RIGOL®is the trademark of RIGOL TECHNOLOGIES CO., LTD.
Notices
•RIGOL products are covered by P.R.C. and foreign patents, issued and pending.
•RIGOL reserves the right to modify or change parts of or all the specifications and pricing
policies at the company's sole decision.
•Information in this publication replaces all previously released materials.
•Information in this publication is subject to change without notice.
•RIGOL shall not be liable for either incidental or consequential losses in connection with the
furnishing, use, or performance of this manual, as well as any information contained.
•Any part of this document is forbidden to be copied, photocopied, or rearranged without prior
written approval of RIGOL.
Product Certification
RIGOL guarantees that this product conforms to the national and industrial standards in China as
well as the ISO9001:2015 standard and the ISO14001:2015 standard. Other international standard
conformance certifications are in progress.
Contact Us
If you have any problem or requirement when using our products or this manual, please contact
RIGOL.
Please review the following safety precautions carefully before putting the instrument
into operation so as to avoid any personal injury or damage to the instrument and
any product connected to it. To prevent potential hazards, please follow the
instructions specified in this manual to use the instrument properly.
•Use Proper Power Cord.
Only the exclusive power cord designed for the instrument and authorized for
use within the destination country could be used.
•Ground the Instrument.
The power adaptor of this instrument is a three-pin power plug, but the
instrument housing is not connected to the earth lead of the three-pin power
plug. To avoid electric shock, use the ground lead provided in accessories to
connect the instrument housing to an electrical ground before performing
measurements.
•Observe All Terminal Ratings.
To avoid fire or shock hazard, observe all ratings and markers on the instrument
and check your manual for more information about ratings before connecting
the instrument.
•Use Proper Overvoltage Protection.
Ensure that no overvoltage (such as that caused by a bolt of lightning) can reach
the product. Otherwise, the operator might be exposed to the danger of an
electric shock.
•Do Not Operate without Covers.
Do not operate the instrument with covers or panels removed.
•Do Not Insert Objects into the Air Outlet.
Do not insert objects into the air outlet, as doing so may cause damage to the
instrument.
•Use the Proper Fuse.
Please use the specified fuses.
•Avoid Circuit or Wire Exposure.
Do not touch exposed junctions and components when the instrument is
powered on.
If you suspect that any damage may occur to the instrument, have RIGOL
authorized personnel before further operations. Any maintenance, adjustment or
replacement especially to circuits or accessories must be performed by RIGOL
authorized personnel.
Inadequate ventilation may cause an increase of temperature in the instrument,
which would cause damage to the instrument. So please keep the instrument
well ventilated and inspect the air outlet and the fan regularly.
To avoid short circuit inside the instrument or electric shock, never operate the
instrument in a humid environment.
To avoid personal injuries or damage to the instrument, never operate the
instrument in an explosive atmosphere.
•Keep Instrument Surfaces Clean and Dry.
To avoid dust or moisture from affecting the performance of the instrument,
keep the surfaces of the instrument clean and dry.
•Prevent Electrostatic Impact.
Operate the instrument in an electrostatic discharge protective environment to
avoid damage induced by static discharges. Always ground both the internal and
external conductors of cables to release static before making connections.
•Handle with Caution.
Please handle with care during transportation to avoid damage to keys, knobs,
interfaces, and other parts on the panels.
•Only for Measurements within the Specified Measurement Categories.
The measurements are allowed at CAT I 1000 V/CAT II 300 V only. To avoid
damage to the instrument and electric shock, do not make measurements out of
the voltage range or at CAT III/CAT IV.
WARNING
Equipment meeting Class A requirements may not offer adequate protection to broadcast
Indicates a potentially hazardous situation or practice which, if not avoided, will result in
serious injury or death.
CAUTION
Indicates a potentially hazardous situation or practice which, if not avoided, could result
in damage to the product or loss of important data.
Safety Notices on the Product:
•DANGER
Safety Requirement
It calls attention to an operation, if not correctly performed, could result in injury
or hazard immediately.
•WARNING
It calls attention to an operation, if not correctly performed, could result in
potential injury or hazard.
•CAUTION
It calls attention to an operation, if not correctly performed, could result in
damage to the product or other devices connected to the product.
Safety Symbols on the Product:
Hazardous
Voltage
Safety Warning Protective Earth
Terminal
1.3Input Terminal Protection Limits
Chassis GroundTest Ground
Protection Limits are defined for the input terminals:
Main Input (HI and LO) Terminals
The HI and LO input terminals are used for voltage, resistance, capacitance, continuity,
frequency (period), and diode test measurements. The following two Protection Limits
are defined for these terminals:
1. HI to LO Protection Limit: 1000 VDC or 750 VAC, which is also the maximum
voltage measurement. This limit can also be expressed as 1000 Vpk maximum.
2. LO to Ground Protection Limit. The LO input terminal can safely "float" a maximum
of 500 Vpk relative to the ground.
The protection limit for HI terminal is a maximum of 1000 Vpk relative to the ground.
Therefore, the sum of the "float" voltage and the measured voltage cannot exceed
1000 Vpk.
Sense (HI Sense and LO Sense) Terminal
HI Sense and LO Sense terminals are used only for four-wire resistance
measurements. The following two Protection Limits are defined for these terminals:
1. HI Sense to LO Sense Protection Limit. HI Sense to LO Sense Protection Limit: 200
Vpk.
2. LO Sense to LO Protection Limit. LO Sense to LO Protection Limit: 2 Vpk.
Current Input (I) Terminal
The I and LO terminals are used for current measurements. The I terminal has a
protection limit of 10 A(DM858)/3.15 A(DM858E) maximum current flowing through
the terminal by the front-panel current input fuse.
CAUTION
The current input terminal will be at approximately the same voltage as the LO terminal
unless the current input fuse is blown. To maintain protection, replace this fuse only with
a fuse of the specified type and rating.
1.4Measurement Category
Measurement Category
CAT I 1000 V/CAT II 300 V
WARNING
To avoid the danger of electric shock, do not make measurements out of the voltage
range or at CAT III/CAT IV (such as measuring AC motors, motor protectors, frequency
converters, and inverters).
WARNING
For CAT II, the input terminal can be connected to the main circuit (up to 300 VAC). To
avoid electric shock, do not connect the input terminal to the main circuit with a line
To protect against the danger of electric shock, DM858/DM858E Digital Multimeter
provides overvoltage protection for line-voltage mains connections meeting both of
the following conditions:
1. The HI and LO input terminals are connected to the mains under Measurement
Category II conditions, defined below.
2. The mains are limited to a maximum line voltage of 300 VAC.
IEC Measurement Category II includes electrical devices connected to mains at an
outlet on a branch circuit. Such devices include most small appliances, test
equipment, and other devices that plug into a branch outlet or socket. DM858/
DM858E may be used to make measurements with the HI and LO inputs connected to
mains in such devices (up to 300 VAC), or to the branch outlet itself.
WARNING
To avoid electric shock, DM858/DM858E may not be used with its HI and LO inputs
connected to mains in permanently installed electrical devices such as the main circuit-
breaker panel, sub-panel disconnect boxes, or permanently wired motors. Such devices
and circuits are subject to overvoltages that may exceed the protection limits of the
DM858/DM858E.
CAUTION
Voltages above 300 VAC may be measured only in circuits that are isolated from mains.
However, transient overvoltages are also present on circuits that are isolated from mains.
DM858/DM858E is designed to safely withstand occasional transient overvoltage up to
2500 Vpk. Do not use this equipment to measure circuits where transient overvoltage
could exceed this level.
Measurement Category Definitions
•Measurement Category I (CAT I) is for measurements performed on circuits not
directly connected to MAINS. Examples are measurements on batteries,
electronic devices (especially laptops), and circuits not derived from MAINS,
especially protected (internal) MAINS derived circuits.
•Measurement Category II (CAT II) is for measurements performed on circuits
directly connected to low voltage installation. Examples are measurements on
household appliances, portable tools and similar equipment.
•Measurement Category III (CAT III) is for measurements performed in the
building installation. Examples are measurements on distribution boards, circuitbreakers, wiring (including cables, bus-bars, junction boxes, switches and socketoutlets) in the fixed installation, and equipment for industrial use and some
other equipment. For example, stationary motors with permanent connection to
a fixed installation.
•Measurement Category IV (CAT IV) is for measurements performed at the
source of a low-voltage installation. Examples are electricity meters and
measurements on primary overcurrent protection devices and ripple control
units.
1.5Ventilation Requirement
This instrument uses the natural convection for heat dissipation. When using the
instrument in a bench-top or rack setting, provide at least 10 cm clearance beside,
above, and behind the instrument for adequate ventilation.
CAUTION
Inadequate ventilation may cause an increase of temperature in the instrument, which
would cause damage to the instrument. So please keep the instrument well ventilated and
inspect the air outlet regularly.
1.6Working Environment
Temperature
Operating: 0℃ to +50℃
Non-operating: -20℃ to +70℃
Humidity
≤80% RH, 40℃, without condensation
WARNING
To avoid short circuit inside the instrument or electric shock, never operate the
instrument in a humid environment.
Altitude
Below 3 km
Protection Level Against Electric Shock
ESD ±8kV
Installation (Overvoltage) Category
This product is powered by mains conforming to installation (overvoltage) category II.
WARNING
Ensure that no overvoltage (such as that caused by a bolt of lightning) can reach the
product. Otherwise, the operator might be exposed to the danger of an electric shock.
Installation (overvoltage) category I refers to signal level which is applicable to
equipment measurement terminals connected to the source circuit. Among these
terminals, precautions are done to limit the transient voltage to a low level.
Installation (overvoltage) category II refers to the local power distribution level which
is applicable to equipment connected to the AC line (AC power).
Pollution Degree
Pollution Degree 2
Pollution Degree Definition
•Pollution Degree 1: No pollution or only dry, nonconductive pollution occurs.
The pollution has no effect. For example, a clean room or air-conditioned office
environment.
•Pollution Degree 2: Normally only nonconductive pollution occurs. Temporary
conductivity caused by condensation is to be expected. For example, indoor
environment.
•Pollution Degree 3: Conductive pollution or dry nonconductive pollution that
becomes conductive due to condensation occurs. For example, sheltered
outdoor environment.
•Pollution Degree 4: The pollution generates persistent conductivity caused by
conductive dust, rain, or snow. For example, outdoor areas.
Safety Class
Class 2
1.7Care and Cleaning
Care
Do not store or leave the instrument where it may be exposed to direct sunlight for
long periods of time.
Cleaning
Clean the instrument regularly according to its operating conditions.
1. Disconnect the instrument from all power sources.
2. Clean the external surfaces of the instrument with a soft cloth dampened with mild
detergent or water. Avoid having any water or other objects into the chassis via the
heat dissipation hole. When cleaning the LCD, take care to avoid scarifying it.
CAUTION
To avoid damage to the instrument, do not expose it to caustic liquids.
To avoid short-circuit resulting from moisture or personal injuries, ensure that the
instrument is completely dry before connecting it to the power supply.
1.8Environmental Considerations
The following symbol indicates that this product complies with the WEEE Directive
2012/19/EU.
The equipment may contain substances that could be harmful to the environment or
human health. To avoid the release of such substances into the environment and
avoid harm to human health, we recommend you to recycle this product
appropriately to ensure that most materials are reused or recycled properly. Please
contact your local authorities for disposal or recycling information.
•Up to 500,000 points logging memory to record and analyze more data
•5½-digit reading resolution (120,000 counts)
•True-RMS AC Voltage and AC Current measuring
•Built-in 10 groups of data storage and 10 groups of setup storage
•Strong Math functions, supporting various application protocols
•7'' color touch screen, displaying dual-measurement results concurrently
•Type-C interface for power charge; compact in size, freeing up your bench
space
•Standard USB and LAN interfaces, supporting Web Control
DM858 series benchtop digital multimeter provides a highly cost-effective test tool
for engineers. It features 5.5-digit resolution, max. 125 readings/s, data logging
memory of 500,000 points, and 0.03% DCV accuracy (1 year). It provides 11
measurement functions for input signals, 5 Math operations, and 3 graphical display
types, satisfying what most experiments and tests require of a multimeter. The 7''
touch screen allows a clear view on measurement results. Its provides USB and LAN
interfaces (standard) as well as the Web Control function for you to control the
instrument directly, bringing smoother interaction. It is equipped with a Type-C
interface for power charge. In addition, it has a compact size and supports the
attachment to a bracket, saving valuable bench space for engineers.
This manual gives you a quick overview of the front and rear panel, user interface as
well as basic operation methods of DM858 series digital multimeter.
TIP
For the latest version of this manual, download it from RIGOL official website (
www.rigol.com
Publication Number
UGC11103-1110
Software Version
Software upgrade might change or add product features. Please acquire the latest
version of the manual from RIGOL website or contact RIGOL to upgrade the software.
Format Conventions in this Manual
1. Key
).
http://
The front panel key is denoted by the menu key icon. For example,
indicates the "Trig" key.
2. Menu
The menu item is denoted by the format of "Menu Name (Bold) + Character
Shading" in the manual. For example,
item. You can click or tap
3. Operation Procedures
The next step of the operation is denoted by ">" in the manual. For example,
> Storage indicates first clicking or tapping and then clicking or tapping
Storage.
4. Connector
The front/rear panel connector is denoted by "Brackets + Connector Name (Bold)",
for example, [EXT TRIG].
Content Conventions in this Manual
Measure to access the "Measure" menu.
Measure indicates the "Measure" menu
DM858 series digital multimeter includes DM858 and DM858E models. The table
below shows the main differences between the two models. Their operation methods
are the same. Unless otherwise specified, this manual takes DM858 as an example to
illustrate the basic operation methods of DM858 series.
If the packaging has been damaged, do not dispose the damaged packaging or
cushioning materials until the shipment has been checked for completeness and
has passed both electrical and mechanical tests.
The consigner or carrier shall be liable for the damage to the instrument resulting
from shipment. RIGOL would not be responsible for free maintenance/rework or
replacement of the instrument.
2. Inspect the instrument
In case of any mechanical damage, missing parts, or failure in passing the electrical
and mechanical tests, contact your RIGOL sales representative.
3. Check the accessories
Please check the accessories according to the packing lists. If the accessories are
damaged or incomplete, please contact your RIGOL sales representative.
Recommended Calibration Interval
RIGOL suggests that the instrument should be calibrated every 12 months.
Adjust the supporting legs properly to use them as stands to tilt the multimeter
upwards for stable placement of the multimeter as well as better operation and
observation. You can also fold the supporting legs when the instrument is not in use
for easier storage or shipment, as shown in the figure below.
(a) Unfold the supporting legs
Figure 4.3 Adjusting the Supporting Legs
4.3.2To Connect to Power
The power requirements of the multimeter are DC, 12 V, 3 A. Please use the power
adaptor provided in the accessories to connect the multimeter to the AC power
source (100 V to 240 V, 50 Hz to 60 Hz), as shown in the figure below.
The power adaptor provided in accessories can only be used to power RIGOL instruments.
Do not use it for mobile phone and other devices.
WARNING
To avoid electric shock, please ensure that the instrument is correctly grounded.
4.3.3Turn-on Checkout
After the instrument is connected to the power source, press the power key at the
lower-left corner of the front panel to power on the instrument. During the start-up
process, the instrument performs a series of self-tests. After the self-test, the splash
screen is displayed.
Input: AC, 100 V to 240 V, 50 Hz to 60 Hz, 1.6 A
Output: 3 A@5 V/9 V/12 V/15 V, 3.25 A@20 V
Input: AC, 100 V to 240 V, 50 Hz to 60 Hz, 1.7 A
Output: 3 A@5 V/9 V/12 V/15 V, 3.25 A@20 V
•Reboot: Click or tap >Reboot. Then a prompt message "Are you sure to
reboot?" is displayed. Click or tap
OK to restart the instrument.
•Shutdown:
-Click or tap > Shutdown. Then a prompt message "Are you sure toshutdown?" is displayed. Click or tap OK to shut down the instrument.
-Press and a prompt message "Are you sure to shutdown?" is displayed.
Click or tap
OK to shut down the instrument.
-Press twice to shut down the instrument.
-Press for three seconds to shut down the instrument.
You can also click or tap >System>Settings and set the "Power Status" to "ON". The
instrument powers on once connected to power.
4.3.4To Set the System Language
This product supports multiple languages. You can click or tap > System >
Settings > Language to select the system language.
4.4Product Overview
Unless otherwise specified, this chapter takes DM858 as an example to introduce the
appearance and dimensions, front and rear panels, and user interface of DM858 and
DM858E.
4.4.1Front Panel Overview
1
12
11
2
3
4
5
6
7
8
10
9
1. 7-inch Capacitive Touch Screen
Displays the menu and parameter settings of the currently enabled function,
system state, prompt messages, and other information.
: indicates the trigger setting key. Press this key to open the trigger setting
menu. You can set the trigger mode (auto, single, and external), enable the reading
hold function, set the number of samples per trigger, set the delay time prior to
reading, the edge of the trigger input signal, and trigger output.
: indicates the system utility function setting. Press this key to open the
system utility setting menu. You can set the basic system settings and view the
system information.
: indicates the math operation key. Press this key to open the math
operation menu. You can perform the math operation (limit, statistics, dBm, and
dB) on the measurement results.
: indicates the storage operation key. Press this key to open the storage
setting key. You can save, load, upgrade the file; perform the disk management
operation to copy, cut, or delete files.
: indicates the measurement setting key. Press this key to open the
measurement parameter setting menu. You can configure the parameters for the
current measurement item.
: indicates the Help key. Press this key to enter the Help documentation.
You can get its help information by clicking on the link for the introduction of the
specified function.
3. Measurement Function Keys
: Press this key to enable the DC voltage (DCV) measurement.
: Press this key to enable the AC voltage (ACV) measurement.
: Press this key to enable the DC current (DCI) measurement.
: Press this key to enable the AC current (ACI) measurement.
: Press this key to enable the 2-wire resistance (Ω 2W) measurement. Press
> to enable the 4-wire resistance (Ω 4W) measurement.
: Press this key to enable the capacitance measurement. Press >
to enable any sensor measurement.
: Press this key to enable the continuity measurement. Press >
: Press this key to enable the frequency measurement. Press >
to enable the period measurement.
: Press this key and then press the specified secondary function to enable
the secondary measurement function. Press > to preset the current
instrument configuration, save it, and apply the configuration.
Under some of the measurement function keys are printed with the
specified alternative function names or labels. To enable the specified function
displayed under the key, first press , and then press the specified function
key. For example, as shown on the front panel, 4-wire resistance measurement
label is displayed under the 2-wire resistance measurement key. To enable the 4-
wire impedance measurement, first press
, and then press .
4. Confirmation Key
: Press this key to confirm the current parameter settings.
5. Back Key
: Press this key to exit the current operation or close the current dialog box
(if exists).
6. Range Selection and Arrow Keys
: Press this key to enable the auto range.
: indicates the Up/Down arrow keys.
Press the Up/Down arrow keys to increase/decrease the range manually.
In measuring DCV, DCI, ACV, ACI, Frequency or Period, you can use the Left/Right
arrow keys to increase/decrease the measurement speed.
7. Single/Local Key
unless you press it once again. To make the instrument to switch to the local mode,
press the Shift key first, and then press this key.
8. Auto Trigger/Reading Hold Key
DM858 Series User Guide
18
: indicates the Left/Right arrow keys.
: Press Single to trigger one single acquisition and will not trigger again
: Press this key to switch between auto trigger and readings hold functions.
- Run: The key is always illuminated and remains the auto-trig state unchanged.
In this state, the instrument continues acquiring data at the specified speed
based on the current settings.
- Hold: The key is blinking constantly and the acquisition stops. The
measurement data are displayed on the screen.
9. Signal Input Terminals
The signals (devices) under test are input to the instrument via these terminals. The
measurement connection methods are different for different objects under test.
10. Current Input Fuse
The front-panel current input fuse allows a max. of 10 A(DM858)/3.15 A(DM858E)
protection limit. Once the input current exceeds 13.5 A(DM858)/4.05 A(DM858E),
the fuse blows out.
11. USB HOST Interface
Allows to connect the USB storage device (FAT32 or NTFS format) or the mouse.
- USB storage device: imports or exports data (e.g. software update, waveforms,
setups, or screenshot files).
- Mouse: It can be connected to the instrument to realize remote control.
The multimeter outputs a pulse via [VM Comp] connector after each
7
3
4
5
6
measurement when VM output is enabled. To enable the VM output, press
on the front panel (you can also click or tap the Trigger label at the bottom of the
screen to enter the trigger setting interface), then click or tap "ON" for the Output
menu to enable the VMC output.
2. External Trigger Input Interface
Triggers the multimeter by connecting a pulse via the [EXT TRIG] connector after
external trigger is selected. To enable the external trigger, click or tap to select
External.
3. LAN Interface
Connects the instrument to network via this interface. The instrument is in
compliance with the standards specified in
used to set up a test system with other standard devices. Then you can use the
Web Control to send the SCPI commands to control the instrument.
4. USB DEVICE Interface
Connects the instrument to the PC via this interface. Then you can use the PC
software to send the SCPI commands or use the user-defined programming to
control the instrument.
The power requirements of the multimeter are DC, 12 V, 3 A. Please use the power
adaptor provided in the accessories to connect the multimeter to the AC power
source (100 V to 240 V, 50 Hz to 60 Hz).
6. Security Lock Hole
You can lock the instrument to a fixed location by using the security lock (please
purchase it by yourself) via the lock hole.
7. Ground Terminal
Connects with the instrument chassis and ground lead (power cord ground
terminal), which is in the grounded state.
8. Mounting Screw Holes (VESA 100 x 100)
Mounts the multimeter to the bracket with the screws.
Displays the measurement result of the current main measurement item.
2. Main Measurement Item
Displays the main measurement item. Click or tap the label, the main measurement
items are displayed. Click or tap to select the specified main measurement item to
be displayed in the main display area.
3. Main Measurement Range Selection
Displays the measurement range for the specified main measurement item.
4. Operating Status
Displays the current operating status of the instrument. It can be Auto Trig, Single
Trig, Ext Trig, Hold, or Stop.
5. Measurement Speed Selection
Displays the current measurement speed. Click or tap the label to select Slow,
Medium, or Fast from the drop-down list.
6. "Relative" Tab
Click or tap the tab to enable or disable the relative operation.
7. "Shift" Tab
Press
highlighted.
8. "2ND" Tab
DM858 Series User Guide
22
on the front panel to enable the Shift key. The "Shift" tab is
When you press the 2ND key on the front panel, the "2ND" tab is highlighted.
9. "Run/Hold" Tab
Click or tap the "Run/Hold" tab to switch the status from Run to Hold. You can also
press Run/Hold on the front panel to switch the status from Run to Hold.
10. "Single" Tab
Click or tap the "Single" tab to initiate one single trigger. You can also press Single
on the front panel to initiate one single trigger.
11. Notification Area
Displays the USB storage device icon, LAN connection, sound icon, and remote
control icon. Click or tap the area, the system setting menu is displayed.
-
USB storage device icon: When a USB storage device is detected,
displayed.
-
LAN connection icon: When the LAN interface is successfully connected, is
displayed.
- Sound icon: In the "System" menu, click or tap
Settings > Beeper to enable or
disable the sound. When the sound is turned on, will be displayed; when off,
will be displayed. Clicking or tapping the sound icon can also enable or
disable the sound.
- Remote control icon: When you control the instrument remotely,
displayed.
- Time: When you enable to display the time information, the system time is
displayed.
12. Shortcut Menus
Provides four fixed shortcut menus (
Measure, Display, Math, and Trigger) and
will be
will be
two customized menus. You can click or tap to customize the shortcut menu
that you want to add here.
13. Function Navigation Icon
Click or tap the icon to enter the function navigation menu. Click or tap the
desired menu to enter the specified menu setting interface.
Displays the measurement result of the 2ND measurement item when the 2ND
measurement is enabled.
17. Secondary (2ND) Measurement Item
Displays the secondary (2ND) measurement item. When you enable the 2ND
measurement, click or tap the label, then the 2ND measurement menu is
displayed.
4.5Touch Screen Gestures
The instrument's capacitive touch screen makes operation and configuration easy and
flexible. It features great convenience, high flexibility, and great sensitivity. The actions
supported by the touch screen controls include dragging and tapping.
4.5.1Drag
Use one finger to select the object, and then drag the object to a destination place, as
shown in
position (e.g. numeric keypad).
4.5.2Tap
Use one finger to tap the symbol or characters on the screen slightly, as shown in
Figure 4.10
Figure 4.9
. With the Tap gesture, you can perform the following operations:
. You can also drag the window controls to change the window
Figure 4.9 Drag Gesture
•Tap the menu displayed on the screen to operate on the menu.
•Tap the function navigation icon at the lower-left corner of the touch screen to
•Tap the displayed numeric keypad to set the parameters.
First check the input method type. If it shows "En/中", then go to Step 2; if it
shows "中/En", click or tap the input method switchover key to switch to "En/
中" (English input method).
2. Clear the name input area
If there is no character in the "Name Input Area", please go to the next step. If
there are characters in the "Name Input Area", click or tap the Backspace key
to delete all the characters from the "Name Input Area" in order.
3. Input the upper-case letter
If you want to input an upper-case letter, first use the Caps key to switch
between the upper-case and lower-case mode. If the Caps key is selected,
input the upper-case letter with the virtual keypad. If not, first click or tap the
Caps key to ensure it is selected, then input the upper-case letter. All the input
letters will be displayed in the "Name Input Area".
Input the lower-case letter
4.
Refer to the operation specified in the previous step. If the Caps key is not
selected, directly input the lower-case letter.
5. Input numbers or symbols
If the letter keypad is displayed, you need to click or tap the numeric
switchover key to switch to the numeric keypad, and input numbers or
symbols with the numeric keypad. All the input letters will be displayed in the
"Name Input Area".
6. Modify or delete the unwanted characters that have been input
During the name input process, you can modify or delete the unwanted
character if necessary. To delete the characters that have been input, click or
tap the Backspace key in the virtual keypad to delete the characters. To
modify the input characters, delete the unwanted characters and input the
desired characters again.
You can directly move the cursor to the character to be modified or deleted,
delete the desired character or input the new characters after deleting the
unwanted character.
•Input Chinese Characters
DM858 Series User Guide
26
7. Confirm the input
After completing the input operation, click or tap OK.
First check the Chinese and English Characters type. If it shows "中/En", then
go to Step 2; if it shows "En/中", click or tap the input method switchover key
to switch to "中/En" (Chinese input method).
2. Clear the name input area
If there is no character in the "Name Input Area", please go to the next step. If
there are characters in the "Name Input Area", click or tap the Backspace key
to delete all the characters denoted "Name Input Area" in order.
3. Input Chinese characters
Click or tap the characters in the virtual keypad to input Pinyin into the input
area, then the characters to be selected will be displayed in the Chinese
character selection area. Slide to view more Chinese characters for you to
choose. Select the desired Chinese character, and then the selected character
will be displayed in the input area.
4. Modify or delete the unwanted characters that have been input
During the name input process, you can modify or delete the unwanted
character if necessary. To delete the characters that have been input, click or
tap the Backspace key in the virtual keypad to delete the characters. To
modify the input characters, delete the unwanted characters and input the
desired characters again.
5. Confirm the input
After completing the input operation, click or tap OK.
Input a Value
When setting or modifying a parameter, use the pop-up numeric keypad to input an
appropriate value.
•Click or tap the value or unit in the virtual numeric keypad to complete the input.
Input Area
Delete
Clear
Enter
Decimal
Point
Number
Key
Plus or
minus sign
Unit(Optional)
Figure 4.13 Numeric Keypad Input Interface
After you input all the values and select the desired units, the virtual numeric keypad
is turned off automatically. This indicates that you have completed parameter setting.
In virtual the numeric keypad, you can perform the following operation:
1. Input Area
Click or tap the desired number and the desired value is input in this area.
2. Number Key
There are 10 numbers in the numeric key area. Click or tap the number to input the
desired value.
3. Decimal Point
- To input a decimal number less than 1, input the decimal point first and then
add a number.
- To input a real number with decimals, input an real number first and then add a
- The dot can be added to the specified position when you input a value that is
- It is not allowed to input several decimal points.
4. Plus or minus sign (±)
DM858 Series User Guide
28
decimal point. The number of zero digits is automatically added after the
decimal point according to the rules of the specified parameter.
Your first click or tap of the sign will input a minus (-) sign, the second time of click
or tap of it will input a plus sign, but the plus sign is always omitted for the
number display.
5. Unit
Click or tap the desired unit.
6. Enter
After inputting the file or folder name, click or tap Enter directly to confirm the
name input. Then it automatically goes to the previous page.
7. Clear (C)
Click or tap C to clear all the input.
8. Delete (x)
Click or tap this icon to clear the input. Move the cursor to the position of the
number that needs to be deleted, then click or tap this icon to clear the digit
ahead of the cursor.
4.7To Use the Security Lock
If necessary, you can lock the instrument to a fixed location by using a standard
laptop security lock (please purchase it by yourself), as shown in the figure below.
The method is as follows: align the lock with the lock hole and plug it into the lock
hole vertically, turn the key clockwise to lock the instrument, and then pull the key
out.
For the specified measurement item, the range can be set to Auto or specified values.
You can set the range to Auto, and then the proper range for the current
measurement will be automatically selected. It is convenient for you to make
measurements. You can also select a specified range manually to obtain higher
reading precision.
The range selection keys are located at the right part of the front panel, as shown in
the figure below.
Specified range selection: Press or to select the available range.
You can also set the range for each measurement item in the specified measurement
interface.
For each measurement item displayed on the screen, click or tap the Measure label at
the bottom of the screen to enter the specified measurement item interface. Click or
tap the drop-down list of
Range to select AUTO or the specified range value.
Remarks:
•When the input signal exceeds the current range (except the Frequency
DM858 Series User Guide
32
Figure 5.2 Range Selection Menu
measurement), a message "OVERLOAD" is displayed on the screen.
Table 5.2 DC Voltage Measurement Parameter Setting
ParameterDescription
Range*Auto, 100 mV, 1 V, 10V, 100 V, 1000V
Available to set when the Range is a specified value rather than
Auto.
10 GΩ, 11.2 MΩ
Impedance
- When you select "11.2 MΩ", the input impedance for all the
ranges is 11.2 MΩ.
- When you select "10 GΩ", the input impedance for 100 mV
and 1 V ranges is "10 GΩ", but the input impedance for the
10V, 100 V, 1000V ranges is "11.2 MΩ".
RateFast, Medium, Slow
NOTE
-20% over range on all ranges except 1000 V range.
-The range can be set to Auto or a specified value (available range options for Range
menu).
-A 1000 V input protection is available in all ranges and a 20% overrange is available on
all ranges except 1000 V.
-If the reading exceeds 1000 V in 1000 V range, "OVER LOAD" will be displayed.
4. Click or tap to enable or disable the relative operation function. When
enabled, the displayed reading value is the result obtained by subtracting the set
relative value from the actual measurement value. For the relative value setting
methods, refer to descriptions in "
Math Operation
".
5. Read the measured value
When reading the measurement results, you can use the arrow keys to switch the
reading rate.
6. Hold the reading data
Press
to run and hold the reading data displayed on the screen.
5.3.2To Measure DC Current
The instrument can measure up to 10 A(DM858)/3 A(DM858E) DC current. The
following section introduces the connection methods and measurement methods of
DC Current.
Figure 5.6 DC Current Measurement Connection Diagram
3. Click or tap to set the measurement parameters of DC current: Range and
Rate.
Table 5.3 DC Current Measurement Parameter Setting
ParameterRemarks
Range*
Auto, 100 μA, 1 mA, 10 mA, 100 mA, 1 A, 10 A (DM858)/3 A
(DM858E)
RateFast, Medium, Slow
NOTE
20% over range on all ranges except 10 A range. The range can be set to Auto or the given
specified values.
4. Click or tap to enable or disable the relative operation function. When
enabled, the displayed reading value is the result obtained by subtracting the set
relative value from the actual measurement value. For the relative value setting
methods, refer to descriptions in "
Math Operation
".
5. Read the measured value
When reading the measurement results, you can use the arrow keys to switch the
reading rate.
6. Hold the reading data
Press
DM858 Series User Guide
38
to run and hold the reading data displayed on the screen.
Figure 5.8 AC Voltage Measurement Connection Diagram
3. Click or tap to set the measurement parameters of AC voltage: Range and
Rate
Table 5.4 AC Voltage Measurement Parameter Setting
ParameterDescription
Range*Auto, 100 mV, 1 V, 10 V, 100 V, 750 V
RateFast, Medium, Slow
NOTE
-20% over range on all ranges except 750 V range.
-The range can be set to Auto or a specified value (available range options for Range
menu).
-A 750 V input protection is available in all ranges and a 20% overrange is available on all
ranges except 750 V.
-If the reading exceeds 750 V in 750 V range, "OVER LOAD" will be displayed.
4. Click or tap to enable or disable the relative operation function. When
enabled, the displayed reading value is the result obtained by subtracting the set
relative value from the actual measurement value. For the relative value setting
methods, refer to descriptions in "
When reading the measurement results, you can use the arrow keys to switch the
reading rate.
6. Hold the reading data
Press to run and hold the reading data displayed on the screen.
5.3.4To Measure AC Current
The instrument can measure up to 10 A(DM858)/3 A(DM858E) AC current . The
following section introduces the connection methods and measurement methods of
AC current.
Operation Procedures:
1. To enter the AC current measurement interface, perform either one of the
following two operations.
Front Panel Operation
- Press on the front panel;
- Click or tap the main measurement item shown at the left part of the main
display area. The measurement items are displayed. Click or tap to select
to enter the AC current measurement interface.
Figure 5.9 AC Current Measurement Interface
2. Connect red lead to terminal Input-I and black lead to terminal Input-LO, as shown
20% over range on all ranges except 10 A range. The range can be set to Auto or the given
specified values.
4. Click or tap to enable or disable the relative operation function. When
enabled, the displayed reading value is the result obtained by subtracting the set
relative value from the actual measurement value. For the relative value setting
methods, refer to descriptions in "
Math Operation
".
5. Read the measured value
When reading the measurement results, you can use the arrow keys to switch the
reading rate.
6. Hold the reading data
Press
DM858 Series User Guide
42
to run and hold the reading data displayed on the screen.
The instrument provides 2-wire (2WR) and 4-wire (4WR) resistance measurements.
The following section introduces the connection methods and measurement methods
of 2WR and 4WR resistance measurements.
5.3.5.12WR Measurement
Operation Procedures:
1. To enter the 2WR measurement interface, perform either one of the following two
operations.
- Press on the front panel;
- Click or tap the main measurement item shown at the left part of the main
Front Panel Operation
display area. The measurement items are displayed. Click or tap to select
to enter the 2WR measurement interface.
Figure 5.11 2WR Measurement Interface
2. Connect red lead to terminal Input-HI and black lead to terminal Input-LO, as
, 5
when the range is
set to 1 MΩ or
above and when
the measurement
rate is Slow)
4. Click or tap to enable or disable the relative operation function.
When enabled, the displayed reading value is the result obtained by subtracting
the set relative value from the actual measurement value. For the relative value
setting methods, refer to descriptions in "
DM858 Series User Guide
44
NOTE
20% overrange in all ranges. You can select Auto or a specified range value.
If the measured resistance is small, Relative operation is recommended to reduce the error
caused by test lead. If no oRelative operation is used, an additional error of 0.2 Ω is added.
To test the resistance above 1 MΩ, avoid external interference.
5. Read the measured value
When reading the measurement results, you can use the arrow keys to switch the
reading rate.
6. Hold the reading data
Press to run and hold the reading data displayed on the screen.
5.3.5.24WR Measurement
Front Panel Operation
When the measured resistance is lower than 100 kΩ, the 4-wire resistance
measurement is recommended to reduce the measurement error caused by test lead
resistance and contact resistance between the probe and the testing point because
these two resistances cannot be ignored comparing with the measured resistance.
Operation Procedures:
1. To enter the 4WR measurement interface, perform either one of the following two
operations.
- Press > to enter the 4WR measurement interface.
- Click or tap the main measurement item shown at the left part of the main
display area. The measurement items are displayed. Click or tap to select
20% overrange in all ranges. You can select Auto or a specified range value.
4. Click or tap to enable or disable the relative operation function.
When enabled, the displayed reading value is the result obtained by subtracting
the set relative value from the actual measurement value. For the relative value
setting methods, refer to descriptions in "
Math Operation
".
5. Read the measured value
When reading the measurement results, you can use the arrow keys to switch the
reading rate.
6.
Hold the reading data
Press to run and hold the reading data displayed on the screen.
NOTE
Please do not put the terminals of the resistance on the conductive plane or in your hand
to avoid error. The larger the resistance, the more the impact on the measurement results.
5.3.6To Measure Capacitance
The instrument can measure up to 1 mF/10 mF (only available for DM858)
capacitance. The following section introduces the connection methods and
measurement methods of capacitance.
Operation Procedures:
1. To enter the capacitance measurement interface, perform either one of the
Auto, 1 nF, 10 nF, 100 nF, 1 μF, 10 μF, 100 μF, 1 mF, and 10 mF
(only available for DM858)
NOTE
20% overrange on all ranges.
4. Click or tap to enable or disable the relative operation function. When
enabled, the displayed reading value is the result obtained by subtracting the set
relative value from the actual measurement value. For the relative value setting
methods, refer to descriptions in "
Math Operation
".
5. Read the measured value
The capacitance measurement is measured at a fixed "slow" rate with a 3.5-digit
display. Therefore, the reading rate cannot be adjusted during measurement
results reading.
6.
Hold the reading data
Press
to run and hold the reading data displayed on the screen.
TIP
Before measuring the electrolytic capacitor, you should first short-circuit the two pins of the
electrolytic capacitor to discharge it.
5.3.7To Test Continuity
The continuity test checks that the resistance in the DUT is low enough to be
considered continuous. In this case, the beeper sounds, if enabled, to inform you of
the continuity situation.
Operation Procedures:
1. To enter the continuity measurement interface, perform either one of the following
perform the continuity test. If you do not need to modify the parameter, go to the
next step.
Table 5.9 Continuity Test Parameter Settings
ParameterDescription
Short Circuit Resistance10 Ω (default); from 1 Ω to 1,000 Ω
Enables or disables the beeper. If the
Beeper
beeper is enabled, beeper sounds when
the circuit is considered to be
continuous.
NOTE
The short-circuit resistance value is stored in the non-volatile memory and will remain
unchanged at power-off.
4. Measurement Display Results
For the open circuit or measured value exceeds 1.2 kΩ, OPEN is displayed on the
screen; When the measured value is smaller than the set short-circuit resistance,
the actual resistance value is displayed on the screen. At this time, if the beeper is
enabled, the beeper sounds. When the measured value is greater than the set
short-circuit resistance and less than 1.2 kΩ, the beeper does not sound whether it
is enabled or not, and the actual resistance value is displayed on the screen.
5.3.8To Check Diode
Operation Procedures:
1. To enter the diode measurement interface, perform either one of the following two
operations.
- Press > on the front panel to enter the diode measurement
Interface.
- Click or tap the main measurement item shown at the left part of the main
display area. The measurement items are displayed. Click or tap to select
- OPEN: when OPEN is displayed on the screen, it indicates that the diode is
open-circuit or the measured forward voltage drop is less than 0.1 V or greater
than 2.1 V.
- Measured Value: when the measured forward voltage drop is within the range
of 0.1 V and 2.1 V.
- Measured Value and with Beeper On: when the measured values are displayed
on the screen, with beeper on, the measured forward voltage drop is within the
range of 0.1 V and 2.1 V.
5.3.9To Measure Frequency/Period
Front Panel Operation
The frequency/period of a signal could be measured with the frequency/period
measurement function. The following section introduces how to perform Frequency
and Period measurement with the instrument.
5.3.9.1Measure Frequency
Operation Procedures:
1. To enter the frequency measurement interface, perform either one of the
following two operations.
- Press on the front panel;
- Click or tap the main measurement item shown at the left part of the main
display area. The measurement items are displayed. Click or tap to select
4. Click or tap to enable or disable the relative operation function.
When enabled, the displayed reading value is the result obtained by subtracting
the set relative value from the actual measurement value. For the relative value
setting methods, refer to descriptions in "
Math Operation
".
5. Read the measured value
The frequency measurement is measured at a fixed "slow" rate with a 5.5-digit
display. Therefore, the reading rate cannot be adjusted during measurement
results reading.
6. Hold the reading data
Press
5.3.9.2Measure Period
Operation Procedures:
1. To enter the period measurement interface, perform either one of the following
two operations.
- Press > on the front panel;
- Click or tap the main measurement item shown at the left part of the main
display area. The measurement items are displayed. Click or tap to select
to run and hold the reading data displayed on the screen.
4. Click or tap to enable or disable the relative operation function.
When enabled, the displayed reading value is the result obtained by subtracting
the set relative value from the actual measurement value. For the relative value
setting methods, refer to descriptions in "
5. Read the measured value
The period measurement is measured at a fixed "slow" rate with a 5.5-digit
display. Therefore, the reading rate cannot be adjusted during measurement
results reading.
6. Hold the reading data
Math Operation
".
Press
to run and hold the reading data displayed on the screen.
5.3.10Any Sensor Measurement
By using any sensor measurement, you can connect Pressure Sensor, Flux Sensor or
Temperature Sensor easily. This function enables you to easily convert a measured
physical quantity (such as pressure, flow rate and temperature) into a quantity that
can be easily measured (such as voltage, current or resistance). By pre-inputting a
response curve, the multimeter converts and amends the data according to the
internal arithmetic and displays the physical quantity measured on the screen. Users
can edit and modify the display unit of the measured physical quantity at will.
The multimeter supports DCV, DCI, Freq, 2WR, 4WR, and temperature sensor
measurement (TC, RTD, Therm). Meanwhile, there are 10 groups of basic Sensor
preset within the instrument.
Connection
Different sensors have different connection methods. For the Voltage Sensor,
Resistance Sensor, Thermocouple Sensor and Frequency Sensor, please connect the
red lead to terminal Input-HI and black lead to terminal Input-LO, as shown in the
figure below.
, click or tap Temperature Sensor to enter the temperature sensor
setting interface.
Select the temperature sensor type. As shown in
down list of
Measurement Type to select "TC", "RTD", or "Therm". By default, TC is
Figure 5.28
, click or tap the drop-
selected.
•TC
-The TC (thermocouple) is one of the most commonly used industrial
temperature detection device. It converts the temperature value into
voltage reading, and has a very large temperature measurement range.
The common thermocouples are B (Pt Rh 30- Pt Rh6), E (NI CR-WRCK), J
(Fe-WRCK), K (NI CR-NiSi), N (NiCrNi-NiSi), R (Pt Rh13 -Pt), S (Pt Rh10-Pt)
and T (Cu-WRCK). Among them, B, R and S are precious metal
thermocouple, while E, J, K, N and T are cheap metal thermocouples. For
the reference table about each thermocouple, please see the ITS-90
standards. By default, B-ITS-90 is selected.
DM858 Series User Guide
60
Click or tap the drop-down list of
Measurement Type to select "TC".
-User-defined Cold Junction Temperature
-Enables or disables the user-defined cold junction temperature. If the
not checked, the input field is grayed out and disabled. You can only
use the built-in cold junction temperature.
-If checked, click or tap the input field of Cold Junction Temperature,
then input the user-defined value with the pop-up numeric keypad.
NOTE
Theoretically, TC is measured at 0°C at the cold side. However, the instrument is
usually at room temperature during measurement rather than 0°C.
This has caused the thermopotential difference to decrease, resulting in
measurement inaccuracy or error.
Therefore, it is necessary to make the cold-junction compensation to reduce the
error.
-Temperature Unit: sets the temperature unit. The available units are ℃, ℉,
and K. By default, the unit is ℃.
-Display Meas. Value:
enables or disables the display of the measurement value on the screen.
-ON: enables the display of the measurement value on the screen.
-OFF: disables the display of the measurement value on the screen.
-Apply
Click or tap Apply to apply the current setting.
•RTD
-RTD is a temperature detector commonly used in medium-low
temperature range. It features in high measurement accuracy and stable
performance.
It measures the temperature depending on the changing resistance of the
measured substance with the temperature. The multimeter will display the
temperature corresponding to the resistance when the resistance changes.
This kind of sensor has good linearity.
The multimeter adopts the approximate arithmetic of IEC751 standards
and converts the resistance signal from the temperature sensor into the
related temperature for displaying.
For the reference table about each RTD, please refer to the related
standards.
-ON: enables the display of the measurement value on the
screen.
-OFF: disables the display of the measurement value on the
screen.
-Apply
Click or tap Apply to apply the current setting.
•Therm
-The THERM (thermistor) converts the temperature variation information
into resistance variation information. It is especially sensitive to
temperature.
In this measurement, the multimeter adopts Steinhart-Hart approximate
arithmetic and converts the resistance signal from the temperature sensor
into the related temperature for displaying.
To use the THERM sensor, you need to set the resistance parameter of the
multimeter and select the connection mode.
Click or tap the drop-down list of
Measurement Type to select "Therm".
Figure 5.30 Therm Temperature Sensor Setting Interface
-Resistance: sets the resistance at temperature 25℃. The available
choices are 2.2 kΩ, 3 kΩ, 5 kΩ, 10 kΩ, and 30 kΩ. By default, it is 2.2
Sets the name of the sensor. Click or tap the input field of Name to enter the
name setting interface.
Figure 5.32 User-defined Sensor Name Setting
a. Input the sensor name. The length of the name shall not exceed 15 characters
and cannot be the same as the existing sensor name.
b. Click or tap OK to confirm input.
2. Sensor Type
Sets the sensor type. The available sensors are DCV, DCI, 2WR, 4WR, FREQ, and
TC. By default, it is DCV.
3. Sensor Unit
Sets the sensor unit. The available units are ℃ (degree Celsius), ℉(degree
Fahrenheit), K, %, °(degree), and Customize. By default, it is °. When you select
"Customize", click or tap the edit field next to the Customize field, then a virtual
keypad is displayed. Input the user-defined unit with the virtual keypad.
4. Display Meas. Value
Enables or disables the display of the measurement value on the screen.
- ON: enables the display of the measurement value on the screen.
- OFF: disables the display of the measurement value on the screen.
5.
Reference Value
Sets the reference value. Click or tap
to enter the reference value setting
interface.
Click or tap to delete the specified line of the reference values.
- Displayed Value: indicates the displayed value of the sensor. The settable
range is from -999999 to +999999.
Figure 5.34 Reference Value Setting
After setting the measured value and displayed value, click or tap OK to confirm
the reference value setting, as shown in the figure below.
6. Apply the setting
Click or tap Apply to apply the current setting.
7. Sensor management
Click or tap Sensor Manager to enter the user-defined sensor setting interface.
The instrument supports at most 10 user-defined sensors. Each time only one
sensor can be selected.
- Save the Sensor
Click or tap
Save to save the selected sensor. After being saved, "saved" in
bracket is added following the saved sensor name.
Figure 5.35 Customize the Sensor
- Apply the Sensor
1. Click or tap Apply to apply the user-defined sensor setting.
When you enable the 2ND measurement, press , the results of the 2ND
measurement is displayed in the secondary display area of the screen, as shown in the
following figure.
Figure 5.37 Main and Secondary Display Areas
As shown in the figure above, the section where the measured item and values are
displayed in yellow is considered as the main display area, and the section below the
main display area is considered as the secondary display area.
Click or tap
measurement item and the secondary measurement item. For details, refer to
Interface Overview
in the secondary display area to switch the position of the basic
User
.
The basic measurement types displayed in the main display area are all supported by
the 2ND measurement. In some specific conditions, some of the basic measurement
types are disabled for 2ND measurement. For details, see the table below.
Table 5.14 Basic Measurement Types and 2ND Measurement Types
1. If the measurement type for the basic measurement and 2ND measurement is the
same:
- The measurement data are updated synchronously in both the main display
area and the secondary display area.
- If the math operations are adopted for the basic measurement types, then the
operation results are displayed in the main display area and the measurement
values before the operation are displayed in the secondary display area. For
details, refer to
User Interface Overview
.
2. If the measurement type for the basic measurement and 2ND measurement is
different:
- The two types of measurements are carried out alternatively. The measurement
data in the main display area and the secondary display area are updated
respectively.
- If math operations such as Stat, Limit, dBm, and dB are adopted and Relative
operation is disabled for the basic measurement types, then the measurement
values for the main measurement types are displayed in the main display area
and the measurement values for the secondary measurement types are
displayed in the secondary display area.
- If the math operations such as Stat, Limit, dBm, and dB are adopted and
Relative operation is enabled for the basic measurement types, then the
Relative operation results are displayed in the main display area and the
measurement values for the secondary measurement types are displayed in the
secondary display area.
3. When the basic measurement type is Sensor, after setting the sensor parameters,
the operation results (Relative Value) of the sensor is displayed in the main display
area, and the current measurement value is displayed in the secondary display
area. For details, refer to
4. When the basic measurement type is ACV and the secondary measurement type is
FREQ/PERIOD; or when the basic measurement type is FREQ/PERIOD and the
secondary measurement type is ACV, the measurement values are mainly displayed
for the basic measurement types.
5. By default, the secondary measurement types adopts Auto range. If the main and
secondary measurement types are the same, then the secondary measurement
type use the same range as the basic meaurement type.
6. The measurement data of the secondary measurement types do not support
Holding operation.
5.5Preset Mode
To Measure User-defined Sensor
.
The instrument provides the preset setting. Preset mode applies to production lines
to avoid misoperations.
The preset mode allows you to preset 10 user-defined settings.
Figure 5.38 Preset Interface
Preset Settings
If User1 is defined to preset DCV measurement configurations and User2 is defined to
preset ACV measurement configurations:
After configuring the DCV measurement parameters, click or tap User1>Save,
"saved" in brackets is added following User1, which indicates that the DCV
configurations are saved as User1(saved) preset file. At this time,
Apply and Delete
menu are added to the interface. To load the DCV configurations that you set, click or
Apply to load the DCV measurement configurations. Click or tap Delete to delete
tap
the user-defined DCV presetting.
After configuring the ACV measurement parameters, click or tap
User2>Save,
"saved" in brackets is added following User2, which indicates that the ACV
configurations are saved as User2(saved) preset file.
At this time,
configurations that you set, click or tap
Apply and Delete menu are added to the interface. To load the ACV
Apply to load the ACV measurement
configurations. Click or tap Delete to delete the user-defined ACV presetting.
The multimeter provides five math functions: Statistics (Max, Min, Avg, standard
deviation, and count), Limit, dBm, dB, and Relative operation. You can select different
math functions for different measurement items.
The math operations are applicable to the voltage, current, resistance, capacitance,
frequency/period, and sensor measurements. However, dB and dBm operations are
only applicable to the voltage measurement.
NOTE
The math operation is diabled when the range is AUTO.
Press or click/tap on the screen, then the math operation interface is
displayed, as shown in
Figure 5.40
Front Panel Operation
.
Figure 5.40 Math Operation Setting
Table 5.15 Description of Math Operations
ParameterRemarks
Makes statistics of the current measurement, including maximum
value, minimum value, average value, standard deviation, and
count (current count of the samples).
73
Page 86
Front Panel Operation
ParameterRemarks
Limit
dBm
Sets the upper and lower limits and makes limit operation to
identify those signals that are beyond the limits by sounding an
alarm. It is applicable to DCV, ACV, DCI, ACI, Resistance,
Frequency, Capacitance, and temperature measurements.
The dBm function is logarithmic, comparing power delivered to a
reference resistance,
relative to 1 milliwatt:
dB
RelativeEnables or disables the Relative operation.
5.6.1Statistics
The Statistics operation calculates the min. value, max. value, average value, standard
deviation, and reading count during the measurement.
The multimeter can perform Statistics operation on the following measurement items:
DCV, ACV, DCI, ACI, Resistance, Frequency, Period, Capacitance, and any sensors.
Press
enter the math operation interface. Then, select Stat from the drop-down list of
Operation Type.
Each dB measurement is the difference between the input signal
and a stored reference value, with both values converted to dBm.
to enter the math operation interface. You can also click or tap to
The operation results are displayed in the secondary display area, as shown in the
following figure.
Figure 5.42 Statistics Operation Result
•Max: indicates the max. reading.
•Min: indicates the min. reading.
•Avg: indicates the average value of the reading.
•Std: indicates the standard deviation of the reading.
•Count: indicates the reading count.
NOTE
•When statistics operation is selected, the multimeter takes the first reading as the
•The max. value, min. value, average value, and sample count are all stored in the volatile
5.6.2Limit
The limit operation allows you to set the specified upper limit and lower limit. Once
the signal beyond the set limits appears, the beeper sounds. The multimeter can
perform limit operation on the following measurement items: DCV, ACV, DCI, ACI,
Resistance, Frequency, Period, Capacitance, and temperature sensor.
Press to enter the math operation interface. You can also click or tap to
enter the math operation interface. Then, select Limit from the drop-down list of
Operation Type.
maximum (or minimum) for displaying. The reading shown on the screen is always the
maximum (or minimum) within the current readings when measuring multiple
When inputting a positive or negative value, click or tap the "±" sign to switch
between the positive and negative value sign.
The limit value unit is determined by the currently selected measurement item.
When the limit value you set exceeds the specified one, the beeper sounds (if
enabled), and an error message "-222 Data out of range" is displayed.
2. Math operation result display
When you perform the math operation, the math operation results are displayed in
the secondary display area, as shown in the figure below.
Figure 5.44 Limit Operation
Lower Failures: indicates the counts of below the defined lower limit.
Upper Failures: indicates the counts of above the defined upper limit.
Status: indicates the status of the limit operation.
Pass count: indicates the counts of values that are within the defined upper limit
and lower limit.
When the count exceeds the upper limit, the secondary display area displays "high
fail"; when the count exceeds the lower limit, the secondary display area displays
"low fail"; when the count does not exceed the upper limit or lower limit, it shows
"pass". If the beeper is enabled, when the count exceeds the upper limit or is
below the lower limit, the beeper sounds an alarm.
5.6.3dBm
dBm represents the absolute value of the power. The dBm operation calculates the
power of the reference resistance according to the measured voltage. The dBm
function applies to AC voltage and DC voltage measurements only.
Press to enter the math operation interface. You can also click or tap to
enter the math operation interface. Then, select dBm from the drop-down list of
Operation Type.
dB represents the relative value which is used in the Relative operation of dBm value.
The dB function applies to AC voltage and DC voltage measurements only.
Press to enter the math operation interface. You can also click or tap to
enter the math operation interface. Then, select dB from the drop-down list of
Operation Type.
When enabled, the multimeter calculates the dBm value of the next reading and
subtracts the relative dB value from this value and then displays the result.
= 10 x Log10 [(Reading2 /R
dB
The range of the relative dB value is from
-120.00 dBm to 120.00 dBm. By default,
it is 0 dBm.
)/0.001W] - dB relative value
REF
Wherein, Reading denotes the measured voltage and R
denotes the reference
REF
resistance.
Take DCV measurement as an example, the calculation results are displayed in the
secondary display area, as shown in the following figure.
NOTE
Relative dB Value
•Input a relative dB value with a pop-up numeric keypad, and save it as a relative dB value.
•The relative dB value is stored in the volatile memory and will be cleared at power-off.
DM858 Series User Guide
80
Figure 5.48 dB Operation Results
The range of the relative dB value is from -120 dBm to 120 dBm. By default, it is 0 dB.
The reading displayed on the screen in Relative operation is the difference between
measured and relative values.
Enable the relative operationTo enable the relative operation, perform the following
operations:
Front Panel Operation
•Click or tap
is highlighted in yellow box.
•In the math operation menu, click or tap ON for the Relative Switch menu.
Displayed Value = Measured Value - Relative Value
to enable the relative operation. The Relative menu label
Figure 5.49 Relative operation
The multimeter can perform relative operation on the following measurement items:
DCV, ACV, DCI, ACI, Resistance, Frequency, Period, Capacitance, and temperature
sensor.
The ranges of the relative values in Relative operation are shown in the table below:
Table 5.20 Relative Value Setting Range
Measurement ItemSetting Range
DCV
ACV
DCI/ACI
2WR/4WR
(-120% of the current range) to (120% of
the current range)
(-120% of the current range) to (120% of
the current range)
(-120% of the current range) to (120% of
the current range)
(-120% of the current range) to (120% of
the current range)
As shown in the figure above, data are grouped in bins represented by vertical
bars in the histogram display. The data are displayed on the left side of the
histogram. Read the data from left downward.
- 82.6%: indicates the percentage of samples in the largest bin.
- 419: indicates the number of samples in the largest bin.
- Total 507: indicates the total number of samples.
- #bin 10: indicates the total number of bins.
For different measurement modes, the max. number of samples supported by the
histogram is different. See the following table for details.
Table 5.21 Max. Number of Samples Displayed in the Histogram
Measurement Rate Max. Number of Samples
Capacitance
Measurement
Frequency/Period
Measurement
Other
Measurements
Slow6145150
MediumN/AN/A1050
FastN/AN/A
The measurement rate for Capacitance, Period, and Frequency measurements is Slow.