Rice Lake Secura User Manual

Application Guide for the Secura, Quintix and Practum Series
Getting Started Quickly
Density
Checkweighing
Counting
Statistics
Components
Unstable Conditions
Conversion
Peak Hold
Percentage
The Touch Screen on Your Balance
Selecting the Language
Leveling
13
12
11 10
9
8
7
1 Toolbar showing currently available buttons:
Zero, Tare, Print (data output), GLP Printout 2 Line for metrological specifications 3 Alert icon indicating calculated values 4 Current weight or value measured 5 Key for toggling between the weight unit and display accuracy 6 Key for running an application 7 Application toolbar: Settings for the selected application 8 Menu key: For toggling to the selection mode for applications
and functions 9 Secura models only: USP minimum sample weight, SQmin 10 Bar graph: Scaled display showing capacity usage (in percent) 11 Plus (or minus) sign of the weight 12 Key for toggling between various display modes of a weight 13 Field for displaying the operating status and warnings
Each balance has been factory preset to English.
1
You can select your preferred language as follows:
2
3 4
5
6
Touch the menu key to enter the selection mode for applications and functions.
Touch the open-end wrench icon.
Select Language.
Touch the desired language and confirm selection by touching the checkmark.
To ensure accurate measurement results, you need to level the balance each time after you have moved it to a different location.
Secura Balances
As soon as LEVEL appears on the display, you will need to level the balance. Make sure that the weighing pan is unloaded before you begin. Touch LEVEL. Then follow the prompts shown on the interactive display.
Quintix and Practum Balances
Turn the two feet as shown in the picture until the air bubble is centered within the circle of the level indicator.
Quick Access to
Getting Started Quickly Calibration and Adjustment
You will need to perform the combined calibration and adjustment procedure the first time you operate the balance, each time after changing it to a different location and at regular intervals during operation. If you are using a Practum series balance, you will have to do this manually as described below. The Secura and Quintix series perform this procedure fully automatically.
Touch CAL to select the combined calibration and adjust­ment mode.
To perform external adjustment, touch CAL-Extern.
First touch A, then w to calibrate the balance. Place a calibration weight equal to that displayed on the balance on the pan. The balance will now perform auto­matic calibration and adjustment, then it will change to the operat­ing mode.
Initial Operation of Balances with a Draft Shield
Install the following parts in the order listed: – In-use dust cover – Turn the draft shield lock
to “unlocked”
– Place the draft shield (1) on
the balance. Note: Use both hands to hold the top struts of the draft shield.
– Slide the side doors all the
way to the back.
– Turn the draft shield lock
to “locked” .
– Position the base plate (2),
centering ring* (3), pan support (4), weighing pan (5)
Secura Balances
Quintix and Practum Balances
and high-rim shield ring* (6) on the balance.
Note: For information about how to clean the draft shield, see the Care and Maintenance chapter in the Installation Instructions.
* only for Secura with 0.1 mg
readability
Initial Operation of Balances Without a Draft Shield
Install the following parts in the order listed:
1) In-use dust cover
.
For detailed instructions, please refer to our Installation Instructions and our detailed User Manual.
2) Position the shield disk (1) and turn it clock­wise until it is locked securely in place.
3) Position the pan support (2) and the weighing pan (3).
Quick Six-Step Guide to Applications
Counting
Familiarize yourself with the capabilities of your Sartorius balance and start off by counting identical parts. The balance will show you the quantity and the total weight.
Mixing**
Say you have a formula and need to weigh in various components quickly with 100% trace­ability, without losing track of the total weight. Just use the Mixing application.
1
2
*
Touch the menu key to access the selection mode for applications and select “Counting”.
1
Place an empty container for your sample on the balance and press D to tare (zero the display).
2
*
Touch the menu key to access the selection mode for applications and select “Mixing”.
Place an empty container on the balance and press D to tare (zero) the container weight.
Components**
Do you need to compound components according to a formula, but don’t want to mix them immediately? The application Components enables you to weigh your individual ingredients into different containers and retrieve the total amount weighed-in – anytime at the touch of a key.
* To view videos of these applications, scan in our QR code or visit
http://www.sartorius.com/manual-secura
** Available in Quintix and Secura
1
2
*
Touch the menu key to access the selection mode for applications and select “Components”.
Place the first empty container on the balance and press D to tare (zero). Then weigh-in component 1 and touch v to save the weight in the memory.
3
4
5 6
Enter the reference sample quantity by touching the application toolbar (factory setting: 10 pcs); place this quantity in the container and press w.
3
Add the first component to the container and touch v to save the component weight. The display will change to 0.
3
The reference sample quantity you entered will now appear on the display. Place your total sample quantity in the container.
4
Add the next component and save its weight by touching v.
4
On the balance display, the piece count of your sample will now be shown. Touch Qnt to view the total weight.
5
If necessary, you can add further components. Touch Net1 to view the total weight.
6
5 6
The total weight of your sample will now appear on the display.
Now the total weight is displayed with a triangle (= calculated value). You can continue to add further components, if necessary.
Remove the container with your first sample from the balance, the weight reading will change to a negative value.
Now place another empty container on the balance and press D to tare (zero). Afterwards, you can add component 2 and confirm using v.
To weigh-in further samples, repeat this procedure. To display the total weight of the components added so far, touch Net.
The total of all saved components will now be displayed (“Sum”) with a triangle (= calculated value). Now you can remove the container with the last sample.
Statistics**
Would you like to save statistics on an entire group of different samples? Your Sartorius balance will transfer the readings to its statistics memory and generate a report on the key statistical parameters.
1
2
*
Touch the menu key to access the selection mode for applications and select “Statistics”.
Touch application toolbar to select whether you want to have the result printed after each weighing operation and whether the balance is to be tared automatically after a value has been saved. Touch the checkmark to confirm.
Density
Let’s suppose you have a solid, irregularly­shaped object and want to precisely determine its density. Just use the Density application along with the Sartorius density determina­tion kit to calculate the quality of, for example, precious metals and plastic parts with high accuracy.
Unstable Conditions
Do you have to weigh in an extremely unstable environment or measure the weight of an inquisitive little mouse that just won’t sit still? No problem. This application lets you deter­mine the weight reliably, giving you peace of mind about the accuracy and traceability of your results.
* To view videos of these applications, scan in our QR code or visit
http://www.sartorius.com/manual-secura
** Available in Quintix and Secura
1
2
*
First install the density kit on your balance by following the instructions for the kit. Touch the menu key and select “Density”.
1
Tare (zero) your balance. Touch the application toolbar liquid (“Medium density”) and the desired accuracy of the density result.
to set the density of the
2
*
Touch the menu key to access the selection mode for applications and select “Unstable Cond.” for unstable conditions or objects.
Touch the the number of subweighing operations (“Measurements”).
application toolbar
to define
3
4
5 6
Touch A to zero the balance. Place your first sample on the balance and press v to save the reading in the memory.
3
Select the desired settings and confirm these by touching the checkmark. Then touch w on the next screen display.
3
Remove the first sample from the balance. Now the display will change to 0.
4
Place your sample on the upper pan of the density kit to view its weight in air and press v.
4
To weigh-in further samples, proceed in the same manner. Touch n to display the results.
5 6
Immerse your sample in the sample holder in the beaker filled with your reference liquid; then touch v.
5 6
You will now see your results as a report.
The balance will now show you the calculated density of your solid sample. The result will be displayed with a triangle (= calculated value).
This number has been factory-set to five measurements. You can enter a different number by touching the three dots.
For example, press 1 and 5 to enter 15 measurements and touch the checkmark to confirm this number.
Place your sample or an animal on the pan and press w to begin weighing under unstable conditions.
The averaged result will be displayed with a triangle (= calculated value).
Sartorius Weighing Technology GmbH Weender Landstrasse 94–108 37075 Goettingen, Germany
Phone +49(0)551.308.0 Fax +49(0)551.308.3289
www.sartorius.com
Specifications subject to change without notice. Printed in Germany on paper that has been bleached without any use of chlorine. | W
Publication No.: WSE6003-e12112 · Order No.: 98649-013-53