIGOL
9
1 2 3
4 5 6 7 8

Rear Panel Overview

Figure 11 Rear Panel Overview
1 Handle
Pull up the handle vertically for easy carrying of the instrument. When you do not need the handle, press it down.
2 LAN
Connect the instrument to the network via this interface fo r remote control . This oscilloscope conforms to the LXI CORE 2011 DEVICE class instrument standards and can quickly build test system with other instruments.
3 USB DEVICE
You can con nect the oscill oscope to a PictBridge printer or PC via this interface. When a PC is connected, users can send SCPI commands using the PC software or control the oscilloscope via user-defined programming. When a printer is connected, users can print the waveforms displayed on the screen using the printer.
4 Trigger Out and Pass/Fail
Trigger Out:
The oscilloscope can output a signal that can reflect the current capture rate of the oscilloscope at each trigger via this interf ace. Connect the signal to a waveform d isplay device to measure the frequency of the signal. The measurement result is equal to the current capture rate.
DS1000Z Quick Guide
Page 20
RIGOL
Pass/Fail:
The instrument will output a negative puls e via this connector whe n a failed waveform is detecte d during the pass/fail test. The instrument continuously outputs a low level via this conne ctor when no f aile d w a ve form is detected .
5 Source Output
The output terminals of the built-in dual-channel source of the oscilloscope. When Source1 or Source2 is enabled, the signal currently set can be output through the [Source1] or [Source2] connector on the rear panel.
6 Lock Hole
You ca n lock the instrume nt to a fixed location by using the s ecurity lock (please purchase it yourself) via the lock hole.
7 Fuse
If a new fuse is required, please use the specified fuse (250V, T2A). The operation method is as follows: a) Turn off the instrument, disconnect the power, and then remove the power
cord.
b) Insert a small straight screwdriver into the slot at the power cord connector
and pry out the fuse holder gently.
c) Take out the f use an d replace it with a spe cified fuse. Then, install the fuse
holder.
8 AC Power Cord Connector
AC power input terminal. The power requirements of this oscill oscope are 100-240 V, 45-440 Hz. Use the power cord provided with the accessories to connect the instrument to AC power. Then, you can p ress t he Power key on the front panel to start the instrument.
10 DS1000Z Quick Guide
Page 21
RIGOL
11
CH1, CH2, CH3, CH4: analog channel setting
the decoding options.
Tip

Front Panel Function Overview

VERTICAL
REF: press this key to enable the reference waveform function to compare the
waveform actually measured with the reference waveform.
Vertical
waveform. Turn clockwise to increase the position and turn counterclockwise to decrease. During the modification, the waveform would move up and down a nd the position message (e.g. change accordingly. Press down this knob to quickly reset the vertical position to zero.
VERTICAL
clockwis e to decrease the scale and turn counterclockwise to increase. During the modification, the amplitude of the waveform would enlarge or reduce and the scale information (e.g. accordingly. Press down this knob to quickly switch the vertical scale adjustment modes between "Coarse" and "Fine".
How to set the vertical scale and vertical position of each channel? The 4 channels of DS1000Z use the same VERTICAL VERTICAL
position of a certain channel, please press CH1, CH2, CH3 or CH4 at firs t to select the desired channel. Then rotate the VERTICAL POSITION and
VERTICAL
POSITION: modify the vertical position of the current channel
SCALE: modify the vertical scale of the current channel. Turn
) at the lower side of the screen would change
SCALE knobs. If you want to set the vertical scale and vertical
SCALE knobs to set the values.
keys. The 4 channels are marked by different colors which are also used to mark both the corresponding waveforms on the screen and the channel input connectors. Press any key to open the corresponding channel menu and press again to turn off the channel.
MATH: press MATH Math, you can open the math operation menu under which A+B, A-B, A× B, A/B, FFT, A&&B, A||B, A^B, !A, Intg, Diff, Sqrt, Lg, Ln, Exp and Abs are provided. You can also press MATH to open the decoding menu and set
) at the lower-left corner of the screen would
POSITION and
DS1000Z Quick Guide
Page 22
RIGOL
Press this key to open the logic analyzer control menu. Y ou ca n
can also set a label for each digital channel.
Press this key to enter the source setting interface. You can
signal.
Logic Analyzer
Note:
This function is only applicable to DS1000Z Plus with the MSO upgrade option. Press LA D7-D0; when "On" is selected, CH4 function is automatically
disabled; when "Off" is selected, CH4 function recovers automatically. Press LA D15-D8; when "On" is selected, CH3 function is automatically disabled; when "Off" is selected, CH3 function recovers automatically.
turn on or off any channel or channel group, modify the display size of the digital channel, modify the logic threshold of the digital channel as well as group the 16 digital channels. You
Signal Source
Note: This function is only applicable to digital oscilloscopes with source channels.
enable or disable the output of the [Source1] or [Source2] connector on the rear panel, set the output signal waveform and parameters, turn on or off the state display of the current
12 DS1000Z Quick Guide
Page 23
RIGOL
13
HORIZONTAL POSITION: modify th e horizontal
delayed sweep state.
MODE: press this key to switch the trigger mode to Auto,
value in the trigger level message
FORCE: pres s t his key to generate a trigger sign al forcibly.
HORIZONTAL
position. The trigger point would move left or right relative to the center of the screen when you rotate the knob . During the modification, waveforms of all the cha nnels would m ove left or right and the horizontal position message (e.g.
) at the upper-right corner of the screen would change accordingly. Press down this knob to quickly reset the horizontal position (or the delayed sweep position).
MENU: press this key to open the horizontal contro l men u where you can turn on or off the delayed sweep function and switch between different time base modes.
HORIZONTAL SCALE: modify the horizontal time base. Turn clockwise to reduce the time base and turn counte r clo ckwise to increase the time base. During the modif ication, waveforms of all the channels will be displayed in expanded or compressed mode and the time base message (e.g. the screen would change accordingly. Press down this knob to quickly switch to
) at the upper side of
TRIGGER
Normal or Single and the corresponding state backlight of
the current trigger mode would be illuminated. TRIGGER
to increase the level and turn counterclockwise to reduce the level. During the modification, the trigger level line would move up and down and the box (e.g. ) at the lower-left corner of the screen would change accordingly. Press down the knob to quickly reset the trigger level to zero point.
MENU: press this key to open the trigger operation menu. This oscilloscope pro vides various trigger types; for the details, please refer to the introduction in
LEVEL: modify the trigger level. Turn clockwise
DS1000Z User Guide.
DS1000Z Quick Guide
Page 24
RIGOL
Press this key to clear all the waveforms on the screen. If the displayed.
Press this key to enable the waveform auto setting function. Press this key to "RUN" or "STOP" wave f orm sampling. In the
state, the key is illuminated in red.
Press this key to set the trigger mode to "Single". In single
immediately.
CLEAR
oscilloscope is in "RUN" state, new waveforms will still be
AUTO
Note: W av eform auto setting f unction re quires that t he frequency of sine is no lowe r
than 41 Hz; the duty cycle should be greater than 1% and t he amplitude must be at least 20 mVpp for square. Otherwise, the Waveform auto setting function may be invalid and the quick parameter measurement function displayed in the menu will also be unavailable.
The oscilloscope will automatically adjust the vertical scale, horizontal time base and trigger mode according to the input signal to realize optimum wav eform display.
RUN/STOP
"RUN" state, the key is illuminated in yellow. In the "STOP"
SINGLE
14 DS1000Z Quick Guide
trigger mode, press FORCE to generate a trigger signal
Page 25
RIGOL
15
Adjust waveform brightness:
Multifunctional:
filename.
Multifunction Knob
In non-menu-operation mode, turn this knob to adjust the brightness of waveform display. The adjustable range is from 0% to 100%. Turn clockwise to increase the brightness and counterclockwise to reduce. Press down this k n ob to reset the brightness to 60%. You can also press Display Intensity and use the knob to adjust the waveform brightness.
In menu operation, the backlight of the knob goes on. Press a ny menu softkey a nd rotate the knob to select the submenus under this menu and then press down the knob to select the current submenu. It can also be used to modify parameters (please refer to the introduction in "Parameter Set t in g Method
") and input
Function Menus
Measure: press this key to open the measurement setting menu. You can set the
measurement source, turn on or off the frequency counter, all measure, statistic function and etc. Press MENU at the left of the screen to switch the measurement menus of 37 waveform parameters. Then, press down the corresponding menu softkey to quickly realize one-key measurement and the measurement result will be displayed at the bottom of the screen.
Acquire: press this key to enter the sample setting menu to set the acquisition mode, Sin(x)/x and mem ory depth of the oscilloscope.
Storage: press this key to enter file store and recall interface. The storable file types include picture, traces, waveforms, setups, CSV and parameters. Internal and external storage as well as disk management are also supported.
Cursor: press this key to enter the cursor measurement menu. The oscilloscope provides four cu rsor modes: manu al, track, a uto and XY. Note that XY cursor m ode is only available when the horizontal time base is set to XY.
DS1000Z Quick Guide
Page 26
RIGOL
Press this key to print the screen or save the screen to a USB
pressing this key.
Display: press this key to enter display setting menu to set the display type, persistence time, wave intensity, grid type and grid brightness.
Utility: press th i s key to enter the system function setting menu to set the system-related functions or parameters, such as the I/O, sound and language. Besides, some advanced functions (such as the pass/fail test, waveform record, etc.) are also supported.
Print
storage device. If a PictBridge printer is connected currently and the
printer is in idle state, pressing this key will execute the print operation.
If no printer is connected but a USB storage device is
inserted, pressing this key can save the screen to the USB storage device in the specified format.
If both a prin t er and a USB st o rage device are connected
at the same time, the printer enjoys higher priority when
16 DS1000Z Quick Guide
Page 27
RIGOL
17
12 13 14 15 16 17 18 19 20 21 22
1 2 3 4 5 6 7 8 9 10 11

User Interface

DS1000Z provides 7.0 inch WVGA (800*480) TFT LCD.
1. Auto Measurement Items
Provide 20 horizontal (HORIZO NTAL) and 17 vertical (VERTICAL) measur ement parameters. Press the softkey at the left of the screen to activate the corresponding measurement item. Press MENU continuously t o switch between the horizontal and vertical parameters.
2. Digital Chan n el Label/Waveform The logic high level of the digital wa vefor m is displa yed i n blue and the logic lo w level in green. Its edge is displayed in white. The label and waveform of the digital channel currently selected are display ed in re d. The gr oup set functi on in logic analyzer menu can divide the digital channels into four groups, channel labels in the same group are mark e d with the same color and labels in different groups are marked with different colors.
Note: This function is only app licable to DS1000Z Plus with the MSO upgrade
option.
Figure 12 User Interface
DS1000Z Quick Guide
Page 28
RIGOL
waveform on the screen
waveform in the memory
3. Status
Available states include RUN, STOP, T’D (triggered), W AIT and AUTO.
4. Horizontal Time Base
Represent the time per grid on the horizontal axis on the screen. Use HORIZONTAL
SCALE to modify this parameter. The range
available is from 5 ns to 50 s.
5. Sample Rate /Memory Dep th
Display the current sample rate and memory depth of the oscilloscope. The sample rate and memory depth will change in accordance with the
horizontal time base.
6. Waveform Memory
Provide the schematic diagram of the memory position of the waveform currently on the screen.
7. Trigger Position
Display the trigger position of the waveform in the waveform memory and on the screen.
8. Horizontal Position
Use HORIZONTAL
POSITION to modify this parameter. Press down the
knob to automatically set the parameter to zero.
9. Trigger Type
Display the currently selected trigger type and trigger condition setting. Different labels are displayed when different trigger types are selected. For example:
represents triggering on the rising edge in “Edge” trigger.
10. Trigger Source
Display the trigger source currently selected (CH1-CH4, AC Line or any channel of D0-D15). Different labels are displayed when different trigger sources are selected and the color of the trigger parameter area will change accordingly. For example:
denotes that CH1 is sel ected as th e trigger source.
11. Trigger Level
When the trigger source is set to analog channel, you need to set proper
trigger level.
at the right of the screen is the trigger level label and the trigger level
value is displayed at the upper-right corner of the screen.
18 DS1000Z Quick Guide
Page 29
RIGOL
19
When using TRIGGER LEVEL to modify the trigger level, the trigger
level value will change with the up and down of
.
Note: In slope trigger, runt trigger and windows trigger, there are two
trigger level labels (
and ).
12. CH1 Vertical Scale
Display the voltage value per grid of CH1 waveform vertically. Use VERTICAL
SCALE to modify this parameter.
The following labels will be displayed according to the current channel
setting: channel coupling (e.g.
) and bandwidt h limit (e.g. ).
13. Analog Channel Label/Waveform
Different channels are ma rked with diffe rent colors and the col ors of the channel label and waveform are the same.
14. CH2 Vertical Scale
Display the voltage value per grid of CH2 waveform vertica lly. Use VERTICAL
SCALE to modify this parameter.
The following labels will be displayed according to the current channel
setting: channel coupling (e.g.
) and bandwidt h limit (e.g. ).
15. CH3 Vertical Scale
Display the voltage value per grid of CH3 waveform vertically. Use VERTICAL
SCALE to modify this parameter.
The following labels will be displayed according to the current channel
setting: channel coupling (e.g.
) and bandwidt h limit (e.g. ).
16. CH4 Vertical Scale
Display the vo ltage value per grid of CH4 waveform vertically. Use VERTICAL
SCALE to modify this parameter.
The following labels will be displayed according to the current channel
setting: channel coupling (e.g.
) and bandwidt h limit (e.g. ).
17. Message Box
Display prompt messages.
18. Digital Channel Status Area
Display the current status of the 16 digital channels. The digital channels currently turned on a re displa yed in green, the digital channel currently sele cted is displayed in red and the digital channels turned off are displayed in grey.
Note: This function is only app licable to DS1000Z Plus with the MSO upgrade
option.
19. Source1 Waveform
Display the type of waveform currently set for Source1. When the mo dulation is enabled,
will be displayed at the bottom of the
DS1000Z Quick Guide
Page 30
RIGOL
Denote that you can rotate the multifunction knob to modify modification.
Denote that you can rotate to select the desired item, the item
with this symbol are selected.
Denote that press to input desired paramrter values directly using menus with this symbol are selected.
Denote that the current menu has several options.
Denote that the current menu has a lower level menu.
Press this key to return to the previous menu.
The number of the dots denotes that the number of the pages the current menu has.
Source1 Waveform.
When the impedance of si gnal source is set to 50 Ω,
will be displayed at
the bottom of the Source1 Waveform.
Only applicable to digital oscilloscopes with sour ce channels.
20. Source2 Waveform
Display the type of waveform currently set for Source2. When the mo dulation is enabled,
will be displayed at the bottom of the
Source2 Waveform.
When the impedance of si gnal source is set to 50 Ω,
will be displayed at
the bottom of the Source2 Waveform.
Only applicable to digital oscilloscopes with source channels.
21. Notification Area
Display sound icon and USB disk icon.  Sound Icon: Press UtilitySound to enable or dis a ble the sound. When
the sound is enabled, will be displayed; when the sound is disabled,
will be displayed.
USB Disk Icon: when a USB disk is detected,
will be displayed.
22. Operation Menu
Press any softkey to activate the corresponding menu. The following symbols might be displayed in the menu:
parameters. The backlight of turns on during the parameter
currently selected is displayed in b lue, press d own the knob to enter the corresponding menu. The backlight of
is constant on after menus
the pop-up numeric keyboard. The backlight of is constant on after
20 DS1000Z Quick Guide
Page 31
RIGOL
21
Security Lock Hole

Parameter Setting Method

This oscilloscope provides two methods to set the basic parameters. Method 1:
When directly.
Method 2: When
shown below. Rotate the knob to select the desired value, and then press down the knob to input the value. Aft er all the value i nput, rotate the kn ob to select the desired unit, press down the knob to finish your input.
is displayed on the menu, you can rotate to set the desired value
is displayed, press down and the numeric keyboard is displayed, as
Figure 13 Numeric Keyboard

To Use the Security Lock

If needed, you can use the security lock (please buy it yourself) to lock the oscilloscope to a fixed location. The method is as f ollows, align the lock wi th the lo ck hole and plug it into the lock hole vertically, turn the key clockwise to lock the oscilloscope and then pull the key out.
Figure 14 To Use the Security Lock
Note: Please do not insert other articles into the security lock hole to avoid
damaging the instrument.
DS1000Z Quick Guide
Page 32
RIGOL
Help Options Help Display Area

To Use the Built-in Help System

The help system of this oscilloscope provides instructions for all the function keys (including menu keys ) on the front panel. Press Help to open the help interfa ce a nd press ag ain to clos e the inter face. The help interface mainly consists of two parts. The left is "Help Options" and the right is "Help Display Area".
Figure 15 Help Information
You can press the button (except the Power key Page Up/Down key information in the "Help Display Area".
/ ) on the front panel directly to get the corresponding help
, the knob and the menu
22 DS1000Z Quick Guide
Page 33
RIGOL
23

Troubleshooting

The commonly encountered failures and their solutions are listed below. When you encounter those failures, please s olve the m followin g the cor responding ste ps. If the problem remains still, please contact RIGOL and provide your device information (acquisition method: Utility System System Info).
1. The screen is still dark (no display) after you powered on the instrument.
(1) Check whether the power switch is really on. (2) Check whether the power is correctly connected. (3) Check whether the fuse is burned out. If the fuse needs to be changed,
please use the specified fuse. (4) Restart the instrument after finishing the above inspections. (5) If it still persists, please contact RIGOL.
2. The signal is sampled but no waveform of the signal is displayed.
(1) Check whether the probe is correctly connected to the item under test. (2) Check whether signals are generated from the item under test (you can
connect the probe compensat ion signal to the proble matic channel to locate
the problem). (3) Resample the signal.
3. The voltage amplitude measured is greater or smaller than the actual value (note that this failure usually only occurs when probe is used).
Check whether the probe ratio of the channel compli es with the at tenuation rat io of the probe.
4. Waveforms are found to be displayed but not stable.
(1) Check the trigger sign a l source: press MENU (in the trigger control area
(TRIGGER) on the front panel) Source to check whether the trigger source selected complies with the signal channel actually used.
(2) Check the trigger type: general signals sh oul d use "Edge" trigger and vi deo
signal should use "Video" trigger. Only when t he prop er trigger t ype is used,
can the waveform be displayed stably. (3) Check the trigger level: adjust the trigger level to the middle of the signal. (4) Change the trigger holdoff setting.
5. No display is found after pressing RUN/STOP.
Check whether the trigger mode in the trigger control area (TRIGGER) on the front panel is "Normal" or "Single" and whether the trigger level exceeds the waveform range. If yes, set the trigger level t o the middl e or press MODE to set the trigger mode to "Auto". Note: Pressing AUTO could automatically finish the above setting.
DS1000Z Quick Guide
Page 34
RIGOL
6. The display of waveform is ladder-like.
(1) The horizontal time base might be too low. Increase the horizontal time
base to increase the horizontal resolution and improve the display.
(2) If the display type is "Vectors", the lines between the sample points may
cause ladder-like display. Press Display Type to set the display type to "Dots" to solve the problem.
7. I Fail to connect PC or PictBridge printer through USB. (1) Press Utility IO Setting USB Device to check whether the current
setting matches the device currently connected.
(2) Check whether the USB cable is correctly connected to the instrument and
PC.
(3) Check whether the USB cable is in good condition. If needed, restart the
oscilloscope.
8. The USB storage device cannot be recognized.
(1) Check whether the USB storage device can work normally. (2) Check whether the USB storage device being used is flash type. This
oscilloscope does not support USB3.0 a nd hardware type USB stor a ge device.
(3) Check whether the capacity of the USB storage device is too large. It is
recommended that the capacity of the USB storage device being used with
this oscilloscope should not exceed 8 GBytes. (4) Restart the instrument and then insert the USB storage device to check it. (5) If the USB storage device still cannot be used normally, please contact
RIGOL.
24 DS1000Z Quick Guide
Page 35
RIGOL
快速指南
DS1000Z 系列数字示波器
2019 4
RIGOL (SUZHOU) TECHNOLOGIES INC.
Page 36
Page 37
RIGOL
I

保证和声明

版权
© 2013 苏州普源精电科技有限公司
商标信息
RIGOL 是苏州普源精电科技有限公司的注册商标。
文档编号
QGA19008-1110
声明
本公司产品受中国及其它国家和地区的专利(包括已取得的和正在申请的专利)
保护。
本公司保留改变规格及价格的权利。 本手册提供的信息取代以往出版的所有资料。 本手册提供的信息如有变更,恕不另行通知。 对于本手册可能包含的错误,或因手册所提供的信息及演绎的功能以及因使用本
手册而导致的任何偶然或继发的损失,RIGOL 概不负责。
未经 RIGOL 事先书面许可,不得影印、复制或改编本手册的任何部分。
产品认证
RIGOL 认证本产品符合中国国家产品标准和行业产品标准及 ISO9001:2015 标准和 ISO14001:2015 标准,并进一步认证本产品符合其它国际标准组织成员的相关标准。
联系我们
如您在使用此产品或本手册的过程中有任何问题或需求,可与 RIGOL 联系: 电子邮箱:service@rigol.com 网址:www.rigol.com
DS1000Z 快速指南
Page 38
RIGOL
1.
9.
警告
注意
DANGER
WARNING
CAUTION

安全要求

一般安全概要

请使用所在国家认可的本产品专 用电源线。
2. 请确保产品可靠接地。
3. 正确连接探头。
4. 查看所有终端额定值。
5. 请使用合适的过压保护。
6. 请勿开盖操作。
7. 请勿将异物插入排风口。
8.
请使用合适的保险丝。

安全术语和符号

本手册中的安全术语:
避免电路外露。
10. 怀疑产品出故障时,请勿进行操作。
11. 请保持适当的通风。
12. 请勿在潮湿环境下操作。
13. 请勿在易燃易爆的环境下操作。
14. 请保持产品表面的清洁和干燥。
15. 请注意防静电保护。
16. 请注意搬运安全。
警告性声明指出可能会造成人身伤害或危及生命安全的情况或操作。
注意性声明指出可能导致本产品损坏或数据丢失的情况或操作。
产品上的安全术语:
表示您如果不进行此操作,可能会立即对您造成危害。 表示您如果不进行此操作,可能会对您造成潜在的危害。 表示您如果不进行此操作,可能会对本产品或连接到本产品的其他设 备造成损坏。
产品上的安全符号:
高电压
安全警告
保护性接地端
壳体接地端
测量接地端
II DS1000Z 快速指南
Page 39
RIGOL
III
警告
警告

测量类别

测量类别 DS1000Z 系列数字示波器可在测量类别 I 下进行测量。
本示波器仅允许在指定的测量类别中使用。
测量类别定义 测量类别 I 是指在没有直接连接到主电源的电路上进行测量。例如,对不是从主电源 导出的电路,特别是受保护(内部)的主电源导出的电路进行测量。在后一种情况下, 瞬间应力会发生变化。因此,用户应了解设备的瞬间承受能力。
测量类别 II 是指在直接连接到低压设备的电路上进行测量。例如,对家用电器、便携 式工具和类似的设备进行测量。
测量类别 III 是指在建筑设备中进行测量。例如,在固定设备中的配电板、断路器、
线路(包括电缆、母线、接线盒、开关、插座)以及工业用途的设备和某些其它设备
(例如,永久连接到固定装置的固定电机)上进行测量。
测量类别 IV 是指在低压设备的源上进行测量。例如,电表、在主要过电保护设备上的 测量以及在脉冲控制单元上的测量。

通风要求

本示波器通过风扇强制冷却。请确保进气和排气区域无阻塞并有自由流动的空气。为 保证充分的通风,在工作台或机架中使用示波器时,请确保其两侧、上方、后面应留
出至少 10 厘米的间隙。
通风不良会引起仪器温度升高,进而引起仪器损坏。使用时应保持良好的 通风,定期检查通风口和风扇。
DS1000Z 快速指南
Page 40
RIGOL
警告
警告

工作环境

温度 操作时:0℃至+50℃ 非操作时:-40℃至+70℃
湿度
0℃至+30℃:≤95%相对湿度 +30℃至+40℃:75%相对湿度 +40℃至+50℃:45%相对湿度
为避免仪器内部电路短路或发生电击的危险,请勿在潮湿环境下操作仪 器。
海拔高度 操作时:3000 米以下 非操作时:15000 米以下
安装(过电压)类别 本产品由符合安装(过电压)类别 II 的主电源供电。
确保没有过电压(如由雷电造成的电压)到达该产品。否则操作人员可 能有遭受电击的危险。
安装(过电压)类别定义 安装(过电压)类别 I 是指信号电平,其适用于连接到源电路中的设备测量端子,其 中已经采取措施,把瞬时电压限定在相应的低水平。 安装(过电压)类别 II 是指本地配电电平,其适用于连接到市电(交流电源)的设备。
污染程度 2
污染程度定义 污染度 1:无污染,或仅发生干燥的非传导性污染。此污染级别没有影响。例如:清 洁的房间或有空调控制的办公环境。 污染度 2:一般只发生干燥的非传导性污染。有时可能发生由于冷凝而造成的暂时性 传导。例如:一般室内环境。 污染度 3:发生传导性污染,或干燥的非传导性污染由于冷凝而变为具有传导性。例 如:有遮棚的室外环境。
IV DS1000Z 快速指南
Page 41
RIGOL
V
注意
警告
污染度 4:通过传导性的尘埃、雨水或雪产生永久的可导性污染。例如:户外场所。
安全级别 1 级 – 接地产品

保养与清洁

保养
请勿将仪器放置在长时间受到日照的地方。
清洁
请根据使用情况经常对仪器进行清洁。方法如下:
1. 断开电源。
2. 用柔和的清洁剂或清水浸湿软布擦拭仪器外部,请注意不要将水或其他异物通过
散热孔进入机箱内。清洁带有液晶显示屏的仪器时,请注意不要划伤 LCD 显示屏。
请勿使任何腐蚀性的液体沾到仪器上,以免损坏仪器。
重新通电之前,请确认仪器已经干透,避免因水分造成电气短路甚至人身 伤害。

环境注意事项

以下符号表明本产品符合 WEEE Directiv e 2002/96/EC 所制定的要求。
设备回收
本产品中包含的某些物质可能会对环境或人体健康有害,为避免将有害物质释放到环 境中或危害人体健康,建议采用适当的方法回收本产品,以确保大部分材料可正确地
重复使用或回收。有关处理或回收的信息,请与当地权威机构联系。
DS1000Z 快速指南
Page 42
RIGOL
标识
旋钮
标识
旋钮
水平
垂直
水平
POSITION
垂直
POSITION
触发
型号
模拟带宽
模拟通道数
信号源通道数
数字通道数
DS1104Z-S Plus
100 MHz
4 2 16
[1]
DS1074Z-S Plus
70 MHz
4 2 16
[1]
DS1104Z Plus
100 MHz
4
16
[1]
DS1074Z Plus
70 MHz
4
16
[1]
DS1054Z
50 MHz
4

文档概述

本文档用于指导用户快速了解DS1000Z系列数字示波器的前后面板、用户界面及基本 操作方法。您可登录RIGOL官网(www.rigol.com
文档格式的约定:
1. 按键
用“按键字符(加粗)+文本框”表示前面板功能按键,如 Utility 表 示“ Utility” 按键。
2. 菜单
用“菜单文字(加粗)+字符底纹”表示一个菜单选项,如 系统 表示 Utility 按 键下的“系统”菜单选项。
3. 操作步骤
用箭头“”表示下一步操作,如 Utility 系统 表示按下前面板上的 Utility 按键后,再按 系统 菜单键。
4. 旋钮
)下载所需文档的最新版本。
SCALE
LEVEL
水平档位旋钮
水平位移旋钮
SCALE
触发电平旋钮 —— ——
通道垂直档位旋 钮
通道垂直位移旋
文档内容的约定:
DS1000Z 系列包含以下型号。如无特殊说明,本手册以 DS1104Z-S Plus为例说明 DS1000Z 系列的前后面板、用户界面及基本操作方法等。
—— ——
[1]
——
:需通过MSO升级选件(包含逻辑分析仪探头RPL1116和型号标签)升级。
——
VI DS1000Z 快速指南
Page 43
RIGOL
VII
目录
保证和声明 .................................................................................................. I
安全要求 ................................................................................................... II
一般安全概要 ........................................................................................... II
安全术语和符号 ....................................................................................... II
测量类别 ............................................................................................... III
通风要求 ............................................................................................... III
工作环境 ................................................................................................ IV
保养与清洁 .............................................................................................. V
环境注意事项 ........................................................................................... V
文档概述 .................................................................................................. VI
快速入门 .................................................................................................... 1
一般性检查 .............................................................................................. 1
外观尺寸 ................................................................................................. 2
使用前准备 .............................................................................................. 3
调节支撑脚 ....................................................................................... 3
连接电源 ........................................................................................... 3
开机检查 ........................................................................................... 3
连接探头 ........................................................................................... 4
功能检查 ........................................................................................... 5
探头补偿 ........................................................................................... 6
前面板总览 .............................................................................................. 7
后面板总览 .............................................................................................. 8
前面板功能概述 ...................................................................................... 10
垂直控制 .......................................................................................... 10
逻辑分析仪 ...................................................................................... 11
信号源 ............................................................................................. 11
水平控制 .......................................................................................... 11
触发控制 .......................................................................................... 12
全部清除 .......................................................................................... 12
波形自动显示 ................................................................................... 12
运行控制 .......................................................................................... 12
单次触发 .......................................................................................... 13
多功能旋钮 ...................................................................................... 13
功能菜单 .......................................................................................... 13
打印 ................................................................................................ 14
用户界面
参数设置方法 .......................................................................................... 19
使用安全锁 ............................................................................................. 19
使用内置帮助系统 ................................................................................... 20
故障处理 .................................................................................................. 21
................................................................................................ 15
DS1000Z 快速指南
Page 44
Page 45
RIGOL

快速入门

一般性检查

1. 检查运输包装
如运输包装已损坏,请保留被损坏的包装或防震材料,直到货物经过完全检查且
仪器通过电性和机械测试。
因运输造成仪器损坏,由发货方和承运方联系赔偿事宜。RIGOL公司恕不进行
免费维修或更换。
2. 检查整机
若存在机械损坏或缺失,或者仪器未通过电性和机械测试,请联系您的 RIGOL
经销商。
3. 检查随机附件
请根据装箱单检查随机附件,如有损坏或缺失,请联系您的RIGOL经销商。
DS1000Z 快速指南 1
Page 46
RIGOL

外观尺寸

1 正视图 单位:mm
图 2 俯视图 单位:mm
2 DS1000Z 快速指南
Page 47
RIGOL
3

使用前准备

调节支撑脚
适当调整支撑脚,将其作为支架使示波器向上倾斜,以稳定放置示波器,便于更好地
操作和观察显示屏。
3 调节支撑脚
连接电源
本示波器可输入的交流电源的规格为:100-240 V45-440 Hz。请使用附件提供的电 源线如下图所示将示波器连接至 AC 电源。
电源插孔
4 连接电源
开机检查
当示波器处于通电状态时,按前面板左下角的电源键
程中示波器执行一系列自检,自检结束后出现开机画面。仪器出厂时我们将为用户提
供选件的试用版本,剩余时间约为 2000 分钟。如果您的仪器当前安装了试用版本的
选件,屏幕将弹出“已安装选件”对话框,您可以查看当前已安装选件的选件名称、
所属模块、选件版本和剩余时间。
DS1000Z 快速指南
即可启动示波器。开机过
Page 48
RIGOL
连接探头
RIGOL DS1000Z 系列提供无源探头和逻辑探头。探头的具体型号请参考《DS1000Z 系列数据手册》。有关探头的详细技术信息请参考相应的探头用户手册。
连接无源探头:
1. 将探头的 BNC 端连接至示波器前面板的模拟通道输入端。
2. 将探头接地鳄鱼夹连接至电路接地端,然后将探针连接至待测电路测试点中。
5 连接无源探头
连接无源探头后,您需要在测量前进行探头功能检查和探头补偿调节,具体步骤请参 考本手册中“
连接逻辑探头:
1. 将逻辑探头单线端按正确的方向连接至示波器前面板的数字通道输入接口。
2. 将逻辑探头另一端连接至被测信号端。RIGOL DS1000Z Plus系列选配 RPL1116
逻辑探头。为方便用户灵活探测信号,RPL1116提供了两种连接被测信号的方法。 具体请参考《RPL1116 逻辑探头用户手册》。
功能检查”和“探头补偿”一节介绍的内容。
数字通道输入接口
6 连接逻辑探头
注意:数字通道输入接口不支持热插拔,请勿在仪器带电的情况下插入或拔出逻辑探
头。
4 DS1000Z 快速指南
Page 49
5
功能检查
警告
提示
1. Storage 默认设置,将示波器恢复为默认配置。
2. 将探头的接地鳄鱼夹连接至如下图所示的“接地端”。
3. 使用探头连接示波器的通道 1CH1)输入端和“补偿信号输出端”。
补偿信号输出端 接地端
7 使用补偿信号
4. 将探头衰减比设定为 10X,然后按 AUTO 键。
5. 观察示波器显示屏上的波形,正常情况下应显示下图所示的方波:
RIGOL
8 方波信号
6. 用同样方法检查其他通道。如实际显示的方波形状与上图不相符,请执行下一节
探头补偿”。
为避免使用探头时被电击,请首先确保探头的绝缘导线完好,并且在连 接高压源时不要接触探头的金属部分。
探头补偿连接器上输出的信号仅作探头补偿调整之用,不可用于校准。
DS1000Z 快速指南
Page 50
RIGOL
探头补偿
首次使用探头时,应进行探头补偿调节,使探头与示波器输入通道匹配。未经补偿或
补偿偏差的探头会导致测量误差或错误。探头补偿步骤如下:
1. 执行上一节“
2. 检查所显示的波形形状并与下图对比。
3. 用非金属质地的改锥调整探头上的低频补偿调节孔,直到显示的波形如上图“补
偿正确”。
功能检查”中的步骤 1、2、3和 4。
补偿过度
补偿正确
9 探头补偿
补偿不足
6 DS1000Z 快速指南
Page 51
7

前面板总览

编号
说明
编号
说明
1
11
2
LCD
12
USB Host
3
13
[1]
4
14
5
15
[1]
6
16
[2]
7
17
8
18
9
19
10
20
3 4 5 6 7 8 9 10
1 2
11 12 13 14 15 16 17 18 19 20
1 前面板说明
RIGOL
10 前面板总览
测量菜单操作键
功能菜单操作键 多功能旋钮 常用操作键 全部清除键 波形自动显示 运行/停止控制键 单次触发控制键 内置帮助/打印键
[1]
:仅适用于 DS1000Z Plus
[2]
:仅适用于带有信号源通道的数字示波器。
DS1000Z 快速指南
电源键
接口 数字通道输入 模拟通道输入 逻辑分析仪操作键 信号源操作键 垂直控制 水平控制 触发控制 探头补偿信号输出端/接地端
Page 52
RIGOL
1 2 3
4 5 6 7 8

后面板总览

11 后面板总览
1. 手柄 垂直拉起该手柄,可方便提携示波器。不需要时,向下轻按手柄即可。
2. LAN
通过该接口将示波器连接到网络中,对其进行远程控制。本示波器符合 LXI CORE 2011 DEVICE 类仪器标准,可快速搭建测试系统。
3. USB DEVICE
通过该接口可将示波器连接至计算机或 PictBridge 打印机。连接计算机时,用户 可通过上位机软件发送 SCPI 命令或自定义编程控制示波器。连接打印机时,用 户通过打印机打印屏幕显示的波形。
4. 触发输出/通过失败 触发输出:
示波器产生一次触发时,可通过该接口输出一个反映示波器当前捕获率的信 号,将该信号连接至波形显示设备,测量该信号的频率,测量结果与当前捕
获率相同。
通过失败:
在通过/失败测试中,当示波器监测到一次失败时,将通过该连接器输出一个 负脉冲,未监测到失败时,通过该连接器持续输出低电平。
5. 信号源输出 示波器内置的 2 个信号源通道的输出端。当示波器中对应的源 1 输出或源 2 输出
8 DS1000Z 快速指南
Page 53
9
打开时,后面板[Source1][Source2]连接器输出当前设置的信号。
6. 锁孔
可以使用安全锁(请用户自行购买)通过该锁孔将示波器锁定在固定位置。
7. 保险丝
如需更换保险丝,请使用符合规格的保险丝。本示波器的保险丝规格为 250V
T2A。操作方法如下: a) 关闭仪器,断开电源,拔出电源线。 b) 使用小一字螺丝刀插入电源插口处的凹槽,轻轻撬出保险丝座。 c) 取出保险丝,更换指定规格的保险丝,然后将保险丝座安装回原处。
8. AC 电源插孔
AC 电源输入端。本示波器的供电要求为 100-240 V45-440 Hz。请使用附件提 供的电源线将示波器连接到 AC 电源中,按下前面板电源键即可开机。
RIGOL
DS1000Z 快速指南
Page 54
RIGOL
CH1、CH2、CH3、CH4
打开相应通道菜单,再次按下关闭通道。
、 、
垂直
垂直
提示
如何设置各通道的垂直档位和垂直位移?
垂直
垂直
垂直
垂直

前面板功能概述

垂直控制
:模拟通道设置键。
4 个通道标签用不同颜色标识,并且屏幕中的波
形和通道输入连接器的颜色也与之对应。按下任 一按键
MATH:按 MATH Math 可打开 A+B、A-B A×B、A/B、FFT、A&&B、A||B、A^B、!A、Intg DiffSqrtLgLnExp Abs等多种运算
按下 MATH 您还可以打开解码菜单,设置解码 选项。
POSITION:修改当前通道波形的垂直位移。顺时针转动增大位移,逆时
针转动减小位移。修改过程中波形会上下移动,同时屏幕左下角弹出的位移信息(如
)实时变化。按下该旋钮可快速将垂直位移归零。
SCALE:修改当前通道的垂直档位。顺时针转动减小档位,逆时针转动增大
档位。修改过程中波形显示幅度会增大或减小,同时屏幕下方的档位信息(如
)实时变化。按下该旋钮可快速切换垂直档位调节方式为“粗调”或“微
调 ”。
REF:按下该键打开参考波形功能。可将实测波 形和参考波形比较。
DS1000Z 系列数字示波器的 4 个通道复用同一组
SCALE 旋钮。如需设置某一通道的垂直档位和垂直位移,请首先按 CH1CH2 CH3 CH4 键选中该通道,然后旋转
钮进行设置。
10 DS1000Z 快速指南
POSITION
POSITION
SCALE
Page 55
RIGOL
11
[Source1]
水平 POSITION
水平
逻辑分析仪
按下该键打开逻辑分析仪控制菜单。您可以打开或关闭任意通道或通
道组、更改数字通道的显示大小、更改数字通道的逻辑阈值、对 16
数字通道分组等。您还可以为每一个数字通道设置标签。
注意:
― 该功能仅适用于带有 MSO 升级选件的 DS1000Z Plus。 ― 按 LA D7-D0,选择“打开”时,CH4 功能自动禁用;选择“关闭”时,CH4
功能自动恢复。按 LA D15-D8,选择“打开”时,CH3 功能自动禁用;选择 “关闭”时,CH3 功能自动恢复。
信号源
按下该键进入信号源设置界面。可打开或关闭后面板 [Source2]连接器的输出、设置信号源输出信号的波形及参数、打开 或关闭当前信号的状态显示。
注意:该功能仅适用于带有信号源通道的数字示波器。
水平控制
:修改水平位移。转动旋钮时触发点相对
屏幕中心左右移动。修改过程中,所有通道的波形左右移动,
同时屏幕右上角的水平位移信息(如 化。按下该旋钮可快速复位水平位移(或延迟扫描位移)。
MENU:按下该键打开水平控制菜单。可开关延迟扫描功能, 切换不同的时基模式。
SCALE:修改水平时基。顺时针转动减小时基,逆时
针转动增大时基。修改过程中,所有通道的波形被扩展或压缩
显示,同时屏幕上方的时基信息(如 下该旋钮可快速切换至延迟扫描状态。
)实时变化。按
)实 时 变
DS1000Z 快速指南
Page 56
RIGOL
MODE
Auto、Normal或Single
触发
RUN
状态,
整垂直档位、水平时基以及触发方式,使波形显示达到最佳状态。
RUN
触发控制
:按下该键切换触发方式为
当前触发方式对应的状态背光灯会变亮。
LEVEL:修改触发电平。顺时针转动增大电平,逆时针
转动减小电平。修改过程中,触发电平线上下移动,同时屏幕左
下角的触发电平消息框(如 变化。按下该旋钮可快速将触发电平恢复至零点。
MENU:按下该键打开触发操作菜单。本示波器提供丰富的触发
类型,详见《DS1000Z用户手册》中的介绍。
FORCE:按下该键将强制产生一个触发信号。
)中的值实时
全部清除
按下该键清除屏幕上所有的波形。如果示波器处于“
则继续显示新波形。
波形自动显示
按下该键启用波形自动设置功能。示波器将根据输入信号自动调
注意:应用波形自动设置功能时,若被测信号为正弦波,要求其频率不小于 41 Hz
若被测信号为方波,则要求其占空比大于 1%且幅度不小于 20 mVpp。如果不满足此 参数条件,则波形自动设置功能可能无效,且菜单显示的快速参数测量功能不可用。
运行控制
按下该键“运行”或“停止”波形采样。运行(
该键黄色背光灯点亮;停止(STOP)状态下,该键红色背光灯点 亮。
)状态下,
12 DS1000Z 快速指南
Page 57
RIGOL
13
Single”。单次触发方式下,
调节波形亮度:
形亮度
单次触发
按下该键将示波器的触发方式设置为“
FORCE 键立即产生一个触发信号。
多功能旋钮
非操作时,转动该旋钮可调整波形显示的亮度。亮度可调节范围
0%100%。顺时针转动增大波形亮度,逆时针转动减小波 形亮度。按下旋钮将波形亮度恢复至 60%。也可按 Display
多功能: 菜单操作时,该旋钮背光灯变亮,按下某个菜单软键后,转动该旋钮可选择该菜单下 的子菜单,然后按下旋钮可选中当前选择的子菜单。该旋钮还可以用于修改参数(具 体请参考“
参数设置方法”中的详细介绍)、输入文件名等。
,使用该旋钮调节波形亮度。
功能菜单
Measure:按下该键进入测量设置菜单。可设置测量信源、打开或关闭频率计、全部
测量、统计功能等。按下屏幕左侧的 MENU,可打开 37 种波形参数测量菜单,然后 按下相应的菜单软键快速实现“一键”测量,测量结果将出现在屏幕底部。
Acquire:按下该键进入采样设置菜单。可设置示波器的获取方式、Sin(x)/x 和存储 深度。
Storage:按下该键进入文件存储和调用界面。可存储的文件类型包括:图像存储、 轨迹存储、波形存储、设置存储、CSV 存储和参数存储。支持内、外部存储和磁盘管 理。
Cursor:按下该键进入光标测量菜单。示波器提供手动、追踪、自动和 XY 四种光标 模式。其中,XY 模式仅在时基模式为“XY”时有效。
DS1000Z 快速指南
Page 58
RIGOL
Display:按下该键进入显示设置菜单。设置波形显示类型、余辉时间、波形亮度、 屏幕网格和网格亮度。
Utility:按下该键进入系统功能设置菜单。设置系统相关功能或参数,例如接口、声 音、语言等。此外,还支持一些高级功能,例如通过/失败测试、波形录制等。
打印
按下该键打印屏幕或将屏幕保存到U盘中。
― 若当前已连接 PictBridge 打印机,并且打印机处于闲置状态,
按下该键将执行打印功能。
― 若当前未连接打印机,但连接 U 盘,按下该键则将屏幕图形以
指定格式保存到 U 盘中。
― 若同时连接打印机和 U 盘时,打印机优先级较高。
14 DS1000Z 快速指南
Page 59
15

用户界面

12 13 14 15 16 17 18 19 20 21 22
1 2 3 4 5 6 7 8 9 10 11
DS1000Z 示波器提供 7.0 英寸 WVGA800*480TFT LCD
RIGOL
12 用户界面
1. 自动测量选项
提供 20 种水平(HORIZONTAL)测量参数和 17 种垂直(VERTICAL)测量参数。 按下屏幕左侧的软键即可打开相应的测量项。连续按下 MENU 键,可切换水平 和垂直测量参数。
2. 数字通道标记/波形
数字波形的逻辑高电平显示为蓝色,逻辑低电平显示为绿色,边沿呈白色。当前 选中的数字通道波形和通道标签一致,显示为红色。逻辑分析仪功能菜单中的分 组设置功能可以将数字通道分为四个通道组,同一通道组的通道标记显示为同一
种颜色,不同通道组用不同的颜色表示。
注意:该功能仅适用于带有 MSO 升级选件的 DS1000Z Plus。
3. 运行状态
可能的状态包括:RUN(运行)、STOP(停止)、T’D(已触发)、WAIT(等待) 和 AUTO(自动)。
DS1000Z 快速指南
Page 60
RIGOL
4. 水平时基
表示屏幕水平轴上每格所代表的时间长度。 使用水平 SCALE 可以修改该参数,可设置范围为 5 ns 50 s
5. 采样率/存储深度
显示当前示波器使用的采样率以及存储深度。 采样率和存储深度会随着水平时基的变化而改变。
6. 波形存储器 提供当前屏幕中的波形在存储器中的位置示意图。
存储器中的波形
屏幕中的波形
7. 触发位置 显示波形存储器和屏幕中波形的触发位置。
8. 水平位移 使用水平 POSITION可以调节该参数。按下旋钮时参数自动设置为 0
9. 触发类型 显示当前选择的触发类型及触发条件设置。选择不同触发类型时显示不同的标 识。 例如,
表示在“边沿触发”的上升沿处触发。
10. 触发源 显示当前选择的触发源(CH1-CH4AC D0-D15)。选择不同触发源时,显示 不同的标识,并改变触发参数区的颜色。 例如,
表示选择 CH1 作为触发源。
11. 触发电平
触发信源选择模拟通道时,您需要设置合适的触发电平。 屏幕右侧的
为触发电平标记,右上角为触发电平值。
使用触发 LEVEL 修改触发电平时,触发电平值会随 的上下移动而改
变。
注意:斜率触发、欠幅脉冲触发和超幅触发时,有两个触发电平标记( 和 )。
12. CH1 垂直档位
显示屏幕垂直方向 CH1 每格波形所代表的电压。 CH1 选中 CH1 通道后,使用垂直 SCALE 可以修改该参数。 此外还会根据当前的通道设置给出如下标记:通道耦合(如 )、带宽限制
16 DS1000Z 快速指南
Page 61
17
(如 )。
13. 模拟通道标记/波形
不同通道用不同的颜色表示,通道标记和波形的颜色一致。
14. CH2 垂直档位
显示屏幕垂直方向 CH2 每格波形所代表的电压。 CH2 选中 CH2 通道后,使用垂直 SCALE 可以修改该参数。 此外还会根据当前的通道设置给出如下标记:通道耦合(如 )、 带宽限制(如
)。
15. CH3 垂直档位
显示屏幕垂直方向 CH1 每格波形所代表的电压。 CH3 选中 CH3 通道后,使用垂直 SCALE 可以修改该参数。 此外还会根据当前的通道设置给出如下标记:通道耦合(如 )、 带宽限制(如
)。
16. CH4 垂直档位
显示屏幕垂直方向 CH2 每格波形所代表的电压。 CH4 选中 CH4 通道后,使用垂直 SCALE 可以修改该参数。 此外还会根据当前的通道设置给出如下标记:通道耦合(如 )、带宽限制(如
)。
17. 消息框
显示提示消息。
18. 数字通道状态区
显示 16 个数字通道当前的状态。当前打开的数字通道显示为绿色,当前选中的 数字通道突出显示为红色,任何已关闭的数字通道均显示为灰色。
注意:该功能仅适用于带有 MSO 升级选件的 DS1000Z Plus。
19.1 波形
显示当前源 1 设置中的波形类型。 当源 1 的调制打开时,源 1 波形的下方会显示 当源 1 的阻抗设置为 50 Ω 时,源 1 波形的下方会显示
标识。
标识。
仅适用于带有信号源通道的数字示波器。
20.2 波形
显示当前源 2 设置中的波形类型。 当源 2 的调制打开时,源 2 波形的下方会显示 当源 2 的阻抗设置为 50 Ω 时,源 2 波形的下方会显示
标识。
标识。
仅适用于带有信号源通道的数字示波器。
RIGOL
DS1000Z 快速指南
Page 62
RIGOL
21. 通知区域 显示声音图标和 U 盘图标。  声音图标:按 Utility 声音 可以打开或关闭声音。声音打开时,该区域
显示 ;声音关闭时,显示
U 盘图标:当示波器检测到 U 盘时,该区域显示
22. 操作菜单 按下任一软键可激活相应的菜单。下面的符号可能显示在菜单中:
表示可以旋转多功能旋钮
修改参数值。多功能旋钮
的背光灯在
参数修改状态下变亮。 表示可以旋转多功能旋钮 色,按下
进入所选项对应的菜单栏。带有该符号的菜单被选中后,
选择所需选项,当前选中的选项显示为蓝
的背光灯常亮。
表示按下 的菜单被选中后, 表示当前菜单有若干选项。
表示当前菜单有下一层菜单。
将弹出数字键盘,可直接输入所需的参数值。带有该符号
的背光灯常亮。
按下该键可以返回上一级菜单。
圆点数表示当前菜单的页数。
18 DS1000Z 快速指南
Page 63
RIGOL
19

参数设置方法

DS1000Z 支持通过如下两种方法设置参数。
方法一: 对于菜单上显示
方法二: 对于菜单上显示
转旋钮选择所需的数值,按下旋钮输入该数值。输入全部数值后,旋转旋钮选择所需
的单位,按下旋钮即可完成参数设置。
的参数,直接旋转多功能旋钮 即可设置所需的数值。
的参数,按下多功能旋钮 ,弹出如下图所示的数字键盘。旋
13 数字键盘

使用安全锁

如有必要,您可以使用安全锁(请自行购买)将示波器锁在固定位置。方法如下,沿 与后面板垂直的方向对准锁孔将锁头插入,顺时针旋转钥匙锁定示波器,然后拔出钥
匙。
安全锁孔
14 使用安全锁
注意:请勿将其它物品插入安全锁孔以免损坏仪器。
DS1000Z 快速指南
Page 64
RIGOL

使用内置帮助系统

本示波器的帮助系统提供了前面板各功能键(包括菜单键)的说明。按 Help 键打开
帮助界面,再次按下则关闭。帮助界面主要分两部分,左边为“帮助选项”,右边为
“帮助显示区”。
帮助选项
帮助显示区
15 帮助信息
您可以直接按面板上的按键(电源键
、多功能旋钮 和菜单翻页键 /
外)即可在“帮助显示区”中获得相应的帮助信息。
20 DS1000Z 快速指南
Page 65
RIGOL
21

故障处理

下面列举了示波器在使用过程中可能出现的故障及排查方法。当您遇到这些故障时, 请按照相应的步骤进行处理,如不能处理,请与 RIGOL 联系,同时请提供您机器的
设备信息(获取方法:Utility 系统 系统信息)。
1. 如果按下电源键示波器仍然黑屏,没有任何显示:
(1) 检查电源开关是否打开。 (2) 检查电源接头是否接好。 (3) 检查保险丝是否熔断。如需更换保险丝,请使用符合本产品规格的保险丝。 (4) 做完上述检查后,重新启动仪器。 (5) 如果仍然无法正常使用本产品,请与 RIGOL 联系。
2. 采集信号后,画面中并未出现信号的波形:
(1) 检查探头是否与示波器和待测物正常连接。 (2) 检查待测物是否有信号产生(可将探头补偿输出信号连接到有问题的通道确
定是通道还是待测物的问题)。
(3) 再重新采集信号一次。
3. 测量的电压幅度值比实际值大或者小(一般在使用探头时才会出现):
检查通道的探头比设置是否与实际使用的探头衰减比例相符。
4. 有波形显示,但不能稳定下来:
(1) 检查触发信源:按前面板触发控制区(TRIGGER)的 MENU 信源,确
认所选的触发信源是否与实际使用的信号通道相符。
(2) 检查触发类型:一般的信号应使用“边沿触发”方式,视频信号应使用“视
频触发”方式。只有应用适合的触发方式,波形才能稳定显示。
(3) 检查触发电平:将触发电平调整至信号的中间位置。 (4) 改变触发释抑设置。
5. 按下 RUN/STOP 键无任何显示:
检查前面板触发控制区(TRIGGER),确认触发方式是否为“普通”或“单次” 档,且触发电平是否超出波形范围。如果是,将触发电平居中或者按 MODE 将 触发方式设置为“自动”档。 注:使用自动设置 AUTO 按钮可自动完成以上设置。
6. 波形显示呈阶梯状:
(1) 水平时基档位可能过低,增大水平时基以提高水平分辨率,可以改善显示。 (2) 若显示类型为“矢量”,采样点间的连线,可能造成波形阶梯状显示。按
Display 显示类型,设置为“点”显示方式,即可解决。
DS1000Z 快速指南
Page 66
RIGOL
7. 通过 USB 连接 PC PictBridge 打印机失败: (1) Utility 接口设置 USB 设备,检查当前设置是否与当前连接的设
备匹配。 (2) 检查 USB 数据线是否与示波器和 PC 连接正常。 (3) 检查 USB 数据线是否完好,必要时重启示波器。
8. U 盘不能被识别:
(1) 检查 U 盘是否可以正常工作。 (2) 确认使用的为 Flash U 盘,本仪器不支持 USB3.0 U 盘和硬盘型 U 盘。 (3) 确认使用的 U 盘容量是否过大,本示波器推荐使用不超过 8 GBytes 的 U 盘。 (4) 重新启动仪器后,再插入 U 盘设备进行检查。 (5) 如果仍然无法正常使用 U 盘,请与 RIGOL 联系。
22 DS1000Z 快速指南