Review the following saf et y precautio ns caref ully before ope rating the instrument to
avoid any personal injuries or damages to the instrument and any products
connected to it. To avoid potential hazards use the instrument as specif ied by this
user’s guide only.
The instrument should be serviced by qualified personnel only.
Avoid Fire or Personal Injury.
Use Proper Power Cord. Use the power cord designed for the instrument and
authorized in your country only.
Connect and Disconnect Accessories. Do not connect or disconnect probes or
test leads while they are connected to a voltage source.
Ground The Instrument. The oscilloscope is grounded through the grounding
conductor of the power cord. To avoid electric shock the instrument grounding
conductor(s) must be grounded properly . Bef ore making connections t o the in put or
output terminals of the instrument.
Connect The Probe. The probes’ grou nd terminals are at the same v oltage lev el of
the instrument ground. Do not connect the ground terminals to a high voltage.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and marks on the instru ment . F ollow the user’ s guide for further r atin gs inf ormatio n
before making connections to the instrument.
Do Not Operate Without C ove rs. Do n o t op e ra te the instrument with covers or
panels removed.
Use Proper Fuse. Use the f use of the type, voltage and current ratings as specified
for the instrument.
Avoid Circuit or Wire Exposure. Do not touch exposed connections and
components when power is on.
Do Not Operate With Suspected Failures. If suspected damage occurs with the
instrument, have it inspected by qualified service personnel before further
operations.
Provide Proper Ventilation. Refer to the installation instructions for proper
ventilation of the instrument.
Do not operate in wet/damp conditions
Do not operate in an explosive atmosphere
Keep product surfaces clean and dry
The disturbance test of all the models meet the limit values of A in the
standard of EN 61326: 1997+A1+A2+A3, but can’t meet the limit values
of B.
Measurement Category
The DS1000E, DS1000D series Digital Oscilloscope is intended to be used for
measurements in Measurement Category I.
Measurement Category Definitions
Measurement Category I is for measurements performed on circuits not directly
connected to MAINS. Examples are measurements on circuits not derived from
MAINS, and specially protected (inte rnal ) MAI NS de riv ed circuits. In the latter case,
transient stresses are variable; for that reason, the transient withstand capabil ity of
the equipment is made known to the user.
WARNING
IEC Measurement Category I, The input terminals may be connected to circuit
terminal in IEC Category I installations for voltages up to 300 VAC. To avoid the
danger of electric shock, do not connect the inputs to circuit’s voltages above 300
VAC. Transient overvoltage is also present on circuits that are isolated from mains.
The DS1000E, DS1000D series Digital Oscilloscopes is designed to safely withstand
occasional transient overvoltage up to 1000Vpk. Do not use this equipment to
measure circuits where transient overvoltage could exceed this level.
Terms in This Guide: These terms may appear in this guide:
WARNING: Warning statement s identify conditions or practices that could result in injury or loss of life.
CAUTION: Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the Product: These terms may appear on the product:
DANGER: indicates an injury hazard may be immediately accessible.
WARNING: indicates a potential injury hazard may be immediately accessible.
CAUTION: indicates that a potential damage to the instrument or other property
might occur.
Symbols on the Product: These symbols may appear on the product:
This book covers the following four types of DS1000E, DS1000D Series Digital
Oscilloscopes:
DS1102E, DS1052E
DS1102D, DS1052D (With Logic Analyzer)
DS1000E, DS1000D series are kinds of economical digital oscilloscope with
high-performance. DS1000E series are designed with dual channels and 1 external
trigger channel. DS1000D series are design ed with dual channels and 1 external
trigger channel as well as 16 channels logic analyzer.
The front panel of DS1000E, DS100 0D series digital oscilloscopes are clear, direct
viewing and easy to use, which consistent with the traditional instrument use fully.
To speed up the adjustment and measurement, please press AUTO button to get
suitable wav eform display and set tings. Besides, 1GSa/s maximum real-time sample
rate, 25GSa/s maximum equivalent sample rate and the powerful trigger and
analyzer functions enable users view or capture the wav e form more detailed and
faster.
Main Features
Dual analog channels, 1GSa/s maximum real-time sample rate and 25GSa/s
maximum equivalent sample rate, Bandwidth of per channel:
100MHz (DS1102E, DS1102D)
50MHz (DS1052E, DS1052D)
Optional 16 digital channels (DS1000D series), each channel can be tu rn ed on
or off independently, or in a 8 bit group
5.6 inch and 64 k TFT LCD
Abundant trigger types: Edge, Pulse Width, Video , Slope , Alternate, Pattern and
Duration Trigger (Only for DS1000D series)
Unique adjustable trigger sensitivity enables to meet different demands
Enable t o measur e 22 types of wave p arameter s and track measurements via
cursor automatically
Unique waveform record a nd replay function
Fine delayed scan function
Built-in FFT function, hold practical digital filters
Provide 4 types of digital filters: LPF, HPF, BPF, BRF
Pass/Fail detection function enables to output testing results
Math operations available to multiple waves
Powerful PC applic ation soft ware UltraScope
Standard configuration interface: USB Device, USB Host ,RS-232 and support
USB storage device storage and PictBridge print standards
The new function “Special Mode” can meet the needs of industrial production
Support f or remote command control
Built-in help menu enables information getting more convenient
Multiple Language User Interface, support Chinese & English input
Can store files into USB storage device or the internal memory
Waveform intensity can be adjusted
To display a signal automatically by AUTO
Pop-up menu makes it easy to read and use
To Inspect the Instrument
The Panels and User Interface
Appearanc e and Dimen s io ns
To Perform a Functional Check
To Compensate Probes
Digital Leads (Only for DS1000D Series)
To Display a Signal Automatically
To Understand the Vertical System
To Understand the Horizontal System
When you get a new DS1000E, DS1000D series oscilloscope, please inspect the
instrument according to the following steps:
1. Inspect the shipping container for damage.
Keep a damaged shipping container or cushioning material until the contents of
the shipment have been checked for completeness and the instrument has been
checked mechanically and electrically.
2. Inspect the instrument.
In case t here is any mechanical damage or defect, or the instrument does not
operate properly or fails performance tests, please notify the RIGOL Sales
Representative.
If the shipping container is damaged, or the cushioning materials show signs of
stress, please notify the carrier as well as the RIGOL sales off ice. Keep the
shipping materials for the carrier’s inspection. RIGOL offices will arrange for
repair or replacement at RIGOL’s option without waiting for claim settlement.
3. Check the accessories.
Accessories supplied with the instrument are listed in "Appendix A: Accessories"
in this guide.
If the contents are incomplete or damaged, please notify your RIGOL Sales
Representative.
The f irst thing to do with a new oscilloscope is to know its front panel. This chapter
helps to be familiar with the layout of the knobs and buttons and how to use them.
Read the chapter carefully before further operations.
1. Front Panel
Figure 1- 1, Front P anel; the knobs a re use d most often a nd are si milar t o the kno bs
on other oscilloscopes. The buttons allow you to use some of the functions directly
but also bring up soft but t on men us on the screen, which enable the access to many
measurement features associated with advanced functions, mathematics, and
reference or to run control fea t u res.
The front panel of DS1000E:
The front panel of DS1000D:
Figure 1- 1
Front Panel Figure of DS1000E, DS1000D Series Oscilloscope
Rear Panel of DS1000E, DS10000D series contains the following parts:
① Pass/Fail output port: The Pass /Fail tes ting results can be output by the port.
② RS232 port: To connect the oscilloscopes and external equipment as serial
port.
③USB Device port: It is used to transfer data when external USB equipment
connects to the oscilloscope regarded as “Device equipment”. For example: to
use this port when connect PictBridge printer to the instrument by USB.
Notation definitions in this User’s Guide:
Throughout this manual, notation symbols of buttons and knobs are the same to
those on front-panel.
A box around the name of the button denotes MENU function buttons on
front-panel, such as Measure.
denotes the multi-function knob.
denotes the two POSITION knobs.
denotes the two SCALE knobs.
denotes the LEVEL knob.
The name with a drop shadow denotes the menu operating button, such as
Storage denoting the "Storage" menu item under Measure.
Perform this quick functional check to verify that the instrument is operating
correctly.
1. Turn on the instrument.
Use the power cord designed for the oscilloscope only.
The power requirements are 100 to 240 VAC
T urn on the instruments, a nd wait until the display shows the waveform window.
Press the Storagebutton, select Storage in the menu. Rotate the multiple
function knob to select Factory and press the knob down. Then, press Load .
, 45Hz to 440Hz.
RMS
Storage button
Figure 1- 9
Turn on and Check the instrument
WARNING:
To avoid electric shock, be sure the oscilloscope is properly grounded.
2. Input a signal to a channel of the oscilloscope
Align the slot in the probe connector with the button on the CH1 BNC.
Push to connect, and twist to the right to lock the probe.
Attach the probe tip and ground lead to the PROBE COMP connector.
Probe compensator
Figure 1- 10
Attach the probe
② Set the probe attenuation to 10X. To do this, push CH1→Probe→10X.
RIGOL
Probe scale
Figure 1- 11 Figure 1- 12
Set attenuation on the probe Set attenuation in the menu
③ Push the AUTO button. Within a few seconds, a square wave will display.
④ Push the CH1 button a gain to turn off Channel1.
Push the CH2 butt on to turn on c hannel2, r epeat step s 2 and 3.
NOTE: The signal output from Prob e compensator should only be used for probe
compensation, not for calibration.
Perform this adjustment to match the characteristics of the probe and the channel
input. This should be performed whenever attaching a probe to any input channel for
the f i r st time.
1. From CH1 menu, set the Probe attenuation to 10X (press CH1→Probe→10X).
Set the switch to 10X on the probe and connect it to CH1 of the oscilloscope.
When using the probe hook-tip, inserting the tip onto the probe firmly to ensure
a proper connection.
Attach the probe tip to the Probe compensa tor connector and the reference lead
to the ground pin, Select CH1, and t hen press AUTO.
2. Ch eck the shape of the displayed waveform.
Over compensated Correctly Compensated Under Compensated
Figure 1-9
Figure 1- 13
Probe Compensation
3. If necessary , use a non-metallic tool to a djust the t rimmer ca pacitor on the pro be
for the flattest square wave possible as displayed on the oscilloscope.
4. Repeat as necessary.
WARNING: To avoid electric shock while using the probe, be sure the
perfection of the insulated cable, and do not touch the metallic portions of
the probe head while it is connected with a voltage source.
Digital leads are provided only for DS10 0 0D series which have Logic Analyzer.
1. Switch off power supply of the device under test if necessary to avoid short
circuit. Since no voltage is applied to the leads at this step, you may keep the
oscilloscope on.
2. Connect one end of the flat cable FC1868 to the Logic Analyzer Input; connect
the other end to Logic Head LH1116. An identifier is located on each end of the
flat cable; it can only be connected in one way. It is unnecessary to switch off
power supply of your oscilloscope when connecting the cable.
Figure 1- 14
Connect the digital leads
CAUTION: Use only FC1868, LH1116, TC1100 and LC1150 made by
RIGOL for specified DS1000D series.
3. Connect a test clip to one lead wire; make sure it’s connection good.
The oscilloscope has an automated feature to display the input signal best-fit. The
input signal should be 50Hz or higher and a duty cycle is greater than 1%.
Use Auto Setting:
1. Connect a signal to the oscilloscope as described above.
2. Press AUTO.
The oscilloscope will automatically set up VERTICAL, HORIZONTAL and TRIGGER
controls to display the input signal. Adjust the controls manually to get the best
results if necessary.
Figure 1- 18 shows the VERTICAL controls, CH1, CH2, MATH, REF and OFF buttons
and vertical , knobs. Following the exercise of the buttons,
knobs, and the status bar to be familiar with the vertical parameters settings.
Figure 1- 18
The vertical system
1. Moves the signal vertically on the display with the
knob.
When turning the
knob, a voltage value is displayed for a short time
indicating its value with respect to the ground reference located at the center of the
screen. Also notice that the ground symbol on the left side of the display moves in
conjunction with the
knob.
Measurement hints
If the channel is DC coupled, measuring the DC component of the signal by
simply noting its distance from the ground.
If the channel is AC coupled, the DC comp onent of th e signal is blocke d, allow
you to use greater sensitivity to display the AC component of the signal.
, this is especially helpf ul when t he tr ace positio n is
pressing the vertical
Vertical offset back to 0 shortcut key
RIGOL
Turn the
channel and press the knob to set the vertical display position
back to 0 as a shortcut key
far out of the sc reen and want it to get ba ck to the screen center immediately.
2. Change the vertical setup and notice that each change affects the
status bar differently.
View the status bar which is on the bottom of the screen to understand the
vertical scale.
Change the ve rtical scale by turning the
knob and notice the change
in the status bar.
Press CH1 , CH2 , MATH , REF , LA (only for DS1000D series) to display
the menu, sign, waveform of corresponding channel. Press OFF button to turn
off the channel.
Coarse/Fine Shortcut key
The Coarse/Fine vertical control can be set by simply
is another way to enter or exit Delayed Scan mode and it is equal to the
To Understand the Horizontal System
Figure 1-19 shows the HORIZONTAL controls: MENU button, and
knobs of horizontal system. Following the exercise to familiarize with the
buttons, knobs, and status bar.
Figure 1- 19
The horizontal system
1. Turn the
knob and notice the change in the status bar.
The horizontal
knob changes the sweep speed in a 1-2-5 step
sequence, and displays the value in the status bar. The time base rang e s o f th e
oscilloscope is from 2ns/div* to 50s/div.
Delayed Scan Shortcut key
To press the
knob in the horizontal contr ol a rea on the f ront -panel
following menu operations, MENU→Delayed→ON.
* NOTE: The speed of horizontal scan varies by different models.
Figur e 1-20 shows the trigger control: MENU, 50%, FORCE and a trigger
level knob. Following the exercise to familiarize with the buttons, trigger level knob
and status bar.
Figure 1-20
The trigger control window
1. Turn the trigger Level knob and notice the changes on the display.
As you turn the
knob or pressing the 50% button, two things will happen
on the display for a short time.
First, the trigger level value is displayed at the bottom-left of the screen.
Second, a line is displayed showing the location of the trigger level (as long as
AC coupling or low frequency reject are not selected).
Trigger Level to 0 Shortcut key
Turn the knob to change trigger level value and press the
knob to set tri gger level back to 0 as a sho rtcut key.
A time interval before the oscilloscope response to next trigger
2. Change the trigger setup and notice these changes in the statu s ba r.
Press MENU button in the Trigger control. A soft button me nu appears on the display
showing the trigger setting choices as shown in Figure 1 -21.
Figure 1- 21
· Press the trigger Mode button and choose Edge.
· Press the trigger Source button to select CH1.
· Press the trigger Slope button to choos e .
· Press the trigger Sweep button to select Auto.
· Press the trigger Set Up button to enter secondary menu.
NOTE: The trigger type, slope and source cha nge in conjun ction with the status bar
on the top-right of the screen.
3. Press 50%
The 50% button sets the trigger level to the center of the signal.
4. Press FORCE
Starting an acquisition regardless of an adequate trigger signal, usually used in
“Normal” or “Single” trigger mode. This button has no effect if the acquisition is
already stopped.
Key point:
Holdoff:
signal. During this holdoff period, the trigger system becomes “blind” to trigger
signals. This function helps to view complex signals such as an AM waveform.
Press Holdoff button to activate (