Nicol Scales & Measurement is an ISO Accredited Calibration
Company that has provided calibration, repair and sales of all types
of weighing and measurement products since 1931.
Page 2
480 Legend Series
Disponible
en Español
Visite ricelake.com/spanish
para ver todos los materiales
RLWS disponibles en Español
Digital Weight Indicator
Version 1.05
Technical Manual
PN 119201 Rev K
Page 3
Page 4
Contents
Technical training seminars are available through Rice Lake Weighing Systems.
Course descriptions and dates can be viewed at www.ricelake.com/training
or obtained by calling 715-234-9171 and asking for the training department.
This manual is intended for use by service technicians responsible for installing and servicing 480 digital weight
indicators. This manual applies to indicators using Version 1.05 of the 480 software.
Configuration and calibration of the indicator can be accomplished using the Revolution
the indicator front panel keys. See Section 3.2 on page 32 for information about configuration methods.
Manuals can be viewed or downloaded from the Rice Lake Weighing Systems website at
www.ricelake.com.
An Operation Manual is available on the website, that provides basic operating instructions for users.
Warranty information can be found on the website at
www.ricelake.com/warranties
1. 1S af e ty
Safety Symbol Definitions
Indicates a potentially hazardous situation that, if not avoided, could result in serious injury or death, and
includes hazards that are exposed when guards are removed.
Indicates information about procedures that, if not observed, could result in damage to equipment or
corruption to and loss of data.
Safety Precautions
Do not operate or work on this equipment unless you have read and understand the instructions and
warnings in this manual. Failure to follow the instructions or heed the warnings could result in injury or
death. Contact any Rice Lake Weighing Systems dealer for replacement manuals. Proper care is your
responsibility.
®
configuration utility or
General Safety
Failure to heed may result in serious injury or death.
Some procedures described in this manual require work inside the indicator enclosure. These procedures are to be
performed by qualified service personnel only.
Do not allow minors (children) or inexperienced persons to operate this unit.
Do not operate without the enclosure completely assembled.
Do not use for purposes other than weight taking.
Do not place fingers into slots or possible pinch points.
Do not use this product if any of the components are cracked.
Do not exceed the rated specification of the unit. See Section 9.17 on page 75.
Do not make alterations or modifications to the unit.
Do not remove or obscure warning labels.
Do not submerge.
Before opening the unit, ensure the power cord is disconnected from the outlet.
Introduction1
Page 9
1. 2O ve r vi e w
The 480 is a single-channel digital weight indicator housed in a NEMA Type 4X/IP66-rated stainless steel
enclosure. The indicator front panel consists of a large (.8 in/20 mm), six-digit, seven-segment LED display , with a
seven-button keypad (480) or an 18-button full numeric keypad (480 Plus).
Features
• Auto switching AC power supply 115 VAC to 230 VAC (50-60 Hz)
• Drives up to ten 350 ohmor twenty 700 ohm load cells
• Supports four and six wire load cell connections
• Two communications ports with demand or continuous (COM 2 only) outputs
• Unit ID, operator can enter up to six numeric digits
• Accumulator with report and clear
• Time and date
• Audit trail tracking
• Optional Ethernet/USB connects to PC or printer in operation mode, or flash drive to save/load
configuration
• Optional analog output module provides 0–10/2-10 VDC or 0–20/4–20 mA tracking of gross/net weight
• Optional digital I/O card, four outputs/two inputs for setpoints and key functions
Supported Applications
• Custom Ticket Printing: gross, net and setpoint format can be customized up to 300 characters and print
time and date, unit id and consecutive ticket number.
• Basic Weighing: Gross or net mode with operator menu to other functions.
• Accumulation: Weights are totaled, with armed print function.
• Batching: Up to eight batch steps with latched or continuous outputs for gross, net or delay setpoint.
Actions include trip high or low, wait for standstill, print, accumulate and tare.
• Keyed Tare: Preset tare value can be entered when the gross weight is at zero.
• Local/Remote: Remote unit displays weight and transmits key press commands to the local unit.
1.3Operating Modes
The 480 has four modes of operation.
Weigh Mode
Weigh mode is the default mode of the indicator . The indicator displays gross or net weights as required, using the
annunciators described in Section 1.6 on page 4 to indicate scale status and the type of weight value displayed.
Configuration Mode
Most of the procedures described in this manual, including calibration, require the indicator to be in configuration
mode.
To enter configuration mode:
1. Remove the setup switch access screw from the back of the enclosure. See Figure 2-1 on page 10.
2. Insert a non-conductive tool into the access hole and press the configuration switch. Indicator display
changes to show
User Menu Setup Mode
The user menu setup mode is used to access the accumulator functions, audit trail, display the tare, unit id, time and
date, setpoints, data communications parameters, print formats and view the firmware version.
It is accessible by pressing the
See Section 3.1 for more information about the user setup mode.
Test Mode
Test mode provides a number of diagnostic functions for the 480 indicator. See Section 9.13 on page 70 for more
information about entering and using test mode.
CONFIG.
Menu key on the front panel.
2480 Legend Series
Page 10
1.4Front Panel Keypad
PRELIMINARY
The up, down, enter, left and right
arrows by the keys describe the
functions assigned in the
operating modes.
Keys are also used to navigate
through menus, select digits within
numeric values and increment/
decrement values.
Numeric Keypad – Enter numeric
values. See Section 1.7.1 on
page 5.
POWER
Note
MENU
ZERO
UNIT
PRINT
TARE
GROSS
NET
B/N
See Section 1.7 for information about using the front panel keys in configuration mode.
Figure 1-1. 480 Front Panel, Showing LED Annunciators and Key Functions (480Plus Shown)
1.5Keypad Functions
KeyFunction
Turns t h e un i t on / off.
If power mode is set to manual, the Power button must be used to turn the unit on and off. If
power mode is set to auto, the unit will automatically power on when it’s plugged in and the only
way to turn it off is to unplug power. See Section 3.1.7.
The Menu key is used to access the User Setup menu.
Sets the current gross weight to zero, provided the amount of weight to be removed or added is within the
specified zero range and the scale is not in motion. The zero band is defaulted to two percent of full scale, but can
be configured for up to 100 percent of full scale.
Also used as a move up key to navigate to different menu levels or used to increment a number when editing a
value.
Switches the weight display to an alternate unit. The alternate unit is defined in the Configuration menu, and could
be kg, g, lb, oz, tn or t.
Also used as a scroll left key to navigate to different menus or to toggle to another digit when editing a value.
Sends on-demand print format out the serial port, provided the conditions for standstill are met. Print may be
displayed while printing.
Also used as a scroll right key to navigate to different menus or to toggle to another digit when editing a value.
Performs one of several predetermined Tare functions dependent on the mode of operation selected in the
TAREFN parameter. See Section 1.8.7 for more information on viewing stored tares.
Also acts as an enter key for numeric or parameter entry.
Switches the display mode from gross to net, or from net to gross. If a tare value has been entered or acquired,
the net value is the gross weight minus the tare.
Gross mode is shown by the Gross/Brutto annunciator; net mode is shown by the Net annunciator.
Also used as a move down key to navigate to different menu levels or to decrement a number when editing a
value.
During a numeric entry, sets the currently selected digit to 0, then selects one digit to the right.
Introduction3
Page 11
1.6LED Annunciators
PRINT
MENU
UNIT
POWER
TARE
GROSS
NET
B/N
ZERO
Move Right/Next
Move Up/Increment Value
Access User Setup
ENTER Value
Move Left/Previous
Move Down/Decrement Value
Power On/Off
The 480 display uses a set of eight LED annunciators to provide additional information about the value being
displayed.
LEDDescription
Gross/Brutto LED
Gross weight display mode (or Brutto in OIML mode)
Net LED
Net weight display mode
Zero (Center of Zero) LED
The Center of Zero LED indicates that the current gross weight reading is within +/- 0.25 display divisions of the
acquired zero, or is within the center of zero band.
A display division is the resolution of the displayed weight value, or the smallest incremental increase or decrease
that can be displayed or printed.
Standstill LED
Scale is at standstill or within the specified motion band. Some operations, including zero, tare and print, can only
be done when the standstill LED is on.
lb/kg LED
Displays which unit of measure is being used.
lb and kg annunciators indicate the units associated with the displayed value: lb = pounds, kg = kilograms.
The displayed units can also be set to short tons (tn), metric tons (t), ounces (oz), grams (g) or none (no units
information displayed). The lb and kg LEDs function as primary and secondary units annunciators. If neither
primary nor secondary units are lb or kg, the lb annunciator is lit for primary units and kg is lit for secondary units.
T LED
Indicates that a push-button tare weight has been acquired and stored in memory.
PT LED
Indicates that a preset tare weight has been keyed in or entered and stored in memory.
Table 1-1. LED Annunciators
See Section 3.2.3 for more information about configuring primary and secondary display units.
1.7Front Panel Navigation
Figure 1-2. Front Panel Key Functions
Four front panel keys are used as directional keys to navigate through the menus (see Figure 1-2).
•
Unit () and Print ( ) scroll left and right on the same menu level.
•
Zero () and Gross/Net ( ) move up and down to different menu levels.
•The
•The
Tare key serves as an enter key () for selecting parameter values within the menus.
Menu key allows front panel access to user setup and configuration mode.
4480 Legend Series
Page 12
Navigating Through Levels
Default valueValueValue
When moving through values below the first menu level, press to return to the level above.
Value
1st Level
Parameter
2nd Level
Parameter
1st Level
Parameter
2nd Level
Parameter
00 0 0 0 0
When editing numeric values, press or to change the
digit selected. Press or to increment or decrement the
value of the selected digit. Press to save the value entered
and return to the level above.
Note
Figure 1-3. Menu Navigation
To select a parameter, press or to scroll left or right until the desired menu group appears on the display, then
press to move down to the desired sub-menu or parameter. When moving through the menu parameters, the
present value appears first on the display.
Edit Parameter Values
To change a parameter value, scroll left or right to view the values for that parameter. When the desired value
appears on the display, press Tar e to select the value and move back up one level. T o edit numerical values, use
the navigation keys to select the digit and to increment or decrement the value.
Figure 1-4. Editing Procedure for Numeric Values
1.7.1Numeric Keypad - Editing Procedure for Numeric Values (480 Plus Only)
Figure 1-5. Numeric Keypad for the 480Plus
With the numeric keypad option, numeric values can be edited with the numbers on the keypad rather than using
the navigation arrows. Numeric keys are functional only if the display is prompting for an input with a flashing
digit.
1. When editing numeric values, insert the required value using the numeric keypad.
2. Press
Tare to save the value entered and return to the level above.
• Press to set the currently selected digit to 0.
• Press to enter a decimal point.
When editing fractional numeric values, the decimal point must be positioned in accordance with the primary
units formatting, otherwise the keyed number may be rejected by the software.
Introduction5
Page 13
1.8Indicator Operations
MENU
Note
ZERO
UNIT
TARE
Basic 480 operations are summarized below.
1.8.1Menu
Press , Menu will be displayed.
Press , then or to select the following parameters.
• Audit Trail (See Section 1.8.10)
• Display T are (See Section 1.8.7)
• Unit ID (See Section 1.8.11)
• Accumulator (See Section 1.8.12)
• Time and Date (See Section 1.8.13)
• Setpoints (See Section 1.8.14)
• Serial (See Section 3.1.2)
• Print Formats (See Section 7.0)
• USB (See Section 9.8)
• Ethernet (See Section 9.9)
• Version (See Section 1.8.15)
• Miscellaneous (Power Option) (See Section 3.1.7)
See Section 3.1 for more information.
Press repeatedly to return to the weigh mode.
Menus may be password protected. See Section 1.8.16 to setup a password.
1.8.2Status Lights While in Various Menus
Sub-menu levels are indicated by the LEDs as shown below.
First sub-menu level
Second sub-menu level
Third sub-menu level
Fourth sub-menu level
1.8.3Zero Scale
1. In gross mode, remove all weight from the scale and wait for the LED to light.
2. Press . The LED lights to indicate the scale is zeroed.
1.8.4Toggle Units
Press to toggle between primary and secondary units. The current unit LED will be lit.
1.8.5Acquire Tare
1. Place container on scale and wait for the LED to light.
2. Press to acquire the tare weight of the container. Net weight is displayed and the
show the tare value was entered.
See Section 9.14 for Regulatory Mode Functions.
T LED lights to
6480 Legend Series
Page 14
1.8.6Preset Tare (Keyed Tare)
TARE
TARE
MENU
PRINT
PRINT
PRINT
PRINT
GROSS
NET
B/N
MENU
1. With the scale empty and display showing zero weight, press . 000000 is displayed with the focused
digit flashing.
2. Edit the value using the following method; or with the 480 PLUS, use the keypad. See Section 1.7 on page 4.
• Press or to select the digit.
• Press or to increment or decrement the value.
3. Press when the value is correct. The display will change to the Net mode and the
PT LED lights to
show the preset tare was entered.
1.8.7Display Tare
When a stored tare value is displayed, the gross and net LEDs will be off and the will be lit.
To display a stored tare:
1. Press .
2. Press to
3. Press to
AUDIT.
TAR E and press .
4. Press repeatedly to return to weighing mode.
If there is no tare in the system,
0 will displayed and the Gross and Net LED will be off.
See Section 9.14 for more information pertaining to the regulatory mode of operation.
1.8.8Print Ticket
1. Press to print either the gross or net format.
2. Wait for the LED to light.
3. Press to send data to the serial port.
If the LED is not lit and is pressed, the print action will take place only if the scale comes out of
motion within three seconds. If the scale stays in motion for over three seconds, the press is ignored.
1.8.9Toggle Gross/Net Mode
1. Press to switch the display mode between gross and net. If a tare value has been entered or acquired,
the net value is the gross weight minus the tare.
Gross mode — Gross/Brutto LED is lit.
Net mode —
Net LED is lit.
1.8.10View Audit Trail
See Section 3.1 on page 22 for audit trail settings.
1. Press .
2. Press to
3. Press . The audit trail
4. Press then or to
5. Press to view selected parameter.
6. Press twice to return to
7. Press to the audit trail
8. Press repeatedly to return to weighing mode.
AUDIT.
CALIB is displayed.
CNT, TIME or DATE.
CALIB.
CONFIG and repeat steps 5 and 6 to view configuration number.
Introduction7
Page 15
1.8.11Enter New Unit ID
MENU
TARE
MENU
TARE
TARE
Note
MENU
TARE
Note
1. Press .
2. Press to
3. Press until display reads
AUDIT.
UNIT ID.
4. Press to view the current value.
5. Edit the value using the following method; or with the 480 PLUS, use the keypad. See Section 1.7 on page 4.
• Press or to select the digit.
• Press or to increment or decrement the value.
6. Press when the value is correct.
7. Press repeatedly to return to weighing mode.
1.8.12Display Accumulator
1. Press .
2. Press to
3. Press until display reads
4. Press to display
5. Press or to select desired parameter (
•For
•To
selected parameter
6. Press repeatedly to return to weighing mode.
See Section 3.1 for the accumulator menu structure.
AUDIT.
ACCUM.
VIEW.
VIEW, TIME, DATE, PRINT, CLR Y).
VIEW, TIME or DATE, press to view the value. Press or to return to selected parameter.
PRINT or CLEAR, press , then press to print or clear the accumulator. Press to return to
If the accumulated value exceeds 999999, display shows EE ACC. The value will still be correct and will print
correctly up to 1,000,000,000.
1.8.13Display or Change Time and Date
To set the date and time:
1. Press .
2. Press to
3. Press until display reads
4. Press and select time or date with or .
5. Press to view the current setting.
6. To edit the value of the time, in 24 hour or 12 hour format (hh.mm.ss), use the following method.
• Press or to select hours, minutes, or seconds – the selected value will be flashing
• Press or to increment or decrement the value.
7. Press when the value is correct.
Use the same procedure to enter the date in the same format configured for the indicator.
See Section 3.2.5 for available formats.
8. Press repeatedly to return to weighing mode.
AUDIT.
TIMDAT (T ime/ Date).
The time and date are backed up with an internal battery. If the main power is interrupted, time and date will
not be lost.
When in 12 hour format, the PT LED indicates pm setting.
8480 Legend Series
Page 16
1.8.14Display, Edit and Set Setpoint Value
MENU
TARE
TARE
MENU
TARE
Important
See Section 8.0 on page 59 more information.
1. Press .
2. Press to AUDIT.
3. Press until display reads
4. Press and navigate across to desired setpoint number (1-8).
5. Press and navigate across to select
6. Press and navigate across to select
7. Press to view and edit the value.
• To edit Value, use the following method; or with the 480 PLUS, use the keypad. See Section 1.7 on page 4.
- Press or to select the digit.
- Press or to increment or decrement the value.
- Press when the value is correct.
SETPNT.
USER.
VALUE or ENABLE.
• To edit
- Press or to select
ENABLE:
On/Off.
- Press when the value is correct.
8. Press repeatedly to return to weighing mode.
See Section 3.1.1 for the
SETPNT menu layout.
1.8.15View Firmware Version
1. Press .
2. Press .
3. Press until display reads
4. Press .
AUDIT is displayed.
VERS.
FIRMW is displayed.
5. Press to view version.
6. Press repeatedly to return to weighing mode.
1.8.16Enter User Password
1. Remove the setup switch access screw from the back of the enclosure. See Figure 2-1 on page 10.
2. Insert a non-conductive tool into the access hole and press the configuration switch. Indicator display
changes to show
3. Press or until
4. Press .
CNFG is displayed.
5. Press to
6. Press .
000000 is displayed.
7. To edit the password, use the following method; or with the 480 PLUS, use the keypad. See Section 1.7 on
page 4
• Press or to select the digit.
• Press or to increment or decrement the value.
CONFIG.
PAS WRD is displayed.
USER.
• Press when the value is correct.
8. Press to return to
9. Press to
CONFIG.
PASWRD.
10. Press to return to weighing mode.
When entering a user function, the operator will now be required to enter the password.
Enter 999999 to reset password, this will also reset the configuration back to default values.
Introduction9
Page 17
2.0Installation
WARNING
Important
Setup switch
access screw
Load cell cable
cord grip (open)
Communications
access cord grip
(plugged)
Power cord
Back of indicator
Important
2.1 Unpacking
Immediately after unpacking, visually inspect the 480 to ensure all components are included and undamaged . The
shipping carton should contain the indicator, this manual, and a parts kit. If any parts were damaged in shipment,
notify Rice Lake Weighing Systems and the shipper immediately.
See Section 2.8 on page 18 for parts kit contents.
2.2 Enclosure Disassembly
The indicator enclosure must be opened to connect the scale load cell cable and any other interface connection.
Before opening the unit, ensure the power cord is disconnected from the power outlet.
Ensure power to the indicator is disconnected, then place the indicator face-down on an anti-static work mat.
Remove the screws that hold the backplate to the enclosure body. Then lift the backplate away from the enclosure
and turn it over to access boards.
The display cable will still be attached. Be careful when lifting and turning over the backplate so it does
not become damaged or dislodged.
2.3 Cable Connections
The 480 provides four cord grips for cabling into the indicator: one for the power cord, three to accommodate load
cell, communications, digital inputs and outputs, and analog output cables. Two of the three free cord grips come
with a plug installed to prevent moisture from entering the enclosure. Depending on the application, remove the
plug from any cord grip that will be used and install cables as required. Figure 2-1 shows the recommended
assignments for the 480 cord grips.
10480 Legend Series
Figure 2-1. Recommended Cord Grip Assignments
When installing cords through cord grips, ensure the cord grip is tightened snuggly to prevent moisture
from entering the enclosure.
Page 18
2.3.1Cable Grounding
Grounding
clamp
Braid
Braided Cable
Cut insulation here
Foil Insulated Wire
Silver
side out
Shield wire (cut)
Cut insulation here
for foil sheilded cables
Length of foil before
folding back on
cable insulation
Grounding
bracket
Grounding
clamp
480 Indicator
back panel
Ground
clamp screw
Ground
clamp screw
Except for the power cord, all cables routed through the cord grips should be grounded against the indicator
enclosure. Do the following to ground shielded cables:
• Use the ground clamp screws to install grounding clamps on the grounding bar. Do not tighten screws yet.
There is a larger clamp available in the parts kit; the larger clamp is for a cable diameter of 0.25 inch, the
smaller clamp is for a cable diameter of 0.18 inch
• Route cables through cord grips and grounding clamps to determine cable lengths required to reach cable
connectors. Mark cables to remove insulation and shield as described below:
• For cables with foil shielding, strip insulation and foil from the cable 1/2
grounding clamp (see Figure 2-2). Fold the foil shield back on the cable where the cable passes through
the clamp. Ensure silver (conductiv e) side of foil is turned outward for contact with the grounding clamp.
• For cables with braided shielding, strip cable insulation and braided shield from a point just past the
grounding clamp. Strip another 1/2 inch (15 mm) of insulation to only expose the braid where the cable
passes through the clamp (see Figure 2-2).
• For load cell cables, cut the shield wire just past the grounding clamp. Shield wire function is provided
by contact between the cable shield and the grounding clamp.
• Route stripped cables through cord grips and grounding clamps. Ensure shields contact grounding
clamps as shown in Figure 2-2. Tighten ground clamp screws.
inch (15 mm) past the
Figure 2-2. Grounding Clamp Attachment for Foil-Shielded and Braided Cabling
Cable Connections11
Page 19
2.3.2Wiring
Relay (DIO) Board
Analog Output Board
A/D Board
CPU Board
Power Supply
Analog Output
Connection
1 I/O 15
20mA Comm 2
Connection
Comm 1 & 2
Connection
Option Card Bracket
Ethernet/USB Board
Battery and Battery Holder
Analog Output and Relay Board Options
Power Supply, CPU and A/D Board
Ethernet/USB Board Option
Battery Option
Charger Board
Digital I/O Connection
Ethernet RJ45 Plug
USB 5-Pin Plug
12480 Legend Series
Figure 2-3. 480 Board Options
Page 20
20mA_Out
RXD1
TXD1
GND
RXD2
TXD2
GND
GND
Sig+
SIG-
SEN+
SEN-
EXC+
EXC-
Load Cell Connection
RS-232
Connection
20mA
Connection
L
GND
N
Power Supply
CPU Board PN 131318
A/D Board PN 131319
Power Supply
PN 131316
Analog Output
PN 131341
Relay Board
PN 131342
J2
J1
J2
J13
J17
J5
J6
J15
J16
J14
J7 to display board
J2
J10
R1_NC
R1_C
R1_NO
R2_C
R2_NO
R3_C
R3_NO
R4_C
R4_NO
IN-
IN1+
IN-
IN2+
+5V_OUT
GND
I_OUT
RETURN
V_OUT
RETURN
Analog Output Terminal
I/O Terminal
J3 J4
Sense Jumpers
Ethernet/USB Board
PN 164600
Heartbeat
LED
LED1
Activity
USB
5-Pin Plug
J2
RJ45
Ethernet
Jack
J1
J4
RL4_ON
RL1_ON
RL2_ON RL3_ON
IN1
IN2
+5V
OUT
RL4
RL1
RL2
RL3
J2
Mode Select
Jumpers
JMP2
JMP3 JMP1
A2 A1 A0
J3
Battery Charger
PN 131317
Battery
PN 131323
Standard
Option
Key
J9
Battery and Battery Charger are optional.
Analog, Ethernet/USB and Relay boards require an option card bracket (PN131340) for mounting.
Relay Board Information:
To use outputs 1-4 and inputs 1-2, short jumper A1 (JP2)
To use outputs 5-8 and inputs 3-4, short jumpers A0 (JP1) and A1 (JP2)
Output relay rating of relay 2 AMPS at 30 VDC dry contact relay closure
Digital Input is opto-isolated, activated with 5-24VDC.
Note
Figure 2-4. Wiring Diagram
Cable Connections13
Page 21
2.4 Battery Operation
Note
The 480 can be equipped with an optional internal Lithium-Ion battery to provide up to:
• 18 hours of continuous use (single load cell, 350 ohms, no options)
• 15 hours of continuous use (four load cell, 350 ohms, no options)
When operated on AC power, the internal battery is automatically charged by means of an internal battery charger circuit.
Battery life depends on several factors, including number of load cells, frequency of use, options installed,
temperature and other accessories being used. Battery charge time is approximately 16 hours to 100 percent.
To extend battery life, the indicator features a sleep mode that allows the indicator to go into a low power mode
while it’s operating on battery power. See Section 9.6 on page 65 for more information about the sleep mode.
When installing a 480 with a battery option, it is recommended that the CPU board jumper 4 be removed.
Removing jumper 4 will maximize the life of the battery when the indicator is turned off due to reduced
standby current. See the Power parameter in Section 3-8 on page 31 for more information.
When the Standby parameter is set to any setting other than NONE, any key press will work to wake the
indicator from standby.
14480 Legend Series
Page 22
2.5 Option Card Installation
Note
Note
Use the following instructions to install or replace the Analog Output Module (PN 131341), Relay Board (PN
131342) or Ethernet/USB option card (PN 164600).
1. Disconnect p
ower to the indicator.
2. Remove backplate as described in Section 2.2 on page 10.
3. If not already in place, install an option card bracket (PN 131340). See Figure 2-3.
4. Mount the analog output module, relay card or the Ethernet/USB option card on the option card bracket. See
Figure 2-3. The Ethernet/USB option card has to be installed on the option card bracket above the A/D card,
not in the location near the power supply.
5. Connect the option card ribbon to the option card(s).
6. Connect the option card ribbon to J9 on the CPU board.
7. Connect output cables, as needed, to the option card(s).
Either option card position can be used for the Analog Output Module, or the Relay Card.
The Ethernet/USB Option Card must be installed in the location above the A/D Card, not next to the Power
Supply.
2.6 Board Removal
If you must remove any 480 board, use the following procedure:
1. Disconnect power to the indicator.
2. Remove backplate as described in Section 2.2 on p age 10.
3. Disconnect all cables from the board.
4. Label connections for re-installation of board.
5. Remove the four screws from the board, then lift the board off the backplate.
To install a board, reverse the above procedure. Be sure to reinstall cable ties to secure all cables inside the
indicator enclosure.
When removing lower boards, the upper boards and option bracket, if installed, will need to be removed first.
Option Card Installation15
Page 23
2.7 Enclosure Reassembly
1
2
4
3
11
5
7
9
6
8
12
10
Step 1.
Torque 1-4 in the order shown
13
42
Step 2.
Torque 5-8 in the order shown
76
58
Step 3.
Torque 9-12 in the order shown
1110
912
To rq u ein-lbNm
Backplate screws
101
Cable Glands
222.5
Cable Gland Caps
13.31.5
Vent
5-70.6-0.8
Table 2-1. Torque Values
Note
Once cabling is complete, position the backplate over the enclosure and reinstall the backplate screws. Use the
torque pattern shown in Figure 2-5 to prevent distorting the backplate gasket. Torque screws to 10 in-lb (1 N-m).
Figure 2-5. 480 Enclosure Backplate
Torqued screws may become less tight as the gasket is compressed during the torque pattern; a second
torque is required using the same pattern and torque value.
16480 Legend Series
Page 24
2.7.1Seal the Indicator (Optional)
Note
Without A/D cover
11 - Screw, Pan Cross Head
8-32NC
1 - Screw, 8-32 drilled
fillister head
1 - Screw, 8-32 drilled
fillister head
1 - Screw, 1/4'' drilled
fillister head
Important
With A/D cover
12 - Screw, Pan Cross Head
8-32NC
Calibration Cover
for Hardware Sealing
Setup Switch
Access
1 - Screw, 4-40 drilled
fillister head
1 - Screw, 8-32 drilled
fillister head
For access to configuration parameters, the setup switch must be pressed.
Prevents Service Access
Insert a lead wire seal through three fillister screws. This prevents access to the electronics, electrical contacts and
Legal for Trade configuration parameters.
Figure 2-6. Sealing the Indicator to Prevent Access
Only use the 1/4 inch screw in the top hole for sealing, using the longer one will damage the CPU board.
Allows Service Access – Optional
The calibration cover is used for inserting a lead wire seal through both fillister screws and the plastic cover. The
cover allows access to the electronics and electrical contacts, while preventing access into Legal for Trade
configuration parameters.
There is an alternative cover over the A/D to ensure the A/D and load cell connection cannot be changed after the
unit is sealed. This cover is secured to the A/D with the bottom fillister screw in the calibration cover.
There is also an audit trail counter to track calibration and configuration changes made to legally relevant
parameters.
Place the calibration cover on the backplate and secure with the existing fillister head screws and o-rings to seal the
indicator for Legal for Trade approval.
Figure 2-7. Sealing the Indicator to Allow Access Option
Enclosure Reassembly17
Page 25
2.8 Replacement Parts
4
1
1
2
3
7
7
7
8
4
9
19
16
18
17
6
10
23
25
3
20
21
27
15
14
13
Representation only.
Actual boards may
look dierent.
164604Ethernet cable with cord grip, 6 ft1
164602USB Cable, Type A Female, 20 cm1
167707USB Cable, Type A Male, 200 cm1
167708USB Cable, Type B Male, 200 cm1
88733Vent, breather vent1
88734Nut, breather vent1
1034622-position screw terminal 1
765134-position screw terminal 1
765146-position screw terminal 2
15722315-position screw terminal1
155230Ribbon cable CPU to display1
155231Ribbon cable CPU to A/D1
155232Ribbon cable PS to CPU Board1
155233Ribbon cable CPU to Option 1
155234Pwr cord 115VAC, NEMA 5-15 plug1
155235Pwr cord 230VAC, Euro CEE 7/7 1
Part No.DescriptionQty
94422Label, Capacity1
53374Label, Annunicators1
14862Screw, 8-32NCx3/88
45042Washer,Bonded Sealing #8 x 0.3758
76514Conn, 6-Pin for LC and RS-232 ports2
103462Conn, 2- Pin for 20 mA port1
131325Screw, Fillister 8-32NC x ¼ 1
131320Screw, Fillister 8-32NC x ½1
Table 2-3. Parts Kit (PN 131314)
Item
No.
Part No.Description QTY
Table 2-2. Parts List
Replacement Parts19
Page 27
3.0Configuration
Note
There are two types of configuration parameters in the 480: Legal for Trade configuration and non-legal
configuration (or operator param eters). Legal for Trade configuration requires pressing the setup switch (see
Section 3.2). Non-legal configuration parameters do not require pressing the setup switch, but may be password
protected.
The following sections provide graphic representations of the 480 menu structures. In the actual menu structure,
the settings chosen under each parameter are arranged horizontally. To save page space, menu choices are shown in
vertical columns. The factory default setting appears at the top of each column in bold type. Most menu diagrams
are accompanied by a table that describes all parameters and parameter values associated with that menu.
Parameter level is indicated by the left LED position, levels 1-4.
See Section 1.7 on page 4 for navig ation methods.
A menu map is provided on the following page for reference.
20480 Legend Series
Page 28
Shaded Menus
Indicate lower
level values,
selections or
settings are
not shown.
SAVE LOAD
USBMEM
ADTEST NWGMNT WGMNST DIGIN DIGOUT ALGOUT DEFLT
999999TOTAL
OVRCAP
Raw A/D
Counts
1-41-40
25%
50%
100%
VIEW
HI
LOW
Press G/N
TEST
CNFG USER
Enter
Value
Enter
Value
PASWRD
SOURCE OFFSET ERRACT MIN
GROSS
NET
0%
20%
FULLSC
HOLD
ZEROSC
MAX MINNEG
NO
YES
Enter
Value
MAXNEG
Enter
Value
NO
YES
TWZERO TWSPAN OUTPUT
Enter
Value
Enter
Value
CURENT
VOLTAGE
ALGOUT
DIGIN1-4
OFF ZERO TARE NT-GRS UNITS DPSTAR PRINT CLRCN KBDLOC HOLD CLRTAR ACCUM CLRACC BATSTR BATSTP BATRES BATPUS BATRUN NET GROSS PRI SEC
The menu key can be pushed while in the weigh mode. The configuration parameters can be accessed by
pressing the setup switch while in the User Menu. See Section 3.2 for configuration setup.
Figure 3-1. Menu Key User Menu
ParameterChoicesDescription
AUDITCALIB
CONFIG
TARE0View the Tare value.
UNITID0 – 999999View and edit the Unit ID (up to a 6 digit number)
ACCUMVIEW
TIME
DATE
PRINT
CLR Y
TIMDAT
TIME
DATE
SETPNTBATCHG
SERIALCOM-1
SETPNT1-8
COM-2
STREAM
ETHNETDHCP
IPADR
NETMSK
USBDEVICE
DFTGWY
MAC
SERVER
CLIENT
TRIGER
View audit trail including the time and date of last configuration or calibration.
Options: CNT, TIME, DATE
View the current accumulator value.
View the time and date of the last accumulation.
Print or clear the accumulator data.
View/set time and date as currently programmed in the indicator.
See Section 1.8.13.
Format date asMMDDYY, DDMMYY, YYMMDD or YYDDMM.
See Section 3.1.1.
Configure serial ports.
See Section 3.1.2.
Configure optional Ethernet ports.
See Section Section 3.1.3.
View or configure the print format used for gross, net ticket, accumulator and setpoint
print tickets. See Section 7.0 on page 57 for more information.
See Section 3.1.5.
Displays the firmware and legally relevant firmware currently installed.
LR
MISCPOWER
Set miscellaneous parameters, see Section 3.1.7.
BKLGHT
BAT
Table 3-1. Menu Key Parameters (Continued)
3.1.1Setpoint Menu
For more information on setpoints, see Section 8.0 on page 59.
ParameterChoicesDescription
SETPT1-8USER
BATCHGOFF
SUPVSR
AUTO
MANUAL
Figure 3-2. Setpoint Setup Menu
Setpoint options that do not require a user password, see sub-menu below.
Setpoint options that require a user password, see sub-menu below.
Set to OFF batching is disabled. Set to AUTO or MANUAL to allow a batch sequence to run.
MANUAL requires a batch start signal (BATSTRT digital input or BATSTART serial command)
before the batch sequence can run. AUTO allows batch sequences to repeat continuously,
after receiving a single batch start signal.
Table 3-2. Setpoint Setup Menu Parameters
User Menu Setup23
Page 31
ParameterChoicesDescription
Note
Note
User submenus
VALUEnumberDisplay and edit the setpoint target value.
• For weight-based setpoints: specifies the target weight value, 0 – 999999.
• For time-based setpoints (delay): specifies, in 0.1-second intervals, a time value in the
range 0 – 65535.
ENABLEOFF
Enable or disable the setpoint.
ON
Supervisor Submenus
KINDSpecifies the setpoint kind and determines whether function is based on GROSS or NET
weight or based on time value for delay.
OFFSetpoint turned off/ignored.
GROSSGross setpoint. Performs functions based on the gross weight. The target weight entered is
considered a positive gross weight.
NETNet setpoint. Performs functions based on the net weight. The target weight entered is
considered a positive net weight value.
-GROSSNegative gross weight. Performs functions based on the gross weight. The target weight
entered is considered a negative gross weight.
-NETNegative net weight. Performs functions based on the net weight. The target weight entered is
considered a negative net weight value.
DELAYDelays the batch sequence for a specified time. The length of the delay (in tenths of a second)
is specified by the VALUE parameter.
TRIPHIGHER
LOWER
Trips the setpoint when the weight is higher or lower than the setpoint value.
If batch sequence is off (Continuous Setpoint) with:
• TRIP = HIGHER – the associated digital output will become active when the weight value
is higher than the setpoint value
• TRIP = LOWER – the output is active when the weight is below the setpoint value.
If batch sequence is on (Batching Setpoint) with:
• TRIP = HIGHER – the associated digital output is active until the setpoint value is reached
or exceeded
• TRIP = LOWER – the output is active until the weight goes below the setpoint value.
WAITSSOFF
Wait for Standstill – Value must be stable to satisfy this action.
ON
PUSHPROFF
ON
Push Print – Specify ON to perform a print operation when the setpoint is satisfied. This action
does not wait for standstill, it prints as soon as the setpoint is satisfied.
To wait for standstill before printing, set the WAITSS parameter to ON.
This action uses the Setpoint Print Format.
PUSHACOFF
ON
Push Accumulator – Specify ON to update the accumulator when the setpoint is satisfied. This
action does not wait for standstill, it accumulates as soon as the setpoint is satisfied.
To wait for standstill before accumulating, also set the WAITSS parameter to ON.
PUSHTROFF
ON
Push Tare – Specify ON to perform an acquire tare operation when the setpoint is satisfied.
This action does not wait for standstill, it tares the weight as soon as the setpoint is satisfied.
To wait for standstill before performing the tare, also set the WAITSS parameter to ON.
PUSHTR acquires the tare regardless of the value specified for the REGULAT
parameter on the PROGRAM menu.
DIGOUTNONE
1-8
Digital Output – specify the digital output associated with this setpoint.
To use outputs 1-4 and inputs 1-2, short jumper A1 (JP2)
To use outputs 5-8 and inputs 3-4, short jumpers A0 (JP1) and A1 (JP2)
BATSEQOFF
ON
Determines if the setpoint is a continuously running setpoint (OFF), or part of a batch sequence
(ON).
PREACTnumberAllows a setpoint to shut off before the setpoint is satisfied to allow for material in suspension.
HYSTERnumberSpecifies a band around the setpoint value that must be exceeded before the setpoint, once
off, can trip on.
Table 3-2. Setpoint Setup Menu Parameters (Continued)
See Section 9.3 for information about the 480 serial data format.
Figure 3-3. Serial Menu
User Menu Setup25
Page 33
ParameterChoicesDescription
Note
COM-1TRIGER
Specifies settings for COM-1.
BAUD
BITS
SBITS
TERMIN
EOLDLY
ECHO
PRNMSG
COM-2TRIGER
Specifies settings for COM-2.
BAUD
BITS
SBITS
TERMIN
EOLDLY
ECHO
PRNMSG
STRUR
STREAMSpecifies the operating mode of the indicator. See Section 9.4 on page 65.
NONEIndicator operates normally, COM-2 will not stream, even if the TRIGER is set to a streaming
parameter (STR1-5)
LOCALIndicator operates normally, data streaming occurs on COM-2 if the TRIGER is set to stream
data (STR1-5)
REMOTEIndicator acts as a Serial Data scale, displaying values based on a received steam of Serial
Data into COM-2. It also sends keypress commands back through COM-2 to the Local
indicator.
Submenus COM-1 and COM-2
TRIGER
(COM-1)
DEMANDDemand port – printing is performed by pressing the Print key or when a KPRINT EDP
Specifies the ports mechanism for triggering a transmission of data.
command is received. This port will also allow operation and EDP commands.
PRNPrinter Port – printing is performed by pressing the Print key. The port does not allow
operation of EDP commands.
NONEPort is inactive.
COMANDCommand port – allows operation of EDP commands, but will not print.
TRIGER
(COM-2)
NONEPort is inactive.
PRNPrinter Port – Printing is performed by pressing the Print key. The port does not allow
operation of EDP commands.
STR1Port is used to transmit a continuous stream of data according to RLWS stream format.**
STR2Port is used to transmit a continuous stream of data according to Toledo8142 stream
format.*
STR3Port is used to transmit a continuous stream of data according to Cardinal738 stream
format.*
STR4Port is used to transmit a continuous stream of data according to Weightronix WI-120
stream format.*
STR5Port is used to transmit a continuous stream of data according to consolidated stream
format.*
*STREAM must be set to local to enable streaming.
See Section 9.3 on page 62 for more information on Stream formats.
Com-2 streams simultaneously on both the RS-232 and 20mA outputs.
BAUD9600
Baud rate. Selects the transmission speed of data.
19200
38400
1200
2400
4800
BITS8NONE
Selects number of data bits and parity of data.
7EVEN
7ODD
Table 3-3. Serial Menu Parameters
26480 Legend Series
Page 34
ParameterChoicesDescription
Note
Note
SBITS1 STOP
Stop bits – Sets the number of stop bits to 1 or 2.
2 STOP
TERMINCR/LF
Termination character – selects line termination character(s) for data sent.
CR
EOLDLY000000
0 – 255
End-of-line delay – specifies, in 0.1 second intervals, the delay between transmitted lines of
data.
An end-of-line delay may be required for continuous transmission at slower
baud rates to ensure the receiving buffer is empty before another string is
transmitted.
ECHOON
OFF
PRNMSGOFF
Specifies whether characters received by the port are echoed back to the sending unit.
Port 1 default is ON, Port 2 default is OFF
Print displays when a demand print is performed.
ON
STRURCom 2 only – defines the stream data update rate when one of the stream formats is
selected.
STRLFTStream Legal for Trade – the weight information in the data stream is updated at the same
rate as the indicator's display.
STRINDStream industrial – the weight information in the data stream is updated at up to the A/D
Dynamic Host Configuration Protocol (DHCP)
ON – Dynamic allocation of IP address
Figure 3-4. Ethernet Menu Layout
OFF – Static allocation of IP address
IPADRR0.0.0.0IP address – a four field value, the range of each field is 0-255
NETMSK0.0.0.0SUBNET mask – a four field value, the range of each field is 0-255
DFTGWY0.0.0.0Default gateway – a four field value, the range of each field is 0-255
MAC00.00.00.00.00.00 View the MAC address in hexadecimal base (read only)
SERVERSee Submenu below.
CLIENTSee Submenu below.
DEMANDAllows operation of EDP commands and will print.
COMANDAllows operation of EDP commands only, does not print.
STRLFTStream legal for trade scale data – data is transmitted at the configured display update
rate. Will also accept EDP commands and print.
STRINDStream industrial scale data – data is transmitted up to the configured A/D sample rate.
Will also accept EDP commands and print.
TIMOUT0
0-65535
TERMINCR/LF
CR
SFMTSTR1-5Stream format – specifies the stream format used for streaming output of scale data
PORT10001
1-65535
RMOTIP0.0.0.0Remote Host IP address – IP address of the Remote Host (Server) the 480 Client will
RMPORT1
1-65535
EOLDLY0
0-255
ECHOOFF
ON
Timeout – inactivity disconnect timeout. A connection (either client or server) is closed if
there is no activity before the timeout expires. Time is in seconds. A timeout value of 0
disables the inactivity disconnect.
Termination. Selects the termination character(s) for data sent from the port.
(TRIGER=STRLFT or STRIND). See Section 9.3 on page 62.
The TCP port number of the 480 server.
Server level only.
connect to.
Client level only.
Remote Host TCP Port number – The TCP Port of the Remote Host (Server) the 480 Client
will connect to
Client level only.
Port end-of-line delay – specifies (in 0.1 second intervals) the delay between transmitted
lines of data.
Specifies whether characters received by the port are echoed back to the sending unit.
Table 3-4. Ethernet Menu Parameters (Continued)
3.1.4USB Menu
Figure 3-5. USB Menu Layout
ParameterChoicesDescription
DEVICEHOSTPCThe port is assigned to a HOSTPC, the protocol is a demand type similar to serial
communications demand mode, meaning that the port supports EDP commands and Printing.
DRIVEThe port is assigned to a flash drive, the data is written in a FAT32 system, no special driver is
PRINTRWhen set to PRINTR mode, the USB port can be connected to a USB printer using an
TRIGERCOMANDAllows operation of EDP commands only. Does not print.
required.
Use the USBMEM Load and Save functions, found in the Configuration Menu (Section 3.2.1 on
page 33), to load configuration data to or from a flash drive.
appropriate USB cable. See Section 9.8 on page 66.
DEMANDAllows operation of EDP commands and prints.
Table 3-5. USB Menu
User Menu Setup29
Page 37
3.1.5Print Format Menu
TIMDAT
ACCUM
SERIALSETPNT
PFRMAT
......
Display first 6
characters of format
Display and edit
active character and
ASCII value
Delete active
character
Scroll right in format string
Scroll left in format string
Increment ASCII value of active character
Decrement ASCII value of active character
Pressto insert a space
before the active character
POWERAUTOThe power up functionality depends on whether the battery option is installed:
• If the battery option is not installed, the indicator turns on automatically when connected
to AC power. It turns off by disconnecting the AC power or by pressing the power key.
• If the battery option is installed, and CPU Jumper 4 is connected, the indicator turns on
automatically when connected to AC power. It turns off by disconnecting the AC power or
by pressing the power key. This configuration is not recommended due to increased
leakage current from the battery when the indicator is turned off.
• If the battery option is installed, and CPU Jumper 4 is disconnected, the indicator only
turns on and off by pressing the power key.
MANUALIndicator powers On/Off by pressing Power key.
BKLGHTOFF
ON
AUTO
BATLEVELAllows reading the battery level in mV units.
Allows control of the LCD backlight.
OFF – backlight always off.
ON – backlight always on.
AUTO - backlight will turn off after the configured amount of time, if no buttons are pressed
and there is no change in the weight. Once either of these occurs, the backlight will turn on.
Auto time Options: 10 sec, 30 sec, 1 min, 5 min,10 min.
If there is not a battery option installed, NOBATT will display.
When the battery is charging, CHRNG will display.
Table 3-8. MISC Menu Parameters
User Menu Setup31
Page 39
3.2 Configuration Using the Front Panel (Legal for Trade)
Note
The 480 indicator can be configured using a series of menus accessed through the indicator front panel when the
indicator is in configuration mode.
When the indicator is placed in configuration mode,
first of eight main menus used to configure the indicator. Detailed descriptions of these menus are given in
Section 3.2.1 on page 33. When configuration is complete, return to
configuration mode and save changes. Then replace the configuration switch access screw.
T o place the indicator into the configuration mode, a configuration switch is accessed by removing the setup switch
access head screw on the enclosure back, see Figure 2-1 on page 10. The switch is activated by inserting a
non-conductive tool into the access hole after pressing the menu switch.
1. Remove setup switch access screw from back of indicator.
2. Using a non-conductive tool, press the button located in the access hole. This allows access to configuration
menus.
CONFIG is shown on the display. The CONFIG menu is the
CONFIG and press the (Zero) key to exit
Menu
CONFIGConfigurationConfigGraduations (scale capacity), Configure zero tracking, zero range, motion band,
FORMATFormatConfigSet primary and secondary units.
CALIBRCalibrationConfigCalibrate the indicator. See Section 4.0 on page 46 for calibration procedures.
PROGRMProgramConfigSet power-up and regulatory modes, consecutive numbering and sleep
DIG INDigital InputUserAssign digital input functions. Used only if the Relay option is installed.
ALGOUTAnalog OutputConfigConfigure the analog output module. Used only if analog output option is
PASWRDPasswordConfigEnable and edit configuration and user passwords.
USBMEMUSB MemoryConfigAllows the saving and loading of configuration settings from a USB flash
TESTTestConfigUsed to test A/D, test and set Digital I/O, reset the indicator to default settings
MENUMenuUserNon Legal for Trade settings, such as Audit Trail, view Tare, set Unit ID, view/
* Password required for all menus if a password for that level has been entered. See Section 3.2.8.
Required
Password*
Menu Function
overload, tare function, sample rate and digital filtering parameters.
parameters.
installed.
memory device.
and set over and under weighment characteristics.
print Accumulator, set Time/Date, and Non Legal for Trade setting for
Communication Port, Setpoints and Print Format parameters. See Section 3.1.
Can be accessed without pressing the Setup Switch, by pressing
the Menu key on the front panel.
Table 3-9. 480 Menu Summary
32480 Legend Series
Page 40
3.2.1Configuration Menu Structures and Parameter Descriptions
Graduations – specifies the number of full scale graduations. The value entered must be in
the range 1-100 000 and should be consistent with legal requirements and environmental
limits on system resolution.
Zero track band – automatically zeros the scale when within the range specified, as long as
the input is within the configured zero range (ZRANGE parameter). Selections are ± display
divisions. Maximum legal value varies depending on local regulations.
ZRANGE
2%
1% – 100%
INIZR0 %
0 – 100%
MOTBAN0.1d
0.0d – 100d
Zero range – selects the range within which the scale can be zeroed. The 2% selection is ±
2% around the calibrated zero point, for a total range of 4%. Indicator must be at standstill
to zero the scale.
Initial Zero Range at power up – see Error messages HINOFF and LINOFF in Section 9.1.1
If on power up the weight value is between the ±% range specified of Calibrated Zero, the
indicator will zero off that weight. If outside that range, the indicator will display HINOFF or
LINOFF, depending on whether the initial reading is above or below the range.
If not set at zero, hopper/scale MUST be empty when powered up.
Motion band – sets the level, in display divisions, at which scale motion is detected. If
motion is not detected, the standstill symbol lights.
Some operations, including print, tare and zero, require the scale to be at standstill.
Maximum legal value varies depending on local regulations.
Setting the MOTBAN setting to 0.0 disables motion sensing, the scale will
always be considered stable.
Table 3-10. Configuration Menu Parameters
Description
Configuration Using the Front Panel (Legal for Trade)33
Overload – determines the point at which the display blanks and an out-of-range error
message is displayed. Maximum legal value varies depending on local regulations.
Display rate – sets the update rate for displayed values. Values are in seconds (SEC).
No display filtering – the display is updated at up to the A/D sample rate.
A/D Sample rate – selects measurement rate, in samples per second, of the
analog-to-digital converter. Lower sample rate values provide greater signal noise immunity.
Adaptive Filter – selects the adaptive digital filter. See Figure 3-10.
Rolling Average Filter - selects the rolling average filter. See Figure 3-10.
OFF- Disables the digital filter.
See Section 9.11 on page 68 for more information on Digital Filtering.
Tare function – enables or disables push-button and keyed tares.
Both push-button and keyed tares are enabled
Table 3-10. Configuration Menu Parameters (Continued)
3.2.2Adaptive Filter and Rolling Filter Sub-menu Parameters
Figure 3-10. Adaptive Filter and Rolling Averaging Filter Sub-menu
34480 Legend Series
Page 42
ParameterChoices
Description
ADPFIL Sub-menu
AFSENSLIGHT
MEDIUM
HEAVY
Digital filter cutout sensitivity – the light setting responds quickly to small weight changes and
is less stable. The heavy setting responds slowly to small weight changes and is more stable.
Use the heavy setting in an unstable environment.
AFTHRH000002Digital filter cutout threshold – specifies the filter threshold, in weight value. When a weight
change exceeds the value set for threshold, digital filtering is suspended. A setting of zero
disables the filter. The decimal point in the entered value is assumed, so if the display division
is 1lb, a value of 1 is equal to 1lb. If the dd is 0.1lb, a value of 1 is equal to 0.1lb.
ROLFIL Sub-menu
RFSTG1
RFSTG2
RFSTG3
2, 4, 8, 16,
32, 64, 1
Rolling filter stages – the number of A/D samples that are averaged in each stage.
A higher number gives a more stable display by minimizing the effect of a few noisy readings,
but slows down the settling rate of the indicator. See Section 9.11 on page 68 for more
information on digital filtering.
RFSENS8OUT
16OUT
32OUT
Rolling filter cutout sensitivity – specifies the number of consecutive readings that must fall
outside the rolling filter threshold (RFTHRH parameter) before digital filtering is suspended. If
NONE is selected, the filter is always enabled.
64OUT
128OUT
2OUT
4OUT
RFTHRHNONE
2 dd
5 dd
Rolling filter cutout threshold – specifies the rolling filter threshold, in display divisions. When a
specified number of consecutive scale readings (RFSENS parameter) fall outside of this
threshold, digital filtering is suspended. If NONE is selected, the filter is always enabled.
10 dd
20 dd
50 dd
100 dd
200 dd
250 dd
Table 3-11. Adaptive Filter and Rolling Filter Submenu Parameters
Configuration Using the Front Panel (Legal for Trade)35
Primary Units – specifies the capacity, decimal position, display divisions, and units used.
Secondary Units – specifies the capacity, decimal position, display divisions, units and
conversion multiplier used. See below for submenu parameter descriptions.
Decimal point location – specifies the location of the decimal point in the primary unit
display. Value should be consistent with local legal requirements.
EDIT is the current calibrated value that can be edited.
Note
See Section 4.0 on page 46 for Calibration p r ocedures.
Figure 3-12. Calibration Menu
ParameterChoicesDescription
WZERO
CAL
EDIT
WVALnumber
WSPANCAL
EDIT
REZEROZERO
(Flashing)
Do not adjust this value after WSPAN has been set.
Press
to display ENTER, then press to capture the zero value.
Press to display the current captured zero mV value. The value may be edited as required.
Press to save the edited value.
Press to display and edit the test weight value.
Press
to display ENTER, then press to capture the net span value.
Press to display the current captured net span mV value. The value may be edited as
required. Press to save the edited value.
Press to remove the weight of the testing apparatus like hooks or chains used during the
calibration process.
Always use this parameter after WZERO and WSPAN have been set to recapture a new zero
value. See Section 4.1 on page 46 for more information about using this parameter.
Table 3-13. Calibrat i on Menu Parameters
Configuration Using the Front Panel (Legal for Trade)37
Power up mode – when set to GO, the indicator goes into operation immediately after a brief
power up display test.
When set to DELAY, the indicator performs a power up display test and then enters a 30 second
warm-up period. If no motion is detected during the warm-up period, the indicator becomes
operational when the warm-up period ends; if motion is detected, the delay timer is reset and the
warm-up period repeated.
REGULANTEP
OIML
CANADA
NONE
Regulatory mode – specifies the regulatory agency having jurisdiction over the scale site.
OIML, NTEP, and CANADA modes allow a tare to be acquired at any weight greater than zero.
NONE allows tares to be acquired at any weight value.
OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight is at no
load. NONE allows tares to be cleared at any weight value.
NTEP and OIML modes allow a new tare to be acquired even if a tare is already present. In
CANADA mode, the previous tare must be cleared before a new tare can be acquired.
NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net mode as
long as the current weight is within the specified ZRANGE. In OIML mode, the scale must be in
gross mode before it can be zeroed; pressing the ZERO key in net mode clears the tare.
CONSNU000000
0 – 999999
Consecutive numbering. Allows sequential numbering for print operations. The consecutive
number value will increment following each print operation when the print format used contains
the <CN>, or consecutive number formatting token. This value is read only.
The initial value of this parameter is set to the start up value specified on the
CONSTU parameter. Changing CONSTU immediately resets the consecutive
number used for printing.
CONSTU000000
0 – 999999
Consecutive number start up value. Specifies the initial consecutive number (CONSNU) value
used when the indicator is reset or configured.
Consecutive numbering starts with number entered.
Example: If 100 is entered the first ticketed printed will be 100.
ACCUMOFF
ON
RTZGRD0
0 – 999999
Turns the accumulator on and off. Stores the accumulated weight, time and date of last
accumulation.
Return to zero grads to re-arm the accumulator. Default = 0.
Defines the number of graduations from Zero that the weight has to go below before the
accumulator will be re-armed.
DATEDATFMT
DATSEP
TIMETIMFMT
Allows selection of date format and date separator. See Date sub-menu for parameter
descriptions.
Allows selection of time format and separator. See Time sub-menu for parameter descriptions.
TIMSEP
SLEEPNEVER
1MIN
2MIN
5MIN
10MIN
Sleep mode. When weight is stable, no buttons have been pressed, and the weight is below the
sleep weight threshold (WGTHRH), the indicator will enter the sleep mode in the time defined in
the SLEEP Parameter. A NEVER setting will disable the sleep mode.
See Section 9.6 on page 65.
The SLEEP mode will only operate if the indicator is operating on Internal Battery
Power - disconnected from AC power.
Setting the WGTHRH or DDTHRH parameters to 0 also disables sleep mode.
Tare is removed if the indicator goes into sleep mode.
STDBY1SEC
5SEC
10SEC
15SEC
30SEC
60SEC
Controls the Standby mode.
When in the sleep mode, at the number of STDBY seconds set, the indicator will wake if:
• there is weight on the scale of more than the WGTHRH value
• the weight has is changed greater than DDTHRH value
• it detects that a button is being pressed
OFF
Table 3-14. Program Menu Parameters
Configuration Using the Front Panel (Legal for Trade)39
Page 47
ParameterChoicesDescription
WGTHRH0 – FS
(in primary
units )
DDTHRH999999
(in display
divisions )
DATE sub-menus
DATFMTMMDDYY
DDMMYY
YYMMDD
YYDDMM
DATSEPSLASH
DASH
SEMI
TIME sub-menus
TIMFMT24HOUR
12HOUR
TIMSEPCOLON
COMMA
Controls the Weight Threshold of the Sleep Mode.
If the weight is above WGTHRH, the indicator will not go into sleep mode.
If the weight is below WGTHRH, it will go into sleep mode if the weight remains stable and no
buttons are pressed, for the time set in the Sleep mode parameter.
Controls the weight change detection feature of SLEEP mode, once the indicator detects that
the weight level on the scale is changed more than DDTHRH it will wake up.
Specifies the format used to display or print the date.
Specifies the date separator character.
Specifies the format used to display or print the time.
Specifies the time separator character.
Table 3-14. Program Menu Parameters (Continued)
40480 Legend Series
Page 48
3.2.6Digital Input Menu
DIGIN1-4
OFF
PRINT
ZERO
TARE
NT-GRS
UNITS
DPSTAR
CLRCN
CONFIGFORMATCALIBRPROGRMDIG INALGOUTPASWRD
USBMEM
MENU
TEST
KBDLOC
BATSTR
BATSTP
BATRES
HOLD
CLRTAR
ACCUM
CLRACC
BATRUN
BATPUS
NET
GROSS
PRI
SEC
Figure 3-14. Digital Input Menu
ParameterChoicesDescription
DIGIN1
DIGIN2
DIGIN3
DIGIN4
OFFSets the Digital Input to perform no function.
ZEROSame function as the front panel key.
TARESame function as the front panel key.
NT-GRSNet-Gross – same function as the front panel key.
UNITSSame function as the front panel key.
DSPTARDisplays the current tare value.
PRINTSame function as the front panel key.
CLRCNResets the consecutive number to the value specified on the CONSTU parameter (PROGRM
KBDLOCDisables the keypad while the digital input is held active.
HOLDHolds the current display. Releasing this input returns to normal display update mode.
CLRTARClears the tare.
ACCUMDisplays the accumulator.
CLRACCClears the accumulator.
BATSTRStarts a batch process.
BATSTPStops a batch; any restart starts the batch at the beginning.
BATRESResets the batch process so the pointers are aimed at first batching setpoint.
BATPUSPauses a batch, but allows it to restart at the point where it was paused.
BATRUNA switch that allows a batch to be started. It must be active to allow any batch to run.
NETPuts the scale in net mode.
GROSSPuts the scale in gross mode.
PRIPuts the scale in primary units.
SECPuts the scale in secondary units.
Specifies the function activated by digital inputs 1, 2, 3 and 4.
menu).
If not active, and a BATSTRT input is activated, the batch will abort or reset.
Table 3-15. Digital Input Menu Parameters
Configuration Using the Front Panel (Legal for Trade)41
If the analog output option is installed, configure all other indicator functions and calibrate the indicator (see
Section 4.0 on page 46) before configuring the analo g output. See Section 9.12 on p age 70 for analog output
calibration procedures.
Figure 3-15. Analog Output Menu
ParameterChoicesDescription
SOURCEGROSS
OFFSET0
ERRACT
MIN000000
MAX010000
MINNEGNO
MAXNEGNO
TWZERO000000
TWSPAN000000
OUTPUT
NET
20
FULLSC
HOLDHold current value.
ZEROSCSet to zero value (0 V or 0 mA) or (2 V or 4 mA)
0 – 999999
0 – 999999
Yes
Yes
0 – 65535
0 – 65535
CURENT
VOLTAG0-10 V or 2-10 V, depends on OFFSET.
Specifies the weigh data, gross or net, tracked by the analog output.
Zero offset in %. Select 0% for 0–10 V / 0-20 mA; select 20% for 2-10 V / 4-20 mA output.
Set OFFSET after calibrating 0-20 mA.
Error action – specifies how the analog output responds to system error conditions.
Set to full value (10 V or 20 mA).
Specifies the minimum weight value tracked by the analog output. Specify a weight value (in
primary units) in the range 0 – 999999.
Specifies the maximum weight value tracked by the analog output. Specify a weight value (in
primary units) in the range 0 – 999999.
YES implies that the value in MIN is a negative value. NO is the default and implies that MIN is
a positive value.
YES implies that the value in MAX is a negative value. NO is the default and implies that MAX is
a positive value.
Tweak zero – adjust the analog output zero calibration. Use a multimeter to monitor the analog
output value. Press
(flashing) digit. Press to save the new value.
Tweak span – adjust the analog output span calibration. Use a multimeter to monitor the
analog output value. Press
selected (flashing) digit. Press to save the new value.
Defines the output type.
0-20 mA or 4-20 mA, depends on OFFSET.
Table 3-16. Analog Output Menu Parameters
orto select the digit to edit. Press or to edit the selected
orto select the digit to edit. Press or to edit the
Enter 999999 to reset password, this will also reset the configuration back to default values.
Edit configuration password. 0 = no password
If enabled, the configuration password is used for electronic sealing of the
indicator. When configuration password has been entered, the existing
configuration password is required to edit. Entering this menu requires the
SETUP switch to be pressed.
Edit user password. 0 = no password
If enabled, the user password restricts access to the setpoint, serial and print
formatting menus. Once a user password has been entered, the existing user
password is required to edit parameters.
Table 3-17. Password Menu Parameters
3.2.9USBMEM Menu
USBMEM features will be available in future versions.
Figure 3-17. USBMEM Menu
ParameterChoicesDescription
SAVESave the configuration parameters to a file on a flash drive.
LOADLoad the configuration parameters from a file on a flash drive.
Table 3-18. USBMEM Menu Parameters
For more information on using a flash drive to load and save configuration settings, see Section 9.8 on
page 66.
Configuration Using the Front Panel (Legal for Trade)43
Access to this menu requires the configuration password if it has been enabled.
Figure 3-18. Test Menu
Digital output relay test.
HI = Active = ON
LO = Inactive = OFF
ParameterChoicesDescription
ADTESTDisplays raw A/D counts.
NWGMNTTOTALThe number of weighments that exceed value set in WGMNST.
OVECAPThe number of weighments that exceed the capacity of the scale as set in the config and
format menus. See Table 3-10 and Table 3-12.
WGMNST0 – 999999Weighment Threshold – Sets the weight threshold over which WGMNST/Over value will be
counted.
DIGIN1 – 4
VIEW
Access to display status. Allows selection of each Digital Input to display its current state, HI
or LO.
If the setting cannot viewed, that option is not installed.
DIGOUT1 – 8
HI
LO
ALGOUT0
25%
50%
100%
Access to each of the Digital Output and to manually set them individually to HI or LO. HI is
active.
If the setting cannot viewed, that option is not installed.
Allows analog output to be manually set to certain output values for calibration and test of
external equipment
The analog output must not be in an error state as that will override these
settings. When the display division has a decimal, the test outputs will be
offset by the decimal place.
If the setting cannot viewed, that option is not installed.
DEFLTResets all parameters to factory defaults. At DEFLT, press Gross/Net, indicator will display
SAVING and cycle power.
Table 3-19. Test Menu Parameters
3.2.11 User Menu Setup
See Section 3.1 on page 22.
44480 Legend Series
Page 52
3.3 Revolution® Configuration
The Revolution configuration utility provides another method for configuring the 480 indicator. Revolution runs on
a computer to set configuration parameters for the indicator. When Revolution configuration is complete,
configuration data is downloaded to the indicator. Revolution supports both uploading and downloading of
indicator configuration data. This capability allows configuration data to be retrieved from one indicator, edited
and then downloaded to another indicator with an identical hardware configuration. See Section 4.3 on page 47 and
Section 5.0 on page 49 for more information about using Revolution.
Revolution® Configuration45
Page 53
4.0Calibration
Note
The 480 can be calibrated using the front panel or the Revolution® configuration utility. Each method consists of
the following steps:
• Zero calibration
• Entering the test weight value
• Span calibration
• Optional rezero calibration for test weights using hooks or chains.
The following sections describe the calibration procedure for each of the calibration methods.
4.1 Front Panel Calibration
The CALIBR sub-menu is used to calibrate the 480. See Section 3.2.4 for the calibration menu structure and
parameter descriptions. To calibrate the indicator using the front panel:
1. Remove the setup switch access screw from the back of the enclosure.
2. Insert a non-conductive tool into the access hole and press the configuration switch. Indicator display
changes to show
3. Press until
4. Remove all weight from the scale platform. If the test weights require hooks or chains, place the hooks or
chains on the scale for zero calibration.
5. Press to
6. Press to
7. Press to
8. Press to calibrate zero. The indicator displays
mV value for the zero calibration is displayed.
9. Press to save the zero calibration value and go to
10. Press to
11. Press to
12. Press to show the test weight value.
13. Press to edit value using the procedure shown in Sectio n 1.7 on page 4 to enter the actual value of the test
weights to be used for the span calibration. Press to store the entered
14. Press to
15. Press to
16. Press to
17. Press to
18. Place test weights on the scale and press to calibrate span. The indicator displays *CAL* while calibration
is in progress. When complete, the net millivolt value of the test weight for the span calibration is displayed.
CONFIG (see Figure 2-1).
CALIBR is displayed.
WZERO.
CAL or press to view or edit the existing value.
ENTER.
*CAL* while calibration is in progress. When complete, the
CAL.
WZERO.
WVAL.
WVAL value.
WVAL.
WSPAN.
CAL or press to view or edit the existing value.
ENTER.
Net mV value is equal to the raw mV value at the span minus the mV value captured for zero.
19. Press to save value and go to CAL.
20. Press to WSPAN. If using chains or hooks for your test weights, proceed with step 21.
If no chains or hooks were used, proceed with step 23.
21. Press to REZERO. The REZERO function is used to remove a calibration offset when hooks or chains are
used to hang the test weights.
22. Remove the hooks or chains used during calibration and remove the test weights from the scale. Press .
ZERO will flash. With all weight removed, press to rezero the scale. This function adjusts the zero and
span calibration values. The indicator displays *CAL* while the zero and span calibrations are adjusted.
23. Press to return to the CALIBR menu.
24. Remove all weight from the scale. Press until the display reads CONFIG. Press to exit configuration
mode.
46480 Legend Series
Page 54
4.2 EDP Command Calibration
Note
To calibrate the indicator using EDP commands, the indicator COM 1 port must be connected to a terminal or
computer. See Section 6.0 on pa ge 51 for more information about using EDP commands.
Once the indicator is connected to the sending device, use the following steps:
1. Place the indicator in configuration mode (display reads
CONFIG) and remove all weight from the scale
platform. If your test weights require hooks or chains, place them on the scale for zero calibration.
2. Send the SC.WZERO EDP command to calibrate zero. The indicator displays
*CAL* while calibration is in
progress.
3. Place test weights on the scale and use the SC.WVAL command to enter the test weight value in the
following format:
WVAL=nnnnnn<CR>
4. Send the SC.WSPAN EDP command to calibrate span. The indicator displays *CAL* while calibration is in
progress.
5. T o remove an offset value, clear all weight from the scale, including items used to hang test weights, and then
send the SC.REZERO EDP command. The indicator displays
*CAL* while the zero and span calibrations are
adjusted.
6. Send the KUPARROW EDP command to exit configuration mode.
4.3 Revolution Calibration
To calibrate the indicator using Revolution, the indicator COM 1 port must be connected to a PC running the
Revolution configuration utility.
Use the following procedure to calibrate the indicator:
1. Select
2. Revolution uploads calibration data from the indicator and presents the information in the calibration wizard,
as shown in Figure 4-1.
Calibration Wizard from the Revolution tools menu.
Figure 4-1. Revolution Calibration Display
3. Enter the Value of Test Weight to be used for span calibration and click OK.
4. The zero calibration dialog box prompts to remove all weight from the scale. Clear the scale and click
OK to
begin zero calibration.
If the test weights require hooks or chains, place them on the scale for zero calibration.
5. When zero calibration is complete, the span calibration dialog box prompts to place test weights on the scale
for span calibration. Place test weights on the scale and click
6. When calibration is complete, the
Exit to save the new values and return to the Revolution main menu; to restore the previous calibration values,
New Settings fields of the indicator calibration display are filled in. Click
OK.
click Restore Settings.
EDP Command Calibration47
Page 55
4.4 More About Calibration
The following provides additional information about WVAL and how to manually tweak a calibration up or down.
WVAL
When a WVAL number is entered that uses a decimal point, the value will be rounded when displayed.
Example:
A precision value such as 2455.23 is entered by the operator, but when the operator returns to the WVAL
display, the number is rounded to 2455. A higher precision value can be entered for WVAL when the Display
division is greater than 1, such as X2, X5 or X10.
Millivolt Values
Zero is the dead weight millivolt value and Span is the test weight millivolt value, so the total millivolts equals Zero
plus Span.
4.4.1Adjusting Final Calibration
Calibration may be affected by environmental factors including wind, vibration and angular loading.
Example:
A strain test may determine that at 2000 lb the calibration is 2 lb high. In this case, final calibration can be
tweaked by changing WVAL to 999 lb. This adjustment provides a linear correction of 1 lb per 1000 lb.
T o adjust the final calibration, return to the
to adjust calibration up or down. Press to save the value. Press to return to the CALIBR menu.
WVAL prompt and press to show the test weight value. Press or
48480 Legend Series
Page 56
5.0Revolution
Note
The Revolution® utility provides a suite of functions used to support configuration, calibration, customization and
backup of the 480 software.
Calibration values and scale configuration can both be saved and restored to the 480 using Revolution.
Hardware and Software Requirements
•Minimum system requirements: 166 MHz, x86-compatible, with 32MB RAM (64MB for NT4/2000/XP),
40MB disk space
•Recommended system: 233 MHz, x86-compatible or greater, with 64MB RAM, 300 MB disk space
®
Revolution runs on most Windows
8.1 and Windows 10 (both 32 bit and 64 bit architectures are supported).
System requirements to run Revolution are as follows:
•1 GHz CPU (x86-32 bit or x86-64 bit)
•512 MB RAM
•850 MB disk space (32 bit) or 2 GB disk space (64 bit)
5.1Connecting to the Indicator
Connect the PC serial port to com 1 of the 480, then click Connect in the toolbar. Revolution attempts to establish
communications to the indicator. If communications settings need to be adjusted, select
menu.
Downloading to the Indicator
The Download Configuration function on the Revolution Communications menu allows a Revolution configuration
file (with or without scale calibration data) or ticket formats to be downloaded to a connected indicator in
configuration mode.
The
Download Section function on the Communications menu allows the download of only the currently displayed
object, such as a scale configuration.
Because less data is transferred using the
but there is an increased possibility that the download may fail due to dependencies on other objects. If the
download fails, try performing a complete download using the
operating systems, including Windows V ista SP2, Windows 7 SP1, Windows
Options… from the Tools
Download Section, it is typically faster than a full configuration download,
Download Configuration function.
Uploading Configuration to Revolution
The Upload Configuration function on the Revolution Communications menu allows the existing configuration of a
connected indicator to be saved to a file on the PC. Once saved, the configuration file provides a backup that can be
quickly restored to the indicator if needed. Alternatively, the file can be edited within Revolution and downloaded
back to the indicator.
5.2Saving and Transferring Data
Revolution has a module to save and transfer data. It is the preferred method over using ProComm or Hyper
Term in al .
5.2.1Saving Indicator Data to a Personal Computer
Configuration data can be saved to a computer connected to the selected port. The PC must be running a
communications program such as PROCOMMPLUS
When configuring the indicator, ensure that the values set for the baud and bits parameters on the serial menu
match the baud rate, bits and parity settings configured for the serial port on the PC.
To save all configuration data, first put the communications program into data capture mode, then
indicator in configuration mode and send the DUMPALL command to the indicator. The 480 responds by sending
all configuration parameters to the PC as ASCII-formatted text.
5.2.2Downloading Configuration Data from PC to Indicator
Configuration data saved on a PC or disc can be downloaded from the PC to an indicator. This procedure is useful
when a number of indicators with similar configurations are set up or when an indicator is replaced.
®
.
place the
Connecting to the Indicator49
Page 57
To download configuration data, connect the PC to the selected port as described in Section 5.2.1. Place the
Note
Note
indicator in configuration mode and use the PC communications software to send the saved configuration data to
the indicator. When transfer is co mp lete, calibrate the indicator as described in Section 4.0.
5.3Reflash Indicator
The 480/482 flash loader program must be downloaded on the PC being used.
Reflash cable (PN 165379) is required for this process.
5.3.1480/482 Legend Series Flash Loader
1. Download the 480/482 Legend Series Flash Loader file from the website at: www.ricelake.com.
2. Save the file to a PC or select open from the pop-up.
3. Double click on the file
4. Select
RUN. The license agreement box will display.
5. Check the box to accept the license. A box will display showing the installation.
6. When installation is complete, select
5.3.2480/482 Legend Series Firmware Update
1. Download the latest firmware .hex file to a PC from: www.ricelake.com
2. Remove the back of the indicator. See Section 2.2 on page 10.
3. Connect the flash cable (PN 165379) betwee n the PC and the indicator by plugging into J14 on the CPU
board.
If there are option cards in place, they will need to be removed to gain access to the CPU board connection.
See Section 2.5 on page 15.
Ensure the cable is plugged in correctly, otherwise the reflash will not work. The indicator display will shut
down when the flash cable is plugged in.
4. Open the 480FlashLoader.exe program.
5. Select the serial port the 480 is connected to on the PC.
6. Navigate to the .hex file needed to be loaded.
7. Click the
Load key in the program to start the reflash.
Setup480FlashLoader.exe. An open file box will display.
FINISH.
50480 Legend Series
Page 58
6.0EDP Commands
Note
EDP commands are case sensitive. Commands must be entered in upper case.
The 480 indicator can be controlled by a computer or remote keyboard connected to one of the indicator’s
communication ports. Control is provided by a set of EDP commands that can simulate front panel key press
functions, display and change setup parameters, and perform reporting functions. This section describes the EDP
command set.
6.1The EDP Command Set
The EDP command set can be divided into five groups: key press commands, reporting commands, the
RESETCONFIGURATION special function command, parameter setting commands and transmit weight data
commands.
When the indicator processes an EDP command, it responds with the message
the command was received and has been executed. If the command is unrecognized or cannot be executed, the
indicator responds with
??.
The following sections list the commands and command syntax used for each of these groups.
6.1.1Key Press Commands
Key press EDP commands (see Table 6-1) simulate pressing the keys on the front panel of the indicator. Most
commands can be used in both setup and weighing mode. Several of the commands serv e as pseudo keys,
providing functions that are not represented by a key on the front panel.
For example, to enter a 15 pound tare weight using EDP commands:
1. Type K1 and press
2. Type K5 and press
3. Type KTARE and press
Enter (or Return).
Enter.
Enter. The display shifts to net mode when the tare is entered.
OK. The OK response verifies that
CommandFunction
KZEROIn normal mode, press the ZERO key.
KGROSSNETIn normal mode, press the GROSS/NET key.
KGROSSGo to gross mode.
KNETGo to net mode.
KTAREPress the TARE key.
KUNITSIn weighing mode, press the UNIT key.
KPRIMGo to primary units.
KSECGo to secondary units.
KPRINTIn normal mode, press the PRINT key.
KDISPACCUMDisplays current accumulator value.
KCLRACCUMClears accumulator.
KDISPTAREDisplays tare.
KMENUPresses the MENU key.
KCLRCNReset consecutive number.
KCLRTARClear tare from system.
KLEFTARROWIn setup mode, move left in the menu.
KRIGHTARROW In setup mode, move right in the menu.
KUPARROWIn setup mode, move up in the menu.
KDOWNARROW In setup mode, move down in the menu.
KSAVEIn setup mode, saves the current configuration. Command works only if a parameter has been changed.
KLOCKLock specified front panel key. For example, to lock the ZERO key, enter KLOCK=KZERO. Setting for KLOCK
KUNLOCKUnlock specified front panel key. For example, to unlock the PRINT key, enter KUNLOCK=KPRINT.
KCLRPress the CLEAR key.
KDATEDisplays date.
KTIMEDisplays time.
KTIMEDATEDisplays time and date.
is lost when power cycles.
Table 6-1. Serial Key Press Commands
The EDP Command Set51
Page 59
6.1.2Reporting Commands
Note
Reporting commands (see Table 6-2) send specific information to the selected port. These commands can be used
in both setup mode and weigh mode.
CommandFunction
DUMPALLLists all parameter values.
DUMPAUDITLists audit data information.
PWrites current displayed weight with units identifier. See Section 9.2 on page 61 for more information.
ZZWrites current weight and annunciator status. See Section 9.2 on page 61 for more information.
SWrites one frame of stream format.
VERSIONReturns the currently installed firmware version.
Table 6-2. EDP Reporting Commands
6.1.3The RESETCONFIGURATION Command
The RESETCONFIGURATION command can be used to restore all configuration parameters to their default
values. Before issuing this command, the indicator must be placed in configuration mode.
This command is equivalent to using the DEFLT function on the TEST menu. See Section 9.1.2 for more
information about test mode and using the TEST menu.
All load cell calibration settings are lost when the RESETCONFIGURATION command is run.
6.1.4Parameter Setting Commands
Parameter setting commands allow the display or change of the current value for a particular configuration
parameter.
Current configuration parameter settings can be displayed in either setup mode or weigh mode using the following
syntax:
command<ENTER>
Most parameter values can be changed in setup mode only; setpoint parameters listed in Table 6-10 can be changed
when in weighing mode.
Use the following command syntax when changing parameter values:
command=value<ENTER>, where value is
either a number or a parameter value. Use no spaces before or after the equal (=) sign. If an incorrect command is
typed, the display reads
6.1.5Soft Reset
??.
The command RS will reset the indicator without losing configuration and calibration. It does reset consecutive
number and count mode.
CommandDescriptionValues
SC.GRADSGraduations1-100000
SC.ZTRKBNDZero track band0.0-3.0
SC.ZRANGEZero range1 - 100 (example: to make it 5%, enter "SC.ZRANGE=5").
SC.INIZRInitial zero range0-100 (if set to zero , no initial zero is performed)
SC.MOTBANDMotion band0.0 - 100 (in display divisions)
Example: to enter 5 divisions, send "SC.MOTBAND=5.0".
SC.OVRLOADOverloadFS+2%, FS+1D, FS+9D, FS
SC.DSPRATDisplay Rate0.1 SEC, 0.25 SEC, 0.5 SEC, 0.75 SEC, 1 SEC, NONE
SC.SMPRATSample rate5HZ, 40HZ, 20HZ, 10HZ
SC.DIGFILSelect Digital Filter TypeADAPTIVE, ROLLING, OFF
SC.AFILTER.SENSITIVITYAdaptive digital filter cutout
sensitivity
SC.AFILTER.THRESHOLDAdaptive digital filter cutout
SC.PRI.DECPNTPrimary units decimal position8.88888, 88.8888, 888.888, 8888.88, 88888.8, 888888
SC.PRI.DSPDIVPrimary units display divisions1D, 2D, 5D, 10D, 20D , 50D
SC.PRI.UNITSPrimary unitsLB, KG, OZ, TN, T, G
SC.SEC.DECPNTSecondary units decimal position8.88888, 88.8888, 888.888, 8888.88, 88888.8, 888888
SC.SEC.DSPDIVSecondary units display divisions1D, 2D, 5D, 10D, 20D , 50D
SC.SEC.UNITSSecondary unitsLB, KG, OZ, TN, T, G
Table 6-4. FORMAT EDP Commands
CommandDescriptionValues
SC.WZEROZero calibration—
SC.WVALTest weight valuevalue
SC.WSPANSpan calibration—
SC.REZERORezero —
SC.LC.CDRead or set dead load countsvalue
SC.LC.CWRead or set span countsvalue
ALG.SOURCEAnalog output source GROSS, NET
ALG.OFFSETZero offset0%, 20%
ALG.ERRACTError actionFULLSC, HOLD, ZEROSC
ALG.MINMinimum value tracked0 – 999999
ALG.MAXMaximum value tracked0 – 999999
ALG.MINNEGMinimum negativeNO, YES
ALG.MAXNEGMaximum negativeNO, YES
ALG.TWZEROZero calibration0 – 65535
ALG.TWSPANSpan calibration0 – 65535
ALG.OUTPUTVoltage or current outputCURRENT, VOLTAGE
Table 6-9. ALGOUT EDP Commands
Command
(n = setpoint number)
SP.BATCHGBatchingOFF, AUTO, MANUAL
SP.ENABLE#nSetpoint enableOFF, ON
SP.KIND#nSetpoint kindGROSS, NET, -GROSS, -NET, OFF, DELAY
SP.VALUE#nSetpoint valuenumber
SP.TRIP#nTripHIGHER, LOWER
SP.PREACT#nBand valuenumber
SP.HYSTER#nHysteresisnumber
SP.PUSHPR#nPerform print when setpoint is satisfiedOFF, ON
SP.PUSHAC#nUpdates accumulator and performs a print
when setpoint is satisfied
SP.PUSHTR#nPerform Tare when setpoint is satisfiedOFF, ON
SP.DIGOUT#nDigital outputsNONE, 1 – 8
SP.BATSEQ#nBatch type setpointOFF, ON
SP.WAITSS#nValue must be stableOFF, ON
DescriptionValues
OFF, ON
Table 6-10. SETPNTS 1– 8 EDP Commands
Command
(p = port number)
EDP.TRIGER#pEDP port triggerPort 1 – COMAND, DEMAND, PRN, NONE
EDP.BAUD#pEDP port baud rate1200, 2400, 4800, 9600, 19200, 38400
EDP.BITS#pEDP port data bits/parity8NONE, 7EVEN, 7ODD
EDP.SBITS#pEDP port stop bits1STOP, 2STOP
EDP.TERMIN#pEDP port termination characterCR/LF, CR
EDP.EOLDLY#pEDP port end-of-line delay0 – 255 (0.1-second intervals)
EDP.ECHO#pEDP port echoON, OFF
EDP.PRNMSG#pEnable print message when a
demand print is performed
EDP.STRUR#pLegal for Trade or industrial stream STRLFT,STRIND
EDP.STREAMSpecifies the operating mode of
the indicator (see Section 9.4)
DescriptionValues
Port 2 – NONE, PRN, STR-1, STR-2, STR-3, STR-4, STR-5
OFF, ON
NONE, LOCAL, REMOTE
Table 6-11. SERIAL EDP Commands
54480 Legend Series
Page 62
Ethernet Functionality
The Ethernet port works in COMMAND mode meaning that it is possible to send any EDP command to the
indicator and receive a response from the indicator accordingly. It is not possible to print from the port.
CommandDescriptionValu es
ETH.DHCPDynamic Host Configuration Protocol
(DHCP)
ETH.IPADDRESSSource IP address, a 4 field valueThe range of each field is 0-255
ETH.NETMASKSUBNET mask, a 4 field valueThe range of each field is 0-255
ETH.DEFAULTGATEWAYDefault gate way, a 4 field valueRange of each field is 0-255
ETH.MACReturns the MAC address, read only6 fields, 8 bit each, in hexadecimal base
ETH.SERVER.PORTThe ethernet server TCP port1-65535
ETH.SERVER.ECHOSets the Server Port EchoON, OFF
ETH.SERVER.TRIGGERServer Port Trigger DEMAND, COMAND, STRLFT, STRIND
ETH.SERVER.TIMEOUTServer Inactivity Timeout in seconds0-65535
ETH.SERVER.TERMIN Server Line Termination CR or CR/LF
ETH.SERVER.SFMTServer Stream FormatSTR-1, STR-2, STR-3, STR-4, STR-5
ETH.SERVER.EOLDLYServer End of Line Delay0-255
ETH.CLIENT.REMOTESERVERIPClient Remote Server IP AddressRange of each field is 0-255
ETH.CLIENT.REMOTESERVERPORT Client Remote Server TCP Port number1-65535
ETH.CLIENT.TRIGGERClient Port TriggerDEMAND, COMAND, STRLFT, STRIND
ETH.CLIENT.TIMEOUTClient Inactivity Timeout in seconds0-65535
ETH.CLIENT.TERMIN Client Line TerminationCR or CR/LF
ETH.CLIENT.SFMTClient Stream FormatSTR-1, STR-2, STR-3, STR-4, STR-5
ETH.CLIENT.EOLDLY Client End of Line Delay0-255
ETH.CLIENT.ECHOSets the Client EchoON, OFF
ON – Dynamic allocation of IP address
OFF – Static allocation of IP address
(0-255).(0-255).(0-255).(0-255)
(0-255).(0-255).(0-255).(0-255)
(0-255).(0-255).(0-255).(0-255)
(0-255).(0-255).(0-255).(0-255)
Table 6-12. Ethernet Commands
CommandDescriptionDefault Values
GFMTGross demand print format stringGROSS<G><NL2><TD><NL>
NFMTNet demand print format stringGROSS<G><NL>TARE<SP><T><NL>NET<SP2><N><NL2><TD><NL>
ACCFMTAccumulator demand print format string ACCUM <A><NL><DA> <TI><NL>
SPFMTSetpoint print format<SCV><SP><SPM><NL>
Table 6-13. PFORMT EDP Commands
CommandDescriptionValues
POWERDetermines if indicator automatically powers on
when plugged in or if power key must be pressed.
Read only in Weighing Mode.
BKLGHTControls the LCD backlight display.
Read only in Weighing Mode.
BKLGHT.AUTOControls the LCD backlight display auto off time.
Read only in Weighing Mode.
BATReads the battery level. Read only.Current battery level in millivolts.
USB.DEVICE Select the USB Device.HOST, DRIVE, PRINTER
USB.TRIGGERSet the USB Port Trigger.DEMAND, COMAND
AUTO
MANUAL
ON, OFF, AUTO
10SEC, 30SEC, 1MIN, 5MIN, 10MIN
Table 6-14. MISC Commands
The EDP Command Set55
Page 63
6.1.6Normal Mode Commands
Note
Note
The serial transmit weight data commands (see Table 6-15) transmit data to the selected port on demand. The SX
and EX commands are valid only in weigh mode; all other commands are valid in either setup or weigh mode.
CommandDescriptionResponse Format
UIDView and set unit ID when in setup mode.
In Weigh mode, the value can be changed temporarily. It
will reverts to last saved value at next power up.
DATESETView or set date.The response will be in the format, and uses the
TIMESETView or set time.The response will be in the format, and uses the
SXStart streaming on Port 2.
The Trigger Mode of Port 2 must be set to one
of the five stream modes for this command to
function.
EXStop streaming on Port 2
The Trigger Mode of Port 2 must be set to one
of the five stream modes for this command to
function.
RSSoft reboot—
XATransmit accumulator valuennnnnn UU
XGTransmit gross weight in displayed units
XNTransmit net weight in displayed units
XTTransmit tare weight in displayed units
XG2Transmit gross weight in non-displayed units
XN2Transmit net weight in non-displayed units
XT2Transmit tare weight in non-displayed units
XEQuery system error conditionsnnnnn nnnnn
Table 6-15. Normal Mode Commands
nnnnnn
separators as defined in configuration. To Set, first
query the date, then send DATESET=, in exactly the
same format (except use only 2 digits for the year).
separators as defined in configuration. To Set, first
query the time, then send TIMESET= followed by the
time in exactly the same format as the query,
including seconds, and AM or PM if 12 hour.
OK or ??
OK or ??
Where nnnnnn is the weight value, UU is the units.
See Section 9.0 for detailed information about the XE
command response format.
56480 Legend Series
Page 64
7.0Print Formatting
Note
The 480 provides multiple print formats, GFMT, NFMT, ACCFMT and SPFMT, that determine the format of the
printed output when the
GFMT is used.
Each print format can be customized to include up to 300 characters of information, such as company name and
address, on printed tickets. Use the indicator front panel (PFORMT menu) to customize the print formats.
Display will only show capital letters. Press the down arrow to view the ASCII character decimal value on the
second menu level. Refer to Table 9-6 on page 72 and Table 9-7 on page 73 to view the ASCII character
charts.
7.1Print Formatting Tokens
T able 7-1 lists tokens that can be used to configure the print formats. Tokens included in the format strings must be
enclosed between < and > delimiters. Any characters outside of the delimiters are printed as text on the ticket. Text
characters can include any ASCII character that can be printed by the output device.
<G>Gross weight in displayed units.
<G2>Gross weight in non-displayed units. ***
<N>Net weight in displayed units.
<N2>Net weight in non-displayed units. ***
<T>Tare weight in displayed units.
<T2>Tare weight in non-displayed units. ***
<A>Accumulated weight in displayed units.
<AC>Number of accumulator events (5 digit counter).
<AD>Date of last accumulator event.
<AT>Time of last accumulator event.
<UID>Unit ID number
<CN>Consecutive number
<NLnn>New line (nn = number of termination (<CR/LF> or <CR>) characters).*
<SPnn>Space (nn = number of spaces).*
<SU>Toggle weight data format (formatted/unformatted). ** ***
<TI>Time
<DA>Date (prints 4 digit year)
<TD>Time & Date
<SCV>Setpoint captured value.
<STV>Setpoint target value.
<SPM>Setpoint mode (Net or Gross)
<SN>Setpoint number
Gross, net, and tare weights are 9 digits in length, including sign (10 digits with decimal point)
followed by a space and a two-digit units identifier. Total field length with units identifier is 12
(or 13) characters.
UID and consecutive number (CN) fields are 6 characters in length.
* If nn is not specified, 1 is assumed. Value must be in the range 1–99.
** After receiving an SU command, the indicator sends unformatted data until the next SU
command is received. Unformatted data omits decimal points, and leading and trailing
characters.
*** Not available if regulatory mode is set to OIML.
Print key is pressed. If a tare has been entered or acquired, NFMT is used; otherwise,
To ke nDescription
Table 7-1. Print Format Tokens
Print Formatting Tokens57
Page 65
The default 480 print formats are shown in Table 7-2:
Note
Note
TIMDAT
ACCUM
SERIALSETPNT
PFRMAT
......
Display first 6
characters of format
Display and edit
active character and
ASCII value
Delete active
character
Scroll right in format string
Scroll left in format string
Increment ASCII value of active character
Decrement ASCII value of active character
Pressto insert a space
before the active character
GFMTGROSS<G><NL2><TD><NL>Weigh mode – no tare in system
NFMTGROSS<G><NL>TARE<SP><T><NL>NET<SP2><N>
Weigh mode – tare in system
<NL2><TD><NL>
ACCFMT
ACCUM <A><NL><DA><SP><TI><NL>
Accumulator demand print format string
SPFMT<SCV><SP><SPM><NL>Setpoint print operation with PSHPRNT=ON
Table 7-2. GFMT and NFMT Formats
The <G2>, <N2> and <T2> tokens (Table 7-1) print the gross, net and tare weights in the units not currently
displayed on the indicator.
The 300 character limit of each print format string includes the output field length of the print formatting
tokens, not the token length. For example, if the indicator is configured to show a decimal point, the <G> token
generates an output field of 13 characters: the 10 character weight value (including decimal point), one space,
and a two-digit units identifier.
PT (preset tare) is added to the tare weight if tare was keyed in.
7.2Customizing Print Formats
The following sections describe procedures for customizing the GFMT, NFMT, ACCFMT and SPFMT formats
using the front panel (PFORMT menu). See Figure 3-6 for print format menu structure.
7.2.1Using the Front Panel
Use the PFORMT menu to customize the print formats.
Using the PFORMT menu (see Figure 3-6), edit the print format strings by changing the decimal values of the
ASCII characters in the format string.
Some characters cannot be displayed on the 480 front panel (see the ASCII character charts on page 72 and
page 73) and are shown as blanks. The
480 can send or receive any ASCII character; the character printed
depends on the particular ASCII character set implemented for the receiving device.
58480 Legend Series
Figure 7-1. Print Format Menu
Page 66
8.0Setpoints
Note
Important
Note
The 480 indicator provides eight configurable setpoints for control of both indicator and external equipment
functions. Setpoints can be configured to perform actions or functions based on specified parameter conditions.
Parameters associated with various setpoint kinds can, for example, be configured to perform functions (print, tare,
accumulate), to change the state of a digital output controlling indicator or external equipment functions, or to
make conditional decisions.
Weight-based setpoints are tripped by values specified in primary units only.
Firmware versions 1.00 and 1.01 require having a relay board installed for the setpoints to operate.
8.1Batch and Continuous Setpoints
480 setpoints can be either continuous or batch.
Continuous setpoints are free-running; the indicator constantly monitors the condition of free-running setpoints at
each A/D update. The specified setpoint action or function is performed when the designated setpoint parameter
conditions are met. A digital output or function assigned to a free-running setpoint continuously changes state,
becoming active or inactive, as defined by the setpoint parameters.
Batch setpoint (BATSEQ = ON) is a digital output associated with a batch setpoint is act ive until the setpoint condition
is met. Then it is latched for the remainder of the batch sequence.
To use batch setpoints, set the batching (BATCHG) parameter on the setpoints (SETPTS) menu. This parameter
defines whether a batch sequence is automatic or manual. AUTO sequences repeat continuously after receiving a
single batch start signal, while MANUAL sequences only run through once for each Batch Start signal. The batch
start signal can be initiated by a digital input (set to BATSTR), or an EDP command (BATSTART).
For a setpoint to be used as part of a batch sequence, its batch sequence (BATSEQ) parameter must be set to ON. If
a setpoint is defined and enabled, but its Batch Sequence parameter is set to OFF, the setpoint operates as a
continuous setpoint even during batch sequences.
In applications that contain both batch setpoint routines and continuous setpoints, it is good practice to keep
continuous setpoints separate from the batch sequence.
Do not assign them the same digital output.
Batch and Continuous Setpoints59
Page 67
9.0Appendix
Note
Note
9.1Error Messages
The 480 indicator provides a number of error messages. When an error occurs, the message is shown on the
indicator LED display.
9.1.1Displayed Error Messages
The 480 provides a number of front panel error messages to assist in problem diagnosis. Table 9-1 lists these
messages and their meanings.
Error MessageDescriptionSolution
E A/DA/D physical errorCall Rice Lake Weighing Systems Service at 800-472-6703.
EEEROMEEPROM physical error
EVIREEVirgin EEPROMUse TEST menu to perform DEFLT (restore defaults) procedure, then
EPCKSMParameter checksum error
EACKSMA/D calibration checksum errorA/D converter requires recalibration. Call Rice Lake Weighing
EFCKSMPrinter format checksum errorCall Rice Lake Weighing Systems Service at 800-472-6703.
ELCKSMLoad cell calibration checksum errorRecalibrate load cells.
EIDATA Internal RAM checksum errorCall Rice Lake Weighing Systems Service at 800-472-6703.
E REFA/D reference errorA/D converter requires recalibration. Call Rice Lake Weighing
ERRORInternal program errorCheck configuration.
OVERFLOverflow errorWeight value too large to be displayed.
- - - - - -
- - - - - -
- - - - - -
- - - - - -
RNGERRGRADS > 100,000
EEPERREEPROM errorCall Rice Lake Weighing Systems for service at 800-472-6703.
HINOFF?High offsetZero load at powerup is more than initial zero range (INIZR) setting of
LINOFFLow offsetZero load at power up is less than initial zero range (INIZR) setting of
NOBATTNo batteryThe RTC lost time/date tracking at previous power off state due to
Gross > overload limitGross value exceeds overload limit. Check configuration or signal
Gross < 20d behind zeroGross value is more than 20 divisions behind zero (OIML).
WVAL > 100,000
recalibrate load cells.
Systems Service.
Systems Service.
Call Rice Lake Weighing Systems Service if unable to clear error by
cycling power or if error recurs.
input level. Overload can be caused by input signal > 45 mV or
common mode voltage > 950 mV.
Only shows up in configuration mode.
calibration zero – remove the extra load.
calibration zero – add the missing load.
low battery or no battery condition. The printer, accumulator and
audit functions will fail to get time and date.
Refers to internal coin battery only, not the
rechargeable battery option.
EUCKSMConfiguration checksumThe checksum value of configuration has changed from that stored
in memory.
OIMLEROIML parameter errorParameter set incorrectly for use in the OIML mode.
Example: Primary units set for lb or oz.
EE-ACCAccumulator errorError with the accumulator such as attempting to display an
accumulated value greater than six digits.
Table 9-1. 480 Error Messages
Shorting the excitation voltage shuts the excitation voltage off. The only way to restore excitation voltage is to
cycle power.
60480 Legend Series
Page 68
9.1.2Using the XE EDP Command
The XE EDP command can be used to remotely query the 480 for the error conditions shown on the front panel.
The XE command returns two five digit numbers in the format:
xxxxx yyyyy
where xxxxx contains a decimal representation of any existing error conditions as described in Table 9-2.
If more than one error condition exists, the number returned is the sum of the values representing the error
conditions. For example, if the XE command returns the number 1040, this value represents the sum of an A/D
reference error (1024) and an A/D calibration checksum error (16).
The second number returned (
test for the error condition was run. For example, the value
yyyyy) uses the same bit assignments as shown in Table 9-2 to indicate whether the
yyyyy = 50815 represents the decimal equivalent of the
binary value 1100 0110 0111 1111. Using the bit assignments in Table 9-2, this value indicates all tests were run.
Two EDP commands, P and ZZ, can be used to provide status about the indicator. These commands are described
in the following sections.
9.2.1P EDP Command
The P EDP command returns the current displayed weight value, along with the units identifier. If the indicator is
in an underrange or overload condition, the weight value is replaced with
9.2.2ZZ EDP Command
The ZZ EDP command can be used to remotely query which annunciators are currently displayed on the indicator
front panel. The ZZ command returns the currently displayed weight and a decimal number representing the LED
annunciators currently lit. The format of the returned data is:
wwwwww uuzzz
where wwwwww uu is the current displayed weight and units, and zzz is the annunciator status value (see Table 9-3).
If more than one annunciator is lit, ZZZ is the sum of the values representing the active annunciators.
&&&&&& (overload) or :::::: (underrange).
Status Messages61
Page 69
Example:
<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>
ASCII 02
decimal
Polarity:
<space> = Positive
<–> = Negative
Weight: 7 digits, right-justified, dummy
zeroes, decimal point with no leading
zeroes except for leading zero immediately
preceding the decimal point. Leading
zeroes transmitted as spaces.
L = pounds
K = kilograms
T = tons
G = grams
O = ounces
G = Gross
N = Net
Status:
<space> = valid
I = Invalid
M = Motion
O = Over/under range
Z = COZ
<CR> <LF>
or <CR>
Note
If the annunciator status value returned on the ZZ command is 145, the gr oss, standstill and lb annunciators are lit.
The number 145 represents the sum of the values for the standstill annunciator (128), gross mode annunciator (16)
and the lb/primary units annunciator (1).
Decimal
Val ue
1lb/primary units
2kg/secondary units
4Tare entered
8Keyed tare entered
16Gross
32Net
64Center of zero
128Standstill
Annunciator
Table 9-3. Status Codes Returned on the ZZ Command
9.3Continuous Data (Stream) Output Formats
When the trigger setting for COM-2 is set to one of the stream formats (STR1-STR5), or the Ethernet client or
server trigger setting is STRLFT or STRIND, data is continuously streamed from the appropriate port in one of the
five fixed formats that follow.
Figure 9-1. STR-1 Rice Lake Stream Data Format
Set format 5 if receiving device cannot read a status bit Z.
62480 Legend Series
Page 70
<STX> <SWA> <SWB> <SWC> <wwwwww> <tttttt> <CR>
STX (02h)
Status Word A
<CR> entered at end
of string.
Status Word B
Status Word C
MUL1 MUL0Parity
DP2 DP1 DP0 Fixed Zero/Dec. Pt. Location
0 0 0 x 100 (2 fixed zeroes)
0 0 1 x 10 (1 fixed zero)
0 1 0 x 1 (No decimal point)
0 1 1 x 0.1 (1 decimal place)
1 0 0 x 0.01 (2 decimal places)
1 0 1 x 0.001 (3 decimal places)
1 1 0 x 0.0001 (4 decimal places)
1 1 1 x 0.00001 (5 decimal places)
Disp. Div.
MUL1
MUL0
Multiplier
0 1 x 1
1 0 x 2
1 1 x 5
Even parity bit.
Requires EDP Port setup.
Status Word A:
Status Word C: ASCII 20h (space)
Even parity bit:
Requires EDP Port setup.
Status Word B:
0 = Gross Display Mode
1 = Net Display Mode
0 = Positive indicated weight value
1 = Negative indicated weight value
0 = In range
1 = Out of range
0 = Stable
1 = Motion
0 = lb
1 = kg
Even parity bit:
Requires EDP Port
setup.
Six digits, right justified, dummy zeros, decimal
point with no leading zeroes except for immediately
preceding the decimal point. Leading zeroes
transmitted as spaces.
Tare Weight: Six digits, right justified,
dummy zeros, decimal point with no
leading zeroes except for immediately
preceding the decimal point.
Leading zeroes transmitted as spaces.
76543210
DP2 DP1DP010
Parity
76543210
1000000
Parity
76543210
10
Figure 9-2. STR-2 Toledo Stream Data Format
Continuous Data (Stream) Output Formats63
Page 71
Figure 9-3. STR-3:Cardinal Stream Data Format
Carriage
Return
Polarity:
<+> for positive
"–" for negative
Six digits, right justified, floating decimal
(follows weight if not embeded),
leading zeroes are present.
g for Gross
n for Net
Hex value of <03>
is entered in string
lb = pound
kg = kilogram
g = gram
tn = ton (short)
t = ton (metric)
oz = ounce
<space> = none
Status:
<space> = valid or invalid
m = motion
o = over/under range
Space
Polarity:
<+> for positive
"–" for negative
Six digits, right justified, floating decimal,
leading zeros are present
g for Gross
n for Net
<CR><LF>
or
<CR>
lb = pound
kg = kilogram
g = gram
tn = ton (short)
t = ton (metric)
oz = ounce
<space> = none
Space
<SP> <G/N> <POL> <wwwwww> <SP> <UNIT> <TERM>
<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>
ASCII 02
decimal
Polarity:
<space> = Positive
<–> = Negative
Weight: 7 digits, right-justified, dummy
zeroes, decimal point with no leading
zeroes except for leading zero immediately
preceding the decimal point. Leading
zeroes transmitted as spaces.
L = pounds
K = kilograms
T = tons
G = grams
O = ounces
G = Gross
N = Net
Status:
<space> = valid
I = Invalid
M = Motion
O = Over/under range
<CR> <LF>
or <CR>
Note
Figure 9-4. STR-4 Weightronix Stream Data Format
Port 1 cannot be set to stream.
64480 Legend Series
Figure 9-5. STR-5: Consolidated Controls Stream Data Format
Page 72
9.4Local/Remote Operation
For systems that require two locations, local/remote support provides function equivalent to that of a Legal for
Trade remote display with keypad. Scale data from the local indicator is also displayed at the remote unit, and
keypad input from the remote allows transactions to be initiated from either the local or remote unit.
To configure for local/remote operation, set STREAM to local or remote. Once the local unit scale has been
configured, use the SERIAL menu, serial commands or Revolution to set the serial parameters for local/remote
operation. See Table 3-3 on page 26 for comport option s.
Serial
Configuration
Parameter
STREAMLocalStream Continuous Data RemoteSets indicator to display the
COM2.TRIGERSelect STR-1Streams the local unit
COM2.BAUDLocal and remote must match
COM2.ECHOOFFOFF
Parameter ValueFunctionParameter ValueFunction
Printing operation can be performed at the remote unit and/or at the local unit. When the operator presses the
Local UnitRemote Unit
weight data sent from the
Local Unit.
weight data to the
remote unit
Table 9-4. Local/Remote Configuration
Select STR-1Transmits demand key
presses, zero, gross/net,
tare, units
Print
key at the local unit, a printing ticket will be issued only locally. When the operator presses the Print key at the
remote unit, print will only occur at the remote unit.
9.5Audit Trail Support
Audit trail support provides tracking information for configuration and calibration events. To prevent potential
misuse, all configuration and calibration changes are counted as change events.
Audit trail information can be printed from Revolution or by sending the DUMPAUDIT serial command. See
Figure 3-1 on page 22. Revolution can be used to display audit trail information. The audit trail display includes the
legally relevant (LR) version number (firmware version for the code that provides audit trail information), a
calibration count and, if REGULA=NTEP or Canada, a configuration count. The exact format of the information
shown depends on the regulatory agency specified for the REGULA (PROGRM menu) parameter.
See Section 1.8.10 on page 7 for steps to view the Audit Trail.
For NTEP and Measurement Canada modes, there are two numbers displayed: configuration and calibration. The
configuration number tracks changes to the legally relevant configuration parameter changes. The calibration
number tracks changes to the scale calibration.
In OIML mode, the only number displayed is audit trail. The audit trail number is incremental for any legally
relevant configuration parameter changes and all calibration changes. The last date the scale was calibrated is also
displayed.
9.6Sleep Mode
Firmware version 1.03 and newer of the 480 features a sleep mode. It is designed to place the indicators into a low
power mode to reduce power consumption.
The sleep mode will only activate if the indicator is operating on its internal optional battery. If connected to
external AC power, the indicator will not go into sleep mode regardless of the parameter settings.
The indicator will enter the sleep mode, after the time specified in the
detected and the displayed weight is below the weight threshold (
disables the sleep mode.
Standby
While in sleep mode, at a time specified in the standby (STDBY) parameter, the indicator wakes up for about 0.5
seconds. The range of the standby time is 1 to 60 seconds. During the awake time, it checks for a change in weight
greater than the display division threshold (DDTHRH), or if the weight has gone above the weight threshold
(WGTHRH). It also checks to see if a key is being pressed and performs zero tracking.
SLEEP parameter, when no keypad activity is
WGTHRH). Setting the sleep parameter to never
Local/Remote Operation65
Page 73
Example:
Note
If the standby parameter set to 1 second, the weight threshold to 100, and the display division threshold is set to
10d, once in sleep mode, the indicator will wake up every second for about half a second to check to see if a key is
being pressed, the weight threshold is over 100 (WGTHRH), or the weight has change more than 10d (DDTHRH)
since the last time it checked.
Setting the standby parameter to a higher value reduces the amount of time the indicator is awake, this will increase
battery life.
OFF setting stops the indicator from waking up. The only way to wake up the indicator is to press the Power key.
An
The Power key is the only key that will wake the indicator, unless the Standby parameter is set to any setting
other than NONE. Then any key press will work to wake the indicator from standby.
9.7Internal Battery Option
In Firmware version 1.03 and later, an optional internal rechargeable Li-ion battery is available. The battery can
replace the AC power when needed. Typical run time is 12 hours (single load cell, no options, with LED display).
The amount of time the indicator will operate on the battery depends on:
•Type of display (LED)
•If sleep mode is used or not
•How may load cells are attached
•What options are installed
The battery is automatically charged when the indicator is connected to AC power. It automatically switches to
battery power if the AC power is removed. The battery is designed to only be charged by connecting to AC power.
When operating on battery power, the indicator features a sleep mode that allows the indicator to go into a low
power mode. Using the sleep mode helps improve battery life. See Section 9.6 for more information about the sleep
mode.
When the battery life reaches 5-10 percent of capacity, the
less than 5 percent, it will flash faster, at 0 percent, the indicator will turn off. Battery char ge time is approximately
16 hours to 100 percent.
Lo Bat flashes on the display. When the battery life is
9.8USB Option
In firmware version 1.03 and later, an optional Ethernet/USB board is supported.
The USB connection is capable of being either of the following:
•USB device (connected to a PC and appearing as an available communication port)
•USB host interface to a flash drive
•USB interface to printer
The mode of the USB option is selected in the User Menu.
9.8.1HOST
When set to HOST mode, the USB port can be connected to a USB connection on a PC. An HCC Driver is required
on the PC.
The HCC driver is available on a CD (PN 171153) that is included with the USB/Ethernet Option. Use the
documentation (PN 171152) for the option to install the driver on the PC.
When connected to a PC, the connection will appear to applications as another serial communications port. Using
the USB trigger parameter in the 482 user menu, it is possible to set the port to act as both a serial command port
and demand print port (
used through the USB HOSTPC port.The USB port cannot be used to stream data.
9.8.2Flash Drive
When set to Drive mode, the USB port can be connected to a flash drive to allow the upload and download of the
indicator configuration parameters. The data is written to the drive in a FA T32 system format; no special drivers are
required.
To use with a flash drive, first configure the
the drive, or loading it from the drive, can only be done in the configuration menu.
DEMAND), or as a serial command port only (COMAND). All EDP serial commands can be
USB parameter in the user menu to DRIVE. Saving the configuration to
66480 Legend Series
Page 74
Save the Configuration
1. Attach a flash drive to the USB port.
2. Enter the configuration menu and navigate go to
3. Press , then press or to select
4. Press the
Tare key. A copy of the configuration DUMPALL, will be written to the drive in an ASCII text
SAVE. See Section 3.2.9 on page 43.
USBMEM. See Section 3.2 on page 32.
file with the name “480_<UID>.txt”, where <UID> is replaced with the unit ID configured in the indicator.
Load a Configuration File
1. Install the file on the flash drive. The file can be either of two formats, a configuration dumpall text file, or
a configuration file created by the Revolution indicator configuration utility. The file name must be
“480_<UID>.txt” for a text file, or “480_<UID>.rev” for a Revolution file, where <UID> is the currently
configured Unit ID of the indicator.
2. Attach the flash drive to the USB port.
3. Enter the configuration menu and navigate go to
USBMEM. See Section 3.2 on page 32.
4. Press , then press or to select LOAD. See Section 3.2.9 on page 43
5. Press the
Tare key. If a file is found on the flash drive with a <UID> that matches that currently in the
indicator, it will be loaded. If the loading of the file fails for any reason, an error will be displayed.
Edit and Save File on Flash Drive with Revolution
1. To open the file, select the appropriate 480 module and open it.
2. Use the Import function in the File menu to import the configuration parameters into Revolution.
9.8.3USB Printing
When set to PRINTR mode, the USB port can be connected to a USB printer using an appropriate USB cable.
To configure (see Section 3.1.4 on page 29):
1. Set the USB port to
2. Set the trigger to
3. Press the
Print key or call a remote print command to print to the USB printer.
PRINTR.
DEMAND.
9.9Ethernet Option
In firmware version 1.03 and later, an optional Ethernet/USB board is supported.
With the Ethernet option installed, the 480 can either be connected directly to a PC via the Ethernet connection, or
to an Ethernet network.
The 480 Ethernet option will support two simultaneous connections, one as a server, the other a client. It can be set
up with a static, or DHCP-provided IP address. Both ports can be used to communicate to the indicator using all the
available serial EDP commands. Both ports can be set to print on demand and can stream data.
The server port has a configurable TCP port number, and can accept a single remote client connection.
The client port can be configured to connect to a remote server’s IP and TCP port number upon power up, or
whenever there is data to send.
Connection is made via a RJ45 connector inside the indicator on the Ethernet/USB option board.
For details on all the configurable parameters for the optional Ethernet ports, see Section 3.1.3 on page 28.
Ethernet Option67
Page 75
9.10Conversion Factors for Secondary Units
Note
The 480 has the capability to mathematically convert a weight into many different types of units and instantly
display those results with a press of the
Secondary units can be specified on the Format menu using the
Multipliers are preconfigured within the indicator.
Ensure that the secondary decimal point position is set appropriately for the scale capacity in the secondary
units. If the converted value requires more digits than are available, the indicator will display an overflow
message (OVERFL).
For example, if the primary units are short tons, secondary units are pounds, and the secondary decimal point is set to
8888.88, the indicator will overflow if five tons or more are applied to the scale. With 5 tons applied, and a conversion
factor of 2,000, the secondary units display needs five digits to the left of the decimal point to display the 10,000 pound
secondary units value.
Unit key.
SECNDR parameter.
9.11Digital Filtering (DIGFIL)
Digital filtering is used to create a stable scale reading in challenging environments. The 480 (firmware version
1.03 and later) allows selection of
for no filtering.
The A/D sample rate must be selected first (see Section 9.11.1). Then select the type of digital filter using the
DIGFIL parameter.
9.11.1Sample Rate (SMPRAT)
The A/D sample rate is the number of weight readings the indicator takes per second (
20 or 40 Hertz, or readings per second.
Set the A/D sample rate to the lowest setting required f or the application. Lo wer settings result in better stability, so
5 Hertz is more stable than 40 Hertz.
Adaptive Filtering (see Section 9.11.2), Rolling Filtering (see Section 9.11.3) or OFF
SMPRAT). This can be 5, 10,
9.11.2Adaptive Filter
The adaptive filter has two settings, sensitivity (
AFSENS) and threshold (AFTHRH). It maintains a running average
of the A/D readings when the weight change remains less than the defined Threshold value. The filter
automatically applies less value to each successive A/D reading the longer the weight change is less than the
Threshold value. The amount of value given to the most recent A/D reading is determined by the sensitivity setting.
Adaptive Filter Sensitivity (AFSENS)
Adaptive filter sensitivity (AFSENS) can be set to HEAVY, MEDIUM or LIGHT. A heavy setting will result in an output
that is more stable to weight changes, than that of light. However , small changes in weight data (a few grads) on the
scale base will not be seen as quickly.
If the difference in typical subsequent weight values on the scale will only be a few grads, use a light setting. On a
truck scale where the changes in subsequent weight values will be 100s of grads, a heavy setting will be more
appropriate. When the weight has not exceeded the Threshold (
approximately 30 A/D cycles for
Adaptive Filter Threshold (AFTHRH)
LIGHT, 90 A/D cycles for MEDIUM and 180 A/D cycles for HEAVY.
AFTHRH) setting, the effective response is
The adaptive filtering threshold (AFTHRH) should be set for the amount of observed instability in the system. This
parameter can be set in the range of 0 to 99999 and is entered as a weight value. When a new sampled weight value
is acquired, the adaptive filter compares the new value to the previous (filtered) output value.
If the difference between the new value and the previous output value is greater than the
AFTHRH parameter the
adaptive filter is reset to the new weight value.
If the difference between the new value and the previous output value is less than the
AFTHRH parameter, the two
values are averaged together using a weighted average. The weighted average is based on the amount of time the
system has been stable, and selected
AFSENS sensitivity.
With the adaptive filter threshold set at zero, determine the amount of instability that is present. Enter this amount
of weight instability to set the threshold of the adaptive filter. The adaptive filter is set to
OFF the AFTHRH
parameter is set to zero.
68480 Legend Series
Page 76
9.11.3Rolling Averaging Filter
The rolling averaging filter uses mathematical averaging with three stages. These configurable stages control the
effect of a single A/D reading on the displayed weight. When an A/D reading is encountered that is outside a
predetermined band, the rolling averaging filter is overridden and the display jumps directly to the new value.
Filter Stages (RFSTG1-3)
The filter stages can each be set to a value of 1 to 64. The value assigned to each stage sets the number of readings
received from the preceding filter stage before averaging. Setting the filter stages to 1 effectively disables the
rolling filter.
A rolling average is passed to successive filtering stages for an overall filtering effect that is effectively a weighted
average of the product of the values assigned to the filter stages (
corresponding to the sum of the values (
Filter Sensitivity (RFSENS) and Threshold (RFTHRH)
RFSTG1 + RFSTG2 + RFSTG3).
RFSTG1 x RFSTG2 x RFSTG3) within a time frame
The rolling averaging filter can be used by itself to eliminate vibration effects, but heavy filtering also increases
settling time. The
RFSENS and RFTHRH parameters can be used to temporarily override filter averaging and
improve settling time.
RFSENS specifies the number of consecutive A/D readings that must fall outside the filter threshold
•
(RFTHRH) before filtering is suspended.
•
RFTHRH sets a threshold value, in display divisions. When the specified number of consecutive A/D
readings (RFSENS) fall outside of this threshold, filtering is suspended. Set RFTHRH to NONE to turn off the
filter override.
Rolling Averaging Filter Parameters
1. In setup mode, set the rolling filter stage parameters (RFSTG1-3) to 1.
2. Set
RFTHRH to NONE.
3. Return to weigh mode.
4. Remove all weight from the scale, then watch the indicator to determine the magnitude of vibration effect
on the sc
5. Record the weight below which all but a few readings fall. This value is used to calculate the
ale.
RFTHRH
parameter in step 8.
Example – if a heavy-capacity scale (10000x5lb) produces vibration-related readings of up to 50 lb, with
occasional spikes to 75 pound, record 50 pound as the threshold weight value.
6. Place the indicator in setup mode and set the filter stage parameters (
ef
fects on the scale (leave
7. Find the lowest effect value for the
If needed, the rolling filter cutout sensitivity (
RFTHRH set to NONE).
RFSTG1-3 parameters.
RFSENS) and the rolling filter cutout threshold (RFTHRH) can be used
RFSTG1-3) to eliminate the vibration
to reset the averaging filter so the response to a rate change is faster.
8. Calculate the
div
isions (
RFTHRH parameter value by converting the weight value recorded in step 5 to display
Threshold_Weight_Value / Display_Divisions).
In the example in step 5, with a threshold value of 50 pound, and a display divisions value o f 5 pound,
RFTHRH should be set to 10 display divisions.
9. Set the
RFSENS parameter high enough to ignore transient peaks. Longer transients (typically caused by
lower vibration frequencies) will cause more consecutive out-of-band readings, so RFSENS should be set
higher to counter low frequency transients.
s
Digital Filtering (DIGFIL)69
Page 77
9.12Analog Output Calibration
Note
Note
The following calibration procedure requires a multimeter to measure voltage or current output from the analog
output module. See Figure 3-15 for analog output menu structure.
The analog output must be calibrated after the indicator itself has been configured (see Section 3.0) and
calibrated (see Section 4.0).
1. Enter configuration mode and go to the ALGOUT menu (see Figure 3-15).
•S et source to
•Set
•Set
•Set
OFFSET to 0% for 0–10V/0-20mA output, 20% for 2-10V/4–20 mA output
MIN to lowest weight value to be tracked by the analog output
MAX to highest weight value to be tracked by the analog output
Calibrate the minimum weight value for 0 mA, even if a 4-20 mA output will be used. After the calibration is
complete, select the 20% offset.
If set to a 20% offset, the lowest value tracked will be 4 mA. To use a value lower than 4mA, use the 0-20 mA
setting.
2. Connect multimeter to analog output:
•For voltage output, connect voltmeter leads to pins three and four
•For current output, connect ammeter leads to pins one and two
3. Adjust the zero calibration: Scroll to the
multimeter. Press or to select the digit to edit and or to adjust the value of selected (flashing)
digit to set the zero value.
4. Adjust span calibration: Scroll to the
Press or to select the digit to edit and or to adjust the value of selected (flashing) digit to set the
span value.
5. Final zero calibration: Return to the
Press or to select the digit to edit and or to adjust the value of selected (flashing) digit to set the
zero value.
6. Return to weigh mode. Analog output function can be verified using test weights.
GROSS or NET, depending on the weight value to be tracked by the analog output
TWZERO parameter. Check voltage or current reading on
TWSPAN parameter. Check voltage or current reading on multimeter.
TWZERO parameter and verify that the zero calibration has not drifted.
9.13Test Mode
In addition to weigh and setup modes, test mode provides a number of diagnostic functions for the 480, including:
•Display raw A/D count
•Reset configuration parameters to default values
•Number of weighments over threshold
•View the state of the digital inputs
•Control the state of the digital outputs
•Test the analog output
1. Enter configuration mode by removing the setup switch access screw from the back of the enclosure.
2. Insert a non-conductive tool into the access hole to press the configuration switch. Indicator display
changes to show
3. Navigate over to
Figure 3-18 shows the Test Menu structure and Table 3-19 summarizes the test menu functions.
CONFIG.
TEST and then down into the TEST menu.
70480 Legend Series
Page 78
9.14Regulatory Mode Functions
Note
Regulatory
Parameter
NTEP
Canada
OIML
None
Weight On
Scale
Zero
Negative
Positive
Zero
Negative
Positive
Zero
Negative
Positive
Zero
Negative
Positive
Tare I n
System
No“000000”Zero
Yes Clear tare Zero
No No action Zero
YesClear tare Zero
No TareZero
Yes TareZero
No “000000”Zero
Yes Clear tare Clear tare
No No action Zero
YesClear tare Clear tare
No Tare Zero
Yes No action Clear tare
No“000000”Zero
YesClear tareZero & Clear tare
NoNo action Zero
YesClear tare Zero & Clear tare
No“000000”Zero
Yes Tare Z ero & Clea r Tare
No“000000”Zero
YesClear tare Clear tare
NoNo action Zero
YesClear tare Clear tare
NoTare Zero
YesClear tare Clear tare
Front Panel
Key Tare
Table 9-5. Tare and Zero Key Functions for REGULA Parameter Settings
Front Panel
Key Zero
At zero weight, the push-button tare will prompt for a keyed tare when tare function is set to keyed or both.
Regulatory Mode Functions71
Page 79
9.15ASCII Character Chart
Use the decimal values for ASCII characters listed in Tables 9-6 and 9-7 when specifying print format strings on
the 480 PFORMT menu. The actual character printed depends on the character mapping used by the output device.
The 480 can send or receive any ASCII character value (decimal 0–255), but the indicator display is limited to
numbers, upper-case, unaccented letters and a few special characters. See Section 9.16 for information about the
480 LED display.