Review the following safety precautions carefully before operating the instrument
to avoid any personal injuries or damages to the instrument and any products
connected to it.
To avoid potential hazards use the instrument as specified by this user’s guide
only.
The instrument should be serviced by qualified personnel only.
To Avoid Fire or Personal Injury.
Use Proper Power Cord. Use the power cord designed for the instrument as
authorized in your country only.
Ground The Instrument. The instrument is grounded through the grounding
conductor of the power cord. To avoid electric shock the instrument grounding
conductor(s) must be grounded properly before making connections to the input
or output terminals of the instrument.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all
ratings and marks on the instrument. Follow the user’s guide for further ratings
information before making connections to the instrument.
Do Not Operate Without Covers. Do not operate the instrument with covers
or panels removed.
Use Proper Fuse. Use the fuse of the type, voltage and current ratings as
specified for the instrument.
Avoid Circuit or Wire Exposure. Do not touch exposed connections and
components when power is on.
Do Not Operate With Suspected Failures. If suspected damage occurs with
the instrument, have it inspected by qualified service personnel before further
operations.
The disturbance test of all the models meet the limit values of A in the
standard of EN 61326: 1997+A1+A2+A3, but can’t meet the limit
values of B.
WARNING
IEC Measurement Category II. The HI and LO input terminals may be connected
to mains in IEC Category II installations for line voltages up to 300 VAC. To avoid
the danger of electric shock, do not connect the inputs to mains for line voltages
above 300 VAC.
Protection Limits: To avoid instrument damage and the risk of electric shock, do
not exceed any of the Protection Limits defined in the following section.
IEC Measurement Category II Overvoltage Protection
To protect against the danger of electric shock, the RIGOL DM3000 series
Digital Multimeter provides overvoltage protection for line-voltage mains
connections meeting both of the following conditions: The HI and LO input
terminals are connected to the mains under Measurement Category II conditions,
defined below, and The mains are limited to a maximum line voltage of 300 VAC.
IEC Measurement Category II includes electrical devices connected to mains at
an outlet on a branch circuit.
Such devices include most small appliances, test equipment, and other devices
that plug into a branch outlet or socket. The DM3000 series Digital Multimeter
may be used to make measurements with the HI and LO inputs connected to
mains in such devices, or to the branch outlet itself (up to 300 VAC). However,
the DM3000 series Digital Multimeter may not be used with its HI and LO inputs
connected to mains in permanently installed electrical devices such as the main
circuit-breaker panel, sub-panel disconnect boxes, or permanently wired motors.
Such devices and circuits are subject to overvoltage that may exceed the
protection limits of the DM3000 series Digital Multimeter.
Note: Voltages above 300 VAC may be measured only in circuits that are isolated
from mains. However, transient overvoltage are also present on circuits that are
isolated from mains. The DM3000 series Digital Multimeter is designed to safely
withstand occasional transient overvoltage up to 2500 Vpk. Do not use this
equipment to measure circuits where transient overvoltage could exceed this
level.
The book covers the following description and six models DM3000 Series Digital
Multimeter:
DM3061, DM3062, DM3064;
DM3051, DM3052, DM3054.
DM3000 Series Digital Multimeter naming rules:
Prefix desktop Digital Multimeter
Serial Number
6-6½, 5-5¾ digit
No.
1-Basic; 2-LAN/GPIB interface;
4-Inspection plate with the model and LAN/GPIB interface.
Application examples:
DM3061- 6½ DM3000 series, Basic type.
DM3062- 6½ DM3000 series, Basic type, equipped with LAN/GPIB module.
DM3064 - 6½ DM3000 series, Basic type, equipped with LAN/GPIB and
inspection module.
DM3051- 5¾ DM3000 series, Basic type.
DM3052- 5¾ DM3000 series, Basic type, equipped with LAN/GPIB module.
DM3054 - 5¾DM3000 series, Basic type, equipped with LAN/GPIB and
inspection module.
RIGOL DM3000-Series Digital Multimeter is a equipment designed for
high-precision, multifunction, automation measurements. The series includes 6½
digits multimeter, with high-speed data acquisition, automatic measurements,
multiplexer, mathematical operations, and flexible user sensor configurations etc.
Interface includes RS-232, USB, LAN, GPIB for disk storage and print.
The DM3000 has a high-resolution monochrome LCD display system for simple
waveform display and recording. The concise and user-friendly layout of the front
panel has a keyboard, and back lighted functional buttons, embedded with
operating instructions makes the instrument more flexible, and capable. The
50kSa/s high data sampling rate allows to capture precision audio waveforms and
high speed data. It has 2Mbyte of internal memory depth while the external
memory depth is expandable as preferred. The AC voltage and current
measurement is true RMS. It supports virtual terminal display and control, and
remote network access.
With the performance and characteristics given below, you will understand how a
DM3000 can satisfy your measurement requirements.
z50kSa/s data sampling rate can be used, such as the rapidly changing
high-precision audio waveform data. Meanwhile waveform can be displayed
on LCD Screen
z Resolving Index: > 6½ digits and 2,400,000 Count
z 24 measurement functions
DC voltage and current, AC voltage and current, two-wire and four-wire
Upper limit and lower limit on threshold measurement
Arithmetic include: maximum, minimum, limit, average, dBm, dB
Data acquisition functions include : data records, inspection, automatic
measurement.
z True RMS AC voltage and current measurement
z 16-Channels inspection functional measurement and control software
(optional)
z DC voltage >10GΩ input impedance to achieve the range of 48V (±24V)
z 10 groups measuring set-up storage and unlimited setup through PC
z 256 x 64 pixel monochrome LCD
z I/O: RS-232, USB, LAN and GPIB
z Built-in USB Host to support USB disk and USB printer
z Simple, convenient, flexible control software: Ultralogger, Ultrasensor and
UltraDMM Supports for Microsoft® Windows 98/Me/2000/XP
Note: The chapter one, chapter two, chapter three will be described according to
6½ digits. For 5¾ digits, please refer to ‘’Chapter five: Characteristics for
DM306x’’.
General Inspection
Handle Adjustment
The Front Panel and User Interface
To Measure DC Voltage
To Measure AC Voltage
To Measure DC Current
To Measure AC Current
To Measure Resistance
To M ea su re Ca pa ci ta nc e
To Test Continuity
To Check Diodes
To Measure Frequency and Period
To Make an Sensor measurement
To Choose Digits resolving index
To Choose Data Digit Display
To Choose Range Options
To Control Trigger Options
Inspect a new DM3000 Digital Multimeter with the following steps:
1. Inspect the shipping container for damage.
Keep the damaged shipping container or cushioning material until the contents of the
shipment have been checked for completeness and the instrument has been checked
mechanically and electrically.
2. Check the accessories.
Accessories supplied with the instrument are listed in "Accessories" at Appendix B of
this guide.
If the contents are incomplete or damaged, please notify the RIGOL Sales
Representative.
3. Inspect the instrument.
In case of any damage, or defect, or failure, notify the RIGOL Sales Representative.
If the shipping container is damaged, or the protective material shows signs of stress,
notify the carrier as well as your RIGOL sales office. Keep the shipping materials for
the carrier’s inspection.
RIGOL offices will arrange reparation or replacement at RIGOL’s option without
waiting for claim settlement.
To adjust the handle position of DM3000 Digital Multimeter, please grip the handle by
the sides and pull it outward. Then, rotate the handle to the desired position as
shown in
Figure 1- 1, Figure 1- 2.
Figure 1- 1
Figure 1- 2 Viewing Positions and Carrying Position
It is important to get familiar with the front panel of a new DM3000. This chapter
gives an introduction of the operation and functions of the Front Panel.
The front panel of the DM3000 is user friendly as shown below. It includes 4
Direction buttons and 12 Function keys, 6 Menu keys and 2 Control keys as shows
below:
How the definitions express in this book:
In this manual, the regarding keys writing expression has the same log with the keys
on the front panel. It is noteworthy that the menu operates keys, marking with the
belt shadow. For example, Conti indicates the short circuit option in menu .
The following shows the system connections and selection of measurement functions.
This practice provides a guide to get familiar with the DC Voltage measurement
technique.
Figure 1- 6 DC Voltage Measurement Data Interface
Table1- 1 DC Voltage Measurement Characteristics
Five Ranges 200mV, 2V, 20V, 200V, 1000V
Max Resolution 100nV
Input Protection 1000V on all ranges (HITer mi na l)
Configurable
Parameters
Range, DC impedance, Null value
Basic measurement:
1. Connect the test leads as shown in
Figure 1- 7; red test lead to the HI Terminal,
black test lead to the LO Terminal.
2. Press
to select the DC Voltage measurement function.
3. Choose the appropriate measurement range.
4. Setup the DC impedance.
Press Æ Res, to setup the DC input impedance (Default value: 10MΩ).
5. Set the Null value.
Null computing will be an option operation, it could be setup in accordance with
user demand. If user does not implement Null computing, this parameter is not
required. (To know the specific setting methods of the Null value setting, please
refer to Chapter 2 “To Set Up Measurement Parameters”, Null computing)Lead
Figure 1- 8 The History Data
Use the history function to review or save the data that has acquired by the current
measurement function. The data can be display “Info” (information), “List” and
“HistoG” formats.
Press Update softkey to update the History data.
Press Save softkey to save data.
Note
Select Auto range if the measurement range is uncertain to get more accurate
measurement data.
The following shows the system connections and selection of measurement functions.
This practice provides a guide to get familiar with the AC Voltage measurement
technique. (The AC functions only support 5½ digits measurement.)
Figure 1- 9 AC Voltage Measurement Data Interface
Table1- 2 AC Voltage Measurement Characteristics
Five Ranges 200mV, 2V, 20V, 200V, 750V
Max Resolution 100nV
Input Protection 750VRMS on all ranges (HITer mi na l)
Configurable
Parameters
Range, DC impedance, Null value
Steps:
1. Connect test leads as shown in
Figure 1- 10; red test lead to the HI Terminal,
black test lead to the LO Terminal.
2. Press
to select the AC Voltage measurement function.
3. Choose the appropriate measurement range.
4. Setup the AC Filter.
Press Æ Filter, to setup the AC Filter Bandwidth (Default value: Mid).
5. Set the Null value.
Null computing will be an option operation, could be setup in accordance with
user demand. If user does not implement Null computing, this parameter is not
required, direct implementation of the next step.
(To know the specific setting methods of the Null value setting, please refer to
Chapter 2 “To Set Up Measurement Parameters”, Null computing)
6. Lead test leads into circuit and start to measure.
AC Voltage
Figure 1- 10 AC Voltage Measurement
7. Use history function.
Press History, the menu shows as below:
Figure 1- 11 The History Data
Use the history function to review or save the data that has acquired by the current
measurement function. The data can be display “Info” (information), “List” and
“HistoG” formats.
Press Update softkey to update the History data.
Press Save softkey to save data.
The following shows the system connections and selection of measurement functions.
This practice provides a guide to get familiar with the DC Current measurement
technique.
Figure 1- 12 DC Current Measurement Data Interface
Table1- 3 DC Current Measurement Characteristics
Five Ranges 2mA, 20mA, 200mA, 1A, 10A
Max Resolution 10nA
Input Protection
Configurable
Parameters
10A, 250V Current Input Fuse on rear
panel
Range, Null value
Steps:
1. Connect test leads as shown in
Figure 1- 13; red test lead to the HI Terminal,
black test lead to the LO terminal.
2. Press
to select the DC Current measurement function.
3. Choose the appropriate measurement range.
4. Set the Null value.
Null computing will be an option operation, could be setup in accordance with user
demand. If user does not implement Null computing, this parameter is not required,
direct implementation of the next step.
(To know the specific setting methods of the Null value setting, please refer to
Chapter 2 “To Set Up Measurement Parameters”, Null computing)
5. Lead test leads into circuit, start to measure.
Use the history function to review or save the data that has acquired by the current
measurement function. The data can be display “Info” (information), “List” and
“HistoG” formats.
Press Update softkey to update the History data.
Press Save softkey to save data.
The following shows the system connections and selection of measurement functions.
The practice provides as guide to be familiar with the AC Current measurement
technique. (The AC functions only support 5½ digits measurement.)
Figure 1- 15 AC Current Measurement Data Interface
Table1- 4 AC Current Measurement Characteristics
Five Ranges 20mA, 200mA, 1A, 10A
Max Resolution 100nA
Input Protection
Configurable
Parameters
10A, 250V Current Input Fuse on rear
panel
Range, AC Filter, Null value
Steps:
1. Connect test leads as shown in
Figure 1- 16; red test lead to the HI Terminal,
black test lead to LO Terminal.
2. Press
to select the AC Current measurement function.
3. Choose the appropriate measurement range.
4. Setup the AC Filter.
Press Æ Filter, to setup the AC Filter Bandwidth (Default value: “Mid” (Middle)).
5. Set the Null setting value.
Null computing will be an option operation, could be setup in accordance with
user
demand. If user does not implement Null computing, this parameter is not
required, direct implementation of the next step.
(To know the specific setting methods of the Null value setting, please refer to
Chapter 2 “To Set Up Measurement Parameters”, Null computing)
6. Lead test leads into circuit and start to measure.
AC Current
Figure 1- 16 AC Current Measurement
7. Use history function.
Press History, the menu shows as below:
Figure 1- 17 The History Data
Use the history function to review or save the data that has acquired by the current
measurement function. The data can be display “Info” (information), “List” and
“HistoG” formats.
Press Update softkey to update the History data.
Press Save softkey to save data.
The following shows the system connections and selection of measurement functions.
The practice provides a guide get familiar with the Resistance measurement
technique. Resistance measurement methods include 2-Wire Resistance Measurement and 4-Wire Resistance Measurement, and will explain
separately.
Null computing will be an option operation, could be setup in accordance with user
demand. If user does not implement Null computing, this parameter is not
required, direct implementation of the next step.
(To know the specific setting methods of the Null value setting, please refer to
Chapter 2 “To Set Up Measurement Parameters”, Null computing)
5. Lead test leads into circuit and start to measure.
Resistance
Figure 1- 19 2-Wire Resistance Measurement
6. Use history function.
Press History, the menu shows as below:
Figure 1- 20 The History Data
Use the history function to review or save the data that has acquired by the current
measurement function. The data can be display “Info” (information), “List” and
“HistoG” formats.
Press Update softkey to update the History data.
Press Save softkey to save data.
NOTE
When measuring small value resistance, Null operation will be recommended, the
test wire impedance error could be eliminated.
twice to select the 4-Wire Resistance Measurement.
3. Choose the appropriate measurement range.
4. Set the Null setting value.
Null computing will be an optional operation, it could be setup in accordance
with users’ demand. If user does not implement Null computing, this parameter
is not required, direct implementation of the next step.
(To know the specific setting methods of the Null value setting, please refer to
Chapter 2 “To Set Up Measurement Parameters”, Null computing)
5. Lead test leads into circuit, start to measure.
Use the history function to review or save the data that has acquired by the current
measurement function. The data can be display “Info” (information), “List” and
“HistoG” formats.
Press Update softkey to update the History data.
Press Save softkey to save data.
NOTE
When measuring resistances,
measurement
.
avoid contacting both ends of the resistor for accurate
The following shows the system connections and selection of measurement functions.
The practice provides a guide to get familiar with the Capacitance measurement
technique.
Figure 1- 24 Capacitance Measurement Data Interface
Table1- 7 Capacitance Measurement Characteristics
Six Ranges 2nF, 20nF, 200nF, 2uF, 20uF, 200uF
Max Resolution 0.1pF
Input Protection 1000V on all ranges (HITer mi na l)
Configurable
Parameters
Basic measurement:
1. Connect test leads as shown in
black test lead to the LO Terminal.
2. Press
3. Choose the appropriate measurement range.
4. Set the Null value.
Null computing will be an optional operation, could be setup in accordance with
userdemand. If user does not implement Null computing, this parameter is not
required, direct implementation of the next step.
(To know the specific setting methods of the Null value setting, please refer to
Chapter 2 “To Set Up Measurement Parameters”, Null computing)
5. Lead test leads into circuit, start to measure.
Use the history function to review or save the data that has acquired by the current
measurement function. The data can be display “Info” (information), “List” and
“HistoG” formats.
Press Update softkey to update the History data.
Press Save softkey to save data.
NOTE
Before measuring the electrolytic capacitance, you should make the two legs of the
electrolytic capacitance short circuit and let it be discharged, and then you can
measure it.
The following shows the system connections and the selection of measurement
functions. The practice provides a guide to get familiar with the Continuity
measurement technique.
Figure 1- 27 Continuity Measurement Data Interface
Table1- 8 Continuity Measurement Characteristics
Tests Current 1mA
Max Resolution Range fixed at 2KΩ
Open-circuit Voltage<7V
Input Protection 1000V (HITer mi na l)
Configurable
Parameters
0≤R
≤Short-circuit impedance
testing
(0Ω≤Short-circuit impedance≤2kΩ)
Steps:
1. Connect test leads as
Figure 1- 28 shown. Red test lead connects the HI
Terminal, Black test lead connects the LO Terminal.
2. Press
to select the Continuity Measurement.
3. Setup the Short-circuit resistance.
Press Set button to set up the Short-circuit Impedance.
The default value is 10Ω. User may carry on the Continuity measurement
directly without modification.
4. Lead test leads into circuit, start to measure.
The following shows the system connections and selection of measurement functions.
The practice provides a guide to get familiar with the Check Diodes technique.
The following shows the system connections and selection of measurement functions.
The practice provides a guide to get familiar with the Frequency and Period
Measurement technique.
Frequency Test
Figure 1- 31 Frequency Measurement Data Interface
Table1- 10 Frequency Test Characteristics
Ranges 200mV, 2V, 20V, 200V, 750V
Measurement Range 3Hz~300kHz
Input Signal Range 100mVAC ~ 750VAC
Input Protection 750VRMS on all ranges (HITer mi na l)
Configurable ParametersNull value
Basic measurement:
1. Connect test leads as
Terminal, Black test lead connects the LO Terminal.
2. Press
3. Set the Null value.
Null computing will be an option operation, could be setup in accordance with
user demand. If user does not implement Null computing, this parameter is not
required, direct implementation of the next step.
(To know the specific setting methods of the Null value setting, please refer to
Chapter 2 “To Set Up Measurement Parameters”, Null computing)
4. Lead test leads into circuit and start to check.
AC Signal
Figure 1- 32 Frequency Test
5. Use history function..
Press History, the menu shows as below:
Figure 1- 33 The History Data
Use the history function to review or save the data that has acquired by the current
measurement function. The data can be display “Info” (information), “List” and
“HistoG” formats.
Press Update softkey to update the History data.
Press Save softkey to save data.
Import Protection 750VRMS on all ranges (HITer mi na l)
Configurable ParametersNull value
Steps:
1. Connect test leads as
Figure 1- 35 shown. Red test lead connects the HI
Terminal, Black test lead connects the LO Terminal.
2. Press
twice to select the Period Test.
3. Set the Null value.
Null computing will be an optional operation, could be setup in accordance with
users’ demand. If user does not implement Null computing, this parameter is
not required, direct implementation of the next step.
(To know the specific setting methods of the Null value setting, please refer to
Chapter 2 “To Set Up Measurement Parameters”, Null computing)
Use the history function to review or save the data that has acquired by the current
measurement function. The data can be display “Info” (information), “List” and
“HistoG” formats.
Press Update softkey to update the History data.
Press Save softkey to save data.
The DM3000 converts the sensor physical properties into electrical voltage,
resistance, current for measurement. So it needs the sensor name, sensor type,
sensor physical unit, sensor reference data, and arithmetic.
Figure 1- 37
Table1- 12 Sensor Building Options
New Newly built sensor reference data file
Edit Edit a sensor reference data file
Load Load a sensor reference data file
Display Set display mode
Steps:
1. Connect test leads as shown in
Figure 1- 54, Figure 1- 55, red test leads to the
HI Terminal, black test leads to the LO Terminal.
2. Press
to select the Sensor function.
3. Press New, the display shows:
Figure 1- 38
(1). In New function interface, press Prpty to edit the sensor Name, sensor Type and
Press Add button, to input the Measured and Corresponding value to the reference
value data.
Figure 1- 44
Figure 1- 45
Figure 1- 46
Press SEG button, you are allowed to segment the reference value with different
arithmetic.
Press Arith button select the algorithms to Linear or Curvilinear.
Press return to New interface then press Done button, you have finished the
input work, then you can use this sensor reference immediately, or you can save it
into the built-in storage space or your U-disk for future work.
Figure 1- 49
Press Apply button, to use this reference value file.
Press Save to save the file.
Figure 1- 50
Press Save button, to finish the save operation.
Figure 1- 51
Press Apply button, to start the sensor measurement.
Figure 1- 52
(2). Press Edit button, to edit the saved sensor reference values.
(3). Press Load button, to load the saved sensor reference file.
(4). Press Disp button, to choose the value to be shown on the display interface.
(5). Press History, enter the menu shown below:
Use the history function to review or save the data that has acquired by the current
measurement function. The data can be display “Info” (information), “List” and
“HistoG” formats.
Press Update softkey to update the History data.
Press Save softkey to save data.
4. Lead test leads into circuit, start to check.
Sensor
1-32
Figure 1- 54 Voltage, Resistance, and Frequency Mode Sensor
The measurement reading resolutions (the accuracy) are 4 1/2, 5 1/2, 6 1/2 digits.
Reduce the index
Increases the index
Figure 1- 57 The Digits Resolution Control Keys
Methods:
Use the left and/or right direction key to adjust the desired measurement resolution.
Press left button to decrease accuracy, press right button to increase accuracy.
The digits resolving index Selection
(1). Each precision of the measure function can be set separately without influence.
(2). Choose the reading precision of 6 1/2 bit when measuring AC for best results.
(3). Save the digits resolving index in volatile memory.
Use “manual” selection or “automatic” to choose measurement range. The
“automatic” allows the instrument to determine the most appropriate range while for
better performance choose the “manual” method.
Increased range
Automatic
selection range
Reduced range
Figure 1- 61 Choice Range Options Keys
Methods 1:
Use up and down direction keys to adjust the Range. Press Up to increase the range,
press Down to reduce the range.
Press key, to select the automatic method.
Methods 2:
Use the menu option keys to adjust the range as shown in the following
By now the front/rear panel, the function control area and keys, and the ways to set
up the multimeter have been introduced.
This chapter goes through all groups of front-panel buttons and menus, and extends
the knowledge the operation instructions.
Follow the exercises to get the most of the powerful measurement capabilities of the
multimeter.
This chapter covers the following topics:
To Set up Measurement Parameters ( )
To Make Mathematics Operation ( )
To Set up Trigger System ( )
To Save and Recall ( )
To Set up Utility ( )
To Set up High-speed data acquisition ( )
Press key to enter the Measurement Menu for setting to set up the
measurement parameters. Users may use the factory defaults or establish desired
configurations.
The Measurement parameters Menu includes: Conti, Filter, Res, Null, and Ratio. To
change these parameters, satisfy the dissimilar condition of the measurement
request.
Table 2- 1 Menu Description
Function
Menu
Description
Conti Set up the resistance value in continuity test.
Filter Choose the AC filter bandwidth.
Res Choose the DC voltage input impedance.
Null Set up null value.
Ratio Measured the ratio of two DC voltage signal.
Set up the continue resistance value in the short test menu. When the measured
resistance is below limit, the DM3000 will beep to indicate the circuit continuation.
The continue resistance is only using at Continue Test.
Press Æ Conti, enter the menu shown below:
Figure 2- 1
Use direction keys to change the parameter values:
Press left and right directional keys to choose different digits. Press up and down
keys to change the current digit value.
Continue Resistance
The range of continue resistance is 1Ω~2000Ω. The default value is 10Ω.
The continue resistance value stored in the nonvolatile memory, the resistance still
keep when the power is off.
There are three settings at the AC Filter menu. Choose the appropriate setting for
more accurate measurements. This function applies AC Voltage and AC Current
measurement only.
Press Æ Filter, enter the menu shown below:
Figure 2- 2
Table 2- 2 AC Filter Menu Description
Function
Menu
Description
Slow Set up the filter with low speed.
Mid Set up the filter with to middle speed.
Fast
Set up the filter with high speed.
Save all changes, back to a higher level menu.
Table 2- 3 AC Filter Parameters Characteristics
AC Filter Options Input Frequency Setting Timer
Slow
Mid
Fast
3Hz~300kHz
20Hz~300kHz
200Hz~300kHz
1.2 reading/s
0.5 reading/s
0.3 reading/s
AC Filter
The AC Filter Parameters are saved in the volatile memory, the data will lose when
the power is off.
The default value of AC Filter Parameters is “Mid” (middle).
The DM3000 null settings is available for DC voltage, AC voltage, DC current, AC
current, resistance, frequency/period, and capacitance measurements.
With null setting each measurement is the difference between a stored null value
and input signal. The null measurement enhance the accuracy by off set test lead
impedance. It is particularly important prior for making capacitance measurements.
The formula used for calculating null measurements is:
Result = reading - null value
The null value is adjustable, and can be set to any value between 0 and
the highest range, for the present function.
(1). In operation interface press Null button, use the current value to be Null value.
(2). To select null function. Start null function, the multimeter will use the current
value to Null value.
(3). In Null setting display interface, it uses the Direction Keys to input null value.
Figure 2- 11
The DM3000 provides five math functions: Null, statistic, dB, dBm and Limit testing.
Only one of these math functions can be enabled at a time, and remains in effect
until change.
In Math function interface, choose the required math function. Press On to start the
Math function.
Math functions are the combination of mathematical and the basic measurement
operation. However, not all combinations are supported. In such case the math
function will automatically turn off.
Table 2- 8 Math Function Menu Description
2-12
Function
Menu
Statistic
dB
dBm
Limit
ON/OFF
SettingsDescription
Reading statistic functions, including: Max, Min,
Average, and number of measurement.
The dB measurement is the difference between
the input signal and a stored relative value.
The dBm function is logarithmic, and is based on
a calculation of power delivered to reference
impedance.
The limit test function performs pass/fail testing
The DM3000 provides five math functions: Null measurements, Total measurements,
dB measurements, dBm measurements, and Limit testing. Only one of these math
functions can be enabled at the same time, and remains in effect until you turn it off
or change it.
Press
key, the display shows:
Figure 2- 12
Table 2- 10 Math Functions Menu Function Description
Function
Menu
Statistic
dB
SettingsDescription
Reading statistic functions, including: Max, Min,
Average, and number of measurement.
The dB measurement is the difference between
the input signal and a stored relative value.
The dBm function is logarithmic, and is based on
dBm
a calculation of power delivered to reference
impedance.
Limit
The limit test function performs pass/fail testing
The Statistic function is for DC voltage, AC voltage, DC current, AC current,
resistance, frequency/period, and capacitance measurement.
The front panel can display the statistical data for any set of readings: average (Ave),
maximum (Max), minimum (Min), and which can read with All functionand the
number of samples taken (Total).
Figure 2- 13
Press Æ Stats, the display shows:
Figure 2- 14
Table 2- 11 Statistic Measurement Menu Function Description
Function MenuDescription
Max (Maximum)
Maximum value from a set of measurements.
Min (Minimum) Statistical measurement all reading Min value.
Ave (Average) Statistical measurement all reading Average value.
Use the Limit test function to perform pass/fail testing with respect to specified
upper and lower limits. The limits can be set to any value between 0 and ±120% of
the highest range of the current function. The upper limit must be positive than the
lower limit.
Press Æ Limit, the display shows:
Figure 2- 15
Table 2- 12 Limit Measurement Menu Function Description
Function
Menu
SettingsDescription
High Set the desired Upper limit.
Low Set the desired Lower limit.
Save all changes, back to a higher level menu.
The range of Limit function:
(1). The limit value scope is 0%~±120% of the current measurement range.
(2). The upper limit value should be always bigger than the lower limit value.
The dB function applies to AC voltage and DC voltage measurements only.
Each dB measurement is different between the input signal and a stored relative
value, with both values converted to dBm.
Press Æ dB, enter the menu shown below:
Figure 2- 16
Table 2- 13 dB Measurement Function Menu Function Description
Function
Menu
Description
Default Use the default value.
Save all changes, back to a higher level menu.
dB =10xLog
[ (Reading2 / R
10
) / 0.001W ] – (dB setting value)
REF
expressed measuring the resistance value in the actual electric circuit.
R
REF
Range of the dB setting value: -120 dBm ~ +120 dBm. The default is 0 dBm.
You can either let the instrument automatically measure this value, or you can enter
a specified value.
This function applies to AC voltage and DC voltage measurements only.
The dBm function is logarithmic, and is based on a calculation of power delivered to
a reference resistance, relative to 1 milliwatt.
Press Æ dBm, the display shows:
Figure 2- 17
Table 2- 14 dB Measurement Function Menu Function Description
Function
Menu
Description
Default Use the default value.
Save all changes, back to a higher level menu.
The computation method of the dBm:
dBm = 10 x Log
expressed measuring the resistance value in the actual electric circuit.
The DM3000 triggering system allows generation of triggers either manually or
automatically, and taking multiple readings per trigger. The DM3000 also allows
setting of an internal triggering level, and to set up pre-triggering.
Selecting a Trigger Source
Specify the source of a trigger. The power–on default is auto triggering. Other types
of triggering are described in the following sections.
The power–on trigger default mode was auto trigger (RUN) mode. Press
to the hold trigger mode. Press
reading is taken, and another reading is taken each time, the
to go to the single trigger mode, a single
button you
pressed, or when a hardware trigger is received on the Ext Trig connector.
Figure 2- 18
Table 2- 15 Trigger Parameters Setting Menu Function Description
Function
Menu
Auto
Setting system Auto trigger and reading Hold meter
reading measurement schemes..
Setting the output signal pulse width at sampling
ending output signal.
Save all changes, back to a higher level menu.
User’s Guide for DM3000 Series
2-19
RIGOL
Auto Triggering
Auto triggering takes continuous measurement at the highest sample rate for the
specified measurement configuration (function, range, resolution, and so forth).
Auto trigger is the default trigger mode at system power up.
Press Æ Auto, the display shows:
Figure 2- 19
Table 2- 16 Auto Trigger Function Menu Description
Function
Menu
Interval
Setting Description
Set interval time in 400~2000ms.*
Hold ON/OFF Set turn the Reading Hold function ON or OFF.
Save all changes, back to a higher level menu.
Interval time:
The delay between the trigger signal and the first sample taken. This is useful in
applications when the input signal settles before taking a reading, or for pacing a
burst of readings.
z The trigger delay may be set from 400 to 2000 ms.*
z The continuity and diode test functions ignore the trigger delay.
z The default trigger delay automatically set if no specified value entered.
z If a trigger delay entered, it will is used for all measurement functions (except
continuity and diode test).
*NOTE:
The trigger interval time is 400~2000ms (6 1/2), 200~2000ms (5 1/2) and
100~2000ms (4 1/2). The range of the interval time will be different in different
resolving index.
The reading hold mode captures and holds a reading on the front panel display when
a stable reading is detected. The system Hold ranges are 0.01%, 0.1%, 1%, and
10%.
Press Æ Auto Æ Hold, the display shows:
Figure 2- 20
Table 2- 17 Reading Hold Function Menu Description
Start the Reading Hold Function, the hold measurement use the following rules
judge the reading count:
When Max() - Min() ≤ hold range x ReadingN, the multimeter hold ReadingN on the
display.
The display update a new reading basing on the current value and the three
preceding measurements before the reading was hold:
Max (ReadingN, ReadingN-1, ReadingN-2, ReadingN-3)
Min (ReadingN, ReadingN-1, ReadingN-2, ReadingN-3)
NOTE
When Reading Hold started, the input resistance was set to 10MΩ for all DC voltage
range to reduce noise arising from the open-loop testing.
At external triggering mode, when the data sampling is over, the instrument outputs
a pulse signal at VM Comp located on the rear panel. The output pulse width can be
adjusted.
Press Æ VMC, the display shows:
Figure 2- 24
Table 2- 19 The Explanation of the External Triggering(polarity: positive)
Function
Menu
Polar
Setting Explanation
Pos
Neg
Setting the pulse signal’s polarity.
PWidth Setting the pulse width.
Store the changing and back to
the higher menu.
The VMC function output
(1). At external triggering mode, when the data sampling is over, the instrument will
output a pulse signal to indicate operation completed.
(2). At external triggering mode, when operating math limited value, the instrument
will export a pulse signal to indicate out of limits detected.
To Storage and Recall function enable save, load, and delete the measurement data,
parameters and sensor files in the local storage as well as in a USB storage.
Press
key, the display shows:
Figure 2- 25
Table 2- 20 Storage and Recall Function Menu Description
Each device on the GPIB (IEEE-488) interface must have a unique address between
0 and 30. The factory set is “1” when shipped.
Press Æ I/O, the display shows:
Figure 2- 34
Table 2- 23 I/O Setting Function Menu Description
Function
Menu
Description
LAN Set up LAN interface.
GPIB Set up GPIB I/O interface.
USB Check USB interface ID.
RS232 Set up RS-232 I/O interface.
Save all changes, back to a higher level menu.
Operation introduction:
Remote control through LAN, GPIB (IEEE-488), USB and RS-232 interface can be
configured. The I/O interface of GPIB, USB and RS-232 only one could be used at the
same time.
Configure the strings of the host name, including letters and numbers. Suggest you
ask you network administrator for host name .Host name can be fixed by moving the
direction button to choose the letters. Host name can be saved in non-violet
memory.
Domain Name:
Configure the strings of the domain name, including letters and numbers. Suggest
you ask you network administrator for domain name. Domain name can be fixed by
moving the direction button to choose the letters. Domain name can be saved in
non-violet memory.
DNS server:
Suggest you ask you network administrator for available DNS server. It could be
saved in non-violet memory.
Each device on the GPIB (IEEE–488) interface must have a unique address. You can
set the address of multimeter to any integral value between 0 and 30. The default
address is “1” when the instrument is shipped from the factory.
Press Æ I/O Æ GPIB, enter the menu shown below:
Figure 2- 39
Table 2- 27 GPIB I/O Setting Function Menu Explanation
Choose RS-232 (serial data interface) interface. Proper configure the baud rate and
parity to make parameters the same as that settled in your computer. Make sure the
connection line is unbroken. The configuration is saved in inviolate memory and
shown in according menu.
Press Æ I/O Æ RS232, enter the menu shown below:
Figure 2- 40
Table 2- 28 RS-232 Parameter Function Menu Explanation
Function
Menu
DisplayExplanation
1200
Set RS-232 baud rate as 1200,
2400, 4800, 9600, 19200, 38400,
57600 or 115200.
Baud
.
.
115200
Parity
None
Odd
Even
The parity check include: None, Odd
check, and Even check.
Save all changes, back to a higher
level menu.
Baud rate:
Proper configure the baud rate and parity to make parameters the same as that
settled in your computer.
Available baud rate values are 1200, 2400, 4800, 9600, 19200, 38400, 57600 and
115200. The default value is 2400.
Parity:
Proper configure the parity to make it same as that settled in your computer.
The decimal and the separator cannot be the same mode, if the decimal is , then
the separator can only be none of space; in contrarily, if the decimal is , then the
separator can only be