Rice Lake SCT-2200 Quick Setup Manual

SCT-2200
When settings are complete press until the indicator displays SAVE?. Press to save set up and return to the weigh mode.
Pressing any other key exits the setup and discards changes.
Press and to set decimals and minimum division. Press , and to set the total capacity of the load cells. Press , and to set the mV/V sensitivity of the load cell. Press , and to set the weight value present on the load cell. Press , and to set the scale capacity. Quick calibration of zero. Press , and to set the weight value used for the calibration. Check the load cell signal in millivolts.
Set the analog output on net or gross weight. Set the maximum value of the analog output. Set the analog output when the scale shows zero weight. Set the minimum value of the analog output.
Set the key to be emulated by the input.
Set the output normal state. Set the output switching condition.
Enable time to disable the output. Enable time to enable the output.
Enable or disable the hysteresis.
INPUTS
0 NONE 1 GROS
2 NET 3 PCS
4 GRO.0
5 NET.0
6 MOTI
23 K.PR
25.K.MOD 26 K.C
27.K.ZER
28.K.TAR 29 ERR
DIV.DEC CEL.CAP CEL.SEN DEAD.LD
RANGE

ZERO

WEIGHT
ADC.MVV
AN.OUT
AO ZER

MODE

AO MAX
INP.02
AO MIN
INP.01
OUT.02
OUTPUT
Disabled. Setpoint for gross weight. Setpoint for net weight, even with no tare. Setpoint for pieces.
Gross weight at zero.
Net weight at zero. Active on instability.
Active when the C key is pressed.
Error indication.
Active when is pressed.
Active when is pressed.
Active when is pressed.
Active when is pressed.
Setpoint for net weight when tare exists.
NO/NC
ONSTAT
RL.IST
ENAB.TM
DELAY
OUT.01
FUNC
1 2 3 4 5 6 7 8
9
10
11
30.NET.T
Optional Analog Model Parameter
Quick Setup Guide
To enter the setup menu, turn the instrument on and briefly press the mode key ( ) when the software version displays.
Press the print key ( ) in the setup menu to view/select the parameter choices.
The instrument has a default calibration with the following features:
Capacity = 10,000 kg Load cell sensitivity = 2.000mV/V
Default Factory Calibration
Divisions = 1
Theoretical Calibration
Use parameters 1,2,3,4 to perform a theoretical calibration.
Use parameters 1,5,6,7 to perform a standard calibration with a known weight. For parameters 6 & 7, press and
Calibration Using Known Weight
wait for ERNOT, STORE?. Then press to accept. For parameter 7, apply test weight before pressing .
Setpoint Programming
In the weigh mode, press and hold to access output’s setpoint programming.
Key Function
In setup: scroll forward through parameters In numeric input: decreases the digit to be modified
ZERO
In setup: scroll back through the functions In numeric input: increases the digit to be modified

TARE

In setup: quickly position the first step of a menu In numeric input: selects the digit to be modified, from
MODE
left to right
In setup: enter into a parameter or to confirm a setting In numeric input: confirms the entry made

PRINT

Turn the instrument on and off
C
In setup: exit a step without confirming the setting

ON/OFF

In numeric input: clears the present value
November 7, 2018
PN 184974 Rev A
230 W. Coleman St. • Rice Lake, WI 54868 • USA
U.S. 800-472-6703 • Canada/Mexico 800-321-6703 • International 715-234-9171 • Europe +31 (0)26 472 1319
Rice Lake Weighing Systems is an ISO 9001 registered company.
© Rice Lake Weighing Systems Specifications subject to change without notice.
www.ricelake.com

Wiring Diagram

SCT-2200
VE 12 / 24 Vdc POWER SUPPLY
1. GND 0 Vdc (GND)
2. +Vdc +12 / 24 Vdc
LOAD CELL RECEPTORS
CELL
:
15. SIG+ SIGNAL +
16. SIG- SIGNAL -
17. SEN+ SENSE +
18. SEN- SENSE -
19. EXC+ EXCITATION +
20. EXC- EXCITATION -
ANALOG OUTPUT (Analog Option Only)
On current
: On voltage:
10. I+ +20 mA 8. V+ +10 V
11. I- 0 mA (GND) 9. V- 0 V (GND)
Note: the maximum resistance applicable on the output current is 350 and the minimum resistance applicable on the output
voltage is 10 N.
INPUTS AND OUTPUTS
Outputs (48Vac or 60Vdc, 150mA max):
6. IN1 input 1 3. OUT1 output 1
7. IN2 input 2 4. OUT2 output 2
Input common normally connected to ground. 5. COMOUT
SERIAL PORT
RS 485
RS 232 RJ-IN 485 Line 12. TX Transmission RJ-OUT 485 Line 13. RX Reception Pin 4 = A(+) , Pin5 = 14. GND GND
RJ OUT
IN 2 IN 1
COMOUT
12/24 Vdc
Analog Output
RS-232
Load Cell
CAL
Optoisolated Inputs positive logic (12–24Vdc, 5–20mA max)
output common
B(-)
OUT 1 OUT 2
V
I
CALIBRATION JUMPER
J1 = close to enable access to the metrological parameters
J1
RJ IN
November 7, 2018
PN 184974 Rev A