Rice Lake RL-MIS-20 User Manual

Rice Lake Mechanical Baby Scale
RL-MIS and RL-MIS-20
Operation Instructions
132019 - Rev A
Contents
5FDIOJDBMUSBJOJOHTFNJOBSTBSFBWBJMBCMFUISPVHI3JDF-BLF8FJHIJOH4ZTUFNT
$PVSTFEFTDSJQUJPOTBOEEBUFTDBOCFWJFXFEBUXXXSMXTDPNPSPCUBJOFE
CZDBMMJOHBOEBTLJOHGPSUIFUSBJOJOHEFQBSUNFOU
1.0 Introduction.................................................................................................................................. 1
2.0 Installation Instructions...............................................................................................................
2.1 Pillar Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.2 Steelyard Rod Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . 4
3.0 Weigh Platter Installation............................................................................................................ 5
4.0 Zero Adjustment........................................................................................................................... 6
5.0 Weighing a Baby.......................................................................................................................... 6
6.0 Troubleshooting ........................................................................................................................... 7
7.0 Specifications.............................................................................................................................. 7
© Rice Lake Weighing Systems. All rights reserved. Specifications subject to change without notice.
Rice Lake Weighing Systems is an ISO 9001 registered company
July 2013
i
ii
Rice Lake continually offers web-based video training on a growing selection
of product-related topics at no cost. Visit www.ricelake.com/webinars.

1.0 Introduction

Weigh beam
Pillar
Molded plastic weighing tray
Frame
Small poise weight
Large poise weight
Internet
The Rice Lake Mechanical Baby Scale provides precise weighing of babies and toddlers. The scale comes in two configurations. Those configurations include:
RL-MIS (Lb-oz & Kg) with measuring tape in inches and centimeters
R
L-MIS-20 (Kg only) with measuring tape in centimeters
The scale is durable, having a sturdy enameled steel body and a wipe down, molded plastic weighing tray with measuring ruler. pillar assembly, plastic weighing tray installation, and the zero adjustment of the scale.
These installation and operating instructions will guide the user through basic instructions on
Figure 1-1. Rice Lake Mechanical Baby Scale Parts Breakdown
This manual can be viewed and downloaded from the Rice Lake Weighing Systems web site at www.ricelake.com/health. Rice Lake Weighing Systems is an company.
ISO 9001 registered
1

2.0 Installation Instructions

Remove cable tie from steelyard rod
Remove twisty ties from scale parts
You will receive your Rice Lake Mechanical Baby Scale partially assembled. Those items that need additional assembly are:
Pillar assembly
Steelyard rod connection
Molded plastic weighing tray assembly
Remove all components from the shipping box and lay out in a convenient pla

2.1 Pillar Installation

Prior to assembly, you’ll need the following tools to finish assembling the Rice Lake Mechanical Baby Scale.
Phillips head screwdriver
Multi-purpose tool (included)
The pillar comes separate from the scale base and must be attached prior to use. Use the following steps to attach the
pillar to the scale base.
1. Set the scale base on a table or other convenient place with the photo in Figure 2-1).
2. Remove twisty tie off of the scale base as shown on the left hand side that holds the steelyard rod from shifting during transit (shown in right side of Figure 2-1).
bottom facing upwards (shown in left side
ce.
of Figure 2-1. Remove the cable tie
Figure 2-1. Remove Cable Ties
2 Mechanical Baby Scale Installation and Operation Instructions
3. Insert the pillar into the scale base assembly ensuring that the RLWS logo located on the weigh beam
Scale base
Pillar
faces the
scale base.
Figure 2-2. Attach the Pillar to the Scale Base
4. Using a phillips head screwdriver, screw in the eight bolts and washers to secure the pillar to scale base.
Figure 2-3. Attach Eight Bolts and Washers to Secure the Pillar to the Scale Base
Mechanical Baby Scale Installation and Operation Instructions - Installation Instructions 3

2.2 Steelyard Rod Connection

Multi-purpose tool
Note
Force lever assembly linkage must be properly aligned in order to weigh
The steelyard rod is located inside the scale pillar. Once the pillar is attached to the scale base, the steelyard rod must be attached to the bottom of the scale.
1. Lay the scale pillar on a table so that the scale platform hangs over the edge of the table and allows you to access
2. Insert the hook part of the multi-purpose tool (included and shown below) to the small hole in bottom the steelyard rod, and pull the hook of the steelyard rod with the hook of the multi-purpose tool to connect it to the scale base lever as shown in Figure 2-4 then push the scale base lever towards the pillar head, and hook the bearing of the steelyard rod on the pivot of the long lever.
the bottom of the scale.
of
Figure 2-4. Connecting the Steelyard Rod to the Scale Base
Visually ensure that the force lever assembly linkage (shown in Figure 2-5) is centered and properly
aligned. Occasionally during shipping, the alignment wil the scale to weigh properly. The linkage can be seen by looking up underneath the top weigh beam if the scale is slightly tipped to one side.
l become skewed The linkage must be free floating in order for
Figure 2-5. View Looking up at Linkage up Underneath the Top Weigh Beam
3. Place the scale upright gently back on the table.
4 Mechanical Baby Scale Installation and Operation Instructions

3.0 Weigh Platter Installation

Inch/cm measuring ruler located next to the scale pillar.
The weighing platter comes separate from the scale base and must be attached for proper operation. To attach the weigh platter to the scale base, use the following instructions.
1. Position the weigh platter on top of the scale platform with the inch/cm marks located next to shown in Figure 3-2.
2. Line up the screw holes as shown in Figure 3-1.
Figure 3-1. Weigh Platter Screw Hole Location
3. Using a Phillips head screwdriver, tighten up the four screws to secure it to the scale platform.
the pillar as
Figure 3-2. Weigh Platter Attached to the Scale Base
Mechanical Baby Scale Installation and Operation Instructions - Weigh Platter Installation 5

4.0 Zero Adjustment

Zero adjusting screw
Weigh beam
Scale pointer
Eye loop area
Poise weights
To ensure accurate weighments, a zero adjustment should be done to the scale upon arrival and setup. To perform a zero adjustment, perform the following steps.
1. Ensure the scale is sitting upright on a level surface.
2. Make sure that the poise weights are pushed all the way to the left hand side (at zero) making sure that the large poise is firmly seated in the notch and the small poise is firmly against the shoulder of the beam.
3. Check the eye loop area of the scale to ensure that the scale pointer is centered.

5.0 Weighing a Baby

To weigh a baby, place the baby on the weigh platter and slide the larger poise (on bottom) to the approximate target weight. Next move the smaller poise on top until the weigh beam balances and the scale pointer is centered within the eye loop area as shown in Figure 4-1.
Figure 4-1. Eye Loop Area and Zero Adjusting Screw Location (Lb and kg version shown)
4. If the scale is not balancing properly, then the small zero balance weight must be adjusted. Turn the zero adjusting screw (shown in Figure 4-2) using the enclosed multi-purpose tool. Depending on the direction the zero balance weight will move accordingly.
Figure 4-2. Zero Adjusting Screw Location
6 Mechanical Baby Scale Installation and Operation Instructions

6.0 Troubleshooting

For the most accurate weight, always use the scale on a hard, level surface and weigh in the center of the platform with the weight distributed as evenly as possible. If an error occurs or seems excessive, check the following:
Problem Possible Fix
Zero balance of beam The weigh beam must be balanced so the pointer comes to
center of the eye loop (shown in Figure 4-1) when both poise weights are set at zero (see Figure 4-1 for poise weight location). Adjust the balance by
rning the zero adjusting screw at the left end of the weigh beam (see
tu Figure 4-2).
Beam does not move freely Check if the beam pointer is touching the side of the ey
travel. Visually ensure that the linkage is centered and pr 2-6). Occasionally during shipping, the alignment will become skewed. The linkage must be free floating in order for the scale to weigh properly. The linkage can be seen by looking up underneath the top weigh beam if the scale is slightly tipped to one side.
Platform rocks excessively When you push down on any c
significant rocking. Ensure scale is setting on a level surface.
Beam does not move at all during a weighment This can be caused by the poise
person’s actual weight. Reset the weights to a lower weight. Also check to see that the steelyard rod is properly connected and aligned.
or
ner of the platform, you should not feel any
weight
s being set to a higher weight than the
operly aligned (see Figure
Table 6-1. Troubleshooting Guide
a rest in the
e loop in its range of

7.0 Specifications

Maximum Capacity 135 Lb (62 Kg) 70 Kg
Graduation 1.0 Oz (20 G) 20 G
Measuring Tape Length 23 in/58 cm 58 cm
RL-MIS RL-MIS-20
Table 7-1. Model Specifications
Mechanical Baby Scale Installation and Operation Instructions - Troubleshooting 7

For More Information

Web Site
http://www.ricelake.com/health
Contact Information
Hours of Operation
Knowledgeable customer service representatives are available 6:30 a.m. - 6:30 p.m. Monday through Friday and 8 a.m. to 12 noon on Saturday. (CST)
Telephone
Sales/Technical Support 800-472-6703
Canadian and Mexican Customers 800-321-6703
International 715-234-9171
Fax
Fax Number 715-234-6967
Email
U.S. sales and product information at prodinfo@ricelake.com
International (non-U.S.) sales and product
Mailing Address
Rice Lake Weighing Systems
230 West Coleman Street
Rice Lake, WI 54868 USA
information at intlsales@ricelake.com
8 Mechanical Baby Scale Installation and Operation Instructions

Mechanical Baby Scale Limited Warranty

Rice Lake Weighing Systems warrants that all RLWS equipment and systems properly installed by a Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by the Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects in materials and workmanship for two (2) years.
RLWS warrants that the equipment sold hereunder will conform
to the current by RLWS. RLWS warrants the equipment against faulty workmanship and defective materials. If any equipment fails to conform to these warranties, RLWS will, at its option, repair or replace such goods returned within the warranty period subject to the following conditions:
Upon discovery by Buyer of such nonconformity, RL
WS will be given prompt written notice with a
detailed explanation of the alleged deficiencies.
Examination of such equipment by RLWS confirms that the nonconformity actually not RLWS shall be the sole judge of all alleged non-conformities.
Such equipment has not been modified, altered, or changed
by any person other than RLWS or its
duly authorized repair agents.
RLWS will have a reasonable time to repair or replace the
defective equipment. Buyer is responsible
for shipping charges both ways.
In no event will RLWS be responsible for travel time or on-location re disassembly of equipment, nor will RLWS be liable for the cost of any repairs made by others.
HESE WARRANTIES EXCLUDE ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WITHOUT
T
LIMITATION WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NEITHER
RLWS
RLWS
REPAIR OR REPLACEMENT OF SUCH GOODS. IN ACCEPTING THIS WARRANTY, THE BUYER WAIVES ANY AND ALL OTHER CLAIMS TO WARRANTY.
S
WARRANTY CLAIMS.
N
WARRANTY SHALL HAVE ANY LEGAL EFFECT UNLESS MADE IN WRITING AND SIGNED BY A CORPORATE OFFICER OF RLWS AND THE BUYER.
NOR DISTRIBUTOR WILL, IN ANY EVENT, BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.
AND BUYER AGREE THAT RLWS’S SOLE AND EXCLUSIVE LIABILITY HEREUNDER IS LIMITED TO
HOULD THE SELLER BE OTHER THAN RLWS, THE BUYER AGREES TO LOOK ONLY TO THE SELLER FOR
O TERMS, CONDITIONS, UNDERSTANDING, OR AGREEMENTS PURPORTING TO MODIFY THE TERMS OF THIS
written specifications authorized
exists, and was
pairs, including assembly or
© 2012 Rice Lake Weighing Systems, Inc. Rice Lake, WI USA. All Rights Reserved.
RICE LAKE WEIGHING SYSTEMS • 230 WEST COLEMAN STREET • RICE LAKE, WISCONSIN 54868 • USA
Mechanical Baby Scale Installation and Operation Instructions - Specifications 9
10 Mechanical Baby Scale Installation and Operation Instructions
PN 132019- Rev A 07/2013
Loading...