Rigol DP1308A User’s Guide

RIGOL
User’s Guide
DP1308A Programmable
Linear DC Power Supply
May 2009
RIGOL Technologies, Inc.
RIGOL
I
Copyright
© 2009 RIGOL Technologies, Inc. All Rights Reserved.
Trademark Information
RIGOL is registered trademark of RIGOL Technologies, Inc.
Notices
RIGOL products are protected by patent law in and outside of P.R.C.. RIGOL Technologies, Inc. reserves the right to modify or change parts of or all
the specifications and pricing policies at company’s sole decision.
Information in this publication replaces all previously corresponding material. RIGOL shall not be liable for losses caused by either incidental or consequential
in connection with the furnishing, use or performance of this manual as well as any information contained.
Any part of this document is forbidden to copy or photocopy or rearrange without
prior written approval of RIGOL.
Product Certification
RIGOL guarantees this product conforms to the standards of national and industrial.
Meanwhile, the related standards conform to other ISO will get further. At present, DP1308A has passed CE, cTUVus and LXI certification.
Users Guide for DP1308A
RIGOL
II
Safety Notices
Review the following safety precautions carefully before operating the instrument to avoid any personal injuries or damages to the instrument and any product connected to it. To prevent potential hazards, do use the instrument specified by this user’s guide only.
Use Proper Power Cord.
Only the power cord designed for the instrument and authorized by local country could be used.
Ground the Instrument.
The instrument is grounded through the grounding conductor of the power cord. To avoid electric shock, the instrument grounding conductor(s) must be grounded properly before connecting the input or output terminals.
Connect Output Wire Correctly.
Please connect the output wire in strict accordance with operation methods in this guide.
Observe all Terminal Ratings.
To avoid fire or shock hazard, observe all ratings and marks on the instrument. Consult your manual for more information about ratings before connecting.
Do Not Operate Without Covers.
Do not operate the instrument with
covers or panels removed.
Use Proper Fuse.
Please use the fuse with specified voltage and current rating.
Avoid Circuit or Wire Exposure.
Do not touch exposed connections and components when power-on.
Do not Operate with Suspected Failures.
If suspected damage occurs to the instrument, have it inspected by qualified service personnel before further operations.
Keep Well Ventilation.
Inadequately ventilated will cause the temperature rises or damages to the device. Please keep well ventilation and inspect the intake and fan regularly.
Do not Operate in Wet/Damp Conditions.
In order to avoid short circuit to the interior of the device or electric shock, please do not operate in a humid environment.
Do not Operate in an Explosive atmosphere.
In order to avoid damages to the device or personal injury, please operate far away from an explosive atmosphere.
Users Guide for DP1308A
RIGOL
III
Keep Product Surfaces Clean and Dry.
In order to prevent the performance of the device from influencing by dust or water in air, please keep the surface of device clean and dry.
Handling Safety
Please handle with care during transporting to avoid damages to buttons, and knobs or interfaces as well as other parts on the panel resulting from the falling from your hand.
Users Guide for DP1308A
RIGOL
IV
!
!
Protective Earth Terminal
Hazardous Voltage
Refer to Instructions
Test Ground
Chassis Ground
!
Safety Terms and Symbols
Terms in this Guide. These terms may appear in this manual:
WARNING
Warning statements indicate the conditions or practices that could result in injury or loss of life.
CAUTION
Caution statements indicate the conditions or practices that could result in damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury or hazard that may immediately happen. WARNING indicates a potential injury or hazard that may happen. CAUTION indicates that a potential damage to the instrument or other property
might occur.
Symbols on the Product. These symbols may appear on the product:
Users Guide for DP1308A
RIGOL
V
DP1308A Overview
DP1308A programmable linear DC power supply is designed with high performance and 80W triple-output. It has clear user interface, excellent specifications and multiple standard interfaces, which can meet diversified testing demands.
Main Features:
4.3 inch, 16M true color TFT LCD; Individual and controllable triple-channel enables to output power up to 80W in
total;
Low ripples noise, the output noise and ripples<350μVrms/2mVpp; Excellent load regulation rate and linear regulation rate: <0.01%+2mV (voltage),
<0.01%+250μA (current);
To implement effective protection for loads through setting up the corresponding
parameters based on overvoltage/overcurrent protection function;
Two-level over-temperature protection; Store and recall system parameters; Timing output; Waveform display: enable to display voltage/current waveform in real time and
make the state of instrument understand fully at one glance coordinating with digital voltage, current and power value;
Standard interfaces: USB Device, USB Host, LAN and GPIB; Support U disk store; Support remote control via Web; Support remote control via commands; Conform to LXI-C class instrument standard; One button help function; Support Chinese & English interface, input method.
Users Guide for DP1308A
RIGOL
VI
Hint:
For the newest edition of this manual please go to
http://www.rigolna.com/download_manual.aspx to download.
Structure of This Document
Chapter 1 Quick Start
This chapter introduces you the panels and display interface of DP1308A as well as some considerations when first use.
Chapter 2 Front Panel Operations
This chapter gives detailed information on function and operating methods about each button of the front panel.
Chapter 3 Troubleshooting
This chapter offers some methods to help you solve the problems during operations.
Chapter 4 Specifications
This chapter lists the specifications of DP1308A power supply.
Chapter 5 Appendix
Information about accessories, warranties, services and supports and the like.
Conventions about how to denote the buttons and menu softkeys from front panel in this document:
Buttons: button characters + textbox, for example: +6V denotes the selection
button of + 6V channel.
Menu softkeys: Menu character+ character shading, for example: Volt denotes the voltage option in Channel settings.
Users Guide for DP1308A
RIGOL
VII
Contents
Safety Notices ...................................................................................... II
DP1308A Overview ....................................................................................... V
Chapter 1 Quick Start ............................................................................ 1-1
General Inspection .................................................................................... 1-2
Front/Rear Panels and User Interface .......................................................... 1-3
Front Panel ......................................................................................... 1-3
Rear Panel.......................................................................................... 1-5
User Interface .................................................................................... 1-6
First Use DP1308A ..................................................................................... 1-7
Power Requirement ............................................................................. 1-7
Power-On Inspection ........................................................................... 1-7
Output Inspection ............................................................................... 1-8
Display Modes ......................................................................................... 1-11
Chapter 2 Front Panel Operations ......................................................... 2-1
Parameter Input ........................................................................................ 2-2
Constant Voltage Output ............................................................................ 2-4
Constant Current Output ............................................................................ 2-8
Timing Output ......................................................................................... 2-12
Multiplexed Output .................................................................................. 2-14
Tracking Output ...................................................................................... 2-16
Save and Recall ....................................................................................... 2-18
Set up the Utility ..................................................................................... 2-21
I/O Setting ....................................................................................... 2-22
System Setting ................................................................................. 2-24
Calibrate .......................................................................................... 2-27
Timer Setting ................................................................................... 2-28
Online Help System ................................................................................. 2-29
Key Lock ................................................................................................. 2-30
Chapter 3 Troubleshooting .................................................................... 3-1
Chapter 4 Specifications ........................................................................ 4-1
Chapter 5 Appendix ............................................................................... 5-1
Appendix A: DP1308A Accessories .............................................................. 5-1
Appendix B: Warranty ................................................................................ 5-2
Users Guide for DP1308A
RIGOL
VIII
Appendix C: General Care and Cleaning ...................................................... 5-3
Appendix D: Contact RIGOL ....................................................................... 5-4
Index ........................................................................................................... 1
Users Guide for DP1308A
Chapter 1 Quick Start RIGOL
1-1
Chapter 1 Quick Start
This chapter introduces you the panels and display interface of DP1308A as well as some considerations when first use. From this part, you can learn how to use this instrument quickly and correctly.
Following topics are mainly throughout this chapter:
General Inspection
Front/Rear Panels and User Interface
Front Panel Rear Panel User Interface
First Use DP1308A
Power Requirement Power-On Inspection Output Inspection
Display Modes
Users Guide for DP1308A
RIGOL Chapter 1 Quick Start
1-2
General Inspection
Please inspect your new DP1308A programmable linear DC power supply according to the following steps:
1 Inspect the shipping container for damage.
Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically. If the shipping container is damaged, or the protective material shows signs of stress, notify the carrier as well as your RIGOL sales representative. Keep the shipping materials for the carrier’s inspection.
2 Inspect the instrument.
In case of any damage, or defect, or failure, notify the RIGOL sales representative.
RIGOL offices will arrange reparation or replacement at RIGOL’s option without waiting for claim settlement.
3 Check the accessories.
Accessories supplied with the instrument are listed in “Appendix A: DP1308A Accessories at of this guide.
If the contents are incomplete or damaged, please notify the RIGOL sales representative.
Users Guide for DP1308A
Chapter 1 Quick Start RIGOL
1-3
LCD
Display user interface
Parameters input area
Include digital keyboard, unit key (direction key) and enter key
Power button
Turn on (or off) the instrument
USB Host port
Connect with external USB device as a main equipment
Menu softkey
With different functions in different menus.
Display modes switching button
Switch display modes between General and Focus
Channel setting button
Switch channels, enable/disable channel output
Output terminal
Channel output connector
Advanced function
Wave Disp: Set up whether to display the signal
V
mV
mA
A
Programmable DC Power Supply
③ ④ ⑤ ⑥ ⑦ ⑧ ⑨
① ②
Front/Rear Panels and User Interface
Front Panel
Figure 1-1 Front panel overview
Users Guide for DP1308A
RIGOL Chapter 1 Quick Start
1-4
area
through waveform
Store/Recall: Store and recall the settings of
instrument and files
Timer: Enable timing output Utility: Set up system configuration Track: Enable track function Help: System online help
USB Host: Connect as a main equipment with an external USB device such as U disc.
Output terminals of +6V channel
This terminal is connected with the chassis and ground wire when power on, which could act as an additional ground terminal for the devices connected to.
*
Output terminals of ± 25V channels
1. Button Indicator
The indicator lights when press the button to start-up the instrument.
On/Off The corresponding button light will go on if turn on a channel.
Wave Disp The button light will go on if enable waveform display.
Timer The button light will go on if enable timing output function. ·1
Track The button light will go on if enable track function.
Store/Recall The button light will go on if enter into store/recall menu.
Utility The button light will go on if enter into system settings.
Help The button light will go on if enter into online help.
2. Rear panel connector
① ② ③ ④
NOTE*: COM is a public port.
Users Guide for DP1308A
Chapter 1 Quick Start RIGOL
1-5
FUSE:
250V
T2.5 A
T2A
LINE
100V + /-10% 115V +/ -10% 220V + /-10% 230V + /-10%
HAZA RDOUS V OLTAGE INS IDE, DO NOT RE MOVE TH E COVER UNLE SS BY SPE CIFIED PERS ONNEL . WARNI NG: MAI NTAIN GR OUND TO AVOID EL ECTRI C SHOCK
40
RIGO L Technolo gies, In c.
MADE IN CHINA
GPI B
LXI C lass C Non Au to-MD IX
1
0
0
1
1
5
2
2
0
2
3
0
GPIB Port
Conform to IEEE 488.2 protocol
Fan
Air outlet of the fan
Supply Hub
AC input interface
Power Voltage Selector
Select the proper level according to the input power supply
LAN Port
Access LAN via RJ45 port
USB Device Port
Connect as a slave unit with an external USB device.
Fuse
Offer two types of specifications: 250V, T3A and 250V, T2A
Power Switch
Power-on/off the power
① ②
⑤ ⑥ ⑦ ⑧
Rear Panel
Figure 1-2 Rear panel overview
Users Guide for DP1308A
RIGOL Chapter 1 Quick Start
1-6
Channel output state
Output voltage
Channel mark
Output current
Selected channel*
Output power
Constant voltage output
O.V.P input box
LAN connecting
O.C.P input box
Keylock
Operation menu
Self-test failed
Timing output mark
O.T.P mark
Time for timing output
Remote control connecting
Voltage input box
U disk detected
Current input box
Help system mark
Track mark
The beeper is enabled
O.V.P/O.C.P state
21
22
⑯ ⑰
⑱ ⑲ ⑳
① ② ③ ④ ⑤ ⑥ ⑦ ⑧ ⑨ ⑪ ⑫
23
24
21
22
23
24
User Interface
Figure 1-3 User interface of DP1308A
NOTE*: The channel is highlighted on the screen (in breath state).
Users Guide for DP1308A
Chapter 1 Quick Start RIGOL
1-7
!
!
First Use DP1308A
Before operating, please do perform necessary inspections to ensure the instrument is in a good position to work normally.
Power Requirement
Four types of AC power within 50Hz-60Hz of frequency are provided by DP1308A: 100V, 115V, 220V, 230V. You can choose an appropriate input power as required through switching the Power Voltage Selector on the rear panel.
WARNING
Before switching the Power Voltage Selector”, make sure that the power
supply has been cut off and the instrument has installed specified fuse.
Power-On Inspection
Please use the power cord listed in accessories to connect the instrument with AC supply, and then perform the following instructions:
1. Turn on the instrument
Turn on the power switch on the rear panel. (At present, the instrument has connected to AC supply.)
WARNING
To avoid electric shock, make sure that the instrument has properly grounded.
2. Turn on the power button
Press down the power button on the front panel to enable the self-test. If
pass test, the user interface will appear on screen, or else a clue System power-on self-test has a mistake, view the error or not? will pop up, press OK to get more information about fail, or press Cancel to exit clue.
Users Guide for DP1308A
RIGOL Chapter 1 Quick Start
1-8
Output Inspection
In virtue of this function, the instrument enables to respond to front panel correctly and output rated value. The inspections are mainly centered on the voltage output on the occasion of no loads for each channel and the current output when short circuit.
1. Channel selection and output switch
Choose an appropriate output channel using selection buttons. The selected channel will be highlighted on screen.
Figure 1-4 Output channel selection
Turn on/off the output by pressing On/Off button under the selection button of each channel. After you enable the output of a channel, the corresponding button will light until you disable the output regardless of whether it is the current channel or not.
2. Voltage output inspection
(1) When the instrument has no load: Press power button and make sure that
the setting value of current in each channel is not 0;
(2) Inspect all on and all off function of channels;
Press All On button, then a clue about whether to open output of all
channels will appear on, if click ok, all the channel buttons light will go on that denotes they are ready to output; on the contrary, the instrument backs to the state before.
Press All Off buttons to disable the output of all channels. Meanwhile, the
corresponding On/Off button is black out.
Users Guide for DP1308A
Chapter 1 Quick Start RIGOL
1-9
(3) Inspect the voltage function of +6V channel;
Press +6V button and the corresponding On/Off button, the channel will be in constant voltage output state after On/Off button light goes on, then check whether the “voltage” can be adjusted from 0 to the maximum rated value(6.3V). See the Parameter Input section in chapter 2.
(4) Inspect the voltage function of +25V channel;
Press +25V button and the corresponding On/Off button, the channel will be in constant voltage output state after On/Off button light goes on, then check whether the voltage can be adjusted from 0 to the maximum rated value(26.25V).
(5) Inspect the voltage function of -25V channel;
Press -25V button and the corresponding On/Off button, the channel will be in constant voltage output state after On/Off button light goes on, then check whether the voltage can be adjusted from 0 to the minimum rated value(-26.25V).
3. Current output inspection
(1) Press the power button .
(2) Press All Off button to disable the output of all channels.
(3) Inspect the current function of +6V channel.
Short the terminal (+) and (-) of +6V channel using an insulating output
lead;
Press +6V button; Set the voltage as 6V; Press the corresponding On/Off button of +6V channel; After On/Off button light goes on, check whether the current can be
adjusted from 0 to the maximum rated value (5.25A).
(4) Inspect the current function of +25V channel.
Short the terminal (+) and (COM) of +25V channel using an insulating
testing lead;
Press +25V button; Set the voltage value as 25V;
Users Guide for DP1308A
RIGOL Chapter 1 Quick Start
1-10
Press the corresponding On/Off button of +25V channel; After On/Off button light goes on, check whether the current can be
adjusted from 0 to the maximum rated value (1.05A).
(5) Inspect the current function of -25V channel.
Short the terminal (-) and (COM) of -25V channel using an insulating testing
lead;
Press -25V button; set the voltage value as -25V; Press the corresponding On/Off button of -25V channel; After On/Off button light goes on, check whether the current can be
adjusted from 0 to the maximum rated value (1.05A).
-
Users Guide for DP1308A
Chapter 1 Quick Start RIGOL
1-11
Display Modes
DP1308A provides four display modes: general display mode, focus display mode, general waveform display mode and focus waveform display mode.
The output current and voltage can be displayed in waveforms by DP1308A, for the details , please see general waveform display mode (Figure 1-7) and focus waveform display mode (Figure 1-8). The Wave Disp button light will go on regardless of which waveform display mode is active until this function is disabled.
1. General display mode
In the following figure, the three channel outputs are shown on screen abreast, and the selected channel is highlighted and stay in breath state.
Figure 1-5 General display mode
2. Focus display mode
Press to enter into focus display mode from general display mode; to exit,
press the button again. In this mode, the parameters and output state of selected channel are displayed clearly.
Figure 1-6 Focus display mode
Users Guide for DP1308A
RIGOL Chapter 1 Quick Start
1-12
3. General waveform display mode
Press Wave Disp button in general display mode or press in focus waveform
display mode to enter into general waveform display mode. The state of Wave Disp button light varies with channel display mode. The waveforms from all three channels could be displayed simultaneously in this mode.
Figure 1-7 General waveform display mode
4. Focus waveform display mode
Press in general waveform display mode or press Wave Disp button in focus
display mode to enter into focus waveform display mode. In this mode, the parameters and channel output state are displayed clearly.
Figure 1-8 Focus waveform display mode
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-1
Chapter 2 Front Panel Operations
This chapter gives detailed information on function and operating methods about each button on the front panel, among them, you will learn more about DP1308A and make primary work.
Following topics are mainly throughout this chapter:
Parameter Input
Constant Voltage Output
Constant Current Output
Timing Output
Multiplexed Output
Tracking Output
Save and Recall
Set up the Utility
I/O Setting System Setting Calibrate Timer Setting
Online Help System
Key Lock
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-2
Parameter Input
DP1308A has two methods to input parameters: dialog box input and direct input (modify). Parameters can be inputted by the digital keyboard and unit key (direction
key).
Figure 2-1 Digital keyboard and unit key (direction key)
1. Dialog box input
When the selected menu is Volt or Curr, input desired numerical value using the digital keyboard and then press V, mV”, “A” or “mA” on the panel or in the menu, the instrument will automatically set the voltage or current according to the value and unit you have selected. When the selected menu is O.C.P or O.V.P, use the same way to set O.C.P or O.V.P.
For either Calibrate or Timer Setting mode, the unit must be defined based on practical needs after inputting numerical value, for instance, when input a voltage value, only mV and V can be chosen as its unit.
2. Direct input(modify)
When the cursor is in the input box of voltage or current, press the up (V) direction key to increase the value or press the “down (mV) direction key to decrease. To move the position of the cursor, press the “left (mA) or right (A) direction key.
For I/O Setting, respectively use the numeral key and direction key to input numerical value and move the cursor; to backspace, press ”.
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-3
NOTE
A long press on up or down direction key may change the setting
value of current or voltage quickly and linearly; if on left or right direction, the position of the cursor could be moved quickly.
While inputting the file name, the position of the cursor can be moved
quickly If press direction keys for long time.
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-4
Constant Voltage Output
DP1308A provides two kinds of output mode for each channel including Constant Voltage Output (CV) and Constant Current Output (CC), which is depends on the specified voltage and current as well as the loads carried by the instrument. The instrument will not output voltage or current beyond setting value. In CV mode, the output voltage equals to the specified value and the output current equals to the specified value in CC mode.
For example: Choose the +6V channel; set the voltage to 6V, the current to 5A and the load to 5Ω separately. Because 5Ω× 5A=25V>6V and 6V/5Ω=1.2A<5A, so the Constant Voltage Output of the channel is 6V, 1.2A.
Operation Steps:
1. Power on
Press to start-up the instrument.
2. Choose the channel
Press +6V to enable +6V channel, at present, the channel is highlighted on the screen (in breath state).
3. Set the voltage
Press Volt and set the voltage to 6 V.
Figure 2-2 Set the voltage
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-5
!
4. Set the current
Press Curr and set the current to 5 A.
Figure 2-3 Set the current
5. Connect the output lead
Connect the output terminal of+6V as the following figure:
Figure 2-4 Connect the terminal of +6V
CAUTION
Wrong connection may lead damages to the instrument or the equipment
connected to.
6. Turn on the output
Press the On/Off button below +6V button, and which will be lightened during the output until you press again.
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-6
NOTE
In CV state, the instrument will switch to the CC state in terms of the preset setting value of the current automatically if the output current exceeds the setting value resulting from the decrease of load, meanwhile, the output voltage will decrease pro rata. To resume CV state, increase the load or the current setting value.
NOTE
If channel output is disabled, the O.V.P function can not be enabled
but only for inputting value.
If the output voltage beyond the value of overvoltage protection, the
output will be disabled automatically.
*: Please use O.V.P function as required.
7. O.V.P*
In channel output mode, press O.V.P to enter into the setting menu, at present, the cursor is flashing in O.V.P voltage input box, here, you can input needed value using number keyboard and then press O.V.P to enable (or disable) O.V.P protection. If the parameters of O.V.P have already been set, press O.V.P to directly enable (or disable) O.V.P protection.
Figure 2-5 Set O.V.P
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-7
NOTE
If channel output is disabled, the O.C.P function can not be enabled but
only for inputting value.
If the output current beyond the value of over current protection, the
output will be disabled automatically.
*: Please use O.C.P function as required.
8. O.C.P*
In channel output mode, press O.C.P to enter into the setting state, at present, the cursor is flashing in O.C.P current input box, here, you can input needed current value using number keyboard and then press O.C.P to enable (or disable) O.C.P protection. If the parameters of O.C.P have already been set, press O.C.P to directly enable (or disable) O.C.P protection.
Figure 2-6 Set O.C.P
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-8
Constant Current Output
For example:
Choose the +6V channel; set the voltage to 6V, the current to 5A and the load to 1Ω separately. Because 6V/1Ω=6A>5A and 5A× 1Ω5V<6V, so the Constant Current Output of the channel is 5A, 5V.
Operation Steps:
1. Power on
Press to start-up the instrument.
2. Choose the Channel
Press +6V to enable +6V channel, at present, the channel is highlighted on the screen (in breath state).
3. Set the voltage
Press Volt and set the voltage to 6 V.
Figure 2-7 Set the voltage
4. Set the current
Press Curr and set the current to 5 A.
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-9
!
Figure 2-8 Set the current
5. Connect the output lead
Connect the output terminal of+6V as the following figure:
Figure 2-9 Connect the terminal of +6V
CAUTION
Wrong connection may lead damages to the instrument or the equipment connected to.
6. Turn on the output
Press the On/Off button below +6V button and which will be lightened during the output until you press again.
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-10
NOTE
In CC state, the instrument will switch to the CV state in terms of the preset setting value of the voltage automatically if the output voltage exceeds the setting value resulting from the increase of load, meanwhile, the output current will decrease pro rata. To resume CC state, decrease the load or increase the voltage setting value.
NOTE
If channel output is disabled, the O.V.P function can not be enabled
but only for inputting value.
If the output voltage beyond the value of overvoltage protection, the
output will be disabled automatically.
*: Please use O.V.P function as required.
7. O.V.P*
In channel output mode, press O.V.P to enter into the setting state, at present, the cursor is flashing in O.V.P voltage input box, here, you can input needed value using number keyboard and then press O.V.P to enable (or disable) O.V.P protection. If the parameters of O.V.P have already been set, press O.V.P to directly enable (or disable) O.V.P protection.
Figure 2-10 Set the O.V.P
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-11
NOTE
If channel output is disabled, the O.C.P function cannot be enabled but
only for inputting value.
If the output current beyond the value of overcurrent protection, the
output will be disabled automatically.
*: Please use O.C.P function as required.
8. O.C.P*
In channel output mode, press O.C.P to enter into the setting state, at present, the cursor is flashing in O.C.P current input box, here, you can input needed current value using number keyboard and then press O.C.P to enable (or disable) O.C.P protection. If the parameters of O.C.P has already been set, press O.C.P to directly enable (or disable) O.C.P protection.
Figure 2-11 Set O.C.P
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-12
Clock icon
Specified
time
Timing Output
Press Timer to enable the timing output of current channel, at this moment, the button light goes on and a Clock icon and the corresponding timing parameters are shown on the screen. See the figure below. The instrument will start output after you turn on the current channel until finishing all timing outputs. (at the same time, the light of Timer button goes out.)
If no timing parameters have been set before enabling the timer, the system will enter Timer Setting” Interface automatically after you press Timer (the light goes on). In timing mode, if you turn off the channel under timing, the timer will stop until it is enabled again. However, the timer would be disabled once you press Timer during timing.
Figure 2-12 Timing output (in focus display mode)
In focus display mode, the character Set on the screen denotes the current output voltage and current setting value while Next denotes the next output voltage and current setting value. The next output value does not displayed in normal display mode, see the figure below.
Figure 2-13 Timing output (normal display mode)
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-13
NOTE
If the timing output is enabled on a channel, of which Volt and Curr values
cant be set.
Timing Output can not be used if tracking function of +25V (-25V) channel is
active.
To change the voltage, current or the time that has been set, press Utility to enter the TimerSet menu.
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-14
+25V
COM
COM
-25V
Multiplexed Output
Typically, all three channels of DP1308A enable to output synchronously, and each of them could be controlled independently.
Operation Steps:
1. Power on
Press to start-up the instrument.
2. Choose and set channels
Press the channel selection button to choose the channel required to output, then separately set the voltage and current for each of them.
3. Connect the output lead
The connection method of output terminals are given as follow, you can use a cable with banana plug or a common cable to connect them with customers.
4. Turn on the output
Press All On, and the system will prompt whether to turn on the output of all channels. If you select “ok”, the On/Off button of the corresponding channel will be light, which indicates that all channel outputs are enabled; conversely the instrument returns the state before.
5. Set O.V.P and O.C.P
In output state, you can set O.V.P and O.C.P parameters for each channel if necessary.
Users Guide for DP1308A
Figure 2-14 Connect the multiplexed terminals
Chapter 2 Front Panel Operations RIGOL
2-15
+25V
-25V
NOTE*
When +25V channel is connecting with -25V channel in series, the output voltage is the sum of the two channels output voltage, while the output current may be one of the two channels output current.
6. How to output 0V~50V power*
DP1308A can output 0V~50V power by connecting +25V channel in series with
-25V channel, see the figure below.
Figure 2-15 Connect +25V and -25V channel in series
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-16
Tracking Output
The ± 25V channel provides tracking output function, which denotes that the voltage of a channel can vary with the one of another channel.
Select a channel such as +25V, and then press Track, at present, the button lights
and a track icon appears at the lower left of +25V that denotes +25V is in tracking mode.
Figure 2-16 Tracking output (before tracking)
Next, change the voltage of -25V channel, and then the voltage of +25V channel changes along with it*. For example: Change the voltage of -25V channel to 2.5V, the voltage of +25V channel is also changed to 2.5V, please see the figure below.
Figure 2-17 Tracking output (after tracking)
Press Track again to disable track function and the light is out.
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-17
NOTE
*: Track function is only available for the voltage setting value but has nothing
to do with the output value. The voltage of channel under tracking varies with the value of channel to be tracked only when both channels are in CV state.
Track function is forbidden when Timing Output of +25V (-25V) channel is
active.
The voltage of channel under tracking can not be set manually when track
function is enabled.
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-18
Menu
Explanations
Browser
Shift cursor between directory and file window
Store
Enter the Save interface
Recall
Recall the selected system configuration file
Delete
Delete the selected file
Enter next page
Back to previous page
Update
Select the desired file and update
Exit
Exit the save and recall interface
Save and Recall
DP1308A supports two types of document storage ways: U disc storage and local storage. Among the internal memory, 4 groups of settings could be saved or recalled and more settings are available for USB device.
Press Store/Recall to enter the Save and Recall interface (the button is light). To exist, press this button again. See as figure 2-18.
Figure 2-18 Save and recall interface
Table 2-1 Explanations of save and recall menus
Press Store to enter the interface below, press InType to switch the input method.
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-19
Menu
Explanations
Input Method
Chinese or English
Focus Cursor
Shift cursor among the input box of file name, virtual keyboard and Chinese show box
ABC(abc)
Switch capital and lower case. If display ABC”, lowercase letter and pinyin could be inputted. If display “abc”, capital letter could be inputted.
Save
Save system configuration file
Back to save and recall interface
Figure 2-19 Input interface of the file name
Table 2-2 Explanations of file name input menu
1. Chinese Input
Press InType and select “Ch” to enter the Chinese input interface as the figure above. Take file name 北京普源 as an example:
(1) Input 北京普源
Select character b in soft keyboard using direction keys and press OK to
input, then, input beiin the same way.
Press Focus and switch cursor to Chinese input box, then press direction
keys up or down to shift characters until the character appears. After
that, input the whole name file 北京普源 in the same way.
(2) Delete the wrong characters you input
In the process of editing, if press ”, the characters in display box will be deleted at first.
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-20
NOTE
The length of Chinese file name must be less than 6 characters and the English name must be less than 12 letters (including number).
(3) Save the file
Press Save to save the file as the specified name and finish editing.
2. English Input
Press InType and select En to enter the English input interface as the figure below. Take file name newfile as an example:
Figure 2-20 English Input
(1) Input “newfile”
Select character “n” in soft keyboard using direction keys and press OK to input, then, input “newfile” in the same way.
(2) Delete the wrong letters you input
Shift the focus cursor to the input box, then select the letter to be deleted using left/right direction key and press to delete.
(3) Save the file
Press Save to save the file as the specified name and finish editing.
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-21
Menu
Explanation
I/O
Set the LAN parameters and GPIB address and view the ID of USB device
Set system parameters
Calibrate
Enter the calibration interface
TimerSet
Set the timing parameters for each channel
Exit
Exist system menu
Set up the Utility
Press Utility (the light is on) to enter the utility interface as the figure below.
Figure 2-21 Utility Setting interface
Table 2-3 Explanations of utility setting menu
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-22
Net Connection:
Link/Unlink
IP configuration mode:
DHCP/ Auto IP/Manual IP
MAC Address
VISA Name
I/O Setting
DP1308A provides LAN, USB and GPIB three interfaces for communication; you can set and review the related parameters through setting interface.
1. LAN
Press I/O LAN, enter the following interface.
Figure 2-22 Set the LAN
General condensed summaries:
If “DHCP is selected, the instrument will automatically assign IP Address”,
Mask, “Gateway and DNS Server”.
If “Auto IP” is selected, the instrument will automatically assign IP Address and
Mask”.
If Manual IP is selected, IP Address”, Mask”, “Gateway and “DNS Server has
to be set manually.
If both DHCP”, “Auto IP and Manual IP” are selected, the priority is from high to
low: DHCP”, “Auto IP” and “Manual IP”.
Default settings:
In the second page of I/O menu, press Default OK to recall the default setting of LAN. See the following figure.
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-23
Figure 2-23 Recall the default setting of LAN
2. USB
Press I/O USB, enter the following interface. This option shows the ID information about USB device with setup-free.
Figure 2-24 USB ID information
3. GPIB
Press I/O GPIB, enter the following interface. The range of GPIB address is 1~30. See the figure below, directly input the desired address by number keyboard and then press OK to save. Press or the left direction key to delete the number before the cursor.
Figure 2-25 Set the GPIB address
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-24
System Setting
Press to enter the system setting menu, which including language, power-on
and brightness, over-temperature protection, beeper, self-test and system information.
Figure 2-26 System settings
1. Language Setting
Press Language to enter the setting interface of system language. The optional languages are “Chinese” and “English”.
2. PowerOn Setting
Press PowerOn to enter the setting interface. The optional settings are “Default” and “Last”.
“Default”: Factory settings will be recalled when reset the instrument. “Last”: Settings before last power-off will be recalled when reset the
instrument.
3. Brightness Setting
Press Bright to enter the brightness setting interface. The brightness of display screen could be regulated by pressing “+”/ - key (in menu) or left/right direction key.
Figure 2-27 Brightness setting
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-25
4. Over-temperature Protection Setting
Press O.T.P to enter the over-temperature protection setting interface. The over-temperature protection includes two grades. Users can turn ON or OFF the
first-grade protection. A mark will appear at the Status Bar when the
protection is turned on, see as follow:
Figure 2-28 Over-temperature Protection Settings
The second-grade protection is a mandatory protection which is always on.
5. Beeper Setting
Press Beeper to enter into setting interface. A sign would shown at Status
Bar once you turn on beeper, which indicates that the instrument may buzzed once pop up a hint message or some button is pressed down.
Figure 2-29 Beeper settings
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-26
6. SelfTest
Press SelfTest to enter the following interface. Users can perform self-test and view its result. If do not carry out self-test, the display is the results when power on.
Figure 2-30 Self-Test menu
Figure 2-31 View Self-Test Result
7. View System Information
Press Sysinfo to view the Model, Serial Number, Hardware Version, Logic Version and Software Version of the instrument.
Figure 2-32 System information settings
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-27
NOTE
Please do not calibrate privately without permission.
For calibration help, please refer to DP1308A Service Guide” or contact
RIGOL technical support.
Calibrate
Before enter the calibration interface, correct password must be input. The instrument has been calibrated before leaving factory.
Figure 2-33 Input interface of calibration code
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-28
Timer Setting
In timer setting interface, you can set user-defined voltage, current and output time as well as up to five groups timing output for each channel.
See the figure below: the interface is divided into two parts: parameter setting area (right) and waveform display area (left).
Figure 2-34 Timing Settings
In parameter setting area, three kinds of parameters for each output are available to set: voltage (Volt), current (Curr) and time (Set). In waveform display area, the system displays waveform based on specified voltage (Volt) and current (Curr) for user to view the output state more clearly and directly.
General condensed summaries:
If Timing output is disabled, the parameter Left(s) in parameter setting area is
equal to Set(s)”.
If Timing Output is enabled, the instrument starts output with first group of
settings after you enable relative channel, and the value of Left(s) is on the decrease until to 0, at this moment, the instrument enters next group and so on until finish outputs of all settings.
Users Guide for DP1308A
Chapter 2 Front Panel Operations RIGOL
2-29
Online Help System
Online help system provides the help information about buttons on the front panel (except for numeric keys and direction keys), which make users acquire prompt or help information about each function and the operation methods more conveniently and quickly.
After you press Help (the light is on), a letter HELP will shown at status bar; then, press desired help button and the related help information will shown on the screen, meanwhile, Help light goes out and the letter HELP disappears. Press Exit to quit help menu.
Figure 2-35 Help information
Users Guide for DP1308A
RIGOL Chapter 2 Front Panel Operations
2-30
Key Lock
This function is widely used in product line of industry to avoid misoperation.
1. Local Mode
In Local mode, press and hold “Little Key ” (number 7) in digital keyboard to lock
all buttons except power button ,“Little Key” and All Off, at the moment, a
sign appears on the screen. To unlock the keyboard, press “Little Key” again.
2. Remote Mode
In Remote mode, all buttons are locked automatically except power button ,
“Little Key” and All Off. To exit remote mode, press “Little Key” again.
Users Guide for DP1308A
Chapter 3 Troubleshooting RIGOL
3-1
Chapter 3 Troubleshooting
1. After the instrument is powered on, the screen still dark (no display):
(1) Check the power cord connection. (2) Ensure the power indicator is lighting and flickering. (3) Pull out the power cord from instrument and ensure the voltage selector is in right
scale and the fuse is suitable and intact.
(4) If the problem still remains, please ask RIGOL for help.
2. Constant voltage output is abnormal:
(1) Check whether the maximum power from selected channel meets load
requirement.
(2) If meet, please do:
Verify whether the current value of this channel is appropriate, if lower, please
increase it in moderation.
Ensure the cord used for connecting load and power contracts well and no
short circuit occurs.
Check if the load goes wrong.
(3) If the problem still remains, please ask RIGOL for help.
3. Constant current output is abnormal:
(1) Check whether the maximum power of selected channel meets load requirement. (2) If meet, please do:
Verify whether the voltage value of channel is appropriate, if lower, please
increase it in moderation.
Ensure the cord used for connecting load and power contracts well and no
short circuit occurs.
Check if the load goes wrong.
(3) If the problem still remains, please ask RIGOL for help.
4. U disk cant be identified:
(1) Check whether the U disk works normally. (2) Ensure the U disk used now is Flash, as the hard disk can not be supported. (3) Restart the instrument and insert U disk again. (4) If the problem still remains, please ask RIGOL for help.
Users Guide for DP1308A
Chapter 4 Specifications RIGOL
4-1
Model
DP1308A
Channels
+6V
+25V
-25V
DC output (@ 0℃~40)
Voltage
0+6V
0+25V
0-25V
Current
05A
01A
01A
Overvoltage Protection
0.1V6.5V
0.1V~27V
-0.1V-27V
Overcurrent Protection
0.1A~5.5A
0.1A1.2A
0.1A1.2A
Load Regulation± (output percentage + offset)
Voltage
<0.01%+2mV
Current
<0.01%+250μA
Linear Regulation± (output percentage + offset)
Voltage
<0.01%+2mV
Current
<0.01%+250μA
Ripple and Noise (20 Hz~20 MHz)
Normal Mode Voltage
<350μVrms/2mVpp
Normal Mode Current
<2mArms
<500μArms
Common Mode Current
<1.5μArms
Accuracy 12 Months
[1]
(25± 5)± (output percentage + offset)
Program ming
Voltage
0.1%+5mV
0.05%+20mV
Current
0.2%+10mA
0.15%+4mA
Read Back
Voltage
0.1%+5mV
0.05%+10mV
Current
0.2%+10mA
0.15%+4mA
Resolution
Programming
0.5mV/0.5mA
1.5mV/0.1mA
Read Back
0.5mV/0.5mA
1.5mV/0.1mA
Meter
1mV/1mA
10mV/1mA
Transient Response Time
Chapter 4 Specifications
DP1308A Programmable Linear DC Power Supply specif ication and operating characteristics are based on the instrument having been operated continuously for 30 minutes under the specified operating temperature.
NOTE:
All the specifications below apply to all the three channels output unless where noted.
Users Guide for DP1308A
RIGOL Chapter 4 Specifications
4-2
Less than 50μs is spent on recovering the voltage within 15mV during the output current changes from full load to half load or half to full.
Command Processing Time
[2]
<50ms
Temperature Coefficient per (output percentage + offset)
Voltage
0.01%+2mV
0.01%+3mV
Current
0.02%+3mA
0.01%+0.5mA
Stability
[3]
± (output percentage + offset)
Voltage
0.03%+1mV
0.02%+2mV
Current
0.1%+3mA
0.05%+1mA
Voltage Programming Speed(to within 1% of total variation range)
Rising Full Load
11ms
50ms
No Load
10ms
45ms
Falling Full Load
13ms
20ms
No Load
200ms
400ms
OVP/OCP
Accuracy ± (output percentage + offset)
0.5%+0.5V/0.5%+0.5A
Activation Time
1.5ms (OVP≥ 3V) <10ms (OVP<3V and OCP)
Machine
Dimension
235 mm (W) x 155 mm (H) x 384 mm (D)
Weight
8.5 kg
Power Supply
AC Output (50Hz-60Hz)
100Vac+10%, 115Vac+10%, 220Vac+10%, 230Vac+10%
Environment
Working Temperature
Full rated value output: 0℃~40 At higher temperature: the output current falls into 50% at the maximum temperature 55
Cooling Method
Fan cooling
Remarks:
[1] Specifications are for one hour warm-up and at 25℃.
[2] The maximum time required for regulating corresponding output when received APPLy and
SOURce commands.
[3] The variation of output within 8 hours after warm-up 30 minutes and both the load circuit and
environment temperature are in constant conditions.
Users Guide for DP1308A
Chapter 5 Appendix RIGOL
5-1
Chapter 5 Appendix
Appendix A: DP1308A Accessories
Standard Accessories:
A power cord that fits the standard of destination country An USB data cable A CD (including Users Guide) An User Instructions An INSTRUCTION Four fuses (two of 250V/T3A and two of 250V/T2A)
NOTE: All the accessories are available by contacting your local RIGOL office.
Users Guide for DP1308A
RIGOL Chapter 5 Appendix
5-2
Appendix B: Warranty
RIGOL warrants that its products mainframe and accessories will be free from defects in materials and workmanship within the warranty period. If a product proves defective within the respective period. RIGOL guarantees the free replacement or repair of products which are approved defective.
To get repair service or obtain a copy of the whole warranty statement, please contact with your nearest RIGOL sales and service office.
RIGOL does not provide any other warranty items except the one being provided by this summary and the warranty statement. The warranty items include but not being subjected to the hint guarantee items related to tradable characteristic and any particular purpose.
RIGOL will not take any responsibility in cases regarding to indirect, particular and ensuing damage.
Users Guide for DP1308A
Chapter 5 Appendix RIGOL
5-3
!
Appendix C: General Care and Cleaning
General Care
Do not store or leave the instrument in where the instrument exposed to direct sunlight for long periods of time.
Caution
To avoid damages to the instrument, do not expose it to liquids which have causticity.
Cleaning
Clean the instrument often based on the operating conditions. To clean the exterior surface, perform the following steps:
1. Disconnect the instrument from all power sources.
2. Clean the loose dust on the outside of the instrument with a lint- free cloth (with
a mild detergent and water). When clean the LCD, take care to avoid scarifying it.
WARNING
To avoid injury resulting from short circuit, make sure the instrument is completely dry before reconnecting into a power source.
Users Guide for DP1308A
RIGOL Chapter 5 Appendix
5-4
Appendix D: Contact RIGOL
If you have any problem or requirement during using our products, please contact RIGOL Technologies, Inc. or the local distributors.
Domestic: Please call Tel: (86-10) 8070 6688 Fax: (86-10) 8070 5070
Service & Support Hotline: 800 810 0002
9:00 am –5: 00 pm from Monday to Friday
Or by e-mail:
service@rigol.com
Or mail to: RIGOL Technologies, Inc. 156# CaiHe Village, ShaHe Town, ChangPing District, Beijing, China Post Code: 102206
Overseas: Contact the local RIGOL distributors or sales office. For the latest product information and service, visit our website:
http://www.rigolna.com/
Users Guide for DP1308A
RIGOL
1
Index
Brightness ................................. 2-24
Calibration ................................. 2-27
CC .............................................. 2-4
Cleaning ...................................... 5-3
Command Processing Time ........... 4-2
CV .............................................. 2-4
DC output ................................... 4-1
GPIB ......................................... 2-23
LAN .......................................... 2-22
Language .................................. 2-24
Linear Regulation ......................... 4-1
Load Regulation ........................... 4-1
Local Mode ................................ 2-30
Noise ........................................... 4-1
O.C.P........................................... 2-7
O.V.P ........................................... 2-6
Power-On .................................. 2-24
Resolution .................................... 4-1
Ripple .......................................... 4-1
Stability ....................................... 4-2
Temperature Coefficient ................ 4-2
Timing ................. 2-1, 2-2, 2-12, 2-28
Transient Response Time .............. 4-1
USB ........................................... 2-23
Voltage Programming Speed ......... 4-2
Users Guide for DP1308A