Rigol DS1000E Series, DS1000D Series User Manual

Page 1
User’s Guide RIGOL
Publication number UGA07118-1110 Dec. 2015
DS1000E, DS1000D Series
Digital Oscilloscopes
DS1102E, DS1052E, DS1102D, DS1052D
© 2008 RIGOL Technologies, Inc. All R ights Rese rved
www.GlobalTestSupply.com
Page 2
www.GlobalTestSupply.com
Page 3
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
I
1. © 2008 RIGOL Technologies, Inc. All Rights Reserved
2. RIGOL products are protected by patent law in and outside of P.R. China.
3. Information in this publication replaces all previ ously corresponding material.
4. RIGOL Technologies, Inc. reserves the right to modify or change part of or all the specifications and pricing policies at company’s sole decision.
NOTE: RIGOL is registered trademark of RIGOL Technologies, Inc.
www.GlobalTestSupply.com
Page 4
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
II
Safety Notices
Review the followin g safet y pre cautions caref ully before operating the instrument to avoid any personal injuries or damages to the instrument and any products connected to it. To avoid potential hazards use the instrument as specified by this user’s guide only.
The instrument should be serviced by qualified personnel only. Avoid Fire or Personal Injury. Use Proper Power Cord. Use the power cord designed for the instrument and
authorized in your country only. Connect and Disconnect Accessories. Do not connect or disconnect probes or
test leads while they are connected to a voltage source. Ground The Instrument. The oscilloscope is grounded through the grounding
conductor of the power cord. To avoid electric shock the instrument grounding conductor(s) must be grounded properly. Before ma king connections to the in put o r output terminals of the instrument.
Connect The Probe. The probes’ g round terminals are at t he same v oltage lev el of the instrument ground. Do not connect the ground terminals to a high voltage.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and marks on the inst rum ent. Follow the user’ s guide for further r atin gs inf ormatio n before making connections to the instrument.
Do Not Operate Withou t C over s. Do no t o p era te the instrument with covers or panels removed.
Use Proper Fuse. Use the fuse of the type, voltage and current ratings as specified for the instrument.
Avoid Circuit or Wire Exposure. Do not touch exposed connections and components when power is on.
Do Not Operate With Suspected Failures. If suspected damage occurs with the instrument, have it inspected by qualified service personnel before further operations.
Provide Proper Ventilation. Refer to the installation instructions for proper ventilation of the instrument.
www.GlobalTestSupply.com
Page 5
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
III
Do not operate in wet/damp conditions Do not operate in an explosive atmosphere Keep product surfaces clean and dry The disturbance test of all the models meet the limit values of A in the
standard of EN 61326: 1997+A1+A2+A3, but can’t meet the limit values of B.
Measurement Category
The DS1000E, DS1000D series Digital Oscilloscope is intended to be used for measurements in Measurement Category I.
Measurement Category Definitions
Measurement Category I is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (inte rnal ) MAI NS de riv ed circuits. In the latter case , transient stresses are variable; for that reason, the transient withstand capability of the equipment is made known to the user.
WARNING
IEC Measurement Category I, The input terminals may be connected to circuit terminal in IEC Category I installations for voltages up to 300 VAC. To avoid the danger of electric shock, do not connect the inputs to circuit’s voltages above 300 VAC. Transient overvoltage is also present on circuits that are isolated from mains. The DS1000E, DS1000D series Digital Oscilloscopes is designed to safely withstand occasional transient overvoltage up to 1000Vpk. Do not use this equipment to measure circuits where transient overvoltage could exceed this level.
www.GlobalTestSupply.com
Page 6
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
IV
Safety Terms and Symbols
Terms in This Guide: These terms may appear in this guide:
WARNING: Warning stat ements identify conditions or practices that could result in injury or loss of life.
CAUTION: Caution statements identify conditions or practices that could result in dama ge to this product or other p roperty.
Terms on the Product: These terms may appear on the product:
DANGER: ind icates an injury hazard may be immediately accessible. WARNING: indicates a potential injury hazard may be immediately accessible. CAUTION: indicates that a potential damage to the instrument or other property
might occur.
Symbols on the Product: These symbols may appear on the product:
!
Hazardous
Voltage
Safety
Warning
Protective
Earth Terminal
Test
Grounding Terminal
Grounding
Terminal of Chassis
!
!
www.GlobalTestSupply.com
Page 7
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
V
General-Purpose Oscillo scop e s
This book covers the following four types of DS1000E, DS1000D Series Digital Oscilloscopes: DS1102E, DS1052E DS1102D, DS1052D (With Logic Analyzer)
DS1000E, DS1000D series are kinds of economical digital oscilloscope with high-performance. DS1000E series are designed with dual channels and 1 external trigger channel. DS1000D series are designed with dual channels and 1 external trigger channel as well as 16 channels logic analyzer.
The front panel of DS1000E , DS100 0D series digital oscilloscopes are clear, direct viewing and easy to use, which consistent with the traditional instrument use fully. To speed up the adjustment and measurement, please press AUTO button to get suitable wa vefo rm display and settings. Besides, 1GSa/s maximum real-time sample rate, 25GSa/s maximum equivalent sample rate and the powerful trigger and analyzer functions enable users view or capture the waveform more detailed and faster.
Main Features
Dual analog channels, 1GSa/s maximum real-time sample rate and 25GSa/s
maximum equivalent sample rate, Bandwidth of per channel: 100MHz (DS1102E, DS1102D) 50MHz (DS1052E, DS1052D)
Optional 16 digital channels (DS1000D series), each channel can be tu rn ed on
or off independently, or in a 8 bit group
5.6 inch and 64 k TFT LCD Abundant trigger types: Edge, Pulse Width, Vi deo, Slope, Alternate, Pattern and
Duration Trigger (Only for DS1000D series)
Unique adjustable trigger sensitivit y enables to meet different demands Enabl e to meas ure 22 types o f wave paramete rs and track measurements via
cursor automatically
Unique waveform record a n d replay function Fine delayed scan function Built-in FFT function, hold practical digital filters
www.GlobalTestSupply.com
Page 8
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
VI
Provide 4 types of digital filters: LPF, HPF, BPF, BRF Pass/Fail detection function enables to output testing results Math operations available to multiple waves Powerful PC applic ation software UltraScope Standard configuration interface: USB Device, USB Host ,RS-232 and support
USB storage device storage and PictBridge print standards
The new function “Special Mode” can meet the needs of industrial production Support for remote command control Built-in help menu enables information getting more convenient Multiple Language User Interface, support Chinese & English input Can store files into USB storage device or the internal memory Waveform intensity can be adjusted To display a signal automatically by AUTO Pop-up menu makes it easy to read and use
www.GlobalTestSupply.com
Page 9
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
VII
Content
Safety Notices ......................................................................................... II
General-Purpose Oscilloscopes .................................................................. V
Chapter 1 Quick Start ........................................................................ 1-1
To Inspect the Instrument ..................................................................... 1-2
The Panels and User Interface ................................................................ 1-3
Appearance and Dimensions ................................................................... 1-7
To Perform a Functional Check ............................................................... 1-8
To Compensate Probes ........................................................................ 1-10
Digital Leads (Only for DS1000D Series) ................................................ 1-11
To Display a Signal Automatically .......................................................... 1-13
To Understand the Vertical System ....................................................... 1-14
To Understand the Horizontal System .................................................... 1-16
To Understand the Trigger System ........................................................ 1-18
Chapter 2 Operating Your Oscilloscope ............................................. 2-1
To Set up the Vertical System ................................................................. 2-2
To Set up the Horizontal System ........................................................... 2-27
To Set up the Trigger System ............................................................... 2-34
To Set up the Sampling System ............................................................ 2-53
To Set up the Display System ............................................................... 2-57
To Store and Recall ............................................................................. 2-59
To Set up the Utility System ................................................................. 2-67
To Measure Automatically .................................................................... 2-88
To Measure with Cursors ...................................................................... 2-95
To Use Run Control Buttons ............................................................... 2-102
Chapter 3 Application & Examples ..................................................... 3-1
Example 1: Taking Simple Measurements ................................................ 3-1
Example 2: View a Signal Delay Caused by a Circuit.................................. 3-2
Example 3: Capture a Single-Shot Signal ................................................. 3-3
Example 4: To Reduce the Random Noise on a Signal ............................... 3-4
Example 5: Making Cursor Measurements ................................................ 3-6
Example 6: The application of the X-Y operation ...................................... 3-8
Example 7: Triggering on a Video Signal ................................................ 3-10
Example 8: FFT Cursor measurement .................................................... 3-12
www.GlobalTestSupply.com
Page 10
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
VIII
Example 9: Pass/Fail Test .................................................................... 3-13
Example 10: Triggering on a Digital Signal ............................................ 3-14
Chapter 4 Troubleshooting ................................................................ 4-1
Chapter 5 Specifications .................................................................... 5-1
Specifications ....................................................................................... 5-2
General Specifications ........................................................................... 5-6
Chapter 6 Appendix ........................................................................... 6-1
Appendix A: Ac ce s s ories ........................................................................ 6-1
Appendix B: Warranty ........................................................................... 6-2
Appendix C: Care and Cleaning .............................................................. 6-3
Appendix D: Contact Us ......................................................................... 6-4
Index ...................................................................................................... 1
www.GlobalTestSupply.com
Page 11
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-1
Chapter 1 Quick Start
This chapter covers the following topics:
To Inspect the Instrument The Panels and User Interface Appearance and Dimension s To Perform a Functional Check To Compensate Probes Digital Leads (Only for DS1000D Series) To Display a Signal Automatically To Understand the Vert ical System To Understand the Horizontal System To U nderstand the Trigger Sy stem
www.GlobalTestSupply.com
Page 12
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-2
To Inspect the Instrument
When you get a new DS1000E, DS1000D series oscilloscope, please inspect the instrument according to the following steps:
1. Inspect the shipping container for damage.
Keep a damaged shipping container or cushioning material until the contents of the shipment have been checked f or completeness and the instrument has be en checked mechanically and electrically.
2. Inspect the instrument.
In cas e there is any mechanical damage or defect, or the instrument does not operate properly or fails performance tests, please notify the RIGOL Sales Representative.
If the shipping container is damaged, or the cushioning materials show signs of stress, please notify the carrier as well as the RIGOL sales off ice. Keep the shipping materials for the carrier’s inspection. RIGOL offices will arrange for repair or replacement at RIGOL’s option without waiting for claim settlement.
3. Check the accessories.
Accessories supplied with the instrument are listed in "Appendix A: Accessories" in this guide. If the contents are incomplete or damaged, please notify your RIGOL Sales Representative.
www.GlobalTestSupply.com
Page 13
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-3
The Panels and User Interface
The f irst thi ng to do with a new oscilloscope is to know its front panel. This chapter helps to be familiar with the layout of the knobs and buttons and how to use them. Read the chapter carefully before further operations.
1. Front Panel
Figure 1- 1, Front P anel; the knobs a re used most often and a re si milar to t he kno bs on other oscilloscopes. The buttons allow you to use some of the functions directly but also bring up s oft b ut ton men us on the screen, which enable the access to many measurement features associated with advanced functions, mathematics, and reference or to run control fe a tures.
The front panel of DS1000E:
The front panel of DS1000D:
Figure 1- 1
Front Panel Figure of DS1000E, DS1000D Series Oscilloscope
www.GlobalTestSupply.com
Page 14
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-4
Multi-function Common Menu Run Control
Knob Buttons Buttons
Trigger Control
Horizontal Control
Vertical Control
USB Host Logic Analyzer Port
Signal Input EXT Trigger Probe
Channel Input Compensation
Figure 1- 2
Front Panel Instruction
2. Rear Panel
Figure 1- 3
Rear Panel Figure of DS1000E, DS1000D Series Oscilloscope
www.GlobalTestSupply.com
Page 15
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-5
Figure 1- 4
Rear Panel Instruction
Rear Panel of DS1000E, DS10000D series contains the following parts:
Pass/Fail output port: The P ass/F ail testing results can be output by the port. RS232 port: To connect the oscilloscopes and external equipment as serial
port.
USB Device port: It is used to transfer data when external USB equipment
connects to the oscilloscope regarded as “Device equipment”. For examp le: to use this port when connect PictBridge printer to the instrument by USB.
Notation definitions in this User’s Guide:
Throughout this manual, notation symbols of buttons and knobs are the same to those on front-panel.
A box around the name of the button denotes MENU function buttons on
front-panel, such as Measure.
denotes the multi-function knob .
denotes the two POSITION knobs.
denotes the two SCALE knobs.
denotes the LEVEL knob.
The name with a drop shadow denotes the menu operating button, such as
Storage denoting the "Storage" menu item under Measure.
www.GlobalTestSupply.com
Page 16
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-6
3. User Interface
Figure 1- 5
Display screen (Analog cha nnels only)
Figure 1- 6
User Interface (Analog and Digital channels)
Waveform
display window
Menu
Channel 1
Channel 2
Running status
Location of
waveform window
in memory
Trigger point in waveform window
Trigger point in memory
Turn off
digital channels
Running status
Channel 1
Digital channels
Horizontal
time base div
Trigger
offset
Channel coupling
and vertical div
Turn on
digital channels
www.GlobalTestSupply.com
Page 17
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-7
Appearance and Dimensions
Figure 1- 7 Front View
Figure 1- 8 Side View
www.GlobalTestSupply.com
Page 18
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-8
To Perform a Functional Ch eck
Perform this quick functional check to verify that the instrument is operating correctly.
1. Turn on the instrument.
Use the power cord designed for the oscilloscope only. The power requirements are 100 to 240 VAC
RMS
, 45Hz to 440Hz.
T urn on the instruments, a nd wait until the display shows the waveform window. Press the Storage button, select Storage in the menu. Rotate the multiple
function knob to select Factory and press the knob down. Then, press Load .
Figure 1- 9
Turn on and Check the instrument
WARNING:
To avoid electric shock, be sure the oscilloscope is properly grounded.
2. Input a signal to a channel of the oscilloscope
DS1000E series: 2 channels input + 1 external trigger channel input DS1000D series: 2 channels input + 1 external trigger channel input +16 channels digital input
Do the following steps:
Set the switch on the probe to 10X and connect the probe to Channel1 on the
oscilloscope:
Power button
Storage button
!
www.GlobalTestSupply.com
Page 19
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-9
Align the slot in the probe connector with the button on the CH1 BNC. Push to connect, and twist to the right to lock the probe. Attach the probe tip and ground lead to the PROBE COMP connector.
Figure 1- 10
Attach the probe
Set the probe attenuation to 10X. To do this, push CH1Probe10X.
Figure 1- 11 Figure 1- 12 Set attenuation on the probe Set attenuation in the menu
Push the AUTO button. Within a few seconds, a square wave will display.
Push the CH1 button agai n to turn off Channel1.
Push the CH2 button to turn on channel2, repeat steps 2 a nd 3.
NOTE: The signal output from Pr obe compensator should only be used for probe compensation, not for calibration.
Probe compensator
Probe scale
www.GlobalTestSupply.com
Page 20
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-10
To Compensate Probes
Perform this adjustment to match the characteristics of the probe and the channel input. This should be performed whenever attaching a probe to any input channel for the first time.
1. From CH1 menu, set the Probe attenuation to 10X (press CH1Probe→10X). Set the switch to 10X on the probe and connect it to CH1 of t he oscilloscope. When using the probe hook-tip, inserting the tip onto the pr obe firmly to ensure a proper connection. Attach the probe tip to the Probe compens ator connector and the reference lead to the ground pin, Select CH1, and then press AUTO.
2. C he ck the shape of the displayed waveform.
Over compensated Correctly Compensated Under Compensated
Figure 1-9
Figure 1- 13
Probe Compensation
3. If necessary , use a non-metallic tool to a djust t he trimmer capacitor on the p robe
for the flattest square wave possible as displayed on the oscilloscope.
4. Repeat as necessary.
WARNING: To avoid electric shock while using the probe, be sure the perfection of the insulated cable, and do not touch the metallic portions of the probe head while it is connected with a voltage source.
!
www.GlobalTestSupply.com
Page 21
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-11
Digital Leads (O nly for DS1000D Series)
Digital leads are provided only f or DS1000D series which have Logic Analyzer.
1. Switch off power supply of the device under test if necessary to avoid short circuit. Since no voltage is applied to the leads at this step, you may keep the oscilloscope on.
2. Connect one end of the flat cable FC1868 to the Logic Analyzer Input; connect the other end to Logic Head LH1116. An identifier is located on each end of the flat cable; it can only be connected in one way. It is unnecessary to switch off power supply of your oscilloscope when connecting the cable.
Figure 1- 14
Connect the digital leads
CAUTION: Use only FC1868, LH1116, TC1100 and LC1150 made by RIGOL for specified DS1000D series.
3. Connect a test clip to one lead wire; m ake sure it’s connection good.
Figure 1- 15
Input Interface of Digital Channel
Test clip
!
www.GlobalTestSupply.com
Page 22
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-12
4. Test your device with the clip.
Figure 1- 16 Digital Channel Measurement
5. Remember to connect Ground Channel to the DUT’s ground terminal.
Figure 1- 17
Connecting to the Ground
GND
www.GlobalTestSupply.com
Page 23
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-13
To Display a S ignal Automatic ally
The oscilloscope has an automated feature to display the input signal best-fit. The input signal should be 50Hz or higher and a duty cycle is greater than 1%.
Use Auto Setting:
1. Connect a signal to the oscilloscope as described above.
2. Press AUTO.
The oscilloscope will automatically set up VERTICAL, HORIZONTAL and TRIGGER controls to display the input signal. Adjust the controls manually to get the best results if necessary.
www.GlobalTestSupply.com
Page 24
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-14
To Understand the Vertical System
Figure 1- 18 shows the VERTICAL controls, CH1, CH2, MATH, REF and OFF buttons and vertical , knobs. Following the exercise of the buttons, knobs, and the status bar to be familiar with the vertical parameters settings.
Figure 1- 18
The vertical system
1. Moves the signal vertically on the display with the
knob.
When turning the
knob, a voltage value is displayed for a short time indicating its value with respect to the ground reference located at the center of the screen. Also notice that the ground symbol on the left side of the display moves in conjunction with the
knob.
Measurement hints
If the channel is DC coupled, measuring the DC component of the signal by simply noting its distance from the ground.
If the channel is AC couple d, the DC component of th e signal is bl ocked, allow you to use greater sensitivity to display the AC component of the signal.
www.GlobalTestSupply.com
Page 25
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-15
Vertical offset back to 0 shortcut key
Turn the
knob to change the vertical display position of
channel and press the knob to set the vertical display position back to 0 as a shortcut key
, this is especially helpful when the t race positio n i s
far out of the s creen and want it to get back to the screen center im mediately.
2. Change the vertical setup and notice that each change affects the status bar differently.
View the status bar which is on the bottom of the screen to understand the
vertical scale.
Change the vertical scale by turning the
knob and notice the change
in the status bar.
Press CH1 , CH2 , MATH , REF , LA (only for DS1000D series) to display
the menu, sign, waveform o f corresponding channel. Press OFF button to turn off the channel.
Coarse/Fine Shortcut key
The Coarse/Fine vertical control can be set by simply
pressing the vertical
knob.
www.GlobalTestSupply.com
Page 26
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-16
To Understand the Horizontal System
Figure 1-19 shows the HORIZONTAL controls: MENU button, and
knobs of horizontal system. Following the exercise to familiarize with the
buttons, knobs, and status bar.
Figure 1- 19
The horizontal system
1. Turn the
knob and notice the change in the status bar.
The horizontal
knob changes the sweep speed in a 1-2-5 step sequence, and displays the value in the status bar. The time base rang es of th e oscilloscope is from 2ns/div* to 50s/div.
Delayed Scan Shortcut key
To press the
knob in the horizontal cont rol area on the fr ont -panel
is another way to enter or exit Delayed Scan mode and it is equal to the
following menu operations, MENUDelayedON.
* NOTE: The speed of horizontal scan varies by diff erent models.
www.GlobalTestSupply.com
Page 27
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-17
2. The horizontal
knob moves displayed signal horizontally
on waveform window
Turn the horizontal
knob to adjust trigger offset. The waveform
will move horizontally at the s ame time.
Horizontal offset back to 0 shortcut key
Press the knob to set the horizontal offset to 0 as a sho rtcut key
, this is especially he lpful w hen the t rigger point is far out of the screen a nd
want it to get back to the screen center immediately.
3. Press the MENU button to display the TIME menu.
To enter or exit the Delayed Scan mode, set the display to Y -T, X-Y or ROLL mode, or reset the trigger offset.
Horizontal position control
Trig-Offset: The position of the practical trigger point relative to cen ter of
the memory. Adjust the horizontal
knob
to move the trigger
point horizontally.
www.GlobalTestSupply.com
Page 28
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-18
To Understand the Trigger System
Figur e 1-20 shows the trigger control: MENU, 50%, FORCE and a trigger level knob. Following the exercise to familiarize with the buttons, trigger level knob and status bar.
Figure 1-20
The trigger control window
1. Turn the trigger Level knob and notice the changes on the display.
As you turn the
knob or pressing the 50% button, two things will happen
on the display for a short time.
First, the trigger level value is displayed at the bottom-left of the screen. Second, a line is displayed showing the location of the trigger level (as long as
AC coupling or low frequency reject are not selected).
Trigger Level to 0 Shortcut key
Turn the knob to change trigger level value and p r e ss t he knob to set tr i gger level back to 0 a s a sho rtcut key.
www.GlobalTestSupply.com
Page 29
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-19
2. Change the trigger setup and notice these changes in the sta tu s bar.
Press MENU button in the Trigger control. A soft button me nu appears on the display showing the trigger setting choices as shown in Fi g ure 1-21.
Figure 1- 21
· Press the trigger Mode button and choose Edge.
· Press t he trigger Source button to select CH1.
· Press the trigger Slope butto n to choose .
· Press the trigger Sweep button to select Auto.
· Press the trigger Set Up button to enter secondary menu.
NOTE: The trigger type, slope and source change i n conjun ction w ith the st atus bar
on the top-right of the screen.
3. Press 50%
The 50% button sets the trigger level to the center of the signal.
4. Press FORCE
Starting an acquisition regardless of an adequate trigger signal, usually used in “Normal” or “Single” trigger mode. This button has no effect if the acquisition is already stopped.
Key point: Holdoff:
A time interval before the oscilloscope response to next trigger
signal. During this holdoff period, the trigger system becomes “blind” to trigger signals. This function helps to view complex signals such as an AM waveform. Press Holdoff button to activate (
) knob, then turn it to adjust Holdoff time.
www.GlobalTestSupply.com
Page 30
www.GlobalTestSupply.com
Page 31
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-1
Chapter 2 Operating Your Oscilloscope
By now, a user should understand the VERTICAL, HORIZONTAL and TRIGGER control systems and knows how to determine the system setup by status bar of a DS1000E, DS1000D series digital oscilloscope.
This chapter will go through all groups of front-panel buttons, knobs and menus; and further the knowledge of the operation by hints in this gui de. It is recommended to perform all of the following exercises to get the most of the powerful measurement capabilities of the oscilloscope.
This chapter contains the following topics:
To Set up the Vertical System
( CH1, CH2, MATH, REF, LA (for DS1000D series) OFF, Vertical
,
Vertical )
To Set up the Horizontal System
( MENU, Horizontal
, Horizontal )
To Set up the Trigger System ( , MENU, 50%, FORCE) To Set up the Sampling Syst em ( Acquire) To Set up the Display System ( Display) To Store and Recall ( Storage) To Set up the Utility System ( Utility) To Measure Automatically ( Measure) To Measure with Cursors ( Cursor) To Use Run Control Buttons ( AUTO, RUN/STOP)
Please refer to this chapter more carefully to get more information of the measurement and operat ion methods of DS1000E and DS1 000D.
www.GlobalTestSupply.com
Page 32
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-2
To Set up the Vertical System
Settings of the Channels
This series oscilloscopes offer dual channels. Each channel has an operation menu and it will pop up after pressing CH1 or CH2 button. The settings of all items in the menu is sho wn in the table belo w.
Figure 2- 1 Table 2- 1 The Channel menu (Page 1/2)
Menu Settings Comments
Coupling
AC
DC
GND
Blo
cks the DC component of the
input Signal Passes both AC and DC components of the input signal Disconnects the input signal
BW Limit
ON
OFF
Limits the channel bandwidth to
20MHz to reduce display noise. Get full bandwidth.
Probe
1X 5X 10X 50X 100X 500X 1000X
Set this to match your probe
attenuation factor to make the
vertical scale readout correct
Digital filter
Setup digital filter (See table 2-4)
1/2
Go to the next menu page (The
followings are the same, no
more explanation)
www.GlobalTestSupply.com
Page 33
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-3
Figure 2- 2 Table 2- 2 The Channel Menu (Page 2/2)
Menu Settings Comments
2/2
Back to the previous menu page
(The followings are the same, no
more explanation)
Volts/Div
Coarse
Fine
Selects the resolution of the
knob Defines a 1-2-5 sequence. To c hange the resolution to small steps between the coarse settings.
Invert
ON OFF
Turn on the invert function. Restore
original display of the
waveform.
1. Channel coupling To use Channel 1 as an example, input a sine wave signal with DC shift.
Press CH1CouplingAC to set “AC” coupling. It will pass AC component and blocks the DC component of the input signal. The wavef orm is displayed as Figure 2­3:
Figure 2- 3
AC coupling setting
AC coupling
setup
AC coupling
status symbol
www.GlobalTestSupply.com
Page 34
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-4
Press CH1CouplingDC, to set “DC” coupling. It will pass both AC and DC components of the input signal. The waveform is displayed as Figure 2- 4:
Figure 2- 4
DC coupling setting
Press CH1CouplingGND, to set “ GND” coupling, it blocks the input signal. The screen displays as Figure 2-5:
Figure 2- 5
GND coupling setting
GND coupling
setup
GND coupling status symbo l
DC coupling
setup
DC coupling
status symbo l
www.GlobalTestSupply.com
Page 35
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-5
2. Set up the channel bandwid th limit
To take Channel1 as an example, input a signal containing high frequency component. Press CH1BW Lim i tOFF, to set up bandwidth limit to “OFF” status. The oscilloscope is set to f ull bandwidth and passing the high fre quency c omponent in the signal. The waveform is displ ayed a s
Figure 2- 6:
Figure 2- 6
Turn off the BW limit
Press CH1BW LimitON, to set up bandwidth limit to “ON” status. It will reject the frequency component higher than 20MHz.
Figure 2- 7
Turn on the BW limit
Turn off the BW limit
20M BW limit
Mark of
BW limit
www.GlobalTestSupply.com
Page 36
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-6
3. Probe Attenuation Setting
The oscilloscope allows selecting the attenuation factor for the probe. The attenuation factor changes the vertical scaling of the oscilloscope so that the measurement results reflect the actual voltage levels at the probe tip.
To change (or check) the probe attenuation setting, press the CH1 or CH2 button (according to which channel in using). Press the Probe so ft button to match the attenuation factor of the probe. This setting remains in effect until changed again.
Figure 2-8 shows an example for using a 1000:1 probe and its attenuation factor.
Figure 2- 8
Use the 1000:1 attenuation
Table 2- 3 Probe setting
Probe attenuation factors Corresponding settings 1:1 1X 5:1 5X 10:1 10X 50:1 50X 100:1 100X 500:1 500X 1000:1 1000X
Probe attenuation
Vertical volt/div.
www.GlobalTestSupply.com
Page 37
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-7
4. Digital Filter settings
DS1000E, DS1000D series offer 4 types of digital filters (Low Pass Filter, High Pass Filter, Band Pass Filter and Band Reje ct Filter). In order to achieve a good filter effect, you can filter the special waveform frequency by setting up the range of filter bandwidth. Press CH1Digital f ilter, display the digital filter menu. Turn ( ) knob to set high and low limit of frequency.
Figure 2- 9
The waveform when turning off digital filter
Figure 2- 10
The waveform when turning on digital filter
Turn off digital filter
Turn on digital filter
Mark of digital filter
www.GlobalTestSupply.com
Page 38
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-8
Figure 2- 11 Table 2- 4 The Filter menu
Menu Settings Comments
Digital Filter
ON OFF
Turn on the digital filter Turn off the digital filter
Filter Type
Setup as LPF (Low Pass Fil t er) Setup as HPF (High Pass Filter) Setup as BPF (Band Pass Filter) Setup as BRF (Band Reject Filter)
Upper limit
<frequency>
Turn (
) knob to set high limit
Lower limit
<frequency>
Turn (
) knob to set low limit
Back to higher level menu (The followings are the same)
5. Volts/Div settings
The Volts/Div control has Coarse or Fine configuration. The Vertical Sensitivit y is 2mV/div - 10V/div. Coarse: It is the default setting of Volts/Div in a 1-2-5-step sequence from 2mV/div, 5mV/div, 10mV/div, 20mV/div……10V/div. Fine: This setting changes the vertical scale to small steps between the coarse settings. It will be helpful to adjust the waveform in smooth steps.
Figure 2- 12 Fine configurations
Fine adjustment
Fine adjustment data
www.GlobalTestSupply.com
Page 39
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-9
Coarse/Fine Shortcut key: To change Coarse/Fine setting, not only by menu but also by pressing vertical
knob
6. To invert a waveform
Inver t turns the displayed waveform 180 degrees, as respect to the ground level. Figure 2- 13 and Fig ure 2- 14 show the changes after inversion.
Figure 2- 13 The waveform before inversion
Figure 2- 14 The wavef orm af ter inversion
Invert ON
Invert OFF
www.GlobalTestSupply.com
Page 40
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-10
Math Functions
The mathematic functions include “Add”, “Subtract”, “Multiply” and “FFT” for Channel 1 and Channel 2. The mathematic result can be measured by grid and cursor. Press MATH button, the system enter the MATH interface as follows:
Figure 2- 15
The Math function
Figure 2- 16 Table 2- 5 The Math menu
Menu Settings Comments
Operation
A+B
A-B
A×B
FFT
Add source A and source B
Subtract source B from source A Multiply source B by source A
Fast Fourier Transform
Source A
CH1 CH2
Define CH1 or CH2 as source A
Source B
CH1 CH2
Define CH1 or CH2 as source B
Invert
ON
OFF
Invert the MATH waveform. Restore to orig inal waveform display.
MATH scale
www.GlobalTestSupply.com
Page 41
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-11
1. Using the FFT
FFT (Fast Fourier Transform) transforms time domain signals to frequency domain signals. The horizontal axis shows the frequency and the vertical axis shows the amplitude dBVrms or Vrms. FFT function can not only to help you find crossfire problem and analog waveform distortion caused by n onlinear a mplifier, but also can adjust the analog filter.
FFT waveforms are useful in the following applications:
Measuring harmonic content and distortion in systems Characterizing noise in DC power supplies Analyzing vibration
Press MATHOperateFFT, to display FFT menu.
Figure 2- 17 Table 2- 6 The FFT menu
Menu
Settings Comments
Operate
A+B A-B A x B FFT
Add source A to source B Subtract source B from source A Multiply source B by source A Fast Fourier Transform
Source
CH1 CH2
Define CH1 or CH2 as FFT source
Window
Rectangle Hanning Hamming Blackman
Select window for FFT
Display
Split Full screen
Display FFT waveform on half screen Display FFT waveform on full screen
Scale
Vrms dBVrms
Set “Vrms ” as vertical unit Set “dBVrms ” as vertical unit
www.GlobalTestSupply.com
Page 42
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-12
2. Selecting an FFT Window
The oscilloscopes provide four FFT windows. Each window is a trade-off between freque ncy re so lut ion a nd am pl itud e ac curac y. What you want to me asur e and your source signals characteristics help determine which window to use. Use the following guidelines to select the best window.
Table 2- 7 FFT Windows
Window Features Best for measuring
Rectangle
Best frequency Resolution and worst magnitude resolution. This is essentially the same as no window.
Transients or bursts, the signal levels before and after the event are nearly equal. Equal-amplitude sine waves with fixed frequencies. Broadband random noise with a relatively slow varying spectrum.
Hanning Hamming
Better frequency, poorer magnitude accuracy than Rectangular.
Hamming has slightly
better frequency resolution than Hanning.
Sine, periodic, and narrow-band random noise. Transients or bursts where the signal levels before and after the events are significantly different.
Blackman
Best magnitu de, worst frequency resolution.
Single frequency waveforms, to Find higher order harmonics.
Key points for FFT
Signals that have a DC component or offset can cause incorrect FFT waveform
component magnitude values. To minimize the DC component, choose AC Coupli ng on the sour c e signal.
To reduce random noise and aliases components in repetitive or single-shot events, set the oscilloscope acquisition mode to average.
T o display FFT wa veforms with a large dynamic ran ge, use the dBVrms scale. T he
dBVrms scale displays component magnitudes using a log scale.
www.GlobalTestSupply.com
Page 43
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-13
Key points: FFT Resolution: the quotient between
sampling rate and number of FFT
points. With a fixed FFT points, the lower sampling rate can insure better resolution.
Nyquist Frequency
The highest frequency that any real-
time digitizing oscilloscope can acquire
with
out aliasing. It’s normally half of the sample rate. This frequency is called
the Nyquist frequency. Frequency above the Nyquist frequency will be under
sampled, causing a situation known as aliasing.
www.GlobalTestSupply.com
Page 44
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-14
REF Function
Reference Wa veforms are sav ed waveforms to be selected for display. The reference function will be av ailable after sa ving the selected w aveform to no n-volatile me mory .
Press REF button to display reference waveform menu.
Figure 2- 18 Table 2- 8 REF Menu when using internal memory
Menu Settings Comments
Source
CH1 CH2 MATH/FFT LA
Select channel1 as REF channel Select channel2 as REF channel Select MATH/FFT as REF channel Select LA as REF channel (DS1000D series)
Location
Internal External
Select mem ory location in scope
Select mem ory location out s cope Save Save REF waveform Imp./Exp. Go to import/export menu(see table 2-10) Reset Reset REF waveform
Figure 2- 19 Table 2- 9 REF Menu when using external memory
Menu Settings Comments
Source
CH1 CH2 MATH/FFT LA
Select channel1 as REF channel
Select channel2 as REF channel
Select Math/FFT as REF channel
Select LA as REF channel (DS1000D series)
Location
Internal External
Select internal memory location
Select external memory location Save
Save REF waveform to outer memory
location Import Go t o import menu(see table 2-14) Reset Reset REF waveform
www.GlobalTestSupply.com
Page 45
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-15
1. Import and Export
Press REF Imp./Exp. To enter the following menu.
Figure 2- 20 Table 2- 10 T he Imp. /Exp. menu
Menu
Settings Comments
Explorer
Path Directory File
Switch to Path, directory or file
Export
Export the REF file from internal
memory
to export memory (see
table 2-11)
Import
Import the REF file to internal
memory Delete File
Delete file
Figure 2- 21
Import /export interface
www.GlobalTestSupply.com
Page 46
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-16
2. Export
Press REFImp./Exp.Export to enter the following menu.
Figure 2- 22 Table 2- 11 The Export menu
Menu Settings Comments
Move the cursor up Move the cursor down To delete chosen letter
Save Execute the operation
Figure 2- 23
Export interface
www.GlobalTestSupply.com
Page 47
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-17
3. Save
Press REFSave to enter the following menu.
Figure 2- 24 Table 2- 12 The Save Menu
Menu Settings Comments
Explorer
Path Directory File
Switch among Path, Directory and File
New File (Folder)
Set up new file in Path and File.
Set up new folder in directory. Delete File(Folder)
Delete f il e(Folder)
Figure 2- 25
Save interface
www.GlobalTestSupply.com
Page 48
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-18
New File (or New Folder)
Press REFSaveNew File (or New Folder) and go to the following menu.
Figure 2- 26 Table 2- 13 The New File menu
Menu Settings Comments
Move the cursor up Move the cursor down To delete chosen letter
Save Execute the operation
Figure 2- 27
Chinese Input interface
Upper/Lower Case Switch
Switch Chinese/English
File name input box
www.GlobalTestSupply.com
Page 49
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-19
4. Import
Press REFImport and go to the following menu.
Figure 2- 28 Table 2- 14 The Import menu
Menu Settings Comments
Explorer
Path Directory File
Switch among Path, Directory and File
Import
Import the REF file into internal
memory
Figure 2- 29
Import interface
www.GlobalTestSupply.com
Page 50
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-20
5. Displaying a Reference Waveform
Figure 2- 30
Reference waveform display
1. Push REF button to show the reference waveform menu.
2. Press soft button No.1 to select the reference channel: CH1, CH2, MATH, FFT or LA (DS1000D series).
3. Turn vertical and vertical to adjust the REF waveform to a suitable position.
4. Press soft button No.2 to select the save location of REF wavefor m.
5. Press soft button No.3 to save the waveform as REF.
NOTE: The reference function is not available in X-Y mode.
www.GlobalTestSupply.com
Page 51
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-21
Set up LA Channel (Only for DS1000D Series)
16 digital channels logical analyzer are provided for DS1000D series, which can achieve mixed-signal measure in line with the dual analog channels. Single channel or group channels can be chosen ON or OFF, and also can set the size of waveform. Change display location of digital channel on screen and select threshold type.
Press LA funct ion butto n and go to the following menu.
Figure 2- 31 Table 2- 15 The LA m enu
Menu Settings Comments
D7-D0
Set up channel group D7-
D0 (see table
2-16)
D15-D8 set up channel group D15-D8 (see 2-17)
Current
<D15-D0>
Select channel by turning (
) knob
Threshold
TTL
CMOS
ECL
User
Select mode of whole digital channels.
The threshold voltage can set by user
when in user-defined style.
User
<Threshold
Voltage>
Set threshold voltage by turning (
)
knob.
1. Display and re-line up the digital channels (1) Press LAD7-D0 or D15-D8 and go to the g roup channel setting menu. Turn on
or turn off the display of the digital channels.
(2) Press LAcurrent and choose digital channel by turning ( ) knob. The chosen
channel will display in red color.
(3) Turn vertical
knob to re-position the channel in screen.
The figure of menu shows as follow.
www.GlobalTestSupply.com
Page 52
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-22
Figure 2- 32 Turn on the digital channel
2. Set threshold mode of digital channels
Press LAThreshold, select logic standard or User to define your own threshold voltage. The figure of menu shows as follow.
Figure 2- 33 Set the threshold
Threshold explanation
LOGIC STANDARD TTL CMOS ECL USER THRESHOULD VLOTAGE 1.4V 2.5V -1.3V -8V to 8V
Setup threshold
Setup data of
threshold by user
Select
running channel
Running channel
www.GlobalTestSupply.com
Page 53
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-23
Set up Channel Group
Press LAD7-D0 or D15-D8; Turn on/off the channel single, or in a group . Als o you can change the size of waveforms in 8 bits as a group. See table 2-1 6 and 2-17.
Figure 2- 34 Table 2- 16 The Digital Channel menu (Page 1)
Menu Settings Comments
channel D7-D0 Turn on or off single channel of D7-D0
D7-D0
Turn on Turn off
Turn on or off all 8 channels tog eth er
Size
Display 8 channels in a single screen Display 16 channels in a single screen
Reset Reset waveform of channel D7-D0
Figure 2- 35 Table 2- 17 The Digital Channel menu (Page 2)
Menu Settings Comments
channel D15-D8 Turn on or off single channel of D15-D8
D15-D8
Turn on Turn off
Turn on or off 8 channels together
Size
Display 8 channels in a single screen Display 16 channels in a single screen
Reset Reset waveform of channel D15-D8
www.GlobalTestSupply.com
Page 54
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-24
1. Turn on or off a digital channel
Press LAD7-D0→Channel, and choose the wanted c hannel by turning ( ) knob. Press No. 1 soft button or push down ( ) knob to t urn o n /off the channel. When
the channel is on, we can see the mark (
). When the channel is turned off; the
mark will display as (
).
As figure 2-36 shows.
Figure 2- 36
Turn on or off the digital channel
2. Force turn on or off all digital channels
Press LAD7-D0Turn On / Tu rn Of f (or D15-D8 Tur n O n / Tur n Off) will force to turn all the channe ls on /off. If you want to turn on/off any single channel i nstead, select the Channel by turning ( ) knob, then pre ss No. 1 soft butt on or ( ) knob.
3. Set up the viewing size of digital channels
Press LAD7-D0Size, or D15-D8Size, to select wave size of logic channels. Select to view 8 channels on the screen: Select to view all of the 16
channels on the screen.
4. Reset the digital channels display
Press LAD7-D0Reset, or D15-D8Reset to reset the display of digital channels.
On-off
channels
Channel list
and status of on-off
www.GlobalTestSupply.com
Page 55
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-25
Turn on/off Channels
The CH1, CH2, Ex t. Trigger and LA (DS1000D series) channels are input channels. All functionalities applied will be based on operating the instrument with channels. So MATH and REF can be reg a rded as relatively isolated channels.
To turn on/off any one of the channels, press the corresponding button on the front panel. The key backlight indicates the channel is currently active. Press th e button again to turn the channel off. Or when channel is currently selected, press OFF will turn the channel off as well, and the key backlight also goes off.
The channel status symbol is displayed at the lower-left of the screen, which can help users to judge the channel status quickly.
Table 2- 18 S ta tu s of the channels
Channel Mode Settings Status Indicator
Channel 1 (CH1)
ON Selected OFF
CH1 (yellow letter)
CH1 (black letter) No indicat or
Channel 2 (CH2)
ON Selected OFF
CH2 ( blue letter)
CH2 (black letter) No indicat or
MATH
ON Selected OFF
Math (purple letter)
Math (black letter) No indicat or
NOTE: Pressing LA will turn all the digital channels on/off.
www.GlobalTestSupply.com
Page 56
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-26
Set up Vertical Position and Scale
You can use the vertical controls to display waveforms, adjust vertical and
, and set input parameters.
1. Using vertical
knob.
The vertical
control changes the position of signal waveforms in all channels (including MATH and REF). The resolution chan ges accordin g to the vertical level set. Pressing this knob will clear the channel offset to zero. (The function is available for DS1000D series, but not includes digital channel.)
2. Using vertical
knob.
The vertical
can change the vertical sensitivity of waveforms in all channels (including MATH and REF, excluding LA). If the Volts/Div is set to “Coarse”, the wav eform scales in a 1-2-5 step sequence from 2 mV to 5 V. If the Volts/Div is set to “Fine”, it scales to small steps between the coarse settings.
3. Channels can be adjusted by the vertical
and only when
they ar e selected.
4. During the v erti cal position, a position message is displayed on the left bottom of the screen, in the same color as the corresponding channel. The unit is V (Volts).
www.GlobalTestSupply.com
Page 57
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-27
To Set up the Horizontal Sys t e m
The oscilloscope shows the time per division in the scale readout. Since all active waveforms use the same time base, the oscilloscope only displays one value for all the active channels, except when using Delayed Scan, or Alternate Trigger.
The horizontal controls can change the horizontal scale and position of waveforms. The horizontal center of the screen is the time reference for waveforms. Changing the horizontal scale causes the waveform to expand or contract about the screen center.
Horizontal position changes the displayed waveform position, relative to the trigger point. Press the horizontal MENU button to display the horizontal menu as follows.
Figure 2- 37 Table 2- 19 The Horizontal menu
The parameter status symbol will be displayed on the screen during horizontal system settings, which can help users to judge the channel status quickly.
Menu Settings Comments
Delayed
ON OFF
Turn on Delayed Scan mode Turn off the Delayed Scan mode
Time Base
Y-T
X-Y
Roll
Show the relative relation between
vertical voltage and horizontal
time.
Show CH1 value at X axis; CH2
value at Y axis. In Roll Mode, the wav eform display updates from right to left.
Sa Rate
Show system sample rate
Trig-offset Reset
Adjust to the cente r
www.GlobalTestSupply.com
Page 58
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-28
① ② ③
Figure 2- 38
Status bar and mark for Horizontal control
Marks Indicator:
The position of current waveform window in the memory The trigger position in the memory. The trigger position in the current waveform windo w s. The horizontal time base (main time base). The trigger’s horizontal offset according to the center of the window.
www.GlobalTestSupply.com
Page 59
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-29
Key Points Y-T: The conventional oscilloscope display format. It shows the voltage of a
waveform record (on the vertical axis) as it varies over time (on the horizontal axis).
X-Y: XY format displays channel 1 in the horizontal axis and channel 2 in the vertical axis.
Roll Mode: In this mode, the waveform display rolls from right to left. No trigger or horizontal offset control of waveforms is available during Roll Mode, and it’s only available when set to 500 ms/div or slower.
Slow Scan Mode: This mode is available when the horizontal time base is set to 50ms/div or slower. In this mode, the oscilloscope acquires sufficient data for the left part to the trigger point, then wait for trigger, when trigger occurs, it continues to draw the rest part fro m the trigger point to the end of the right side. When choosing this mode to view low frequency signals, it is recommended that the channel coupling be set as DC.
Time/Div: Horizontal scale. The Time/Div control expands or compresses the waveform.
www.GlobalTestSupply.com
Page 60
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-30
Delayed Scan
The Delayed Scan is a magnified portion of the main waveform window. Use Delayed Scan to locate and horizontally expand part of the main waveform window for a more detailed (higher horizontal resolution) analysis of signal. The Delayed Scan time base setting cannot be set slower than the Main time base setting.
Press hori zo ntal MENUDelayedON or press hori z ontal knob to enter Delayed Scan mode.
Figure 2- 39
Delayed Scan window
The screen splits into two parts. The upper half displays the main waveform wi ndo w and t he lower half displays an
expanded portion of the main wa veform window . Thi s expanded portion of the main window is called the Dela yed Scan win dow. Two blocks shaded at the upper hal f; the un-shaded portion is expanded in the lower half. The horizontal
and
knobs control the size and position of the Delayed Scan. The value at bottom of the scr een is the main time base and the value on the center bottom means the Delayed Scan time.
Waveform to be horizontally expanded
Expanded waveform in horiz o ntal
Main timebase
Time base of
Delayed Scan
www.GlobalTestSupply.com
Page 61
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-31
Use the horizontal
knob to change the position of the expan de d
portion.
Use the horizontal
knob to adjust the Delayed Scan resolution.
To change the main time base, turn off the Delayed Scan mode. Since both the main and Delayed Scan are displayed; there are half as many
vertical divisions so the vertical scaling is doubled. Notice the changes in the status bar.
Delayed Scan Shortcut Key
:
Delayed Scan function can be activated not only by menu but also by pressing
horizontal knob.
www.GlobalTestSupply.com
Page 62
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-32
X-Y Format
This format is useful for studying phase relationships between two signals. Channel 1 in the horizontal axis(X) and channel 2 in t he vertical a xis(Y), the oscilloscope uses a none-trigger acquisition mode, data is displayed as dots.
Press horizontal MENUTime BaseX-Y to enter the X-Y mo de .
Figure 2- 40
X-Y displa y f or mat
NOTE: In Y-T format, all sample rates are available. But in X-Y format, 100 MSa/s is not available. In common, deceasing the sample rate can display the wavefrom better.
The following modes or functions will not work in X-Y format.
LA Function (DS1000D series) Automatic Measurements Cursor Measurements REF and MATH Operations Delayed Scan Mode Vector Display Mode Horizontal
knob
Trigger Controls
www.GlobalTestSupply.com
Page 63
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-33
The Horizontal Knobs
:
The horizontal
knob adjusts the horizontal position of all channels (including Math) wavefo rms. The resol ution of this control va ries with the tim e base. Pressing this button clears trigger offset and moves the trigger point to the horizontal center of the screen.
:
Use
to select the horizontal time/div (scale factor) for the main or the time base of Delayed Scan. When Delayed Scan is enabled, it changes the width of the window zone by changing the Delayed Scan time base.
www.GlobalTestSupply.com
Page 64
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-34
To Set up the Trigger System
The trigger determines when the oscilloscope starts to acquire data and display a waveform. When a trigger is set up properly, it can convert unstable displays or blank screens into meaningful waveforms.
When the oscilloscope st arts to acquire a wave form, it collects enough data so that it can draw the w av eform to the left of the trigger point. The oscillosco pe continu es to acquire data while waiting for the trigger condition to occur. After it detects a trigger, the oscilloscope continues to a cquire enou gh data s o that it can dr aw the waveform to the right of the trigger point .
The trigger control area on the front panel includes a knob and three buttons:
: The knob that set the trigger level; press the knob and the level will
reset to zero.
50%: The instant execute button setting the trigger level to the vertical
midpoint between the peaks of the trigger signal
FORCE: Force to create a trigger signal and the function is mainly used in
Normal and Single mode
MENU: The button that activates the trigger controls menu. Press the Trigger MENU key to enter the trigger system setting interface.
Figure 2- 41
Trigger controls interface
www.GlobalTestSupply.com
Page 65
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-35
Trigger Modes
The oscilloscope provides seven trigger modes: Edge, Pulse Width, Slope, Vi deo, Alternate, Pattern (only for DS1000D series) and Dur ation trigge r (only for DS1000D series).
Edge: An edge trigger occurs when the trigger input passes through a specified voltage level at the specified slope direction.
Pulse Width: Use th is trigger type to catch pulses with certain pulse width. Slope: The oscilloscope begins to trigger according to the signal rising or falling
speed.
Video: Use video trigger on fields or lines for standard video signals. Alternate: Trigger on non-synchronized signals of the dual channels. Pattern: To Trigger through detecting a specified code. Duration: To trigger within a specified time on the conditions of a specified code.
www.GlobalTestSupply.com
Page 66
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-36
Edge Trigger
Edge Trigger Mode; An edge trigger determines whether the oscilloscope finds
the trigger point on the rising or the falling edge of a signal. Select Edge trigger Mode to trigger on Rising edge, falling edge or rising & falling edge.
Press Trigger MENUModeEdge to display the edge trigger menu as follows.
Figure 2- 42 Table 2- 20 The Edge Trigg er Menu
Menu Settings Comments
Source
CH1 CH2 EXT AC Line D15-D0
Select CH1 as trigger signal Select CH2 as trigger signal Select EXT TRIG as trigger signal Select power line as trigger signal Select a digital channel in D15-
D0 as
trigger sourc e (for DS1000D series)
Slope
Rising Falling
Rising &
Falling
Trigger on rising edge Trigger on falling edge Trigger on both ring & falling edge
Sweep
Auto
Normal Single
Acquire waveform even no trigger
occurred Acquire waveform when trigger occurred. When trigger occurs,
acquire one
waveform then stop
Set up To go to Set Up menu, s ee tabl e 2-38
www.GlobalTestSupply.com
Page 67
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-37
Pulse Width Trigger
Pulse Width Trigger Mode: Pulse trigger occurs according to the width of pulse.
The abnormal signals can be detected through setting up the p ulse wi dth condition.
Press Trigger MENUModePulse to display the pulse width trigger menu as follows.
Figure 2- 43 Table 2- 21 The Pulse Width Trigger Menu (page 1/2)
Menu Settings Comments
Source
CH1 CH2 EXT D15-D0
Selects CH1 as trigger signal Select CH2 as trigger signal Select EXT TRIG as trigger signal Select a
digital channel in
D15-
D0 as trigger source( Only
for DS1000D series)
When
(
+Pulse width
less than)
(+Pulse width
more than)
(+Pulse width
equal to)
(-
Pulse width
less than)
(-
Pulse width
more than)
(-
Pulse width
equal to)
To select pulse condition
Settings
<Width>
Set required pulse width
www.GlobalTestSupply.com
Page 68
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-38
Figure 2- 44 Table 2- 22 The Pulse Width Trigger Menu (page 2/2)
Menu Settings Comments
Sweep
Auto
Normal Single
Acquire waveform even no trigger
occurred Acquire waveform when trigger occurred. When trigger occurs,
acquire one
waveform and then stop
Set Up
To g o to Set Up menu, see table 2-38
NOTE: The Pulse width a djust range is 20ns ~ 10s. When the condition is met, it wi ll trigger and acquire the waveform.
www.GlobalTestSupply.com
Page 69
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-39
Slope Trigger
Slop Trigger Mode: Slope trigger sets the oscilloscope as the positive/negative
slope trigger within the specified time. Press Trigger MENUModeSlope to display the slope trigger menu as follows.
Figure 2- 45 Table 2- 23 The Slope Trigger Menu (Page 1/2)
Menu
Settings Comments
Source
CH1 CH2 EXT
Set channel 1 as trigger source Set channel 2 as trigger source Set EXT. channel as trigger source
When
To select the slope condition
Time
[1]
To set slope time
NOTE
[1]
:
Slope time can be set from 20ns to 10s. When a signal mee ts the trigger condition, The oscilloscope will execute the acquisition.
Figure 2- 46 Table 2- 24 The Slope Trigger Menu (Page2/2)
Menu
Settings Comments
Vertical
[2]
Select the
level that can be adjusted
by
Sweep
Auto
Normal
Single
Acquire waveform even when no trigger condition is met. Acquire waveform when trigger condition is met. When trigger condition is met, acquire one waveform and then stop
Set Up To go to set up menu. See t able 2-38
NOTE
[2]
:
You can adjust LEVEL A/ LEVEL B or both simultaneous by turning the knob.
www.GlobalTestSupply.com
Page 70
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-40
Video Trigger
Video Trigger Mode: Video trigger is used to capture the complicated waveforms
of television. The trigger circuitry detects the vertical and horizontal interval of the waveform and produces triggers based on the video trigger settings you have selected. Choose video trigger to trigger on fields or lines of NTSC, PAL/SECAM standard video signals. Trigger coupling preset to DC.
Figure 2- 47 Table 2- 25 The Video Tri gge r Men u (Pa ge 1/2)
Menu Settings Comments
Source
CH1 CH2 EXT
Selects CH1 as trigger source Select CH2 as trigger source Select EXT TRIG as trigger source
Polarity
positive
negative
Triggers on positive going sync pulses
Triggers on negative going sync pulses
Sync
All Lines Line Num
Trigger on all lines
Trigger on an specified line Odd field Even f iel d
Select to trigger on odd field
Select to trigger on even f iel d
Figure 2- 48 Table 2- 26 The Vid eo Trigger menu (Page 2/2, Sync is line)
Menu Settings Comments
Line Num
[1]
< Line sync >
Select the specified line number
for sync
Standard
PAL/SECM NTSC
Select Video sta ndard
Sweep
Auto
Normal Single
Acquire waveform even no trigger occurred
Acquire waveform when trigger
occurred.
When trigger occurs, acquire one
waveform and then stop
Set Up
To go to set
up menu, see table
2-39
www.GlobalTestSupply.com
Page 71
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-41
NOTE
[1]
: The line number ranges can be set from 1 to 525 under NTSC standard,
and from 1 to 625 under PAL/ SECAM.
Figure 2- 49 Table 2 - 27 The Video Menu (When the Sync is set as All lines, Odd
f ie ld and Even field)
Key points Sync Pulses: When positive polarity is selected, the trigger always occurs on
negative-going sync pulses. If the video signal has positive-going sync pulses, use the negative polarity selection.
Figure 2- 50 Figure 2- 51
Video Trigger: Line Synchronization Video Trigger: Field Synchronization
Menu Settings Comments
Standard
PAL/SECAM NTSC
Select Video sta ndard
Sweep
Auto
Normal Single
Acquire waveform even no trigger occurred
Acquire waveform when trigger
occurred.
When trigger occurs, acquire one
waveform and then stop
Set Up To g o to set up menu, see table 2-39
www.GlobalTestSupply.com
Page 72
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-42
Alternate Trigger
Alternate Trigger Mode: When alternate trigger is on, the trigger sources come
from two vertical channels. This mode can be used to observe two non-related signals. You can choose two different trigger modes for the two vertical channels. The options are as follows: Edge, Pulse, Slop e and video. The information of t he trigger level of the two channels will be displa y e d on the upper-right of the screen.
Press Trigger MENUModeAlternate to display the alternate trigger menu as follows.
Figure 2- 52 Table 2- 28 The Alternate menu (Trigger Type: Edge)
Menu
Settings Comments
Select
CH1 CH2
Set trigger mode for Channel 1 Set trigger mode for Channel 2
Type Edge
Set Edge Trigger
as the trigger
type
Slope
(Rising) (Falling)
Trigger on rising edge Trigger on falling edge
Set Up
To go to set up menu. See table
2-38
Figure 2- 53 Table 2- 29 (Trigger Type: Pulse, Page 1/2)
Menu
Settings Comments
Select
CH1
CH2
Set trigger mode for Channel 1 Set trigger mode for Channel 2
Type Pulse
Set Pulse Trigger for the channel
When
(+Pulse width less than)
(+Pulse width more than) (+Pulse width equal to)
(-Pulse width less than)
(-Pulse width more than) (-Pulse width equal to)
To select pulse condition
www.GlobalTestSupply.com
Page 73
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-43
Figure 2- 54 Table 2- 30 The Alternate menu (T rigger Type: Pulse, Page 2/2)
Menu Settings Comments
Setting
<
pulse width>
Set the width of the pulse
Set Up
To g o to set up menu. See tabl e 2-38
Figure 2- 55 Table 2- 31 The Alternate menu (Trigger Type: Slope, Page 1/2)
Menu Settings Comments
Select
CH1 CH2
Set trigger mode for C hannel 1 Set trigger mode for Channel 2
Type Slope
Set Slope Trigger for the vertical channel
When
Set trigger condition
Figure 2- 56 Table 2- 32 The Alternate menu (Trigger Type: Slope Page 2/2)
Menu Settings Comments
Time
<
Time Set >
Set slope time
Vertical
Select the level to be adjusted by
Set Up
To go to set up menu
. See table
2-38
www.GlobalTestSupply.com
Page 74
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-44
Figure 2- 57 Table 2- 33 The Alternate menu (Trigger Type: Video, Page 1/2)
Menu Settings Comments
Select
CH1 CH2
Set trigger mode for C hannel 1
Set trigger mode for Channel 2
Type Video Video Trigger f or the cha nnel
Polarity
positive
negative
Triggers on positive going sync pulses
Triggers on negative going sync pulses
Figure 2- 58 Table 2- 34 The Alternate menu (Video, Line Num Page 2/2)
Menu Settings Comments
Sync
ALL lines Line Num
Trigger on all lines
Trigger on an specified line Odd field Even f iel d
Select to trigger on odd field or even
field
Line Num
<
Lines Set >
Select the specified line number for sync
Standard
PAL/SECM NTSC
Select Video sta ndard
Set Up
To g o to set up menu, se e t able 2 -39
www.GlobalTestSupply.com
Page 75
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-45
Figure 2- 59 Table 2- 35 The Alternate menu (Trigger Mode: Video, All Lines, Odd
Field or Even Filed, Page 2 /2)
Menu
Settings Comments
Sync
ALL lines Line Num
Trigger on all lines
Trigger on an specified line Odd fie ld Even field
Select to trigger on odd field or even
field
Standard
PAL/SECM NTSC
Select Video sta ndard
Set Up
To g o to set up menu, see ta ble 2-39
www.GlobalTestSupply.com
Page 76
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-46
Pattern Trigger (Only for DS1000D Series)
Pattern Trigger Mode: Pattern trigger identifies trigger terms by checking
appointed code. The code is logical relationshi p of all channels, w ith high (H), low ( L) and ignore(X). Press Trigger MENUModePattern to display the patter n trigg er menu as follows.
Figure 2- 60 Table 2- 36 The Pattern Trigger Menu
Menu Settings Comments
Select
D15-D0 Choose digital channel for Pattern trigger
Code
H L X
High Low Ignore Rising Edge Falling Edge
Sweep
Auto Normal Single
Acquire waveform even no trigger occurred Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform
and then stop
Set Up
To go to set up menu, see table 2-40
Key Points:
1. H (High): Logic high: voltage is higher than threshold setting.
2. L (Low): Logic lo w: voltage is lower than threshold setting.
3. X (Ignore): Don’t car e. If all the channels are ignored, the oscilloscope won’t
be triggered.
4. Rising Edge (
) or Falling Edge ( ): Set the code as an edge of the
channel, rising edge or falling edge. When the edge is appointed, if code settings of other channels are all true, the oscilloscope will be triggered on the appointed edge. If no edge is appointed, the oscilloscope will be triggered on the la st edge whose code is true.
Edge of appointed code
You can only appoint one code as edge. If you have appointed an edge, then appointed another edge in a different channel , and the first appointed ed ge will be set to X (Ignore).
www.GlobalTestSupply.com
Page 77
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-47
Duration Trigger (Only for DS1000D Series)
Duration Trigger Mode: Trigger in appointed time when code ter ms are satisfied.
Press Trigger MENUModeDuration to display the duration trigger menu as follows.
Figure 2- 61 Table 2- 37 The Duration Trigger Menu (Page 1/2)
Menu Settings Comments
Select D15-D0 Choose digital channel for Duration Trigger
Code
H L X
High Low Ignore
Qualifier < >
=
Set time limit terms
Figure 2- 62 Table 2- 38 The Duration Trigger Menu (Page 2/2)
Menu Settings Comments
Time
<Time Setting>
Set duration and limit symbol time
Sweep
Auto Normal Single
Acquire waveform even no trigger occurred Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform
and then stop
Set Up To go to set up menu, see table 2-40
Key Points:
1. H (High): Logic high: voltage is higher than threshold setting.
2. L (Low): Logic lo w: voltage is lower than threshold setting.
3. X (Ignore): Don’t car e. If all the channels are ignored, the oscilloscope won’t
be triggered.
4. Qualifier: A timer begins when code terms are satisfied. Duration trigger occurred in the time set by the qualifier.
www.GlobalTestSupply.com
Page 78
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-48
Trigger Setup
Set up different trigge r sett ings ac cording to dif ferent trigge r modes . When choosing source as D15-D0 in the mode of Edge and Pulse (DS1000D series), only Holdoff is adjustable. When source is non-digital channel and in slope trigger, only trigger coupling, trigger sensitivity and Holdoff can be set. For video trigge r, Sensitivity and Holdoff can be set. For the pat tern tri gger and the durat ion trigge r (DS1000D series), only Holdoff is adjustable.
Figure 2- 63 Ta b l e 2 - 39 The Trigger Set Up Menu (Settings for trigger coupling,
trigger sensitivity and holdoff)
Menu Settings Comments
Coupling
DC AC HF Reject LF Reject
Allow all signals pass Block DC signals Reject high f requency signals Reject DC and low frequency signals
Sensitivity
<Sensitivity Setting>
Set trigger sensitivity
, the
range is from 0.1div to 1div
Holdoff
<Holdoff Setting>
Set time slot before another trigger event
, the range is
from 500ns to 1.5s
Holdoff
Reset
Reset Holdoff time to 500ns
Figure 2- 64 Ta b le 2 - 40 The Trigger Set Up Menu (Settings for sensitivity and
holdoff)
Menu Settings Comments
Sensitivity
<Sensitivity Setting>
Set trigger sensitivity
Holdoff
<Holdoff Setting>
Set time slot before
another trigger event
Holdoff
Reset
Reset Holdoff time to
500ns
www.GlobalTestSupply.com
Page 79
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-49
Figure 2- 65 Table 2- 41 T he Trigger Set Up Menu (Settings only for holdoff)
Menu Settings Comments
Holdoff
<Holdoff Setting>
Set time slot before another
trigger event
Holdoff Reset Reset Holdoff time to 500ns
www.GlobalTestSupply.com
Page 80
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-50
Trigger Holdoff
Trigger Holdoff can stabilize complex waveform, such as the pulse range. Holdoff time is the oscilloscope’ s w aiting period bef ore starting a new trigger. During Holdoff, oscilloscope will not trigger until Holdoff ends. For instance: To trigger on the first pulse on a group of them, users can set the holdoff time to Pulse cluster width.
Figure 2- 66 Trigger Holdoff display
To use trigger Holdoff:
1. Press the trigger MENU key to display Trigger Menu.
2. Press Set Up key to display trigger set up menu.
3. Turn the multi function knob ( ) to change Holdoff time until waveform is stable.
4. Press Reset can reset the Holdoff time to its default value.
Time of holdoff
Trigger holdoff
Trigger site
www.GlobalTestSupply.com
Page 81
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-51
Trigger Key points
1. Trigger Source: Trigger can occur from several sources: Input channels (CH1
and CH2), AC Line, Ext.
CH1 or CH2: It is the most commonly used trigger sou rce . The channel works
when selected as a trigger source whatever displayed or not.
Ext Trig: The instrument can be triggered from a third source while acquiring
data from CH1 and CH2. For example, to trigger from an external clock or with a signal from another part of the test circuit. The Ext trigger sources use an external trigger signal con nected to the EXT T RIG connector. Ext uses the signal directly; it has a trigger level range of -1.2V to +1.2V.
AC Line: AC power can be used to display signals related to the power line
frequency, such as lighting equipment and power supply devices. The oscilloscope gets triggered on its AC power input; an AC trigger signal i s not required. When AC Line is selected as trigger source, the oscilloscope automatically set coupling to DC, set trigger level to 0V.
2. Sweep Mode: The sweep mode determines how the oscilloscope behaves in the abse nce of a tr igger event. The oscilloscope provides three trigger modes: Auto, Normal, and Single.
Auto: This sweep mode allows the oscilloscope to acquire waveforms even
when it does not detect a trigger condition. If no trigge r condition occurs while the oscilloscope is waiting for a specific period (as determined by the time-base setting), it will force itself to trigger.
When forcing invalid triggers, the oscilloscope cannot synchronize the waveform, and the waveform seems to roll across the display. If valid triggers occur, the display becomes stable on the screen.
Any factor results in the un-stability of waveforms can be detected by Auto Trigger, such as the output of Power supply.
NOTE: When horizontal control is set under 50 ms/div, Auto mode allows the oscilloscope not to capture trigger signal.
Normal: The Normal mode allows the oscilloscope to acquire a wavef orm only
when it is triggere d. If no trig ger occurs, the os cilloscope keeps waiting, and the
www.GlobalTestSupply.com
Page 82
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-52
previous waveform, if any, will remain on the display.
Single: In Single mode, after pressing the RUN/STOP key, the oscilloscope
waits for trigger. While the trigger occurs, the oscilloscope acquires one waveform then stop.
3. Coupling: Trigger coupling determines which signal component passing to the trigger circuit. Coupling types include AC, DC, LF Reject and HF Reject.
AC: AC coupling blocks DC components and attenuates the signal below 10Hz. DC: DC coupling passes both AC and DC components. LF Reject: LF Reject coupling blocks DC component, and attenuates all
signal with a frequency lower than 8 kHz.
HF Reject: HF Reject coupling attenuates all signals with a frequency higher
than 150 kHz.
4. Pre-trigger/delayed trigger: The data collected before and after trigger.
The trigger position is typically set at the horizontal center of the screen. In the full-screen display the 6div data of pre-trigger and delayed trigger can be surveyed. Mor e data of pr e-trigger and 1s delayed trigger can be surveyed by adjusting the horizontal
knob.
This feature is very useful to study th e events that led up to the trigger point. Everything to the right of the trigger point is called post-trigger informati on. Th e delay range (pre-trigger and post-trigger information) depends on the sweep speed selected.
5. Adjustable trigger sensitivity: To avoid the influence of noise from the
physical world, and get the stable trigger, the trigger circuit has adopted Stickiness. In DS1000E, DS1000D series, the stickiness is adjustable from
0.1div-1.0div, which means when it sets to 1.0div, the trigger circuit will not affect a ny sig na l wit h pe ak-peak amplitude less than 1.0div, so as to avoid the influence of the noise.
www.GlobalTestSupply.com
Page 83
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-53
To Set up the Sampling System
Acquire shows the menu button for the acquire system on the front panel.
Figure 2- 67
Acquire setup button Press the Acquire button, the interface menu as follows: Figure 2- 68 Table 2- 42 The Acquire menu
NOTE
[1]
: This function is used in “Average Acquisit ion”;
NOTE
[2]
: This function is used in “Real Time” sampling.
The waveform display will be different when selecting different acquisition and sampling:
Menu Settings Comments
Acquisition
Normal Average Peak Detect
Normal Acquisition mode Average Acquisition mode Peak Detect Acquisition mode
Averages
[1]
2 to 256
Step by 2 times the power of N. Set average times from 2 to 256
Sampling
Real-Time Equ-Time
Real-time sampling mode Equivalent sampling mode
Mem Depth
Long Mem Normal
Set up memory as 512kpts or 1Mpts Set up memory as 8kpts or 16kpts
Sinx/x
[2]
ON OFF
Set the insert mode to sinx/x Set the insert mode to line
Acquire setup button
www.GlobalTestSupply.com
Page 84
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-54
Select Real-time acquisition to observe the single-shot or pulse signals. Select Equ-Time to observe high frequency repetitive signals. To reduce the displayed random noise, select Average Acquisition. And this
mode would make the screen refresh slower.
To Avoid signal aliasing, select Pea k Detect Acquisition.
The waveform displaye d on t he screen wil l change i n conjunctio n with the sett ing of Acquire menu.
Figure 2- 69
Signal that contains noise, and without average sampling
Figure 2- 70
Display signal after average sampling
www.GlobalTestSupply.com
Page 85
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-55
Figure 2- 71
Signal without Peak Detect Acquisition
Figure 2- 72
Signal with Peak Detect Acquisition
Stop Acqu isition: When the s cope is a cquiring w av eforms, the wa vefor ms is in a live status; when acquisition is stopped, frozen waveform will be displayed, the position and scale can still be adjusted by vertical control and horizontal control.
www.GlobalTestSupply.com
Page 86
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-56
Key Points
Real-time Sampling: The oscilloscope has Real-
time sampling rate up to
1GSa/s. At the time base 5
0ns or faster, the oscilloscopes use the sine(x)/x
interpolation to expand the horizontal time base.
Equivalent Sampling: Known as Repetitive sampling to get up to 4
0ps of
horizontal resolution (equivalent 25
Gsa/s). This mode is good for observing
repetitive signals, and it is not recommended for single-shot or pulse.
Normal Acquisition: Oscilloscope acquires signal by equal time interval. Average Acquisition: Apply averaging to your signal to remove uncorrelated
noise an d improve measur e ment acc uracy. Reduces random or u ncorrelated noise
in the signal displa y. The averaged w av ef orm is a run ning a v er a ge o ver a specified
number of acquisitions from 2 to 256.
Peak Detect Acquisition:
Peak Detect mode captures the maximum and
minimum values of a signal. Finds highest and lowest record points over many acquisitions.
www.GlobalTestSupply.com
Page 87
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-57
To Set up the D isplay System
Display shows the menu button for the display system on the front panel.
Figure 2- 73
Display setup button Press the Display button to pop up the menu for the settings of the display system.
Figure 2- 74 Table 2- 43 The Display menu (Page 1/2)
Menu Setting Comments
Type
Vectors Dots
Display waveforms as vectors Display waveforms as dots
Clear
Clear all existing waveforms from screen
persist
Infinite
OFF
The sample points remain displayed until turn the persistence “OFF”. Turn off the persistence function
Intensity
<
percentage >
Set up waveform intensity, the range is from 0% to 100%
Display setup button
www.GlobalTestSupply.com
Page 88
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-58
Figure 2- 75 Table 2- 44 The Display menu (Page 2/2)
Key points:
1. Display type: Display type includes Vector and Dot. In vectors type, the
oscilloscopes connect dots through digital interpolation including both linearity and sin(x)/x. Sin(x)/x interpolation is s uita ble f or R ea l-time sampling an d will b e more effective at 50ns or faster time base.
2. Refresh rate: It is an important performance of digital oscilloscopes. It means the number of display refreshing per second and it will affect the ability to observ e signal.
Menu Settings Comments
Grid
Display grids and coordin
ates on
the screen Turn off the grids Turn off the grids and coordinates
Brightness
<
percentage >
Set up grid bri gh t n e s s
Menu Display
1s 2s 5s 10s 20s Infinite
Set the time before menu fades away. T
he menu will be hidden
after the set time of last button pressing.
Adjusting waveform intensity
Default setup of multi-function knob (
) is adjusting wavef orm intensity.
www.GlobalTestSupply.com
Page 89
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-59
To Store and Recall
Storage shows the menu button for the storage system on the front panel.
Figure 2- 76
Storage setup button
Press the Storage butto n to show the menu for the settings of the storage system. Waveforms and setups can be stored in and recalled from , both internal memory and external mem ory. The wavef o rm file, setup file, bitmap and CSV file in external memory can be created and deleted (Note: internal file can be deleted or overwritten). System supports English/Chinese key in.
Waveform and setup, the menu as following: Figure 2- 77 Table 2- 45 The Storage menu (For Waveform and Setups)
Menu Settings Comments
Storage
Waveform Setups
Store or recall waveform or instrument setups
Internal
Go to menu for internal memory operation (see table 2-48)
External
Go to menu for e xtern al m emory operation (see table 2-49)
Disk Mana.
Go to disk manage
menu (see
table 2-50)
Storage setup button
www.GlobalTestSupply.com
Page 90
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-60
Figure 2- 78 Table 2- 46 The Storage menu (For Bitmap)
Menu Settings Comments
Storage Bit map Create or delete bit map files
Para Save
ON OFF
Save the current oscillos cope
settings in different format
with the same file name
External
Go to menu for external
memory operat i on(see table
2-49)
Disk Mana.
Go to disk manage menu (see table 2-50)
Figure 2- 79 Table 2- 47 The Storage menu (For CSV)
Menu Settings Comments
Storage CSV Create or delete CSV files
Data Depth
Displayed
Maximum
Save currently displayed
waveform data to CSV file Save the whole waveform data
in memory to CSV file
Para Save
On Off
Save the current oscilloscope
settings in different format with
the same file name
LA Data Type
[1]
Hex
Decimal
Binary
Save data in hexadecimal format
Save data in decimal format
Save data in binary format
External
Go to menu for exte rnal me mory operation (see table 2-49)
Disk Manage
Go to disk manage menu (see table 2-50)
NOTE
[1]
: Applicable to DS1000D Series only.
www.GlobalTestSupply.com
Page 91
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-61
Figure 2- 80 Table 2- 48 The Storage menu ( For factory)
Menu Settings Comments
Storage Factory Recall factory setups Load Recall facto r y setups or files
Disk Mana.
Go to disk manage menu (see table 2-50)
www.GlobalTestSupply.com
Page 92
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-62
Internal Memory
Press StorageInternal to enter the following menu.
Figure 2- 81 Table 2- 49 Th e Int e rnal Memory Menu
Menu Settings Comments
Internal
Int_00 (S) . . . Int_09 (N)
Set up the location of files in
internal memory
Load
Recall wavef orm files and setup files
from the internal memory
location
Save
Save waveform files and setup files
to the internal memory
location Delete File (Folder)
Delete the selected location f il e
(folder)
www.GlobalTestSupply.com
Page 93
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-63
External Memory
Press StorageExternal to enter the following menu.
Figure 2- 82 Table 2- 50 T he Memory menu
Menu
Settings Comments
Explorer
Path Directory File
Switch among Path, Directory and File
New File (Folder)
To create new file or folder.
Delete File
(Folder)
Delete the selected location file (folder)
Load
Recall waveform and setup from USB storage device.
Figure 2- 83
File system display
Files
File information.
Current directory
Back to previous folder
Path
www.GlobalTestSupply.com
Page 94
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-64
Disk Manage
Press StorageDisk Mana to display the following menu.
Figure 2- 84 Table 2- 51 The Disk Manage Menu (Page 1/2)
Menu Settings Comments
Explorer
Path Directory File
Switch among Path, Directory and File
New folder
To create new folder( same as new f il es, se e table 2-13)
Delete File Delete file
Load
Recall waveform, setup
, recorded
waveform, P ass / Fail file
Figure 2- 85 Table 2- 52 The Disk Manage Menu (Page 2/2)
Menu Settings Comments
Rename To rename a file (see table 2-52)
Disk info Display disk information
www.GlobalTestSupply.com
Page 95
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-65
Rename
Press StorageDisk Manage.Rename to go to the following menu. Figure 2- 86 Table 2- 53 The Rename Menu
Menu Settings Comments
To move the cursor up To move the cursor down To delete chosen letter
OK Rename the file
Figure 2- 87
Rename the file
Key Points: Factory: The oscilloscope has default settings and can be recalled at anytime.
Memory location: Specify memory location to save/recall the waves and setups. Load: Recall saved wave forms, setups and default settings. Save: Save waveforms and setups.
www.GlobalTestSupply.com
Page 96
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-66
NOTE:
1. Select Save stores not only the waveforms, but also the current settings.
2. To ensure the setups being saved properly, only after the settings are changed for more than 5 seconds can the user turn off the instrument. The oscilloscope can store 10 settings permanently and can restore at anytime.
www.GlobalTestSupply.com
Page 97
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-67
To Set up the U tility System
Utility shows the menu button for the Utility on the front panel.
Figure 2- 88
Utility setup button
Press the Utility button to pop up the menu for the settings of the Utility system.
Figure 2- 89 Table 2- 54 The Utility Menu (Page 1/3)
Menu Setting Comments
I/O setting Setup I/O configuration
Sound
(ON)
(OFF)
Switch beeper sound on/off
Counter
OFF ON
Turn off Frequency Counter Turn on Frequency Counter
Language
简体中文 繁体中文
English 日本語
François
Select languages
(More lan guages may be adde d in l ater
firmware versions)
Utility setup button
www.GlobalTestSupply.com
Page 98
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-68
Figure 2- 90 Table 2- 55 The Utility menu (Page 2/3)
Menu Settings Comments
Pass/Fail Setup Pass/Fail test Record Setup Waveform Recorder Print set Setup printing
Figure 2- 91 Table 2- 56 The Utility menu (Page 3/3)
NOTE: Self-Calibration: Oscilloscope will calibrate parameter of vertical system (CH1,
CH2, and Ext), horizontal system and trigger system.
Menu
Settings
Comments
Preference Go to preference menu Self-Cal Execute Self-calibration
System Info Display system inf ormation
Special Mode Lock/Unlock Lock the keyboard
www.GlobalTestSupply.com
Page 99
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-69
I/O setting
Press UtilityI/O setting to go to the following menu.
Figure 2- 92 Table 2- 57 The I/O Setup Menu
NOTE
[1]
: Users must set the USB device type manually first to ensure the
equipment connected is the same as the settings when switch the device from computer (PictBridge printer) to P ictBridge printer (computer) by the USB Device port.
Menu
Display Comments
RS-232 Baud
300 . .
38400
Set RS-232
baud rate as
300, 2400, 4800, 9600, 19200 or 38400.
USB Device
[1]
Computer PictBridge
Set the USB device
GPIB# 0-30 Set the GPIB address
www.GlobalTestSupply.com
Page 100
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-70
Language
The oscilloscopes have multi-language user menu, choose as your desire.
Press UtilityLanguage to select the language.
Figure 2- 93
The Language menu
Please use the fourth menu button or use the knob
to select the language you
need.
www.GlobalTestSupply.com